Agilent 16048G, 16048H test leads Operation and Service Manual

Agilent 16048G, 16048H test leads Operation and Service Manual

Below you will find brief information for test leads 16048G, test leads 16048H. These test leads are used to attach user-fabricated test fixtures. The 16048G is 1 meter long, the 16048H is 2 meter long. They can be connected to the 4294A LCR meter and Impedance Analyzer. For accurate readings, fixture compensation is required. To compensate stray capacitance and residual impedance, fixture compensation should be performed, which is explained in the 4294A Operation Manual.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Agilent 16048G, 16048H Test Leads Operation and Service Manual | Manualzz
Agilent 16048G/H Test Leads
Operation and Service Manual
Third Edition
Agilent Part No. 16048-90050
January 2001
Printed in: Japan
Notices
The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright.All
rights are reserved. No part of this document may be photocopied, reproduced, or
translated to another language without the prior written consent of the Agilent
Technologies.
Agilent Technologies Japan, Ltd.
Component Test PGU-Kobe
1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan
© Copyright Agilent Technologies Japan, Ltd. 1999, 2001
Manual Printing History
The manual’s printing date and part number indicate its current edition. The
printing date changes when a new edition is printed. (Minor corrections and
updates that are incorporated at reprint do not cause the date to change.) The
manual part number changes when extensive technical changes are incorporated.
April 1999
First Edition (part number: 16048-90050)
December 1999
Second Edition (part number: 16048-90050)
January 2001
Third Edition (part number: 16048-90050)
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific WARNINGS elsewhere in this manual may impair the
protection provided by the equipment. In addition it violates safety standards of
design, manufacture, and intended use of the instrument.
The Agilent Technologies assumes no liability for the customer’s failure to comply
with these requirements.
NOTE
16048G/H comply with INSTALLATION CATEGORY I and POLLUTION
DEGREE 2 in IEC61010-1. 16048G/H are INDOOR USE product.
2
•
DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of flammable gasses or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
•
Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with the power cable connected. Under
certain conditions, dangerous voltages may exist even with the power cable
removed. To avoid injuries, always disconnect power and discharge circuits
before touching them.
•
DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present.
•
DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not install
substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Agilent Technologies Sales and Service Office for
service and repair to ensure that safety features are maintained.
•
Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the warnings must
be followed.
WARNING
Dangerous voltages, capable of causing death, are presenting this instrument.
Use extreme caution when handling, testing, and adjusting this instrument.
Certification
Agilent Technologies certifies that this product met its published specifications at
the time of shipment from the factory. Agilent Technologies further certifies that
its calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institution’s calibration
facility, or to the calibration facilities of other International Standards Organization
members.
3
Warranty
This Agilent Technologies instrument product is warranted against defects in
material and workmanship for a period corresponding to the individual warranty
periods of its component products. Instruments are warranted for a period of one
year. Fixtures and adapters are warranted for a period of 90 days. During the
warranty period, Agilent Technologies will, at its option, either repair or replace
products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to
Agilent Technologies and Agilent Technologies shall pay shipping charges to
return the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by
