Mindray Accutorr Plus Brite Service manual

Mindray Accutorr Plus Brite Service manual
Add to My manuals

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Manual
Mindray Accutorr Plus Brite Service manual | Manualzz

Accutorr Plus

is a U.S. trademark of Mindray DS USA, Inc.

Masimo SET

®

and LNOP

®

are registered trademarks of Masimo Corporation.

Nellcor

®

, OxiMax

®

, Oxiband

®

, and Durasensor

®

are U.S. registered trademarks of Nellcor Puritan Bennett Inc.

Oxisensor

is a trademark of Nellcor Puritan Bennett Inc.

Velcro

®

is a registered trademark of Velcro Industries B.V.

Copyright © Mindray DS USA, Inc., 2008. All rights reserved. Contents of this publication may not be reproduced in any form without permission of Mindray DS USA, Inc.

0070-10-0691 Accutorr Plus™ Service Manual

Table of Contents

Contents..................................................................................................................................................... vii

Foreword ................................................................................................................................................... vii

Notes and Precautions ............................................................................................................................... viii

Notes.......................................................................................................................................................... ix

Precautions ................................................................................................................................................ x

Theory of Operation...................................................................................................................................... 1 - 1

Major Functional Blocks of the Accutorr Plus......................................................................................... 1 - 2

CPU/NIBP Module Assembly with LCD Display................................................................................... 1 - 3

Software Overview ........................................................................................................................... 1 - 3

CPU Hardware Controls ................................................................................................................... 1 - 4

NIBP Hardware functions................................................................................................................. 1 - 4

Main pressure transducer circuitry ........................................................................................... 1 - 4

Over pressure transducer circuitry ............................................................................................ 1 - 4

Five A/D channels .................................................................................................................... 1 - 5

Pulse Channel Filter .................................................................................................................. 1 - 5

Inflation Pump .......................................................................................................................... 1 - 5

Bleed Valve ............................................................................................................................... 1 - 5

Dump Valve .............................................................................................................................. 1 - 5

Cuff Inflation Requirements ..................................................................................................... 1 - 5

Pulse Acquisition Requirements ............................................................................................... 1 - 6

Artifact Rejection Requirements .............................................................................................. 1 - 6

Parameter determination Requirements .................................................................................... 1 - 6

Major Hardware Components........................................................................................................... 1 - 6

CPU ........................................................................................................................................... 1 - 6

Flash Memory ........................................................................................................................... 1 - 6

Battery Backed Up Static RAM ................................................................................................ 1 - 6

Nonvolatile RAM ..................................................................................................................... 1 - 6

Micromanager ........................................................................................................................... 1 - 7

Battery Backup Voltage Regulator Circuitry ............................................................................ 1 - 7

Real Time Clock ....................................................................................................................... 1 - 7

Recorder Interface ..................................................................................................................... 1 - 7

Communication Block .............................................................................................................. 1 - 8

Backlight LED Driver ............................................................................................................... 1 - 8

LCD Trend Display .................................................................................................................. 1 - 8

Temperature Interface ............................................................................................................... 1 - 9

SpO

2

Processor Circuit Boards ................................................................................................................ 1 - 11

SpO

2

SpO

2

Nellcor

®

Nellcor

®

Interface ........................................................................................................................... 1 - 9

General Theory Of Operation.................................................................................................. 1 - 11

Nell-3 SpO

2

Circuit Board Theory of Operation .............................................................. 1 - 11

Nell-3 Interface Board Theory of Operation ..................................................................... 1 - 12

Masimo SET

® Technology............................................................................................................... 1 - 12

Masimo ®

Interface Board Theory of Operation ............................................................................... 1 - 12

Main Power Supply.................................................................................................................................. 1 - 14

Description ........................................................................................................................................ 1 - 14

Power Supply Output Voltages......................................................................................................... 1 - 14

Lithium Ion Battery and Charger ...................................................................................................... 1 - 14

LED and Tone Module............................................................................................................................. 1 - 16

Accutorr Plus Service Manual 0070-10-0691 iii

Table of Contents iv

Keypad Control ................................................................................................................................. 1 - 17

Specifications.................................................................................................................................................. 2 - 1

Performance Specifications...................................................................................................................... 2 - 1

NIBP Measurement Cycle Time ....................................................................................................... 2 - 1

Pulse Rate ................................................................................................................................. 2 - 2

Maximum Cuff Pressure ................................................................................................................... 2 - 2

Inflation Source................................................................................................................................. 2 - 2

Leak Rate .......................................................................................................................................... 2 - 2

Cuff Vent Rate .................................................................................................................................. 2 - 2

Temperature (Predictive) .................................................................................................................. 2 - 2

Nellcor

®

Masimo

SpO

2

®

SpO

2

.................................................................................................................................. 2 - 3

................................................................................................................................. 2 - 3

Battery............................................................................................................................................... 2 - 5

Real-Time Clock ............................................................................................................................... 2 - 5

Safety Characteristics............................................................................................................................... 2 - 6

Risk (Leakage) Currents ................................................................................................................... 2 - 6

Dielectric Withstand ......................................................................................................................... 2 - 6

Ground Resistance ............................................................................................................................ 2 - 6

Type of Protection Against Electric Shock....................................................................................... 2 - 6

Degree of Protection Against Electric Shock ................................................................................... 2 - 6

Physical Characteristics............................................................................................................................ 2 - 7

Environmental Characteristics ................................................................................................................. 2 - 7

Electrical Ratings ..................................................................................................................................... 2 - 8

Agency Compliance ................................................................................................................................. 2 - 8

Electromagnetic Compatibility ................................................................................................................ 2 - 9

Repair Information ....................................................................................................................................... 3 - 1

Introduction .............................................................................................................................................. 3 - 1

Safety Precautions .................................................................................................................................... 3 - 2

General Troubleshooting Guidelines ....................................................................................................... 3 - 3

Troubleshooting Tips ........................................................................................................................ 3 - 3

Exchange Program ............................................................................................................................ 3 - 3

Test Equipment Required......................................................................................................................... 3 - 4

Troubleshooting (Problem Isolation) ....................................................................................................... 3 - 5

Error Codes ............................................................................................................................... 3 - 5

Isolating the Problem, System Level ................................................................................................ 3 - 6

Isolating the Problem within the Main Unit...................................................................................... 3 - 7

Isolating the Problems with Optional Accessory Modules............................................................... 3 - 7

Clinical Issues with AccuTemp IR Ear Thermometer Applications................................................. 3 - 7

Multiple Temperature Measurements ....................................................................................... 3 - 7

How the AccuTemp IR Measures Arterial Temperature (Adults/Pediatrics) ........................... 3 - 8

Troubleshooting ........................................................................................................................ 3 - 9

Maintenance .............................................................................................................................. 3 - 9

Disassembly Instructions.......................................................................................................................... 3 - 10

Removal of the Rear Housing........................................................................................................... 3 - 10

Removal of the Front Bezel .............................................................................................................. 3 - 10

Removal of the Keyboard Assembly ................................................................................................ 3 - 10

Removal of the LED/Tone board...................................................................................................... 3 - 10

0070-10-0691 Accutorr Plus Service Manual

Table of Contents

Removal of the CPU / NIBP board................................................................................................... 3 - 11

Removal of the Power Supply Assembly ......................................................................................... 3 - 12

Removal of the LCD Display (Advanced Model) ............................................................................ 3 - 13

Removal and replacement of the Internal Li-Ion Battery ................................................................. 3 - 13

Removal of the AC Input Receptacle Assembly .............................................................................. 3 - 13

Thermal Printer (optional module) ................................................................................................... 3 - 13

Thermometer, Predictive (optional module)..................................................................................... 3 - 13

AccuTemp IR, Infrared Thermometer (optional module) ................................................................ 3 - 14

AccuTemp IR Mounting Cradle ....................................................................................................... 3 - 14

Interconnect Cables Run List ................................................................................................................... 3 - 15

Interface Specifications, Connectors and Functions. ........................................................................ 3 - 15

Connector to LED / Tone PCB, J-2 .......................................................................................... 3 - 15

Power Supply Header, J-502 .................................................................................................... 3 - 17

J-2 Keypad Connector .............................................................................................................. 3 - 17

J-4 Speaker Connector .............................................................................................................. 3 - 18

Recorder Connector, J-5 ........................................................................................................... 3 - 18

Thermometer Connector, J-4 .................................................................................................... 3 - 19

SpO

2

Connector, J-7 ................................................................................................................. 3 - 19

NIBP Pump Connector, J-505 .................................................................................................. 3 - 20

LED Backlight Connector, J-651 .............................................................................................. 3 - 20

Communication Connector, J-603 ............................................................................................ 3 - 20

LCD Connector, J-8 .................................................................................................................. 3 - 20

Power Supply Connectors................................................................................................................. 3 - 21

AC input connector, J-2 ............................................................................................................ 3 - 21

DC output connector, J-3 .......................................................................................................... 3 - 21

Battery connector, J-4 ............................................................................................................... 3 - 22

Replacement Parts and Isometric Drawings............................................................................................... 4 - 1

Introduction .............................................................................................................................................. 4 - 1

Available Replacement Parts And Sub-Assemblies................................................................................. 4 - 1

Product Variations And Options .............................................................................................................. 4 - 1

Exchange Program ................................................................................................................................... 4 - 2

Replacement Parts Pricing Information ................................................................................................... 4 - 2

Ordering Information ............................................................................................................................... 4 - 2

Isometric Drawings and Parts Lists.......................................................................................................... 4 - 3

Accessory Modules and Mounting Parts Lists......................................................................................... 4 - 10

Recorder Module Replacement Parts List ........................................................................................ 4 - 10

Predictive Temperature Module Replacement Parts List ................................................................. 4 - 10

Universal Rolling Stand (P/N ACCTROLLSTD, Assembly Parts List) .......................................... 4 - 11

Mounting Options ............................................................................................................................. 4 - 12

Calibration ..................................................................................................................................................... 5 - 1

Introduction .............................................................................................................................................. 5 - 2

Status and Error Code Table ............................................................................................................. 5 - 2

Error Codes ............................................................................................................................... 5 - 2

Precautions And Guidelines ..................................................................................................................... 5 - 4

Test Equipment And Special Tools Required .......................................................................................... 5 - 4

Power-up Sequence, Internal Testing....................................................................................................... 5 - 4

Service Diagnostics .................................................................................................................................. 5 - 5

Accutorr Plus Service Manual 0070-10-0691 v

Table of Contents

Introduction (Hidden Key)................................................................................................................ 5 - 5

Software Version Test (0a, 0b) ......................................................................................................... 5 - 6

Keypad Test ...................................................................................................................................... 5 - 7

LED Test (2a, 2b).............................................................................................................................. 5 - 8

Communication Test (3a, 3b)............................................................................................................ 5 - 8

Recorder Test (4) .............................................................................................................................. 5 - 8

Pump Test (5).................................................................................................................................... 5 - 9

Bleed Rate Test (6a, 6b) ........................................................................................................... 5 - 9

Leak Test (7) ..................................................................................................................................... 5 - 10

Over Pressure Test (8a, 8b, 8c) ......................................................................................................... 5 - 10

Pulse Channel DC Offset Test (9a, 9b, 9c) ....................................................................................... 5 - 11

Pulse Channel Average Noise Test (10a, 10b, 10c).......................................................................... 5 - 11

Main Pressure Transducer Verification Test (11a, 11b, 11c) ........................................................... 5 - 11

Verification of Accutorr Plus Pneumatic Performance using an NIBP Simulator (11d).................. 5 - 12

Overpressure Transducer Verification <12c, 12a, 12b> ................................................................... 5 - 13

Main Pressure Transducer Calibration.............................................................................................. 5 - 14

Overpressure Transducer Calibration ............................................................................................... 5 - 15

Predictive Thermometer Verification And Calibration............................................................................ 5 - 16

Water Bath Method........................................................................................................................... 5 - 16

Temperature Verification Test, Infrared Thermometer (Applicable Only To Units Equipped With A

Temperature Option) ................................................................................................................... 5 - 16

Low Battery Sensing......................................................................................................................... 5 - 17

Setting the Current Time .......................................................................................................................... 5 - 18

NIBP Normal Operation........................................................................................................................... 5 - 19

Trend Memory Initialization .................................................................................................................... 5 - 19

SpO

2

Normal Operation (Accutorr Plus model with SpO

2

) .................................................................... 5 - 20

Preventive Maintenance................................................................................................................................ 6 - 1

Introduction .............................................................................................................................................. 6 - 1

Limitations Of Physiological Simulators ................................................................................................. 6 - 1

Preventive Maintenance Schedule ........................................................................................................... 6 - 2

Mechanical and Physical Visual Inspection (One Year Interval) ..................................................... 6 - 2

Electrical Safety and Performance Checks (One Year Interval)....................................................... 6 - 2

vi 0070-10-0691 Accutorr Plus Service Manual

Contents Introduction

Contents

1.0 Theory of Operation ........................................................................ 1-1

2.0 Specifications .................................................................................. 2-1

3.0 Repair Information ........................................................................... 3-1

4.0 Replacement Parts and Isometric Drawings ................................. 4-1

5.0 Calibration ........................................................................................ 5-1

6.0 Preventive Maintenance .................................................................. 6-1

On the first page of each chapter a table of contents for that chapter is provided.

This is a Service Manual for the Accutorr Plus monitor. There are no Operating

Instructions in this manual. Complete operating instructions and listing of available accessories are published in the Accutorr Plus Operating Instructions—

P/N 0070-00-0692-02 (English) and P/N 0070-00-0695 (international languages).

Foreword

This Service Manual (P/N 0070-00-0691) is intended as a guide, for technically qualified personnel, to use during repair and calibration procedures for the Accutorr Plus (part number 0998-00-0444-9XX).

Use this Accutorr Plus Service Manual ONLY for the monitor part numbers listed below. This manual provides specific information that may not be applicable to other part numbers.

The part number (P/N) and serial number are necessary for positive identification of the correct parts. They are located on the rear panel label of the Accutorr Plus unit.

MONITOR PART NUMBER (P/N) MONITOR CONFIGURATION

0998-00-0444-91A Basic (NIBP only) (IEC, English)

0998-00-0444-91G

0998-00-0444-91P

Basic (NIBP only) (IEC, German)

Basic (NIBP only) (IEC, Spanish)

0998-00-0444-91F

0998-00-0444-91T

0998-00-0444-93A

Basic (NIBP only) (IEC, French)

Basic (NIBP only) (IEC, Italian)

Advanced (Nellcor ® ) (IEC, English)

0998-00-0444-93G Advanced (Nellcor ® ) (IEC, German)

0998-00-0444-93P

Advanced (Nellcor

®

) (IEC, Spanish)

0998-00-0444-93F

Advanced (Nellcor ® ) (IEC, French)

0998-00-0444-93T

Advanced (Nellcor ® ) (IEC, Italian)

0998-00-0444-94A

Advanced (Masimo ® ) (IEC, English)

0998-00-0444-94G

Advanced (Masimo ® ) (IEC, German)

0998-00-0444-94P

Advanced (Masimo ® ) (IEC, Spanish)

Accutorr Plus Service Manual 0070-10-0691 vii

Introduction Notes and Precautions

MONITOR PART NUMBER (P/N) MONITOR CONFIGURATION

0998-00-0444-94F Advanced (Masimo ® ) (IEC, French)

0998-00-0444-94T Advanced (Masimo ® ) (IEC, Italian)

Notes and Precautions

NOTES are a general information statement concerning the Accutorr Plus.

A CAUTION is provided to alert the user to use special care necessary for the safe and effective use of the device. They may include actions to be taken to avoid effects on patients or users that may not be potentially life threatening or result in serious injury, but about which the user should be aware. Cautions are also provided to alert the user to adverse effects on this device of use or misuse and the care necessary to avoid such effects.

Please read and adhere to all notes and precautions listed here and in the appropriate areas throughout this manual.

viii 0070-10-0691 Accutorr Plus Service Manual

Notes

Notes

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

See the serial and product number label on the rear panel of the unit for part number identification. This manual also includes information on the optional Recorder and

Predictive Temperature modules and Infrared

Temperature module.

