Breas Vivo 30, 40 Ventilator Service Manual
Breas Vivo 30 and Vivo 40 are pressure-supported and pressure-controlled ventilators for patients who require intermittent support by mechanical ventilation. These ventilators have three modes of operation: PSV, PCV, and CPAP. PSV and PCV modes have adjustable trigger sensitivity settings which allow the patient to initiate ventilator-assisted breaths. The PCV mode provides assisted or controlled pressure-regulated breathing, while the PSV mode offers expiratory sense adjustment for patient-specific needs. The CPAP mode provides continuous positive airway pressure. The logs of the Vivo 30/40 can be downloaded to a PC for analysis with the Vivo 30/40 PC Software.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Table of Contents
Intended Use .................................................................................... 5
Service Personnel’s Training Requirements..................................... 6
Intended Audience ........................................................................... 6
Safety Precautions .......................................................................... 8
Service Record................................................................................. 8
Inspection Equipment and Tools ...................................................... 8
Replacement Parts ........................................................................... 9
Initial Recording................................................................................ 9
Checking additional services............................................................ 9
Inspecting the Markings ................................................................... 9
Information from the Patient/User .................................................. 10
Validity of the Documentation ......................................................... 10
Visual Inspection for External Damage and Wear.......................... 10
Checking the HA 01 Humidifier (if used) ........................................ 10
Checking the Power Connection .................................................... 10
Inspecting the Patient Circuit.......................................................... 10
Inspecting the Ventilator Accessories............................................. 10
Changing/Washing the Patient Filters ............................................ 10
Minimum Function Check............................................................... 11
Cleaning the Inside of the Ventilator............................................... 11
Checking the Cables ...................................................................... 11
Checking the Fastening of Components ........................................ 11
Checking the Power Supply ........................................................... 11
Reassembling the Casing .............................................................. 11
Checking the Breath Rate and Inspiration Time............................. 12
Checking the Pressure ................................................................... 13
2.7.6
Checking the Tidal Volume Indication ............................................ 13
Preparing the Ventilator for Alarm Checks ..................................... 13
Checking the Low Leakage Alarm and High Leakage Alarm ......... 14
Checking the High Pressure Alarm ................................................ 14
Doc. 003823 En Z-1
Table of Contents
Breas Vivo 30/40 Service Manual
1
Checking the Inspiratory Trigger .................................................... 15
Checking the Low Pressure Alarm ................................................. 15
Checking the High Breath Rate Alarm (Vivo 40 only)..................... 15
Checking the Low Breath Rate Alarm (Vivo 40 only) ..................... 15
Checking the Low Volume Alarm .................................................. 16
Checking the Audio Pause ............................................................. 16
Checking External Battery Operation ............................................. 16
Checking the Internal Battery Operation (Vivo 40 only) ................. 16
Adjusting the Settings for the Patient ............................................. 16
Parts Drawing 1 – Main Components ............................................................. 19
Parts Drawing 2 – Vivo 30 Back Casing, Cables, Screws .............................. 20
Parts Drawing 3 – Vivo 40 Back Casing, Cables, Screws .............................. 21
Parts Drawing 4 - Blower Assembly................................................................ 21
CPU Boards revision 9 and earlier ................................................. 22
CPU Boards revision 10 and later .................................................. 23
Parts Drawing 7 – Vivo 40 Front Casing ......................................................... 25
Vivo 30/40 with magnetic valve (CPU board rev 9 and earlier)....................... 30
Vivo 30/40 without magnetic valve (CPU board rev 10 and later)................... 32
Vivo 30/40 Measuring and Display Devices.................................................... 33
Opening the Vivo 30/40 and Replacing the Main Components .................................... 34
Replacing the Complete Blower Assembly ..................................................... 35
Replacing the Push Button Silicone Pad......................................................... 45
6.2
2 Table of Contents
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
Main Cabling Diagram of the Vivo 30/40......................................................... 47
PSU Board ..................................................................................... 48
CPU Board ..................................................................................... 50
Test Point Locations ....................................................................... 52
Test Point Matrix............................................................................. 53
Ventilator Switch-over Operating Voltages...................................................... 55
Mains Power Supply Operation...................................................... 55
External Battery Operation ............................................................. 55
Internal Battery Operation (Vivo 40 only) ....................................... 56
Checking the Internal Battery (Vivo 40 only)................................................... 56
Reading the Error Codes................................................................ 58
Error Code Table ............................................................................ 58
9.4
Emission and Immunity Declaration................................................................ 64
Guidance and Manufacturer's Declaration – Electromagnetic
Immunity......................................................................................... 64
Guidance and Manufacturer's Declaration – Electromagnetic
Emission......................................................................................... 65
Product Return Form – Breas Vivo 30/40 ventilator........................................ 70
Doc. 003823 En Z-1
Table of Contents
Breas Vivo 30/40 Service Manual
3
4 Table of Contents
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
1 Introduction
This chapter gives an overview of the Vivo 30/40 ventilator and this service manual.
WARNING!
This product must be:
• Subjected to regular service, maintenance and control and any applicable upgrades, in accordance with Breas service instructions.
• Repaired and/or modified in accordance with Breas service manuals, technical bulletins, and any special service instructions, by service technicians that have been authorised after the Breas Vivo 30/40 service training.
Deviation from these service instructions may lead to risk of personal injury!
1.1
About the Vivo 30/40 Ventilator
1.1.1
Function
The Vivo 30/40 is a pressure-supported and pressure-controlled ventilator for patients who require intermittent support by mechanical ventilation.
It has three modes of operation: PSV, PCV, and CPAP. The PSV and PCV modes have an adjustable trigger sensitivity setting which allows the patient to initiate ventilator-assisted breaths.
In the PCV mode (Pressure Controlled Ventilation), the ventilator provides assisted or controlled pressure-regulated breathing.
In the PSV mode (Pressure Support Ventilation), the ventilator’s expiratory sense can also be adjusted allowing the ventilator to more easily match each patient’s needs.
In the CPAP mode (Continuous Positive Airway Pressure), the ventilator provides a continuous positive airway pressure.
The logs of the Vivo 30/40 can be downloaded to a PC, printed out, and analysed via the
Vivo 30/40 PC Software. For more information about the Vivo 30/40 PC Software, please contact your Breas representative.
1.1.2
Intended Use
For a detailed description on intended use, refer to the Vivo 30/40 Clinician's or User Manuals.
1.1.3
Design
The Vivo 30/40 is constructed around a blower assembly.
An electronically controlled stepmotor rotates a valve to control the pressure. A microprocessor controls the correct speed of the blower motor and its power supply by means of calculations based on the settings for pressure, rate, inspiration time, etc. The pressure and trigger settings are monitored at the same time.
Doc. 003823 En Z-1
Introduction
Breas Vivo 30/40 Service Manual
5
1.1.4
Service Personnel’s Training Requirements
Service personnel working with the Vivo 30/40 should have medical/technical training and a good knowledge of the construction and function of respiratory devices. Authorisation by Breas after
Vivo 30/40 service training is mandatory.
Always contact your Breas representative if you have any questions or if any training is required.
1.2
About this Manual
1.2.1
Scope
This manual describes all the routine maintenance checks and the additional service actions for the
Vivo 30/40. The manual contains all the documentation that is required for the maintenance and the service of the ventilator, such as replacement parts lists, exploded drawings, wiring diagrams, component location guides, etc.
Breas Medical AB reserves the right to make changes to the products and/or the contents of this manual without any prior notice.
1.2.2
Intended Audience
This service manual is intended for service technicians who have medical/technical training and who have a good knowledge of the construction and function of respiratory devices. Authorisation by Breas after Vivo 30/40 service training is mandatory.
Always contact your Breas representative if you have any questions or if any training is required.
The service manual is not intended for clinical personnel or patients, who will find all the information they need in the Vivo 30/40 Clinician's and User Manual.
1.2.3
Icons
In this manual, icons are used to highlight specific information. The meaning of each icon is explained in the table below.
I
CON
E
XPLANATION
Warning!
Risk of death and serious personal injury.
Caution!
Risk of minor or moderate injury. Risk of equipment damage, loss of data, extra work, or unexpected results.
Note
Information that may be valuable but is not of critical importance, tips.
Reference
Reference to other manuals with additional information on a specific topic.
6 Introduction
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
2 Maintenance Instructions
This chapter describes all the routine maintenance checks and additional service instructions for the
Vivo 30/40.
WARNING!
This product must be:
• Subjected to regular service, maintenance and control and any applicable upgrades, in accordance with Breas service instructions.
• Repaired and/or modified in accordance with Breas service manuals, technical bulletins, and any special service instructions, by service technicians that have been authorized after Breas Vivo 30/40 service training.
Deviation from these service instructions may lead to risk of personal injury!
The patient and care provider should follow the checks that are described in the Vivo 30/40 User and Clinician's Manual.
2.1
Purpose
The Vivo 30/40 is designed to give users many years of trouble-free breathing assistance, provided that the preventive maintenance is carried out at the intervals specified in this manual. Well-performed maintenance services will considerably increase the expected lifetime of the ventilator.
It is also important that any peripheral equipment is checked at the same time as the services are carried out.
2.2
Service Schedule
The maintenance service includes all the checks listed in the schedule below.
