Bovie MI-750, Wall Mount, Floor Model, Single Ceiling Mount, Dual Ceiling Mount Surgical Light Installation and Service Manual

Bovie MI-750, Wall Mount, Floor Model, Single Ceiling Mount, Dual Ceiling Mount Surgical Light Installation and Service Manual
Add to My manuals

Below you will find brief information for Surgical Light MI-750, Surgical Light MI-750 Wall Mount, Surgical Light MI-750 Floor Model, Surgical Light MI-750 Single Ceiling Mount, Surgical Light MI-750 Dual Ceiling Mount. The MI-750 series of surgical lights are designed for procedural medical lighting in a variety of settings. This light system is mounted centrally in a room such that access is available all around the operating environment. The lamp head can be rotated approximately 190°. An ON/OFF button is provided with 3 levels of dimming.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

MI-750 Installation and Service Manual | Manualzz

INSTALLATION AND

SERVICE MANUAL

MI-750

MI-750

Agency Approvals

Medical Electrical Equipment

With respect to electric shock, fire

And mechanical hazards only

In accordance with IEC60601-1:2006 / CAN/CSA C22.2

No.601.1-M90 with updates 1 & 2

Classifications:

1. Protection against electrical shock (5.1, 5.2). Class I permanently connected,

2. Protection against harmful ingress of water (5.3). None.

3. Degree of safety in the presence of flammable anesthetics or oxygen (5.5). Not suitable for use in the presence of flammable anesthetics or oxygen.

4. Mode of operation (5.6). Continuous

5. Surgical luminaries (IEC60601-2-41)

Electromagnetic compatibility for immunity

And emissions in accordance with

EN-60601-1-2(2001) Class B and CISPR 22 (1997) Class B

Medical Electrical Equipment

Particular requirements for the safety of surgical luminaires and luminaires for diagnosis

In accordance with IEC-60601-2-41

Intended use:

Procedural Medical Lighting for Hospital, Clinic, Veterinary, Minor Surgery,

Examination or Diagnosis, within suitable facilities designated for such purposes. This light system is mounted centrally in a room such that access is available all around the operating environment.

User interface:

The MI-750 lighting system is a procedural light intended to be used by medical professionals in examination rooms. The functional interfaces are 600° positioning horizontally and ± 40° vertically. The lamp head can be rotated approximately 190°. An

ON/OFF button is provided with 3 levels of dimming.

Misuses:

For any purpose or use at any facility other than stated above. Caution shall be exercised when positioning the light head and avoid contact or collision with the patent, other medical professionals or other lights/equipment.

MI-750 Installation and Service Manual (1003136 Rev A) Page 2

MI-750

Installation and Service Manual

Table of Contents

Section 1: Terms, Symbols & Warranty ................................................................................. 5

Definition of Terms .................................................................................................................. 5

List of Symbols ....................................................................................................................... 7

Disposal of Waste ................................................................................................................... 8

MI-750 Models ........................................................................................................................ 9

Limited Warranty ................................................................................................................... 10

Section 2: Specifications Mechanical Specifications ......................................................... 12

Electrical Specifications ........................................................................................................ 12

Optical Specifications............................................................................................................ 13

Environmental Specifications ................................................................................................ 13

Section 3: Installation Ceiling Calculation: Single Mount Dimensions ............................. 14

Ceiling Calculation: Double Mount Dimensions .................................................................... 15

Ceiling Mount Pre-Installation Guidelines ............................................................................. 16

Ceiling Mount Pre-Installation Guidelines (cont’d) ................................................................ 17

Extended Ceiling Mount Installation ...................................................................................... 19

Single Ceiling Mount - General Information .......................................................................... 21

Double Ceiling Mount Installation ......................................................................................... 24

Wall Mounted Light Installation ............................................................................................. 29

Floor Stand Light Installation ................................................................................................ 33

Light Head to Arm Installation ............................................................................................... 37

Section 4: Operating Instructions Light Head ..................................................................... 39

Light Head Intensity Indicators, Control and Indicator Display .............................................. 40

Floor Model Positioning and Detachable Parts ..................................................................... 41

Section 5: Warnings & Safety Instructions ......................................................................... 42

Section 6: Maintenance ......................................................................................................... 44

Fuse Replacement ................................................................................................................ 44

LED Maintenance ................................................................................................................. 44

Arm Adjustment .................................................................................................................... 45

Head/Yoke Adjustment ......................................................................................................... 46

Handle Sterilization ............................................................................................................... 47

Cleaning Instructions ............................................................................................................ 48

Maintenance Schedule ......................................................................................................... 49

Section 7: Troubleshooting .................................................................................................. 50

MI-750 Installation and Service Manual (1003136 Rev A) Page 3

MI-750

Installation and Service Manual

Table of Illustrations

Figure 1: Single Ceiling Mount Ceiling Rod Calculation ........................................................ 14

Figure 2: Double Ceiling Mount Ceiling Rod Calculation ...................................................... 15

Figure 3: Recommended Ceiling Structure Construction ...................................................... 17

Figure 4: Ceiling Casting Mount Diagram ............................................................................. 18

Figure 5: Extended Ceiling Mount Installation ....................................................................... 19

Figure 6: Single Ceiling Mount Components ......................................................................... 20

Figure 7: Single Ceiling Mount Installation ............................................................................ 22

Figure 8: Double Ceiling Mount Components ....................................................................... 24

Figure 9: Double Ceiling Mount Installation .......................................................................... 25

Figure 10: Wire Harness Positioning..................................................................................... 27

Figure 11: Wire Channel ....................................................................................................... 27

Figure 12: Wall Mount Dimensions ....................................................................................... 29

Figure 13: Wall Mount Installation ......................................................................................... 30

Figure 14: Wall Bracket Mounting Diagram .......................................................................... 30

Figure 15: Stud Anchor Installation ....................................................................................... 32

Figure 16: Floor Mount Dimensions ...................................................................................... 33

Figure 17: Floor Stand Base Assembly................................................................................. 34

Figure 18: Arm Assembly to Upright Pole Installation ........................................................... 35

Figure 19: Light Head-Arm Installation.................................................................................. 37

Figure 20: Light Head to Arm Electrical Connections ........................................................... 38

Figure 21: Light Head Components ...................................................................................... 39

Figure 22: Arm Adjustment ................................................................................................... 45

Figure 23: Head / Yoke Adjustment ...................................................................................... 46

Figure 24: Handle Sterilization .............................................................................................. 47

MI-750 Installation and Service Manual (1003136 Rev A) Page 4

Section 1: Terms, Symbols & Warranty

Definition of Terms

Section 1

I.E.C.

International Electrotechnical Commission

ETL

Edison Testing Laboratories

Medical Electrical Equipment

Electrical equipment intended to diagnose, treat the patient under medical supervision.

Electrical equipment that transfers light energy to the patient.

Central Illuminance

Illuminance of light head measured at 1 meter from the light emitting area with no obstructions.

The value is expressed in Foot-candles or Lux.

