ELCART 9/9420 User manual
ELCART 9/9420
ELCART 9/9420 is a 3 ½ digit LCD display autorange digital multimeter. It complies with the IEC 61010-1, CAT. III 1000V and CAT.IV 600V standards. It has a variety of features and possible use cases, including:
- Voltage measurement: AC and DC voltage measurements up to 1000V.
- Current measurement: AC and DC current measurements up to 10A.
- Resistance measurement: Resistance measurements up to 40MΩ.
- Capacitance measurement: Capacitance measurements up to 40mF.
- Frequency measurement: Frequency measurements up to 10MHz.
- Temperature measurement: Temperature measurements from -40°C to 1000°C.
Advertisement
Advertisement
ART. 9/9450
ELCART
Manuale di istruzioni/Scheda tecnica
MULTIMETRO DIGITALE AUTORANGE CON DISPLAY 3 ½ DIGIT LCD
ART. 9/9450 NI9450
MANUALE DI ISTRUZIONI
PAGINA 1 DI 9
1. ISTRUZIONI GENERALI
Questo strumento rispetta le normative IEC 61010-1, CAT. III 1000V e CAT.IV 600V.
Per utilizzarlo al meglio, seguite le indicazioni presenti in questo manuale e rispettate le precauzioni di sicurezza in esso contenute.
I simboli internazionali usati per questo strumento sono spiegate al capitolo 1.1.3
1.1 Precauzioni di sicurezza
1.1.1 Preliminari
* Gli strumenti di categoria IV sono progettati per proteggere l’utilizzatore dai rischi derivanti dal primo livello di alimentazione (protezione ambientale e di terra).
* La categoria di misurazione III è relativa alle misurazioni effettuate nell’ambito domestico.
* La categoria di misurazione II è relativa alle misurazioni effettuate su circuiti collegati direttamente a una bassa tensione.
* La categoria di misurazione I è relativa alle misurazioni effettuate su circuiti non alimentati.
* Quando utilizzate questo multimetro dovete osservare tutte le normali regole di sicurezza:
- protezione contro il pericolo di scariche elettriche.
- protezione contro usi impropri del multimetro.
* per sicurezza utilizzate i puntali forniti nella confezione.
Prima di utilizzare i puntali, controllate che siano in buone condizioni.
1. ISTRUZIONI GENERALI
1.1 Precauzioni di sicurezza
1.1.1 Istruzioni preliminari
1.1.2 Durante l’uso
1.1.3 Simboli
1.1.4 Istruzioni
1.2 Meccanismo di protezione
2. DESCRIZIONE
2.1 Familiarizzare con lo strumento
2.2 Display LCD
2.3 Pulsanti
2.4 Terminali
2.5 Accessori
3. DESCRIZIONE DELLE FUNZIONI
3.1 Funzioni generali
3.1.1 Modalità DATA HOLD
3.1.2 Risparmio batteria
3.2 Funzioni di misura
3.2.1 Misurazioni Tensioni CA e CC
3.2.2 Misurazione Resistenza
3.2.3 Test Diodo
3.2.4 Test Continuità
3.2.5 Misurazione Capacità
3.2.6 Misurazione Transistor
3.2.7 Misurazione Frequenza
3.2.8 Misurazione Temperatura
3.2.9 Misurazione Corrente
4. SPECIFICHE TECNICHE
4.1 Specifiche generali
4.2 Specifiche di misurazione
4.2.1 Tensione CC
4.2.2 Tensione CA
4.2.3 Frequenza
4.2.4 Resistenza
4.2.5 Test Diodi
4.2.6 Test Continuità
4.2.7 Transistor
4.2.8 Temperatura
4.2.9 Capacità
4.2.10 Corrente CC
4.2.11 Corrente CA
5. MANUTENZIONE
5.1 Manutenzione Generale
5.2 Sostituzione Batteria
1.1.2 Durante l’uso
* Se utilizzate il multimetro vicino a apparecchi che generano interferenze, controllate che il display non diventi instabile o che indichi errori.
* Non utilizzate il multimetro o i puntali se sembrano danneggiati.
* Utilizzate il multimetro solo per le funzioni specificate in questo manuale
* Prestate attenzione nel caso in cui lavoriate vicino a conduttori non isolati
* Non utilizzate il multimetro vicino a gas esplosivi o polvere
* Verificate periodicamente il funzionamento del multimetro con segnali conosciuti. Non utilizzatelo se da risultati anormali.
In caso di malfunzionamento contattate un tecnico per la riparazione.
* Nel caso in cui non sappiate quale gamma di misurazione utilizzare, scegliete la più alta possibile, oppure utilizzate la funzione autorange.
* Per evitare danneggiamenti, non superate il limite massimo di ingresso, specificato in questo manuale.
* Quando il multimetro è collegato al circuito da misurare, non toccate i puntali.
* Prestate attenzione quando lavorate con tensioni superiori a 60VCC o
30VCA (Rms).
* Quando procedete alla misurazione collegate prima il puntale nero
(COM) e poi quello rosso (+).
Quando finite la misurazione scollegate prima il puntale rosso e poi quello nero.
* Prima di cambiare funzione, scollegate i puntali dal circuito.
* Se state effettuando una misurazione CC, prima di passare a una misurazione in CA, assicuratevi di aver tolto qualsiasi alimentazione per evitare scosse o guasti allo strumento.
Per tornare a una misurazione CC, selezionate una gamma superiore all’ultimo valore di CA misurato.
* Togliete l’alimentazione al circuito e scaricate tutti i condensatori prima di misurare resistenza, continuità, diode e condensatori.
