Intell-Check Service Manual

© Intelligent Weighing Technology Inc. 2008 All rights reserved Worldwide.

The information contained herein is the property of Intelligent Weighing Technology and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.

Intell-Check Service Manual SME 300000111

TABLE OF CONTENTS

INDICATOR HOUSING ......................................................................................... 2

BEFORE USING THE SCALE.............................................................................. 3

CHAPTER 1 INTRODUCTION.............................................................................. 4

1-1 Features....................................................................................................... 4

1-2 Specifications ............................................................................................... 4

1-3 Scale Appearance ........................................................................................ 5

1-4 Power Supply ............................................................................................... 5

1-5 Explanation of Display.................................................................................. 6

1-6 Explanation of Keyboard .............................................................................. 7

1-7 Self-diagnosis Mode..................................................................................... 9

1-8 Error Message ........................................................................................... 10

1-9 Weight Unit ................................................................................................ 10

CHATPER 2 SERVICE MODE ACCESS ............................................................ 11

2-1 Capacity Configuration

2-1-1 Weight Units Setting

2-1-2 Customized Weight Unit Setting

2-1-3 Dividing Point Setting of Each Unit

.................................................... 12

................................................. 13

(no resolution limit).. 15

............................... 21

2-2 Weight Calibration

2-2-1 Weight Calibration Setting

........................................................... 22

........................................... 23

2-2-2 Manufacturing Location Gravity Setting

2-3 Linearity Calibration

..................... 24

......................................................... 25

2-4 Function Setting .............................................................. 27

2-4-1 Parameter Setting for the Approval Model ................... 28

2-4-2 Zero Tracking Range Setting

2-4-3 Parameter Setting for 24C08

..................................... 29

....................................... 30

2-4-4 Tuning On Zero Range Setting

2-5 Gravity Setting

.................................. 31

................................................................ 32

Appendix 1 RS232 SERIAL INTERFACE.......................................................... 33

Appendix 2 BUTTON CODE.............................................................................. 41

Appendix 3 7 SEGMENT DISPLAY CHARACTERS.......................................... 42

Intell-Check Service Manual 1 SME300000111

INDICATOR HOUSING

Before opening the stainless steel indicator housing, make sure it is dry.

If necessary wipe the housing dry, paying particular attention to the joint between the front panel and rear housing.

The re-assembly instructions should be accurately followed to ensure the IP rating is maintained in service.

NOTE: There are no user serviceable parts inside the indicator and all service work should be carried out by the technical staff of your supplier.

RE-ASSEMBLY INSTRUCTIONS:

Position the front panel onto the rear housing and fit all six fixing screws finger tight.

Tighten the screws in the order shown in the diagram below to a torque of 12 kgf-cm.

NOTE: Over tightening the screws could impair the IP rating.

Tightening order:

3

rd

1

st

6

th

5

th

Intell-Check Service Manual

2

nd

2

4

th

SME300000111

BEFORE USING THE SCALE

To enable you to use this scale correctly, we suggest that you read this manual carefully.

INSTRUCTIONS FOR USE

1. The load placed on the weigh pan must NOT exceed the maximum weighing capacity of the scale.

2. Protect the scale from high temperatures.

3. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the weigh pan to any strong shock loads.

PREPARING TO USE THE SCALE

1. Put the scale on a firm level surface for accurate weight readings.

2. Protect the scale from direct sunlight and drafts.

3. Use a separate power socket, if possible, when re-charging the indicator.

4. For maximum stability and accuracy switch on the indicator for 15~20 minutes before operation.

5. Once the scale has been powered on, it will go through a LCD display test and re-zero, ready for use.

6. All goods weighed / counted should be placed in the centre of weigh pan for best accuracy. The overall dimensions of the goods being weighed / counted should not exceed the dimensions of the weigh pan.

7. Please note when recharged.

symbol flashes on the display, the battery needs to be

Intell-Check Service Manual 3 SME300000111

CHAPTER 1 INTRODUCTION

1-1 Features

IP 67 waterproof:

moisture-proof, mist-proof, with IP67 waterproof and dustproof

High resolution:

the maximum external resolution is 1/30000, and the internal is 1/300000 (Input voltage should be over 6mV)

Adjustable weighing speed:

three weighing speed modes (fast, moderate,

 and slow) provided

Selectable units modes:

Kilogram (kg), gram (g), pound (lb), and ounce (oz) weighing modes

HI, OK, LO indication:

Large LCD display with HI, OK, LO indication and the signals for Relay output with max. current 7A, max. voltage 265Vac, 30Vdc

Practical functions:

10 preset settings, Percentage display, Simple / Limit /

Target operation modes, Pre-tare function, Net/Gross switch function, Auto backlight, RTC setting and Auto power-off timer etc.

Others:

Auto calibration, Auto zero tracking, Low battery indication, Adjustable

 gravity value, and Linearity calibration.