This manual does not provide operating instructions. Consult the Operating Instructions

Manual for proper operation of this monitor.

Unauthorized servicing may void the remainder of the warranty. Check with the factory or with a local authorized representative to determine the warranty status of a particular instrument.

Do not change Room Number and /or Bed Letter on the Accutorr Plus during the transmission of temperature data.

The AccuTemp IR thermometer should be at room temperature when it is used. If it has been in a very warm or very cold environment it may need time for the thermometer temperature to equalize to ambient temperature.

Special care should be taken to insure that the front panel and glare screen are not scratched.

New batteries are shipped in a discharged state for safety reasons. A new battery must be charged for 4 hours for Li-Ion before first use. The monitor may be used with AC mains power during the charge cycle but battery operation could be limited during this time.

Mindray DS USA, Inc. maintains a policy of continuous development for product improvement and reserves the right to change materials, specifications, and prices without notice.

When replacing a Keyboard, you must replace the

Graphic Panel.

Monitors with Nellcor report SpO

2

®

and Masimo

®

SpO

2

do not

Runtime and Boot Software Versions.

The Accutorr Plus always displays time in a 24 hour format.

Introduction

Accutorr Plus Service Manual 0070-10-0691 ix

Introduction Precautions

Precautions

CAUTION: Li-lon batteries used in this device may present a risk of fire or chemical burn if mistreated. Do not disassemble, heat above 100°C (212°F), or incinerate. Replace battery with P/N: 0146-00-

0069 only. Use of another battery may present a risk of fire or explosion.

CAUTION: Dispose of used battery properly. Keep away from children. Do not disassemble and do not dispose of in fire.

CAUTION: Soldering and solder removal equipment must be low voltage operated and grounded to avoid static charge and stray current induced component damage. Maximum wattage: 25 W.

CAUTION: The Inside Of This Instrument Contains Static

Sensitive Components. Use correct static protection safeguards.

This publication may have been updated to reflect product design changes and/or manual improvements. This manual is correct to the best of our knowledge at the time of publication and reflects the product variations shown in this manual. You may contact

Technical Support to confirm that the manual is applicable for your unit. Updated manuals are available for sale.

x 0070-10-0691 Accutorr Plus Service Manual

1.0

Theory of Operation

1.1 Major Functional Blocks of the Accutorr Plus .............................. 1-2

1.2 CPU/NIBP Module Assembly with LCD Display ............................ 1-3

1.3 SpO2 Processor Circuit Boards ..................................................... 1-11

1.4 Main Power Supply .......................................................................... 1-14

1.5 LED and Tone Module ..................................................................... 1-16

This section of the manual explains the basic architecture of the Accutorr Plus. It is an overview, not a detailed circuit analysis. It is intended to be a tool, that if used properly, will lead to an understanding of the functions executed by a major functional block, and thus enabling a logical isolation and resolution of malfunctions.

The Accutorr Plus is offered in many functional configurations, including the options licensed and manufactured by third parties. The third party circuits will be treated solely as “black box” devices, with only the relevant input, output and power sources described.

The Theory of Operation section describes only the highest configuration level. Units built with fewer or different options may not use all of the capabilities of the Accutorr Plus, yet the over-all theories will remain constant.

Accutorr Plus Service Manual 0070-10-0691 1 - 1

Major Functional Blocks of the Accutorr Plus

1.1

Major Functional Blocks of the Accutorr Plus

• CPU/NIBP Module Assembly with LCD Display

• Interface and SpO

2

Processor Board

• Main Power Supply and Battery Charger Assembly

• LED/Tone Module Assembly

Theory of Operation

1 - 2 0070-10-0691 Accutorr Plus Service Manual

Theory of Operation

1.2

1.2.1

CPU/NIBP Module Assembly with LCD Display

CPU/NIBP Module Assembly with LCD Display

The Accutorr Plus CPU/NIBP module controls the functions of the NIBP, temperature probes, and SpO

2

module; transmits output data to the recorder; and provides a serial I/O port connection for PC communication.

The CPU/NIBP module distributes power from the power supply, which includes power and control signals to the LED/Tone module, to the remaining components of the system such as the recorder and SpO

2

modules. The CPU/NIBP module also supplies separate power to the LED backlight and provides trend data to the display.

The CPU/NIBP circuit board is comprised of both software and hardware. The following is a summary of the software control exercised and the major hardware affected. This is for informational purposes only, and both software and hardware controls may be changed at any time without notice.

Software Overview

The software performs the following functions:

1. Initializes hardware and software modules

• Microcontroller ports and communication (serial) interfaces

• Software entry points and variables

• Read configuration from NVRAM

2. Performs power up diagnostic

• Microcontroller, FLASH, and STATIC RAM test

3. Performs NIBP functions

• Start measurement

• Repeat measurement

• Over pressure limit

• Enter or exit test mode

• Abort measurement

• Enter or exit calibration mode

• Update alarm limits

• Report alarm status

• Report data

• Determine inflation pressure

• Perform controlled cuff deflation

• Monitor DC and AC pressure and acquire pulsations

• Reject artifacts

• Decide on termination point and exhaust cuff pressure

• Smooth pulsation data

• Determine systolic, mean, diastolic pressure and heart rate

• Check alarm conditions

Accutorr Plus Service Manual 0070-10-0691 1 - 3

CPU/NIBP Module Assembly with LCD Display Theory of Operation

1.2.2

1.2.3

1.2.3.1

1.2.3.2

4. Transmits/Receives data to/from Infrared/Predictive Temperature modules

5. Manages Recorder functions

6. Manages Keypad interface

7. Provides RS232 communication to PC for software download

8. Supports Service Diagnostics

9. Supports Nurse Call feature

10. Checks error conditions

• Maintain watchdog timer

• Run-time diagnostic

11. Transmits/Receives data to/from SpO

2

module, if present

12. Controls LCD view angle

CPU Hardware Controls

1. Microcontroller executes the above software functions, including:

• NIBP

• Recorder

• Temperature module, PTM or IR

• SpO

2

• LCD

2. Battery Back Up Circuitry for Power Savings

• Static memory

• Non-volatile RAM for configuration and future serial number

3. Flash ROM for Program Storage and Execution

4. Static RAM for Data Storage

5. RS232 level translation for PC Download and second interface for CIS/HIS

6. Real time clock with 16K of Lithium backed up NV RAM

NIBP Hardware functions

Main pressure transducer circuitry

A pressure transducer and amplifier combination is provided to acquire the cuff pressure signal.

Over pressure transducer circuitry

An over pressure transducer, independent of the main pressure transducer. Over pressure limits are set according to the patient size.

1 - 4 0070-10-0691 Accutorr Plus Service Manual

Theory of Operation CPU/NIBP Module Assembly with LCD Display

1.2.3.3

1.2.3.4

1.2.3.5

1.2.3.6

1.2.3.7

1.2.3.8

Five A/D channels

All channels are at 12-bit resolution and digitized at 101 Hz:

1. DC pressure

2. AC pressure (pulse)

3. Hardware over pressure transducer

4. Hardware over pressure cooperator reference

5. Battery voltage monitoring, when the Accutorr is in battery operation

Pulse Channel Filter

A high-pass filter is provided to isolate the pulsatile component from the DC cuff pressure. This pulsatile component is digitized as the pulse channel. The gain of this AC coupled channel is programmable in a manner consistent with the specified range of pulse amplitudes.

Inflation Pump

A pump capable of inflating the cuff within the specified time limit.

Bleed Valve

A valve and control system capable of bleeding down the cuff pressure.

Dump Valve

A valve to rapidly discharge the residual cuff pressure at the end of the measurement cycle.

Cuff Inflation Requirements

The cuff is initially inflated to a software-determined pressure dependent upon patient size. In timer mode, it should be a predetermined increment, traditionally 50 mmHg (40 mmHg for Neonates) (determined by the CPU), above the previous systolic reading, after the first measurement. The software detects error conditions in the pneumatic system such as disconnected cuffs or kinked hoses. To reduce the possibility of over inflating the cuff, the pump may be driven at a lower speed when pressurizing small neonatal cuffs. This is achieved by pulse width modulating the pump control signal. Further control of the neonatal inflation rate may be achieved by using the bleed valve to relieve some portion of the pump's delivery.

Accutorr Plus Service Manual 0070-10-0691 1 - 5

CPU/NIBP Module Assembly with LCD Display Theory of Operation

1.2.3.9

1.2.3.10

1.2.3.11

1.2.4

1.2.4.1

1.2.4.2

1.2.4.3

1.2.4.4

Pulse Acquisition Requirements

Pulse amplitudes from 0.06 mmHg to 12 mmHg are recognized. Pulse rates can range from 30 bpm to 250 bpm. The software discriminates pulses from the digitized pulse channel signal, and characterizes the pulses for amplitude, DC pressure, and pulse period.

False triggering, due to the dicrotic notch or noise, is minimized.

Artifact Rejection Requirements

The software provides means of artifact rejection, both during acquisition and during subsequent processing of the pulse array. Means of rejecting artifacts include attention to the timing and amplitude of pulses relative to adjacent pulses. If motion artifact is detected during acquisition, the software may cause temporary suspension of the bleed down process until a quiescent patient condition returns.

Parameter determination Requirements

Pulses that are deemed acceptable after artifact processing are entered in the pulse array.

Heart rate is taken to be the average rate of the pulse array data. Pulse periods differing excessively from the median are ignored. Pulses that do not appear to fit the amplitude envelope are removed. The systolic, diastolic, and mean pressures are then calculated in accordance with standard algorithms. A fair amount of data manipulation and computation is required in the parameter determination. The results are reported no more than one second from the termination of the pulse acquisition phase.

Major Hardware Components

CPU

A microcontroller (min. 16MHz) with three (3) serial communication controllers (SCC), a serial communication port (SCP), programmable chip select lines (to minimize glue logic), a IDMA controller, and parallel I/O ports for discrete signals.

Flash Memory

512K x 16 Flash ROM used for program storage and execution. A Flash ROM is used also for downloading new revisions of software via the CPU's serial I/O lines (SCC).

Battery Backed Up Static RAM

512K x 16 static ram is used for data storage, trend information, etc. This memory is used for data storage. Therefore, a battery back up is provided.

Nonvolatile RAM

Nonvolatile ram is used for configuration and serial number information. This memory requires a battery back up.

1 - 6 0070-10-0691 Accutorr Plus Service Manual

Theory of Operation CPU/NIBP Module Assembly with LCD Display

1.2.4.5

1.2.4.6

1.2.4.7

1.2.4.8

Micromanager

This device provides a number of features that are useful for management of this board.

They are listed as follows:

A. ON/STANDBY capability through keyboard control

B. Watchdog timer

C. Conversion of CMOS SRAM to pseudo nonvolatile memory

Battery Backup Voltage Regulator Circuitry

Implements regulation of the STANDBY VOLTAGE from the main power supply to 4V

+/- 0.2V dc with a current draw of 821 uA max.

Real Time Clock

This device provide seconds, minutes, hours, date of the month, month, and year with leap year compensation. The device is memory mapped to be pointed to as software requires.

Also, provisions for running this device include main power supply and battery operation in order for the RTC to keep time when the system is not powered. A lithium battery is included, having a minimum of 10 years of operation, for configuration and clock data retention during battery changes and main battery failure. Accuracy is +/- 1 minute/week.

Recorder Interface

IDMA capability for writing 8-bit buffered parallel data to the recorder. Handshaking signals to specify new line to be printed, request for next byte of data.

Accutorr Plus Service Manual 0070-10-0691 1 - 7

CPU/NIBP Module Assembly with LCD Display Theory of Operation

1 - 8

1.2.4.9

1.2.4.10

1.2.4.11

Communication Block

The module provides the following interfaces and functions:

• RS-232E drive from TTL levels. RS-232E driver for CIS/HIS that will provide minimum of 9V signal level.

• IC based DC/DC converter capable of converting VCC to 12VDC, with leakage current when off < 100uA. High frequency operation to limit emitted fields. Capable of providing an output current of 125 mA @ 12Vdc, with an efficiency of 80% or higher.

• Signal driver for an external isolated relay that will be used for Nurse Call, provided on pins of the 9-pin RS-232 connector.

The RS-232E interface required is implemented using a Maxim, MAX236 Multi-Channel

Transceiver (U602). This chip was chosen because it has a minimum ±9 volt output signal level. Communication is to support a 5 wire interface, 4 signals and a ground. The signals are described below:

SIGNAL

RXD

TXD

RTS

CTS

DESCRIPTION

Received Data from DCE to

DTE device

Transmitted Data from DTE to DCE device

Request to send from DTE to DCE device

Clear to Send from DCE device

Signal Ground

INPUT/OUTPUT

Input

Output

Output

Input

Input GND

All RS-232 signals are ESD protected by a U603, an SM14M24C Transient Voltage

Suppressor array. The +12 VT dc power required by the Nurse Call cable is supplied from the onboard DC/DC converter. The MAX236 receivers are controlled by logic signal

CSTAT*. Logic 0 enables the receivers; otherwise, they are tri-stated.

The RS-232 signals appear on connector J603, a dual row 14 pin header which will have a mating 14 pin pendant ribbon cable attached. The other end of the ribbon cable assembly is a 9 pin panel mount ‘D’ connector.

Backlight LED Driver

The LED driver module is implemented with Linear LT3474, step-down 1A LED driver. It can source constant current maximum up to 1A. The current can be programmed by setting the voltage at Vadj pin (pin 13) through resistors R42 and R14. The switching frequency is programmed to 300KHz for best duty cycle (i.e, 96%) for higher efficiency.

The typical LED drop voltage of 3.5V is achieved by programming the constant output

LED current of 71mA. The Vadj is sourced from external voltage source (i.e., +5V VCC) by external resistor divider.

LCD Trend Display

DAC control provided from the LED/Tone Board controls the LCD View Angle.

0070-10-0691 Accutorr Plus Service Manual

Theory of Operation CPU/NIBP Module Assembly with LCD Display

1.2.4.12

1.2.4.13

Temperature Interface

The board provides a Serial Interface to either Temperature module.

SpO

2

Interface

The board provides a Serial Interface to SPO

2

and reset control of the module.

FIGURE 1-1 CPU Block Diagram

Accutorr Plus Service Manual 0070-10-0691 1 - 9

CPU/NIBP Module Assembly with LCD Display Theory of Operation

1 - 10

FIGURE 1-2 NIBP Block Diagram

0070-10-0691 Accutorr Plus Service Manual

Theory of Operation

1.3

1.3.1

1.3.2

SpO

2

Processor Circuit Boards

SpO

2

Processor Circuit Boards

The Accutorr Plus can be equipped with one of two optional technologies:

• Nellcor® SpO

2

system

• Masimo® SpO

2

system

Each technology uses proprietary sensors and interface cables. There are no “adapters” possible or available.

SpO

2

General Theory Of Operation

Regardless of manufacturer, SpO

2

determination is based on the relative attenuation of a red and an infrared light source. A sensor from the instrument contains two sets of LEDs to illuminate a portion of the body (for example, a fingertip) and a single photo sensor detects the light returned. The LEDs are alternately pulsed, red and infrared. Since hemoglobin is the carrier of oxygen in the blood and the relative attenuation of the light sources is proportional to the amount of oxygen carried by the hemoglobin, a relative level of saturated hemoglobin can be determined. Since we are measuring the ratio of red and infrared return from the circulatory system, the detection method is substantially independent of skin color or total skin attenuation. Since the blood flow has a pulsatile characteristic, a pulse rate can also be computed from this information.

NOTE:

NOTE:

There is no calibration possible for the determination of the SpO

2

% or of the pulse rate derived. As mentioned above, we are measuring a ratio of intensity of two light sources to determine the % of saturated hemoglobin. The determination is based on proprietary software interpretation of the reflected light ratios, and it is validated against strict clinical trials where fresh arterial blood is examined with a clinical blood gas analyzer. Once the software is finalized, the SpO

2

monitor accuracy is fixed and there are no possible adjustments.

The wavelengths of the light emitted are nominally 660nm and 890nm with the energy not exceeding 15mV.

