SciCan STATMATIC smart 2 Instructions For Use Manual
Advertisement
Advertisement
Instructions For Use
STATMATIC
smart 1, 2, 3
Cleaning and Maintenance System
EN/DE
Brugsvejledning DK via: www.scican.uk.com/downloads
Manuel d'instructions FR via: www.scican.uk.com/downloads
Istruzioni per l’uso IT via: www.scican.uk.com/downloads
Gebruiksaanwijzing NL via: www.scican.uk.com/downloads
www.scican.com
STAT MATIC smart 1, 2, 3
Copyright 2017 SciCan. All rights reserved.
Doc. No. 2.001.8462
– Rev. 2017-11
Table of Contents
1 Important Information .................................................. 3
Type description .......................................................... 3
Scope of delivery ........................................................ 3
General ....................................................................... 3
Intended use ............................................................... 4
Disposal ...................................................................... 4
Qualifications of the operator ...................................... 5
Safety Notes SciCan "Dental Oil" and SciCan
"Cleaner" ..................................................................... 5
Used symbols ............................................................. 5
Disclaimer ................................................................... 6
Device overview .......................................................... 7
2 Commissioning ........................................................... 9
Environmental conditions ............................................ 9
Electrical Requirements .............................................. 9
Media connections ...................................................... 9
Filling the media system ............................................10
Setting the language ..................................................10
Setting the date/time ..................................................11
Assemble and check the adapter ...............................11
3 Operation ...................................................................12
Manual pre-cleaning of instruments ...........................12
"Collet care" program .................................................12
STAT MATIC PLUS is a trademark; OPT IM and STAT IM are registered trademarks of SciCan Ltd. All other trademarks referred to in this manual are the property of their respective owners.
Manufactured by:
SycoTec GmbH & Co. KG
Wangener Strasse 78
88299 Leutkirch
GERMANY
Phone:
Fax:
+49 7561 86-0
+49 7561 86-266
"Special instrument" program .................................... 15
4 Maintenance .............................................................. 16
Checking adapter fixation .......................................... 16
Checking the O-rings ................................................. 16
Changing the adapter ................................................ 16
Refilling of the media containers ............................... 17
Replacing the absorption pads .................................. 18
Empty the collecting tray ........................................... 18
Recommendation ...................................................... 18
Data backup of the log file ......................................... 19
Deletion of the log file on the STAT MATIC smart ...... 19
Emptying the media system ...................................... 19
Warning message: Servicing required soon .............. 20
Error message: Servicing required ............................ 20
Transport ................................................................... 20
5 Troubleshooting ......................................................... 21
Replacing fuses ......................................................... 22
6 Information on Electromagnetic Compatibility (EMC) 23
7 Accessories ............................................................... 25
8 Technical Data .......................................................... 26
Type plate .................................................................. 27
9 Documentation, Log File............................................ 28
10 Guarantee ................................................................. 29
SciCan GmbH
Wangener Strasse 78
88299 Leutkirch, GERMANY
Phone:
Fax:
+49-7561-98343-0
+49-7561-98343-699
SciCan Ltd.
1440 Don Mills Road
Toronto, ON M3B 3P9, CANADA
Phone:
Fax:
Toll free:
E-mail:
+1-416-445-1600
+1-416-445-2727
+1-800-667-7733 [email protected]
SciCan Medtech AG
Alpenstrasse 14
6300 Zug, SWITZERLAND
Phone: +41-41-727-70-27
Fax:
SciCan Inc.
+41-41-727-70-29
701 Technology Drive
Canonsburg, PA 15317, USA
Phone:
Fax:
+1-724-820-1600
+1-724-820-1479
Page 2 STAT MATIC smart 1,2,3 EN
1 Important Information
1 Important Information
Congratulations on purchasing a STAT MATIC smart. We are convinced that you have acquired the best product of its kind. The STAT MATIC smart is a compact cleaning and maintenance system that can satisfy all your needs and expectations regarding the cleaning and maintenance of instruments thanks to its diverse programs and capabilities.
This instruction manual contains all details on the installation, operation and maintenance of your STAT MATIC smart system. In order to guarantee many years of safe and problem-free operation, you should read the instructions prior to commissioning the device and store them for future reference.
The instructions for the operation, maintenance and replacement of replacement parts must be followed so that the device can function properly. The content of this instruction manual may change without notice in order to reflect changes and improvements to the STAT MATIC smart system.
Type description
The STAT MATIC smart is manufactured in 3 different versions.
STAT MATIC smart Type 1 is equipped with a DIN EN ISO 3964 coupling, which is operated by a motor as well as
2 neutral couplings that can be individually equipped with the required turbine adapters (see "Accessories" section).
STAT MATIC smart Type 2 is equipped with two DIN EN ISO 3964 couplings, which are operated by a motor as well as
1 neutral coupling that can be individually equipped with the required turbine adapter (see "Accessories" section).
STAT MATIC smart Type 3 is equipped with three DIN EN ISO 3964 couplings, which are operated by a motor.
Scope of delivery
Base STAT MATIC smart unit
Power supply connection cable
Instructions for use
Absorption pad (set)
Compressed air hose
SciCan "Dental Oil" (250 ml)
SciCan "Cleaner" (500 ml)
O-Ring Set
2 seal plugs (red) for media container
USB stick (2 GB)
General
The STAT MATIC smart has been developed and manufactured in accordance with the most recent quality and safety standards.
However, some precautions should be taken during installation and operation.
•
Disconnect the STAT MATIC smart system from the power supply when leaving the practice for a longer period of time.
•
Protect the STAT MATIC smart from temperatures above 50°C.
•
Any changes or modifications made by a third party will invalidate the warranty as well as operating certification.
•
Only use original replacement parts for operation or repair, as these have been tested extensively for safety, function and specific suitability. SciCan therefore also assumes any product responsibility for these parts and components.
•
Observe all warnings on the device and in the instructions for use.
•
The device may only be supplied with the voltage indicated on the type plate.
•
The power cable must not be replaced by a weak or underpowered power cable.
•
Operators are not allowed to perform maintenance on the unit in other respects than those stated in the instruction manual.
•
Medical gloves have to be worn during commissioning, operation and maintenance of the device.
•
The device may only be operated with original SciCan "Cleaner", SciCan "Dental Oil" and accessories.
•
Prior to transport or before prolonged periods of standstill (> 30 days), the containers for cleaner and dental oil must be emptied.
•
The operator must make sure that the device is in the proper operating condition prior to using the device as well as confirm its functional safety.
•
A damaged device or a device with a damaged power cable has to be disconnected from the mains immediately.
Operating the device is not permitted in this case.
•
The device may only be operated inside buildings and areas that are not subject to explosive hazards.
•
Never open the protective housing of the device or insert objects through the openings of the protective housing.
Page 3 STAT MATIC smart 1,2,3 EN
1 Important Information
Intended use
•
The STAT MATIC smart is a cleaning and maintenance system used for the machine-based treatment (internal cleaning and maintenance) of dental turbines as well as regular and contra-angle handpieces.
•
Refer to the manufacturer instructions for further information on the suitability of certain instruments for cleaning and maintenance purposes.
The STAT MATIC smart treatment is restricted to the internal cleaning of the instruments. The outer surfaces of the instruments have to be cleaned manually or using machines prior to any internal cleaning. Following treatment with the cleaning and maintenance system, continue treating the instrument immediately afterwards and in accordance with the manufacturer's instructions.
NOTE: For all preceding and subsequent processing steps, the RKI guidelines as well as the Medical Products Law must be observed (applicable only for Germany).
Misuse
All applications not listed as part of the intend use constitute a case of misuse and may lead to damage to the device and the medical products to be reprocessed and/or a potential danger to the operator himself. SciCan shall accept no liability in such an event.
Disposal
Consumables
Any waste resulting from use of the device have to be disposed of or recycled without presenting a danger to human beings or the environment while existing national regulations must also be observed in this regard.
Devices and accessories after use
Based on EU Directive (WEEE 2012/19/EU) on waste electrical and electronic equipment we hereby inform you that this product is subject to the aforementioned directive and must be disposed of through special channels.
Contact the manufacturer or dental supply shop for further information.
This product must be disposed of or recycled without presenting a danger to human beings or the environment while existing national regulations must also be observed in this regard.
Page 4 STAT MATIC smart 1,2,3 EN
1 Important Information
Qualifications of the operator
The STAT MATIC smart system may only be used by trained personnel, including dentists and their assistants who have been instructed in the use of this product.
Safety Notes SciCan "Dental Oil" and SciCan "Cleaner"
First-Aid-Measures - Dental oil
General information Never give any oral medication to unconscious persons or persons suffering from
When inhaled cramps. Move the person to safety
Bring the person to the fresh air and provide unobstructed breathing
Contact with skin
Wash with plenty of water. Immediately changed out of contaminated clothes. Medical treatment required
Contact with eyes
When swallowed
Immediately, carefully and thoroughly rinse with eye-bath or water. Consult an eye specialist in case of eye irritation.
Vomiting may pose a danger of aspiration. Immediately rinse the mouth and drink plenty of water. Induce vomiting if the person is conscious. Medical treatment required.
Fire-fighting measures
The product is not flammable.
First-Aid-Measures - Cleaner
General information
Never give any oral medication to unconscious persons or persons suffering from
When inhaled cramps. Move the person to safety
Provide fresh air.
Contact with skin
Contact with eyes
When swallowed
Wash with plenty of water. Change out of contaminated clothing.
Immediately, carefully and thoroughly rinse with eye-bath or water.
Immediately rinse the mouth and drink plenty of water.
Fire-fighting measures
The product is not flammable.
Used symbols
In the manual:
A potential danger to the operator
A situation that may lead to a mechanical defect
Important information
Instructions for disposal
Wear medical gloves
On the device:
LED ready to operate
LED error
Manufacturer
VDE certification mark
Program selection / selection up
Start program / selection down
USB port
CE marking
Fuse
Follow instruction manual
MET certification mark
Menu / select / save
Collet care / back
Page 5 STAT MATIC smart 1,2,3 EN
1 Important Information
Disclaimer
The installation of replacement parts, maintenance and upkeep may only be carried out by authorized personnel.
SciCan assumes no liability whatsoever for coincidental, special or consequential damage resulting from upkeep or maintenance of the STAT MATIC smart carried out by third parties or for the use of equipment or individual parts manufactured by third parties, including loss of profits, commercial losses of any kind, financial losses or losses due to the injury of persons.
Never remove the rear panel of the device and never insert any objects through the holes or openings on the housing.
Noncompliance may result in damage to the device and/or danger to the operator.
Page 6 STAT MATIC smart 1,2,3 EN
1 Important Information
Device overview
(1) Cover
(2) Membrane keypad
(3) Plug-in position for turbine adapter
(2x Type 1 / 1x Type 2) or
INTRAmatic ISO 3964 maintenance couplings
(3x Type 3 / 2x Type 2 / 1x Type 1)
(4) Collecting tray
(5) Flap
(6) Absorption pad (2x)
(7) Collet care nozzle
(8) Display
(9) USB stick
(10) Filling opening for dental oil
(11) Filling opening for cleaner
(12) Holding bracket for testing adapter
(21) USB port
(22) Media aeration
(13) Power switch
(14) Mains fuse drawer
(15) Power line input
(16) Drain hose for dental oil and cleaner
(17) Pressure indicator
(18) Pressure setting
(19) Compressed air input
(20) Type plate
Page 7 STAT MATIC smart 1,2,3 EN
1 Important Information
M embrane keypad
(L1) LED stand-by
(L2) LED error
(S1) Program selection / selection up
(S2) Start program / selection down
(D1) Display
(S3) Collet care / back
(S4) Menu / select / save
Display
Date / Time
Status lines
Slots
Page 8
Slot number
Slot is finished
Slot is occupied
Frame is flashing:
Slot is being processed
STAT MATIC smart 1,2,3 EN
2 Commissioning
2 Commissioning
Environmental conditions
There are a number of factors that can influence the performance of your STAT MATIC smart system. Please consider these factors carefully and choose a suitable location for installing the device.
• Temperature and air humidity
Do not set up your STAT MATIC smart system at a location where it is exposed to direct sunlight or a heat source
(for example a heater). The recommended operating temperature lies between 5 and 35°C with air humidity between 25 and 85%.
• Environment
The STAT MATIC smart should be operated in a clean, dust-free and well-ventilated environment.
• Work surface
The STAT MATIC smart should be installed on a flat, level and water-resistant as well as oil-resistant surface. Do not install or operate your unit on a sloped surface.
• Electromagnetic environment
The STAT MATIC smart has been tested for electromagnetic emissions and meets the decisive standards. While the product itself emits no radiation, it can be influenced by other radiating devices. We therefore recommend that you keep the device away from potential sources of interference.
Electrical Requirements
DANGER! Mortal danger with improper handling of power sources! Danger of electrical shock
•
For connection to the mains, use a properly grounded and secure socket providing the stated voltage (see type plate on the back of the STAT MATIC smart).
•
Avoid using multiple sockets. When using power bars with surge protection, always connect only a single device.
• The mains input connector on the back of the device is to be used as a separator from the power supply.
Media connections
Plug the mains plug into the socket.
Attach the compressed air hose to the coupling and tighten.
The set pressure should be between 4.5 bar and 4.3 bar!
In the event of a deviating display, pull out the grip for regulating pressure and turn it to the right or the left until the manometer shows 4.5 bar.
Push the grip back into the pressure regulator and fixate it.
In order to ensure a safe and trouble-free operation, only use compressed air in accordance with the information in the "Technical Data" section.
Page 9 STAT MATIC smart 1,2,3 EN
2 Commissioning
Commissioning
Turn on the mains switch.
Select the desired language with the arrow keys next to the display (Page 10).
Set the date and time (Page 11).
"WARNING Fill with dental oil" and/or "WARNING Fill with cleaner" will be displayed.
Pull off top cover with both hands approx. 15 mm (overcome magnetic resistance) and pull off upwards.
Refill the media containers (also see chapter 4 "Maintenance").
Observe the color code on the media container! Blue = dental oil / green = cleaner
Only use original SciCan "Dental Oil" and SciCan "Cleaner".
Use of other media may put patients and users at risk and damage the device and the instruments.