Built-in bi-directional RS-232 interface

1-2 Specifications

Analog Input and A/D Conversion: Input Sensitivity 0.18

V/d or more.

Input Signal Range: -2mV~+16mV

Adjustable Zero Range: -1mV~+5mV

Load Cell Excitation: 5V DC

Non-linearity: 0.01% of full scale

A/D Resolution: 1,000,000 counts (Max.)

Intell-Check Service Manual 4 SME300000111

1-3 Scale Appearance

256 mm

170 mm

The package includes:

1.Indicator

1 off

2.Power Cable 1 off

3.User Manual 1 off

When you unpack the indicator and find any of the above items are missing, please contact your supplier.

90 mm

RS-232 interface connection

Load cell connection

Power cable

Only the cables with

ψ

3~

ψ

5.5mm

caliber can be used, otherwise the waterproof design will be damaged

1-4 Power Supply

Power Supply Selection

1. 6V / 4.5Ah Rechargeable battery

2. AC adaptor DC 9V

Power Consumption

Approximately DC 19 mA (Indicator)

Approximately DC 38 mA (Indicator + Display backlight)

Approximately DC 70~80 mA (Indicator + Display backlight + Relay)

Low Battery Warning

Please note when the ( ) symbol keeps flashing on the display, the batteries should be recharged.

Intell-Check Service Manual 5 SME300000111

1-5 Explanation of Display

LIMIT indication

LCD1

Analog indication column with maximum

50 graduation

LCD2

Digital Display column with maximum 6 digits

HI, LO, OK status indication

Zero point indication

NET indication

Battery indication

Stable reading indication

GROSS indication

TARGET indication

In the Simple mode, if the weight is over the maximum graduation, the “+” zone will be flashing.

PRE TARE indication Weigh unit indication: kg, g, lb, lb-oz, %

Current memory reading

(HI, LO, OK, RANGE) data

Keypad

Intell-Check Service Manual 6 SME300000111

1-6 Explanation of Keyboard

Press to switch the indicator On/Off.

Unload the weight pan, and then press the key to zero the scale.

Zero setting is possible only within a certain range. The auxiliary function of the zero key is the tare function.

Press this button to clear the setting or the values entered from the keypad

.

Press this button to accept data entered from the keypad.

Digit: 9

Press this button to switch between the “Net value” and the “Gross value” (This button is functional only under the tare or pre-tare modes).

When the indication shows upon “GROSS” on the display, all the keypads are not functional except the key and the key.

Digit: 6

Press this button to deduct the weight of the container via entering the weight value directly.

Digit: 3

Press this button to deduct the weight of container.

Press this button to recall the memory.

Intell-Check Service Manual 7 SME300000111

Digit 8

Press this button to output the displayed values via the built-in interface into a printer or a computer.

Digit 5

In Target mode, press this button to pre-set a “Range” of weight values, which are acceptable upper and lower limits.

(target weight + upper range

≒ range

≒ target weight – lower range).

Digit 2

In Limit mode, press this button to input “Low” value, so that the weighing result cannot be smaller than this value.

Press this button to store HI, LO, OK and Range settings or exit the setting mode.

Digit 7

Press this button to choose different weighing modes, including Simple,

Limit, and Target weight modes.

Digit 4

In Target mode, press this button to input weight value which meets the user’s requirement.

Digit 1

In Limit mode, press this button to input “HI” value, so that the weighing result cannot be greater than this value.

Digit 0

Press this button to select the different weight units.

Intell-Check Service Manual 8 SME300000111

1-7 Self-diagnosis Mode

(1) Set the jumper SWA1 on the main board to the ADJ position.

Press

ON/OFF

key to test the LCD display if the LCD display normally.

(2) Press

ON/OFF

key, and hold

MODE

key then press

ON/OFF

key.

The LCD shows and enter the Self-diagnosis mode.

Press

MR

Press

MR

Press

MR

Press

MR

Internal value mode

(Before testing, the Load cell should be connected)

Press

Enter

key and the LCD shows the internal value.

(1) Check the internal value is in the normal range (the internal value should be in the range of 100000 ~ 400000 when free of loading)

(2) Check the backlight

Keypad testing mode

Press

Enter

key and the LCD shows

Refer to Appendix 2 Button code for the keypad code

Press

MS

key to back to the menu level

Software version

Press

Enter

key and the LCD shows the software version

Press

MS

key to back to the menu level

Exit the self-diagnosis mode

Press

Enter

key to exit the self-diagnosis mode and the system restarts.

Intell-Check Service Manual 9 SME300000111

1-8 Error Message

EEPROM is not on board or the wiring around EEPROM is malfunctioned.

Zero is higher than the zero rang when switching the indicator on.

Zero is lower than the zero rang when switching the indicator on.

Internal value is unstable.

ADIC malfunction (cannot read AD value). The cable of the load cell does NOT connected appropriately to the indicator. EX: Sensing inputs

(+/-) are not connected to the load cell.