Nellcor

®

Nell-3 SpO

2

Circuit Board Theory of Operation

The Nell-3 is a complete SpO

2

detection and determination circuit, proprietary of

Nellcor® Puritan Bennett. Patient data is processed for serial transmission to the host system.

Accutorr Plus Service Manual 0070-10-0691 1 - 11

SpO

2

Processor Circuit Boards

1.3.3

1.3.4

1.3.5

Theory of Operation

Nellcor

®

Nell-3 Interface Board Theory of Operation

The Nellcor® Nell-3 interface circuit board (0670-00-0740) establishes bilateral communication and control of the Nellcor® Nell-3 oximeter circuit board. The circuit board consists of a power supply section that converts the system +5 VDC into the logic and analog supplies required by the oximeter, and optically isolated logic control for the processing circuits of the oximeter. A 5V to 10V DC/DC converter (T1) and two voltage regulators (U4 and U5) generate the required digital (+5VD) and analog (+5VA) voltages.

Regulator U4 also provides optocoupler U1 and protection array U6 with power. The logic control section, consisting of U1, U2 and U3, provide the buffer and isolation from the main CPU circuit to the Nellcor® Nell-3 oximeter section.

Masimo SET

®

Technology

Masimo SET® Technology combines advances in fundamental sensor technology, digital signal processing, an easily portable platform, and a low noise hardware platform incorporating multiple wavelengths applicable to many physiological parameters. Masimo

SET® Technology is composed of three components: (1) new signal processing apparatus,

(2) a new method for in vivo measurement, and (3) new sensor technologies. The main benefit of Masimo® technology is the effective cancellation of unpredictable “in-band noise” (noise that is in the same frequency, phase, and/or amplitude space as the desired signal). Conventional fixed filters, whether analog or digital form, have been designed to pass certain frequencies while rejecting others.

For example, a bandpass filter with a lower cut-off frequency of 1Hz and an upper cut-off frequency of 10Hz will pass all frequencies between 1Hz and 10Hz and reject frequencies below 1Hz and above 10Hz. Therefore, any “noise” that has a frequency outside the pass band is rejected and not considered (i.e., 120Hz due to room light and 60Hz noise from electrical lines).

However, if noise exists between the pass band of 1Hz and 10Hz, it will enter the system with the desired signal and corrupt the measurement (i.e., patient motion at 2Hz). Fixed filters will always have this limitation. Masimo® has overcome these problems by developing a novel technique to accurately determine the noise reference (patents issued and pending); thus, making adaptive filters work in real-time even when applied to medical monitoring of widely variable and unpredictable patients. With this new technique, Masimo® is able to solve the perennial problem of motion artifact in pulse oximetry. Further technical information is available from Masimo® on the Internet at www.masimo.com.

Masimo

®

Interface Board Theory of Operation

The Accutorr Plus Masimo® interface board (P/N: 0670-00-0716) provides an isolated data and power interface between the Accutorr Plus and the Masimo® MS-3 Pulse

Oximetry module. The patient isolation requirement for the Masimo® section is

1500VAC, less than or equal to 1mA leakage current, from input connectors J1 and J4-all terminals to J2-all terminals.

1 - 12 0070-10-0691 Accutorr Plus Service Manual

Theory of Operation SpO

2

Processor Circuit Boards

The isolated DC/DC converter, T1, supplies the regulated +5DIG as required by the

Masimo® MS-3 Pulse Oximetry module. The digital 5V is then utilized to generate the analog 15V used by the MS-3. The optocouplers (U1 and U2) isolate the signals used for bidirectional serial communication with the Masimo® MS-3 Pulse Oximetry module.

Control signal SpO

2

COMM at J4 will allow the Accutorr Plus software to interrupt the serial transmission for future options. All data and control lines interfaced to the SpO

2 module are opto-isolated. Power to the SpO

2 with the input coming directly from the CPU/LED module.

module is isolated by a DC/DC converter

Accutorr Plus Service Manual 0070-10-0691 1 - 13

Main Power Supply

1.4

1.4.1

1.4.2

1.4.3

Theory of Operation

Main Power Supply

There is one power supply used in the Accutorr Plus. P/N 0014-00-0083-02 is configured for the use and charge of a Lithium Ion battery, P/N 0146-00-0069.

Description

The power supply provides four regulated DC output voltages for the Accutorr Plus. The power supply can operate from an AC line or a lithium ion battery and will automatically select the source of power available to the monitor, with priority given to the AC line. In addition, the power supply can operate normally from the AC line without the battery or with a battery cell shorted.

The power supply consists of two basic blocks. The first block is an AC to DC converter, which consists of a universal AC input switching regulator and a battery charger for the lithium ion battery. The second block consists of DC to DC converters that will supply

+5V, +12V, +/-15V. Power to the DC to DC converters is "ORed" from either the battery or the output of the AC to DC converter. The power supply continuously monitors the battery voltage and temperature to charge the battery as required.

Power Supply Output Voltages

VOLTAGE

+5 volts

+12.5 volts

±15 volts

Standby

TOLERANCE

±2%

±5%

±3%

7.7V to 13.6V

MAXIMUM CURRENT

2.63 amps

0.55 amps

±100mA

17mA

Lithium Ion Battery and Charger

• Battery: 11.1V Lithium Ion battery (3S2P, 4.4 Ah min), P/N: 0146-00-0069

• Battery charger: fully charges the Lithium Ion battery (P/N 0146-00-0069) within a max. of 4 hours (with the monitor in STANDBY mode), and not overcharge if left on continuously. The power supply automatically reduces the charging current during peak power demands on the +5V and +12.5V outputs.

• Battery temperature: monitored with NTC thermistor Semitec P/N 103AT-2.

• Charge temperature: 0ºC to 50ºC

• Discharge temperature: -10ºC to 60ºC

• Charger power requirements: Nominal power (during current-limit mode) decreases linearly from 28W to 0W; maximum power (during fast charge) is 31.5 W max.

1 - 14 0070-10-0691 Accutorr Plus Service Manual

Theory of Operation Main Power Supply

• Low battery shutdown: Automatic shutdown of all DC outputs when the battery voltage is 9.1V ±4%. Remaining load on the battery is less than 5 mA.

When the remote On/Off switch is enabled and the AC input is not available, power is delivered regardless of the charge on the battery as long as battery voltage is greater than the low battery shutdown voltage. This insures that the system could be operated with a battery that is not fully charged.

• Enable after shutdown: If the AC input is not available after battery replacement, power is delivered when the remote On/Off switch is enabled as long as the battery voltage is 9.9V minimum.

Accutorr Plus Service Manual 0070-10-0691 1 - 15

LED and Tone Module

1.5

Theory of Operation

LED and Tone Module

The Accutorr Plus LED/Tone module performs four main tasks: 1) It drives all the 7segments and discrete LED user displays on the keypad; 2) it generates an IEC60601-1-8 compliant alarm signal via the onboard Digital-to-Analog-Converter (DAC) and drives advisory tones to the speaker by the onboard power amplifier; 3) it controls the interface between the main CPU and the keypad; the functionality on the module is accomplished by the use of a Field Programmable Gate Array (FPGA) device; and 4) DAC drives view angle of trend display.

The LED/Tone module is under direct control of the CPU/NIBP module for inputs, and converts the input data into LED segment drive as well as the audio tones for advisory and alarm conditions.

1 - 16

FIGURE 1-3 LED/Tone Module, Functional Block Diagram

0070-10-0691 Accutorr Plus Service Manual

Theory of Operation

1.5.1

LED and Tone Module

Keypad Control

Circuitry for scanning the keypad for key presses. The board scans a 3x8 matrix of keys.

In the case of simultaneous key presses, a key is recognized by the software. The other key will not be recognized, until the initial key is released.

Accutorr Plus Service Manual 0070-10-0691 1 - 17

LED and Tone Module Theory of Operation

This page intentionally left blank.

1 - 18 0070-10-0691 Accutorr Plus Service Manual

2.0

Specifications

2.1

2.1.1

2.1 Performance Specifications ........................................................... 2-1

2.2 Safety Characteristics ..................................................................... 2-6

2.3 Physical Characteristics ................................................................. 2-7

2.4 Environmental Characteristics ....................................................... 2-7

2.5 Electrical Ratings ............................................................................. 2-8

2.6 Agency Compliance ........................................................................ 2-8

2.7 Electromagnetic Compatibility ....................................................... 2-9

Performance Specifications

Systolic and Diastolic Pressure Readout

Number of Digits:

Accuracy*:

3

Mean error < ±5 mmHg

Standard deviation < ±8 mmHg

Range:

*

Adult Mode: 55 to 260 mmHg

Pediatric Mode: 55 to 160 mmHg

Neonatal Mode: 45 to 120 mmHg

Tested per ANSI/AAMI SP10-1992 methods.

NIBP Measurement Cycle Time

Less than 30 seconds average at 80 BPM with 180 mmHg pump up pressure, without retries, motion artifact, or arrhythmia with standard adult cuff on a healthy individual.

Cycle time is affected by arm size and wrapping technique.

Accutorr Plus Service Manual 0070-10-0691 2 - 1

Performance Specifications Specifications

2.1.1.1

2.1.2

2.1.3

2.1.4

2.1.5

2.1.6

Pulse Rate

Range:

Display Resolution:

Accuracy:

35-245 BPM for Adult and Pediatric

70-245 BPM for Neonate

1 BPM

±3 BPM or ±3%, whichever is greater

Maximum Cuff Pressure

Two means of limiting cuff pressure are provided: 1) a hardware over pressure monitor that limits the pressure to 330 mmHg for Adults, 220 mmHg for Pediatrics, and 165mmHg for Neonates; 2) a software overpressure monitor that vents if the pressure exceeds 300 mmHg for Adults, 200 mmHg for Pediatrics, and 157 mmHg for Neonates. If the hardware over pressure circuit is tripped in normal operation, then the unit must be turned off and on to reset the system.

Inflation Source

This inflation source is capable of supplying sufficient air to bring a volume of 700 cc’s to a pressure of 300 mmHg in no more than 35 seconds. If the cuff is not inflated to the desired pressure within 60 seconds, then the cuff is vented and a retry cycle is initiated.

Leak Rate

With the bleed valve closed, the maximum pressure drop is 10 mmHg in 90 seconds measured with a 700 cc volume at a differential pressure of 250 mmHg.

Cuff Vent Rate

When the unit is vented, a volume of at least 700 cc’s is reduced from a pressure of 250 mmHg to a pressure of 20 mmHg in a maximum of 14 seconds.

Temperature (Predictive)

Range:

Display Resolution:

Accuracy:

90-110°F, 32-43°C

0.1°F, 0.1°C

Meets ASTM E1112-2006 for accuracy

2 - 2 0070-10-0691 Accutorr Plus Service Manual

Specifications Performance Specifications

2.1.7

2.1.8

Nellcor

®

SpO

2

Range:

Display Resolution:

SpO2 Response Time:

Display Update:

Calibration:

70-100% SpO

2

1% SpO

2

4.5 to 6.5 seconds

Less than 4 seconds

Factory Calibrated to Functional Saturation

Accuracy*: ± 2 digits from 70 - 100% SpO

2

± 3 digits from 70 - 100% SpO

2

<70% unspecified

- Adult

- Neonates

Pulse Rate Range: 21 to 249 BPM

* Neonatal accuracy specifications are based upon testing the N-3000 and N-25 neonatal sensors on healthy adult volunteers in induced hypoxia studies, in the range of 70 - 100% SpO2. The specified accuracy also takes into account published literature which predicts that there may be a small difference in % of

SpO2 reported by the oximeter when measurements from adult and fetal blood with 100% fetal hemoglobin are compared. Fetal hemoglobin is present in concentrations varying from 10% to 90% in neonatal blood, and this percentage declines over time. As the percentage of fetal hemoglobin in neonatal blood declines, the theoretical effect on accuracy due to this source is reduced.

Masimo

®

SpO

2

Range:

Display Resolution:

Display Update:

70-100% SpO2

1% SpO

2

< 4 seconds

SpO2 Accuracy Saturation during No Motion Conditions

1

Adults / Pediatrics: 70% to 100% ± 2 digits

5

<70% unspecified

Neonates (LNOP/LNCS): 70% to 100% ± 3 digits

SpO2 Accuracy Saturation during Motion Conditions

6

Adults / Pediatrics

2

: 70% to 100% ± 3 digits

<70% unspecified

Neonates

Neonates

3

3

(LNOP):

(LNCS):

70% to 100% ± 4 digits

70% to 100% ± 3 digits

<70% unspecified

SpO

2

Response Time: 10 secs (Display Averaging Time is not user-selectable for Masimo®. It will be set to the default value of 8 seconds.)

Accutorr Plus Service Manual 0070-10-0691 2 - 3

Performance Specifications Specifications

NOTE: This time was measured with post average time at 8 seconds.

Low Perfusion Performance

4

: >0.02% Pulse Amplitude

%Transmission >5%

Saturation (%SpO

2

Pulse ±3 digits

) ±2 digits

Wavelengths Emitted:

Maximum Emitted Energy:

660nm and 905nm

30mW at 50mA pulsed

Interfering Substances: Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present.

Dyes, or any substance containing dyes, that change usual arterial pigmentation may cause erroneous readings.

Pulse Rate Performance Specifications

Pulse Rate During No Motion Conditions

1

:

Adult/Pediatric/Neonates: 26 to 239 BPM ±3 digits

Pulse Rate During Motion Conditions

2,3

:

Adult/Pediatric/Neonates: 26 to 239 BPM ±5 digits

Update Rate: Less than 4 seconds

1 The Masimo® MS-3 Pulse Oximetry module with LNOP®•Adt sensors have been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of

70% to 100% SpO2 against a laboratory co-Oximetry and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

2

The Masimo® MS-3 Pulse Oximetry module with LNOP®•Adt sensors has been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz. At an amplitude of 1 to 2 cm and non-repetitive motion between

1 to 5 Hz. At an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70% to 100% SpO2 against a laboratory co-Oximetry and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

3

The Masimo® MS-3 Pulse Oximetry module with LNOP®•Neo and LNOP®•NeoPt sensors has been validated for motion accuracy in human blood studies on neonates while moving the neonates foot at 2 to 4 Hz at an amplitude of 1 to 2 cm against a laboratory co-Oximetry and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

4 The Masimo® MS-3 Pulse Oximetry module has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than

0.02% and a % transmission of greater than 5% for saturation’s ranging from 70% to 100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

5 The LNOP®•Ear Sensors have an SpO2 accuracy of 70% to 100% ±3.5 digits for adults during no motion conditions, however, since the monitor cannot display ½ digits, the accuracy shall be rounded to ±4 digits.

6

The SpO2 accuracy during motion conditions is not specified for the LNOP®•Ear Sensors.

2 - 4 0070-10-0691 Accutorr Plus Service Manual

Specifications Performance Specifications

2.1.9

2.1.10

Battery

BATTERY TYPE:

Number of Batteries:

Battery Voltage:

Battery Capacity:

LITHIUM-ION

1

11.1 VDC nominal

4.4 Amp-Hour

Battery Run Times from full charge with a new battery at 25° C with 1

NIBP measurement every 5 minutes, continuous SpO

2

monitoring, and recorder not in use.

Accutorr Plus, basic model - 9.5 hours

Recharge Time:

Accutorr Plus, advanced models:

Nellcor® - 7 hours

Masimo® - 7 hours

4 hours max., in STANDBY only

Real-Time Clock

RESOLUTION:

Accuracy:

Display Format:

Power:

1 MINUTE

±1 minute/week

24 hours

The real-time clock maintains the time and date when the instrument is ON or in STANDBY mode, connected to AC mains, or running from the internal battery for at least ten years from original assembly. The real-time clock will maintain time and date even if the instrument’s main battery is disconnected.

Accutorr Plus Service Manual 0070-10-0691 2 - 5

Safety Characteristics

2.2

2.2.1

2.2.2

2.2.3

2.2.4

2.2.5

Specifications

Safety Characteristics

Risk (Leakage) Currents

Enclosure Risk Current (leakage): < 100uA in normal operating conditions

< 300uA in any single fault condition

Patient Source Current: < 10uA in normal operating conditions

< 50uA in any single fault condition

Patient Sink Current: < 50uA

Dielectric Withstand

2500V RMS at 50 or 60 Hz for 1 minute from any patient lead or combination of patient leads to the chassis.