A complete maintenance service (as described in this chapter) must be carried out at least every 12 months.
I
NTERVAL
Every 12 months
Every 2 years
Every 5 years
Every 20,000 total operating hours
S
ERVICE CHECK
Perform external inspection.
Perform a complete function test.
Change the internal battery kit (Vivo 40 only).
Change the clock battery.
Change the alarm battery.
Perform an internal inspection.
Check the power failure alarm.
Change the complete blower assembly.
Perform an internal inspection.
S
ECTION
Doc. 003823 En Z-1
Maintenance Instructions
Breas Vivo 30/40 Service Manual
7
2.3
Introduction
Before you start a maintenance service, read the safety precautions and make sure you have a new service record and all the necessary equipment, tools, and replacement parts at hand.
2.3.1
Safety Precautions
Follow the safety precautions below when working with the Vivo 30/40:
• Do not work on the ventilator with the casing removed and the power supply connected, unless the instructions in this manual clearly says so.
• Always use caution when working with the ventilator connected to the mains and the casing removed.
• Do not use explosive gases and/or fluids near the ventilator.
• Make sure that all precautions to prevent electrostatic discharge (ESD) have been taken. Follow all regulations regarding ESD.
The Vivo 30/40 User and Clinician's Manuals contains extended lists of safety precautions.
2.3.2
Service Record
The Breas service record is found in chapter 9 “Appendices”.
Copy the service record and use it for noting the service checks while performing the yearly service.
2.3.3
Inspection Equipment and Tools
Before starting the service of the Vivo 30/40, make sure you have the following equipment at hand:
• Test lung or reservoir bag (for example Breas part no 001917).
• Measuring instrument for tidal volume and minute volume/rate (ventilator tester, spirometer, or equivalent, for example Fluke VT Plus).
Breas part no 001937).
• Digital voltmeter.
• Standard toolkit containing screwdrivers, including small potentiometer screwdriver, and
Torx keys, TX 10 and TX 20.
• Vivo to PC communication cable, Breas part no 003588.
• Vivo 30/40 Service Software (compatible with Fluke VT Plus, Thommen HM 28 and
TSI 4040), available for download from Breas Extranet. Contact Breas technical support for more information.
• Vivo Training CD, Breas part no 004278.
• External battery (24 or 12 V, for example Breas EB 2).
• Flow meter (for example TSI 4040).
• Test connector kit, Breas part no 004829 (connectors and leakageports used when testing).
8 Maintenance Instructions
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
2.3.4
Replacement Parts
The following replacement parts should be available when servicing the ventilator:
D
ESCRIPTION
Patient air filter, grey, washable
Patient air filter, white, disposable
If required:
Internal battery kit (Vivo 40)
Alarm battery, NiMH 4.8 V, 70 mAh
Clock battery, CR 2032
Blower assembly kit for replacement after a total operating time of
20,000 hours
P
ART
N
O
.
003563
003564
003452
004007
003524
2.4
Preparing for Inspection
2.4.1
Initial Recording
1
2
3
4
5
Copy a new service record (see “Service Record Vivo 30/40” on page 67).
Identify the Vivo 30 or Vivo 40.
Note the model and serial number and any inventory number on the service record.
Check any comments recorded on the previous service records.
Document the current patient settings.
2.4.2
Checking additional services
1
Note the number of Total operating hours on the service record.
The operating hours are found at the “Device information” screen. Access the “Device information” using the menu as described in the Vivo 30/40 Clinician’s Manual.
2
Check the service schedule to see whether the alarm/clock batteries, the internal battery kit
(Vivo 40 only), or the complete motor assembly needs to be replaced.
2.4.3
Inspecting the Markings
Make sure that all markings on the ventilator’s information labels can be read:
• Model description, serial number
• Warning texts
• Any inventory marking
• Any other texts
Doc. 003823 En Z-1
Maintenance Instructions
Breas Vivo 30/40 Service Manual
9
2.4.4
Information from the Patient/User
Check the following with the patient:
• Has the ventilator functioned without any problems? If not, what were they?
• How does the patient/care provider check the function of the ventilator? How often?
• How often is the filter replaced?
• How many filters will be required before the next service?
• Other observations?
2.4.5
Validity of the Documentation
1
Check the validity of the User or Clinician’s Manual enclosed with the ventilator.
2
Check if any modification or upgrading of the ventilator needs to be done at the same time as the service.
2.5
External Inspection
2.5.1
Visual Inspection for External Damage and Wear
3
4
1
2
Clean the outside of the ventilator using a mild soap solution.
Check for any visible damage to the casing and the other components.
Check that nothing has become loose.
Check that the rear lid is connected properly to the ventilator.
2.5.2
Checking the HA 01 Humidifier (if used)
1
2
3
4
Remove and open the HA 01.
Check that there is no visible damage.
Check that the water container is clean.
Check that the humidifier is connected properly to the ventilator.
2.5.3
Checking the Power Connection
3
4
1
2
Check the plugs on the power cord, the cord itself, and the ventilator’s power socket.
Make sure that the strain-relief clamp for the power cord is not damaged.
Inspect the external battery cable, if used.
Check the external battery socket in the ventilator.
2.5.4
Inspecting the Patient Circuit
Inspect the patient circuit and replace it if necessary.
2.5.5
Inspecting the Ventilator Accessories
Check any other accessories that are used with the ventilator.
2.5.6
Changing/Washing the Patient Filters
1
2
3
Change the white air filter.
Change the grey filter, if necessary.
Make sure the patient has enough filters to last until the next service.
10 Maintenance Instructions
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
2.5.7
Minimum Function Check
1
2
3
Connect the power cord.
Connect the patient circuit.
Switch on the ventilator and make sure it operates normally.
2.6
Internal Inspection
Perform the internal inspection after having opened the Vivo 30/40.
Make sure to disconnect the power supply before opening the casing of the ventilator.
2.6.1
Cleaning the Inside of the Ventilator
1
2
Remove any dirt or dust that has collected in the ventilator.
2.6.2
Checking the Cables
Inspect all the cables and their connectors. Check the front and rear panels to make sure that the cables and the wires are not pinched or kinked.
2.6.3
Checking the Fastening of Components
Make sure that all the components, such as the blower assembly, the circuit boards and the connectors are securely fastened.
2.6.4
Checking the Power Supply
1
2
3
Make sure that the power connector is undamaged and that it is securely in place.
Check the wiring to the CPU board.
Check that the PSU cooling fan is securely in place.
2.6.5
Reassembling the Casing
See “Opening the Vivo 30/40 and Replacing the Main Components” on page 34 for instructions.
Doc. 003823 En Z-1
Maintenance Instructions
Breas Vivo 30/40 Service Manual
11
2.7
Complete Function Test
You can also perform the complete function test using the Vivo 30/40 Service Software.
To perform a complete function test, all tasks in this section need to be completed in the written order.
2.7.1
Checking the Breath Rate and Inspiration Time
1
Adjust the settings as follows:
S
ETTING
Mode
Target Volume
IPAP
EPAP
Set Rate
Inspiration Time
Rise Time
Inspiration Trig
V
ALUE
PCV
Inactive
20 cmH
2
O
4 cmH
2
O
10 bpm
3.0 s
3
3
2
Connect a pressure/flow meter and test lung or a reservoir bag to the ventilator.
Leakage
4 mm hole
Pressure/flow meter
Test lung
3
Measure and check that the measured values for rate and inspiration time are correct
(tolerance ±10%).
12 Maintenance Instructions
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
2.7.2
Checking the Pressure
1
Adjust the settings as follows:
S
ETTING
Mode
IPAP
EPAP
Inspiration Time
V
ALUE
PCV
4 cmH
2
O
2 cmH
2
O
5.0 s
2
Connect a pressure manometer or ventilator tester to the ventilator.
3
Measure and check that the measured IPAP value for each pressure is correct.
EPAP IPAP T
OLERANCE
5 cmH
2
O 10 cmH
2
O ±1.8 cmH
2
O
10 cmH
2
O 20 cmH
2
O ±2.8 cmH
2
O
15 cmH
2
O 30 cmH
2
O ±3.8 cmH
2
O
If the measured pressure values are not within tolerance, perform a pressure calibration/adjustment
of the ventilator (see “Pressure and Flow Calibration” on page 46).
2.7.3
Checking the Tidal Volume Indication
1
Adjust the settings as follows:
S
ETTING
Mode
IPAP
EPAP
Set Rate
Inspiration Time
V
ALUE
PCV
30 cmH
2
O
4 cmH
2
O
8 bpm
5.0 s
2
3
Connect a ventilator tester to the ventilator.
Measure and check that the measured value for volume is correct (tolerance ±20%).
If the measured volume value is not within tolerance, perform a volume calibration/adjustment of
the ventilator (see “Pressure and Flow Calibration” on page 46).
2.7.4
Preparing the Ventilator for Alarm Checks
1
Connect the ventilator to the mains power supply and switch it on by pressing the On/Off button on the side panel.
The ventilator should be in stand-by mode (not operating) and the mains LED panel should be lit.
on the front
Do not connect the patient circuit.