The weight of the object is over the max. capacity + 9 division

ADIC is over the maximum resolution range.

1-9 Weight Unit

(kg)

(g)

(lb)

(lb,oz)

(oz)

1 g = 0.001 kg

1 g = 1 g

1 g = 0.002204623 lb

1 g = 0.03527396 oz

1 g = 0.03527396 oz

Intell-Check Service Manual 10 SME300000111

CHATPER 2 SERVICE MODE ACCESS

Set the jumper SWA1 on the main board to the ADJ position (EEPROM UNLOCK), Switch the scale on. The display showing calibration.

means the scale is ready for the

After finishing the calibration, please set the jumper SWA1 to the LOCK position.

If the user set the jumper SWA1 to the LOCK position during the calibration, the scale will exit the service mode automatically.

Press the

MR

key

Press the

MR

key

Press the

MR

key

Press the

MR

key

Pres the s

MR

key

Capacity Configuration

Weight Calibration

Linearity Calibration

Intell-Check Service Manual 11

Function Setting

Gravity Setting

SME300000111

2-1 Capacity Configuration

Capacity configuration includes

Weight Units Setting

Customized Weight Units Setting

Dividing Point Setting of Each Unit Range

Press

ENTER

key

Input a parameter 00,

01, 02, 03

Input “00”,

Press

ENTER

key to return to the upper level

C

key

Clear the parameters

ENTER

key

Press to confirm

Intell-Check Service Manual

Return to the Upper Level

Weight Units Setting

Customized Weight Units Setting

Dividing Point Setting of Each Unit

12 SME300000111

2-1-1 Weight Units Setting

Press

ENTER

key

Refer to “Note 1” for operating

Press

ENTER

key

Refer to “Note 2” for operating

Press

ENTER

key

Refer to “Note 3” for operating

Press

ENTER

key

.

.

kg

The display shows the max.

capacity of the scale + minimum division and the unit weight for 2 seconds.

Example:

Choose “kg”, “lb” two unit weights and “kg” is the pre-set weight unit for weight calibration

Example:

015000

15.000 kg

kg

For example:

1

 minimum division

3

 decimal point position

2

 range setting

The display shows the max. capacity of the scale + minimum division and the unit weight for 2 seconds.

Input the parameter “00”

Press

ENTER

key to complete the setting

Intell-Check Service Manual 13

C

key

Clear the parameters

ENTER

key

Press to confirm

MS

key

Back to

SME300000111

Note 1

(a) (b) (c) (d) (e) (f)

The users can set up the different weight units in various orders according to their preference.

The description of the parameters

0

 kg Parameter “0” represents “kg” weight unit

1

 g

2

 lb

Parameter “1” represents “g” weight unit

Parameter “2” represents “lb” weight unit

4

 lb, oz

Parameter “4” represents “lb/oz” weight unit

5

 oz

Parameter “5” represents “oz” weight unit

(a)

The first weight unit (only “kg”, “g”, or “lb” are available to choose from. Please select one of the parameters 0, 1, or 2)

(b)

The second weight unit (select one of the parameters 0 ~ 5)

(c)

The third weight unit (select one of the parameters 0 ~ 5)

(d)

The fourth weight unit (select one of the parameters 0 ~ 5)

(e)

The fifth weight unit (select one of the parameters 0 ~ 5)

(f)

The amount of the weight units selected (depends on the total number of units)

For example:

Choose “kg”, “lb” two unit weights and “kg” is the pre-set weight unit for weight calibration.

In this step, the parameters “

020002”

should be input.

Note 2

(g) (h ) ( i ) ( j ) ( k ) ( l )

Please enter the maximum capacity of the scale, total are 6 digits (exclude 9d)

For example:

15.000 kg

 enter 015000 1500.0 g

 enter 015000 6.000 lb

 enter 006000

Note 3

(m) ( n) ( o) ( p)

(m)

Minimum division selection( select one of parameters 1, 2, or 5)

(n)

The decimal point position ( 0 ~ 5)

For example:

15.000 kg

 enter 3 1500.0 g

 enter 1 6.000 lb

 enter 3

(o)

 range setting (select one of parameters 0, 1, 2, or 3)

For example:

0

 full range, 1

 full range, 2

 dual range(divided at 1/2 of the full scale), 3

 triple range( divided at 1/3 of the full scale, 2/3 of the full scale )

(p)

Parameter description: 0

 multi-interval, 2

 multi-range

Intell-Check Service Manual 14 SME300000111

2-1-2 Customized Weight Unit Setting

Press

ENTER

key

Refer to “Note 4” for operating

Press

ENTER

key

Refer to “Note 5” for operating

Press

ENTER

key

.

(no resolution limit) kg

The display shows the max.

capacity of the scale + minimum division set preciously for 2 seconds.

