Allied Healthcare Products Aridyne 3500 medical air compressor OPERATION/SERVICE MANUAL
The Allied Healthcare Aridyne 3500 is a medical air compressor designed for heavy workloads. It features a special quieting material and drying system to eliminate noise and reduce condensation in air lines. The compressor provides cool, clean air and is easy to move with its casters.
PDF
Download
Document
Advertisement
Advertisement
OPERATION/SERVICE MANUAL ARIDYNE 3500 MEDICAL AIR COMPRESSOR FOR 120 VOLT, 60 HZ MODELS FORM NO. D-006 ARIDYNE 3500 DECEMBER 1992 Allied Healthcare Products, Inc. TIMETER INSTRUMENT CORPORATION READ ALL INSTRUCTIONS BEFORE OPERATING OR SERVICING THE ARIDYNE 3500 INTRODUCTION Allied Healthcare's ARIDYNE ™ Air Compressors are designed to withstand a heavy workload. Their construction includes both a special quieting material and drying system. The special quieting material in the ARIDYNE ™ eliminates annoying pump and fan noise. The drying system reduces condensation in air lines and provides cool, clean air. Casters make the air compressor easy to move from room to room. TABLE OF CONTENTS Specifications .................. 2...” .... 1 Precautions .............. 00... 00.0. 000oreo. 1 Important Unpacking Instructions ................ 2 Controls and Other Features .................... 2 Operation ........ 22... ea aa ea ane 2 Routine Maintenance .......................... 3 Service os 4 internal View of the Aridyne 3500 ............. 5 ReplacementParts ............................ 6 Troubleshooting ............................... 7 WINING ...............0.00arraciaoeareconev. 8 WARRANTY Because the ARIDYNE ™ 3500 Air Compressor is quality designed, Allied Healthcare Products, Inc. promises to repair or replace any defective parts at no cost, including labor, for one full year from date of purchase. This is the extent of AHP responsibility. All consequential damages are specifically excluded. PRECAUTIONS e DO NOT use the compressor in the presence of flammable anesthetics. eo CAUTION: To prevent electrical shock hazard refer all servicing to qualified service persons. ® Electrical components inside the compressor system are isolated from ground. e The air compressor system is grounded. DO NOT plug it into ungrounded outlets or adaptors. e® The compressor is of the oil-less type and MUST NOT be lubricated. ® Never put oxygen into the compressor. The electrical components are not approved for oxygen use. SPECIFICATIONS Output coer, 42 pm at 50 + 5 psig at 5°F dew point depression POWET ..............eneenenernes 120 Volt, 60 Hz 1 Phase, 8.5 Amr Power Cordes. #16-3 x 9'4" (2.8m) (with hospital grade plug DIMENSIONS: WIA eee, 11 1/2" (29 cm) DEP cee 15 3/4" (40cm) Height... eee vere, 33 1/2" (87 cm) Wei cee eee verona 90 165. (41 Ка) Outlets over, Two Grounded DISS Male Air Outlets W/Internal Check Valves Includes ee 10 Amp Internal Circuit Breaker Precautions Continued e It is recommended that you install a transparent water trap at the inlet of your blenders, respirators, etc. to protect them from moisture in case the drying system of your compressed air source (portable compressor or piped) fails. The water trap should be checked and emptied during each patient visit. If your equipment did not come equipped with a trap, contact your local dealer for information on AHP's F-100 Condensation Trap. e Saline solutions can cause the valve plates inside the air pump to corrode. This lowers the compressors operating pressure and performance. Whenever applicable, isolate the air compressor from environments in which saline solutions are used. e Do not substitute parts without permission from the Allied Healthcare Products, Inc. Service Manager. Such substitutions can cause premature product failure and may void warranty. FIG. 1 BRACKETS IMPORTANT UNPACKING INSTRUCTIONS 1. Unthread the six Phillips screws on the side panel and remove the panel. 2. Loosen and remove three nuts and bolts attached to pump bracket, on the bottom of the pump. 3, Loosen, butdo not remove the three 5/1 6-18 hex nuts securing the tie down straps to the angle on the side of the cabinet. 4. Swing the tie down straps (either up or down) toward the cabinet. 