Scanivalve ZOC33 64Px, 64PxX2 Electronic Pressure Scanner Instruction and Service Manual
The ZOC33 64Px and ZOC33 64PxX2 are electronic pressure scanners which can accept up to 128 pneumatic inputs. Each ZOC33 module incorporates 64 individual piezoresistive pressure sensors, designed to facilitate field replacement. The ZOC33 electronic pressure scanning module is specifically designed for use in wind tunnels, flight tests or applications where space is at a premium and pressures will not exceed 50 psi.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
ZOC 33/64Px and
ZOC 33/64PxX2
Electronic Pressure Scanning Module
Instruction and Service Manual
PREFACE
Warnings, Cautions and Notes
The WARNING! symbol indicates that danger of injury for persons and the environment and/or considerable damage (mortal danger, danger of injury) will occur if the respective safety precautions are not taken.
The CAUTION ! symbol indicate danger for the system and material if the respective safety precautions are not taken.
The ESD note symbol indicates that proper precautions for handling Electrostatic Sensitive Devices needs to be taken when performing the related operation. This includes the use of grounded work surfaces and personal wrist straps to prevent damage to sensitive electronic components.
Warranty
Scanivalve Corporation, Liberty Lake, Washington, hereafter referred to as Seller, warrants to the Buyer and the first end user that its products will be free from defects in workmanship and materi al for a period of twelve
(12) months from date of delivery. Written notice of any claimed defect must be received by Seller within thirty
(30) days after such defect is first discovered. The claimed defective product must be returned by prepaid transportation to Seller within ninety (90) days after the defect is first discovered. Seller’s obligations under this Warranty are limited to repairing or replacing, at its option, any product or component part there of that is proven to be other than as herein warranted.
Surface transportation charges covering any repaired or replace ment product or component part shall be at Seller’s expense; however, inspection, testing and return transportation charges covering any product or component part returned and redelivered, which proves not to be defective, shall be at the expense of Buyer or the end user, whichever has returned such product or component part.
This Warranty does not extend to any Seller product or component part thereof which has been subjected to misuse, accident or improper installation, maintenance or application; or to any product or component part thereof which has been repaired or altered outside of Seller’s facilities unless authorized in writi ng by Seller, or unless such installation, repair or alteration is performed by Seller; or to any labor charges whatsoever, whether for removal and/ or reinstallation of the defective pro duct or component part or otherwise, except for Seller’s labor charges for repair or replacement in accordance with the Warran ty.
Any repaired or replacement product or component part there of provided by Seller under this Warranty shall, upon redelivery to Buyer, be warranted for the unexpired portion of the original product warranty.
THIS WARRANTY IS IN LIEU OF AND EXCLUDES ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, ARISING BY
OPERATION OF LAW OR OTHERWISE, INCLUDING THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE, AND IN NO EVENT SHALL
SELLER BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES.
In the event of a failure:
1) Notify Scanivalve Corporation, Customer Service
De partment. Include model number and serial number.
On receipt of this information, service data or shipping instructions will be forwarded. This may be transacted by telephone or e-mail.
2) On receipt of shipping instructions, forward the pro duct, transportation prepaid. Repairs will be made and the product returned.
3) All shipments should be made via “Best Way”.
The product should be shipped in the original packing container or wrapped in protective material and surrounded by a minimum of four (4) inches of a shock absorbing material.
Trademarks ® and Copyrights ©
Scanivalve is a registered trademark of Scanivalve Corporation.
All other brand and product names are trademarks or registered trademarks of their respective companies.
Packaging for Shipment
If the product must be shipped, whether being returned to Scanivalve or relocated to another location it must be packaged properly to minimize the risk of damage. The recommended method of packing is to place the instrument in a container, surrounded on all sides with at lease four inches of shock attenuating material such as Styrofoam peanuts.
Important Notice
Please note that the product specifications and other information contained in this manual are subject to change without notice. Scanivalve Corporation makes an effort and strives to provide complete and current information for the proper use of the equipment. If there are any questions regarding this manual or the proper use of the equipment, contact Scanivalve Corporation.
