TSI 9565 Series Air Velocity Meter Operation and Service Manual
Below you will find brief information for Air Velocity Meter 9565 Series. The 9565 Series Air Velocity Meter is a versatile tool for measuring air velocity, temperature, humidity, and pressure. The device can be used to measure air flow in ducts, to assess the effectiveness of ventilation systems, and to monitor indoor air quality. The device is also capable of measuring carbon monoxide and volatile organic compounds, and it can be used to perform draft rate, heat flow, turbulence, percent outside air, and leakage tests thanks to the different probes.
Advertisement
Advertisement
VELOCICALC
®
AIR VELOCITY METER
MODEL 9565 SERIES
OPERATION AND SERVICE MANUAL
P/N 6004851, REVISION D
OCTOBER 2014
Copyright
TSI Incorporated / 2011-2014 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
Limitation Of Warranty And Liability (effective April 2014)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com
.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions: a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in specifications, are warranted for 90 days from the date of shipment; b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals; c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material, under normal use, for 90 days from the date of shipment; d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies; e. This warranty does not cover calibration requirements, and seller warrants only that the instrument or product is properly calibrated at the time of its manufacture.
Instruments returned for calibration are not covered by this warranty; f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning; g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND
EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE
PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS
BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR i
OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE
REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE
REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE,
SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO
DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO
EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS
INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR
INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR
INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No
Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY
AND LIABILITY, which contains the complete and exclusive limited warranty of Seller.
This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call Customer Service department at (800) 874-2811 (USA) or (1)
651-490-2811 (International). ii
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
CHAPTER 2 SETTING-UP ................................................................. 3
Supplying Power to the Model 9565 Series .................................... 3
Installing the Batteries ............................................................... 3
DIP Switch Settings ................................................................... 3
Using the AC Adapter ................................................................ 4
Connecting Ventilation or IAQ Probes ............................................ 5
Using the Telescoping Thermoanemometer Probes ...................... 5
Extending the Probe .................................................................. 5
Retracting the Probe .................................................................. 5
Using the Pressure Ports (9565-P) ................................................. 6
Connecting the Static Pressure Probe ...................................... 6
Connecting an Optional Pitot Probe or Airflow (straight pitot) Probe ............................................................................ 7
Thermocouple Ports ....................................................................... 8
Connecting the Thermocouples ................................................ 8
Connecting the Optional Bluetooth Portable Printer Device ........... 9
Connecting to a Computer .............................................................. 9
CHAPTER 3 OPERATION ............................................................... 11
Keypad Functions ......................................................................... 11
Common Terms ............................................................................ 12
Menus ........................................................................................... 13
ZERO PRESSURE ............................................................................. 13
DISPLAY SETUP ............................................................................... 14
SETTINGS ....................................................................................... 15
FLOW SET UP .................................................................................. 15
VOC SETUP ..................................................................................... 17
ACTUAL/STANDARD SETUP .............................................................. 18
DATA LOGGING ............................................................................... 19
Measurements ......................................................................... 19
Log Mode/Log Settings ............................................................ 19
Choose Test ............................................................................ 25
Name Test ............................................................................... 25
View Data ................................................................................ 26
View Stats ................................................................................ 26
View Samples .......................................................................... 27
Print Test ................................................................................. 28
Delete Data .............................................................................. 29
% Memory................................................................................ 30
ZERO CO ........................................................................................ 31
APPLICATIONS ................................................................................ 32 iii
CALIBRATION .................................................................................. 33
BLUETOOTH FUNCTIONS .................................................................. 33
Discover Devices ..................................................................... 33
Discoverability .......................................................................... 34
PINcode ................................................................................... 34
# AutoConnects ....................................................................... 34
Printing Data Using the Portable Printer ....................................... 34
T
RAK
P
RO
™ Data Analysis Software ............................................. 35
LogDat2™ Downloading Software ................................................ 35
CHAPTER 4 MAINTENANCE .......................................................... 37
Recalibration ................................................................................. 37
Cases ............................................................................................ 37
Storage .......................................................................................... 37
CHAPTER 5 TROUBLESHOOTING ................................................ 39
APPENDIX A SPECIFICATIONS ..................................................... 41
APPENDIX B OPTIONAL PLUG-IN PROBES ................................ 45
These Application Notes can be found under TSI’s web site:
http://www.tsi.com
TSI-106 Traversing a Duct to Determine Average Air Velocity or
Volume
TSI-107 Using the V
ELOCI
C
ALC
®
Meter to Measure Pressure
TSI-109 Converting Standard Velocity to Actual Velocity
TSI-114 Use of K-Factors with the V
ELOCI
C
ALC
®
Meter
TSI-124 Heat Flow Calculations
TSI-138 Percent Outdoor Air Calculation and Its Use
TSI-141 Turbulence Intensity Measurements
TSI-142 Draft Rate: A Determining Factor in the Quantification of
Human Comfort
TSI-147 Photo-Ionization Detection (PID) Technology
TSI-150 Using Bluetooth
®
Communications iv
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. Static pressure tip (9565-P model only)
4. Rubber tubing (9565-P model only)
5. USB cable
6. T
RAK
P
RO
CD-ROM with data analysis software
7. LogDat2 CD-ROM with downloading software
8. AC adapter
1
(This page intentionally left blank)
2 Chapter 1
Chapter 2
Setting-up
Supplying Power to the Model 9565 Series
The Model 9565 V
ELOCI
C
ALC
®
Air Velocity Meter can be powered in one of two ways: four size AA batteries or the AC adapter.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Model 9565 is designed to operate with either alkaline or NiMH rechargeable batteries. Battery life will be shorter if NiMH batteries are used. If
NiMH batteries are used the DIP switch will need to be changed.
