Precision and Analytical Balances
ML-T
Table of Contents
Installation and Putting into Operation 17
Leveling the balance using the leveling assistant ........................................ 23
Precision and Analytical Balances Table of Contents 1
Counting - Main configuration................................................................... 61
Defining reference piece weight in standard mode ....................................... 62
Defining reference piece weight in advanced mode...................................... 63
Check Weighing - Main configurations....................................................... 65
Before performing check weighing............................................................. 66
Performing check weighing ...................................................................... 67
Fine adjustment (model-dependent)................................................................ 83
2 Table of Contents Precision and Analytical Balances
Communication with Peripheral Devices 87
Installing SerialPortToKeyboard software .................................................... 89
Collecting measurement results and balance details with EasyDirect Balance ..................... 91
Cleaning the glass draft shield (0.1 mg and 1 mg models) ............................... 102
Analytical balances with a readability of 0.1 mg with draft shield ....................... 117
Precision balances with a readability of 1 mg with draft shield ........................... 119
Precision balances with a readability of 10 mg ................................................ 120
Precision balances with a readability of 100 mg .............................................. 122
Balances with a readability of 0.1 mg with draft shield (235 mm) ..................... 123
Balances with a readability of 1 mg with draft shield (235 mm) ........................ 124
Balances with a readability of 100 mg with square weighing pan....................... 126
MT-SICS interface commands and functions..................................................... 129
12 Accessories and Spare Parts 130
Precision and Analytical Balances Table of Contents 3
4 Table of Contents Precision and Analytical Balances
1 Introduction
Thank you for choosing a METTLER TOLEDO balance. The balance combines high performance with ease of use.
This document is based on the software version V 4.20.
EULA
The software in this product is licensed under the METTLER TOLEDO End
User License Agreement (EULA) for Software.
When using this product you agree to the terms of the EULA.
u www.mt.com/EULA
1.1 Further documents and information
u www.mt.com/mlt-analytical u www.mt.com/mlt-precision
This document is available in other languages online.
Search for software downloads
Search for documents u u www.mt.com/ml-t-RM www.mt.com/labweighing-software-download u www.mt.com/library
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
u www.mt.com/contact
1.2 Explanation of conventions and symbols used
Conventions and symbols
Key and/or button designations and display texts are shown in graphic or bold text, e.g., , Language .
Note For useful information about the product.
Refers to an external document.
Elements of instructions
In this manual, step-by-step instructions are presented as follows. The action steps are numbered and can contain prerequisites, intermediate results and results, as shown in the example. Sequences with less than two steps are not numbered.
§ Prerequisites that must be fulfilled before the individual steps can be executed.
1 Step 1
ð Intermediate result
2 Step 2
ð Result
1.3 Compliance information
The compliance information related to this product can be found in appendix.
Precision and Analytical Balances Introduction 5
2 Safety Information
Two documents named "User Manual" and "Reference Manual" are available for this instrument.
• The User Manual is printed and delivered with the instrument.
• The electronic Reference Manual contains a full description of the instrument and its use.
•
•
Keep both documents for future reference.
Include both documents if you transfer the instrument to other parties.
Only use the instrument according to the User Manual and the Reference Manual. If you do not use the instrument according to these documents or if the instrument is modified, the safety of the instrument may be impaired and Mettler-Toledo GmbH assumes no liability.
2.1 Definitions of signal warnings and warning symbols
Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following signal words and warning symbols:
Signal words
DANGER A hazardous situation with high risk, resulting in death or severe injury if not avoided.
WARNING A hazardous situation with medium risk, possibly resulting in death or severe injury if not avoided.
A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
CAUTION
NOTICE A hazardous situation with low risk, resulting in damage to the instrument, other material damage, malfunctions and erroneous results, or loss of data.
Warning symbols
General hazard: read the User Manual or the Reference Manual for information about the hazards and the resulting measures.
Electrical shock Notice
2.2 Product specific safety notes
Intended use
This instrument is designed to be used in laboratories by trained staff. The instrument is intended for weighing purposes.
Any other type of use and operation beyond the limits of technical specifications without written consent from
Mettler-Toledo GmbH is considered as not intended.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The instrument owner is responsible for the safety of all users of the instrument and third parties.
METTLER TOLEDO assumes that the instrument owner trains users to safely use the instrument in their workplace and deal with potential hazards. METTLER TOLEDO assumes that the instrument owner provides the necessary protective gear.
6 Safety Information Precision and Analytical Balances
Safety notes
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to death or injury.
1 Only use the METTLER TOLEDO power supply cable and AC/DC adapter designed for your instrument.
2 Connect the power cable to a grounded power outlet.
3 Keep all electrical cables and connections away from liquids and moisture.
4 Check the cables and power plug for damage and replace damaged cables and power plugs.
NOTICE
Damage to the instrument due to the use of unsuitable parts
Using unsuitable parts with the instrument can damage the instrument or cause it to malfunction.
− Only use parts from METTLER TOLEDO that are intended to be used with your instrument.
Precision and Analytical Balances Safety Information 7
5
4
3
2
1
10 mg
4
3
2
1
3 Design and Function
3.1 Overview balance
0.1 mg
6
1 mg
6
8
7
5
2
1
5
4
9
2
1
4
100 mg
8
8
8
7
5
9
16 15
1 Capacitive color TFT touch screen
3 Draft shield element
5 Handle for operation of the draft shield door
7 Level indicator
9 Safety feet (10 mg and 100 mg models)
11 USB host port
13 Bat.ON to switch the balance on during battery operation
15 RS232C serial interface
14 13 1211 10
2 Operating keys
4 Weighing pan
6 Glass draft shield
8 Leveling feet
10 Ethernet port
12 USB device port
14 Kensington slot for anti-theft purposes
16 Socket for AC/DC adapter
8 Design and Function Precision and Analytical Balances
3.2 Overview peripheral devices
LAN
RS232C
USB Host
USB Device
Interfaces and possible connectivity of peripheral devices:
USB DEVICE
PC
RS232C
RS-P20 / P-50 printer
Barcode reader
RS 2 nd display
PC
USB HOST
Barcode reader
USB-P25 printer
P-50 printer
USB memory stick
USB Keyboard
Wireless dongle
MTICWD-100
For more information about peripheral devices, see [Accessories } Page 130].
LAN
LAN
P-50 printer
3.3 Overview type plate
The balance type plate is located at the side of the balance, see example illustration and contains the following information:
1 2 3 4 5
7
1 Model designation
3 Readability
5 Manufacturer
7 Power supply
6
2 Year of manufacture
4 Maximum capacity
6 Serial number (SNR)
Precision and Analytical Balances Design and Function 9
3.4 Overview operation keys
1
2
3
4
5
Key
1 2
Name
ON/OFF
Capacitive color TFT touch screen
Tare
Zero
Home
3 4 5
Description
Switches the balance on or off.
General navigation
Tares the balance.
Zeros the balance.
Returns from any menu level, or other window to the application home screen.
3.5 User Interface
The screen displays information and allows the user to enter commands by tapping certain areas on its surface. You can choose the information displayed on the screen, change the balance settings and perform certain operations on the balance.
10 Design and Function Precision and Analytical Balances
3.5.1 Main settings and activities at a glance
Depending on the application, the options available to be selected and their content may differ.
Application main configuration
Balance information
General settings Activities
Precision and Analytical Balances Design and Function 11
3.5.2 Application home screen
The application home screen appears after switching the balance on. It always displays the last application that was in use before the balance was switched off. The application home screen is the main screen of the balance. Every function can be accessed from here. You can return to the application home screen at any time by pressing the home button in the lower right corner of the screen.
9 5 6 10
1
2
15
3
Prepare weight and start
4
14 8 13 7 11 12
Information and work bars
Name
1 Weighing information bar
2 Work title bar
3 Value bar
4 Main navigation
Information fields
Name
5 Weighing-in aid
Description
Displays the weighing-in aid and general balance information.
Displays information about the current activity.
Displays information about the current weighing process.
Work-related functions.
6 Short balance information
7 Weighing value field
8 Coach text field
Description
A dynamic graphic indicator displays how much of the total weighing range is in use.
Readability and capacity of the balance.*
Displays the value of the current weighing process (model-specific).
Displays instructions for the current weighing process.
* For legal-for-trade approved balances: Min (minimum capacity) and e (verification of scale interval) are shown in the upper left corner.
Action buttons
13 Print
14 Settings/preferences
15
Name
Status information field
Description
9 Main activity configuration To configure the current application, e.g., Weighing .
10 Detailed balance information Displays detailed technical data about the balance.
11 Weighing unit
12 Activities
Displays the unit of the current weighing process (model- and countryspecific).
Opens the activities selection.
Prints out results and/or settings (printer required).
Configures balance and user settings/preferences (application independent).
Displays information about the system status.
12 Design and Function Precision and Analytical Balances
3.5.3 Entering characters and numbers
The keyboard allows the user to enter characters, including letters, numbers and a range of special characters.
If a barcode reader is connected to your balance and your sample provides a barcode, scan the product barcode instead of entering the designation manually (e.g. the ID can be scanned via barcode reader to ensure that the sample is clearly assigned to the corresponding product). Additionally, it is possible to connect a USB keyboard to enter the information.
1 2
3
8 ID 1 - Label
7
4
Name
1 Input field
2 Delete all
3 Discard
4 Delete
5 Confirm
6 Shift
7 Specialized tabs
8 Explanation field
6 5
Description
Displays all characters that have been entered.
Deletes all entered characters.
Discards the entered data and exits the dialog.
Deletes the last character.
Confirms the data entered.
Switches between lower and upper case letters.
Switches keyboard mode for entering letters, numbers or special characters.
Extra information about the value to be entered.
Precision and Analytical Balances Design and Function 13
3.5.4 Lists and tables
Navigation: > General configuration
The basic elements in a simple list include a content title and a list of sub-elements. Tapping an element opens a list of sub-elements or an input dialog.
1 2
6
General configuration
Weighing options
Publishing
Devices
Network and Bluetooth
3
4
5
Name
1 List title
2 Contextual help
3 Back button
4 List element title
5 Scroll position
6 Selection tabs
Description
Title of the current list.
Additional information about the current process
Moves go one step back.
Title of the list element.
Scrolls through the list.
Tabs of the selectable sub-categories.
3.5.5 Detailed balance information
− Tap to open the general balance information menu.
Balance identification
The balance identification section contains identification data for both hardware and software.
− Tap to display Balance identification .
ð The display shows the Balance identification
as defined by the user. See [System settings } Page 43]
and information about the software and hardware.
Balance support information
The balance support information section contains information about last and next service checks and provides support contact information.
− Tap to display Support and service .
ð The display shows the Service information , Support information , and Quick support request .
14 Design and Function Precision and Analytical Balances
Quick support request
The Quick support request option contains a unique QR (Quick Response) code. If you have a QR code reader application on your smart phone, scan the QR code on the balance screen. The smartphone will automatically create an email with all relevant service information.
Support and service
Note
Make sure that the QR code can be identified by the smart phone. A program to read the QR codes must be installed. Make sure that there are no access restrictions, which could block your email program in some way.
3.5.6 Basic navigation
3.5.6.1 Touch screen navigation
To interact with the balance, use the screen and the operating keys at the bottom of the screen.
Opening an application
Navigation: > Activities - Weighing applications
To open settings or applications, tap with your finger on the application symbol, e.g., Weighing .
There is not enough space to show all the weighing applications on one screen. Scroll horizontally to reach the applications on the second page.
Activities
Weighing applications
Formulation
Counting Check weighing
Backweighing
Precision and Analytical Balances Design and Function 15
Scrolling
Navigation: > Quick settings
When the options are too numerous to be displayed on one screen, a blue bar on the right side appears. This means that the user can scroll up and down. To scroll simply place your finger anywhere on the list and drag it up or down.
Quick settings
Brightness
50%
Sound
ON (3 of 3 beeps)
Language
English
Glove mode
OFF
Using shortcuts
To simplify navigation on the touch screen, there are a few shortcuts that provide quick access to key areas of the balance. For example, the weighing value field on the application home screen works as a shortcut (see screen below), as does the weighing unit next to the weighing value field. Other shortcuts may be available to use depending on the application.
Every setting that can be changed directly via shortcut, can also be changed in the main configuration settings for that application.
16 Design and Function Precision and Analytical Balances
4 Installation and Putting into Operation
4.1 Selecting the location
A balance is a sensitive precision instrument. The location where it is placed will have a profound effect on the accuracy of the weighing results.
Requirements of the location
Place indoors on stable table
Ensure sufficient spacing Level the instrument Provide adequate lighting
Avoid direct sunlight Avoid vibrations Avoid strong drafts Avoid temperature fluctuations
Sufficient spacing for balances: > 15 cm all around the instrument
Take into account the environmental conditions. See "Technical Data".
4.2 Unpacking
Open the balance packaging. Check the balance for transport damage. Immediately inform a METTLER TOLEDO representative in the event of complaints or missing accessories.
Retain all parts of the packaging. This packaging offers the best possible protection for transporting the balance.
Precision and Analytical Balances Installation and Putting into Operation 17
4.3 Scope of delivery
Components
Balance with draft shield
Balance
235 mm
Weighing pan ø 90 mm
ø 120 mm
170×190 mm
Draft shield element
Pan support
Bottom plate
Protective cover
Universal AC/DC adapter
User Manual
Declaration of Conformity
0.1 mg
–
–
–
1 mg
–
–
–
–
10 mg
–
–
–
–
4.4 Assembling the balance
CAUTION
Injury due to sharp objects or broken glass
Instrument components, e.g., glass, can break and lead to injuries.
− Always proceed with focus and care.
Balances with a readability of 0.1 mg with draft shield (235 mm)
Place the following components on the balance in the specified order:
1 Push the side glass doors (1) back as far as they will go.
2 Insert the bottom plate (2).
3 Insert the draft shield element (3) and weighing pan
(5) with pan support (4).
100 mg
–
–
–
–
–
5
4
3
2
1
18 Installation and Putting into Operation Precision and Analytical Balances
Balances with a readability of 1 mg, with draft shield (235 mm)
Place the following components on the balance in the specified order:
1 Push the side glass doors (1) back as far as they will go.
2 Insert the bottom plate (2).
3 Insert the weighing pan (4) with pan support (3).
Balances with a readability of 10 mg with square weighing pan and draft shield element
1 Remove the two screws (1).
2 Remove the plate (2) and retain it.
3 Place the draft shield element (3) and fix it with the two screws.
4 Place the pan support (4) with weighing pan (5).
5
4
3
2
1
Balances with a readability of 100 mg with square weighing pan
Place the following components on the balance in the specified order:
• Place the pan support (1)
• Place the weighing pan (2)
2
1
4
3
2
1
Precision and Analytical Balances Installation and Putting into Operation 19
4.5 Putting into operation
4.5.1 Connecting the balance
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to death or injury.
1 Only use the METTLER TOLEDO power supply cable and AC/DC adapter designed for your instrument.
2 Connect the power cable to a grounded power outlet.
3 Keep all electrical cables and connections away from liquids and moisture.
4 Check the cables and power plug for damage and replace damaged cables and power plugs.
NOTICE
Damage to the AC/DC adapter due to overheating
If the AC/DC adapter is covered or in a container, it is not sufficiently cooled and will overheat.
1 Do not cover the AC/DC adapter.
2 Do not put the AC/DC adapter in a container.
1 Install the cables in such a way that they cannot be damaged or interfere with operation.
2 Insert the plug of the AC/DC adapter ( 1 ) in the power inlet of the instrument.
3 Screw the plug into the balance if applicable.
4 Insert the plug of the power cable into a grounded power outlet that is easily accessible.
ð The balance is ready for use.
1
Note
Always connect the AC/DC adapter to the balance before connecting to the power.
Do not connect the instrument to a power outlet controlled by a switch. After switching on the instrument, it must warm up before giving accurate results.
See also
2
4.5.2 Battery operation
The balance can also operate with batteries, especially useful in the case of regular power outages. Under normal operation conditions, the balance works independently of the AC power line for up to 8 hours for balances with a readability of 1 - 100 mg and 2 hours for balances with a readability of 0.1 mg (using alkaline batteries). If the battery run time is not sufficient it is recommended to use lithium batteries (e.g.,
Energizer™ ULTIMATE LITHIUM), achieving a battery run time of more than 8 hours.
Immediately after the AC power supply is interrupted, e.g., by withdrawing the power plug or if there is a power failure, the balance switches automatically to battery operation. Once the AC power supply is restored, the balance reverts automatically to AC operation.
20 Installation and Putting into Operation Precision and Analytical Balances
It is also possible to use rechargeable batteries. Charging batteries inside the balance is not possible.
Rechargeable batteries have a lower voltage of 1.2 V. Therefore, the battery indication shown on the balance might differ from the actual battery status.
Your balance uses 8 standard AA batteries (alkaline or lithium batteries preferred).
The Bluetooth dongle or other options only work when the balance is connected to the power supply and is not only powered by batteries.
100% full In battery operation mode, a battery symbol appears in the status information field. The size of the white bar indicates the battery condition. When the batteries status reaches 25%, part of the symbol becomes red. If the status is lower than 15%, the whole symbol becomes red.
75%
50%
25%
15%
Switch the balance on and off in battery operation
The balance must be switched on in battery operation with the Bat.ON switch on the rear of the balance. The key on the touch screen does not work, since the touch screen in the battery operation has no power supply in the extended state.
1 Press the Bat.ON switch on the rear of the balance to switch on the balance.
2 Press to switch off the balance.
Inserting / replacing batteries
WARNING
Death or serious injury due to electric shock
Contact with parts that contain a live current can lead to injury and death.
− Disconnect the instrument from the power supply when replacing batteries.
•
•
•
•
Read and follow all warnings and instructions supplied by the battery manufacturer.
Do not mix different types or brands of batteries. Performance of batteries varies depending on the manufacturer.
Remove the batteries from the balance if the balance is not used for a long period of time.
Batteries must be disposed of properly, according to local regulations.
§ Make sure that the balance is off before removing or inserting batteries.
1 Remove weighing pan, pan support and draft shield element or draft shield "100 mm" if present.
2 NOTICE: Damage to the instrument. Do not place the instrument on the pan support location bolt.
Turn the balance carefully on its side.
Precision and Analytical Balances Installation and Putting into Operation 21
3 Open and remove the battery-chamber cover.
4 Insert / replace the batteries with the correct polarity as shown in the battery holder.
5 Insert and close the battery-chamber cover.
6 Turn the balance carefully to its normal position.
7 Reinstall all components in the reverse order.
8 Press the Bat.ON switch on the rear of the balance to switch on the balance.
4.5.3 Switching on the balance
Before using the balance, it must be warmed up in order to obtain accurate weighing results. To reach operating temperature, the balance must be connected to the power supply for at least 30 minutes (60 minutes for 0.1 mg models).
§ The balance is connected to the power supply.
§ The balance is warmed up.
− Press .
ð After the start screen has disappeared, the application home screen will open.
When the balance is switched on for the first time, the Weighing application home screen will open. If the balance is switched on again, it will always start with the screen of the application last used before switching it off.
4.5.4 Changing the date and time
Navigation: > General configuration > System settings > Date and Time
The dialog (Picker view) allows the user to set the date and time.
Tap for Time and
25
for Date . The format can be selected by tapping .
Time
22 Installation and Putting into Operation Precision and Analytical Balances
Name
1 Change date/time format
2 Pick button
3 Picker field
4 Pick button
5 Selection tabs
Description
Various date/time formats can be selected.
Increment.
Displays the defined time/date.
Decrement.
Tabs of the selectable sub-categories.
4.5.5 Leveling the balance
Exact horizontal and stable positioning are essential for repeatable and accurate weighing results.
The balance can be levelled using the leveling assistant and/or the level indicator in front of the balance. There are two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench.
The balance must be leveled and adjusted each time it is moved to a new location.
Balances with two levelling feet
1 Position the balance at the selected location.
2 Align the balance horizontally.
3 Turning the two front leveling feet of the housing until the air bubble is in the middle of the glass.
È
Balances with four levelling feet
1 First turn the two rear leveling feet all the way in.
2 Adjust the two front leveling feet as previously described.
3 Turn the rear leveling feet down onto the surface for extra stabilizing safety, so the balance cannot tilt over under eccentric loads.
Example
Air bubble at 12 o'clock: turn both feet clockwise.
Air bubble at 3 o'clock:
Air bubble at 6 o'clock:
Air bubble at 9 o'clock: turn left foot clockwise, right foot counterclockwise.
turn both feet counterclockwise.
turn left foot counterclockwise, right foot clockwise.
4.5.5.1 Leveling the balance using the leveling assistant
When the balance is switched on at its new location, the symbol The instrument is out of level appears in the status information field on the left of the screen.
Precision and Analytical Balances Installation and Putting into Operation 23
1 Tap .
ð The screen Notifications appears.
2 Select The instrument is out of level .
ð The function Leveling assistant appears.
The function Leveling assistant is a step-by-step guide that helps to level the balance.
Navigation: Quick settings/Preferences > Leveling assistant
Instrument is out of level
After following the instructions, the leveling assistant will display the next steps. Follow the steps until the balance is level.
Note
Always use the physical air bubble on the level indicator as a reference. If the physical air bubble is centered, but the symbol The instrument is out of level still appears on the screen, please consider performing a central adjustment of the level indicator. See System settings.
4.5.6 Adjusting the balance
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location. This is also dependent on the ambient conditions. After reaching the operating temperature, it is important to adjust the balance in the following cases:
•
•
•
Before the balance is used for the first time.
If the balance has been disconnected from the power supply or in the event of power failure.
After significant environmental changes, e.g., temperature, humidity, air draft or vibrations.
• At regular intervals during weighing service.
4.6 Performing a simple weighing
This section describes how to perform a simple weighing operation. The basic navigation concept and the basic functions of the balance are also explained.
When you switch on the balance for the first time, the Weighing application home screen opens automatically.
If the balance has already been used, the application that was last used before the balance was switched off will open. If another application is running, switch to Weighing .
Navigation: > Activities - Weighing applications > Weighing
1 Press to zero the balance.
ð The application home screen appears.
2 Place the sample on the weighing pan.
ð The instability symbol appears and the value in the weighing value field becomes light blue .
24 Installation and Putting into Operation Precision and Analytical Balances
3 Wait until the instability symbol disappears and the value in the weighing value field becomes dark blue again.
ð The weighing process is complete.