Doc. 003823 En Z-1
Maintenance Instructions
Breas Vivo 30/40 Service Manual
13
2
Adjust the settings as follows:
S
ETTING
Mode
IPAP
EPAP
Set Rate
Inspiration Time
Rise Time
Inspiration Trig
Low Pressure Alarm
Low Volume Alarm
High Breath Rate Alarm
(Vivo 40 only)
Low Breath Rate Alarm
(Vivo 40 only)
V
ALUE
PCV
14 cmH
2
O
2 cmH
2
O
10 bpm
1.5 s
4
Off
2 cmH
2
O
0.05 l
Off
4 bpm
2.7.5
Checking the Low Leakage Alarm and High Leakage Alarm
1
Connect a test lung or a reservoir bag to the ventilator.
Leakage
4 mm hole
Test lung
2
Enter operating mode by switching the ventilator on with the Start/Stop button.
3
Block the leakage hole.
The low leakage Alarm shall occur within 60 seconds.
4
Disconnect the hose from the air outlet.
The low leakage Alarm shall disappear and the high leakage alarm shall occur within 30 seconds.
2.7.6
Checking the High Pressure Alarm
1
2
Enter stand-by mode by switching the ventilator off with the Start/Stop button.
Connect a patient circuit to the ventilator.
3
Enter operating mode by switching the ventilator on with the Start/Stop button.
4
Create pressure towards the ventilator by blowing air into the mask or patient tube.
The high pressure alarm shall be activated after 3 consecutive high pressure breaths.
14 Maintenance Instructions
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
2.7.7
Checking the Inspiratory Trigger
1
Set the inspiratory trigger to 3.
2
Create a spontaneous breath (inspiratory flow) and make sure a triggered breath is given.
The green Inspiratory Trigger (T) LED shall be lit.
2.7.8
Checking the Low Pressure Alarm
1
2
Enter stand-by mode by switching the ventilator off with the Start/Stop button.
Connect the patient circuit to a test lung (<1.5 l) and a leakage connector.
3
Adjust the settings as follows:
S
ETTING
IPAP
Low Pressure Alarm
V
ALUE
12 cmH
2
O
10 cmH
2
O
4
Enter operating mode by switching the ventilator on with the Start/Stop button.
5
Adjust the IPAP to 5 cmH
2
O.
The low pressure alarm shall occur within 30 seconds.
6
Adjust the IPAP to 12 cmH
2
O again.
The low pressure alarm shall disappear.
2.7.9
Checking the High Breath Rate Alarm (Vivo 40 only)
1
Adjust the settings as follows:
S
ETTING
IPAP
EPAP
Set Rate
High Breath Rate Alarm
The high breath rate alarm shall occur within 30 seconds.
2
Adjust the set rate to 5 bpm.
The high breath rate alarm shall disappear within 30 seconds.
2.7.10 Checking the Low Breath Rate Alarm (Vivo 40 only)
V
ALUE
15 cmH
2
O
10 cmH
2
O
20 bpm
10 bpm
1
Adjust the settings as follows:
S
ETTING
Set Rate
High Breath Rate Alarm
Low Breath Rate Alarm
The low breath rate alarm shall occur within 30 seconds.
2
Adjust the set rate to 20 bpm.
The low breath rate alarm shall disappear in 30 seconds.
V
ALUE
5 bpm
Off
15 bpm
Doc. 003823 En Z-1
Maintenance Instructions
Breas Vivo 30/40 Service Manual
15
2.7.11 Checking the Low Volume Alarm
1
Adjust the settings as follows:
S
ETTING
Low Volume Alarm
V
ALUE
2.0 l
The low volume alarm shall occur within 20 seconds.
2
Adjust the low volume alarm to 0.05 l.
The low volume alarm shall disappear within 30 seconds.
2.7.12 Checking the Audio Pause
1
Switch on the ventilator. Disconnect the patient circuit.
The low pressure alarm shall occur in 15 seconds.
2
Press the audio pause button and make sure the audible alarm is paused.
3
Check that the audible alarm starts again after approximately 1 minute.
2.7.13 Checking External Battery Operation
1
2
Start the ventilator.
Connect an external battery to the ventilator.
3
Disconnect the mains power cord while the ventilator is running.
4
Check that the ventilator automatically switches over to external battery operation.
The following shall occur:
• A short audible alarm sounds (Vivo 30 only).
• The message “Changed Power Source Switched to External Battery” is shown in the display.
• The power source LED is changed to external DC.
5
Reconnect the power cord and check that the power source LED is changed to mains.
2.7.14 Checking the Internal Battery Operation (Vivo 40 only)
1
Start the ventilator.
2
Disconnect any external battery from the ventilator.
3
Disconnect the mains power cord while the ventilator is running.
4
Check that the ventilator automatically switches over to internal battery operation.
The following should occur:
• A short audible alarm sounds.
• The message “Changed Power Source Switched to Internal DC” is shown in the display.
• The power source LED is changed to internal DC.
5
Reconnect the power cord and check that the power source LED is changed to mains.
2.7.15 Adjusting the Settings for the Patient
Adjust the settings as prescribed for the patient.
16 Maintenance Instructions
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
2.8
Electrical Safety Precautions
Electrical safety measurements must be made in accordance with IEC 601.
Use an automatic electrical safety tester to make the measurements. All tests must be performed in accordance with class II type BF.
Supply Voltage
Note the power voltage reading.
The voltage must be noted at each service check, as the currents measured are directly in relation to the supply voltage. This allows all measurements made on the same ventilator to be compared with measurements made on different occasions.
Insulation
The insulation resistance is measured using a 500 V DC power supply. The most suitable method is to connect the plus lead to the two ventilator power socket pins, and the minus lead to the casing or the patient air connector. The measurements made during the delivery inspection can be used as reference values for measurements made during future services. If no reference values are available, the value for the insulation resistance should be >20 M
Ω
.
Leakage Currents
The leakage currents are measured at different parts of the ventilator using an RC circuit to earth.
Make the measurements partly at normal case (NC) and at the single fault condition (SFC). Reverse the polarity of the power supply and note the highest value.
Leakage currents to earth must not exceed the stated limit values.
Leakage Currents from the Casing
The leakage current of the casing is measured at an unpainted point, for example, the head of a screw.
Limit values: NC <0.1 mA
SFC <0.5 mA
Break neutral for SFC.
Patient Leakage Currents
The patient leakage current is measured between the patient connector and earth.
Limit values: NC <0.1 mA
SFC <0.5 mA
Break neutral for SFC.
Leakage Currents with Mains Power Supply at the Patient-connected Part
This test must be done using an automatic electrical safety tester with this function. See the safety instructions for the tester.
Limit value: SFC <5 mA
Doc. 003823 En Z-1
Maintenance Instructions
Breas Vivo 30/40 Service Manual
17
2.9
Power Failure Alarm Check
2.9.1
Vivo 30
1
Start the Vivo 30 with the power cord connected.
2
Disconnect the power cord.
The power failure alarm shall occur immediately.
2.9.2
Vivo 40
1
Remove the internal battery as in “Opening the Vivo 30/40” on page 34, 1-2.
2
Start the Vivo 40 with the power cord connected.
3
Disconnect the power cord.
The power failure alarm shall occur immediately.
18 Maintenance Instructions
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
3 Parts Location
This chapter contains part-number lists and drawings of the parts for the ventilator.
3.1
Parts Drawing 1 – Main Components
Accessory:
Humidifier
HA 01
7
(Vivo 30)
8
(Vivo 40)
9 5
14
3-1
(4x)
2-5
(2x)
Vivo 30
3-4
(2x)
Vivo 40
13
16
(Vivo 30)
17
(Vivo 40)
10
(Vivo 40 only)
Doc. 003823 En Z-1
Parts Location
Breas Vivo 30/40 Service Manual
19
3.2
Parts Drawing 2 – Vivo 30 Back Casing, Cables, Screws
9-1 9
9-3 9-4
1-3 2-4
(2x)
2-5
20-2
(2x)
20-1
19
21
2-8
(2x)
16
18
4-10
4-11
2-2
2-6
2-9
2-7
1-2
1-1
20 Parts Location
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
3.3
Parts Drawing 3 – Vivo 40 Back Casing, Cables, Screws
9-1
9
9-3 9-4
1-3 3-5
(2x)
3-6
20-2
(2x)
20-1
19
21
3-9
(2x)
17
18
4-10
4-11
3-2
3-7
3-10
3-8
1-2
1-1
3.4
Parts Drawing 4 - Blower Assembly
9-3
9-4
Doc. 003823 En Z-1
9-2
4-10
4-11
Parts Location
Breas Vivo 30/40 Service Manual
21
3.5
Parts Drawing 5 – Batteries, Tubings
3.5.1
CPU Boards revision 9 and earlier
6-1
6-2
9-4
CPU Board revision number
4-1
9-3
4-7
(2x)
4-2 4-4
4-3
4-5
4-6
22 Parts Location
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
3.5.2
CPU Boards revision 10 and later
6-1
6-2
9-4
CPU Board revision number
9-3
4-7
(2x)
4-2 4-4 4-3 4-12
Doc. 003823 En Z-1
Parts Location
Breas Vivo 30/40 Service Manual
23
3.6
Parts Drawing 6 – Vivo 30 Front Casing
7-2
7-4 7-1
7-3
11 2-3
(4x)
15
24 Parts Location
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
3.7
Parts Drawing 7 – Vivo 40 Front Casing
8-2
8-4 8-1
8-3
12 3-3
(4x)
15
Doc. 003823 En Z-1
Parts Location
Breas Vivo 30/40 Service Manual
25
3.8
Parts List for the Vivo 30 and Vivo 40
The parts of the Vivo 30 and Vivo 40 are listed in the table below.