The display shows the amount of weight units and the weight unit for the weight calibration set previously

For example:

5

5 weight units

2

“lb” is the weight unit for weight calibration

Show the code of weight unit for weight calibration for 2 sec

“ lb ”

“ g ”

“ kg ”

For example:

012000

12.000 lb

Press

ENTER

key

Refer to “Note 6” for operating

Press

ENTER

key

.

Cont

Intell-Check Service Manual

lb

Please key in the mass value

012000

12.000 lb

For example:

5

 minimum division

3

 decimal point position

2

“ lb ”

0

 full range / 0

 multi-interval

1

 save the setting at the first position

The display shows the max. capacity of the scale + minimum division set preciously for 2 seconds, then back to upper level

C

key

Clear the parameters

ENTER

key

Press to confirm

MS

key

Back to

15 SME300000111

Cont

Press

ENTER

key

.

lb

The display shows the max.

capacity of the scale + minimum division set preciously for 2 seconds.

Refer to “Note 5” for operating

Press

ENTER

key

Refer to “Note 5” for operating

Press

ENTER

key

Refer to “Note 6” for operating

Press

ENTER

key

Set all weight units according to the preference

For example:

006000

6000 g

Please key in the mass value

12lb = 5443 g

5443g

005443

For example:

1

 minimum division

0

 decimal point position

1

“ g ”

0

 full range / 0

 multi-interval

2

 save the setting at the second position

g

The display shows the max.

capacity of the scale + minimum division set preciously for 2 seconds.

After all weight units desired have been set

Input the parameter “00”

Press

ENTER

key to complete the setting

Intell-Check Service Manual 16

C

key

Clear the parameters

ENTER

key

Press to confirm

MS

key

Back to

SME300000111

Note 4

(a) (b)

(a)

The amount of the weight units (Max: 5, key in 1 ~ 5)

(b)

The weight unit for weight calibration (choose from “kg”, “g”, “lb”, key in 0, 1, or 2)

Parameter description:

0

 kg, 1

 g, 2

 lb

Note 5

(c) (d ) ( e) (f ) ( g) ( h)

c ~ h

set the max. capacity in 6 digits

(i) (j ) (k ) ( l ) ( m) ( n )

i ~ n

set the mass value for weight calibration in 6digits

The max. capacity needs to be presented based on decimal system.

For example: How to calculate the maximum capacity and the mass value based on the different types of weight unit.

A. Choose “lb” as the weight unit for weight calibration

The first weight unit setting: 12.000 l b / 0.005 lb (the max. capacity/ division)

Enter the max capacity 012000 at (c) ~ (h)

Enter the mass value 012000 at (i) ~ (n)

The second weight unit setting “g” :

0.002204623 lb = 1 g

12 lb = 5443 g. Take l 6000 g as the max capacity

6000 g / 2 g (the max. capacity/ division),

Enter the max capacity 006000 at (c) ~ (h)

Enter the mass value 005443 at (i) ~ (n)

”lb/oz” (hexadecimal notation system):

12.

0.00 lb / 0.05 oz (the max. capacity/ division)

12 lb = 12

Enter the max capacity 019200 at (c) ~ (h)

Enter the mass value 019200 at (i) ~ (n)

”oz” :

12 lb = 192.00 oz ,

Take 200.00 oz as the max. capacity

200.00 oz / 0.05 oz (the max. capacity/ division)

Enter the max capacity 020000 at (c) ~ (h)

Enter the mass value 019200 at (i) ~ (n)

Intell-Check Service Manual 17 SME300000111

B. Choose “kg” as the weight unit for the weight calibration

The first weight unit setting: 6.000 kg / 0.002 kg

Enter the max capacity 006000 at (c) ~ (h)

Enter the mass value 006000 at (i) ~ (n)

The second weight unit setting “lb/oz” (hexadecimal notation system):

0.001 kg = 0.002204623 lb

6 kg = 6 2.204623 (lb) = 13.227738 lb. Take 12 lb as the max capacity

12.

12 lb

13.227738 lb

Enter the max capacity 019200 at (c) ~ (h)

16 (oz) = 211.64 oz

Enter the mass value 021164 at (i) ~ (n)

Intell-Check Service Manual 18 SME300000111

Note 6

(o)

Minimum division

Parameter description:

Decimal system:

Input 1, 2, or 5 as the minimum division for the weight value

Hexadecimal notation system:

Input 1, 2, 4, or 8 as the minimum division for the weight value

(p)

Decimal point position

Parameter description:

Decimal system:

0

1

2

3

0

0.0

0.00

0.000

4

0.0000

5

0.00000

Hexadecimal notation system:

0

1

2

3

0.

0.

0.

(q)

Weight unit displayed

Parameter

Unit

Notation system symbol

0 1 2 4 kg g lb lb,oz

5 oz

10 10 10 16 kg g lb

10 lb

The 7 th

 indication

0.