Light Field Center

The point of maximum illuminance in lighted area. This is the reference point for light field size and light distribution measurements.

Light Field Diameter

The diameter of the circle where illuminance reaches 10% of light field center illuminance.

Depth of Illumination

The distance above and below 1 meter to where the central Illuminance is reduced to 20%.

Shadow Dilution

The ability of the equipment to minimize the impact of shadows in the working area due to partial obstruction by the operator or other medical personnel.

Correlated Color Temperature (CCT)

The color temperature of the light fixture when compared to a blackbody radiator expressed in degrees Kelvin.

Total Irradiance

The total amount of energy imparted to the patient by the lighting system expressed in

Watts/meter squared.

Color Rendering Index (CRI)

A method of how well a light source will render other colors when illuminating them based upon eight CIE chromaticity coordinates measured with a spectro-radiometer.

Sterilizable Handle

An easily removable device that when properly sterilized maintains a sterile area in order to handle it under aseptic conditions when attached to the equipment.

Head/Yoke Assembly

The part of the device which includes the light source, heat removal system and light focusing system.

MI-750 Installation and Service Manual (1003136 Rev A) Page 5

Section 1

Definition of Terms (cont’d)

Protective earth ground

The conductor used to connect the non-current-carrying metal parts of equipment, raceways, and other enclosures to the system grounded conductor, the grounding electrode conductor, or both, of the circuit at the service equipment or at the source of a separately derived system.

Off Center Moment

The unit of measurement for torque which is caused by an off-center load. This is measured in foot-pounds.

Arm Assembly – Extension/Articulating Arm

Horizontal section of the positioning arm with pivots that is used to increase the area covered by the light head and articulating arm. The articulating arm allows for vertical positioning of the light head.

Light Mounting

Support apparatus used to connect arm assembly/light head to a fixed surface, consisting of either a single, double or triple ceiling mount.

Neutral Conductor (common)

In an AC circuit, the return line for current.

Means of Isolation

The Means of Isolation, disconnects (Isolates) the light from the source of power. This isolation makes it safe to work on the light. In the case of a ceiling mounted light, the means of isolation is the circuit breaker that supplies the power to the circuit. In the case of the Wall, floor or Floor UPS, the means of isolation is the cord, unplugging the cord isolates the light from power.

MI-750 Installation and Service Manual (1003136 Rev A) Page 6

List of Symbols

ETL Listing marking

Read accompanying documents

C.E. Marking

Fuse marking

Protective earth ground

Neutral conductor

Caution

Electric shock hazard

MI-750 Installation and Service Manual (1003136 Rev A)

Section 1

Page 7

Disposal of Waste

Section 1

This product must not be disposed of with your other waste. Instead, it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment, or by returning it to Medical Illumination

International, Inc for reprocessing. The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. For more information about where you can drop off your waste equipment for recycling, please contact your local city office, your waste disposal service, or your product distributor or retailer.

MI-750 Installation and Service Manual (1003136 Rev A) Page 8

Section 1

MI-750 Models

The following MI-750 models are covered in this manual:

061412 Wall Mount ....................................................................... 100-240 VAC, 50/60 Hz, 65 W

061413 Floor Model ...................................................................... 100-240 VAC, 50/60 Hz, 65 W

061414 Single Ceiling Mount ........................................................ 100-240 VAC, 50/60 Hz, 65 W

061415 Dual Ceiling Mount ........................................................ 100-240 VAC, 50/60 Hz, 130 W

MI-750 Installation and Service Manual (1003136 Rev A) Page 9

Section 1

Medical Illumination International, Inc.

Limited Warranty

This document comprises the general terms of your product’s Limited Warranty. By accepting the shipment of the product, the owner/purchaser agrees to adhere to the warranty terms and conditions expressed herein.

Medical Illumination International, Inc. (“Medical Illumination”) Lighting Equipment is warranted against defective material and/or workmanship, excluding normal replacement parts (e.g. bulbs, sterilizable handles or glass items), for a period of three (3) years from the date of shipment. This

Limited Warranty applies exclusively to the repair or replacement of parts recognized as defective by Medical Illumination that are in normal use and have not been modified or repaired by unauthorized personnel.

This Limited Warranty extends only to the first retail purchaser of a product, and is not transferable or assignable. This warranty supersedes all other guarantees or warranties, expressed or implied.

WARRANTY SERVICE & REPAIRS

Medical Illumination does not provide (or give any compensation for) outside repair services or field labor. Therefore, in the event of a failure covered under this warranty, please take the following immediate action:

1. Contact Medical Illumination via phone at (818) 838-3025, through our website at http://www.medillum.com, or by facsimile at (818) 838-3725.

A. Be prepared to give the model number, serial number, and full description of the failure.

B. Our Customer Service department will attempt to solve the problem over the phone. If it becomes necessary to send the product to the factory for repair, you will be provided with a

Return Authorization number. Products sent to the factory without a Return Authorization number will not be accepted.

2. It is the retail purchaser’s obligation to arrange for shipment return of a product to the factory for warranty service, which shall be at the retail purchaser’s expense. Carefully package the light component (light head, arm assembly, mount assembly, etc.) and return it, freight prepaid and insured, with the Return Authorization number clearly marked on the outside of the box, to:

Medical Illumination

547 Library Street

San Fernando, CA 91340

RA#______

Damage resulting from inadequate packing is not covered by this warranty, and shipping insurance does not cover damage due to inadequate packing. We recommend that the package be insured against loss or in-transit damage. Medical Illumination cannot be held responsible for in-transit loss or damage. In the event that freight-related damage should occur, Medical Illumination will notify you immediately so that you can file a damage claim with the proper freight carrier.

MI-750 Installation and Service Manual (1003136 Rev A) Page 10

Section 1

Medical Illumination

Limited Warranty (cont’d)

Within the aforementioned time period of three (3) years from date of shipment, Medical

Illumination will evaluate the returned product, repair as appropriate, and ship the product back to you with freight costs prepaid by the Company. In the event that non-warranty damage or failure is discovered, you will be contacted before any repairs are performed.

EXCLUSIONS

This Limited Warranty does not cover the following:

• Any field labor or outside services (electricians, contractors, installation services, routine maintenance or other repair services)

• Damage to the product resulting from tampering, accident, abuse, negligence, alteration, or other causes unrelated to problems with material and/or workmanship

• Damage due to improper installation, use, cleaning or maintenance, as outlined in the

Installation and Service Manual for the product

• Labor costs associated with removing, re-packaging for shipment or reinstalling product

PRODUCT RETURNS:

Please contact the authorized dealer from whom the product was purchased to inquire about a product return. Additional terms and conditions set by the dealer may apply for any returned items.

SHIPMENT DAMAGE:

If the initial shipment of your purchased product arrived in damaged condition, please leave the packaging and its contents intact and contact Medical Illumination immediately.