Refer to the next section, DIP Switch Settings . Carbon-zinc
batteries are not recommended because of the danger of battery acid leakage.
DIP Switch Settings
To access the DIP switch, remove the batteries from the battery compartment. On the inside of the battery compartment, there is a window with a single DIP switch (see figure below). The table below shows the functions for the switch.
Caution: Make certain that power is turned off before changing the
DIP switch settings.
Switch Function Settings
1 NiMH OFF: Alkaline Batteries
ON: NiMH Rechargeable Batteries
V
ELOCI
C
ALC
is a registered trademark of TSI Incorporated.
3
Do not attempt to charge alkaline batteries.
Using the AC Adapter
The AC adapter can be used to power the instrument or to charge the NiMH batteries when the DIP switch in the battery compartment is set to NiMH. If the DIP switch is set to
Alkaline, and the AC power adapter is connected, then the batteries will be bypassed and the meter will be powered by the AC adapter. Be sure to provide the correct voltage and frequency, which is marked on the back of the AC adapter.
AC Adapter input
4 Chapter 2
Connecting Ventilation or IAQ Probes
The ventilation and IAQ probes have a
“D” shape overmolding on the mini-DIN connector which must align with the connector at the base of the 9565 series meter. This will ensure the probe is properly connected and remains so during use. Once connected and turned on, refer to the DISPLAY SETUP for displaying the desired measurements.
“D” Shaped mini-DIN connector
Using the Telescoping Thermoanemometer Probes
The telescoping probe contains the velocity, temperature, and humidity sensors. When using the probe, make sure the sensor window is fully exposed and the orientation dimple is facing upstream.
N O T E
For temperature and humidity measurements, make sure that at least 3 inches (7.5 cm) of the probe is in the flow to allow the temperature and humidity sensors to be in the air stream.
Extending the Probe
To extend the probe, hold the handle in one hand while pulling on the probe tip with the other hand. Do not hold the cable while extending the probe as this prevents the robe from extending.
Retracting the Probe
To retract the probe, hold the handle in one hand while gently pulling the probe cable until the smallest antenna section is retracted.
Setting-up 5
Do not use the instrument or probes near hazardous voltage sources since serious injury could result.
Using the Pressure Ports (9565-P)
The 9565-P includes pressure ports that can be used to measure static and differential pressures in ductwork. For more information, see Application Note TSI-107 .
Negative (-) Pressure Port Positive (+) Pressure Port
Connecting the Static Pressure Probe
The Static Pressure probe included with the 9565-P is connected to the + port on the 9565-P using the included tubing. The Static
Pressure probe is used to measure the duct static pressure and features a magnet which holds the probe to the ductwork.
6 Chapter 2
Connecting an Optional Pitot Probe or Airflow (straight pitot) Probe
When connected to a pitot probe, air velocity or air volume can be measured. A pitot probe can be connected to the
“+” and “-” pressure ports on the Model 9565-P using two pieces of tubing of equal length. The total pressure port of the pitot probe connects to the “+” port on the meter, and the static pressure port of the pitot probe connects to the “-” port on the meter.
For information on how to perform a duct traverse, refer to
Application Note TSI-106 .
N o t e
The pitot velocity needs a valid temperature to perform the standard or actual velocity correction. This is accomplished in the “Actual/Std Setup” menu. If no probe capable of measuring temperature (plug in probe or thermocouple) is connected, the“Temp Source” must be set to “Entered”. The duct air temperature must then be manually inputted by the user using the “Entered Temp” setting. If the “Temp Source” is set to Probe or Thermocouple 1 or 2, and no probe is connected, dashes (------) will appear on the display.
For more information on entering the temperature manually,
refer to the Actual/Standard Setup section of this manual.
Setting-up 7
Do not use the instrument or probes near hazardous voltage sources since serious injury could result.
Thermocouple Ports
The 9565 series includes two thermocouple ports at the base of the meter. Any K-Alloy thermocouple with mini-connector can be
attached. See Display Setup for setting the thermocouple
temperature readings to be displayed as TC1, TC2, or TC1-TC2.
Connecting the Thermocouples
The K-Alloy thermocouple with mini-connector has one terminal wider than the other. The wider terminal will be inserted into the bottom of the TC1 or TC2 connector port.
K-alloy Thermocouple
TC1 TC2
Thermocouples from an alternate TSI supplier must have the metal sheath electrically isolated from the wires inside. Failure to meet this requirement may result in false readings, electrical shock, or fire hazard.
Do not use the instrument or probes near hazardous voltage sources since serious injury could result.
8 Chapter 2
Connecting the Optional Bluetooth
®
Portable Printer Device
To connect the Bluetooth printer to the Model 9565, power on the unit and the printer. Then press the
MENU soft key. From the Menu use the
and
keys to highlight
Bluetooth Functions and press the key. Highlight
Discover Devices and press the
key. If other TSI
Bluetooth-printers are in the area, turn them off before searching. The
Model 9565 will then search for and list all available Bluetooth devices.
For more information on establishing Bluetooth connections, refer to
TSI Applications Note TSI-150 .
Connecting to a Computer
Use the Computer Interface USB Cable provided with the Model 9565 to connect the instrument to a computer for downloading stored data.
USB Communications Port
C a u t i o n
This symbol is used to indicate that the data port of the
Model 9565 is not intended for connection to a public telecommunications network. Connect the USB data port only to another USB port on a safety certified computing device.
