Fiberglass 30
Laboratory Fume Hoods
Copyright © 2018 Labconco Corporation. The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
Warranty
Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply, unauthorized repair and products used with incompatible chemicals.
The warranty for Fiberglass 30 Laboratory Fume Hoods will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only applies to the owner (organization) of record.
Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3 rd party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit this warranty based on a service agent’s travel, working hours, the site’s entry restrictions and unobstructed access to serviceable components of the product.
Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier.
Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:30 a.m. and 5:30 p.m., Central Standard Time.
Part #2831600, Rev. H
ECO L354
Original instructions
T
A B L E
O
F
C
O N T E N T S
CHAPTER 1: INTRODUCTION
About This Manual
Typographical Conventions
Your Next Step
CHAPTER 2: PREREQUISITES
Location Requirements
Support Requirements
Exhaust Requirements
Electrical Requirements
Service Line Requirements
Space Requirements
Your Next Step
CHAPTER 3: GETTING STARTED
Unpacking Your Fume Hood
Removing the Shipping Skid
Sash Weight Release
Install the Hood on a Supporting Structure & Work Surface
Connecting to the Hood Exhaust System
Connecting the Electrical Supply Source to the Hood
Connecting the Service Lines to the Hood
Sealing the Fiberglass 30 Hood to the Work Surface
Certifying the Fiberglass 30 Fume Hood
Your Next Step
CHAPTER 4: PERFORMANCE FEATURES AND
SAFETY PRECAUTIONS
Performance Features
Safety Precautions
Your Next Step
5
6
6
6
7
7
7
7
1
1
2
4
10
12
13
15
8
8
9
10
16
16
17
18
18
21
24
CHAPTER 5: USING YOUR FIBERGLASS 30
Operating the Vertical-Rising Sash
Operating the Blower
Operating the Lights
Working in Your Fiberglass 30 Fume Hood
Your Next Step
CHAPTER 6: MAINTAINING YOUR FIBERGLASS 30
Routine Maintenance Schedule
Routine Service Operations
Your Next Step
CHAPTER 7: MODIFYING YOUR FIBERGLASS 30
Installing Work Surfaces and Storage Cabinets
Installing Additional Service Fixtures
Installing Guardian Digital Airflow Monitor or Guardian Jr.
Installing an Electrical Duplex Receptacle
Your Next Step
CHAPTER 8: TROUBLESHOOTING
APPENDIX A: FIBERGLASS 30 HOOD COMPONENTS
APPENDIX B: FIBERGLASS 30 HOOD DIMENSIONS
APPENDIX C: FIBERGLASS 30 HOOD SPECIFICATIONS
APPENDIX D: REFERENCES
35
38
43
46
27
27
28
29
30
30
30
31
32
32
33
25
25
25
25
25
26
Original instructions
C
H A P T E R
1
I I
N T R O D U C T I I O N
Congratulations on your purchase of a Labconco Fiberglass 30 Laboratory Fume
Hood. Your Fiberglass 30 Laboratory Fume Hood is designed to protect you. It is the result of Labconco’s more than 50 years experience in manufacturing fume hoods, and users like you suggested many of its features to us.
The Labconco Fiberglass 30 Fume Hood has been engineered to effectively contain toxic, noxious, or other harmful materials when properly installed. The
Fiberglass 30 offers many unique features to enhance safety, performance, and visibility. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how fume hoods operate, please review Chapter 4: Performance Features and Safety
Precautions before you begin working in the fume hood. Even if you are an experienced fume hood user, please review Chapter 5: Using Your Fume Hood, which describes your Fiberglass 30 features so that you can use the hood efficiently.
About This Manual
This manual is designed to help you install, use, and maintain your laboratory fume hood. Instructions for installing optional equipment on your hood are also included.
Chapter 1: Introduction provides a brief overview of the laboratory fume hood, explains the organization of the manual, and defines the typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your laboratory fume hood. Electrical and service requirements are discussed.
Product Service 1-800-522-7658 1
Original instructions
Chapter 1: Introduction
Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify your laboratory fume hood.
Chapter 4: Performance Features and Safety Precautions explains how the laboratory fume hood operates and the appropriate precautions you should take when using the fume hood.
Chapter 5: Using Your Fiberglass 30 discusses the basic operation of your fume hood. Information on how to prepare, use and shut down your fume hood is included.
Chapter 6: Maintaining Your Fiberglass 30 explains how to perform routine maintenance on your fume hood.
Chapter 7: Modifying Your Fiberglass 30 explains how to modify the fume hood or add accessories.
Chapter 8: Troubleshooting contains a table of problems you may encounter while using your laboratory fume hood including the probable causes and suggested corrective actions.
Appendix A: Fiberglass 30 Hood Components contains labeled diagrams of all of the components of the fume hoods.
Appendix B: Fiberglass 30 Hood Dimensions contains comprehensive diagrams showing all of the dimensions for the laboratory fume hoods.
Appendix C: Fiberglass 30 Hood Specifications contains the electrical requirements for laboratory fume hood. Wiring diagrams are also included.
Appendix D: References lists the various resources available that deal with laboratory fume hoods.
!
Typographical Conventions
Recognizing the following typographical conventions will help you understand and use this manual:
Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started ).
Steps required to perform a task are presented in a numbered format.
Comments located in the margins provide suggestions, reminders, and references.
Critical information is presented in boldface type in paragraphs that are preceded by the exclamation icon. Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to fume hood.
2 Product Service 1-800-522-7658
Original instructions
Chapter 1: Introduction
Les informations critiques sont présentées en gras dans les paragraphes qui sont précédés par l'icône d'exclamation. Ne pas se conformer aux informations qui suivent une icône d'exclamation peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte aspirante.
Critical information is presented in boldface type in paragraphs that are preceded by the wrench icon. These operations should only be performed by a trained certifier or contractor. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your hood.
Les informations critiques sont présentées en gras dans les paragraphes qui sont précédés par l'icône de clé plate. Ces opérations devraient être seulement exécutées par un professionnel agrée. L'échec pour se conformer aux informations qui suivent une icône de clé plate peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte.
Important information is presented in capitalized type in paragraphs that are preceded by the pointer icon. It is imperative that the information contained in these paragraphs be thoroughly read and understood by the user.
!
CAUTION – See Manual. When this symbol is on a fume hood it indicates a caution that is detailed in this manual.
PRUDENCE – Consulter le Manuel. Quand ce symbole est sur une hotte aspirante, il indique une prudence qui est détaillée dans ce manuel.
CAUTION – Hot Surface.
AVERTIR – Surface Chaude
CAUTION – See Manual. This symbol on the fume hood indicates the possibility of a pinch hazard.
PRUDENCE – Consulter le Manuel. Ce symbole sur la hotte indique la possibilité d'un risque de pincement.
Product Service 1-800-522-7658 3
Original instructions
Chapter 1: Introduction
Your Next Step
If your Fume Hood needs to be installed, proceed to Chapter 2: Prerequisites to ensure your installation site meets all of the requirements. Then, go to Chapter 3:
Getting Started for instructions on how to install your laboratory fume hood and make all of the necessary connections.
If you would like to review how laboratory fume hoods operate, go to Chapter 4:
Performance Features and Safety Precautions .
For information on the operational characteristics of your laboratory fume hood, go to Chapter 5: Using Your Fiberglass 30 Hood .