Contact Information
If there are any questions or concerns regarding any Scanivalve product please do not hesitate to contact us at the following:
Scanivalve Corp.
1722 N. Madson Street
Liberty Lake, WA 99019
Telephone: (800)935-5151 (509)891-9970
Fax: (509)891-9481 [email protected]
www.scanivalve.com
Scanivalve Corporation is an ISO9001:2000 certified company.
ZOC33
Table of Contents
TablE Of CONTENTS
PREfaCE
Warnings, Cautions and Notes
iv
iv
Warranty iv
Trademarks ® and Copyrights ©
Packaging for Shipment v v
Important Notice
Contact Information v v
Section 1: Specifications
General Specifications
Environment Specifications
Section 2: Introduction
General Description
ZOC33/64Px
ZOC33/64PxX2
ZOC33/64Px - Valveless
Thermal Control Unit (ZOCTCU)
Section 3: Installation & Operation 10
Unpacking 10
Overview 10
Electrical Inputs & Outputs 10
Pneumatic Inputs
Calibration Valve Operation
ZOC33 ZOCTCU Connectors
Sensor Installation
11
12
13
14
4
6
8
4
5
9
2
2
2
Section 4: Electronic Components
Amplifier Board
Multiplexer Board
Decoder Board
15
15
16
17
1
Section 1: Specifications
SECTION 1: SPECIfICaTIONS
Resolution
General Specifications
Size (WxHxD) 1.43” x 1.69” x 4.98”
(36.42cm x 42.8cm x 126.37cm)
Inputs (Px)
Weight
ZOC33/64Px
ZOC33/64PxX2
ZOC33TCU
(including ZOC33 module)
11 oz (312g)
13 oz (369g)
4.0 lbs (1.8kg) infinite
64 or 128 (PxX2)
Pneumatic Connectors
ZOC Px .042” tubulations
ZOC Control Pressures 0.063” tubulations
ZOCTCU 1-2 73 port connectors
(73ZOCM-063)
ZOC33
Full Scale Ranges ±10 inH
2
O, ±20 inH
2
O,
1psid, 2.5psid, 5psid,
15psid and 50psid
(2.5kPa, 5kPa, 7kPa,
17kPa, 35kPa, 100kPa, and 350kPa)
Electrical Connectors
ZOC Module
ZOCTCU
Power Requirements
ZOC33 Module
ZOCTCU
15 pin MDM 15SL2P
MS3100A Type, Size 20
+15Vdc @ 120mA
-15Vdc @ 30mA
+24VDC @ 1.0A
Accuracy (after calibration)
10 inH
20 inH
2
2
O
O
1psid
2.5psid
5psid
15psid
50psid
±0.15% FS
±0.12% FS
±0.10% FS
±0.10% FS
±0.08% FS
±0.08% FS
±0.08% FS
Full Scale Output
Standard
Optional
Sensor Excitation
±2.5Vdc
±5.0Vdc
+5Vdc Constant Voltage
Environment Specifications
Operating Temperature 0 to 60 °C
Overpressure (No damage)
10 inH
20 inH
2
2
O
O
1psid
2.5psid
5psid
15psid
50psid
5psi
5psi
5psi
200%
200%
200%
200%
Humidity
Operation
Storage
Shock & Vibration
Shock
Vibration
5 to 95% RH, Non-Condensing
5 to 95% RH, Non-Condensing
Temperature Sensitivity
MIL-STD-810D Curve H
10G
10G
Acceleration 10G
10 inH
2
20 inH
2
O
O
1psid
2.5psid
5psid
15psid
50psid
0.25%
0.25%
0.10%
0.10%
0.10%
0.10%
0.10%
0.10%
0.08%
0.05%
0.05%
0.05%
0.05%
0.05%
Media Gasses compatible with
Aluminum and Buna-N
Max Reference Pressure
Scan Rate
50 psig (350kPa)
40kHz (standalone)
2
ZOC33
3
Section 2: Introduction
SECTION 2: INTRODuCTION
General Description
The ZOC33 is an electronic pressure scanner which can accept up to 128 pneumatic inputs. Each ZOC33 module incorporates 64 individual piezoresistive pressure sensors.