1500V RMS at 50 or 60 Hz for 1 minute AC mains hot or neutral to the chassis.

Ground Resistance

Ground resistance < 0.1 ohm from the AC mains power inlet module’s ground contact pin to any exposed metal part, which may become energized when measured per UL 60601 and IEC 60601-1. A ground resistance of up to 0.2 ohm is allowed when measured from the U-blade of the supplied AC line cord to any exposed metal part, which may become energized.

Type of Protection Against Electric Shock

Class 1 with internal electric power source. Where the integrity of the external protective earth (ground) in the installation of its conductors is in doubt, the equipment is operated from its internal electric power source.

Degree of Protection Against Electric Shock

Monitor – Type B applied part

NIBP – Type BF defibrillation protected applied part

SpO

2

– Type BF defibrillation protected applied part

Protection Against Hazards of Explosion

Not protected (ordinary)

Protection Against Ingress of Liquids

Meets the requirements specified by IEC 60601-1, clause 44.3 and IEC 60601-2-30: Non-

Protected Equipment (IPX1) as specified in EN 60529.

2 - 6 0070-10-0691 Accutorr Plus Service Manual

Specifications

2.3

2.4

Physical Characteristics

Physical Characteristics

Size (maximum):

Main Unit:

Recorder Module:

Predictive Module:

Weight:

19 cm(W) x 26.93 cm(H) x 20.83 cm (D)

7.5" (W) x 10.6" (H) x 8.2" (D)

5.33 cm(W) x 23 cm(H) x 11 cm (D)

2.1" (W) x 9" (H) x 4.25" (D)

5.7 cm(W) x 15.9 cm(H) x 11.8 cm (D)

2.25" (W) x 6.25" (H) x 4.63" (D)

<4.95 kg (11 pounds), depending on configuration

Environmental Characteristics

Operating

Temperature:

Humidity:

Altitude:

Shock and Vibration:

10°C to 40°C, 50°F to 104°F (Accutorr Plus

& Recorder)

10°C to 32°C, 50°F to 90°F (Predictive

Thermometer)

15 to 90% max, non-condensing.

1013 hPa to 697 hPa (0 to 10,000 ft.)

Meets IEC 60068-2-27 for shock with peak acceleration of 15g, 11 mSec duration, half sine.

Meets EN 60068-2-64 for random vibration with frequencies of 10 to 2000 Hz, resolution of

10Hz, 10 minutes per axis, acceleration spectral density of:

10 Hz to 100 Hz : 1.0 (m/s2)2/Hz

100 Hz to 200 Hz : -3 dB per octave

200 Hz to 2000 Hz : 0.5 (m/s2)2/Hz

Meets ISTA Test Procedure 1A (less than 100 lbs) Shipping:

Storage

Temperature:

Humidity:

-15°C to +60°C

+5°F to 140°F

10 to 95%, non-condensing

Accutorr Plus Service Manual 0070-10-0691 2 - 7

Electrical Ratings

2.5

2 - 8

2.6

Specifications

Electrical Ratings

Voltage:

Current:

Frequency:

Power Consumption:

100 - 120 / 220 - 240 VAC

0.85 / 0.5 A

60 / 50 Hz

40 W, maximum

Agency Compliance

The Accutorr Plus is designed to comply with the following agency standards:

Safety, IEC EN 60601-1:1990 +A1:1993 + A2:1995 +A13:1996

IEC 60601-1:1988 +A1:1991 +A2:1995

Safety, UL

Safety, Canada

UL 60601-1:2003

CAN/CSA C22.2 No. 601.1-M90 (R2005)

Safety, collateral

Multifunction

SpO2 particular

NIBP, particular

EN 60601-1-1:2001 / IEC 60601-1-1:2000

EN 60601-2-49:2001 / IEC 6060 1-2-49:2001

ISO 9919:2005

EN 60601-2-30:2000 / IEC 60601-2-30:1999

NIBP, IEC

NIBP, supplementary

NIBP, USA

Programmable

EN 1060-1:1995 +A1:2002

EN 1060-3:1997 +A1:2005

ANSI/ AAMI SP10:2002 +A1:2003 +A2:2006

EN 60601-1-4:1996 +A1:1999

IEC 60601-1-4:1996 +A1:1999

Temperature

Biological

Risk

Acoustics

Random Vibe

Shock

Ingress

Shipping

ASTM E1112-00:2006

ISO 10993-1:2003

EN 14971:2000 +A3:2003

EN 60601-1-8:2004 +A1:2006

EN 60068-2-64:1994 / IEC 60068-2-64 +corrig:1993

IEC 60068-2-27:1987

EN 60529:1991 +A1:2001 +Corr.1:2003

ISTA Procedure 1A:2001

0070-10-0691 Accutorr Plus Service Manual

Specifications

2.7

Electromagnetic Compatibility

Drop and Impact

EMC

Magnetic Fields

Voltage Dips

ECRI PB 29689:1979

IEC 60601-1-2:2001

EN 61000-4-8:1993 +A1:2001

EN 61000-4-11:2004

Volt Change, Fluctuations, Flicker EN 61000-3-3:1995 +A1:2001 +A2:2005 +IS1:2005

Conducted Immunity EN 61000-4-6:1996 +A1:2001 +A2:2006

Radiated Immunity

ESD

EN 61000-4-3:2006

EN 61000-4-2:1995 +A1:1998 +A2:2001

EFT

Surge

EN 61000-4-4:2004

EN 61000-4-5:2006

Radiated and Conducted Emissions EN 55011:1998 +A1:1999 +A2:2002

Harmonic Emissions EN 61000-3-2:2000 +A1:2001 +A2:2005

Electromagnetic Compatibility

The Accutorr Plus has been tested and found to comply with the requirements of IEC

60601-1-2: 2001 “Collateral Standard Electromagnetic Compatibility Requirements and

Tests.”

To ensure Electromagnetic Compatibility (EMC), operate the Accutorr Plus in accordance with the instructions for use.

To prevent the possibility of other equipment affecting the Accutorr Plus, insure adequate distances are maintained between devices.

Accutorr Plus Service Manual 0070-10-0691 2 - 9

Electromagnetic Compatibility Specifications

This page intentionally left blank.

2 - 10 0070-10-0691 Accutorr Plus Service Manual

3.0

Repair Information

3.1

3.1 Introduction ...................................................................................... 3-1

3.2 Safety Precautions .......................................................................... 3-2

3.3 General Troubleshooting Guidelines ............................................. 3-3

3.4 Test Equipment Required ............................................................... 3-4

3.5 Troubleshooting (Problem Isolation) ............................................. 3-5

3.6 Disassembly Instructions ............................................................... 3-10

Introduction

This chapter of the Service Manual provides the technical information necessary to resolve most instrument malfunctions. Important prerequisites for effective troubleshooting are a thorough understanding of the instrument functions and

specifications, as well as an understanding of the theory of operation. (See “Theory of

Operation” on page 1-1. See “Specifications” on page 2-1.)

Functional overviews of the major circuit blocks are provided in this manual. (See

“Theory of Operation” on page 1-1.) Technically qualified individuals may prefer to

isolate problems to a sub-circuit or component level. Component level repair activity is not always possible. Isolation of defective components is frequently impossible without the functional emulation of operating software. The equipment and processes required are not cost-effective on a single unit basis.

The procedures detailed in this chapter are for use by qualified technical personnel only.

Mindray DS USA, Inc. offers a comprehensive selection of Technical Training Seminars for this and other products. Contact the Registrar, Technical Support Department, or your regional sales or service representative for course offerings and dates.

Accutorr Plus Service Manual 0070-10-0691 3 - 1

Safety Precautions

3.2

Repair Information

Safety Precautions

There are areas where improper repair techniques will cause further damage. When the instrument covers are removed, observe the following precautions and guidelines.

1. Power line voltages will be exposed; identify hazard points and avoid direct contact.

2. Battery terminals are accessible; do not short to each other or the positive terminal to ground circuits, as the capacity of the battery is sufficient to create an over heating safety hazard.

3. Fuses must not be by passed or replaced by different current, voltage or time delay ratings.

4. The circuit boards contain static sensitive components; use proper static safeguard techniques.

CAUTION: Li-lon batteries used in this device may present a risk of fire or chemical burn if mistreated. Do not disassemble, heat above 100°C (212°F), or incinerate. Replace battery with P/N: 0146-00-

0069 only. Use of another battery may present a risk of fire or explosion.

CAUTION: Dispose of used battery properly. Keep away from children. Do not disassemble and do not dispose of in fire.

3 - 2 0070-10-0691 Accutorr Plus Service Manual

Repair Information

3.3

3.3.1

3.3.2

General Troubleshooting Guidelines

General Troubleshooting Guidelines

This procedure does not cover all possible problems. The intent of troubleshooting is to provide quick isolation and remedy of malfunctions and to return the instrument promptly to safe patient care. A logical approach will quickly identify the problem and suggest costeffective solutions.

Troubleshooting Tips

1. Isolate the cause of the problem to be either clinical, instrument, or external. If a clinical cause is suspected, take the physiological measurement by a manual method or use an alternate measuring device. Check the physiological value against the instrument specifications; if it exceeds specification, continue with manual or alternate method.

If a problem with the instrument is suspected, first examine and / or replace patient connected accessories. Check instrument set-up / configuration. If all set-ups are OK, replace instrument with similar device.

If the environment is suspected, observe if the malfunction coincides with other equipment cycling (especially high-power motors, RF generators, etc.). Replace problem instrument with an identical substitute. Place the problem instrument in a low-disturbance environment. If the problem persists in the low-disturbance environment, the instrument performance may be suspect. If the substitute instrument exhibits similar poor performance in the original environment, the environmental causes must be eliminated.

2. Use the proper equipment. See “Test Equipment Required” in section 3-4 for problem

isolation and resolution. Similar devices may be used if they equal or exceed performance standards of the equipment suggested.

CAUTION: Soldering and solder removal equipment must be low voltage operated and grounded to avoid static charge and stray current induced component damage. Maximum wattage: 25 W.

3. Clean the repair area. After soldering operations, clean off the repaired area with ethyl or methyl alcohol and a stiff hair brush. This will remove residual solder flux, making the repaired area more visible for an inspection and returning the instrument to its original, neat appearance. Removal of the flux will also facilitate making electrical measurements in the affected area as the flux itself is not conductive.

Exchange Program

A comprehensive circuit board and electro-mechanical module exchange program is available. The exchange circuits and modules are warranted, factory pre-tested and calibrated. Final calibration of the exchange item is strongly suggested to match the new

part to the host system. See section 4.4 for details concerning the exchange program.

Accutorr Plus Service Manual 0070-10-0691 3 - 3

Test Equipment Required

3.4

Repair Information

Test Equipment Required

Equipment types other than these listed may be utilized if they equal or exceed the listed equipment performance.

• Non-Invasive Blood Pressure Simulator

• Stopwatch

• SpO

2

Simulator

• RS232 Test Connector (pins 2 and 3 shorted), 9-pin male “D” and software download port tester

• Phono plug (tip and ring shorted)

• 700 cc Dummy Cuff (P/N 0138-00-0001-01)

• Chart Paper (P/N 0683-00-0447-01)

• Water Bath, stirred, heated 80° to 120° F

• Reference Thermometer, 0.1%

3 - 4 0070-10-0691 Accutorr Plus Service Manual

Repair Information Troubleshooting (Problem Isolation)

3.5

3.5.0.1

Troubleshooting (Problem Isolation)

This section has been divided into two major areas: clinical and technical. Clinical problems can generally be resolved by the user. Technical problems may require resolution by a qualified technical service person. The error code table below indicates

Technical problems with an asterisk (*). Error codes are displayed on the front panel

LEDs of the Accutorr Plus.

Error Codes

TYPE

NIBP

TEMP

(PTM)

SpO

2

8830

8831

8850

8851

8852

8853

8854

8856

8857

8857

8858

8858

CODE

8810

8811

8812

8813

DESCRIPTION REASON

Retry - Unable to Measure Motion artifact, cycle time-out, weak pulsations or no pulsations.

A triple beep tone is generated.

Retry - Pump Higher Insufficient cuff pressure. A triple beep tone is generated.

Stop - Cuff Overpressure Excessive cuff pressure detected by the software. A triple beep tone is generated.

Stop - Unable to Measure 4 successive measurement attempts failed. A triple beep tone is generated.

Check Probe

Replace Probe

Tissue contact may have been lost.

Defective probe or connection.

No Sensor

Sensor Off

No sensor connected.

(Masimo SpO

2

only)

Interference

Pulse Search

(Masimo SpO

2

only)

Unit cannot find signal. (Nellcor

®

SpO

2

module will report ‘‘Pulse

Search" -8853- when the sensor is not on the patient.)

Weak Pulse

(Masimo

®

SpO

2

only)

Check Sensor

(Masimo SpO

2

only)

PR<21

(Nellcor SpO

2

only)

PR<26

(Masimo SpO

2

only)

PR>249 bpm. (Nellcor SpO

2

only)

PR>239 bpm. (Masimo SpO

2

only)

Accutorr Plus Service Manual 0070-10-0691 3 - 5

Troubleshooting (Problem Isolation) Repair Information

3.5.1

TYPE

SYSTEM

988*

990*

991*

995*

996*

CODE

984*

985*

986*

987*

DESCRIPTION

NIBP Hardware Failure

NIBP Overpressure

Circuit not Programmed

NIBP Overpressure

Circuit not Tracking

Stop - Hardware

Overpressure

TEMP Bad Calibration

TEMP Illegal Mode

TEMP Module Failed

SpO

2

Uncalibrated

SpO

2

Failure

REASON

NIBP A/D failure detected.

The overpressure circuit is not set to the current patient size.

The two pressure transducers are not tracking each other.

Excessive cuff pressure detected by hardware overpressure sensor.

A triple beep tone is generated.

Thermometer needs calibration.

Thermometer switch is set wrong.

Thermometer internal failure.

SpO

2

fails calibration check.

SpO

2

failed self-test.

Isolating the Problem, System Level

1. Determine if the problem is in the main NIBP unit or in one of the add-on accessories.

2. If the problem is in the accessory, check the recorder unit for correct paper, correct installation, and paper jams.

3. If the problem is in the main NIBP and/or SpO

2

section, rule out these simple problems first:

If AC power is available, then the green LED adjacent to the front panel

ON/STANDBY keys should be illuminated. If not, connect the power cord to a live power outlet. If the green LED fails to illuminate, there is a probable power supply malfunction.

Check the battery operation LED. Steady ON indicates battery operation. A flashing

LED indicates a near depleted battery. If the green LED is NOT illuminated, then the battery or power supply/charger may be suspect.

Confirm correct operating parameters selection (i.e., correct patient size selection,

(Adult, Neonate, Ped.) and NIBP cuffs to match. In units with SpO

2 sensor is connected to the unit and it is of the proper type/size.

, verify that a

3 - 6 0070-10-0691 Accutorr Plus Service Manual

Repair Information Troubleshooting (Problem Isolation)

3.5.2

3.5.3

3.5.4

3.5.4.1

Isolating the Problem within the Main Unit

Problems within the main NIBP and/or SpO

2 supply problems and NIBP and/or SpO

2

unit are categorized broadly into power

detection and display. For a monitor that will not even turn on, proceed to Power Supply and Battery Tests, otherwise skip to Diagnostics

Test and beyond.

1. Power Supply and Battery test: The main power supply output voltages should be verified under normal instrument loads. Below are the tolerances for the voltages shown:

5V = 4.9 - 5.1

12.5V = 12 - 13

±15V = 14.5 - 15.5

2. Perform Diagnostics Test 01 through 12. Repair or replace circuits or components that fail the Diagnostic Test.

3. Trend circuits test: Obtain a minimum of five NIBP measurements and simultaneous

SpO

2

and Temperature readings if the Accutorr has those options. Confirm Trend storage function as explained in the operating instructions.

