Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Microsoft
is a U.S. registered trademark of Microsoft Corp.
© Copyright 1999 - 2001, 2003 Agilent Technologies, Inc.
Introduction
This procedure can be used to retrofit Low Emissions, Option 060, into an E4401B, E4402B, E4404B, E4405B, or E4407B spectrum analyzer. Separate kits are provided for each model number. Several kits contain two or more versions of a part, such as a nameplate or connector label, so that each of these kits can support all options the analyzer may already have. As a result, not all parts in each kit will be used.
Products Affected: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E4401B
E4402B
E4404B
E4405B
E4407B
Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . US00000000/US99999999
MY00000000/MY99999999
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . . . . . . . . (X) Agilent Technologies Service Center
(X) Personnel Qualified by Agilent
( ) Customer
Estimated Installation Time:. . . . . . . . . . . . . . . . . . . . . . . 2.0 Hours
Estimated Verification Time: . . . . . . . . . . . . . . . . . . . . . . 0.75 Hours
Installation Note E4401-90456 3
E4401B Installation Kit Parts List
Qty
9
1
1
1
1
11
1
1
2
1
1
1
Description
Chassis Cover
Dress Cover
Front Frame Assembly, Opt 060
Cap, Connector Cover
Plug, TG
Nameplate, E4401B, 50 Ohm
Connector Label, 50 Ohm
Screw, 3.0 x 8 mm, panhead torx
Screw, 3.0 x 10 mm, flathead torx
Option 060 Retrofit Kit Installation Note
Option Upgrade Entitlement Certificate
Serial Number Label
Item Number
E4401-00067
E4401-00073
E4401-60259
E4401-40006
E4401-40020
E4401-80027
E4401-80003
0515-0372
0515-2600
This note
---
9320-6050
4 Installation Note E4401-90456
E4402B Installation Kit Parts List
Qty
1
1
1
1
1
1
1
2
1
1
1
1
1
1
1
11
9
Description
Chassis Cover
Dress Cover
Front Frame Assembly, Opt 060
Cap, Connector Cover
Plug, TG
Conductive Washer
Wavy Washer, 0.641 ID
Type-N Spacer
Nameplate, E4402B, Std
Nameplate, E4402B, Opt UKB
Connector Label, Std
Connector Label, Opt UKB
Screw, 3.0 x 8 mm, panhead torx
Screw, 3.0 x 10 mm, flathead torx
Option 060 Retrofit Kit Installation Note
Option Upgrade Entitlement Certificate
Serial Number Label
Item Number
E4401-00067
E4401-00073
E4401-60259
E4401-40006
E4401-40020
E4402-20010
3050-2127
5022-6164
E4402-80002
E4403-80016
E4403-80002
E4403-80015
0515-0372
0515-2600
This note
---
9320-6050
Installation Note E4401-90456 5
E4404B Installation Kit Parts List
Qty
1
1
1
1
1
1
1
2
1
1
1
1
1
1
1
11
9
Description
Chassis Cover
Dress Cover
Front Frame Assembly, Opt 060
Cap, Connector Cover
Plug, TG
Conductive Washer
Wavy Washer, 0.641 ID
Type-N Spacer
Nameplate, E4404B, Std
Nameplate, E4404B, Opt UKB
Connector Label, Std
Connector Label, Opt UKB
Screw, 3.0 x 8 mm, panhead torx
Screw, 3.0 x 10 mm, flathead torx
Option 060 Retrofit Kit Installation Note
Option Upgrade Entitlement Certificate
Serial Number Label
Item Number
E4401-00067
E4401-00073
E4401-60259
E4401-40006
E4401-40020
E4402-20010
3050-2127
5022-6164
E4404-80004
E4404-80008
E4404-80001
E4404-80007
0515-0372
0515-2600
This note
---
9320-6050
6 Installation Note E4401-90456
E4405B Installation Kit Parts List
Qty
1
1
1
1
1
1
1
2
1
1
1
1
1
1
1
11
9
Description
Chassis Cover
Dress Cover
Front Frame Assembly, Opt 060
Cap, Connector Cover
Plug, TG
Conductive Washer
Wavy Washer, 0.641 ID
Type-N Spacer
Nameplate, E4405B, Std
Nameplate, E4405B, Opt UKB
Connector Label, Std
Connector Label, Opt UKB
Screw, 3.0 x 8 mm, panhead torx
Screw, 3.0 x 10 mm, flathead torx
Option 060 Retrofit Kit Installation Note
Option Upgrade Entitlement Certificate
Serial Number Label
Item Number
E4401-00067
E4401-00073
E4401-60259
E4401-40006
E4401-40020
E4402-20010
3050-2127
5022-6164
E4405-80003
E4405-80006
E4405-80001