Turn the metal flaps to the back in order to open the media container.
Fill up the container to the filling level marking. Upon exceeding the minimum filling level, a 1-second beep will sound. If the container is full, a 2-second beep will sound 2 times. The beep is always followed by a visual sign on the container.
Close the media container.
Attach the top cover and close until it stops.
Turn the device off using the power switch and then back on again.
The message "Fill media system" will be displayed after filling the dental oil and cleaner containers. The device will automatically start this function and run for about 3 minutes.
Filling the media system
This is only necessary if the function did not start automatically or if a container has been completely empty for more than 3 days. The containers must be filled already for this function.
Turn on the device
Press the button. "Export log data" will appear on the display.
Press the button once until "Fill media system" is highlighted in black on the display.
Using the button, the "Fill media system" menu is selected.
The media system will be filled automatically. During this time, a clock will run backwards until 00:00.
Press the button to return to the "Cleaning + Maintenance" main menu
Setting the language
Turn on the device
Press the button. "Export log data" will appear on the display.
Press the button 4x until "Language setting" is highlighted in black on the display.
The "Language setting" menu is selected using the button.
The desired language is selected using the button.
Save the language selection using the button.
Return to the "Language setting" menu using the button.
Press the button to return to the "Cleaning + Maintenance" main menu.
Page 10 STAT MATIC smart 1,2,3 EN
2 Commissioning
Setting the date/time
Turn on the device
Press the button. "Export log data" will appear on the display.
Press the button 3x until "Date/time" is highlighted in black on the display.
The "Date/time" menu is selected using the button.
Set the desired value using the and buttons
Using the button, the cursor is moved to the next possible setting. The values are set in the following order: Year, month, day, hour, minute.
After final actuation using the , the program jumps back again.
Press the button to return to the "Cleaning + Maintenance" main menu.
Assemble and check the adapter
Assemble the adapter, if required
Check the MULTIflex receptacle (and/or other turbine adapter) for firm seating.
The clamping nut must be tight.
Retighten clamping nut, if required.
Please note that only white STAT MATIC smart / STAT MATIC PLUS adapters can be used. Adapters (black) of the STAT MATIC maintenance unit are not suitable.
Check the output quantity on all cleaning slots using a testing adapter intended for the coupling.
Functional testing of flow rate using the testing adapter
Attach the matching testing adapter to the corresponding instrument couplings
(such as INTRAmatic, DIN ISO EN coupling).
Close the front cover.
Press the START button.
After completion of the program, check the filling quantity in each of the testing adapters.
The fluid must be above the mark ("Min.") !
If the filling quantity is below the mark, please consult the SciCan customer service.
Please note that only STAT MATIC smart testing adapters can be used. STAT MATIC
PLUS testing devices are not suitable.
After this process has completed, the STATMATIC smart system is operational.
Page 11 STAT MATIC smart 1,2,3 EN
3 Operation
3 Operation
Manual pre-cleaning of instruments
•
Clean the instrument primarily in the head area with a medium-hard toothbrush for 20 sec. under running water (non strong water jet) until the instrument looks fully clean. (drinking water, temperature 30 °C
5 °C).
•
Leave the instrument for 5 minutes.
"Cleaning + Maintenance" program - for regular and contra-angle handpieces and turbines
Turn on the mains switch.
LED (Stand-by) lights up green. "Cleaning + Maintenance" will appear on the display.
Open the front door.
Place regular and contra-angle handpieces and turbines on the adapters provided for this purpose (instruments must snap into place). A frame will appear on the display around the station number per occupied slot.
Close the front cover.
Push the (Program Start) button on the control membrane keyboard.
Program runs automatically. During the "Cleaning + Maintenance" program, a clock will count backwards until 00:00 on the display.
All attached regular and contra-angle handpieces and turbines are cleaned and maintained for inside in turn. After a successful process, the program slots will be displayed inversely. The display shows: "Cleaning + Maintenance successful".
Open the front door.
Remove regular and contra-angle handpieces and turbines by pressing the locking button (A). Pull off turbines downwards (press support lever down).
The frame around the slot number will disappear from the display.
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants listed by the VAH (such as SciCan OPT IM BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's instructions on reprocessing (such as disinfection/sterilization in the STAT IM ).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
"Collet care" program
Turn on the mains switch.
LED (Stand-by) lights up green. "Cleaning + Maintenance" will appear on the display.
Open the front door.
Push instrument with chuck opening into the collet care nozzle.
Activate (Collet care) key on the control membrane keyboard.
Program runs automatically. A clock on the display runs backwards until 00:00
The following is then shown on the display: " Collet care measure successful".
Remove instrument from the nozzle.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants listed by the VAH (such as SciCan OPT IM BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's instructions on reprocessing (such as disinfection/sterilization in the STAT IM ).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
Page 12 STAT MATIC smart 1,2,3 EN
3 Operation
The processes "Clean only" and "Maintenance only" can also be carried out individually.
"Maintenance only" program - for regular and contra-angle handpieces and turbines
Turn on the mains switch.
LED (Stand-by) lights up green. "Cleaning +
Maintenance" will appear on the display.
Open the front door.
Place regular and contra-angle handpieces and turbines on the adapters provided for this purpose (instruments must snap into place). A frame will appear on the display around the station number per occupied slot.
Close the front door.
Press the button 1x. "Maintenance only" will appear on the display.
Push the (Program Start) button on the control membrane keyboard.
Program runs automatically. During the maintenance program, a clock will count backwards until 00:00 on the display.
All attached regular and contra-angle handpieces and turbines are maintained for inside in turn. After a successful process, the program slots will be displayed inversely. The display shows: "Maintenance successful".
Open the front door.
Remove regular and contra-angle handpieces and turbines by pressing the locking button.
Pull off turbines downwards (press support lever).
The frame around the maintenance slot number will disappear from the display.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants listed by the VAH (such as SciCan OPT IM BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's instructions on reprocessing (such as disinfection/sterilization in the STAT IM ).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
"Clean only" program - for regular and contra-angle handpieces and turbines
Turn on the mains switch.
LED (Stand-by) lights up green. "Cleaning + Maintenance" will appear on the display.
Open the front door.
Place regular and contra-angle handpieces and turbines on the adapters provided for this purpose
(instruments must snap into place). A frame will appear on the display around the station number per occupied slot.
Close the front cover.
Press the button 2x. "Clean only" will appear on the display.
Push the (Program Start) button on the control membrane keyboard.
Program runs automatically. During the cleaning program, a clock will count backwards until 00:00 on the display
Page 13 STAT MATIC smart 1,2,3 EN
3 Operation
All attached regular and contra-angle handpieces and turbines are cleaned inside in turn. After a successful process, the program slots will be displayed inversely. The display shows: "Cleaning successful".
Open the front door.
Remove regular and contra-angle handpieces by pressing the locking button.
Pull off turbines downwards (press support lever).
The frame around the maintenance slot number will disappear from the display.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants listed by the VAH (such as SciCan OPT IM BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's instructions on reprocessing (such as disinfection/sterilization in the STAT IM ).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
Page 14 STAT MATIC smart 1,2,3 EN
3 Operation
"Special instrument" program
Special instruments can be maintained and cleaned using this program (such as prophylaxis). The times require for maintenance and cleaning can be individually adjusted by a service technician.
Turn on the mains switch.
LED (Stand-by) lights up green. "Cleaning + Maintenance" will appear on the display
Open the front door.
Place the instruments on the adapters provided for this purpose (instruments must snap into place). A frame will appear on the display around the station number per occupied slot.
Close the front cover.
Press the button 3x. "Special instrument" will appear on the display.
Push the (Program Start) button on the control membrane keyboard.
Program runs automatically. During the "Cleaning" program, a clock will count backwards until 00:00 on the display
All attached instruments are cleaned and maintained for inside in turn. After a successful process, the program slots will be displayed inversely. The display shows: "Special instrument successful".
Open the front door.
Remove regular and contra-angle handpieces by pressing the locking button.
Pull off turbines downwards (press support lever).
The frame around the slot number will disappear from the display.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants listed by the VAH (such as SciCan OPT IM BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's instructions on reprocessing (such as disinfection/sterilization in the STAT IM ).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
Page 15 STAT MATIC smart 1,2,3 EN
4 Maintenance
4 Maintenance
Checking adapter fixation
Recommended:
- After installation
- During operation: 1x week
Check the MULTIflex receptacle (and/or other turbine adapter) for firm seating..
The clamping nut must be tight.
Retighten clamping nut, if required.
Check the output quantity on all cleaning slots using a testing adapter intended for the coupling.
Please note that only STAT MATIC smart testing adapters can be used. STAT MATIC
PLUS testing devices are not suitable.
Checking the O-rings
Recommended:
- During operation: 1x week
Check all O-rings on the adapters for damage
Replace damaged O-rings
Type: 2
Functional testing of flow rate using the testing adapter
Recommended:
Upon initial commissioning
After changing an instrument coupling
During operation: 1x day before use
Attach the matching testing adapter to the corresponding instrument couplings
(such as INTRAmatic, DIN ISO EN Coupling).
Close the front cover.
Press the START button.
After completion of the program, check the filling quantity in each of the testing adapters.
The fluid must be above the mark ("Min.") !
If the filling quantity is below the mark, please consult the SciCan customer service.
Please note that only STAT MATIC smart testing adapters can be used. STAT MATIC
PLUS testing devices are not suitable.
Pressure test:
Recommended:
- Upon initial commissioning
- during operation: 1x day, preferably before starting work
Read off the "operating pressure" using the pressure gage.
Pressure should be between 4.5 bar and 4.3 bar!
Type: 3
Changing the adapter
The loosening of the clamping nut allows for the exchangeable adapter to be removed. Insert the desired adapter and tighten it again using the clamping nut.
Please note that only white STAT MATIC smart / STAT MATIC PLUS adapters can be used. Adapters (black) of the
STAT MATIC maintenance unit are not suitable.
Check the output quantity on the adapter using a testing adapter intended for the coupling.
Please note that only STAT MATIC smart testing adapters can be used. STAT MATIC PLUS testing devices are not suitable.
Page 16 STAT MATIC smart 1,2,3 EN
4 Maintenance
Refilling of the media containers
Check the filling levels of the containers every day prior to commencing work
Turn on the mains switch
The following will appear on the display: "WARNING Fill
Dental oil" or "WARNING Fill Cleaner"
Pull off top cover with both hands approx. 15 mm
(overcome magnetic resistance) and pull off upwards.
Observe the color code on the media container!
Blue = dental oil (SciCan "Dental Oil")
Green = cleaner (SciCan "Cleaner")
Turn the metal flaps to the back in order to open the media container.
Fill up the container to the filling level marking.
Upon exceeding the minimum filling level, a 1-second beep will sound.
If the container is full, a 2-second beep will sound 2 times.
The beep is always followed by a visual sign on the container.
Attach the top cover and close until it stops.
Only use original SciCan "Dental Oil" and SciCan "Cleaner"!
Use of other media may put patients and users at risk and damage the device and the instruments.
Page 17 STAT MATIC smart 1,2,3 EN
4 Maintenance
Replacing the absorption pads
Replace both pads after 35 instrument cleaning processes or at least 1x week (more often if needed).
Close the front door.
Pull the pads from the pins.
Remove pads upwards from fastening clips.
Dispose of pads in household waste.
Hook new fleeces into the fastening clips.
Fixate fleeces on the pins.
Close the front cover.
Only use original SciCan absorption fleeces.
Empty the collecting tray
Empty the collecting tray every day and clean/disinfect it
Open the front door.
Pull out tray forwards (horizontal) (overcome slight catch).
Empty the tray and clean/disinfect it with a disinfectant wipe.
For Germany we recommend using disinfectants listed by the VAH (such as
SciCan OPT IM BLUE).
Push tray into the side guides until it locks.
The rear fleece must not be pinched in, but rather has to hang loosely inside the tray.
Close the front cover.
Inside cleaning and disinfection of the maintenance area
Clean the maintenance area and door with disinfectant wipes every day. For Germany we recommend using disinfectants listed by the VAH (such as SciCan OPT IM BLUE).
Outside cleaning and outside disinfection of the device
The surface and control panel of the STAT MATIC smart system can be cleaned and disinfected using disinfectant wipes. For Germany we recommend using disinfectants listed by the VAH
(such as SciCan OPT IM BLUE).
Recommendation
SciCan recommends an annual maintenance measure to be performed by an authorized dealer or the manufacturer.
NOTE: For users in countries with a legal validation requirement
•
An annual maintenance by the manufacturer or an authorized specialist company is recommended.
•
Please observe the national specifications for validation. An authorized specialist company must carry out a first validation with performance qualification (PQ) in accordance with the legal requirements if validation is required. Please observe the validation instructions for the device.
•
SciCan recommends that a revalidation be carried out at intervals of one year (acc. to EN ISO 15883-1), unless shorter cycles are required by local authorities or the validator. These revalidations should be carried out during or following maintenance work.
Page 18 STAT MATIC smart 1,2,3 EN
4 Maintenance
Data backup of the log file
Switch off the device
Insert the USB stick on the back of the device
Turn on the device
Press the button. "Export log data" will appear on the display.
Use the button to start exporting the log files to the USB stick.
When the process is finished, the display shows the following: Log data export complete.
USB connection
(rear panel of the device)
Use the button to return to the menu.
Switch off the device
Remove the USB stick.
•
If the USB port is to be extended using a USB cable, this cable must be equipped with a ferrite core. The cable must be inserted with the ferrite core facing the device side.
•
Please note that the data on the USB stick must be copied to a suitable permanent storage device. Please check that the data on the storage medium can also be read correctly.
Deletion of the log file on the STATMATIC smart
Deletion of the log file causes the data to be permanently lost. Please create a backup of the log file first and check whether the data on the USB stick can also be read correctly.
Turn on the device
Press the button. "Export log data" will appear on the display.
Press the button 6x until "Delete log file" is highlighted in black on the display.
"Deletion" is selected using the button.
Press the button once more. The deletion of the log data will be performed.
Use the button to return to the menu.
Press the button to return back to the "Cleaning + Maintenance" main menu
Emptying the media system
This is only required if the system is to be emptied completely - for example for transport purposes or longer standstill periods (>30 days).