Protection provided by the instrument could be impaired if used in a manner other than specified in this user manual.
Bluetooth is a registered trademark of Bluetooth SIG.
Setting-up 9
(This page intentionally left blank)
10 Chapter 2
Chapter 3
Operation
Keypad Functions
ON/OFF ( ) Key
Arrow () Keys
Enter ( ) Key
Press the
ON/OFF key to turn the Model 9565 on and off. During the power up sequence the display will show the following: Model Number, Serial
Number, and Software Revision. To turn the instrument off, press and hold the
ON/OFF Key for 3 seconds. The instrument will count down (OFF2,
OFF1, OFF). If AC Adapter is attached, the Battery and
ON/OFF Key is bypassed. If the ON/OFF Key is pressed while the AC adapter is attached, the instrument instructs you to “Unplug the instrument t o turn off unit”. To turn the instrument on again, attach the AC adapter or press the
ON/OFF Key.
Press to scroll through choices while setting a parameter. Pressing the keys simultaneously will lock the keypad to prevent unauthorized adjustments to the instruments. To unlock the keypad, press the keys simultaneously.
Press to accept a menu selection, value or condition. Press to Start or Stop datalogging when in Continuous Key mode.
11
Arrow (or ) and
Menu Soft Keys
Press arrow keys to change choices while setting a parameter. Press the Menu soft key to select the
Menu selections, which are Pressure Zero, Display
Setup, Settings, Flow Setup, VOC Setup,
Actual/Std Setup, Data Logging, Zero CO,
Applications, Calibration, and Bluetooth Functions.
Common Terms
In this manual there are several terms that are used in different places. The following is a brief explanation of the meanings of those terms.
Sample
Consists of all of the measurement parameters stored at the same time.
Test ID
Time Constant
Log Interval
Test Length
A group of samples. The statistics (average, minimum, maximum, and count) are calculated for each test ID. The maximum number of test IDs is 100.
The time constant is an averaging period. It is used to dampen the display. If you are experiencing fluctuating flows, a longer time constant will slow down those fluctuations. The display will update every second, but the displayed reading will be the average over the last time constant period. For example, if the time constant is 10 seconds, the display will update every second, but the displayed reading will be the average from the last 10 seconds. This is also referred to as a “moving average”.
The logging interval is the period over which the instrument will average the logged sample. For example, if the logging interval is set to 30 minutes, each sample will be the average over the previous
30 minutes.
This is the time over which the data will be logged in the “Continuous-Time” mode of data logging.
12 Chapter 3
Menus
The menu structure is organized to allow easy navigation and instrument setup utilizing the arrow keys and button. To exit a menu or menu item, press the
ESC key.
To access the Menu items, press the
Menu soft key.
To select a parameter, use the Arrow keys to highlight the selection and press the button.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
ZERO PRESSURE
To zero the pressure reading, leave the positive and negative ports open to atmosphere and select the
Zero Press. The pressure sensor should be zeroed after initial setup and checked periodically. The instrument will indicate if the pressure zero was successful.
Operation 13
DISPLAY SETUP
Display Setup menu is where you will setup the desired parameters to be displayed on the instrument screen. With a parameter highlighted you can then use the
ON soft key to have it show up on the instrument screen or select the
OFF soft key to turn off the parameter. Use
PRIMARY soft key to have a parameter show up on the instrument screen in a larger display. A total of 5 parameters can be shown on the display, 1 primary (large font) and 4 secondary. Parameters shown in the Display Setup screen are dependent on the type of probe currently connected.
When set to
PRIMARY, measurement will be the large font on the display.
When set to
ON, measurement will be displayed as a secondary parameter (up to 4 can be displayed).
When set to
OFF, measurement will not be displayed.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
DISPLAY SETUP
Velocity
Flow
*ON
OFF
Pitot Velocity OFF
Pressure ON
Temperature ON
Baro Press OFF
ON PRIMARY OFF
N o t e
Pitot Velocity and AFProbe Velocity cannot both be ON at the same time, nor can one be set to PRIMARY and the other to ON. Only one can be ON at any given time.
14 Chapter 3
SETTINGS
Settings menu is where you can set the general settings. These include Language, Beeper, Select Units, Time Constant, Contrast,
Set Time, Set Date, Time Format, Date Format, Number Format,
Backlight and Auto Off. Use the or
keys to select an option, and the or soft keys to change the settings for each option. Press the
key to accept settings.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
SETTINGS
Language English
Beeper Disable
Select Units
Time Constant 1 Second
Contrast 5
Set Time 09:14 AM
Set Date 10/31/08
Time Format 12 hr
Date Format
MM/DD/YY
Number Format XX,XXX.YY
Backlight Auto
Auto Off Enable
-------
FLOW SET UP
In
Flow Setup mode, there are six types: Round Duct, Rectangle Duct,
Duct Area, Pressure/Kfactor, Horn and Air Cone. Use the or soft keys to scroll through the types and then press the key to accept the desired type. To change the value, highlight the
Enter Settings option and press the key.
FLOW TYPE MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
FLOW SETUP
Flow Type
Round Duct
Enter Settings
Round duct
Rect duct
Duct area
Press/Kfact
Horn
Air cone
ENTER SETTINGS
12.0 in dia
Round duct, Rect duct and Duct area are used to perform a duct traverse using a pitot probe or thermoanemometer probe. For more
Operation 15
information on how to perform a duct traverse, refer to Application
Note TSI-106 .
