Mindray PM 7000 Patient Monitor Service Manual
The PM-7000 Patient Monitor (hereinafter called PM-7000 for short), a portable and accessible patient monitor, is supplied by rechargeable batteries or external AC/DC power, which applies to adults, pediatric and neonates. You can select different configurations as required. Besides, the PM-7000 can be connected with the central monitoring system whereby a monitoring network will be formed. Parameters that the PM-7000 can monitor include: ECG, RESP, SpO2, NIBP, 2-channel TEMP, 2-channel IBP, CO, CO2 and AG. It, integrating the functions of parameter measurement, waveform monitoring, freezing and recording, is a compact and lightweight patient monitor. Its color TFT LCD is able to show patient parameters and 7 waveforms clearly. The compact control panel and knob control, and the easy-to-use menu system enable you to freeze, record, or perform other operations conveniently.
Advertisement
Advertisement
PM-7000
Patient Monitor
Service Manual
Service Manual
Copyright
Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns all rights to this unpublished work and intends to maintain this work as confidential.
Mindray may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purposes of reference, operation, maintenance, or repair of Mindray equipment. No part of this can be disseminated for other purposes.
In the event of inadvertent or deliberate publication, Mindray intends to enforce its rights to this work under copyright laws as a published work. Those having access to this work may not copy, use, or disclose the information in this work unless expressly authorized by Mindray to do so.
All information contained in this publication is believed to be correct. Mindray shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this material. This publication may refer to information and protected by copyrights or patents and does not convey any license under the patent rights of
Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties.
Content of this manual is subject to changes without prior notice.
PROPERTY OF SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
ALL RIGHTS RESERVED
Responsibility on the manufacturer party
Mindray is responsible for safety, reliability and performance of this equipment only in the condition that:
• all installation, expansion, change, modification and repair of this equipment are conducted by Mindray qualified personnel;
•
applied electrical appliance is in compliance with relevant National Standards;
• the monitor is operated under strict observance of this manual.
Warning
For continued safe use of this equipment, it is necessary that the listed instructions are followed. However, instructions listed in this manual in no way supersede established medical practices concerning patient care. z Do not rely only on audible alarm system to monitor patient. When monitoring
I
Service Manual
adjusting the volume to very low or completely muting the sound may result in the disaster to the patient. The most reliable way of monitoring the patient is at the same time of using monitoring equipment correctly, manual monitoring should be carried out.
z This multi-parameter patient monitor is intended for use only by medical
professionals in health care institutions.
z To avoid electrical shock, you shall not open any cover by yourself. Service must be
carried out by qualified personnel.
z Use of this device may affect ultrasonic imaging system in the presence of the
interfering signal on the screen of ultrasonic imaging system. Keep the distance between the monitor and the ultrasonic imaging system as far as possible.
z It is dangerous to expose electrical contact or applicant coupler to normal saline,
other liquid or conductive adhesive. Electrical contact and coupler such as cable connector, power supply and parameter module socket-inlet and frame must be kept clean and dry. Once being polluted by liquid, they must be thoroughly dried. If to further remove the pollution, please contact your biomedical department or
Mindray.
It is important for the hospital or organization that employs this equipment to carry out a reasonable maintenance schedule. Neglect of this may result in machine breakdown or injury of human health.
II
Service Manual
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANT ABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the substitution upon it of parts or accessories not approved by Mindray or repaired by anyone other than a Mindray authorized representative.
This warranty shall not extend to any instrument which has been subjected to misuse, negligence or accident; any instrument from which Mindray's original serial number tag or product identification markings have been altered or removed, or any product of any other manufacturer.
Safety, Reliability and Performance
Mindray is not responsible for the effects on safety, reliability and performance of the
PM-7000 Patient Monitor if:
■ assembly operations, extensions, re-adjusts, modifications or repairs are carried out by persons other than those authorized by Mindray.
■ the PM-7000 is not used in accordance with the instructions for use, or the electrical installation of the relevant room does not comply with NFPA 70: National Electric
Code or NFPA 99: Standard for Health Care Facilities (Outside the United States, the relevant room must comply with all electrical installation regulations mandated by the local and regional bodies of government).
III
Service Manual
Return Policy
Return Procedure
In the event that it becomes necessary to return a unit to Mindray, the following procedure should be followed:
1. Obtain return authorization. Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping container. Return shipments will not be accepted if the Mindray number is not clearly visible. Please provide the model number, serial number, and a brief description of the reason for return.
2. Freight policy. The customer is responsible for freight charges when equipment is shipped to Mindray for service (this includes customs charges).
Company Contact
Mindray
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial
Park, Nanshan, Shenzhen 518057, P.R.China
Phone:
+86 755 26582479 26582888
Fax:
+86 755 26582934 26582500
IV
Service Manual
Safety Precautions
1. Meaning of Signal Words
In this manual, the signal words
DANGER
, WARNING
, and
CAUTION
are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual.
Signal word
DANGER
WARNING
CAUTION
Meaning
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
2. Meaning of Safety Symbols
Symbol Description
Type-BF applied part
"Attention" (Refer to the operation manual.)
Safety Precautions
Please observe the following precautions to ensure the safety of service engineers as well as operators when using this system.
DANGER:
Do not use flammable gases such as anesthetics, or flammable
liquids such as ethanol, near this product, because there is danger of explosion.
WARNING:
Do not connect this system to outlets with the same circuit breakers and fuses that control current to devices such as life-support systems. If this system malfunctions and generates an over current, or when there is an instantaneous
V
Service Manual
current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped.
CAUTION:
1. Malfunctions due to radio waves
(1) Use of radio-wave-emitting devices in the proximity of this kind of medical electronic system may interfere with its operation. Do not bring or use devices which generate radio waves, such as cellular telephones, transceivers, and radio controlled toys, in the room where the system is installed.
(2) If a user brings a device which generates radio waves near the system, they must be instructed to immediately turn OFF the device. This is necessary to ensure the proper operation of the system.
2. Do not allow fluids such as water to contact the system or peripheral devices. Electric shock may result.
VI
Service Manual
Symbols
Equipotential grounding terminal
Be Careful
Direct current and alternating current (DC&AC)
Alternating current (AC)
CE mark 93/42/EEC a directive of the European
Economic Community
Protective earth ground
Direct current (DC)
Battery indicator
ESD sensitivity Power ON/OFF
Network connector
High voltage
Indicates that the instrument is IEC-60601-1 Type CF equipment. The unit displaying this symbol contains an F-Type isolated (floating) patient applied part providing a high degree of protection against shock, and is suitable for use during defibrillation.
VII
Service Manual
Contents
Chapter 1 About the Product .....................................................................................................1
1.1 Introduction......................................................................................................................1
1.2 Application .......................................................................................................................1
1.2.1 General ......................................................................................................................1
1.3 Environment ....................................................................................................................3
1.3.2 Humidity ...................................................................................................................3
Chapter 2 Principles....................................................................................................................4
2.1 General ............................................................................................................................4
2.2.2 ECG/RESP/TEMP Module .......................................................................................9
2.2.4 SpO
2
Module...........................................................................................................12
2.2.7 Button Panel............................................................................................................15
2.2.8 Power PCB..............................................................................................................16
2.3.1 General ....................................................................................................................17
2.3.2 System Task.............................................................................................................18
2.3.3 System Function.............................................................................................................19
2.4.1 General ....................................................................................................................20
2.4.2 ECG/RESP ..............................................................................................................20
2.4.3 NIBP........................................................................................................................21
2.4.4 SpO
2
........................................................................................................................22
2.4.5 TEMP ......................................................................................................................22
2.4.6 IBP ..........................................................................................................................22
2.4.7 CO ...........................................................................................................................23
2.4.8 CO
2
..........................................................................................................................23
2.4.9 AG...........................................................................................................................23
I
Service Manual
Chapter 3 Product Specifications .................................................................................................25
3.1 Safety Classifications............................................................................................................25
3.2 Environmental Specifications ...............................................................................................26
3.3 Power Source Specifications.................................................................................................27
3.4 Hardware Specifications .......................................................................................................28
3.5 Wireless network...................................................................................................................29
3.6 Data Storage..........................................................................................................................30
3.9 RESP Specifications..............................................................................................................34
3.10 SpO
2
Specifications.............................................................................................................35
1.1.1. Mindray SpO
2
Module ..................................................................................................35
1.1.2. Masimo SpO
2
Specifications.........................................................................................35
1.1.3. Nellcor SpO
2
Specifications..........................................................................................36
3.12 TEMP Specifications...........................................................................................................38
3.13 IBP Specifications...............................................................................................................39
3.14 CO Specifications ...............................................................................................................40
3.15 CO
2
Specifications ..............................................................................................................41
1.1.4. Mindray CO
2
Specifications .........................................................................................41
1.1.5. Oridion CO
2
Specifications...........................................................................................42
1.1.6. Welch Allyn CO
2
Specifications ...................................................................................43
3.16 AG Specifications ...............................................................................................................44
Chapter 4 Disassembling/Assembling & Troubleshooting .............................................................47
4.1.2 PM-7000 Support Assembly ..........................................................................................48
4.1.3 Front Bezel Assembly ....................................................................................................50
4.1.4 Rear Housing Assembly.................................................................................................51
4.1.5 Microstream CO
2
Assembly...........................................................................................52
4.2 Troubleshooting ............................................................................................................53
4.2.1 Black Screen, Startup Failure ...................................................................................53
4.2.2 White Screen & Other Abnormal Screen ................................................................54
4.2.3 Encoder Faults............................................................................................................54
4.2.4 No Audio Alarm ...........................................................................................................54
4.2.5 Printing Failure............................................................................................................54
4.2.6 Abnormal Paper Drive................................................................................................54
Chapter 5 Test and Material List .............................................................................................55
5.1 Test Procedure..............................................................................................................55
5.1.1 Connection and Checking .........................................................................................55
5.1.2 Functions of Buttons ..................................................................................................55
5.1.3 ECG/RESP ..................................................................................................................55
5.1.4 Temperature ................................................................................................................56
II
Service Manual
5.1.5 NIBP...............................................................................................................................56
5.1.6 SpO
2
...............................................................................................................................56
5.1.7 IBP .................................................................................................................................56
5.1.8 CO ..................................................................................................................................57
5.1.9 CO
2
.................................................................................................................................57
5.1.10 Recorder .......................................................................................................................58
5.1.12 Power Supply ...............................................................................................................58
5.1.13 Clock ............................................................................................................................58
5.1.14 System Test ..................................................................................................................58
5.3 IBP CALIBRATE..............................................................................................................59
5.4 CO2 CHECK.......................................................................................................................62
5.5 AG CALIBRATE .................................................................................................................64
5.5.1 AG Check.....................................................................................................................64
5.5.2 AG CALIBRATE ..........................................................................................................65
5.6 Bill of Materials for PM-7000 Main Unit .......................................................................66
Chapter 6 Maintenance and Cleaning .............................................................................................67
6.1 Maintenance ..........................................................................................................................67
6.1.1Checking Before Using ...................................................................................................67
6.1.2 Regular Checking...........................................................................................................67
6.2 Cleaning ..........................................................................................................................67
6.4 Sterilization .....................................................................................................................68
6.5 Disinfection.....................................................................................................................68
III
Chapter 1 About the Product
Chapter 1 About the Product
1.1 Introduction
The PM-7000 Patient Monitor (hereinafter called PM-7000 for short), a portable and accessible patient monitor, is supplied by rechargeable batteries or external AC/DC power, which applies to adults, pediatric and neonates. You can select different configurations as required. Besides, the PM-7000 can be connected with the central monitoring system whereby a monitoring network will be formed. Parameters that the PM-7000 can monitor include: ECG,
RESP, SpO
2
, NIBP, 2-channel TEMP, 2-channel IBP, CO, CO
2
and AG. It, integrating the functions of parameter measurement, waveform monitoring, freezing and recording, is a compact and lightweight patient monitor. Its color TFT LCD is able to show patient parameters and 7 waveforms clearly. The compact control panel and knob control, and the easy-to-use menu system enable you to freeze, record, or perform other operations conveniently.