K
IT
N
O
.
D
ESCRIPTION
3-10
4
4-1
4-2
4-3
4-4
3-6
3-7
3-8
3-9
3-2
3-3
3-4
3-5
2-8
2-9
3
3-1
2-4
2-5
2-6
2-7
2
2-1
2-2
2-3
1
1-1
1-2
1-3
Cable kit –Vivo 30/40
Cable AC input PSU
Cable AC connection - switch
Cable DC input
Screw kit – Vivo 30
Screw MRT M4x10 FZB
M4- Serrated washer
Screw MRT M3x6 FZB
Screw PT WN 1413 KA 30x8
Screw Tx20 M4x20 FZB
Spacer M4x8
Knurled nut
Screw STS-KN1039 TX T30x10 FZB
Cable restrainer
Screw kit - Vivo 40
Screw MRT M4x10 FZB
M4 Serrated washer
Screw MRT M3x6 FZB
Screw Tx20 M4x30 FZB
Screw PT WN 1413 KA 30x8
Screw Tx20 M4x20 FZB
Spacer M4x8
Knurled nut
Screw STS-KN1039 TX T30x10 FZB
Cable restrainer
Tube kit CPU - Vivo 30/40
Tube 1 silicone 40 mm (yellow)
Tube 2 silicone 85 mm (yellow)
Tube 3 silicone 60 mm (white)
Tube 4 silicone (grey)
B
REAS
P
ART NO
.
003567
A
MOUNT
1
1
1
003568
1
2
3
1
1
4
2
4
1
004012
1
1
2
1
2
1
4
2
4
1
003533
1
1
1
1
26 Parts Location
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
K
IT
N
O
.
D
ESCRIPTION
9
9-1
9-2
9-3
8-1
8-2
8-3
8-4
9-4
10
11
7-2
7-3
7-4
8
6-1
6-2
7
7-1
5
6
4-11
4-12
4-5
4-6
4-7
4-10
Tube 5 silicone 45 mm (white)
Tube 6 silicone (blue)
Tube connector
Drain-tube 235 mm (white)
Drain-tube 185 mm (white)
Tube 12 silicone 75 mm (white)
Filter carrier
Battery kit - Vivo 30/40
Battery, alarm
Battery, CR2032 3 V 230 mAh, lithium
Case front kit - Vivo 30
Window Vivo 30
Overlay logotype Vivo
Keypad
Case front
Case front kit - Vivo 40
Window Vivo 40
Overlay logotype Vivo
Keypad
Case front
Blower complete - Vivo 30/40
Tube inlet
Tube outlet
Tube sense 2
Tube sense 1
Internal battery complete - Vivo 40
PCB CPU card - Vivo 30 (Language)
PCB CPU card Vivo 30DA
PCB CPU card Vivo 30 DE
PCB CPU card Vivo 30 EL
PCB CPU card Vivo 30 EN
PCB CPU card Vivo 30 ES
PCB CPU card Vivo 30 FI
Doc. 003823 En Z-1
003625
003561
003628
003620
003622
003562
B
REAS
P
ART NO
.
003465
004007
A
MOUNT
2
1
1
1
1
1
1
1
1
003565
1
1
1
1
003566
003524
003452
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Parts Location
Breas Vivo 30/40 Service Manual
27
15
16
13
14
17
18
K
IT
N
O
.
D
ESCRIPTION
12
PCB CPU card Vivo 30 FR
PCB CPU card Vivo 30 IT
PCB CPU card Vivo 30 JA
PCB CPU card Vivo 30 NL
PCB CPU card Vivo 30 NO
PCB CPU card Vivo 30 PL
PCB CPU card Vivo 30 PT
PCB CPU card Vivo 30 SV
PCB CPU card Vivo 30 US
PCB CPU card Vivo 30 ZH
PCB CPU card - Vivo 40 (Language)
PCB CPU card Vivo 40 DA
PCB CPU card Vivo 40 DE
PCB CPU card Vivo 40 EL
PCB CPU card Vivo 40 EN
PCB CPU card Vivo 40 ES
PCB CPU card Vivo 40 FI
PCB CPU card Vivo 40 FR
PCB CPU card Vivo 40 IT
PCB CPU card Vivo 40 JA
PCB CPU card Vivo 40 NL
PCB CPU card Vivo 40 NO
PCB CPU card Vivo 40 PL
PCB CPU card Vivo 40 PT
PCB CPU card Vivo 40 SV
PCB CPU card Vivo 40 US
PCB CPU card Vivo 40 ZH
Plug memory card - Vivo 30/40
Air outlet patient assembly - Vivo 30/40
Beeper + cable - Vivo 30/40
PCB PSU card Vivo 30
PCB PSU card Vivo 40
Fuse T2A, mains
28 Parts Location
Breas Vivo 30/40 Service Manual
B
REAS
P
ART NO
.
003626
003623
003630
003624
003621
003629
003627
003560
003631
003632
A
MOUNT
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
10
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
003897
003640
003637
003645
003643
003633
003898
003899
003641
003634
003644
003636
003638
003635
003642
003639
003542
003591
003509
003492
003602
003900
Doc. 003823 En Z-1
K
IT
N
O
.
D
ESCRIPTION
21
21-1
21-2
21-3
21-4
19
20
20-1
20-2
Fuse T8A, external DC
Lock device kit
Lock device
Spring conical
Cooling fan kit
Elina fan
Fan bracket
Screw MRX-H M2,5x25
Nut M2,5 DIN985
B
REAS
P
ART NO
.
003901
004289
A
MOUNT
10
1
2
004827
2
2
1
1
Doc. 003823 En Z-1
Parts Location
Breas Vivo 30/40 Service Manual
29
4 Functional Diagrams
This chapter contains a diagram of the pneumatic system of the ventilator and a block diagram of the Vivo 30/40’s functions.
The functional block diagram below shows how the electronics and pneumatics are designed and how they are connected to the other components.
4.1
Vivo 30/40 with magnetic valve (CPU board rev 9 and earlier)
Charger
LCD
Display controller
Power control
Mains
Buttons
Int DC
Ext DC
LED
Blower control
Control logic
Step-up conv.
Alarm batt.
Alarm ctrl.
Memory card
Com port analog out
Valve control
Pressure transducer
Flow transducer
Pressure transducer
1
8
Blue
2
4
3
5
6
Grey
Pink
7
Blue
Yellow
13
Pink
6
9
Magnetic valve MV1 (13):
Not powered Powered
Electrical connection
Pneumatic connection
30 Functional Diagrams
Breas Vivo 30/40 Service Manual
10
11
12
Doc. 003823 En Z-1
The table below describes the components of the pneumatic diagram.
N
O
.
7
8
5
6
3
4
1
2
11
12
9
10
13
D
ESCRIPTION
Air inlet filter
Turbine
Silencer box
Vivo 30/40 casing
Pressure regulating valve
Pressure sensor
Flow sensor
Ambient
Humidifier
Patient tube
Leak holes
Mask
Magnetic valve MV1
Doc. 003823 En Z-1
Functional Diagrams
Breas Vivo 30/40 Service Manual
31
4.2
Vivo 30/40 without magnetic valve (CPU board rev 10 and later)
Charger
LCD
Display controller
Power control
Mains
Int DC
Buttons
Ext DC
LED
Blower control
Control logic
Step-up conv.
Alarm batt.
Alarm ctrl.
Memory card
Com port analog out
Valve control
Pressure transducer
Flow transducer
Pressure transducer
1
8
2
4
3
5
6
Grey
Pink
7
Blue
Yellow
6
Pink
9
10
11
12
Electrical connection
Pneumatic connection
32 Functional Diagrams
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
The table below describes the components of the pneumatic diagram.
N
O
.
7
8
5
6
3
4
1
2
11
12
9
10
D
ESCRIPTION
Air inlet filter
Turbine
Silencer box
Vivo 30/40 casing
Pressure regulating valve
Pressure sensor
Flow sensor
Ambient
Humidifier
Patient tube
Leak holes
Mask
4.3
Filtering/Smoothing Techniques
F
UNCTION
Pressure
Inspiration trigger
Expiration trigger
T
ECHNIQUE DESCRIPTION
Low pass average time constant 16 ms
Differential mass flow resolution 4 ms
Flow low pass filtering with level sensing
4.4
Vivo 30/40 Measuring and Display Devices
D
EVICE
P
URPOSE
T
YPE
Pressure sensing
Flow sensing
Pressure display
Regulate pressure
Regulate pressure
Indicate patient pressure
Differential to ambient
Mass flow
Graphic
LCD
R
ANGE
-10 to
+60 cmH
2
O
0 to 300 l/min 1 l/min
0 to 40 cmH
2
(Vivo 40)
0 to 30 cmH
(Vivo 30)
2
O
O
R
ESOLUTION
A
CCURACY
0.1 cmH
1 cmH
2
2
O ±0.1% FSS
O
±1 l/min
S
ENSING
P
OSITION
Air outlet of
Vivo 30/40
Air outlet of
Vivo 30/40
±0.5 cmH
2
O Air outlet of
Vivo 30/40
Doc. 003823 En Z-1
Functional Diagrams
Breas Vivo 30/40 Service Manual
33
5 Opening the Vivo 30/40 and Replacing the Main
Components
Make sure to disconnect the power supply before removing the casing of the ventilator.
