advertisement
AVL 9120, 9130,
9140, 9180, 9181
Electrolyte Analyzers
Service Manual
ii
Tenth Edition
November 1997
Copyright, 1997, AVL Scientific Corporation. All rights reserved. Unless otherwise noted, no part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in any form without the written permission of AVL Scientific Corporation.
For information contact:
AVL Scientific Corporation
50 Mansell Court
P.O. Box 337
Roswell, Georgia USA 30077
1-800-526-2272
AVL MEDICAL INSTRUMENTS AG
Stettemerstrasse 28
CH-8207 Schaffhausen
Switzerland
41-848-800-885
AVL LIST GmbH
Kleiststrasse 48
A-8020 Graz
Austria
43-316-787-0
Printed in USA
US0285 REV K
Important Information!
This Operator´s Manual contains important warnings and safety information to be observed by the user.
This instrument is only intended for one area of application which is described in the instructions. The most important prerequisites for application, operation and safety, are explained to ensure smooth operation. No warranty or liability claims will be covered if the instrument is applied in areas other than those described or if the necessary prerequisites and safety measures are not observed.
The instrument is only to be operated by qualified personnel capable of observing these prerequisites.
Only accessories and supplies either delivered by or approved by AVL are to be used with the instrument.
Due to the operating principle of the instrument, analytical accuracy not only depends on correct operation and function, but also upon a variety of external influences beyond the manufacturers control. Therefore, the test results from this instrument must be carefully examined by an expert, before further measures are taken based on the analytical results.
Instrument adjustment and maintenance with removed covers and connected power mains are only to be performed by a qualified technician who is aware of the dangers involved.
Instrument repairs are only to be performed by the manufacturer or qualified service personnel.
Symbol Explanation
Attention symbol - Refer to the Operator's Manual or Service
Manual for further instructions. This symbol is located on the inside of the instrument.
Type B instrument symbol - An instrument of the B type falls under safety categories I, II, or III, or has an internal power supply providing the required insulation against discharge current and reliable ground connections.
Important Information!
iii
iv
Operating Safety Information
• This instrument falls under Safety Category I.
• This instrument is a Class B instrument.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference’s, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Warning: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communication. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does not cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna
• Increase the separation between the equipment and receiver
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio TV technician for help
Caution:
• The instrument is designed as a conventional device (closed, not waterproof type).
• Do not operate the instrument in an explosive environment or in the vicinity of explosive anesthetic mixtures containing oxygen or nitrous oxide.
• This instrument is suitable for continuous operation.
• The power plug is to be plugged into a ground socket only. When using an extension cord, make sure that it is of the proper size and is properly grounded.
• Any breakage of the ground lead inside or outside the instrument or a loose ground connection can cause a hazardous condition when operating the instrument.
Intentional disconnection of the grounding is not permitted.
• When replacing the fuses, make sure that they are of the same type and rating as the original fuses. Never use repaired fuses or short-circuit the fuse holders.
Operating Safety Information
Contents
2.1 Reported Parameters .................................................................. 2
2.2 Operating Parameters ................................................................. 2
3.1 System Software Flow Charts 9120, 30, 40 ............................... 7
3.2 System Software Flow Charts 9180 ......................................... 17
3.3 System Software Flow Charts 9181 ......................................... 29
3.4 Service Functions .................................................................... 42
4 Description of Modules ....................................................................... 48
4.1 Mechanical Assemblies ............................................................ 48
4.2 Fluidic Module ........................................................................ 52
4.3 Electronics ............................................................................... 56
5 Maintenance and Troubleshooting ..................................................... 59
5.1 Maintenance Procedures ......................................................... 59
5.2 Error Messages and Troubleshooting ...................................... 60
6.1 Sample Sensor Adjustment ...................................................... 71
6.2 Temperature Adjustment .......................................................... 71
System Block Diagram 9120, 30, 40, 80 ........................................ 74
System Block Diagram 9181 .......................................................... 75
System Interconnect 9120, 30, 40 .................................................. 76
SBC PCB - Power Supply 9120, 30, 40 ........................................ 77
SBC PCB - Valve Drivers 9120, 30, 40 ......................................... 78
SBC PCB - Pump Motor Driver/Door Detect 9120, 30, 40 ............ 79
SBC PCB - Input Amplifiers 9120, 30, 40 ...................................... 80
SBC PCB - Temperature Circuit 9120, 30, 40 ............................... 81
SBC PCB - Analog Selector and A/D Converter 9120, 30, 40 ....... 82
v
vi
SBC PCB - Microprocessor 9120, 30, 40 ..................................... 83
SBC PCB - Printer/Display Drivers 9120, 30, 40 ........................... 84
Printer Daughter PCB 9120, 30, 40, 80, 81 .................................... 85
Lamp PCB 9120, 30, 40 ................................................................ 86
Display Daughter PCB 9120, 30, 40 ............................................... 87
System Interconnect 9180/9181 ..................................................... 88
SBC PCB - Power Supply 9180/9181 ........................................... 89
SBC PCB - Valve Drivers 9180/9181 ............................................ 90
SBC PCB - Pump Motor Driver/Door Detect 9180/9181 ............... 91
SBC PCB - Input Amplifiers 9180/9181 ......................................... 92
SBC PCB - Temperature Circuit 9180/9181 .................................. 93
SBC PCB - Analog Selector and A/D Converter 9180/9181 .......... 94
SBC PCB - Microprocessor 9180/9181 ........................................ 95
SBC PCB - Printer/Display Drivers 9180/9181 .............................. 96
SBC PCB - Fluid Pack Detector 9180/9181 .................................. 97
Lamp PCB 9180/9181 ................................................................... 98
Display Daughter PCB 9180 ........................................................... 99
Main Assembly Front ................................................................... 101
Main Assembly Rear .................................................................... 102
Electrode Holder Assembly .......................................................... 103
Needle Unit Assembly .................................................................. 104
Rear Panel Assembly .................................................................... 105
Front Door Assembly ................................................................... 106
9 Interface Specifications ..................................................................... 107
9.1 Interface Information.............................................................. 107
9.2 Example Data String Information ............................................ 108
9.3 Data Link Information ............................................................ 108
1 Introduction
The Service Manual for the 9120, 30, 40, 80, and 81 Electrolyte Analyzers contains the technical information needed to ensure easy fault identification.
This manual is intended to be complementary to the Operator's Manual where detailed instructions for operation, maintenance and troubleshooting are provided.
As for all clinical instrumentation, a thorough understanding of the principles of operation is prerequisite to attempting service of this product. Training along with experience will enhance the use of this manual.
Service and repair of 9120, 30, 40, 80, and 81 analyzers should be performed only by qualified repair technicians. Care should be taken when removing the covers as hazardous voltages are exposed. Use only accepted electronic test procedures and static protection when replacing and handling all electronic parts.
This manual is divided into 9 chapters to facilitate location of technical information. Chapter 2 provides specifications and operating parameter information. Chapter 3 includes flow charts for all system functions and detailed operation of system test procedures. In Chapter 4, all mechanical, fluidic and electronic assemblies are described. Chapter 5 outlines routine maintenance and troubleshooting procedures. Chapter 6 includes electronic adjustments and Chapter 7 provides the system block diagram and all circuit schematic diagrams and wiring interconnection information. In Chapter 8, part identification, location and description is provided by the illustrated parts list.
Finally, in Chapter 9 interface specifications are provided for the RS232 output.
1
2
2 General Description
Specifications
2.1 Reported Parameters
Whole Blood
Serum
Plasma
Measuring Range Resolution
Sodium Na + 40 - 205 mmol/L
Potassium K + 1.5 - 15 mmol/L
0.1 mmol/L
0.01 mmol/L
(0.8 - 15 mmol/L dialysate)
Chloride Cl 50 - 200 mmol/L
Calcium Ca ++ 0.2 - 5.0 mmol/L
Lithium Li + 0.1 - 6.0 mmol/L
0.1 mmol/L
0.01 mmol/L
0.001 mmol/L
Urine
Sodium Na
Potassium K +
+ 1 - 300 mmol/L 1.0 mmol/L
4.5 - 120 mmol/L 0.1 mmol/L
(60 - 120 mmol/L w. addit. dilution)
Chloride Cl 1 - 300 mmol/L 1.0 mmol/L
Calcium and Lithium are not measured in urine samples.
Lithium is not measured in dialysate samples.
2.2 Operating Parameters
Sample type Whole blood, serum, plasma, urine, dialysate (acetate or bicarbonate)
Sample device Syringe, sample cup, collection tube, capillary
Sample size 95 µL
Analysis time
Sample rate
50 seconds
45 per hour with printout
60 per hour without printout
Measurement
Sodium (Na
+
) sensor
Potassium (K
Chloride (Cl
Calcium (Ca
Lithium (Li
Reference System
Calibration
Standby Mode
Warm-up time
Temperature
Humidity
+
-
+
) sensor
) sensor
++
) sensor
) sensor
Data management
Diagnostic Programs
Electronics
Display
Printer
Autosampler (9181 only)
Ion-selective, flow-through, glass capillary electrode
Ion-selective, flow-through, liquid membrane electrode
Ion-selective, flow-through, liquid membrane electrode
Ion-selective, flow-through, liquid membrane electrode
Ion-selective, flow-through, liquid membrane electrode
Open liquid junction, flow-through electrode
Fully automatic
1 point with each sample
2 point every 3 hours (9180/9181 every 4 hrs.) suspends calibrations
1 minute
Room temperature, 15 - 32° C, 60 - 90° F maximum 85% relative humidity, non-condensing
Quality Control memory storage, 3 levels,
35 values; calculation of mean, standard deviation, and coefficient of variation (CV) user-controlled diagnostics,
YES/NO operation via the display
Microprocessor-controlled
LCD dot-matrix, 2 line, 16 characters per line
Integral thermal printer, 16 character width
Integral turntable - 18 positions,
2 mL or 0.25 mL sample cups
3
4
Interface
Data Link
Electrical requirements
Nominal power consumption
RS232C Serial Port
Data link to Compact 2/3 (9180/9181)
100 - 240 V~, 50/60 Hz,
1.4 A max.
Self-adjusting
30 W
Dimensions height width depth weight
Classifications
Safety category
Device type
Mode of operation
Protection classification
Explosion protection
Approvals
12.2"; 335mm
12.4"; 315m
12.0"; 295mm approx. 13 lbs; 6kg
I
B (according to ÖVE-MG/EN 60601-1,
IEC 601-1) continuous operation
IP 20 the device is not designed for operation in explosive environments.
CSA, CE, FCC Class B
CLIA Complexity Category: Moderate
Test System Code: 04739
FDA 510(k) number: K972673
Analyte Codes: Na + : 5805
K + : 4910
Cl
-
: 1018 iCa ++ : 1004
Li + : 3712
Data subject to change without notice. Technical information is supplied for general informational purposes only.
3 Software Operation
3.1 Software Operation - 9120, 9130, 9140
Software operation of the 9120, 9130, 9140 analyzers can be accessed by YES/
NO selection on the analyzer front panel. The following flow chart diagrams are provided to identify operating sequences of the 9120, 9130, 9140 Analyzers:
1. Overall Program Operating Flow
2. Main Menu
3. Measurement Sequence
4. Print Functions Menu
5. QC/Standard/Urine Sample Menu
6. Daily Maintenance Menu
7. Operator Functions Menu
8. Program Instrument Menu
9. Service Functions Menu
10. Calibration Sequence
3.2 Software Operation - 9180
1. Overall Program Operating Flow
2. Main Menu
3. Measurement Sequence
4. Print Functions Menu
5. QC/Std/Dialysate/Urine Sample Menu
6. Daily Maintenance Menu
7. Operator Functions Menu
8. Program Instrument Menu
9. Service Functions Menu
10. Calibration Sequence
11. Power-up Sequence
5
6
3.3 Software Operation - 9181
1. Overall Program Operating Flow
2. Main Menu
3. Measurement Sequence
4. Print Functions Menu
5. QC/Std/Dialysate/Urine Sample Menu
6. Daily Maintenance Menu
7. Operator Functions Menu
8. Program Instrument Menu
9. Service Functions Menu
10. Calibration Sequence
11. Power-up Sequence
The Service Functions Menu is also provided with a more detailed description of each test parameter for use in testing instrument subassemblies.
3.1 Software Operation - 9120, 30, 40
1 Overall Program Operating Flow - 9120, 30, 40
7
2 Main Menu - 9120, 30, 40
MAIN MENU
READY 12:45
NO
FUNCTIONS?
NO
QC/STANDARD/
URINE SAMPLE?
NO
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
NO
PROGRAM
INSTRUMENT?
NO
SERVICE
FUNCTIONS?
NO
NO
CALIBRATION?
8
LEGEND
Display prompt
Action performed by analyzer
2-second display prompt
Decision-making by analyzer
3 Measurement Sequence - 9120, 30, 40
READY 12:45
YES
NO
OPEN
FLAP
FUNCTIONS?
Open Sample Door
Introduce Sample
NO
DOOR OPENED
WITHIN 2s?
YES
Please
Introduce Sample
CHECK SS
VOLTAGE
> air voltage
< air voltage
Remove Sample and Wipe Probe
Remove Sample and Wipe Probe
2s
Please Close
Sample Door
FLAP
CLOSED
Serum Sample in Process 49
2s
Please Close
Sample Door
FLAP
CLOSED
Blood Sample in Process 49
Na K Cl
151 5.1 114
Na K Cl
151 5.1 114
YES/NO
1 min.
YES/NO
1 min.
9
4 Print Functions Menu - 9120, 30, 40
FUNCTIONS?
YES
Print Last
Sample Report?
NO
Cal. Report?
NO
Print QC Values and Statistics?
NO
YES
Remain in Print
Functions?
NO
YES
YES
YES
NO
QC/STANDARD/
URINE SAMPLE?
PRINTING
...
PRINTING
...
PRINTING
...
READY 12:45
READY 12:45
10
5 QC/Standard/Urine Sample Menu - 9120, 30, 40
QC/STANDARD/
URINE SAMPLE?
NO
YES
QC Level 1
Sample?
YES
DAILY
MAINTENANCE?
Open Sample Door
Introduce Sample
OPEN
2 min.