* Non effettuate mai misurazioni di resistenza e continuità su circuiti alimentati.
* Prima di misurare la corrente, controllate il fusibile del multimetro e togliete l’alimentazione prima della misurazione.
* Utilizzate una batteria 9V per alimentare lo strumento.
* Sostituite la batteria quando apparirà il simbolo ( ).
* Non misurate tensioni superiori a 600V in Categoria IV, o 1000V in
Categoria III.
* Non utilizzate il multimetro senza la custodia
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
ELCART
Manuale di istruzioni/Scheda tecnica
1.1.3 Simboli:
Attenzione: fate riferimento a questo manuale.
Un uso non corretto potrebbe provocare guasti allo strumento o ai circuiti da misurare.
SIMBOLO
~
CA (corrente alternata) CC (corrente continua)
-
PAGINA 2 DI 9
SIGNIFICATO
La batteria è scarica
Per evitari guasti e malfunzionamenti, sostituite la batteria
Indicatore letture negative
Indicatore per tensione e corrente alternata.
Viene visualizzata la media del valore assoluto di ingresso
(valore efficace dell’onda sinusoidale)
DC
Terra Doppio isolamento
Indicatore per tensione e corrente continua
Fusibile Conforme alle direttive Europee
1.1.4 Istruzioni
* Togliete i puntali prima di aprire il coperchio della batteria.
* In caso di guasti, utilizzate parti di ricambio di stesso livello di qualità.
* Prima di utilizzare lo strumento, scollegate ogni alimentazione dal circuito da misurare.
* Per qualsiasi manutenzione o riparazione contattate personale qualificato
* Con personale qualificato si intende qualsiasi persona che abbia familiarità con istallazione e costruzione di questo tipo di apparecchiatura.
E’ l’unico personale autorizzato a compiere riparazioni su questi articoli.
* Quando aprite lo strumento ricordate che ci potrebbero essere dei condensatori carichi che potrebbero danneggiare il circuito interno.
* Se non utilizzate lo strumento a lungo, togliete la batteria e non conservatelo in luoghi con alta temperatura o umidità.
1.2 Meccanismi di protezione
* Utilizza fusibile (FF400mA/1000V) durante la misurazione di capacità,
temperatura, mA e hFE.
* Protezione a PTC permanente fino a 250V durante la misurazione di
resistenza, frequenza, continuità e test di diodi.
2. DESCRIZIONE
2.1 Familiarizzare con lo strumento
2.2 Display LCD
1. Area allerta tensione
2. Luce allerta tensione
3. LCD display
4. Pulsanti
5. Manopola
6. Terminali
H
°C, °F
V,mV
Modalità Test diodi
Modalità Test continuità
Modalità Data Hold
°C: scala Celsius / °F: scala Fahrenheit
V: Volts, unità di misura della tensione mV: Millivolts 1x10 -3 o 0.001 volts
A,mA,µA
A: Ampere, unità di misura della corrente mA: Milliampere 1x10 -3 o 0.001 ampere
µA: Microampere 1x10 -6 o 0.000001 ampere
Ω,KΩ,MΩ
Ω: Ohm, unità di misura della resistenza
KΩ: Kilo ohm 1x10 3 o 1000 ohm
MΩ: Mega ohm 1x10 6 o 1.000.000 ohm
KHz
F,µF,nF
KHz: Kilohertz 1x10 3 o 1000 hertz
F: Farad unità di misura della capacità
µF: Micro Farad 1x10 -6 o 0.000001 farad nF: Nano Farad 1x10 -9 o 0.000000001 farad
2.3 Pulsanti
PULSANTE OPERAZIONI
Tasto di accensione e spegnimento
FUNC
H
MAX/
MIN
RANGE
REL
Hz/Duty
Tasto “FUNC” per selezionare le diverse funzioni.
CC/CA, resistenza/continuità/diodi/capacità e °C/°F.
Tasto “HOLD” per entrare e uscire dalla modalità Hold.
Premendo il tasto per 2", viene attivata la retroilluminazione
Premendolo una volta verrà visualizzato il valore massimo (MAX sul display); premendolo nuovamente verrà visualizzato il valore minimo (MIN sul display).
Mantenendolo premuto per più di 2", il multimetro tornerà alla funzione normale.
Durante la visualizzazione, il multimetro continuerà con la misura e i due valori verranno aggiornati
Consente di scegliere la modalità manuale o autorange.
La modalità di default è autorange.
Premendo il tasto per più di 2", cambierete la modalità di misurazione. Nella modalità manuale, premendo il tasto potrete cambiare la gamma di misurazione.
Premendo il tasto “REL” memorizzerete il valore da cui partire per la misurazione relativa.
Il valore che visualizzerete successivamente, sarà relativo a quello memorizzato e non allo 0.
Pertanto il valore sarà dato da: Vx-Vref = V display
Premendo il tasto “Hz/Duty” potrete selezionare le funzioni Frequenza e duty cicle.
COM
Terminale di ritorno per ogni misurazione (COM o massa)
Ingresso per misurazione di tensione, resistenza, capacità, frequenza, diodi e continuità
MAX/
MIN
Ingresso per misurazione di capacità, temperatura, hFE e corrente mA
10 A
Ingresso per misurazione di corrente da 400mA a 10A
(Inserire puntale rosso)
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
3.2.9 Misurazione Corrente
ATTENZIONE
ELCART
Manuale di istruzioni/Scheda tecnica
Per evitare danneggiamenti dello strumento non tentate di effettuare misurazioni su circuiti con potenziale di circuito aperto superiore a 250V.
Prima di effettuare misurazioni controllate l’integrità del fusibile dello strumento.