E4405-80005
0515-0372
0515-2600
This note
---
9320-6050
Installation Note E4401-90456 7
E4407B Installation Kit Parts List
Qty
1
1
1
1
1
1
1
2
1
1
1
1
1
11
9
1
1
1
1
Description
Chassis Cover
Dress Cover
Front Frame Assembly, Opt 060
Cap, Connector Cover
Plug, TG
Conductive Washer
Wavy Washer, 0.641 ID
Type-N Spacer
Nameplate, E4407B, Std
Nameplate, E4407B, Opt UKB
Connector Label, Std
Connector Label, Opt AYZ
Connector Label, Opt UKB
Connector Label, Opt UKB and AYZ
Screw, 3.0 x 8 mm, panhead torx
Screw, 3.0 x 10 mm, flathead torx
Option 060 Retrofit Kit Installation Note
Option Upgrade Entitlement Certificate
Serial Number Label
Item Number
E4401-00067
E4401-00073
E4401-60259
E4401-40006
E4401-40020
E4402-20010
3050-2127
5022-6164
E4407-80003
E4407-80006
E4407-80001
E4407-80005
E4407-80007
E4407-80008
0515-0372
0515-2600
This note
---
9320-6050
8 Installation Note E4401-90456
Tools Required
T-10 TORX screwdriver
T-15 TORX screwdriver
Flat-blade screwdriver
Heat gun or >1200 W blow dryer
Razor blade
Torque Settings
To avoid potential RFI leakage and prevent connector damage, tighten to the following torque limits:
Torque
Item
SMA Connectors
3-mm, T-10 TORX screws
4-mm, T-15 TORX screws
In-lb
8.5
9
21
N-cm
95
102
236
Installation Note E4401-90456 9
Procedure
WARNING Before you disassemble the instrument, turn the power switch to Standby and unplug the instrument. Failure to unplug the instrument can result in personal injury.
CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
Removing the Dress Cover
1. Referring to Figure 1 , disconnect the analyzer from ac power
(1) .
2. Remove any adapters or cables (2) connected to the front frame.
3. Carefully place the analyzer on the work surface with the front frame (3) facing down.
4. Position the handle (4) as shown.
5. Remove the four screws (5) that hold the rear frame and dress cover in place.
6. Remove the rear frame (6) .
7. Pull the dress cover (7) off towards the rear of the analyzer.
10 Installation Note E4401-90456
Figure 1 Dress Cover and Rear Frame Removal
Installation Note E4401-90456 11
Removing the Chassis Cover
1. Lay the analyzer flat as shown in Figure 2 .
2. Remove the 17 screws (2) and (3) attaching the chassis cover to the chassis. Note that the number of screws attaching the chassis cover may vary with option mixes.
3. Remove the chassis cover (1) from the chassis.
Figure 2 Chassis Cover Removal
12 Installation Note E4401-90456
A1 Front Frame Assembly
CAUTION Use ESD precautions when performing this replacement procedure.
Extending the Front Frame
The A1 front frame assembly can be extended from the instrument without detaching any connections.
1. With the instrument still on its face, remove the two screws (1) on the bottom side of the instrument, that secure the front frame to the RF assembly.
Figure 3 A1 Front Frame Assembly Removal
2. Refer to
Figure 3 . Place the instrument with the top side facing up and remove the remaining two screws
(2) that secure the front frame subpanel to the chassis.
3. Slide the front frame forward until it catches on the tabs on the sides of the chassis.
Installation Note E4401-90456 13
Removing the Front Frame
procedure, then continue with the following steps:
1. Refer to
Figure 4 . Disconnect the ribbon cable
(1) from the A1A1 front panel interface board. Analyzers with Option
B7B will have a second ribbon cable connecting to A1A1. Disconnect the second ribbon cable if present.
Figure 4 Front Frame Ribbon Cable
2. Carefully pull the sides of the front frame subpanel away from the chassis and over the tabs on the chassis.
3. Slide the front frame forward to disengage from the chassis assembly.
14 Installation Note E4401-90456
Removing the A1A2 Display
Refer to
Figure 5 for the following procedure.