(r)

Dividing point (Input the parameter 0, 1, 2, or 3)

Parameter description:

0

 full range

1

 full range

2

 dual range( divided at 1/2 of the full scale )

3

 triple range( divided at 1/6 of the full scale

、 and 2/3 of the full scale)

(s)

Save the weight units at preferred positions (no more than the number of the set weight units)

Parameter description:

1

 the first position (the weight calibration unit)

2

 the second position

3

 the third position

4

 the fourth position

5

 the fifth position

(f)

Parameter description: 0

 multi-interval, 2

 multi-range

Intell-Check Service Manual 19 SME300000111

Modify “Customized Weight Units” Setting

Press

ENTER

key

Refer to “Note 4”

Press

ENTER

key

.

lb

The display shows the max.

capacity of the scale + minimum division set preciously for 2 seconds.

The display shows the amount of weight units and the weight unit for the weight calibration set previously

For example:

5

5 weight units

2

 set “lb” as weight unit for weight calibration

Show the code of weight unit for weight calibration for 2 sec

“ lb ”

“ g ”

“ kg ”

For example:

019200

192.00 oz

Refer to “Note 5”

Press

ENTER

key

Refer to “Note 5”

Press

ENTER

key

Refer to “Note 6”

Press

ENTER

key

Users can either exit the setting or continue to modify it

Intell-Check Service Manual

.

.

lb

20

Enter the mass value

12 lb = 192 oz

019200

192.00 oz

For example:

5

 minimum division

2

0.

0.00

4

“ lb ”

0

 full range / 0

 multi-interval

3

 save at the third position

The display shows the max. capacity of the scale + minimum division set preciously for 2 seconds., then back to upper level.

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

SME300000111

2-1-3 Dividing Point Setting of Each Unit

Press

Enter

key

Select unit parameter

Press

Enter

key

Enter “the 1st dividing point”

Press

Enter

key

Enter “the 2nd dividing point”

Press

Enter

key

Select the unit intended to set dividing point:

All units set in and can be set dividing point. Refer to “Note 1” about unit parameter.

Screen displays the default of the first point:

Dual range

 dividing point 1/2 full scale

Triple range

 dividing point 1/6 full scale

2/3 full scale

Enter the 1 st dividing point intended:

Only dual and triple range models can do the 1 st dividing point setting.

Screen displays the default of the first point:

Triple range

 dividing point 1/ 6 full scale

2/3 full scale

Enter the 2 nd dividing point intended:

Only triple range model can do the 2 nd dividing point setting.

Flow the steps to finish the rest of the setting

Press

Enter

key to complete the dividing point setting

After all dividing points has been set, enter the parameter “00”

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

Setting through and , if the model is dual range, the original dividing point is 1/2 full scale; if the model is triple range, the original dividing point is 1/6 or 2/3 full scale. The dividing point of each unit can be re-set in .

After completed

, if or again, the dividing point will recover the original default.

is re-set

Intell-Check Service Manual 21 SME300000111

2-2 Weight Calibration

Press

TARE

key

Input a parameter “00”, “01”, or ”02”

After input “00”, press

ENTER

key to return to the upper level. (Exit from weight calibration)

C

key

Clear all parameters

ENTER

key

Press to confirm

Return to Upper Level

Weight Calibration Setting

Manufacturing Location Gravity Setting

Intell-Check Service Manual 22 SME300000111

2-2-1 Weight Calibration Setting

The value of the mass for weight calibration is better equal to the max. capacity for better performance of weight calibration

The display shows the calibration mass value and the weight unit set at the first position

The calibration mass value can be modified according to the user’s needs

Press

ENTER

key

kg

.

Unload the weight pan

Press

ENTER

key

Press

MR

key.

The indicator reads zero point.

kg

After 1 seconds

Press

MR

key

kg

Place full load (30kg in this example) on the weight pan then press

ENTER

key

Remove the weight from the weight pan and press any key to exit the setting.

The indicator shows the mass value. The buzzer sounds “beep, beep, beep” when stable.

Unload the weight pan.

Input a parameter “00”

.

.

kg

After 2 seconds

Press

ENTER

key to complete weight calibration setting

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

Intell-Check Service Manual 23 SME300000111

2-2-2 Manufacturing Location Gravity Setting

The gravity value is between the equator and the polar region.

Gravity value at the equator G

E

= 9.7803184558 m/sec

2

Gravity value at the polar region G

P

= 9.8321772792 m/sec

2

Taipei

9.78914 m/sec

2

Shanghai

9.79423 m/sec

2

Press

Factory: Taipei

Press

ENTER

key

Factory: Shanghai

ENTER

key

Press

ENTER

key

.

.

Factory: Other regions

Press

ENTER

key

.

Display the gravity value set previously

Re-set the gravity value

.

Press

ENTER

key

Input a parameter “00”

Press

ENTER

key to complete the gravity setting

Intell-Check Service Manual 24

C

Clear all parameters

ENTER

Press to confirm

MS

Back to

SME300000111

2-3 Linearity Calibration

After completing the linearity calibration, the weight calibration needs to be re-done again.

The weight value for linearity calibration should be at least equal to the weight value for weight calibration.