MI-750 Installation and Service Manual (1003136 Rev A) Page 11

Section 2

Section 2: Specifications

Mechanical Specifications

Parameter

Weights:

Light Head/Yoke Assembly

Arm

Arm (Floor)

Wall Bracket Assembly

Floor Stand Assembly

Single Ceiling Mount Assembly

Dual Ceiling Mount Assembly

Dimensions:

Light Head/Yoke Assembly

Arm (Wall/Ceiling)

Arm (Floor)

Floor Stand

Value

Approximately 8.6 lbs (3.9 Kg)

Approximately 12.5 lbs (5.7 Kg)

Approximately 11.5 lbs (5.2 Kg)

Approximately 3.0 lbs (1.4 Kg)

Approximately 43.0 lbs (19.5 Kg)

Approximately 14.5 lbs (6.6 Kg)

Approximately 16.0 lbs (7.3 Kg)

14.5” (368 mm) Dia. 3.28” (83.3 mm) deep

58.5” (1,485 mm)Long

39.5” (1,003 mm) Long

75.5” (1,917 mm) High

Rotations:

Articulating Arm Vertical Movement

Articulating Arm Horizontal Movement

Articulating Arm/Yoke Interface

Yoke/Lamp Head Interface

± 40° Travel

Approximately 345°

Approximately 540°

Approximately 190°

Electrical Specifications

Parameters Value

Supply Circuit

100 ~ 240 VAC 50/60 Hz

Single Phase (NOT 208 compatible)

0.5 AMP (Dual Lights)

Fuse Rating

Power Supply power handling capacity

Light Head power consumption

1.5 Amps @ 230VAX, Slo-Blo (all supply voltages)

100 ~ 240 VAC 50/60 Hz

1.3 Amps, 65W

27W @ 24VDC

MI-750 Installation and Service Manual (1003136 Rev A) Page 12

Optical Specifications

Performance Value

Color temperature

Focal length

Central illuminance (full intensity)

Dimming – 3 Intensity Levels

(Indicated by 3 LED’s on 2

indicator strips located 180° apart)

Light field diameter (d10)

Light field diameter (d50)

Depth of illumination

Shadow Dilution

Illuminance (one mask)

Illuminance (two masks)

Illuminance at bottom of standard tube

Illuminance (tube with one mask)

Illuminance (tube with two masks)

Irradiance

CRI

4,300° Kelvin ±300°K

39.4” (1 meter)

50,000 +/- 5K LUX

100% 75% 50%

50,000 37,500 25,000

Approximately Ø 9.0” (Ø 229mm)

Approximately Ø 5.0” (Ø 127mm)

Approximately 50”

Approximately 240 LUX

Approximately 23,960 LUX

Approximately 51,950 LUX

Approximately 100 LUX

Approximately 23,070 LUX

Approximately 141.2 W/m

≥ 87

2

Environmental Specifications

Parameter

Ambient / Operating temperature

Storage temperature range

Humidity

Pressure

Value

41

F to 104F 5 to 40° Celsius)

41

F to 113F (5 to 45° Celsius)

10 - 90% relative humidity

100 kPA (@<2K meters)

Section 2

MI-750 Installation and Service Manual (1003136 Rev A) Page 13

Section 3

Section 3: Installation

Ceiling Calculation: Single Mount Dimensions

Use the following table to select the correct length ceiling rod for your application.

Ceiling Mounting

Height (Y-Value)

8’0” - 8’9”

8’10” - 9’7”

9’8” - 10’4”

10'5" - 11'1"

11’2" - 11'11"

12'0" - 12'3"

12'4" - 13'1"

13’2" - 14'0"

Ceiling Rod

Length

16”

25”

35”

44”

53”

63”

67”

77”

“X” Value

22”

31”

41”

50”

59”

69”

73”

83”

Head room to bottom of extension arm

(y-value – x-value)

6'2" – 6'11”

6'3" - 7'0"

6'3" – 6'11”

6'3" – 6'11”

6'3" - 7'0"

6'3" - 6'6"

6'3" - 7'0"

6'2" - 7'1"

NOTE: Longer ceiling rods are available for higher ceilings.

Figure 1: Single Ceiling Mount Ceiling Rod Calculation

MI-750 Installation and Service Manual (1003136 Rev A) Page 14

Section 3

Ceiling Calculation: Double Mount Dimensions

Use the following table to select the correct length ceiling rod for your application.

Ceiling Mounting

Height (Y Value)

8'0" -8'10"

8’11” - 9’8”

9’9” - 10’5”

10'6" - 11'2"

11'3" - 12'0"

12'1" - 12'3"

12'4" - 13'2"

13'3" - 14'0"

Ceiling Rod

Length

16”

25”

35”

44”

53”

63”

67”

77”

“X” Value

23”

32”

42”

51”

60”

70”

74”

84”

Head room to bottom of extension arm

(y-value – x-value)

6'3" - 6'11"

6'3" - 7'0"

6'3" - 6'11"

6'3" - 6'11"

6'3" - 7'0"

6'3" - 6'5"

6'3" - 7'0"

6'1" - 7'0"

NOTE: Longer ceiling rods are available for

Figure 2: Double Ceiling Mount Ceiling Rod Calculation

higher ceilings.

MI-750 Installation and Service Manual (1003136 Rev A) Page 15

Section 3

Ceiling Mount Pre-Installation Guidelines

SPECIAL NOTE: Installation and repair of this equipment should be performed by qualified persons only. Medical Illumination International, Inc. does not warranty any damage occurring as a result of improper installation. It is recommended that this installation manual be completely reviewed prior to installation.

Before installation, check to insure the following minimum conditions are provided:

Ceiling Mount Weight: Lb Moment: Ft Lb

Single Ceiling 70 215

Double Ceiling 115 350 ← recommended

Medical Illumination strongly recommends that the ceiling structure be designed to the weights and moments for the worst case (double ceiling). Designing for the heaviest model with the highest load/torque will provide adequate support for the heaviest, and will not significantly increase the cost of the lightest. The design margin will also allow for flexibility for future product upgrades.

A structural mount that does not meet these minimum conditions can cause serious injury and/or property damage.

A sloped or vaulted ceiling will require a level mounting surface be constructed that meets the above listed requirements.

The ceiling structure must be strong enough to support the weight and rigid enough to constrain rotation to less than 0.1° at the ceiling casting.

The equipment may be mounted directly over a 4-0 junction box. Input power supply lines should be wired in accordance with all applicable building codes.

The Single Ceiling mount supply circuit line must have as a minimum:

100-240VAC, 50/60 Hz, single phase, three wire, capable of supplying 65 Watts @ 1.5A.

The Dual Ceiling mount supply circuit line must have as a minimum:

100-240VAC, 50/60 Hz, single phase, three wire, capable of supplying 130 Watts @ 3A.

The equipment is not deemed compatible with any sort of electrical dimming device. Use line voltage only.

The power supply circuit line must be routed and wired to the wire harness in compliance with all applicable building codes.

Failure to provide a circuit meeting these minimum standards or complying with local building codes can cause a shock hazard.

WARNING: Secure the source of power using the means of isolation, before proceeding with electrical work.