First, empty the containers completely using drain hoses. Then, tilt the device to the rear by approx. 45 degrees in order to also empty the residual liquids located at the bottom of the container using the drain hoses. Media can only be re-used again if they have been kept in their original bottles (mind the best before dates on the bottom or on the sides of the original bottles).
Turn on the device
Press the button. "Export log data" will appear on the display.
Press the button 2x until "Empty media system" is highlighted in black on the display.
Using the button, the "Empty media system" menu is selected.
The media system will be automatically emptied. During this time, a clock will run backwards until 00:00.
Press the button to return to the "Cleaning + Maintenance" main menu
Page 19 STAT MATIC smart 1,2,3 EN
4 Maintenance
Warning message: Servicing required soon
This warning message appears if more than 25,000 regular and contra-angle handpieces and turbines have been cleaned and maintained for with the STAT MATIC smart. Please contact the service department. You can clean a further
3,300 regular and contra-angle handpieces and turbines.
Error message: Servicing required
This error message appears if more than 28,300 regular and contra-angle handpieces and turbines have been cleaned and maintained for with the STAT MATIC smart. Please contact the service department.
Transport
•
If the device was already filled with dental oil and cleaner, the media must be emptied out fully through the drain hoses.
•
Empty the media system (see p. 20)
•
Unscrew the safety screw of the media aeration and blow compressed air into the valve several times. Tighten the screw again.
•
Place the accompanying seal plugs in the filling funnel of the media container (see fig.)
•
Close the flaps.
•
If possible, send in the original box.
Page 20 STAT MATIC smart 1,2,3 EN
5 Troubleshooting
5 Troubleshooting
Error
LEDs don't light up
Compressed air
Top cover open
Program doesn't start
Process interrupted
Undefined fault
Memory full
Display
No display
ATTENTION
Compressed air error
Fault LED (L2) flashes
WARNING
Close the top cover
Fault LED (L2) flashes
ATTENTION
Close the door
ATTENTION
No slots occupied
Acoustic signal
–
Interval
0.5 seconds on,
8 seconds off;
10 times
–
–
–
– Beep
ATTENTION
Process interrupted
–
ATTENTION
Please consult the customer service.
ATTENTION
Not enough memory available
Interval
– Beep
2 seconds on,
4 seconds off
–
Cause Correction
No power supply
Defective fuse
Compressed air hose not connected
Compressed air hose pinched
Activate mains switch
Plug in the power supply plug
Replace the fuse
(see Chapter 6
"Troubleshooting")
Connect compressed air hose
Remove flow disruption
Mains supply defective Create supply
Pressure reducer Please consult the SciCan defective
Pressure switch customer service.
Please consult the SciCan defective
If available; filter in supply line full customer service.
Replace the filter
Top cover not closed correctly
Magnet on top cover is missing
Close top cover tightly
Please consult the SciCan customer service.
Front doors are not correctly closed
No instruments in the device
Switching flap does not recognize instrument
Front doors are not correctly closed
Instrument is not attached to the coupling correctly
Internal device fault
Close front doors tightly
Attach the instruments to the couplings
Turn the instrument until a frame is displayed around the station number
Close front doors tightly.
Push the instrument firmly onto the coupling
Please consult the SciCan customer service.
Internal memory full Export log data to USB stick
Page 21 STAT MATIC smart 1,2,3 EN
5 Troubleshooting
Error
Cleaner empty
Oil empty
WARNING
Fill Dental oil or
Cleaner, even though the respective container is full.
Display
WARNING
Fill the cleaner
Fault LED (L2) flashes
WARNING
Fill in dental oil
Fault LED (L2) flashes
WARNING
Fill Dental oil or
Cleaner
Fault LED (L2) flashes
Acoustic signal
Beeping sound
2 seconds
Beeping sound
2 seconds
Beeping sound
2 seconds
Cause
Cleaner container empty
Oil container empty
Air in the fill measurement system
Export data
Liquids (dental oil and cleaner) have been switched around
ATTENTION
Storage medium not readable
Fault LED (L2) flashes
–
–
–
USB stick not recognized
USB stick not formatted correctly
Correction
Fill the cleaner container with
SciCan "Cleaner"
Fill the oil container with SciCan
"Dental Oil"
Fully empty out the affected container through the drain hose.
Tilt the device back 45°.
Unscrew the safety screw of the media aeration and blow compressed air into the valve several times. Tighten the screw again and fill the container.
Turn off the device, insert the
USB stick, turn on the device
- Perform NTFS formatting
- Use another USB stick
Empty media container:
Drain liquids from the container using a drain hose.
Draw any residual liquid from the corresponding media container using a 50 ml syringe.
Completely remove residue of the "false "medium from the media container with a highly absorbent textile cloth that is free of fibers.
Turn on the device
Select "Empty media system" on the display (see Chapter 4
"Maintenance")
Refill dental oil and cleaner
(see Chapter 4
"Maintenance")
Open the front door
Select the "Fill media system" menu item in the display
Close the front door
After this process has completed, the device is operational.
Replacing fuses
Remove the STAT MATIC smart device plug.
Remove fuse tray from the power input module (push both tabs on the fuse tray inward).
Replace fuses.
Insert fuse tray again.
Only use the fuse type indicated in the technical data.
Page 22 STAT MATIC smart 1,2,3 EN
6 Information on Electromagnetic Compatibility (EMC)
6 Information on Electromagnetic Compatibility (EMC)
The STAT MATIC smart is designed for operation in an environment as described below. The customer or the user of the STAT MATIC smart should ensure that it is operated in an environment of this kind.
Electromagnetic interference measurements
HF emissions according to CISPR 11
Compliance
Group 1
Electromagnetic environment
The STAT MATIC
– guidelines
smart utilizes HF energy exclusively for its internal function. Its HF emissions are therefore very low and it is unlikely that neighboring electronic devices will be disrupted.
HF emissions according to CISPR 11
Emissions from harmonics according to IEC
61000-3-2
Emissions from voltage fluctuations / flickers according to IEC 61000-3-3
Class B
Class A
Conforms
The STAT MATIC smart is suitable for usage in all facilities, including those located in living areas and those directly connected to a public power supply also supplying buildings used for residential purposes.
Susceptibility tests
Discharge of static electricity (ESD) according to IEC 61000-4-2
Rapid, transient electrical disturbances/ bursts according to IEC
61000-4-4
Surges according to IEC 61000-4-5
Voltage drops, short-term interruptions and fluctuations of the supply voltage according to IEC 61000-4-11
IEC 60601 test level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for power cables
± 2 kV for input and output lines
± 1 kV normal mode voltage
± 2 kV common mode voltage
0 % UT for 0.5 periods
40 % UT for 5 periods
70 % UT for 25 periods
0 % UT for 250 periods
3 A/m
Compliance level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV, not applicable for power cables
Not applicable
± 1 kV normal mode voltage
± 2 kV common mode voltage
0 % UT for 0.5 periods
40 % UT for 5 periods
70 % UT for 25 periods
0 % UT for 250 periods
Not applicable
Electromagnetic environment – guidelines
Flooring should be of wood or concrete, or be covered with ceramic tiles. When the floor is covered with synthetic material, the relative air humidity must be at least 30 %.
The quality of the supply voltage should correspond to that of a typical business or hospital environment.
The quality of the supply voltage should correspond to that of a typical business or hospital environment.
The quality of the supply voltage should correspond to that of a typical business or hospital environment. When the user of the STAT MATIC smart also requires continuing function with the occurrence of interruptions of the power supply, it is recommended that the STAT MATIC smart draw power from an uninterruptable power supply.
Not applicable Magnetic field of the supply frequency
(50/60 Hz) according to IEC 61000-4-8
Conducted HF disturbance according to IEC 61000-4-6
Radiated
HF disturbances according to IEC 61000-4-3
3 Veff
150 kHz to 80 MHz outside of the ISM bands a
10 V/m
80 MHz to 2.5 GHz
10 Veff
10 V/m
Portable and mobile radio devices should be used at a distance from the STAT MATIC , including the cable, no less than the recommended protective distance calculated according to the equation applicable to the transmission frequency.
Recommended protective distance: d = 1.17
P d = 0.35
P for 80 MHz to 800 MHz d = 0.70
P for 800 MHz to 2.5 GHz with P as the maximum power rating of the transmitter in watts (W) in accordance with the information of the transmitter manufacturer and d as the recommended protective distance in meters
(m). b
The field strength of stationary radio transmitters should be less than the compliance level for all frequencies according to experiments carried out on location. d Interference is possible in the vicinity of devices bearing the icon shown below.
Page 23 STAT MATIC smart 1,2,3 EN
6 Information on Electromagnetic Compatibility (EMC)
NOTE U
T
: is the alternating supply voltage prior to application of the test levels.
NOTE 1: The higher frequency range applies for 80 MHz and 800 MHz.
NOTE 2: These guidelines may not be applicable in all cases. The spread of electromagnetic disturbances is influenced by absorption and reflection of the building, objects and people. a
The ISM frequency bands (for industrial, scientific and medical applications) between 150 kHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz. b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range between 80 MHz and 2.5
GHz are intended to reduce the probability of mobile/portable communication devices causing any interference whenever they are inadvertently brought into patient areas. For this reason, the additional factor of 10/3 is sued for the calculation of the recommended protective distances in these frequency ranges. c The field strengths of fixed transmitters, such as base stations for radio telephones and mobile radios, amateur radio stations, AM and
FM radio broadcasts as well as TV broadcasts cannot be predicted in a fully accurate manner. In order to determine the electromagnetic environment with respect to the stationary transmitter, a study of the location should be considered. When the measured field strengths at the location at which the STAT MATIC smart is used exceeds the compliance level above, the STAT MATIC smart should be observed in order to demonstrate correct functioning. When unusual performance features are observed, additional measures may be necessary, such as a changed orientation or another location for the STAT MATIC smart. d Above the frequency range of 150 kHz to 80 MHz, the field strength should be less than 3 Veff V/m.
Recommended protective distances between portable and mobile HF telecommunications devices and the STAT MATIC smart.
The STAT MATIC smart is designed for operation in an electromagnetic environment in which the HF disturbances are controlled. The customer or the user of the STAT MATIC smart can help avoid electromagnetic disturbances by observing the minimum distance between portable and mobile HF telecommunications devices (transmitters) and the STAT MATIC smart, depending upon the output power of the communications device, as indicated below.
Rated power of the transmitter / W
Protective distance depending upon the transmission frequency in m
150 kHz to 80 MHz d=0.35
P 80 MHz to 800 MHz d=0.35
P 800 MHz to 2.5 GHz d=0.70
P
0.01
0.1
1
10
0.04
0.11
0.35
1.11
0.04
0.11
0.35
1.11
0.07
0.22
0.70
2.21
100 3.50 3.50 7.00
For transmitters with a maximum rated power not shown in the table above, the recommended protective distance d can be determined in meters (m) using the equation form the respective column, whereby P is the maximum rated power of the transmitter in watts (W) in accordance with the information provided by the transmitter manufacturer.
NOTE 1: The higher frequency range applies for 80 MHz and 800 MHz.
NOTE 2: These guidelines may not be applicable in all cases. The spread of electromagnetic disturbances is influenced by absorption and reflections of the building, objects and people.
Page 24 STAT MATIC smart 1,2,3 EN
7 Accessories
7 Accessories
Please order accessories only from your authorized dealer.
Article no. Description
S32102
SciCan "Cleaner"
(6x 500 ml)
S32101
SciCan "Dental Oil"
(2x 500 ml)
S32100
Absorption pad (set)
(10 pieces)
S32114
Compressed air hose
(line)
S30116 MULTIflex* O-Ring set
S32118 INTRAmatic O-Ring set
S32101-CX
Maintenance coupling
MULTIflex*/ CONNEX
S32101-ST1T
Maintenance coupling
Sirona T1 turbine
S32101-NSK
Maintenance coupling
NSK turbine
S32101-W&H
Maintenance coupling
W&H turbine
S32101-BA
Maintenance coupling
BienAir turbine
S33126 Seal plugs
*MULTIflex is a registered trademark of
Kaltenbach & Voigt
Article no. Description
S30146 Chuck tip
S32117 Fuse, T 2.0 A (H)
S30108 Rubber base
S30122
Maintenance coupling gasket
S33120 Front door seal
S32201-E
S32201-CX
Testing adapter
INTRAmatic
DIN EN ISO 3964-
Coupling
Testing adapter
MULTIflex*
S32201-ST1T
Testing adapter
Sirona T1 turbine
S32201-NSK
Testing adapter
NSK turbine
S32201-W&H
Testing adapter
W&H turbine
S32201-BA
Testing adapter
BienAir turbine
S32202
Spare test tube
STAT MATIC smart
Page 25 STAT MATIC smart 1,2,3 EN
8 Technical Data
8 Technical Data
Device:
Device dimensions:
Weight:
Power ratings:
Protection class:
Protection:
Installation category:
Contamination level:
Ambient temperature:
Noise level:
Relative air humidity:
Max. operating altitude:
Operating location:
Compressed air:
Air consumption:
Compressed air quality:
Power fuses:
Number of ISO 3964 integrated couplings:
Width:
Depth:
Height:
Rated voltage:
Rated frequency:
Rated power:
Residual oil content:
Residual dust:
Residual water:
Rated current:
Rated voltage:
Triggering characteristic:
Breaking capacity:
Type 1
Type 2
Type 3
Device:
Ambient temperature:
Relative air humidity:
Compressed air:
Storage conditions
5
– 35 °C C (41 – 95 °F)
5
– 95 %
700 - 1060 hPa
Empty the media containers prior to transport! (see Chapter 4 "Maintenance")
Media: Storage conditions
Ambient temperature
– dental oil:
5
– 50 °C (41 – 122 °F)
Ambient temperature – cleaner: 5 – 35 °C (41 – 95 °F)
Relative air humidity: 5 - 95 %
Compressed air: 800 - 2500 hPa
Transport conditions
-30
– 50 °C (-22 – 122 °F)
-30 – 50 °C (-22 – 122 °F)
5 - 95 %
800 - 2500 hPa
300 mm (11.8 inches)
280 mm (11.0 inches)
345 mm (13.6 inches)
7.3 kg (15.8 lbs)
100
– 230 V
10%
50 / 60 Hz
45 VA max.
I covered
Cat. II
P 2
5 – 35 °C (41 – 95 °F)
60 (65) dB(A)
85 % max.