Press/Kfact allows for calculating flow rate from diffusers or flow stations with pressure taps using the instruments pressure ports and
Kfactors. The Kfactors are obtained from the diffuser or flow station manufacturer. For more information, refer to Application Note TSI-
114 .
Up to 5 rectangular ducts, 5 round ducts, 5 duct areas and 5 Kfactors can be pre-programmed for quick use on a jobsite:
When Flow is set as the Primary measurement in the Display Setup menu, the dimensions or Kfactor will also be displayed:
N o t e
For flow measurements with round, rectangular, or duct area, velocity must be set as a secondary measurement in order for flow readings to appear when using the pitot or AF probe.
16 Chapter 3
When measuring Flow as the Primary measurement, the parameters can be quickly changed by pressing the or key while on the main measurement screen:
Make adjustments with the or arrow keys and press to accept, or enter the Select Duct or Select Kfactor menu to choose a different pre-programmed flow value or dimension.
The Horn numbers are the models of the horns. For example, 100 refers to a horn model number AM 100. Only horns with Model numbers as follows can be used with this function: AM 100, AM 300,
AM 600 and AM 1200. If a horn model number is chosen, the instrument will return to measuring mode and use a preprogrammed curve to calculate flow rate from velocity when using a thermoanemometer probe.
The Air cone selection applies to the Model 995 Rotating Vane
Anemometer and the Air Cone kit p/n 801749 which are used to take flow measurements on small outlets.
VOC SETUP
This menu item applies to TSI probes that measure Volatile Organic
Compounds (VOC). Refer to the manual included with VOC probe
Models 984, 985, 986, and 987 for information on usage and setup.
Operation 17
ACTUAL/STANDARD SETUP
Choose
Actual/Standard measurements and parameters in the Act/Std
Setup menu. The Model 9565 measures the actual barometric pressure using an internal sensor. The temperature source can be entered manually or taken from a probe that measures temperature
(plug in probe or thermocouple).
For more information on Actual and Standard conditions, refer to
Application Note TSI-109 .
SETTING
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
ACTUAL/STD SETUP
Setting Standard
Temp source Entered
Standard
Actual
TEMP SOURCE
Entered Temp
Probe
Thermocouple 1
Thermocouple 2
70
F
18 Chapter 3
DATA LOGGING
Measurements
Measurements to be logged to memory are independent of measurements on the display, and must therefore be selected under
DATA LOGGING Measurements.
When set to
ON, measurement will be logged to memory.
When set to
DISPLAY, measurement will be logged to memory if it is visible on the main running screen.
When set to
OFF, measurement will not be logged to memory.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
DATA LOGGING
Measurements
Log Mode Cont.-time
Log Settings
Temperature
DISPLAY
DISPLAY
Choose Test Test 001
Name Test
Flow
View Data
Heatflow
Delete Data
DISPLAY
OFF
Pressure
DISPLAY
% Memory
Baro Press
ON
ON DISPLAY OFF
N o t e
Pitot Velocity and AFProbe Velocity cannot both be ON at the same time, nor can one be set to PRIMARY and the other to ON. Only one can be ON at any given time.
Log Mode/Log Settings
The 9565 can be programmed to store measurements to memory in several different logging formats:
Manual
Auto-save
Cont-key
Cont-time
Program 1
Program 2
Operation 19
Manual Logging
Manual mode does not automatically save data, but instead prompts the user to
SAVE a sample or ESC to not save. To start logging, press the key.
N o t e
To adjust the averaging period for a sample, change the
Time Constant (increase or decrease in seconds) which is located in the Settings Menu.
DATA LOGGING
Measurements
Log Mode Manual
Log Settings
Choose Test Test 001
Name Test
View Data
Delete Data
% Memory
LOG MODE
Manual
Auto-save
Cont.-key
Cont.-time
Program 1
Program 2
Auto Save Logging
In
Auto-save mode, the user samples are automatically logged to memory at the end of the sampling period. To start logging, press the key.
DATA LOGGING
Measurements
Log Mode Auto-save
Log Settings
LOG MODE
Manual
Auto-save
Choose Test Test 001
Name Test
View Data
Cont.-key
Cont.-time
Delete Data
% Memory
Program 1
Program 2
20 Chapter 3
When set to
Auto-save, the Sample Time can be adjusted. Sample
Time is the time period over which the Sample will be averaged.
DATA LOGGING
Measurements
Log Mode Auto-save
Log Settings
Choose Test Test 001
Name Test
SAMPLE TIME
00:05
Min:Sec
View Data
Delete Data
% Memory
Cont-key Logging
In
Cont-key mode, the user starts logging by pressing the
key.
The instrument will continue logging until the key is pressed again.
DATA LOGGING
Measurements
Log Mode Cont.-key
LOG MODE
Manual
Log Settings
Choose Test Test 001
Name Test
Auto-save
Cont.-key
Cont.-time
View Data
Program 1
Delete Data
% Memory
Program 2
Operation 21
When set to
Cont. key, the log interval can be adjusted.
DATA LOGGING
Measurements
LOG SETTINGS
Log Mode Cont.-key
Log Settings
Log Interval 00:01
Choose Test Test 001
Name Test
LOG INTERVAL
00:05
Min:Sec
View Data
Delete Data
% Memory
N o t e
Pressing the keys simultaneously will lock the keypad to prevent unauthorized adjustments to the instruments during unattended logging. A “Lock” symbol will appear on the display. To unlock the keypad, press the keys simultaneously. The “Lock” symbol will disappear.