Isolating the Problems with Optional Accessory Modules

1. Predictive Temperature Module Test: Obtain one or more temperature readings from the optional temperature modules. They will be displayed in the temperature display window. If the readings are not displayed, replace the temperature probe first, then the complete module.

2. Recorder Module Test: Obtain a print out of stored data, or perform diagnostics test

(#4) to verify recorder integrity. Substitute recorder with a known good device to confirm diagnosis.

3. Infrared (AccuTemp IR) temperature module test: The following paragraphs will assist you to separate those clinical issues that can result in inaccurate or inconsistent readings, as well as instrument malfunctions.

Since there are no user serviceable components (except 9V battery) inside the

AccuTemp IR, all service requests should be directed to the factory.

Clinical Issues with AccuTemp IR Ear Thermometer Applications

Multiple Temperature Measurements

The AccuTemp IR employs an easy and quick method to taking temperature measurements. Because of this, it is tempting to take immediate repetitive measurements.

However, taking immediate repetitive measurements in the same ear may result in a decrease in temperature. This is because the probe of an ear thermometer is cooler than the ear canal and will draw the heat from the ear canal, cooling the canal slightly. Waiting a few minutes before taking another temperature in the same ear will give the ear canal time to recover, and the readings should be consistent with the initial measurement.

Accutorr Plus Service Manual 0070-10-0691 3 - 7

Troubleshooting (Problem Isolation) Repair Information

3.5.4.2

How the AccuTemp IR Measures Arterial Temperature (Adults/

Pediatrics)

The AccuTemp IR is a self-calibrating infrared ear thermometer that accurately measures the infrared heat energy radiating from the ear canal opening. The AccuTemp IR determines the highest temperature in the ear canal opening, measures the ambient temperature, and then calculates arterial (core) temperature through its patented heat balance system. This measurement will respond instantly to changes in arterial temperature without the delays and artifacts inherent in oral and rectal methods, providing the means to identify a fever faster and more reliably than other methods.

Temperature gradients in the ear canal vary with the amount of environmental exposure, perfusion patterns, ambient temperature, etc. The least variable and therefore most accurate is the small deep area of warmest, but still visible tissue. A conventional ear thermometer takes many readings of the same large area, which includes the high gradient distal ear tissue, resulting in low, non-reproducible, and inaccurate temperature readings.

The AccuTemp IR utilizes a narrow field-of-view and scanning capability to search the ear for the small deep area of warmest tissue required for heat balance calculation to ensure highly accurate and reproducible arterial temperatures.

The temperature measurement will remain on the display for a minimum of 55 and a maximum of 65 seconds after the Start button is released. If using the AccuTemp IR with and Accutorr Plus, the temperature measurement can be transferred from the AccuTemp

IR to the Accutorr Plus for display and entry in to the trend database, by placing the

AccuTemp IR in its holder within the 60 seconds after releasing the Start button. A beep tone is sounded from the Accutorr Plus once the temperature measurement information has been received by the Accutorr Plus.

NOTE:

Do not change Room Number and /or Bed Letter on the Accutorr Plus during the transmission of temperature data.

The AccuTemp IR can be used to take temperature measurements for up to 128 power cycles without having to be put back into its holder. After 128 power cycles, the

AccuTemp IR must be put back into the holder to be reset. When the 129th power cycle is reached and the Start button is pressed, 4 L’s are displayed, indicating that the AccuTemp

IR is locked and must be returned to its holder.

NOTE:

The AccuTemp IR thermometer should be at room temperature when it is used. If it has been in a very warm or very cold environment it may need time for the thermometer temperature to equalize to ambient temperature.

3 - 8 0070-10-0691 Accutorr Plus Service Manual

Repair Information Troubleshooting (Problem Isolation)

3.5.4.3

3.5.4.4

Troubleshooting

The following table provides suggestions to resolve problems when taking temperature measurements:

SYMPTOM

Various readings when taking measurements in the same location, ear or axilla.

Various readings when taking measurements in opposite ears.

No measurement or inaccurate measurements.

SUGGESTION

Quick repetitive measurements in the same location can result in a decrease in temperature. This can occur because the optical head of AccuTemp IR may slightly cool the area. Wait a few minutes for the temperature of the area to equilibrate. Replace cover and clean lens with an alcohol swab.

The patient may by lying down on one ear. The measurement of the ear that was down may be a higher temperature. Wait a few minutes for the temperature of the ear canal to equilibrate or use opposite ear. This is also a factor if the area has been covered for any reason (for example, with a hat, blanket, bandage, telephone, etc.).

This can be caused by technique errors:-Not keeping the button pressed throughout the measurement.-Not scanning side-to-side

(adults and pediatrics).-Twisting or rotating the probe instead of scanning side-to side.-Not changing the protective optical film cover with each reading. (Even on the same patient.)-Debris on the lens.-Not covering the opening of the ear canal.-Presence of a visible wax plug, or dense hair blocking the view into the ear canal.-This can also be caused by low battery voltage.

ACCUTEMP IR ERROR

AccuTemp IR double beeps and displays "——" while holding the

Start button. The AccuTemp IR then powers down.

AccuTemp IR double beeps while holding the Start button

AccuTemp IR displays "E-O0"or

"E-o1" while holding the Start button.

REASON

Battery voltage < 5.7 VDC. Battery requires replacement.

Low battery condition. Battery voltage is between

6.2 and 5.7 VDC.

Return AccuTemp IR for service.

Maintenance

The AccuTemp IR can be wiped down with any hospital approved disinfectant, including bleach. With normal use, the only maintenance required is to keep the lens on the end of the probe clean. The lens is made of polycarbonate. Dirt, greasy films or moisture on the lens will interfere with the passage of infrared heat and affect the accuracy of the measurement. If necessary, clean the lens with a cotton swab dipped in alcohol. Periodic cleaning is recommended. To keep the lens clean when not is use, store the AccuTemp IR with an optical film cover in place.

Accutorr Plus Service Manual 0070-10-0691 3 - 9

Disassembly Instructions

3.6

3.6.1

3.6.2

3.6.3

3.6.4

Repair Information

Disassembly Instructions

CAUTION: The Inside Of This Instrument Contains Static

Sensitive Components. Use correct static protection safeguards.

1. Power down the Accutorr Plus and remove the AC power cable.

2. Remove all cables and hoses from the front and rear of the instrument.

3. Remove temperature measuring module and recorder module, if equipped.

Removal of the Rear Housing

1. Place the Accutorr Plus with the display side down, onto a protective surface.

NOTE:

Special care should be taken to insure that the front panel and glare screen are not scratched.

2. Remove the 4 screws from the corners of the back housing recess and the two screws in the center channel of the rear case.

3. Slowly lift the rear housing up and place in a safe location.

4. Disconnect battery connector (J4) from the Power Supply board.

Removal of the Front Bezel

1. Place the front housing with the display side down, onto a protective surface.

2. Disconnect the SpO

2

ribbon cable (J3) from the SpO

2

circuit board and the tubing from the NIBP front panel connector.

3. Slowly separate the assembly up and away from the front housing.

Removal of the Keyboard Assembly

1. Place the instrument with the display side down, onto a protective surface.

2. Remove the Front Bezel, as described above.

3. Remove the screws (6) securing the Keyboard Assembly and lift the keyboard, supporting and separating the connector to the LED/Tone board.

Removal of the LED/Tone board

1. Remove front bezel and keyboard.

2. Disconnect cable from J4, the speaker connection.

3. Remove the four screws holding the circuit board. Pull circuit board forward, disconnecting the rear J-1 connector.

3 - 10 0070-10-0691 Accutorr Plus Service Manual

Repair Information

3.6.5

Disassembly Instructions

Removal of the CPU / NIBP board

1. Disconnect J-505 NIBP pump connector, J-7 SpO

2

interface connector (advanced model), and J-502 Power supply connector, all on the top edge of the board.

2. Disengage the captive cable connectors for the Recorder and Temperature options at the bottom left and right edges of the board.

3. Disconnect gray ribbon cable at J-603 to the external interface connectors.

4. Disconnect J-651 the LED backlight connector (advanced model).

5. Disconnect silicon hose from pump, at the right angle elbow.

6. Remove the four hexagonal spacers and one Phillips head screw on the board.

7. Lift board up gently and disconnect data ribbon cable from J-8 to the LCD display

(advanced model).

Accutorr Plus Service Manual 0070-10-0691 3 - 11

Disassembly Instructions

3.6.6

Repair Information

Removal of the Power Supply Assembly

1. Remove AC mains power cord and internal battery. Remove rear case.

2. Disconnect ribbon cable at J-3 on the power supply. Disconnect battery J-4 connector cable. Disconnect input cable at J-2.

3. Remove screws as shown in Figure 3-1.

3 - 12

FIGURE 3-1 Power Supply Assembly Diagram

0070-10-0691 Accutorr Plus Service Manual

Repair Information Disassembly Instructions

3.6.7

3.6.8

3.6.9

3.6.10

3.6.11

Removal of the LCD Display (Advanced Model)

1. Remove the right side support bracket for the LCD. Disconnect data ribbon cable.

2. Disengage LCD circuit board from left side support bracket and disconnect backlight cable from the CPU / NIBP board.

Removal and replacement of the Internal Li-Ion Battery

To remove the externally accessible battery, press the quick release tab on the battery access door and open door. The battery is retained internally in a sliding compartment, captured by a spring loaded release tab. Press the tab away from the battery; the battery will eject with a spring assist mechanism.

Reverse the process to install the replacement battery.

NOTE:

New batteries are shipped in a discharged state for safety reasons. A new battery must be charged for 4 hours for Li-Ion before first use. The monitor may be used with AC mains power during the charge cycle but battery operation could be limited during this time.

Removal of the AC Input Receptacle Assembly

1. Press the plastic locking tabs on the rear of the AC receptacle. A small screwdriver may be inserted at the cut out on the left of the AC connector, to release the locking tabs.

2. From the front side of the receptacle, starting at the edge closest to the Interface connector, press upwards, while pulling the connector out.

3. Disconnect the cable end from the Power Supply circuit board.

4. Remove the locking nuts from the safety ground terminal stud; remove ground conductor.

Thermal Printer (optional module)

1. Detach the complete module from the main NIBP section, by removing the two screws in the plastic housing. Grasp the plastic housing and pull to the rear to disengage the Interface connector.

2. Remove the two screws retaining the circuit board and the anti static shield assembly.

Remove the shield.

3. Remove the remaining two screws and lift the circuit board out.

Thermometer, Predictive (optional module)

1. Detach the complete module from the main NIBP section by removing the two screws in the plastic housing. Grasp the plastic housing and pull to the rear to disengage the interface connector.

Accutorr Plus Service Manual 0070-10-0691 3 - 13

Disassembly Instructions Repair Information

3.6.12

3.6.13

2. Remove J2 and J3 connectors from the circuit board to front panel input receptacle.

NOTE: The connectors may be reversed or interchanged without affecting final function.

3. Remove the three screws securing the circuit board and anti static shield. Remove the assembly.

AccuTemp IR, Infrared Thermometer (optional module)

The AccuTemp IR has no user accessible adjustments or replaceable parts, except for the 9

Volt battery. If problems arise, contact Technical Service.

AccuTemp IR Mounting Cradle

The mounting cradle has no adjustments or replaceable parts. The mounting cradle houses a small circuit board and a photocell sensor.

3 - 14 0070-10-0691 Accutorr Plus Service Manual

Repair Information Interconnect Cables Run List

3.7

3.7.1

3.7.1.1

Interconnect Cables Run List

Interface Specifications, Connectors and Functions.

Connector to LED / Tone PCB, J-2

27

28

29

30

23

24

25

26

31

32

33

34

19

20

21

22

15

16

17

18

11

12

13

14

7

8

9

10

5

6

3

4

PIN #

1

2

INPUT/OUTPUT

Output

Input/Output

Input

Input/Output

Output

Output

Input/Output

Output

Input

Input/Output

Input

Input

Input/Output

Input

Input

Input/Output

Output

N/C

Input/Output

Input

N/C

Input/Output

DESCRIPTION

Vcc return

Reset

Damped Data Bus D(0)

Write Recorder

Vcc return

FPGA Not Done

Damped Data Bus D(1)

Read/ Write

Vcc return

LDS

Damped Data Bus D(2)

Chip Select I/O

Vcc return

Write Delay

Damped Data Bus D(3)

ST Watchdog

Vcc return

LCD Angle

Damped Data Bus D(4)

LCD Angle Feed Back

Vcc return

LCD Angle Reference

Damped Data Bus D(5)

TONE

Vcc return

N/C

Damped Data Bus D(6)

Read Configure

Vcc return

N/C

Damped Data Bus D(7)

+5V Supply

Vcc return

+5V Supply

NAME

D(digital ground)

RST_CONN*

D_DAMP_CONN(0)

WR_RECD*

D(digital ground)

FPGA_NOT DONE*

D_DAMP_CONN(1)

R/W_CONN*

D(digital ground)

LDS*

D_DAMP_CONN(2)

CS_IO*

D(digital ground)

WR_DLY*

D_DAMP_CONN(3)

ST_WDOG*

D(digital ground)

LCD_ANGLE

D_DAMP_CONN(4)

LCD_ANGLE_FB

D(digital ground)

LCD_REF

D_DAMP_CONN(5)

TONE

D(digital ground)

N/C

D_DAMP_CONN(6)

RE_CONFG*

D(digital ground)

N/C

D_DAMP_CONN(7)

VCC

D(digital ground)

VCC

Accutorr Plus Service Manual 0070-10-0691 3 - 15

Interconnect Cables Run List

3 - 16

Repair Information

61

62

63

64

57

58

59

60

53

54

55

56

49

50

51

52

69

70

71

72

65

66

67

68

73

74

45

46

47

48

41

42

43

44

37

38

39

40

PIN #

35

36

N/C

Output

N/C

Output

N/C

N/C

Output

INPUT/OUTPUT

Input/Output

N/C

Output

Input

Output

N/C

N/C

Output

Input

N/C

Output

N/C

N/C

Output

N/C

Output

Output

Input

Input

Input

N/C

N/C

DESCRIPTION

Damped Data Bus D(8)

N/C

Vcc return

+5V Supply

Clock O

+5V Supply

Vcc return

Keypad Data Bus (D0)

Damped Address(2)

N/C

Vcc return

N/C

Damped Address(3)

Keypad Data Bus (D3)

Vcc return

N/C

Damped Address(4)

N/C

Vcc return

N/C

Damped Address(5)

N/C

Vcc return

AC ON

Damped Address(6)

Switched ON/OFF

Vcc return

N/C

Damped Address(7)

N/C

Vcc return

N/C

N/C

AC LED ON

+5V Supply

High Low

+5V Supply

LCD Unit

N/C

N/C

NAME

D_DAMP_CONN(8)

N/C

D(digital ground)

VCC

CLKO_CONN

VCC

D(digital ground)

KEYD0

A_DAMP_CONN(2)

N/C

D(digital ground)

N/C

A_DAMP_CONN(3)

KEYD3

D(digital ground)

N/C

A_DAMP_CONN(4)

N/C

D(digital ground)

N/C

A_DAMP_CONN(5)

N/C

D(digital ground)

ACON*

A_DAMP_CONN(6)

ON/OFFSW*

D(digital ground)

N/C

A_DAMP_CONN(7)

N/C

D(digital ground)

N/C

N/C

ACON_LED

VCC

HIGH/LOW*

VCC

LCD_UNIT*

N/C

N/C

0070-10-0691 Accutorr Plus Service Manual

Repair Information Interconnect Cables Run List

3.7.1.2

3.7.1.3

7

8

9

10

11

5

6

3

4

PIN #

1

2

12

13

14

8

9

10

11

6

7

4

5

2

3

PIN #

1

77

78

79

80

PIN #

75

76

INPUT/OUTPUT

N/C

Output

DESCRIPTION

SpO

2

RESET

+12V -15V ON

CHARGE OFF

N/C

A/D Chip Select

PWM CONTROL

Power Supply Header, J-502

DESCRIPTION

+5V Supply

+5V Supply

+5V Return

+5V Return

+15V Supply

-15V Supply

+/- 15V Return

+12V Return

+12V Supply

Standby

AC ON

+12V -15V ON

OFF/ON

Charge Off

J-2 Keypad Connector

FUNCTION

Battery LED

Interval LED

Neonatal LED

Pediatric LED

Adult LED

Memory Full LED

NIBP Ready LED

Print LED

Mute LED

Data Scan LED

AC On LED

22

23

24

25

26

18

19

20

21

PIN #

16

17

NAME

+5V

+5V

D(Digital Ground)

