CareFusion LTV 900, LTV 950, LTV 1000 Ventilator Quick Reference Guide
Below you will find brief information for Ventilator LTV 900, Ventilator LTV 950, Ventilator LTV 1000. These ventilators offer a range of features to support patients in a variety of settings. They feature multiple modes of ventilation, including Control, Assist/Control, SIMV, and CPAP. The devices can be adjusted to meet the individual needs of each patient, with variable settings for pressure, volume, and flow. The ventilators also feature a variety of alarms to alert the user to potential problems, and the device can be configured to communicate with external devices like printers, graphics monitors, and modems.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
LTV
®
Series Ventilators
(LTV
®
900, 950, and 1000)
Quick Reference Guide
ASSISTANCE
CareFusion Respiratory Systems
22745 Savi Ranch Parkway
Yorba Linda, California 92887-4645, USA
Tel: 763.398.8500
800.754.1914
Fax: 763.398.8403
Website: www.carefusion.com
Email: [email protected]
P/N 19819-001 Rev. A LTV
®
Series Ventilators ii
TABLE OF CONTENTS
Front and Side Panel Reference ................................................................................. 1
Front Panel Display and Description .......................................................................... 1
Side Panel Descriptions............................................................................................. 3
Turning the Ventilator On and Off .............................................................................. 5
Turning the Ventilator On .......................................................................................... 5
Turning the Ventilator Off .......................................................................................... 6
Ventilator Checkout Tests ........................................................................................... 7
Alarm Test ................................................................................................................. 8
Display Test ............................................................................................................... 9
Control Test ............................................................................................................. 10
Leak Test ................................................................................................................. 11
Vent Inop Alarm Test ............................................................................................... 12
Set Defaults ............................................................................................................. 13
Exit .......................................................................................................................... 14
Variable Controls........................................................................................................ 15
Setting Up Modes of Ventilation ............................................................................... 17
Setting Up Control Mode ......................................................................................... 17
Setting Up Assist/Control Mode ............................................................................... 19
Setting Up SIMV Mode ............................................................................................ 21
P/N 19819-001 Rev. A LTV
®
Series Ventilators iii
Setting Up CPAP Mode ........................................................................................... 23
Setting Up NPPV Mode ........................................................................................... 25
Monitored Data ........................................................................................................... 27
Extended Features ..................................................................................................... 29
Using AC/DC Power ................................................................................................... 57
Using the AC Adapter .............................................................................................. 57
Using an External DC Power Source ....................................................................... 58
Power Displays and Indicators ................................................................................. 59
Attaching a Breathing Circuit .................................................................................... 63
Oxygen Computer Chart ............................................................................................ 65
Alarms ......................................................................................................................... 67
P/N 19819-001 Rev. A LTV
®
Series Ventilators iv
FRONT AND SIDE PANEL REFERENCE
Front Panel Display and Description
P/N 19819-001 Rev. A LTV
®
Series Ventilators 1
A - Mode and Breath Selection – Selects ventilation modes. Selects breath types.
B - Power – Turns ventilator “On” or to “Standby.”
C - Variable Control Settings – Sets ventilation characteristics, such as Tidal Volume and Breath Rate.
D - Display Window – Displays Alarm Messages, Monitored Data, Extended Features menu.
E - Airway Pressure Display – Displays real-time airway circuit pressure.
F - Patient Effort Indicator – LED is lit briefly each time a patient trigger is detected.
G - Power Source – Displays power source and charge levels.
H - Variable Alarm Settings – Sets variable alarm levels.
I - Alarm Silence/Reset – Silences audible alarms. Clears visual alarms.
J - Set Value Knob – Changes variable control settings. Navigates Extended
Features menu.
K - Special Controls – Activates special controls such as Manual Breath, Low
Pressure O
2
Source, Insp/Exp Hold or Control Lock feature.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 2
FRONT AND SIDE PANEL REFERENCE
Side Panel Descriptions
P/N 19819-001 Rev. A LTV
®
Series Ventilators 3
A - 22mm Outlet Port – Patient Breathing Circuit outlet port.
B - Flow Xducer – Flow Transducer high pressure sensing port.
C - Flow Xducer – Flow Transducer low pressure sensing port.
D - Exh Valve – Exhalation Valve drive line port.
E - Alarm Sounder Port
F - Cooling Fan
G - DC Input – External DC power port (earlier version) or DC power port pigtail connector (current version).
H - Patient Assist – Patient Assist Call jack.
I - Comm Port – Communications port.
J - O
2
Inlet – Oxygen Inlet fitting.
K - Filter – Air Inlet.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 4
TURNING THE VENTILATOR ON AND OFF
Turning the Ventilator On
To turn the LTV
®
ventilator on:
1) Connect the ventilator to an external power source:
• The AC power adapter may be used or the ventilator may be connected to an external battery.
• If you do not connect the ventilator to an external power
source, it will operate from the internal battery.
2) Press and release the On/Standby button. The ventilator will commence operation:
During POST;
• The front panel displays are illuminated.
• Verify the audible alarm is activated for 1 second (only on ventilators with a
symbol on the back panel label).
• Verify a confirming audible chirp is activated (only on ventilators with a symbol on the back panel label).