Cont-time Logging
In
Cont-time mode, the user starts taking readings by pressing the
key. The instrument will continue taking samples until the time as set in “Test Length” has elapsed.
DATA LOGGING
Measurements
Log Mode Cont.-time
LOG MODE
Manual
Log Settings
Choose Test Test 001
Auto-save
Name Test
View Data
Cont.-key
Cont.-time
Program 1
Delete Data
% Memory
Program 2
22 Chapter 3
When set to
Cont.-time, the log interval and test length can be adjusted.
DATA LOGGING
Measurements
LOG SETTINGS
Log Mode Cont.-time
Log Settings
Log Interval 00:01
LOG INTERVAL
00:05
Min:Sec
Name Test
View Data
Delete Data
% Memory
TEST LENGTH
00 : 00 : 05
Day:Hour:Min
N o t e
Pressing the keys simultaneously will lock the keypad to prevent unauthorized adjustments to the instruments during unattended logging. A “Lock” symbol will appear on the display. To unlock the keypad, press the keys simultane ously. The “Lock” symbol will disappear.
Operation 23
Program 1 and Program 2
Program 1 and Program 2 are customized data logging setup programs. Setting them up is performed using TSI’s T
RAK
P
RO
™
Data Analysis software.
DATA LOGGING
Measurements
Log Mode Cont.-time
Log Settings
Choose Test Test 001
Name Test
LOG MODE
Manual
Auto-save
Cont.-key
Cont.-time
View Data
Delete Data
% Memory
Program 1
Program 2
For more information, refer to the T
RAK
P
RO
Data Analysis
Software User’s Guide which can be found on the T
RAK
P
RO software CD which is included with the 9565.
T
RAK
P
RO
is a trademark of TSI Incorporated.
24 Chapter 3
Choose Test
Test IDs consist of a group of Samples that are used to determine statistics (average, minimum, and maximum) of a measurement application. The 9565 can store 26,500+ samples and 100 test
IDs (one sample can contain fourteen measurement types).
Example: Each duct traverse will have its own Test ID consisting of several Samples.
Pressing
NEW
will advance to the next available Test ID. Pressing
DATES will list the date the Test was taken.
DATA LOGGING
Measurements
Delete Data
CHOOSE TEST
Log Mode Cont.-time
Test 001 9 Samples
Test 002 7 Samples
Log Settings
Test 003 0 Samples
Choose Test Test 001
Name Test
Test 005 0 Samples
View Data
Test 006 0 Samples
Test 007 0 Samples
% Memory
Test 008 0 Samples
Test 009 0 Samples
Test 010 0 Samples
--- ----
NEW DATES
Name Test
This option allows for customizing the Test ID name using 8 characters maximum. Use the arrow keys to move the cursor to a desired location, press to accept. Repeat until the desired name appears. Press
SAVE to store custom ID name.
DATA LOGGING
Measurements
Log Mode Cont.-time
Log Settings
Choose Test Test 001
Name Test
View Data
Delete Data
% Memory
Operation 25
View Data
Choose Test
To view stored data, first select the Test ID that contains the data to be recalled.
This is accomplished in the “Choose Test” menu.
DATA LOGGING
Measurements
VIEW DATA
Log Mode Auto-save
Test 001
Log Settings
View Stats
CHOOSE TEST
Test 001 9 Samples
Test 002 7 Samples
Choose Test Test 001
Name Test
Print Test
Test 003 0 Samples
Test 004 0 Samples
View Data
Delete Data
% Memory
Test 005 0 Samples
Test 006 0 Samples
Test 007 0 Samples
Test 008 0 Samples
Test 009 0 Samples
Test 010 0 Samples
--- ----
DATES
View Stats
Displays statistics (average, minimum, and maximum) of a selected Test ID and the number of samples, date and time the samples were taken.
DATA LOGGING
Measurements
VIEW DATA
Log Mode Auto-save
Choose Test
Test 001
Log Settings
View Stats
Choose Test Test 001
Name Test
Print Test
View Data
Delete Data
% Memory
26 Chapter 3
Use the
and
arrow keys to view statistics of all the measurement parameters stored in a Test ID.
TEST 001
Pressure
Avg 1.739 in. H2O
Min 1.665 in. H2O
Max 1.812 in. H2O
# Samples 9
10/31/08 07:01:39 AM
TEST 001
Temperature
Avg
Min
Max
78.2
F
78.1
F
78.3
F
# Samples 9
10/31/08 07:01:39 AM
TEST 001
%RH
Avg 12.2 %RH
Min 11.1 %RH
Max 12.9 %RH
# Samples 9
10/31/08 07:01:39 AM
PRINT PRINT PRINT
Example: TEST 001 has 9 samples, each sample consists of a pressure, temperature, and relative humidity reading. Use the < or > keys to view statistics of each measurement parameter.
The 9565 can send this data to the optional Model 8934 wireless printer or PC capable of Bluetooth communications. To use the
PRINT command, Bluetooth communications must be established between the 9565 and the Model 8934 wireless printer or PC set up with Bluetooth communications.
View Samples
VIEW DATA
Choose Test
Test 001
View Stats
View Samples
Print Test
Operation 27
Use the
and
arrow keys to view samples of all the measurement parameters stored in a Test ID.