5.1
Opening the Vivo 30/40
• Always perform a complete function test after reassembling the Vivo 30/40, see “Complete
• Always perform an electrical safety test after reassembling the Vivo 30/40, see “Electrical
Safety Precautions” on page 17.
• Make sure that the Vivo 30/40 is placed on a non-scratching surface.
• Make sure that the tubes are not pinched or kinked when closing the Vivo 30/40.
1
Remove the two screws for the base plate
(Vivo 30) or the internal battery (Vivo 40).
2
Remove the base plate (Vivo 30) or the internal battery (Vivo 40) by pulling it slightly backwards and down.
34 Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
3
Remove the four screws for the casing.
4
Gently fold the front casing forward.
5
Reassemble in reverse order.
• Make sure that the hooks on the Vivo 40 internal battery is properly fitted in the Vivo 40 back casing.
5.2
Replacing the Complete Blower Assembly
1
2
Open the Vivo 30/40 as in section 5.1.
Disconnect the cabling for the flow valve.
Doc. 003823 En Z-1
Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
35
3
Disconnect the yellow and the grey tubing at the T-connectors.
4
Loosen the lower right screw on the blower assembly to remove the small L-shaped board.
5
Push the tube outlet through the hole at the rear of the Vivo 30/40.
36 Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
6
out.
Pull the complete blower assembly straight
7
Disconnect the two drainage tubes from the rear side of the blower assembly.
8
Disconnect the BLDC motor control cable.
Doc. 003823 En Z-1
Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
37
9
Cut the cable tie securing the blower cabling connector and disconnect it.
10
Reassemble in reverse order.
• Make sure that a new cable tie is securing the blower cabling connector. You need to remove the
PSU board as in section 5.3 to replace the cable tie.
• Make sure you fit the tube outlet to the blower assembly, pointing slightly upwards. Gently pull the tube outlet through the hole at the rear of the Vivo 30/40.
• When reinserting the blower into the back casing, make sure that the cabling does not disrupt the PSU cooling fan air ways. Do this by pulling the cablings out and to the right when inserting the blower unit.
5.3
Replace the PSU Board
1
2
Remove the blower assembly as in section 5.2.
Disconnect the external DC connector.
38 Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
3
Disconnect the AC supply connector.
4
Remove the two screws holding the PSU board.
5
Remove the ground cable.
6
Remove the PSU board.
Be careful so you do not damage the Preci-dip contacts at the back of the PSU board
(Vivo 40 only).
Doc. 003823 En Z-1
7
Reassemble in reverse order.
Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
39
5.4
Replace the PSU Cooling Fan
1
Remove the PSU board as in section 5.3.
2
Disconnect the cooling fan from the PSU board.
3
fan.
Remove the two nuts holding the cooling
4
Remove the cooling fan.
5
Remove the fan bracket.
You might have an earlier revision of the PSU cooling fan that do not have any fan bracket.
If so, you shall install a bracket when replacing the PSU cooling fan to improve the
Vivo 30/40 cooling air flow.
6
Reassemble in reverse order.
40 Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
• Make sure that the opening in the fan bracket is facing towards the heatsinks.
• Make sure that the cooling fan is placed with the text upwards.
5.5
Replace the CPU Board
The Vivo 30/40 serial number needs to be programmed into the new CPU board.
You need the Vivo 30/40 Service Software to program the CPU Board with the correct serial number after replacement.
1
2
Open the Vivo 30/40 as in section 5.1.
Disconnect the cabling for the flow valve.
3
Disconnect the yellow and the grey tubing at the T-connectors.
Doc. 003823 En Z-1
Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
41
4
Disconnect the beeper.
5
Disconnect the two ribbon cables by loosening the connectors.
• When reconnecting the two ribbon cables - be sure that they are fitted straight, otherwise there is risk of a short-circuit over the connectors.
• Make sure that the ribbon cables are properly fastened.
6
Remove the four screws and lift the CPU board straight up.
7
Reassemble in reverse order.
42 Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
5.6
Replace the Clock Battery
Save the patient settings to a CF card before changing the clock battery. It will then be easier to restore all the patient settings after the procedure.
Study the Vivo 30/40 clinician’s manual for instructions on how to use the CF card.
1
Open the Vivo 30/40 and disconnect the flow valve cabling and the tubings as in
2
Remove the clock battery by pulling it straight up. Press a new clock battery into place.
Make sure that the new battery is not handled with bare hands.
3
4
Reassemble in reverse order.
Restore the patient settings.
Doc. 003823 En Z-1
Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
43
5.7
Replace the Alarm Battery
1
2
Open the Vivo 30/40 and remove the CPU board as in section 5.5.
Remove the jumper.
+
3
Unsolder the two pins on the alarm battery and remove the battery.
+
4
Remove any remaining solder from the holes. Fit a new battery, checking the polarity, and solder it in place.
+
5
Put the jumper back.
+
6
Reassemble in reverse order.
44 Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
5.8
Replacing the Push Button Silicone Pad
1
2
Remove the CPU board as in section 5.5.
Remove the beeper.
3
Lift the push button silicone pad.
4
Reassemble in reverse order.
• Make sure that the silicon push pad’s corner fits properly into the beeper holder.
Doc. 003823 En Z-1
Opening the Vivo 30/40 and Replacing the Main Components
Breas Vivo 30/40 Service Manual
45
6 Upgrade and Calibration
6.1
Firmware Upgrade
To upgrade the Vivo 30/40 firmware you need the “Firmware Upgrade Tool iSleep/Vivo”.
Contact Breas technical support for more information.
6.2
Pressure and Flow Calibration
To calibrate the Vivo 30/40 you need the “Vivo 30/40 Service Software”.
Contact Breas technical support for more information.
46 Upgrade and Calibration
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
7 Electronics
Always perform a complete function test if you have opened the Vivo 30/40.
The electronics, optics, mechanics, and pneumatics of the Vivo 30/40 are integrated. To understand fully the electronics of the Vivo 30/40, you must know how to use the ventilator, study the pneumatic diagram and acquaint yourself with the mechanical construction.
7.1
Main Cabling Diagram of the Vivo 30/40
The diagram below illustrates the main cabling of the Vivo 30/40.
Humidifier
X2
Blower assembly
PSU
Brown
Yellow
Purple
Ribbon cable
Hall 1
Hall 2
GND
Hall 3
NTC
+6 V
1
1
X3
X10
Brown
Brown
Mains inlet
External battery inlet
1 + Red
2 - Black
X1
X5 X7 X8
X9
X6
Battery charger
X3
X6
X5
CPU
X3
X10
Red
Black
X7
1
Orange
Blue
Yellow
Red
B
B
A
A
Stepper motor
X4
LCD
Palm
Compact flash
Internal battery
Vivo 40 only
7.2
Circuit Board Descriptions
This section contains descriptions and figures of the circuit boards, the LCD display, and the pushbutton membrane panel of the Vivo 30/40.
Doc. 003823 En Z-1
Electronics
Breas Vivo 30/40 Service Manual
47
7.2.1
PSU Board
The diagram below is an overview of the PSU board
CPU connector Mains connector
Mains flyback converter
Power source selection
CPU board interface
12 V generator
Humidifier interface
DC voltage converter
Internal battery interface
Current control
BLDC motor controller
Ext DC connector
Int DC connector
Humidifier connector
BCU connector BLDC motor connector
PSU board contains the switch mode power supply (SMPS) built on the double ended flyback topology. The converter is powered with mains 100-240 V AC, +10%, -20% (i.e. 80 V AC-264 V AC).
Power Source Selection
There are three power inputs (two for Vivo 30), Mains, External DC (EXT DC) and internal battery
(INT DC, Vivo 40 only).
The three power sources are controlled by the main processor on the CPU board. The processor senses status for each source to decides if to disable mains and enable any DC source.
48 Electronics
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
Voltage Converter
The converter is a boost converter that steps up EXT DC or INT DC to +30 V DC, the nominal value for net VM (Voltage Motor). When Vivo 30/40 is running on mains, the DC step up converter is disabled because of the already +30 V DC from the SMPS.
Brush less DC (BLDC) Motor Controller
The BLDC amplifier is based on a three-phase inverter controlled by a MC33035 motor control circuit. The main processor regulates the speed with a 0-5 V analogue signal into the MC33035P. The main processor can also brake the BLDC motor via the control circuit. To determine if the fan is running, the main processor senses a hall element from the BLDC motor. The control circuit also has a built-in over current protection.
Internal Battery Pack Connector (Vivo 40 only)
Vivo 40 is equipped with an internal battery pack, accessed from the PSU-board via 2x6 pole
Preci-dip connectors. The battery pack contains its own charging processor and either supply
Vivo 40 with power, or gets charging power from Vivo 40.
The main unit supplies the battery pack with +5 V from the CPU-board through this board.
HA 01 Humidifier Interface
The Vivo 30/40 can be equipped with an integrated HA 01 humidifier.