Utilizzate la giusta gamma di misurazione, non collegate mai i puntali in parallelo a un componente quando siete nella funzione di misurazione di corrente.
PAGINA 3 DI 9
4.2 Specifiche di misurazione
La precisione dello strumento è valida fino a un anno dopo la calibrazione, a una temperatura di lavoro compresa tra i 18°C e i 28°C con umidità relativa compresa tra 0% e 75%.
4.2.1 Tensione VCC
RANGE
200mV
2V
20V
200V
1000V
RISOLUZIONE
0.1mV
1mV
10mV
100mV
1V
PRECISIONE
±(0.5% of rdg+1digits)
±(0.5% of rdg+1digits)
±(0.5% of rdg+1digits)
±(0.5% of rdg+1digits)
±(0.8% of rdg+2digits)
Gamme di misurazione: CC: 200mA, 10A
Per misurazioni di corrente:
CA: 200mA, 10A
1 Togliete l’alimentazione al circuito da misurare e scaricate tutti i
condensatori
2. Ruotate la manopola sulla gamma di misurazione desiderata
3. Collegate il puntale nero al terminale COM e quello rosso al terminale
mA (per misurazioni fino a 200mA).
Per misurazioni fino a 10A, utilizzate il terminale 10A
4. Collegate il multimetro in serie al circuito
5. Accendete l’alimentazione del circuito e sul display vi verrà visualizzato
il valore misurato
Se sul display sarà visualizzato il simbolo “1”, significherà che dovete
selezionare una gamma di misurazione più alta
6. Spegnete l’alimentazione, scaricate i condensatori e poi rimuovete i
puntali
4 SPECIFICHE TECNICHE
4.1 SPECIFICHE GENERALI
- Condizioni ambientali:
- Grado di inquinamento:
- Altitudine di utilizzo:
- Temperatura di utilizzo:
- Temperatura di conservazione:
1000V CAT. II e 600V CAT. III
2
< 2000m
0°C÷40°C, 32°F÷122°F
(<80% RH, <10°C senza condensa)
-10°C÷+60°C, 14°F÷140°F
(<70% RH, rimuovere la batteria)
- Coefficiente di temperatura: 0.1%(precisione specificata)/°C
(<18°C o >28°C)
- MAX tensione tra terminali e massa:
- Protezione fusibile:
750VCA rms o 1000VCC mA: fusibile (FF400mA/1000V)
- Display:
A: fusibile (FF10A/500V)
3 1/2" digits LCD con indicazione automatica di funzioni e simboli
- Indicazione sovraccarico: LCD visualizzerà "1"
- Indicatore batteria scarica: simbolo " " visualizzato sul display
- Indicazione polarità:
- Alimentazione:
- Tipo Batteria:
- Dimensioni:
- Peso:
"-" visualizzato automaticamente
9V
NEDA 1604, 6F22, or 006P
188(L)x92(W)x50(H) mm
380gr. circa (batteria inclusa)
Impedenza di ingresso:
Max tensione di ingresso:
10MΩ
250Vcc/Vca rms per la gamma 200mV
1000Vcc o 750Vca rms per le altre gamme
4.2.2 Tensione VCA
RANGE
2V
20V
200V
750V
RISOLUZIONE
1mV
10mV
100mV
1V
PRECISIONE
±(0.8% of rdg+3digits)
±(0.8% of rdg+3digits)
±(0.8% of rdg+3digits)
±(1.2% of rdg+3digits)
Impedenza di ingresso:
Max tensione di ingresso:
Gamma di Frequenza:
10MΩ
250Vcc/Vca rms per la gamma 200mV
1000Vcc o 750Vca rms per le altre gamme
40Hz-400Hz
Risposta: media, calibrazione in valore efficace di un’onda sinusoidale
4.2.3 Resistenza
RANGE
200Ω
2KΩ
20KΩ
200KΩ
2MΩ
20MΩ
200MΩ
RISOLUZIONE
0.1Ω
1Ω
10Ω
100Ω
1KΩ
10KΩ
0.1MΩ
PRECISIONE
±(0.8% of rdg+3digits)
±(0.8% of rdg+1digits)
±(0.8% of rdg+1digits)
±(0.8% of rdg+1digits)
±(0.8% of rdg+1digits)
±(1.0% of rdg+2digits)
±(5.0% of rdg+10digits)
Protezione sovraccarico:
Tensione circuito aperto:
4.2.4 Diodi
RANGE RISOLUZIONE
1mV
Tensione di caduta CC:
Tensione di caduta inversa CC:
Protezione sovraccarico:
4.2.5 Continuità
RANGE FUNZIONE
Se viene misurata una resistenza ≤30Ω il buzzer suonerà
Corrente di test:
Tensione di caduta inversa CC:
Protezione sovraccarico:
250Vcc o 250Vca rms
<700mV
FUNZIONE
Il Display visualizzerà la tensione di caduta del diodo.
circa 1mA circa 2.8V
250Vcc o 250Vca rms circa 1mA circa 2.8V
250Vcc o 250Vca rms
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
4.2.6 Transistor
RANGE hFE
DESCRIZIONE
Il valore approssimativo Hfe
(0÷1000) del transistor verrà visualizzato sul display
4.2.7 Induttanza
RANGE
20mH
200mH
2H
20H
4.2.8 Capacità
RANGE
20nF
200nF
2µF
200µF
RISOLUZIONE
0.01mH
0.1mH
1mH
10mH
Protezione sovraccarico:
RISOLUZIONE
10pF
0.1nF
1nF
100nF
Protezione sovraccarico:
4.2.9 Corrente CC
RANGE
200mA
10A
RISOLUZIONE
0.1mA
10mA
ELCART
Manuale di istruzioni/Scheda tecnica
CONDIZIONI DEL TEST
Corrente di base approssimativa 10μA,
Vce approssimativa 2.8V
PRECISIONE
±(3.0% of rdg+5digits)
±(3.0% of rdg+5digits)
±(3.0% of rdg+5digits)
±(3.0% of rdg+5digits)
PAGINA 4 DI 9
5. MANUTENZIONE
Non tentate di riparare lo strumento. Solo persone qualificate possono effettuare riparazioni tecniche sul multimetro.