Each eight pressure sensors are manufactured in a housing designed to facilitate field replacement. No special tools are required to access the sensors. The ZOC33 electronic pressure scanning module is specifically designed for use in wind tunnels, flight tests or applications where space is at a premium and pressures will not exceed 50 psi.
The ZOC33 is powered by ±15Vdc. The module is manufactured in a 64 channel model, a 128 channel duplexed model and a 64 channel valveless model. All models are available in multiplexed versions only.
The ZOC33 pressure sensors are arranged in blocks of 8.
In all variations except the valveless model, each block of eight sensors has its own individual calibration valve. This valve had four modes of operations:
(1) Operate
(2) Calibrate
(3) Purge
(4) Isolate
The modes are selected by applying control pressures in a predetermined and logical order. The ZOC33 calibration valves utilizes valve logic where the valve defaults to the purge mode when no control pressures are applied.
Beginning with serial number 144, all ZOC33 modules have a 500 ohm platinum RTD installed. This RTD is used by a
DSM3000/3200/3400 module or RAD3200/4000 module to determine the temperature of the ZOC33 module. This feature will be added to any older ZOC33 module received for repairs.
Beginning with serial number 390, all ZOC33 modules have a TEDS ID chip installed and programmed to interface with a RAD3200/4000 or DSM3400 pressure measuring system.
This feature will be added to all older ZOC33 modules received for repair.
Beginning with serial number 399, the ZOC33 valves were redesigned. This results in a small change to the module dimensions. This change only applies to valved models. A
ZOC33 valved model that requires repairs to the valves will have to be upgraded to the newest configuration.
4
ZOC33
ZOC33
ZOC33/64Px
This module contains sixty four (64) pressure sensors in eight sensor packs. Each sensor pack contains:
• eight sensors
• a calibration valve
• an excitation board
• a high speed multiplexer
The output of each sensor is directed to the multiplexer/ amplifier. The channel to be output is selected by a CMOS level binary address supplied by either a Scanivalve data acquisition system or by the customer’s data acquisition system.
Figure 2.1 depicts a ZOC33/64Px.
Section 2: Introduction figure 2.1 - ZOC33/64Px Dimensions [inches(cm)]
5
Section 2: Introduction
ZOC33/64PxX2
This module contains sixty four (64) pressure sensors in eight sensor packs. Each sensor pack contains:
• eight sensors
• a calibration valve
• a duplexing valve
• an excitation board
• a high speed multiplexer
• 16 input tubes (Bank A and Bank B, 8 inputs each)
The sensors are arranged in eight groups of eight. Each sensor has two pneumatic inputs: Bank A and Bank B. The inputs are switched pneumatically by enabling the duplexing valve. Even though there are 128 pressure inputs only one Bank of 64 pressure inputs can be measured at one time. Each group of eight sensors may be a different range.
The standard output of the module is ±2.5Vdc corresponding to the channel selected by a CMOS level 6 bit binary address.
ZOC33
Figure 2.2 depicts a ZOC33/64PxX2, Serial number 398 and earlier.
Figure 2.3 depicts a ZOC33/64PxX2, Serial number 399 and later.
figure 2.2 - Early ZOC33/64PxX2 Dimensions [inches(cm)]
6
ZOC33
Section 2: Introduction figure 2.3 - Current ZOC33/64PxX2 Dimensions [inches(mm)]
7
Section 2: Introduction
ZOC33/64Px - Valveless
This module contains sixty four (64) pressure sensors in eight sensor packs. Each sensor pack contains:
• eight sensors
• an excitation board
• a high speed multiplexer
The output of each sensor is directed to the multiplexer/ amplifier. The channel to be output is selected by a CMOS level binary address supplied by either a Scanivalve data acquisition system or by the customer’s data acquisition system.
ZOC33
This version does not have a calibration valve. All calibrations are performed using the reference port. In wind tunnel applications, a zero offset correction is performed
“wind off.” The valveless design is also preferred for rotating applications as no control pressures are required.