The PM-7000 measures patient’s ECG, NIBP, SpO
2
, TEMP, RESP, IBP, CO and CO
2 physiological signals through the ECG electrode, SpO
2
sensor, cuff, temperature sensor and pressure transducer. During the measurement, the patient monitor does not get energy or any substance from the human body, and does not release any substance to the human body.
However, it releases sine wave signals to the patient when measuring the respiration rate. The patient monitor converts the measured physiological signals to the digital signals, waveforms and values, and then displays them on the screen. You can control the patient monitor through the control panel. For example, you can set different alarm limits for different patients. Thus, when the patient monitor detects any physiological parameter exceeding the preset alarm limit, it will enable the audio and visual alarm.
1.2 Application
1.2.1 General
In the treatment processes, it is necessary to monitor important physiological information of patients. Therefore, the patient monitor has been playing an outstanding role among medical devices. The development of technology does not only help medical staff get the important physiological information, but also simplifies the procedures and makes it more effective. For patients in hospital, the basic and important physiological information is required, including
ECG, SpO
2
, RESP, IBP, CO, CO
2
, TEMP, etc. In recent years, the development of science and technology helping measure and get important physiological information of patients has made the patient monitor more comprehensive in performance and better in quality. Today, multi-parameter patient monitors are widely used.
1.2.2 Usage
Parameters that the PM-7000 include: ECG, RESP, SpO
2
, NIBP, TEMP, IBP, CO AG and CO
2
.
PM-7000 converts these physiological signals to digital signals, processes them and displays them on the screen. You can set the alarm limit as required. When the monitored parameter
1
Chapter 1 About the Product exceeds the preset alarm limit, the patient monitor will start the alarm function. In addition, you can control the patient monitor through the control panel. Usually, patient monitors are seen in some clinical areas in hospital, such as ICU, CCU, intensive care units for heart disease patients, operating rooms, emergency departments and observation wards. They can also be used in clinics. The PM-7000 should be run under the control of clinical staff.
PM-7000 has the following functions:
ECG Heart Rate (HR)
RESP
2-channel ECG waveform
Arrhythmia analysis and S-T analysis (optional)
Respiration Rate (RR)
SpO
TEMP
IBP
CO
2
2
NIBP
CO
AG
Respiration waveform
Pulse Oxygen Saturation(SpO
2
), Pulse Rate (PR)
SpO
2
Plethysmogram
Systolic pressure (NS), diastolic pressure (ND), mean pressure
(NM)
T1, T2, TD
CH1: SYS, DIA
CH2: SYS, DIA
IBP waveform
Temperature of blood (TB)
Cardiac Output (CO)
End-tidal carbon dioxide (EtCO
2
)
Inspired minimum CO
2
(InsCO
2
)
Airway Respiration Rate (AwRR)
Inhaled and exhaled CO
2
(FiCO
2
, EtCO
2
)
Inhaled and exhaled N
2
O (FiN
2
O, EtN
2
O)
Inhaled and exhaled O
2
(FiO
2
, EtO
2
)
Inhaled and exhaled anesthetic agent (FiAA, EtAA, where AA refers to any of the following anesthetic agents.)
HAL (Halothane)
ISO (Isoflurane)
ENF (Enflurane)
2
Chapter 1 About the Product
SEV (Sevoflurane)
DES (desflurane)
Airway Respiration Rate (rpm: Respiration Per Minute): AwRR
Minimum Alveolar Concentration (MAC)
4 AG waveforms (CO
2
, N
2
O, O
2
, AA)
The PM-7000 provides the functions of audio/visual alarm, trend graphic storage and output,
NIBP measurement, alarm event identification, large font screen, defibrillator synchronization, oxyCRG recall, drug calculation, etc.
1.3 Environment
1.3.1 Temperature
Work mode 0 ~40℃
MINDRAY CO
2
module +5 ~ +35℃
Welch Allyn mainstream CO
2
module +10 ~ +35℃
Microstream CO
2
module +5 ~ +35℃
Artema AION AG module +10 ~ +35℃
Transportation & Storage -20 ~ 60℃
1.3.2 Humidity
Work mode 15% – 95 % (non-condensing)
Transportation & Storage 10% – 95 % (non-condensing)
Atmospheric pressure 70.0kPa – 106.0kPa
AC power supply: 100 to 240 V AC, 50/60 Hz, Maximum input power: 140VA
DC power supply: 12 V (nominal), 10 to 16 V, Power: 80W
2.3 Ah 12V lead-acid rechargeable battery
Working time of fully-charged batteries in normal status: 75 minutes (1 battery).
From the first low-battery alarm, the batteries can supply power to the patient monitor for 5
~ 15 minutes.
Maximum charging time: about 8h
4.4Ah 11.1V lithium battery
Working time of fully-charged batteries in normal status: 150 minutes (1 battery).
From the first low-battery alarm, the batteries can supply power to the patient monitor for 5
~ 15 minutes.
Maximum charging time: about 8h
3
Chapter 2 Principles
Chapter 2 Principles
2.1 General
The intended use of the PM-7000 is to monitor a fixed set of parameters including ECG, RESP,
SpO
2
, NIBP, TEMP, IBP, CO and CO
2
(IBP, CO and CO
2
are optional). It consists of the following functional parts:
Parameter measurement;
Main control part;
Man-machine interface;
Power supply;
Other auxiliary functions;
These functional units are respectively detailed below.
Figure 2-1 Structure of the PM-7000
4
Chapter 2 Principles
The parameter measurement and monitoring are the core functions of the patient monitor. The parameter measurement part of the PM-7000 consists of the measurement probe, parameter input socket assembly, NIBP assembly and the main control board.
This part converts the physiological signals to electric signals, processes the those signals and conducts the calculation by the preset program or command delivered from the main control board, and then sends the values, waveforms and alarm information (which will be displayed by using the man-machine interface) to the main control board.
2.1.2 Main Control Part
In the PM-7000, the main control part refers to the main control part of the main control board.
It drives the man-machine interface, manages the parameter measurement and provides users with other special functions, such as storage, recall of waveforms and data. (See Figure
2-1)
The man-machine interface of the PM-7000 includes the TFT display, recorder, speaker, indicator, buttons and control knob.
The TFT display is the main output interface. It, with the high resolution, provides users with abundant real-time and history data and waveforms as well as various information and alarm information.
The recorder is a subsidiary of the display, which is used for the user to print data.
The speaker provides the auditory alarm function.
The indicator provides additional information about the power supply, batteries, alarms and so on.
The buttons and control knob are the input interface, which are used for the user to input the information and commands to the patient monitor.
The power supply part is an important part of the patient monitor. It includes the main power
PCB, backlight board, batteries and fan.
The main power PCB converts the external AC current respectively to the 5V DC and 12V DC current, which are supplied for the whole system. For the TFT display, there is a special requirement on the power supply, so a backlight board is used. The batteries supply power for the system for a short time when there is no external AC current. The fan is used for the heat sink of the system.
The PM-7000 also provides the network upgrade function for the service engineers to upgrade the system software without disassembling the enclosure.
5
Chapter 2 Principles
2.2 Hardware Description
The structure of the PM-7000 is shown in the following figure.
Figure 2-2 Functional structure of the PM-7000
6
Chapter 2 Principles
The PM-7000 PCB connection is shown in the following figure.
Figure 2-3 PCB connection
Basic functions and working principles of modules are described in the following sections.
2.2.1.1 General
The main board is the heart of the patient monitor. It implements a series of tasks, including the system control, system scheduling, system management, data processing, file management, display processing, printing management, data storage, system diagnosis and alarm.
7
2.2.1.2 Principle diagram
Chapter 2 Principles
Figure 2-4 Working principle of the main board
2.2.1.3 Principle
The main board is connected with external ports, including the power input port, multi-way serial port, TFT display interface, analog VGA interface, network port and analog output port.
Besides, on the main board is also a BDM interface reserved for the software debugging and software downloading.
CPU System
CPU is the core part of the main board. It, connected with other peripheral modules through the bus and I/O cable, implements the data communication, data processing, logical control and other functions.
RTC
RTC provides the calendar information (such as second, minute, hour, day, month and year). CPU can read and modify the calendar information from RTC.
Ethernet Controller
Ethernet Controller supports the IEEE802.3/IEEE802.3u LAN standard, and supports two data transmission rate: 10Mbps and 100Mbps. CPU exchanges data with the Ethernet through the Ethernet Controller.
Analog Output
The D/A converter converts the digital ECG/IBP signals sent from CPU to the analog signals, which are provided for the external after low-pass filtered by the filter and amplified by the amplifier.
FPGA and VRAM
VRAM stores the displayed data. CPU stores the displayed data to VRAM through FPGA.
FPGA gets data from VRAM, processes them, and then sends them to the relevant graphic display device.
8
Chapter 2 Principles
In addition, FPGA also extends multiple serial ports, which communicate with peripheral modules. FPGA transfers the received data to CPU through the bus; CPU delivers data to
FPGA through the bus, and then the FPGA transfers those data to the peripheral modules.
Watchdog
When powered on, watchdog provides reset signals for CPU, FPGA and Ethernet
Controller.
The patient monitor provides the watchdog timer output and voltage detection functions.
2.2.2.1 General
This module provides the function of measuring three parameters: electrocardiograph (ECG), respiration (RESP) and temperature (TEMP).
2.2.2.2 Principle diagram
Figure 2-5 Working principle of the ECG/RESP/TEMP module
2.2.2.3 Principle
This module collects the ECG, RESP and TEMP signals through the transducer, processes the signals, and sends the data to the main board through the serial port.
ECG Signal Input Circuit
The input protection and filtering circuits receive the ECG signal from the transducer, and filter the high-frequency interference signal to protect the circuit against the damage by defibrillator high-voltage and ESD.
The right-leg drive circuit gets the 50/60Hz power common-mode signal from the lead cable, and sends the negative feedback signal to the human body to reject the common-mode interference signal on the lead cable, which helps the detection of the ECG signal.
The lead-off detecting circuit checks whether the ECG lead is off, and sends the information to
CPU.
ECG Signal Process Circuit
The difference amplifying circuit conducts the primary amplification of the ECG signal and
9
Chapter 2 Principles rejects the common-mode interference signal.
The low-pas filtering circuit filters the high-frequency interference signal beyond the frequency band of the ECG signal.
The PACE signal refers to the ECG pace signal. It has significant interference to the ECG signal detection. The PACE rejection circuit can rejects the PACE signal, which helps the ECG signal detection.