ð The results are now displayed.
Zeroing
Use the zeroing key before starting to weigh anything.
1 Unload the balance.
2 Press to zero the balance.
ð All weight values are measured in relation to this zero point.
Taring
If you are working with a weighing container, tare the balance.
1 Place a container on the weighing pan.
ð The weight is displayed.
2 Press to tare the balance.
ð 0.000 g and Net appears in the display. Net indicates that all weight values displayed are net values.
Weighing
− Place the sample in the container.
ð The results are now displayed.
•
•
If the container is removed from the balance, the tare weight will be shown as a negative value.
The tare weight remains stored until the key is pressed again or the balance is switched off.
Switching weight units
There are several weighing units available. The default value is country-specific.
The weighing unit can be selected using the main configuration of the current application or via shortcut. This example describes how to change the weighing unit via shortcut.
Legal-for-trade
With approved balances, the menu topic has a fixed setting and cannot be changed.
1 Tap the weighing process unit (shortcut) gram (g) .
ð The screen Main unit appears.
2 Place your finger somewhere on the list and swipe up to scroll down.
3 Select another weighing unit (e.g. ounce (oz) ) by tapping it.
4 Tap to confirm.
ð The weighing unit gram (g) has now been changed to ounce (oz) .
Main unit g (Default)
Precision and Analytical Balances Installation and Putting into Operation 25
Changing the readability
There are several readabilities available. The default readability (d) is model-specific.
1 Tap the weighing value field.
2 Tap 10d - 0.001 g .
3 Confirm the selected readability by tapping .
ð The readability has now been changed.
Readability g (Default)
Print / transmit data
The balance can send data to a printer or a PC. Press the key to transmit the weighing results or settings via the interface. The procedure for activating and configuring a printer is described in "Publishing" and "Devices".
§ Printer is connected to the balance.
§ Printer is switched on.
§ Printer is activated and configured.
− Tap .
ð The data are transmitted.
Switching off
1 Press and hold until the dialog Switch-off appears.
2 Tap to confirm.
•
•
ð The balance switches off and enters standby mode.
After switching on from standby mode, the balance does not need to warm up. It is immediately ready to start weighing.
If the balance has been switched off manually, the display will also be off.
To switch off the balance fully, it must be disconnected from the power supply.
4.7 Transporting, packaging and storage
CAUTION
Injury due to sharp objects or broken glass
Instrument components, e.g., glass, can break and lead to injuries.
− Always proceed with focus and care.
26 Installation and Putting into Operation Precision and Analytical Balances
1 Press and hold the key.
2 Disconnect the balance from the AC/DC adapter.
3 Disconnect all interface cables.
4.7.1 Transporting over short distances
To move the balance over a short distance to a new location, follow the instructions below.
1 Hold the balance with both hands as shown.
2 Carefully lift the balance and carry the balance in horizontal position to the new location.
If you want the balance put into operation, proceed as follows:
1 Connect in reverse order.
2 Level the balance.
3 Perform an internal adjustment.
4.7.2 Transporting over long distances
To transport the balance over long distances, always use the original packaging.
4.7.3 Packaging and storage
Packaging
Store all parts of packaging in a save place. The elements of the original packaging are developed specifically for the balance and its components to ensure maximum protection during transportation or storing.
Storage
Store the balance under following conditions:
• Indoor and in the original packaging.
•
•
According to the environmental condition, see "Technical data".
When storing for longer than six months, the rechargeable battery maybe down (date and time get lost).
4.8 Weighing below the balance
Your balance is equipped with a weighing hook for performing weighing operations below the work surface
(weighing below the balance).
Note
Models with a glass draft shield: carefully lift the draft shield from the weighing platform and put it aside.
1 Press and hold the key.
2 Disconnect the balance from the AC/DC adapter.
3 Disconnect all interface cables.
Precision and Analytical Balances Installation and Putting into Operation 27
1 Remove weighing pan, pan support and draft shield or
"Easy draft shield" element if present.
2 NOTICE: Damage to the instrument. Do not place the instrument on the pan support location bolt.
Turn the balance carefully on its side.
3 Remove one of the weighing caps (1) depending on the model. Keep it for later use.
4 Turn the balance to its usual position and simply reinstall all components in the reverse order.
1
28 Installation and Putting into Operation Precision and Analytical Balances
5 General Settings
Navigation:
This section describes the procedure for adapting the balance to suit specific requirements. The settings apply to the entire weighing system and therefore to all applications.
The section Settings is divided into two sub-sections:
•
•
Quick settings
General configuration
All settings and configuration data can be printed out.
§ Printer is connected and configured.
1 Tap .
ð The screen Print and export report appears.
2 Select the settings, e.g., Quick settings and Weighing options for printing and confirm with .
ð A list of the selected settings will be printed.
Parameter
Description
Select the settings for printing.
Values
Quick settings* |
Weighing options* |
Publishing* | Devices* |
Network and Bluetooth*
| System settings* | User management* | ISO-Log
– Adjustments | ISO-Log
– Balance | ISO-Log –
Settings and status |
ISO-Log – Activity fails
5.1 Quick settings
Navigation: > Quick settings
The following options are available:
• Brightness
• Sound
• Language
• Glove mode
•
•
•
•
Quick adjustment
Date/Time widget
Leveling assistant
Change user password… (only if User management is activated)
5.1.1 Brightness
Navigation: > Quick settings > Brightness
This menu item can be used to adjust the display brightness. Each time you tap the bar, the brightness adjusts in increments of 10%.
The following options can be set:
Parameter
Brightness
* Factory setting
Description
Sets the brightness of the display (in 10% increments).
Values
10…100% (80%*)
Precision and Analytical Balances General Settings 29
5.1.2 Sound
Navigation: > Quick settings > Sound
This menu item can be used to adjust the sound and the sound volume.
The following options can be set:
Parameter
Stability beep
Description
Indicates when an unstable weight becomes stable.
Workflow feedback beep
Touch beep
Provides additional feedback in the event of input errors, messages and status notifications.
Notifies every touch of the interactive elements on the touch display and zero/tare bar.
Values
OFF | Low* | Medium |
High
OFF | Low* | Medium |
High
OFF* | Low | Medium |
High
* Factory setting
5.1.3 Language
Navigation: > Quick settings > Language
This menu item can be used to set the dialog language. The language changes immediately. All windows and messages are displayed in the selected language.
The following options can be set:
Parameter
Language
Description
Sets the preferred language.
The language is normally preset for the country of use.
Values
English | Deutsch |
Français | Español |
Italiano | Русский |
Polski | Česky |
Magyar | Nederlands |
Português PT. |
Português BRA. |
Türkçe | 中文 |
日本語 | 한국어
5.1.4 Glove mode
Navigation: > Quick settings > Glove mode
If Glove mode activated, the touch screen becomes more sensitive and easier to navigate when wearing gloves.
The following options can be set:
Parameter
Glove mode
* Factory setting
Description
Activates or deactivates the function Glove mode .
Values
ON | OFF*
5.1.5 Quick adjustment
Navigation: > Quick settings > Quick adjustment
If Quick adjustment is activated, the symbol appears in the main navigation of the application. You can start the adjustment directly from the work area.
The following options can be set:
Parameter Description
Quick adjustment Activates or deactivates Quick adjustment .
Starts the adjustment directly from work area.
Values
ON* | OFF
Internal adjustment
(model-specific) |
External adjustment
30 General Settings Precision and Analytical Balances
5.1.6 Date/Time widget
Navigation: > Quick settings > Date/Time widget
This option can be activated to permanently display the current date and time in the work area in the value bar above the weighing value field.
The following options can be set:
Parameter
* Factory setting
Description
Date/Time widget Activates or deactivates the permanent display of the current date and time in the work area.
Values
ON | OFF*
5.1.7 Leveling assistant
Navigation: > Quick settings > Leveling assistant
The function Leveling assistant helps to adjust the balance, e.g., if the balance has changed location.
For more information, see [Leveling the balance } Page 23].
5.1.8 Change password
Navigation: > Quick settings/Preferences > Change password…
If user management is activated and the user is protected with a password, the user password can be changed in this section.
1 Tap Change password… .
ð The screen Old password appears.
2 Enter the password and confirm with .
3 The screen New password appears.
4 Enter the password and confirm with .
ð The screen Confirm password appears.
5 Enter the password again and confirm with .
ð The screen Password changed appears.
6 Tap to confirm.
7 Tap to return to the application screen.
Precision and Analytical Balances General Settings 31
5.2 General configuration
Navigation: > General configuration
The following options are available:
• Weighing options
•
•
•
•
Publishing
Devices
Network and Bluetooth
System settings
•
•
User management
ISO-Log
5.2.1 Weighing options
Navigation: > General configuration > Weighing options
This menu item can be used to adapt the balance to suit specific requirements.
5.2.1.1 Weighing mode
Navigation: > General configuration > Weighing options > Weighing mode
This setting can be used to configure the balance to the weighing mode.
The following options can be set:
Parameter
Weighing mode
* Factory setting
Description
Sets the weighing mode.
Universal = for all standard weighing applications.
Dosing = for dosing liquid or powdery samples.
Values
Universal* | Dosing
5.2.1.2 Environment
Navigation: > General configuration > Weighing options > Environment
This setting is used to adapt the balance optimally to the ambient conditions at specific locations.
The following options can be set:
Parameter
Environment
Description
Sets the ambient conditions.
Values
Stable | Standard* |
Unstable | Very unstable
* Factory setting
5.2.1.3 Autozero
Navigation: > General configuration > Weighing options > Autozero
This menu item can be used to switch the automatic zero correction ON or OFF . It corrects the zero deviation, that may occur, e.g., due to slight fouling of the weighing pan.
Legal-for-trade
The Autozero cannot be deactivated for approved balances (except for some selected countries).
The following options can be set:
Parameter
Autozero
* Factory setting
Description
Activates or deactivates the automatic zero correction.
Values
ON* | OFF
32 General Settings Precision and Analytical Balances
5.2.1.4 Auto tare
Navigation: > General configuration > Weighing options > Auto tare
Tare automatically the first weight loaded on the empty weighing pan (zeroed). This function applies to all applications except for Formulation and Back-weighing .
The following options can be set:
Parameter
Auto tare
* Factory setting
Description
Activates or deactivates the auto tare function.
Values
ON | OFF*
5.2.1.5 Auto clear tare
Navigation: > General configuration > Weighing options > Auto clear tare
If the function Auto clear tare activated, the current tare is automatically cleared after removing all loaded weight from the weighing pan. This function applies to all applications except for Formulation and Backweighing .
The following options can be set:
Parameter
* Factory setting
Description
Auto clear tare Activates or deactivates the auto clear tare function.
Values
ON | OFF*
5.2.1.6 MinWeigh
Navigation: > General configuration > Weighing options > MinWeigh
The MinWeigh method function notifies when a measurement value does not meet the required weighing accuracy. The usage of MinWeigh is application-specific (contextual settings).
Setting the MinWeigh method
1 Tap MinWeigh .
ð The screen MinWeigh configuration appears.
2 Tap MinWeigh method .
ð The screen MinWeigh method appears.
3 Select method and confirm with .
ð The screen MinWeigh configuration appears.
4 Tap Minimum weight .
ð The dialog screen Minimum weight in g appears.
5 Enter the weight for the minimum weight and confirm with .
ð The screen MinWeigh configuration appears.
6 Tap to return to the previous screen.
The following options can be set:
Parameter
MinWeigh
Description
Activates or deactivates the MinWeigh method.
Sets the MinWeigh method.
Values
Not defined*
Customized | Certificate
* Factory setting
The selectable options in the dialog MinWeigh configuration vary depending on the chosen method.
The following methods are available:
• Customized
•
•
Certificate
OIML
Precision and Analytical Balances General Settings 33
Legal-for-trade
The OIML method is only available for approved balances.
5.2.1.7 Recall
Navigation: > General configuration > Weighing options > Recall
To retain and recall the last stable weight.
The following options can be set:
Parameter
Recall
* Factory setting
Description
Activates or deactivates recall.
Values
ON | OFF*
5.2.2 Publishing
Navigation: > General configuration > Publishing
The publishing options can be defined individually for all corresponding receiving devices/services available in the general settings area Devices .
Define which work output you want to produce, in which ways and in which format.
The available options may vary depending on the peripheral device connected and configured in "Devices". Not every option described is available for the selected peripheral.
The publishing menu is structured in the following sections:
• Print and export report
•
•
Send data
Advanced options
Publishing
Print and export report
Strip printout
RS232 printer csv file
Memory stick (USB Host)
Send data
Single value
Command host (Network)
5.2.2.1 Print and export report
Navigation: > General configuration > Publishing > Print and export
The following sections are available. While the main configuration is always available, the print report and export report file(s) are only available depending if a device/service is setup.
• General configuration
•
•
Print report
Export report file
34 General Settings Precision and Analytical Balances
Print and export – General configuration
Print and export
General configuration
Single values reports
Manual, stable
Workflow reports
Automatic
FACT report
Automatic
The following options can be set, which are valid for both the Print report and the Export report file .
Parameter
Single values reports
Description
Sets the behaviour of the printer for single values.
Values
Manual, stable* |
Manual, all values |
Automatic, stable (zero excluded) | Automatic, stable (zero included)
Automatic* | Manual Workflow reports Sets the workflow for the reports.
Automatic = the report is printed at the end of the workflow automatically.
FACT report
Manual = the report is printed at will.
Defines whether the FACT report is being printed automatically.
Automatic* | OFF
* Factory setting
Depending on the application the behavior, when a report/result is published, can be different.
• Single values reports (S): Each generated weight value can be manually or is automatically published
•
– Weighing
– Counting
– Check weighing
– Dynamic weighing
– Percent weighing
– Factor weighing
Workflow reports (W): Several weight values are stored during a workflow within the application and the final result can be manually or is automatically published at the very end of the workflow.
– Formulation
– Totaling
– Back-weighing
– Density
– Adjustment and tests
An example workflow of how a report could be set up correctly to publish the data is shown below. As a prerequisite, the device/service needs to be configured correctly beforehand in Devices .
1 Tap Print and export report .
ð The screen Print and export – General configuration appears.
2 Tap Single values reports .
ð The screen Single values reports appears.
3 Select, e.g., Automatic, stable and confirm with .
ð The next stable weight is automatically printed and/or exported.
Precision and Analytical Balances General Settings 35
36
4 Tap Workflow reports .
ð The screen Workflow reports appears.
5 Select, e.g., Automatic and confirm with .
ð The workflow report is automatically printed and/or exported.
Print report
Print out manually or automatically application, adjustment and test results on strip paper.
The following options can be set:
Parameter
Printout type
Description
Activates or deactivates the printer.
Values
ON* | OFF
Export report file
Export the weighing results and additional data according the report configuration in each application to a USB memory stick (CSV, TXT or XML file) or a FTP file server (XML file).
Please be aware the content of the file is always defined in the report configuration of each application (for more details go to section reports configuration).
The following options can be set:
Parameter
File type
File name
Date/Time of file creation
File location
Description
Defines the file type for export.
Each result generates a new XML file. However, for CSV files in single value reports (e.g., Weighing) all measurement results are stored in the same file, as long as the application, the user or any meta data is not changed.
Activates or deactivates the Standard (Activity name) file name.
Custom… = defines the file name (max. 16 characters).
Values csv | txt | xml
Depending on export mode.
Standard (Default)* |
Custom…
Activates or deactivates the date/time of the file creation at the end or beginning of the report.
ON* | OFF
Add at the end* | Add at the beginning
Describes the location on the USB memory stick or FTP server.
any
Create a folder for the publication, e.g.,
"METTLER TOLEDO
\Reports" on the storage device. The corresponding file will be saved in the created folder.
Below and example how to configure the publishing option to export a CSV file to USB memory stick which has to be attached.
Export report file
Memory stick (USB Host)
File type csv
File name
Standard (Activity name)
Date/Time of file creation
Add at the end
File location
\METTLER TOLEDO\Reports\
General Settings Precision and Analytical Balances
1 Tap Export report file .
ð The screen Export report file appears.
2 Tap File type .
ð The screen File type appears.
3 Select, e.g., csv and confirm with .
4 Tap File name .
ð The screen File name appears.
5 Select, e.g., Custom… by tapping.
ð The screen File name appears.
6 Enter a name and confirm with .
7 Tap to confirm.
ð The screen Date/Time of file creation appears.
8 Activate, e.g., Add at the beginning and confirm with .
ð The data and time is printed at the beginning of the report.
9 Tap File location .
ð The screen File location appears.
10 Enter a name for the path of the storage folder and confirm with .
11 Tap to confirm.
5.2.2.2 Send data
Navigation: > General configuration > Publishing > Send data
Send manually or automatically single weight- or result values to a PC with the PC-Direct, EasyDirect Balance or Command host service via RS232, USB or LAN/WLAN. We do not recommend to use WLAN in the mode automatic/continous.
For more information about PC-Direct, EasyDirect Balance and Command host and how to set it up, please refer to "Devices".
The following options can be set for PC-Direct, EasyDirect Balance and the Command host service.
PC-Direct or EasyDirect Balance
Parameter
ON* | OFF
Single values
Description
Activates or deactivates the option send data.
Send next stable weight.
Result values
Manual, stable* |
Manual, all values |
Automatic, stable (zero excluded) | Automatic, stable (zero included)
Automatic* | Manual Send result value.
Automatic = the result is sent on finishing automatically.
Manual = the result is sent at will.
* Factory setting
Command host
Parameter
ON* | OFF
Description
Activates or deactivates the option send data.
Manual, stable* Send next stable weight at will
Manual, all values
Send any stable or unstable weight at will
Precision and Analytical Balances General Settings 37
Automatic, stable Send next stable weight automatically
The next stable weight is sent after a minimal deviation, see table relationship between readability and deviation below.
Automatic, continuous
Send any stable or unstable weight automatically . (The weight value is sent with an update rate of 22.9 per seconds).
* Factory setting
Stability criteria: Relationship between readability and deviation
Readability
0.1 mg
Min. deviation
0.1 g
0.001 g
0.01 g
0.1 g
1 g
1 g
1 g
1 g
5 g
5.2.2.3 Advanced options
Navigation: > General configuration > Publishing > Advanced options
The following options can be set:
Parameter
Autopublish
Commands options
* Factory setting
Description
Activates or deactivates the publishing time interval.
Sets the publishing time interval of single values.
Please be aware that exporting a file to a USB memory stick or a
FTP server takes some time and we therefore do not recommend to set this value in these cases below 10 seconds.
Activates or deactivates the function Print and tare . The balance will tare automatically after publishing.
Values
ON | OFF* numerical values (1…
65535 seconds)
ON | OFF*
5.2.3 Devices
Navigation: > General configuration > Devices
Add a specific device and service, e.g., printer for RS232, barcode reader for USB or FTP server for WLAN.
Added Devices can be temporarily made unavailable for usage (turn OFF ).
Before adding devices or services using network or Bluetooth make sure that LAN, WLAN or Bluetooth are available and correctly configured and that Bluetooth devices are discoverable and in range. Please compare
Network and Bluetooth in General configuration .
Only one device/service set up per interface is possible. On creating a new device/service the existing will be replaced. The last saved configuration of the replaced device will be used as default when the same device type is added again at a later time.
The following devices and services can be added and configured via the different interfaces.
Parameter
P-20
P-52
Description
Connect to a Standard lab printer.
Connect to a Excellence lab printer (Dot-matrix technology).
Values
RS232* | USB Host
RS232 | USB Host |
LAN/WLAN | Bluetooth
P-56/58
Printer
Connect to a Excellence lab printer (Direct thermal technology).
RS232 | USB Host |
LAN/WLAN | Bluetooth
Connect to a custom printer.
Barcode reader Connect to a barcode reader.
RS232
RS232 | USB Host
38 General Settings Precision and Analytical Balances
Keyboard Connect a wired USB keyboard. Only few wireless keyboards with a USB/Bluetooth converter working according USB HID class are supported (e.g., Logitec K270).
Second display Connect a auxiliary weight display.
Memory stick
PC-Direct
Note : If you select Second display make sure that no other device is connected at RS232. Other devices could be damaged because of the voltage.
Connect a USB memory stick. Please use a high-quality USB
Memory stick, which is FAT32 formatted and empty for reliable operation. Wait at least 10 seconds after the last transaction, before you remove the USB memory stick.
Establishes a service to send the weight value to a PC. E.g., to
Microsoft Excel by placing the cursor in the required cell. PC-
Direct sends the weight value like a numeric keypad. Please make sure the NumLock is on. Via the USB device interface there is no additional software on the PC needed except a METTLER
TOLEDO USB driver.
EasyDirect
Balance
Establish this service to send measurement and device data to the
METTLER TOLEDO data management software EasyDirect Balance running on a PC ( www.mt.com/EasyDirectBalance ). To benefit from the full functionality, please connect the balance via the LAN interface. We do not recommend to use EasyDirect Balance if the balance is connected to the WLAN only.
File server (FTP) Establishes the service to connect to a FTP Server.
Command host Establish the Command host service to communicate with the balance via MT-SICS.
USB Host
RS232
USB Host
RS232 | USB Device
RS232 | LAN
Network: LAN/WLAN
RS232 | USB Device* |
Network: LAN/WLAN
* Factory setting
Adding a new device or service
By default a P-20 printer on RS232 and the command host service on USB device are already available. To add other or more devices and services there are two possibilities:
Adding a new device/service via the menu General configuration > Devices .
1 Tap .
2 Select the required device/service.
3 By following the step-by-step guidance the correct connection/interface is selected and the device and interface specific configuration parameters can be set. At the very end of the process all settings need to be confirmed and saved by pressing .
For printers there is the possibility to check if everything is working correctly by pressing when re-entering into the printer configuration.
Adding devices via USB Host
By connecting a, e.g., a printer on the USB host interface, the balance recognizes the new device and adds it automatically to the devices and service list. If the balance needs any additional information regarding the settings the new device symbol will blink in the home screen. In any case please check the settings of the new device/service by going into General configuration > Devices and press on the newly added device/service.
Before adding a printer via USB host, please make sure it is powered on correctly. Otherwise, it will not be recognized by the balance.
Deactivating or deleting a device/service
A device/service can be deactivated any time by switching them off in the device configuration General configuration > Devices .