NO
2s
2s
Introduce
Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
SAMPLE
Wipe Probe/Close
Sample Door
CLOSED
IN
QC Level 1 in Process 49
2s
NO
2s
Store
Values?
Na K Cl
99.65.34 88.2
2 min
YES
READY 12:45
QC Level 2
Sample?
2 min.
NO
YES
Open Sample Door
Introduce Sample
OPEN
2s
2s
Introduce
Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
SAMPLE
IN
Wipe Probe/Close
Sample Door
CLOSED
QC Level 2 in Process 49
2s
NO
2s
Store
Values?
Na K Cl
99.6 5.34 88.2
2 min
YES
READY 12:45
QC Level 3
Sample?
2 min.
NO
YES
Open Sample Door
Introduce Sample
OPEN
2s
2s
Introduce
Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
SAMPLE
Wipe Probe/Close
Sample Door
CLOSED
QC Level 3 in Process 49
IN
2s
NO
2s
Store
Values?
Na K Cl
99.6 5.34 88.2
2 min
YES
READY 12:45
YES
5s
Open Sample Door
Introduce Sample
OPEN
Introduce
Sample
SAMPLE
IN
Wipe Probe/Close
CLOSED
Sample Door
Standard Sample in Process 49
2 min.
Standard
Sample?
NO
2 min.
Urine
Sample?
NO
YES
5s
Open Sample Door
Introduce Sample
OPEN
Introduce
Sample
SAMPLE
IN
Wipe Probe/Close
Sample Door
CLOSED
Urine Sample in Process 49
YES
Remain in QC/
Std/Urine Sample?
NO/
2 min.
READY 12:45
Na K Cl
158.8 5.82 113.2
1 min. NO/
YES
Na K Cl
238 18.8 198
1 min.
NO/
YES
11
6 Daily Maintenance Menu - 9120, 30, 40
DAILY
MAINTENANCE?
YES
NO
Daily
Cleaning?
NO
OPERATOR
FUNCTIONS?
YES
OPEN
Open Sample Door
Introduce Sample
OPEN
Wipe Probe/Close
CLOSE
Sample Door
Cleaning Solution detected?
YES
NO
Cleaning Solution not detected!
CLEAN
Daily
Conditioning?
YES
OPEN
Open Sample Door
Introduce Sample
OPEN
Wipe Probe/Close
Sample Door
CLOSE
Condition. Solution detected?
YES
CONDITION
NO
Condition. Solution not detected!
YES
Remain in Daily
Maintenance?
NO/1min.
CLEANING AND/OR
COND. PERFORMED?
YES
30 SECOND
STABILIZATION
NO
Calibration due?
NO
YES
CALIBRATE
READY 12:45
12
7 Operator Functions Menu - 9120, 30, 40
OPERATOR
FUNCTIONS?
YES
Check/Change
SnapPak?
NO
Go to
Standby Mode
NO
NO
YES
YES
PROGRAM
INSTRUMENT?
37% remaining
***STANDBY!***
YES -> READY
YES/NO
Change
SnapPak?
NO
YES
New SnapPak installed?
NO
YES
Fluid Counter reset to 100%
Maint. due?
NO
New Pack
Prime ...
YES
Set
Time/Date?
NO
NO
Leave
Standby Mode?
YES
PRESS YES
WITHIN 2s
YES
IS
CALIBRATION
DUE?
YES
Daily
Cleaning?
YES
NO
NO
CLEAN
YES
NO
Date: 01-FEB-94
Time: 12:00 ok ?
NO
Date: 01-FEB-94
Time: 12:00 ok
YES
YES
READY 12:45
Daily
Conditioning?
YES
CONDITION
30 SECOND
STABILIZATION
CALIBRATE
CALIBRATE
READY 12:45
NO
YES
Remain in Daily
Maintenance?
1min/NO
IS
CALIBRATION
DUE?
NO
YES
NO
Reset
Sample Number?
NO
Take out of
Operation?
NO
Remain in
Operator Func?
YES
YES
YES
Are you sure?
YES
Shutdown Plug /
Water installed?
YES/
NO
YES
ShutdownComplete
Turn Power Off
YES
SET SAMPLE
NUMBER TO 0
Transp Housing
Ref installed?
Do You Wish to
DELETE all Data?
YES
YES
Remove Water and Press YES
Pump Windings
Relieved?
YES
YES
All Electrodes
& Plug Removed?
YES
4 Relief Clamps
Inserted?
13
8 Program Instrument Menu - 9120, 30, 40
PROGRAM
INSTRUMENT?
YES
Enter Code:
AAA
CORRECT
NO
YES
IN-
CORRECT CODE ERROR!
RETRY?
NO
Program QC
Level 1 Ranges?
NO
YES
Current Lot: 0000
Change Lot?
NO
YES
YES
Cl low = 050
Cl high = 200
YES
SERVICE
FUNCTIONS?
Print old Values and Statistics?
NO
K low = 1.5
K high = 15.0
YES
NO
New Lot!
Delete old Data?
YES
Na low = 040
Na high = 205
YES
Enter last 4 Dig of Lot#: 0000
YES
Lot Number ok?
NO
Program QC
Level 2 Ranges?
NO
YES
Program QC
Level 3 Ranges?
NO
YES
"see Level
1"
"see Level
1"
Program
Normal Ranges?
NO
YES
YES
Program
Corr Factors
NO
Na low = 135.0
Na high = 150.0
Reset Corr
Factor (default)?
YES
YES
Cl (b) = +00.0
Cl (m) = 1.000
YES
NO
K low = 3.50
K high = 5.50
Input/Verify
Corr Factors?
YES
NO
YES
YES
K (b) = +00.0
K (m) = 1.000
YES
Cl low = 95.0
Cl high = 108.0
YES
Remain in
Corr Factors?
NO
Na (b) = +00.0
Na (m) = 1.000
YES
Print two
Patient Reports?
YES/
NO Automatic
Cal Report?
YES/
NO
Program Printer
Setup?
NO
YES
Remain in
Program Func?
NO
READY 12:45
YES
Patient Report off?
YES
NO
14
9 Service Functions Menu - 9120, 30, 40
SERVICE
FUNCTIONS?
YES
Test
Electrodes?
NO
NO
CALIBRATION?
YES
Test
Standard A?
NO
Test
Standard B?
NO
Test
External Sample?
YES
YES
YES
Std. A: Na =0000
K =0000 Cl =0000
NO
Std. B: Na =0000
K =0000 Cl =0000
NO
Open Door and
Introduce Sample
OPEN
Introduce Sample
Test
Sample Sensor?
NO
Test
Sample Door?
NO
Test Pump?
NO
Test Valves?
NO
Test Interface?
Jump Pins 2&3
NO
Test
Amplifier?
NO
Enter
Service Code?
NO
YES
Remain in
Test Electrodes?
NO
NO
Ext.: Na =0000
K =0000 Cl =0000
YES
DRY: 80-120: 102
YES=Pump/NO=Exit
YES
NO
External: 102
Pump: NO=OFF
CLOSED
Wipe Probe and
Close Sample Door wash...
YES
YES
Sample Door: C
C:Closed O:Open
NO
Test Pump!
YES
YES
Test Valve A:
YES/NO=ON/OFF
-- Test --
-- Interface --
NO
Test Valve B:
YES/NO=ON/OFF
NO
Interface ok!
Test Valve V:
YES/NO=ON/OFF
YES
Remain in
Test Interface?
NO
YES
Na=-1030 K=-1200
Cl = -1093
NO
Ground = 2
-2500mV = -2496
YES
Very slow speed
Enter Code:
AAA ok?
2s
YES
Test Pump!
Slow speed
Is Code valid?
NO
Code
Error!
2s
YES
Test Pump!
Medium speed
Code
Accepted!
2s
NO
Test Pump!
Fast speed
2s
Test Valve R:
YES/NO=ON/OFF
NO
YES
Enter Other
Service Code?
NO
YES
Remain in
Service Func?
NO
Calibration?
Calibrate...
READY 12:45
15
10 Calibration Sequence - 9120, 30, 40
CALIBRATION?
YES
Calibration in
Process 124
Clear Sample
Path
Calibrate
Sample Sensor
Sample Sensor
Voltage ok?
NO
YES
Prime
Standard B
***CHECK***
SAMPLE SENSOR
Standard B detected?
YES
Prime
Standard A
NO
**STANDARD B**
NOT DETECTED
Standard A detected?
YES
Pressurize Ref.
Asp. Std. A into
Meas. Chamber
NO
**STANDARD A**
NOT DETECTED
Reference
Housing ok?
YES
A
NO
***CHECK***
REF HOUSING
A
Asp. Std. A into
Meas. Chamber
Standard A detected?
YES
Obtain Electrode
Voltages Std. A
NO
**STANDARD A**
NOT DETECTED
3 Std. A Values obtained?
YES
NO
Are Values reproducible?
YES
Aspir. Std. B into
Meas. Chamber
NO
Standard B detected?
YES
Obtain Electrode
Voltages Std. B
NO
**STANDARD B**
NOT DETECTED
NO
6 Aspir. of
Std.A performed?
YES
***CHECK***
ELECTRODES
Perform
Wash Cycle
Aspir. Std. A into Meas.
Chamber
Standard A detected?
YES
Calculation of
Std.A-Std.B
(slope)
NO
**STANDARD A**
NOT DETECTED
Slope of all
Electrodes in
Range?
YES
READY 12:45
NO
At least 1 slope in Range?
N O
YES
READY 12:45
XX NOT CAL'D
***CHECK***
ELECTRODES
16
3.2 Software Operation - 9180
1 Overall Program Operating Flow - 9180
17
2 Main Menu - 9180
NO
MAIN MENU
NA K CL
READY
NO
FUNCTIONS?
NO
QC/STD/DIALYSATE
/URINE SAMPLE?
NO
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
NO
PROGRAM
INSTRUMENT?
NO
SERVICE
FUNCTIONS?
NO
CALIBRATION?
18
LEGEND
Display prompt
Action performed by analyzer
2-second display prompt
Decision-making by analyzer
3 Measurement Sequence - 9180
NA K CL READY
YES
NO
OPEN
FLAP
FUNCTIONS?
Open Sample Door
Introduce Sample
NO
DOOR OPENED
WITHIN 2s?
YES
Please
Introduce Sample
CHECK SS
VOLTAGE
> air voltage
< air voltage
Remove Sample and
Wipe Probe
Remove Sample and
Wipe Probe
2s
Please Close
Sample Door
FLAP
CLOSED
Serum Sample in Process 49
2s
Please Close
Sample Door
FLAP
CLOSED
Blood Sample in Process 49
Na K Cl
151 5.1 114
Na K Cl
151 5.1 114
YES/NO
1 min.
YES/NO
1 min.
19
4 Print Functions Menu - 9180
YES
FUNCTIONS?
YES
Print Last
Sample Report?
NO
Cal. Report?
NO
Print QC Values and Statistics?
NO
Remain in Print
Functions?
NO
YES
YES
YES
NO
QC/STD/DIALYSATE
/URINE SAMPLE?
PRINTING...
PRINTING...
PRINTING...
NA K CL READY
NA K CL READY
20
5 QC/Standard/Dialysate/Urine Sample Menu - 9180
QC/STD/DIALYSATE
/URINE SAMPLE?
NO
2 min.
YES
QC Level 1
Sample?
NO
YES
OPEN
DAILY
MAINTENANCE?
Open Sample Door
Introduce Sample
OPEN
2 s
2 s
Introduce Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
SAMPLE
IN
Wipe Probe/Close
Sample Door
2s
NO
Store
Values?
YES
2 min
NA K CA READY
CLOSED
2s
QC Level 1 in Process 49
Na K Ca
99.6 5.34 1.250
2 min.
QC Level 2
Sample?
NO
YES
OPEN
Open Sample Door
Introduce Sample
OPEN
2 s
2 s
Introduce Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
SAMPLE
IN
Wipe Probe/Close
Sample Door
2s
NO
Store
Values?
2 min
YES
NA K CA READY
CLOSED
2s
QC Level 2 in Process 49
Na K Ca
99.6 5.34 1.250
2 min.
QC Level 3
Sample?
NO
YES
OPEN
Open Sample Door
Introduce Sample
OPEN
2 s
2 s
Introduce Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
SAMPLE
IN
Wipe Probe/Close
Sample Door
NO
YES
2 min
2s
Store
Values?
NA K CA READY
CLOSED
2s
QC Level 3 in Process 49
Na K Ca
99.6 5.34 1.250
2 min.
Standard
Sample?
NO
2 min.
Bicarbonate
Sample?
NO
2 min.
Acetate
Sample?
NO
2 min.
Urine
Sample?
NO
YES
YES
5 s
Open Sample Door
Introduce Sample
OPEN
YES
OPEN
5 s
Open Sample Door
Introduce Sample
5 s
Open Sample Door
Introduce Sample
OPEN
YES
OPEN
5 s
Open Sample Door
Introduce Sample
YES
Remain in QC/
Std/Urine Sample?
NO/2 min.
NA K CA READY
OPEN
OPEN
OPEN
OPEN
Introduce Sample
Introduce Sample
Introduce Sample
Introduce Sample
SAMPLE
IN
SAMPLE
IN
SAMPLE
IN
SAMPLE
IN
Wipe Probe/Close
Sample Door
Wipe Probe/Close
Sample Door
Wipe Probe/Close
Sample Door
Wipe Probe/Close
Sample Door
CLOSED
CLOSED
CLOSED
CLOSED
Standard Sample in Process 49
Bicarb. Sample in Process 49
Acetate Sample in Process 49
Urine Sample in Process 49
Na K Ca
158.8 5.82 1.250
1 min.
NO/YES
Na K Ca
158.8 5.82 1.250
1 min.
NO/YES
Na K Ca
158.8 5.82 1.250
1 min.
NO/YES
Na K Cl
238 18.8 198
1 min.
NO/YES
21
6 Daily Maintenance Menu - 9180
DAILY
MAINTENANCE?
YES
Perform Daily
Cleaning?
NO
NO
OPERATOR
FUNCTIONS?
YES
OPEN
Open Sample Door
Introduce Sample
OPEN
Wipe Probe/Close
Sample Door
CLOSE
Cleaning Solution detected?
NO
Cleaning Solution not detected!