MI-750 Installation and Service Manual (1003136 Rev A) Page 16

Section 3

Ceiling Mount Pre-Installation Guidelines (cont’d)

Check the length of the ceiling rod supplied to make sure that it is the proper length to install and operate the light without interference or over reach

The Ceiling Mount system will experience various level of dynamic off center moment during regular use. Therefore, it is crucial that the ceiling structure be strong enough to uphold the weight of the system and support the positioning arms and light head without deflection. The owner and/or owner’s contractor has the final responsibility for the strength and rigidity of the ceiling structure. An inadequate ceiling may result in unintended drift and/or equipment damage.

Because the ceiling structure is the owner’s responsibility, the design and construction recommendation shown below covers only one of the many possible alternatives that can be used. Medical Illumination highly recommends that the owner consult a structural engineer prior to designing and installing the ceiling structure.

Recommended Ceiling Structure Construction Details

The illustrations below are suggested mounting schemes per 2001 California Building Code –

Section 1632A: Anchorage and Seismic. For any other mounting scheme, please consult a structural engineer and/or professional contractor for the best solution for your situation.

Installation and repair of this equipment should be performed by qualified persons only.

Medical Illumination does not warranty any damage occurring as a result of improper installation.

Figure 3: Recommended Ceiling Structure Construction

MI-750 Installation and Service Manual (1003136 Rev A) Page 17

Section 3

Ceiling Structure Construction and Mounting (cont’d)

Improper fastening of the ceiling casting can cause serious injury and/or property damage.

Mount the Ceiling casting to the ceiling structure using four 3/8” DIA A325 bolts with nuts and washers.

Safety Label

Figure 4: Ceiling Casting Mount Diagram

WARNING: To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth.

WARNING: Secure the source of power using the means of isolation, before proceeding with electrical work.

MI-750 Installation and Service Manual (1003136 Rev A) Page 18

Extended Ceiling Mount Installation

For 10’2” – 14’ (3.23m-4.27m) Ceilings

Section 3

Lower Ceiling

Rod

Figure 5: Extended Ceiling Mount Installation

Insert the ceiling rod up through the ceiling casting until the hole in the ceiling rod becomes visible. Insert the dowel pin into the hole in the ceiling rod and lower the rod in the ceiling casting, making sure the pin is seated securely in the indentation on the ceiling casting.

Securely tighten the two set-screws located in the casting.

Failure to install the dowel pin can cause the arm/head assembly to fall from the ceiling causing serious injury and/or property damage.

The ceiling casting must be properly grounded during installation. This can be done by mounting the casting to a suitable material that will function as a ground conductor, or a wire lead that must be attached to the ground screw on the casting then routed to proper ground.

Note: Ceiling rod must be plumb. Adjust the nuts/washers accordingly.

Note: The ceiling casting itself must be electrically grounded to maintain proper grounding reliability.

MI-750 Installation and Service Manual (1003136 Rev A) Page 19

Section 3

Extended Ceiling Mount Installation (cont’d)

If ceiling rod is not plumb, unwanted arm drifting may occur.

Feed the wire harness through the ceiling rod and route to a junction box. Leave sufficient wire to extend slightly beyond the bottom of the ceiling rod. Important: To achieve proper

grounding reliability, the green ground wire from the wire harness MUST be properly fastened to the grounding screw located on the ceiling casting. Make all electrical connections in compliance with all applicable electrical codes.

Failure to comply with local electrical codes can cause a shock hazard.

WARNING: Secure the source of power using the Means of Isolation, before proceeding with electrical work.

Slide the casting cover up the rod extension and over the casting. Similarly, slide the collar to the casting cover, then tighten the set screw to hold the cover in place.

Feed the wire harness through the lower ceiling rod with the rod in the direction shown. Slide ceiling rod up through ceiling rod extension until the holes in the ceiling rod align with the upper holes in the ceiling rod extension. Insert a 1/4-20 x 2” bolt carefully, making certain not to damage the wire harness, and then fasten the cap nut. Secure rods in place with the two set screws provided.

Figure 6: Single Ceiling Mount Components

MI-750 Installation and Service Manual (1003136 Rev A) Page 20

Section 3

Single Ceiling Mount - General Information

Installation and repair of this equipment should be performed by qualified persons only. Medical Illumination does not warranty any damage occurring as a result of improper installation.

The shipping cartons contain a light head assembly, an arm assembly, a ceiling casting, casting cover, collar, bolt cover, ceiling rod, hardware kit, wire harness, and an Installation and

Service Manual.

Notes:

There are 3 standard length rods for different ceiling heights. Extended ceiling rod kits are available for ceilings over 10’2”. Verify that your ceiling rod length is correct for your ceiling height (See Ceiling Rod calculations on Page 14 & 15). If not correct please contact customer service.

When removing parts from the shipping carton, be careful not to damage the components. Important: thoroughly check each box for parts that may be located in areas that can be overlooked.

MI-750 Installation and Service Manual (1003136 Rev A) Page 21

Section 3

Single Ceiling Mount Installation (cont’d)

Figure 7: Single Ceiling Mount Installation

 Insert the ceiling rod up through the ceiling casting until the hole in the ceiling rod becomes visible (see Figure 7: “Single Ceiling Mount Installation”). Insert the dowel pin into the hole in the ceiling rod and lower the rod in the ceiling casting, making sure the pin is seated securely in the indentation on the ceiling casting. Securely tighten the two set-screws located in the casting.

Failure to install the dowel pin can cause the arm/head assembly to fall from the ceiling causing serious injury and/or property damage.

MI-750 Installation and Service Manual (1003136 Rev A) Page 22

Section 3

Single Ceiling Mount Installation (cont’d)

The ceiling casting must be properly grounded during installation. This can be done by mounting the casting to a suitable material that will function as a ground conductor, or a wire lead that must be attached to the ground screw on the casting then routed to proper ground.

Note: Ceiling rod must be plumb. Adjust the nuts/washers accordingly.

Note: The ceiling casting itself must be electrically grounded to maintain proper grounding reliability.

If ceiling rod is not plumb, unwanted arm drifting may occur.

 Feed the wire harness through the ceiling rod extension and route to a junction box.

Leave sufficient wire to extend slightly beyond the bottom of the ceiling rod. Important:

To achieve proper grounding reliability, the green ground wire from the wire harness MUST be properly fastened to the grounding screw located on the ceiling casting. Make all electrical connections in compliance with all applicable electrical codes.

Failure to comply with local electrical codes can cause a shock hazard.

WARNING: Secure the source of power using the means of isolation, before proceeding with electrical work.

 Slide the casting cover up the rod over the casting. Similarly, slide the collar to the casting cover, then tighten the set screw to hold the cover in place. Feed the wire harness through the ceiling rod with the rod in the direction shown.

Do not install the extension arm with the light head attached. Installation with the light head attached can cause damage to the light. Refer to the procedure for installing the head to the arm after the bracket and arm are installed.

