Keithley 2520 is a Pulsed Laser Diode Test System designed to provide a safe and convenient way to test pulsed laser diodes. It features a key interlock that must be inserted and turned to the ENABLED position before the source outputs can be energized. Additionally, a remote interlock connector allows for the connection of a remote interlock switch, which can be used to disable the source outputs if the lid of a test fixture is opened.
Advertisement
Advertisement
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
(440) 248-0400
Model 2520 Pulsed Laser Diode Test System
Packing List
Introduction
This packing list contains information on using the supplied interlock key and remote interlock connector for the Model 2520
Pulsed Laser Diode Testhead. For the testhead output to be energized, the key must be inserted, and the remote interlock connector must be installed and wired as covered below:
WARNING It is the responsibility of the customer to operate instruments in a safe manner. Follow all applicable safety regulations for installing, configuring, and using the Model 2520.
The Model 2520, as installed, should be approved by the appropriate safety personnel, such as the responsible Laser Safety Officer or equivalent.
Suggested starting points for workplace regulations and standards: ANSI Z136.1, IEC
825, OSHA 29 CFR 1910.
In short, always be aware of workplace hazards, strive to minimize them, and work safely.
Using the key interlock
To use the key interlock, simply insert the key into KEY INTERLOCK (see Figure 1), and then rotate it clockwise to the
ENABLED position. To inhibit the source outputs, rotate the key to the DISABLED position.
WARNING If at any time the indicators provided for INTERLOCK STATUS or LASER POWER
ON should fail to light or to properly indicate status, immediately contact a Keithley service representative for repair. Failure to do so may expose the user to hazards without proper warning.
Figure 1
Key interlock and remote interlock connector locations
REMOTE INTERLOCK
Connector
KEY INTERLOCK
Insert Key, Rotate
Clockwise to Enable
!
BOTH INTERLOCKS MUST BE ENABLED TO OPERATE
DISABLED
(PULL TO
REMOVE)
KEY
INTERLOCK
1
REMOTE
INTERLOCK
DISABLED
9
ENABLED
GREEN=ENABLED
RED=DISABLED
INTERLOCK
STATUS
ENABLED
LASER
POWER
ON
MAINFRAME
CONN 2
MAINFRAME
CONN 1
PA-787 Rev. A / 4-01
2
Using the remote interlock connector
A remote interlock switch should be wired to pins 1 and 9 of the interlock connector, as shown in Figure 2. The connector should then be connected to the REMOTE INTERLOCK connector of the testhead (see Figure 1 for location).
Figure 3 shows typical interlock switch connections to the connector. If the test fixture switch is closed (Figure 3A), all three source outputs are enabled and can be turned on. If the lid of the test fixture opens (Figure 3B), the switch opens, and the three source outputs are disabled and cannot be turned on.
Figure 2
Remote interlock connector
1
REMOTE
INTERLOCK
DISABLED
9
ENABLED
Pin 1 Pin 9
Figure 3
Remote interlock operation
Model 2520 Testhead
Interlock
Connector
Pin 1
Pin 9
Test Fixture
Interlock Switch
Closed
A. 2520 outputs can be turned on.
Model 2520 Testhead
Interlock
Connector
Pin 1
Pin 9
Test Fixture
Interlock Switch
Open
B. 2520 outputs cannot be turned on.