Keysight M9180 Series PXI
6½ Digit Digital Multimeters
M9181A, M9182A, M9183A
Startup Guide
2
Notices
Copyright Notice
© Keysight Technologies 2011 - 2021
No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Manual Part Number
M9182-90001
Edition
Edition 5, July 2021
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from the
Web. Go to http://www.keysight.com/ go/conformity . You can then search by product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly,
Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License
Agreement (EULA), a copy of which can be found at http://www.keysight.com/ find/sweula . The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight:
(1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DIS-
CLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMA-
TION CONTAINED HEREIN, INCLUD-
ING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MER-
CHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSE-
QUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CON-
TAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WAR-
RANTY TERMS COVERING THE MATE-
RIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death.
Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Keysight M9180 Series Startup Guide
Trademarks
Microsoft ® and Windows ® are U.S. registered trademarks of Microsoft
Corporation.
MATLAB is a U.S registered Trademark of The MathWorks, Inc.
PXI is a registered trademark of the PXI Systems Alliance.
Visual Studio is a registered trademark of Microsoft Corporation in the United
States.
LabVIEW is a registered trademark of National Instruments.
Keysight M9180 Series Startup Guide 3
Safety Symbols
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Alternating current (AC)
Earth (ground) terminal
Direct current (DC)
Indicates that antistatic precautions should be taken.
Frame or chassis (ground) terminal
Standby Power. Unit is not completely disconnected from AC mains when power switch is in standby position
4 Keysight M9180 Series Startup Guide
Safety Considerations
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these requirements.
General
Do not use this product in any manner not specified by the manufacturer. The protective features of this product must not be impaired if it is used in a manner specified in the operation instructions.
Before applying power
Verify that all safety precautions are taken. Make all connections to the unit before
applying power. Note the instrument's external markings described under Safety
Ground the instrument
PXI chassis are provided with a grounding-type power plug. The instrument chassis and cover must be connected to an electrical ground to minimize shock hazard. The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Do not operate in an explosive atmosphere or wet environments
Do not operate the Keysight module/chassis in the presence of flammable gases or fumes, vapor or wet environments.
Keysight M9180 Series Startup Guide 5
6
Do not operate near flammable liquids
Do not operate the Keysight module/chassis in the presence of flammable liquids or near containers of such liquids.
Do not remove instrument cover
Only qualified, service-trained personnel who are aware of the hazards involved should remove instrument covers. Always disconnect the power cable and any external circuits before removing the instrument cover.
Cleaning
Clean the outside of the Keysight module, chassis, or accessory with a soft, lint-free, slightly dampened cloth. Do not use detergent or chemical solvents.
Keep away from live circuits
Operating personnel must not remove equipment covers or shields. Procedures involving the removal of covers and shields are for use by service-trained personnel only. Under certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous electrical shock, DO NOT perform procedures involving cover or shield removal unless you are qualified to do so.
Do not operate damaged equipment
Whenever it is possible that the safety protection features built into this product have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to an
Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
Keysight M9180 Series Startup Guide
Do not block the primary disconnect
The primary disconnect device is the appliance connector/power cord when a chassis used by itself, but when installed into a rack or system the disconnect may be impaired and must be considered part of the installation.
Do not modify the instrument
Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an Keysight Sales and Service Office to ensure that safety features are maintained.
In case of damage
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
WARNING
– IEC Measurement Category II. The HI and LO input terminals may be connected to mains in IEC Category II installations for line voltages up to
300 VAC rms.
To avoid the danger of electric shock, do not connect the inputs to mains for line voltages above 300 VAC rms limited to 240 VAC rms
. The M9181A DMM is
. See below for further information.
– The DMM comes installed with four shields (bottom, top and two edge strips). These shields must be in place for performance and safety reasons. Removal or improper assembly of these shields can result in lethal voltages occurring within your PC. Be sure to check your installation before closing the cover on your chassis.
Keysight M9180 Series Startup Guide 7
8
WARNING
– During and after installing your DMM, verify that no loose wires or ribbon cables infringe upon any of the internal circuits of the DMM, as this may apply measurement voltages to your chassis, causing electrocution and/ or damage to your chassis!
– To avoid shock hazard, install the DMM only into a chassis that has its power connector connected to a power receptacle with an earth safety ground.
– When making any measurements above 50 VDC or 40 VAC rms safety test leads, such as Keysight test leads.