TEST 001
Velocity
Sample 1 218 ft/min
Sample 2 280 ft/min
Sample 3 316 ft/min
Sample 4 399 ft/min
Sample 5 188 ft/min
Sample 6 306 ft/min
Sample 7 313 ft/min
Sample 8 294 ft/min
Sample 9 309 ft/min
TEST 001
Temperature
Sample 1 73.5
F
Sample 2 73.7
F
Sample 3 73.8
F
Sample 4 73.8
F
Sample 5 73.6
F
Sample 6 73.6
F
Sample 7 73.5
F
Sample 8 73.4
F
Sample 9 73.4
F
TEST 001
%rh
Sample 1 15.1%rh
Sample 2 14.2%rh
Sample 3 13.8%rh
Sample 4 13.8%rh
Sample 5 13.5%rh
Sample 6 13.6%rh
Sample 7 13.6%rh
Sample 8 13.5%rh
Sample 9 13.5%rh
PRINT PRINT PRINT
The 9565 can send this data to the optional Model 8934 wireless printer or PC capable of Bluetooth communications. To use the
PRINT command, Bluetooth communications must be established between the 9565 and the Model 8934 wireless printer or PC set up with Bluetooth communications.
Print Test
Press to print all statistics and samples for the selected Test
ID.
VIEW DATA
Choose Test
Test 001
View Stats
View Samples
Print Test
The 9565 can send this data to the optional Model 8934 wireless printer or PC capable of Bluetooth communications. To use the
PRINT command, Bluetooth communications must be established between the 9565 and the Model 8934 wireless printer or PC set up with Bluetooth communications.
For more information on establishing Bluetooth connections, refer to TSI Applications Note TSI-150 .
28 Chapter 3
Delete Data
Use this to delete all data, delete test or delete sample.
DATA LOGGING
Measurements
DELETE DATA
Log Mode Cont.-time
Delete All
Log Settings Delete Test
Choose Test Test 001
Delete Sample
Name Test
View Data
Delete Data
% Memory
Delete All will clear stored data in all Test IDs.
DELETE DATA
Delete All
Delete Test
Delete Sample
DELETE ALL
Are you sure?
YES NO
Delete Test will clear stored data in an individual Test ID selected by the user.
DELETE DATA
Delete All
Delete Test
Delete Sample
DELETE TEST
Test 001 14 Samples
Test 002 10 Samples
DELETE TEST
Test 003 12 Samples
Test 004 8 Samples
Test 005 7 Samples
Test 006 15 Samples
Test 007 0 Samples
Test 008 0 Samples
Test 009 0 Samples
Test 010 0 Samples
Delete Test 005.
Are you sure?
YES NO
--- ---DATES
Operation 29
Delete Sample will clear the last sample in an individual Test ID selected by the user.
DELETE DATA
Delete All
Delete Test
Delete Sample
DELETE SAMPLE
Test 001 14 Samples
Test 002 10 Samples
Test 003 12 Samples
Test 004 8 Samples
Test 005 7 Samples
Test 006 15 Samples
Test 007 0 Samples
Test 008 0 Samples
Test 009 0 Samples
Test 010 0 Samples
--- ---DATES
% Memory
This option displays the memory available.
Delete All, under Delete
Data, will clear memory and reset the memory available to 100%.
MEMORY
Test ID
Sample
83 %
92 %
30 Chapter 3
ZERO CO
This menu item applies to TSI probe Model 982 which can measure carbon monoxide (CO). Zero CO will zero the CO sensor readings that may have drifted. Initiating the Zero CO function will show the sensor CO reading and the time remaining.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
ZERO CO
0.0 ppm
N o t e
The Zero CO function should be performed in an area where no combustion is taking place which may affect zeroing of the sensor.
Operation 31
APPLICATIONS
This menu option includes specialized measurement protocols used to perform various tests or investigations. You can choose Draft Rate,
Heat flow, Turbulence, % Outside Air, and Leakage Test in the
Applications menu. For more information on these applications, refer to the following information:
Draft Rate: Application Note TSI-142
Heat Flow: Application Note TSI-124
Turbulence Intensity: Application Note TSI-141
Percent Outside Air: Application Note TSI-138
Leakage Test: PANDA Manual
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
APPLICATIONS
Draft Rate
Heatflow
Turbulence
% Outside Air
Leakage Test
32 Chapter 3
CALIBRATION
The Calibration Menu lists measurement parameters that can be adjusted in the field. The appropriate detachable probes must be attached to the 9565 before field calibration can be undertaken except for pressure and barometric pressure calibration.
For more information on performing field calibrations, refer to TSI
Applications Note TSI-146 .
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
CALIBRATION
Calibrate Temp
Calibrate Vel
Calibrate %RH
Calibrate B.P.
Calibrate CO2
Calibrate CO
Calibrate VOC
Restore Factory Cal
BLUETOOTH FUNCTIONS
The V
ELOCI
C
ALC
Model 9565 meter contains a Bluetooth Functions menu which is used to adjust parameters to assist with wireless connections to other Bluetooth capable devices.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
BLUETOOTH FUNCT
.
Discover Devices
Discoverability Enable
PINcode 0000
# AutoConnects 1
Discover Devices
Start the Bluetooth process of finding other devices from the
V
ELOCI
C
ALC
Model 9565.
Operation 33
Discoverability
Describes whether another device can discover the V
ELOCI
C
ALC
Model 9565 meter. Options include:
Disable
Temporary
Enable
The instrument is not discoverable by other devices.
Allows the instrument to be discoverable until another device pairs with it or until the instrument power is turned off and back on.
Makes the instrument discoverable indefinitely.
PINcode
The PINcode is a security key to be entered into the computer if prompted. The factory default PINcode is 0000.
NOTE: PINcode must be set to 0000 in order to use 8934 printer.