Agilent Technologies for use with an instrument will execute its programming
instruction when property installed on that instrument. Agilent Technologies does
not warrant that the operation of the instrument, or software, or firmware will be
uninterrupted or error free.
Limitation Of Warranty
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modification or misuse, operation outside the environmental
specifications for the product, or improper site preparation or maintenance.
IMPORTANT
No other warranty is expressed or implied. Agilent Technologies specifically
disclaims the implied warranties of merchantability and fitness for a particular
purpose.
Exclusive Remedies
The remedies provided herein are buyer’s sole and exclusive remedies. Agilent
Technologies shall not be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other legal theory.
4
Assistance
Product maintenance agreements and other customer assistance agreements are
available for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service
Office. Addresses are provided at the back of this manual.
Safety Symbol
General definitions of safety symbols used on the instrument or in manuals are
listed below.
Instruction Manual symbol: the product is marked with this symbol when it is
necessary for the user to refer to the instrument manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
In position of push-button switch.
Out position of push-button switch.
Frame (or chassis) terminal. A connection to the frame (chassis) of the equipment
which normally include all exposed metal structure.
WARNING
This warning sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could
result in injury or death to personnel.
CAUTION
This Caution sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result
in damage to or destruction of part or all of the product.
NOTE
Note denotes important information. It calls attention to a procedure, practice,
condition or the like, which is essential to highlight.
5
6
Contents
1. Installation Guide
Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Connecting the 16048G/H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2. Overview
Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3. Operation
4294A Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Adapter Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Performing Fixture Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.
Specifications
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5. Service
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7
Contents
8
1
Installation Guide
9
Installation Guide
Incoming Inspection
Incoming Inspection
Inspect the shipping container for damage. If the shipping container or cushioning
material is damaged, it should be kept until the contents of the shipment have been
checked for completeness and the 16048G/H has been checked mechanically and
electrically. The contents of the shipment should be as listed in Table 1-1. If the
contents are incomplete, if there is mechanical damage or defect, notify the nearest
Agilent Technologies office. If the shipping container is damaged, or the
cushioning material shows signs of unusual stress, notify the carrier as well as the
Agilent Technologies office. Keep the shipping materials for the carrier’s
inspection.
Table 1-1
Contents
Description
Part Number
Qty.
Test Lead (16048G or 16048H)
-
1
Mounting Plate
-
1
BNC Bracketa
16048-60003
1
Operation and Service Manual
16048-90050
1
a. Only furnished with option 001.
10
Chapter 1
Installation Guide
Connecting the 16048G/H
Connecting the 16048G/H
Follow these steps below to connect the 16048G or 16048H to the 4294A.
Step 1. Set the 16048G/H to the test connectors on the front panel of the HP 4294A by
gradually coupling the four BNC connectors and fastening screws of the fixture
with the test connectors and accessory mounting holes of the instrument until they
come to complete contact.
Step 2. Fasten two of the four BNC connectors to the mating test connectors by gradually
turning the BNC connectors' rotation levers until each pair of connectors are
securely connected. Be sure to align the grooves on both sides.
