OsteoSys DEXXUM T Osteoporosis diagnostic User Manual
DEXXUM T is an osteoporosis diagnostic using x-ray. It uses x-rays having two different energies and calculates the value of bone mineral density, by analyzing images acquired from scanning of spines and femurs in pencil-beam method and comparing them with collected data. Scanning method: Pencil beam. Measurement points (measuring time): AP spine (fast: 1 min 25 sec normal: 3min 12sec) Femur (Fast: 1min 5sec normal: 1min 59sec). Measurement method: Non-stop Scan (AP Spine, Dual Femur) Simultaneous measurement of 3 points is possible. Reference Data: Korean standard data. Reproducibility: below 1%. Modification of specific regions: Users can add or delete bones and tissues to reduce the errors in case of BMD calculation of fracture, implant or operating regions. Auto ROI: After measurement, ROI (region of interest) are divided automatically. Radiation dose: below 10mRem per dose. System linkage: DICOM support. Multi-print function: Results, subject information, and trend are printed. Trend comparison: comparison of trend data by region.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
DEXXUM T
User Manual
1
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
DEXXUM T
User Manual
Contents of User manual
2 safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
User Manual
Safety Precautions
1. General Use
........................................................................................................................................................... 4
2. Safety in Use and Measures ..........................................................................................................12
3. Washing and Maintenance
..................................................................................................................... 17
Contents of User manual
3 safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
General Use
1. Notation Method of Manual ............................................................................................................. 5
1) Notation Method ........................................................................................................................................... 5
2) Label .................................................................................................................................................................... 6
3) Mark .................................................................................................................................................................... 7
2. Product Instructions
......................................................................................................................................... 8
3. Software
................................................................................................................................................................... 9
4. Personal Data Protection
.............................................................................................................................. 9
5. Installation and Starting of Equipment
............................................................................................. 10
6. Equipment Maintenance
............................................................................................................................ 11
Contents of General Use
4 safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Notation Method of Manual.
1) Notation
User manual and installation manual are described according to the following display method and consist of following notation methods.
- Methods of sequence description, symbolic notation and icon notation
Description contents with sequences are described in alphabetical order. a. Contents of sequence No.1 b. Contents of sequence No.2 c. Contents of sequence No.3
Examples on the descriptions in manual are described in the box marked example.
Example) a. Place the patient on the bed of measurement equipment. b. Check if measurement points are within measurement range and place them in the exact position. c. Select yes/no when display window of user program asks, “Will you move to standard position?”
Additional descriptions are described as‘ - ‘ and they show the information to supplement brief information.
Information on important things or marks to be paid attention to among the manual descriptions
: Contents to be noted concerning equipment use
: Contents to be known thoroughly concerning equipment use
: Pages to be referred to
5
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2) Label
Label 1.
“ CAUTION! “ warning label against laser radiation
- Staring directly at the outlet of laser radiation without protection equipment when the laser is turned on can cause serious damage to eyeballs.
Label 2.
“ CAUTION “ warning label against High voltage
- Users can be exposed to risks due to important equipments with high voltage.
Label 3.
“ WARNING “ warning label against equipment movements
When fingers are pinched between moving measurement arm of equipment and patient table, accidents likely to occur.
6
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3) Marks
: X-ray is turned on
It is the mark indicating X-ray is being radiated at present. It appears on monitor as well as controller of equipment during measurement or daily test.
: X-ray is turned off
It is the mark indicating X-ray is not being radiated. It appears on monitor in case of stand by mode, standing ready to be measured, or measurement completion.
: laser pointer
- It is a mark indicating the dangers associated with laser pointer.
: Power is turned on
- It indicates power is turned on. (power-on)
: Power is turned off
- It indicates power is turned off. (power-off)
: Start mark
- It indicates X-ray generation is in the preparation stage.
: Communication state
- It indicates the communication state of equipment.
: Caution
- It indicates when caution is needed for safety or the case when instructions are
- needed.
7
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2. Product Instructions
The users of this equipment need knowledge of equipment and train for equipment use.
They shall read and understand the operation manual correctly and then operate the equipment according to the sequences.
Especially attention to the “caution” mark as it appears when there are directions or warnings in each partial item.
Installers and users of equipment shall take the responsibility for equipment operation according to the related laws and regulations of regions where the equipment is installed and operated.
Users shall take adequate education on adequate use of equipment and know well the use procedures by repetitive learning of manual, periodically or non periodically.
It is recommended that users should simulate to provide against emergencies, while training the equipment use repeatedly.
8
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3. Software
User programs and application programs for equipment use are protected by copyright.