D(Digital Ground)

+15V

-15V

A(Analog Ground)

P(Power Ground)

+12V

STDBY

ACON*

12+-15ON*

OFF/ON*

CHRGOFF*

NAME

SpO2RST*

+12V -15V ON

CHRGOFF

N/C

ADC_CS*

PWM_CNTRL

FUNCTION

Key D7

Key D6

Key D5

Key D4

Key D3

Key D2

Key D1

Key D0

On/Off Sw*

Ground

LED_EN0

Accutorr Plus Service Manual 0070-10-0691 3 - 17

Interconnect Cables Run List Repair Information

3.7.1.4

3.7.1.5

PIN #

12

13

14

15

FUNCTION

AC On*

Key 2

Key 1

Key 0

J-4 Speaker Connector

2

3

PIN #

1

PIN #

27

28

29

30

FUNCTION

Speaker terminal

Speaker terminal no connection

FUNCTION

High_Low*

LCD_Unit*

LED_EN1

LED_EN2

Recorder Connector, J-5

7

8

9

10

11

5

6

3

4

PIN #

1

2

16

17

18

19

20

12

13

14

15

Output

Output

Output

Input

Output

Output

Output

Output

Output

INPUT/OUTPUT

Output

Output

Output

Output

Output

Output

Output

Output

Output

Output

Input

DESCRIPTION

TTL, recorder data

TTL, recorder data

TTL, recorder data

TTL, recorder data

TTL, recorder data

TTL, recorder data

NAME

Buffered Data Bus, DX<7>

Buffered Data Bus, DX<6>

Buffered Data Bus, DX<5>

Buffered Data Bus, DX<4>

Buffered Data Bus, DX<3>

Buffered Data Bus, DX<2>

TTL, recorder data

TTL, recorder data ground

TTL, recorder write logic low: Recorder module attached

TTL, Reset recorder

TTL, Recorder home

TTL, Run print head

TTL, DMA request

Buffered Data Bus, DX<1>

Buffered Data Bus, DX<0>

DGND

WR_RECD*

RECD*

RESETSRV*

HOME*

RUN*

DREQ*

+5V supply (+/-2%) VCC

+5V printer supply (+/-1V) +5V

+5V return PGND

+5V printer supply (+/-1V) +5V

+5V return PGND

3 - 18 0070-10-0691 Accutorr Plus Service Manual

Repair Information Interconnect Cables Run List

3.7.1.6

3.7.1.7

Thermometer Connector, J-4

6

7

8

9

10

4

5

2

3

PIN #

1

19

20

11

12

13

14

7

8

9

10

15

16

5

6

3

4

PIN #

1

2

17

18

N/C

N/C

N/C

N/C

Output

Output

Output

Output

Input

Input

INPUT/OUTPUT

Input

Input

Output

Output

Output

Output

Output

Input

Input

Input

DESCRIPTION

TTL, serial in

TTL, serial in ground ground

TTL, serial out

TTL, serial out

N/C

N/C

N/C

N/C

9 - 17.3 VDC ground

+5VDC +/- 2% supply

NAME

TEMP_TXD

TEMP_TXD

DGND

DGND

TEMP_RXD

TEMP_RXD

N/C

N/C

N/C

N/C

STDBY aground

VCC

VCC

PRED*

PRED* logic low: 0.8V max

Predictive Temp module attached ground logic low: 0.8V max

SpO

2

option 2 installed logic low: 0.8V max IR

Temp module attached logic low: 0.8V max

SpO

2

option installed

DGND

SpO2_OPT2*

TYMP*

SpO2_OPT1*

SpO

2

Connector, J-7

INPUT/OUTPUT

Output

Output

Output

Output

Input

Input

Output

Output

Output

Output

DESCRIPTION ground

TTL, module reset ground

TTL, serial out

SpO

2

module installed: logic low @ 0.8V max

TTL, serial in

+5V supply (+/-5%)

-15V supply (+/-5%)

Analog ground

+15V supply (+/-5%)

NAME

DGND

SpO2RST*

DGND

SpO2_RX

SpO2*

SpO2_TX

+5V

-15V

AGND

+15V

Accutorr Plus Service Manual 0070-10-0691 3 - 19

Interconnect Cables Run List Repair Information

3 - 20

3.7.1.8

3.7.1.9

3.7.1.10

3.7.1.11

NIBP Pump Connector, J-505

LED

PIN #

1

2

3

INPUT/OUTPUT

Output

Input

Input

DESCRIPTION pump supply pump return power ground

LED Backlight Connector, J-651

PIN #

1

2

INPUT/OUTPUT

Output

Input

DESCRIPTION

Min 3.4V - Max 3.6V

Return

Communication Connector, J-603

NAME

+12VF

PUMP*

P(power ground)

NAME

BL+

BL-

12

13

14

8

9

10

11

6

7

4

5

2

3

PIN # INPUT/OUTPUT

1 N/C

Input

Output

Output

N/C

Output

Output

Input

Output

Output

Output

Output

Input

Output

DESCRIPTION

N/C

RS-232

RS-232

Ground

N/C

Ground

RS-232, Transmit

RS-232, Receive

Ground

Ground

+7Vdc min

RS-232 or RS-485

RS-232 or RS-485

12VDC +/- 5%@ 100mA

LCD Connector, J-8

5

6

3

4

PIN # INPUT/OUTPUT

1 Out

2 Out

Out

Out

Out

Out

DESCRIPTION

Chassis Ground ground

+5V supply (+/-5%)

View angle control -15V max

TTL, control line

TTL, control line

NAME

CHASSIS

DGND

VCC

VEE

WR_LCD*

RD_LCD*

NAME

N/C

RTS

CTS

DGND

N/C

DGND

EXT_TXD

EXT_TXD

DGND

DGND

NCall

TXD

RXD

+12VT

0070-10-0691 Accutorr Plus Service Manual

Repair Information

3.7.2

3.7.2.1

3.7.2.2

21

22

23

24

25

26

17

18

19

20

13

14

15

16

9

10

11

12

PIN # INPUT/OUTPUT

7

8

Out

Out

In

Out

In/Out

In/Out

In/Out

In/Out

In/Out

In/Out

In/Out

In/Out

Out

N/C

Out

Out

Out

Out

Out

Out

DESCRIPTION

TTL, control line

TTL, address line

-10V supply

TTL, LCD reset

TTL, data line

TTL, data line

TTL, data line

TTL, data line

TTL, data line

TTL, data line

TTL, data line

TTL, data line ground

N/C

TTL Level

TTL Level

TTL Level

TTL Level

+15VDC

-15VDC

Power Supply Connectors

AC input connector, J-2

2

3

PIN #

1

FUNCTION

AC Ground

N/C (pin removed)

AC Neutral

PIN #

4

5

DC output connector, J-3

4

5

2

3

PIN #

1

FUNCTION

+5 V

+5 V

Common

Common

+15 V

PIN #

8

9

10

11

12

Accutorr Plus Service Manual 0070-10-0691

Interconnect Cables Run List

NAME

CS_LCD*

A_DAMP_CONN<2>

-10V

RESETSRV*

D_DAMP_CONN<0>

D_DAMP_CONN<1>

D_DAMP_CONN<2>

D_DAMP_CONN<3>

D_DAMP_CONN<4>

D_DAMP_CONN<5>

D_DAMP_CONN<6>

D_DAMP_CONN<7>

DGND

N/C

A_DAMP_CONN<3>

A_DAMP_CONN<4>

DSP_CONT*

A_DAMP_CONN<5>

+15VV

-15VV

FUNCTION

N/C (pin removed)

AC Hot

FUNCTION

Common

+12.5 V

STDBY

ACON*

+12-15ON*

3 - 21

Interconnect Cables Run List

3.7.2.3

PIN #

6

7

FUNCTION

-15 V

Common

Battery connector, J-4

Connector Pin Out

PIN #

1

2

3

CONNECTION

TS (temperature sense)

Battery +11.1 V

Common

PIN #

13

14

FUNCTION

OFF/ON*

CHRGOFF*

Repair Information

3 - 22 0070-10-0691 Accutorr Plus Service Manual

4.0

Replacement Parts and

Isometric Drawings

4.1

4.2

4.3

4.1 Introduction ...................................................................................... 4-1

4.2 Available Replacement Parts And Sub-Assemblies ..................... 4-1

4.3 Product Variations And Options .................................................... 4-1

4.4 Exchange Program .......................................................................... 4-2

4.5 Replacement Parts Pricing Information ........................................ 4-2

4.6 Ordering Information ....................................................................... 4-2

4.7 Isometric Drawings and Parts Lists ............................................... 4-3

4.8 Accessory Modules and Mounting Parts Lists ............................. 4-10

Introduction

This chapter of the Service Manual provides information necessary to identify the replacement parts and assemblies of the instrument.

Available Replacement Parts And Sub-Assemblies

The parts listings which follow are divided into two sections. The Isometric Drawings and the accompanying parts lists identify the available chassis mounted components.

Product Variations And Options

Product variations, due to differences for various line voltages, options and languages may require different components. These variations are reflected, where necessary, on the parts lists.

Accutorr Plus Service Manual 0070-10-0691 4 - 1

Exchange Program

4.4

4.5

4.6

Replacement Parts and Isometric Drawings

Exchange Program

An exchange policy for most of the printed circuit board assemblies is available. This program may provide the most expedient method of servicing the equipment. A standard charge for this service is made. Contact the Service Department for details concerning this exchange program.

Many circuit boards make extensive use of multi-layer technology and high density packaging. Individual component replacement is not recommended on these boards unless the technician is properly equipped to repair multi-layer circuit boards.

Circuit boards, returned as parts of the exchange program, that show evidence of improper repair techniques and are damaged in the process are not considered for exchange.

Damaged boards will be invoiced at full value and no exchange credit will be applied.

Replacement Parts Pricing Information

Current parts prices and exchange charges can be determined by contacting the Order

Entry Department.

Ordering Information

Replacement parts and assemblies are available from Mindray DS USA, Inc. Please follow these guidelines when ordering replacement items for the instrument.

1. Include the Model and Serial Number of the instrument. (The model number of the

Accutorr Plus starts with the numbers “0998-” and consists of 13 or 14 digits. All digits must be supplied.)

2. Include the Part Number exactly as it appears in the Parts List under the column, “Part

Number.”

3. Include a description of the item.

Example Orders:

NOTE:

(1) ea. P/N 0014-00-0083-02

Main Power Supply Board

Serial No. XXXX-XX

(2) ea. P/N 0212-12-0404

Screw, #4-40 x .25 lg. Pan Head,

Serial No. XXXX-XX

Mindray DS USA, Inc. maintains a policy of continuous development for product improvement and reserves the right to change materials, specifications, and prices without notice.

4 - 2 0070-10-0691 Accutorr Plus Service Manual

Replacement Parts and Isometric Drawings

4.7

Isometric Drawings and Parts Lists

Isometric Drawings and Parts Lists

The following figures show exploded views of the Accutorr Plus components, part numbers, and descriptions.

Accutorr Plus Service Manual 0070-10-0691 4 - 3

FIGURE 4-1

FIGURE 4-2

FIGURE 4-3

31

ITEM PART NUMBER

7

9

0012-00-1721

0012-00-1690

28

30

31

0012-00-1689

DESCRIPTION

Cable, LCD Interface

Cable, SpO2 to CPU/NIBP Interface

Cable, Power Supply to CPU/NIBP Interface

0670-00-0814-01 PCB, CPU/NIBP

0014-00-0083-02 Power Supply (Li-Ion)

FIGURE 4-4

FIGURE 4-5

FIGURE 4-6 Non-invasive Tubing Diagram

Accessory Modules and Mounting Parts Lists

4.8

4.8.1

4.8.2

Replacement Parts and Isometric Drawings

Accessory Modules and Mounting Parts Lists

Recorder Module Replacement Parts List

DESCRIPTION

Chart Paper Door

Chart Cutter

Recorder Main Housing

Printer Assembly with printhead

Recorder, complete

PART NUMBER

0380-00-0274

0380-00-0275

0380-00-0273-01

0161-00-0015

0998-00-0127

Predictive Temperature Module Replacement Parts List

DESCRIPTION

Temperature probe input connector

Temperature probe, Adult, Oral

Temperature probe, Adult, Rectal

Temperature probe covers, box of 500

Complete replacement Predictive

Temperature module with 2 probes and 1 box of probe covers

PART NUMBER

0012-00-1335

0206-00-0725-01

0206-00-0725-02

0198-00-0012-03

0998-00-0129

4 - 10 0070-10-0691 Accutorr Plus Service Manual

Replacement Parts and Isometric Drawings

4.8.3

Accessory Modules and Mounting Parts Lists

Universal Rolling Stand (P/N ACCTROLLSTD, Assembly Parts List)

3

1

2

CALLOUT

NUMBER DESCRIPTION

Basket

Casters, Non-Locking

Casters, Locking

Mounting Bracket for Accutorr

Plus, includes mounting screws

Complete Rolling Stand and

Mounting Bracket Assembly

PART NUMBER

0202-00-0166

0401-00-0045

0401-00-0046

0406-00-0855-01

ACCTROLLSTD

3

1

2

FIGURE 4-7 Accutorr Plus Rolling Stand (P/N ACCTROLLSTD)

Accutorr Plus Service Manual 0070-10-0691 4 - 11

Accessory Modules and Mounting Parts Lists

4.8.4

Mounting Options

Accutorr Plus

Wall Mount Bracket

0436-00-0247

Replacement Parts and Isometric Drawings

Accutorr Plus

(Basic/Advanced)

FIGURE 4-8 Accutorr Plus Mounting Options

Sample Wall Mount

4 - 12 0070-10-0691 Accutorr Plus Service Manual

5.0

Calibration

5.1 Introduction .................................................................................. 5-2

5.2 Precautions And Guidelines ....................................................... 5-4

5.3 Test Equipment And Special Tools Required ........................... 5-4

5.4 Power-up Sequence, Internal Testing ........................................ 5-4

5.5 Service Diagnostics .................................................................... 5-5

5.6 Predictive Thermometer Verification And Calibration ............. 5-16

5.7 Setting the Current Time ............................................................. 5-18

5.8 NIBP Normal Operation ............................................................... 5-19

5.9 Trend Memory Initialization ........................................................ 5-19

5.10 SpO2 Normal Operation (Accutorr Plus model with SpO2) ..... 5-20

Accutorr Plus Service Manual 0070-10-0691 5 - 1

Introduction Calibration

5.1

5.1.1

5.1.1.1

Introduction

The Accutorr Plus is a state of the art device employing digital determination and verification systems to obtain superior performance for the life of the product. Most function calibration constants are written into the operating software, therefore, for the most part, there are no means to calibrate or adjust system functions. The instrument contains extensive Service Diagnostics that will, in most cases, isolate a malfunction to the lowest serviceable module.

NOTE: This procedure refers to this instrument as the

UUT (Unit Under Test).

All internal verifications and adjustments should be performed with the internal battery disconnected. Operate the UUT on the main power supply. This will insure a reliable power source for the test, as well as providing a measure of safety in case of accidental overload of the power source. To re-initialize the low battery detector circuits, reconnect the battery at the conclusion of all tests while the UUT is on A.C. power source. Reset the real-time clock and date, if required.

Status and Error Code Table

The Accutorr Plus uses the various displays on the front panel to display the operational status. Error codes listed below can generally be resolved by the user however, error codes with an asterisk (*) may require resolution by a qualified technical service person.

Error Codes

Table 5-1

TYPE

NIBP

CODE

8810

8811

8812

8813

TEMP

(PTM)

8830

8831

DESCRIPTION REASON

Retry - Unable to

Measure

Motion artifact, cycle time-out, weak pulsations or no pulsations. A triple beep tone is generated.