Press the device/service you would like to deactivate and change the on/off switch to OFF .
To delete a device go to General configuration > Devices press like to remove and confirm with .
and select the device/service you would
Precision and Analytical Balances General Settings 39
40
Note
Please also consult Publishing and Reports configuration for further information.
RS232
The following options can be set depending on the connected device or service:
Parameter Description
Command set MT-SICS * | MT-PM | Sartorius 22 | Sartorius 16
MT-SICS = MT-SICS data transfer format is used.
MT-PM = emulates the data format of PM balances.
Baudrate
Bit/Parity
Sartorius 22/Sartorius 16 = emulates the data format of Sartorius balances.
600 | 1200 | 2400 | 4800 | 9600* | 19200 | 38400 | 57600 | 115200
(available values are device-specific)
8/No * | 7/No | 7/Mark | 7/Space | 7/Even | 7/Odd
Stop bits
Handshake
Character set
End of line
1 bit * |
Xon/Xoff
2 bits
IBM/DOS
* |
|
RTS/CTS
ANSI/WIN
| None
| UTF-8 *
UTF-8 = is a character encoding capable of encoding all possible characters, or code points, defined by unicode (device-specific).
<CR><LF> * | <CR> | <LF> | <TAB>
<CR><LF> = writes in the same column, e.g., in Excel.
<TAB> = writes in the same row, e.g., in Excel.
* Factory setting
USB Device (type B)
The following options can be set depending on the connected device or service:
Parameter Description
Command set MT-SICS * | MT-PM | Sartorius 22 | Sartorius 16
MT-SICS = MT-SICS data transfer format is used.
MT-PM = emulates the data format of PM balances.
Sartorius 22/Sartorius 16 = emulates the data format of Sartorius balances.
Character set
End of line
ANSI/WIN | UTF-8
Cannot be changed (device-specific).
<CR><LF> * | <CR> | <LF> | <TAB>
<CR><LF> = writes in the same column, e.g., in Excel.
<TAB> = writes in the same row, e.g., in Excel.
* Factory setting
USB Host (type A)
The following options can be set depending on the connected device or service:
Parameter
Character set
End of line
Description
IBM/DOS | ANSI/WIN | UTF-8
UTF-8 = is a character encoding capable of encoding all possible characters, or code points, defined by unicode (device-specific).
<CR><LF> * | <CR> | <LF>
*
* Factory setting
Bluetooth
The following options can be set depending on the connected device or service:
General Settings Precision and Analytical Balances
Parameter Description
Connection type not selectable
Device name
Character set
Select the device to be configured.
End of line
UTF-8
UTF-8 = is a character encoding capable of encoding all possible characters, or code points, defined by unicode (device-specific).
<CR><LF>
LAN / WLAN
The following options can be set depending on the connected device or service:
Parameter Description
Connection type not selectable
IP address
Port
Defines the IP address of the device.
Defines the device port.
Command set MT-SICS * | MT-PM | Sartorius 22 | Sartorius 16
MT-SICS = MT-SICS data transfer format is used.
MT-PM = emulates the data format of PM balances.
Character set
Sartorius 22/Sartorius 16 = emulates the data format of Sartorius balances.
ANSI/WIN | UTF-8 *
End of line <CR><LF> * | <CR> | <LF> | <TAB>
<CR><LF> = writes in the same column, e.g., in Excel.
<TAB> = writes in the same row, e.g., in Excel.
* Factory setting
Precision and Analytical Balances General Settings 41
5.2.4 Network and Bluetooth
Navigation: > General configuration > Network and Bluetooth
For wireless communication (WLAN and Bluetooth) a METTLER TOLEDO wireless dongle (MTICWD-100) is needed; WLAN and Bluetooth are only listed in the menu, when the wireless dongle is connected to the balance. Please contact your METTLER TOLEDO-sales representative for more information.
For installing and configuring a LAN/WLAN you need to have basic knowledge of TCP/IP-based networks and of network technology in general. If necessary, contact the network administrator of your IT department or your IT support.
Note
While WLAN and Bluetooth can be activate simultaneously, LAN and WLAN are exclusive. E.g., when switching
WLAN on, an existing LAN connection is switched off automatically.
The section has the following sub-sections:
• LAN
•
•
WLAN
Bluetooth
LAN
Ethernet interface for connecting the instrument to a TCP/IP network. The most simple network can be implemented by connecting the instrument directly to the PC with a cross-over cable (RJ45 socket).
We recommend to use the default setting DHCP (automatic network settings).
The following options can be set:
Parameter
Configuration mode
IP address
Subnet
Standard
Gateway
MAC-Address
Description
Defines the parameter for the WLAN connection.
DHCP = the parameters for the Ethernet connection will be automatically set.
Manual = the options for the Ethernet connection must be set manually by the user.
Values
DHCP* | Manual
If the IP is not to be automatically obtained, you can enter it here. 000.000.000.000 …
255.255.255.255
Defines the subnet mask that is used by the TCP/IP protocol to determine whether a host is on the local subnet or on a remote network.
Defines the address of the standard gateway that links the host's subnet to other networks.
000.000.000.000 …
255.255.255.255
000.000.000.000 …
255.255.255.255
Information on the MAC address (Media Access Control) that is used to uniquely identify the balance in the network.
not selectable
* Factory setting
Connecting LAN automatically (DHCP)
§ Balance is connected via a Ethernet cable to the LAN.
1 Tap LAN .
ð The screen LAN appears.
2 Tap ON to activate LAN and confirm with .
ð LAN connection is established.
WLAN
The following options can be set:
42 General Settings Precision and Analytical Balances
Parameter
No network connected
Configuration mode
IP address
Subnet
Standard
Gateway
MAC-Address
Description
Select a network or search for a non-visible one. Please repeat the network search , if the network you are looking for is not listed initially.
Defines the parameter for the WLAN connection.
DHCP = the parameters for the Ethernet connection will be automatically set.
Manual = the options for the Ethernet connection must be set manually by the user.
Values
DHCP* | Manual
If the IP is not to be automatically obtained, you can enter it here. 000.000.000.000 …
255.255.255.255
Defines the subnet mask that is used by the TCP/IP protocol to determine whether a host is on the local subnet or on a remote network.
000.000.000.000 …
255.255.255.255
Defines the address of the standard gateway that links the host's subnet to other networks.
000.000.000.000 …
255.255.255.255
Information on the MAC address (Media Access Control) that is used to uniquely identify the balance in the network.
not selectable
Bluetooth
The following options can be set:
Parameter
Bluetooth identification
Secure mode
* Factory setting
Description
With this identification the balance is shown to other Bluetooth devices.
Activates or deactivates the secure mode.
Values
Freely configurable
(balance model name*)
ON* | OFF
§ Wireless dongle MTICWD-100 is connected to USB Host interface.
1 Tap Bluetooth .
ð The screen Bluetooth appears.
2 Tap ON and confirm with .
ð The screen Network and Bluetooth appears.
ð Bluetooth is available.
5.2.5 System settings
Navigation: > General configuration > System settings
This section describes the procedure for adapting the balance to suit specific requirements.
The following options can be set:
Parameter Description
Date and Time Sets the time format and the date.
The format can be selected by tapping .
Values
Time:
24:MM* | 12:MM |
24.MM | 12.MM
Date:
DD.MM.YYYY* |
D.MMM YYYY |
MM/DD/YYYY |
MMM DD YYYY |
YYYY-MM-DD |
YYYY/MM/DD
Precision and Analytical Balances General Settings 43
44
Balance identification
Sleep mode
Backlight OFF
Defines a balance identification.
By default the balance identification consists of the balance model and the serial number.
Values can be defined individually.
Activates or deactivates the screen saver.
Determines when the screen saver appears with date and time.
Tap on the screen to exit the screen saver.
Activates or deactivates the display turns off.
Determines when the display turns off.
Tap the screen to end Backlight OFF mode.
ON* | OFF
After 30 seconds
After 1 minute
After 2 minutes
After 5 minutes
After 10 minutes*
ON | OFF*
After 30 seconds
After 1 minute
After 2 minutes
After 5 minutes
After 10 minutes
ON* | OFF Quick wake up Terminates Sleep mode and/or Backlight OFF .
Out of level notification
By placing the sample on the weighing pan, the Sleep mode and/or Backlight OFF will be terminated.
Activates or deactivates the function Out of level notification .
Service due notification
Activates or deactivates the function Service due notification .
ON* | OFF
ON* | OFF
* Factory setting
If Sleep mode and Backlight OFF have the same value, the screen saver briefly appears before the backlight switches off.
Advanced options
The following options can be set:
Parameter Description
Reset balance… Resets the balance to factory settings.
Backup and restore...
Generates a backup of the current balance settings (excluding MinWeigh, service due notification and ISO-Log).
To execute a backup, an external storage device (USB memory stick, FAT32) must be connected to the USB host (type A). Additionally, if an external storage device (USB memory stick, FAT32) is attached with a back-up file, it can be restored. We do not recommend to back-up/restore balance settings between approved and non-approved balance models.
Software update...
Communication
OFF…
Activates or deactivates the option to update the balance software.
The balance software can be updated by a METTLER TOLEDO service technician only.
ON* | OFF
Activates or deactivates communication from/to the balance via any interface except the touch display.
If turn off, any device connected to as well as any service/function using any balance interface will not be available for usage anymore, e.g., Printer, Backup/Restore function.
Network and Bluetooth-, Devices- and Publishing settings will not be available anymore.
Note
By resetting the balance, any changes to the general settings and contextual settings that have been made along with any temporary collected data, e.g., paused applications or statistics will be lost.
General Settings Precision and Analytical Balances
5.2.6 User management
Navigation: > General configuration > User management
In the section User management rights for users and user groups can be defined. User management simplifies operations by customizing the user interface according the requirements of every user. Additionally, it helps in regulatory environments that actions and results on a device can be traced to the individual user. If user management is activated, access to the system is user-specific (user logon). The access to the system can be protected by a password. Create up to 10 users and assign them a previously defined access authorization
(group). According the access authorization only the relevant information is shown to a specific user. After switching on the balance, the user must login with the password, depending on the settings. The user must logout to switch off.
1 Tap User management .
ð The screen User management – General appears.
2 Activate the User management and confirm with .
3 Tap to return to the General configuration screen.
The section User management is divided into the following sub-sections:
• User management – General
defines general settings for all user profiles, see [General } Page 45].
•
•
User management – Groups
defines the settings for the user groups, see [Groups } Page 46].
User management – User (1)
defines the settings for a single users, see [Users } Page 47].
User password
The user password can be defined by the user. The length of the configurable user password is restricted to 12 digits.
What if you forget the password?
If you forget or lose the password, please ask the user with the administrator rights to provide you a new password. If also the administrator forgets or loses its password, please contact a METTLER TOLEDO representative. For more information please tap on the "?" icon on the Login – User screen of the administrator and generate an email with the relevant information by tapping on the on the right.
5.2.6.1 General
Navigation: > General configuration > User management > User management – General
In this section you can activate the Auto Lock function to block access to the work area for the currently logged in user after a certain time of inactivity or a specific balance condition.
1 Tap User management .
ð The screen User management – General appears.
2 Activate the User management .
3 Tap Auto Lock .
ð The screen Auto Lock appears.
4 Activate Auto Lock .
5 If necessary, change the conditions, when Auto Lock is activated (after a certain time, On Sleep , On
Backlight OFF ) and confirm with .
The following options can be set:
Parameter
Auto Lock
Description
Activates or deactivates the automatic lock access to work area.
Defines the screen lock time.
Values
ON | OFF*
After…* | On Sleep | On
Backlight OFF
* Factory setting
Precision and Analytical Balances General Settings 45
5.2.6.2 Groups
Navigation: > General configuration > User management > User management – Groups
In this section the access rights of the different user groups can be defined. While the access rights of the
Administrators group are fix, the access rights for the Operators group can be changed. By customizing the available functions and settings for each user group, the complexity for the user can be reduced and the easeof-use, productivity and process security can be increased.
The following 2 groups with different pre-defined access rights and group properties are available:
• Administrators
• Operators
Group properties
1 Tap User management .
ð The screen User management – General appears.
2 Tap .
ð The screen User management – Groups appears.
3 Tap, e.g., Operators .
4 If necessary, edit the group parameter and confirm with .
5 Tap to return to the User management – Groups screen.
Access rights - Activities
For the Administrator group, all available rights are activated by default and cannot be changed.
The following options can be set:
Parameter
Run applications
(All)
Description
Defines which of the available applications the group can access and execute.
Values can be (deactivated) or (activated)*.
Values
Weighing* | Counting* |
Check weighing* |
Formulation* | Totaling*
| Back-weighing* |
Dynamic weighing* |
Percent weighing * |
Density* | Factor weighing*
ON* | OFF Run adjustments
(All)
Activates or deactivates the adjustments. Enables the group to access and execute all available adjustments.
Values can be (deactivated) or (activated)*.
Run tests (All) Activates or deactivates the tests. Enables the group to access and execute all available tests.
Values can be (deactivated) or (activated)*.
Configure applications
Configure the applications. Enables the group to access and edit the contextual settings of the application. The group has the right to access and execute.
Configure adjustments
Values can be (deactivated) or (activated)*.
Configure the adjustments. Enables the group to access and change the settings of adjustments.
Values can be (deactivated) or (activated)*.
Configure tests Configure the tests. Enables the group to access and change the settings of tests.
Values can be (deactivated) or (activated)*.
* Factory setting
ON* | OFF
ON* | OFF
Main configuration* |
Reports* | Statistics*
ON* | OFF
ON* | OFF
Access rights – Settings
The Administrator group have all rights activated by default.
46 General Settings Precision and Analytical Balances
Parameter
Quick settings/
Preferences
Weighing options Activates or deactivates the access to the weighing options.
Values can be (deactivated) or (activated)*.
Publishing
Devices and services
Activates or deactivates the access to the publishing settings.
Values can be (deactivated) or (activated)*.
Activates or deactivates the access to the device and service settings.
Values can be (deactivated) or (activated)*.
Network and
Bluetooth
System settings
Standard
Description
Activates or deactivates the access to the quick settings/ preferences.
Values can be (deactivated) or (activated)*.
Activates or deactivates the access to the network and Bluetooth settings.
Values can be (deactivated) or (activated)*.
Activates or deactivates the access to the standard system settings.
Values can be (deactivated) or (activated)*.
System settings
Advanced
User management
ISO-Log
Activates or deactivates the access to advanced system settings.
Values can be (deactivated) or (activated)*.
Activates or deactivates the access to the user management settings.
Values can be (deactivated) or (activated)*.
Activates or deactivates the access to the ISO-Log.
Values can be (deactivated) or (activated)*.
* Factory setting
Editing a group
1 Tap User management .
ð The screen User management – General appears.
2 Tap .
ð The screen User management – Groups (4) appears.
3 Tap, e.g., Operators .
4 Edit the group parameter (see table above) and confirm with .
5 Tap to return to the User management – Groups (4) screen.
Values
ON* | OFF
ON* | OFF
ON* | OFF
ON* | OFF
ON* | OFF
ON* | OFF
ON* | OFF
ON* | OFF
ON* | OFF
5.2.6.3 Users
Navigation: > General configuration > User management > User management – User (1)
In this section, you can create, edit or delete user accounts and set a user password. Password protection is deactivated by default.
Create a user
1 Tap User management .
ð The screen User management – General appears.
2 Tap .
ð The screen User management – User (1) appears.
3 Tap .
ð The screen User name appears.
Precision and Analytical Balances General Settings 47
48
4 Enter a name and confirm with .
ð The screen Assigned group appears.
5 Select a configuration of the access rights by tapping, e.g., Operators and confirm with .
ð The screen Password appears.
6 If necessary, activate the password and confirm with .
ð The screen New password appears.
7 Enter the password and confirm with .
ð The screen Confirm password appears.
8 Enter the password again and confirm with .
ð The screen Password defined appears.
9 Tap to confirm.
ð The new user is displayed in the user management.
10 Tap to return to the User management – General screen.
The following options can be set:
Parameter
User name
Description
Assigned group Displays the group assignment.
Password
* Factory setting
Displays or changes the name of the logged in user.
Activates or deactivates the password.
Values any (1 … 12 chars) not selectable
ON | OFF*
Editing a user
1 Tap User management .
ð The screen User management – General appears.
2 Tap .
ð The screen User management – User (1) appears.
3 Select the user you want to edit by tapping.
4 If necessary, change the assigned group, password or user name and confirm with .
5 Tap to return to the User management – User (1) screen.
Deleting a user
Note
The user Administrators and the currently active user cannot be deleted.
1 Tap User management .
ð The screen User management – General appears.
2 Tap .
ð The screen User management – User (1) appears.
3 Tap .
ð The screen Remove users appears.
4 Select the user you want to delete by tapping.
ð The screen Confirm remove appears.
5 Tap to return to the User management – User (1) screen.
ð The screen User management – User (1) appears.
Activate password
Activate the password for the administrator and/or user.
1 Tap User management .
ð The screen User management – General appears.
General Settings Precision and Analytical Balances
2 Tap .
ð The screen User management – User (1) appears.
3 Select, e.g., Administrator by tapping.
ð The screen Administrator appears.
4 Tap Password .
ð The screen Password appears.
5 Activate Password and confirm with .
6 Enter the password and confirm with .
ð The screen Confirm password appears.
7 Enter the password again and confirm with .
ð The screen Password defined appears.
8 Tap to confirm.
9 Tap to return to the User management – User (1) screen.
ð The screen User management – User (1) appears.
10 Tap to return to the User management – General screen.
5.2.6.4 User login / lock / logout
Navigation: > General configuration > User management > User management – User (1)
If the user password is activated, the user must login with the defined password. To release the balance for a different user, the user needs to logout first. Please be aware that the data and results generated by a specific user are not saved after a logout.
5.2.6.4.1 Login
There are different starting points to login to your user account.
Another user is still active and the screen is not locked
§ User management is activated.
§ Users are defined.
§ Application screen appears.
1 Press and hold until the dialog screen appears.
ð The screen Quit balance appears.
2 Tap Logout (logged in user).
ð The screen Logout appears.
3 Tap to confirm.
ð The screen Login – User (4) appears.
4 Select the user, e.g., LEA by tapping.
ð The application screen appears directly or after the password of the user LEA is entered. LEA is the logged in user.
Quit balance
Lock
(Password OFF)
Logout
Switch-off
Precision and Analytical Balances General Settings 49
Another user is still active and the screen is locked
§ User management is activated.
§ The screen is locked (either manually by the user or automatic as Auto Lock is activated).
§ Users are defined.
§ Screen with logged in user appears.
1 Tap anywhere on the screen or the terminal bar to unlock.
ð The application screen appears.
ð If the currently logged in user is protected by a password, the Password screen appears.
"JOHN" logged in
2 Tap .
ð The screen Login – User (4) appears.
ð If the currently logged in user is not protected by a password the application screen appears.
3 Press and hold until the dialog screen and follow the instructions as explained in the chapter logout later in this chapter.
ð The screen Login – User (4) appears.
4 Select the user, e.g., LEA by tapping.
ð The application screen appears directly or after the password of the user LEA is entered. LEA is the logged in user.
No user is active and the login screen appears
§ User management is activated.
§ Users are defined.
§ Screen with Login – User (4) appears.
− Tap, e.g., LEA .
ð The application screen appears. LEA is the logged in user.
Enter
Password
Login
User (4)
5.2.6.4.2 Lock
With the lock function the access to the work area of the current user is blocked. The lock function can be activated either automatically, see Auto Lock
[General } Page 45] for the configuration or can be activated
manually anytime. Below the manual process is described.
50 General Settings Precision and Analytical Balances
§ User is logged in.
§ Screen with the logged in user appears.
1 Press and hold until the dialog screen appears.
ð The screen Quit balance appears.
2 Tap Lock (Password OFF) .
ð The screen with the logged in users appears.
"LEA" logged in
Quit balance
Lock
(Password OFF)
Logout
Switch-off
5.2.6.4.3 Logout
§ User is logged in.
1 Press and hold until the dialog screen appears.
ð The screen Quit balance appears.
2 Tap, e.g., Log out "LEA" ("LEA" is the active logged in user).
ð The screen Logout appears.
3 Tap to confirm.
ð The screen Login – User (4) appears.
Quit balance
Lock
(Password OFF)
Logout
Switch-off
5.2.7 ISO-Log
Navigation: > General configuration > ISO-Log
In this section, detailed information about adjustments already performed, intensity of use and settings can be displayed. ISO-log can store up to 999 events until the oldest one is overwritten.
1 Tap ISO-Log .
ð The screen ISO-Log appears.
2 Tap, e.g., ISO-Log – Adjustments .
ð The screen ISO-Log – Adjustments with all results appears.
3 Tap to return to the configuration screen.
4 Tap to return to the application home screen.
The following options can be set:
Precision and Analytical Balances General Settings 51
Parameter
ISO-Log –
Adjustments
ISO-Log –
Balance
ISO-Log –
Settings and status
ISO-Log – Activity fails
Description
Displays detailed information about adjustments already performed.
Displays detailed information about the balance history.
Displays detailed information about setting changes.
Displays detailed information about communication errors.
52 General Settings Precision and Analytical Balances
6 Application Settings
Navigation: > Activities - Weighing applications
Each application can be defined manually via its application settings. Select the application and tap the application settings symbol in the upper left corner to define the application, e.g., in the application Percent weighing .
The application settings can only be changed if no measurement is in progress.
The available options may differ depending on the application. Most applications have the following options:
•
•
•
Main configuration
Reports configuration
Statistics
6.1 Main configuration
Navigation: > Activities - Weighing applications >
In this section, the current application individually be defined individually. The available options may vary depending on the application.
More information about the available options can be found in the activities section.
6.2 Reports configuration
Navigation: > Activities - Weighing applications >
In this section, the report content can be configured. By default only the weighing value and the weight unit is published. Any additional information can be configured by the user.
The available options are model and country-specific and may vary depending on the application. The report configuration is valid independently of the report type. As far as possible the same content is published on a
XML, CSV or TXT file or when printed on the strip printer. However, each report type has certain limitations.
Identifications
Define if and how task- and sample identifications are used by the application.
The following options can be set:
Parameter
Identifications
ID 1 … ID 4
Description
Activates or deactivates the identifications.
Activates or deactivates ID 1 … ID 4
ID 4 = for sample identification.