3) Once POST is successfully completed, the ventilator begins operating using the stored control settings.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 5
Turning the Ventilator Off
To turn the LTV
®
ventilator off:
1) Disconnect the patient from the ventilator.
2) Press and hold the On/Standby button for 3 seconds. The ventilator ceases operating, the audible alarm sounds continuously and the Vent
Inop LED is lit.
• Verify a confirming audible chirp is activated immediately after the alarm is silenced (only on ventilators with a symbol on the back panel label).
4) The ventilator continues to charge the internal battery as long as it is connected to an external power source.
Note: The Vent Inop LED will remain lit for a minimum of 5 minutes and does not affect battery life.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 6
VENTILATOR CHECKOUT TESTS
WARNING - Disconnect the patient from the ventilator prior to running the Ventilator
Checkout tests and ventilate the patient using an alternative method. The ventilator does not deliver gas during the Ventilator Checkout tests.
To enable the Ventilator Checkout menu:
1) Begin with the ventilator in Standby mode (off) and connected to a valid AC power source.
• Verify that the External Power and Charge Status LEDs are illuminated.
2) Press and hold the Monitor Select button. While holding the Select button, press the On/Standby button.
• REMOVE PTNT alarm message is displayed and an audible alarm is sounded.
3) Clear the alarm by pressing the Silence/Reset button.
• Audible alarm is silenced and VENT CHECK is displayed.
• The first Ventilator Checkout Test, ALARM, is displayed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 7
Alarm Test
The alarm Test is used to verify that the audible alarm is working correctly.
1) Press the Select button while ALARM is displayed.
2) Verify the audible alarm is sounded.
• If a Patient Assist Call System or Remote Alarm is connected via the ventilator’s Patient Assist Port, verify the device also activates (audible/visual), as specified by its manufacturer.
3) When the alarm has sounded for at least 2 seconds, press the
Select button again.
• The audible alarm is silenced and the next menu item is displayed.
4) For ventilators with an audio sound symbol ( ) on the back panel label, verify a confirming audible chirp occurs after the alarm is silenced.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 8
Ventilator Checkout Tests
Display Test
The display Test is used to verify that the ventilator displays are working correctly.
To run the Display Test:
2) All segments of the 7-segment control displays, all dots of the dot-matrix window displays and all LEDs are illuminated. the AC adapter is connected to the ventilator (see page 7). page 12).
3) To end the display test, press the Select button again and the next menu item is displayed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 9
Control Test
The Control Test is used to verify that the ventilator buttons and controls are working correctly.
To run the Control Test:
2) SELECT is displayed in the display windows.
3) To test each control, press the button. The name of the button is displayed in the display window. To test the Set Value knob, turn it clockwise and counterclockwise. The direction of rotation is displayed in the display window.
4) To exit the control test, press the Select button again and the next menu item is displayed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 10
Ventilator Checkout Tests
Leak Test
The Leak Test is used to test the patient circuit for leaks. The patient circuit should be tested with all accessories, such as humidifiers or water traps, in place.
To run the Leak Test:
1) Cap or otherwise occlude the patient circuit wye.
• To perform the Leak Test, the ventilator closes the exhalation valve, sets the flow valve to a near-closed state, elevates the turbine motor speed and elevates the circuit pressure.
• At the conclusion of the test, the display shows LEAK xx.x pass or fail, where
xx.x is the measured leak.
3) To exit the Leak Test, press the Select button again and the next menu item is displayed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 11
Vent Inop Alarm Test
The Vent Inop Alarm Test is used to verify that the Inop Alarm is working correctly.
To run the Vent Inop Alarm Test:
1) To run the Vent Inop Alarm Test, the ventilator must be on (running) for at least 60 seconds and the Ventilator Checkout menu must be enabled.
2) Turn the ventilator off by pressing and holding the On/Standby button for a minimum of 3 seconds. DO NOT press the Silence/Reset button.
3) Observe the ventilator for 15 seconds.
• Listen for the alarm tone
• Watch the Vent Inop LED
4) For all ventilators, verify that both of the following conditions existed;
• The alarm tone sounded continuously for the full 15-second duration.
• The Vent Inop LED illuminated continuously for the full 15-second duration.
5) If a Patient Assist Call System or Remote Alarm is connected via the ventilator’s Patient
Assist Port, verify the device also activates (audible/visual), as specified by its manufacturer.
6) Silence the alarm by pressing the Silence/Reset button.
7) For ventilators with an a audio sound symbol ( ) on the back panel label, verify the following condition existed;
• A confirming audible chirp occurred after the alarm was silenced.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 12
Ventilator Checkout Tests
When the Ventilator Checkout Tests have been completed, proceed to Exit for instructions to exit the vent check mode, or see below concerning the use of the Set Defaults option.
Set Defaults
The Set Defaults option is used to reset user settable Controls and Extended Features settings to their factory-set default values (see the LTV
®
Series Ventilators Operator’s
Manual for factory-set default values).
To set the default values:
• VENT CHECK is displayed
• SET DEFAULTS is displayed.
•
•
Except for the Language selected and the Date/Time settings and format, all user settable Controls and Extended Features options are reset to their factory-set default values.
A DEFAULTS SET alarm will be generated the next time the ventilator is powered up in normal ventilation mode (see Alarms, DEFAULTS SET for additional information).