The main amplifying/filtering circuit conducts the secondary amplification of the ECG signal, filters the signal, and then sends the ECG signal to the A/D conversion part.
Pace Detect
This part detects the PACE signal from the ECG signal and sends it to CPU.
Temperature Detect Circuit
This circuit receives the signal from the temperature transducer, amplifies and filters it, and then sends it to the A/D conversion part.
Carrier Generate Circuit
The RESP measurement is based on the impedance method. While a man is breathing, the action of the breast leads to changes of the thoracic impedance, which modulates the amplitude of the high-frequency carrier signal. Finally, the modulated signal is sent to the measurement circuit. The purpose of this module is generating the high-frequency carrier.
RESP Signal Input Circuit
This circuit couples the RESP signal to the detecting circuit.
RESP Signal Process Circuit
The pre-amplifying circuit conducts the primary amplification of the RESP signal and filters it.
The detecting circuit detects the RESP wave that has been modulated on the actuating signal.
The level shifting circuit removes the DC component from the RESP signal.
The main amplifying/filtering circuit conducts the secondary amplification of the RESP signal, filters the signal, and then sends it to the A/D conversion part.
A/D
The A/D conversion part converts the analog signal to the digital signal, and sends the signal to CPU for further processing.
CPU System
Implementing the logical control of all parameter parts and A/D conversion parts;
Implementing the data processing for all parameters;
Implementing the communication with the main board.
Power & Signal isolate Circuit
Isolating the external circuits to ensure the safety of human body;
Supplying power for all circuits;
Implementing the isolation communication between the CPU System and the main board.
10
Chapter 2 Principles
2.2.3.1 General
This module provides the function of measuring two parameters: Cardiac Output (CO) and
Invasive Blood Pressure (IBP).
2.2.3.2 Principle diagram
Figure 2-6 Working principle of the CO/IBP module
2.2.3.3 Principle
This module collects the CO/IBP signal through the transducers, processes it and sends it to the main board throgh the serial port.
CO Signal Process Network
The CO parameter is measured with the thermal dilution method. The transducer sends two signals (TI: Temperature of Injectate; TB: Temperature of Blood) to the CO
Signal Process Network. After that, the signals are amplified and low-pass filtered, and then sent to the CPU System for processing.
IBP Signal Process Network
The IBP signal is the differential signal. After the common-mode filtering, the difference signal is amplified by the difference amplifying circuit which changes the dual-end signal to the single-end signal. After the low-pass filtering, the IBP signal is sent to the CPU
System for processing.
CPU System
Converting the analog signal obtained by the circuit to the digital signal;
Implementing the logical control of all parameter parts;
Implementing the data processing for the two parameters;
Implementing the communication with the main board.
Power & Signal isolate Circuit
Isolating the external circuits to ensure the safety of human body;
Supplying power for all circuits;
Implementing the isolation communication between the CPU System and the main board.
11
Chapter 2 Principles
2.2.4 SpO
2
Module
2.2.4.1 General
This module provides the function of measuring the Pulse Oxygen Saturation (SPO
2
).
2.2.4.2 Principle diagram
Figure2-7 Working principle of the SpO
2
module
2.2.4.3 Principle
The SpO
2
measurement principle
1. Collecting the light signal of the red light and infrared transmitting through the finger or toe which is pulsing;
2. Processing the collected signal to get the measured result.
The drive circuit of the LED and the gain of the amplifying circuit should be controlled according to the different perfusions and transmittances of the tested object.
Led Drive Circuit
This circuit supplies the LED with the drive current, which can be regulated.
SPO2 Signal Process Network
The pre-amplifying circuit converts the photoelectric signal to the voltage signal and conducts the primary amplification.
The gain adjusting and amplifying circuit conducts the secondary signal amplification and adjusts the gain.
The biasing circuit adjusts the dynamic range of the signal, and sends it to the A/D conversion part.
A/D
The A/D conversion part converts the analog signal to the digital signal, and then sends it to
CPU for further processing.
D/A
The D/A conversion part converts the digital signal received from CPU to the analog signal, and provides the control signal for the Led Drive Circuit and SPO2 Signal Process Network.
12
Chapter 2 Principles
CPU System
Implementing the logical control of all the circuits;
Implementing the data processing for the SpO
2
parameter;
Implementing the communication with the main board.
Power & Signal isolate Circuit
Isolating the external circuits to ensure the safety of human body;
Supplying power for all circuits;
Implementing the isolation communication between the CPU System and the main board.
2.2.5.1 General
This module provides the function of measuring the Non-Invasive Blood Pressure (NIBP) parameter.
2.2.5.2 Principle diagram
Figure 2-8 Working principle of the NIBP module
2.2.5.3 Principle
The NIBP is measured based on the pulse vibration principle. Inflate the cuff which is on the forearm till the cuff pressure blocks the arterial blood, and then deflate the cuff according to a specified algorithm. While the cuff pressure is decreasing, the arterial blood has pulses, which are sensed by the pressure transducer in the cuff. Consequently, the pressure transducer, connected with the windpipe of the cuff, generates a pulsation signal, which is then processed by the NIBP module to get the NIBP value.
13
Chapter 2 Principles
Valve Drive Circuit
This circuit controls the status (ON/OFF) of valves. It, together with the Motor Drive Circuit, implements the inflation and deflation of the cuff.
Motor Drive Circuit
This circuit controls the action of the air pump. It, together with the Valve Drive Circuit, implements the inflation and deflation of the cuff. Besides, it provides the status signal of the motor for the A/D conversion part.
NIBP Signal Process Network
The NIBP signal is the differential input signal. The difference amplifying circuit amplifies the dual-end difference signal and converts it to the single-end signal; meanwhile, this circuit sends a channel of signal to the A/D conversion part, and the other to the DC isolating and amplifying circuit.
The DC isolating and amplifying circuit removes DC components from the signal, amplifies the signal, and then sends it to the A/D conversion part.
A/D
The A/D conversion part converts the analog signal to the digital signal, and sends it to the
CPU System for further processing.
Over Pressure Detect
The circuit detects the NIBP pressure signal. Once the pressure value exceeds the protected pressure value, it will send a message to the CPU System, which asks the Valve Drive Circuit to open the valve to deflate the cuff.
CPU System
Implementing the logical control of all the circuits;
Implementing the data processing for the NIBP parameter;
Implementing the communication with the main board.
2.2.6.1 General
This module is used to drive the heat-sensitive printer.
2.2.6.2 Principle diagram
Figure 2-9 Working principle of the recorder module
14
Chapter 2 Principles
2.2.6.3 Principle
This module receives the to-be-printed data from the main board, converts them to the dot matrix data, sends them to the heat-sensitive printer, and drives the printer.
Step Motor Drive Circuit
There is a step motor on the heat-sensitive printer. The step motor drives the paper. This circuit is used to drive the step motor.
Printer Status Detect Circuit
This circuit detects the status of the heat-sensitive printer, and sends the status information to the CPU system. The status information includes the position of the paper roller, status of the heat-sensitive recorder paper and the temperature of the heat-sensitive head.
CPU System
Processing the data to be printed;
Controlling the heat-sensitive printer and step motor;
Collecting data about the status of the heat-sensitive printer, and controlling the printer;
Implementing the communication with the main board.
2.2.7.1 General
This module provides a man-machine interactive interface.
2.2.7.2 Principle diagram
Figure 2-10 Working principle of the button panel
2.2.7.3 Principle
This module detects the input signals of the button panel and control knob, converts the detected input signals to codes and then sends to the main board. The main board sends commands to the button panel, which, according to the commands, controls the status of the
LED and the audio process circuit to give auditory/visual alarms.
15
Chapter 2 Principles
CPU
Detecting the input signal of the button panel and control knob;
Controlling the status of LED;
Controlling the audio process circuit;
Regularly resetting the Watchdog timer;
Communicating with the main board.
Audio Process Circuit
This circuit generates audio signals and drives the speaker.
Watchdog
When powered on, the Watchdog provides the reset signal for CPU.
The patient monitor provides the watchdog timer output and voltage detection functions.
2.2.8.1 General
This module provides DC working current for other boards.
2.2.8.2 Principle diagram
Figure 2-11 Working principle of the power PCB
2.2.8.3 Principle
This module can convert 220V AC or the battery voltage to 5V DC and 12V DC voltages, which are supplied for other boards. When the AC voltage and batteries coexist, the AC voltage is supplied for the system and used to charge the batteries.
AC/DC
This part converts the AC voltage to the low DC voltage for the subsequent circuits; besides, it supplies the power for charging the batteries.
Battery Control Circuit
When the AC voltage and batteries coexist, this circuit controls the process of charging the batteries with the DC voltage converted by the AC/DC part. When the AC voltage is unavailable, this circuit controls the batteries to supply power for the subsequent circuits.
16
Chapter 2 Principles
5V DC/DC
This part converts the DC voltage to the stable 5V DC voltage and supplies it for the external boards.
12V DC/DC
This part converts the DC voltage to the stable 12V DC voltage and supplies it for the external boards.
Power Switch Circuit
This circuit controls the status of the 5V DC/DC part and the 12V DC/DC part, thus to control the switch of the patient monitor.
Voltage Detect Circuit
This circuit detects the output voltages of the circuits, converts the analog signal to the digital signal, and sends the digital signal to the main board for processing.
2.3 Software Description
2.3.1 General
Figure 2-12 System function
As shown in Figure 2-12, in the red frame is the software system, on the left to the red frame are the inputs of the software system, and on the right to the red frame are the outputs. The parameter measurement module exchanges data with the software through the serial port, while the user interacts with the system through the button panel. Among the output devices, the recorder and alarm device receive data through the serial ports, the analog output component is an MBUS component, and the LCD and network controller are controlled directly by CPU.
17
Chapter 2 Principles
NO Task
1 System initialization
Function
Initializing the system
Period
In case of a startup
2
Data processing
3
Analyzing and saving the data
Display of timer
Implementing the timed refreshing information
5
6
7
System monitoring
1 second
1 second
Switchover of modules and screens
Switching over between waveforms and
In case of a screen change parameters on the screen event
Processing of user commands and displaying them on the screen. screens
In case of a button event
System monitoring, voltage monitoring and battery management
1 second
1 second 8 Network connection Implementing the network connection
9
Network data sending Sending the network data
10 Network data receiving Receiving the network data (viewbed)
11 ECG analysis
Analyzing ECG signal, calculating ECG values
(HR, ARR and ST), and saving the analysis results.
1 second
1 second
1 second
12
13
Record output
NIBP processing
14
WATCHDOG task
Outputting records
Implementing NIBP-related processing
Managing the system watchdog
In case of a record event
1 second
1 second
18
Chapter 2 Principles
2.3.3 System Function
The system tasks can be classified as follows.
Figure 2-13 System task
19
Chapter 2 Principles
2.4
System Parameter
2.4.1 General
For the PM-7000, signals are collected by modules, and the results are transferred to the main board through the adapter board, thus to process and display the data and waveforms.
Commands from the main board, as well as the status information of modules, are transferred through the adapter board. In addition, the adapter board adapts and changes the power supply. The structure of the whole system is shown in the following figure.