If your laboratory fume hood is installed and you need to perform routine maintenance on the cabinet, proceed to Chapter 6: Maintaining Your Fiberglass
30 Hood .
For information on making modifications to the configuration of your fume hood, go to Chapter 7: Modifying Your Fiberglass 30 Fume Hood .
Refer to Chapter 8: Troubleshooting if you are experiencing problems with your fume hood.
4 Product Service 1-800-522-7658
Original instructions
C
H A P T E R
2
P
R E R E Q U I I S I I T E S
Before you install your laboratory fume hood, you need to prepare your site for installation. Carefully examine the location where you intend to install your hood. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn the requirements for your installation site:
The location requirements.
The support requirements.
The exhaust requirements.
The electrical power requirements.
The service line requirements.
The space requirements.
Refer to Appendix B: Fiberglass 30 Hood Dimensions for complete fume hood dimensions.
Refer to Appendix C: Fiberglass 30 Hood Specifications for complete laboratory fume hood electrical and environmental conditions, specifications and requirements.
! If the unit is not operated as specified in this manual it may impair the protection provided by the unit.
Si l'unité n'est pas utilisée comme spécifié dans ce manuel il peut diminuer la protection fournie par l'unité.
Product Service 1-800-522-7658 5
6
Original instructions
Chapter 2: Prerequisites
Location Requirements
! The fume hood should be located away from traffic patterns, doors, windows, fans, ventilation registers, and any other airhandling device that could disrupt its airflow patterns. All windows in the room should be closed.
La hotte aspirante devrait être localisé loin des voies de circulation, des portes, des fenêtres, des ventilateurs, des bouches de ventilation, et de tout appareil qui pourrait interrompre ses voies de flux d'air. Toutes les fenêtres dans la pièce devraient être fermées.
Support Requirements
! DO NOT install the fume hood on a cart, dolly, or mobile bench. ALL Fiberglass 30 Hood installations must be permanent and stationary. The supporting structure usually consists of a base cabinet and chemically resistant work surface. See Chapter 7 Modifying Your Fiberglass 30 Hood.
NE PAS installer la hotte sur un panier, chariot, ou banc mobile. TOUS fibre de verre de 30 installations hotte doit être permanent et stationnaire. La structure de support est généralement constitué d'une armoire de base et la surface de travail résistant aux produits chimiques. Voir Chapitre 7
Modification de la fibre de verre 30 Hood.
Exhaust Requirements
The exhaust duct connection has been designed for 6" nominal duct (6.625" OD) to allow for minimum static pressure loss while operating at 100 fpm face velocities. The 6" diameter exhaust duct also allows for proper transport velocities away from the hood in the 1000 fpm to 2500 fpm range. The proper exhaust volume and static pressure loss are listed next for each hood model:
Fiberglass 30 Hood with Integral Blower, 100 fpm, 310 cfm, capable of overcoming 0.12" external static pressure.
Fiberglass 30 Hood for use with Remote Blower, 100 fpm, 310 cfm, 0.14" static pressure loss.
Product Service 1-800-522-7658
Original instructions
Chapter 2: Prerequisites
Fiberglass 30 Hood with Explosion Proof Blower, 100 fpm, 310 cfm, capable of overcoming 0.12" external static pressure.
Proper blower selection can be determined from these exhaust requirements and the total system static pressure loss. Contact Labconco Customer Service for assistance in sizing a blower.
Electrical Requirements
Other than explosion proof models, the Fiberglass 30 Hood models feature internal wiring for the incandescent light assembly, light switch and blower switch. If your unit has an integral blower, then internal wiring to the blower motor and blower switch is included. Explosion Proof blowers require on site wiring. All internal wiring is terminated at the single point wiring junction box for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C : Fiberglass 30 Specifications for the wiring diagram for proper electrical installation.
Autre que les modèles d'explosion preuve, les modèles en fibre de verre 30 capot disposent câblage interne pour l'ensemble de lumière incandescente, interrupteur de lumière et commutateur de soufflerie. Si votre appareil dispose d'une soufflerie intégrée, alors le câblage interne du moteur du ventilateur et l'interrupteur du ventilateur est inclus. Explosion Proof ventilateurs exigent le câblage du site. Tout le câblage interne est terminé à l'câblage boîte de jonction unique point pour le branchement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en route et à l'Annexe C: fibre de verre 30 Spécifications pour le schéma de câblage pour l'installation électrique appropriée.
Service Line Requirements
All service lines to the laboratory fume hood should be ¼ inch outside diameter, copper (brass for natural gas), and equipped with an easily accessible shut-off valve, should disconnection be required. Recommended operating pressure is
40 PSI, with a maximum allowable pressure of 200 PSI. Consider a pressure regulator to reduce line pressure to 40 PSI. Please check with local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in Appendix B: Fiberglass 30
Hood Dimensions .
Your Next Step
After you have determined that the location you have selected accommodates the installation and operational requirements of your fume hood, you are ready to begin installation. Proceed to Chapter 3: Getting Started .
Product Service 1-800-522-7658 7
Original instructions
C
H A P T E R
3
G
E T T I I N G
S
T A R T E D
Now that the site for your laboratory fume hood is properly prepared, you are ready to unpack, inspect, install, and certify your unit. Read this chapter to learn how to:
Unpack and move your Fiberglass 30 Hood.
Set up the fume hood with the supporting structure and work surface.
Connect to an exhaust system.
Connect the electrical supply source.
Connect the service lines.
Seal the Fiberglass 30 Hood to the work surface.
Arrange certification of your Fiberglass 30 Hood.
Depending upon which model you are installing, you may need common plumbing and electrical installation tools in addition to 5/16", 3/8", 7/16", and 1/2" wrenches, ratchets, sockets, a nut driver set, a flat-blade screwdriver, a Phillips screwdriver, and a carpenter level to complete the instructions in the chapter.
8
Unpacking Your Laboratory Fume Hood
Carefully remove the shrink-wrap or carton on your fume hood and inspect it for damage that may have occurred in transit. If your unit is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR
AUTHORIZATION OF LABCONCO. UNAUTHORIZED
RETURNS WILL NOT BE ACCEPTED.
IF YOUR HOOD WAS DAMAGED IN TRANSIT, YOU MUST
FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER.
LABCONCO CORPORATION AND ITS DEALERS ARE NOT
RESPONSIBLE FOR SHIPPING DAMAGES.
The United States
Interstate
Commerce
Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
Do not discard the shipping skid or packing material for your fume hood until you have checked all of the components and installed and tested the unit.
Do not remove the fume hood from its shipping skid until it is ready to be placed into its final location. Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
! THE FIBERGLASS 30 HOOD MODELS WEIGH BETWEEN
96 TO 129 LBS. (44-59 KG). THE SHIPPING SKID ALLOWS
FOR LIFTING WITH A MECHANICAL LIFT TRUCK OR
FLOOR JACK. IF YOU MUST LIFT THE FUME HOOD
MANUALLY, FOLLOW SAFE-LIFTING GUIDELINES.
LA FIBRE DE VERRE DE 30 MODELES HOOD PESENT
ENTRE 96 A 129 LBS. (44-59 KG). LA PALETTE
D'EXPÉDITION PERMET LEVAGE AVEC UN CHARIOT
ELEVATEUR MECANIQUE DES SOLS JACK. SI VOUS
DEVEZ DECROCHER LE HOTTE MANUELLE, SUIVRE
LES DIRECTIVES SAFE-LEVAGE.