P/N 19819-001 Rev. A LTV
®
Series Ventilators 13
Exit
To return to any of the VENT CHECK tests, turn the Set Value knob until the desired test is displayed.
To Exit: is displayed.
The Ventilator performs a Self Test (POST) and resumes normal operation.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 14
VARIABLE CONTROLS
P/N 19819-001 Rev. A LTV
®
Series Ventilators 15
To set a variable control:
1) Select the control by pressing the associated button. The display for the selected control will be displayed at normal brightness and all other control displays will be dimmed.
2) Change the control value by rotating the Set Value Knob. Rotate clockwise to increase and counter-clockwise to decrease the value.
3) The new control value goes into effect when the operator:
•
•
•
•
Presses the selected button again, or
Selects another control, or
Presses the Control Lock button, or
Waits 5 seconds
All controls will then return to their normal brightness.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 16
SETTING UP MODES OF VENTILATION
Setting Up Control Mode
P/N 19819-001 Rev. A LTV
®
Series Ventilators 17
To set the ventilator up in Control mode:
1)
2)
3)
4)
5)
Press the Select button twice to toggle the modes between Assist/Control and
SIMV/CPAP. Select the Assist/Control mode.
Press the Select button twice to toggle between Volume and Pressure ventilation.
Select Volume or Pressure, as desired. (Not available on the LTV
®
900.)
Set the Breath Rate.
If Volume ventilation is selected, set the Tidal Volume. The calculated peak flow
Vcalc is displayed in the window while Tidal Volume is being changed.
If Pressure ventilation is selected, set the Pressure Control. (Not available on the
LTV
®
900.)
6)
7)
8)
9)
Set the Inspiratory Time. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation.
Set O
2
% (LTV
®
1000 only).
Set the Sensitivity to Off (dash “-“).
Set the High Pressure Limit alarm.
10) Set the Low Pressure alarm.
11)
Set the Low Minute Volume alarm.
12) Set the PEEP control on the exhalation valve.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 18
SETTING UP MODES OF VENTILATION
Setting Up Assist/Control Mode
P/N 19819-001 Rev. A LTV
®
Series Ventilators 19
To set the ventilator up in Assist/Control mode:
1)
Press the Select button twice to toggle the modes between Assist/Control and
SIMV/CPAP. Select the Assist/Control mode.
2)
Press the Select button twice to toggle between Volume and Pressure ventilation.
Select Volume or Pressure, as desired. (Not available on the LTV
®
900).
Set the Breath Rate.
3)
4)
5)
If Volume ventilation is selected, set the Tidal Volume. The calculated peak flow
Vcalc is displayed in the window while Tidal Volume is being changed.
If Pressure ventilation is selected, set the Pressure Control. (Not available on the
LTV
®
900.)
6)
7)
8)
Set the Inspiratory Time. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation.
Set O
2
%, (LTV
®
1000 only).
Set the Sensitivity to a setting from 1 to 9.
9)
Set the High Pressure Limit alarm.
10)
Set the Low Pressure alarm.
11) Set the Low Minute Volume alarm.
12)
Set the PEEP control on the exhalation valve
.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 20
SETTING UP MODES OF VENTILATION
Setting Up SIMV Mode
P/N 19819-001 Rev. A LTV
®
Series Ventilators 21
To set the Ventilator up in SIMV mode:
1)
2)
3)
4)
5)
Press the Select button twice to toggle the modes between Assist/Control and
SIMV/CPAP. Select the SIMV/CPAP mode.
Press the Select button to toggle between Volume and Pressure ventilation. Select
®
Volume or Pressure, as desired. (Not available on the LTV 900).
Set the Breath Rate.
If Volume ventilation is selected, set the Tidal Volume. The calculated peak flow
Vcalc is displayed in the window while Tidal Volume is being changed.
If Pressure ventilation is selected, set the Pressure Control. (Not available on the
LTV
®
900.)
6)
7)
8)
9)
Set the Inspiratory Time. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation.
Set the Pressure Support, if desired.
Set O
2
% (LTV
®
1000 only).
Set the Sensitivity to a setting from 1 to 9.
10)
Set the High Pressure Limit alarm.
11) Set the Low Pressure alarm.
12) Set the Low Minute Volume alarm.
13)
Set the PEEP control on the exhalation valve.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 22
SETTING UP MODES OF VENTILATION
Setting Up CPAP Mode
P/N 19819-001 Rev. A LTV
®
Series Ventilators 23
To set the ventilator up in CPAP mode:
SIMV/CPAP. Select the SIMV/CPAP mode.
Apnea backup. Select Volume or Pressure for Apnea backup. (Not available on the
LTV
®
900).
3) Set the Breath Rate to Off (dashes “- -“).
4) If Volume ventilation is selected, set the Tidal Volume for Apnea backup. The calculated peak flow Vcalc is displayed in the window while Tidal Volume is being changed.
5) If Pressure ventilation is selected, set the Pressure Control for Apnea backup. (Not available on the LTV
®
900.)
6) Set the Inspiratory Time for Apnea backup. The calculated peak flow Vcalc is displayed in the window while Inspiratory Time is being changed. Vcalc only applies to volume ventilation.
7)
9)
Set the Pressure Support, if desired.
O
2
% (LTV
®
1000 only).
Set the Sensitivity to a setting from 1 to 9.