Defines content and behavior of the single IDs.
Activate Input prompt if you want to change the ID value for each result.
Values
ON | OFF*
ON | OFF*
ID Label | ID Value |
Autoincrement | Input prompt
Optional report data
Define additional information to be published in report printouts and report files.
The following options can be set:
Parameter Description
Title information Activates or deactivates the title information.
Defines if the application title, data and time appears on the report.
Values
ON | OFF*
Report title | Date, time
Precision and Analytical Balances Application Settings 53
Balance information
Activates or deactivates the balance information.
Defines which balance information appears on the report.
Task information Activates or deactivates the task information.
Sample information
Footer
Defines which task information appears on the report.
Activates or deactivates the sample information.
Defines which sample information appears on the report.
Activates or deactivates the footer.
Defines which information appears in the footer.
ON | OFF*
Balance type | Balance identification | Software version | Serial number
(SNR) | FACT status |
Last adjustment
ON | OFF*
Application parameters |
MinWeigh parameters
ON | OFF*
Gross/Tare weight |
Additional unit | Leveling status | Tolerance status
ON | OFF*
Date, time | User name |
Signature line | Empty lines
* Factory setting
Advanced options
The following options can be set:
Parameter
Print series
* Factory setting
Description
Activates or deactivates the print series.
Values
ON | OFF*
This option allows you to print a series of weight values using the same header and footer, e.g., header, sample 1, sample 2, sample 3, footer. The series starts with the printout of the first value and can be terminated at any time.
This option is only available for Weighing , Counting , Check weighing , Dynamic weighing , Percent weighing and Factor weighing applications and works only when publishing on a strip printer or generating a TXT file.
6.2.1 Working with IDs
Identifications (IDs) contain descriptive text for measurements, enabling samples to be easily allocated to specific tasks or customers. This feature defines identifications in order to be able comment on measurements, such as company ID, batch ID or sample ID.
Identifications must be defined under the application settings in Reports configuration . Usage and definition of the ID differ depending on the application for which the ID is used.
Identification dialog screen
The Identification dialog screen may vary slightly depending on the application for which the ID is used. The dialog screen always consists of two parts:
•
•
The table with the ID definitions on the upper part of the screen.
The Workflow handling options on the lower part of the dialog screen.
54 Application Settings Precision and Analytical Balances
6.2.2 Defining an ID
The maximum length of an ID is 12 characters.
1 Open an application, e. g., Weighing .
2 Tap the application settings symbol in the upper left corner.
3 Tap Reports configuration .
4 Tap e.g., Header .
5 Tap ID 1 .
ð The input dialog ID 1 appears. The input dialog is inactive.
6 Activate ID 1 with the switch in the title bar.
ð The input dialog ID 1 is activated.
7 Define ID 1 - Label .
8 Tap to confirm.
9 Define ID 1 - Value .
10 Tap to confirm.
ð The screen ID 1 shows the values for ID - Label and ID 1 - Value .
11 Tap to accept the defined values.
Header
ð The screen Header Reports configuration appears the defined ID.
12 Tap to confirm.
ð The screen Header Reports configuration appears.
13 Tap to confirm.
6.2.3 Workflow handling options
The workflow handling options differ depending on the application in which they are used. The following functions are available:
•
•
Autoincrement
Input prompt
Autoincrement
The function Autoincrement specifies that the last part of the ID is incremented with each use of that ID. There are two basic functions depending on how the ID is defined:
• If there is no counter in the ID, the system automatically adds a counter to the ID starting with 1 (e.g., the ID
Process will be Process1 in the next use).
•
•
•
If a counter is part of the ID, the system automatically increments the ID starting at that counter (e.g., the ID
Process 1 will be Process 2 in the next use).
The counter must be set at the end of the ID, otherwise the system doesn't recognize the number as a counter (e.g. in 567Apple the system doesn't recognize 567 as a counter).
If the ID has no counter and a maximum length of 12 characters, the last few characters will be overwritten by the counter.
Input prompt
The Input prompt function can be used for every ID. If Input prompt is activated, the ID will be prompted on the display before it is used. The user can decide whether to use the default value that has been defined with the ID or define an individual value. The value can either be define with the keyboard on the touch screen, by reading in the information with a barcode reader or by attaching an external keyboard to the balance. For more information please refer to Devices .
Precision and Analytical Balances Application Settings 55
6.3 Statistics
Navigation: > Activities - Weighing applications >
The Statistics function generates statistics for a series of results. The Statistics function is not available for the applications Totaling and Formulation .
The Automatic setting is used to automatically transfer the result to the statistics. If using the Manual setting, the key must be pressed to transfer the result.
If there are more than 3 values from a series in the statistics, the result is also displayed as a Gaussian curve.
If statistics are activated, the following options are available during the weighing process:
• Finish and publish
•
•
•
Pause
View result
Discard data
Statistics configuration
The following options can be set:
Parameter Description Values
Acceptance range Defines the acceptable deviation in relation to the average value. 1%...100% (30%*)
Accepting mode Defines if a weight sample is added automatically to the result.
Automatic | Manual*
* Factory setting
Stability criteria: Relationship between readability and deviation
Readability Min. deviation
0.1 mg
0.001 g
0.1 g
1 g
0.01 g
0.1 g
1 g
1 g
1 g
5 g
Defining statistics
1 Open an application, e.g., Weighing .
2 Tap the main configuration symbol of the application, e.g., .
ð The main configuration screen appears.
3 Tap .
ð The screen Statistics appears.
4 Activate the Statistics .
5 Define the available options.
6 Tap to confirm.
Discard value
If a weighing value was incorrect, it can be discarded from the result. Discard is only possible up to the last time the balance was tared.
1 Tap .
ð The dialog screen Confirm discard appears. An overview with all values within the current weighing process is displayed.
2 Tap to discard the last value from the result.
ð The incorrect value has been deleted. The weighing process can be continued.
56 Application Settings Precision and Analytical Balances
Terminate the application
1 Tap .
ð A dialog screen appears.
2 Tap Finish and publish .
ð The results are published according to the publishing configurations and the data are deleted.
ð The application home screen appears.
Discard
All results are deleted.
1 Tap .
ð A dialog screen appears.
2 Tap Discard .
ð All data are deleted.
ð The application home screen appears.
View result
1 Tap .
ð A dialog screen appears.
2 Tap View result .
ð The results are now displayed.
3 Tap to publish the results according to the configurations.
4 Tap to return to the previous screen.
Precision and Analytical Balances Application Settings 57
7 Activities
Navigation:
The Activities section includes the following two sub-sections:
•
•
Activities - Weighing applications
Activities - Adjustments and tests
7.1 Activities - Weighing applications
Navigation: > Activities - Weighing applications
There is not enough space to show all the weighing applications on one screen. Scroll horizontally to reach the other applications on the second page.
Activities - Weighing applications includes the following applications:
• Weighing
, see Weighing and [Performing a simple weighing } Page 24]
•
•
•
•
•
•
•
•
•
F
Counting
Check weighing
, see [Check weighing } Page 65]
Formulation
Totaling
Back-weighing
, see [Back-weighing } Page 73]
Dynamic weighing
, see [Dynamic weighing } Page 76]
Percent weighing
, see [Percent weighing } Page 78]
Density
Factor weighing
, see [Factor weighing } Page 81]
58 Activities Precision and Analytical Balances
7.1.1 Weighing
Navigation
Activities > Activities - Weighing applications > Activities - Weighing applications
The Weighing application allows the user to perform simple weighing operations.
Weighing – Main configuration
The following options can be set:
Parameter
Main unit
Secondary information
Description
Sets the main unit of the weighing process.
The available units are model- and country-specific.
Values g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
ON | OFF*
Readability
Activates or deactivates the secondary information displayed on screen.
Additional unit *
Selects the secondary information displayed on screen. The available units are model- and country-specific.
Current tare
Current tare weight.
Defines the readability (d) of the weighing process.
The available readabilities are model-specific.
MinWeigh
Target and tolerances
Activates or deactivates the MinWeigh function.
Defines the target weight and tolerances.
Target weight
Predefines a target weight. The value can be selected manually or by weighing.
Upper tolerance
Defines the upper tolerance.
Lower tolerance
Defines the lower tolerance.
If the value for Target weight , Upper tolerance or Lower tolerance has been defined, the option title Target and tolerances will be replaced by the defined values.
1d - 0.0001 g* |
2d - 0.0002 g |
5d - 0.0005 g |
10d - 0.001 g |
100d - 0.01 g |
1000d - 0.1 g
ON | OFF*
ON | OFF*
Numerical value
(depending on the balance type)
* Factory setting
Precision and Analytical Balances Activities 59
Setup the weighing application
1 Tap .
ð The screen Weighing – Main configuration appears.
2 Tap e.g. Main unit .
ð The screen Main unit appears.
3 Select the desired unit by tapping it.
4 Tap to confirm.
ð The screen Weighing – Main configuration appears.
5 Tap to confirm the configuration.
60 Activities Precision and Analytical Balances
7.1.2 Counting
Navigation: Activities > Activities - Weighing applications > Counting
The Counting application determines a specific number of pieces based on a predetermined reference piece weight.
There are two counting modes available: Advanced and Standard . With the additional features of Advanced mode, the entire process is more comfortable and more secure due to an automatic workflow. The mode can be changed in the section Counting – Main configuration . Default mode: Advanced .
Legal-for-trade
A fixed minimum reference piece number of 10 and inactive reference weight options are predetermined for approved balances for selected countries.
7.1.2.1 Counting - Main configuration
The following options can be set:
Parameter Description
Counting mode Select the Counting mode .
Reference piece weight
Sets the number and weight of the reference piece(s).
Secondary information
Target and tolerances
Activates or deactivates the secondary information displayed on screen.
Additional unit *
Selects the secondary information displayed on screen. The available units are model- and country-specific.
Current tare
Current tare weight.
Defines the target weight and tolerances.
Target weight
Predefines a target weight. The value can be selected manually or by weighing.
Upper tolerance
Defines the upper tolerance.
Lower tolerance
Defines the lower tolerance.
If the value for Target weight , Upper tolerance or Lower tolerance has been defined, the option title Target and tolerances will be replaced by the defined values.
* Factory setting
Values
Advanced* | Standard
1...999
(10*)
ON* | OFF
ON | OFF*
Numerical value
(depending on the balance type)
Advanced mode options
The following options can be set:
Parameter Description
Reference mode Selects the Reference mode .
Automatic
The next stable weight is automatically accepted as a reference weight according to the defined piece number.
Manual
The reference can be defined manually.
Values
Automatic* | Manual
Precision and Analytical Balances Activities 61
Auto clear reference
Reference optimization
Activates or deactivates the Auto clear reference .
The current value of the option Reference piece weight will be automatically deleted after zeroing or removing all loaded weights from the weighing pan.
Activates or deactivates the Reference optimization .
The current reference will be continuously optimized during operation by accepting additional pieces automatically or manually.
Reference check Activates or deactivates the Reference check .
Accuracy information
Activates or deactivates the Accuracy information .
Counting accuracy can be shown in percent (default mode) or piece(s).
ON | OFF*
ON | OFF*
ON | OFF*
ON | OFF*
* Factory setting
7.1.2.2 Defining reference piece weight in standard mode
To define the Reference piece weight , the Reference piece number and Reference weight must be successively defined. The system will automatically navigate from one option to the other.
Defining reference piece number
The reference piece number must be a number between 1 and 999.
1 Tap .
ð The screen Counting – Main configuration appears.
2 Tap Counting mode .
ð The screen Counting mode appears.
3 Activate Standard .
4 Tap to confirm.
5 Tap Reference piece weight .
ð The dialog screen Reference piece number appears.
6 Tap to delete the value.
7 Enter the number of reference pieces.
8 Tap to confirm.
ð The dialog screen Reference weight appears.
Defining the reference weight
There are two ways to define the reference weight. The reference weight can be defined manually by entering the value or weighing the reference weight.
Defining reference weight manually
1 Tap to delete the value.
2 Enter the new reference weight.
3 Tap to confirm.
ð The reference piece weight has been defined.
4 Tap to confirm.
62 Activities Precision and Analytical Balances
Defining the reference weight by weighing
1 Tap .
ð A dialog screen appears.
2 Place the reference weight on the weighing pan.
3 Tap to confirm.
ð The screen Reference weight appears.
4 Tap to confirm.
ð The screen Counting – Main configuration appears.
5 Tap to confirm the configuration.
7.1.2.3 Defining reference piece weight in advanced mode
To define the Reference piece weight , the Reference piece number and Reference weight can be defined directly via the shortcuts.
Defining the reference piece number
The reference piece number must be a number between 1 and 999.
§ Counting mode Advanced is activated.
1 Tap 1 piece in the work title bar.
ð A dialog screen appears.
2 Tap Piece number .
3 Tap to delete the value.
4 Enter the number of reference pieces.
5 Tap to confirm.
ð The defined value for the option Reference piece number appears in the work title bar.
Defining the reference piece weight manually
1 Tap Piece weight in the work title bar.
ð The screen Reference piece weight appears.
2 Tap to delete the value.
3 Enter the new value.
4 Tap to confirm.
ð The defined value for the option Reference piece weight appears in the work title bar.
Defining reference piece weight by weighing
If no reference weight has been defined yet, the work title bar shows Piece weight Not defined .
1 Place the reference sample weight on the weighing pan.
2 Depending on whether the option Reference mode is set to Automatic (Default) or Manual , the value will be automatically accepted or must be confirmed.
ð The balance returns to the application main screen and shows the defined value for the option
Reference piece weight in the work title bar.
Once the reference weight has been defined in Advanced mode, it appears on the right side of the work title bar.
The reference weight can be changed in Main configuration under Reference piece weight or using the shortcut on the left side of the work title bar.
Counting with reference weight check
The reference weight check ensures that the reference weight is high enough for the resulting counting accuracy to fit the customer process tolerance. Activate a reference weight check and define the process tolerance in percent. The percentage factor range is 0.01 - 30.00%. The higher the factor, the smaller the required minimum reference weight. Factory setting: 2%. The minimum reference weight is equal to d / factor.
Precision and Analytical Balances Activities 63
Example d = 0.1 g
Factor = 20%
Minimum reference weight = 0.1 g / 20% = 0.5 g
If the reference calculation is activated manually or automatically, the minimum reference weight is checked to ensure the desired accuracy. If it is not sufficient, the user is prompted to add the number of additional parts needed. The number of additional parts needed is reduced to zero, after which the user adds additional parts. At zero, the reference calculation is activated automatically. If too many parts are added, the user is prompted to remove the number of parts until zero is reached.
64 Activities Precision and Analytical Balances
7.1.3 Check weighing
Navigation: Activities > Activities - Weighing applications > Check weighing
The Check weighing application allows the user to check the deviation of a sample weight within a tolerance limit against a reference target weight. The target weight can be determined manually or by weighing; the tolerance limit must be defined manually.
7.1.3.1 Check Weighing - Main configurations
Legal-for-trade
For approved balances, this menu item has fixed settings and cannot be changed.
The following options can be set:
Parameter
Target and tolerances
Tolerance threshold
Within tolerance beeps
Main unit
Readability
Description
Defines the target weight and tolerances.
Target weight
Predefines a target weight. The value can be selected manually or by weighing.
Upper tolerance
Defines the upper tolerance.
Lower tolerance
Defines the lower tolerance.
If the value for Target weight , Upper tolerance or Lower tolerance has been defined, the option title Target and tolerances will be replaced by the defined values.
Defines the tolerance threshold. Values below the defined threshold are not checked.
Activates or deactivates the acoustic signal.
Gives an acoustic signal when the result is within tolerance.
Sets the main unit of the weighing process.
The available units are model- and country-specific.
Defines the readability (d) of the weighing process.
The available readabilities are model-specific.
Values
Target weight |
Upper tolerance |
Lower tolerance
Numerical value
(depending on the balance type)
1%…100% (1%*)
ON | OFF* g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
1d - 0.0001 g* |
2d - 0.0002 g |
5d - 0.0005 g |
10d - 0.001 g |
100d - 0.01 g |
1000d - 0.1 g
* Factory setting
Precision and Analytical Balances Activities 65
7.1.3.2 Before performing check weighing
Before performing a check weighing operation, the following options can be set:
• Target weight
•
•
•
Upper tolerance limit
Lower tolerance limit
Tolerance threshold
Defining the target weight manually by entering the nominal weight
1 Tap .
ð The screen Check weighing - Main configuration appears.
2 Tap Target and tolerances .
ð The dialog screen Target weight in g appears.
3 Tap to delete the value.
4 Enter the value of the target weight.
5 Tap to confirm.
ð The screen Check weighing - Main configuration appears.
6 Tap to confirm, and return to the application screen.
Defining the target weight by weighing
1 Tap .
ð The screen Check weighing - Main configuration appears.
2 Tap Target and tolerances .
ð The dialog screen Target weight in g appears.
3 Tap .
ð The dialog screen Target weight in g appears.
4 Place the reference weight on the weighing pan.
5 Tap to confirm.
ð The dialog screen Target weight in g appears.
6 Tap to confirm.
ð The screen Check weighing - Main configuration appears.
7 Tap to confirm, and return to the application screen.
Defining the lower and upper limits manually by entering a percentage value or weight
1 Tap .
ð The screen Check weighing - Main configuration appears.
2 Tap Target and tolerances .
ð The dialog screen Target weight in g appears.
3 Tap Upper tolerance limit or tap Lower tolerance limit .
ð The dialog screen Upper tolerance in g or Lower tolerance in g appears.
4 Activate the option using the switch at the upper right.
5 Tap to delete the value.
6 Enter the tolerance limit.
7 Tap to confirm.
ð The screen Check weighing - Main configuration appears.
8 Tap to confirm, and return to the application screen.
66 Activities Precision and Analytical Balances
Defining tolerance threshold
With the option Tolerance threshold a value limit can be set using the Tolerance threshold option. If the value of the check weight is below the defined threshold, it will not be checked.
1 Tap .
ð The screen Check weighing - Main configuration appears.
2 Tap Tolerance threshold .
ð The dialog screen Tolerance threshold in % appears.
3 Activate the option using the switch at the upper right.
4 Tap to delete the value.
5 Enter the value for the Tolerance threshold .
6 Tap to confirm.
ð The screen Check weighing - Main configuration appears.
7 Tap to confirm, and return to the application screen.
The Tolerance threshold option always refers to the lower tolerance limit.
7.1.3.3 Performing check weighing
After defining the target weight and tolerance limits, the Check weighing application can be performed. The color weighing-in aid bar at the top bar indicates whether the weight sample is within the defined tolerances.
Example: the defined target weight is 100.0000 g and the tolerance limit is ±2.5%. The sample weight is
97.0000 g.
− Place the sample weight on the weighing pan.
ð The weight is stable and the unstability symbol disappears.
ð The value is out of tolerance, and the weighing-in aid bar and weighing value field are red.
Example: the defined target weight is still 100.0000 g and the tolerance limit is ±2.5%. The sample weight is
99.0000 g.
Value out of tolerance
− Place the sample weight on the weighing pan.
ð The weight is stable and the unstability symbol disappears.
ð The value is within the tolerance limit, and the weighing-in aid bar and weighing value field are green.
If the weight is below a defined tolerance threshold, the background color of the screen does not change.
Value in tolerance
Precision and Analytical Balances Activities 67
7.1.4 Formulation
Navigation: Activities > Activities - Weighing applications > Formulation
The Formulation application allows the user to:
• Weigh-in (add and store) up to 999 individual component weights without tare container and display the total.
•
•
•
Tare/pre-tare and store up to 799 container weights and display the total.
If tare containers need to be stored, the maximum number of permitted tares is 200.
Fill up the sum of all component net weight values by adding a further component to a higher value.
Formulation – Main configuration
The following options can be set:
Parameter
Main unit
Readability
Description
Sets the main unit of the weighing process.
The available units are model- and country-specific.
Defines the readability (d) of the weighing process.
The available readabilities are model-specific.
Values g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
1d - 0.0001 g* |
2d - 0.0002 g |
5d - 0.0005 g |
10d - 0.001 g |
100d - 0.01 g |
1000d - 0.1 g
* Factory setting
Performing formulation
1 Press to zero the balance.
2 If using a container: place container on the weighing pan and press
ð The status information field displays Net .
3 Place the first component weight.
to tare the balance.
ð The weighing value field displays the value of the first component weight.
4 Tap to add the first component weight.
5 Place the second component weight.
ð The weighing value field displays the value of the second component weight.
6 Tap to add the second component weight.
7 Continue adding components until all components are weighed.
Defining fill up function
Fill up sample allows an additional component weight to be added to the total weight of all components in order to reach a desired target weight.
§ The weighing value field displays the total net weight.
1 Tap .
ð A dialog screen appears.
2 Place the fill up sample.
ð The weighing value field displays the total weight.
3 Tap to confirm.
4 Tap to exit the application or to view the result.
68 Activities Precision and Analytical Balances
The following options are available during the weighing process:
• Finish and publish
•
•
•
Pause
View result
Discard
Discard value
If a weighing value was incorrect, it can be discarded from the result. Discard is only possible up to the last time the balance was tared.
1 Tap .
ð The dialog screen Confirm discard appears. An overview with all values within the current weighing process is displayed.
2 Tap to discard the last value from the result.
ð The incorrect value has been deleted. The weighing process can be continued.
Terminate the application
1 Tap .
ð A dialog screen appears.
2 Tap Finish and publish .
ð The results are published according to the publishing configurations and the data are deleted.
ð The application home screen appears.
Pause the application
1 Tap .
ð A dialog screen appears.
2 Tap Pause .
ð The application is paused and another application can be used in the meantime.
ð The application home screen appears.
3 Open the application again.
4 Tap .
ð The process can be continued.
View result
1 Tap .
ð A dialog screen appears.
2 Tap View result .
ð The results are now displayed.
3 Tap to publish the results according to the configurations.
4 Tap to return to the previous screen.
Discard
All results are deleted.
1 Tap .
ð A dialog screen appears.
2 Tap Discard .
ð All data are deleted.
ð The application home screen appears.
Precision and Analytical Balances Activities 69
7.1.5 Totaling
Navigation: Activities > Activities - Weighing applications > Totaling
The Totaling application allows the user to weigh different samples, add their weight values and totalize them.