Only authorized users of OsteoSys shall have rights to use software of this company.
Warning
We shall not take the responsibility of death accidents by equipment or equipment damages when persons who are not authorized to use the software/programs use software, or the damages occur by fabricating, modifying, illegally altering software/programs without our company’s permission.
4. Personal Data Protection
Personal information or figures about patients shall be protected under related laws.
9
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
5. Installation and Starting of Equipment
Equipment installation and position change of installed equipment shall be based on related laws.
Please note that manufacturer, seller, installer, importer and OsteoSys shall not take any responsibility for malfunction of equipment or measurement values in the following cases.
* If the equipment was not used according to the directions of manual.
* If the person who has not been authorized by our company performed the software install, upgrade, modification or sales after service.
* If the service parts affecting equipment safety are not formal ones of our company.
* If the environment of equipment installation and starting is out of standard.
10
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
6. Equipment Maintenance
It is recommended to periodically perform examinations and maintenances of equipment, and check up safety devices and equipment, for the safety of equipment users, patients and the third.
The details about examination and maintenance of equipment are indicated in
“contents” of equipment operating procedure in manual, and it is recommended to periodically perform procedures on examination and maintenance.
When you need services, please contact our AS team or local agencies.
Contact number of Osteosys A/S dept.: +82-2-2025-1663(4) / 1650
Periodical examination and required specific examination can be added according to the related laws.
11
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Safety in Use and Measures
1. Equipment Operation .........................................................................................................................13
2. Use of Emergency Switch .................................................................................................................13
3. Interruption of Operations of Equipment in Case of Emergency Evacuation ........15
4. Interruption of Operations of Equipment in Case of Fire ..................................................15
5. Safety instructions on Measurement Assistant .......................................................................16
Contents of Safety in Use and Measure
12
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Equipment Operation
DEXXUM T is equipment that generates X-ray and measures the spine and femur of patients using the movements by motor drive. Subjects and patients need to be careful of motor movements for safety reasons.
Please master the equipment operating method.
If users lean against the equipment arm or touch it during the stand-by phase or measuring, safety problems can occur in users or subjects.
Do not let subjects sit down or stand up while measuring.
Do not approach the moving arm when turning on the equipment in “OFF” state. If power is turned “ON”, the equipment goes to the initial position automatically.
When abnormal operating or noise occurs during operation, press the emergency switch and contact related agencies or our company’s AS team.
2. Use of Emergency Switch
When using emergency switch due to unexpected situation, follow the procedures below. a. Press the emergency switch in case of emergency or abnormal operating. b. Push the arm in one direction with hands so that a subject may come down. c. Have the subject come down from the equipment and then change the main switch of equipment to off-state. d. Return to the original state by pressing emergency switch once more. e. Take the necessary measures and turn on the main switch. f. Check if equipment operates.
13
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Press it
2. Move it with hands so that a patient can come down.
14
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3. Interruption of Operations of Equipment in Case of Emergency Evacuation
In case of fire incidences of buildings or evacuations due to natural calamities, follow the procedures below. a. Press the emergency switch and evacuate the subject by pushing arm in one direction. b. Separate the main switch or power and then promptly lower and evacuate the subject.
4. Interruption of Operations of Equipment in Case of Fire
a. In case of fire, flip the main switch down and separate the power connected to equipment. b. Lower the subject and promptly evacuate.
If the equipment is smoked or burned, gas can be generated.
Osteosys recommends that you should have measures for fires.
15
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
5. Safety instructions on Measurement Assistant
Measurement assistant made or approved by Osteosys shall be used for safety use of equipment.
’
Please note that if any errors of measurement value occur by using unapproved assistant, users shall take the whole responsibility for that.
16
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Washing and Maintenance
1. Washing of Equipment ......................................................................................................................18
2. Washing of Monitors ..........................................................................................................................19
3. Washing of Keyboards .......................................................................................................................19
Contents of Maintenance
17
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1.
Washing of Equipment
Warning
Turn off the main switch before washing.
Separate the AC power cable from power source.
Notice
Do not eject the spray type cleansing solution or cleansing solution in a spray container onto equipment.
If spray type cleansing solution or cleansing solution is ejected to equipment, it has a very high possibility to cause damage to electronic machines within equipment.
Be sure that cleansing solution might not be flowed into the equipment inside.
- Contact areas between patients and equipment shall be kept clean.
- When cleaning cover of equipment, wipe it using soft cloth dipped in cleansing solution mixed with water or neutral detergents.
Sprinkling water on equipment itself or soaking the equipment in water can cause damage it.
- Do not use solvent or paint remover.