10) Set the High Pressure Limit alarm.
NOTE: Although Tidal Volume, Pressure
Control and Insp Time are dimmed, they should be set to clinically
11) Set the Low Pressure alarm for Apnea backup. appropriate levels as the ventilator uses these settings for Apnea back-up ventilation.
12) Set the Low Minute Volume alarm.
13) Set the PEEP control on the exhalation valve.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 24
SETTING UP MODES OF VENTILATION
Setting Up NPPV Mode
P/N 19819-001 Rev. A LTV
®
Series Ventilators 25
3)
4)
5)
To set the Ventilator up in NPPV mode:
1) Set the ventilator controls for Control, Assist/Control, SIMV, or CPAP mode, as described in the preceding section.
2)
Set the ventilator controls for Volume or Pressure ventilation, as described in the preceding section.
Set all other ventilation parameters, as described in the previous section.
Set the High Pressure Limit alarm.
Enter Extended Features by pressing and holding the Monitor Select button for 3 seconds.
6)
7)
8)
Turn the Set Value knob until VENT OP is displayed.
Press the Monitor Select button.
Turn the Set Value knob until NPPV Mode is displayed.
9)
Press the Monitor Select button.
10)
Turn the Set Value knob until NPPV On is displayed.
11)
Press Monitor Select button.
12) The NPPV LED will be illuminated.
13) Exit the Extended Features menus by turning the Set Value knob until Exit is displayed, and pressing Select button until monitored data is displayed in the window.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 26
MONITORED DATA
The monitored data displays may be automatically scrolled or manually scrolled. To cycle through the available monitored data automatically from a halted scan, press the
Monitor Select button twice. Pressing the Select button once while scan is active shall halt scanning and the currently display monitor shall remain in the display window. Each time you press the button once; the next data item in the list will be displayed. To resume scan, press the scan button twice. The monitored data is displayed in the following order.
Display Description
PIP
MAP
PEEP
f
Vte
Displays the Peak Inspiratory Pressure measured during the inspiratory phase. PIP is not updated for spontaneous breaths.
Displays a running average of the airway pressure for the last 60 seconds.
Displays the pressure in the airway circuit at the end of exhalation.
Displays the breaths per minute and includes all breath types.
Displays the exhaled tidal volume as measured at the patient wye.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 27
Display Description
VE
I:E
Vcalc
Displays the exhaled tidal volume for the last 60 seconds as calculated from the last 8 breaths.
Displays the ratio between measured inspiratory time and measured exhalation time. Both normal and inverse I:E Ratios are displayed.
Is based on the Tidal Volume and Inspiratory Time settings. Displayed when selected and whenever Tidal Volume or Inspiratory Time is selected for change.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 28
EXTENDED FEATURES
Navigating the Extended Features Menus:
To enter the Extended Features menu (in normal ventilation mode), press and hold the Monitor Select button for three seconds.
To view the next item in a menu, turn the Set Value knob clockwise.
To view the previous item, turn the Set Value knob counterclockwise.
To enter a menu item or select a setting, press the
Select button.
To exit a menu, turn the Set Value knob until the EXIT option is displayed, then press the Select button or press Control Lock.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 29
P/N 19819-001 Rev. A LTV
®
Series Ventilators 30
EXTENDED FEATURES
Alarm Operations
2)
3)
4)
Alarm Volume
After accessing Extended Features, ALARM OP is displayed. Press the Select button and ALARM VOL is displayed.
1) Press the Select button.
VOL xx dBA is displayed, where xx is the currently set volume.
Turn the Set Value knob until the desired setting is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 31
Alarm Operations
Apnea Interval
After accessing Extended Features, ALARM OP is displayed. Press the Select button and ALARM VOL is displayed. Turn the Set Value knob until APNEA INT is displayed.
1)
2)
Press the Select button.
APNEA xx sec is displayed, where xx is the currently set Apnea interval.
3)
Turn the Set Value knob until the desired setting is displayed.
4)
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 32
EXTENDED FEATURES
Alarm Operations
High Pressure Alarm Delay
This menu item is used to select immediate or delayed audible notification for High
Pressure alarms.
After accessing Extended Features, ALARM OP is displayed. Press the Select button and ALARM VOL is displayed. Turn the Set Value knob until HP DELAY is displayed.
1)
2)
Press the Select button.
Turn the Set Value knob until the desired setting is displayed,
NO DELAY, DELAY 1 BRTH, or DELAY 2 BRTH.
3) Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 33
Alarm Operations
Low Peak Pressure Alarm
1)
2)
This item is used to select the type of breaths that the Low Pressure Alarm applies to.
After accessing Extended Features, ALARM OP is displayed. Press the Select button and ALARM VOL is displayed. Turn the Set Value knob until LPP ALARM is displayed.
Press the Select button.
Turn the Set Value knob until the desired setting is displayed,
ALL BREATHS, VC/PC ONLY.
3)
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 34
EXTENDED FEATURES
Alarm Operations
High PEEP Alarm
1
3)
4)
This menu item is used to set a high PEEP alarm value. When the current PEEP value exceeds the set high PEEP alarm value, an audible alarm will be sounded and a flashing HIGH PEEP message will be displayed.
After accessing Extended Features, ALARM OP is displayed. Press the Select button and ALARM VOL is displayed. Turn the Set Value knob until HIGH PEEP is displayed.