The application allows the user to:
• Weigh-in (add and store) up to 999 individual component weights without tare container and display the total.
Totaling – Main configuration
The following options can be set:
Parameter Description Values
Accepting mode Defines if a weight sample is added automatically to the result.
Automatic | Manual*
Main unit Sets the main unit of the weighing process.
The available units are model- and country-specific.
g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
Readability Defines the readability (d) of the weighing process.
The available readabilities are model-specific.
1d - 0.0001 g* |
2d - 0.0002 g |
5d - 0.0005 g |
10d - 0.001 g |
100d - 0.01 g |
1000d - 0.1 g
* Factory setting
Stability criteria: Relationship between readability and deviation
Readability
0.1 mg
Min. deviation
0.1 g
0.001 g
0.01 g
0.1 g
1 g
1 g
1 g
1 g
5 g
70 Activities Precision and Analytical Balances
Performing totaling
1 Press to zero the balance.
2 If using a container, place container on the weighing pan and press
3 Place the first sample on the weighing pan.
to tare the balance.
4 Wait until the unstability symbol disappears.
ð When the balance is stable, the weighing value becomes dark blue.
5 Tap to accept the weight and start the procedure.
6 Place the next sample.
7 Tap to accept the second sample weight.
ð The work title bar shows the number of samples (2 samples) and the total weight of the samples, e.g.,
= 30.0000 g.
The following options are available during the weighing process:
• Finish and publish
•
•
•
Pause
View result
Discard
Discard value
If a weighing value was incorrect, it can be discarded from the result. Discard is only possible up to the last time the balance was tared.
1 Tap .
ð The dialog screen Confirm discard appears. An overview with all values within the current weighing process is displayed.
2 Tap to discard the last value from the result.
ð The incorrect value has been deleted. The weighing process can be continued.
Terminate the application
1 Tap .
ð A dialog screen appears.
2 Tap Finish and publish .
ð The results are published according to the publishing configurations and the data are deleted.
ð The application home screen appears.
Pause the application
1 Tap .
ð A dialog screen appears.
2 Tap Pause .
ð The application is paused and another application can be used in the meantime.
ð The application home screen appears.
3 Open the application again.
4 Tap .
ð The process can be continued.
View result
1 Tap .
ð A dialog screen appears.
2 Tap View result .
ð The results are now displayed.
3 Tap to publish the results according to the configurations.
4 Tap to return to the previous screen.
Precision and Analytical Balances Activities 71
Discard
All results are deleted.
1 Tap .
ð A dialog screen appears.
2 Tap Discard .
ð All data are deleted.
ð The application home screen appears.
72 Activities Precision and Analytical Balances
7.1.6 Back-weighing
Navigation: Activities > Activities - Weighing applications > Back-weighing
The balance displays and prints the automatically calculated difference of 2 measured weights. Automatic
(default) and manual modes are possible. The use of a tare container can be activated (default) or deactivated.
As result, the tare, initial weight, final weight and difference can be displayed and printed. The difference can be displayed and printed as absolute values (in main units), percentage (%), percentage (Abs. %), Atro AM, or
Atro AD.
If there are no function statistics, the following options are available during the weighing process:
•
•
•
•
Finish and publish
Pause
View result
Discard
Back-weighing – Main configuration
The following options can be set:
Parameter Description
Accepting mode Defines if a weight sample is added automatically to the result.
The next stable weight of at least 10 * readability is accepted following minimal deviation; see relationship between readability and deviation in the table below.
Use tare container Activates or deactivates the use of a tare vessel.
Result value as... Select the result view for the calculated difference.
Percentage = Reports the difference between back-weighing and initial weighing as a percentage of the initial weight.
Absolute percentage = Reports the back-weighing as a percentage of the initial weight.
Moisture content = Reports the moisture content of the sample as a percentage of the dry weight.
Dry content = Reports the wet weight of the sample as a percentage of the dry weight.
Result decimals Defines the number of decimal places of the percentage (option only available if % result is activated).
Show difference…
Main unit
Displays the calculated difference in work area and result view.
Signed (Default) = Displays the value via algebraic sign.
Unsigned = Displays the absolute value.
Sets the main unit of the weighing process.
The available units are model- and country-specific.
Values
Automatic* | Manual
ON* | OFF
Weight (Default)* |
Percentage | Absolute percentage | Moisture content | Dry content
1 | 2 | 3* | 4 | 5
Signed (Default)* |
Unsigned
Readability Defines the readability (d) of the weighing process.
The available readabilities are model-specific.
g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
1d - 0.0001 g* |
2d - 0.0002 g |
5d - 0.0005 g |
10d - 0.001 g |
100d - 0.01 g |
1000d - 0.1 g
* Factory setting
Precision and Analytical Balances Activities 73
74 Activities
Stability criteria: Relationship between readability and deviation
Readability Min. deviation
0.1 mg
0.001 g
0.1 g
1 g
0.01 g
0.1 g
1 g
1 g
1 g
5 g
Setting up the back-weighing application
1 Tap .
ð The screen Back-weighing – Main configuration appears.
2 Tap Accepting mode .
ð The screen Accepting mode appears.
3 Select Automatic (Default) or Manual and confirm with .
4 Activate the option Use tare container if required.
5 Tap Result value as...
.
ð The screen Result value as...
appears.
6 Select the values for the result view, e.g., Percentage (%) and the print-out.
7 Tap to confirm.
8 Tap Result decimals after percent value.
ð The screen Result decimals appears.
9 Select the number of decimals for the difference in % and confirm with .
10 Tap to return to the application screen.
Performing back-weighing with automatic accept mode and using a tare container
§ Use tare container is activated.
§ Accepting mode Automatic is selected.
1 Place a container on the weighing pan.
ð The tare weight appears in the work title bar.
2 Place the initial sample in the container.
ð The initial weight appears in the value bar.
3 Remove the container with the sample.
4 Place the container with the treated sample on the weighing pan.
ð The final weight appears in the value bar.
5 Remove the container with the sample.
ð The back-weighing result will be displayed and can be published according to the publishing configurations.
6 Tap to return to the application screen.
Precision and Analytical Balances
Performing back-weighing with manual accept mode without using a tare container
§ Use tare container is deactivated.
§ Accepting mode Manual is selected.
1 Place the initial sample on the weighing pan.
ð The initial weight appears in the value bar.
2 Tap to confirm.
3 Remove the sample to continue.
4 Place the treated sample on the weighing pan.
ð The final weight appears in the value bar.
5 Tap to confirm.
6 Push checkmark to access the result report.
ð The back-weighing result will be displayed and can be published according to the publishing configurations.
7 Tap to return to the application screen.
Precision and Analytical Balances Activities 75
7.1.7 Dynamic weighing
Navigation: Activities > Activities - Weighing applications > Dynamic weighing
The Dynamic weighing application determines the weights of unstable samples or when the weighing process is being executed under unstable ambient conditions. The balance calculates the weight as an average of a number of weighing operations over a defined time.
The displayed measuring time on the application home screen can be used as shortcut to define the measuring time.
Dynamic weighing – Main configuration
The following options can be set:
Parameter Description
Measuring time Defines the measuring time in seconds.
Start mode
Main unit
Readability
Secondary information
Defines the Start mode .
Sets the main unit of the weighing process.
The available units are model- and country-specific.
Defines the readability (d) of the weighing process.
The available readabilities are model-specific.
Activates or deactivates the secondary information displayed on screen.
Additional unit *
Selects the secondary information displayed on screen. The available units are model- and country-specific.
Current tare
Current tare weight.
Values
3…120
(3 seconds*)
Automatic* | Automatic
With time delay |
Manual g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
1d - 0.0001 g* |
2d - 0.0002 g |
5d - 0.0005 g |
10d - 0.001 g |
100d - 0.01 g |
1000d - 0.1 g
ON | OFF*
* Factory setting
Defining measuring time
1 Tap .
ð The screen Dynamic weighing – Main configuration appears.
2 Tap Measuring time .
ð The dialog screen Measuring time in seconds appears.
3 Tap to delete the value.
4 Enter a value between 3 and 120 seconds.
5 Tap to confirm.
ð The screen Dynamic weighing – Main configuration appears.
6 Tap to confirm.
76 Activities Precision and Analytical Balances
Defining start mode
1 Tap .
2 Tap Start mode .
3 Select Automatic *, Automatic With time delay or Manual .
4 Tap to confirm.
5 Tap to return to the application screen.
Performing dynamic weighing
When the measuring time and the start mode have been defined, the dynamic weighing process can be started.
The weighing process will automatically be aborted when an overload or underload is detected.
1 Press to zero the balance.
2 If using a container: place container on the weighing pan and press
"Weighing option" menu [Auto tare } Page 33].
to tare the balance or use the
3 Place sample weight.
ð If Start mode is set to Automatic , the weighing process starts automatically with relative stability.
ð If Start mode is set to Automatic With time delay , the weighing process starts automatically after 3 seconds.
ð It Start mode is set to Manual , tap to start the weighing process.
ð The weighing process starts. The defined measuring time in the work title bar is counting down.
ð The result is displayed in a blue weighing value field.
Precision and Analytical Balances Activities 77
7.1.8 Percent weighing
Navigation: Activities > Activities - Weighing applications > Percent weighing
Percent weighing allows a sample weight to be checked as a percentage of a reference target weight.
Percent weighing – Main configuration
The following options can be set:
Parameter Description
Reference weight Defines the reference weight manually or by weighing.
Secondary information
Values
Available range is model-specific.
ON* | OFF Activates or deactivates the secondary information displayed on screen.
Additional unit *
Selects the secondary information displayed on screen. The available units are model- and country-specific.
Current tare
Current tare weight.
* Factory setting
Defining the reference weight
There are two ways to define the reference weight. The reference weight can be defined manually by entering the value or weighing the reference weight.
Defining the reference weight by weighing
1 Tap .
ð The screen Percent weighing – Main configuration appears.
2 Tap Reference weight .
ð The screen Reference weight appears.
3 Tap .
4 Press to zero the balance.
5 Place the reference weight on the weighing pan.
6 Tap to confirm.
ð The screen Reference weight appears.
7 Tap to confirm.
8 Tap to return to the application screen.
Defining the reference weight manually
1 Tap .
ð The screen Percent weighing – Main configuration appears.
2 Tap Reference weight .
ð The screen Reference weight appears.
3 Tap to delete the value.
4 Enter the value of the reference weight and confirm with .
5 Tap to return to the application screen.
78 Activities Precision and Analytical Balances
7.1.9 Density
Navigation: Activities > Activities - Weighing applications > Density
Density allows the density of solid bodies and liquids to be determined.
The density is determined via the Archimedes principle, according to which a body immersed in a fluid undergoes an apparent loss in weight which is equal to the weight of the fluid it displaces.
To determine the density of solid bodies, we recommend using the optional density kit, which contains all attachments and aids needed for convenient and precise density determination. To determine the density of liquids, a sinker is also required. This can be obtained from your METTLER TOLEDO dealer.
Density – Main configuration
The following options can be set:
Parameter
Method
Description
Depending if the weight type is solid or liquid.
Auxiliary liquid Select the auxiliary liquid.
Sinker volume This option is only available when Liquid is active.
Result decimals Defines the number of decimal places density in g/cm 3 .
Main unit Sets the main unit of the weighing process.
The available units are model- and country-specific.
Values
Solid* | Liquid
H2O* | Ethanol | Free...
(0.1…500.0 cm 3 )
1 | 2 | 3* | 4
(depending on the balance type) g* | kg | mg | μg | ct | N
| lb | oz | ozt | GN | dwt | mom | msg | tlh | tls | tlt
| tcl | tola | baht | lb:oz
* Factory setting
Determination of density of solids
1 Tap .
ð The screen Density – Main configuration appears.
2 Tap Method .
ð When Solid has been activated (default value), the Auxiliary liquid appears in the list.
3 Tap Auxiliary liquid .
ð The screen Auxiliary liquid appears.
4 Define the Auxiliary liquid that is used. Select between H2O for distilled water, Ethanol or Free...
for a freely definable auxiliary liquid.
5 Tap to confirm.
ð The selected Auxiliary liquid defines the next steps:
ð The dialog screen Temperature in °C appears.
6 Enter the Temperature in °C for Ethanol and H2O .
7 Auxiliary liquid name and Density in g/cm3 must be defined for the option Free...
.
8 Tap to confirm.
ð The screen Density – Main configuration appears.
9 Tap to return to the application screen.
ð The balance is prepared for determining the density of solids.
Precision and Analytical Balances Activities 79
80 Activities
Determining the density for solids
§ The balance has been configured to determine the density of solids.
1 Tap to start the process.
ð The dialog screen Sample weight in air appears.
2 Place the solid mass on the weighing pan.
3 Tap to confirm.
ð The dialog screen Sample weight in liquid appears.
4 Immerse solid in the liquid.
5 Tap to confirm.
ð The results are now displayed.
6 Tap to publish the results according to the configurations.
7 Tap to return to the application screen.
Determination of density for liquids
1 Tap .
ð The screen Density – Main configuration appears.
2 Tap Method .
3 Tap Liquid .
4 Tap to confirm.
ð When Liquid has been activated, Sinker volume appears in the list.
5 Tap Sinker volume .
ð The dialog screen Sinker volume in cm³ appears.
6 Tap to delete the value.
7 Enter the volume of the sinker.
8 Tap to confirm.
ð The screen Density – Main configuration appears.
9 Tap to return to the application screen.
Determining the density of liquids
§ The balance has been configured to determine the density of liquids.
1 Tap to start the process.
ð The dialog screen Sinker in air appears.
2 Place the sinker on the weighing pan.
3 Tap to confirm.
ð The dialog screen Sinker in liquid appears.
4 Immerse the sinker.
5 Tap to confirm.
ð The results are now displayed.
6 Tap to publish the results according to the configurations.
7 Tap to return to the application screen.
Precision and Analytical Balances
7.1.10 Factor weighing
Navigation: Activities > Activities - Weighing applications >
F
Factor weighing
The application Factor weighing multiplies or divides a pre-defined factor by the measured weight value (in grams) and calculates it to a predefined number of decimal places.
The permitted range for the steps depends on the pre-defined factor and the resolution of the balance.
Factor weighing – Main configuration
The following options can be set:
Parameter
Factor, step
Secondary information
Target and tolerances
Description
Defines the factor and the step.
Activates or deactivates the secondary information displayed on screen.
Additional unit *
Selects the secondary information displayed on screen. The available units are model- and country-specific.
Current tare
Current tare weight.
Defines the target weight and tolerances.
Target weight
Predefines a target weight. The value can be selected manually or by weighing.
Upper tolerance
Defines the upper tolerance.
Lower tolerance
Defines the lower tolerance.
If the value for Target weight , Upper tolerance or Lower tolerance has been defined, the option title Target and tolerances will be replaced by the defined values.
Values
Factor | Step
ON* | OFF
ON | OFF*
Numerical value
(depending on the balance type)
* Factory setting
Defining factor and step
1 Tap .
ð The screen Factor weighing – Main configuration appears.
2 Tap Factor, step .
ð The dialog screen Factor – Multiplication appears.
3 Tap to delete the value.
4 Define Factor .
5 Tap to change the operation from Multiplication to Division or vice versa.
6 Tap .
7 Tap .
ð The dialog screen Step appears.
8 Define Step .
9 Tap to confirm.
ð The screen Factor weighing – Main configuration appears.
10 Tap to return to the application screen.
Precision and Analytical Balances Activities 81
7.2 Activities - Adjustments and tests
Navigation: >
Activities - Adjustments and tests consists of the following elements:
• FACT
, see [Fully automatic adjustment (FACT) } Page 82]
•
•
Internal Adjustment
, see [Internal adjustment } Page 82]
External Adjustment
, see [External adjustment } Page 83]
•
•
•
•
•
Fine adjustment
, see [Fine adjustment (model-dependent) } Page 83]
Adjust touch
, see [Touch screen adjustment } Page 84]
Adjust level
, see [Level center adjustment } Page 84]
Routine test
, see [Routine test } Page 84]
Repeatability test
, see [Repeatability test } Page 85]
7.2.1 Fully automatic adjustment (FACT)
Navigation: Activities > Activities - Adjustments and tests > FACT
FACT is activated as a default value. If the FACT function is not activated, all functionalities, such as temperature and time, are inactive.
FACT means that the balance adjusts itself based on the following criteria:
•
• if the conditions change (temperature difference min. 2 °C), which may lead to a noticeable deviation in the measurement.
on a predefined date and time programmed by the user.
Defining FACT
The date and time of FACT can be defined as follows:
1 Tap FACT .
2 Activate Fully automatic adjustment (FACT) .
ð The dialog screen Fully automatic adjustment (FACT) appears.
3 Select the time (hours : minutes) using the pick buttons.
4 Tap to confirm.
ð The time underneath FACT has been updated and displays the time of the daily adjustment.
5 Tap to return to the application screen.
When setting the time, hold the pick button to scroll faster.
If a predefined criterion is due, the flashing FACT status icon appears in the display. The balance indicates that it wishes to perform a FACT adjustment.
1 Unload the balance.
2 Do not select any key.
ð Adjustment starts automatically.
ð The status icon extinguishes after successful adjustment.
7.2.2 Internal adjustment
Navigation: Activities > Activities - Adjustments and tests > Adjust internal
If the balance is configured and connected to a printer, the results of the adjustment process will be printed.
Perform an internal adjustment manually
1 Unload the balance.
2 Tap Adjust internal .
82 Activities Precision and Analytical Balances
3 Tap .
ð The internal adjustment procedure will start. The screen displays Adjustment ongoing… .
ð When the internal adjustment procedure is successfully completed, the results of the internal adjustment will appear.
4 Tap to confirm.
ð The screen Activities - Adjustments and tests appears.
5 Tap to return to the application screen.
7.2.3 External adjustment
Navigation: Activities > Activities - Adjustments and tests > Adjust external
Legal-for-trade
Because of certification legislation, approved balances cannot be adjusted with an external weight (depending on the certification legislation of the selected country).
1 Tap Adjust external .
ð The dialog screen Adjustment weight appears.
2 Tap to define the adjustment weight according to the weight certificate.
3 Tap to delete the value.
4 Enter the new value and confirm with .
5 Prepare the adjustment weight and tap to start the adjustment process.
6 Load the adjustment weight in the center of the weighing pan.
7 Unload the adjustment weight from the weighing pan.
ð When the external adjustment procedure is successfully completed, the result will appear.
8 Tap to confirm.
ð The screen Activities - Adjustments and tests appears.
9 Tap to return to the application screen.
7.2.4 Fine adjustment (model-dependent)
Navigation: > Activities - Adjustments and tests > Adjust fine
The value of the internal adjustment weight can be individually adjusted in a very small range via the Fine adjustment function.
•
•
Use only certificated weights.
Make sure that the environmental conditions are correct.
•
•
•
The balance must be leveled.
Balance and test weights must observe the operating temperature.
For fine adjustments, we recommend contacting a balance expert or a METTLER TOLEDO representative.
Legal-for-trade
Approved models are not able to be adjusted with this function.
Performing fine adjustment
§ Adjusting weight is prepared.
1 Tap Adjust fine .
ð The dialog screen Reference weight appears.
2 Tap to delete the value.
3 Enter the weight value according to the certificate.
4 Tap to confirm.
5 Tap to start the process.
6 Place the adjustment weight in the center of the weighing pan.
Precision and Analytical Balances Activities 83
7 Remove the adjustment weight.
ð The result will appear once the fine adjustment procedure is successfully completed.
8 Tap to confirm.
ð The screen Activities - Adjustments and tests appears.
9 Tap to return to the application screen.
Tap to reset the defined reference weight back to the default value.
7.2.5 Touch screen adjustment
Navigation: Activities > Activities - Adjustments and tests > Adjust touch
If the balance does not respond correctly when a certain area of the display is tapped, the touch screen can be adjusted with Adjust touch .
1 Tap Adjust touch .
ð Touch screen ongoing. Please wait until the screen Done appears.
2 Tap to confirm.
ð The screen Activities - Adjustments and tests appears.
3 Tap to return to the application screen.
7.2.6 Level center adjustment
Navigation: Activities > Activities - Adjustments and tests > Adjust level
The adjustment sets the exact center position of the level indicator and corrects any possible shifts.
Note
Run the level indicator center adjustment only when the electronic in/out level notification (status icon) and visual bubble position do not match.
1 Tap Adjust level .
ð The screen Level center adjustment appears.
2 Tap to confirm.
ð The screen Level center adjustment appears.
3 Follow the instructions and confirm with .
ð The screen Done Level center adjustment appears.
4 Tap to confirm.
5 Tap to return to the application screen.
7.2.7 Routine test
Navigation: Activities > Activities - Adjustments and tests > Routine test
The Routine test function allows the sensitivity of the balance to be set for periodic tests.
The set values appear at the top of the screen in the weighing information bar. The bar works as a shortcut.
Routine test – Main configuration
The following options can be set:
Parameter Description
Test weight in g Defines the test weight.
± Control limit in g
Defines the control limit.
Values
Numerical value
(depending on the balance type)
Numerical value
(depending on the balance model)
84 Activities Precision and Analytical Balances
± Warning limit in g
Activates or deactivates the warning limit.
ON* numerical value
(depending on the balance model) | OFF
ON | OFF* Use tare container Activates or deactivates the use of a tare vessel.
* Factory setting
Setting the test weight, control limits and warning limits
1 Tap .
ð The screen Routine test – Main configuration appears.
2 Tap Test Weight .
ð The dialog screen Test weight in g appears.
3 Tap to delete the value.
4 Enter the new value.
5 Tap .
ð The dialog screen ± Control limit in g appears.
6 Tap to delete the value.
7 Enter the new value.
8 Tap .
ð The dialog screen ± Warning limit in g appears.
9 Tap to delete the value.
10 Enter the new value and confirm with .
11 Activate or deactivate the option Use tare container if required.
12 Tap to confirm.
13 Tap to return to the application screen.
Performing a routine test
§ The routine test options Test weight in g , ± Control limit in g and ± Warning limit in g are defined.
§ The test weight is prepared.
1 Tap to start the process.
2 Place the test weight in the center of the weighing pan.
ð During the test, the screen shows Waiting for stable weight...
.
ð When the test is finished, the screen shows Please unload weight .
3 Remove the test weight from the weighing pan.
ð When the routine test is successfully completed, the result appears.