NORMALLY, THE FUME HOOD CAN BE SLID OFF A
HYDRAULIC LIFT TABLE AND BE PLACED INTO
POSITION ON TOP OF THE WORK SURFACE. DO NOT
LIFT BY THE FRONT AIR FOIL.
NORMALEMENT, LES HOTTE PEUT ETRE GLISSE OFF
UN ASCENSEUR HYDRAULIQUE ET TABLE ETRE
PLACE DANS POSITION SUR LE DESSUS DE LA
SURFACE DE TRAVAIL. NE PAS LEVER DU FRONT AIR
FOIL.
! Do not move the hood by tilting it onto a hand truck.
Ne déplacez pas le capot en l'inclinant sur un camion de main.
Removing the Shipping Skid
LEAVE THE FUME HOOD ATTACHED TO ITS SHIPPING
SKID UNTIL IT IS AS CLOSE TO ITS FINAL LOCATION AS
POSSIBLE. MOVE THE HOOD BY USING A SUITABLE
FLOOR JACK, OR BY PLACING A FURNITURE DOLLY
UNDERNEATH THE SKID. DO NOT MOVE THE HOOD BY
TILTING IT ONTO A HAND TRUCK.
Product Service 1-800-522-7658 9
Original instructions
Chapter 3: Getting Started
After you verify the fume hood components, move your hood to the location where you want to install it.
Sash Weight Release
To protect the fume hood from damage in shipment, the sash weight has been secured to the back of the fume hood with four (4) screws. Simply remove the screws and make sure the sash cables are on the pulleys before operation of the sash.
NOTE : THE SASH WEIGHT ITSELF WAS INDIVIDUALLY
MATCHED FOR THIS SPECIFIC HOOD AND SHOULD NOT
BE EXCHANGED ON ANY OTHER UNIT.
Install the Fiberglass 30 Hood on a
Supporting Structure and Work Surface
! The Fiberglass 30 Hood is heavy! Use caution when lifting or moving the unit.
La fibre de verre 30 Hood est lourd! Soyez prudent lorsque vous soulever ou déplacer l'unité
When installing the Fiberglass 30 Fume Hood onto a chemically resistant work surface or benchtop, ensure that the structure can safely support the combined weight of the fume hood and any related equipment. The work surface should be at least as wide as the hood to properly support it. The work surface is aligned with the back of the fume hood. This will provide the correct spacing under the air foil for proper bypass airflow.
! WARNING: It is important to support the rear of the work surface and fume hood. The cross support provides support for the bottom of the work surface. Install the cross support after the base cabinets and work surface are leveled and before installing the hood.
AVERTISSEMENT : Il est important de soutenir l'arrière de la surface de travail et la hotte aspirante. Le support tranversal soutient le bas de la surface de travail. Installer le support transversal après que les meubles et la surface de travail soient nivelés et avant d'installer la hotte.
The following are instructions for mounting a cross support:
1.
Level the base cabinets and the work surface. Work surface should be placed flush with the back of the fume hood as shown in Figure 3-1.
2.
Scribe a line on the wall or back of the base cabinet to locate the support under the work surface.
10 Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
3.
Mount the support by attaching it to the wall or base cabinet.
4.
Place the hood on top of the work surface and cross support.
The work surface should be smooth and durable, such as a chemically-resistant epoxy resin. The surface should be nonporous and resistant to the acids, solvents, and chemicals used.
Place Cross
Support Here
Product Service 1-800-522-7658
Figure 3-1
11
Original instructions
Chapter 3: Getting Started
Connecting to the Fiberglass 30 Hood
Exhaust System
! WARNING: The weight of the exhaust ductwork system must be supported independently of the hood superstructure. Do not allow this weight to be supported by the hood structure as damage to the hood may occur.
AVERTISSEMENT : Le poids du système d'aspiration de ductwork doit être soutenu d'une manière indépendante de la superstructure de la hotte. Au cas où ce poids est supporté par la structure de la hotte, des dommages à la hotte peuvent arriver.
!
The exhaust connection should be installed by a qualified
HVAC contractor. The exhaust connection on your hood has been designed for 6" nominal pipe (6.625" OD) to allow for minimum static pressure loss with proper transport velocities away from the hood. Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and total system static pressure loss. See Chapter 2 for the hood airflow specifications.
Le raccord d'évacuation doit être installé par un entrepreneur en CVC qualifié. La connexion d'échappement sur votre hotte a
été conçue pour 6 "diamètre nominal (6.625 "OD) pour permettre la perte de pression statique minimale avec des vitesses de transport appropriés à l'extérieur de la hotte.
Consultez Labconco Service à la clientèle si vous avez besoin d'aide dimensionnement de votre ventilateur pour le volume de gaz d'échappement et système global de perte de pression statique. Voir le chapitre 2 pour le flux d'air de capot spécifications.
The selected exhaust duct material should match the hood procedures and chamicals used to ensure compatibility.
Le matériau de conduit d'évacuation choisi doit correspondre aux procédures de capot et chamicals utilisés pour assurer la compatibilité.
12 Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
Connecting the Electrical Supply Source to the Fiberglass 30 Fume Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the hood identification plate for the proper electrical requirements of your specific model.
WARNING : The building electrical supply system for
Fiberglass 30 Hoods is required to include overload protection. !
A switch or circuit breaker shall be in close proximity to the equipment and within easy reach of the operator. The switch or circuit breaker is to be marked as the disconnecting device for the equipment. Do not position the unit so that it is difficult to operate the main disconnect device.
AVERTISSEMENT : Le système d'alimentation électrique pour la construction en fibre de verre 30 Hoods est nécessaire d'inclure la protection de surcharge . Un commutateur ou disjoncteur doit être tout près de l'équipement et à portée facile de l'opérateur. Le commutateur ou le disjoncteur doit être marqué comme l'appareil débranchant pour l'équipement. Ne pas placer l'appareil de sorte qu'il est difficile de faire fonctionner le dispositif de déconnexion principal.
The identification plate, model number, serial number, and electrical connection boxes are accessible from the front of the fume hood by removing the front panel.
The Fiberglass 30 Hood is normally wired for 115 Volt, 60 Hz, 20 Amp or 230
Volt, 50 Hz, 10 Amp electrical services. The number of circuits varies depending on the model. All of the electrical connections are terminated at the single point internal junction box for hook-up by a qualified electrician. Refer to the wiring diagram for your Fiberglass 30 Hood in Appendix C: Fiberglass 30 Fume Hood
Specifications.
Product Service 1-800-522-7658 13
Chapter 3: Getting Started
Original instructions
Internal Junction
Box
Figure 3-2
All wiring for the fume hood SHOULD be performed by a licensed electrician and conform to all local and national electrical codes. In most cases, the hood will require the use of shielded conduit to protect the wiring into the hood. The grounding connection shall not be made to the terminal box cover.
Tout le câblage pour la hotte doit être effectuée par un
électricien agréé et conforme à tous les codes électriques locaux et nationaux. Dans la plupart des cas, le capot ne nécessiter l'utilisation d'un conduit blindé pour protéger le câblage dans la hotte. La connexion de mise à la terre ne doit pas être fait à la couverture de la boîte à bornes.