Press the Select button.
1)
2) Turn the Set Value knob until the desired setting is displayed,
HI PEEP OFF or PEEP xx cmH
2
O.
Turn the Set Value knob until the desired setting is displayed.
Press the Select button.
1
The HIGH PEEP alarm is only available on ventilators with software version 3.15 or higher installed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 35
Alarm Operations
PNT Assist
2
This menu item is used to configure the patient Assist Port output signal to be generated for use with remote alarm systems.
After accessing Extended Features, ALARM OP is displayed. Press the Select button and ALARM VOL is displayed. Turn the Set Value knob until PNT ASSIST is displayed.
1) Press the Select button.
2) Turn the Set Value knob until the desired setting is displayed,
NORMAL or PULSE.
3)
Press the Select button.
2
The PNT ASSIST option is only available on ventilators with software version 3.15 or higher installed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 36
EXTENDED FEATURES
Alarm Operations
Exit
To return to the top of the ALARM OP menu:
1)
Turn the Set Value knob until EXIT is displayed.
Press the Select button while EXIT is displayed 2)
P/N 19819-001 Rev. A LTV
®
Series Ventilators 37
Ventilator Operations
Variable Rise Time
1)
2)
3)
4)
The variable Rise Time option is used to select the rise time profile for Pressure Control and Pressure Support breaths. The rise time profiles are numbered 1 through 9, where
1 is the fastest rise time and 9 is the slowest rise time.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button, and RISE TIME is displayed.
Press the Select button.
PROFILE x is displayed, where x is the currently set value.
Turn the Set Value knob until the desired Rise Time Profile is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 38
EXTENDED FEATURES
Ventilator Operations
Variable Flow Termination
The Variable Flow Termination is used to select the percentage of peak flow used for cycling
Pressure Support breaths. Pressure Support breaths are cycled from inspiration to exhalation when the flow reaches the set percentage of the peak flow, or when flow goes below 2 lpm.
When Pressure Control Flow Termination is enabled, the Variable Flow Termination setting is used for flow termination of Pressure Control breaths as well.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until
VENT OP is displayed. Press the Select button. the
3) % OF PEAK xx is displayed, where xx is the current Flow Termination setting. is displayed.
Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 39
Ventilator Operations
Variable Time Termination
The Variable Time Termination is used to select maximum inspiratory time for cycling
Pressure Support breaths. Pressure Support breaths are cycled from inspiration to exhalation, if this time is reached before the flow reaches the set percentage of the peak flow. When a breath is cycled based on the time setting, the Pressure Support display is flashed briefly.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
4)
Turn the Set Value knob until TIME TERM is displayed.
Press the Select button.
TERM x.x sec is displayed, where xx is the current Time Termination setting.
5)
Turn the Set Value knob until the desired Time Termination is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 40
EXTENDED FEATURES
Ventilator Operations
Pressure Control Flow Termination
The Pressure Control Flow Termination option is used to enable or disable flow termination for Pressure Control breaths.
When this option is on, Pressure Control breaths are cycled at the set percentage of peak flow, if it is reached before the set Inspiratory Time elapses. The percentage of peak flow is set in the Variable Flow Termination option.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
4)
5)
Turn the Set Value knob until PC FLOW TERM is displayed.
Press the Select button.
PC FLOW ON or PC FLOW OFF is displayed.
Turn the Set Value knob until the desired state is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 41
Ventilator Operations
Leak Compensation
Use the Leak Compensation option to enable or disable tracking of the Baseline Flow to improve triggering when a circuit leak is present.
When Leak Compensation is on, the system is gradually adjusted to maintain set sensitivity, if the leak is stable and there is no auto cycling.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
Turn the Set Value knob until LEAK COMP is displayed.
Press the Select button.
LEAK COMP ON or LEAK COMP OFF is displayed.
4)
5)
Turn the Set Value knob until the desired state is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 42
EXTENDED FEATURES
Ventilator Operations
NPPV Mode
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
Turn the Set Value knob until the NPPV MODE is displayed.
Press the Select button.
3) NPPV MODE ON or NPPV MODE OFF is displayed.
4) Turn the Set Value knob until the desired state is displayed.
5) Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 43
Ventilator Operations
Control Unlock
When the Easy method is selected, unlock the controls by pressing and releasing the
Control Lock button.
When the Hard method is selected, unlock the controls by pressing and holding the
Control Lock button for 3 seconds.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
Turn the Set Value knob until CTRL UNLOCK is displayed.
Press the Select button.
UNLOCK EASY or UNLOCK HARD is displayed.
4)
5)
Turn the Set Value knob until the desired setting is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 44
EXTENDED FEATURES
Ventilator Operations
Language Selection
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
Turn the Set Value knob until LANGUAGE is displayed.
Press the Select button.
3) ENGLISH or the currently selected language is displayed.
4)
5)
Turn the Set Value knob until the desired language is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 45
Ventilator Operations
Software Versions
After accessing Extended Features, ALARM OP is displayed. Turn the
Set Value knob until VENT OP is displayed. Press the Select button.
Turn the Set Value knob until VER xx.xx.xx is displayed, where
xx.xx.xx is the current software version.