Retry - Pump

Higher

Stop - Cuff

Overpressure

Insufficient cuff pressure. A triple beep tone is generated.

Excessive cuff pressure detected by the software.

A triple beep tone is generated.

Stop - Unable to

Measure

Check Probe

Replace Probe

4 successive measurement attempts failed. A triple beep tone is generated.

Tissue contact may have been lost.

Defective probe or connection.

5 - 2 0070-10-0691 Accutorr Plus Service Manual

Calibration Introduction

Table 5-1

TYPE

SpO

2

CODE

8850

8851

8852

8853

SYSTEM

988*

990*

991*

995*

996*

8854

8856

8857

8857

8858

8858

984*

985*

986*

987*

DESCRIPTION REASON

No Sensor

Sensor Off

No sensor connected.

Interference

Pulse Search

Sensor not on patient. (Masimo® SpO

2

only)

Interference on signal. (Masimo® SpO

2

only)

Unit cannot find signal. (Nellcor® SpO

2 report

module will

“ Pulse Search ” -8853- when the sensor is not on the patient.)

Weak Pulse

Check Sensor

PR<21

Weak pulse detected. (Masimo® SpO

2

only)

Sensor problem. (Masimo® SpO

2

only)

Pulse rate is less than 21 bpm. (Nellcor® SpO only)

2

PR<26

PR>249

PR>239

Pulse rate is less than 26 bpm. (Masimo® SpO

2 only)

Pulse rate is greater than 249 bpm. (Nellcor® SpO only)

2

Pulse rate is greater than 239 bpm. (Masimo®

SpO

2

only)

NIBP A/D failure detected.

NIBP Hardware

Failure

NIBP

Overpressure

Circuit not

Programmed

NIBP

Overpressure

Circuit not

Tracking

Stop - Hardware

Overpressure

The overpressure circuit is not set to the current patient size.

The two pressure transducers are not tracking each other.

Excessive cuff pressure detected by hardware overpressure sensor. A triple beep tone is generated.

Thermometer needs calibration.

TEMP Bad

Calibration

TEMP Illegal

Mode

TEMP Module

Failed

SpO

2

Uncalibrated

SpO

2

Failure

Thermometer switch is set wrong.

Thermometer internal failure.

SpO

2

fails calibration check.

SpO

2

failed self-test.

Accutorr Plus Service Manual 0070-10-0691 5 - 3

Precautions And Guidelines

5.2

5.3

5.4

Calibration

Precautions And Guidelines

In the event that the instrument covers are removed, observe these following precautions and general guidelines:

• Do not short component leads together.

• Perform all steps in the order given.

• Use extreme care when reaching inside the opened instrument. Do not contact exposed metal parts which may become live.

Read through each step in the procedure so it is understood prior to beginning the step.

Test Equipment And Special Tools Required

Equipment types other than these listed may be utilized if they equal or exceed the listed equipment performance.

• Non-Invasive Blood Pressure Simulator

• Stopwatch

• SpO

2

Simulator

• RS232 Test Connector (pins 2 and 3 shorted), 9-pin male “D” and software download port tester

• Phono plug (tip and ring shorted)

• 700 cc Dummy Cuff (P/N 0138-00-0001-01)

• Chart Paper (P/N 0683-00-0447-01)

• Water Bath, stirred, heated 80° to 120° F

• Reference Thermometer, 0.1%

Power-up Sequence, Internal Testing

1. Attach the AC power cord into the UUT rear panel AC Input module.

2. Press the ON/STANDBY key to turn the UUT On. The UUT automatically performs internal self diagnostics. The Timer / Temp. display starts a count down from 20. This is the initial software verification. Allow the unit to complete the start up sequence until there is a single LED light steadily illuminated above the Patient Setup key.

After the initial power up sequence, “a” is displayed in the bed LED, and “0” is displayed in the Room LEDs.

3. Press the ON/STANDBY key to turn the UUT off.

5 - 4 0070-10-0691 Accutorr Plus Service Manual

Calibration

5.5

5.5.1

Service Diagnostics

Service Diagnostics

The Service Diagnostics tests must be run in the sequence indicated. The numbers and letters in parentheses after the title of each test, represent the room number and bed letter used to access the test.

Introduction (Hidden Key)

To enter the Service Diagnostics mode, while powering on the unit, press and hold the hidden key until two beeps are heard. The hidden key is the Adult symbol above the

Patient SETUP key.

Verify that after the initial power up sequence a “0” is displayed in the Room LED, “a” is displayed in the Bed LED, and “ddd” is displayed in the Systolic LEDs.

Accutorr Plus Service Manual 0070-10-0691 5 - 5

Service Diagnostics

5.5.2

Software Version Test (0a, 0b)

This test displays the installed software version (s) in the monitor. The results are

displayed in the read outs illustrated in Figure 5-1 on page 5-6.

Calibration

Installed Software

Example: AO1.02

Software Module

Selection

Example: ACM

Runtime software identifier

FIGURE 5-1 - Software Version Test

Confirm that the ROOM number is at “0”; otherwise use the “ROOM/BED” key to select the “ROOM”, and the UP/DOWN keys to set Room number to “0”.

Press ROOM/BED key to select BED. Use the UP/DOWN ARROW keys to select test

“a” or “b”. Monitors without SpO

2 option may not display results for tests “0c” and “0d”.

Press ROOM/BED again to lock in selection.

Press the START NIBP key.

Press the DEFLATE key until a beep is heard. This allows a selection of the next test.

Repeat process for test “b”.

Table 5-2

DESCRIPTION OF SOFTWARE MODULE

ACM Runtime Software Version

ACM Boot Software Version

0

0

TEST ROOM

NUMBER a b

SUB TEST

PATIENT BED

LETTER

5 - 6 0070-10-0691 Accutorr Plus Service Manual

Calibration

5.5.3

Service Diagnostics

Keypad Test

1.

Press the Room Up arrow to change the Room Number to “1”.

2.

Press the Start NIBP key to run the keypad test.

3.

Press any key on the keyboard. A number will appear in the Patient Bed Letter

Display. This number is identified in Figure 5-1 on page 5-6. Verify that the correct

number appears in the Patient Bed Letter Display when each key is pressed.

Keyboard Test,

Accutorr Plus (Advanced Model)

FIGURE 5-2

4.

Press the DEFLATE key for 3 seconds to exit keypad test.

Accutorr Plus Service Manual 0070-10-0691 5 - 7

Service Diagnostics

5.5.4

5.5.5

5.5.6

Calibration

LED Test (2a, 2b)

This test allows the LED function to be verified by driving all seven LEDs with numbers from 0 to 9, in sequential order. For this test, all annunciator LEDs turn on for 1 second.

Table 5-3

TEST DESCRIPTION

7 Segment LED Test

Annunciator LED Test

TEST ROOM NUMBER

2

2 a b

SUB TEST PATIENT

BED LETTER

Communication Test (3a, 3b)

This test checks the integrity of the external communication systems. Test 3a checks the external RS232 interface and requires that the transmit and receive pins be connected together. Test 3b checks the Download port and requires that the tip (transmit) and ring

(receive) pins be connected together. The CRC table is transmitted and received. If the

CRC data received matches the CRC table transmitted, the test passes; otherwise a failure is reported.

Table 5-4

TEST DESCRIPTION

RS232 Port Test

Download Port Test

SHORT

CONNECTOR PINS

Transmit (2) and

Receive (3)

Tip and Ring

TEST ROOM

NUMBER

3

3

SUB TEST

PATIENT BED

LETTER a b

Recorder Test (4)

1. Select Room Number 4 and press the NIBP START key to activate the test. This test prints out a pattern on the recorder if a recorder module is attached. A strip similar to

Figure 5-3 will be printed.

FIGURE 5-3

5 - 8 0070-10-0691 Accutorr Plus Service Manual

Calibration Service Diagnostics

5.5.7

5.5.7.1

Pump Test (5)

NOTE: 700 cc dummy cuff must be used.

Select Room Number 5 and press NIBP START key to activate the test.

This test evaluates the ability of the pump to pressurize a fixed volume representative of an adult cuff to a typical adult pressure.

A fixed volume of 700 cc is pumped to a pressure of 300 mmHg. While the pump is operational, the elapsed time is displayed in the Timer/Temperature display. When the target pressure is reached, the elapsed time display holds its current reading and the pump deflates the chamber for another test. The test passes if the chamber is inflated in less than

35 seconds. Pressing the DEFLATE key at any time during the test causes the NIBP module to open all valves to release the pressure, clear the MAP display, and exit the test.

Pressing the DEFLATE key after the test is completed will clear the MAP display and exit the test.

Bleed Rate Test (6a, 6b)

NOTE: 700 cc dummy cuff must be used.

Press the NIBP START key to activate the selected test. See Table 5-5 on page 5-9.

This test evaluates the ability of the instrument to maintain a constant bleed rate, for the purpose of accurate and repeatable NIBP measurements. This test also evaluates the ability of the system to reduce cuff pressure quickly.

A fixed volume of 700 cc is pumped to approximately 20 mmHg over the specified start

pressure (see Table 5-5 on page 5-9), then vented to atmosphere through one of two

different valves. During the test, the MAP display indicates the real-time system pressure and the Timer/Temperature display indicates the elapsed time in seconds. When the stop pressure is reached, the MAP display is blanked and the elapsed time displays the current reading. Pressing the DEFLATE key deflates the chamber, clears the MAP display, and exits the test. Each test is uniquely identified in the Patient Bed Letter Display as

described in Table 5-5 on page 5-9. The table identifies each test, the corresponding start

and stop pressures and required completion times.

Table 5-5

TEST

DESCRIPTION

Linear Bleed Valve

Test

Dump Valve Test

START

PRESSURE

(MMHG)

180

STOP

PRESSURE

(MMHG)

60

250 20

TIME

(SEC.)

15-25 secs.

14 sec. max.

TEST

ROOM

NUMBER

6

6

SUB

TEST

PATIENT

BED

LETTER a b

Accutorr Plus Service Manual 0070-10-0691 5 - 9

Service Diagnostics

5.5.8

5.5.9

Calibration

Leak Test (7)

NOTE: 700 cc dummy cuff must be used.

Select Room Number 7 and press the NIBP START key to activate the test.

This test evaluates the ability of the system to maintain a steady pressure.

A fixed volume is pumped up to an initial differential pressure of approximately 250 mmHg and monitored for a fixed period of time. While the pump is operational, peak pressure is displayed in the MAP display. When the target pump pressure is reached, the peak pressure is frozen in the MAP display during a five second dwell. During leakage measurement, the current pressure leakage is displayed in the MAP display and the elapsed test time (in seconds) is displayed in the Temperature/Time/Interval display. After the leakage measurement time has elapsed, the MAP and Temperature/Time/Interval displays are blanked and a “00” (Pass) or “FF” (Fail) will be displayed in the Bed window.

Pressing the DEFLATE key at any time during the test causes the NIBP module to open all valves to release the pressure, clear the MAP display, and end the test. The maximum pressure drop is no more than 10 mmHg in any 90 second period when connected to a 700 cc volume. Pressing the DEFLATE key after the test is completed will clear the MAP display and exit the test.

Over Pressure Test (8a, 8b, 8c)

NOTE: 700 cc dummy cuff must be used.

Select Room Number and Bed Letter per Table 5-6 on page 5-10. Press the NIBP START

key to activate the test.

This test evaluates the ability of the system to prevent an over pressure condition, independent of software control, for each of 3 over pressure settings. A fixed volume of

700 cc’s is pumped to a target pressure that is above the maximum over pressure specified

for the respective over pressure limit (see Table 5-6 on page 5-10), while displaying the

real-time system pressure in the MAP display. For this test, the software over pressure monitor is disabled. On completion, the MAP display indicates the maximum pressure reached and the Patient Bed Letter Display indicates if the over pressure sensor was activated (00) or not (FF). Pressing the DEFLATE key at any time during the test causes the NIBP module to open all valves to release the pressure and end the test.

Following each overpressure hardware test, the UUT must be turned off and rebooted in the diagnostic mode in order to re-enable the pump.

Table 5-6

TEST SIZE/MODE

Low

Medium

High

TARGET PUMP

PRESSURE (MMHG)

157, ±5

210, ±10

315, ±15

8

8

TEST ROOM

NUMBER

8 a b

SUB TEST PATIENT

BED LETTER c

5 - 10 0070-10-0691 Accutorr Plus Service Manual

Calibration Service Diagnostics

5.5.10

5.5.11

5.5.12

Pulse Channel DC Offset Test (9a, 9b, 9c)

Select Room Number and Bed Letter per Table 5-7 on page 5-11, and press the NIBP

START key to activate the test.

This test measures the DC offset for each of the three electronic gain settings available on the pulse channel. Results for each gain setting are displayed in the Timer/Temperature

display (in Volts). Table 5-7 on page 5-11 describes the tests performed.

Press the DEFLATE key after each test to exit.

Table 5-7

TEST DESCRIPTION

Gain 1 (Low) Test

Gain 2 (Med) Test

Gain 3 (High) Test

ACCEPTABLE

LIMIT (V)

1.50 to 1.83

1.50 to 1.83

1.50 to 1.83

9

9

TEST ROOM

NUMBER

9 a b

SUB TEST PATIENT

BED LETTER c

Pulse Channel Average Noise Test (10a, 10b, 10c)

Select Room Number and Bed Letter per Table 5-8 on page 5-11, and press the NIBP

START key to activate the test.

This test measures the average noise for each of three electronic gain settings available on the pulse channel. Results for each gain setting are displayed in the Timer/Temperature

display (in millivolts). Table 5-8 on page 5-11 describes the tests performed.

Press the DEFLATE key after each test to exit.

Table 5-8

TEST DESCRIPTION

Gain 1 (Low) Test

Gain 2 (Med) Test

Gain 3 (High) Test

ACCEPTABLE

LIMIT (MV)

<25

<50

<75

TEST ROOM

NUMBER

10

10

10 a b

SUB TEST PATIENT

BED LETTER c

Main Pressure Transducer Verification Test (11a, 11b, 11c)

NOTE: 700 cc dummy cuff must be used.

This tests the calibration of the main pressure transducer in LOW, MED, and HIGH modes. This is done by comparing the readings of the UUT to an external pressure gauge in LOW mode below 100 mmHg, in MED mode between 100 mmHg to 200 mmHg, and in HIGH mode between 200 mmHg to 300 mmHg. In the event of a disagreement between the device and an external gauge, the transducer must be calibrated by performing the

calibration in section 5.5.15 “Main Pressure Transducer Calibration” on page 5-14.

1. Select diagnostic test “11” (Room) “a” (Bed).

Accutorr Plus Service Manual 0070-10-0691 5 - 11

Service Diagnostics Calibration

5.5.13

2. Press the START NIBP key to start the LOW calibration test. Verify that the pump comes on, the pressure increases, and then the pump stops. After 10 seconds, verify that the pressure settles below 100 mmHg. Compare the final pressure displayed to an accurate external pressure gauge. The MAP LEDs reading should agree within ±1% with the external pressure gauge. If it does not agree, a failure has occurred and

proceed to section 5.5.15 “Main Pressure Transducer Calibration” on page 5-14.

3. Press the DEFLATE key to release the pressure and exit the test.

4. Select diagnostic test “11” (Room) “b” (Bed).

5. Press the START NIBP key to start the MED calibration test. Verify that the pump comes on, the pressure increases, and then the pump stops. After 10 seconds, verify that the pressure settles between 100 mmHg and 200 mmHg. Compare the pressure displayed to an accurate external pressure gauge. The pressure MAP LEDs reading should agree within ±1% with the external pressure gauge. If it does not agree, a

failure has occurred and proceed to section 5.5.15 “Main Pressure Transducer

Calibration” on page 5-14.