 Slide the bolt cover up onto the ceiling rod as shown in Figure 7, then raise the extension arm to the ceiling rod and plug the Molex connectors together. Push the wire into the ceiling rod while inserting the bushing on the extension arm assembly into the rod. Secure the assembly to the ceiling rod using the four supplied bolts as shown. Two 1/4-20 x 1/4" screws must be installed on one side of the ceiling rod. Two 1/4-20 x 3/8” screws and lock washers must be used to mount the stop bracket on the other side of the rod. Be sure the bend on the stop faces down, towards the arm.

Failure to install the stop clip subassembly can cause damage to the light fixture.

Failure to install or tighten the 1/4-20 socket head cap screw can cause the arm/head assembly to fall causing serious injury and/or property damage.

 Complete the installation by lowering the bolt cover onto the extension arm assembly.

 See the instructions for “Section 3: Light Head to Arm Installation” on page 37 to

complete the assembly.

MI-750 Installation and Service Manual (1003136 Rev A) Page 23

Double Ceiling Mount Installation

Section 3

Figure 8: Double Ceiling Mount Components

GENERAL INFORMATION

Installation and repair of this equipment should be performed by qualified persons only. Medical Illumination does not warranty any damage occurring as a result of improper installation.

The shipping cartons each contain a light head assembly and an arm assembly. One of the cartons also contain a ceiling casting, casting cover, collar, ceiling rod, double mount assembly, hardware kit, wire harness, and an Installation and Service Manual.

Notes:

There are 3 standard length rods for different ceiling heights. Extended ceiling rod kits are available for ceilings over 10’6”. Verify that your ceiling rod length is correct for your ceiling height (see “Ceiling Rod Calculation” on page 15). If not correct please contact customer service.

When removing parts from the shipping carton, be careful not to damage the components. Important: thoroughly check each box for parts that may be located in areas that can be overlooked.

MI-750 Installation and Service Manual (1003136 Rev A) Page 24

Section 3

Single Ceiling Mount Installation (cont’d)

Figure 9: Double Ceiling Mount Installation

 Insert the ceiling rod up through the ceiling casting until the hole in the ceiling rod becomes visible (see Figure 9: “Double Ceiling Mount Installation”). Insert the dowel pin into the hole in the ceiling rod and lower the rod in the ceiling casting, making sure the pin is seated securely in the indentation on the ceiling casting. Securely tighten the two set-screws located in the casting.

Failure to install the dowel pin can cause the arm/head assembly to fall from the ceiling causing serious injury and/or property damage.

MI-750 Installation and Service Manual (1003136 Rev A) Page 25

Section 3

Double Ceiling Mount Installation (cont’d)

 The ceiling casting must be properly grounded during installation. This can be done by mounting the casting to a suitable material that will function as a ground conductor, or a wire lead that must be attached to the ground screw on the casting then routed to proper ground.

Note: Ceiling rod must be plumb. Adjust the nuts/washers in accordingly.

Note: The ceiling casting itself must be electrically grounded to maintain proper grounding reliability.

If ceiling rod is not plumb, unwanted arm drifting may occur.

 Feed the wire harness through the ceiling rod and route to a junction box. Leave sufficient wire to extend slightly beyond the bottom of the ceiling rod. Important: To achieve

proper grounding reliability, the green ground wire from the wire harness MUST be properly fastened to the grounding screw located on the ceiling casting. Make all electrical connections in compliance with all applicable electrical codes.

Failure to comply with local electrical codes can cause a shock hazard.

WARNING: Secure the source of power using the Means of Isolation, before proceeding with electrical work.

 Slide the casting cover up the rod over the casting. Similarly, slide the collar to the casting cover, then tighten the set screw to hold the cover in place.

Do not install the extension arm with the light head attached. Installation with the light head attached can cause damage to the light. Refer to the procedure for installing the head to the arm after the bracket and arm are installed

 Bring one section of the double ceiling mount casting to the ceiling rod and fasten it to the rod with the 2” long bolts (P/N 0001311) and cap nuts (P/N 0001312). Tighten the cap nuts just enough to hold the section in place; they do not need to be fully tightened at this time. Bring an arm assembly to the double mount section and fasten the bushing on the extension arm into the extension arm mounting cavity with two 3/8” long mounting screws

(P/N 0001291). Install the other arm assembly to the section with two more 3/8” long screws. Route the wire and Molex connector from the ceiling rod into the wire channel as shown in Figure 10: “Wire Harness Positioning” and Figure 11: “Wire Channel”. Plug

the connector at the end of each Arm assembly to the mating Moles connector from the wire harness. Place each set of wires within the channel area.

MI-750 Installation and Service Manual (1003136 Rev A) Page 26

Section 3

Double Ceiling Mount Installation (cont’d)

Figure 10: Wire Harness Positioning

Figure 11: Wire Channel

 Position the two extension arms as shown in Figure 9: Double Ceiling Mount Installation” to allow for easier mounting of the remaining double mount section. Firmly press the fastened double ceiling mount section against the ceiling rod. While firmly holding the double ceiling mount section with arms attached in place, remove only the cap nuts from the 2” long bolts. It is strongly recommended that a second person hold the arms in place while the nuts are removed. Place the remaining double ceiling mount section onto the bolts. Be sure the wires are enclosed within the wire channel to avoid pinching them.

Replace the cap nuts and fully tighten. Install the remaining four 3/8” long mounting screws and then fully tighten all eight screws.

Failure to tighten the cap nuts for the 2” long bolts can cause the arm/head assembly to fall causing serious injury and/or property damage.

MI-750 Installation and Service Manual (1003136 Rev A) Page 27

Section 3

Double Ceiling Mount Installation (cont’d)

Failure to install or tighten all eight of the 3/8” mounting screws can cause the arm/head assembly to fall causing serious injury and/or property damage.

 Check to make sure that the wire connections are seated far enough into the double casting enclosure. Fasten the cover plates to the casting enclosure using the four 3/8” long pan-head screws (P/N 0001319) and plastic grommets (P/N 0001306) supplied in the hardware kit.

 See the instructions for “Section 3: Light Head to Arm Installation” on page 37 to

complete the assembly.

MI-750 Installation and Service Manual (1003136 Rev A) Page 28

Wall Mounted Light Installation

Section 3

Figure 12: Wall Mount Dimensions

GENERAL INFORMATION

The shipping cartons contain a light head assembly, arm assembly, a wall bracket assembly with a hospital grade plug attached, hardware kit, and an Installation and Service Manual.

(Mounting hardware for attaching the wall bracket to the wall is not supplied).

Prior to installation ensure that all components shown on Figure 13: “Wall Mount

Installation” are present.

When removing parts from the shipping carton, be careful not to damage the components. Important: thoroughly check each box for parts that may be located in areas that can be overlooked.

WARNING: To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth.

MI-750 Installation and Service Manual (1003136 Rev A) Page 29

Wall Mounted Light Installation (cont’d)

Section 3

WARNING: Secure the source of power using the Means of Isolation, the power cord, before proceeding with electrical work.