4 Tap to confirm.
5 Tap to return to the application screen.
7.2.8 Repeatability test
Navigation: Activities > Activities - Adjustments and tests > Repeatability test
A specific number of internal weight tests can be defined via the Repeatability test function.
The set number of tests appears at the top of the screen in the weighing information bar. The bar works as a shortcut.
Setting the number of repetitions
1 Tap .
ð The dialog screen Repeatability test - Repetitions appears.
2 Tap to delete the value.
3 Enter the number of repetitions. The number must be between 5 and 100.
4 Tap to confirm the number of repetitions.
Precision and Analytical Balances Activities 85
5 Tap to start the process.
ð The balance executes the defined number of tests. The message Test ongoing, please wait… appears on the display during the process. The process can be aborted by tapping .
ð After the test is finished, an overview with the test results appears on the screen.
6 Tap to confirm.
7 Tap to return to the application screen.
86 Activities Precision and Analytical Balances
8 Communication with Peripheral Devices
This section lists some typical examples of where the balance can communicate with peripheral devices and networks. In most cases, it is possible to do so via USB or RS232C. The chapters are primarily restricted to a
USB setup.
8.1 USB - interface and installation
Before connecting the balance via the USB device interface to a PC and using either the HOST or PC-Direct function, the appropriate METTLER TOLEDO USB driver has to be assigned to the PC first. The USB driver can be found on www.mt.com/labweighing-software-download . If you have any questions, please contact a METTLER
TOLEDO representative.
Note
If you connect the balance via USB to the PC before installing the METTLER TOLEDO USB driver, Windows will automatically install the wrong driver.
Requirements
•
•
•
•
Balance with USB device interface
PC with one of the following Microsoft Windows ®
Win 10
32-bit/64-bit operating systems: Win 7 (SP1), Win 8 or
Administrator rights for installing software
USB connection cable to link PC to balance
Download USB driver
1 Connect to the internet.
2 Go to the site www.mt.com/labweighing-software-download .
3 Click Download Driver in section USB Driver for Advanced and Standard level laboratory balances .
ð A pop-up window with interactions appears.
4 Click, e.g., Open .
ð The extract screen appears.
5 Extract the file MT_Lab_Advanced-Standard_USB_Driver_Installer_v1.xx.zip
to your specified location.
6 Right-click on the downloaded installation program MT_Lab_Advanced-
Standard_USB_Driver_Installer_v1.xx.exe
and select Run as Administrator .
7 If a safety warning appears, confirm windows to perform the installation.
8 Click Next and follow the installer's instructions.
Installing the balance
1 Switch the balance off .
2 Connect the balance to the preferred USB port on the PC.
3 Switch the balance on .
Precision and Analytical Balances Communication with Peripheral Devices 87
8.2 Send weight value via USB or RS232C to a PC using PC-Direct
The numerical value displayed at the balance can be transferred to the cursor position in windows applications, e.g., Excel, Word in the same way as typing with the keyboard.
The data is transferred via USB or via the serial RS232C interface.
The weight value without the unit will be transferred.
Requirements
• PC with one of the following Microsoft Windows ®
Win 10
32-bit/64-bit operating systems: Win 7 (SP1), Win 8 or
•
•
•
•
Serial interface RS232C or USB
Administrator rights for installing the SerialPortToKeyboard software (if data transfer is via RS232C)
Windows application, e.g., Excel
Connection between balance and PC via cable RS232C or USB
8.2.1 PC-Direct via USB
The balance can send data (as a keyboard) to the PC used for PC applications, e.g., Excel. The balance sends the weight value without the unit to the PC.
Use the USB connection cable to connect the balance with the PC. Connect the USB cable to the USB device
(type B) on the balance.
§ The balance must be disconnected from the PC.
1 Tap .
2 Tap General configuration .
3 Tap Devices .
4 The screen Devices appears.
5 Tap .
ð The screen Device/service type appears.
6 Select PC-Direct and confirm with .
ð The screen Connection type appears.
7 Tap USB Device and confirm with .
ð The screen PC-Direct – Document on PC appears.
8 If necessary, change the other settings, e.g., End of line and confirm with .
9 Tap to confirm.
10 The screen Devices appears.
11 Tap to return to the previous screen.
12 Tap Publishing .
ð The screen Publishing appears.
13 Tap Send data .
ð The screen Send data appears.
14 Select the transmission mode for single values and result values, e.g., Manual, stable and confirm with .
15 Tap to return to the previous screen.
16 Connect the balance to the PC.
17 Place the sample on the weighing pan.
18 Press , the next stable weight will be sent to the cursor position of your application.
88 Communication with Peripheral Devices Precision and Analytical Balances
8.2.2 PC-Direct via RS232C
8.2.2.1 Installing SerialPortToKeyboard software
The operation of PC-Direct via serial port RS232C requires the installation of SerialPortToKeyboard on your host computer. The file SerialPortToKeyboard can be found on www.mt.com/labweighing-software-download .
If you have any questions, please contact a METTLER TOLEDO representative.
Download SerialPortToKeyboard
1 Connect to the internet.
2 Go to the site www.mt.com/labweighing-software-download .
3 Click Download Software and Instructions in section SerialPortToKeyboard software for Advanced and
Standard level laboratory balances .
ð A pop-up window with interactions appears.
4 Click, e.g., Open .
ð The extract screen appears.
5 Extract the file SerialPortToKeyboard_V_x.xx_installer_and_instructions.zip
to your specified location.
6 Right-click on the downloaded installation program SerialPortToKeyboard_V_x.xx.exe
and select Run as
Administrator .
7 If a safety warning appears, confirm windows to perform the installation.
8 Click Next and follow the installer's instructions.
Checking operation
1 Start SerialPortToKeyboard (RS232C)
2 Start Excel (or another application) at the PC.
3 Activate a cell in Excel.
Settings at the PC
Settings for SerialPortToKeyboard
1 Select the serial port COM for the connection with the balance.
2 Set the Baud Rate to 9600 .
3 Activate Connect .
• Closing the window terminates the session.
➂
➀
➁
According to your selected End of line option, the displayed values will appear, e.g., in the column one after the other one in the different rows.
8.2.2.2 Settings on the balance
§ The balance is connected to the PC via an RS232 cable.
1 Tap .
2 Tap General configuration .
3 Tap Devices .
4 The screen Devices appears.
Precision and Analytical Balances Communication with Peripheral Devices 89
5 Tap .
ð The screen Device/service type appears.
6 Select PC-Direct and confirm with .
ð The screen Connection type appears.
7 Tap RS232 and confirm with .
ð The screen PC-Direct – Document on PC appears.
8 If necessary, change the other settings, e.g., End of line and confirm with .
9 Tap to confirm.
10 Tap to return to the previous screen.
11 Tap Publishing .
ð The screen Publishing appears.
12 Tap PC-Direct (RS232) .
ð The screen Send data appears.
13 Select the transmission mode for single values and result values, e.g., Automatic, stable and confirm with
.
14 Tap to confirm.
ð The screen Publishing appears.
15 Tap to return to the previous screen.
16 Place the sample on the weighing pan.
ð The next stable weight will be sent automatically to the cursor position of your application.
90 Communication with Peripheral Devices Precision and Analytical Balances
8.3 Collecting measurement results and balance details with EasyDirect Balance
EasyDirect Balance from METTLER TOLEDO is a PC software to collect, analyze, store and export measurement results and balance details from up to 10 balances. EasyDirect Balance supports all Advanced and Standard
Level Laboratory Balances and many legacy models from METTLER TOLEDO. For more information and to download a trial version of the software, please refer to www.mt.com/EasyDirectBalance .
The balance can be connected either via LAN or RS232, but the full functionality is only available with a LAN connection. When using an RS232 connection, only a limited set of data can be collected by EasyDirect
Balance. Please refer to the table "Data available to EasyDirect Balance, depending on the data interface" for more details.
Configure the balance
§ The balance is connected with an Ethernet cable to the LAN.
§ LAN is activated on the balance. For more information, please refer to " Network and Bluetooth ".
1 Tap .
2 Tap General configuration .
3 Tap Devices .
4 The screen Devices appears.
5 Tap .
ð The screen Device/service type appears.
6 Select EasyDirect Balance and confirm with .
ð The screen Connection type appears.
7 Tap Network: LAN and confirm with .
ð The screen Port appears.
8 If necessary, change the port number and confirm with .
ð The screen EasyDirect Balance – Software on PC appears.
9 Tap to confirm.
10 Tap to return to the General configuration screen.
11 Tap Publishing .
ð The screen Publishing appears.
12 Tap Send data .
ð The screen Send data appears.
13 Select the transmission mode, e.g., Automatic, stable and confirm with .
14 Tap to confirm.
15 Tap to return to the previous screen.
Please be aware that the report configuration does not apply to EasyDirect Balance; all available data is sent according the details on the table "Data available to EasyDirect Balance, depending on the data interface".
Transfer data
1 Install EasyDirect Balance on your PC.
2 Follow the instructions to add the balance in EasyDirect Balance.
3 Publish the data from the balance either manually or automatically.
•
•
Note
To connect the balance manually via LAN, you need to know the IP address and port number of your balance, which can be displayed when selecting the EasyDirect Balance under "Services" in the Devices menu.
The PC with EasyDirect Balance and the balances must be connected to the same LAN, e.g.,
168.125.x.xxx
.
Precision and Analytical Balances Communication with Peripheral Devices 91
• When connecting the balance via LAN, certain information is also sent to EasyDirect Balance, independently of any publishing activity.
Data available to EasyDirect Balance, depending on the data interface
LAN RS232
Balance details Balance model
Balance ID
Balance serial number
Balance capacity
Balance readability
Level status
Adjustment status
Service status
Measurement results
Connection status
Gross/Tare/Net weight
Unit 1 and Unit 2 (incl. pcs, %)
Stability condition
Date and Time
Sample and task IDs
Target and tolerances
User name
Application specific results and parameters
Supported activities Weighing
Counting
Percent weighing
Factor weighing
Check weighing
Dynamic weighing
Formulation
Totaling
Back-weighing
Density
Adjustments
Routine test
Repeatability test
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
92 Communication with Peripheral Devices Precision and Analytical Balances
8.4 Connect via USB to a printer and print weighing results
Prerequisite
•
•
The printer is connected to the power supply.
Printer is switched on.
• The printer is connected to the balance via USB cable. Do not connect the printer with the balance, before it is correctly powered on.
The following example shows the option to automatically print the next stable value.
1 Tap .
2 Tap General configuration .
3 Tap Devices .
4 The screen Devices appears.
5 The connected USB device appears automatically.
6 Tap the connected printer, e.g., P-56/58 and confirm with .
ð The screen P-56/58 – Excellence lab printer appears.
7 Tap for the printer test.
ð Test page printed. Check the printout.
8 Tap to confirm.
9 Tap to return to the previous screen.
ð The screen Device/service type appears.
10 Tap to return to the General configuration screen.
11 Tap Publishing .
ð The screen Publishing appears.
12 Tap Print and export report .
ð The screen Print and export – General configuration appears.
13 Tap Single values reports .
ð The screen Single values reports appears.
14 Select the transmission mode, e.g., Automatic, stable and confirm with .
15 Tap to confirm.
16 Tap to return to the previous screen.
17 Place the sample on the weighing pan.
ð The next stable weight will be sent automatically.
Precision and Analytical Balances Communication with Peripheral Devices 93
8.5 Connect via LAN to a P-50 printer and print weighing results
The following example shows the option to automatically print the next stable value.
§ Balance and P-50 printer are connected via Ethernet cables to the same LAN.
1 Tap .
2 Tap General configuration .
3 Tap Network and Bluetooth .
ð The screen Network and Bluetooth appears.
4 Tap LAN .
ð The screen LAN appears.
5 Tap ON to activate LAN and confirm with .
ð LAN connection is established.
6 Tap to return to the General configuration screen.
7 Tap Devices .
8 The screen Devices appears.
9 Tap .
ð The screen Device/service type appears.
10 Select the device to be configured as P-56/58 by tapping.
11 Tap to confirm.
ð The screen Connection type appears.
12 Tap Network: LAN and confirm with .
ð The screen IP address appears.
13 Enter the IP address of the printer and confirm with .
ð The screen Port appears.
14 If necessary, change the port number and confirm with .
ð The screen P-56/58 – Excellence lab printer appears.
15 Tap for the printer test.
ð Test page printed. Check the printout.
16 Tap to confirm.
17 Tap to return to the General configuration screen.
18 Tap Publishing .
ð The screen Publishing appears.
19 Tap Print and export report .
ð The screen Print and export – General configuration appears.
20 Tap Single values reports .
ð The screen Single values reports appears.
21 Select the transmission mode, e.g., Automatic, stable and confirm with .
22 Tap to confirm.
23 Tap to return to the previous screen.
24 Place the sample on the weighing pan.
ð The next stable weight will be sent automatically.
94 Communication with Peripheral Devices Precision and Analytical Balances
8.6 Connect via Bluetooth to a P-50 printer and print weighing results
The following example shows the option to automatically print the next stable value.
§ The wireless dongle MTICWD-100 is connected to the balance and the Bluetooth to RS32 adapter is connected to the printer.
§ Bluetooth is activated on the balance. For more information, please refer to " Network and Bluetooth ".
§ Printer is switched on.
1 Tap .
2 Tap General configuration .
3 Tap Devices .
4 The screen Devices appears.
5 Tap .
ð The screen Device/service type appears.
6 Tap the printer you would like to connect to, e.g., P-56/58 and confirm with .
ð The screen Connection type appears.
7 Tap Bluetooth and confirm with .
ð The screen Bluetooth device appears.
8 Select the device to be configured as P-56/58 by tapping.
9 Tap to confirm.
ð The screen Password appears.
10 Tap to confirm.
ð The screen P-56/58 – Excellence lab printer appears.
11 Tap to confirm.
12 The screen Devices appears.
13 Tap to return to the General configuration screen.
14 Tap Publishing .
ð The screen Publishing appears.
15 Tap Print and export report .
ð The screen Print and export – General configuration appears.
16 Tap Single values reports .
ð The screen Single values reports appears.
17 Select the transmission mode, e.g., Automatic, stable and confirm with .
18 Tap to confirm.
19 Tap to return to the previous screen.
20 Place the sample on the weighing pan.
ð The next stable weight will be sent automatically.
Precision and Analytical Balances Communication with Peripheral Devices 95
8.7 Connect a USB barcode reader and scan the barcode
The following example shows how to scan the sample ID via a barcode reader.
Use the USB connection cable to connect the barcode reader to a balance.
Only the balance settings are changed for this menu item.
Refer to your barcode reader's documentation for information about the barcode reader settings.
The barcode reader must be configured as USB keyboard (with standard key encoding).
Settings on the balance
§ USB barcode reader is connected to the balance.
1 Tap in the status information field on the balance screen.
ð The screen Notifications appears.
2 Tap Input device connected .
ð The screen Input device connected appears.
3 Tap Barcode reader and confirm with .
4 Tap to return to the previous screen.
5 Tap .
6 Tap General configuration .
7 Tap Devices .
8 The screen Devices appears.
9 Tap Barcode reader .
ð The screen Barcode reader – External input device appears.
10 Check the End of line setting. The setting must be the same as the barcode reader.
11 Tap to confirm.
12 Tap to return to the previous screen.
Typical setup to use the barcode reader
1 Tap .
2 Select an application, e.g., Weighing
3 Tap .
ð The screen Weighing – Main configuration appears.
4 Tap .
ð The screen Weighing - reports configuration appears.
5 Tap Identifications .
ð The screen Identifications appears.
6 Tap ID 4 .
7 Activate ID 4 .
8 Select Input prompt and confirm with .
9 Tap to return to the previous screen.
10 Place the sample on the weighing pan.
11 Tap .
ð The screen Sample ID appears.
12 Scan the sample ID with the barcode reader.
ð The sample ID is entered on the Sample ID screen and the screen is then closed.
96 Communication with Peripheral Devices Precision and Analytical Balances
8.8 Connect a USB keyboard
Use a USB keyboard with US, German and French layout (QWERTY, QWERTZ, AZERTY). An external keyboard can be used to enter the IDs. IDs can consist of numeric, alphanumeric and special characters.
§ USB keyboard is connected to the balance.
1 Tap in the status information field on the balance screen.
ð The screen Notifications appears.
2 Tap Input device connected .
ð The screen Input device connected appears.
3 Tap Keyboard and confirm with .
4 Tap to return to the previous screen.
5 Tap .
6 Tap General configuration .
7 Tap Devices .
8 The screen Devices appears.
9 Tap Keyboard .
ð The screen Keyboard – External input device appears.
10 Tap Keyboard .
ð The screen Keyboard language appears.
11 Select your language by tapping and confirm with .
12 Tap to confirm.
13 Tap to return to the previous screen.
ð The IDs can be entered via the external keyboard.
Precision and Analytical Balances Communication with Peripheral Devices 97
8.9 Export measurement results to a USB memory stick
Connect a USB memory stick (FAT32 formatted, high quality, if possible empty) to the USB host interface.
The USB memory stick appears automatically as a new device in the Menu > General configuration >
Devices .
1 Tap .
2 Tap General configuration .
3 Tap Devices .
4 The screen Devices appears.
5 Tap Publishing .
ð The screen Publishing appears.
6 Tap Print and export report .
ð The screen Print and export – General configuration appears.
7 Tap Single values reports .
ð The screen Single values reports appears.
8 Select the transmission mode, e.g., Automatic, stable and confirm with .
9 Tap to confirm.
10 Tap .
ð The screen Export report file appears.
11 Tap File type
ð The screen File type appears.
12 Select, e.g., csv and confirm with .
13 Tap File name .
ð The screen File name appears.
14 If necessary, change the default name and confirm with .
ð The screen Date/Time of file creation appears.
15 Select the Date/Time of file creation at the end or beginning of the report and confirm with .
16 Tap File location .
ð The screen File location appears.
17 Enter a name for the path of the storage folder and confirm with .
18 Tap to confirm.
ð The screen Publishing appears.
19 Tap to confirm.
20 Tap to return to the application screen.
View exported data
1 Connect memory stick to PC.
2 Open the METTLER TOLEDO folder and then the "Reports folder".
3 Open the PDF with your relevant measurement results.
98 Communication with Peripheral Devices Precision and Analytical Balances
8.10 Export XML files to a FTP server and import them with the help of a XSD file
Exporting comprehensive weighing data on a regular basis to an Excel spreadsheet can be done in a very convenient way by exporting/importing the data as an XML file in combination with the use of a XSD file. The
XSD file contains the data structure of an XML document, which allows easy integration of weighing results into an Excel sheet. It maps the elements and attributes from the XML file to your Excel spreadsheet/template.
§ The balance is connected with an Ethernet cable to the LAN.
1 Tap .
2 Tap General configuration .
3 Tap Network and Bluetooth .
ð The screen Network and Bluetooth appears.
4 Tap LAN .
ð The screen LAN appears.
5 Tap ON to activate LAN and confirm with .
ð LAN connection is established.
6 Tap to return to the General configuration screen.
7 Tap Devices .
8 The screen Devices appears.
9 Tap .
ð The screen Device/service type appears.
10 Tap File server (FTP) and confirm with .
ð The screen Connection type appears.
11 Tap Network: LAN and confirm with .
ð The screen IP address appears.
12 Enter the IP address of the FTP server and confirm with .
ð The screen Port appears.
13 If necessary, change the port number and confirm with .
ð The screen Credentials appears.
14 If necessary tap Credentials and activate the option by tapping ON .
15 Enter a user name and password and confirm with .
ð The screen File server (FTP) – Remote storage appears.
16 Tap to confirm.
17 Tap to return to the previous screen.
18 Tap Publishing .
ð The screen Publishing appears.
19 Tap Print and export report .
ð The screen Print and export – General configuration appears.
20 Tap .
ð The screen Export report file appears.
21 Tap File type
ð The screen File type: xml appears.
22 Tap to return to the previous screen.
23 If necessary, change the other settings and confirm with .
24 Tap to return to the previous screen.
Make sure that in the application used, the report content is correctly configured.
1 Tap .
ð The screen Activities - Weighing applications appears.
Precision and Analytical Balances Communication with Peripheral Devices 99
2 Select, e.g., Percent weighing .
3 Tap symbol in the upper left corner to define the application.
ð The screen Percent weighing – Main configuration appears.
4 Tap .
ð The screen Percent weighing – Reports configuration appears.
5 Configure the report and conform all you settings with .
• Download the relevant XSD file for Advanced level balances from www.mt.com/labweighing-softwaredownload .
•
•
Import the XSD file into MS Excel via the developer tab.
Drag and drop the relevant elements and attributes to your Excel spreadsheet.
•
•
Import the XML weighing report you published on the balance from the FTP server.
The Excel spread sheet gets populated with the required data at the pre-defined place.
For more information on how to use XSD and XML files with MS Excel, please consult the internet.
100 Communication with Peripheral Devices Precision and Analytical Balances
8.11 Communication with MT-SICS commands via LAN
To enable you to integrate balances in your IT or data management system in a simple manner most balance functions are also available as appropriate commands (MT-SICS) via the data interfaces. The complete set of available commands are available in the "Reference Manual for MT-SICS commands", which can be found in the expertise library www.mt.com/library .
To communicate with MT-SICS commands with the balance over, e.g., LAN the following configuration is required on the balance.
§ The balance is connected with an Ethernet cable to the LAN.
1 Tap .
2 Tap General configuration .
3 Tap Network and Bluetooth .
ð The screen Network and Bluetooth appears.
4 Tap LAN .
ð The screen LAN appears.
5 Tap ON to activate LAN and confirm with .
ð LAN connection is established.
6 Tap to return to the General configuration screen.
7 Tap Devices .
8 The screen Devices appears.
9 Tap .
ð The screen Device/service type appears.
10 Select Command host and confirm with .
ð The screen Connection type appears.
11 Tap Network: LAN and confirm with .
ð The screen Port appears.
12 If necessary, change the port number and confirm with .
ð The screen Command host – Hosting system appears.
13 Configure the Command set , Character set and End of line according your requirements and confirm with
.
14 Tap to confirm.
15 Tap to return to the General configuration screen.
16 Tap Publishing .
ð The screen Publishing appears.
17 Tap Send data .
ð The screen Send data appears.