---
As shown in Figure 2-14, the five modules and measurement cables monitor and measure
NIBP, SpO
2
, ECG/RESP/TEMP, IBP/CO and CO
2
in real time, and send the results to the main board for processing and displaying. If necessary, the results are sent to the recorder for printing.
The parameter monitoring functions are described respectively in the following sections.
2.4.2 ECG/RESP
■
ECG
The PM-7000 has the following ECG functions:
1) Lead type: 3-lead, 5-lead, 12-lead
2) Lead way:
3-lead (1 channel):
5-lead (2 channels):
I, II, III
I, II, III, aVR, aVL, aVF, V
20
Chapter 2 Principles
12-lead (8 channels): I, II, III, avR, avL, avF, V1-V6
3) Floating input
4) Right-foot drive
5) Lead-off detection
6) 2-channel ECG waveform amplification; processing ECG signals of any two leads
■ The ECG circuit processes the ECG signals. It consists of the following parts:
1) Input circuit: The input circuit protects the ECG input level, and filters the ECG signals and external interference. The ECG electrode is connected to the input circuit through the cable.
2) Buffer amplifying circuit: This circuit ensures extremely high input impedance and low output resistance for ECG.
3) Right-foot drive circuit: The output midpoint of the buffer amplifying circuit is fed to the
RL end of the 5-lead after the inverse amplification, so as to ensure that the human body is in the equipotential state, decrease the interference, and increase the common-mode rejection ratio of the circuit.
4) Lead-off detection: The lead-off causes changes in the output level of the buffer amplifying circuit. Therefore, the lead-off can be detected with a comparator, and the state of lead-off can be converted TTL level for the Micro Controller Unit (MCU) to detect it.
5) Lead circuit: Under the control of MCU, the lead electrodes should be connected to the main amplification circuit.
6) Main amplification circuit: The measurement amplifier is composed of 3 standard operation amplifiers.
7) Subsequent processing circuit: This circuit couples the ECG signals, remotely controls the gains, filters the waves, shifts the level, amplifies the signal to the specified amplitude, and sends the signal to the A/D converter.
■
RESP
The PM-7000 measures the RESP based on the impedance principle. While a man is breathing, the action of the breast leads to impedance changes between RL and LL.
Change the high-frequency signal passing the RL and LL to amplitude-modulation high-frequency signal (AM high-frequency signal), which is converted to the electric signal after being detected and amplified and then sent to the A/D converter. The RESP module consists of the RESP circuit board and coupling transformer. The circuit has several functions: vibration, coupling, wave-detection, primary amplification and high-gain amplification.
2.4.3 NIBP
The NIBP is measured based on the pulse vibration principle. Inflate the cuff which is on the forearm till the cuff pressure blocks the arterial blood, and then deflate the cuff according to a specified algorithm. While the cuff pressure is decreasing, the arterial blood has pulses, which are sensed by the pressure transducer in the cuff.
Consequently, the pressure transducer, connected with the windpipe of the cuff, generates a pulsation signal. Then, the pulsation signal is filtered by a high-pass filter
(about 1Hz), amplified, converted to the digital signal by the A/D converter, and finally processed by the MCU. After that, the systolic pressure, diastolic pressure and mean
21
Chapter 2 Principles pressure can be obtained. For neonates, pediatric and adults, it is necessary to select the cuffs of a proper size to avoid possible measurement errors. In the NIBP measurement, there is a protection circuit used to protect patient from over-high pressure.
The NIBP measurement modes include:
1) Adult/pediatric/neonate mode: To be selected according to the build, weight and age of the patient;
2) Manual/Auto/Continuous mode: The manual measurement is also called single measurement; in this mode, only one measurement is done after being started. In the auto measurement mode, the measurement can be done once within the selected period, with the interval being 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240 or 480 minutes. In the continuous measurement mode, quick continuous measurement will be done within 5 minutes after being started; it detects the changes in blood pressure effectively.
2.4.4 SpO
2
The SpO
2
value is obtained through the pulse waves of the finger tips based on specific algorithm and clinical data. The SpO
2
probe is the measurement transducer. It has two inbuilt LEDs and an inbuilt light receiver. The two LEDs include one red-light diode and one infrared diode, which emit light in turns. When the capillaries in the finger tip are iteratively congested with blood pumped by the heart, the light emitted by the LEDs, after absorbed by the capillaries and tissue, casts on the light receiver, which can sense, in the form of electric signal, the light strength changing with the pulsated blood. The DC/AC ratio of the two photoelectric signals corresponds to the content of the oxygen in the blood.
Therefore, the correct pulse oxygen saturation can be obtained with specific algorithm.
Moreover, the pulse rate can be obtained according to the pulse waveform.
The circuit of the SpO
2
module is involved in four parts: SpO
2
probe, signal processing unit, LED-driven sequencing control part and the MCU.
2.4.5 TEMP
Temperature measurement principle:
1. The transducer converts the body temperature to the electric signal;
2. The amplifier amplifies the electric signal;
3. The CPU processes the data.
The circuit is a proportional amplifier consisting of operation amplifiers. When the temperature reaches the heat-sensitive probe, the heat-sensitive probe generates the voltage signal, which is sent to the A/D converter after being amplified. The probe detecting circuit is a voltage comparator consisting of operation amplifiers. When the probe is disconnected, the voltage input is lower than the comparing voltage, so the voltage comparator outputs the low level; when the probe is connected, the voltage input is higher than the comparing voltage, so the voltage comparator outputs the high level.
2.4.6 IBP
The IBP module can monitor the arterial pressure, central venous pressure and pulmonary arterial pressure.
22
Chapter 2 Principles
Measurement principle: Introduce a catheter, of which the external end is connected to the pressure transducer, into the blood vessel under test, inject the physiological saline.
Since the liquid can be transferred by pressure, the pressure inside the blood pressure is transferred by liquid to the pressure transducer, and the dynamic waveform of the pressure inside the blood pressure is obtained in real time. Thus, the arterial pressure, central venous pressure and pulmonary arterial pressure are obtained based on specific algorithm.
2.4.7 CO
CO measurement principle: The thermal dilution method is widely used in the clinical CO monitoring. Introduce a floating catheter into the pulmonary artery through the right atrium, and inject the physiological saline into the right atrium through the catheter whose front end is connected to the temperature transducer. When the cold liquid mixes with the blood, there will be a change of temperature. Thus, when the blood mixed with the physiological saline flows into the pulmonary artery, its temperature will be sensed by the temperature transducer. According to the injection time and temperature change, the patient monitor can analyze the CO, and calculate the Cardiac Index (CI), Stroke Volume
Index (SVI), SVIs of the left atrium and right atrium, Pulmonary Vascular Resistance (PVR) and so on.
2.4.8 CO
2
The CO
2
module works based on the infrared spectrum absorption principle. According to different connection methods, the infrared light transducer is classified as the mainstream infrared light transducer and sidestream infrared light transducer. The sidestream CO
2 module is composed of the circuit board, inbuilt sidestream infrared light transducer, deflation pump and control. When used, this module requires the external water trap, drying pipe and sampling tube. The mainstream CO
2
module is composed of the circuit board and external mainstream infrared light transducer. The infrared light transducer needs to be preheated. In the sidestream mode, the deflation rate can be set to
100ml/min, 150ml/min or 200ml/min according to the patient situation. In the AG monitoring, multiple compensations can be set, such as hydrosphere, oxygen, temperature and desflurane (Des). When the CO
2
measurement is not being conducted, the sidestream deflation pump, the transducer of the mainstream module, and the infrared source are expected to be shut down, thus to extend the service life and reduce the power consumption of the module. In the mainstream mode, the infrared light transducer takes a longer time to be preheated, and there is no windpipe which is available in the sidestream mode.
2.4.9 AG
The Anesthesia Gas (AG) can be used to measure the AG and respiration gas of the anesthetized patient. The AG concentration is measured based on the principle that the
AG has the property of absorbing the infrared. All gases that the AG module can measure have the property of absorbing the infrared, and every gas has their own specific absorption peculiarity.
AG measurement procedure:
23
Chapter 2 Principles
1. Send the gas to be measured to a sampling chamber;
2. Use an optical infrared filter, select a specific band of infrared, and transmit it through the gas;
3. Measure the infrared that gets through the gas to obtain the gas concentration.
For a given volume, the higher the gas concentration is, the more absorbed infrared is, and the less the infrared that gets through the gas is. For the measurement of multiple gases, multiple infrared filters are required in the AG module.
The oxygen does not absorb the infrared within the above-mentioned wave band.
Therefore, the oxygen is measured based on its paramagnetism. Inside the transducer of the O
2
module, there are two crystal balls full of nitrogen. They are suspended in the symmetrical magnetic field, and designed to point to the strongest outgoing part of the magnetic field. Outside the balls is the paramagnetic oxygen. Therefore, the balls are forced, by the relatively stronger paramagnetic oxygen, out of the magnetic field. The moment of the force acting on the balls is proportional to the paramagnetic strength as well as to the concentration of the paramagnetic oxygen.
24
Chapter 3 Product Specifications
Chapter 3 Product Specifications
3.1 Safety Classifications
Type of protection against electric shock
Degree of protection against electric shock
Class I with internal electric power supply.
Where the integrity of the external protective earth (ground) in the installation or its conductors is in doubt, the equipment shall be operated from its internal electric power supply (batteries)
Monitor:
Sidestream CO
2
/AG:
ECG/RESP/TEMP/SpO
2
/NIBP/IBP/
CO/mainstream CO
2
:
B
BF (defibrillation proof)
CF (defibrillation proof)
Degree of protection against hazards of ignition of flammable anesthetic mixtures
Degree of protection against harmful ingress of water
Mode of operation
Equipment type
Not protected (ordinary)
Not protected (ordinary)
Continuous
Portable
25
Chapter 3 Product Specifications
3.2 Environmental Specifications
Operating temperature
Operating humidity
Atmospheric pressure
Storage temperature
Storage humidity
0 to 40℃
5 to 35℃ (With Mindray CO
2
module)
10 to 35℃ (With Welch Allyn CO
2
module)
5 to 35℃ (With Oridion CO
2
module)
10 to 35℃ (With AION AG module)
15 to 95%, noncondensing
70.0 kPa~106.0 kPa
86.0 kPa~106.0 kPa(With Oridion CO
2
module)
–20 to 60℃
10 to 95%, noncondensing
26
Chapter 3 Product Specifications
3.3 Power Source Specifications
AC mains
Input voltage 100 to 240V
Frequency 50/60Hz
Power 140VA
DC mains
Input voltage 10 to 16VDC(rating-12 VDC)
Power 80W
Internal battery
Number of batteries
Type
Time to shutdown
Lithium battery
Rated voltage
Capacity
Operating time
Charge time
1
Sealed lead-acid battery or lithium-ion battery
5 to 15min (after the first low-power alarm)
Sealed lead-acid battery
Nominal voltage 12VDC
Capacity 2.3Ah
Operating time
Charge time
75 minutes typical when powered by one new fully-charged battery (25℃, ECG, SpO2, NIBP measurement per 15 minutes).
About 8 h (in the running status or standby mode)
11.1VDC
4.4Ah
150 minutes typical when powered by one new fully-charged battery (25℃, ECG, SpO2, NIBP measurement per 15 minutes).