# AutoConnects
Specifies how many times the instrument will attempt to reattach to a paired device after the power is turned on. For this option, the instrument Discoverability setting must be enabled. Settings are 0 to 5 times.
For more information on establishing Bluetooth connections, refer to
TSI Applications Note TSI-150 .
Printing Data Using the Portable Printer
To print logged data, first enter the DATALOGGING menu. Then, use the
CHOOSE TEST item to select the data to be printed. After the test is selected, use the
VIEW STATS and VIEW SAMPLES items to select statistics or individual data points to view and print. After selecting
VIEW STATS or VIEW SAMPLES, press the PRINT key to print the data.
34 Chapter 3
T
RAK
P
RO
™ Data Analysis Software
The V
ELOCI
C
ALC
Model 9565 comes with special software called
T
RAK
P
RO
™ Data Analysis Software, which is designed to provide you with maximum flexibility and power.
Follow the instructions on the label of the T
RAK
P
RO
software to install the software on your computer. T
RAK
P
RO
software contains a very comprehensive Help Function. This utility provides all the necessary information to guide you in all aspects of software operation. The software is shipped on a CD-ROM. Updates are available from the
TSI website at http://www.tsi.com/SoftwareDownloads .
To download data from the Model 9565, connect the supplied computer interface USB cable to the Model 9565 and to a computer
USB port. Any USB port can be used.
LogDat2 ™ Downloading Software
The V
ELOCI
C
ALC
Model 9565 also includes downloading software called LogDat2.
LogDat2 software transfers the stored data from the Model 9565 to a computer as a spreadsheet file. This software is useful for applications such as duct traverses, fume hood, and filter face velocity testing.
Follow the instructions on the label of the LogDat2 software to install the software on your computer. The software is shipped on a CD-
ROM and includes a User’s Guide and Installation Instructions.
Updates are available from the TSI website at http://www.tsi.com/SoftwareDownloads .
To download data from the Model 9565, connect the supplied computer interface USB cable to the Model 9565 and to a computer
USB port. Any USB port can be used.
Operation 35
(This page intentionally left blank)
36 Chapter 3
Chapter 4
Maintenance
The Model 9565 and probe accessories require very little maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we recommend that you return your Model 9565, 960 series thermoanemometer probes, IAQ and VOC probes to TSI for annual recalibration. Please contact one of TSI’s offices or your local distributor to make service arrangements and to receive a Return
Material Authorization (RMA) number. To fill out an online RMA form, visit TSI’s website at http://service.tsi.com
.
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: (800) 874-2811
(651) 490-2811
Fax: (651) 490-3824
The Model 9565 and accessory probes can also be recalibrated in the field using the CALIBRATION menu. These field adjustments are intended to make minor changes in calibration to match a user's calibration standards. The field adjustment is NOT intended as a complete calibration capability. For complete, multiple-point calibration and certification, the instrument must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft cloth and isopropyl alcohol or a mild detergent. Never immerse the Model 9565. If the enclosure of the Model 9565 or the AC adapter becomes broken, it must be replaced immediately to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage.
37
(This page intentionally left blank)
38 Chapter 4
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Model 9565. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Model 9565
Symptom Possible Causes Corrective Action
No Display Unit not turned on Switch unit on.
Velocity reading fluctuates unstable
No response to keypad
Instrument Error message appears
Low or dead batteries
Dirty battery contacts
Fluctuating flow
Keypad locked out
Memory is full
Fault in instrument
Replace batteries or plug in AC adapter.
Clean the battery contacts.
Reposition probe in less turbulent flow or use longer time constant.
Unlock keypad by pressing keys simultaneously.
Download data if desired, then
DELETE ALL memory.
Factory service required on instrument.
Factory service required on probe.
Turn instrument OFF and then turn it back ON.
Probe Error message appears
Probe is plugged in, but the instrument does not recognize it
Fault in probe
Probe was plugged in when the instrument was
ON
W A R N I N G !
Remove the probe from excessive temperature immediately: excessive heat can damage the sensor. Operating temperature limits
can be found in Appendix A, Specifications . The pressure sensor is
protected from damage up to 7 psi (48 kPa or 360 mmHg). At higher pressure it can burst!
39
(This page intentionally left blank)
40 Chapter 5
Appendix A
Specifications
Specifications are subject to change without notice.
Velocity (TA Probe):
Range:
Accuracy
1&2
0 to 9999 ft/min (0 to 50 m/s)
: ±3% of reading or ±3 ft/min (±0.015 m/s), whichever is greater
Resolution: 1 ft/min (0.01 m/s)
Velocity (Pitot Tube):
Range
3
: 250 to 15500 ft/min (1.27 to 78.7 m/s)
Accuracy
4
: ±1.5% at 2000 ft/min (10.16 m/s)
Resolution: 1 ft/min (0.01 m/s)
Duct Size:
Range: 1 to 500 inches in increments of 0.1 in.