The HA 01 humidifier consumes 30-40 W depending on feeding voltage.
The power to the HA 01 humidifier is controlled from the main CPU via a transistor.
The other two signals in the interface beside power and ground are an analogue temperature signal from the HA 01 humidifier and a control signal to the HA 01 humidifier.
Doc. 003823 En Z-1
Electronics
Breas Vivo 30/40 Service Manual
49
7.2.2
CPU Board
The diagram below is an overview of the CPU board
Alarm battery
Display
Buttons
LEDs
UI processor
Display controller
Reset
V
ref
Beeper
Pressure sensor
Reset
Memory
Flow sensor
Main processor
Pressure sensor
Serial interface
Analog outputs
Stepper motor amplifier
Real time clock
Compact flash memory interface
PSU connector
RJ45 connector
Stepper motor connector
50 Electronics
Breas Vivo 30/40 Service Manual
Compact flash memory connector
Main Processor and User Interface (UI) Processor
The processing core of the main unit is two M16C62P Microcontrollers.
The main processor has a supply voltage of +5 V and is not operating when Vivo 30/40 is shut of.
The UI processor is supplied with both +5 V (signal side) and +3,3 V (data side).
The UI processor is battery backed up and will operate in sleep mode when the Vivo 30/40 is switched off.
Compact Flash Memory Interface
The main processor can transfer the data log from the memory to the compact flash memory.
Doc. 003823 En Z-1
Display
The display used is a graphic type and have a visible area of approx. 60x80 mm, with a resolution of
240x320 pixels.
To drive the display a display controller is used.
RJ45 Connector
To communicate with a PC, the Vivo 30/40 uses a RS232 interface.
Pressure Sensors
Vivo 30/40 is equipped with two pressure sensors for redundancy. The main processor gets input from both pressure sensors, but the UI processor only gets input from one of the sensors. This means that both processors supervise the actual pressure and have the availability to stop the blower in case of an error condition.
Mass Flow Sensor
The mass flow sensor is used for air flow measurements.
Alarm Beeper
The alarm beeper is a 25 mm plastic membrane loudspeaker, powered by a 1 W bridge-connected audio amplifier. The alarm is generated by hardware and either of the two CPU:s can start and stop the alarm.
Stepper Motor Amplifier
The Vivo 30/40 uses a valve for controlling the exhalation phase. This valve is controlled by a stepper motor which is controlled by the main processor.
Voltage Regulators
The CPU board contains voltage regulators for 3,3 V, 5 V, 8 V, 10 V and 24 V.
Doc. 003823 En Z-1
Electronics
Breas Vivo 30/40 Service Manual
51
7.3
Test Points – CPU board
7.3.1
Test Point Locations
The locations of the test points for the CPU board are shown in the figure below.
The white test points are for production use only.
52 Electronics
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
7.3.2
Test Point Matrix
The following are the test points on the CPU board of the Vivo 30/40.
TP1
TP2
TP3
TP4
TP5
TP6-7
TP8
TP9
TP10
T
EST
POINT
TP11
TP12
TP13
TP14
TP15
TP16
TP17
TP18
TP19
TP20-39
V
OLTAGE
30 V
5.5 V
3.3 V
3 V
5 V
–
0–10 V
0–10 V
3.3 V
10 V
0–5 V
3.3 V
5.5 V
5.5 V
T
OLER
ANCE
D
ESCRIPTION
P
OWER
SUPPLY
WHEN
MEASURE
M
ACHINE
STATUS
±
2 V VM Power good signal for U1
For production use
±
1 V 5 V/LA
For production use
±
0.5 V +3.3 V
For production use
±
0.3 V Lithium battery CR2032
±
1 V Reset signal for U15
±
±
0.5 V
0.6 V
Pressure sensor 1 amplifier output:
0 mbar = 0.5 V
30 mbar = 2 V
Mains
Mains
Mains
No supply
Mains
Mains
±
±
±
±
0.15V
0.2 V
0.15V
0.2 V
Flow sensor amplifier (FlowL) output:
0 l/m = 2.5 V Adjust with P4
50 l/m = 3 V Adjust with P1
Be sure to reseal the potentiometer after adjusting
Flow sensor amplifier (FlowH) output:
0 l/m = 2.5 V Adjust with P3
50 l/m = 4 V Adjust with P2
Be sure to reseal the potentiometer after adjusting
±
0.5 V Regulated 3.3 V
Mains
Mains
Mains
±
1 V Regulated 10 V for flow sensor ampl.
Mains
Control signal for magnetic valve.
5 V=powered
Mains
±
0.5 V
±
0.6 V
Pressure sensor 2 amplifier output:
0 mbar = 0.5 V
30 mbar = 2 V
Mains
±
±
0.5 V Regulated 3.3 V
1 V 5 V/LA Reset signal for U1
±
1 V VM divided for power good reset circuit(U2)
For production use
Mains
Mains
Mains
Stand by
Stand by
Stand by
Off
Stand by
Stand by
On
On
Stand by
Stand by
On
Stand by
Stand by
Stand by
Stand by
Doc. 003823 En Z-1
Electronics
Breas Vivo 30/40 Service Manual
53
TP50
TP51
TP52
TP53
TP54
TP55
TP56
TP57
TP58
TP59
TP60
TP61
T
EST
POINT
TP40
TP41-47
TP48
TP49
TP62
TP63
TP64
TP65
TP66
TP67
TP68-73
TP74
TP75
TP76
TP77
TP78
V
OLTAGE
T
OLER
ANCE
D
ESCRIPTION
P
OWER
SUPPLY
WHEN
MEASURE
M
ACHINE
STATUS
12 V
25 V
4 V
±
1 V +12 V Regulated 12 V
For production use
±
5 V Contrast voltage for LCD
±
2 V Backlight voltage depending on intensity
Mains
Mains
Mains
Stand by
Stand by
Stand by
0–4 V
0–4 V
5 V
30 V
5 V
12–30 V
0 V
0 V
0 V
0–5 V
3.3 V
5 V
24 V
0–5 V
0–5 V
5.5 V
5 V
5 V
12 V
±
±
0 or 5 V –
0 or 5 V –
±
±
PWM (not used)
PWM (Pam LCD)
0.5 V Regulated 5 V
2 V VM Regulated 30 V
Mains
Mains
Mains
Mains
±
0.5 V +5 V Regulated 5 V
VCC CPU. Depending on power source Mains/DC On
EXT BAT
Stand by
Stand by
Stand by
Stand by
Ground
Ground
Ground
For production use
Beeper current measure voltage Mains
Mains
On.
beeping
Stand by
±
0.5 V Regulated 3.3 V
±
0.1 V Reference voltage for A/D in U15
±
2 V Regulated 24 V for valve motor
Beeper amplifier speaker output
Beeper amplifier speaker output
Mains
Mains
Mains
Mains
Stand by
On, beeping
On, beeping
Stand by
±
1 V + 5 V Regulator (CPU BAT)
For production use
±
1 V 5 V when powered. 3 V when off
0.5 V Regulated 5 V
1 V Regulated 12 V
Mute 0 V if pushed
Soft key 1 button. 0 V if pushed
Mains/
No mains
Mains
Mains
Mains
Mains
Stand by/
Off
Stand by
Stand by
Stand by
Stand by
54 Electronics
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
T
EST
POINT
TP79
TP80
TP81
TP82
TP83
TP84
TP85
TP86
V
OLTAGE
T
OLER
ANCE
D
ESCRIPTION
0 or 5 V –
0 or 5 V –
0 or 5 V –
0 or 5 V –
0 or 5 V –
0 or 5 V –
0 or 5 V –
0 or 5 V –
Soft key 2 button. 0 V if pushed
Soft key 3 button. 0 V if pushed
Soft key 4 button. 0 V if pushed
Up button. 0 V if pushed
Down button. 0 V if pushed
On/Off button. 0 V if pushed
Plus button. 0 V if pushed
Minus button. 0 V if pushed
P
OWER
SUPPLY
WHEN
MEASURE
Mains
Mains
Mains
Mains
Mains
Mains
Mains
Mains
M
ACHINE
STATUS
Stand by
Stand by
Stand by
Stand by
Stand by
Stand by
Stand by
Stand by
7.4
Ventilator Switch-over Operating Voltages
The Vivo 30/40 will issue alarm and switch between the various power sources available if any of the conditions described below occur.
7.4.1
Mains Power Supply Operation
If a voltage drop occurs while running from an mains power supply the ventilator will react as follows:
If the mains drops below 65 ±15 V AC the Vivo 30/40 will issue a low power alarm or switch over to the external battery, if available, or internal battery (Vivo 40 only).
7.4.2
External Battery Operation
24 V battery
Vivo 30:
If the ventilator is running from an external battery and the voltage drops below 22.7 V the ventilator will issue a low external DC warning and continue to run until the voltage drops below
20 V. The ventilator will then issue a power-failure alarm and switch off.
Vivo 40:
If the ventilator is running from an external battery and the voltage drops below 20 V the ventilator will switch to the internal battery supply.
When the external battery voltage returns to 24 V the ventilator will switch back to the external battery supply. The external battery voltage is checked continuously.
12 V battery
Vivo 30:
If the ventilator is running from an external battery and the voltage drops below 11.3 V the ventilator will issue a low external DC warning and continue to run until the voltage drops to 10 V.