The maximum point of linearity calibration is up to 9. The more points of linearity calibration are done, the better stability of linearity performance is.

* Multiple of the mass value

Example 1: linearity calibration for 15kg (the point of linearity is 5)

1. Remove all objects on the load cell then press the

MR key to recode the “first point”

2. If the weight of the first mass is 1kg (the first mass value must be a factor of the full capacity, ex: 1, 3 or 5), the multiple of the mass is 1. Press

MR key to record the “second point”.

3. If the weight of the second mass is 5kg, the second mass is 5 times value of the first mass.

Input 5 and press

MR key to record the “third point”.

4. The weight of the third mass is 5kg, the third mass is 5 time value of the first mass. Input 5 and press

MR key to record the “fourth point” and save the setting.

5. The weight of the third mass is 5kg, the third mass is 5 time value of the first mass. Input 5 and press

MR key to record the “fourth point” and save the setting.

6. The weight of the third mass is 1kg, the third mass is 1 time value of the first mass. Input 1 and press

ENTER key to record the “fifth point” and save the setting.

Note:

During linearity calibration step

During linearity calibration step and exit the setting.

, or , press

ENTER

key to abort the setting.

During linearity calibration step

ENTER

, , , , , or ,, press key to save 2, 3, 4, 5, 6, 7, or 8 points linearity calibration and exit the setting.

, press either

MR to save 9 points linearity calibration

Intell-Check Service Manual 25 SME300000111

Unload the weight pan. Press

ENTER

key to enter the linearity calibration mode

Input a parameter “01”

Press

ENTER

key

Remove the steel weight pan and all objects on the load cell, then press

MR

key to record the “first point”

Place the first mass (1kg), then input

“1” as the multiple of the mass value,

Press

MR

key to record the “second point”

Add the 2 nd mass (5kg), then input “5” as the multiple of the mass value,

Press

MR

key to record the “3 rd point”

Add the 3 rd mass (5kg), then input the multiple of the mass value,

Press

MR

key to record the “4 th point”

Add the “4 th mass (1kg), then input the multiple of the first mass value.

Press

ENTER

key to confirm 5 points of linearity calibration

Input a parameter “00”

Press

ENTER

key to complete the linearity calibration

Intell-Check Service Manual 26

* Multiple of mass value

Add the second mass (5kg), then input “5” as the multiple of the first mass value mass value (refer to

EX1, p 24). Press

MR

key to record the “third point”

SME300000111

2-4 Function Setting

Parameter setting for the approval model

Zero point tracking range

Parameter setting for 24C08

Tuning on zero range setting

Press

ENTER

key

Input a parameter 00, 01, 02,

03, or 04

After inputting “00”,

Press

ENTER key back to the upper level

C

key

Clear all parameters

ENTER

key

Press to confirm

Intell-Check Service Manual

Back to Upper Level

Parameter Setting for the Approval Model

Zero Point Tracking Range

Parameter Setting for 24C08

Tuning On Zero Range Setting

27 SME300000111

2-4-1 Parameter Setting for the Approval Model

Press

ENTER

key

Input a parameter “00”,

“01”, “02”, or “03”

Press

ENTER

key

Input the parameter “00”

Press

ENTER

key to exit the setting

Parameter description

00

Non-approval model

01

OIML or NTEP approval model, external weight calibration prohibited

02

Sri Lanka non- approval model, external weight calibration allowed

03

Sri Lanka approval model, external weight calibration prohibited

The models allow external weight calibration will not detect E1 and E2.

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

Intell-Check Service Manual 28 SME300000111

2-4-2 Zero Tracking Range Setting

Press

ENTER

key

Input a parameter

“00” ~ “04”

Press

ENTER

key

Input the parameter “00”

Press

ENTER

key to complete the setting

Parameter description

00

 default, <1/4 e/ sec

01

< 1/2e / sec

02

< 1 e / sec

03

< 2 e / sec

04

< 4 e / sec

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

Intell-Check Service Manual 29 SME300000111

2-4-3 Parameter Setting for 24C08

Press

ENTER

key

Input a parameter

“00”

“01”

Press the

ENTER

Input the parameter “00”

Press

ENTER

key to complete the setting

Parameter description

00

NOT connect 24C08 to the indicator

01

 connect 24C08 to the indicator

If the indicator is not connected to 24C08, there is NO 11 th item for selection in setting.

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

Intell-Check Service Manual 30 SME300000111

2-4-4 Tuning On Zero Range Setting

Press

Enter

key

Input the parameter

“00” ~ “09” by pressing numerical keys.