YES
CLEAN
Perform Daily
Conditioning?
YES
OPEN
Open Sample Door
Introduce Sample
OPEN
Wipe Probe/Close
Sample Door
CLOSE
Condition. Solution detected?
NO
Condition. Solution not detected!
YES
CONDITION
YES
Remain in Daily
Maintenance?
NO/1min.
CLEANING AND/OR
COND. PERFORMED?
YES
NO
Calibration due?
NO
YES
30 SECOND
STABILIZATION
CALIBRATE
NA K CA READY
22
7 Operator Functions Menu - 9180
OPERATOR
FUNCTIONS?
YES
Check/Change
SnapPak?
NO
Go to
Standby Mode
NO
NO
YES
YES
PROGRAM
INSTRUMENT?
37% remaining
***STANDBY!***
YES -> READY
YES/NO
Change
SnapPak?
NO
YES
*****STATUS*****
**NO SnapPak**
Pack detected?
GLOBAL*
New SnapPak installed?
NO
YES
NO
*Background check in all menus except "Take out of Operation?"
YES
Fluid Counter reset to 100%
YES
Maint. due?
NO
New Pack
Prime ...
NO
Set
Time/Date?
NO
Select Parameter
Configuration?
NO
Reset
Sample Number?
NO
Take out of
Operation?
NO
Remain in
Operator Func?
YES
NO
YES
NO
YES
YES
YES
Leave
Standby Mode?
YES
PRESS YES
WITHIN 2s
Date: 01-FEB-94
Time: 12:00 ok ?
NO
Date: 01-FEB-94
Time: 12:00 ok
YES
YES
YES
Sel. Parameter:
[Na] [K] [ ] ok?
Are you sure?
YES/
NO
YES
Shutdown Plug and
Water installed?
YES
Shutdown
Complete
Turn Power Off
CALIBRATE
YES
NO
IS
CALIBRATION
DUE?
YES
Daily
Cleaning?
YES
CLEAN
NA K READY
YES
Sel. Parameter:
[Na] [K] [Cl] ok?
SET SAMPLE
NUMBER TO 0
NO
Na K READY
NO
Daily
Conditioning?
YES
CONDITION
30 SECOND
STABILIZATION
CALIBRATE
NO
NO
YES
Sel. Parameter:
[Na] [K] [Ca] ok?
YES
Sel. Parameter:
([Na]) [ ] [Li] ok?
NO
YES
Remain in Daily
Maintenance?
1min/NO
IS
CALIBRATION
NO
YES
DUE?
NO
New
Parameter added?
YES
Electrodes in place?
NO
NO
YES
Sel. Parameter:
[Na] [K] [Li] ok?
YES
NO
Sel. Parameter:
[Na] [ ] [Li] ok?
YES
YES
Transp Housing
Ref installed?
Do You Wish to
DELETE all Data?
YES
YES
Remove Water and Press YES
Pump Windings
Relieved?
YES
YES
All Electrodes
& Plug Removed?
YES
5 Relief Clamps
Inserted?
23
8 Program Instrument Menu - 9180
PROGRAM
INSTRUMENT?
YES
Enter Code:
AAA
CORRECT
Program QC
Level 1 Ranges?
NO
NO
IN-
CORRECT
YES
NO
SERVICE
FUNCTIONS?
YES
CODE ERROR!
RETRY?
NO
Current Lot: 0000
Change Lot?
NO
Additional
Parameters?
YES
YES
Print old Values and Statistics?
NO
NO
New Lot!
Delete old Data?
YES
YES
K low = 1.5
K high = 15.0 ok?
YES
Na low = 040
Na high = 205 ok?
YES
Enter last 4 Dig of Lot#: 0000
YES
Enter last 4 Dig of Lot#: 0000 ok?
NO
Cl low = 050
Cl high = 200 ok?
YES
Ca low = 0.2
Ca high = 5.0 ok?
YES
Li low = 0.1
Li high = 6.0 ok?
YES
Program QC
Level 2 Ranges?
NO
Program QC
Level 3 Ranges?
NO
Normal Ranges?
NO
Corr Factors
NO
Program
Program
YES
YES
Program Bicarb.
Corr Factors
NO
YES
Program Acetate
Corr Factors
NO
YES
Program Urine
Corr Factors
NO
YES
Program Printer
Setup?
NO
YES
YES
NO
Program
Interface?
Remain in
Program Func?
YES
NO
"see Level 1"
"see Level 1"
YES
YES
Na low = 136
Na high = 145 ok?
YES
K low = 03.5
K high = 05.1 ok?
YES
Reset Corr
Factor (default)?
YES
Reset Bicarb.
Factor (default)?
YES
Reset Acetate
Factor (default)?
YES
Reset Urine
Factor (default)?
YES
NO
NO
NO
NO
Input/Verify
Corr Factors?
Input/Verify
Bicarb. Factors?
Input/Verify
Acetate Factors?
Input/Verify
Urine Factors?
NO
YES
NO
YES
NO
YES
NO
YES
YES
Remain in
Corr Factors?
NO
YES
Remain in
Bicarb Factors?
NO
YES
Remain in
Acetate Factors?
NO
YES
Remain in
Urine Factors?
NO
Patient Report off?
YES
NO
Print two
Patient Reports?
YES/
NO Automatic
Cal Report?
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
YES
YES
YES
K (b) = +00.0
K (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
YES
Na (b) = +00.0
Na (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
YES
YES/
NO Enter Comment
Line?
NO
YES
Enter Comment:
_
YES
Activate
Data Link?
YES/NO
NA K READY
24
9 Service Functions Menu - 9180
SERVICE
FUNCTIONS?
YES
NO
Test
Electrodes?
Test
Sample Sensor?
NO
Test
SnapPak Sensor?
NO
Test
Language Switch?
NO
Test
Sample Door?
NO
Test Pump?
NO
YES
YES
NO
YES
CALIBRATION?
NO
Test
Standard A?
Test
Standard C?
NO
Test
External Sample?
NO
Remain in
Test Electrodes?
DRY: 80-120: 102
YES=Pump/NO=Exit
YES
YES
YES
Std. A: Na = 0000
K = 0000 Ca = 0000
NO
Std. C: Na = 0000
K = 0000 Ca = 0000
NO
Open Door and
Introduce Sample
NO
OPEN
Test
Standard B?
NO
Introduce Sample
NO
YES
Ext.: Na = 0000
K = 0000 Ca = 0000
NO
External: 102
Pump: NO=OFF
CLOSED
Wipe Probe and
Close Sample Door wash...
YES
YES
SnapPak: I
I=In O=Out
Language Switch
English
YES/NO
YES/NO
NO
Test Valves?
NO
Test Interface?
Jump Pins 2&3
NO
NO
Test
Amplifier?
NO
Enter
Service Code?
YES
Std. B: Na = 0000
K = 0000 Ca = 0000
NO
YES
YES
YES
YES
Sample Door: C
C:Closed O:Open
NO
Test Pump!
Very slow speed
Test Valve A:
YES/NO=ON/OFF
-- Test --
-- Interface --
2s
NO
Test Pump!
Slow speed
Test Valve B:
YES/NO=ON/OFF
Interface ok!
YES
Na = -1030 K = -1200
Cl = -1093
YES
Enter Code:
AAA ok?
NO
YES
YES
Ground = 2
-2500mV = -2496
NO
NO
YES
Is Code valid?
NO
Code Error!
Retry?
2s
NO
Test Pump!
Medium speed
Test Valve C:
YES/NO=ON/OFF
YES
Remain in
Test Interface?
NO
Temp (mV) = -2670
Temp (°C) = 26.2
NO
Code
Accepted!
2s
Test Pump!
Fast speed
2s
NO
NO
Test Valve V:
YES/NO=ON/OFF
NO
Test Valve R:
YES/NO=ON/OFF
YES
Enter Other
Service Code?
NO
YES
Remain in
Service Func?
NO
Calibration?
Calibrate...
NA K CA READY
25
10 Calibration Sequence - 9180 (1)
(Example Na/K/Li Calibration)
26
CALIBRATION?
YES
Calibration in
Process 124
Clear Sample
Path
Calibrate
Sample Sensor
Sample Sensor
Voltage ok?
YES
Prime
Standard B
NO
***CHECK***
SAMPLE SENSOR
Prime
Standard C
Prime
Standard A
NO
1st try
All 3 Standards not detected?
NO
Two Standards not detected?
NO
Standard A,B,C not detected?
NO
Pressurize Ref.
Asp. Std. A into
Meas. Chamber
YES
YES
YES
*********CLOG********
CHECK FLUID PATH
**STANDARD X/Y**
NOT DETECTED
**STANDARD X**
NOT DETECTED
Reference
Housing ok?
A
YES
NO
2nd try
***CHECK***
REF HOUSING
A
Asp. Std. A into
Meas. Chamber
Standard A detected?
YES
Obtain Electrode
Voltages Std. A
NO
3 Std. A Values obtained?
YES
Are Values reproducible?
YES
Aspir. Std. B into
Meas. Chamber
NO
NO
Standard B detected?
YES
Obtain Electrode
Voltages Std. B
NO
**STANDARD A**
NOT DETECTED
**STANDARD B**
NOT DETECTED
YES
NO
6 Aspir. of
Std.A performed?
YES
At least 1
Channel reprod. ?
NO
***CHECK***
ELECTRODES
Calc. of Std.A-
Std.B (slope)
At least 1 slope in Range?
NO
***CHECK***
ELECTRODES
Perform
Wash Cycle
B
10 Calibration Sequence - 9180 (2)
B
Aspir. Std. A into
Meas. Chamber
Standard A detected?
YES
Aspir. Std. C into
Meas. Chamber
NO
**STANDARD A**
NOT DETECTED
Standard C detected?
YES
Obtain Electrode
Voltages Std. C
NO
**STANDARD C**
NOT DETECTED
Perform
Wash Cycle
Aspir. Std. A into
Meas. Chamber
Standard A detected?
YES
Obtain Electrode
Voltages Std. A
NO
**STANDARD A**
NOT DETECTED
Calculation of
Std.A-Std.C
(Use last Std. A)
C
C
At least 1 slope in Range?
YES
NO
Slope A-B and A-C ok for all Electrodes?
YES
All Std.A
volt. reprod.?
YES
NA K LI READY
NO
NO
***CHECK***
ELECTRODES
NA K LI READY
XX, XX NOT CAL'D
NA K LI READY
XX, XX NOT CAL'D
27
11 Power-up - 9180
Power-up
AVL9180 A80X1.AA
WARM UP
All dots lit for 1/2 sec.
Cold start?
NO
Power Fail
1-Nov-95 1:12
(Print)
YES
Set
Time/Date
Date: 1-JAN-95
Time: 12:00
YES
NO
Pack in place?
Global*
YES
*****PLEASE*****
*******WAIT*******
(Wash)
*Background check in all menus except "Take out of Operation" and "Service Functions"
Perform Daily
Cleaning?
YES
NO
Perform Daily
Conditioning?
YES
NO
Clean/Cond.
needed?
NO
AVL9180 A80X1.AA
WARM UP 59
YES
Clean...
Condition...
Calibration...
Pack in place?
Global*
NO
YES
YES
NO
Clean/Cond.
needed?
NO
Pack inserted?
Global*
****STATUS****
**NO SnapPak**
NO
YES
New SnapPak installed?
NO
*Background check in all menus except "Take out of Operation" and "Service Functions"
YES
FUNCTIONS?
NO
NA K CA READY
NO
Reset
100%
YES
Remain in Daily
Maintenance?
counter to
NO
1min
QC/STD/DIALYSATE/
URINE SAMPLE?
NO
NO
CALIBRATION?
NO
SERVICE
FUNCTIONS?
NO
PROGRAM
INSTRUMENT?
NO
Is Calibration due?
YES
Stabilizing
30 sec.
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
NO
28
3.3 Software Operation - 9181
1 Overall Program Operating Flow - 9181
29
2 Main Menu - 9181
MAIN MENU
NO
NA K CL READY
NO
AUTOMATED
MEASUREMENT?
NO
FUNCTIONS?
NO
QC/STD/DIALYSATE/
URINE SAMPLE?
NO
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
NO
PROGRAM
INSTRUMENT?
NO
SERVICE
FUNCTIONS?
NO
CALIBRATION?
30
LEGEND
Display prompt
Action performed by analyzer
2-second display prompt
Decision-making by analyzer
3 Manual Measurement Sequence - 9181
Na K Cl READY
YES
NO
AUTOMATED
MEASUREMENT?
Probe moves to upper position
Please Wait
Introduce Sample
YES/NO
1 min.
> air voltage
CHECK SS
VOLTAGE
< air voltage
Remove Sample and
Wipe Probe
Remove Sample and
Wipe Probe
Retract Probe ->
YES
Probe retracts automatically in 30 sec.
Retract Probe ->
YES
Probe moves to lower position
Probe moves to lower position
Thank You!
Thank You!
Serum Sample in Process 49
Na K Cl
151 5.1 114
Blood Sample in Process 49
Na K Cl
151 5.1 114
YES/NO
1 min.
31
4 Automated Measurement Sequence - 9181
Na K Cl READY
YES
Manual
Measurement
NO
AUTOMATED
MEASUREMENT?
YES
Probe moves to upper position
Probe moves to upper position
Tray homes
Tray moves to applicable cup position
Probe moves down to cup position
Probe moves to upper position
Turntable moves to mid position
NO SAMPLE!
No sample < 2
Probe moves to lower position
No sample = 2
CHECK SS
VOLTAGE
End of Run
Sample * in Process
Na K Cl
151 5.1 114
Home Tray
Na K Cl READY
* applicable sample number
32
5 Print Functions Menu - 9181
YES
FUNCTIONS?
YES
Print Last
Sample Report?
NO
Print Last
Automated Run?
NO
Cal. Report?
NO
Print QC Values and Statistics?
NO
Remain in Print
Functions?
NO
YES
YES
YES
YES
NO
QC/STD/DIALYSATE/
URINE SAMPLE?
PRINTING...
PRINTING...
PRINTING...
PRINTING...
Na K Cl READY
Na K Cl READY
33
6 QC/Standard/Dialysate/Urine Sample Menu - 9181
QC/STD/DIALYSATE
/URINE SAMPLE?