Figure 2.4 depicts a ZOC33/64Px Valveless.
figure 2.4 - ZOC33/64Px Valveless Dimensions [inches(cm)]
8
ZOC33
Thermal Control unit (ZOCTCu)
An optional Thermal Control Unit (ZOCTCU) (Figure 2.5) is available for applications where temperature swings my be great enough to exceed the compensated range of the sensors. Exceeding the compensated temperature range can induce errors in the pressure measurements. See Section
1, Specifications for more information on the compensated temperature range. The ZOCTCU consists of a housing, insulation and a proportional heater which will maintain the temperature for the module at 40°C ±0.3°C.
Section 2: Introduction figure 2.5 - ZOC33 Thermal Control unit (128 channel shown)
9
Section 3: Installation & Operation
ZOC33
SECTION 3: INSTallaTION & OPERaTION
unpacking
All ZOC33 modules have been extensively tested prior to shipment. All modules are packed to minimize the chances of shipping damage. However, damage can still occur. The customer must inspect the modules and shipping materials for obvious signs of damage. If it is suspected that damage may have occurred, contact Scanivalve Corporation immediately.
Once you have unpacked the module, do an inventory check of the shipment. Each shipment should at least include a mating connector and this manual.
all other ZOC cable serviced modules. The ZOC33 may be installed into existing HyScan systems without changing configurations.
The user is cautioned to follow safe instrument handling practices when handling the ZOC33 modules. This includes:
(1) Make and break all connections to the module with the power off.
(2) Recommended power input to the module is
±15Vdc.
Figure 3.2 shows the output connector pin assignments for all of the ZOC33 variations. This figure shows the RTD connections (+Tem and -Temp) and the TEDS ID chip output
(ID). These connections may not be active on all modules.
Refer to Section 2, Introduction for more information.
Overview
ZOC33 modules are designed to function best when used with one of Scanivalve Corporation’s data acquisition systems, either a DSM3000/3200/3400 or RAD3200/4000.
All ZOC33 modules will function with older data acquisition systems such as HyScan 2000/1000 as well. They can also be used as a standalone module with another high speed data acquisition system.
Electrical Inputs & Outputs
The Electrical Input and Output wiring is compatible with
CAUTION! Not following standard safe instrumentation handling practices could permanently damage the modules.
figure 3.1 - ZOC33/64Px Exploded View
10
ZOC33
Section 3: Installation & Operation
Manifold blocks are provided with the module to facilitate plumbing. This also permits easy removal of a sensor pack in the event that field repairs are required.
figure 3.2 - ZOC33 Input/Output Connector
Pneumatic Inputs
Pneumatic inputs consist of: Px Inputs(8 or 16), Control
Pressure Inputs ( PxA Ctl, Cal Ctl), a Calibration Input, and a Reference Input. The duplex version also has a duplex control pressure input (PxB). Valveless units do not have
Control Pressure and Calibration Inputs
All Px inputs are .040 inch(1.067mm) bulged tubulations.
These tubulations are designed to accept any .042 inch tubing manufactured by Scanivalve Corp. Each sensor pack valve block contains eight (8) or, in the duplex version, sixteen(16) Px inputs. ZOC33 modules are capable of measuring pressures up to 50 psid.
Control pressure inputs consist of: PxA CTL and CAL CTL.
The duplex module has a third control pressure input, PxB
CTL. These inputs are used to switch the valve logic to each of the four (4) states: Operate, Calibrate, Purge, and Isolate.
The control pressures must be 65 psi. Figure 3.3 shows the
64Px valve logic. Figure 3.4 shows the 64PxX2 Valve Logic.
Calibration/Reference Inputs consist of a Calibration input and a Reference input. The Calibration input is an .042 inch
O.D. tubulation. It is normally connected to a source of calibration pressures. Internally, this input is manifolded to all of the sensors through the calibration valving. The
Reference input is an .063 inch (1.6mm)O.D. tubulation. It provides a point of reference for the transducers. All of the sensors in each block of eight share a single reference.
figure 3.3 - ZOC33/64Px Pneumatic Inputs figure 3.4 - ZOC33/64PxX2 Pneumatic Inputs
11
Section 3: Installation & Operation figure 3.5 - ZOC33/64Px Valveless Pneumatic
Inputs figure 3.6 - ZOC33 Channel Identification
Calibration Valve Operation
As discussed in Section 2, Introduction, ZOC33/64Px and
ZOC33/64PxX2 modules are equipped with an internal calibration valve in each sensor pack. The calibration valve can take one of four states:
It is important that all control pressures should be dry, filtered instrument air or nitrogen.