About 8 h (in the running status or standby mode)
27
Chapter 3 Product Specifications
3.4 Hardware Specifications
Physical
Size
Weight
310 × 280 × 150mm (width×height×depth)
Different due to different configurations
Standard configuration: 5.5kg
Display
Size 10.4 inches (diagonal)
Recorder
Type
Horizontal resolution
Vertical resolution
Thermal dot array
160 dots/cm (at 25 mm/s recording rate)
80 dots/cm
Width of the recorder paper 50 mm
Length of the recorder paper 30 m
Recording rate
Recorded waveforms
25 mm/s, 50 mm/s
2
LED indicator
Alarm indicator
Running status indicator
AC/DC power indicator
Battery indicator
Audio indicator
1 (yellow and red)
1 (green)
1 (green)
1 (green)
Speaker
Giving audio alarms, keypad tones, and heartbeat/pulse tone.
Supporting PITCH TONE and multi-level volume.
Audio alarms comply with EN475 and IEC60601-1-8.
Control
Control knob
1 knob, which can be rotated clockwise/counterclockwise or pressed.
28
Chapter 3 Product Specifications
Button
7 buttons: POWER, MAIN, FREEZE, SILENCE, RECORD,
NIBP, MENU
Connectors
Power supply
Parameter
Network
VGA
Auxiliary output
Equipotentiality
1 AC power connector
1 DC power connector
ECG, RESP, TEMP, SPO2, NIBP, IBP, CO, CO2, AG
1 standard RJ45 network connector, 100 BASE-TX
1 standard color VGA monitor connector, 15-PIN D-sub
1 BNC connector
1 equipotential grounding connector
3.5 Wireless network
Standards
Frequenct range
Operating channel
Safe distance
Maximum data rate
IEEE 802.11b, Wi-Fi compatible
2.412 to 2.462GHz
China America Canada Europe Spain France Japan
1 to 11 10, 11 2
For other country, please refer to your local law.
10m (a circle centering AP with the diameter of 10m)
11Mbps
29
Chapter 3 Product Specifications
3.6 Data Storage
Trend data
Alarm events
ARR events
NIBP measurements
Long trend: 96 hours, resolution 1min, 5 min or 10 min.
Short trend: 1 hour, resolution 1 s or 5 s.
70 alarm events and associated waveforms (with user selectable waveform length 8s, 16 or 32).
80 ARR events and associated waveforms with 8s wavelength.
800 NIBP groups, including systolic pressures, mean pressures, diastolic pressures and measurement time.
30
Chapter 3 Product Specifications
3.7 Signal Output Specifications
Standards
Output impedance
Meets the requirements of EC60601-1 for short-circuit protection and leakage current
50Ω
ECG analog output
Bandwidth
(-3dB; reference frequency: 10Hz)
Diagnostic mode:
Monitor mode:
0.05 to 100Hz(12-lead: 0.05 to
150Hz)
0.5 to 40Hz
Signal delay
Surgery mode:
≤ 25ms
1 to 20Hz
Maximum propagation delay 25ms (In DIAGNOSTIC mode, NOTCH is OFF)
Sensitivity 1V/mV±5%
PACE rejection/enhancement No pace rejection or enhancement
IBP analog output
Bandwidth 0 to 12.5 Hz (-3dB, reference frequency: 1Hz)
Maximum propagation Delay 55ms (the filter function is disabled)
Nurse call output
Driver Relay
Electrical specifications ≤60W, ≤2A, ≤36VDC, ≤25VAC
Isolation voltage
Signal type
> 1500 VAC
Normally open or normally closed, selectable
Defibrillator synchronization pulse
Maximum time delay 35ms (R-wave peak to leading edge of the pulse)
Amplitude 3.5 V (min) at 3 mA sourcing; 0.8 V (max) at 1 mA sinking
Pulse width
Rising and falling time
100 ms ±10%
< 3ms
VGA
Signal
RGB: 0.7 Vp-p/75Ω;
Horizontal/vertical synchronization: TTL level
31
Chapter 3 Product Specifications
3.8 ECG Specifications
Lead type
Lead naming style
Sensitivity selection
Sweep speed
Bandwidth (– 3dB)
Common mode rejection
3-lead (1 channel):
5-lead (2 channels):
12-lead (8 channels):
I, II, III
I, II, III, aVR, aVL, aVF and V
I, II, III, avR, avL, avF, V1-V6
AHA, EURO
1.25mm/mV (×0.125), 2.5mm/mV (×0.25), 5mm/mV (×0.5),
10mm/mV (×1), 20mm/mV (×2) and auto
12.5mm/s, 25mm/s, 50mm/s
Diagnostic mode:
Monitor mode:
Surgery mode:
Diagnostic mode:
Monitor mode:
Surgery mode:
(The notch filter is turned off.)
0.05 to 100Hz (12-lead: 0.05 to 150Hz)
0.5 to 40Hz
1 to 20Hz
≥90 dB
≥105 dB
≥105 dB
Differential input impedance
Input signal range
DC offset voltage
Patient leakage current
Recovery time after defibrillation
Calibration signal
HR
≥ 5MΩ
±8mV (peak-to-peak value)
±300mV (12-lead: ±500mV)
< 10uA
< 5s
1mV (peak-to-peak value), precision:
±5%
Measurement range
Neonate:
Pediatric:
Adult:
15 to 350 BPM
15 to 350 BPM
15 to 300 BPM
Precision ±1BPM or ±1%, whichever is greater.
Sensitivity 200μV (lead II)
32
Chapter 3 Product Specifications
Response time to heart rate changes
Meets the requirement of ANSI/AAMI EC13-2002: Section
4.1.2.1 f).
Less than 11 sec for a step increase from 80 to 120 BPM
Less than 11 sec for a step decrease from 80 to 40 BPM
Response time of tachycardia alarm
When tested in accordance with ANSI/AAMI EC13-2002 Section
4.1.2.1 g, the response time is as follows.
Figure 4ah – range:
4a – range:
4ad – range:
Figure 4bh – range:
4b – range:
4bd – range:
15.7 to 19.2s, average: 17.4s
5.7 to 8.5s, average: 7.5s
3.6 to 5.1s, average: 4.2s
11.5 to 14.7s, average: 12.9s
4 to 14s, average: 7.2s
6.6 to 14.5s, average: 10.5s
Pace pulse
Pulse indicator
Pulse rejection
Pace pulses meeting the following conditions are marked by the
PACE indicator.
Amplitude:
Width:
Rise time:
±4 to ±700mV
0.1ms to 2ms
10us to 100µs
When tested in accordance with the ANSI/AAMI EC13-2002:
Sections 4.1.4.1 and 4.1.4.3, the heart rate meter rejects all pulses meeting the following conditions.
Amplitude:
Width:
Rise time:
Min. input slew rate:
±2 to ±700mV
0.1ms to 2ms
10us to 100µs
50V/s RTI
ST segment measurement
Measurement range – 2.0 to +2.0 mV
Precision
Update period
– 0.8 to +0.8mV:
Beyond this range:
10s
Arrhythmia analysis
Type
±0.02mV or ±10%, whichever is greater.
Undefined.
ASYSTOLE, VFIB/VTAC, PVC, COUPLET, VT>2,
BIGEMINY, TRIGEMINY, R ON T, MISSED BEATS, TACHY,
BRADY, PNC and PNP
33
Chapter 3 Product Specifications
3.9 RESP Specifications
Measurement technique Thoracic impedance
Lead
Differential input impedance
Optional: lead I and lead II; default lead II
> 2.5MΩ
Respiration impedance test range
0.3 to 3Ω
Excitation current < 300µA
Baseline impedance range 200 to 2500Ω (using an ECG cable with 1kΩ resistance)
Bandwidth
Sweep speed
0.2 to 2Hz (-3 dB)
6.25 mm/s, 12.5 mm/s, 25 mm/s
RR
Measurement range
Adult:
Pediatric/neonate:
0 to 120 BrPM
0 to 150 BrPM
Precision
Apnea alarm delay
7 to 150 BrPM:
0 to 6 BrPM:
10 to 40s
±2 BrPM or ±2%, whichever is greater.
Undefined.
34
Chapter 3 Product Specifications
3.10 SpO
2
Specifications
1.1.1. Mindray SpO
2
Module
SpO
2
Measurement range 0 to 100%
Resolution 1%
Precision
Refreshing rate
70 to 100%:
70 to 100%:
70 to 100%:
0% to 69%:
1s
±2 % (adult/pediatric, non-motion conditions)
±3 % (neonate, non-motion conditions)
±3 % (in motion conditions)
Undefined.
PR
Measurement range 20 to 254bpm
Resolution 1bpm
Precision
±3 bpm (non-motion conditions)
±5 bpm (in motion conditions)
1s Refreshing rate
1.1.2. Masimo SpO
2
Specifications
SpO
2
Measurement range 1 to 100%
Resolution 1%
Precision
Refreshing rate
70 to 100%:
70 to 100%:
70 to 100%:
0% to 69%:
1s
±2% (adult/pediatric, non-motion conditions)
±3% (neonate, non-motion conditions)
±3% (in motion conditions)
Undefined.
PR
35
Chapter 3 Product Specifications
Measurement range 25 to 240bpm
Resolution 1bpm
Precision
±3bpm (non-motion conditions)
±5bpm (in motion conditions)
1s Refreshing rate
1.1.3. Nellcor SpO
2
Specifications
SpO
2
measurement range and precision
MAX-A, MAX-AL, MAX-N,
MAX-P, MAX-I and MAX-FAST
OxiCliq A, OxiCliq N, OxiCliq P,
OxiCliq I
D-YS, DS-100A, OXI-A/N and
OXI-P/I
MAX-R, D-YSE and D-YSPD
70 to 100%
0% to 69%
70 to 100%
0% to 69%
70 to 100%
0% to 69%
70 to 100%
0% to 69%
±2%
Undefined
±2.5%
Undefined
±3%
Undefined
±3.5%
Undefined
PR measurement range and precision
20 to 250bpm: ±3bpm
251 to 300bpm: Undefined
1s Refreshing rate
*: When sensors are used on neonatal subjects as recommended, the specified precision range is increased by ±1%, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood.