2000 m above sea level only indoors
4.5 – 6 bar (65 – 87 psi)
40 Nl/min max. 0.1 mg/m³ max. 1 µm; max. 1 mg/m³ max. 0.1 g/m³ at -40 °C (-40 °F)
2 A
250 V
Large (T)
H
1
2
3
Transport conditions
-30
– 50 °C (-22 – 122 °F)
5 - 95 %
700 - 1060 hPa
Page 26 STAT MATIC smart 1,2,3 EN
8 Technical Data
Type plate
STATMATIC smart 1
STATMATIC smart 2
STATMATIC smart 3
The symbols used are described in Chapter 1 "Important instructions"
Page 27 STAT MATIC smart 1,2,3 EN
9 Documentation, Log File
9 Documentation, Log File
Page 28 STAT MATIC smart 1,2,3 EN
10 Guarantee
10 Guarantee
Limited warranty
(DE = original)
SciCan extends a guarantee of two years (from the date of purchase) against any malfunction of the
STATMATIC smart due to material or processing errors not resulting from obvious misuse, false use or accidents, with the foregoing applying to devices in a new and unused condition under normal conditions of use.
The two-year guarantee applies to the functional capability of all parts of the device, provided that the device is used and maintained in accordance with the description included in the user manual.
In the event of any malfunction that can be traced back to such defects arising during the guarantee period,
SciCan is obligated only to either carry out repairs at no charge or to replace any defective parts (with the exception of O-rings and seals) at no charge at our own discretion, on condition that we are notified in writing within 30 days of the occurrence of such a malfunction and that the defective parts are also returned freight paid.
This guarantee will be considered to have been agreed upon when the product is accompanied by the original invoice issued by our authorized SciCan dealer and when that invoice identifies the item with an item number and clearly indicates the purchase date. No other type of receipt is acceptable. After the expiration of two years, all of the guarantees provided by SciCan , as well as other obligations with respect to the quality of the product are considered to be conclusively fulfilled. Any liability is thus terminated and no legal proceedings may thereafter be initiated against SciCan on account of a breach of any such guarantee or obligation.
We exclude and reject any specific guarantee not indicated here and any implied guarantee or representation with reference to performance capability, as well as any legal remedy in the event of a contractual violation, which, apart from this provision, could arise due to implication, legal interpretation, commercial practice or custom of trade, including any implied guarantee with respect to saleability or suitability for a particular purpose with reference to one or all products manufactured by SciCan . To find out more about SciCan products and services, please visit our website at www.scican.com.
Page 29 STAT MATIC smart 1,2,3 EN
STAT MATIC smart 1,2,3 EN
Gebrauchsanweisung
STATMATIC
smart 1, 2, 3
Reinigungs- und Pflegesystem
DE
Brugsvejledning DK via: www.scican.uk.com/downloads
Manuel d'instructions FR via: www.scican.uk.com/downloads
Istruzioni per l’uso IT via: www.scican.uk.com/downloads
Gebruiksaanwijzing NL via: www.scican.uk.com/downloads
www.scican.com
STAT MATIC smart 1, 2, 3
Copyright 2017 SciCan. Alle Rechte vorbehalten.
Doc. No. 2.001.8462
– Rev. 2017-11
Inhaltsverzeichnis
1 Wichtige Hinweise ......................................................33
Typenbeschreibung ...................................................33
Lieferumfang ..............................................................33
Allgemein ...................................................................33
Zweckbestimmung .....................................................34
Entsorgung ................................................................34
Qualifikation des Anwenders .....................................35
Sicherheitshinweise SciCan Dentalöl und SciCan
Reiniger .....................................................................35
Verwendete Symbole .................................................35
Haftungsausschluss ...................................................36
Vor Inbetriebnahme ...................................................36
Geräteübersicht .........................................................37
2 Inbetriebnahme ..........................................................39
Umgebungsbedingungen ...........................................39
Elektrische Anforderungen .........................................39
Medienanschlüsse .....................................................39
Mediensystem füllen ..................................................40
Sprache einstellen .....................................................40
Datum / Uhrzeit einstellen ..........................................41
Adapter montieren und überprüfen ............................41
Funktionsprüfung der Durchflussmenge mit
Prüfadapter ................................................................41
3 Bedienung ..................................................................42
Manuelle Vorreinigung der Instrumente .....................42
Programm "Reinigen + Pflege" - für Hand-
/Winkelstücke und Turbinen .......................................42
Programm "Spannzangenpflege" ...............................42
Programm "Nur Pflege - für Hand-/Winkelstücke und
Turbinen .....................................................................43
STAT MATIC ist ein Warenzeichen; OPT IM und STAT IM sind eingetragene
Warenzeichen von SciCan Ltd. Alle anderen in diesem Handbuch erwähnten
Warenzeichen sind Eigentum ihrer jeweiligen Besitzer.
Hergestellt von:
SycoTec GmbH & Co. KG
Wangener Strasse 78
88299 Leutkirch
DEUTSCHLAND
Telefon:
Fax:
+49 7561 86-0
+49 7561 86-266
Programm "Sonderinstrument" .................................. 45
4 Wartung ..................................................................... 46
Adapterfixierung überprüfen ...................................... 46
O-Ringe prüfen .......................................................... 46
Funktionsprüfung der Durchflussmenge mit
Prüfadapter ................................................................ 46
Adapter wechseln ...................................................... 46
Nachfüllen der Medienbehälter .................................. 47
Absorptions-Vliese wechseln ..................................... 48
Auffangschale entleeren ............................................ 48
Innenreinigung und Innendesinfektion des
Pflegeraums .............................................................. 48
Außenreinigung und Außendesinfektion des Gerätes 48
Empfehlung ............................................................... 48
Hinweis für Benutzer in Ländern mit gesetzlicher
Validierungspflicht: .................................................... 48
Datensicherung der Log-Datei ................................... 49
Löschen der Log-Datei auf dem STAT MATIC smart . 49
Mediensystem leeren ................................................ 49
Warnmeldung: "Service wird bald benötigt" ............... 50
Fehlermeldung: "Service wird benötigt" ..................... 50
Transport ................................................................... 50
5 Fehlerbehebung ........................................................ 51
Sicherungswechsel ................................................... 52
6 Informationen zur elektromagnetischen
Verträglichkeit (EMV) ................................................ 53
7 Zubehör ..................................................................... 55
8 Technische Daten ..................................................... 56
Typenschild ............................................................... 57
9 Dokumentation Log-Datei .......................................... 58
10 Garantie..................................................................... 59
SciCan GmbH
Wangener Strasse 78
88299 Leutkirch, DEUTSCHLAND
Telefon:
Fax:
+49-7561-98343-0
+49-7561-98343-699
SciCan Ltd.
1440 Don Mills Road
Toronto, ON M3B 3P9, KANADA
Telefon:
Fax:
Gebührenfrei:
E-Mail:
+1-416-445-1600
+1-416-445-2727
+1-800-667-7733 [email protected]
SciCan Medtech AG
Alpenstrasse 14
6300 Zug, SCHWEIZ
Telefon:
Fax:
SciCan Inc.
+41-41-727-70-27
+41-41-727-70-29
701 Technology Drive
Canonsburg, PA 15317, USA
Telefon:
Fax:
+1-724-820-1600
+1-724-820-1479
Seite 32 STAT MATIC smart 1,2,3 DE
1 Wichtige Hinweise
1 Wichtige Hinweise
Gratulation zum Kauf Ihres STAT MATIC smart. Wir sind davon überzeugt, dass Sie das beste Produkt seiner Art erworben haben. Das STAT MATIC smart ist ein kompaktes Reinigungs- und Pflegegerät, das mit seinen verschiedenen
Programmen all Ihren Bedürfnissen und Erwartungen bei der Instrumentenreinigung und -pflege gerecht wird.
Diese Gebrauchsanweisung enthält alle Einzelheiten zu Einrichtung, Betrieb und Wartung Ihres STAT MATIC smart. Zur
Gewährleistung eines langjährigen, sicheren und problemfreien Betriebs sollten Sie die Anweisungen vor
Inbetriebnahme des Gerätes lesen und sie für künftige Verwendung aufbewahren.
Die Anweisungen zu Betrieb, Wartung und Austausch von Ersatzteilen müssen befolgt werden, damit das Gerät ordnungsgemäß arbeiten kann. Der Inhalt dieser Gebrauchsanweisung kann ohne Vorankündigung geändert werden, um Veränderungen und Verbesserungen am STAT MATIC smart wiederzugeben.
Typenbeschreibung
Das STAT MATIC smart wird in 3 unterschiedlichen Typen hergestellt.
STAT MATIC smart Typ 1 hat eine DIN EN ISO 3964 Kupplung, welche motorisch betrieben wird und 2 neutrale
Kupplungen, die individuell mit den benötigten Turbinenadaptern bestückt werden können (siehe Zubehör).
STAT MATIC smart Typ 2 hat zwei DIN EN ISO 3964 Kupplungen, welche motorisch betrieben werden und 1 neutrale
Kupplung, die individuell mit dem benötigten Turbinenadapter bestückt werden kann (siehe Zubehör).
STAT MATIC smart Typ 3 hat drei DIN EN ISO 3964 Kupplungen, welche motorisch betrieben werden.
Lieferumfang
Basisgerät STAT MATIC smart
Netzanschlussleitung
Gebrauchsanweisung
Absorptions-Vlies (Set)
Druckluftschlauch
SciCan "Dental Oil" (Dentalöl) (250 ml)
SciCan "Cleaner" (Dentalreiniger) (500 ml)
O-Ring Set
2 Verschlussstopfen (rot) für Medienbehälter
USB-Stick (2GB)
Allgemein
Das STAT MATIC smart wurde nach den neuesten Qualitäts- und Sicherheitsstandards entwickelt und hergestellt.
Einige Vorkehrungen sollten Sie bei Installation und Betrieb jedoch treffen.
•
Beim Verlassen der Praxis für eine längere Zeit ist das STAT MATIC smart vom Stromnetz zu trennen.
•
STAT MATIC smart vor Erwärmung über 50 °C schützen.
•
Bei Änderungen durch Dritte erlischt die Garantie und die Zulassung.
•
Zum Betrieb bzw. zur Reparatur dürfen nur SciCan Original-Ersatzteile verwendet werden, da diese in umfangreichen
Versuchen auf ihre Sicherheit, Funktion und spezifische Tauglichkeit geprüft wurden. SciCan übernimmt damit für diese Teile auch die Produktverantwortung.
•
Beachten Sie alle Warnhinweise am Gerät und in der Gebrauchsanweisung.
•
Das Gerät darf nur mit der Spannung versorgt werden, die auf dem Typenschild vermerkt ist.
•
Das Netzkabel darf nicht durch ein zu schwach dimensioniertes Netzkabel ersetzt werden.
•
Anwender dürfen über die in der Gebrauchsanweisung hinausgehenden Punkte nicht versuchen das Gerät selbst zu warten.
•
Bei Inbetriebnahme, Bedienung und Wartung des Gerätes sind medizinische Handschuhe zu tragen.
•
Das Gerät darf nur mit Original SciCan "Cleaner" (Reiniger) und SciCan "Dental Oil" (Dentalöl) sowie Zubehör betrieben werden.
•
Vor dem Transport oder vor längerem Stillstand (> 30 Tage) des Gerätes sind die Behälter für Reiniger und Dentalöl zu entleeren.
•
Der Anwender hat sich vor jeder Anwendung des Geräts von dem ordnungsgemäßen Zustand, sowie der
Funktionssicherheit des Geräts, zu überzeugen.
•
Ein beschädigtes Gerät, oder ein Gerät mit beschädigter Netzleitung muss sofort vom Stromnetz getrennt werden. Ein weiterer Betrieb ist nicht zulässig.
•
Die Verwendung des Gerätes ist nur in Innenräumen und nicht explosionsgefährdeten Bereichen zugelassen.
•
Niemals das Schutzgehäuse des Gerätes öffnen oder Gegenstände durch die Öffnungen des Schutzgehäuses einführen.
Seite 33 STAT MATIC smart 1,2,3 DE
1 Wichtige Hinweise
Zweckbestimmung
•
Das STAT MATIC smart ist ein Reinigungs- und Pflegegerät für die maschinelle Wiederaufbereitung (Innenreinigung und Pflege) von dentalen Turbinen sowie Hand- und Winkelstücken.
•
Für weitere Informationen über die Eignung bestimmter Instrumente für die Reinigung/Pflege sind die
Herstelleranweisungen zu beachten.
Die Aufbereitung im STAT MATIC smart beschränkt sich auf die Innenreinigung und Pflege der Instrumente. Die
Außenflächen der Instrumente müssen vor der Innenreinigung und Pflege manuell oder maschinell gereinigt werden.
Nach der Behandlung mit dem Reinigungs- und Pflegesystem, das Instrument unmittelbar, gemäß den jeweiligen
Herstellerangaben, weiterbehandeln.
Hinweis: Für die vorangestellten und die nachfolgenden Aufbereitungsschritte sind die RKI-Richtlinien und das
Medizinproduktegesetz zu beachten.
Missbrauch
Alle nicht in der Zweckbestimmung aufgeführten Anwendungen stellen einen Missbrauch dar und können zu Schäden am Gerät und den aufzubereitenden Medizinprodukten und/oder zu einer Gefährdung des Anwenders führen. Hierfür
übernimmt SciCan keine Haftung.
Entsorgung
Verbrauchsmaterialien:
Die entstehenden Abfälle sind für Mensch und Umwelt gefahrfrei der stofflichen Verwertung oder der Beseitigung zuzuführen, wobei die geltenden nationalen Vorschriften einzuhalten sind.
Geräte sowie Zubehör am Ende der Nutzungsdauer:
Auf Basis der EU-Richtlinie (WEEE 2012/19/EU) über Elektro- und Elektronik-Altgeräte weisen wir darauf hin, dass das vorliegende Produkt der genannten Richtlinie unterliegt und innerhalb Europas einer speziellen Entsorgung zugeführt werden muss.