Operate Mode
This connects each Px input to its associated pressure sensor. The ZOC33/64PxX2 module allows the customer to select one of two banks for input.
Calibrate Mode
This mode connects all the pressure sensors to the calibration input.
ZOC33
Purge Mode
This mode connects the Px inputs to the pressure sensors and the calibration input. A safe purge pressure can be applied to purge input lines.
Isolate Mode
This mode isolates the pressure sensors from the
Px and calibration lines.
Tables 3.1, 3.2 and 3.3 are state tables that describe pneumatic logic for each state of the valves.
Table 3.1 - ZOC33/64Px Valve
Logic
Mode Px CTL CAL CTL
Operate
Calibrate
Purge
X
65psi
X
65psi
X
X
Isolate 65psi 65psi
Table 3.2 - ZOC33/64PxX2 Valve Logic
Mode
Operate A
Operate B
Calibrate
PxA CTL
X
65psi
65psi
PxB CTL
65psi
X
65psi
CAL CTL
65psi
65psi
X
Purge
Isolate
X
65psi
X
65psi
X
65psi
Table 3.3 - ZOC33/64Px Valveless
Valve Logic
Mode Px CTL CAL CTL
Operate X X
Calibrate
Purge
N/A
N/A
N/A
N/A
Isolate N/A N/A
NOTE: Valveless Modules do not have control pressure inputs
12
ZOC33
ZOC33 ZOCTCu Connectors
Figure 3.7 shows the pneumatic and electrical connectors on a ZOC33 ZOCTCU. View B shows the electrical connector, View A shows the pneumatic connector for inputs 1-64, control pressures, the reference input and the calibration inputs. View C shows the pneumatic inputs for channels
65-128. This connector is not installed on a non-duplexed module. The drawing has been modified from the actual configuration for clarity.
figure 3.7 - ZOC33 ZOCTCu Connectors
Section 3: Installation & Operation
Channels
Table 3.4 - ZOCTCU Pneumatic Inputs
Connector Tubes
1-64
PxA CTL
View A
View A
1-64
PxA CTL
PxB CTL View A PxB CTL
Notes
PxB is not connected in a nonduplexed module
CAL CTL
CAL HI
CAL LOW
View A
View A
View A
CAL CTL
CAL HI
CAL LOW
REF HI
REF LOW
View A
View A
REF HI
REF LOW
1-64 View C 1-64
This connector is only installed for duplex modules
Pin
Table 3.5 - ZOCTCU Electrical Inputs/Outputs
Function Notes
G
H
E
F
C
D
A
B
L
M
J
K
P
R
S
T
U
V
Address 0
Address 1
Address 2
Address 3
Address 4
+15Vdc
-15Vdc
+ Temp
- Temp
±15Vdc Return
+ Output
- Output
+ Temp
(Heater)
ID
Address 5
- Temp
(Heater)
+28Vdc
+28Vdc Return
ZOC33 Internal RTD
ZOC33 Internal RTD
±2.5Vdc Nominal
Heater Temperature - Pre
March 2000 only
TEDS Chip information
Heater Temperature - Pre
March 2000 only
Heater Power
13
Section 3: Installation & Operation
Sensor Installation
The sensors used in a ZOC33 module are piezoresistive.
They are a standard Scanivalve Corporation “S-sensor”.
They are mounted in chip carriers which are then installed in 8-channel sensor packs. The sensor packs are field replaceable with standard hand tools. Replacing a sensor in the field is possible but not recommended without prior training from Scanivalve. If a sensor must be replaced, follow these instructions.
CAUTION! ESD PROTECTION
REQUIRED. The proper use of grounded work surfaces and personal wrist straps are required when coming into contact with exposed circuits to prevent static discharge from damaging sensitive electronic components.
ZOC33
To replace a sensor, follow this procedure:
(1) Remove appropriate sensor pack assembly from the
ZOC33 (screws for each sensor pack are located at the base of the module).