The incandescent light has been mounted inside the top liner panel and is sealed from vapors inside the hood structure. To change the incandescent light bulbs in your hood, you must first remove the protective glass.
Replace the defective bulbs, and reassemble.
14 Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
Connecting the Service Lines to the
Fiberglass 30 Fume Hood
The hoods with service fixture kits are plumbed from the valve to the hose connector. Supply tubing shall be provided by the qualified installer. Tubing can enter the hood from above, through the back, or through the work surface to make these connections to the service fixtures.
! NOTE: Inspect all fittings for leakage. Tighten the fittings slightly if needed.
NOTE: Inspectez tous les raccords pour les fuites. Serrer les raccords légèrement si nécessaire.
! CAUTION : Do not use oxygen with any standard service fixture. Contact Labconco Customer Service for oxygen fixture information.
ATTENTION: Ne pas utiliser d'oxygène avec n'importe quel appareil de service standard. Contactez Labconco Service à la clientèle pour obtenir des informations de fixation de l'oxygène.
Should access to the hood plumbing fixture bodies be required, remove the service access plate on the side of the hood by loosening screws. The valve body will be fully exposed for any service work that may be necessary. The service fixture kits on your laboratory hood are designed for use with the following services:
Air
Hot Water
Vacuum
Cold Water
Natural Gas – See caution below
! WARNING : Contact Labconco Customer Service directly before using any service other than those listed above in these valves to assure full compatibility.
AVERTISSEMENT: Contact Labconco directement au service client avant d'utiliser tout service autre que ceux énumérés cidessus dans ces vannes pour assurer une compatibilité totale.
Product Service 1-800-522-7658 15
Original instructions
Chapter 3: Getting Started
! CAUTION: Natural gas should be used only in the service fixture that has been pre-plumbed with brass tubing. Sulfur content of the gas could cause deterioration of standard copper supply lines.
PRUDENCE : Le gaz naturel devrait être seulement utilisé dans l'accessoire de service qui a été pré soudé avec des tuyaux de cuivre. Le contenu soufré du gaz pourrait causer la détérioration des lignes d’alimentation en cuivre standard.
Sealing the Fiberglass 30 Hood to the
Work Surface
When the hood has been set in place, ducted, wired, and plumbed, it should be sealed at the work surface to prevent spilled materials from collecting under the walls of the hood. Materials such as silicone sealants are recommended to seal the hood structure.
Certifying the Fiberglass 30 Fume Hood
The combination of your laboratory hood, exhaust ductwork, and exhaust blower give you the flexibility to change the airflow at the sash opening of your hood. To determine the actual face velocity at the sash opening, airflow velocity readings will need to be taken. This should be done across the sash opening of the hood in accordance with the Industrial Ventilation Manual section on laboratory hoods.
(See Appendix D – Reference) Labconco recommends an average face velocity at the sash opening of 80 to 100 feet per minute. Consult Chapter 2 for proper airflow volumes for your particular model.
The Fiberglass 30 Fume Hood has been tested per ASHRAE 110-1995. All hoods achieve an “as manufactured rating” of less than 0.10 part per million (ppm) at 4 liters per minute (lpm); AM<0.10 (consult Labconco for individual fume hood ratings). For “field use” ASHRAE testing by a certified technician, contact
Labconco Customer Service.
! NOTE: Face velocity profiles and smoke testing should be done periodically to ensure safe performance .
NOTE : Les profils de flux frontal et les tests de fumée devraient être régulièrement faits pour garantir une utilisation en toute securité.
16 Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
Your Next Step
After your fume hood has been installed and certified, you are ready to proceed to
Chapter 4: Performance Features and Safety Precautions .
Product Service 1-800-522-7658 17
C
H A P T E R
4
P
E R F O R M A N C E
F
E A T U R E S
A
N D
S
A F E T Y
P
R E C A U T I I O N S
Performance Features
The Fiberglass 30 Laboratory Hood is designed to meet the needs of the laboratory scientist. A fully featured by-pass hood with baffle and air foil can effectively contain toxic, noxious, or other harmful materials when properly installed. The by-pass airflow feature promotes full containment as the sash is moved.
The air foil sweeps the worksurface to remove heavier than air contaminants.
1.
Unique sash provides maximum visibility of 28" high while conserving energy by limiting sash travel to 20" . Vertical-rising sash may be raised from a closed to 20" operating height. Exhaust volume, and blower sizing is based on the 20" height.
2.
By-pass airflow design ensures containment as the sash moves.
3.
Large usable interior work depth and interior height of 36" provides ample working space.
4.
Baffle (not shown) directs airflow to the rear of the interior to provide efficient airflow. The baffle may be removed for cleaning purposes only.
5.
Lift-Away™ front panel provides easy access to electrical wiring, and sash weights.
6.
Incandescent lighting provides ample illumination inside the fume hood.
No lighting is provided on explosion proof models.
7.
Low mounted, factory-wired light and blower switches are ADA compliant. No switches are provided on explosion-proof models.
Product Service 1-800-522-7658 18
Original instructions
Chapter 4: Performance Features & Safety Precautions
8.
Curved Air Foil allows air to sweep the work surface for maximum containment.
!
WARNING: Do not lift hood in this area. Hood is to be lifted only by side or back.
AVERTISSEMENT: Ne pas soulever le capot dans ce domaine. Hotte doit être levée que par côté ou à l'arrière.
9.
Corner posts provide maximum visibility and the flexibility to add services after installation.
10.
All hoods are factory prepared for up to 3 service fixtures . Field installed by removing side access panel.
11.
Removable side access panel for servicing valves and electrical connections.
12.
All hoods are factory prepared for one electrical duplex. A duplex can be field mounted on the right corner post. Receptacles are field-wired to the single point junction box.
13.
Shipped fully assembled and eliminates the need for costly onsite assembly.
14.
Accessory Guardian™ Digital Airflow Monitor or Guardian Jr.
Monitor continuously monitors face velocity. An audio/visual alarm alerts the user to low airflow conditions. The right corner post is factory prepared to accommodate the Guardian Monitor (sold separately).
15.
Frame of epoxy-coated steel and aluminum is durable and corrosion resistant.
16.
Exhaust connection . The hood features 6" (6.625" OD pipe) exhaust connections sized to allow for a minimum static pressure loss through the hood structure while providing a good transport velocity through the exhaust system.
17.
Single point junction box provides a single connection point for electrical wiring connections.
18.
Convenient integral blower motor is readily accessible on integral blower models.
Product Service 1-800-522-7658 19
1
5
15
Chapter 4: Performance Features & Safety Precautions
16
2
8
12
18
7
7
Figure 4-1
Original instructions
17
9
14
10
11
20 Product Service 1-800-522-7658
Original instructions
Chapter 4: Performance Features & Safety Precautions
Safety Precautions
! Although the laboratory hood has been engineered to maintain optimum operator safety, caution should always be used while working in the hood. Prior to using the hood, check to make sure that the exhaust blower is operating and
! that air is entering the hood at its specified face velocity.
Bien que la hotte de laboratoire ait été réglée pour maintenir la sécurité optimale de l’opérateur, la prudence devrait toujours être utilisée en travaillant sous la hotte.