5.1 Manutenzione generale
ATTENZIONE
Per evitare danneggiamenti, non utilizzate getti d’acqua vicino allo strumento. Rimuovete i puntali e qualsiasi segnale di ingresso prima di aprire lo strumento.
Periodicamente strofinate lo strumento con un panno umido e un detergente delicato, non utilizzate panni abrasivi o solventi.
Sporcizia e umidità possono provocare errate misurazioni.
Per pulire i terminali:
- Spegnete lo strumento e rimuovete i puntali
- Scuotetelo per far uscire lo sporco dai terminali
- Impregnate un tampone con un detergente per pulizia
- Utilizzate il tampone intorno a ogni terminale. Il detergente creerà una patina di protezione contro le contaminazioni da umidità.
fusibile (FF400mA/1000V)
PRECISIONE
±(4.0% of rdg+3digits)
±(4.0% of rdg+3digits)
±(4.0% of rdg+3digits)
±(5.0% of rdg+10digits)
5.2 Sostituzione Batteria e fusibile
ATTENZIONE
Per evitare errori di lettura, scosse elettriche o guasti, sostituite la batteria appena verrà visualizzato il simbolo.
Utilizzate fusibili di ricambio identici a quello in dotazione.
FF400mA/1000V Min interrupt rating 10000A
FF10A/500V Min interrupt rating 10000A fusibile (FF400mA/1000V)
Per sostituire la batteria o il fusibile:
- Spegnete lo strumento
- Scollegate ogni puntali o qualsiasi connettore collegato al multimetro
- Con un cacciavite, svitate le due viti che chiudono il coperchio della
batteria
- Togliete il coperchio dallo strumento
- Rimuovete la batteria esaurita o il fusibile danneggiato
- Sostituite la batteria o il fusibile
- Richiudete il coperchio e fissatelo con le due viti al multimetro
PRECISIONE
±(1.5% of rdg+1digits)
±(2.0% of rdg+5digits)
Protezione sovraccarico: fusibile (FF400mA/1000V) gamma 10A fusibile(FF10A/500V)
Max corrente di ingresso: 200mA CC o 200mA CA rms per gamma mA
10A CC o 10A CA rms per gamma 10A
Per misurazioni >5A: 10 secondi max di durata, 1 minuto di stop; sopra 10A non specificato
4.2.10 Corrente CA
RANGE
200mA
10A
RISOLUZIONE
0.1mA
10mA
PRECISIONE
±(1.8% of rdg+3digits)
±(3.0% of rdg+7digits)
Protezione sovraccarico: fusibile (FF400mA/1000V) gamma 10A fusibile(FF10A/500V)
Max corrente di ingresso: 200mA CC o 200mA CA rms per gamma mA
10A CC o 10A CA rms per gamma 10A
Per misurazioni >5A: 10 secondi max di durata, 1 minuto di stop; sopra 10A non specificato
Gamma di Frequenza: 40Hz-400Hz
Risposta: media, calibrazione in valore efficace di un’onda sinusoidale
Informazioni agli utenti
Il simbolo riportato sull’apparecchiatura indica che il rifiuto deve essere oggetto di
“raccolta separata”.
Pertanto, l’utente dovrà conferire (o far conferire) il rifiuto ai centri di raccolta differenziata predisposti dalle amministrazioni comunali, oppure consegnarlo al rivenditore contro acquisto di una nuova apparecchiatura di tipo equivalente.
La raccolta differenziata del rifiuto e le successive operazioni di trattamento, recupero e smaltimento favoriscono la produzione di apparecchiature con materiali riciclati e limitano gli effetti negativi sull’ambiente e sulla salute eventualmente causati da una gestione impropria del rifiuto.
Lo smaltimento abusivo del prodotto da parte dell’utente comporta l’applicazione delle sanzioni amministrative di cui l’articolo 50 e seguenti del D. Lgs. N° 22/1997.
IMPORTATO E DISTRIBUITO DA
ELCART DISTRIBUTION SPA
Via Michelangelo Buonarroti, 46
20093 COLOGNO MONZESE (MI)
ITALY www.elcart.com - [email protected]