Nähere Informationen erhalten Sie vom Hersteller oder vom dentalen Fachhandel.
Dieses Produkt für Mensch und Umwelt gefahrfrei der stofflichen Verwertung oder der Beseitigung zuführen, dabei die geltenden nationalen Vorschriften einhalten.
Seite 34 STAT MATIC smart 1,2,3 DE
1 Wichtige Hinweise
Qualifikation des Anwenders
Das STAT MATIC smart darf nur von ausgebildetem Personal angewendet werden, welches Zahnärzte/Zahnärztinnen sowie deren Assistent/Innen einschließt, die eine Einweisung für dieses Produkt erhalten haben.
Sicherheitshinweise SciCan "Dental Oil" (Dentalöl) und SciCan "Cleaner" (Reiniger)
Erste-Hilfe-Maßnahmen - Dentalöl
Allgemeine Hinweise Niemals einer bewusstlosen Person oder bei auftretenden Krämpfen etwas über den
Mund verabreichen. Person in Sicherheit bringen.
Nach Einatmen
Die Person an die frische Luft bringen und für ungehinderte Atmung sorgen
Nach Hautkontakt
Mit reichlich Wasser abwaschen. Kontaminierte Kleidung ist sofort zu wechseln.
Ärztliche Behandlung notwendig
Nach Augenkontakt
Nach Verschlucken
Sofort vorsichtig und gründlich mit Augendusche oder mit Wasser spülen. Bei
Augenreizung einen Augenarzt aufsuchen.
Bei Erbrechen Aspirationsgefahr beachten. Sofort Mund ausspülen und reichlich
Wasser nachtrinken. Erbrechen herbeiführen, wenn die betroffene Person bei
Bewusstsein ist. Ärztliche Behandlung notwendig.
Maßnahmen zur Brandbekämpfung
Das Produkt ist nicht brennbar.
Erste-Hilfe-Maßnahmen - Reiniger
Allgemeine Hinweise
Niemals einer bewusstlosen Person oder bei auftretenden Krämpfen etwas über den
Mund verabreichen. Person in Sicherheit bringen.
Nach Einatmen
Für Frischluft sorgen.
Nach Hautkontakt
Mit reichlich Wasser abwaschen. Kontaminierte Kleidung wechseln..
Nach Augenkontakt
Nach Verschlucken
Sofort vorsichtig und gründlich mit Augendusche oder mit Wasser spülen.
Sofort Mund ausspülen und reichlich Wasser nachtrinken.
Maßnahmen zur Brandbekämpfung
Das Produkt ist nicht brennbar.
Verwendete Symbole
Im Handbuch:
Eine potenzielle Gefahr für den
Bediener.
Eine Situation, die zu einem mechanischen Defekt führen kann
Auf dem Gerät:
LED Betriebsbereit
LED Fehler
CE Kennzeichnung
Sicherung
Gebrauchsanweisung befolgen
Wichtige Information Hersteller
Entsorgungshinweis, s. Wichtige
Informationen "Entsorgung"
VDE Prüfzeichen MET-Prüfzeichen
Medizinische Handschuhe tragen
Menü / Auswählen /
Speichern
Auswahl nach oben
Programmstart /
Auswahl nach unten
Spannzangenpflege /
Zurück
USB-Anschluss
Seite 35 STAT MATIC smart 1,2,3 DE
1 Wichtige Hinweise
Haftungsausschluss
Der Einbau von Ersatzteilen, die Wartung und die Instandhaltung dürfen nur durch autorisiertes Personal erfolgen.
SciCan übernimmt keinerlei Haftung für zufällige, besondere oder Folgeschäden, die durch Instandsetzung oder
Wartung des STAT MATIC smart durch Dritte entstehen, oder für die Verwendung von Ausrüstung oder Einzelteilen aus der Herstellung Dritter, einschließlich Gewinneinbußen, jeglicher gewerblicher Verluste, wirtschaftlicher Verluste oder
Verluste durch Verletzungen von Personen.
Entfernen Sie niemals die Rückwand des Gerätes und führen Sie auch keine Gegenstände durch die Löcher oder
Öffnungen am Gehäusekasten ein. Zuwiderhandlung kann zu Schäden am Gerät und/oder zur Gefährdung des
Bedieners führen.
Vor Inbetriebnahme
Das STAT MATIC smart muss vor Inbetriebnahme gemäß der Checkliste "Installationsqualifikation" (siehe beiliegendes Medizinproduktebuch) ordnungsgemäß aufgestellt werden.
Seite 36 STAT MATIC smart 1,2,3 DE
1 Wichtige Hinweise
Geräteübersicht
(1) Abdeckung
(2) Bedienfolie
(3) Steckplatz für Turbinenadapter
(2x Typ 1 / 1x Typ 2) oder
Pflegekupplungen INTRAmatic ISO 3964
(3x Typ 3 / 2x Typ 2 / 1x Typ 1)
(4) Auffangschale
(5) Klappe
(6) Absorptions-Vlies (2x)
(7) Spannzangenpflegedüse
(8) Displayanzeige
(9) USB Stick
(10) Füllöffnung Dentalöl
(11) Füllöffnung Reiniger
(12) Halter für Prüfadapter
(21) USB-Anschluss
(22) Medienbelüftung
(13) Netzschalter
(14) Netzsicherungsschublade
(15) Eingang Netzleitung
(16) Ablassschlauch Dentalöl und Reiniger
(17) Druckanzeige
(18) Druckeinstellung
(19) Drucklufteingang
(20) Typenschild
Seite 37 STAT MATIC smart 1,2,3 DE
1 Wichtige Hinweise
Bedienfolientastatur
(L1) LED Stand-by
(L2) LED Fehler
(S1) Programmauswahl / Auswahl nach oben
(S2) Programmstart / Auswahl nach unten
(D1) Display
(S3) Spannzangenpflege / Zurück
(S4) Menü / Auswählen / Speichern
Displayanzeige
Datum / Uhrzeit
Statuszeilen
Plätze
Seite 38
Platznummer
Platz ist fertig bearbeitet
Platz ist belegt
Rahmen blinkt:
Platz in Bearbeitung
STAT MATIC smart 1,2,3 DE
2 Inbetriebnahme
2 Inbetriebnahme
Umgebungsbedingungen
Es gibt eine Reihe von Faktoren, die die Leistung Ihres STAT MATIC smart beeinflussen können. Schauen Sie sich bitte diese Faktoren genau an und wählen Sie daraufhin einen passenden Ort zum Aufstellen des Gerätes.
• Temperatur und Luftfeuchtigkeit
Stellen Sie Ihr STAT MATIC smart nicht an einem Ort mit direkter Sonneneinstrahlung oder in der Nähe einer
Heizquelle (z.B. Heizkörper) auf. Die empfohlene Betriebstemperatur liegt zwischen 5 und 35 °C bei einer
Luftfeuchtigkeit zwischen 25 und 85 %.
• Umgebung
Das STAT MATIC smart sollte in einer sauberen, staubfreien und gut belüfteten Umgebung betrieben werden.
• Arbeitsfläche
Das STAT MATIC smart sollte auf einer flachen, ebenen, wasserabweisenden und ölunempfindlichen Oberfläche aufgestellt werden. Installieren und betreiben Sie das Gerät nicht auf einer geneigten Fläche.
• Elektromagnetisches Umfeld
Das STAT MATIC smart ist auf elektromagnetische Emissionen getestet worden und erfüllt die maßgeblichen
Normen. Während das Gerät selbst keine Strahlung abgibt, kann es jedoch von anderen strahlenden Geräten beeinflusst werden. Wir empfehlen Ihnen deshalb, das Gerät von potenziellen Interferenzquellen fernzuhalten.
Elektrische Anforderungen
GEFAHR! Lebensgefahr bei unsachgemäßem Umgang mit Stromquellen! Stromschlaggefahr
•
Verwenden Sie für den Anschluss an das Stromnetz eine ordnungsgemäß abgesicherte und geerdete Steckdose mit der angegebenen Nennspannung (siehe Typenschild an der Rückseite des STAT MATIC smart).
•
Vermeiden Sie die Verwendung von Mehrfachsteckdosen. Bei der Verwendung von Stromleisten mit
Überspannungsschutz stecken Sie immer nur ein Gerät ein.
•
Der Netzeingangsstecker auf der Geräterückseite ist als Trennvorrichtung zum Versorgungsnetz zu verwenden.
Medienanschlüsse
Netzstecker in Steckdose einstecken.
Druckluftschlauch auf Kupplung aufstecken und befestigen.
Der eingestellte Druck muss zwischen 4,5 bar und 4,3 bar liegen!
Bei abweichender Anzeige Griff an der Druckregulierung herausziehen und nach rechts oder links drehen bis der
Zeiger vom Manometer auf 4,5 bar steht.
Griff wieder an Druckregulierung eindrücken und fixieren.
Um eine sichere und störungsfreie Funktion zu gewährleisten, nur Druckluft gemäß den Angaben in den
"Technischen Daten" verwenden.
Seite 39 STAT MATIC smart 1,2,3 DE
2 Inbetriebnahme
Inbetriebnahme
Netzschalter einschalten.
Mit den Pfeiltasten neben dem Display die Sprache auswählen (Seite 40).
Datum und Uhrzeit einstellen (Seite 41).
Im Display erscheint "WARNUNG Dentalöl füllen", und/oder "WARNUNG Reiniger füllen".
Obere Abdeckung mit beiden Händen ca. 15 mm nach hinten ziehen (Magnetwiderstand überwinden) und nach oben abnehmen.
Nachfüllen der Medienbehälter (siehe auch Kapitel 4 "Wartung").
Medienbehälter Farbcode beachten! blau = Dentalöl / grün = Reiniger
Nur original SciCan "Dental Oil" und SciCan "Cleaner" verwenden.
Die Verwendung von anderen Medien kann zur Gefährdung von Patienten und Anwendern, sowie zu Schäden am Gerät und den Instrumenten führen.
Zum Öffnen der Medienbehälter Metallklappen nach hinten drehen.
Bis zur Füllstandsmarkierung auffüllen. Beim Überschreiten der Minimummarke ertönt ein Piep-Ton für 1 Sek.
Wenn der Behälter ganz voll ist, ertönt 2 Mal ein Piep-Ton von 2 Sek. Der Piep-Ton kommt jeweils zeitversetzt zur optischen Anzeige am Behälter.
Medienbehälter schließen.
Obere Abdeckung aufsetzen und bis Anschlag schließen.
Gerät am Netzschalter ausschalten und wieder einschalten.
Im Display erscheint nach dem Füllen der Behälter "Dental oil" und "Cleaner" die Meldung "Mediensystem füllen".
Das Gerät startet diese Funktion automatisch und läuft dabei ca. 3 min.
Mediensystem füllen
Ist nur notwendig, sofern die Funktion bei der Inbetriebnahme nicht automatisch gestartet ist, oder wenn ein Behälter länger als 3 Tage vollständig geleert war. Die Behälter müssen für diese Funktion bereits gefüllt sein.
Gerät einschalten
Taste drücken. Es erscheint "Log-Daten exportieren" im Display.
Taste 1x drücken bis im Display "Mediensystem füllen" schwarz hinterlegt ist.
Mit Taste wird das Menü "Mediensystem füllen" ausgewählt.
Das Mediensystem wird automatisch befüllt. Während dieser Zeit läuft eine Uhr rückwärts bis auf 00:00.
Mit Taste zurück ins Hauptmenü "Reinigung + Pflege"
Sprache einstellen
Gerät einschalten
Taste drücken. Es erscheint "Log-Daten exportieren" im Display.
Taste 4x drücken bis im Display "Sprache einstellen" schwarz hinterlegt ist.
Mit Taste wird das Menü "Sprache einstellen" ausgewählt.
Mit Taste die gewünschte Sprache auswählen.
Mit Taste die Sprachauswahl sichern.
Mit Taste zurück ins Menü "Sprache einstellen".
Mit Taste zurück ins Hauptmenü "Reinigung + Pflege".
Seite 40 STAT MATIC smart 1,2,3 DE
2 Inbetriebnahme
Datum / Uhrzeit einstellen
Gerät einschalten
Taste drücken. Es erscheint "Log-Daten exportieren" im Display.
Taste 3x drücken bis im Display "Datum / Uhrzeit" schwarz hinterlegt ist.
Mit Taste wird das Menü "Datum / Uhrzeit" ausgewählt.
Mit den Tasten und die gewünschten Werte einstellen
Mit Taste springt der Cursor weiter auf die nächste Einstellmöglichkeit. Die Werte werden in folgender
Reihenfolge gesetzt: Jahr, Monat, Tag, Stunde, Minute.
Nach der letzten Bestätigung mit der Taste springt das Programm wieder zurück.
Mit Taste zurück ins Hauptmenü "Reinigung + Pflege".
Adapter montieren und überprüfen
Ggf. Adapter montieren
Den festen Sitz der MULTIflex-Aufnahme (bzw. anderer Turbinenadapter) prüfen.
Spannmutter muss fest sitzen.
Ggf. Spannmutter nachziehen.
Bitte beachten Sie, dass nur weiße STAT MATIC smart / STAT MATIC PLUS Adapter verwendet werden dürfen.
Adapter (schwarz) des STAT MATIC Pflegegerätes sind nicht geeignet.
Mit einem für die Kupplung vorgesehenen Prüfadapter die Ausbringungsmenge an allen Pflegeplätzen überprüfen.
Funktionsprüfung der Durchflussmenge mit Prüfadapter
Passenden Prüfadapter auf die entsprechenden Instrumentenkupplungen (z.B.
INTRAmatic, DIN ISO EN Kupplung) aufstecken.
Fronttüre schließen.
START Taste drücken.
Nach Ablauf des Programms Überprüfung der Füllmenge in den einzelnen
Prüfadaptern.
Flüssigkeit muss über der Markierung ("Min.") liegen!
Falls die Füllmenge unter der Markierung liegt, bitte an SciCan Kundendienst wenden.
Bitte beachten Sie, dass nur STAT MATIC smart Prüfadapter verwendet werden dürfen. Prüfadapter des STAT MATIC PLUS sind nicht geeignet.
Nach Ablauf dieses Prozesses ist das STATMATIC smart einsatzbereit.