18
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2. Washing of Monitors
Notice
When cleaning monitors, do not use acid or alkaline detergents. It can cause damage on monitor surface.
- When cleaning monitors, brush the dust off with a soft brush or wipe it off with a soft cloth dipped in a neutral detergent when need wiping.
It is recommended to wipe the surface of monitors with a soft cloth or use a cleaner for monitor screens.
3. Washing of Keyboards
- When cleaning keyboards, brush the dust off with a soft brush or wipe them off with a soft cloth dipped in a neutral detergent when need wiping.
19
Safety
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
DEXXUM T User Manual
1.
Product Introduction of DEXXUM T .................................................................................................. 21
2.
Basic Installation Method of DEXXUM T ......................................................................................... 26
3.
Instructions of DEXXUM T Program .................................................................................................. 24
4.
Documents Related to DEXXUM T ..................................................................................................... 78
Contents of User Manual
20
System Overview
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
DEXXUM T Product Introduction
1.
Product Description .................................................................................................................................. 22
2.
Outlines of DEXXUM T .......................................................................................................................... 24
3.
System Configuration of DEXXUM T ............................................................................................... 25
Contents of DEXXUM T introduction
21
System Overview
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Product Description
Osteosys’ DEXXUM T is an osteoporosis diagnostic using x-ray.
It uses x-rays having two different energies and calculates the value of bone mineral density, by analyzing images acquired from scanning of spines and femurs in pencil-beam method and comparing them with collected data.
- Scanning method: Pencil beam
- Measurement points (measuring time):
AP spine (fast: 1 min 25 sec normal: 3min 12sec)
Femur (Fast: 1min 5sec normal: 1min 59sec)
- Measurement method:
Non-stop Scan (AP Spine, Dual Femur)
Simultaneous measurement of 3 points is possible
- Reference Data: Korean standard data
- Reproducibility: below 1%
- Modification of specific regions: Users can add or delete bones and tissues to reduce the errors in case of BMD calculation of fracture, implant or operating regions.
- Auto ROI: After measurement, ROI (region of interest) are divided automatically
- Radiation dose: below 10mRem per dose
- System linkage: DICOM support
- Multi-print function: Results, subject information, and trend are printed.
- Trend comparison: comparison of trend data by region
22
System Overview
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2. Outlines of DEXXUM T
1
8
2
7
6
3
4
5
1. Emergency switch
2. Carriage
3. bed wings(top)
4. bed wings(top) – folded
5. front cover
6. subject table
7. bed wings (bottom)
8. Operation switch for users
23
System Overview
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3. Outlines of User Operation Switch
13
12
1
2
8
7
11
10
9
3
4
5
6
1. Device “ON” : When the power of DEXXUMT equipment is turned “on”, Green LED is turned on.
2. Laser pointer “ON” LED : When laser pointer is turned on, red light is turned on.
3. Move “Right”: It is a direction key making it move to the right hand direction of a subject..
4. Move “Bottom”: It is a direction key making it move to the feet direction of a subject.
5. Move “Top” : It is a direction key making it move to the head direction of a subject.
6. Move “Left” : It is a direction key making it move to the left hand direction of a subject.
7. Left Forearm : Select the left forearm of a subject. Green LED is turned on.
8. Left Femu r: Select the left femur of a subject. Green LED is turned on.
9. Spine : Select the spine of a subject. Green LED is turned on.
10. Right Femur : Select the right femur of a subject. Green LED is turned on.
11. Right Forearm : Select the right forearm of a subject. Green LED is turned on.
12. Shutter Open : When the shutter is opened, yellow LED is turned on.
13. X-ray ON : It is in the course of X-ray radiation and yellow red LED is turned on.
24
System Overview
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
4. DEXXUM T System Configuration
- DEXXUM T
- Computer system
1) Main frame
2) LCD monitor
3) Printer
- Protective wall (Domestic)
- Measurement assistant (tripod)
- Daily check Phantom
- User manual
- Service manual
- Power cable
- LAN cable for communication
- Software installation CD
- Supplies: A4 paper, printer ink
25
System Overview
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Basic Installation Method of DEXXUM T
1.
Equipment Movement .......................................................................................................................... 27
2.
Equipment Installation .......................................................................................................................... 29
3.
Software Installation............................................................................................................................... 30
1) PC LAN CARD NETWORK IP setting
2) Installation of User Program
3) Installation and Setting of Database
Contents of Basic installation Method
26
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Equipment Movement
- DEXXUM T has 4 wheels and a fixed supporter in the lower part of equipment.