YES
NO
QC Level *
Sample?
YES
DAILY
MAINTENANCE?
Probe moves to upper position
2 min.
NO
2s
VALUES
REJECTED!
* same for level 1 - 3
Please
Wait!
NO
YES
2 s
VALUES
ACCEPTED!
Introduce Sample
2 min
Store
Values?
NA K CA READY
2 min.
Standard
Sample?
NO
YES
Automated
Measurement?
YES
NO
Probe moves to upper position
Please
Wait!
Na K Ca
99.6 5.34 1.250
Bicarbonate
Sample?
2 min.
NO
YES
Automated
Measurement?
YES
NO
Probe moves to upper position
Please
Wait!
Na K Ca
99.6 5.34 1.250
Acetate
Sample?
2 min.
NO
YES
Automated
Measurement?
YES
NO
Probe moves to upper position
Please
Wait!
Na K Ca
99.6 5.34 1.250
Urine
Sample?
2 min.
NO
YES
Automated
Measurement?
YES
NO
Probe moves to upper position
Please
Wait!
Na K Ca
99.6 5.34 1.250
YES
Remain in QC/
Std/Urine Sample?
NO/
2 min.
NA K CA READY
Remove Sample
Wipe Probe
Na K Ca
99.6 5.34 1.250
Introduce Sample
Standard Sample in Process 49
Introduce Sample
Bicarb. Sample in Process 49
Introduce Sample
Acetate Sample in Process 49
Introduce Sample
Urine Sample in Process 49
Retract Probe
-> YES
QC Level * in Process 49
Remove Sample
Wipe Probe
Thank
You!
Remove Sample
Wipe Probe
Thank
You!
Remove Sample
Wipe Probe
Thank
You!
Remove Sample
Wipe Probe
Thank
You!
Probe moves to lower position
Thank
You!
Retract Probe
-> YES
Probe moves to lower position
Retract Probe
-> YES
Probe moves to lower position
Retract Probe
-> YES
Probe moves to lower position
Retract Probe
-> YES
Probe moves to lower position
34
7 Daily Maintenance Menu - 9181
DAILY
MAINTENANCE?
YES
Perform Daily
Cleaning?
NO
NO
YES
OPERATOR
FUNCTIONS?
Probe moves to upper position
Perform Daily
Conditioning?
NO
YES
Probe moves to upper position
YES
Remain in Daily
Maintenance?
NO/1min.
CLEANING AND/OR
COND. PERFORMED?
YES
NO
Calibration due?
NO
YES
30 SECOND
STABILIZATION
Introduce Sample
Cleaning Solution not detected!
Introduce Sample
Condition. Solution not detected!
Cleaning Solution detected?
NO
YES
CLEAN
Condition. Solution detected?
YES
CONDITION
NO
CALIBRATE
NA K CA READY
35
8 Operator Functions Menu - 9181
*****STATUS*****
**NO SnapPak**
OPERATOR
FUNCTIONS?
YES
Check/Change
SnapPak?
NO
Go to
Standby Mode
NO
NO
YES
YES
PROGRAM
INSTRUMENT?
37% remaining
***STANDBY!***
YES -> READY
YES/NO
Change
SnapPak?
NO
YES
Pack detected?
GLOBAL*
YES
New SnapPak installed?
NO
NO
*Background check in all menus except "Take out of Operation?"
YES
Fluid Counter reset to 100%
YES
Maint. due?
NO
New Pack
Prime ...
NO
Set
Time/Date?
NO
Select Parameter
Configuration?
NO
Reset
Sample Number?
NO
Take out of
Operation?
NO
Remain in
Operator Func?
YES
NO
YES
NO
YES
YES
YES
Leave
Standby Mode?
YES
PRESS YES
WITHIN 2s
Date: 01-FEB-94
Time: 12:00 ok ?
NO
Date: 01-FEB-94
Time: 12:00 ok
YES
YES
YES
Sel. Parameter:
[Na] [K] [ ] ok?
Are you sure?
YES/
NO
YES
Shutdown Plug and
Water installed?
YES
Shutdown
Complete
Turn Power Off
CALIBRATE
YES
NO
IS
CALIBRATION
DUE?
YES
Daily
Cleaning?
YES
CLEAN
NA K READY
YES
Sel. Parameter:
[Na] [K] [Cl] ok?
SET SAMPLE
NUMBER TO 0
NO
Na K READY
NO
NO
NO
Daily
Conditioning?
YES
CONDITION
30 SECOND
STABILIZATION
CALIBRATE
YES
Sel. Parameter:
[Na] [K] [Ca] ok?
YES
Sel. Parameter:
([Na]) [ ] [Li] ok?
NO
NO
YES
Remain in Daily
Maintenance?
1min/NO
IS
CALIBRATION
YES
DUE?
NO
New
Parameter added?
YES
Electrodes in place?
YES
NO
YES
Sel. Parameter:
[Na] [K] [Li] ok?
NO
YES
NO
Sel. Parameter:
[Na] [ ] [Li] ok?
YES
Transp Housing
Ref installed?
Do You Wish to
DELETE all Data?
YES
YES
Remove Water and Press YES
Pump Windings
Relieved?
YES
YES
All Electrodes
& Plug Removed?
YES
5 Relief Clamps
Inserted?
36
9 Program Instrument Menu - 9181
PROGRAM
INSTRUMENT?
YES
Enter Code:
AAA
CORRECT
Program QC
Level 1 Ranges?
NO
NO
IN-
CORRECT
YES
NO
SERVICE
FUNCTIONS?
YES
CODE ERROR!
RETRY?
NO
Current Lot: 0000
Change Lot?
NO
Additional
Parameters?
YES
YES
Print old Values and Statistics?
NO
NO
New Lot!
Delete old Data?
YES
YES
K low = 1.5
K high = 15.0 ok?
YES
Na low = 040
Na high = 205 ok?
YES
Enter last 4 Dig of Lot#: 0000
YES
Enter last 4 Dig of Lot#: 0000 ok?
NO
Cl low = 050
Cl high = 200 ok?
YES
Ca low = 0.2
Ca high = 5.0 ok?
YES
Li low = 0.1
Li high = 6.0 ok?
YES
Program QC
Level 2 Ranges?
NO
Program QC
Level 3 Ranges?
NO
Program
Normal Ranges?
NO
Program
Corr Factors
NO
YES
YES
Program Bicarb.
Corr Factors
NO
YES
Program Acetate
Corr Factors
NO
YES
Program Urine
Corr Factors
NO
YES
Program Printer
Setup?
NO
YES
YES
NO
Program
Interface?
Remain in
Program Func?
YES
NO
"see Level 1"
"see Level 1"
YES
YES
Na low = 136
Na high = 145 ok?
YES
K low = 03.5
K high = 05.1 ok?
YES
Reset Corr
Factor (default)?
YES
Reset Bicarb.
Factor (default)?
YES
Reset Acetate
Factor (default)?
YES
Reset Urine
Factor (default)?
YES
NO
NO
NO
NO
Input/Verify
Corr Factors?
Input/Verify
Bicarb. Factors?
Input/Verify
Acetate Factors?
Input/Verify
Urine Factors?
NO
YES
NO
YES
NO
YES
NO
YES
YES
Remain in
Corr Factors?
NO
YES
Remain in
Bicarb Factors?
NO
YES
Remain in
Acetate Factors?
NO
YES
Remain in
Urine Factors?
NO
Patient Report off?
YES
NO
Print two
Patient Reports?
YES/
NO Automatic
Cal Report?
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
YES
YES
YES
K (b) = +00.0
K (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
YES
Na (b) = +00.0
Na (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
YES
YES/
NO Enter Comment
Line?
NO
YES
Enter Comment:
_
YES
Activate
Data Link?
YES/NO
NA K READY
37
10 Service Functions Menu - 9181
YES
SERVICE
FUNCTIONS?
YES
Test
Electrodes?
NO
CALIBRATION?
*same for Std. A, B, C
NO
YES
NO
Test
Standard A *?
YES
Std. A*: Na = 0000
K = 0000 Ca = 0000
NO
YES
NO
Test
External Sample?
NO
Remain in
Test Electrodes?
Test
Sample Sensor?
NO
Test
SnapPak Sensor?
NO
Test
Language Switch?
NO
Test
Probe?
NO
YES
DRY: 80-120: 102
YES=Pump/NO=Exit
YES
NO
External: 102
Pump: NO=OFF
YES
YES
YES
SnapPak: I
I=In O=Out
Language Switch
English
U = 0 L = 1
Sample Position?
YES/NO
YES/NO
Probe moves to extended position
Tray moves to mid position
Probe moves to upper position wash...
U = 0 L = 0
Upper Position?
U = 0 L = 0
Lower Position?
YES
Introduce Sample
NO
Ext.: Na = 0000
K = 0000 Ca = 0000
Probe moves to upper position
Probe moves to cup position
Retract Probe
-> YES
YES
U = 0 L = 0
Cup Position?
Tray moves to cup position
Probe moves to lower position
U = 0 L = 1
Exit Probe Test?
NO
YES
Probe moves to upper position
Tray homes
Probe moves to lower position
NO
Test
Sampler?
Test Pump?
NO
Test Valves?
NO
Test Interface?
Jump Pins 2&3
NO
NO
Test
Amplifier?
Enter
Service Code?
NO
YES
YES
YES
Test Pump!
Very slow speed
Test Valve A:
YES/NO=ON/OFF
-- Test --
-- Interface --
2s
NO
Test Pump!
Slow speed
Test Valve B:
YES/NO=ON/OFF
Interface ok!
YES
Na = -1030 K = -1200
Cl = -1093
YES
Enter Code:
AAA ok?
NO
YES
YES
Ground = 2
-2500mV = -2496
NO
NO
YES
Is Code valid?
NO
Code Error!
Retry?
2s
NO
Test Pump!
Medium speed
Test Valve C:
YES/NO=ON/OFF
YES
Remain in
Test Interface?
NO
Temp (mV) = -2670
Temp (°C) = 26.2
NO
2s
Test Pump!
Fast speed
2s
NO
NO
Test Valve V:
YES/NO=ON/OFF
NO
Test Valve R:
YES/NO=ON/OFF
Code
Accepted!
YES
Enter Other
Service Code?
NO
Remain in
Service Func?
NO
Calibration?
Calibrate...
NA K CA READY
38
11 Calibration Sequence - 9181 (1)
(Example Na/K/Li Calibration)
CALIBRATION?
YES
Calibration in
Process 124
Clear Sample
Path
Calibrate
Sample Sensor
Sample Sensor
Voltage ok?
YES
Prime
Standard B
NO
***CHECK***
SAMPLE SENSOR
Prime
Standard C
Prime
Standard A
NO
1st try
All 3 Standards not detected?
NO
Two Standards not detected?
NO
Standard A,B,C not detected?
NO
Pressurize Ref.
Asp. Std. A into
Meas. Chamber
YES
YES
YES
*********CLOG********
CHECK FLUID PATH
**STANDARD X/Y**
NOT DETECTED
**STANDARD X**
NOT DETECTED
Reference
Housing ok?
A
YES
NO
2nd try
***CHECK***
REF HOUSING
A
Asp. Std. A into
Meas.
Chamber
Standard A detected?
YES
Obtain Electrode
Voltages Std. A
NO
3 Std. A Values obtained?
YES
Are Values reproducible?
YES
Aspir. Std. B into
Meas. Chamber
NO
NO
Standard B detected?
YES
Obtain Electrode
Voltages Std. B
NO
**STANDARD A**
NOT DETECTED
**STANDARD B**
NOT DETECTED
YES
NO
6 Aspir. of
Std.A performed?
YES
At least 1
Channel reprod. ?
NO
***CHECK***
ELECTRODES
Calc. of Std.A-
Std.B (slope)
At least 1 slope in Range?
NO
***CHECK***
ELECTRODES
Perform
Wash Cycle
B
39
11 Calibration Sequence - 9181 (2)
B
Aspir. Std. A into
Meas. Chamber
Standard A detected?
YES
Aspir. Std. C into
Meas. Chamber
NO
**STANDARD A**
NOT DETECTED
Standard C detected?
YES
Obtain Electrode
Voltages Std. C
NO
**STANDARD C**
NOT DETECTED
Perform
Wash Cycle
Aspir. Std. A into
Meas. Chamber
Standard A detected?
YES
Obtain Electrode
Voltages Std. A
NO
**STANDARD A**
NOT DETECTED
Calculation of
Std.A-Std.C
(Use last Std. A)
C
C
At least 1 slope in Range?
YES
NO
Slope A-B and A-C ok for all Electrodes?
NO
YES
All Std.A
volt. reprod.?
YES
NA K LI READY
NO
***CHECK***
ELECTRODES
NA K LI READY
XX, XX NOT CAL'D
NA K LI READY
XX, XX NOT CAL'D
40
12 Power-up - 9181
Power-up
AVL9180
A80X1.AA
WARM UP
All dots lit for 1/2 sec.
Cold start?
NO
Power Fail
1-Nov-95 1:12
(Print)
YES
Set
Time/Date
Date: 1-JAN-95
Time: 12:00
YES
NO
Pack in place?
Global*
YES
*****PLEASE*****
*******WAIT*******
(Wash)
*Background check in all menus except "Take out of Operation" and "Service Functions"
Perform Daily
Cleaning?
YES
NO
Perform Daily
Conditioning?
YES
NO
Clean/Cond.
needed?
NO
AVL9180
A80X1.AA
WARM UP 59
YES
Clean...
Condition...
Calibration...
Pack in place?
Global*
NO
YES
YES
NO
Clean/Cond.
needed?
NO
Pack inserted?
Global*
****STATUS****
**NO SnapPak**
NO
YES
New SnapPak installed?
NO
*Background check in all menus except "Take out of Operation" and "Service Functions"
YES
FUNCTIONS?
NO
NA K CA READY
NO
Reset
100%
YES
Remain in Daily
Maintenance?
counter to
NO
1min
QC/STD/DAILYSATE
/URINE SAMPLE?
NO
NO
CALIBRATION?
NO
SERVICE
FUNCTIONS?
NO
PROGRAM
INSTRUMENT?
NO
Is Calibration due?
YES
Stabilizing
30 sec.