Figure 13: Wall Mount Installation

Figure 14: Wall Bracket Mounting Diagram

MI-750 Installation and Service Manual (1003136 Rev A) Page 30

Wall Mounted Light Installation (cont’d)

Section 3

Improper fastening of the wall bracket can cause serious injury and/or property damage. Make certain the installation is capable of supporting a load of at least 65 pounds and an off center moment of 195 ft-lbs.

The supply circuit line must be as follows:

100-240VAC, 50/60 Hz, single phase, three wire, capable of supplying 65 Watts @ 1.5A.

WARNING: Secure the source of power using the Means of Isolation, the power cord, before proceeding with electrical work.

The equipment is not deemed compatible with any sort of electrical dimming device. Use line voltage only. To maintain proper grounding reliability, the ground wire connections with the wall bracket must be kept properly fastened at all times. The power cord is permanently attached and is not intended to be replaceable.

Do not install the extension arm with the light head attached. Installation with the light head attached can cause damage to the light. Refer to the procedure for installing the head to the arm after the bracket and arm are installed.

Recommended Wall Structure Construction Details

The illustrations below are suggested mounting schemes per 2001 California Building Code –

Section 1632A: Anchorage and Seismic. For any other mounting scheme, please consult a structural engineer and/or professional contractor for the best solution for your situation. Installation and repair of this equipment should be performed by qualified persons only. Medical Illumination does not warranty any damage occurring as a result of improper installation.

Wall mounting with load anchor or stud anchor Wall mounting with Backing Plate (not provided)

Concrete Wall (3000 PSI min., 6” thick min.):

 Use two 3/8” Dia HILTI KB3 Expansion

Anchors or approved equivalent bolts (min. embed. 3”). See Figure 15: “Stud Anchor

Installation”.

Also see HILTI technical document ESR-2302 for additional concrete anchoring information.

Lattice stone, cinder block, etc. (min. 5/8” wall board thickness):

 Use two 3/8” Dia A307 Bolts with nuts/washers to Backing Plate

MI-750 Installation and Service Manual (1003136 Rev A) Page 31

Wall Mounted Light Installation (cont’d)

Section 3

 Drill hole according to the diameter of the stud anchor

 Thoroughly clean hole – removing all debris

 Insert anchor into hole (min. embed 3”)

 Torque the nut down against the washer to 20 ft-lb (see HILTI tech. doc.

ESR-2302 Table 1 for reference)

Figure 15: Stud Anchor Installation

 Remove the four screws holding the cord retainer plate in place. Carefully pull the plate away from the wall bracket to expose the Molex connector. DO NOT use excessive

force during this procedure or in any way loosen or disconnect the green wires within the bracket.

Failure to comply with local electrical codes can cause a shock hazard.

WARNING: Secure the source of power using the means of isolation, the power cord, before proceeding with electrical work.

 Bring the extension arm over the wall bracket and feed the Molex connector from the transformer housing into the bracket while inserting the bushing into the bracket. Secure the arm subassembly to the bracket with the flat head cap screw provided.

 Bring the cord retainer plate up to the bottom of the wall bracket and plug the Molex connector at the end of the power cord to the connector from the arm subassembly. Route the connectors and wire to the rear of the bracket (not shown). Refasten the retainer plate to the bracket with the four screws provided.

 See the instructions for “Section 3: Light Head to Arm Installation” on page 37 to

complete the assembly.

MI-750 Installation and Service Manual (1003136 Rev A) Page 32

Floor Stand Light Installation

Section 3

Figure 16: Floor Mount Dimensions

GENERAL INFORMATION

The shipping carton contains a light head assembly, arm assembly, floor base casting, two leg assemblies, a hardware kit, and an Installation and Service Manual. A second carton contains the upright pole assembly. Prior to installation insure that all components are present.

When removing parts from the shipping carton, be careful not to damage the components. Important: thoroughly check each box for parts that may be located in areas that can be overlooked.

WARNING: To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth. The power cord is permanently attached and is not intended to be replaceable.

MI-750 Installation and Service Manual (1003136 Rev A) Page 33

Floor Stand Light Installation (cont’d)

Section 3

WARNING: Secure the source of power using the Means of Isolation, the power cord, before proceeding with electrical work.

Figure 17: Floor Stand Base Assembly

 Bolt the assembled legs onto the cast iron base using four 1/4-20 x 1 3/4” bolts and 1/4 lock washers. Position the floor base casting so that the set-screw in the base is at the rear of the assembly. See Figure 17: “Floor Stand Base Assembly”.

 Insert the upright pole fully into the floor base casting so that the anti-rotation slot sits over the pin in the casting. Rotate the pole until the slots slides fully over the pin, and lower the pole completely into the base. The power cord (not shown) should be in line with the setscrew. Once the pole is properly positioned, securely tighten the set-screw using the 1/8 hex key provided.

MI-750 Installation and Service Manual (1003136 Rev A) Page 34

Floor Stand Light Installation (cont’d)

Section 3

Do not install the extension arm with the light head attached. Installation with the light head attached can cause damage to the light. Refer to the procedure for installing the head to the arm after the floor mount is assembled and the arm is in place.

Figure 18: Arm Assembly to Upright Pole Installation

While securely supporting the arm assembly above the upright pole, firmly grasp the cable at the end of the pole. DO NOT allow the cable to slip into the pole as it may be difficult to retrieve it. Connect the Molex connector from the PSU housing to the connector at the end of the pole. Remove the cable tie holding the harness in place and discard it. Carefully lower the extension arm and insert the PSU housing into the tube while pushing the wiring into the pole.

Secure the housing to the pole utilizing the two 1/4-20 x 3/8” bolts supplied. Tighten the bolts securely.

MI-750 Installation and Service Manual (1003136 Rev A) Page 35

Section 3

Floor Stand Light Installation (cont’d)

The light arm should be between the extended legs as shown in Figure 18: “Arm Assembly

to Upright Pole Installation”.

To level the base, remove the end caps at the end of each leg and adjust the caster bolts and nuts. Once the base has been leveled, tighten the nuts securely and insert the leg cap in the end of each leg. See Figure 17: “Floor Stand Base Installation”.

See the instructions for Section 3: “Light Head to Arm Installation” on page 37 to

complete the assembly.

MI-750 Installation and Service Manual (1003136 Rev A) Page 36

Light Head to Arm Installation

Section 3

Figure 19: Light Head-Arm Installation

Installation and repair of this equipment should be performed by qualified persons only. Medical Illumination does not warranty any damage occurring as a result of improper installation.

Insert the wire set from the yoke / arm connector through the hole in the arm/wire junction.

MI-750 Installation and Service Manual (1003136 Rev A) Page 37

Section 3

Light Head to Arm Installation (cont’d)

 Insert the yoke/arm connector into the arm until the bolt-holes are aligned. Insert the

1/4-28 x 1 9/16” bolt as shown being careful not to damage the wires, and secure with the

1/4-28 hex nut. Tighten the nut securely, but do not over tighten.

 Tighten the anti-rotation screw into the yoke/arm connector using the hex key provided.