Made in China
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
1. GENERAL INSTRUCTIONS
1.1 Precaution safety measures
1.1.1 Preliminary
1.1.2 During use
1.1.3 Symbols
1.1.4 Instructions
1.2 Protection mechanisms
2. DESCRIPTION
2.1 Instrument Familiarization
2.2 LCD Display
2.3 Key pad
2.4 Terminals
2.5 Accessories
3. FUNCTION DESCRIPTION
3.1 General Functions
3.1.1 DATA HOLD mode
3.1.2 Battery saver
3.1.3 Non-Contact AC Voltage detection
3.2 Measurement Functions
3.2.1 AC and DC Voltage measurement
3.2.2 Resistance measurement
3.2.3 Diode Test
3.2.4 Continuity Check
3.2.5 Capacitance measurement
3.2.6 Inductance measurement
3.2.7 Transistor measurement
3.2.8 Current measurement
4. TECHNICAL SPECIFICATIONS
4.1 General specifications..
4.2 Measurement specifications
4.2.1 DC Voltage
4.2.2 AC Voltage
4.2.3 Resistance
4.2.4 Diode Test
4.2.5 Continuity Check
4.2.6 Transistor
4.2.7 Inductance
4.2.8 Capacitance
4.2.9 DC Current
4.2.10 AC Current
5. MAINTENANCE
5.1 General maintenance
5.2 Battery and Fuse replacement
ELCART
Manuale di istruzioni/Scheda tecnica
DIGITAL MULTIMETER WITH DISPLAY 3 ½ DIGIT LCD NIMEX
ART. 9/9420 NI 9420
USER'S MANUAL
PAGINA 5 DI 9
1. GENERAL INSTRUCTIONS
This instrument complies with IEC 61010-1, CAT. III 1000V and CAT. IV
600V overvoltage standards. See Specifications.
To get the best service from this instrument, read carefully this user's manual and respect the detailed safety precautions.
International symbols used on the Meter and in this manual are explained in chapter 1.1.3
1.1 Precautions safety measures
1.1.1 Preliminary
* Measurement category IV meters are designed to protect against transients
from the primary supply level (overhead or underground utility service).
* Measurement category III is for measurements performed in the
building installation.
* Measurement category II is for measurements performed on circuits
directly connected to the low voltage installation.
* Measurement category I is for measurements performed on circuits
not directly connected to MAINS.
* When using this Multimeter, the user must observe all normal safety
rules concerning:
- protection against the dangers of electric current
- protection of the Multimeter against misuse
* For your own safety, only use the test probes supplied with the instrument
Before use, check that they are in good condition
1.1.2 During use
* If the meter is used near noise generating equipment, be aware that
display may become unstable or indicate large errors.
* Do not use the meter or test leads if they look damaged.
* Use the meter only as specified in this manual;
* Use extreme caution when working around bare conductors or bus bars
* Do not operate the meter around explosive gas, vapor, or dust.
* Verify a Meter's operation by measuring a known voltage.
Do not use the Meter if it operates abnormally, protection may be impaired
When in doubt, have the Meter serviced.
* Uses the proper terminals, function, and range for your measurements
* When the range of the value to be measured is unknown, check that
the range initially set on the multimeter is the highest possible or,
wherever possible, choose the auto-ranging mode.
* To avoid damages to the instrument, do not exceed the maximum
limits of the input values shown in the technical specification tables.
* When the multimeter is linked to measurement circuits, do not touch
unused terminals.
* Caution when working with voltages above 60Vdc or 30Vac rms.
Such voltages pose a shock hazard.
* When making connections, connect the common test lead before
connecting the live test lead; when disconnecting, disconnect the live
test lead before disconnecting the common test lead.
* Before changing functions, disconnect the test leads from the circuit.
* For all dc functions, including manual or auto-ranging, to avoid the risk
of shock due to possible improper reading, verify the presence of any
ac voltages by first using the ac function.
Then select a dc voltage range equal to or greater than the ac range.
* Disconnect circuits power and discharge all high-voltage capacitors
before testing resistance, continuity, diodes, or capacitance.
* Never perform resistance or continuity measurements on live circuits.
* Before measuring current, check the meter's fuse and turn off power
to the circuit before connecting the meter to the circuit.
* In TV repair work, remember that high amplitude voltage pulses at the
test points can damage the multimeter.
Use of a TV filter will attenuate any such pulses.
* Use the 9V NEDA battery to power the Meter.
* Replace the battery as soon as the battery indicator ( ) appears.
With a low battery, the Meter might produce false readings that can
lead to electric shock and personal injury.
* Do not measure voltages above 600V in Cat. IV, or 1000V in Cat. III.
* Do not operate the Meter with the case (or part of the case) removed.
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
ELCART
Manuale di istruzioni/Scheda tecnica
1.1.3 Symbols:
Caution: refer to the instruction manual.
Incorrect use may result in damage to the device or its components
~
AC (Alternating Current) DC (Direct Current)
SYMBOL
Earth ground Double insulated
-
PAGINA 6 DI 9
MEANING
The battery is low
Warning:
To avoid false readings and malfunction, replace the battery as soon as the battery indicator appears.
Indicates negative readings
Indicator for ac voltage or current.
AC voltage and current are displayed as the average of the absolute value of the input, calibrated to indicate the equivalent rms value of a sine wave.
Fuse Conforms to European
Union directives
1.1.4 Instructions
* Remove test leads from the Meter before opening the battery cover.
* When servicing the Meter, use only specified replacement parts.
* Before opening up the instrument, always disconnect from all sources of
electric current and make sure you are not charged with static electricity
* Any adjustment, maintenance or repair work carried out on the meter while
it is live should be carried out only by appropriately qualified personnel,
after having taken into account the instructions in this present manual.
* A "qualified person" is someone who is familiar with the installation,
construction and operation of the equipment and the hazards involved.
He is trained and authorized to energize and de-energize circuits and
equipment in accordance with established practices.
* When the instrument is opened up, remember that some internal
capacitors can retain a dangerous potential even after the instrument
is switched off.
* If any faults or abnormalities are observed, take the instrument out of
service and ensure that it cannot be used until it has been checked out.
* If the meter is not going to be used for a long time, take out the battery and
do not store the meter in high temperature or high humidity environment.
1.2 Protection mechanisms
* Fused by the fuse (FF400mA/1000V) during capacitance,temperature,
mA and hFE measurements.