, only use
– Measuring voltages greater than 30 VAC rms
, 42 V pk
or 60 VDC is always hazardous. ALL multimeter terminals must be considered as hazardous whenever inputs greater than 30 VAC rms
, 42 V pk
or 60 VDC are connected to ANY input terminal. Regard all terminals as being at the same potential as the highest voltage applied to any terminal.
CAUTION
Use with caution when working with voltages above 30 VAC rms
60 VDC. These voltages pose a shock hazard.
, 42 V pk
or
Keysight M9180 Series Startup Guide
Input terminal measurement limits
Measurement limits are defined for the input terminals:
Main input (HI and LO) terminals: The HI and LO input terminals are used for voltage, resistance, frequency (period), capacitance, and diode test measurements. Two measurement limits are defined for these terminals:
– HI to LO measurement limit.
The measurement limit from HI to LO (Input terminals) is 240 VDC or
240 VAC rms
(M9181A), and 300 VDC or 300 VAC rms
(M9182A/83A), which is also the maximum voltage measurement. This limit can also be expressed as
340 Vpk maximum (M9181A) and 425 Vpk maximum (M9182A/83A).
– LO to ground measurement limit.
The LO input terminal can safely "float" a maximum of 240 VDC or 240 VAC rms
(M9181A), and 330 VDC or 250 VAC rms
(M9182A/83A) relative to ground, where ground is defined as the Protective Earth Conductor in the AC mains power cord connected to the instrument. As implied by the above limits, the measurement limit for the HI input terminal is as shown in table below.
Measurement limit
HI to LO
LO to Ground
HI to Ground
M9181A
340 V pk
240 V pk
580 V pk
M9182A/83A
425 V pk
354 V pk
779 V pk
CAUTION
Current input terminal: The current input ("I") terminal has a measurement limit of
2.4 A (DC or AC) between the HI terminal and the LO input terminal. Note that the current input terminals will always be at approximately the same voltage as the main input LO terminal.
Sense terminal measurement limits: The HI and LO sense terminals are used for four-wire resistance and temperature measurements. The measurement limit is 5
V pk
for all the terminal pairings: LO sense to LO input, HI sense to LO input, and HI sense to LO sense.
Do not connect the V HI and V LO or I HI and I LO terminals to anything when the measurement function does not utilize these terminals. Disconnect these terminals with an external device.
Keysight M9180 Series Startup Guide 9
10
IEC Measurement Category II overvoltage protection
To protect against the danger of electric shock, the Keysight M9180 Series digital multimeters (DMMs) provide overvoltage protection for line-voltage mains connections meeting both of the following conditions:
The HI and LO input terminals are connected to the mains under Measurement
Category II conditions, defined below, and
The mains are limited to a maximum line voltage of 300 VAC rms the M9181A).
(240 VAC rms for
IEC Measurement Category II includes electrical devices connected to mains at an outlet on a branch circuit. Such devices include most small appliances, test equipment, and other devices that plug into a branch outlet or socket. The M9180
Series DMMs may be used to make measurements with the HI and LO inputs connected to mains in such devices, or to the branch outlet itself (up to 300
VAC rms
, or 240 VAC rms
for the M9181A). However, the DMMs may not be used with their HI and LO inputs connected to mains in permanently installed electrical devices such as the main circuit-breaker panel, sub-panel disconnect boxes, or permanently wired motors. Such devices and circuits are subject to overvoltages that may exceed the protection limits of the M9180 Series.
NOTE
The M9182A and M9183 can measure voltages up to 330 VAC rms are isolated from mains (CAT I).
in circuits that
Keysight M9180 Series Startup Guide
Environmental Conditions
This instrument is designed for indoor use and in an area with low condensation.
The table below shows the general environmental requirements for this instrument.
Environmental conditions
Operating temperature
Operating humidity
Storage temperature
Altitude
Measurement category
Pollution degree
Requirements
–10 °C to 55 °C
Up to 80% RH at 40 °C (non-condensing)
Full accuracy up to 40% RH for 41 °C to 55 °C
(non-condensing)
–40 °C to 85 °C up to 2000 m
M9181A: CAT II 240 V
M9182/M9183A: CAT II 300 V
Pollution Degree 2
Product Regulatory and Compliance
The M9180 Series PXI multimeter complies with safety and EMC requirements.
Refer to Declaration of Conformity at http://www.keysight.com/go/conformity for the latest revision.