Step 3. Turn clockwise the fixture's two fastening screws together, so that the fixture is
secured to the instrument.
Step 4. Finally, secure the remaining two BNC connectors of the fixture by turning
clockwise their rotation levers.
Figure 1-1
Connecting the fixture to the instrument
NOTE
Using the connection block securing screws enables the 16048G/H to connect the
GND of the 4294A.
Chapter 1
11
Installation Guide
Connecting the 16048G/H
12
Chapter 1
2
Overview
13
Overview
Product Overview
Product Overview
The 16048G/H consists of a direct attachment, 4-terminal pair interface which is equipped
with four BNC (f) connectors. These test leads are used to attach user-fabricated test
fixtures. Cable length of the 16048G is 1 meter, the 16048H is 2 meter.
Figure 2-1
Product Overview
14
Chapter 2
Overview
Functions
Functions
Figure 2-2 shows name of each part of the 16048G/H and Table 2-2 shows their function.
Figure 2-2
16048G/H Parts
Table 2-1
16048G/H Function
No.
NAME
FUNCTION
1
Instrument Side
Connection Block
Connect to the instrument’s UNKNOWN terminal.
2
Connection Block
Securing Screw
Secures the connection block to the 4294A.
3
Ground Lead
Connect to the chassis or grounded part of the
measurement target.
4
DUT Side Connection
Block
Used to attach BNC Bracket or user-fabricated test
fixtures.
5
Mounting Plate
Mounts on the measurement target to stabilize the
cable.
Chapter 2
15
Overview
Functions
Figure 2-3
Cable Assignment
16
Chapter 2
3
Operation
This chapter describes the proper methods for setting the 4294A, fixture
compensation with the 16048G/H.
17
Operation
4294A Setting
4294A Setting
Before you begin your measurement, you should perform the adapter setup in the
4294A. Also refer to the operation manual of the 4294A about the adapter setup.
Adapter Setup
Connect the 16048G/H to the 4294A and perform the adapter setup described
below.
NOTE
For adapter setup, use the 100 Ω Resistor furnished with the 4294A (Agilent P/N
04294-61001).
1. More than 30 minutes warm-up time is required after turning on the 4294A.
2. Press [Cal] key to bring up the Calibration Menu.
3. Press ADAPTER [ ] key to bring up the adapter setup Menu. [ ] shows current
settings.
4. Select 4TP 1M in the case of 16048G, 4TP 2M in the 16048H. When the
selection is completed, the softkey label will be underlined.
5. Press SETUP key to bring up the Adapter Setup Menu.
6. Connect the Lcur terminal and Lpot terminal on the 16048G or 16048H to the
Hpot terminal and Hcur terminal of the 100 Ω Resistor furnished with the
4294A. The Hcur and Hpot terminal on the 16048G/H should remain open
(nothing should be connected).
Figure 3-1
Connecting 100 Ω Resister (Phase Compensation Data Measurement)
18
Chapter 3
Operation
4294A Setting
7. Press PHASE COMP [-] key to start the phase compensation data measurement.
About 1 minutes later, phase compensation data measurement is completed and
the softkey label changes to PHASE COMP [DONE].
8. Connect the Lcur, Lpot, Hcur, Hpot terminal on the 16048G or 16048H to the
Lcur, Lpot, Hcur, Hpot terminal of the 100 Ω Resistor respectively.
Figure 3-2
Connecting 100 Ω Resister (Load Data Measurement)
9. Press LOAD [-] key to start the load data measurement. When the load data
measurement is completed, the softkey label changes to LOAD [DONE].
10. Press done key.
Chapter 3
19
Operation
Performing Fixture Compensation
Performing Fixture Compensation
To compensate stray capacitance and residual impedance, fixture compensation
should be performed. When you use the optional BNC Bracket (Agilent P/N
16048-60003) or an user fabricated test fixture, fixture compensation is required.
Refer to the 4294A Operation Manual for the fixture compensation procedure.
20
Chapter 3
4
Specifications
This chapter provides specifications of the 16048G/H Test Leads.
21
Specifications
Specifications
Specifications
Applicable Instruments
LCR meters and Impedance Analyzers with
four-terminals
Maximum Voltage
± 42V peak max. (AC+DC)
Operating
Environment
temp.
-20°C to +150°C
(Exclude up to 20 cm from four terminal pair
connection block)
humidity
15% to 95%RH( @ wet bulb temp. < 40°C)