Seite 41 STAT MATIC smart 1,2,3 DE
3 Bedienung
3 Bedienung
Manuelle Vorreinigung der Instrumente
•
Instrument überwiegend in der Kopfregion mit einer mittelharten Zahnbürste mindestens 20 Sek. unter fließendem
Wasser (kein scharfer Wasserstrahl) reinigen, bis das Instrument optisch komplett sauber ist. (Trinkwasserqualität,
Temperatur 30 °C
5 °C).
•
Instrument für 5 Minuten liegen lassen.
Programm "Reinigen + Pflege" - für Hand-/Winkelstücke und Turbinen
Netzschalter einschalten.
LED (Stand-by) leuchtet grün. Im Display erscheint "Reinigung + Pflege"
Fronttüre öffnen.
Hand-/Winkelstücke und Turbinen auf die vorgesehenen Adapter stecken
(Instrumente müssen einrasten). Im Display erscheint pro belegtem Platz um die
Stationsnummer ein Rahmen.
Fronttüre schließen.
Taste (Programmstart) auf Bedienfolientastatur betätigen.
Programm läuft automatisch ab. Während der "Reinigung + Pflege" läuft auf dem
Display eine Uhr rückwärts bis 00:00
Alle aufgesteckten Hand-/Winkelstücke und Turbinen werden nacheinander innen gereinigt und gepflegt. Nach erfolgreichem Prozess werden die Programmplätze invers dargestellt. Das Display zeigt: "Reinigung + Pflege erfolgreich".
Fronttüre öffnen.
Hand- und Winkelstücke durch Drücken des Rastknopfes (A) entnehmen.
Turbinen nach unten abziehen (Unterstützungshebel nach unten drücken).
Im Display erlischt der Rahmen um die Steckplatznummer.
•
Instrumente mit einem feuchten, mit Desinfektionsmittel getränkten Tuch abwischen. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B. SciCan OPT IM BLUE)
•
Unmittelbar nach dieser Behandlung muss das Instrument, entsprechend den Herstellerangaben zur
Wiederaufbereitung, weiterbehandelt werden (z.B. Desinfektion /Sterilisation im STAT IM ).
•
Vor der Verwendung der Instrumente am Patienten, ist immer ein Funktionstest durchzuführen. Dabei kann ggf. etwas Flüssigkeit austreten.
Programm "Spannzangenpflege"
Netzschalter einschalten.
LED (Stand-by) leuchtet grün. Im Display erscheint "Reinigung + Pflege"
Fronttüre öffnen.
Instrument mit Spannzangenöffnung an Spannzangenpflegedüse halten.
Taste (Spannzangenpflege) auf Bedienfolientastatur betätigen.
Programm läuft automatisch ab. Im Display läuft eine Uhr rückwärts bis auf 00:00
Dann wird im Display angezeigt: "Spannzangenpflege erfolgreich"
Instrument von Düse entfernen.
•
Instrumente mit einem feuchten, mit Desinfektionsmittel getränkten Tuch abwischen. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B. SciCan OPT IM BLUE)
•
Unmittelbar nach dieser Behandlung muss das Instrument, entsprechend den Herstellerangaben zur
Wiederaufbereitung, weiterbehandelt werden (z.B. Desinfektion /Sterilisation im STAT IM ).
•
Vor der Verwendung der Instrumente am Patienten, ist immer ein Funktionstest durchzuführen. Dabei kann ggf. etwas Flüssigkeit austreten.
Seite 42 STAT MATIC smart 1,2,3 DE
3 Bedienung
Die Prozesse "Nur Reinigen" und "Nur Pflege" können auch einzeln durchgeführt werden.
Programm "Nur Pflege" - für Hand-/Winkelstücke und Turbinen
Netzschalter einschalten.
LED (Stand-by) leuchtet grün. Im Display erscheint
"Reinigung + Pflege"
Fronttüre öffnen.
Hand-/Winkelstücke und Turbinen auf die vorgesehenen
Adapter stecken (Instrumente müssen einrasten). Im
Display erscheint pro belegtem Platz um die
Stationsnummer ein Rahmen.
Fronttüre schließen
Taste 1 x drücken. Es erscheint "Nur Pflege" im
Display.
Taste (Programmstart) auf Bedienfolientastatur betätigen.
Programm läuft automatisch ab. Während der Pflege läuft auf dem Display eine Uhr rückwärts bis 00:00
Alle aufgesteckten Hand-/Winkelstücke und Turbinen werden nacheinander innen gepflegt. Nach erfolgreichem
Prozess werden die Programmplätze invers dargestellt.
Das Display zeigt: "Nur Pflege erfolgreich".
Fronttüre öffnen.
Hand- und Winkelstücke durch Drücken des Rastknopfes entnehmen.
Turbinen nach unten abziehen (ggf. Unterstützungshebel drücken).
Im Display erlischt der Rahmen um die Pflegeplatznummer.
•
Instrumente mit einem feuchten, mit Desinfektionsmittel getränkten Tuch abwischen. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B. SciCan OPT IM BLUE)
•
Unmittelbar nach dieser Behandlung muss das Instrument, entsprechend den Herstellerangaben zur
Wiederaufbereitung, weiterbehandelt werden (z.B. Desinfektion /Sterilisation im STAT IM ).
•
Vor der Verwendung der Instrumente am Patienten, ist immer ein Funktionstest durchzuführen. Dabei kann ggf. etwas Flüssigkeit austreten.
Programm "Nur Reinigen"- für Hand-/Winkelstücke und Turbinen
Netzschalter einschalten.
LED (Stand-by) leuchtet grün. Im Display erscheint "Reinigung + Pflege"
Fronttüre öffnen.
Hand-/Winkelstücke und Turbinen auf die vorgesehenen Adapter stecken (Instrumente müssen einrasten).
Im Display erscheint pro belegtem Platz um die Stationsnummer ein Rahmen.
Fronttüre schließen.
Taste 2 x drücken. Es erscheint "Nur Reinigen" im Display.
Taste (Programmstart) auf Bedienfolientastatur betätigen.
Programm läuft automatisch ab. Während der Reinigung läuft auf dem Display eine Uhr rückwärts bis 00:00
Seite 43 STAT MATIC smart 1,2,3 DE
3 Bedienung
Alle aufgesteckten Hand-/Winkelstücke und Turbinen werden nacheinander innen gereinigt. Nach erfolgreichem
Prozess werden die Programmplätze invers dargestellt. Das Display zeigt: "Nur Reinigen erfolgreich".
Fronttüre öffnen.
Hand- und Winkelstücke durch Drücken des Rastknopfes entnehmen.
Turbinen nach unten abziehen (ggf. Unterstützungshebel drücken).
Im Display erlischt der Rahmen um die Pflegeplatznummer.
•
Instrumente mit einem feuchten, mit Desinfektionsmittel getränkten Tuch abwischen. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B. SciCan OPT IM BLUE)
•
Unmittelbar nach dieser Behandlung muss das Instrument, entsprechend den Herstellerangaben zur
Wiederaufbereitung, weiterbehandelt werden (z.B. Desinfektion /Sterilisation im STAT IM ).
•
Vor der Verwendung der Instrumente am Patienten, ist immer ein Funktionstest durchzuführen. Dabei kann ggf. etwas Flüssigkeit austreten.
Seite 44 STAT MATIC smart 1,2,3 DE
3 Bedienung
Programm "Sonderinstrument"
Mit diesem Programm können spezielle Instrumente gepflegt und gereinigt werden (z.B. Prophylaxe). Die Zeiten für die
Pflege und Reinigung können von einem Servicetechniker individuell eingestellt werden.
Netzschalter einschalten.
LED (Stand-by) leuchtet grün. Im Display erscheint "Reinigung + Pflege"
Fronttüre öffnen.
Instrumente auf die vorgesehenen Adapter stecken (Instrumente müssen einrasten). Im Display erscheint pro belegtem Platz um die Stationsnummer ein Rahmen.
Fronttüre schließen.
Taste 3 x drücken. Es erscheint "Sonderinstrument" im Display.
Taste (Programmstart) auf Bedienfolientastatur betätigen.
Programm läuft automatisch ab. Während der Reinigung läuft auf dem Display eine Uhr rückwärts bis 00:00
Alle aufgesteckten Instrumente werden nacheinander innen gereinigt und gepflegt. Nach erfolgreichem Prozess werden die Programmplätze invers dargestellt. Das Display zeigt: "Sonderinstrument erfolgreich".
Fronttüre öffnen.
Hand- und Winkelstücke durch Drücken des Rastknopfes entnehmen.
Turbinen nach unten abziehen (Unterstützungshebel drücken).
Im Display erlischt der Rahmen um die Steckplatznummer.
•
Instrumente mit einem feuchten, mit Desinfektionsmittel getränkten Tuch abwischen. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B. SciCan OPT IM BLUE)
•
Unmittelbar nach dieser Behandlung muss das Instrument, entsprechend den Herstellerangaben zur
Wiederaufbereitung, weiterbehandelt werden (z.B. Desinfektion /Sterilisation im STAT IM ).
•
Vor der Verwendung der Instrumente am Patienten, ist immer ein Funktionstest durchzuführen. Dabei kann ggf. etwas Flüssigkeit austreten.
Seite 45 STAT MATIC smart 1,2,3 DE
4 Wartung
4 Wartung
Adapterfixierung überprüfen
Empfohlene Durchführung:
- nach Aufstellung
- bei Betrieb: 1x wöchentlich
Den festen Sitz der MULTIflex-Aufnahme (bzw. anderer Turbinenadapter) prüfen.
Spannmutter muss fest sitzen.
Ggf. Spannmutter nachziehen.
Mit einem für die Kupplung vorgesehenen Prüfadapter die Ausbringungsmenge an allen Pflegeplätzen überprüfen.
Bitte beachten Sie, dass nur STAT MATIC smart Prüfadapter verwendet werden dürfen. Prüfadapter des STAT MATIC PLUS sind nicht geeignet
Typ: 2
O-Ringe prüfen
Empfohlene Durchführung:
- bei Betrieb: 1x wöchentlich
Alle O-Ringe an den Aufnahmeadaptern auf Beschädigung prüfen
Beschädigte O-Ringe ersetzen
Funktionsprüfung der Durchflussmenge mit Prüfadapter
Empfohlene Durchführung:
- bei erster Inbetriebnahme
- nach Wechsel einer Instrumentenkupplung
- bei Betrieb: 1x täglich vor Arbeitsbeginn
Passenden Prüfadapter auf die entsprechenden Instrumentenkupplungen
(z.B. INTRAmatic, DIN ISO EN Kupplung) aufstecken.
Fronttüre schließen.
START Taste drücken.
Nach Ablauf des Programms Überprüfung der Füllmenge in den einzelnen
Prüfadaptern.
Flüssigkeit muss über der Markierung ("Min.") liegen!
Falls Füllmenge unter der Markierung liegt, bitte an SciCan Kundendienst wenden.
Bitte beachten Sie, dass nur STAT MATIC smart Prüfadapter verwendet werden dürfen. Prüfadapter des STAT MATIC PLUS sind nicht geeignet
Druckprüfung:
Empfohlene Durchführung
- bei erster Inbetriebnahme
- bei Betrieb: 1x täglich, am besten vor Arbeitsbeginn
Am Manometer "Betriebsdruck" ablesen.
Druck muss zwischen 4,5 bar und 4,3 bar liegen!
Typ: 3
Adapter wechseln
Durch das Lösen der Spannmutter kann der wechselbare Adapter entfernt werden. Anschließend den vorgesehenen
Adapter mit der Spannmutter wieder festziehen.
Bitte beachten Sie, dass nur weiße STAT MATIC smart / STAT MATIC PLUS Adapter verwendet werden dürfen.
Adapter (schwarz) des STAT MATIC Pflegegerätes sind nicht geeignet.
Mit einem für die Kupplung vorgesehenen Prüfadapter die Ausbringungsmenge an dem Adapter überprüfen.
Bitte beachten Sie, dass nur STAT MATIC smart Prüfadapter verwendet werden dürfen. Prüfadapter des STAT MATIC
PLUS sind nicht geeignet
Seite 46 STAT MATIC smart 1,2,3 DE
4 Wartung
Nachfüllen der Medienbehälter
Täglich vor Arbeitsbeginn Füllstände der Behälter kontrollieren
Netzschalter einschalten
Im Display erscheint: "WARNUNG Dentalöl füllen" oder
"WARNUNG Reiniger füllen"
Obere Abdeckung mit beiden Händen ca. 15 mm nach hinten ziehen (Magnetwiderstand überwinden) und nach oben abnehmen.
Medienbehälter Farbcode beachten! blau = Dentalöl (SciCan "Dental oil") grün = Reiniger (SciCan "Cleaner")
Zum Öffnen der Medienbehälter Metallklappen nach hinten drehen.
Bis zur Füllstandsmarkierung auffüllen.
Beim Überschreiten der Minimum-Marke ertönt ein Piep-
Ton für 1 Sek.
Wenn der Behälter ganz voll ist, ertönt 2 Mal ein Piep-
Ton von 2 Sek.
Der Piep-Ton kommt jeweils zeitversetzt zur optischen
Anzeige am Behälter.
Obere Abdeckung aufsetzen und bis zum Anschlag schließen.
Nur Original SciCan Dentalöl und SciCan Reiniger verwenden!
Die Verwendung von anderen Medien kann zur Gefährdung von Patienten und Anwendern sowie zu Schäden am Gerät und den Instrumenten führen.
Seite 47 STAT MATIC smart 1,2,3 DE
4 Wartung
Absorptions-Vliese wechseln
Beide Vliese nach 35 Instrumentenreinigungen bzw. mindestens 1 x wöchentlich wechseln (bei Bedarf öfters).
Fronttüre öffnen.
Vliese von den Pins abziehen.
Vliese nach oben aus Befestigungslaschen entnehmen.
Vliese im Hausmüll entsorgen.
Neue Vliese in Befestigungslaschen einhängen.
Vliese auf den Pins fixieren.
Fronttüre schließen.
Nur Original SciCan Absorptions-Vlies verwenden.
Auffangschale entleeren
Auffangschale täglich entleeren und reinigen/desinfizieren
Fronttüre öffnen.