Avant utiliser la hotte, le contrôle pour s'assurer que la soufflerie d'aspiration fonctionne et que cet air entre dans la hotte au flux spécifié.
USE GOOD HOUSEKEEPING IN THE HOOD AT ALL
TIMES. CLEAN UP SPILLS IMMEDIATELY.
PERIODICALLY CLEAN HOOD INTERIOR, INCLUDING
LIGHT GLASS PANEL. REPLACE BURNED OUT LIGHT
BULBS TO MAINTAIN MAXIMUM ILLUMINATION.
DO NOT OVERLOAD THE WORK SURFACE WITH
APPARATUS OR WORK MATERIAL. THE SAFE
OPERATION OF THE LABORATORY HOOD IS BASED
UPON HAVING PROPER AIRFLOW THROUGH THE
STRUCTURE. DO NOT PLACE LARGE, BULKY OBJECTS
SUCH AS BLOCK HEATERS, DIRECTLY ON THE HOOD
WORK SURFACE. INSTEAD, ELEVATE THE OBJECT 2"
TO 3" ON BLOCKS TO ALLOW A FLOW OF AIR UNDER
THE OBJECT AND INTO THE LOWER REAR BAFFLE
EXHAUST SLOT. ENSURE BLOCKS ARE LEVEL AND
SECURED IN PLACE.
Blocking the bottom of the baffle at the rear of hood will change the airflow pattern in the hood causing turbulence and possible leakage at the face of the hood. (Don’t store containers or supplies against the baffle, as this will affect airflow through the hood).
Le blocage de la partie inférieure de la chicane à l'arrière du capot va changer le modèle d'écoulement d'air dans la hotte provoquant une turbulence et une fuite possible à la face du capot. (Ne pas entreposer les contenants ou des fournitures contre le déflecteur, car cela aura une incidence sur le débit d'air à travers le capot).
Product Service 1-800-522-7658 21
Original instructions
Chapter 4: Performance Features & Safety Precautions
Avoid placing your head inside hood. Keep hands out of hood as much as practical.
Evitez de placer votre tête dans le capot. Gardez les mains sur le capot autant que possible.
Always work as far back in the hood as possible. It is best to keep all chemicals and apparatus 6" inside the front of the hood.
Toujours travailler aussi loin dans le capot que possible. Il est préférable de garder tous les produits chimiques et appareils 6 "à l'intérieur de l'avant du capot.
Only explosion-proof models feature explosion-proof electrical components. Therefore, use of flammable or explosive materials in quantities above the explosive limit are not recommended.
Modèles antidéflagrants disposent composants électriques antidéflagrants. Par conséquent, l'utilisation de matériaux inflammables ou explosives dans des quantités supérieures à la limite d'explosion ne sont pas recommandés.
Do not work with chemicals in this hood without the exhaust system running. Do not store chemicals in a fume hood.
Ne pas travailler avec des produits chimiques dans la hotte sans que le système d'échappement en marche. Ne pas stocker de produits chimiques dans une hotte.
Perchloric acid use in this hood is prohibited.
Utilisation de l'acide perchlorique dans cette hotte est interdite.
High-level radioisotope materials are prohibited for use in this hood.
Matériaux de radio-isotopes de haut niveau sont interdits d'utilisation dans la hotte.
AVOID CROSS DRAFTS AND LIMIT TRAFFIC IN FRONT
OF THE HOOD. AIR DISTURBANCES CREATED MAY
DRAW FUMES OUT OF THE HOOD.
22 Product Service 1-800-522-7658
Original instructions
!
Chapter 4: Performance Features & Safety Precautions
The use of heat-generating equipment in this hood without the exhaust system operating properly can cause damage to the hood.
L'usage d'équipement chaleur-produiant dans ce capuchon sans l'opération de système d'aspiration peut causer convenablement des dommages à la hotte.
The Fiberglass 30 Laboratory Hood should be certified by a qualified certification technician before it is initially used.
The hood should be re-certified whenever it is relocated, serviced or at least annually thereafter.
La fibre de verre 30 Hotte de laboratoire doit être certifié par un technicien qualifié avant de certification, il est d'abord utilisé. La hotte doit être re-certifié à chaque fois qu'il est déplacé, l'entretien ou au moins chaque année par la suite.
Ensure that the unit is connected to electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the hood from electrical service.
Assurez-vous que l'appareil est connecté au service
électrique conformément aux codes électriques locaux et nationaux. Ne pas le faire peut créer un risque d'incendie ou
électrique. Ne pas enlever ou de réparer des composants
électriques sans d'abord débrancher le capot du service
électrique.
Proper operation of the fume hood depends largely upon the hood’s location and the operator’s work habits. Consult the
Reference Manual in Appendix D.
Le bon fonctionnement de la hotte dépend largement de l'emplacement de la hotte et les habitudes de travail de l'opérateur. Consultez le manuel de référence à l'annexe D.
Product Service 1-800-522-7658 23
Original instructions
Chapter 4: Performance Features & Safety Precautions
!
Do not touch the blower motor. The surfaces of the motor can become hot and could cause burns.
Ne pas toucher le moteur de soufflerie. Les surfaces du moteur peuvent devenir chaudes et pourrait causer des brûlures.
Your Next Step
After you understand the theory of operation and safety precautions, you are ready to proceed to Chapter 5: Using Your Fiberglass 30 Fume Hood .
24 Product Service 1-800-522-7658
C
H A P T E R
5
U
S I I N G
Y
O U R
F
I I B E R G L A S S
3 0
Operating the Vertical-Rising Sash
Because of the Fiberglass 30 Hood counterbalanced sash mechanism, it will take only a few pounds of force to move the sash up or down, and you can operate the sash smoothly with one or two hands positioned anywhere along the handle. The vertical-rising sash may be raised to a maximum 20" operating height. The airflow requirements should be sized for the 20" operating height.
Operating the Blower
Your Fiberglass 30 Fume Hood utilizes a remote style blower or integral blower, which can be activated by turning the blower switch to “ON.” You can validate the hood performance by watching smoke drawn into the hood face opening.
Explosion-proof models do not have a switch or wiring as this is field installed.
Operating the Lights
Your Fiberglass 30 Fume Hood is equipped with a factory-wired incandescent light to illuminate the hood interior. Simply turn the light switch to “ON” to operate. Explosion-proof models do not have a light.
Working in Your Fiberglass 30 Fume
Hood
Planning
Thoroughly understand procedures and equipment required before beginning work.
Arrange for minimal disruptions, such as room traffic or entry into the room while the hood is in use.
Product Service 1-800-522-7658 25
Original instructions
Chapter 5: Using the Protector Perchloric Acid
Start-up
Turn on incandescent light and hood blower.
Slowly raise the sash.
Check the baffle air slots for obstructions.
Allow the hood to operate unobstructed for 5 minutes.
Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate.
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload the hood.
Do not obstruct the front air foil, or rear baffle slots.
Large objects should not be placed close together and spaced above the work surface to permit airflow to sweep under the equipment.
After loading the hood, wait one minute to purge airborne contaminants from the work area.
Work Techniques
Keep all materials at least 6 inches inside of the sash, and perform all contaminated operations as far to the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area.
Avoid using techniques or procedures that disrupt the airflow patterns of the hood.