36
Chapter 3 Product Specifications
3.11 IBP Specifications
Measurement technique
Displayed parameters
Mode of operation
Measurement interval in auto mode
Measurement time in continuous mode
Auto oscillation
Systolic pressure, diastolic pressure and mean pressure
Manual, auto and continuous
1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes
5 minutes
Measurement range in normal mode
Measurement precision
Systolic pressure
Diastolic pressure
40 to 270
10 to 210
Mean pressure 20 to 230
40 to 200
10 to 150
20 to 165
Maximum average error:
±5mmHg
Maximum standard deviation: 8mmHg
Resolution 1mmHg
Over-pressure protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
40 to 135
10 to 100
20 to 110
37
Chapter 3 Product Specifications
3.12 TEMP Specifications
Number of channels
Displayed parameters
Measurement range
Resolution
Precision
Update period
Minimum time for accurate measurement
2
T1, T2 and TD
0 to 50
°C (32 to 122°F)
0.1
°C
0.1
°C (excluding the sensor)
±0.2
°C (including the YSI 400 series sensor)
1s
Body surface: < 100s
Body cavity: < 80s
(YSI 400 series sensor)
38
Chapter 3 Product Specifications
3.13 IBP Specifications
Number of channels
Pressure labels
Measurement range
2
ART, PA, CVP, RAP, LAP, ICP, P1 and P2
ART 0 to 300 mmHg
PA
CVP/RAP/LAP/ICP
P1/P2
– 6 to 120 mmHg
– 10 to 40 mmHg
– 50 to 300 mmHg
Precision
Update period
Pressure transducer
Sensitivity
Impedance range
±2% or ±1mmHg, whichever is greater
1s
5 uV/V/mmHg
300 to 3000Ω
39
Chapter 3 Product Specifications
3.14 CO Specifications
Measurement technique
Calculated parameter
Measurement range
Resolution
Precision
Thermal dilution
CO, hemodynamics
CO
TB
0.1 to 20l/min
23 to 43
°C
TI
0 to 27
°C
CO:
TB, TI:
0.1 l /min
0.1
°C
CO:
TB, TI:
±5% or ± 0.1 l /min
0.1
°C
TB:
23 to 43
°C
Alarm range
40
Chapter 3 Product Specifications
3.15 CO
2
Specifications
Measurement technique Infrared absorption technique
Measurement mode
Displayed parameter
CO2 function
Sidestream, microstream or mainstream (optional)
EtCO2, FiCO2, Respiration Rate
Meet the requirements of EN864 and ISO9918.
1.1.4. Mindray CO
2
Specifications
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
100, 150ml/min
±2mmHg
±5%
±10%
Deflation rate
Precision of deflation rate 15%
Start-up time of CO2 module
< 1min.
The module enters the warming up status after the startup. Ten minutes later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
Response time
Delay time
Apnea alarm delay
0 to 70 BrPM:
> 70 BrPM:
< 240 ms (10 to 90%)
±2 BrPM
±5 BrPM
< 2s (Length of sampling tube: 7 feet; inner diameter: 0.055 inches; sampling flow: 150ml/min)
AwRR: 10 to 40 s
41
Chapter 3 Product Specifications
* Conditions for measurements in typical precision:
The measurement is started after the preheating mode of the module;
Ambient pressure: 750mmHg to 760mmHg; room temperature: 22 to 28 ;
The gas under test is dry, and the balance gas is N2;
The deflation rate is 150ml/min, the respiration rate is no greater than 30BrPM, with a fluctuation less than ±3BrPM, and the inhale interval/exhale interval is 1:2;
In other conditions, the measurement precision should meet the requirements of EN864 or
ISO9918: ±4mmHg (0 to 40mmHg) or ±12% of the reading (41 to 99mmHg)
1.1.5. Oridion CO
2
Specifications
CO
2
measurement range 0 to 99mmHg
Precision*
0 to 38 mmHg:
39 to 99 mmHg:
Resolution
Waveform:
Value:
±2mmHg
±5% + 0.08%× (reading - 38mmHg)
0.1mmHg
1mmHg
Flow rate
50
−
+
7.5
15 ml/min
Initialization time
Response time
Delay time
30s (typical)
2.9s (typical)
2.7s (typical)
AwRR measurement range 0 to 150 BrPM
AwRR measurement precision
0 to 70BrPM:
70 to 120BrPM:
121 to 150BrPM:
±1BrPM
±2BrPM
±3BrPM
Apnea alarm delay AwRR: 10 to 40s
* Precision applies for breath rates of up to 80 bpm. For breath rates above 80 bpm, accuracy complies with EN 864/ISO 9918 (4 mmHg or ±12% of reading whichever is greater) for EtCO
2 values exceeding 18 mmHg. To achieve the specified accuracies for breath rates above 60 breaths/minute, the Microstream® FilterLine H Set for Infant/Neonatal (p/n 006324) must be used.
The accuracy specification is maintained to within 4% of the values indicated in the above table in the presence of interfering gases according to EN864 Section Eleven, Part 101.
42
Chapter 3 Product Specifications
1.1.6. Welch Allyn CO
2
Specifications
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
Resolution
Refreshing rate
1mmHg
1s
Start-up time
Response time transducer: 5W)
100ms (10% to 90 %)
±2mmHg
±5%
±10%
Calibration stability
There is a difference (< 1%) from the precison criteria after a
12-month continuous service time
0 to 99 mmHg Alarm range
AwRR measurement range 0 to 150 BrPM
AwRR alarm range 0 to 150 BrPM
Apnea alarm delay AwRR: 10 to 40 s
60 cc/min.
43
Chapter 3 Product Specifications
3.16 AG Specifications
Measurement technique
Measurement mode
AG functions
Warm-up time
Sampling flow
(sidestream)
Gas type
Measurement range
Resolution
Precision
Infrared absorption
Side stream
Meets requirements of ISO9918, ISO11196, EN12598 and
ISO7767
45 seconds (warming-up status)
10 minutes
(ready-to-measure status)
Adult/Pediatric
Neonatal
120, 150, 200 ml/minute (user-selectable)
70, 90, 120 ml/minute (user-selectable)
CO
2
, N
2
O, O
2
(optional), Des, Iso, Enf, Sev, Hal
CO
2
:
N
2
O:
Des:
Sev:
Enf, Iso, Hal:
O
2
:
AwRR:
0 to 30%
0 to 105%
0 to 30%
0 to 30%
0 to 30%
0 to 105%
2 to 100 BrPM
CO
2
:
AwRR:
Gas
1 BrPM
Range (%REL)
CO2
0 to 1
1 to 5
5 to 7
7 to 10
Precision (%ABS)
±0.1
±0.2
±0.3
±0.5
N2O
Des
> 10
0 to 20
20 to 100
0 to 1
1to 5
Not specified
±2
±3
±0.15
±0.2
44
Chapter 3 Product Specifications
5 to 10
10 to 15
15 to 18
±0.4
±0.6
±1
Sev
Enf, Iso, Hal
O2 (Optional)
AwRR
0 to 1
1 to 5
5 to 8
> 8
0 to 1
1 to 5
> 5
0 to 25
25 to 80
80 to 100
2 to 60 BrPM
> 60 BrPM
±0.15
±0.2
±0.4
Not specified
±0.15
±0.2
Not specified
±1
±2
±3
±1 BrPM
Not specified
Alarm range
Apnea alarm delay
CO2:
AwRR:
AwRR:
0 to 10 % (0 - 76 mmHg)
2 to 100 BrPM
20 - 40 s
Refreshing rate 1s
Calibration
Calibration stability
Rise time (10 % to 90 %)
Sampling flow 120ml/min, using the DRYLINE™ water trap and neonatal
DRYLINE™ sampling line (2.5m)
Yearly calibration requested.
After module being used for 12 consective months, the error is <
1%
CO
2
250 ms (fall time 200 ms)
N
2
O
O
2
600ms
HAL, ISO, SEV, DES
ENF
250 ms
300 ms
350 ms
Rise time (10 % to 90 %)
Sampling flow 200ml/min, using the DRYLINE™ water trap and adult
DRYLINE™ sampling
CO
2
N
2
O
HAL, ISO, SEV, DES
250 ms (fall time 200 ms)
250 ms
O
2
500ms
300 ms
45
line (2.5m)
Delay time
Chapter 3 Product Specifications
< 4s
46
Chapter 4 Disassembling/Assembling & Troubleshooting
Chapter 4 Disassembling/Assembling
& Troubleshooting
4.1 PM-7000 Disassembling/Assembling
4.1.1 Exploded View of PM-7000
Figure 4-1 Exploded view of PM-7000
NO
1
Material code
7001-30-67413
Part & Specification
Front bezel assembly
2 0000-10-10876
3
4
7001-30-67452
7001-30-67450
5 900E-20-05085
Hose (φ1.5mm)
Main support assembly
Rear housing assembly
Label
6 M04-000305--- Cross head tapping screw PT3×12
7 M04-051003--- Cross head tapping screw PT2×6
8 M02A-3025913 Water trap socket assembly
9 M02A-20-25906
10 M04-004012---
11 TR6C-30-16654
12 M04-004014---
13 M04-051002---
Cross head screw
Cross head screw M4×10
Cross head screw M3×16
47
Quantity
1
0.33
1
1
1
6
6
1
1
2
1
2
4
Chapter 4 Disassembling/Assembling & Troubleshooting
4.1.2 PM-7000 Support Assembly
Figure 4-2 PM-7000 support assembly
NO
1
Material Code
0000-10-11021
Part & Specification
LCD, TFT, 10.4 inch
2 7000-20-24508
3 M04-051137--- Stainless steel cross head screw M2×4
4 7000-20-24540 LCD support, LG
5 M04-002405---
6 7000-20-24417
Cross head screw M2×6
Backlight board insulation plate
7 0010-10-12096
8 8000-21-10239
Inverter
Connecting wire for TFT backlight board
9 M90-000002-01 Insulation flat washer
Quantity
1
2
4
1
2
1
1
1
2
48
Chapter 4 Disassembling/Assembling & Troubleshooting
10
11
9300-20-13901
8000-20-10214
12 M03A-30-26050
13 7000-20-24418
14 7000-30-24425
15 7000-30-24424
16 DA8K-20-14426
17
18
9006-30-33900
812A-30-08569
19 M04-060009---
Backlight insulation plate
IBP insulation plate
Insulation plate for main board
1
1
1
1
1
3
1
1
1
1
20 7001-20-67403
21 7000-20-24509
22 M04-051121---
23 M90-000002---
24 M02A-30-25930
25 0000-10-11002
26 9201-30-35920
27
29
32
7001-30-67468
28 M04-002505---
7001-30-67454
30 M04-021024---
31 M04-003012---
9200-21-10442
33
34
8000-21-10141
7001-30-67438
35 7000-20-24387
36 7001-20-67470
37 9210-30-30150
38 9901-10-23920
9006 SPO2 board
812A ECG board assembly
Bolt M3×14
CO
2
module mounting board
Insulation plate for ECG board
Cross head screw M2.5×8
Insulation flat washer
CO
2
module
Wireless network card
CF card interface board
Flexible cable for CF card
Cross head screw M3×6
Power socket assembly
Washer
Cross head screw M3×6
Backlight board connecting wire
Power cord for DC/DC converter
Lead acid battery power board
Power board insulation sheet
1
1
1
1
1
4
4
1
4
1
3
39
1
1
1
1
1
1
0.1m
39 7000-20-24365
40 7000-20-24420
41 7000-20-24366
42 M04-030031---
43 M04-000603---
44 M04-000306---
45 630D-30-09111
Screw for fixing battery latch
Hex bolt M3×20
1
1
1
1
1
1
1
51 7000-20-24442
Bolt M3×16
630D PCB assembly
46 A21-000009---
47 7000-20-24539
48 7000-20-24446
49 7000-20-24447
50 9200-21-10446
Hose
Screen support 1 for LG screen
Communication wire to ECG/TEMP/RESP board
0.4m
1
1
1
1
Communication wire to SPO
2 board 1
49
Chapter 4 Disassembling/Assembling & Troubleshooting
52
53
54
M02A-20-25909
7000-20-24444
7000-20-24439
55 7000-20-24438
56 7000-20-24500
57
58
7000-20-24443
9200-21-10454
Communication wire to CO
2
module
Communication wire to IBP module
Communication wire to power board
Signal wire to recorder
Communication wire to NIBP board
TFT connecting wire
4.1.3 Front Bezel Assembly
1
1
1
1
1
1
1
Figure 4-3 Front bezel assembly
SN P/N
1 7000-20-24354-51
2 9200-30-10701
3 7000-20-24373
4 7000-20-24358
Description
Alarm light board
Alarm light shade
6 7000-20-24369
7 900E-20-04895
8
9
900E-20-04894
6200-30-09774
Knob
Dust pad 2
Dust pad 1
Screen encoder mount
50
Qty
1
1
2
2
1
1
1
1
1
Chapter 4 Disassembling/Assembling & Troubleshooting
10 7000-20-24356-51
11 7000-20-24368
12 M04-003905---
13 DA8K-20-14416
14 7001-30-67441
15 M04-003105---
4.1.4 Rear Housing Assembly
Cross head tapping screw PT3×6
Connecting wires for encoder
Cross head tapping screw PT3×8
1
1
2
1
1
6
SN P/N
1 7000-21-24435
2 M04-000301---
3 7000-20-24435
4 9200-30-10522
5 7000-20-24370
6 M04-000802---
7 M04-000104---
Figure 4-4 Rear housing assembly
Description
Stainless steel nut
8 7000-20-24359
9 DA8K-10-14410
10
14
9200-21-10633
11 9200-20-10620
12 7000-20-24388
13 7000-20-24361
7000-20-24376
Handle
Buffer
Speaker and connecting wires
Battery door link
51
Qty
1
2
2
1
2
3
2
1
2
1
1
1
1
1
Chapter 4 Disassembling/Assembling & Troubleshooting
15 9200-20-10622
16 M04-003105---
17 7000-20-24377
4.1.5 Microstream CO
2
Assembly
1
Cross head tapping screw PT3×8 11
Cushion 2
NO
1
2
3
4
5
6
M04-003105---
9201-20-36010
9201-20-35961
9201-20-35915
9201-20-35914
Figure 4-5 Microstream CO
2
assembly
Material Code Part & Specification
Cross-head self-tapping screw 3*8
Baffle of torsional spring
Retaining torsional spring of microstream CO
2 connector
Mounting plate of CO
2
connector
Baffle of CO
2
connector
52
Chapter 4 Disassembling/Assembling & Troubleshooting
4.2 Troubleshooting
4.2.1 Black Screen, Startup Failure
Y
Figure 4-6 Location flow of faults causing black screen
53
Chapter 4 Disassembling/Assembling & Troubleshooting
4.2.2 White Screen & Other Abnormal Screen
In case of faults causing white screen or other abnormal screens,
■
Check whether the LCD connection wires are in good contact;
■
Replace the LCD connection wires, or replace the LCD if necessary;
■
Replace the main control board if the fault still exists.