18 Select the transmission mode to use when no MT-SICS command is sent from the system and confirm with
.
19 Tap to confirm.
20 Tap to return to the previous screen.
ð The balance is ready to respond on MT-SICS commands via the LAN interface.
Precision and Analytical Balances Communication with Peripheral Devices 101
9 Maintenance
To guarantee the functionality of the balance and the accuracy of the weighing results, a number of maintenance actions must be performed by the user.
9.1 Maintenance tasks
Maintenance action
Performing an internal adjustment
Performing routine tests
(eccentricity test, repeatability test, sensitivity test).
METTLER TOLEDO recommends to at least perform a sensitivity test.
Cleaning
•
•
•
•
Recommended interval
• Daily
•
•
•
After cleaning
After leveling
After changing the location
•
•
•
After cleaning
After assembling the balance
Depending on your internal regulations
(SOP)
After every use
After changing the substance
Depending on the degree of pollution
Depending on your internal regulations
(SOP)
Remarks see "Activities - Adjustments and tests" see "Activities - Adjustments and tests" see "Cleaning"
9.2 Cleaning
9.2.1 Cleaning the glass draft shield (0.1 mg and 1 mg models)
CAUTION
Injury due to sharp objects or broken glass
Instrument components, e.g., glass, can break and lead to injuries.
− Always proceed with focus and care.
102 Maintenance Precision and Analytical Balances
1 Turn the two lock covers (1) on the front.
2 Push the side glass doors back.
1
3 Tilt the front glass.
4 Remove the front glass.
Precision and Analytical Balances Maintenance 103
5 Pull the top glass door out.
6 Lift the side glass doors and pull them out.
104 Maintenance Precision and Analytical Balances
7 Push the lock button to release the rear glass.
8 Remove the rear glass.
Precision and Analytical Balances Maintenance 105
9 Remove weighing pan, pan support and bottom plate.
After cleaning, reinstall all components in the reverse order. For balance mounting, refer to "Assembling the balance".
9.2.2 Cleaning the balance
WARNING
Death or serious injury due to electric shock
Contact with parts carrying a live current can lead to injury and death.
1 Disconnect the instrument from the power supply prior to cleaning and maintenance.
2 Prevent liquid from entering the instrument, terminal or AC/DC adapter.
NOTICE
Damage due to improper cleaning
Improper cleaning can damage the load cell or other essential parts.
1 Do not use any cleaning agents other than the ones specified in the "Reference Manual" or "Cleaning Guide".
2 Do not spray or pour liquids on the instrument. Always use a moistened lint-free cloth or a tissue.
3 Always wipe out from inside to outside of the instrument.
Cleaning around the balance
− Remove any dirt or dust around the balance and avoid further contaminations.
Cleaning the removable parts
− Clean the removed part with a damp cloth or a tissue and a mild cleaning agent.
Cleaning the balance
1 Disconnect the balance from the AC/DC adapter.
2 Use a lint-free cloth moistened with a mild cleaning agent to clean the surface of the balance.
3 Remove powder or dust at first with a disposable tissue.
4 Remove sticky substances with a damp lint-free cloth and a mild solvent.
106 Maintenance Precision and Analytical Balances
Note
Useful details to avoid soiling the instrument are described in the Mettler-Toledo GmbH "SOP for Cleaning a
Balance".
9.2.3 Putting into operation after cleaning
1 Reassemble the balance.
2 Check the functionality of the draft shield.
3 Press to switch on the balance.
4 Warm up the balance. Wait 1h for the acclimatization, before starting the tests.
5 Check the level status, level the balance if necessary.
6 Perform an internal adjustment.
7 Perform a routine test due to the internal regulations of your company. METTLER TOLEDO recommends to perform an repeatability test after cleaning the balance.
8 Press to zero the balance.
ð The balance has been putting into operation and is ready to use.
See also
2
Leveling the balance } Page 23
2
Activities - Adjustments and tests } Page 82
Precision and Analytical Balances Maintenance 107
10 Troubleshooting
Possible errors with their cause and remedy are described in the following chapter. If there are errors that cannot be corrected through these instructions, contact METTLER TOLEDO.
10.1 Error messages
Error message
NO STABILTY
Adjustment aborted
Weight out of range.
EEPROM error.
Wrong cell data.
No standard adjustment.
Program memory defect.
Possible cause
Vibrations at the workplace.
Draft due to untight draft shield and /or open window.
Diagnostic
Place beaker with tap water on the weighing table. Vibrations cause ripples on the water surface.
Make sure draft shield or window is closed.
Remedy
• Protect weighing location against vibrations (vibration absorber, etc.).
• Set weighing parameters coarser
(change Environment from Stable to
Standard or even
Unstable .
•
•
•
Find a different weighing location (by agreement with customer).
Close draft shield or window.
Set weighing parameters coarser
(change Environment from Stable to
Standard or even
Unstable .
The location is not suitable for weighing.
–
Something is touching the weighing pan.
Check for touching parts or dirts.
Wrong adjustment weight. Check weight.
Check and observe the requirements for the location, refer to "Selecting the location".
Remove touching parts or clean the balance.
Place correct weight on the weighing pan.
Data in EEPROM damaged.
Defect load cell data.
–
–
–
–
–
–
Please contact your
METTLER TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
108 Troubleshooting Precision and Analytical Balances
Error message
Temperature sensor defect.
Wrong load cell brand.
Wrong type data set.
Memory full
Possible cause
AC/DC adapter connected to power before connecting to the balance.
Temperature sensor of load cell defect.
Diagnostic
–
Wrong load cell installed. –
Wrong type data set.
Memory full.
–
–
Remedy
Remove the AC/DC adapter from the power and connect first to the balance before connecting to the power if persist please contact your METTLER
TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
Clear the memory by finishing all applications where a measurement is ongoing.
Please contact your
METTLER TOLEDO-Support representative if the problem persists.
A problem occurred while starting the balance.
Some data could not be read correctly from memory. Please proceed and check date and time settings. Please contact your MT-Support representative if the problem persists.
A problem occurred while starting the balance.
Some data could not be read correctly from memory. The instrument will reset and restart.
Please contact your MT-
Support representative if the problem persists.
Weight out of initial zero range
Some data could not be read correctly from memory.
Some data could not be read correctly from memory.
Wrong weighing pan.
Pan is missing.
Pan is not empty.
Weight out of zero range Zero range limit overrun or underrun.
Weight out of tare range Tare range limit overrun or underrun.
Battery backup lost Backup battery is empty.
This battery ensures that the date and time are not lost when the balance is disconnected from power.
Check date and time settings.
–
Check weighing pan.
–
–
Connect the balance to the power supply for charging the battery (full capacity after 2 days of charging).
Please contact your
METTLER TOLEDO-Support representative.
Mount correct weighing pan or unload weighing pan.
Reduce / increase weight on weighing pan.
Reduce / increase weight on weighing pan.
If battery cannot be recharged, contact your
METTLER TOLEDO-Support representative.
Precision and Analytical Balances Troubleshooting 109
Error message
USB device is not recognized when connected to balance
Possible cause
Fluctuation in the external electric grid.
Interference from power line.
Diagnostic
–
Remedy
Restart balance.
10.2 Error symptoms
Error symptom
Display is dark
Touch display is not responding
Touch display is partly not responding
Possible cause Diagnostic
Instrument is switched off. –
Power plug not connected. Check
Power supply not connected to balance.
Power supply is faulty.
Wrong power supply.
Touch display is not correctly adjusted.
Check
Balance must be restarted. –
Connector socket on balance is corroded or faulty.
Check
Display is faulty.
Replace display.
Touch display is faulty.
Replace display.
–
Remedy
Switch on the instrument.
Connect power cable to power supply.
Connect power supply.
Check/test
Check that input data on type plate match the power supply values.
Replace power supply.
Use proper power supply.
Restart balance.
Please contact your
METTLER TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
Please contact your
METTLER TOLEDO-Support representative.
Perform touch screen adjustment.
Reset balance (factory reset).
110 Troubleshooting Precision and Analytical Balances
Error symptom
The value drifts into plus or minus
Possible cause
Room, environment not suitable.
Direct sunlight or other heat source.
Diagnostic
–
Is any sun shade (blinds, curtains, etc.) available?
Weighing sample absorbs moisture or evaporates moisture.
Weighing sample is electrostatically charged.
Weighing sample is hotter or colder than the air in the weighing chamber.
Instrument has not yet reached thermal equilibrium.
•
•
•
•
Is the weighing result with a test weight stable?
Sensitive weighing samples, e.g., paper, cardboard, wood, plastic, rubber, liquids.
Is the weighing result with a test weight stable?
Sensitive weighing samples, e.g., plastic, powder, insulating materials.
Weighing operation with test weight does not show this effect.
•
•
Was there a power outage?
Was power supply disconnected?
Remedy
Environmental recommendations
• Windowless, non airconditioned room, e.g., basement.
•
•
•
•
•
•
Only one person in the weighing room.
Sliding doors.
Standard doors cause pressure changes.
No draft in weighing room (check with suspended threads).
No air conditioning
(temperature oscillates, draft).
Acclimatize balance, take dummy measurements.
Instrument uninterruptedly connected to the power supply (24h per day).
Select location according to "Selecting the location" (customer responsibility).
•
•
Use aids.
Cover weighing sample.
•
•
Increase air humidity in weighing chamber
(45% - 50%).
Use ionizer.
Bring weighing sample to room temperature before weighing.
•
•
Acclimatize instrument for at least 1 hour.
Depending on climatic conditions, extend this period accordingly.
Instrument switched on for at least 1 hour, refer to "General data"
Precision and Analytical Balances Troubleshooting 111
Error symptom
Display shows overload or underload
Possible cause
The weight on the weighing pan exceeds the weighing capacity of the instrument.
Wrong weighing pan.
Diagnostic
Check weight.
No weighing pan.
Incorrect zero point at switch-on.
Remedy
Reduce the weight on the weighing pan.
–
–
Slightly lift or press weighing pan. The weight display appears.
Use proper weighing pan.
Install weighing pan.
•
•
Switch off balance.
Disconnect and reconnect power cable.
10.3 Status messages/Status icons
Status messages are displayed by means of small icons. The status icons indicate the following:
Icon Status description
Automatic FACT adjustment is currently not possible.
Service due.
Diagnostic
Instrument is busy.
–
Remedy
• Unload the balance.
• Do not press any key for 2 minutes. The display stabilizes.
Please contact your METTLER
TOLEDO-Support representative.
The built-in level sensor has detected that the instrument is not correctly leveled.
The balance battery must be replaced. This battery ensures that the date and time are retained when the balance is disconnected from the power supply.
LAN cable disconnected.
LAN connection problem.
Instrument is out of level.
Replace battery.
Check cable.
Settings can be modified.
Immediately level the instrument.
Please contact your METTLER
TOLEDO-Support representative.
WLAN not connected.
Network not selected.
Settings can be modified.
Connect the cable.
Tap icon in the status field.
• View corresponding settings.
Link to corresponding settings in Network and
Bluetooth area by means of
"edit" function.
Select network.
Tap icon in the status field.
• View corresponding settings.
Link to corresponding settings in Network and
Bluetooth area by means of
"edit" function.
112 Troubleshooting Precision and Analytical Balances
Icon Status description
WLAN connected. Three bars are indicating signal strength of connection.
Diagnostic
Settings can be modified.
Remedy
Tap icon in the status field.
• View corresponding settings.
Link to corresponding settings in Network and
Bluetooth area by means of
"edit" function.
Precision and Analytical Balances Troubleshooting 113
Icon Status description
Connection request.
Connection request failed.
Publishing failed
Diagnostic
An external Bluetooth device wants to connect with balance.
Connection with external
Bluetooth device failed.
External input device connected. A keyboard or a barcode reader has been connected to the balance, and the balance could not recognize the device type.
One or more publishing processes have failed.
Remedy
Tap icon in the status field.
• View corresponding workflow.
Connect external device and select the device type in
Devices and services .
Tap icon in the status field.
• View corresponding workflow.
Connect external device and select the device type in
Devices and services .
Tap icon in the status field.
• View corresponding workflow.
Connect external device and select the device type in
Devices and services .
Tap icon in the status field.
View corresponding message(s).
All fails are recorded in the ISO-
Log.
10.4 Putting into operation after fixing an error
After fixing an error, perform the following steps to put the balance into operation:
• Ensure that the balance is completely reassembled and cleaned.
• Reconnect the balance to the AC/DC adapter.
114 Troubleshooting Precision and Analytical Balances
11 Technical Data
11.1 General data
Standard power supply
AC/DC adapter:
Balance power consumption:
Polarity:
Mean see level:
Battery operation:
Optional power supply
AC/DC adapter:
Cable for AC/DC adapter:
Polarity:
Protection and standards
Overvoltage category:
Degree of pollution:
Protection:
Standards for safety and EMC:
Range of application:
Environmental conditions
Height above mean sea level:
Height above mean sea level:
Ambient temperature:
Relative air humidity:
Warm-up time:
Primary: 100 – 240 V~ 0.5 A, 50 – 60 Hz
Secondary: 12 V 1.0 A MAX 24 – 34 VA LPS (Limited Power
Source), SELV (Safety Extra Low Voltage)
12 V DC, 0.84 A
Can be used up to 2000 m above mean sea level
If the balance is used above 2000 m mean sea level, the optional power supply must be used
8 standard AA batteries (alkaline or lithium) for up to 8 hours for balances with a readability of 1 - 100 mg and 2 hours for balances with a readability of 0.1 mg.
Primary: 100 – 240 V~ 0.8 A, 50 – 60 Hz, 60 – 80 VA
Secondary: 12 V DC, 2.5 A LPS (Limited Power Source), SELV
(Safety Extra Low Voltage)
3-core, with country-specific plug
II
2
Protected against dust and water
See Declaration of Conformity
Use only indoors in dry locations
Up to 2000 m (Standard power supply)
Up to 4000 m (Optional power supply)
Operating conditions for ordinary lab application: +10 to 30 °C
(operability guaranteed between +5 and 40 °C)
Max. 80% up to 31 °C, linearly decreasing to 50% at 40 °C, noncondensing
At least 30 minutes (60 minutes for 0.1 mg models) after connecting the balance to the power supply. When switched on from standby, the instrument is ready for operation immediately.
Precision and Analytical Balances Technical Data 115
Materials
Housing:
Weighing pan:
Draft shield element:
Draft shield:
Protective cover:
TFT touch screen surface:
Top Housing: Plastic (ABS)
Bottom housing: Die-cast aluminum, powder coated
170 × 190 mm: Stainless steel X5CrNi18-10 (1.4301)
ø 120 mm: Stainless steel X5CrNi18-10 (1.4301)
ø 90 mm: Stainless steel X2CrNiMo 17-12-2 (1.4404)
Roughness Ra < 0.8 μm
0.1 mg models: Stainless steel X2CrNiMo 17-12-2 (1.4404)
Plastic (ABS), glass
Plastic (PET)
Glass
116 Technical Data Precision and Analytical Balances
11.2 Model-specific data
11.2.1 Analytical balances with a readability of 0.1 mg with draft shield
1)
ML54T
Limit values
Capacity
Nominal load
52 g
50 g
Readability
Repeatability
0.1 mg
0.1 mg
Linearity deviation 0.2 mg
Eccentricity deviation (at test load) 0.3 mg (20 g)
Sensitivity offset (at nominal load) 1)
Sensitivity temperature drift 2)
0.8 mg
0.0002%/°C
Typical values
Repeatability 0.08 mg
Linearity deviation 0.06 mg
Eccentricity deviation (at test load) 0.1 mg (20 g)
Sensitivity offset (at nominal load) 1)
Minimum weight (USP, tolerance = 0.10%) 3)
0.5 mg
160 mg
Minimum weight
(tolerance = 1%) 3)
Settling time
Adjustment
Dimensions & other specifications
16 mg
2 s
Int. Cal/FACT
Balance dimensions (W×D×H) 193×290×331 mm
Weighing pan dimensions (W×D) Ø 90 mm
Usable height of draft shield
Balance weight
Weights for routine testing
Weights (OIML class)
Weights (ASTM class)
235 mm
4.1 kg
2 g (F2)/ 50 g (F2)
2 g (ASTM 1)/ 50 g (ASTM 1) after adjustment with internal weight
ML104T
120 g
100 g
0.1 mg
0.1 mg
0.2 mg
0.3 mg (50 g)
1 mg
0.0002%/°C
0.08 mg
0.06 mg
0.1 mg (50 g)
0.6 mg
160 mg
16 mg
2 s
Int. Cal/FACT
193×290×331 mm
Ø 90 mm
235 mm
4.1 kg
5 g (F2)/ 100 g (F2)
5 g (ASTM 1)/ 100 g (ASTM 1)
2)
3) in the temperature range 10 °C … 30 °C determined at 5% load, k = 2
Precision and Analytical Balances Technical Data 117
118
1)
ML204T
Limit values
Capacity
Nominal load
Readability
Repeatability
Sensitivity offset (at nominal load) 1)
220 g
200 g
Linearity deviation 0.2 mg
Eccentricity deviation (at test load) 0.4 mg (100 g)
Sensitivity temperature drift 2)
Typical values
Repeatability
0.1 mg
0.1 mg
1.2 mg
0.0002%/°C
0.08 mg
Linearity deviation 0.06 mg
Eccentricity deviation (at test load) 0.12 mg (100 g)
Sensitivity offset (at nominal load)
Minimum weight (USP, tolerance = 0.10%) 3)
1) 0.8 mg
160 mg
ML304T
320 g
300 g
0.1 mg
0.1 mg
0.2 mg
0.4 mg (100 g)
2.4 mg
0.0002%/°C
0.08 mg
0.06 mg
0.12 mg (100 g)
1.5 mg
160 mg
Minimum weight
(tolerance = 1%) 3)
Settling time
Adjustment
Usable height of draft shield
16 mg
2 s
Int. Cal/FACT
Dimensions & other specifications
Balance dimensions (W×D×H) 193×290×331 mm
Weighing pan dimensions (W×D) Ø 90 mm
235 mm
4.1 kg Balance weight
Weights for routine testing
Weights (OIML class)
Weights (ASTM class)
10 g (F2)/ 200 g (F2)
16 mg
3 s
Int. Cal/FACT
193×290×331 mm
Ø 90 mm
235 mm
4.1 kg
10 g (F2)/ 200 g (F2)
10 g (ASTM 1)/ 200 g (ASTM 1) 10 g (ASTM 1)/ 200 g (ASTM 1) after adjustment with internal weight
2) in the temperature range 10 °C … 30 °C
3) determined at 5% load, k = 2
Technical Data Precision and Analytical Balances
11.2.2 Precision balances with a readability of 1 mg with draft shield
1)
ML203T
Limit values
Capacity
Nominal load
Readability
Repeatability
220 g
200 g
1 mg
1 mg
Linearity deviation 2 mg
Eccentricity deviation (at test load) 4 mg (100 g)
Sensitivity offset (at nominal load)
Sensitivity temperature drift 2)
1) 8 mg
0.0003%/°C
Typical values
Repeatability 0.7 mg
Linearity deviation 0.6 mg
Eccentricity deviation (at test load) 1.5 mg (100 g)
Sensitivity offset (at nominal load) 1)
Minimum weight (USP, tolerance = 0.10%) 3)
5 mg
1.4 g
ML303T
320 g
300 g
1 mg
1 mg
2 mg
4 mg (100 g)
8 mg
0.0003%/°C
0.7 mg
0.6 mg
1.5 mg (100 g)
5 mg
1.4 g
ML503T
520 g
500 g
1 mg
1 mg
2 mg
4 mg (200 g)
8 mg
0.0003%/°C
0.7 mg
0.6 mg
1.5 mg (200 g)
5 mg
1.4 g
Minimum weight
(tolerance = 1%) 3)
Settling time
Adjustment
Dimensions & other specifications
140 mg
1.5 s
Int. Cal/FACT
140 mg
1.5 s
Int. Cal/FACT
140 mg
1.5 s
Int. Cal/FACT
Balance dimensions (W×D×H) 193×290×331 mm 193×290×331 mm 193×290×331 mm
Weighing pan dimensions (W×D) Ø 120 mm Ø 120 mm Ø 120 mm
Usable height of draft shield
Balance weight
Weights for routine testing
Weights (OIML class)
Weights (ASTM class)
230 mm
4.2 kg
230 mm
4.2 kg
230 mm
4.2 kg
10 g (F2)/ 200 g (F2) 10 g (F2)/ 200 g (F2) 20 g (F2)/ 500 g (F2)
10 g (ASTM 1)/ 200 g
(ASTM 1)
10 g (ASTM 1)/ 200 g
(ASTM 1)
20 g (ASTM 1)/ 500 g
(ASTM 1) after adjustment with internal weight
2)
3) in the temperature range 10 °C … 30 °C determined at 5% load, k = 2
Precision and Analytical Balances Technical Data 119
11.2.3 Precision balances with a readability of 10 mg
1)
ML802T
Limit values
Capacity
Nominal load
Readability
Repeatability
820 g
800 g
10 mg
10 mg
Linearity deviation 20 mg
Eccentricity deviation (at test load) 30 mg (500 g)
Sensitivity offset (at nominal load)
Sensitivity temperature drift 2)
1) 80 mg
0.0003%/°C
Typical values
Repeatability 7 mg
Linearity deviation 6 mg
Eccentricity deviation (at test load) 10 mg (500 g)
Sensitivity offset (at nominal load) 1)
Minimum weight (USP, tolerance = 0.10%) 3)
50 mg
14 g
ML1602T
1620 g
1600 g
10 mg
10 mg
20 mg
30 mg (500 g)
80 mg
0.0003%/°C
7 mg
6 mg
10 mg (500 g)
50 mg
14 g
Minimum weight
(tolerance = 1%) 3)
Settling time
Adjustment
Dimensions & other specifications
1.4 g
1.5 s
Int. Cal/FACT
Balance dimensions (W×D×H) 184×290×84 mm
Weighing pan dimensions (W×D) 170×190 mm
Usable height of draft shield
Balance weight
Weights for routine testing
Weights (OIML class)
–
3.6 kg
Weights (ASTM class)
1.4 g
1.5 s
Int. Cal/FACT
184×290×84 mm
170×190 mm
–
3.6 kg
20 g (F2)/ 500 g (F2) 50 g (F2)/ 1000 g
(F2)
20 g (ASTM 1)/ 500 g
(ASTM 1)
50 g (ASTM 1)/
1000 g (ASTM 1) after adjustment with internal weight
2)
3) in the temperature range 10 °C … 30 °C determined at 5% load, k = 2
ML3002T
3.2 kg
3 kg
10 mg
10 mg
20 mg
40 mg (1000 g)
80 mg
0.0003%/°C
7 mg
6 mg
15 mg (1000 g)
50 mg
14 g
1.4 g
1.5 s
Int. Cal/FACT
184×290×84 mm
170×190 mm
–
3.6 kg
100 g (F2)/ 2000 g
(F2)
100 g (ASTM 1)/
2000 g (ASTM 1)
120 Technical Data Precision and Analytical Balances
1)
ML4002T
Limit values
Capacity
Nominal load
Readability
Repeatability
Sensitivity offset (at nominal load) 1)
4.2 kg
4 kg
Linearity deviation 20 mg
Eccentricity deviation (at test load) 40 mg (2 kg)
Sensitivity temperature drift 2)
Typical values
Repeatability
10 mg
10 mg
80 mg
0.0003%/°C
7 mg
Linearity deviation 6 mg
Eccentricity deviation (at test load) 15 mg (2 kg)
Sensitivity offset (at nominal load)
Minimum weight (USP, tolerance = 0.10%) 3)
1) 50 mg
14 g
ML6002T
6.2 kg
6 kg
10 mg
10 mg
20 mg
60 mg (2 kg)
80 mg
0.0003%/°C
7 mg
6 mg
20 mg (2 kg)
50 mg
14 g
Minimum weight
(tolerance = 1%) 3)
Settling time
Adjustment
Usable height of draft shield
1.4 g
1.5 s
Int. Cal/FACT
Dimensions & other specifications
Balance dimensions (W×D×H) 184×290×84 mm
Weighing pan dimensions (W×D) 170×190 mm
–
3.6 kg Balance weight
Weights for routine testing
Weights (OIML class)
Weights (ASTM class)
200 g (F2)/ 2000 g (F2)
1.4 g
1.5 s
Int. Cal/FACT
184×290×84 mm
170×190 mm
–
3.6 kg
200 g (F2)/ 5000 g (F2)
200 g (ASTM 4)/ 2000 g (ASTM 4) 200 g (ASTM 4)/ 5000 g (ASTM 4) after adjustment with internal weight
2) in the temperature range 10 °C … 30 °C
3) determined at 5% load, k = 2
Precision and Analytical Balances Technical Data 121
11.2.4 Precision balances with a readability of 100 mg
1)
ML3001T
Limit values
Capacity
Nominal load
Readability
Repeatability
3.2 kg
3 kg
100 mg
100 mg
Linearity deviation 50 mg
Eccentricity deviation (at test load) 300 mg (1000 g)
Sensitivity offset (at nominal load)
Sensitivity temperature drift 2)
1) 210 mg
0.0005%/°C
Typical values
Repeatability 70 mg
Linearity deviation 20 mg
Eccentricity deviation (at test load) 100 mg (1000 g)
Sensitivity offset (at nominal load) 1)
Minimum weight (USP, tolerance = 0.10%) 3)
120 mg
140 g
ML6001T
6.2 kg
6 kg
100 mg
100 mg
200 mg
300 mg (2 kg)
300 mg
0.0005%/°C
70 mg
60 mg
100 mg (2 kg)
150 mg
140 g
Minimum weight
(tolerance = 1%) 3)
Settling time
Adjustment
Dimensions & other specifications
14 g
1 s
Int. Cal/FACT
Balance dimensions (W×D×H) 184×290×84 mm
Weighing pan dimensions (W×D) 170×190 mm
Usable height of draft shield
Balance weight
Weights for routine testing
Weights (OIML class)
Weights (ASTM class)
–
3.3 kg
200 g (F2)/ 2000 g (F2)
14 g
1 s
Int. Cal/FACT
184×290×84 mm
170×190 mm
–
3.3 kg
200 g (F2)/ 5000 g (F2)
200 g (ASTM 4)/ 2000 g (ASTM 4) 200 g (ASTM 4)/ 5000 g (ASTM 4) after adjustment with internal weight
2)
3) in the temperature range 10 °C … 30 °C determined at 5% load, k = 2
122 Technical Data Precision and Analytical Balances
11.3 Dimensions
11.3.1 Balances with a readability of 0.1 mg with draft shield (235 mm)
Models:
ML54T
ML104T
ML204T
ML304T
126 97
78
ø 90
54
134
169
216
185 19
290
156
Precision and Analytical Balances Technical Data 123
11.3.2 Balances with a readability of 1 mg with draft shield (235 mm)
Models:
ML203T
ML303T
ML503T
126
78
ø 120
97
54
134
169
216
185
19
290
156
124 Technical Data Precision and Analytical Balances
11.3.3 Balances with a readability of 10 mg with square weighing pan and draft shield element
Models:
ML802T
ML1602T
ML3002T
ML4002T
ML 6002T
200
167
54 200
87 203
290
190
Precision and Analytical Balances Technical Data 125
11.3.4 Balances with a readability of 100 mg with square weighing pan
Models:
ML3001T
ML6001T
167
54 200
87 203
290
190
126 Technical Data Precision and Analytical Balances
11.4 Interface specifications
11.4.1 RS232C interface
Each balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device, e.g., printer or computer.