Final Purging
Upon completion of work, the hood should be allowed to operate for two to three minutes undisturbed, to purge airborne contaminants from the work area before shutting down blower.
Unloading Materials and Equipment
Objects in contact with contaminated material should be surface decontaminated before removal from the hood.
All open trays or containers should be covered before being removed from the hood.
Shutdown
Turn off the incandescent light and hood blower, and then close the sash.
Your Next Step
After you understand how to operate and work in the fume hood, you are ready to proceed to Chapter 6: Maintaining Your Fiberglass 30 Fume Hood .
26 Product Service 1-800-522-7658
C
H A P T E R
6
M
A I I N T A I I N I I N G
Y
O U R
F
I I B E R G L A S S
3 0
Now that you have an understanding of how to work in the fume hood, we will review the suggested maintenance schedule and the common service operations necessary to maintain your fume hood for peak performance.
! Only trained and experienced certification technicians should perform some of the service operations after the fume hood has been properly decontaminated. DO NOT attempt to perform these operations if you are not properly trained. The wrench icon precedes the service operations that require qualified technicians.
Techniciens de certification Seulement formés et expérimentés devraient effectuer certaines des opérations de service après la hotte a été correctement décontaminés. NE PAS tenter de réaliser ces opérations si vous n'êtes pas correctement formés.
L'icône de la clé précède les opérations de service qui nécessitent des techniciens qualifiés
Routine Maintenance Schedule
Weekly
Using ordinary dish soap to clean the surface inside of the fume hood, and the work surface.
Using an appropriate glass cleaner, clean the sash and all glass surfaces.
Operate the fume hood blower, noting the airflow velocity through the hood using a source of visible smoke.
Product Service 1-800-522-7658 27
Original instructions
Chapter 6: Maintaining Your Fiberglass 30
Monthly (or more often as required)
Mensuel (ou plus souvent si nécessaire)
Determine the actual face velocity through the sash opening of the hood where the average reading should be at the specified velocity. (Use calibrated thermal anemometer or other approved apparatus).
Déterminer la vitesse de visage réel à travers l'ouverture de la guillotine de la hotte où la valeur moyenne doit être à la vitesse spécifiée. (Utilisez calibré anémomètre thermique ou un autre appareil approuvé)
Using a damp cloth, clean the exterior surfaces of the hood, particularly the front of the hood, to remove any accumulated dust.
Utiliser un chiffon humide, nettoyez les surfaces extérieures de la hotte, en particulier l'avant du capot, pour enlever toute la poussière accumulée.
Check all service valves, if so equipped, for proper operation.
Vérifiez que toutes les vannes de service, si équipé, pour un bon fonctionnement.
The hood baffles should be checked for blockages behind them to ensure that the hood is maintaining proper airflow.
Les déflecteurs de capot doivent être vérifiés pour les blocages derrière eux pour veiller à ce que le capot est de maintenir une ventilation adéquate.
All weekly activities.
Toutes les activités hebdomadaires.
Annually
Annuellement
Have the fume hood recertified by a qualified certification technician. See
Certifying the Fiberglass 30 Fume Hood in Chapter 3
Avez la hotte recertification par un technicien de certification qualifié. Voir la certification de fibre de verre 30 Hotte au chapitre 3.
All monthly activities.
Toutes les activités mensuelles
Biannually
The sash assembly should be checked to ensure proper operation and to make sure there are no signs of abnormal wear on the sash pulleys, cables and clamps.
Routine Service Operations
Front Panel Removal:
1.
Simply lift the front panel up and then away from the hood to provide access to the top.
28 Product Service 1-800-522-7658
Original instructions
!
Chapter 6: Maintaining Your Fiberglass 30
Changing the Incandescent Lamp:
1.
Turn light switch to “OFF”.
2.
Raise the sash and reach inside the fume hood to unscrew the clear glass globe that protects the incandescent bulb.
3.
Install the new incandescent bulb and then reverse the removal procedure.
Do not contact blower wheel while still in motion
Your Next Step
After you understand the maintenance procedures, you are ready to proceed to
Chapter 7: Modifying Your Fiberglass 30 Fume Hood .
Product Service 1-800-522-7658 29
C
H A P T E R
7
M
O D I I F Y I I N G
Y
O U R
F
I I B E R G L A S S
3 0
There are several ways to modify the fume hood for your individual requirements.
These include the addition of work surfaces, service fixtures, air monitor, and electrical duplex outlets. See Appendix A: Fiberglass 30 Hood Components for installation location.
Installing Work Surfaces and Storage
Cabinets
An optional work surface is available when installing the hood on storage cabinets.
Contact Labconco Customer Service for ordering information on these 30" work surfaces and storage cabinets.
4882806 Work Surface 30" x 30"
9902200 Solvent Cabinet 30"
9901200 Acid Cabinet 30"
9900200 Base Cabinet 30"
Installing Additional Service Fixtures
Additional service fixtures can be installed in the available service fixture holes in the right sidewall and corner post. The fume hood is factory set to accept up to three valves. Contact Labconco Customer Service for information. The following service fixture kits are available:
30 Product Service 1-800-522-7658
Original instructions
Chapter 7: Modifying Your Fiberglass 30
2832600 Cold Water (CW) – Green
2832601 Hot Water (HW) – Red
2832602 Deionized Water (DW) – White
2832603 Vacuum (VAC) – Yellow
2832604 Air (AIR) – Orange
2832605 Gas (GAS) – Blue
2832606 Nitrogen (NIT) – Brown
2832607 Argon (ARG) – Gray
2832608 Steam (STM) - Black
Figure 7-1 - Knob Figure 7-2 – Valve Figure 7-3 – Hose Connector
Installing Guardian™ Digital Airflow
Monitor or Guardian™ Jr. Airflow Monitor
The Guardian Digital Airflow Monitor P/N 9743201 and 9743205 or Guardian Jr.
Airflow Monitor P/N 9743202 and 9743206 allows you to continuously monitor face velocity through the fume hood opening. The fume hood right corner post is factory prepared to mount either monitor. Contact Labconco Customer Service to order.
Figure 7-4 Figure 7-5
Product Service 1-800-522-7658 31
Original instructions
Chapter 7: Modifying Your Fiberglass 30
Installing an Electrical Duplex Receptacle
If you ordered your hood and want an electrical duplex outlet, then you can have one installed in the field by a qualified electrician. Contact Labconco Customer
Service for ordering information. (Not acceptable on explosion-proof hoods). The following duplex kits are available:
Si vous avez commandé votre hotte et que vous voulez une sortie duplex
électrique, alors vous pouvez avoir un installé sur le terrain par un électricien qualifié. Contactez Labconco Service à la clientèle pour les informations de commande. (Non acceptable sur cagoules anti-explosion). Les kits recto verso suivantes sont disponibles:
2834800 115V, 20A Duplex Kit
2834801 115V, 20A GFCI Duplex Kit
Figure 7-6
Your Next Step
After you understand the modifying procedures, you are ready to proceed to
Chapter 8: Troubleshooting .
32 Product Service 1-800-522-7658
C
H A P T E R
8
T
R O U B L E S H O O T I I N G
Refer to the following table if your fume hood fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance.
PROBLEM
Remote blower and light won’t operate
Remote blower won’t operate, but light works
Fume hood blower operates but light will not operate
Fume hood blower operates but light will not operate
CAUSE
Wires not connected at junction box or switches.
