Mindray Accutorr 3 Vital Signs Monitor Service Manual

Below you will find brief information for Vital Signs Monitor Accutorr 3. This device allows you to monitor vital signs, including SpO2, Pulse Rate, NIBP and Temperature, for adult, pediatric, and neonatal patients in healthcare facilities. The monitor includes a variety of features, such as a user-friendly interface and a comprehensive range of monitoring parameters.

PDF Download
Document
Vital Signs Monitor Accutorr 3 Service Manual | Manualzz

Advertisement

Advertisement

Advertisement

Key features

  • SpO2, Pulse Rate, NIBP and Temperature monitoring
  • Adult, Pediatric, and Neonatal patient compatibility
  • User-friendly interface
  • Comprehensive range of monitoring parameters
  • Temperature module for spot-check temperature monitoring
  • Internal battery for portable use
  • Alarm functions for critical parameter values
  • RS-232 port for data transfer to external devices
  • Maintenance mode for device diagnostics
  • Software upgrade capability for new features and bug fixes

Frequently asked questions

The equipment should be inspected and maintained by qualified service personnel every two years. This includes visual inspection, SpO2 test, NIBP test, Temperature test, and electrical safety and other tests.

Connect the NIBP cuff to the monitor, apply the cuff to a rigid vessel (like a cylinder), and press the button to enter the Maintenance mode. Then, select the NIBP leakage test interface and press the button to start the test.

The expected service life of the equipment is five years.

If the equipment or a module fails to perform as specified in any test, repairs or replacements must be done to correct the problem. Contact Mindray Technical Support if you have any questions.

The equipment should be used in an environment free from vibration, dust, and corrosive substances. The temperature should be between 0 to 40°C (without the Temperature module) or 5 to 40°C (with the Temperature module). The relative humidity should be between 15% and 95%.

Connect the equipment to a computer using the RS-232 port and follow the instructions in the software upgrade guide.