Non Operating
temp.
-40°C to +70°C
Environment.
humidity
≤ 90 % RH ( @ wet bulb temp. <65°C)
Cable length (Nominal)
1 m (16048G) / 2 m (16048H)
Weight
Approximately 460 g (16048G)
Approximately 690 g (16048H)
Safety Standards
EN61010-1:1993 +A2:1995
IEC61010-1:1990 +A1:1992 +A2:1995
CSA C22.2 No.1010.1:1992
INSTALLATION CATEGORY I
POLLUTION DEGREE 2
INDOOR USE
22
Chapter 4
5
Service
This chapter provides information on servicing and proper maintenance.
23
Service
Maintenance
Maintenance
An exploded view of the 16048G/H for parts identification is shown in Figure 5-1
to Figure 5-4. Do not disassemble any further than shown. Maintenance consists
principally of cleaning contacts and replacing worn or damaged parts. Take special
care when cleaning contacts.
To order parts, use the Agilent part numbers listed in Table 5-1 to Table 5-4. If a
faulty part is located in an assembly that cannot be disassembled, order the next
higher assembly or return the fixture to the nearest Agilent Technologies
Sales/Service Office for repair or replacement.
Figure 5-1
Replacable Parts (part 1 of 4)
24
Chapter 5
Service
Maintenance
Table 5-1
Replacable Parts (part 1 of 4)
Reference
Designator
Agilent Part No.
Qty.
Description
1
16048-61620
1
CBL ASSY 16048G
(Excludes 2,3,4,5,6)
16048-61623
1
CBL ASSY 16048H
(Excludes 2,3,4,5,6)
2
1400-0493
4
CABLE TIE
3
9170-1465
1
CORE
4
-
1
CBL CLANP
5
0515-1550
1
SCR M3-L 8 P-H
6
3050-0891
1
WASHER M3
7
2950-0035
4
NUT-HEX-DBL-CHAM
8
2190-0102
4
WASHER
9
16048-00623
1
GUIDE
10
1250-0252
4
CONN-RF BNC
11
16048-00626
1
PLATE
12
16048-00627
1
ANGLE
13
0515-1012
2
SCR FL M4L8
Chapter 5
25
Service
Maintenance
Figure 5-2
Replacable Parts (part 2 of 4)
Table 5-2
Replacable Parts (part 2 of 4)
Reference
Designator
Agilent Part No.
Qty.
Description
1
16044-60001
2
CONN BNC-SMB
2
3050-0067
4
WSHR-FL MTLC
3
3050-0789
4
WSHR-FL NM
4
16047-40002
4
INSULATOR
5
1253-0476
2
ADPT BNC-SMB
26
Chapter 5
Service
Maintenance
Figure 5-3
Replacable Parts (part 3 of 4)
Table 5-3
Replacable Parts (part 3 of 4)
Reference
Designator
Agilent Part No.
Qty.
Description
1
16047-24021
2
KNOB
2
16048-04021
1
COVER TOP (16048G)
16048-04025
1
COVER TOP (16048H)
3
16047-24026
2
FLANGE
4
0515-0914
4
SCR-MACH M3X0.5
5
0515-0952
4
SCR-MACH M2X0.4
6
16048-00624
1
ANGLE
7
0515-0914
3
SCR-MACH M3X0.5
Chapter 5
27
Service
Maintenance
Table 5-3
Replacable Parts (part 3 of 4)
Reference
Designator
Agilent Part No.
Qty.
Description
8
16048-04022
1
COVER BOTTOM
9
16048-00625
1
ANGLE
Figure 5-4
Replacable Parts (part 4 of 4)
Table 5-4
Replacable Parts (part 4 of 4)
Reference
Designator
Agilent Part No.
Qty.
Description
1
1250-1985
4
CONNECTOR-BNC
2
16048-00621
1
PLATE
3
2190-0016
4
WASHER
4
2950-0043
4
NUT-HEX-DBL-CHAM
5
16380-24001
4
TERMINAL
6
42036-00601
2
PLATE
7
3050-0893
4
WASHER
8
0535-0043
4
NUT
28
Chapter 5

advertisement

Key Features

  • 4-terminal pair interface.
  • Direct attachment.
  • 1m (16048G) / 2m (16048H) cable length.
  • BNC connectors.
  • Compatible with HP 4294A LCR meter.
  • Fixture compensation required.

Frequently Answers and Questions

How to connect the 16048G/H to the 4294A?
Connect the 16048G/H to the 4294A by gradually coupling the four BNC connectors and fastening screws of the fixture with the test connectors and accessory mounting holes of the instrument until they come to complete contact. Fasten two of the four BNC connectors to the mating test connectors by gradually turning the BNC connectors' rotation levers until each pair of connectors are securely connected. Be sure to align the grooves on both sides. Turn clockwise the fixture's two fastening screws together, so that the fixture is secured to the instrument. Finally, secure the remaining two BNC connectors of the fixture by turning clockwise their rotation levers.
What is fixture compensation and why it is required?
Fixture compensation is required to compensate stray capacitance and residual impedance. This is important for accurate measurements when using the 16048G/H with user-fabricated test fixtures.
What is the maximum voltage the 16048G/H can handle?
The maximum voltage for the 16048G/H is ± 42V peak max. (AC+DC).