(2.5 to 1270 cm in increments of 0.1 cm)
Volumetric Flow Rate:
Range: Actual range is a function of actual velocity, pressure, duct size, and K factor
Temperature (TA Probe):
Range:
Accuracy
5
:
14 to 140°F (-10 to 60°C)
±0.5°F (±0.3°C)
Resolution: 0.1°F (0.1°C)
Relative Humidity (TA Probe):
Range:
Accuracy
6
:
5 to 95% RH
±3% RH
Resolution: 0.1% RH
Wet Bulb Temperature (TA Probe):
Range: 40 to 140°F (5 to 60°C)
Resolution: 0.1°F (0.1°C)
Dew Point (TA Probe):
Range: 5 to 120°F (-15 to 49°C)
Resolution: 0.1°F (0.1°C)
Heat Flow (TA Probe):
Range: Function of velocity, temperature, humidity, and barometric pressure
Measurements available: Sensible heat flow, latent heat flow,
Units measured: total heat flow and sensible heat factor
BTU/hr, kW
41
Static / Differential Pressure:
Range
7
: -15 to +15 in. H
2
O (-28.0 to +28.0 mm Hg, -3735 to
+3735 Pa)
Accuracy: ±1% of reading ±0.005 in. H
2
O (±1 Pa, ±0.01 mm Hg)
Resolution: 0.001 in. H
2
O (0.1 Pa, 0.01 mm Hg)
Barometric Pressure:
Range: 20.36 to 36.648 in. Hg (517.15 to 930.87 mm Hg)
Accuracy: ±2% of reading
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Operating (Probe):
Storage:
14 to 140°F (-10 to 60°C)
-4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Data Storage Capabilities:
Range: 26,500+ samples and 100 test IDs (one sample can contain fourteen measurement types)
Logging Interval:
Intervals: 1 second to 1 hour
Time Constant:
Intervals: User selectable
Response Time:
Velocity: 200 msec
Temperature: 2 minutes (to 66% of final value)
Pressure: 0.1 msec
Humidity: <1 minute (to 66% of final value)
External Meter Dimensions:
3.8 in.
8.3 in.
2.1 in. (9.7 cm
21.1 cm
5.3 cm)
Meter Probe Dimensions:
Probe length: 40 in. (101.6 cm)
Probe diameter of tip: 0.28 in. (7.0 mm)
Probe diameter of base: 0.51 in. (13.0 mm)
Articulating Probe Dimensions:
Articulating section length: 6.0 in. (15.24 cm)
Diameter of articulating knuckle: 0.38 in. (9.5 mm)
Meter Weight:
Weight with batteries: 0.8 lbs (0.36 kg)
42 Appendix A
Power Requirements:
Four AA-size batteries (included) or
AC Adapter p/n 801761
Input: 90 to 240 VAC, 50 to 60 Hz
Output: 9 VDC, 2A
1
Temperature compensated over an air temperature range of 40 to 150°F (5 to
65°C).
2
The accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is greater, begins at 30 ft/min through 9999 ft/min (0.15 m/s through 50 m/s).
3
Pressure velocity measurements are not recommended below 1000 ft/min (5 m/s) and are best suited to velocities over 2000 ft/min (10.00 m/s). Range can vary depending on barometric pressure.
4
Accuracy is a function of converting pressure to velocity. Conversion accuracy improves when actual pressure values increase.
5
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F
(0.03°C/°C) for change in instrument temperature.
6
Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C) for change in probe temperature. Includes 1% hysteresis.
7
Overpressure range = 190 in. H
2
O (7 psi , 360 mmHg, 48 kPa).
Specifications 43
(This page intentionally left blank)
44 Appendix A
Appendix B
Optional Plug-in Probes
Thermoanemometer Probes
Model
960
Description
Air Velocity and Temperature, Straight Probe
962
964
966
Air Velocity and Temperature, Articulating Probe
Air Velocity, Temperature, and Humidity, Straight Probe
Air Velocity, Temperature, and Humidity, Articulating Probe
Rotating Vane Anemometer Probes
Model Description
995 4 in. (100 mm) Rotating Vane, Air Velocity, and Temperature
Indoor Air Quality (IAQ) Probes
Model Description
980
982
Indoor Air Quality Probe, Temperature, Humidity, CO
2
Indoor Air Quality Probe, Temperature, Humidity, CO
Volatile Organic Compounds (VOC) Probes
2
and CO
Model
984
Description
Low Concentration (ppb) VOC and Temperature
985 High Concentration (ppm) VOC and Temperature
986
987
Low Concentration (ppb) VOC, Temperature, CO
2
, and Humidity
High Concentration (ppm) VOC, Temperature, CO
2
, and
Humidity
K-alloy Thermocouple Probes
Model
792
794
Description
Surface Temperature Probe
Air Temperature Probe
45
Pitot Probes
Model Description
634634000 Pitot Probe 12" (30 cm) - 5/16" (8 mm) diameter
634634001 Pitot Probe 18" (46 cm) - 5/16" (8 mm) diameter
634634002 Pitot Probe 24" (61 cm) - 5/16" (8 mm) diameter
634634003 Pitot Probe 36" (91 cm) - 5/16" (8 mm) diameter
634634005 Pitot Probe 60" (152 cm) - 5/16" (8 mm) diameter
634634004 Telescoping Pitot Probe - 8" to 38" (20 cm to 96 cm)
800187 Air Flow Probe (Straight Pitot) 18" (46 cm) - 5/16" (8 mm) diameter
46 Appendix B
TSI Incorporated
– Visit our website www.tsi.com for more information.
USA
UK
Tel: +1 800 874 2811
Tel: +44 149 4 459200
France
Tel: +33 4 91 11 87 64
Germany Tel: +49 241 523030
India
China
Tel: +91 80 67877200
Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
P/N 6004851 Rev D ©2014 TSI Incorporated Printed in U.S.A.
Advertisement
Key features
- Measures air velocity, temperature, humidity, and pressure
- Optional probes for CO, VOC, draft rate, heat flow, turbulence, and leakage testing
- Data logging with various modes and interval options
- Bluetooth communication for data transfer and printing
- User-friendly interface with easy navigation