5. Reattach the three nuts and bolts removed in instruction #1 to the tie down straps. NOTE: Reverse these instructions when preparing the unit for shipping. 6. Tighten all hex nuts. 7. Check to see that the drainage hose is positioned in: the foam in the water tray. 8. Examine the compressor for shipping damage and replace the side panel. CONTROLS AND FEATURES Power Switch: turns the compressor on and off. Elapsed Time Indicator: measures how manv hours the. Aridyne 3500 has been used. Performance Gauge: indicates how well the compressor is operating. Green: Safety — indicates 50 psig outlet pressure and 5*F or greater dew point depression. Unit may run continuously. Hook-up may be remote. Yellow: Waïning — indicates 45-50 psig and low dew point depression. Hook-up should be very close. The needle should only enter the yellow area occasionally and should not stay in the yellow for long or the drainage system will be impaired. Red: Danger — indicates limited drying, less than 50 psig outlet pressure and flow capacity exceeded. Unit should NEVER be operated in the red. FIG. 2 CONTROL PANEL ARIDYNE® 3500 Е [ote ord on | КО | my HO frito E CA & J I hide! / POWER SWITCH ELAPSED TIME INDICATOR PERFORMANCE OPERATION 1. Position the compressor so there is plenty of air space in front and behind it. Good air circulation can prevent overheating. 2 Insert the plug into a proper hospital grade outlet. DO NOT use ungrounded outlets or adaptor plugs. 3. Turn the power switch “ON.” (Check valves in the outlets allow you to attach equipment when the compressor is on.) DO NOT use wrenches or excessive force in attaching equipment to the outlets. This could damage internal tubing. 4. DO NOT operate the compressor for extended periods of time if the performance gauge is in the yellow. NEVER operate the unit if the gauge is in the red. 5. The compressor can be started with pressure in the system. ROUTINE MAINTENANCE A. DAILY 1. Locate the cabinet inlet filter (Fig. 3, Нет 1). 2. Clean the dust and lint from this filter daily. At least once a week remove it and wash it in warm soapy water. FIG. 3 .— % B. QUARTERLY: Clean the pump inlet filter elements, 1. Turn the power switch off and unplug the power cord. 2. Remove the rear panel of the air compressor. 3. Disconnect the high pressure hose (Fig.7, item 7) at the swivel fitting and pull it up through the shelf. 4. Disconnect the tubing from the bottom of the water trap jar. Loosen the clamp and unthread the jar. | 5. Loosen the clamp on the drainage tubing (Fig.7, item 21) and pull it up through the shelf. 6. Carefully slide the motor/fan shelf out enough to expose both pump inlet filter assemblies (Fig.4). Do this slowly to avoid putting excess stress on wires. 7. Clean or replace the pump inlet filter elements (Fig.4). a. Unthread the jars (1) from their lids. b. Pull out the center studs (2). c. Wash the elements (3) with clean water and air dry, or install new filters. d. Reassemble. FIG. 4 C. YEARLY: Install the Preventive Maintenance Kit. (See REPLACEMENT PARTS for part numbers.) It includes the following: 2 Pump Inlet Filter Elements 1 Water Trap Filter Element 1 Water Jar O-Ring 6 Cabinet Inlet Filters 1 Relief Valve Diaphragm 1 Regulator Diaphragm 1. To install the pump inlet filter elements, refer to QUARTERLY. 2. To replace the water jar filter and O-ring (Fig. 5): a. Remove the rear panel. b. Disconnect the tubing from the bottom of the water jar (1). Loosen the clamp (Fig. 7, item A) and unthread the jar. Unthread the retainer cap (2) and remove the filter (3). . Clean the water jar with warm soapy water. . Install the new water jar filter and reassemble. When reassembling, carefully place the new O-ring (4) on the top of the water jar and rethread the jar. о > Da FIG. 5 Routine Maintenance Continued DD O O 3. To install the relief valve and regulator diaphragms (Fig. 6): a. Pull out the relief valve locking ring (6) and unthread the adjustment knob (7) as far as it will go. b. Remove the relief valve bonnet (5) and replace the diaphragm (2). c. Pull out the regulator locking ring (12) and unthread the adjustment knob (13) as far as it will go. d. Remove the regulator bonnet (11) and replace the regulator diaphragm (9). e. Reassemble the relief valve and regulator. RESYNCHRONIZE THE REGULATOR AND