- Methods for equipment movement are as follows. a. Turn off the main power. b. Separate the AC power connector and LAN cable of PC from equipment. c. Draw and lower the both wings of bed of DEXXUM T. d. Leave a space between the supporter and floor by turning the fixed supporter anticlockwise. e. Push the equipment using wheels.
When moving DEXXUM T, leaned or lifted, carriage shall be protected for equipment protection. In this case, contact our AS team or agencies.
Be especially careful when moving equipment, as one might get injured due to a non-fixed carriage.
Positions of main switch, power and LAN cable terminal
Main switch
AC power terminal and
LAN cable terminal
27
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1
2
Fixation block is fixed when turning anti-clockwise direction
Fixation block loosens when turning clockwise
Method for folding bed wings of equipment and methods for fixing and unfixing of
fixation blocks
28
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2. Equipment Installation
- The sequences of equipment installation are in reverse order of equipment movement.
- Methods for equipment installation are as follows. a. Unpack a box and remove the carriage in equipment and HFG fixation block. b. Push the equipment and place it in re-installation space. c. Have four parts of fixation block and ground to be in contact by turning it clockwise. d. Lift up the both wings of bed of DEXXUM T and fix them into the equipment. e. Plug it into AC power and connect LAN cable and PC. f. Turn on the main power.
Installing or storing in unsuitable places can slow down performance of equipment or cause damage or abnormal operating.
Examples of unsuitable places for equipment installment or storing
1) The place with too much humidity
2) The place exposed to direct rays of sun
3) The place whose temperature is too much higher or too much lower than adequate temperature for product (15 ℃ ~ 40 ℃ )
4) The place that can affect products due to impact or vibration
5) The place slanted too sharply
6) Storage of chemicals
7) The place where equipment generating a magnetic field is placed
Be sure to check if main power is turned off and then connect, when reinstalling equipment and authorizing AC power. Otherwise the equipment can be damaged.
Equipment shall be placed on even ground. Otherwise it can be moved to the direction to which carriage is slanted or HFG is slanted, when power is turned off.
29
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3. Software Installation
1) PC LAN CARD Network IP Setting
DEXXUM T communicates over LAN with PC. Therefore, it is necessary to set up specific IP address in general PC to communicate between PC and equipment. a. Right click the network setting icon on the desktop. b. Click the “properties”. c. Click the Internet protocol (TCP/IP) at the bottom of list under the general tap in the properties window and then click the “properties”. d. When the window appears as below, enter the following in the column of IP address and click the “OK”. e. Click the “OK” and finish the setting.
30
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2) Installation of User Program a. Insert your CD into the PC. b. Click the “My Computer” icon and right click the drive where CD was inserted. c. Click the “Explore” and find out and click the “ Setup.exe “ file when window appears. d. Click the “Next” when the following window appears.
31
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 e. Click the “Install“ and start the installation when the following window appears. f. Finish the Installation.
32
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3) Installation of Database
- Install the database storing patient information. a. Go to the “Database” folder in the path where user program is installed.
(Move to C:\Program files\Osteosys\DEXXUM T\Database.) b. Double click the “Setup.exe” of database folder. c. Press the “Next” and select the “Typical Type”.
33
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 d. Click the “Install”. e. Database is being installed. f. When “My-SQL sign up” window appears, click the “Skip Sign-up”and then click the “Next”.
34
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 g. Click the “Finish“ when database has been installed.
It automatically connects to optimal database setting and then click the
“Next”. h. When “MySQL server interface configuration” window appears, select the
“Standard Configuration” and click the “Next”.
35
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 i. Select the “Install As windows Service” checkbox in the next window and press the “Next”. j. Check the “Mysql Secutiry Settings” and insert the following in the blank and click the “Next”.
Current root Password: SA
Next root Password: SA
Confirm: SA
Select the checkbox of “Enable root access from remote machine”.
36
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 k. Configuration of database has been finished. l. Click the “My computer” icon on the desktop and enter the folder path of
C:\Program files\Osteosys\DEXXUM T m. Execute the “DBsetting.bat” file. n. Once “DBsettings.bat” file is executed, “C:\Windows\System32\CMD.exe” window appears. Insert the “ SA “ when the item of “Enter password :
“appears. o. Automatic file is executed and installation has been completed.
Basic System
37
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
The Directions of DEXXUM T Program
1.
Starting of Program .................................................................................................................... 39
2.
Configuration ................................................................................................................................. 39
3.
Registration of Doctors.............................................................................................................. 44
4.
Daily Check ..................................................................................................................................... 45
5.
Registration of Personal Information of Subjects ........................................................... 49
6.