4.2.3 Encoder Faults
1. If all other functions (indicator, alarm, buttons) of the button panel are normal, proceed to step 2; otherwise, replace the button panel;
2. Check whether short-circuit or abnormal open-circuit occurs in the encoder pad;
3. Replace the encoder.
4.2.4 No Audio Alarm
1. Check whether the audio alarm function is disabled in the software settings;
2. Replace the speaker;
3. Replace the button panel.
4.2.5 Printing Failure
1. Check whether there is any alarm about the recorder. If any, eliminate it;
2. Check whether the recorder indictor is on;
3. If not, check the connection wire for inputting signals to the recorder;
4. Check whether the recorder module is enabled in the maintenance menu;
5. Check the power cord of the recorder (including the recorder power PCB);
6. Replace the recorder module.
4.2.6 Abnormal Paper Drive
1. Check whether there are blocks on the paper roller of the recorder;
2. Check whether there are blocks in the gear cluster of thermal assembly of the recorder;
3. Check whether the voltage input of the recorder is larger than 17.6V.
54
Chapter 5 Test and Material List
Chapter 5 Test and Material List
5.1.1 Connection and Checking
Connect the simulators, power supply and test fixture properly to the PM-7000, and power it on. Then, the patient monitor displays the start-up screen on the TFT screen and enters the system screen.
5.1.2 Functions of Buttons
Press every button on the button panel to check their functions as specified in PM-7000
Patient Monitor Operation Manual. Rotate the control knob to check its functions.
5.1.3 ECG/RESP
The TFT screen displays the standard ECG waveform, and the error between the heart rate and the set value of the simulator is no more than ±1, namely 60±1; the RESP waveform is smooth, and the respiration rate is 20±1.
1. Select all leads in order, including Cal, select all the four gains and AUTO, ensure the waveforms are displayed properly, and check whether the 50Hz/60Hz interference can be filtered.
2. Check, in all the above-mentioned cases, the consistency between the heartbeats, the flashes of the red heart-like indicator, and the R-wave.
3. The gain has no impact on the message “ECG signal over weak” in the HR calculation.
4. Verify the range and precision: Suppose that the amplitude of the GCG signal of the simulator is 1mV, the heart rates are respectively 30, 60, 120, 200, 240 and
300. Check leads I, II and III. The results should meet 29-31, 59-61, 119-121,
198-202, 238-242, and 297-303.
5. PACE pulse test: Set the simulator to PACE. You should be able to view the pace.
Change PACE amplitude to ±8 – 700mv, and pulse width to 0.1ms – 2ms. The
PACE should be legible, and LEAD OFF is displayed properly.
6. RESP measurement: Set the baseline impedance to 1K, the respiration impedance to 0.5Ω and 3Ω, and the respiration rate to 30 and 120. The respiration rate should be 29 – 31, 118 –122.
7. PVC test: Set the simulator to the PVC mode, and set the occurrence times. The relevant PVCS should be obtained.
8. Set the simulator as follows: RR: 40, baseline impedance: 2KΩ, RESP waveform:
3:1. Open the apnea alarm, set the respiration resistance to 0Ω, and set various alarm time. Alarms should be given.
55
Chapter 5 Test and Material List
5.1.4 Temperature
1. YSI probe
Select YSI probe from the manufacturer menu, select YSI temperature probe as the test fixture, set the analog resistance to 1.471K, 1.355K and 1.249K. Then the TEMP parameter should be 35±0.1 , 37±0.1 and 39±0.1 .
2. CY-F1 probe
Select CY-F1 probe from the manufacturer menu, select CY-F1 temperature probe as the test fixture, set the analog resistance to 6.534K, 6.018K and
5.548K. Then the TEMP parameter should be 35±0.1 , 37±0.1
℃
and
39±0.1 .
5.1.5 NIBP
Connect the NIBP simulator, adult cuff and accessories, and then connect the module
CUFF and clockwise screw it tightly.
1. After the simulator self-test, press <ENT> to enter the ADULT analog blood pressure mode. Set the blood pressure to the 255/195/215 mmHg level, SHIFT to +15, and the HR to 80BPM. Set PM-7000 to the adult mode. Press <START>.
Then the results will be obtained in about 30s. The measured results should be respectively 270±8mmHg, 210±8mmHg and 230±8mmHg.
2. Press <ESC> and <↓> on the simulator to enter the NEONATE mode. Set the blood pressure to the 120/80/90 mmHg level, HR to 120bmp, and PM-7000 to the pediatric mode. Press <START>. Then the results will be obtained in about
30s. The measured results should be respectively 120±8mmHg, 80±8mmHg and 90±8mmHg.
3. Press <ESC> and <↓> on the simulator to enter the NEONATE mode. Set the blood pressure to the 60/30/40 mmHg level, SHIFT to -20, HR to 120bmp, and
PM-7000 to the neonate mode. Change the simulator accessory to the neonatal cuff. Press <START>. Then the results will be obtained in about 30s. The measured results should be respectively 40±8mmHg, 10±8mmHg and
20±8mmHg.
5.1.6 SpO
2
Select PLETH as the HR source of PM-7000, and put the finger into the SpO
2
sensor.
The screen should display the PR and SpO
2
values normally. The normal SpO
2
value is above 97%.
5.1.7 IBP
1. Test fixture
Physiological signal simulator
2. Test procedure
① IBP1 test:
Set the BP sensitivity of the ECG simulator to 5uv/v/mmHg, BP to 0mmHG, and the
IBP channel 1 to ART. Enter the IBP PRESSURE ZERO menu of the PM-7000, zero
Channel 1, and then return to the main screen. Set the BP of the simulator to
200mmHg. Enter the IBP PRESSURE CALIBRATE menu of the PM-7000, conduct
56
Chapter 5 Test and Material List calibration, and then exit the IBP PRESSURE CALIBRATE menu.
Set the BP value of the simulator respectively to 40mmHg, 100mmHg and
200mmHg. Then the screen of the PM-7000 should display 40±1mmHg,
100±2mmHg and 200±4mmHg.
Set the simulator output to ART wave. Then the screen of the PM-7000 should display relevant waveform properly.
Unplug the IBP probe. Then the screen should prompt “IBP: Transducer 1 OFF!” and “IBP: Transducer 2 OFF!”
Plug the OHMEDA cable to the IBP1 channel. Then the prompting message “IBP:
Transducer 1 OFF!” disappears.
②
IBP2 test:
Plug the IBP cable to the IBP2 channel, and repeat the procedure in Section .
5.1.8 CO
1. Test fixture
Physiological signal simulator
2. Test procedure
Injectate and blood temperature test: Assemble the TB and TI test fixture, output
36.0±0.1 , 37.0± ℃ ℃
2.0±0.1 .
CO measurement: Set the CO.CONST and T
I set the injectate switch to OFF, and then press START. Then the simulator will and 5±0.5L/M.
5.1.9 CO
2
1. Test fixture
CO
2
steel bottle (containing 10% CO
2
)
2. Test procedure
①
Mainstream CO
2
measurement: Set the calculation compensation of PM-7000 to
COMMON.
Plug the mainstream transducer to the CO
2
socket, connect the windpipe connector with the CO
2
steel bottle, and open/close the valve of the CO
2
steel bottle based on the interval of 3s. The CO
2
value should be the calibration gas pressure value:
76±5%mmHg. When the valve is opened permanently, the patient monitor prompts
“APNEA ALARM”.
Unplug the transducer. The patient monitor prompts “CO
2
transducer OFF” on the main screen. Plug the transducer again. The patient monitor prompts “CO
2 transducer pre-heated”.
②
Sidestream CO
2
measurement: Set the calculation compensation of PM-7000 to
COMMON.
Plug the water trap to the water trap socket, connect the sampling tube with the CO
2 steel bottle, and open//close the valve of the CO
2
steel bottle based on the interval of
3s. The CO
2
value should be the calibration gas pressure value: 76±5%mmHg. When
57
Chapter 5 Test and Material List the valve is opened permanently, the patient monitor prompts “APNEA ALARM”.
Unplug the water trap. The patient monitor prompts “CO
2
water trap OFF”. Plug the water trap again. The prompting message disappears.
When the measured value exceeds the high limit of CO
2
, the patient monitor prompts “CO
2
too high” on the main screen. When the measured value is lower than the low limit, the patient monitor prompts “CO
2
too low”.
5.1.10 Recorder
1. Print the ECG waveform. The recorder should print it normally and clearly. Set the recorder to the fault of lack of paper and abnormal clip. There should be relevant prompting messages on the main screen. When the fault is cleared, the patient monitor should become normal.