Circuit breakers tripped in building electrical supply.
Blower wiring is disconnected.
Belt broken.
Blower motor is defective.
Lamp not installed correctly.
Lamp is defective.
Circuit breaker in building is tripped.
Lamp wiring is disconnected.
Defective lamp.
CORRECTIVE ACTION
Check connection of switches.
Check connection to control box on top of unit.
Reset circuit breakers.
Inspect blower wiring and switch.
Replace belt.
Replace blower motor.
Inspect lamp installation.
Replace lamp.
Reset the circuit breaker.
Inspect lamp wiring.
Replace lamp.
Product Service 1-800-522-7658 33
Original instructions
Chapter 8: Troubleshooting
PROBLEM
Contaminants outside of fume hood
CAUSE
Improper user techniques for the fume hood.
Restriction of the baffle air slots or blockage of the exhaust outlet.
External factors are disrupting the fume hood airflow patterns or acting as a source of contamination.
Fume hood has improper face velocity.
Vertical-rising sash no longer operates smoothly
Electrical duplex outlet no longer has power
Cable is frayed or plastic protection is damaged.
Pulley bearing is damaged.
Cable has slipped off the pulleys.
Weight has broken pulleys.
Wires not connected or faulty duplex.
CORRECTIVE ACTION
See “Certifying the Hood” Chapter 3 and
“Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D)
Remove baffles to ensure that all air slots and the exhaust outlet are unobstructed.
See “Location Requirements” Chapter 2,
“Certifying the Hood” Chapter 3, and
“Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D)
Have fume hood re-certified and check remote blower exhaust system. Hood should have average face velocity of 80-
100 fpm.
Inspect cable and replace cable if worn or damaged immediately; otherwise injury could result.
Replace pulley, bearing or add grease.
Re-install, cable must be replaced immediately if damaged.
Replace weight pulleys.
Check wire connection or replace duplex.
Reset circuit breakers.
Service valves no longer operate
Circuit breakers tripped in building electrical supply.
Faulty building supply.
Valve no longer operates.
Supply line or outlet line has leaks.
Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI.
Replace valve and check for leaks.
Inspect line for leaks and fix any leaking plumbing connections.
34 Product Service 1-800-522-7658
A
P P E N D I I X
A
F
I I B E R G L A S S
3 0 H
O O D
C
O M P O N E N T S
Illustration A-1 indicate the location of the following service parts:
Item Quantity
1A
1B
1C
1D
1
1
1
1
2
6E
6F
6G
6H
6I
7A
7B
7C
7D
7E
8A
8B
8C
6A
6B
6C
6D
3I
3J
4
5
3A
3B
3C
3D
3E
3F
3G
3H
8D
Fiberglass 30 Hood Replacement Parts
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Part No.
2832600
2832601
2832602
2832603
2832604
2832605
2832606
2832607
2832608
2834800
2834801
9818200
9818100
9818300
1273600
1273700
4849800
9823700
9817000
9823702
9818000
9818700-01-02-03-
04-05-06-07-08
9818800
9818801
9818802
9818803
9818804
9818805
9818806
9818807
9818808
9819000
9825500
9818900
4849900
Description
Valve, Labconco (Water) 1/4" Compression Fitting
Valve, Labconco 1/4" Compression Fitting (AIR, GAS, VAC, NIT, etc.)
Valve, Labconco Deionized 1/4" Compression Fitting
Nut, Valve Mtg. (Labconco)
Knobs (Gray, Green, Blue, Orange, Yellow, Red, White, Black, Brown)
Hose Barb, Gray – Neutral or Argon – NOT SHOWN
Hose Barb, Green – Cold Water – NOT SHOWN
Hose Barb, Blue – Gas – NOT SHOWN
Hose Barb, Orange – Air – NOT SHOWN
Hose Barb, Yellow – Vacuum – NOT SHOWN
Hose Barb, Red – Hot Water – NOT SHOWN
Hose Barb, White – Deionized Water – NOT SHOWN
Hose Barb, Black – Neutral or Steam – NOT SHOWN
Hose Barb, Brown – Nitrogen – NOT SHOWN
Nut, Hose Barb – NOT SHOWN
Label, Knob (contains all labels)
Lens, Knob
Fixture Kits – (CW)
Fixture Kits – (HW)
Fixture Kits – (DI)
Fixture Kits – (VAC)
Fixture Kits – (AIR)
Fixture Kits – (GAS)
Fixture Kits – (NIT)
Fixture Kits – (ARG)
Fixture Kits – (STM)
115V, 20A Duplex Kit
115V, 20A GFCI Duplex Kit
Cover Plate 115V Duplex
Cover Plate, 115V GFCI
Wallplate, Blank
Bulb, 115V
Bulb, 230V
Light Fixture Assy complete with wiring, parts 8D through 8G
Light Fixture – a) Body & Socket VFB15, b) Pendent Cover 1/2" VP1, c) Globe, 150W, 264–0006-9909
35 Product Service 1-800-522-7658
Original instructions
Item Quantity
21
22A
22B
23
24
25
26
27
15E
15F
16A
16B
17
18
19
20
28
29A
29B
29C
30
31
32
30
31
32
12
13
14A
14B
15A
15B
15C
15D
8E
8F
8G
9A
9B
10A
10B
11
4
2
1
2
1
1
1
2
1
1
1
1
1
4
1
1
1
1
1
1
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Appendix A: Fiberglass 30 Components
Part No.
1852300
1662600
1450000
4716200
2833400
2831900
2832000
2832500
2833800
1861400
3627500
4949902
1920100
1972100
2831500
1934300
4850000
1279700
1279800
1302300
1327500
2834400
2834500
2834000
2834100
2831400
2833000
2831700
2807100
1203600
1200400
1211000
9713300
2833500
2833501
1663200
2831100
2831200
2831300
9724500
1663200
2831300
Description
Wiring Harness, Light
Gasket, Base
Gasket, Globe
Switch, Rocker
Switch, Plug (Fills cutout without switch)
Wiring Harness, Main 115V
Wiring Harness, Main 230V
Label Set, Corner Post
Label, Front Panel
Front Panel
Motor Plate – 115V only
Motor Plate, 230V, 115V, EP - 230V, EP
Motor, 115V
Motor, 230V
Motor, 115 VEP
Motor, 230V EP
Reducing Bushing – 230V only
Rubber Grommet – 115V only
Wheel, Blower – 115V only
Wheel, Blower – 230V, 115V – EP, 230V – EP
Removable Side Panel – Right
Side Panel – Right
Side Panel – Left
Sash Glass Assembly
Sash Weight
Pulley – 1-3/16" Dia.