* A PTC resistor protects against permanent over-voltages of up to 250V
during resistance, frequency, continuity and diode test measurements.
2. DESCRIPTION
2.1 Instrument Familiarization
DC
H mH / H
V,mV
A,mA,µA
Ω,KΩ,MΩ
KHz
F,µF,nF
Indicator for dc voltage or current
The Meter is in the Diode Test mode
The Meter is in the Continuity Check mode
The Meter is in the Data Hold mode
H: Henry, the unit of inductance mH: Milli Henry 1x10 -3 or 0.001 henry
V: Volts, the unit of voltage mV: milliVolts 1x10 -3 or 0.001 volts
A: Amperes (amps), the unit of current mA: Milliampere 1x10 -3 or 0.001 amperes
µA: Microampere 1x10 -6 or 0.000001 amperes
Ω: Ohm, the unit of resistance
KΩ: Kilo ohm 1x10 3 or 1000 ohms
MΩ: Mega ohm 1x10 6 or 1.000.000 ohms
KHz: Kilohertz 1x10 3 or 1000 hertz
F: Farad, the unit of capacitance
µF: Micro Farad 1x10 -6 or 0.000001 farad nF: Nano Farad 1x10 -9 or 0.000000001 farad
2.3 Keypad
KEY
2.2 Display LCD
1. Voltage alert area
2. Voltage alert light
3. LCD display
4. Keypad
5. Rotary switch
6. Terminals
HOLD
2.4 Terminals
TERMINAL
COM
V/Ω
hFE/mA/Lx
10A
OPERATION PERFORMED turn the meter ON or OFF
Press HOLD to enter and exit the Data Hold mode
Press to turn the backlight on.
After about 5 seconds, the backlight is auto-off.
DESCRIPTION
Return terminal for all measurements.
(Receiving the black test lead or the “COM” plug)
Input for voltage, resistance, diode and continuity measurements
Input for capacitance, Inductance, hFE and mA current measurements
Input for 200mA to 10A current measurements
2.5 Accessories
Delivered with the multimeter:
* User's manual
* Test leads
* Carry case
* Special Multi-function socket
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
ELCART
Manuale di istruzioni/Scheda tecnica
3. FUNCTION DESCRIPTION
3.1 General Functions
3.1.1 DATA HOLD mode
Data Hold mode makes the meter stop updating the display.
Data Hold function can be cancelled by changing the measurement mode, or push HOLD key again.
To enter and exit the Data Hold mode:
1. Press HOLD key, fixes the display on the current value, is displayed
2. A second short press returns the meter to normal mode.
H
3.1.2 Battery Saver
The meter will be turned OFF automatic after approx 15 minutes
3.1.3 Non Contact AC Voltage detection
Hold the Meter so that the Meter’s top is vertically and horizontally centered and contacting the conductor, when the live voltage >110V(RMS), the sensing indicator will be on.
Note:
1. Even without LED indication, the voltage may still exist.
Do not rely on non-contact voltage detector to determine the presence of voltage wire; detection operation may be subject to socket design, insulation thickness and different type and other factors.
2. When the meter input terminals presence voltage, due to the influence of presenced voltage, voltage sensing indicator may also be bright.
3. Keep the meter away from electrical noise sources during the tests, i.e., fluorescent lights, dimmable lights, motors, etc..
These sources can trigger NON-Contact AC Voltage Detection Function and invalidate the test.
3.2 Measurement Functions
3.2.1 AC and DC Voltage measurement
WARNING
To avoid electrical shock and/or damage to the instrument, do not attempt to take any voltage measurement that might exceeds 1000Vdc or 750Vac rms.
The polarity of AC (alternating current) voltage varies over time; the polarity of DC (direct current) voltage is constant.
DC voltage ranges are:
AC voltage ranges are:
To measure AC or DC voltage:
1. Set rotary switch to the proper range.
200mV, 2V, 20V, 200V, 1000V
2V, 20V, 200V, 7500V
2. Connect the black test lead to the COM and red to the V terminals
3. Connect the test leads to the circuit being measured
4. Read the displayed value.
The polarity of red test lead connection will be indicated when making a
DCV measurement.
3.2.2 Resistance measurement
WARNING
To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before measuring resistance.
Resistance ranges are: 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 20MΩ, 200MΩ
To measure resistance:
1. Set the rotary switch to proper range.
2. Connect the black test lead to the COM and red to the V terminals
3. Connect the test leads to the circuit being measured and read the
displayed value.
NOTE:
On 20MΩ and 200MΩ ranges, the meter may take a few seconds to stabilize reading; this is normal for high resistance measuring.
3.2.3 Diode Test
WARNING
To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before testing diodes.
Use the diode test to check diodes, and other semi-conductor devices.
The diode test sends a current through the semiconductor junction, and then measures the voltage drop across the junction; a good silicon junction drops between 0.5V and 0.8V.
PAGINA 7 DI 9
To test a diode out of a circuit:
1. Set the rotary switch to range.
2. Connect the black test lead to the COM and red to the terminals
3. For forward-bias readings on any semiconductor component, place
the red test lead on the component's anode and place the black test
lead on the component's cathode.
4. The meter will show the approx. forward voltage of the diode.
If the test lead connection is reversed, only figure "1" displayed.
3.2.4 Continuity Check
WARNING
To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before testing for continuity.
Continuity is a complete path for current flow, the beeper sounds if a circuit is complete. These brief contacts cause the meter to emit a short beep.
To test for continuity:
1. Set the rotary switch to range.
2. Press the yellow key twice to activate Continuity Check.
3. Connect the black test lead to the COM and red to the Ω terminals
4. Connect the test leads to the resistance in the circuit being measured
5. When the test lead to the circuit is below approx. 30Ω, a continuous
beeping will indicate it.