(2) If necessary, remove the valve from the sensor housing.
This should be done only if tubing restrictions do not allow the sensor pack to be pulled.
(3) Remove the sensor screw access plate. DO NOT
REMOVE O-RINGS.
(4) Remove the PC board.
(5) Loosen the sensor hold down screw. DO NOT REMOVE
THE SCREW OR O-RING.
(6) Carefully remove the damaged sensor. Take care not to damage the wire bonds.
CAUTION! Failing to carefully follow this procedure could permanently damage the modules.
figure 3.8 - ZOC33 Sensor Pack Exploded View
14
ZOC33
Section 4: Electronic Components
SECTION 4: ElECTRONIC COMPONENTS
amplifier board
The amplifier board receives an millivolt input from the channel selected by the multiplexer. The signal is amplified to a nominal 2.5 Vdc full scale and output through the decoder board to the I/O connector. The amplifier gain is set by selecting R1 to match the average output of the sensors. The amplifier (INA110) has a settling time of 20 microseconds which means that the module channels can be scanned at 40 kHz in a “stand-alone” configuration.
Gain and zero adjustments are provided to permit the user to better match the ZOC33 to a non-HyScan system
(Figure 4.1). This circuit also receives an input from an RTD mounted on the Decoder Board. The RTD circuit provides a feedback to the amplifier that increases the gain by approximately 0.2%/°C. This will compensate for the 0.2%/°C decrease in output inherent in pieziorestive sensors.
figure 4.1 - Zero and Span adjustments
CAUTION! Adjusting the Zero and
Span adjustments will invalidate any current calibration coefficients for the module.
figure 4.2 - ZOC33 amplifier board Schematic
15
Section 4: Electronic Components
Multiplexer board
Each eight (8) channel sensor pack has its own combination multiplexer and excitation board. The board consists of a precision voltage regulator, eight sensor mounts, and a multiplexer. The board is installed on the sensor housing
(Figure 3.8).
The board is powered by ±15Vdc. This voltage must be regulated, but does not need to be “instrumentation quality”. The precision voltage regulator is an LT1021 which converts the ±15Vdc to a precision 5.00Vdc. The LT1021 has a very tight output voltage tolerance and a very low temperature coefficient. The output of the LT1021 is used as an excitation voltage for the sensors.
ZOC33
The multiplexer receives millivolt input signals from each of the sensors. It also receives inputs from the decoder board which select the channel to be output to the amplifier board. The multiplexer is a DG507. It has a settling time of
400 nanoseconds.
The Multiplexer board plugs into the decoder board.
figure 4.3 - ZOC33 Multiplexer board Schematic
16
ZOC33
Decoder board
The decoder board is the main circuit board in the ZOC33.
All of the sensor packs and the amplifier plug into this board. The decoder board receives the inputs from the data system and converts them into signals to drive the sensor packs.
Address lines A0 through A5 are routed from the I/O connector to two 4042 latches which convert the address input to a 3 bit channel address and a one of eight enable. A channel in the module is selected by the following:
Section 4: Electronic Components
The data system outputs an address for a channel to be read. The address is sensed by the decoder board and converted to an address and enable output. The enable line selects the sensor pack and the address line selects the channel in that pack. The signal output of that sensor is then input to the amplifier board, amplified and routed back through the decoder board to the I/O connector.
figure 4.4 - ZOC33 Decoder board Schematic
17
Section 4: Electronic Components
ZOC33 figure 4.5 - ZOC33 Decoder board Connector Pinouts
18
ZOC33
Section 4: Electronic Components
19
ZOC33
1722 N. Madson St.
liberty lake, Wa 99019
Phone: 1-800-935-5151
1-509-891-9970 fax: 1-509-891-9481 [email protected]
www.scanivalve.com
21
ZOC33 Service Manual
July, 2010
advertisement
Key Features
- 64 pressure sensors
- 128 pneumatic inputs (duplexed model)
- Field replaceable sensor packs
- Designed for wind tunnels and flight tests
- ±2.5Vdc output
- 40kHz scan rate
- Optional Thermal Control Unit (ZOCTCU)
- Internal calibration valve