Thrust Washer
Cable, Sash
Clamp, Cable Replacement
S-Hook
Air Foil
Air Foil Spacer Stop
Rubber Bumper – Sash Handle
Bracket, Cable Front – Right
Bracket, Cable Front – Left
Rubber Bumper – Upper Sash Stop
Corner Post, Right
Corner Post, Left
Header
Sash Stop Kit
Upper Sash Bumper
Header
36 Product Service 1-800-522-7658
Original instructions
12
32
31
20
28
29
27 Not shown
26
19
11
23
9
10 Not shown
Appendix A: Fiberglass 30 Components
Figure A-1
22
16 Not Shown
15
14
24
21 Not Shown
24 Not Shown
25 Not Shown
7
7E for Air
Monitor
1 through 6 Not Shown
8 Not Shown
18
17
30
Product Service 1-800-522-7658 37
A
P P E N D I I X
B
F
I I B E R G L A S S
3 0 H
O O D
D
I I M E N S I I O N S
Figure B-1
38 Product Service 1-800-522-7658
Original instructions
Appendix B: Fiberglass 30 Hood Dimensions
Figure B-2
Product Service 1-800-522-7658 39
Appendix B: Fiberglass 30 Hood Dimensions
Original instructions
DIMENSION IN INCHES
40
Figure B-3
Product Service 1-800-522-7658
Original instructions
Appendix B: Fiberglass 30 Hood Dimensions
DIMENSION IN INCHES
Figure B-4
Product Service 1-800-522-7658 41
Appendix B: Fiberglass 30 Hood Dimensions
Original instructions
42
Figure B-5
Product Service 1-800-522-7658
Original instructions
A
P P E N D I I X
C
F
I I B E R G L A S S
3 0 H
O O D
S
P E C I I F I I C A T I I O N S
Environmental Conditions
Indoor use only.
Maximum altitude: 10,000 feet (3,048 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal voltage.
Transient over-voltages according to Installation Categories II (Overvoltage Categories per IEC 1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only non-conductive atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
Model
3030000
3030001
3030002
Electrical Requirements
115 VAC – 60 Hz, 1 Phase – 5 Amps
230 VAC – 50 Hz, 1 Phase – 5 Amps
115 VAC – 60 Hz, 1 Phase – 1 Amp
3030003
3030004
3030005
230 VAC – 50 Hz, 1 Phase – 1 Amp
115 VAC – 60 Hz, 1 Phase – 1 Amp
230 VAC – 50 Hz, 1 Phase – 1 Amp
Product Service 1-800-522-7658 43
Appendix C: Fiberglass 30 Specifications
Original instructions
44 Product Service 1-800-522-7658
Original instructions
Appendix C: Fiberglass 30 Specifications
Product Service 1-800-522-7658 45
Original instructions
A
P P E N D I I X
D
R
E F E R E N C E S
Many excellent reference texts and booklets are currently available. The following is a brief listing:
Laboratory Ventilation Standards
Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices”.
Fume hoods should have a continuous monitoring device
Face velocities should be between 60-100 linear feet per minute (lfpm)
Average 2.5 linear feet of hood space per person
Occupational Health and Safety
U.S. Department of Labor
200 Constitution Avenue N.W.
Washington, DC 20210
(202) 523-1452
Industrial Ventilation-ACGIH
Fume hood face velocities between 60-100 lfpm
Maximum of 125 lfpm for radioisotope hoods
Duct velocities of 1000-2000 fpm for vapors, gasses and smoke
Stack discharge height 1.3-2.0 x building height
Well designed fume hood containment loss, <0.10 ppm
Industrial Ventilation, A Manual of Recommended Practice.
24 th Edition, 2001
American Conference of Governmental Industrial Hygienists
1330 Kemper Meadow drive
Cincinnati, OH 45240-1634
(513) 742-2020
ASHRAE 110-1995 Method of Testing Performance of Fume Hoods
Evaluates fume hood’s containment characteristics
Three part test: Smoke generation, Face velocity profile, Tracer gas release @ 4 liters per minute
Rated As Manufactured (AM), As Installed (AI) and As Used (AU)
American Society of Heating, Refrigerating, and Air Conditioning Engineers
1791 Tullie Circle N.E.
Atlanta, GA 30329
(404) 636-8400
Product Service 1-800-522-7658 46
Original instructions
Appendix C: Fiberglass 30 Specifications
ANSI Z9.5-2011 Laboratory Standard
Covers entire laboratory ventilation system.
New and remodeled hoods shall have a monitoring device
Ductless hoods should only be used with non-hazardous materials
American Industrial Hygiene Association
2700 Prosperity Avenue, Suite 250
Fairfax, VA 22031
(703) 849-8888
SEFA 1-2002
Fume hood face velocities based on toxicity levels of chemicals
Class A – 125 to 150 fpm
Class B – 80 to100 fpm
Class C – 75-to 80 fpm
Test method – face velocity profile and smoke generation
Scientific Equipment & Furniture Association
1028 Duchess Drive
McLean, VA 22102
(703) 538-6007
NFPA 45 – 2002 Fire Protection for Laboratories Using Chemicals
Laboratory hoods should not be relied on for explosion protection
Exhaust air from fume hoods should not be recirculated
Services should be external to the hood
Canopy hoods only for non-hazardous applications
Materials of construction should have flame spread of 25 or less
80 to 120 fpm to prevent escape
NFPA 30 – 2000 Flammable and Combustible Liquids Code
Approved cabinets may be metal or wood
Vent location on cabinets are required
Venting of cabinets not a requirement
National Fire Protection Association
1 Batterymarch Park
P.O. Box 9101
Quincy, MA 02269-9101
(800) 344-3555
General References
American Conference of Governmental Industrial Hygienists . Industrial Ventilation, A
Manual of Recommended Practice, Cincinnati, OH
ASHRAE Standard Committee. ASHRAE Standard Atlanta: ASHRAE Publications Sales
Department, 1995
British Standards Institution , Laboratory Fume Cupboards . Parts 1, 2 and 3, London: 1990
Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910,
Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule . Vol. 55,
No. 21. Washington D.C.:1990
Product Service 1-800-522-7658 47
Original instructions
Appendix C: Fiberglass 30 Specifications
DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations .
Wiley & Sons, 1987
McDermott, Henry, Handbook of Ventilation for Contaminant Control , 2 nd
Edition .
Butterworth Publishers, 1985.
Miller, Brinton M. et al. Laboratory Safety: Principles and Practices.
American Society for Microbiology, Washington, D.C.: 1986
NIH Guidelines for the Laboratory Use of Chemical Carcinogens . NIH Publication No.
81-2385.
Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical
Laboratory . Springer-Verlag, New York: 1990
Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the
Workplace . Van Nostrand Reinhold, 1987.
Schilt, Alfred A. Perchloric Acid and Perchlorates . The G. Frederick Smith Chemical
Company, Columbus, OH: 1979.
Steere, Norman. CRC Handbook of Laboratory Safety, 2 nd
Edition . CRC Press, 1971.
Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910,
Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule . Vol. 55,
No. 21. Washington D.C.:1990
DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations .
Wiley & Sons, 1987
McDermott, Henry, Handbook of Ventilation for Contaminant Control , 2 nd Edition .
Butterworth Publishers, 1985.
Miller, Brinton M. et al. Laboratory Safety: Principles and Practices.
American Society for Microbiology, Washington, D.C.: 1986
NIH Guidelines for the Laboratory Use of Chemical Carcinogens . NIH Publication No.
81-2385.
Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical
Laboratory . Springer-Verlag, New York: 1990
Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the
Workplace . Van Nostrand Reinhold, 1987.
Schilt, Alfred A. Perchloric Acid and Perchlorates . The G. Frederick Smith Chemical
Company, Columbus, OH: 1979.
Steere, Norman. CRC Handbook of Laboratory Safety, 2 nd
Edition . CRC Press, 1971.