6. Press the DEFLATE key to release the pressure and exit the test.

7. Select diagnostic test “11” (Room) “c” (Bed).

8. Press the START NIBP key to start the HIGH calibration test. Verify that the pump comes on, the pressure increases, and then the pump stops. After 10 seconds, verify that the pressure settles between 200 mmHg and 300 mmHg. Compare the final pressure displayed to an accurate external pressure gauge; the MAP LEDs reading should agree within ±1% with the external pressure gauge. If it does not agree, a

failure has occurred and proceed to section 5.5.15 “Main Pressure Transducer

Calibration” on page 5-14.

9. Press the DEFLATE key to release the pressure and exit the test.

Verification of Accutorr Plus Pneumatic Performance using an NIBP

Simulator (11d)

Select Room Number 11 and Bed Letter d, and press the NIBP START key to activate the test.

For these tests, program the NIBP simulator to perform the desired test. Instead of pumping up the cuff, this test uses the NIBP simulator to verify the Accutorr Plus pressure reading for the NIBP Leak Test, Over Pressure test, and the Main Pressure Transducer

Verification Test. The Accutorr displays the pressure in the MAP display. Pressing the

DEFLATE key at any time during the test causes the NIBP module to open all valves, clear the MAP display, and exit the test. The NIBP simulator must be exited separately.

NOTE: The reading and setting accuracy will be dependent on the accuracy of the NIBP simulator and the condition of the Accutorr Plus.

The NIBP standard dynamic pressure simulation may produce Systolic / Mean / Diastolic read outs on the Accutorr Plus outside of the specified limits. The pressure simulation results may not be accepted as a measure of the Accutorr Plus clinical performance.

5 - 12 0070-10-0691 Accutorr Plus Service Manual

Calibration Service Diagnostics

5.5.14

Overpressure Transducer Verification <12c, 12a, 12b>

This test verifies that the overpressure transducer is calibrated correctly. Calibration is checked in the Low level pressure mode so that the limit displayed in the MAP window should say 158 mmHg (±1). If it does not, the transducer needs to be calibrated by

proceeding to 5.5.16 “Overpressure Transducer Calibration” on page 5-15. The

overpressure transducer should also be verified in the High and Mid level pressure modes.

1. Select diagnostic test “12” (Room) “a” (Bed). Low Limit

2. Press the START NIBP key to start the test.

3. Verify that the overpressure value displayed in the MAP LEDs is 158 mmHg (±1).

4. Repeat steps 1 through 3, but select test “12” “c” to check overpressure limit, 315 mmHg (±15); and test “12” “b” for Mid level overpressure limit, 210 mmHg (±10).

5. Press the DEFLATE key to exit this test.

Accutorr Plus Service Manual 0070-10-0691 5 - 13

Service Diagnostics Calibration

5.5.15

Main Pressure Transducer Calibration

NOTE: This procedure should only be performed in the

event that there was a failure in 5.5.12 “Main

Pressure Transducer Verification Test (11a, 11b,

11c)” on page 5-11.

700 cc dummy cuff must be used.

NOTE:

This is the procedure for calibrating the main pressure transducer in the event that there is a disagreement between the UUT and the external pressure gauge in the proceeding test.

The device is calibrated by comparison with the NIBP simulator at 158 mmHg (±1) and

then by adjusting R540 on the 0670-00-0814 PCB (see Figure 5-5).

1. Connect the NIBP simulator to the Accutorr Plus cuff connector, and connect a 700 cc test chamber to the open port of the simulator. Select the manometer function of the simulator.

2. Select diagnostic test “11” (Room) “d” (Bed).

3. Press the START NIBP Key to start HIGH calibration procedure.

4. Pump up the NIBP simulator to 158 ±1mmHg.

5. Allow pressure to stabilize for 10 seconds. Read the pressure in the MAP display.

6. Adjust R540 on the 0670-00-0814 PCB until the MAP readout is the same as the simulator’s manometer reading.

7. Press the DEFLATE key to exit this test.

8. Repeat 5.5.12 “Main Pressure Transducer Verification Test (11a, 11b, 11c)” on page 5-11.

5 - 14

FIGURE 5-4 Pressure Adjust (R540)

0070-10-0691 Accutorr Plus Service Manual

Calibration Service Diagnostics

5.5.16

Overpressure Transducer Calibration

NOTE: This procedure should only be performed if a

failure occurred in steps 1 through 3 of 5.5.14

“Overpressure Transducer Verification <12c, 12a,

12b>” on page 5-13.

This is the procedure for calibrating the overpressure transducer in the event that the

display does not read 158 mmHg during the test for Low pressure mode in 5.5.14

“Overpressure Transducer Verification <12c, 12a, 12b>” on page 5-13. The transducer is

calibrated using R504 on the 0670-00-0814 PCB (see Figure 5-5).

1. Adjust R504 on the 0670-00-0814 PCB until the MAP readout is 158 mmHg.

2. Repeat 5.5.14 “Overpressure Transducer Verification <12c, 12a, 12b>” on page 5-13.

FIGURE 5-5 Overpressure Adjust (R504)

Accutorr Plus Service Manual 0070-10-0691 5 - 15

Predictive Thermometer Verification And Calibration

5.6

5.6.1

5.6.2

Calibration

Predictive Thermometer Verification And Calibration

The Water Bath Method, utilizes laboratory grade stirred water bath to confirm that the

PTM is operating within the specified limits.

In case of failure or inaccuracy of readings, the System Calibration Procedure may be followed to re-adjust the device to factory specifications.

This procedure requires high accuracy test equipment and therefore it is not recommended for field calibration. A comprehensive factory service program to support this product is available.

Water Bath Method

1. Prepare a water bath set to 89.6 to 110.0°F as a temperature reference. Confirm temperature with a separate precision thermometer.

2. Snap a protective sheath onto the temperature probe, and immerse into the reference water bath for 30 seconds, or until the temperature display stops flashing. Record the temperature shown on the temperature display.

Temperature Verification Test, Infrared Thermometer (Applicable

Only To Units Equipped With A Temperature Option)

1. Pre-heat water bath to 100°F. Place floating Black Body device into water bath and allow to stabilize until step 7, or a minimum of 30 seconds.

2. Prepare the thermometer for the verification process, by removing any disposable probe covers. Clean the probe lens with a swab dipped in alcohol.

3. Completely cover probe tip with a piece of aluminum foil, from the lens to the probe neck portion. Allow 5 minutes for the probe and foil to stabilize at ambient temperature.

4. Activate the thermometer, by pressing the button. Read the ambient temperature displayed and record the number as T

A

=____.

5. Remove the aluminum foil.

6. Check the water bath temperature against a laboratory grade mercury thermometer, with a verified accuracy of 0.4 °F (0.2 °C) or better. Record the water bath temperature as TT=____.

7. Attach a disposable probe cover to the thermometer head. Insert the head into the floating Black Body cavity, such that the instrument is looking into the needle shaped end.

5 - 16 0070-10-0691 Accutorr Plus Service Manual

Calibration

5.6.3

Predictive Thermometer Verification And Calibration

8. Press the button to measure the temperature. Record the temperature as

T

DCALC

=____.

9. Select the appropriate correction factor from the following choices:

10. Core temperature correction: Kcore=1.087

11. Oral temperature correction: Kora=1.058

12. Rectal temperature correction: Krectal=1.095

13. The standard infrared thermometer is calibrated to indicate Core temperature, thus the correction factor to be used is K=1.087.

14. Calculate the display temperature according to the following formula:

K(T

T

-T

A

)+ T

A

= T dcalc

where:

T dcalc

is Calculated Display Temperature.

Kcore=1.087 Correction Factor for Core Temperature.

T

T

is Target Temperature.

T

A

is Ambient Temperature.

Sample Calculation:

Ambient temperature (T

A

) measured: 72 °F

Water Bath temperature (T

T

) measured: 98.6 °F

Correction Factor (K core

)=1.087

1.087 (98.6-72) + 72= 100.9 °F thus,

Temperature displayed by thermometer is 100.9 °F, and calculated temperature is

100.9 °F, the accuracy has been verified.

Temperature value deviations: The calculated value is a function of the accuracy of the water bath temperature. As an example, a ±1% reading error of water temperature will result in a ±1 °F reading input into the formula, resulting in a final calculated values of

99.9 to 101.9 °F.

With a display resolution of ±1 digit and an over all resolution of ±0.2 °F, the final, calculated reading may fall between 98.6 and 102.2.

Low Battery Sensing

The low battery threshold is not adjustable. When the battery has reached the low battery threshold, the power supply will report this condition to the CPU on the Accutorr Plus.

Accutorr Plus Service Manual 0070-10-0691 5 - 17

Setting the Current Time

5.7

Calibration

Setting the Current Time

1. Press and hold the TIMER/TEMP key for 6 seconds. The hour digit only displays.

2. Press the UP or DOWN ARROW key to change the number.

3. Press the TIMER/TEMP key to activate the minute display.

FIGURE 5-6 Hour Display

NOTE:

The Accutorr Plus always displays time in a 24 hour format.

4. Press the UP or DOWN ARROW key (12 or 13) to change the number. Continue pressing the TIMER/TEMP key and the ARROW keys to set the month, day, and year (in that order).

FIGURE 5-7 Minute Display

5 - 18 0070-10-0691 Accutorr Plus Service Manual

Calibration

5.8

5.9

NIBP Normal Operation

NIBP Normal Operation

1. Press the left ARROW key on the NIBP simulator to highlight SELECT BP, then press the ENT key. Set pressure test to 120/80 (90) and press the ENT key.

2. Select a NIBP simulation of 120/80 mmHg with a Heart rate of 60 BPM.

3. Attach the NIBP simulator to the input pressure fitting of UUT.

4. Press the INTERVAL key of the keypad until a 1 is shown in the Timer/Temp./

Interval LED of the UUT. Press the PATIENT SIZE key to select Adult size indicator.

5. Press the Set ALARMS key to enter the alarm setting mode. Use the UP and DOWN

ARROW keys to set the Systolic Low alarm to 130 mmHg.

6. Press the Set ALARMS key again and set the Diastolic Low Alarm to 95 mmHg.

7. Press the Start NIBP key. Verify the monitor takes a measurement and the Systolic and the Diastolic low alarms are violated. The LEDs for these parameters will blink on and off and an alarm tone will sound for the violation. Press the MUTE Key to silence the alarms.

8. After the fifth measurement is complete, allow the cuff to pump up. Press DEFLATE key. Verify cuff deflates.

9. Review the data that was stored in the database, by scrolling through the patient trend information. Use the UP and DOWN ROOM and BED keys or LED display of the

UUT and verify that the Systolic and Diastolic information is available in the LEDs for each stored settings

10. Remove the cuff from the UUT and press the START key. Confirm that the UUT pump runs and that it displays 8810 in the Room/Bed window within 60 seconds and attempts another inflation. Press the DEFLATE key to stop test.

Trend Memory Initialization

1. To clear the UUT Trend Memory for all patient settings, press and hold the DELETE

INFO. key while powering up the unit.

Accutorr Plus Service Manual 0070-10-0691 5 - 19

SpO

2

Normal Operation (Accutorr Plus model with SpO

2

)

5.10

Calibration

SpO

2

SpO

2

)

Normal Operation (Accutorr Plus model with

The purpose of this test is to verify that an SpO

2

signal is recognized, alarm limits are captured and the data is trended. A full operational test may be conducted, if so warranted, within the limits of the published specification. See chapter 3 of this manual for specifications.

1. Set the simulator to an SpO

2

saturation level of 98%, pulse rate of 60 bpm, and a pulse amplitude of 100%.

2. Set the Accutorr Plus SpO

2

HI alarm limit to 95% and the pulse rate HI alarm limit to

55 bpm.

3. Connect an NIBP cuff or a suitable simulator to the Accutorr Plus and start a measurement.

4. Confirm that the Accutorr Plus has determined an SpO

2

saturation level and has violated the SpO

2

HI alarm and the pulse rate HI alarm (flashing digits and alarm tone). Press the MUTE key to silence the alarms.

5. Check the trend screen to confirm that the measurement information has been saved.

6. Set the SpO

2

HI alarm to OFF and SpO

2

LOW alarm to 86.

7. Set the simulator to an SpO

2

saturation level of 90%, pulse rate of 95 bpm, and the pulse amplitude to approximately 10% (low perfusion Masimo® SpO

2

only).

8. The Accutorr Plus may intermittently determine SpO

2

values or indicate a status code of 8853 (Pulse Search) or 8854 (Weak Pulse / Low Perfusion will display). Adjust the simulators pulse amplitude up and down until either status code of 8853 or 8854 displays.

9. Restore normal simulation parameters as described in step 1.

10. Remove the sensor from the simulator to induce a status code of 8851 (Sensor Off)

(Masimo® SpO

2

only).

11. Disconnect the sensor from the Accutorr Plus to produce a status code of 8850 (No

Sensor).

5 - 20 0070-10-0691 Accutorr Plus Service Manual

6.0

Preventive Maintenance

6.1

6.2

6.1 Introduction ...................................................................................... 6-1

6.2 Limitations Of Physiological Simulators ....................................... 6-1

6.3 Preventive Maintenance Schedule ................................................. 6-2

Introduction

The intent of this procedure is to ensure that the product continues to be safe and effective in its intended environment. Complete knowledge of the instrument characteristics and its relationship to the clinical environment are useful in determining changes in the performance of the monitor, and the clinical limitations of the device.

As a general rule, the instrument is not tested to the full limits of it’s capabilities at each service interval. Rather, representative points are tested and the results recorded. It is important to note that while not all functions are tested to the absolute limits, every function and option should be exercised. Declining performance over time is frequently the first indication of the need for more extensive service action.

Limitations Of Physiological Simulators

The physiological simulator is the principle aid for problem solving. Simulators have many of the same limitations as the devices that they will test. It is imperative that the simulator correlate well to a “Reference Standard*”. The electronic simulators available attempt to replicate a broad spectrum of indices representative of normal and abnormal physiological events, within the interpretation of the manufacturer.

Accutorr Plus Service Manual 0070-10-0691 6 - 1

Preventive Maintenance Schedule

6.3

6.3.1

6.3.2

Preventive Maintenance

Most simulators (properly maintained) will provide reasonably accurate and consistent results, but may not produce the exact results when connected to similar monitors from different manufacturers. However, the consistent performance of a simulator is a good indicator of any change in the performance in the monitor.

* For example, an accurate mercury column for pressure verification.

Preventive Maintenance Schedule

The following preventive maintenance steps are required for continued satisfactory performance and safety of the Accutorr Plus and optional accessories. Inspections and replacement of consumable supplies and accessories that are subject to normal wear, must be accomplished at least as frequently as the inspection of the host unit. Read the Warranty statement in the Operating Instructions for description of warranty conditions.

The suggested minimum maintenance schedule is based on normally expected wear and tear of system and components. The intervals may be shortened at the user’s discretion when conditions of use may warrant. In case of internal instrument failure, component replacement, and possible or actual instrument damage, it is strongly recommended that complete performance and safety verification be conducted.

Mechanical and Physical Visual Inspection (One Year Interval)

Check the following items for wear and physical damage: Repair or replace as needed.

• Outer case of Accutorr Plus, Recorder and Temperature module.

• Patient connected accessories such as blood pressure cuffs and hoses and related connectors. Check the SpO

2

sensor and cables. Check the temperature cables and probes. Verify and replenish the disposable items as needed. (i.e.,: probe covers, recorder chart supply).

• Power cords, external safety ground connections and communication cables.

Electrical Safety and Performance Checks (One Year Interval)

• Line power and battery power indicator lights functional in the proper modes.

• Line power and safety grounds are intact and within specifications.

• Non-invasive blood pressure section is functional within specifications.

• SpO

2

functions are operational and within specification.

• Temperature functions are operational and within specifications.

• Recorder is functional and documents available stored data.

• Check recorder paper path and clean as required.

• Error codes, failure messages, end user reported anomalies are collected and analyzed for reduced performance trends. Identify and correct causes as warranted.

• Perform ground resistance and chassis leakage tests.

6 - 2 0070-10-0691 Accutorr Plus Service Manual

Preventive Maintenance Preventive Maintenance Schedule

This page intentionally left blank.

Accutorr Plus Service Manual 0070-10-0691 6 - 3

0070-10-0691 REV C July 12, 2010

advertisement

Related manuals

Download PDF

advertisement