The ventilator will then issue a power-failure alarm and switch off.
Vivo 40:
If the ventilator is running from an external battery and the voltage drops below 10 V the ventilator will switch to the internal battery supply.
When the external battery voltage returns to 12 V the ventilator will switch back to the external battery supply. The external battery voltage is checked continuously.
Doc. 003823 En Z-1
Electronics
Breas Vivo 30/40 Service Manual
55
7.4.3
Internal Battery Operation (Vivo 40 only)
If a voltage drop occurs while running from an internal battery the ventilator will react as follows:
If the voltage drops to 16.3 V the ventilator will issue a low-battery alarm and continue to run until the voltage drops to 14 ±0.75 V. It will then issue a power-failure alarm and switch off.
7.5
Checking the Internal Battery (Vivo 40 only)
1
Make sure the battery is fully charged.
2
Adjust the settings as follows:
S
ETTING
Mode
IPAP
EPAP
Set Rate
Inspiration Time
Study the Vivo 40 Clinician’s manual for how to adjust the settings.
V
ALUE
PCV
20 cmH
2
O
8 cmH
2
O
15 bpm
3.0 s
3
4
5
Disconnect the power cord.
Connect a test lung and start the ventilator.
The ventilator shall run for at least one hour before shutdown.
56 Electronics
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
8 Fault Tracing
Always perform a Complete Function Test if you have opened the Vivo 30/40.
This chapter contains a fault-tracing table and a table of error codes to use when troubleshooting the Vivo 30/40.
8.1
Fault Tracing Table
If the Vivo 30/40 does not work properly try to identify the problem in the table below. Check the possible causes and carry out the suggested remedial actions.
S
YMPTOM
The Vivo 30/40 does not start when running from the mains supply.
P
OSSIBLE CAUSE
The power cord is not properly connected.
The On/Off button on the side panel is switched off.
The mains fuse has blown.
R
EMEDIAL ACTION
Connect the power cord.
Switch on the On/Off button.
1
Replace the fuse on the
PSU board.
2
If it blows again, replace the PSU board.
Charge the internal battery.
The Vivo 40 does not run from the internal battery supply.
The internal battery is discharged.
The Vivo 30/40 does not run from the external battery supply.
The internal battery is faulty.
Replace the internal battery.
External battery discharged.
Charge external battery.
The external battery cable is not connected properly or is faulty.
1
Connect the cable.
2
If the cable is connected, measure the voltage.
3
Replace the cable if faulty.
The external DC is not within
12–24 V DC.
The external DC fuse on the
PSU board has blown.
Check the power source and the cable polarity.
1
Replace the fuse on the
PSU board.
2
If it blows again, replace the PSU board.
Check polarity.
Battery polarity faulty.
May be the case if fuse blows immediately after connecting to external battery cable.
S
EE REF
.
5.2 in the
Vivo 30/40
Clinician’s manual
5.7 in the
Vivo 40 Clinician’s manual
5.7 in the
Vivo 30/40
Clinician’s manual
Doc. 003823 En Z-1
Fault Tracing
Breas Vivo 30/40 Service Manual
57
S
YMPTOM
Ventilator does not give the adequate pressure.
The air filters are dirty.
P
OSSIBLE CAUSE
External leaks from patient circuit or nasal mask.
Internal leaks from tubes, humidifier or air channels.
The air filters are dirty.
R
EMEDIAL ACTION
1
2
Check the tubes, connectors and mask for leaks.
Check the tubes, air channels and the humidifier.
Replace the white filter.
Wash the grey filter.
S
EE REF
.
10.2 in the
Vivo 30/40
Clinician’s manual
Pressure indicator shows no pressure reading.
The internal supply tube is blocked or loose.
The CPU board is faulty.
Check the tubes.
Replace the CPU board.
8.2
Function Failure Error Codes
8.2.1
Reading the Error Codes
The error codes listed in the table in chapter 8.2.2 “Error Code Table” will appear on the display
when occurs. The six most recent error codes and alarms are logged in the alarm history.
Access the alarm history using the menu as described in the Vivo 30/40 Clinician’s Manual.
8.2.2
Error Code Table
The table below lists each error code and the corresponding text that is shown on the LCD display.
The problem is explained together with the action that is necessary to correct the problem.
If more than one action is listed, the actions should be performed in the order in which they are listed. For example, if action no. 1 does not solve the problem you should continue with action no. 2, and so on.
See “Opening the Vivo 30/40 and Replacing the Main Components” on page 34 for information
about how to replace the circuit boards.
E
RROR
CODE
1
T
EXT ON
LCD D
ISPLAY
FUNCTION FAILURE
MAIN PROCESSOR
RAM FAIL
2
FUNCTION FAILURE
MAIN PROCESSOR
FLASH FAIL
P
ROBLEM
Ventilator fails main processor RAM selftest.
Ventilator fails main processor FLASH selftest.
A
CTION
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
58 Fault Tracing
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
E
RROR
CODE
4
5
6
10
T
EXT ON
LCD D
ISPLAY
FUNCTION FAILURE
PRESS G1 FAIL
FUNCTION FAILURE
PRESS G2 FAIL
FUNCTION FAILURE
FLOW G1 FAIL
FUNCTION FAILURE
ALARM BATTERY LOW
CONNECT TO MAINS AND
RESTART IN 30 MIN
P
ROBLEM
A
CTION
Pressure sensor (PG 1) fail.
Pressure sensor (PG 2) fail.
1
Restart the Vivo 30/40.
2
Make sure that all internal tubings are properly fastened and not kinked.
3
Check test point 40 (see chapter 7.3). If it is out of
range - replace the PSU board.
4
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Make sure that all internal tubings are properly fastened and not kinked.
3
Check test point 40 (see chapter 7.3). If it is out of
range - replace the PSU board.
4
Replace the CPU board or send the Vivo 30/40 for service.
Flow sensor (FG 1) fail.
1
Restart the Vivo 30/40.
2
Make sure that all internal tubings are properly fastened and not kinked.
3
Check test point 40 (see chapter 7.3). If it is out of
range - replace the PSU board.
4
Replace the CPU board or send the Vivo 30/40 for service.
Alarm battery status fail
(AD too low).
1
Connect the Vivo 30/40 to mains for 30 minutes.
2
Replace the alarm battery. Make sure that the new battery is fully charged.
3
Replace the CPU board or send the Vivo 30/40 for service.
Doc. 003823 En Z-1
Fault Tracing
Breas Vivo 30/40 Service Manual
59
E
RROR
CODE
13
14
15
16
17
31
32
T
EXT ON
LCD D
ISPLAY
FUNCTION FAILURE
MAIN PROCESSOR
SHUT DOWN FAIL
FUNCTION FAILURE
UI PROCESSOR
FUNCTION FAILURE
MAIN PROCESSOR
DATA FAIL
SHUT DOWN FAIL
FUNCTION FAILURE
DATE TIME
MISMATCH FAIL
FUNCTION FAILURE
BEEPER FAIL
FUNCTION FAILURE
CPU VOLTAGE FAIL
FUNCTION FAILURE
REMOTE POWER FAIL
P
ROBLEM
A
CTION
Main processor unable to turn off motor.
UI processor unable to turn off motor.
Power down time is not correct.
Memory log may be
Corrupt.
Beeper check fail.
10 V ref and 3.3 V ref is not correct.
Inverse not match
Memory log may be
Corrupt.
1
Restart the Vivo 30/40.
2
Replace the PSU board.
3
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the PSU board.
3
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40 and check all settings.
2
Replace the clock battery and restore all settings.
3
Replace the CPU board or send the Vivo 30/40 for service.
1
2
Restart the Vivo 30/40.
Replace the beeper
3
Replace the CPU board or send the Vivo 30/40 for service.
1
Check test point 40 (see chapter 7.3). If it is out of
range - replace the PSU board.
2
Replace the CPU board or send the Vivo 30/40 for service.
1
Disconnect cable from
RJ45 socket.
2
Restart the Vivo 30/40.
3
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
60 Fault Tracing
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
E
RROR
CODE
35
36
37
38
T
EXT ON
LCD D
ISPLAY
FUNCTION FAILURE
PRESSURE SENSOR
HIGH TEMP. FAIL
FUNCTION FAILURE
FAN MOTOR
OVERTEMPERATURE
FUNCTION FAILURE
MAINS TEMPERATURE
HIGH TEMP. FAIL
FUNCTION FAILURE
PATIENT
PARAMETERS FAIL
P
ROBLEM
A
CTION
Pressure sensor temperature >85°C (185°F).
Fan motor temperature
>100°C (212°F).
Main processor temperature >100°C (212°F).
Check Sum fail.
Memory log may be
Corrupt.
1
Let the Vivo 30/40 cool off and restart.
2
Check that the BLDC motor cablings does not disrupt the PSU cooling fan air way.
3
Replace the PSU board.
4
Replace the CPU board or send the Vivo 30/40 for service.
1
Let the Vivo 30/40 cool off and restart. Let the
Vivo 30/40 run for 2 hours
with the settings used in chapter 7.5. Switch to standby
mode and try to listen if the
PSU cooling fan runs. If not - replace the PSU cooling fan.
If it still doesn’t run - replace the PSU board.