Schematic
GND
5
9 6
RxD
TxD
1
CTS
RTS
+12V
2nd display mode only
DATA
IN
OUT
HAND
SHAKE
IN
OUT
POWER
SUPPLY
OUT
Item
Interface type
Specification
Voltage interface according to EIA RS232C/
DIN66020 CCITT V24/V.28)
Max. cable length 15 m
Signal level Outputs:
+5 V … +15 V (RL = 3–7 kΩ)
–5 V … –15 V (RL = 3–7 kΩ)
Inputs:
+3 V … +25 V
–3 V … -25 V
Connector Sub-D, 9-pole, female
Operating mode Full duplex
Transmission mode Bit-serial, asynchronous
Transmission code ASCII
Baud rates 600, 1200, 2400, 4800, 9600, 19200,
38400 (software selectable)
Bits/parity
Stop bits
7-bit/none, 7-bit/even, 7-bit/odd, 8-bit/none
(software selectable)
1 stop bit
Handshake
End-of-line
Power supply for
2nd display
None, XON/XOFF, RTS/CTS (software selectable)
<CR><LF>, <CR>, <LF> (software selectable)
+ 12 V, max 40 mA (software selectable, 2nd display mode only)
Precision and Analytical Balances Technical Data 127
11.4.2 USB host
Each balance is equipped with a USB host as standard for the attachment of a peripheral device, e.g., printer, barcode reader.
Schematic Item
Standard
Specification
In conformity with USB specifications revision
2.0
Speed
Power usage
Connector
Pin assignment
Full-speed 12 Mbps (requires shielded cable)
Max. 500 mA
Type A
1
2
VBUS (+5 V DC)
D- (Data -)
3
4
Shell
D+ (Data +)
GND (Ground)
Shield
11.4.3 USB device
Each balance is equipped with a USB device interface as standard for the attachment of a peripheral device, e.g., computer.
Note
This interface is not able to communicate with a printer.
Schematic Item
Standard
Speed
Specification
In conformity with USB specifications revision
2.0
Full-speed 12 Mbps (requires shielded cable)
Function CDC (Communication Device Class) serial port emulation
Power usage Suspended device: Max 10 mA
Connector Type B
1 VBUS (+5 VDC)
2
3
D– (Data –)
D+ (Data +)
4 GND (Ground)
Shield Shield
11.4.4 Ethernet
Each balance is equipped with an Ethernet interface as standard to connect to a LAN or a P-50 printer.
Item
Connection
Speed
Recommended Ethernet cable
Supported Ethernet standard
Supported network protocols and services
Specification
RJ45
10/100 Mbps (10BASE-T, 100BASE-TX, Auto MDI-X)
Cat 5
IEEE 802.3
TCP/IP, FTP
128 Technical Data Precision and Analytical Balances
11.4.5 Wi-Fi and Bluetooth
There is the optional METTLER TOLEDO wireless dongle (MTICWD-100, part number 30412536) to connect to a WLAN and Bluetooth devices.
Wi-Fi
Item
Connection
Wi-Fi RF transmission standards
Wi-Fi output power
Supported security protocols
Supported network protocols and services
Specification
USB 2.0 (bus powered device)
IEEE 802.11 a/b/g/n (2.4 and 5 GHz) max. 15 dBm
Open / WPA / WPA2 / LEAP / PEAP
TCP/IP, FTP
Bluetooth
Item
Connection
Bluetooth transmission standards
Bluetooth output power
Transmission range
Supported Bluetooth profiles
Supported security modes
Specification
USB 2.0 (bus powered device)
BR/EDR v2.1/Low Energy v4.0 (2.4 GHz) max. 11 dBm up to 100 m
SPP (Serial Port Profile)
BT2.0: fixed-pin BT4.0: Display Yes/No
11.4.6 MT-SICS interface commands and functions
Many of the instruments and balances used have to be able to integrate into a complex computer or data acquisition system.
To easily integrate a balance into a system and utilize its capacity to the full extent, most balance functions are also available as corresponding commands via the data interface.
All new METTLER TOLEDO balances launched on the market support "METTLER TOLEDO Standard Interface
Command Set" (MT-SICS). The commands available depend on the functionality of the balance.
For further information, please contact your METTLER TOLEDO representative.
Refer to the MT-SICS Reference Manual which can be downloaded from the Internet at u www.mt.com/library
Precision and Analytical Balances Technical Data 129
12 Accessories and Spare Parts
12.1 Accessories
Accessories are additional components that could help you in your workflow.
Description
Printers
RS-P25/01 (EMEA) printer with RS232C connection to instrument
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P25/02 (Asia-Pacific) printer with RS232C connection to instrument
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P25/03 (Northern America) printer with RS232C connection to instrument
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P26/01 (EMEA) printer with RS232C connection to instrument (with date and time)
Paper roll (length: 20 m), set of 5 pcs
Paper roll, self-adhesive (length: 13 m), set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P26/02 (Asia-Pacific) printer with RS232C connection to instrument (with date and time)
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P26/03 (Northern America) printer with RS232C connection to instrument (with date and time)
Paper roll (length: 20 m), set of 5 pcs
Paper roll, self-adhesive (length: 13 m), set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
Part No.
11124300
00072456
11600388
00065975
11124310
00072456
11600388
00065975
11124320
00072456
11600388
00065975
11124303
00072456
11600388
00065975
11124313
00072456
11600388
00065975
11124323
00072456
11600388
00065975
130 Accessories and Spare Parts Precision and Analytical Balances
Precision and Analytical Balances
RS-P28/01 (EMEA) printer with RS232C connection to instrument (with date, time and applications)
Paper roll (length: 20 m), set of 5 pcs
Paper roll, self-adhesive (length: 13 m), set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P28/02 (Asia-Pacific) printer with RS232C connection to instrument (with date, time and applications)
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
RS-P28/03 (Northern America) printer with RS232C connection to instrument (with date, time and applications)
Paper roll (length: 20 m), set of 5 pcs
Paper roll, self-adhesive (length: 13 m), set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
USB-P25/01 (EMEA) printer with USB connection to instrument
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
USB-P25/02 (Asia-Pacific) printer with USB connection to instrument
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
USB-P25/03 (Northern America) printer with USB connection to instrument
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
P-52RUE dot matrix printer RS232C, USB and Ethernet connections, simple print-outs
Paper roll (length: 20 m), set of 5 pcs
Paper roll (length: 13 m), self-adhesive, set of
3 pcs
Ribbon cartridge, black, set of 2 pcs
P-56RUE thermal printer with RS232C, USB and Ethernet connections, simple print-outs, date and time
Paper roll, white (length: 27 m), set of 10 pcs
Paper roll, white, self-adhesive (length: 13 m), set of 10 pcs
Accessories and Spare Parts 131
00065975
11124301
00072456
11600388
00065975
11124311
00072456
11600388
00065975
11124321
00072456
11600388
00065975
30237290
00072456
11600388
00065975
30094673
30094723
30094724
11124304
00072456
11600388
00065975
11124314
00072456
11600388
00065975
11124324
00072456
11600388
P-58RUE thermal printer with RS232C, USB and Ethernet connections, simple print-outs, date and time, label printing, balance applications, e.g., statistics, formulation, totaling, SQC
Paper roll, white (length: 27 m), set of 10 pcs
Paper roll, white, self-adhesive (length: 13 m), set of 10 pcs
Paper roll, white, self-adhesive labels (550 labels), set of 6 pcs
Dimension of the label 56×18 mm
Cables for RS232C interface
RS9 – RS9 (m/f): connection cable for PC, length = 1 m
30094674
30094723
30094724
30094725
11101051
RS9 – RS25 (m/f): connection cable for PC, length = 2 m 11101052
RS232C - USB converter cable – cable with converter to connect a balance (RS232C) to a USB port
Cables for USB interface
USB 2.0 High-speed cable to connect balance with PC (USB A to USB B), length = 1 m
30241476
Wireless interfaces
Bluetooth RS232C serial adapter ADP-BT-S for wireless connection between:
•
•
Balance and PC
Printer and balance
Bluetooth RS232C serial adapter set ADP-BT-P for wireless connection between printer and balance.
64088427
30086494
30086495
132 Accessories and Spare Parts Precision and Analytical Balances
Wireless dongle MTICWD-100
Wireless dongle to connect to a WLAN network and/or via
Bluetooth to a P-50RUE Printer or PC/Android Mobile Device
(Balance software version 4.10 or higher required).
The wireless dongle is connected to the USB Host interface.
To connect to a P-50 RUE Printer the Bluetooth RS232C serial adapter ADP-BT-S (30086494) needs to be connected to the printer.
Density determination
Glass beaker, height 100 mm, ø 60 mm
30412536
00238166
Sinker for density of liquids in conjunction with density kit
Calibrated (sinker + certificate)
Recalibrated (new certificate)
Calibrated thermometer with certificate
00210260
00210672
00210674
11132685
Density Kit Advanced & Standard for balance models with a readability of 0.1 mg / 1 mg
Auxiliary displays
RS232C auxiliary display AD-RS-M7
METTLER TOLEDO
000000
30535760
Draft shields
Draft shield ML-DS-21 for models with readability of 10 mg up to 100 mg.
12121015
12122381
Precision and Analytical Balances Accessories and Spare Parts 133
Pan protections
Protective foils, 164x184 mm, set of 20 pcs, pan protection for weighing pan 170x190 mm
Le arn more
www
.m
about our Pre t.com
/precison-balances cisi on
Balances
Protective covers
Protective cover for models with square weighing pan
30113801
30241513
Protective cover for models with draft shield "165/235 mm" 30241514
Anti-theft devices
Anti-theft cable with lock 11600361
Software
EasyDirect Balance
Various
EasyDirect Balance from METTLER TOLEDO is an application software to collect, analyze, store and export balance measurement and device data from up to 10 balances on one
PC.
License EasyDirect Balance for 10 Instruments
License EasyDirect Balance for 3 Instruments
AC/DC universal adapter (EU, USA, AU, UK) 100–240 VAC,
50/60 Hz, 0.5 A, 12 VDC 1 A
30540473
30539323
11120270
AC/DC adapter (without power cable) 100–240 V AC, 0.8 A,
50/60 Hz, 12 V DC 2.5 A
11107909
134 Accessories and Spare Parts Precision and Analytical Balances
Country-specific 3-Pin power cable with grounding conductor.
Power cable AU
Power cable BR
Power cable CH
Power cable CN
Power cable DK
Power cable EU
Power cable GB
Power cable IL
Power cable IN
Power cable IT
Power cable JP
Power cable TH, PE
Power cable US
Power cable ZA
Adjustment weights
OIML / ASTM weights (with calibration certificate) see www.mt.com/weights
00088751
30015268
00087920
30047293
00087452
00087925
00089405
00225297
11600569
00087457
11107881
11107880
00088668
00089728
Precision and Analytical Balances Accessories and Spare Parts 135
12.2 Spare parts
Spare parts are parts that are delivered with the original instrument but that can be replaced, if needed, without the help of a service technician .
Balances with readability of 0.1 mg, with draft shield (235 mm)
1
2
3
4
5
6
7
8
9
6
7
4
5
8
1
2
3
Order no.
12122010
12122042
12122043
12122044
12122033
30241516
12122036
12122035
9 30241515
10 12122029
11 12122041
12 12122040
11
10
12
Designation
Weighing pan Ø 90 mm
Pan support Ø 90 mm
Draft protection ring
Bottom plate
Top glass door
Rear glass
Side glass door
Pair of handles
Front glass
Weighing below balance cap
Battery chamber cover
Leveling foot
–
–
–
–
–
–
–
–
–
–
–
Remarks
Excluding: Pan support
136 Accessories and Spare Parts Precision and Analytical Balances
Balances with readability of 1 mg, with draft shield (235 mm)
1
2
3
4
5
6
7
8
5
6
3
4
1
2
Order no.
12122037
12122045
12122044
12122033
30241516
12122036
7
8
12122035
30241515
9 12122029
10 12122041
11 12122040
11
10
9
Designation
Weighing pan Ø 120 mm
Pan support Ø 120 mm
Bottom plate
Top glass door
Rear glass
Side glass door
Pair of handles
Front glass
Weighing below balance cap
Battery chamber cover
Leveling foot
–
–
–
–
–
–
–
–
–
Remarks
–
–
Precision and Analytical Balances Accessories and Spare Parts 137
Balances with readability of 10 mg with square weighing pan and draft shield element
1
2
3
4
5
6
3
4
1
2
7
Order no.
12122048
12122049
12122050
11131029
12122029
12122041
12122040
6
5
7
Designation
Weighing pan 170×190 mm
Pan support 170×190 mm
Draft shield element
Cap pan support
Weighing below balance cap
Battery chamber cover
Leveling foot
Remarks
–
–
–
–
–
–
–
138 Accessories and Spare Parts Precision and Analytical Balances
Balances with readability of 100 mg with square weighing pan
1
2
3
3
4
1
2
5
6
Order no.
12122048
12122049
11131029
12122029
12122041
12122040
4
5
6
Designation
Weighing pan 170×190 mm
Pan support 170×190 mm
Cap pan support
Weighing below balance cap
Battery chamber cover
Leveling foot
–
–
–
–
Remarks
–
–
Precision and Analytical Balances Accessories and Spare Parts 139
13 Disposal
In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties, the content of this regulation must also be related.
140 Disposal Precision and Analytical Balances
14 Appendix
14.1 Compliance information
European Union
The instrument complies with the directives and standards listed on the EU Declaration of Conformity.
United States of America
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
The FCC Supplier Declaration of Conformity is available online.
u http://www.mt.com/ComplianceSearch
Precision and Analytical Balances Appendix 141
142 Appendix Precision and Analytical Balances
Index
A
AC/DC adapter
Access rights
Accessories
Action buttons
Activities
Back-weighing
Check weighing
Counting
Density
Dynamic weighing
Factor weighing
Formula
Percent weighing
Totaling
Weighing
Applications
Adjusting
Adjustments
Adjustments and tests
Advanced options
Application home screen
Auto lock
Auto lock function
Autoincrement
Autopublish
B
Backlight
Back-weighing
Balance history
Balance information
Barcode reader
Battery operation
Baudrate
Beep
Below-the-balance weighing
Bit/Parity
Bluetooth
Bluetooth devices
C
Changing the readability
Changing the time
Check weighing
Precision and Analytical Balances
Cleaning
Coach text field
Command host
Compliance information
Conventions and symbols
Counting
D
Date
Defining reference piece number
Defining statistics
Defining the measuring time
Defining tolerance threshold
Density
Density kit
Devices and services
Dimensions
Display results
Disposal
Dosing
Draft shield
Dynamic weighing
E
EasyDirectBalance
Entering characters and numbers
Environment
Environmental conditions
Ethernet
Export report file
F
FACT
Factor weighing
Fill up
Fill up sample
Formula
FTP server
Function PC-Direct
G
General configuration and data
Publishing
System settings
USB device
General settings
Groups
Index 143
H
I
Handshake
History
Host
Identification dialog screen
Information and work bars
Information fields
Input prompt
Installing the components
Interface
MT-SICS
USB device
Interface RS232C
Internal adjustment
Introduction
ISO-Log
L
LAN
Language
Level indicator
Leveling
Leveling assistant
Leveling foot
Leveling the balance
Location
M
Main activity configuration
Main navigation
Materials
Measuring time
Memory stick
MinWeigh
MT-SICS
MT-SICS commands
N
Navigation
Net
Network devices
O
Operating temperature
Operation keys
Overview
144 Index
P
Password
User password
PC software
PC-Direct
Percent weighing
Performing a dynamic weighing
Performing a simple weighing
Performing formula
Power supply
Print report
Printer
Protection and standards
Publishing
R
Reference piece number
Reset balance
RS232C interface
S
Safety information
Signal words
Warning symbols
Sample ID
Selecting the location
Setting FACT
Settings/preferences
Short balance information
Signal words
Simple weighing
Sleep mode
Sound
Statistics
Statistics configuration
Status icons
Status information field
Status messages
Switching off
Switching the weighing unit
Symbols and conventions
System settings
T
Taring
Technical data
Precision and Analytical Balances
Technical data dimensions
Threshold
Time
Totaling
Touch beep
Transmit data
Transport over short distances
Transporting the balance
U
USB
USB cable
USB device
USB device interface
USB driver installing
USB host
USB Host interface
USB keyboard
User interface
User management
Users
V
Value bar
W
Warm-up time
Warning symbols
Weighing
Weighing information bar
Weighing mode
Weighing unit
Weighing value field
Weighing-in aid
Wi-Fi
Wireless dongle
WLAN
Work title bar
Workflow feedback beep
Z
Zero setting
Precision and Analytical Balances Index 145
146 Index Precision and Analytical Balances
30203950F 8/20/2019 6:37 AM - Schema ST4 PDF engine - Layout by Victor Mahler
Mettler-Toledo GmbH
Im Langacher 44
8606 Greifensee, Switzerland www.mt.com/contact
Subject to technical changes.
© Mettler-Toledo GmbH 08/2019
30203950F en www.mt.com/balances
For more information
30203950
Advertisement