Usage Meter
After accessing Extended Features, ALARM OP is displayed. Turn the
Set Value knob until VENT OP is displayed. Press the Select button.
Turn the Set Value knob until USAGE xxxxx.x is displayed, where
xxxxx.x is the current number of hours the ventilator has been in operation.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 46
EXTENDED FEATURES
Ventilator Operations
Communications Setting
The ventilator may be connected to printer, a graphics monitor, or a modem. The
Communications Setting option is used to select the communications protocol for data transmission.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1) Turn the Set Value knob until COM SETTING is displayed.
2)
Press the Select button.
3) MONITOR or the currently selected protocol is displayed.
4)
5)
Turn the Set Value knob until the desired protocol is displayed.
Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 47
Ventilator Operations
Set Date
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
4)
5)
6)
7)
1)
2)
3)
Turn the Set Value knob until SET DATE is displayed.
Press the Select button.
The current date is displayed in the currently selected date format.
4)
Press the Control Lock button to exit, or continue to modify the Date.
To modify the Date:
Press the Select button, YEAR xxxx is displayed.
Turn the Set Value knob until the desired year is displayed.
Press the Select button, MONTH xx is displayed.
Turn the Set Value knob until the desired month is displayed.
Press the Select button, DAY xx is displayed.
Turn the Set Value knob until the desired day is displayed.
Press the Select button to accept the new date.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 48
EXTENDED FEATURES
Ventilator Operations
Set Time
2)
3)
4)
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1) Turn the Set Value knob until SET TIME is displayed.
Press the Select button.
The current time is displayed.
Press the Control Lock button to exit, or
To modify the Time:
1)
2)
3)
4)
5)
Press the Select button, HOUR xx is displayed.
Turn the Set Value knob until the desired hour is displayed.
Press the Select button, MIN xx is displayed.
Turn the Set Value knob until the desired minute is displayed.
Press the Select button to accept the new time. The seconds are automatically reset to 00.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 49
Ventilator Operations
Date Format
The Date Format option is used to select the display format for the current date.
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
4)
Turn the Set Value knob until DATE FORMAT is displayed.
Press the Select button.
MM/DD/YYYY or the currently selected date format is displayed.
Turn the Set Value knob until the desired format is displayed.
5) Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 50
EXTENDED FEATURES
Ventilator Operations
PIP LED
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
1)
2)
3)
4)
Turn the Set Value knob until PIP LED is displayed.
Press the Select button.
PIP LED ON or PIP LED OFF is displayed.
Turn the Set Value knob until the desired setting is displayed.
5) Press the Select button.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 51
Ventilator Operations
Model Number / Serial Number
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
To view the LTV
®
model number:
Turn the Set Value knob until LTV XXXX is displayed, where XXXX is the model of the ventilator.
To view the LTV
®
serial number:
1) Press the Select button while LTV XXXX is displayed.
• The serial number is displayed on the left side of the display area as XXXXXX, where XXXXXX is the serial number of the ventilator.
2) Press the Select button to return to the model number option.
To view LTM™ compatibility:
1)
2)
Press the Select button while LTV XXXX is displayed.
• LTM will be displayed if software and internal hardware in the
LTV
®
Ventilator are LTM™ compatible.
Press the Select button to return to the model number.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 52
EXTENDED FEATURES
Ventilator Operations
Valve Home Position
After accessing Extended Features, ALARM OP is displayed. Turn the Set Value knob until VENT OP is displayed. Press the Select button.
To view the valve home position:
Turn the Set Value knob until Vhome XXX is displayed, where XXX is the home position for the flow valve installed in the ventilator.
Set Defaults
The Set Defaults option is only displayed and accessed through the VENT CHECK and
VENT MTNCE menus and is used to reset user settable Controls and Extended
Features settings to their factory-set default values. See Ventilator Checkout Tests,
Set Defaults for instructions on how to set default values and the LTV
®
Series
Ventilators Operator’s Manual for factory-set default values.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 53
Ventilator Operations
Exit
To return to the top of the VENT OP menu:
1)
Turn the Set Value knob until EXIT is displayed.
2)
Press the Select button.
XDCR ZERO
This item is used to view the Transducer Autozero results and schedule the Transducer
Autozero to be run (see the LTV
®
Series Operator’s Manual for more information).
P/N 19819-001 Rev. A LTV
®
Series Ventilators 54
EXTENDED FEATURES
Ventilator Operations
RT XDCR DATA
This menu displays the Real Time Transducer Data (please see the Service Manual for more information).
EVENT TRACE
This menu displays the Events Codes stored by the ventilator (see the LTV
®
Series
Service Manual for more information).
P/N 19819-001 Rev. A LTV
®
Series Ventilators 55
Ventilator Operations
Exiting Extended Features
To return to Monitored Parameters:
1)
Turn the Set Value knob until EXIT is displayed.
2)
3)
Press the Select button.
Repeat Steps 1 and 2 until the Monitored Parameters are displayed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 56
USING AC/DC POWER
Using the AC Adapter
To run the ventilator from an external AC power source.
1) Connect the power jack (straight or 90°) from the AC adapter to the power port
(earlier version ventilators) or power port pigtail connector (current version ventilators) on the left side of the ventilator.
2) Connect the proper AC power cable (110 or 220 V plug) to the AC power adapter.
3) Connect the 110 or 220 V power cable to a suitable power source.