2. Print the alarm messages of all parameters. Set the alarm print switch to ON for all parameters, and set different alarm limits. Then the recorder should print the alarm message in case of an alarm
.
5.1.12 Power Supply
When the patient monitor is supplied with the external AC power, the CHARGE indicator becomes ON. When it is disconnected from the external AC power, the
CHARGE indicator becomes OFF. After the patient monitor is started without assembling the batteries, “x” is displayed in the battery indication frame on the main screen. After the batteries are assembled, the battery electricity is displayed in the battery indication frame on the main screen. The patient monitor can work normally with or without batteries. It, however, should give an alarm when the batteries are exhausted.
5.1.13 Clock
Verify the correctness of the clock in the system test, and then set the clock to the current time.
5.1.14 System Test
Load all parameters, and conduct operations respectively on the loaded parameters.
During the synchronization, no exceptions (for example, mutual interference) occur.
Set all parameter setups in menus to the default values which are those at the time of software loading, and conduct operations on the menus, for example, managing the patient information, recalling data, and so on. All the operations should be done normally, and the corresponding functions should be correct and meet the product requirements.
58
Chapter 5 Test and Material List
5.2 NIBP Calibration
Figure 5-1 NIBP Calibration
Calibration method:
Based on the precision of 50mmHg (6.7kPa), increase the pressure step by step. The maximum error at any pressure point within the NIBP measurement range of the patient monitor should be no more than ±3mmHg (±0.4kPa). Decrease the pressure step by step.
The maximum error at any pressure point within the NIBP measurement range of the patient monitor should be no more than ±3mmHg (±0.4kPa).
5.3 IBP CALIBRATE
5.3.1 IBP Transducer Zero
Press the ZERO button on the IBP module to call up IBP PRESSURE ZERO menu as shown below:
Figure 5-2 IBP PRESSURE ZERO
59
Chapter 5 Test and Material List
Zero Calibration of Transducer
Select CH1, the system will zero IBP1. Select CH2, the system will zero IBP2.
Cautions:
( Use the PM-6000 IBP module as a example)
Turn off patient stopcock before you start the zero procedure.
The transducer must be vented to atmospheric pressure before the zero procedure.
The transducer should be placed at the same height level with the heart, approximately mid-axially line.
Zero procedure should be performed before starting the monitoring and at least once a day after each disconnect-and-connect of the cable.
Figure 5-3 IBP Zero
5.3.2 IBP Calibration
Press CAL button on the IBP module to call up the IBP PRESSURE CALIBRATE menu as shown below:
60
Chapter 5 Test and Material List
Figure 5-4 IBP Calibration Menu
Calibrate the transducer:
Turn the knob to select the item CH1 CAL VALUE, press and turn the knob to select the pressure value to be calibrated for channel 1. Then turn the knob to select the item
CALIBRATE to start calibrating channel 1.
Turn the knob to select the item CH2 CAL VALUE, press and turn the knob to select the pressure value to be calibrated for channel 2. Then turn the knob to select the item
CALIBRATE to start calibrating channel 2.
The pressure calibration of PM-7000:
Pressure transducer
3-way stopcock
Pressure transducer interface cable
T-shape connector
Monitor
Hydrargyrum pressure meter
Figure 5-5 IBP Calibration
You will need the following pieces of equipment:
•
Standard sphygmomanometer
•
3-way stopcock
•
Tubing approximately 25 cm long
61
Chapter 5 Test and Material List
The Calibration Procedure
:
1. Close the stopcock that was open to atmospheric pressure for the zero calibration.
2. Attach the tubing to the sphygmomanometer.
3. Ensure that connection that would lead to patient is off.
4. Connect the 3-way connector to the 3-way stopcock that is not connected to the patient catheter.
5. Open the port of the 3-way stopcock to the sphygmomanometer. .
6. Select the channel to be calibrated in the menu and select the pressure value to which the
IBP is to be adjusted.
7. Inflate to make the mercury bar rise to the setup pressure value.
8. Adjust repeatedly until the value in the menu is equal to the pressure value shown by the mercury calibration.
9. Press the Start button, the device will begin calibrating.
10. Wait for the calibrated result. You should take corresponding measures based on the prompt information.
11. After calibration, disassemble the blood pressure tubing and the attached 3-way valve.
Calibration completion message: “SUCCESSFUL CALIBRATE”
5.4 CO2 CHECK
Check procedure for sidestream module only
Via the PM-7000’s system and maintain menus you are prompted for a password for entering the factory key. After entering the password you get access to the pump rate settings and to check the accuracy of the CO2 measurement. Using the below test set up to verify the accuracy of the CO2 module.
Figure 5-6 Sidestream test set up
62
Chapter 5 Test and Material List
Note
Neither the mainstream nor the sidestream module can be calibrated. Only the overall performance and accuracy is checked. If the Co2 module fails the tests it should be replaced.
Figure 5-7 Factory Maintain Menu
Figure 5-8 CO2 check menu
63
5.5 AG CALIBRATE
Chapter 5 Test and Material List
5.5.1 AG Check
1、Using T-piece to connect the watertrap and Agent steel bottle well.
One of the T-piece ports must be vented to atmospheric pressure.
2、Select ‘MEASURE’ from work mode item in “AG SETUP” menu, then set pump rate ‘low’ and wait for 10 minutes after the warm up information disappears.
3、Enter ‘CALIBRATE’ menu, then open AG bottle and press the ‘VERIFY ACCURACY’ item.
Figure 5-9 AG Check Menu
4、Observe the display value after 1 minute. The agent concentration accurate should be less than
±5%.
5、Choose other pump rate ‘middle’ or ‘high’,and repeat the previous procedures.
(pump rate definition: three pump rate under adult mode: 100/150/200ml/min; neonate:
70/90/110 ml/min)
6、If the accurate over range, please press ‘START CAL’.
64
Chapter 5 Test and Material List
5.5.2 AG CALIBRATE
(
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40% )
1、Press ‘START CAL’, then input password ‘MINDRAY’ entering ‘CALIBRATE’ menu.
Note: Make sure the AG in ‘Measure’ mode not ‘Standby’ mode before you do calibrate.
Figure 5-10
Figure 5-11
2、Input each gas concentration value according to the label on the AG bottle label.
Note: If your monitor do not have O2 module, input ‘0.0’ in O2 item.
3、Open AG cover, wait for the display value stabilization.
4、If the display value does not accord with the input value, please press ‘CALIBRATE’ item and exit.
AG concentration must fit the following requirements:
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40%
65
Chapter 5 Test and Material List
5.6 Bill of Materials for PM-7000 Main Unit
SN P/N
1 M04-004012---
Description Quantity
Cross-head screw M3*6 39
2 TR6C-30-16650 Recorder drive board 1
5 7000-20-24422 Main support (with lithium battery) 1
6 M04-002505---
7 7001-30-67428
8 9210-30-30150
Cross-head screw M3*6 10
CF wireless network card module assembly 1
Main board 1
9 7001-30-67438 Power board for lead acid battery 1
10 M05-302R3R--- Lead acid battery 1
11 0010-10-12329
12 DA8K-20-14520
13 812A-30-08557
Insulation sheet for ECG board
812A ECG board
1
1
1
14 M04-060009---
15 DA8K-20-14426
16 9006-30-33900
17 630D-30-09121
18 7000-30-24511
19 0010-10-12274
20 7000-20-24387
1
SPO2 board mount 1
SPO2 board 1
1
Nellcor SP02 module 1
MASIMO SPO2 module 1
Insulation sheet for power board 1
21 7001-30-67439
22 7001-30-67464
23 7001-30-67461
24 7001-30-67458
Power board for lithium battery 1
AG module (with O
2)
1
AG module (without O
2)
1
Mainstream CO2 module 1
25 9000-20-07289 Mainstream CO2 module mounting board
26 7001-30-67441
27 9200-30-10701
1
Button board 1
Alarm light board
1
28 6200-30-09774 Display encoder mounting board
29 7001-20-67403 CO2 module mounting board
1
1
30 M05-010R03---
31 7001-30-67456
Button cell battery
Display assembly
32 M08A-30-34703 Digiboard for 12 lead ECG module
33 M08A-30-34701 Analog board for 12 lead ECG module
1
1
1
1
34 7001-30-67436
35 7000-30-24574
Oridion CO2 module
Mindray CO2 module
1
1
66
Chapter 6 Maintenance and Cleaning
Chapter 6 Maintenance and Cleaning
6.1 Maintenance
6.1.1Checking Before Using
■
Check the patient monitor for mechanical damages;
■
Check all exposed conductors, connectors and accessories;
■ Check all functions that are possibly enabled for the monitored patient, and ensure the device is in good working status.
In case of any damage, stop using this patient monitor, and contact biomedical engineers of the hospital or Mindray maintenance engineers.
6.1.2 Regular Checking
An all-around check, including the safety check, should be done by qualified personnel every 6-12 months or after maintenance each time.
All checks in which the patient monitor should be disassembled should be done by qualified maintenance personnel. The safety and maintenance checks can be done by
Mindray engineers. The local office of Mindray at your region will be pleased to provide you with the information about the maintenance contract.
6.2 Cleaning
Do switch off the patient monitor and disconnect the AC power supply before cleaning it or the probes.
The PM-7000r should be dust free. To clean the surface of its enclosure and screen, use the cleaning agent that is not corrosive, for example, soap and water.
1. Do not use strong solvent, such as acetone;
2. Most cleaning agents must be diluted before being used, so conduct dilution under the instruction of manufacturers;
3. Do not use any erosive material (such as steel wool or polishing agent);
4. Prevent the ingress of any liquid to the enclosure and any part of the device;
5. Ensure no residue of cleaning liquid on the surface of the device.
6.3 Cleaning Reagent
1. Diluted aqua ammonia
2. Diluted sodium hypochlorite (bleaching powder for washing)
3. Hydrogen peroxide 3%
4. Ethanol
5. Isopropyl alcohol
67
Chapter 6 Maintenance and Cleaning
6.4 Sterilization
To avoid the long-time damage to the patient monitor, we recommend you
9 To conduct only sterilization which is considered necessary in your maintenance plan;
9
To clean the patient monitor before the sterilization;
9 To sterilize the patient monitor with specified sterilization agent: Ethylate, and
Acetaldehyde.
Caution
Conduct dilution or use the liquid of the possibly-lowest concentration under the instructions by the manufacturer.
Prevent the ingress of liquid to the enclosure.
Prevent any part of the system from being dipped.
In sterilization, do not spill the liquid to the patient monitor.
Ensure no residue of sterilization agent on the surface of the patient
monitor. Clean it if any.
6.5 Disinfection
To avoid the long-time damage to the patient monitor, we recommend you
9 To conduct only disinfection which is considered necessary in your maintenance plan;
9 To clean the patient monitor before the disinfection;
Gas (EtO) or formaldehyde are forbidden for the disinfection of the patient monitor.
68
P/N:
7001-20-67407(1.0)
Advertisement
Key Features
- Portable and accessible
- Rechargeable batteries or external AC/DC power
- Adults, pediatric and neonates
- Different configurations
- Central monitoring system
- ECG, RESP, SpO2, NIBP, TEMP, IBP, CO, CO2 and AG
- Waveform monitoring, freezing and recording
- Compact and lightweight
- Color TFT LCD
- Compact control panel