Figure 5 A1A2 Display Removal
1. Place the front frame assembly face down on the work surface.
2. Disconnect the two 2-wire backlight cables (1) from the front panel interface inverter board.
3. Disconnect W3 ribbon cable (2) from the front panel interface board.
CAUTION The W3 display ribbon cable connector (2) is delicate. With a small screwdriver or similar tool, gently push the lock tabs out from the back of the connector. Excessive force on the locking tab can break the retaining clips, and if broken, board replacement will be necessary.
4. Remove the four screws (3) securing the display (in a bracket) (4) to the front frame.
5. Carefully lift the display out of the front frame assembly.
Installation Note E4401-90456 15
Removing the A1A1 Front Panel Interface Board
Figure 6 A1A2 Display Removal
1. Remove the volume knob and the RPG knob.
TIP You may need to exert considerable force in order to remove the RPG knob. It may be necessary to pry
it off using a screwdriver as shown in Figure 6 .
2.
Refer to
. Remove the four screws (10) that secure the A1A1 front panel interface board (11) to the front frame.
3. Remove the front panel interface board from the front frame assembly.
NOTE There is a water-seal gasket placed around the volume control shaft that will need to be repositioned during the replacement procedure.
16 Installation Note E4401-90456
Removing the Keypad Flex Circuit
Figure 7 Front Frame Assembly Parts
1. Refer to
Figure 7 . Remove the support bracket
(3) .
2. Remove the 6 screws (8) securing the subpanel assembly (9) to the front frame (6) .
3. Remove the subpanel assembly (9) .
4. Lift out the keypad/flex circuit (4) and the main keypad (7) . Be careful to avoid touching the keypad contacts with your fingers.
Installation Note E4401-90456 17
Attaching the Nameplate to Front Frame
1. Locate the nameplate in the kit. The nameplate is labelled with the analyzer product number (for example, “E4404B”) and the analyzer’s frequency range. Some kits contain two nameplates, identifying different frequency ranges for the analyzer.
For example, the kit for the E4404B will have one nameplate for the standard instrument where the frequency range is
9 kHz to 6.7 GHz and another nameplate for Option UKB, where the frequency range is 100 Hz to 6.7 GHz. Choose the nameplate that is appropriate for the analyzer being retrofitted.
2.
Lay the new front frame included in the kit on the table with the front facing up. Locate the pocket (2) above the display opening where the nameplate will go. Refer to
3. Clean the nameplate pocket (2) with alcohol.
4. Remove the backing (3) from the nameplate (4) and press the nameplate into the pocket (2) .
Figure 8 Nameplate Replacement
18 Installation Note E4401-90456
Attaching the Connector Label and Connector Covers to Front Frame
1. Locate the connector label in the kit. The connector label has holes to accommodate the various front panel connectors and the volume control. The frequency range of the analyzer appears near the hole next to the INPUT connector. Some kits contain two or more connector labels identifying different frequency ranges and connector configurations. Choose the connector label that is appropriate for the analyzer being retrofitted.
NOTE For the E4407B, there are two connector labels for each frequency range. One label of each pair is for analyzers which also have external mixing, and the other label is for analyzers which do not have external mixing. Be sure to choose the correct label.
2.
Remove the backing (2) off the new label as shown in
3. Align and install the label by placing a finger in the RF OUT and VOLUME holes. Observe the alignment of the INPUT and PROBE POWER holes (3) .
Figure 9 Connector Label Installation
Installation Note E4401-90456 19
4. Install the two connector covers from the front panel by pinching them from the sides (4) and installing them through the front of the assembly. See
TIP
Figure 10
It may be helpful to trim off about 2 mm of the tabs (5) on the connector covers prior to reinstalling them. This will help in aligning the covers. See
Connector Covers
5. Align the connector covers so they open downward as shown.
6. Turn the front frame over so that the inside of the front frame is facing up.
7. If the analyzer is equipped with a tracking generator, a Type-N EMC Seal (included with the front frame) must be installed onto the inside of the front panel behind the RF OUT connector hole. Remove the adhesive backing from the seal and press it into the front frame pocket. The foiled edges must make contact with the top and bottom of the frame pocket.
8. If the analyzer is an E4401B, a Type-N EMC Seal (included with the front frame) must be installed onto the inside of the front panel behind the INPUT connector hole. Remove the adhesive backing from the seal and press it into the front frame pocket. The foiled edges must make contact with the top and bottom of the frame pocket.
20 Installation Note E4401-90456
Replacing the Keypad/Flex Circuit and A1A2 Display
1. Refer to
Figure 11 . Place the main keypad
(7) so it lays flat in the front frame (6) . Be careful to avoid touching the keypad contacts with your fingers.