While the ventilator is plugged in, the internal battery is continuously charged.
CAUTION: Release Button – To avoid damaging the ventilator or the power connector, press the release button on the connector before removing it from the ventilator power port pigtail connector.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 57
Using an External DC Power Source
To run the ventilator from an external DC power source.
1) Connect the power port of the external DC power adapter cable to the power port on the left side of the ventilator (earlier version ventilators), or the power port pigtail connector (current version ventilators).
Connect the DC jack to the DC power source.
2)
P/N 19819-001 Rev. A LTV
®
Series Ventilators 58
POWER DISPLAYS AND INDICATORS
Indicators
Battery Level
The Battery Level indicator shows the level of available internal battery power while running from the internal battery.
LED Color
Green
Battery Level
Internal battery level is acceptable
Approximate Battery Time
@ nominal settings
45 minutes
Amber
Red
Off
Internal battery level is low
Internal battery level is critically low
Ventilator is running on AC or
External Battery
10 minutes
5 minutes
P/N 19819-001 Rev. A LTV
®
Series Ventilators 59
Indicators
Charge Status
When the ventilator is plugged into an External Power source, it automatically charges the internal battery.
LED Color
Flashing
Amber
Green
Amber
Red
Charge Status
The ventilator is performing pre-charge qualification testing of the battery prior to starting the charge process. This happens when external power is first applied to the ventilator. The qualification process normally takes a few seconds but may take up to an hour on a deeply discharged battery.
The internal battery is charged to full level.
The battery has not reached a full charge level and is still charging.
The ventilator has detected a charge fault or internal battery fault.
The internal battery cannot be charged. Contact your CareFusion
Certified Service Technician.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 60
POWER DISPLAYS AND INDICATORS
Indicators
External Power
The External Power indicator shows the level of external power while the ventilator is operating from an external power source. When the ventilator is running from the internal battery, the External Power indicator is off. When running from external power, the indicator shows the following levels.
LED Color Power Level
Green
Amber
External Power level is acceptable
External Power level is low
External power may be provided by connecting the ventilator to an external battery or to an external AC power source.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 61
P/N 19819-001 Rev. A
This page intentionally left blank
LTV
®
Series Ventilators 62
ATTACHING A BREATHING CIRCUIT
How to attach a patient breathing circuit.
1) Connect the main breathing tube to the 22 mm outlet port on the right side of the ventilator.
2)
Connect the two exhalation flow transducer sense lines to the ports marked Flow Xducer on the right side of the ventilator. These are non-interchangeable Luer fittings.
3)
Connect the Exhalation Valve driver line to the port marked Exh
Valve on the right side of the ventilator.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 63
P/N 19819-001 Rev. A
This page intentionally left blank
LTV
®
Series Ventilators 64
OXYGEN COMPUTER CHART
P/N 19819-001 Rev. A LTV
®
Series Ventilators 65
Oxygen Computer Chart
To determine O2 Input Flow:
1)
Find the desired FIO
2
on the horizontal axis.
2) Project up to the minute volume.
3)
Project horizontally to the left vertical axis and read the oxygen flow.
To determine O
2
Concentration:
1)
2)
Find the O
2
input flow on the vertical axis.
Project horizontally right to the minute volume.
3)
Project vertically down to the horizontal axis and read the FIO
2
.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 66
ALARMS
How to Silence and Reset Alarms
To silence an alarm, press the Silence Reset button.
To reset an alarm that has been corrected, press the Silence Reset button again.
Alarm Cause Solution
APNEA XX bpm Occurs when the time since the last breath start exceeds the set Apnea Interval. When an Apnea alarm occurs, the ventilator will enter Apnea Back up ventilation mode.
Reevaluate the patient’s condition.
Reevaluate ventilator settings.
APNEA
An Apnea alarm has occurred and cleared
The ventilator is no longer in Apnea Backup mode.
Reevaluate the patient’s condition.
Reevaluate ventilator settings.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 67
Alarm Cause Solution
BAT EMPTY
Occurs when the ventilator is operating from the internal battery power and the batter charge level is critically low. This alarm can be temporarily silenced but cannot be cleared.
Attach the ventilator to external AC or DC power.
BATTERY LOW
DEFAULTS
Occurs when the ventilator is operating from internal battery power and the battery charge level is low.
Occurs during POST when the ventilator detects an invalid setting stored in nonvolatile memory.
Attach the ventilator to external AC or DC power.
Reevaluate power requirements.
Push the
Silence/Reset button twice to reset alarm.
DEFAULTS SET
Occurs when the ventilator is first powered up after the SET DEFAULTS option has been used to reset all controls and extended features settings to their factoryset default values.
Reevaluate ventilator settings.
Push the
Silence/Reset button twice to reset alarm.
Reevaluate ventilator settings.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 68
Alarms
Alarm Cause Solution
DISC/SENSE
Occurs when the ventilator detects one of the following conditions:
•
The patient circuit or proximal pressure sense line has become
•
• disconnected.
The low side exhalation flow transducer sense line has become disconnected.
The proximal pressure sense line is pinched or occluded.
HIGH O
2
PRES
Occurs when the average oxygen inlet pressure exceeds the acceptable limit for the type of oxygen source.
Check Patient Circuit assembly for disconnects.