Note:
- Continuity test is available to check open/short of the circuit.
3.2.5 Capacitance measurement
WARNING
To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC voltage function to confirm that the capacitor is discharged.
Capacitance ranges are:
To measure capacitance:
20nF, 200nF, 2µF, 200µF
1. Set the rotary switch to proper range.
2. Connect the black test lead to the COM and red to the terminals
3. Connect the test leads to the capacitor being measured
4. Read the displayed value
Note:
* The meter may take a few seconds to stabilize reading.
This is normal for high capacitance measuring.
* To improve the accuracy of measurements less than 20nF, subtract
the residual capacitance of the Meter and leads
3.2.6 Inductance measurement
The unit of inductance is the Henry (H)
Inductance ranges are:
To measure inductance:
20mH, 200mH, 2H, 200H
1. Set the rotary switch to proper range.
2. Connect the black test lead to the COM and red to the LX terminals
3. Connect the test leads to the inductance being measured
4. Read the displayed value
3.2.7 Transistor measurement
WARNING
To avoid electrical shock and/or damage to the instrument, do not apply more than 250Vdc or 250Vac rms between the hFE terminal and the COM terminal.
1. Set the rotary switch to hFE range.
2. Connect the “com” plug and “+” plug of the special multi-function
socket to the COM and hFE terminals.
3. Determine whether the transistor to be tested is NPN or PNP type
and locate the Emitter, Base and Collector leads.
4. Insert leads of the transistor into proper holes of the special
multi-function socket.
5. The meter will show the approx. hFE value at test condition of base
current 10μA and Vce 2.8V.
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
3.2.8 Current measurement
WARNING
ELCART
Manuale di istruzioni/Scheda tecnica
PAGINA 8 DI 9
4.2 Measurement specifications
Accuracy is specified for one year after calibration, at operating temperatures of 18°C to 28°C, with relative humidity at 0% to 75%.
Accuracy specifications take the form of:
±(% of Reading + Number of Least Significant Digits)
To avoid damage to the Meter or injury if the fuse blows, never attempt an in-circuit current measurement where the open-circuit potential to earth is greater than 250V.
To avoid damage to the meter, check the meter's fuse before proceeding.
Use the proper terminals, function, and range for your measurement.
Never place the test leads in parallel with a circuit or component when the leads are plugged into the current terminals.
4.2.1 DC Voltage
RANGE
200mV
2V
20V
200V
1000V
RESOLUTION
0.1mV
1mV
10mV
100mV
1V
Input impedance:
Max input voltage:
ACCURACY
±(0.5% of rdg+1digits)
±(0.5% of rdg+1digits)
±(0.5% of rdg+1digits)
±(0.5% of rdg+1digits)
±(0.8% of rdg+2digits)
10MΩ
250Vdc or ac rms for 200mV range and 1000Vdc or 750V ac rms for other ranges
DC current ranges are:
To measure current:
1. Turn off the power of the measured circuit.
DC: 200mA, 10A
AC: 200mA, 10A
Discharge all the high voltage capacitors.
2. Set the rotary switch to the proper range.
3. Connect the black test lead to the COM terminal and the red to the
mA terminal for a maximum of 200mA.
For a maximum of 10A, move the red test lead to the 10A terminal.
4. Break the circuit path to be tested.
Connect the black test lead to the more negative side of the break;
connect the red test lead to the more positive side of the break.
(Reversing the leads will give a negative reading,but will not damage
the Meter)
5. Turn on the power of the measured circuit, and then read the display.
Be sure to note the measurement units at the right side of the display
(mA or A). When only the figure "1" displayed, it indicates overrange
situation and the higher range has to be selected.
6. Turn off the power of the measured circuit and discharge all the high
voltage capacitors.
Remove the test leads and recover the measured circuit.
4 TECHNICAL SPECIFICATIONS
4.2.2 AC Voltage
RANGE
2V
20V
200V
750V
RESOLUTION
1mV
10mV
100mV
1V
Input impedance:
Max. input voltage:
Frequency Range:
Response:
ACCURACY
±(0.8% of rdg+3digits)
±(0.8% of rdg+3digits)
±(0.8% of rdg+3digits)
±(1.2% of rdg+3digits)
10MΩ
250Vdc or ac rms for 200mV range and 1000Vdc or 750V ac rms for other ranges
40Hz-400Hz
Average, calibrated in rms of sine wave
4.2.3 Resistance
RANGE
200Ω
2KΩ
20KΩ
200KΩ
2MΩ
20MΩ
200MΩ
RESOLUTION
0.1Ω
1Ω
10Ω
100Ω
1KΩ
10KΩ
0.1MΩ
Overload protection:
Open Circuit Voltage:
ACCURACY
±(0.8% of rdg+3digits)
±(0.8% of rdg+1digits)
±(0.8% of rdg+1digits)
±(0.8% of rdg+1digits)
±(0.8% of rdg+1digits)
±(1.0% of rdg+2digits)
±(5.0% of rdg+10digits)
250V dc or 250Vac rms less than 700mV
4.1 GENERAL SPECIFICATIONS
- Environment conditions:
- Pollution degree:
- Operating altitude:
- Operating temperature:
1000V CAT. II e 600V CAT. III
2
< 2000m
0°C÷40°C, 32°F÷122°F
- Storage temperature:
(<80% RH, <10°C non condensing)
-10°C÷+60°C, 14°F÷140°F
(<70% RH, battery removed)
- Temperature coefficient: 0.1% (specified accuracy)/°C (<18°C o >28°C)
- MAX voltage between terminals and earth ground:
750VCA rms o 1000VCC
- Fuse Protection: mA: fuse (FF400mA/1000V) / A: fuse (FF10A/500V)
- Display: 3 1/2" digits LCD automatic indication of functions and symbols
- Over Range indication: LCD will display "1".