 One set of wires exits from the light head assembly, and the other set exits from the arm/wire junction assembly. Connect the wires with the spade connectors: black wire to black wire, red wire to red wire, and green wire to green wire. Replace the Connection

Cover and securely tighten the two screws. See Figure 20: “Light Head to Arm

Electrical Connections”.

 Press the snap cover onto the top of the hex bolt as shown on Figure 19: “Light Head-

Arm Installation”.

Figure 20: Light Head to Arm Electrical Connections

MI-750 Installation and Service Manual (1003136 Rev A) Page 38

Section 4

Section 4: Operating Instructions Light Head

Power / Dimmer Button: Press button to energize or hold button for 5 seconds to de-energize lamp. While energized, press and release button to cycle thru the 3 level of dimming (100% -

75% - 50% - 100% - 75% - …).

To position the light head and arm over the work area, firmly grasp the sterilizable handle and bring the light head or arm to the desired location. Adjust the tilt of the light head by holding the sterilizable handle and rotating forward or backward. The light head can revolve around the

Yoke/Arm connector by tilting the light head and rotating the head using the sterilizable handle.

Figure 21: Light Head Components

MI-750 Installation and Service Manual (1003136 Rev A) Page 39

Section 4

Light Head Intensity Indicators,

Control and Indicator Display

The MI-750 has the dimmer button at the end of the sterile handle. Pushing this button cycles thru the dimming levels (3-2-1-3-2-1, etc…). This continues until the button is held for 5 seconds, this turns off the light. The indicator display on the light head is just above the handle.

The rightmost LED indicates that the light is receiving power. In the Ceiling Mounted units this indicates that the AC to the light is active. In the Wall or Floor Light it indicates that the Cord is plugged in and AC is active. In the Battery Backup unit the LED indicates that the AC is on and that the Battery is good.

Off Power Ready, No light

Level 3 Full Power, 3 LED’s Illuminated

Level 2 Reduced Power, 2 LED’s Illuminated

Level 1 Low Power, 1 LED Illuminated

The right most LED indicates that the light is receiving power. This LED should most always be on.

Fail No Power to the light.

MI-750 Installation and Service Manual (1003136 Rev A) Page 40

Section 4

Floor Model Positioning and Detachable Parts

Floor Model, Transportation Position

When transporting the Floor Model Lighting System from location to location, the base should be resting on all four wheels with the wheel locks disabled. When positioned the wheel locks may be enabled as required to prevent unwanted motion.

Positioning

When placing the light consider not only the procedure and patient but the need to have access to the cord in case of emergency. Also ensure that the cord and base are not a tripping hazard.

Floor Model, Rough Handling & moving over a threshold

When moving the light at least one person shall keep both hands on the upright pole to stabilize and control the light. Care must be taken to protect the light as well as obstacles along the path of travel. Extra care shall be taken when transitions are encountered. The light may be tipped to move one set of wheels over a threshold before the other set of wheels

Warning: Caution when moving the light that the light does not swing and strike people or obstructions!

WARNING: Secure the source of power using the Means of Isolation, the power cord or circuit breaker, before proceeding with electrical work.

Detachable Parts:

This system has only one detachable part, the Sterilizable Handle.

MI-750 Installation and Service Manual (1003136 Rev A) Page 41

Section 5

Section 5: Warnings & Safety Instructions

Safety Tips

Only facility authorized maintenance personnel should troubleshoot the unit.

Troubleshooting by unauthorized personnel could result in personal injury and/or property damage.

Only facility authorized personnel should repair the unit. Repair by unauthorized personnel could result in personal injury and/or property damage and could void warranty.

WARNING: No modification of this equipment is allowed.

After completing a repair of the unit, ensure that the unit is in proper working order. Failure to do so could result in personal injury and/or property damage.

Do not touch the LED’s or lenses directly. Body oils may significantly lower the optical performance of these parts and may cause equipment damage.

Follow the product manufacturer’s instructions. Failure to do so could result in personal injury and/or property damage.

If the unit fails any part of the preventive maintenance functional checks, repair the unit before use on any patient. Failure to do so could result in personal injury and/or property damage.

Do not use harsh cleaners, solvents, or detergents. Failure to do so could result in equipment damage.

Do not use silicone-based lubricants. Equipment damage could occur.

Turn the power off or unplug the power cord before any repairs are started. Failure to do so could result in personal injury and/or property damage.

WARNING: Secure the source of power using the Means of Isolation, the power cord or circuit breaker, before proceeding with electrical work.

Do not pinch any wires during installation or during any repair. Pinched wires can cause an electrical shock hazard, resulting in personal injury and/or property damage.

MI-750 Installation and Service Manual (1003136 Rev A) Page 42

Section 5

Safety Tips (cont’d)

Do not expose the unit to excessive moisture. Failure to do so could result in personal injury and/or property damage.

WARNING: To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth.

WARNING: The MI-750 is designed not to cause interference with other devices. If interference is suspected, increase the distance between the 750 light and the suspected recipient of interference.

MI-750 Installation and Service Manual (1003136 Rev A) Page 43

Section 6

Section 6: Maintenance

WARNING: Secure the source of power using the Means of Isolation, the power cord or circuit breaker, before proceeding with electrical work.

Before any Maintenance Lock-out/Tag-out the power supply or unplug the power cord before anything is started. Failure to do so could result in personal injury and/or property damage.

Fuse Replacement

The Single and Dual Ceiling mounted units require 1 fuse for each arm and the Floor and Wall mounted units require 2 fuses. The fuse holders are located in top cover of the cylindrical enclosure at the end of the arm. For Ceiling mounted lights remove the plastic fuse holder with a screw driver and remove the blown fuse. When replacing the new fuse, insert the new fuse into the plastic fuse holder cap and then insert the fuse into its receptacle without touching the fuse. Rotate the fuse holder cap with a screw driver.

If the fuse(s) need to be replaced, use only 250V/ 1.5A Slow-blow. Break capacity, 150% of

1.5A @ more than 60 min and 100% of 1.5A @ 20-300ms

IMPORTANT: Make sure that the power supply cord is disconnected from the supply outlet before replacing the fuse(s).

LED Maintenance

The LED life (L70) is rated such that the LED will deliver, on average, 70% lumen maintenance at 50,000 hours of operation.

If one or more LED(s) are off and/or generate noticeably dim light output, contact our customer service department at (818) 838-3025. The LED’s are not readily replaceable by end users and should only be serviced by Medical Illumination.

MI-750 Installation and Service Manual (1003136 Rev A) Page 44

Section 6

Arm Adjustment

The arm has been pre-adjusted by the factory. Should further adjustment be necessary, utilize the following procedure.

 Release the arm friction brake by loosening the set screw inside the Friction Adjustment

Hole with an Allen wrench (3 or 4 turns). This allows the spring to move freely for a more accurate adjustment. The arm is expected to droop as the brakes are loosened.

 Remove the white rubber Cover from the spring tension adjustment slot as shown.

 Move the arm vertically until the adjustment nut is visible down inside the Slot.