Check pressure sensing lines for occlusions.
Reduce O
2
inlet pressure.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 69
Alarm Cause Solution
HIGH PEEP
3
Occurs when the ventilator detects one of Reevaluate ventilator the following conditions:
•
The patient circuit positive end expiratory pressure (PEEP) exceeds settings.
Disassemble, clean and reassemble the
• the High PEEP alarm setting.
Patient Circuit, Exhalation valve and/or PEEP valve occluded.
Patient Circuit,
Exhalation Valve and
PEEP Valve.
HIGH PRES
Occurs when the circuit pressure exceeds the set High Pressure Limit setting.
Reevaluate ventilator settings.
Inspect Patient Circuit for occlusions or kinks.
Reevaluate patient.
HW Fault
Occurs when the ventilator detects a problem with the ventilator hardware.
If alarm reoccurs, contact your Service
Rep or CareFusion.
3
The HIGH PEEP alarm is only available on ventilators with software version 3.15 or higher installed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 70
Alarms
Alarm Cause Solution
INOP
A ventilator INOP occurs when:
•
The ventilator is switched from On to
Standby.
•
The ventilator detects any condition that is deemed to make the ventilator unsafe.
If an INOP alarm occurs during operation, remove ventilator from service and contact your
Service Rep.
LOCKED
The LOCKED message is displayed when a button is pressed while the controls are locked. No audible alarm is given.
LOW MIN VOL
Occurs when the exhaled minute volume is less than the set Low Minute Volume.
Press the Control Lock button.
If locked alert continues, press and hold the Control Lock button for three seconds.
Examine Exhalation
Valve body for disconnects.
Reevaluate patient.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 71
Alarm Cause Solution
LOW O
2
PRES
Occurs when the average oxygen inlet pressure is less than the minimum acceptable inlet pressure of 35 PSIG.
LOW PRES
NO CAL DATA,
NO CAL
Occurs when the peak inspiratory pressure for a machine or assist breath is less than the Low Pressure setting.
Occurs when the ventilator detects invalid or missing calibration records on power up.
Increase O
2
inlet pressure.
If using O
2
cylinder, replace used cylinder with a new one.
Examine Patient
Circuit for disconnect.
Reevaluate ventilator settings.
Reevaluate patient.
Remove ventilator from service, perform
Calibration procedure.
POWER LOST
Occurs when the ventilator is operating on external power and the voltage drops below the useable level and switches to internal battery operation.
Evaluate power requirements.
Attach ventilator to an external AC or DC power source.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 72
Alarms
Alarm Cause Solution
POWER LOW
Occurs when the ventilator is operating on external power and the voltage drops to the low level.
Evaluate power requirements.
REMOVE PTNT
Occurs when the ventilator is powered up in the Ventilator Checkout or Ventilator
Maintenance modes. The ventilator is not delivering gas.
RESET
A RESET alarm occurs if the ventilator restarts following a condition other than being shut down by pressing the
On/Standby button.
Ensure patient is disconnected from ventilator and is being ventilated by alternative means.
May be caused by
Internal Battery depletion during operation
4
or ESD. If the problem reoccurs, remove from service and contact your
Service Rep or
CareFusion
4
Only available on ventilators with software version 3.13 or higher installed.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 73
Alarm Cause Solution
XDCR FAULT
Occurs when a transducer autozero test fails.
Press Silence/Reset button twice to reset alarm. If problem occurs frequently, remove from service and contact your
Service Rep. or
CareFusion.
P/N 19819-001 Rev. A LTV
®
Series Ventilators 74
CareFusion Respiratory Systems
22745 Savi Ranch Parkway
Yorba Linda, California 92887-4645, USA
Tel: 763.398.8500
800.754.1914
Fax: 763.398.8403 www.carefusion.com
advertisement
Key Features
- Multiple ventilaton modes
- Variable pressure, volume and flow settings
- Alarms to alert the user to potential problems
- Communicates with external devices
- Control Unlock feature
- NPPV Mode
- Leak Compensation
- Pressure Control Flow Termination
- Variable Time Termination
- Variable Flow Termination
Frequently Answers and Questions
How do I turn the ventilator on?
What are the different modes of ventilation available?
How do I connect the ventilator to an external power source?
Related manuals
advertisement
Table of contents
- 6 Front and Side Panel Reference
- 6 Front Panel Display and Description
- 8 Side Panel Descriptions
- 10 Turning the Ventilator On and Off
- 10 Turning the Ventilator On
- 11 Turning the Ventilator Off
- 12 Ventilator Checkout Tests
- 13 Alarm Test
- 14 Display Test
- 15 Control Test
- 16 Leak Test
- 17 Vent Inop Alarm Test
- 18 Set Defaults
- 20 Variable Controls
- 22 Setting Up Modes of Ventilation
- 22 Setting Up Control Mode
- 24 Setting Up Assist/Control Mode
- 26 Setting Up SIMV Mode
- 28 Setting Up CPAP Mode
- 30 Setting Up NPPV Mode
- 32 Monitored Data
- 34 Extended Features
- 62 Using AC/DC Power
- 62 Using the AC Adapter
- 63 Using an External DC Power Source
- 64 Power Displays and Indicators
- 68 Attaching a Breathing Circuit
- 70 Oxygen Computer Chart
- 72 Alarms