- Low battery indication: the " " is displayed when the battery is
- Polarity indication:
- Power source:
- Battery type:
- Dimensions:
- Weight: under the proper operation range
"-" displayed automatically
9V
NEDA 1604, 6F22, or 006P
188(L)x92(W)x50(H) mm
380g. approx (battery included)
4.2.4 Diode
RANGE RESOLUTION
1mV
Forward DC Current:
Reversed DC Voltage:
Overload protection:
ACCURACY
Display read approx. forward voltage of diode approx. 1mA approx. 2.8V
250Vdc or 250Vac rms
4.2.5 Audible continuity
RANGE FUNCTION
The buzzer sound if the resistance of a circuit under test is less than 30Ω
Test Current:
Reversed DC Voltage:
Overload protection: approx. 1mA approx. 2.8V
250Vdc or 250Vac rms
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.
ART. 9/9450
4.2.6 Transistor
RANGE hFE
DESCRIPTION
Display read approx. HFE value
(0-1000) of transistor under test
(all type)
ELCART
Manuale di istruzioni/Scheda tecnica
TEST CONDITION
PAGINA 9 DI 9
5. MAINTENANCE
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information.
Base Current approx.
10μA, Vce approx. 2.8V
5.1 General Maintenance
4.2.7 Inductance
RANGE
20mH
200mH
2H
20H
Overload protection:
RISOLUZIONE
0.01mH
0.1mH
1mH
10mH
4.2.8 Capacitance
RANGE
20nF
200nF
2µF
200µF
RESOLUTION
10pF
0.1nF
1nF
100nF
PRECISIONE
±(3.0% of rdg+5digits)
±(3.0% of rdg+5digits)
±(3.0% of rdg+5digits)
±(3.0% of rdg+5digits) fuse (FF400mA/1000V)
ACCURACY
±(4.0% of rdg+3digits)
±(4.0% of rdg+3digits)
±(4.0% of rdg+3digits)
±(5.0% of rdg+10digits)
WARNING
To avoid electrical shock or damage to the meter, do not get water inside the case.
Remove the test leads and any input signals before opening the case
Periodically wipe the case with a damp cloth and mild detergent.
Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
- Turn the meter off and remove all test leads.
- Shake out any dirt that may be in the terminals.
- Soak a new swab with a cleaning and oiling agent (such as WD-40).
- Work the swab around in each terminal. The oiling agent insulates the terminals from moisture-related contamination.
Overload protection:
4.2.9 DC Current
RANGE
200mA
10A
RESOLUTION
0.1mA
10mA fuse (FF400mA/1000V)
ACCURACY
±(1.5% of rdg+1digits)
±(2.0% of rdg+5digits)
Overload protection:
Max. input current: fuse (FF400mA/1000V)
10A range fuse(FF10A/500V)
200mA dc or 200mA ac rms for mA range
10A dc or 10A ac rms for 10A ranges
For measurements>5A:
10 Seconds maximum ON to measure 1 minutes OFF
Above 10A unspecified
5.2 Battery and Fuse replacement
WARNING
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator ( ) appears.
Use only fuses with the amperage,interrutp,voltage
And speed ratings specified
FF400mA/1000V Min interrupt rating 10000A
FF10A/500V Min interrupt rating 10000A
To replace the battery of fuse:
- Turn the meter off.
- Disconnect test leads and/or any connectors from the terminals.
- Use a screwdriver to unscrew the two screws secured on the battery
cover.
- Take out the battery cover from the meter.
- Remove the used batteries or damaged fuse.
- Replace with the new 9V batteries (6F22) or new fuse.
- Rejoin the battery cover and secure by the two screws.
4.2.10 AC Current
RANGE
200mA
10A
RESOLUTION
0.1mA
10mA
ACCURACY
±(1.8% of rdg+3digits)
±(3.0% of rdg+7digits)
Overload protection:
Max. input current:
Resettable fuse (FF400mA/1000V)
10A range fuse(FF10A/500V)
200mA dc or 200mA ac rms for mA range
10A dc or 10A ac rms for 10A ranges
For measurements>5A:
10 Seconds maximum ON to measure 1 minutes OFF
Above 10A unspecified
Frequency Range:
Response:
40Hz-400Hz
Average, calibrated in rms of sine wave
Information for users:
The symbol on the equipment indicates that the waste must be “separately collected”.
Therefore, the user must carry (or have it carried) the waste to the separately collected waste centers set up by local governments, or deliver it to the dealer against purchase of a new equivalent-type equipment.
The separate waste collection and the subsequent processing, recovery and disposal operations favour the production of equipment with recycled materials and limit the negative effects on the environment and on health which may be possibly caused by the waste improper management.
The improper product disposal by the user causes the application of administrative sanctions according to the Art. 50 et. seq. of the Law Decree No. 22/1997.
IMPORTED AND DISTRIBUTED BY:
ELCART DISTRIBUTION SPA
Via Michelangelo Buonarroti, 46
20093 COLOGNO MONZESE (MI)
ITALY www.elcart.com - [email protected]
Made in China
ELCART DISTRIBUTION SPA via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY
Tel. +39 02.25117310 Fax +39 02.25117610 sito internet: www.elcart.com e-mail: [email protected]
La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.
The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.
La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.