Press

Enter

key

Input parameter “00”

Press

Enter

key to complete the setting

Turning on zero range setting:

Screen displays the parameter set previously

Turning on zero range setting:

Use

ON/ZERO

or

UNITS

key to input parameter:

Default setting = 0

(OIML or NTEP Proved Model)

Default setting = 9

(Non-approval Model)

00

Zero range :

10% full scale

01

Zero range :

20% full scale

02

Zero range :

30% full scale

03

Zero range :

40% full scale

04

Zero range :

50% full scale

05

Zero range :

60% full scale

06

Zero range :

70% full scale

07

Zero range :

80% full scale

08

Zero range :

90% full scale

09

Zero range :

100% full scale

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

OIML or NTEP approved model ( the parameter set as “1” in

Turning on zero range:

10% full scale

Non-approved model (the parameter set as “0” in

Turning on zero range:

100% full scale

)

)

Intell-Check Service Manual 31 SME300000111

2-5 Gravity Setting

Press

ENTER

key

Input the parameter 01

Press

ENTER

key

.

Input gravity value of user location.

.

Press

ENTER

key

Input the parameter 00

Press

ENTER

key to complete the setting

C

key

Clear all parameters

ENTER

key

Press to confirm

MS

key

Back to

The gravity value setting must be within gravity value at the equator and pole region.

Gravity value at the equator: G

E

= 9.7803184558 m/sec

2

Gravity value at the Pole region: G

P

= 9.8321772792 m/sec

2

Taipei

9.78914 m/sec

2

Shanghai

9.79423 m/sec

2

Intell-Check Service Manual 32 SME300000111

Appendix 1 RS232 SERIAL INTERFACE

RS232 Command format:

(1) The Host (PC) writes the data to the EEPROM of Slave (Scale).

(2) The Host (PC) reads the data in the EEPROM of Slave (Scale), and the Slave

(Scale) has the data feedback to the host (PC).

Content planning

(1)

W , , , CR LF

Write EEPROM address

( 00 ~ 58 )

Length of storage

(

(2)

R ,

Read

EEPROM address

( 00 ~ 58 )

CR LF

1 2 3 ……………….

n

ASCII code n = length of storage

Write above two commands to the Host (PC), the Slave has the following format feedback to the

Host (PC).

L e n g t h : CR LF

Length of storage

A S C I I : CR LF

S y m b o l : 1

1

2

Store ASCII code n = length of storage

3

2

… …

3 …………… n

… … … n CR LF

Symbol corresponding ASCII code n = length of storage

Intell-Check Service Manual 33 SME300000111

10

11

12

13

28

29

30

31

24

25

26

27

14

15

16

17

18

19

20

21

22

23

Code

04

05

06

07

08

09

00

01

02

03

First EEPROM address

08

10

12

14

16

18

00

02

04

06

25

27

29

34

38

40

42

44

45

46

47

48

49

50

51

52

53

54

55

56

57

58

Item

Gross

Net

Tare

Pre Tare

Target Weight

Range

High Limit

Low Limit

EEPROM address

19

20

21

22

23

EEPROM address

Percent (%)

Weight units

First unit EEPROM address

Second unit EEPROM address

Third unit EEPROM address

Forth unit EEPROM address

Fifth unit EEROM address

Time

Date

Lo

Ok

Hi status

30

31

32

EEPROM address

35

36

Lo EEPROM address

Ok EEPROM address

Hi EEPROM address

Stable (Unstable) sign

Stable sigh EEPROM address

Unstable sign EEPROM address

Accumulations quantity

Accumulations total weight

Check the x weight

User’s definition 1

User’s definition 2

User’s definition 3

User’s definition 4

User’s definition 5

User’s definition 6

User’s definition 7

User’s definition 8

User’s definition 9

User’s definition 10

User’s definition 11

User’s definition 12

User’s definition 13

User’s definition 14

User’s definition 15

Final

EEPROM address

09

11

13

15

17

24

01

03

05

07

26

28

33

37

39

41

43

Intell-Check Service Manual 34 SME300000111

Format

(1)

W , ,

Write

EEPROM address

( 59 ~ 63 )

Code numbers

( < 15 )

(2)

R ,

Read

EEPROM address

( 59 ~ 63 )

CR LF

,

1 2 3 ……………….

n code n =number

CR LF

Write above two commands to the Host (PC), the Slave has the following format feedback to the

Host (PC).

L e n g t h : CR LF

Code numbers

I t e m : CR LF

1 2 3 ……………….

n code n = number

Formatted:

Font: 10 pt

Intell-Check Service Manual 35 SME300000111

EEPROM address

59

60

61

62

63

Format

Simple Format

Limit Format

Target Format

Transmission of accumulation clear

Control code

Transmission

Press “Print” button in Simple mode

Press “Print” button in Limit mode

Press “Print” button in Target mode

Press “Enter” button

Transmit one time when turn on the scale

Code function

Code+80

Do not print the first and final letters

Code+40

Display zero by blank

Code+B0

Do not print the first and final letters and display zero by blank

Example:

Tare Code

03

03+80 = 83

Do not print the first and final letters of Tare

03+40 = 43

Display zero by blank, such as 000.00kg

03+B0 = B3

Do not print the first and final letters, and display zero when the content of Tare value is blank.