2. Lay in the flex circuit (4) so it aligns with the pegs on the front frame and lays flat. A small amount of circuit ripple will be unavoidable.
3. Replace the subpanel (9) , and secure with the six screws (8) . Tighten them to 9 inch-pounds.
4. Replace the support bracket (3) , along with the display (2) and secure with the four screws (1) . Tighten them to
9 inch-pounds.
Figure 11 Front Frame Assembly Parts
Installation Note E4401-90456 21
Replacing the A1A1 Front Panel Interface Board
1. Refer to
Figure 11 . Place the front panel interface board
(11) in the correct position in the front frame assembly. Make sure the water seal is in place around the volume control shaft.
2. Replace the four screws that secure the board to the front frame. Tighten them to 9 inch-pounds.
3. Connect the W3 display ribbon cable (2) to the front panel interface board.
TIP An easy way to insert this delicate cable into the connector is to place your finger on the cable, in the center of the LCD display, and gently slide the cable toward the connector until they align. Then, providing guidance with the other hand as necessary, slide the cable until the end slips into the connector.
Ensure the cable end is seated completely and is aligned straight within the connector body.
Continue to hold the cable in place with your finger, and with the other hand gently press the locking tabs
(5)
into place. Refer to Figure 6 .
NOTE If you experience display problems, check this connection.
4. Reconnect the two 2-wire backlight cables (1) to the inverter board, making sure that the cables are dressed away from the openings for the control knobs.
5. Press the volume and RPG knobs onto their control shafts.
Installing the Type-N Gasket (All models except E4401B)
The following procedure applies only to E4402B, E4404B, E4405B, and E4407B. If the analyzer is an E4401B, proceed to
Replace Front Frame.
1. Remove the sleeve and o-ring from the Type N input connector on the A8A1 3 GHz RF assembly. The o-ring will be reused.
2. Locate the wavy washer, conductive washer, and sleeve in the kit.
3. Refer to
Figure 12 . Install the wavy washer
(1) over the Type N connector (2) .
22 Installation Note E4401-90456
Figure 12 Type-N Gasket Installation
4. The new sleeve has a recess on each end. One recess is deep and the other is shallow. Install the o-ring (3) into the deep recess of the sleeve (4) .
5. Install the sleeve (4) over the Type-N connector (2) with the o-ring nearest the wavy washer.
6. Install the conductive washer (5) over the Type-N connector (2) .
Installation Note E4401-90456 23
Replacing the Front Frame Assembly
1. Align the A1 front frame subpanel rails with the chassis as shown in Figure 13 .
Figure 13 Front Frame Assembly Replacement
2. Refer to
. Connect the ribbon cable (1) to the front frame assembly. If Option B7B is installed, there will be two ribbon cables to connect.
24 Installation Note E4401-90456
Figure 14 Front Frame Ribbon Cable
3. Carefully slide the front frame toward the chassis, assuring the ribbon cable(s) are not pinched between assemblies, and the
RF input connector lines up correctly with the opening in the front frame.
4. Refer to
Figure 13 . Replace the screws
(1) that secure the front frame to the chassis. Tighten them to 9 inch-pounds.
Replacing the Chassis Cover
1. A new chassis cover is provided with this kit. However, the original chassis cover removed from the analyzer may be used if it is the newer style. Refer to
. If the holes on the original chassis cover along the rear sides of the chassis cover
(4) have countersinks, the chassis cover is the new style and may be reused. If these holes do not have countersinks, use the new chassis cover provided in the kit.
2. If the original chassis cover is the old style, the serial number label should be removed from it and readhered to the new chassis cover. To remove the serial number label, apply heat to the serial number label using a heat gun or >1200 Watt blow dryer. Use a razor blade to gently peel off the serial number label. While the serial number label is still warm, readhere it in the same location on the new chassis cover.
3. If the serial number label was damaged in the previous step, a replacement serial number label is supplied. Using a typewriter or a fine-point, indelible marking pen, transcribe the analyzer product number and serial number from the old serial number label onto the new label. If space permits, enter all options listed on the original serial number label. Attach the new serial number label onto the new chassis cover in the space marked.
4. Carefully position the chassis cover (1) on the instrument as shown in
Figure 15 , then lower onto the instrument.
5. Using the screws provided in the kit, replace the 17 screws (2) and (3) as indicated on the instructions on the chassis cover. Tighten the screws to 9 inch-pounds.