Keysight M9180 Series Startup Guide 11
Regulatory Markings
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this ISM device complies with the
Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
RECOGNIZED COMPONENT
Conforms to UL STD 61010-1:
Certified to CSA STD C22.2. No.
61010-1
ISM
This is the symbol for an Industrial,
Scientific, and Medical Group 1 Class A product.
The RCM mark is a registered trademark of the Australian
Communications and Media Authority.
12 Keysight M9180 Series Startup Guide
Waste Electrical and Electronic Equipment (WEEE) Directive
This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Product category
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
– www.keysight.com/find/pxi-dmm
(product-specific information and support, software and documentation updates)
– www.keysight.com/find/assist
(worldwide contact information for repair and service)
– www.keysight.com/find/tips
(information on preventing damage to your Keysight equipment)
Keysight M9180 Series Startup Guide 13
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight M9180 Series Startup Guide
Table of Contents
Before applying power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Ground the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Do not operate in an explosive atmosphere or wet environments . . .5
Do not operate near flammable liquids . . . . . . . . . . . . . . . . . . . . . . . .6
Do not remove instrument cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Keep away from live circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Do not operate damaged equipment . . . . . . . . . . . . . . . . . . . . . . . . .6
Do not block the primary disconnect . . . . . . . . . . . . . . . . . . . . . . . . .7
Do not modify the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
In case of damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Input terminal measurement limits . . . . . . . . . . . . . . . . . . . . . . . . . . .9
IEC Measurement Category II overvoltage protection . . . . . . . . . . .10
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Product Regulatory and Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . .13
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Related documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Step 1: Unpack and Inspect the Module . . . . . . . . . . . . . . . . . . . . . . . .18
Step 2: Verify Shipment Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Step 3: Install the Software on the Controller . . . . . . . . . . . . . . . . . . . .21
System requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Power the controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Keysight M9180 Series Startup Guide 15
Install Keysight IO Libraries Suite . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Install instrument drivers and Soft Front Panel . . . . . . . . . . . . . . . . 24
Step 4: Connect the PC to the PXI Chassis . . . . . . . . . . . . . . . . . . . . . . 26
Using a remote PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Using an embedded computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Step 5: Install the DMM Modules in the PXI Chassis . . . . . . . . . . . . . . 28
Step 6: Verify Operation of the DMM . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Run Keysight IO Libraries Connection Expert . . . . . . . . . . . . . . . . . . 30
Module Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
16 Keysight M9180 Series Startup Guide
Introduction
Introduction
The Keysight M9180 Series digital multimeters (DMM) deliver high-performance measurement with fast, easy installation and configuration via the Keysight
Connection Expert and a soft front panel (SFP).
Keysight PXI 6½ digit DMM Modules
– M9181A, PXI 6½ digit Digital Multimeter, basic features
– M9182A, PXI 6½ digit Digital Multimeter, high performance
– M9183A, PXI 6½ digit Digital Multimeter, enhanced high performance
Keysight also supplies software drivers that allow you to support the modules in many popular PXI chassis and programming environments. The soft front panel allows you to exercise the DMMs.
Related documentation
For the latest specifications, documentation, software, and example programs, please see the Keysight web site at: www.keysight.com/find/pxi-dmm .
Keysight M9180 Series Startup Guide 17
Step 1: Unpack and Inspect the Module
Step 1: Unpack and Inspect the Module
CAUTION
Keysight’s PXI DMMs are shipped in materials which prevent static electricity damage. The modules should only be removed from the packaging in an anti-static area ensuring that correct anti-static precautions are taken. Store all modules in anti-static envelopes when not in use.
CAUTION
ESD
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe work station.
Purchase acceptable ESD accessories from your local supplier.
– Conductive table-mat and wrist-strap combination.
– Conductive floor-mat and heel-strap combination.
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone. To ensure user safety, the static-safe accessories must provide at least 1 M Ω of isolation from ground.
Inspect for damage
After unpacking the DMM, carefully inspect it for any shipping damage. Report any damage to the shipping agent immediately, as such damage is not covered by the warranty.
To avoid damage when handling a module; do not touch exposed connector pins.
NOTE
Information on preventing damage to your Keysight equipment can be found at www.keysight.com/find/tips .
18 Keysight M9180 Series Startup Guide
NOTE
Step 1: Unpack and Inspect the Module
Return the module for service
Should it become necessary to return an Keysight DMM for repair or service, follow the steps below:
1 Review the warranty information shipped with your product.
2 Contact Keysightto obtain a Return Material Authorization (RMA) and return address. For assistance finding Keysight contact information go to www.keysight.com/find/assist (worldwide contact information for repair and service) or refer to the Support information on the product web page at: www.keysight.com/find/pxi-dmm .