Intell-Check Service Manual 36 SME300000111

Command Mode

Command Format A

Host

Slave

Command

Command

MZ Zero

MT Tare

MG Gross weight

MN Net weight

MM Switch operation mode

CT Clear Tare value

SM

SO

UA

UB

UC

UD

Manually transmit

Command mode

Switch to the first unit

Switch to the second unit

Switch to the third unit

Switch to the forth unit

CP Clear Pre Tare value UE Switch to the fifth unit

SA Auto transmit %

Stop continuous transmission and enter the command mode

SC Continuous transmission

Note : UA ~ UE settings are dependent on the model of the scale

Command Format B

Host Command

Slave Data

RW Read current weight RH Read Gross (simple)

RG

RN

RT

RB

Read Gross weight

Read Net weight

Read Tare

Read current display of weight

(simple)

RI Read Net (simple)

RJ

Read comparison situation + current display of weight (simple)

RK

RL

Read comparison situation +

Gross (simple)

Read comparison situation + Net

(simple)

RE Read Pre Tare

Note :

i. add % before the command to read continuously ii. add # before the command to transmit a stable value

Read Weight-comparison setting value

RS





: Group number(00 ~ 09)



: Setting items

LO

HI

Read LO setting value

Read HI setting value

RG

PT

Read Range setting value

Read Pre-tare setting value

OK Read Target setting value

Note

: There is no group number in PT (PT only has one group setting)

EX:

RSPT

CR

><

LF

ANS:

RSPTXXXXXX

CR

><

LF

EX:

RS03RG

CR

><

LF

Read Pre-tare setting value

Read Range setting value

Intell-Check Service Manual 37 SME300000111

ANS:

RS03RGXXXXXX

CR

><

LF

Command Format C

Host Command+ Data

Slave Command+ Data

Write the weight comparison value WS

○○□□



: Group number(00 ~ 09)



: Setting items

LO

HI

Write LO setting value

Write HI setting value

RG

PT

Write Range setting value

Write Pre-tare setting value

OK Write Target setting value

Note

: There is no group number in PT (PT only has one group setting)

EX:

WSPTXXXXXX

CR

><

LF

ANS:

WSPTXXXXXX

CR

><

LF

Write Pre-tare setting value

EX:

WS03RGXXXXXX

CR

><

LF

Write Range setting value

ANS:

WS03RGXXXXXX

CR

><

LF

Error messages:

E1: Wrong command

E2: Command format error (Wrong parameters)

E3: Command not recognized

Intell-Check Service Manual 38 SME300000111

Output Data Format

Weight format

Gross S T , G S , + 0 1 2 3 4 5 6 7 SP SP o z

Net S T , N T , + 1 .

2 3 .

4 5 6 t l .

g

Tare S T , T R , + 0 1 2 .

3 4 5 6 SP SP k g

Plus OL O L , G S , + SP SP SP SP SP SP SP SP SP SP SP SP CR LF

Minus OL O L , G S , - SP SP SP SP SP SP SP SP SP SP SP SP

Unstable U S , G S , + 0 1 2 3 4 .

5 6 SP SP l b

Pre Tare S T , P T , + 0 1 2 3 4 5 6 7 SP SP SP g

Simple format

Gross + 0 1 2 3 4 5 6 7

Net + 1 .

2 3 .

4 5 6

Tare + 0 1 2 .

3 4 5 6

Plus OL + SP SP SP SP SP SP SP SP

CR LF

Minus OL - SP SP SP SP SP SP SP SP

Comparison status + Simple format

Byte0 Byte1 Byte2 +/1 .

2 3

Byte0 : HI 30H/31H

Byte1 : OK 30H/31H

Byte2 : LO 30H/31H

.

4 5 6 CR LF

Intell-Check Service Manual 39 SME300000111

Serial Data Transfer/Receive Format

n , 8 , 1 S

LSB

8-bit data e , 7 , 1 o , 7 , 1

S

S

7-bit data

7-bit data

MSB

STOP

1

0

P STOP

1

0

P STOP

1

0

Notes:

S : Start bit

STOP: Stop bit

P : Parity bit

Intell-Check Service Manual 40 SME300000111

Appendix 2 BUTTON CODE

MODE PRINT

OK RANGE

NET

GROSS

PRE-

TARE

ON

OFF

ZERO

HI LO TARE C

UNITS MS MR ENTER

08 07 06

ON/OFF

18 17 16 15

38 37 36 35

28

Exit the code

26 25

Intell-Check Service Manual 41 SME300000111

Appendix 3 7-SEGMENT DISPLAY CHARACTERS

Digit

7 segments letter

Alphabet

7 segments letter

Alphabet

7 segments letter

0 A N

3

4

5

1

2

6

7

8

9 J

I

K

L

M

D

E

F

B

C

G

H

V

W

X

Y

Z

Q

R

S

O

P

T

U

Intell-Check Service Manual 42 SME300000111