Installation Note E4401-90456 25
Figure 15 Chassis Cover Replacement
26 Installation Note E4401-90456
Replacing the Dress Cover
1. If the chassis cover supplied in the kit was installed in the previous procedure, it will be necessary to also use the new dress cover supplied in the kit. If the original chassis cover was installed in the previous procedure, the original dress cover may be used. If the new dress cover is used, it will be necessary to remove the handle from the original dress cover and mount it onto the new dress cover. Proceed to
step 5 if the original dress cover will be reused.
2. To remove the handle from the old dress cover, position the handle towards the rear of the dress cover. Pull outward on the handle’s hubs. Rotate the handle a few degrees up and down as you pull outward to release the handle.
3. The new dress cover can accommodate both old and new handles. The handle removed in
step 2 was the old style handle,
so it will be necessary to remove the backing plate (only used with new handles) from the new dress cover to accommodate the old handle. Use a flat-blade screwdriver or retaining ring pliers to remove the retaining ring securing the backing plate to the dress cover.
4. Position the handle with the hubs over the keyholes on the dress cover and the handle pointed toward the rear of the analyzer. Rotate the handle a few degrees up and down until the handle locks into place. Rotate the handle so that it points toward the top of the dress cover.
5. Refer to
Figure 16 . Carefully place the spectrum analyzer on the work surface with the front frame
(3) facing down.
6. Replace the dress cover, matching the grill (8) on the bottom of the dress cover to the bottom of the analyzer.
7. Fit the leading edge of the dress cover completely into the slot (9) on the back of the front frame assembly.
8. Replace the rear frame assembly (6) using the four screws (5) to fasten the rear frame to the analyzer. Tighten them to
21 inch-pounds.
Installation Note E4401-90456 27
Figure 16 Dress Cover, Rear Frame Replacement
28 Installation Note E4401-90456
Firmware Revision Verification
1. Turn on the analyzer and wait for the power-on alignments to complete. Check that there are no error messages displayed.
2. Press System , More , Show System . Make a note of the firmware revision and see if Option B72, Expansion Memory, is listed on the display.
4. If your firmware revision is prior to A.04.00, you must upgrade your analyzer firmware before the license key can be activated.
5. If Option B72, Expansion Memory is installed, you should upgrade the instrument firmware to the latest version using one of the following methods:
• Download the instrument firmware via the world-wide web at http://www.agilent.com/find/esa_firmware. This entire process of upgrading firmware will take approximately 45 minutes.
Note: If you are upgrading an EMC analyzer, look for the link stating “EMC Firmware” at this website.
• Receive the latest firmware upgrade disk set by ordering upgrade Option UE2.
Note: Special instructions in the Agilent ordering configurator should have made you aware of Option UE2 in case access to the internet wasn’t available.
6. If Option B72, Expansion Memory is not installed, you should upgrade the instrument firmware to version A.07.05. This is the latest version for instruments that do not have Option B72 installed.
• Download the instrument firmware via the world-wide web at http://www.agilent.com/find/esa_firmware. Look for the link stating “ESA Firmware Version A.07.05”. This entire process of upgrading firmware will take approximately 45 minutes.
• Order upgrade Option B72, which includes the expansion memory and the latest ESA firmware upgrade disk set.
Installation Note E4401-90456 29
Obtaining a License Key and Activating the Option
The entitlement certificate supplied in this kit allows you to obtain a license key from our Agilent website so you can enable this upgrade option. Once you have retrieved the license key, you can begin the process of activating the option.
1. Press System , More , More , Licensing , Option . When you press Option , the alpha editor will be activated. For instructions on using the alpha editor, refer to the Agilent Technologies ESA Spectrum Analyzers User’s Guide.
2. Use the alpha editor to enter the three-character designation for the option, in this case, “060”, that you want to activate.
3. Press Enter when done.
4. Press License Key . When you press License Key the alpha editor will be activated.
5. Use the alpha editor and the numeric keypad to enter the 12-character license key number (for example:
D7C374DABD5B) for the option that you want to activate.
6. Press Enter when done.
7. Press Activate to turn on the option.
8. If the option and license key entries were correct, a message stating “ Option Activated ” will appear on the analyzer display.
Functional Testing
1. Verify that Option 060 was activated by pressing System , More , Show System . Verify that there is an entry which reads,
“060: Low Emissions.”
2. Perform the following performance verification tests, located in the calibration guide:
• Displayed Average Noise Level
• Residual Responses
3. If you have problems performing either of these tests, get in touch with the nearest Agilent Technologies sales and service office listed in the troubleshooting chapter of the user’s guide.