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
NO
41
42
3.4 Service Functions
The Service Functions mode provides a menu of service test functions to assist in failure diagnosis and identification. A logic flow chart is provided for reference of the sequences included in the Service Functions mode.
TEST ELECTRODES
This mode allows for the measurement and display of electrode voltages.
Standard A, Standard B or an external sample can be measured to determine electrode performance and provide measurement data useful in locating electrode calibration faults. Each electrode parameter will be displayed and the electrode voltage associated with the sample type selected can be observed.
Example Displays:
Std. A: Na=1582
K= 654 Cl=-904
Std. B: Na=237
K=-988 Cl=647
Ext: Na=1321
K=-598 Cl= -127
TEST SAMPLE SENSOR
In this mode, the sample sensor output is displayed which is useful for electrical adjustment or for troubleshooting sample sensor faults. A dry sample sensor should display a value of 100 with a range of 80 to 120. After verifying a dry sensor, aspiration of a clear fluid (i.e., water) should increase the displayed value by at least
40 units.
Example Display:
DRY 80-120: 102
YES=Pump/No=Exit
External: 153
Pump: NO=OFF
TEST SnapPak SENSOR
(9180 and 9181 only)
In this test, the SnapPak Sensor detects whether the ISE SnapPak™ is in place. "I" indicates that the ISE SnapPak™ is in place, "O" indicates that it is either not present or not completely seated.
Example Display:
SnapPak: I
I = IN O = OUT
TEST LANGUAGE SWITCH
(9180 and 9181 only)
This mode displays the language selected for display and printout. To change the selection, set the language switch (located behind printer) to the desired language. The display will indicate the language choice according to the switch setting. To activate a new language, turn the analyzer off and back on with the switch set to the desired language.
Example Display:
Lanugage Switch
English
TEST SAMPLE DOOR
(except 9181)
This test verifies the operation of the sample door light gate and its associated circuitry. The display indicates an “O” or “C” depending on the position of the door.
Example Display:
Sample Door: C
C:Closed/O:Open
43
44
TEST PROBE
(9181 only)
This test verifies the operation of the sample probe motor, position sensor and associated circuitry. The "U" represents the output of the upper position sensor, the
"L" represents the output of the lower position sensor.
Example Display:
U = 0 L = 1
Sample Position?
TEST AUTOSAMPLER
(9181 only)
This test verifies the operation of the Autosampler turntable assembly. The display indicates the output of both the position and home sensors.
Note: If the Autosampler is not plugged in or the
TABLE PRESENT
bit is not activated, this test will not appear under the
SERVICE FUNCTIONS
menu.
Example Display:
Home: 0
Pos = 1
TEST PUMP
This test turns on the peristaltic pump at each of the four pump speeds. The display indicates for two seconds each pump speed while each pump speed is being tested.
Observe that the peristaltic pump changes speed as is indicated on the display.
*** Pump Test *** *** Pump Test ***
Very Slow Speed Slow Speed
*** Pump Test *** *** Pump Test ***
Medium Speed Fast Speed
TEST PINCH VALVES
This test turns on/off each solenoid valve as indicated and is useful to verify both electronic control and mechanical operation of each solenoid valve.
TEST INTERFACE
The Interface Test can be used to verify correct operation of the RS232C serial port.
To operate this test, jumper pin 2 and pin 3 of the rear interface connector prior to invoking the test. During the interface test, the analyzer sends out a test string and checks if it is received within a set period of time. If correct operation is observed after completion of the test, INTERFACE OK will be displayed.
TEST AMPLIFIER
The Amplifier Test displays ground, reference and output voltage of each input amplifier. The values displayed for each electrode channel will be dependent on the electrode inputs and may vary. Display of the ground should read between -10mV and +10mV with the reference voltage reading between -2490mV and -2510mV.
With all inputs open, the display should read +/- 4095 for each channel. On the 9140/
9180/9181, an additional display indicating the temperature voltage and temperature in degrees C is available to verify the temperature adjustment.
ENTER SERVICE CODE
Service Codes are available and allow additional programming features. Following is a listing of the available Service Codes and functions:
Set Reset Service Code Description
ALL LLA
(9181 only)
This service code will enable the Ca ++ value during automatic serum/blood measurement. Special precautions are necessary to obtain the correct Ca ++ value due to sample exposure to room air.
CDC
Clears all Service Codes.
45
46
DEC CED
ECO OCE
FIF IFI
LEM MEL
(9180/9181 only)
MGL LGM
NOB BON
(9180 /9181only)
QCC CCQ
Provides one additional digit of resolution for each measured parameter for blood and serum samples.
QC and Standard samples are always displayed with high resolution. Urine samples are always displayed in low resolution.
Allows the operator to run the instrument in the economy mode. The instrument will enter the
STANDBY MODE. This occurs at the time of the next 2-point calibration when no samples have been analyzed since the last 2-point calibration.
Allows printing of electrode voltages during printout of measurements and calibrations.
(not available for 9140)
This service code will print the last 20 error messages with the date and time of occurrence.
This service code is available only in the model 9140/
9180/9181 analyzer and enables Ca
++
values to be displayed in mg/dL. The normal mode displays values in mmol/L. QC and Standard results are always displayed in mmol/L.
This service code disables the beep.
SFC
(9180/9181 only)
SLC CLS
(9180/9181 only)
SSC CSS
(9181 only)
When this code is enabled, QC values will be reported in direct ISE measurement instead of flame-equivalent values. In this mode, correlation factors are NOT applied to the measured values.
This code allows the input of a percentage for the ISE
SnapPak TM counter.
When this code is set, the current setting of the leasing counter, which cannot be reset by the operator, is printed on the calibration report.
This code toggles needle height to accommodate smaller sample cups. In the default setting, the analyzer works with 2.0 mL cups. To select 0.25 mL sample cups,
SSC has to be activated. To return to the default setting, CSS has to be activated.
MGL
NOB
QCC
SFC
SLC
SSC
CODE
ALL
CDC
DEC
ECO
FIF
LEM
FUNCTION
Enables Ca
++
during autom.
blood/serum measurement
9120 9130 9140 9180 9181
X
Clears all Service Codes X X X X X
Additional display resolution X X X X X
Economy Mode X X X X X
Cal voltages on display
Last 20 error messages printed
Display mg/dL
No beep
QC values direct
ISE measurement
X X X X
X X
X X
X
X
X
X X X X X
SnapPak Counter input
Set "Leasing Counter"
Small Sample Cups
X X
X X
X
ANALYZER RESET
WARM-UP BYPASS
Press YES and NO keys simultaneously and turn power off and back on to provide a total reset of the instrument software.
CAUTION! All QC Data and Instrument Programming will be lost.
Press YES and NO keys simultaneously during instrument warm-up.
47
48
4 Description of Modules
4.1
Mechanical Assemblies
4.1.1 Front Door Assembly
For the 9120, 30, 40 and 80 analyzers, the front door can be removed by moving the analyzer to the edge of a work surface so the analyzer door, when opened, will extend past the edge of the work surface. With one hand, hold the analyzer door near the right side hinge pin and, with the other hand, gently apply pressure to the middle rear area of the door. This will allow the right hinge pin to clear the retaining hole in the main chassis. The door can then be removed from the analyzer.
The 9181 front panel can be removed by tilting it slightly away from the analyzer and lifting it straight up. The plastic window can also be replaced in the field by gently pressing from the rear of the window and snapping the plastic window out toward the front of the analyzer door. The door magnets cannot be replaced in the field as they are glued in place with conductive adhesive.
4.1.2 Needle Unit Assembly
Both sample needle and fill port are designed for easy replacement by the user.
To remove the fill port assembly for replacement, pull the fill port holder towards the front for easy access. Then, press the two plastic tabs on the fill port to allow removal of the fill port from the fill port holder. The sample needle can also be removed and is supplied as a one-piece assembly including the intake tubing.
To remove the sample needle, grasp the needle near the white holder block and pull up to unsnap the sample needle from its position. To replace the assembly, align a replacement sample needle and press the replacement assembly into place.
Three screws located on the left side of the assembly hold the needle unit assembly in place. To access the two rear screws, turn power off and remove the rear panel.
The rear panel can be moved away from the unit for easier access by unplugging the wire harness from the power supply. The SBC board can now be removed by removing the screws which secure the SBC board and disconnecting the cables on the SBC board. Disconnect the connector to the lamp board which supplies the
Door Open signal. The two rear screws which secure the needle unit can now be removed using an Allen wrench. Then, remove the front securing screw, the sample intake tubing and the reagent supply tubing (red tag) from the fill port. The needle unit can now be removed from the front of the analyzer for replacement. Reference the Illustrated Parts diagram in Section 8 for a detailed parts diagram of the needle unit.
4.1.3 Electrode Holder Assembly
To remove the electrode holder assembly from the analyzer, first unplug the sample intake tubing on the right side of the module, then the sample sensor cable, then remove the reference solution tubing connector and last, take out the waste tube (green tag) at the left side of the electrode holder. The electrode holder can now be removed by pressing the two plastic tabs (right and left side of the module) and sliding the module forward. Press the plastic tabs again to release the second detent and completely remove the module from the analyzer.
The sample sensor and/or the left side electrode holder can be removed and replaced by removing the screws on the under side of this module. The Illustrated Parts List in Section 8 provides a detailed reference of the parts of this assembly.
4.1.4 Peristaltic Pump Assembly
Replacement of the roller assembly of the peristaltic pump can be accomplished by removing the pump tubing and firmly pulling the roller assembly toward the front. To replace the roller assembly, align the roller and the flat on the motor shaft and press the replacement roller into position.
To remove the motor, first remove the rear panel assembly and SBC Board as described in Sections 4.1.7 and 4.1.8. Now, the electrical connector on the lamp board can be accessed to unplug the motor connector. The motor can now be removed by removing four screws located on the front of the housing near the peristaltic pump.
4.1.5 Solenoid Assembly
Removal of each solenoid is identical. Each solenoid valve has a removable pressure piece which is held in place by a solenoid cap. To remove the pressure piece, locate the arrow on the solenoid cap and remove the solenoid cap by sliding it off the solenoid shaft in the direction of the arrow. This exposes the pressure piece which can now be removed from the solenoid shaft. To replace pressure piece and cap, energize the solenoid as described in Section 3.2 (Test
Valves). This extends the solenoid shaft to the outermost position for ease in replacing pressure piece and cap. Removal of the pressure piece and cap is required prior to removing the solenoid and allows for easier replacement of the tubing under each solenoid.
49
50
To remove the solenoid assembly, the rear panel and the SBC Board described in
Section 4.1.8 must be removed first. Each solenoid has an electrical connection to the lamp board which must be unplugged prior to removal of the solenoid. At this point, remove the two screws on the front panel to remove the solenoid assembly.
4.1.6 Printer Assembly
The printer assembly is designed to allow for easy removal by the user which can be accomplished without removal of the electrical power to the analyzer. Slide the paper tray forward to allow access to the printer, tear the paper roll and completely remove it together with the paper tray. Slide fingers under the printer assembly and pull the printer forward. This will disengage the printer from the interconnector and enable removal of the assembly from the front of the analyzer. Removal of the printer should be performed for replacement and for removal of a paper jam.
To replace the printer, locate the printer slide and insert the printer assembly. Press firmly into place to ensure electrical connection of the printer.
Note: Never attempt to dislodge paper from the printer with a paper clip or similar object to avoid damage to the print head or printer platen.
4.1.7 Rear Panel Assembly
To remove the rear panel assembly, ensure that the power cord has been disconnected. Remove the four corner screws to expose the rear panel. The power supply module located on the rear panel assembly can now be removed by removing the three screws securing the circuit board and by disconnecting the wiring interconnections. The main power receptacle can also be replaced by removing the two screws holding this assembly.
4.1.8 SBC Board
After the rear panel (Section 4.1.7) has been removed, the SBC Board is accessible for removal. This circuit board can be removed by removing the seven screws securing the board to the housing and by disconnecting electrical connectors from the power supply module, display board and lamp board.
The electrode push pins can now be replaced by simply pulling the connecting pin from its socket. When installing the board, make sure to place the plastic washers under the screw heads.
4.1.9 Lamp Board
The rear panel and SBC board must be removed to gain access to the lamp board. Remove one screw and washer which hold the board in place and remove the electrical connectors to remove this board.
4.1.10 Display Board
The rear panel and SBC board must be removed to gain access to the display board. Remove three screws and 50 degree spacers and unplug the sample sensor cable from the front and the ribbon cable to the SBC board to remove the display board.
51
52
4.2 Fluidic Module
4.2.1 Standard A
Standard A is drawn to the electrode module by vacuum provided by the peristaltic pump. When Standard A is to be aspirated into the electrode module, solenoid valve A is opened and solenoid valves B, V and R are closed. Standard A is then drawn from the ISE SnapPak TM , to the fill port, through the sample needle and is sensed by the sample sensor. At this point, solenoid valve V is opened and solenoid valve A is closed as the peristaltic pump continues to pump Standard A into the electrode module. As the trailing edge of the Standard A sample is sensed by the sample sensor, the peristaltic pump stops. During the time Standard A is aspirated into the electrode module, the reference housing is pressurized due to the peristaltic pump providing pressure to the reference solution line and solenoid valve
R being closed. An amount of reference solution is forced out through the reference junction to provide contact to the sample.
4.2.2 Standard B/C
The process for Standard B/C is identical to that of Standard A with the exception of the operation of solenoid valves B and C. These solenoid valves are operated in reverse order of Standard A aspiration.
4.2.3 Reference Solution
The reference housing is filled automatically using the second winding of the peristaltic pump and solenoid valve R. The reference solution connector allows for the reference housing tubing to be plugged into the reference solution circuit. As the peristaltic pump aspirates reference solution, solenoid valve R is opened to allow excess solution to be pumped into the reference return line of the
ISE SnapPak TM
.
4.2.4 Waste Liquids
Calibration standard and sample waste are pumped out of the left side of the electrode module (green-banded tubing) through the peristaltic pump to the waste line of the
ISE SnapPak TM
. An internal one-way valve is incorporated into the waste line of each
ISE SnapPak TM
to prevent any waste products from leaking out of the
ISE SnapPak TM
.