3 Write the following information on a tag and attach it to the malfunctioning equipment.
– Name and address of owner. A Post Office box is not acceptable as a return address.
– Product model number (for example, M9182A)
– Product serial number (for example, MYXXXXXXXX). The serial number label is located on the side of the module.
– A description of failure or service required.
4 Carefully pack the module in its original ESD bag and packing carton. If the original carton is not available, use bubble wrap or packing peanuts and place the instrument in a sealed container and mark the container “FRAGILE.”
5 On the shipping label, write ATTENTION REPAIR DEPARTMENT and the RMA number.
If any correspondence is required, refer to the product by serial number and model number.
Keysight M9180 Series Startup Guide 19
Step 2: Verify Shipment Contents
Step 2: Verify Shipment Contents
Your shipment should have included the following:
– The Keysight PXI DMM module that you ordered.
– Any other accessories that you ordered (cables, connectors, etc.).
20 Keysight M9180 Series Startup Guide
Step 3: Install the Software on the Controller
Step 3: Install the Software on the Controller
The following illustration shows typical installations.
System Controller
Laptop PC
M9045A
PCIe Laptop
Adapter
System Interface:
M9021A
System Interface
Module
System Chassis:
M9018B
18 Slot PXIe Chassis
Desktop PC/
Rackmount PC Y1200A
PCIe Cable
PXIe Embedded
System Controller
Y1202A
PCIe Cable
M9047A
PCIe Desktop
Adapter
NOTE
Do not install the PXI modules in the PXI chassis yet! You must install the software prior to installing the modules in the chassis so that Keysight IO
Libraries Connection Expert finds them.
Keysight M9180 Series Startup Guide 21
Step 3: Install the Software on the Controller
System requirements
The following table lists the minimum system requirements for Keysight IO
Libraries. In general, any x86 or x64 (except Itanium) should work but there may be a significant decrease in performance.
Operating
System
Processor
Speed
Windows 7 SP1 (32- and 64-bit)
1 GHz 32-bit (x86),
1 GHz 64-bit (x64), no support for Itanium64
Windows 8 and Windows 8.1 (32- and 64-bit)
1 GHz 32-bit (x86),
1 GHz 64-bit (x64), no support for Itanium64 or
Windows 8 RT
Windows 10 (32- and 64-bit)
1 GHz 32-bit (x86),
1 GHz 64-bit (x64), no support for Itanium64 or
Windows 8 RT
Available memory
1 GB minimum 1 GB minimum 1 GB minimum
Available hard disk space [a]
1.5 GB available hard disk space, includes:
– 1 GB available for Microsoft.NET
Framework 4.6
– 100MB for Keysight IO Libraries
Suite
2.5 GB available hard disk space, includes:
– 2 GB available for Microsoft.NET
Framework 3.5 and 4.6
– 100MB for Keysight IO Libraries
Suite
2.5 GB available hard disk space, includes:
– 2 GB available for Microsoft.NET
Framework 3.5 and 4.6
– 100MB for Keysight IO Libraries
Suite
Minimum display
1024x768, 96 or 120 DPI 1024x768, 96 or 120 DPI 1024x768, 96 or 120 DPI
[a] Because of the installation procedure, less memory may be required for operation than is required for installation.
22 Keysight M9180 Series Startup Guide
Step 3: Install the Software on the Controller
Power the controller
– If you are using a desktop, laptop, or rackmount PC as a controller:
1 Install any peripheral devices in the PC (e.g., PCIe Interface adapters, etc.).
Follow the manufacturers instructions.
2 Power up the computer.
3 Choose the automatic option for any “Found Hardware” dialogs
– If you are using a PXI embedded computer as the controller:
1 Install the embedded computer module into the PXI chassis following the manufacturer’s instructions.
2 Connect peripherals such as a mouse, keyboard, monitor, CD-ROM drive, etc.
3 Power up the chassis.
4 Choose the automatic option for any “Found Hardware” dialogs
Install Keysight IO Libraries Suite
Install Keysight IO Library Suite (IOLS), which includes the Keysight Connection
Expert, the IVI Shared Components, and the VISA Shared Components.
NOTE
Keysight IO Libraries must be installed prior to installing and running any other software and prior to powering up the chassis.