 Engage the tabs of the Adjustment Wrench into the slots in the Adjustment Nut. Turn the nut in the direction of the arrow as shown in the figure to increase the arm tension. Check the progress as you go until the required adjustment is reached.

Do not over tighten the nut. Doing so could result in equipment damage.

 When proper tension is achieved, the arm should travel up or down an equal distance when the same amount of force is applied from above or below. When moved to the maximum Up position, it should travel back down about 1-1.5 inches (until brake is set).

From maximum Down position, it should travel back up about the same.

 After the spring tension is adjusted, re-tighten the brake until the desired resistance is obtained.

 Replace the spring tension adjustment slot cover.

Figure 22: Arm Adjustment

MI-750 Installation and Service Manual (1003136 Rev A) Page 45

Section 6

Head/Yoke Adjustment

The head/yoke has been pre-adjusted by the factory. Should further adjustment be necessary, utilize the following procedure.

 Locate the Friction Break Adjustment access as shown in Figure 35: “Head/Yoke

Adjustment”.

 Use the supplied Allen Wrench to increase the amount of friction applied to the rotation of the light head.

Do not over tighten the Friction Break. Doing so could result in equipment damage.

Figure 23: Head / Yoke Adjustment

MI-750 Installation and Service Manual (1003136 Rev A) Page 46

Section 6

Handle Sterilization

 Remove sterilizable handle by pressing the button near the base of the handle and pulling the handle off the handle post.

 Sterilize the handle utilizing steam sterilization of minimum 250 Fahrenheit for a minimum of 30 minutes in compliance with AAMI-SSSA-1988: Good Hospital Practices, Steam

Sterilization and Sterility Assurance, or an approved equivalent method. The responsible organization (operator/owner of the equipment) shall follow the national requirements

(standards and directives) for hygiene and disinfection.

Figure 24: Handle Sterilization

MI-750 Installation and Service Manual (1003136 Rev A) Page 47

Section 6

Cleaning Instructions

 Clean the lens using glass/plastic cleaner or mild soap and water mix. It is very important to use a clean, soft cloth to avoid any scratching of the front lens. Never spray the cleaning fluid directly onto the lens surface, but instead spray into clean cloth and then wipe the lens.

 Clean the light housing and arm(s) using mild soap and water mixture. Apply this mixture to a clean cloth and wipe down the light-head and arm. Never spray the cleaning fluid directly onto the light head or arm.

Do not use harsh cleaners, solvents, or detergents. Failure to do so could result in equipment damage.

The MI-750 front lens is supplied with a protective hard coat to resist scratching.

Never use abrasive cleaners on front lens. Failure to do so could result in equipment damage.

Do not expose the unit to excessive moisture. Failure to do so could result in

personal injury and/or property damage.

Long term exposure to cleanings and disinfectants will reduce the life expectancy of some of the light and arm components.

The responsible organization (operator/owner of the equipment) shall follow the

national requirements (standards and directives) for hygiene and disinfection.

MI-750 Installation and Service Manual (1003136 Rev A) Page 48

Section 6

Maintenance Schedule

Function Procedure

Front lens/Optics

Check front lens and optics to assure there are no chips, cracks, or other damage. Do not use equipment if parts are damaged. Replace damaged parts immediately.

Bolts and nuts

Check to see that all mounting and attachment screws, washers, etc. are in place and securely tightened. Replace any missing screws and retighten as required.

Casters

Moving joints/Adjustments

Overall appearance

Ensure that casters are seated properly on the base assembly. Examine the base for any damage.

Check to make sure all moving joints function properly along the mounting system and head and arm system. If the articulating arm does not position properly (drifts from original position) refer to Figure 22:

“Arm Adjustment” on page 45.

If this does not solve the problem contact customer service as the unit may require factory repair.

Check the general aesthetics. Units should be kept clean and dust free.

Clean and dust as necessary.

Note: Maintenance schedules vary for each light depending on usage and operating instructions. An annual inspection of the equipment is recommended at a minimum.

Note: Medical Illumination International Inc. recommends that the maintenance records for this equipment be kept on file at the health care facility.

MI-750 Installation and Service Manual (1003136 Rev A) Page 49

Section 7

Section 7: Troubleshooting

General Troubleshooting

Warning: Disconnect the light from the power supply before attempting any of the electrical checks mentioned below.

Problem Cause

Light will not turn on or stay on

Light does not maintain its position vertically

Light does not maintain its position horizontally

Power to unit is off (not plugged in)

Blown (open) fuse

Exposed wires are cut or damage

Wire not connected correctly during installation

Wire connections made during installation have disconnected

No input power to light unit

Disconnected wires at switch or

PSU

No power output from PSU when input power to PSU is measured

Wire connections made during installation have been reversed

Sterilizable handle button not engaging power push button switch properly

Spring tension is incorrect.

Additional equipment was added to unit.

Mounting bracket is not level.

Hex nuts on pivot bolts are loose.

Remedy

Turn on power (plug in unit)

Replace fuse

Replace wire assembly

Check all wiring connections

Reconnect wires per the instructions

Check power input connections and circuit breakers

Reconnect wires

Replace PSU

Connect proper wires together: black-black and red-red

Take off sterilizable handle and inspect handle post assembly with switch. Make sure handle post is tightly assembled. Make sure the button in the sterilizable handle can move freely

Adjust spring.

Remove any additional equipment from arm.

Adjust or shim as necessary.

Remove plastic caps and adjust as necessary.

MI-750 Installation and Service Manual (1003136 Rev A) Page 50

Section 7

General Troubleshooting (cont’d)

Light head will not rotate at yoke interface

Arm cannot be moved any lower

Arm cannot be raised any higher

Arm stopped moving horizontally

Caster/casters cannot be reinstalled

Light head is against internal stop. Rotate head in opposite direction.

Arm is against internal stop.

Arm is against internal stop.

Arm is against internal stop.

Floor base has been damaged.

Rotate arm in opposite direction.

Rotate arm in opposite direction.

Rotate arm in opposite direction.

Contact Customer Service or Field

Representative.

Upright rotates in floor stand

Upright set screw is loose.

Upright is not fully seated on antirotation pin.

Securely tighten set screw.

Reinstall upright. Ensure the upright fully seats on the anti-rotation pin seats.

Light output is irregular in shape or intensity

Input voltage does not correspond to rating label.

The light head is not mounted on the proper arm (Combo models).

Check circuit to which light was installed.

Remount light head on the correct arm.

MI-750 Installation and Service Manual (1003136 Rev A) Page 51

advertisement

Key Features

  • 3 Levels of dimming
  • Adjustable light head
  • Ceiling, wall and floor mounts
  • Central room mounting
  • Hospital grade plug
  • 190° lamp head rotation

Frequently Answers and Questions

What is the central illuminance of the MI-750?
The central illuminance of the MI-750 is 50,000 Lux.
What are the different mounting options for the MI-750?
The MI-750 can be mounted to the ceiling, wall or floor.
What is the color temperature of the MI-750?
The color temperature of the MI-750 is 4,300° Kelvin ±300°K.