Schale nach vorne (horizontal) herausnehmen (leichte Rastung überwinden).
Schale entleeren und mit einem Desinfektionstuch reinigen/desinfizieren.
Für Deutschland empfehlen wir die Verwendung von VAH gelistetem
Desinfektionsmittel (z.B. SciCan OPT IM BLUE)
Schale in die seitlichen Führungen bis zur Rastung einschieben.
Rückseitiges Vlies darf nicht eingeklemmt werden, sondern muss lose in die Schale hängen.
Fronttüre schließen.
Innenreinigung und Desinfektion des Pflegeraums
Pflegeraum und Türe täglich mit Wischdesinfektionstüchern säubern. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B. SciCan OPT IM BLUE).
Außenreinigung und Außendesinfektion des Gerätes
Die Oberfläche und die Bedienfolientastatur des STAT MATIC smart können mittels Desinfektionstüchern gereinigt und desinfiziert werden. Für Deutschland empfehlen wir die Verwendung von VAH gelistetem Desinfektionsmittel (z.B.
SciCan OPT IM BLUE).
Empfehlung
SciCan empfiehlt eine jährliche Wartung durch einen autorisierten Fachbetrieb oder den Hersteller durchführen zu lassen.
Hinweis für Benutzer in Ländern mit gesetzlicher Validierungspflicht
•
Eine jährliche Wartung durch den Hersteller oder eine autorisierte Fachfirma wird empfohlen.
•
Beachten Sie die nationalen Vorgaben hinsichtlich der Validierung. Bei Validierungspflicht ist, durch eine autorisierte
Fachfirma, eine Erstvalidierung mit Leistungsqualifikation (PQ) gemäß den gesetzlichen Bestimmungen durchzuführen.
Dabei ist die Validierungsanweisung des Gerätes zu beachten.
•
SciCan empfiehlt die Durchführung einer Revalidierung in Zeitabständen von 1 Jahr (gem. EN ISO 15883-1), falls von den örtlichen Behörden, oder dem Validierer keine kürzeren Zyklen gefordert werden. Diese sollten im Rahmen einer
Wartung oder im Anschluss an eine Wartung erfolgen.
Seite 48 STAT MATIC smart 1,2,3 DE
4 Wartung
Datensicherung der Log-Datei
Gerät ausschalten
USB
–Stick an Geräterückseite einstecken
Gerät einschalten
Taste drücken. Es erscheint "Log-Daten exportieren" im Display.
Mit Taste Export der Log-Daten auf den USB Stick starten.
Wenn der Vorgang abgeschlossen ist, erscheint im Display: "Log-Daten exportieren erfolgreich".
USB Anschluss (Geräterückseite)
Mit Taste wieder zurück ins Menü.
Gerät ausschalten
USB-Stick abziehen.
•
Wenn der USB-Port mit einem USB Kabel verlängert wird, so muss dieses mit einem Ferritkern ausgestattet sein. Das
Kabel muss mir dem Ferritkern zur Geräteseite hin eingesteckt werden.
•
Bitte beachten Sie, dass Sie die Daten vom USB-Stick auf ein geeignetes dauerhaftes Speichermedium kopieren.
Überprüfen sie, dass die Daten auf dem Speichermedium auch gelesen werden können.
Löschen der Log-Datei auf dem STATMATIC smart
Die Daten gehen beim Löschen der Log-Datei unwiderruflich verloren. Machen Sie zuerst eine Datensicherung der Log-
Datei und überprüfen Sie, ob die Daten auf dem USB-Stick auch gelesen werden können.
Gerät einschalten
Taste drücken. Es erscheint "Log-Daten exportieren" im Display.
Taste 6x drücken bis im Display "Log-Daten löschen" schwarz hinterlegt ist.
Mit Taste wird das Löschen ausgewählt.
Taste nochmals drücken. Das Löschen der Log Daten wird ausgeführt.
Mit Taste zurück ins Menü.
Mit Taste wieder zurück ins Hauptmenü "Reinigung + Pflege".
Mediensystem leeren
Ist nur notwendig, wenn das System - z.B. zum Transport oder für längere Stillstandszeiten (>30 Tage) - vollständig geleert werden soll.
Zuerst die Behälter mittels den Ablassschläuchen komplett leeren. Danach das Gerät um ca. 45 Grad nach hinten kippen um auch die Restflüssigkeit am Boden der Behälter mit den Ablassschläuchen zu entleeren. Medien können für den erneuten Gebrauch nur wieder verwendet werden, wenn sie in Originalflaschen aufbewahrt wurden (Haltbarkeitsdaten unten oder seitlich an den Originalflaschen beachten).
Gerät einschalten
Taste drücken. Es erscheint "Log-Daten exportieren" im Display.
Taste 2x drücken bis im Display "Mediensystem leeren" schwarz hinterlegt ist.
Mit Taste wird das Menü "Mediensystem leeren" ausgewählt.
Das Mediensystem wird automatisch geleert. Während dieser Zeit läuft eine Uhr rückwärts bis auf 00:00.
Mit Taste zurück ins Hauptmenü "Reinigung + Pflege".
Seite 49 STAT MATIC smart 1,2,3 DE
4 Wartung
Warnmeldung: "Service wird bald benötigt"
Diese Warnmeldung erscheint, wenn mehr als 25.000 Hand- und Winkelstücke, sowie Turbinen mit dem STAT MATIC smart gereinigt und gepflegt wurden. Bitte verständigen Sie den Service. Sie können noch 3.300 Hand- und
Winkelstücke, sowie Turbinen reinigen und pflegen.
Fehlermeldung: "Service wird benötigt"
Diese Fehlermeldung erscheint, wenn mehr als 28.300 Hand- und Winkelstücke, sowie Turbinen mit dem STAT MATIC smart gereinigt und gepflegt wurden. Bitte verständigen Sie den Service.
Transport
•
Sofern das Gerät bereits mit Dentalöl und Reiniger befüllt war, die Medien über die Ablassschläuche komplett entleeren.
•
Mediensystem leeren (siehe S. 20)
•
Die Sicherungsschraube an der Medienbelüftung herausdrehen und mit Druckluft mehrfach in das Ventil blasen.
Schraube wieder einsetzen
•
Mitgelieferte Verschlussstopfen in die Befüllungstrichter der Medienbehälter einsetzen (siehe Abb.).
•
Klappen wieder schließen.
•
Nach Möglichkeit im Originalkarton verschicken.
Seite 50 STAT MATIC smart 1,2,3 DE
5 Fehlerbehebung
5 Fehlerbehebung
Fehler
Keine LED leuchtet
Druckluft
Obere Abdeckung offen
Programm startet nicht
Prozess unterbrochen
Undefinierter Fehler
Speicher voll
Anzeige Display
Keine Anzeige
ACHTUNG
Druckluftfehler
Fehler LED (L2) blinkt
WARNUNG
Obere Abdeckung schließen
Fehler LED (L2) blinkt
ACHTUNG
Türe schließen
ACHTUNG
Keine Plätze belegt
ACHTUNG
Prozess unterbrochen
ACHTUNG
Bitte an Kundendienst wenden
ACHTUNG
Nicht genügend
Speicher vorhanden
Akustisches Signal Ursache
–
Intervall
– Piepton
0,5 Sek. an
8 Sek. aus;
10 mal
–
–
–
–
Intervall – Piepton
2 Sek. an
4 Sek. aus
–
Keine Stromversorgung
Sicherung defekt
Druckluftschlauch nicht angeschlossen
Netzschalter betätigen
Netzstecker einstecken
Sicherungswechsel
(siehe Kapitel 6
Fehlerbehebung)
Druckluftschlauch anschließen
Druckluftschlauch abgeknickt
Netzseitige
Versorgung fehlerhaft
Durchflussstörung entfernen
Versorgung herstellen
Druckminderer defekt Bitte an SciCan
Kundenservice wenden.
Druckschalter defekt Bitte an SciCan
Kundenservice wenden.
Sofern vorhanden;
Filter in Zuleitung voll
Filter wechseln
Obere Abdeckung nicht richtig geschlossen
Magnet an oberer
Abdeckung fehlt
Fronttüre nicht richtig geschlossen
Obere Abdeckung fest schließen
Bitte an SciCan
Kundenservice wenden.
Fronttüre fest schließen
Keine Instrumente im
Gerät
Schaltklappe erkennt
Instrument nicht
Instrumente auf Kupplungen stecken
Instrument drehen, bis ein
Rahmen um die
Stationsnummer erscheint
Fronttüre fest schließen. Fronttüre nicht richtig geschlossen
Instrument nicht richtig auf der Kupplung
Instrument fest auf die
Kupplung schieben
Interner Gerätefehler Bitte an SciCan
Kundenservice wenden.
Interner Speicher voll
Behebung
Log-Daten auf USB-Stick exportieren
Seite 51 STAT MATIC smart 1,2,3 DE
5 Fehlerbehebung
Fehler Anzeige Display
Reiniger leer
WARNUNG
Reiniger füllen
Fehler LED (L2) blinkt
WARNUNG
Dentalöl leer Dentalöl füllen
Fehler LED (L2) blinkt
WARNUNG WARNUNG
Dentalöl füllen oder
Reiniger füllen,
Dentalöl füllen oder
Reiniger füllen obwohl der entsprechende Behälter
Fehler LED (L2) blinkt voll ist.
Daten exportieren
Flüssigkeiten (Dentalöl und Reiniger) wurden verwechselt
ACHTUNG
Speicher medium nicht lesbar
Fehler LED (L2) blinkt
–
Akustisches
Signal
Piepton
2 Sek.
Piepton
2 Sek
Piepton
2 Sek
–
–
Ursache
Reiniger-Behälter leer
Dentalöl-Behälter leer
Behebung
Reiniger-Behälter mit SciCan
"Cleaner" auffüllen
Dentalöl-Behälter mit SciCan
"Dental oil" auffüllen
Luft im Füllstands-
Messsystem
USB-Stick nicht erkannt
USB-Stick nicht richtig formatiert
Den betroffenen Behälter am
Ablassschlauch komplett entleeren.
Das Gerät dabei 45° nach hinten kippen. Die Sicherungsschraube an der Medienbelüftung herausdrehen und mit Druckluft mehrfach in das Ventil blasen.
Schraube wieder einsetzen und
Behälter füllen.
Gerät ausschalten, USB-Stick einstecken, Gerät einschalten
- NTFS Formatierung durchführen.
- Anderen USB-Stick verwenden
Medienbehälter entleeren:
Flüssigkeiten mittels
Ablassschlauch aus dem
Behälter ablassen. Evtl.
Restflüssigkeit mit 50 ml-
Spritze aus dem entsprechenden
Medienbehälter herausziehen.
Mit fusselfreiem, gut saugendem Textiltuch Reste des "falschen" Mediums vollständig aus dem
Medienbehälter entfernen.
Gerät einschalten
Am Display Mediensystem leeren wählen
(s. Kapitel 4 "Wartung")
Dentalöl u. Reiniger nachfüllen
(s. Kapitel 4 "Wartung")
- Fronttüre öffnen
- Im Display Menüpunkt
Mediensystem füllen wählen
- Fronttüre schließen
Nach Ablauf dieses Prozesses ist das Gerät betriebsbereit.
Sicherungswechsel
Gerätestecker vom STAT MATIC smart ausstecken.
Sicherungsschublade aus Netzeingangsmodul herausnehmen (beide Laschen der
Sicherungsschublade nach innen drücken).
Sicherungen ersetzen.
Sicherungsschublade wieder einsetzen.
Nur den in den technischen Daten angegebenen Sicherungstyp verwenden.
Seite 52 STAT MATIC smart 1,2,3 DE
6 Informationen zur elektromagnetischen Verträglichkeit (EMV)
6 Informationen zur elektromagnetischen Verträglichkeit (EMV)
Das STAT MATIC smart ist für den Betrieb in einer wie unten angegebenen Umgebung bestimmt. Der Kunde oder der Anwender des
STAT MATIC smart sollte sicherstellen, dass es in einer derartigen Umgebung betrieben wird.
Störaussendungsmessungen
HF-Aussendungen nach CISPR 11
Übereinstimmung
Gruppe 1
Elektromagnetische Umgebung - Leitfaden
Das STAT MATIC smart verwendet HF-Energie ausschließlich zu seiner internen Funktion. Daher ist seine HF-Aussendung sehr gering, und es ist unwahrscheinlich, dass benachbarte elektronische Geräte gestört werden.
HF-Aussendungen nach CISPR 11
Aussendungen von Oberschwingungen nach
IEC 61000-3-2
Aussendungen von Spannungsschwankungen / Flicker nach IEC 61000-3-3
Klasse B
Klasse A
Stimmt überein
Das STAT MATIC smart ist für den Gebrauch in allen
Einrichtungen einschließlich denen im Wohnbereich und solchen geeignet, die unmittelbar an ein öffentliches Versorgungsnetz angeschlossen sind, das auch Gebäude versorgt, die zu
Wohnzwecken benutzt werden.
Störfestigkeitsprüfungen
Entladung statischer
Elektrizität (ESD) nach IEC 61000-4-2 schnelle transiente elektrische Störgrößen/
Bursts nach IEC 61000-4-4
Stoßspannungen (Surges) nach IEC 61000-4-5
Spannungseinbrüche,
Kurzzeitunterbrechungen und Schwankungen der
Versorgungsspannung nach IEC 61000-4-11
IEC 60601-Prüfpegel
± 6 kV Kontaktentladung
± 8 kV Luftentladung
± 2 kV für Netzleitungen
± 2 kV für Eingangs- und
Ausgangsleitungen
± 1 kV Gegentaktspannung
± 2 kV Gleichtaktspannung
0 % UT für 0,5 Perioden
40 % UT für 5 Perioden
70 % UT für 25 Perioden
0 % UT für 250 Perioden
3 A/m
Übereinstimmungspegel Elektromagnetische Umgebung -
Leitlinien
± 6 kV Kontaktentladung
± 8 kV Luftentladung
Fußböden sollten aus Holz oder Beton bestehen oder mit Keramikfliesen versehen sein. Wenn der Fußboden mit synthetischem Material versehen ist, muss die relative Luftfeuchte mindestens
30 % betragen.