1 Download and install IO Libraries software from www.keysight.com/find/ iosuite .
2 Follow the installation instructions that came with the Keysight IO Libraries
Suite.
Keysight M9180 Series Startup Guide 23
Step 3: Install the Software on the Controller
NOTE
If the IVI Shared Components and VISA Shared Components are not already installed on your PC, Keysight IO Libraries Suite installs them in the standard, default locations. If the IVI Shared Components and VISA Shared Components are already installed, the Keysight IO Libraries Suite installer will upgrade them to the latest version, using the same installation location used by the older version. If this is a first-time installation, you will be able to select installation locations for these components by choosing a Custom Installation.
Install instrument drivers and Soft Front Panel
1 Download and install the instrument drivers from wwww.keysight.com/find/ pxi-dmm .
2 Follow the installer prompts. Accept all of the default directories specified during installation if prompted.
3 After the Welcome screen, you will be prompted three times for license agreements. The software installer installs the following drivers:
– Keysight Modular Software License Agreement . The installer program installs the Soft Front Panel (SFP) application and other object code to connect to the switches. Accept the license terms and click Next .
– Keysight IVI Driver Source Code License Agreement . The Interchangeable
Virtual Instrument (IVI) driver is available for programming the Keysight switch modules using Microsoft ® development environments (e.g., Visual
Studio ® , C, C++, C#, Visual Basic), Keysight VEE, MATLAB ® , or National
Instruments ® LabVIEW TM . Accept the license terms and click Next .
– Keysight Software License Agreement for drivers for use with LabVIEW
Software . The LAbVIEW driver provides access to the functionality of the switches through LabVIEW VIs. This driver works in National Instruments
LabVIEW development environments. Before this driver can be installed, your computer must already have the IVI Shared Components installed.
Accept the license terms and click Next .
NOTE
Installing Keysight IO Libraries also installs the IVI Shared Components.
24 Keysight M9180 Series Startup Guide
Step 3: Install the Software on the Controller
4 Next, the installer indicates the LabVIEW installations found on your host computer. Click Next .
5 After accepting the licenses, the driver software is ready to be installed on your host computer. Click Install .
6 The installation program will take several minutes to finish the installation; do not cancel the installation process. When the installation wizard is finished, you will be prompted to reboot your host computer. Do not reboot the host computer at this time! Select “ No, I will restart my computer later .” Power down the host computer.
Keysight M9180 Series Startup Guide 25
Step 4: Connect the PC to the PXI Chassis
Step 4: Connect the PC to the PXI Chassis
NOTE
To ensure proper operation and the PC’s ability to enumerate all of the PXI modules, you must use an approved embedded PC, desktop PC, or laptop PC along with approved PCIe adapter and cable.
Using a remote PC
Refer to the following figure. Make certain that both the PC and the PXI chassis are turned off. If you are using the M9018A 18 Slot PXIe chassis, install the
Keysight M9021 System Interface Module in the chassis.
– If you are using a desktop or rackmount PC, install the M9047 PCIe Desktop
Adapter in the PC. With an Keysight Y1202 cable, connect the adapter to the
System Interface module.
– If you are using a laptop PC then install the M9045 PCIe ExpressCard Adapter in the laptop PC. With an Keysight Y1200 cable, connect the adapter to the
System Interface module or the embedded computer on the PXI chassis.
26 Keysight M9180 Series Startup Guide
System Controller
Laptop PC
Desktop PC/
Rackmount PC
PXIe Embedded
System Controller
M9045A
PCIe Laptop
Adapter
System Interface:
M9021A
System Interface
Module
Y1200A
PCIe Cable
Y1202A
PCIe Cable
M9047A
PCIe Desktop
Adapter
Step 4: Connect the PC to the PXI Chassis
System Chassis:
M9018B
18 Slot PXIe Chassis
Using an embedded computer
Refer to the figure above. If you are using an embedded controller in the PXI chassis, you should have installed it prior to installing the Keysight IO Libraries and instrument drivers. No cables or other adapters are required. After
installation, proceed to Step 5: Install the DMM Modules in the PXI Chassis .
Keysight M9180 Series Startup Guide 27
Step 5: Install the DMM Modules in the PXI Chassis
Step 5: Install the DMM Modules in the PXI Chassis
CAUTION
– PXI hardware does not support “hot-swap” capabilities (changing modules while power is applied to the chassis).
– Before installing Keysight PXI Modules into the chassis, the chassis must be powered off and unplugged to prevent damage to the PXI module.