RELIEF VALVE. DO NOT ATTEMPT TO OPERATE THE ARIDYNE 3500 UNTIL IT HAS BEEN RESYNCHRONIZED. See Service. buno, + AFTER 6,000 HOURS/ PUMP OVERHAUL KIT After 6,00 hours of use, install the pump overhaul kit, even if the Aridyne 3500 is operating properly. The overhaul kit prolongs pump life without major rebuilding. See the replacement parts list for the part number. Service life of Aridyne 3500 will be shortened if it is operated in a hot environment (90°F/32°C or above). If this is the case, install the kit more often than every 6,000 hours. Instructions are included with each kit. Before using the kit, follow these instructions to remove the shelf/pump assembly: 1. Unplug the power cord. 2. Remove the rear panel. 3. Disconnect the copper tubing at the compression fitting (Fig. 7, item B) above the shelf. 4. Disconnect the tubing from the bottom of the water jar (Fig. 7, item 3), loosen the clamp (Fig. 7, item A) and unthread the jar. 5. Disconnect the tubing from the relief valve (Fig. 7, item 2) and pull it down through the shelf (Fig. 7, ltem D). | 6. Slide the shelf entirely out of the cabinet. CAUTION: Be extremely careful not to bend the fan blades. This can severely affect the cooling and performance of the Aridyne 3500. 7. Remove the pump wires from the termina! block (Fig. 7, Item C). 8. Lay the shelf down on its side and spread the springs (Fig. 7, Item 10) open at the four pump eyebolts. Unhook the springs from the pump to free the pump from the shelf. 9. Proceed with the pump overhaul, using instructions included with the kit. NOTES: © When installing rings in the air pump, be sure the piston is exactly flush with the cylinder when the piston is at top dead center. Readjust the cylinder as necessary. e |f you have questions, please contact Allied Healthcare Products, Inc. e |fthere is excessive damage to the pump, such as broken pistons, scored cylinders, etc., contact Allied Healthcare Products, Inc. to make proper arrangements for factory repair. SERVICE RESYNCHRONIZATION 1. With the rear panel off, plug the power cord in and turn the power switch on. 2. Pull the regulator locking ring (Fig. 6, item 12) and turn the regulator adjustment knob (Fig. 6, item 13) clockwise until it is completely closed. 3. If the performance gauge does not read at the “P” of PSI in the GREEN area at a no-flow condition, pull out the relief valve locking ring (Fig. 6, Item 13) and turn the adjustment knob until it does. CAUTION: DO NOT adjust the needie past the - “P” in the green area or pump damage will result!!! 4. Once the relief valve knob is adjusted properly, push the locking ring in to lock the setting. 5. Attach a pressure gauge to either of the Aridyne 3500 outlets. | 6. Adjust the regulator until the outlet pressure is approximately 55 psig at a no-flow condition. After each adjustment, allow a few seconds for the system to stabilize. 7. Push the regulator locking ring in to lock the setting. ITEM PART # O0 SO O1 AS 0 D-— 081-102 903-401 13532 082-110 029-119 990-434 918-401 11256 12154 12300 611-307 12152 327-204 903-402 973-401 FIG. 7 INTERNAL VIEW OF THE ARIDYNE 3500 PART DESCRIPTION Handle w/hardware 15 Relief valve 16 Water jar w/exit bushing & nut 17 Fan assembly w/hardware 18 Pump w/suspension 19 Pop-off valve 20 High pressure hose w/fittings 21 Water tray w/foam block | Caster back Caster, locking front Suspension spring Complete pump inlet filter Circuit breaker Regulator Needle valve 080-109 Z75 974-401 911-007 911-004 974-402 911-003 Also illustrated A Water jar clamp C Temina! block D Shelf Bulkhead w/nut & lock washer Outlet, check DISS air x 1/8” NPT Sealing sleeve 3/8" O.D. Nylon tubing 3/8” О.