Rossmax Suction Tester is a portable and easy-to-use device that allows users to quickly check the performance of a suction unit's pump without requiring a power source. It consists of an oil vacuum gauge, flow meter, PVC tubing, and stainless steel stand. With this device, users can effortlessly assess:

  1. Maximum (free) air flow: Determine the suction unit's maximum air flow capacity without any resistance.

  2. Maximum vacuum: Measure the highest vacuum level that the suction unit can generate.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Manual
Rossmax Suction Tester Instruction Manual | Manualzz

Portable Testing Device

for Suction unit

Manufacturer: Rossmax International Ltd.,

12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan www.rossmax.com

EC Representative: Rossmax Swiss GmbH,

Tramstrasse 16, CH-9442 Berneck, Switzerland

Model: Suction Tester

Instruction Manual

www.rossmax.com

15

30

20

50

40

60

70

76 cm Hg in Hg

30

20

0

10

10

0

30

20

60

70

50

40

15

30

20

0

10

10

0

15

20

50

40

30

60

70

76 cm Hg in Hg

30

20

0

10

10

0

15

20

60

50

70

40

30

20

10

0

10

30

0

20

60

50

15

40

30

70

76 cm Hg in Hg

30

20

0

10

10

0

15

30

20

60

50

40

70

76 cm Hg in Hg

30

20

0

10

10

0

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

0

10

10

0

20

30

15

60

50

40

70

76 cm Hg in Hg

30

20

0

10

10

0

15

20

50

40

30

60

70

76 cm Hg in Hg

30

20

0

10

10

0

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

0

10

10

0

15

30

20

60

50

40

70

76 cm Hg in Hg

30

20

0

10

10

0

30

20

30

15

10

60

70

50

20

30

40

60

50

40

20

70 in Hg

76 cm Hg in Hg

0

30

20

10

0 0

10

0

20

60

50

40

70

15

30

20

10

0

10

0

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

0

10

10

0

20

30

Introduction

15 15

20

10

20

40 40

50 30 50

20

60 60

10

15

70

76 cm Hg in Hg

0 15

60

30

70

20

50

30

60

50

10

10

30

0

20

60

50

15

30

20

40

30

10 30

50

60 20

40

40

20

40

10

60

50

20

Rossmax Suction Tester is a portable and easy-to-use device that allows the users to quickly

10 check the performance of the pump of a suction unit. Power source is not required for

Suction Tester during operation and the test result is easy to read and interpret. Suction

Tester consists of an oil vacuum gauge, flow meter, PVC tubing and stainless steel stand.

20

76 cm Hg

70

76 cm Hg in Hg in Hg

0

0

30

20

30

20

60

50

60

70

70

15

40

76 cm Hg in Hg

30

0

30

20

10

0

10

20

10

0

10

0

30

70

76 cm Hg in Hg

20

30

10

70

76 cm Hg in Hg

0

0

This device is used to check the working conditions of the pump, and you are able to check:

70

76 cm Hg in Hg

30

20

0

10

10

1. Maximum (free) air flow

2. Maximum vacuum

0

20

60

50

40

30

70

76 cm Hg in Hg

30

20

0

10

0

Flow meter

Regulator

15

30

20

60

50

20

70

40

50

30

15

20

40

0 60

70

10

30

20

0

0

10

10

30 0

Vacuum gauge

PVC tubing

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

10

0

10

0

Connector

Operation procedures

1. Check maximum (free) air flow

Step 1: Turn the regulator of flow meter anticlockwise around 2~3 circles.

Step 2: Insert the other end of PVC tubing into the air inlet of the suction unit.

Step 3: Power on the suction unit.

Step 4: Keep the suction unit running for 30 seconds to warm up.

Step 5: Adjust air flow and set the level of vacuum at 0 cmHg by turning the regulator of flow meter.

Step 6: Check the air flow rate.

Step 7: Turn off the suction unit.

Step 8: Disconnect the PVC tubing from suction unit.

2. Check maximum vacuum

Step 1: Turn the regulator of flow meter anticlockwise around 2~3 circles.

Step 2: Insert the other end of PVC tubing into the air inlet of the suction unit.

Step 3: Power on the suction unit.

Step 4: Keep the suction unit running for 30 seconds to warm up.

Step 5: Set the air flow rate at 0 L/min by turning the regulator of flow meter clockwise completely.

Step 6: Check the values on the vacuum gauge.

Installation

15

20

60

50

40

30

20

10

20

30

60

50

30

70

76 cm Hg in Hg

10

40

0

10

30

20

10

70

76 cm Hg in Hg

0

0

0

30

20

Insert PVC tubing into connector

15

10

40

50 30

20

60

10

70

76 cm Hg in Hg

0

0

Screw the nut securely

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

10

0

10

0

15

20

40

30

(2) Suction tester with NK1000 (Nasal Aspirator)

30

60

50

70

76 cm Hg in Hg

20

0

20

0

15

10

40

30 50

20

15

10

60

(1) 20

30

60

70

50

40

30

0

20

10

30

0

70

76 cm Hg in Hg

0

10

0

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

0

10

10

0

15

20

50

40

60

30

70

76 cm Hg in Hg

30

20

0

10

10

0

(2)

15

20

60

50

40

30

70 in Hg

30

20

10

0

10

0

15

20

50

40

30

60

70

76 cm Hg in Hg

30

20

0

10

10

0

15

20

60

50

40

30

70

76 cm Hg in Hg

30

20

0

10

10

0

20

60

70

50

40

15

30

20

10

0

10

30

0

30

20

60

70

50

40

20

30

60

50

40

70

76 cm Hg in Hg

30

20

10

20

30

30 Step 8: Disconnect the PVC tubing from suction unit.

0

0

20

50

15

40

30

10

20

0

10

0

60

15

60

70

50

76 cm Hg in Hg

40

15

Specifications

30

20

30

0

20

60

10

50

10

0

15

40

70

76 cm Hg in Hg

30

0

20

10

10

0

0

10

70

76 cm Hg

15 in Hg

30

0

20

10

30

60

50

70 in Hg

20

0

10

0

Power source

Operating temperature

Storage temperature

15

Air flow

Vacuum

40

30

20

30

60

50

40

70

76 cm Hg in Hg

30

0

20

10

10

0 scale accuracy scale accuracy

Weight

Dimension

No need

0 ~ 54ºC

-10 ~ 60ºC

0~50 L/min

±4% of full scale

0~-76 cmHg

±2.5% of full scale

730g

120(L) x 105(W) x 145(H) mm

70 in Hg

10

20

30

60

50

40

70 in Hg

30

20

10

20

30

60

50

40

30

20

10

0

10

0

2. Place the device and suction unit on a flat, horizontal surface while in use.

15

20

60

50

70

40

30

20 4. Do not completely unscrew regulator stem.

30

10

0 5. Do not attempt to disassemble the device.

6. Avoid excessive vibration, physical impact and dropping.

7. Do not store the device in direct sunlight, high temperature or humidity.

8. Keep the device away from static electricity.

20

60

50

70

15

40

30

20

10

0

10

30

0

15

20

60

50

40

30

70 in Hg

30

20

10

0

10

0

15

20

60

50

70

40

30

20

10

0

10

30

0