4.2.5 Main Tube Set
The main tube set is supplied as a pre-cut tubing harness with the fluidic interconnections preassembled. Replacement of the tube set should be performed annually. For removal of the tubing, first remove the ISE SnapPak
TM
from the reagent compartment. This will allow access for removal of the white TPR block which is the main interconnection between the ISE SnapPak TM and the tubing. This block can be removed by grasping the front surface and pulling forward. Next, remove tubing from each solenoid valve. Follow the procedure outlined in Section 4.1.5 for disassembly of the solenoid valves to enable fluidic tubing to be easily removed at each solenoid. Pull the two pump windings at the peristaltic pump tube set off the pump roller and remove the white reference solution tube connector. Remove the white TPR reference solution block by grasping the front of the block and pulling forward. The three tube connections are labeled with green, blue and red tags. All remaining tubes can now be removed and the entire tubing harness can be discarded.
Fit the replacement tubing harness and assemble in reverse order as disassembly. Reference the Tube Diagram on the next page for correct tube locations.
53
54
55
56
4.3 Electronics
4.3.1 ISE SBC Card
Power Circuits
Power is supplied to this circuit board through connector JP3 from the power supply assembly. Test points are provided to measure supply voltages, and light emitting diodes D1 through D5 are turned on to indicate each supply voltage present on the card. Since each LED is connected in series after the fuse, a blown fuse results in the respective LED turned off.
Four additional voltage regulator circuits are mounted on the ISE SBC II Card. IC
U2 and IC U3 provide -8 VDC and +8 VDC (9180/9181: -5 VDC and +5 VDC) and are used to provide the supply voltages for the analog circuitry. IC U1 develops the supply voltage used to operate the peristaltic pump and provide either +10
VDC or +22 VDC. The +22 VDC is supplied only for the FAST speed and +10
VDC is used for all other pump speeds. Signal FAST/SLOW determines which supply voltage is selected.
U5 switches the voltage between 22VDC and 12.5 VDC which supplies the solenoids. When a solenoid is turned on, the voltage switches to 22VDC for 0.5
sec, then it returns to 12.5 VDC.
The circuit which includes Q7 and R5 provides necessary switching for the measuring chamber illuminator LED’s.
The following list identifies the test points, fuses and LED’s for each respective DC supply voltage:
Test Point Voltage Fuse Rating LED
TP1
TP2
TP3
TP4
TP5
TP6
TP7
TP8
TP9
TP10
TP11
TP12
TP13
TP14
TP15
TP16
+24 VDC
+5 VDC
-12 VDC
+12 VDC
Ground
F3
F1/F2
F4
F5
-8 VDC (9180/9181: -5 VDC)
+22/12 VDC
+8 VDC (9180/9181: +5 VDC)
Na + Channel
K + Channel
Ca ++ /Cl Channel (Li + 9180/9181 only)
Temperature
+10.000 V Ref. Voltage
-2.500 VRef. Voltage
Sample Sensor
+22/12.5VDC
D3
D1/D2
D4
D5
Valve Drivers
The solenoids are activated by transistors Q9 - Q13. To open the solenoid, a
0.5-second pulse of 22 volts is applied, then the voltage drops down to 12.5
volts to hold the solenoid open.
Pump Motor Driver/Cover Sensor
The pump driver circuit IC U9 and IC U10 provide the drive and control for the peristaltic pump. Signals PUMPEN, Pump Enable and PUMPSTEP, Pump
Step Pulses enable IC U10 to provide the necessary control to driver IC U9 which supplies pump drive signals PHASE A through PHASE D to control peristaltic pump operation.
IC U11 operates with signal DCLOSED from the sample door light gate to provide signal DOOR which detects the status of the sample door (open or closed).
Input Amplifiers
IC U13, U14 and U15 (9180/9181: U13 only) provide high to low input impedance matching for each Electrode input. Offset generator IC U12D and amplifiers IC U12A, U12B and U12C provide the necessary gain and offset for each electrode channel. TP9, TP10 and TP11 provide test points for the Na
+
,
K + and Cl /Ca ++ /Li channels respectively.
Temperature Sensor
This is an optional circuit which is used in the 9140, 9180 and 9181.
The circuit creates a temperature-dependent voltage which is converted by software to degrees C. R54 can be adjusted for correct temperature setting.
Analog Channel Selector and A/D Converter
Analog inputs from the input amplifiers are selected via signals ADSEL0 through
ADSEL2 and are multiplexed by IC U19. This multiplexed signal is then fed to the A/D Converter IC U20. IC U21 supplies the necessary reference voltage
(2500mV) for the A/D Converter and can be measured at TP14.
57
58
Microprocessor
Microprocessor IC U31, EPROM IC U25 (9180/9181: U40) and MOSTEK Real
Time Clock/RAM IC U22 form the components of the central processing system.
IC U23 Address Demultiplexer, IC U24 Programmable Array Logic (PAL) and IC
U28 provide serial data buses (9180/9181: U23, U24 and U28 not present). IC
U34 through IC U38 convert serial data to parallel data and provide printer data, display data, solenoid valve control and peristaltic pump control. This fashion of serial to parallel data conversion is used to prevent analyzer lock-up by peripheral devices. The RS232 serial interface is a direct input/output from the microprocessor utilizing interface driver and receiver U29 and U30.
SnapPak Sensor (9180/9181 only)
Optical detector ISO1 provides a signal indicating if the ISE SnapPak™ is in place.
This signal is fed to U41 which converts the light gate input to a digital logic output indicating SnapPak position (IN or OUT).
4.3.2 Lamp Board
The lamp board contains the LED array used to illuminate the measuring chamber and the interconnectors used for all solenoids, the peristaltic pump, door sensor and sample turntable (9181 only). A ribbon cable from the SBC Board supplies the electrical connection for each signal through connector JP6.
4.3.3 Display Board
The display board contains the LCD display, sample sensor connector and the
YES and NO switches used for operation of the analyzer. Power and signal interconnection is accomplished via a ribbon cable from the SBC board through
JP2 on the display board.
Display boards utilized in the 9181 analyzer also contain circuitry for both the needle mechanism motor control and position sensor. The autosampler "Table
Present" signal is also buffered through U1 on this board.
5 Maintenance and Troubleshooting
Maintenance procedures for the 9120, 30, 40, 80, 81 Analyzers require minimal time by the operator to perform. The procedures outlined below should be performed by the schedule indicated. Detailed instructions for these procedures can be found in the Operator's Manual describing the correct method to perform routine analyzer maintenance.
5.1 Maintenance
Daily Maintenance
- Perform cleaning cycle.
- Perform conditioning cycle.
Weekly Maintenance
- Clean sample probe and fill port.
- Clean analyzer surfaces.
Monthly Maintenance
- Clean reference electrode housing.
Six Month Maintenance
- Replace peristaltic pump tubing.
Annual Preventive Maintenance
- Replace complete tubing set.
- Replace peristaltic pump tubing.
- Replace fill port assembly.
- Replace sample probe.
- Replace sample sensor Quad-ring.
- Check sample sensor voltage.
- Check sample temperature (9140/80/81 only).
59
60
5.2 Error Messages and Troubleshooting
STATUS: NOT CALIBRATED
This message is only displayed when a calibration has been interrupted by pressing NO or opening the sample door. Perform a calibration to return to the
READY Mode.
•
•
•
•
STANDARD A NOT DETECTED
This message is displayed when the sample sensor is unable to properly detect the Standard A solution in a programmed time period. As the analyzer draws a sample of Standard A, the leading edge of the solution is detected by the sample sensor. The sample sensor must also detect a continuous flow of
Standard A Solution into the sample chamber, lasting several seconds, without encountering air bubbles or air. After several seconds, the trailing edge of the
Standard A sample is sensed, the peristaltic pump stops and measurement occurs.
•
Possible causes and remedies:
Check for air leaks preventing the Standard from being drawn into the chamber. Ensure that all O-rings are in place. There must be a continuous draw of standard solution, free of air bubbles.
Check solenoid V for proper sealing of the vent line.
Check fill port to ensure the absence of air leaks.
Check for clots or crystals formed in the Standard A tubing or electrode chamber.
Ensure that the sample sensor is plugged in securely and perform
Sample Sensor Test to ensure sensor is operating correctly. Clean, adjust or replace sample sensor to correct sample sensor operation.
Replace peristaltic pump tube set to ensure proper sample aspiration.
Check the fluid remaining in the ISE SnapPak TM . If less than 5% remains, replace the
ISE SnapPak TM
.
STANDARD B NOT DETECTED
STANDARD C NOT DETECTED
(9180/9181 only)
The system for drawing Standard B or C is like that of Standard A. If Standard A is drawn properly and Standard B or C is not detected, check Standard B tubing for crystallization. Perform checks for Standard A detection problems to correct.
During the calibration sequence, Standard B or C is drawn into the measuring chamber prior to Standard A, therefore check if Standard A calibrant can be properly detected to isolate possible fluidic faults. If neither Standard A, B nor C are detected check for air leaks or blockages within the fluidic path.
•
•
•
•
•
CHECK SAMPLE SENSOR
The sample sensor must provide a reading of 80 - 120 when air is detected.
Perform Test Sample Sensor to observe sample sensor response. When clear fluid
(e.g. water) is aspirated, the sample sensor display must indicate an increase of of at least 40 units. For blood samples (not transparent), the reading should decrease by at least 40 units. The sample sensor is calibrated with air during each calibration.
Perform a calibration to remove the error message after troubleshooting.
Possible causes and remedies:
Clean sample sensor using Daily Cleaning procedure.
Check for correct drying of measuring chamber during the wash cycle.
Replace peristaltic pump tube set.
Test and adjust sample sensor to correct adjustment.
Replace sample sensor assembly.
61
62
•
•
•
•
CHECK REFERENCE HOUSING
At the beginning of a calibration, pressure is formed in the reference housing while the R valve is closed and the V valve is open. After Standard A is detected in the sample chamber, the back pressure formed in the reference housing causes the Standard A solution to move back toward the sample sensor. When the standard A solution is detected by the sample sensor, the R valve is opened to release the pressure. If the sample sensor does not detect standard A as it moves back, CHECK REFERENCE HOUSING will be displayed.
Possible causes and remedies:
Check for clogged reference housing and clean per monthly maintenance procedures.
Check and ensure that reference tubing is securely connected at the tubing connector.
Check for proper filling of the reference housing and ensure that reference solution reagent level allows complete filling of the reference housing.
Check for bubble-free aspiration of Standard A.
•
•
•
PLEASE CLOSE SAMPLE DOOR
(except 9181)
This message occurs and prints NO SAMPLE on the print-out, when either of two conditions exists.
1.
Sample door is not closed within 20 seconds after samples are in place.
2.
Sample door has been opened and no sample is detected within 20 seconds.
Possible causes and remedies:
Close sample door within the time allowed.
Check needle mechanism light gate.
Check whether sample sensor is plugged in.
NO SAMPLE
This message is displayed when any of the following conditions occur:
1.
During sampling, either a bubble in the sample is detected or thesample size is smaller than the minimum sample required.
2.
3.
The sample door is opened and no sample is fed into the analyzer.
Sample door is not closed within 20 seconds after sample is detected.
•
•
Possible causes and remedies:
Check whether the sample sensor is plugged in and perform
TEST SAMPLE SENSOR to verify that it operates correctly.
Check sample draw and look for clots or leaks in the sampling system.
•
•
CLEANING FLUID NOT DETECTED
When the cleaning cycle has been initiated, the analyzer checks for proper aspiration of the cleaning solution using the sample sensor. If solution is not detected in the proper time, above error message will be displayed.
•
•
Possible causes and remedies:
Ensure that a sufficient volume of cleaning solution is presented for aspiration and no air bubbles are present as solution is aspirated.
Check to ensure that the solution is being aspirated through the sampling mechanism correctly. The sample path must be free of leaks or clots.
Verify that sample sensor is properly plugged in and perform the
SAMPLE SENSOR TEST to verify the correct operation of the sample sensor.
Check pump tubing and replace as necessary to ensure a correct pump rate.
63
64
•
•
•
CONDITIONING FLUID NOT DETECTED
When the conditioning cycle has been initiated, the analyzer checks for proper aspiration of the conditioning solution using the sample sensor. If the solution is not detected in the proper time, this error message will be displayed.
•
Possible causes and remedies:
Ensure that a sufficient volume of conditioning solution is presented for aspiration and no air bubbles are present as solution is aspirated.
Check to ensure that the solution is being properly aspirated through the sampling mechanism. The sample path must be free of leaks or clots.
Verify that sample sensor is correctly plugged in and perform the
SAMPLE SENSOR TEST to verify its proper operation.
Check pump tubing and replace as necessary to ensure correct pump rate.
•
INTERFACE ERROR
This message is displayed only during the INTERFACE TEST. Pins 2 and 3 must be shorted together to perform the INTERFACE TEST. Make sure that the pins are not shorted to chassis ground.
Possible cause and remedy:
Replace SBC Card.
PAPER JAM OR PRINTER DEFECT
This message is displayed when the printer head is jammed and tries to print. It is a temporary message that is displayed for 2 seconds and then sample results are displayed. To clear paper jam remove printer from the analyzer and remove jammed paper. Replace printer and retry. The printer can be removed by grasping the rear edge of the printer assembly and pulling it to the front. The printer can be removed with the Analyzer power on. To free up jammed paper, turn the small gears on the left side of the printer module.
•
•
CHECK ELECTRODES
This message is displayed when either of the following conditions are present.
1.
More than six aspirations of Standard A are required for all electrode channels.
2.
The difference between voltage A and B is out of specification for all channels.
Possible causes and remedies:
Electrodes are not plugged in. Check the slide mechanism to ensure it is in the correct position and firmly pushed into electrode connectors.
Perform Daily Maintenance.
•
•
•
• Check or replace reference electrode.
• Replace ISE SnapPak TM .
NA NOT CAL’D
K NOT CAL’D
CL NOT CAL’D
CA NOT CAL’D
LI NOT CAL’D
•
•
These messages will be displayed when a specific electrode does not calibrate properly during a calibration sequence. If more than six standard draws are required or the difference between Standard A and Standard B voltages is out of range, the NOT CALIBRATED message will be displayed for each channel which exhibits either of these conditions. A copy of the calibration report should be printed to provide the electrode output voltages measured during calibration.