± 2 kV, für Netzleitungen nicht anwendbar
± 1 kV Gegentaktspannung
± 2 kV Gleichtaktspannung
Die Qualität der Versorgungsspannung sollte der einer typischen Geschäfts- oder
Krankenhausumgebung entsprechen.
Die Qualität der Versorgungsspannung sollte der einer typischen Geschäfts- oder
Krankenhausumgebung entsprechen.
0 % UT für 0,5 Perioden
40 % UT für 5 Perioden
70 % UT für 25 Perioden
0 % UT für 250 Perioden nicht anwendbar
Die Qualität der Versorgungsspannung sollte der einer typischen Geschäfts- oder
Krankenhausumgebung entsprechen.
Wenn der Anwender des STAT MATIC smart fortgesetzte Funktion auch beim
Auftreten von Unterbrechungen der
Energieversorgung fordert, wird empfohlen, das STAT MATIC smart aus einer unterbrechungs-freien
Stromversorgung zu speisen. nicht anwendbar Magnetfeld bei der
Versorgungsfrequenz
(50/60 Hz) nach IEC 61000-4-8
Geleitete HF-Störgröße nach IEC 61000-4-6
Gestrahlte
HF-Störgrößen nach IEC 61000-4-3
3 Veff
150 kHz bis 80 MHz außerhalb der ISM-Bänder a
10 V/m
80 MHz bis 2,5 GHz
10 Veff
10 V/m
Tragbare und mobile Funkgeräte sollten in keinem geringeren Abstand zum STAT MATIC smart einschließlich der Leitungen verwendet werden als dem empfohlenen Schutzabstand, der nach der für die Sendefrequenz zutreffenden Gleichung berechnet wird.
Empfohlener Schutzabstand: d = 1,17
P d = 0,35
P für 80 MHz bis 800 MHz d = 0,70
P für 800 MHz bis 2,5 GHz mit P als der maximalen Nennleistung des Senders in Watt (W) gemäß Angaben des Senderherstellers und d als empfohlenem Schutzabstand in Metern
(m). b
Die Feldstärke stationärer Funksender sollte bei allen Frequenzen gemäß einer Untersuchung vor
Ort geringer als der Übereinstimmungspegel sein. d In der Umgebung von Geräten, die das unten stehende Bildzeichen tragen, sind Störungen möglich.
Seite 53 STAT MATIC smart 1,2,3 DE
6 Informationen zur elektromagnetischen Verträglichkeit (EMV)
ANMERKUNG U
T
: ist die Netzwechselspannung vor der Anwendung der Prüfpegel.
ANMERKUNG 1: Bei 80 MHz und 800 MHz gilt der höhere Frequenzbereich.
ANMERKUNG 2: Diese Leitlinien mögen nicht in allen Fällen anwendbar sein. Die Ausbreitung elektromagnetischer Größen wird durch
Absorption und Reflexionen der Gebäude, Gegenstände und Menschen beeinflusst. a
Die ISM-Frequenzbänder (für industrielle, wissenschaftliche und medizinische Anwendungen) zwischen 150 kHz und 80 MHz sind
6,765 MHz bis 6,795 MHz; 13,553 MHz bis 13,567 MHz; 26,957 MHz bis 27,283 MHz und 40,66 MHz bis 40,70 MHz. b Die Übereinstimmungspegel in den ISM-Frequenzbändern zwischen 150 kHz und 80 MHz und im Frequenzbereich von 80 MHz und
2,5 GHz sind dazu bestimmt die Wahrscheinlichkeit zu verringern, dass mobile/tragbare Kommunikationseinrichtungen Störungen hervorrufen können, wenn sie unbeabsichtigt in den Patientenbereich gebracht werden. Aus diesem Grunde wird der zusätzliche Faktor von 10/3 bei der Berechnung der empfohlenen Schutzabstände in diesen Frequenzbereichen angewandt. c Die Feldstärke stationärer Sender, wie z. B. Basisstationen von Funktelefonen und mobilen Landfunkgeräten, Amateurfunkstationen,
AM- und FM-Rundfunk- und Fernsehsendern können theoretisch nicht genau vorherbestimmt werden. Um die elektromagnetische
Umgebung hinsichtlich der stationären Sender zu ermitteln, sollte eine Studie des Standorts erwogen werden. Wenn die gemessene
Feldstärke an dem Standort, an dem das STAT MATIC smart benutzt wird, die obigen Übereinstimmungspegel überschreitet, sollte das
STAT MATIC smart beobachtet werden, um die bestimmungsgemäße Funktion nachzuweisen. Wenn ungewöhnliche
Leistungsmerkmale beobachtet werden, können zusätzliche Maßnahmen erforderlich sein, wie z. B. eine veränderte Ausrichtung oder ein anderer Standort des STAT MATIC smart. d Über den Frequenzbereich von 150 kHz bis 80 MHz sollte die Feldstärke geringer als 3 Veff V/m sein.
Empfohlene Schutzabstände zwischen tragbaren und mobilen HF-Telekommunikationsgeräten und dem STAT MATIC smart
Das STAT MATIC smart ist für den Betrieb in einer elektromagnetischen Umgebung bestimmt, in der die HF-Störgrößen kontrolliert sind.
Der Kunde oder der Anwender des STAT MATIC smart kann dadurch helfen elektromagnetische Störungen zu vermeiden, indem er den
Mindestabstand zwischen tragbaren und mobilen HF-Telekommunikationsgeräten (Sendern) und dem STAT MATIC smart abhängig von der Ausgangsleistung des Kommunikationsgerätes, wie unten angegeben,- einhält.
Schutzabstand abhängig von der Sendefrequenz in m
Nennleistung des Senders / W
150 kHz bis 80 MHz d=0,35
P 80 MHz bis 800 MHz d=0,35
P 800 MHz bis 2,5 GHz d=0,70
P
0,01
0,1
1
10
0,04
0,11
0,35
1,11
0,04
0,11
0,35
1,11
0,07
0,22
0,70
2,21
100 3,50 3,50 7,00
Für Sender, deren maximale Nennleistung in obiger Tabelle nicht angegeben ist, kann der empfohlene Schutzabstand d in Metern (m) unter Verwendung der Gleichung ermittelt werden, die zur jeweiligen Spalte gehört, wobei P die maximale Nennleistung des Senders in
Watt (W) gemäß Angabe des Senderherstellers ist.
ANMERKUNG 1: Bei 80 MHz und 800 MHz gilt der höhere Frequenzbereich.
ANMERKUNG 2: Diese Leitlinien mögen nicht in allen Fällen anwendbar sein. Die Ausbreitung elektromagnetischer Größen wird durch
Absorptionen und Reflexionen der Gebäude, Gegenstände und Menschen beeinflusst.
Seite 54 STAT MATIC smart 1,2,3 DE
7 Zubehör
7 Zubehör
Zubehör bitte nur bei ihrem Fachhandel bestellen.
Material Nr. Beschreibung
S32102
SciCan "Cleaner"
(6x 500 ml)
S32101
SciCan "Dental Oil"
(2x 500 ml)
S32100
Absorptions-Vlies Set
(10 Stück)
S32114
Druckluftschlauch
(Leitung)
S30116 O-Ring-Set MULTIflex*
S32118 O-Ring-Set INTRAmatic
S32101-CX
Pflegekupplung
MULTIflex* / CONNEX
S32101-ST1T
Pflegekupplung
Sirona T1 Turbine
S32101-NSK
Pflegekupplung
NSK Turbine
S32101-W&H
Pflegekupplung
W&H Turbine
Pflegekupplung
BienAir Turbine
S32101-BA
S33126 Verschlussstopfen
*MULTIflex ist ein eingetragenes
Warenzeichen von
Kaltenbach & Voigt
Material Nr. Beschreibung
S30146 Spannzangenspitze
S32117 Sicherung T 2,0 A (H)
S30108 Gummifuß
S30122
Dichtung
Pflegekupplung
S33120 Fronttürdichtung
S32201-E
Prüfadapter INTRAmatic
DIN EN ISO 3964-
Kupplung
S32201-CX
Prüfadapter
MULTIflex*
S32201-ST1T
Prüfadapter
Sirona T1 Turbine
S32201-NSK
Prüfadapter
NSK Turbine
S32201-W&H
Prüfadapter
W&H Turbine
S32201-BA
Prüfadapter
BienAir Turbine
S32202
Ersatzprüfröhrchen
STAT MATIC smart
Seite 55 STAT MATIC smart 1,2,3 DE
8 Technische Daten
8 Technische Daten
Gerät:
Geräteabmessungen:
Gewicht:
Elektrische Nennwerte:
Schutzklasse:
Schutz:
Installationskategorie:
Verschmutzungsgrad:
Umgebungstemperatur:
Schallpegel:
Relative Luftfeuchtigkeit:
Max. Betriebshöhe über NN:
Betriebsstandort:
Druckluft:
Luftverbrauch:
Druckluftqualität:
Breite:
Tiefe:
Höhe:
Nennspannung:
Nennfrequenz:
Nennleistung:
Restölgehalt:
Reststaub:
Restwasser:
300 mm (11,8 inches)
280 mm (11,0 inches)
345 mm (13,6 inches)
7,3 kg (15,8 lbs)
100
– 230 V
10%
50 / 60 Hz max. 45 VA
I abgedeckt
Cat. II
P 2
5 – 35 °C (41 – 95 °F)
60 (65) dB(A) max. 85 %
2000 m nur Innenraum
4,5 – 6 bar (65 – 87 psi)
40 Nl/min max. 0,1 mg/m³ max. 1 µm; max. 1 mg/m³ max. 0,1 g/m³ bei -40 °C (-40 °F)
Netzsicherungen:
Anzahl integrierter Kupplungen ISO 3964:
Nennstrom:
Nennspannung:
Auslösecharakteristik:
Schaltvermögen:
Typ 1
Typ 2
Typ 3
2 A
250 V
Large (T)
H
1
2
3
Gerät:
Umgebungstemperatur:
Relative Luftfeuchtigkeit:
Luftdruck:
Lagerbedingungen
5
– 35 °C C (41 – 95 °F)
5
– 95 %
700 – 1060 hPa
Transportbedingungen
-30
– 50 °C (-22 – 122 °F)
5
– 95 %
700 – 1060 hPa
Vor dem Transport sind die Medienbehälter zu entleeren! (siehe Kapitel 4 "Wartung")
Medien:
Umgebungstemperatur - Dentalöl:
Lagerbedingungen
5
– 50 °C (41 – 122 °F)
Transportbedingungen
-30
– 50 °C (-22 – 122 °F)
Umgebungstemperatur - Reiniger:
Relative Luftfeuchtigkeit:
Luftdruck:
5 – 35 °C (41 – 95 °F)
5 – 95 %
800
– 2500 hPa
-30
5
800
– 50 °C (-22 – 122 °F)
– 95 %
– 2500 hPa
Seite 56 STAT MATIC smart 1,2,3 DE
8 Technische Daten
Typenschild
STATMATIC smart 1
STATMATIC smart 2
STATMATIC smart 3
Die Erklärungen der Symbole sind beschrieben in Kapitel 1 "Wichtige Hinweise"
Seite 57 STAT MATIC smart 1,2,3 DE
9 Dokumentation Log-Datei
9 Dokumentation Log-Datei
Seite 58 STAT MATIC smart 1,2,3 DE
10 Garantie
10 Garantie
Eingeschränkte Garantie
(DE = Original)
SciCan garantiert für einen Zeitraum von zwei Jahren (ab Kaufdatum), dass das STATMATIC smart im Neu- und nicht gebrauchten Zustand bei normalem Gebrauch nicht auf Grund von Material- oder Verarbeitungsfehlern, die nicht auf offensichtlichen Missbrauch, falsche Anwendung oder Unglücksfälle zurückzuführen sind, ausfallen wird.
Die zweijährige Garantie gilt für die Funktionsfähigkeit aller Teile des Geräts, vorausgesetzt, das Gerät wird gemäß der im Benutzerhandbuch enthaltenen Beschreibung genutzt und gewartet.
Im Falle eines Ausfalles, der während der Garantiezeit auf solche Defekte zurückzuführen ist, besteht die einzige
Verpflichtung von SciCan nach eigenem Ermessen in der kostenlosen Reparatur oder im unentgeltlichen Ersatz jeglicher defekter Teile (mit Ausnahme von O-Ringen und Dichtungen), vorausgesetzt, dass SciCan innerhalb von 30 Tagen nach Auftreten einer solchen Störung schriftlich benachrichtigt wird und dass die defekten Teile außerdem frachtfrei an SciCan zurückgesendet werden.
Diese Garantie gilt als vereinbart, wenn dem Produkt die Originalrechnung des autorisierten SciCan -Händlers beiliegt und diese Rechnung den Artikel durch seine Artikelnummer kennzeichnet und sowie einen eindeutigen
Beleg des Kaufdatums aufweist. Keine andere Art von Beleg ist akzeptabel. Nach Ablauf von zwei Jahren gelten alle von SciCan gegebenen Garantien und anderen Verpflichtungen bezüglich der Qualität des Produktes als endgültig erfüllt. Somit endet jegliche Haftung, und gerichtliche Schritte wegen Verstoßes gegen eine solche
Garantie oder Verpflichtung dürfen hiernach nicht mehr gegen SciCan eingeleitet werden.
Jegliche ausdrückliche Garantie, die hier nicht gegeben wird, und jegliche implizierte Garantie oder Darstellung hinsichtlich der Leistungsfähigkeit sowie jegliche Rechtsmittel bei Vertragsbruch, die abgesehen von dieser
Bestimmung durch Andeutung, Gesetzesauslegung, Handelsbrauch oder -sitte, einschließlich jeglicher implizierter Garantie hinsichtlich Verkaufsfähigkeit oder Eignung für einen bestimmten Zweck in Bezug auf ein oder alle von SciCan hergestellten Produkte entstehen könnten, werden von SciCan ausgeschlossen und abgelehnt. Falls Sie mehr über unsere Produkte und Fähigkeiten erfahren wollen, besuchen Sie die SciCan -
Webseite unter www.scican.com.
Seite 59 STAT MATIC smart 1,2,3 DE
STAT MATIC smart 1,2,3 DE

Public link updated
The public link to your chat has been updated.
Advertisement