2
Check that the BLDC motor cablings does not disrupt the PSU cooling fan air way.
3
Replace the complete blower assembly.
4
Replace the CPU board or send the Vivo 30/40 for service.
1
Let the Vivo 30/40 cool off and restart.
2
Check that the BLDC motor cablings does not disrupt the PSU cooling fan air way.
3
Replace the PSU board.
4
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
Doc. 003823 En Z-1
Fault Tracing
Breas Vivo 30/40 Service Manual
61
E
RROR
CODE
40
T
EXT ON
LCD D
ISPLAY
FUNCTION FAILURE
INTERFACE
COMM. ERROR
42
43
50
FUNCTION FAILURE
VALVE FUSE BROKEN
FUNCTION FAILURE
SENSOR CALIBRATION
FUNCTION FAILURE
UI PROC. RAM FAIL
51
53
54
55
FUNCTION FAILURE
UI PROC. FLASH FAIL
FUNCTION FAILURE
UI LEDS FAIL
FUNCTION FAILURE
UI KEYS FAIL
FUNCTION FAILURE
UI HIGH PRESSURE
P
ROBLEM
A
CTION
Communication fail between main processor and UI processor.
Fuse on CPU board is blown.
A mandatory calibration is needed after firmware upgrade.
RAM fail.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
Replace the CPU board or send the Vivo 30/40 for service.
1
Perform a pressure and flow calibration.
Flash fail.
LED fail.
Key fail.
Pressure fail.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Make sure that the push button silicone pad is properly fitted.
3
Replace the push button silicone pad.
4
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
tion.
Perform Pressure Calibra-
3
Make sure that all internal tubings are properly fastened and not kinked.
4
Replace the CPU board or send the Vivo 30/40 for service.
62 Fault Tracing
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
E
RROR
CODE
56
T
EXT ON
LCD D
ISPLAY
57
FUNCTION FAILURE
UI COMM. FAIL
P
ROBLEM
A
CTION
Communication fail between UI processor and LCD.
Communication fail between main processor and UI processor.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
1
Restart the Vivo 30/40.
2
Replace the CPU board or send the Vivo 30/40 for service.
Doc. 003823 En Z-1
Fault Tracing
Breas Vivo 30/40 Service Manual
63
9 Appendices
9.1
Emission and Immunity Declaration
According to IEC 60601-1-2(2001) + A1(2004).
9.1.1
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Vivo 30/40 are intended for use in the electromagnetic environment specified below. The customer or the user of the Vivo 30/40 should assure that it is used in such an environment.
Portable and mobile RF (radio frequency) communications equipment should not be used no closer to any part of the Vivo 30/40, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
I
MMUNITY
TEST
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
TEST LEVEL
3 V rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
C
OMPLIANCE
LEVEL
13 V
13 V/m
R
ECOMMENDED SEPARATION DISTANCE
d= 0,27*
√
P m d= 0.27* d= 0.54*
√
√
P m at 80 MHz to 800 MHz
P m at 800 MHz to 2.5 GHz
Equation description:
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a
, should be less than the compliance level in each frequency range b
.
Interference may occur in the vicinity of equipment marked with this symbol.
64 Appendices
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
9.1.2
Notes
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Vivo 30/40 is used exceeds the applicable RF compliance level above, the Vivo 30/40 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Vivo 30/40.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
9.1.3
Guidance and Manufacturer's Declaration – Electromagnetic Emission
The Vivo 30/40 are intended for use in the electromagnetic environment specified below. The customer or the user of the Vivo 30/40 should assure that it is used in such an environment.
E
MISSIONS TEST
C
OMPLIANCE
E
LECTROMAGNETIC ENVIRONMENT
– G
UIDANCE
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emission
IEC 61000-3-3
Group 1
Class B
Class A
Complies
The Vivo 30/40 use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Vivo 30/40 are suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Recommended separation distances between portable and mobile RF communications equipment and the Vivo 30/40.
The Vivo 30/40 are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Vivo 30/40 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Vivo 30/40 as recommended below, according to the maximum output power of the communications equipment.
Doc. 003823 En Z-1
Appendices
Breas Vivo 30/40 Service Manual
65
R
ATED MAXIMUM OUTPUT
POWER OF TRANSMITTER
(W)
1
10
0.01
0.1
100
S
EPARATION DISTANCE ACCORDING TO THE
FREQUENCY OF TRANSMITTER
(
M
)
150 kHz to
80 MHz d= 0.27*
√
P m
0.03
0.08
0.27
0.85
2.7
80 MHz to
800 MHz d= 0.27*
0.03
0.08
0.27
0.85
2.7
√
P m
800 MHz to
2.5 GHz d= 0.54*
√
P m
0.05
0.17
0.54
1.7
5.4
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
9.1.4
Notes
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
66 Appendices
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
9.2
Service Record Vivo 30/40
Use a photocopy of this service record for the maintenance inspection described in “Maintenance
Instructions” on page 7. Use the next page for comments and notes.
Service record no.__________
Model:..................................
Serial no. ..............
Inventory no. ....................................
Accessories: ...............................................................................................................................
Delivery date: .......................
Total operating hours:..............................................................
Service started: ....................
Signature:................................................................................
Service completed:...............
Signature:................................................................................
Product returned:..................
Signature:................................................................................
General See instruction ref.
Open new service record and identify ventilator
Note number of total operating hours
Check all markings
Check information from user
Check validity of documentation
External Checks
Inspect for external damage and wear
Check the HA01 humidifier (if used)
Check power connection
Inspect patient circuit
Inspect the ventilator accessories
Change/wash the patient filters
Perform minimum function check
Internal Battery (Vivo 40 only)
Every 2 years:
Replace the Internal Battery
Alarm/Clock Battery
Every 5 years:
Replace the alarm and clock battery
Check the power failure alarm
Blower Assembly
Every 20,000 total operating hours:
Replace the complete blower assembly
Internal Checks (when required)
Clean inside of ventilator
Check cabling
Check fastening of components
Check the power supply
Reassemble the casing
Continued on next page
Check OK
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
Doc. 003823 En Z-1
Appendices
Breas Vivo 30/40 Service Manual
67
Complete Function Check
Check the breath rate and inspiration time
See instruction ref.
Check the pressure
Check the tidal volume indication
Prepare the ventilator for alarm checks
Check the low/high leakage alarm
Check the high pressure alarm
Check the inspiratory trigger
Check the low pressure alarm
Check the high breath rate alarm (Vivo 40 only)
Check the low breath rate alarm (Vivo 40 only)
Check the low volume alarm
Check the audio pause
Check the external battery operation
Check the internal battery operation (Vivo 40 only) 2.7.14
Adjust settings for patient
Check OK
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
Notes
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
68 Appendices
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
9.3
Returning Products to Breas
You may need to return the ventilator, components or accessories to Breas Medical AB for service, warranty, upgrade or repair. In this case, follow the instructions below to ensure that the correct action is taken to avoid unnecessary delays.
1
Pack the product in its original packing. If this is not available, pack the product in packaging suitable for the transportation to Breas.
2
3
Take a photocopy of the delivery report on the next page.
Fill out the delivery report and pack it together with the product to be returned.
Product damage caused by poor packaging or during transportation is not covered by the factory warranty.
Doc. 003823 En Z-1
Appendices
Breas Vivo 30/40 Service Manual
69
9.4
Product Return Form – Breas Vivo 30/40 ventilator
Breas ref. no.:
...............................
Customer information
Customer name
:.................................................................................................................
Address:
.............................................................................................................................
.............................................................................................................................................
Phone:
.................................................................................................................................
Reference person:
.....................................................
Customer ref. no.:
....................
Product information
Model:
..........................
Serial no.:
.........................
Total operating hours:
.............
Error – Complaint – Accessories
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
70 Appendices
Breas Vivo 30/40 Service Manual
Doc. 003823 En Z-1
advertisement
Key Features
- Pressure-supported and controlled ventilation
- Three modes of operation: PSV, PCV, and CPAP
- Adjustable trigger sensitivity
- Expiratory sense adjustment (PSV mode)
- Log download and analysis via PC software
Frequently Answers and Questions
What are the different modes of operation for the Vivo 30/40?
How often should I perform routine maintenance on the Vivo 30/40?
What types of alarms are included in the Vivo 30/40?
Can I download and analyze data from the Vivo 30/40?
What safety precautions should I take when working with the Vivo 30/40?
Related manuals
advertisement
Table of contents
- 48 Power Source Selection
- 49 Voltage Converter
- 49 Brush less DC (BLDC) Motor Controller
- 49 Internal Battery Pack Connector (Vivo 40 only)
- 49 HA 01 Humidifier Interface
- 50 Main Processor and User Interface (UI) Processor
- 50 Compact Flash Memory Interface
- 51 Display
- 51 RJ45 Connector
- 51 Pressure Sensors
- 51 Mass Flow Sensor
- 51 Alarm Beeper
- 51 Stepper Motor Amplifier
- 51 Voltage Regulators
- 55 24 V battery
- 55 Vivo 40: If the ventilator is running from an external battery and the voltage drops below 20 V the ventilator will switch to the internal battery supply.
- 55 12 V battery
- 55 Vivo 40: If the ventilator is running from an external battery and the voltage drops below 10 V the ventilator will switch to the internal battery supply.