The electrode voltage range table identifies correct voltage ranges for each electrode parameter.
If the difference between Standard A and Standard C (9180/9181 only) is out of range for Li
+ or Na
+
, LI NOT CAL’D will be displayed.
Possible causes and remedies:
Check to ensure calibration reagents are transported correctly. Test for leaks, blockages, bubbles, or improper pump flow rate.
Check or replace
ISE SnapPak TM
.
Check whether electrodes are plugged in properly. Check the slide mechanism to ensure it is in the correct position and firmly pushed into electrode connectors. Clean and/or replace electrode.
If A-C is out of range for Na
+
, repeat the calibration or replace the
ISE SnapPak TM
.
Check or replace SBC Card.
65
66
ELECTRODE VOLTAGE RANGES 9120, 30, 40
Electrode Standard A Standard B Allowable Difference A - B
Na
K +
Cl
-
+
Ca
++
(9120/9130) (9140)
-600 to +2400 -1600 to +2200 +720 to +2000 +250 to +680
-700 to +1000 -2500 to +500 +520 to +1800 +470 to +1200
-3100 to -100 -1100 to +1800 -520 to -1200 not used
-2600 to +1400 -2100 to +2700 not used -480 to -900
ELECTRODE VOLTAGE RANGES 9180, 81
Electrode Standard A Standard B
Na
K
+
Cl
-
Ca
Li
+
+
++
Standard C Allowable Difference
A-B A-C
-600 to +2400 -1600 to +2000 -600 to +2400 +250 to +680
-700 to +1000 -2500 to +500 -700 to +1000 +470 to +1200
-3100 to -100 -1000 to +3000 -3100 to -100 -370 to -860
-50 to +50
-40 to +40 not used
-3100 to +1000 -2300 to +2500 -3100 to +1000 -350 to -660 -150 to +150
-3100 to +1900 -3600 to +1400 -2600 to +3400 +1 to +760 -1730 to -285
CLOG CHECK FLUID PATH
After completion of sample measurement or calibration, the analyzer will monitor the sample sensor to determine if the sample path has been cleared. This message will be displayed if the sample path has not been cleared.
•
•
•
Possible causes and remedies:
Check for blockages in the sample path, especially in the sample probe, the tubing to the sample sensor and in the sample sensor.
Check sample sensor is securely plugged in and check sample sensor test to ensure correct operation.
Check peristaltic pump and tubing are correct and all tubing is properly in place.
•
•
↑↑↑↑ ↓↓↓↓
In case the analyzer displays arrows up or arrows down instead of sample results, the concentration of the sample is outside of the measurement range.
•
•
Possible causes and remedies:
If the sample is a urine sample, arrows up instead of the K result indicate that further dilution of the specimen is required.
Check for proper sample preparation.
Ensure that the sample is correctly aspirated into the measuring chamber and ensure small air bubbles are not present.
Check for proper aspiration of Standard A.
↑ ↓
On the 9140, 9180 and 9181 models, the calibration report will print an arrow up or down instead of the actual temperature, if the temperature measured is out of range (10.0°C -40.0°C). The temperature sensor is located in the right side electrode holder.
•
•
•
•
Possible causes and remedies:
Ensure that sample sensor cable is securely plugged in.
Check whether room temperature is within specified limits
(15°C to 32°C / 60°F to 90°F).
Perform temperature adjustment procedure.
Replace SBC board or right side electrode holder.
•
•
ERR
The analyzer will display ERR in place of sample results when the analyzer is unable to obtain valid voltage readings from the electrode (A/D over- or underflow).
•
Possible causes and remedies:
Ensure that the electrodes are securely in place and plugged into the analyzer.
Check to ensure proper filling of the reference electrode has occurred.
Ensure proper sampling and proper sample preparation. Check for air bubbles in the sample.
PERFORM DAILY MAINTENANCE
This message will be printed at the end of a sample report when cleaning or conditioning have not been performed within the last 24 hours. This message is printed only. Perform Daily Maintenance to clear this message.
67
68
REPLACE FLUID PACK
When the monitored fluid level in the
ISE SnapPak TM
reaches 5% remaining, the analyzer will print REPLACE FLUID PACK at the end of each sample report. This message is printed only. Replace the
ISE SnapPak TM
following the instructions outlined in the Operator Manual.
To maximize the life of the
ISE SnapPak TM
, continue running the analyzer until a STANDARD A NOT DETECTED message is displayed. This will ensure that the ISE SnapPak TM is completely depleted before replacement.
•
•
•
CHECK TEMPERATURE
(9140/9180/9181 only)
On the 9140/80/81 models, the analyzer will display CHECK
TEMPERATURE at the end of each measurement, if the temperature measured is out of range (10.0°C to 40°C). The temperature sensor is located in the right side electrode holder.
•
Possible causes and remedies:
Ensure that the sample sensor cable is securely plugged in.
Check whether room temperature is within specified limits
(15°C to 32°C / 60°F to 90°F).
Perform temperature adjustment procedure.
Replace SBC board or right side electrode holder.
VALVE OVERTEMP! CHECK VALVES
In case one of the valve solenoids overheats, the message VALVE
OVERTEMP! CHECK VALVES will be displayed. Each solenoid incorporates a thermo-fuse, which will reset automatically.
•
•
•
Possible causes and remedies:
Turn power off for at least one hour.
Check fuse F3 (24V).
Check whether room temperature is within specified limits
(15°C to 32°C / 60°F to 90°F).
ERROR: UPPER NEEDLE SENSOR
If probe movement to the upper position is not detected within a certain time period, the above message is displayed.
Possible remedies:
• Check for foreign material that may prevent the probe from moving freely.
• Contact AVL for technical assistance.
ERROR: LOWER NEEDLE SENSOR
If probe movement to the lower position is not detected within a certain time period, the above message is displayed.
Possible remedies:
• Make sure the fill port is installed correctly.
• Check for foreign material that may prevent the probe from moving freely.
• Contact AVL for technical assistance.
WHEEL MISSING OR SAMPLER DEFECT
During the homing sequence, the sampler checks for the sample wheel to be present.
Possible remedies:
• Make sure the wheel is present and seated correctly.
• Replace sample wheel.
• Replace sampler.
69
70
SAMPLER JAMMED OR DEFECTIVE
The analyzer checks for correct movement of the sample wheel. If the sample wheel does not move at the expected speed, the above error message is displayed.
Possible remedies:
• Check for obstructions preventing the sample wheel from moving freely.
• Replace sampler.
6 Adjustments
The 9120, 30, 40, 80 and 81 Analyzers have been designed to require minimal electronic adjustment. The sample sensor circuit may require adjustment following a change of the sample sensor or during routine troubleshooting. The temperature adjustment is active only for models 9140/9180/9181 and requires the temperature adjustment test plug for correct calibration of the temperature circuit.
6.1 Sample Sensor Adjustment
Select TEST SAMPLE SENSOR in the SERVICE FUNCTIONS menu which will display the current sample sensor setting. Adjust R93 Sample Sensor
Adjust located at the top of the ISE SBC card to a value of 100. Press YES to turn on peristaltic pump and ensure displayed value remains constant.
Raise sample door and aspirate clear fluid (e.g. water). Displayed reading should now be 150 or greater.
Note: R93 can be accessed through the rear cover vent slot and is located at the top center on the circuit card.
6.2 Temperature Adjustment (9140/9180/9181 only)
Select AMPLIFIER TEST in the SERVICE FUNCTIONS menu which will display the current amplifier voltages. Press NO twice to display the temperature voltage and actual temperature display. Unplug the sample sensor and fit the
Temperature Adjustment Test Plug in place of the sample sensor plug. Adjust R54 located on the SBC Board for a display of 3000mV. Unplug the Temperature
Adjustment Test Plug and fit the sample sensor connector to display the actual temperature of the measuring chamber.
Due to the elevated temperature inside the analyzer, it is normal for the displayed temperature to be several degrees C higher than the actual room temperature.
If no Temperature Adjustment Test Plug is available, place a quality thermometer
(preferably mercury) next to the right side electrode holder. With the front door open, wait approx. 10 min. Then adjust the displayed temperature to match the reading on the thermometer.
71
72
7 Electronic Diagrams
AVL 9120, 30, 40, 80 Analyzers
System Block Diagram 9120, 30, 40, 80 ........................................ 74
System Block Diagram 9181 .......................................................... 75
System Interconnect 9120, 30, 40 .................................................. 76
SBC PCB - Power Supply 9120, 30, 40 ........................................ 77
SBC PCB - Valve Drivers 9120, 30, 40 ......................................... 78
SBC PCB - Pump Motor Driver/Door Detect 9120, 30, 40 ............ 79
SBC PCB - Input Amplifiers 9120, 30, 40 ...................................... 80
SBC PCB - Temperature Circuit 9120, 30, 40 ............................... 81
SBC PCB - Analog Selector and A/D Converter 9120, 30, 40 ....... 82
SBC PCB - Microprocessor 9120, 30, 40 ..................................... 83
SBC PCB - Printer/Display Drivers 9120, 30, 40 ........................... 84
Printer Daughter PCB 9120, 30, 40, 80, 81 .................................... 85
Lamp PCB 9120, 30, 40 ................................................................ 86
Display Daughter PCB 9120, 30, 40 ............................................... 87
AVL 9180/81 Analyzers
System Interconnect 9180/9181 .......................... .......................... 88
SBC PCB - Power Supply 9180/9181 ........................................... 89
SBC PCB - Valve Drivers 9180/9181 ............................................ 90
SBC PCB - Pump Motor Driver/Door Detect 9180/9181 ............... 91
SBC PCB - Input Amplifiers 9180/9181 ......................................... 92
SBC PCB - Temperature Circuit 9180/9181 .................................. 93
SBC PCB - Analog Selector and A/D Converter 9180/9181 .......... 94
SBC PCB - Microprocessor 9180/9181 ........................................ 95
SBC PCB - Printer/Display Drivers 9180/9181 .............................. 96
SBC PCB - Fluid Pack Detector 9180/9181 .................................. 97
Lamp PCB 9180/9181 ................................................................... 98
Display Daughter PCB 9180 ........................................................... 99
73
74
SBC Board
Preamplifier
Na Electrode
K
Cl
Reference
Electrode
GND
Na Amp
K Amp
Cl/Ca/Li Amp
Analog Selector and
A/D Converter
Power Supply
+22VDC / +10VDC
+5VDC / -5VDC (9180)
+8VDC / -8VDC
+22VDC / +12.5VDC
Interface Driver
JP7
RS232
100 - 240 V~
MAIN
50 - 60 Hz
9120, 30, 40, 80 Analyzer Block Diagram
Fluid Pack Detector
Sensor
(9180 only)
Display Driver
Temperature Control
(9140/9180 only)
JP4
JP2
Display Board
Display Module YES/NO Switches
Temp Sensor
(9140/9180 only)
Sample Sensor
JP1
Printer Driver
Sample Sensor/Temp Sensor
JP1
Printer
Assembly
Microprocessor
JP3
Peristaltic Pump Driver
JP5
Solenoid Valve Drivers