Р. Plastic tubing 1/4” O.D. Sealing sleeve 1/4" O.D. Clear drain tube REPLACEMENT PARTS PART # (Fig. 2) 8134 * 8508 510-504 (Fig. 3) 901-409 025-115 (Fig. 4) 901-405 (Fig. 5) 924-401 13530 (Fig. 6) 991-407 923-401 923-402 PART DESCRIPTION Elapsed time indicator Performance gauge Power switch Cabinet inlet filter Splash shield Pump inlet filter element Water jar filter Water jar O-ring Regulator bonnet (Item 1 1) VOG6 Relief valve diaphragm (Item 2) RO6 Regulator diaphragm (item 9) PART # PART DESCRIPTION (Not Illustrated) 2103 12002 613-006 648-301 Preventive maintenance kit Includes: 901-405 Pump inlet filter elements (2) 924-401 Water jar filter 13530 Water jar O-ring 901-409 Cabinet inlet filters (6) 923-401 Relief valve diaphragm 923-402 Regulator diaphragm Pump overhaul kit Shipping carton Air deflector chute COMPRESSOR PUMP PARTS 990-408 990-410 990-412 990-470 990-415 990-467 990-416 990-418 990-421 990-423 990-430 990-462 + Call Allied Healthcare Service Department to see if adaptor plate # indicator. AD35-AF506 Piston only AD35-AF508 Cylinder head AD35-AF510 Cylinder AD35-515E Eccentric AD35-AF523 Wrist pin AD35-AF524 Flat key AD35-AF529 Valve plate AD35-AF533 Fan blade AD35-AF535 Fan shroud AD35-AF536B Manifold AD35-AF560 Piston & rod assy SCE69FK Wrist pin bearings 11181 is needed to install the elapsed time Trouble PERFORMANCE GAUGE: needle sticks, is slow to respond, or vibrates rapidly. TROUBLESHOOTING Cause Needle valve maladjusted Solution Turn needle valve clockwise tO stop vibration: Counterclockwise to prevent sticking and slow response. Proper needle valve setting is necessary to prevent performance gauge damage and ensure accuracy. OVERHEATING - HIGH PRESSURE: LOW PRESSURE: Pump not running: LOW PRESSURE: Pump still running: WATER PROBLEMS: Water coming from outlet(s): Cabinet filter blocked by curtains, bedding, a wall, etc. Cabinet filter clogged Air Compressor running in hot, unventilated area Fan not operating Regulator set too high - Regulator and relief valve diaphragms need to be cleaned/replaced Power cord unplugged Circuit breaker popped Hospital fuse has blown Thermal overload protector in pump has activated. Outiet(s) leaking Water trap improperly sealed Water jar O-ring ill-fitted or missing internal brass fitting leaking Pop-off valve defective Regulator set too low Contract with saline solution has corroded valve plated Pump worn Flow capacity of pump being exceeded Performance gauge needle in red Pump overheated Regulator out of synchronization Pump in shipping position Damaged pump spring(s) Pump worn Fan brackets bent; blade striking shelf or brackets Move obstruction. Clean or replace. Reposition Check electrical connections/ Replace fan. Resynchronize (See SERVICE). See "YEARLY". Plug it in. To reset, press button until red ring no longer shows. If pops again, call dealer or the Allied Healthcare Service Dept. at (314) 771-2400, or toll-free (800) 444-3954. Check and reset. Automatic reset within 20-30 minutes. Leave power switch on so the fan can cool the pump faster. Check the following: | a. Any of the causes listed in OVERHEATING above. b. A motor defect is causing it to draw too much power. This will quickly overheat the motor. Current usage should not exceed 8.5 amps during normal Compressor usage. Tighten or reseal. See ROUTINE MAINTENANCE YEARLY, step 2, Fig. 5. See ROUTINE MAINTENANCE YEARLY, step 2, Fig. 5. Tighten or reseal. Call Allied Healthcare Service. Resynchronize, See "SERVICE". Look for and remove corrosion. In advanced cases install pump service kit. Install pump service kit. Reduce flow requirements. Reduce flow requirements. See OVERHEATING above. See SERVICE. See IMPORTANT UNPACKING INSTRUCTIONS. Replace spring(s). Install pump service Kit. Straighten. WIRING SCHEMATIC CIRCUIT BREAKER BLK COMPRESSOR HOTOR BLK LINE ORD | TERMINAL BLOCK, AIR AIR - OUTLET OUTLET TO CONTROL ALL WIRES ARE 16 GA. UNLESS PANEL GRD. OTHERWISE NOTED. Allied Healthcare Products, Inc. — TIMETER INSTRUMENT CORPORATION 1720 Sublette Avenue St. Louis, MO 63110 Telephone: (314) 771-2400 Toll Free: (800) 444-3954 INSTRUMENTS OF CARE E 1999 Allied Healthcare Products Inc. Printed in U.S.A. ">
Advertisement
Key features
- Special quieting material
- Drying system
- Casters
- 10 Amp Internal Circuit Breaker
- Two Grounded DISS Male Air Outlets
Frequently asked questions
The output is 42 lpm at 50 ± 5 psig at 5°F dew point depression.
It uses 120 Volt, 60 Hz 1 Phase, 8.5 Amp power.
The dimensions are 11 1/2" (29 cm) width, 15 3/4" (40 cm) depth, and 33 1/2" (87 cm) height.
It has two grounded DISS male air outlets with internal check valves.
The warranty covers defective parts for one year from the date of purchase.
Do not use it with flammable anesthetics or oxygen, and do not lubricate it. Refer all servicing to qualified service persons.