NOTE
These modules can be used in a chassis with a cPCI, PXI-1, or PXIh chassis peripheral slot.
The module can be installed in any standard PXI slot marked with a peripheral slot compatibility image (a circle containing the slot number).
The module can also be installed in any hybrid PXI slot marked with a peripheral slot compatibility image (the letter "H" and a solid circle containing the slot number).
1 Make sure the PXI chassis power is turned off.
2 If the chassis has multiple fan speed settings, ensure that the fans are set to automatic. Do not set the fan speed to low or turn them off.
3 Position the chassis so that there is ample space between the chassis fan intake and exhaust vents. Blockage by walls or obstructions affects the air flow needed for cooling. (Refer to the chassis documentation for more information about cooling).
4 Holding the PXI module by the injector/ejector handle, slide it into an available
PXI (or hybrid) slot, as shown in the following figure.
– Install the module into the PXI slot of the chassis by placing the module card edges into the front module guides (top and bottom).
– Slide the module to the rear of the chassis and assure that the injector/ ejector handle is pushed down in the unlatched (downward) position.
28 Keysight M9180 Series Startup Guide
Step 5: Install the DMM Modules in the PXI Chassis
– Slide the module completely into the chassis. When you begin to feel resistance, push up on the injector/ejector handle to fully seat the module into the chassis.
5 Latch the module by pulling up on the injector/ejector handle and secure the front panel to the chassis using the module front-panel mounting screws.
6 Tighten the screws on the module (or remote controller) front panel.
Performance may suffer if the screws are not securely tightened.
7 Install all chassis covers, filler panels, and air scoops after installing the module. Missing filler panels may disrupt necessary air circulation in the chassis.
8 If you are using a remote controller, connect the Cable Interface Card in the chassis to host computer.
9 Plug in and power up the PXI chassis. Verify that the chassis fans are operating and free of obstructions that may restrict airflow.
10 If you are using a remote controller (rackmount, desktop, or laptop PC), reboot the controller.
Keysight M9180 Series Startup Guide 29
Step 6: Verify Operation of the DMM
Step 6: Verify Operation of the DMM
Run Keysight IO Libraries Connection Expert
If Keysight Connection Expert is already running on the controller, click the
Refresh All button to identify any hardware you have just installed or re-connected.
NOTE
If Connection Expert is not already running, run it now to verify your I/O configuration: In the Windows Notification area, click the IO icon ( ), then click
Keysight Connection Expert .
Locate your interfaces and instruments in the Keysight Connection Expert
Explorer Pane. Click on them to see their properties (displayed in the right-hand pane). Observe their state, also displayed in the right-hand pane -- if communication to the interface or instrument has been successfully established, it will be in the Verified state, denoted by a green check mark.The following illustration shows the Keysight Connection Expert.
30 Keysight M9180 Series Startup Guide
Step 6: Verify Operation of the DMM
You can use the Soft Front Panel (SFP) software to open and close the relays and verify operation of the PXI DMM modules. The Soft Front Panel Software was installed as part of the Software installation process.
Refer to the SFP help file for specific detailed information on using the SFP. The following graphic shows an example of the SFP for the DMM modules.
Verify operation
To verify basic operational readiness, use the Soft Front Panel software to configure and take a few basic measurements. For example, you could apply a short across the input terminals and take a DCV measurement or 2W Ω measurement. Both measurements should report values near zero. Module specifications are guaranteed by design.
Conduct self test
Use the Utility menu on the Soft Front Panel software to run a module self test.
Keysight M9180 Series Startup Guide 31
Step 6: Verify Operation of the DMM
32 Keysight M9180 Series Startup Guide
Module Functional Description
Module Functional Description
The Keysight M9180 Series DMM modules conform to the single slot, 3U height
(128 mm) Eurocard standard. These modules meet PXI Specification 2.2. Local bus, trigger bus, and star trigger are not included. The modules also include:
– CPCI Ejector Handle
– The front panel secures to the chassis by two M2.5 × 6 mm pan-head
Posi-drive screws.
There are no user serviceable or field replaceable components on these modules.
Keysight M9180 Series Startup Guide 33
Module Functional Description
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
34 Keysight M9180 Series Startup Guide
This information is subject to change without notice. Always refer to the
Keysight website for the latest revision.
© Keysight Technologies 2011 - 2021
Edition 5, July 2021
Printed in Malaysia
M9182-90001 www.keysight.com