JP8
Lamp Board
P. Pump Motor
JP6
Sample Door
JP7
Valve A JP1
Valve B JP2
Valve C JP3
Valve V JP4
Valve R JP5
Power Receptacle
1.6 AT
1.6 AT
JP1
Power Supply Module
Fuse F1 2.5A/250V
+5VDC/+24VDC
+12VDC/-12VDC
9181 Analyzer Block Diagram
SBC Board
Preamplifier
Na Electrode
K
Cl
Reference
Electrode
GND
Na Amp
K Amp
Cl/Ca/Li Amp
Analog Selector and
A/D Converter
Power Supply
+22VDC / +10VDC
+5VDC / -5VDC
+8VDC / -8VDC
+22VDC / +12.5VDC
Interface Driver
JP7
RS232
100 - 240 V~
MAIN
50 - 60 Hz
Fluid Pack Detector
Sensor
Display Driver
Temperature Control
Display Board
Needle Motor
Driver
Needle Sensor
Position Buffer
JP4
JP2
Display Module YES/NO Switches
JP5
Probe Motor
JP6
Position Sensor
Temp Sensor Sample Sensor
Needle
Mechanism
Assembly
JP1 JP1
Sample Sensor/Temp Sensor
Printer Driver
JP1
Needle Ctrl Lines
Microprocessor
Turntable Ctrl Lines
JP1
Printer
Assembly
Turntable Assembly
Driver BD
Home Sensor
Position Sensor
Peristaltic Pump Driver
JP5
JP8
Solenoid Valve Drivers
JP3
BK5014 BK5013
JP9
Ctrl Lines
Turntable
Bulkhead
Lamp Board
P. Pump Motor
JP6
Sample Door
JP7
Valve A JP1
Valve B JP2
Valve C JP3
Valve V JP4
Valve R JP5
Power Receptacle
1.6 AT
1.6 AT
JP1
Power Supply Module
Fuse F1 2.5A/250V
+5VDC/+24VDC
+12VDC/-12VDC
75
SYSTEM INTERCONNECT - 9120, 30, 40
76
SBC PCB - POWER SUPPLY - 9120, 30, 40
A D J
77
SBC PCB - VALVE DRIVERS - 9120, 30, 40
78
A D J
SBC PCB - PUMP MOTOR DRIVER / DOOR DETECT - 9120, 30, 40
79
SBC PCB - INPUT AMPLIFIERS - 9120, 30, 40
1 1
80
SBC PCB - TEMPERATURE CIRCUIT - 9120, 30, 40
81
SBC PCB - ANALOG SELECTOR AND A/D CONVERTER - 9120, 30, 40
82
SBC PCB - MICROPROCESSOR - 9120, 30, 40
83
SBC PCB - PRINTER/ DISPLAY DRIVERS - 9120, 30, 40
1 2
84
PRINTER DAUGHTER PCB - 9120, 30, 40, 80, 81
85
LAMP PCB - 9120, 30, 40
86
DISPLAY DAUGHTER PCB - 9120, 30, 40
87
SYSTEM INTERCONNECT - 9180, 81
88
SBC PCB - POWER SUPPLY - 9180, 81
A D J
89
SBC PCB - VALVE DRIVERS - 9180, 81
90
A D J
SBC PCB - PUMP MOTOR DRIVER/DOOR DETECT - 9180, 81
91
SBC PCB - INPUT AMPLIFIERS - 9180, 81
1 1
1 1
92
SBC PCB - TEMPERATURE CIRCUIT - 9180, 81
93
SBC PCB - ANALOG SELECTOR AND A/D CONVERTER - 9180, 81
94
SBC PCB - MICROPROCESSOR - 9180, 81
95
SBC PCB - PRINTER/ DISPLAY DRIVERS - 9180, 81
1 2
96
SBC PCB - FLUID PACK DETECTOR - 9180, 81
97
LAMP PCB - 9180, 81
98
DISPLAY DAUGHTER PCB - 9180
99
100
8 Parts List
Item Part
13
14
15
13
13
13
11
12
12
8
9
10
18
18
18
16
17
18
6
7
8
5
5
5
2
3
4
1
2
19
BP5193
BP5027
BP0087
YA0296
ZB5012
ZB5129
ZB5007
ZB5009
ZB5027
ZB5124
DS5004
DZ5002
YB5005
BP5008
BP5084
BP5005
RE5018
BP5009
BP5188
BP5035
BP5004
RE5016
BP5012
ZB5001
ZB5015
ZB5005
ZB5006
ZB5025
ZB5123
ZB5024
MAIN ASSEMBLY FRONT
Description
SLEEVE: BLACK ELECTRODE HOLDER
DOOR: ISE FRONT COMPLETE (OLD COLOR)
DOOR: ISE FRONT COMPLETE (NEW COLOR)
NEEDLE UNIT: COMPLETE MECHANISM
TRAY: ISE ELECTRODE
MODULE: ELECTRODE HOLDER - 9120, 30
MODULE: ELECTRODE HOLDER - 9180
MODULE: ELECTRODE HOLDER - 9140
MODULE: ISE PRINTER
TRAY: ISE PRINTER PAPER
TUBE SET: ISE COMPLETE - 9120, 30, 40
TUBE SET: ISE COMPLETE - 9180
TUBE SET: PERISTALTIC PUMP
CLAMP: PRESSURE PIECE VALVE TUBING
CAP: SOLENOID VALVE LOCK
LABEL: 91XX PAPER INSERTION DIAGRAM
LABEL: 9180 PAPER INSERTION DIAGRAM
LABEL: 9120 ELECTRODE HOLDER
LABEL: 9130 ELECTRODE HOLDER
LABEL: 9140 ELECTRODE HOLDER
LABEL: 9180 ELECTRODE HOLDER
SCREW: ALLEN HEAD 8-32 X .625
WASHER: #8 STAINLESS FLAT
LABEL: SAMPLE SENSOR CABLE
LABEL: ELECTRODE LIGHT WINDOW
LABEL: AVL 9120 OVERLAY
LABEL: AVL 9130 OVERLAY
LABEL: AVL 9140 OVERLAY
LABEL: AVL 9180 OVERLAY
LABEL: PAPER TEAR BAR
101
102
Item Part
BP5007
BP5002
BB5002
BB5016
RE5025
DS5005
DZ0488
EU1019
DS5006
BP5003
DS5001
DS5006
DZ5000
DS5006
DS0970
BP5013
BP5083
BB5005
BB5006
BB5015
US3059
US3060
US3061
US5010
BB5007
BB5014
DS5002
YB5002
13
14
15
10
11
12
16
17
18
7
8
9
5
6
7
3
4
4
3
3
3
2
2
2
1
1
19
20
MAIN ASSEMBLY REAR
Description
PANEL: ISE BACK COMPLETE (OLD COLOR)
PANEL: ISE BACK COMPLETE (NEW COLOR)
PC CARD: ISE SBC COMPLETE 9120/9130
PC CARD: ISE SBC COMPLETE 9140
PC CARD: ISE SBC COMPLETE 9180
PROGRAM SET: 9120 ENGLISH
PROGRAM SET: 9130 ENGLISH
PROGRAM SET: 9140 ENGLISH
PROGRAM SET: 9180
PC CARD: ISE LAMP - 9120, 30, 40
PC CARD: ISE LAMP - 9180
SOLENOID: COMPLETE ISE
MOTOR: ISE PERISTALTIC PUMP
PC CARD: ISE DISPLAY COMPLETE - 9120, 30 , 40
PC CARD: ISE DISPLAY COMPLETE - 9180
SPACER: 50 DEGREE
SCREW: #6 X .625"
WASHER: LOCK M4
CONNECTOR: GROUND LUG
SCREW: #6 X .375"
PUMP ROLLER: ISE PERISTALTIC PUMP
SCREW: 6-32 X .312"
SCREW: #6 X .375"
WASHER: FLAT
SCREW: #6 X .375"
SCREW: PUMP MOTOR MOUNTING
SCREW: SOLENOID VALVE MOUNTING
SPRING PROBE: ELECTRODE CONNECTOR
0
6
6
4
5
6
6
2
3
1
2
0
0
ELECTRODE HOLDER ASSEMBLY
Item Part Description
BP5009
BP5035
BP5188
BP0171
BP5010
BP5036
RE5019
DS0970
DA0156
ZB5007
ZB5009
ZB5027
ZB5124
MODULE: ELECTRODE HOLDER COMPLETE ASSEMBLY
9120/9130
MODULE: ELECTRODE HOLDER COMPLETE ASSEMBLY
9140
MODULE: ELECTRODE HOLDER COMPLETE ASSEMBLY
9180
ELECTRODE HOLDER: LEFT SIDE ASSEMBLY
HOLDER: ISE RIGHT ELECTRODE 9120/9130
HOLDER: ISE RIGHT ELECTRODE 9140/80
HOLDER: BASE ISE ELECTRODE
SCREW: PHILLIPS M4x16
O-RING: ELECTRODE SEAL
LABEL: 9120 ELECTRODE HOLDER
LABEL: 9130 ELECTRODE HOLDER
LABEL: 9140 ELECTRODE HOLDER
LABEL: 9180 ELECTRODE HOLDER
103
104
Item Part
RE5028
RE5009
BP5006
BP9043
YA9023
DS5006
DS5000
DZ0478
SS5003
YB5006
BP5005
RE5011
RE5010
RE5012
RE5014
BP5018
RE5029
YB5001
12
13
14
15
8
9
10
11
16
17
6
7
4
5
2
3
0
1
NEEDLE UNIT ASSEMBLY
Description
NEEDLE UNIT: COMPLETE MECHANISM
PLATE: ISE NEEDLE UNIT RIGHT MOUNTING
PLATE: ISE NEEDLE UNIT LEFT MOUNTING
HOLDER: ISE NEEDLE UNIT PROBE
GEAR: ISE NEEDLE UNIT
LIGHT GATE: ISE NEEDLE UNIT COMPLETE
PLUNGER: ISE NEEDLE UNIT
SPRING: ISE NEEDLE UNIT
ROD: ISE NEEDLE UNIT
FLAP: ISE NEEDLE UNIT
PROBE: ISE COMPLETE
FILLPORT: ASSEMBLY
PLATE: FILL PORT MOUNTING
SCREW: #6 X 0.375"
SCREW: #6 X .312
WASHER: FLAT M3
TUBING: 3/8"OD X 1/4"ID CLEAR SILICON
RUBBER STOP: ISE NEEDLE UNIT
Item Part
BP5013
BP5083
BP5030
BP5081
BN5000
BK5000
EU5004
EV0044
NS0025
DS5000
EU1019
BK5001
BK5012
4
5
2
3
1
1
0
0
8
9
6
7
10
REAR PANEL ASSEMBLY
Description
PANEL: ISE BACK COMPLETE (OLD COLOR)
PANEL: ISE BACK COMPLETE (NEW COLOR)
COVER: PAINTED ISE REAR (OLD COLOR)
COVER: PAINTED ISE REAR (NEW COLOR)
MODULE: SWITCHING ISE POWER SUPPLY
RECEPTACLE: AC POWER WITH CABLE
HOLDER: 2 POLE ISE FUSE
FUSE: 1.6A SLOW
SCREW PLASTIC TAPING
SCREW: #6 X .312
CONNECTOR: FASTON 6.3
CABLE: DC ASSEMBLY
CABLE: GROUND ASSEMBLY
105
106
FRONT DOOR ASSEMBLY
Item Part
1
2
0
0
2
BP5008
BP5084
RE5008
ZB5002
ZB5125
Description
DOOR: ISE FRONT COMPLETE (OLD COLOR)
DOOR: ISE FRONT COMPLETE (NEW COLOR)
WINDOW: ISE FRONT DOOR
LABEL: 91XX ISE TUBING DIAGRAM
LABEL: 9180 ISE TUBING DIAGRAM
9 Interface Specifications
9.1 Interface Information
The AVL 9120, 30, 40, 80 and 81 analyzers are equipped with a standard serial interface output. This interface output is intended to be used with standard commercially available computer systems.
The data transmitted through the serial interface port employs the ASCII code.
The serial interface is terminated on the rear cover with a 9-pin male DB-9 connector.
The signal levels are as follows:
- Binary 1 = -12V to -3V
- Binary 0 = +3V to +12V
Two stop bits follow the eight data bits to complete the 10 bit word.
The baud rate is set at 9600 Baud fixed.
The maximal recommended cable length is 40 feet.
The pin assignment is as follows: pin 1...sample ground...GND
pin 2...receive data....RxD
pin 3...send data.......TxD
pin 4...NC
pin 5...signal ground...GND
pin 6...NC
pin 7...RTS
pin 8...CTS
pin 9...NC
(NC=Not Connected)
107
108
Software
The patient sample data is sent at the end of each measurement, the calibration report is sent at the end of each calibration.
The interface is always on, independent of the printer settings; the data is always sent.
Note: The arrow up (e.g. out of normal range) is sent as HEX 18 (
↑
), the arrow down as HEX19 (
↓
) and the ° (degree) is sent as HEX1A (
→
).
9.2 Example Data String Information
Automatic Calibration Report
sx* AVL 9130 *crcrlfELECTROLYTE ISEcrcrlf03JAN92
10:51crcrlf*CALIBR REPORT*crcrlfcrcrlfDailyMaintenancecrcrlf
Performed Last: crcrlf02JAN92 10:35crcrlfcrcrlf
Standard AcrcrlfNa = -112mV (3)crcrlfK = -1392mV (3) crcrlfCl = -106mV (3)crcrlfcrcrlfDifference A-Bcrcrlf
Na = 1402mV ( )cr crlfK = 1032mV ( )crcrlfCl = -1006mV ( ) crcrlfcrcrlfFluid Pack:crcrlf6 8% Remainingcrcrlfcrcrlfex
Serum Sample Report
sx* AVL 9130 *crcrlfELECTROLYTE ISEcrcrlf03JAN92
10:59crcrlfcrcrlf Name: ..........crcrlf ..........crcrlfSample:
SERUMcrcrlfcrcrlf Sample No.13 crcrlfcrcrlfNa= 159soh mmol/LcrcrlfK
= 5.4 mmol/LcrcrlfCl= 122soh mmol/Lcrcrlfcrcrlf*PERFORM DAILY
*crcrlf*MAINTENANCE ! *crcrlfcrcrlfex
9.3 Data Link Information (9180/9181 only)
The data link with the AVL Compact 2/3 blood gas analyzer allows to combine ISE results with pH/blood gas results on one printout. If Ca ++ is activated on the 9180/9181, a pH-corrected Ca
++
value will be calculated and printed on the combined sample report.
For connection of the 9180/9181 analyzer to the Compact 2/3 analyzer, the optional Interface Kit (BP5202) is required. To install the kit, first turn both instruments off. Connect the interface filter provided in the kit to the
RS232 port on the 9180/9181. Then connect one end of the cable to the interface filter, the other end to the COM 2 port on the Compact 2/3. On the Compact 2/3, select 9180 under the COM2 interface options.
See Compact 2/3 Operator's Manual for details.
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 7 1 Introduction
- 8 2 General Description
- 8 2.1 Reported Parameters
- 8 2.2 Operating Parameters
- 11 3 Software
- 13 3.1 System Software Flow Charts
- 23 3.2 System Software Flow Charts
- 35 3.3 System Software Flow Charts
- 48 3.4 Service Functions
- 54 4 Description of Modules
- 54 4.1 Mechanical Assemblies
- 58 4.2 Fluidic Module
- 62 4.3 Electronics
- 65 5 Maintenance and Troubleshooting
- 65 5.1 Maintenance Procedures
- 66 5.2 Error Messages and Troubleshooting
- 77 6 Adjustments
- 77 6.1 Sample Sensor Adjustment
- 77 6.2 Temperature Adjustment
- 78 7 Electronic Diagrams
- 80 System Block Diagram
- 82 System Interconnect
- 83 SBC PCB - Power Supply
- 84 SBC PCB - Valve Drivers
- 85 SBC PCB - Pump Motor Driver/Door Detect
- 86 SBC PCB - Input Amplifiers
- 87 SBC PCB - Temperature Circuit
- 88 SBC PCB - Analog Selector and A/D Converter
- 89 SBC PCB - Microprocessor
- 90 SBC PCB - Printer/Display Drivers
- 91 Printer Daughter PCB
- 92 Lamp PCB
- 93 Display Daughter PCB
- 94 System Interconnect
- 95 SBC PCB - Power Supply
- 96 SBC PCB - Valve Drivers
- 97 SBC PCB - Pump Motor Driver/Door Detect
- 98 SBC PCB - Input Amplifiers
- 99 SBC PCB - Temperature Circuit
- 100 SBC PCB - Analog Selector and A/D Converter
- 101 SBC PCB - Microprocessor
- 102 SBC PCB - Printer/Display Drivers
- 103 SBC PCB - Fluid Pack Detector
- 104 Lamp PCB
- 105 Display Daughter PCB
- 107 Main Assembly Front
- 108 Main Assembly Rear
- 109 Electrode Holder Assembly
- 110 Needle Unit Assembly
- 111 Rear Panel Assembly
- 112 Front Door Assembly
- 113 9 Interface Specifications
- 113 9.1 Interface Information
- 114 9.2 Example Data String Information
- 114 9.3 Data Link Information