Modification of Personal Information of Subjects ......................................................... 50
7.
Modification of Personal Information of Subjects ......................................................... 50
8.
Searching of Subject Data ........................................................................................................ 51
9.
Measurement of Bone Mineral Density .............................................................................. 52
10.
Interruption of Bone Mineral Density Measurement .................................................. 60
11.
ReScan .............................................................................................................................................. 61
12.
Descriptions of Analysis Screen ............................................................................................. 63
13.
Analysis of Spine Images .......................................................................................................... 67
14.
Analysis of Femur Images ...................................................................................................... 68
15.
Use of Brush ................................................................................................................................... 70
16.
Print of Results and PACS Linkage ..................................................................................... 73
17.
Print of Trend ............................................................................................................................... 76
Contents of The Direction of User Program
38
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Starting of Program
- Turn on the equipment, start the user program by clicking the “DEXXUM T” icon in the desktop. Power LED of user operation switch is turned “on” and equipment arm goes to the initial position if it is not there.
2. Configuration
- Configuration shall be done once at first after installation.
- Configuration is done in a user program for equipment use.
- Select the checkbox according to the descriptions by each item.
1) Settings Tap a. Patient ID Mask z
Consecutive number: Give ID number serially from 1. z
User-definition Blank: Insert “A” (Letters from A to Z can be inserted. Other characters cannot be inserted.)
Insert “9” (Numbers from 0 to 9 can be inserted. Letters other than numbers cannot be inserted.)
Insert “#” (Both alphabetical letters and numbers can be inserted regardless of digit number.)
If ID 8AD123 shall be inserted
Î
“ 9AA999 “ or “ ##### “ or “##########”
“#” can be used regardless of digit number.
39
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 z
Resident registration number of Korea Republic b. Default Settings z
Scan Site: It is an option that sets up more than two regions (Spine / Left Femur
/ Right Femur) in continuous mode
Once it is set up, position is set up at a time without selecting position by site before finishing, using one scan method. z
Gender: It is an option that selects the basic setup of subject gender. z
Menopause: It is an option that selects if the subject experienced menopause when examining a female subject. z
Ethnicity: It is an option that selects the basic set up of subject ethnicity. z
Print Preview: Set the basic setup of print preview. z
Doctor: It is selected when one person uses equipment and in this setting password is not asked when program starts. z
Forearm Scan: It sets up the forearm measurement mode. z
Spine: It sets up if the representative value of spine shown in report will be printed. z
Femur: It sets up if the representative value of femur shown in report will be printed. z
Scan Mode: It sets up the use of T-Scan Mode. c. Language z
Language can be selected among English, Korean, German, Spanish and Chinese.
(Be sure to restart the program so that it might be applied.)
40
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2) QC/DB/Back UP a. Calibration
- Fix the setup on calibration.
- z
Interval of daily test: Set up the calibration interval. z
Image Files Path: Appoint the folder where subject’s images are to be stored.
(value: D:\DT\image) b. Database
- It is a setup related to database and fixed when network mode is used. c. Back up
- Back up File Path: Appoint the folder where subject’s data is to be backed up.
(value: D:\D3N_Backup) d. DICOM
- Set up the DICOM. z
Set up the IP / Port / Atitle suitable for each hospital.
41
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3) Hospital
- Enter the hospital information.
- The hospital information is shown on the top of subject’s report paper. a. Hospital information z
Hospital Name: Enter the hospital name. z
Hospital Address: Enter the hospital address. z
Hospital Telephone: Enter the hospital telephone numbers.
Use LOGO : It sets up if image file of hospital will be shown in report paper. To add logo, you can enter the path of hospital logo image file into path column.
Logo addition
42
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
4) Doctor
- It has password setting on doctors’ name, telephone numbers, fields, and software use. a. Basic information z
Name: Enter the name of doctor in charge who uses it. z
Telephone: Enter the telephone number of doctor in charge. z
Department: Enter the department of doctor in charge. b. Password Change
- Change the password of doctor in charge being used. z
Current Password: Enter the password of doctor in charge at present. z
New Password: Enter the new password of doctor in charge at present to be changed. z
New password Confirmation: Enter the new password one more time.
43
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3. Registration of Doctors
User program of DEXXUM T reports patient records by setting separate password by doctor when many doctors use it. When one doctor uses it, refer to “Doctor” column of configuration at page 35. a. Doctor Selection Menu z
Select: If there are many doctors, click a doctor and Select. z
New: It is used when a new doctor is added.
Delete: It is used to delete the name of doctor and it can be moved to another doctor database or deleted completely, if there is patient data in database. z
Exit: It is selected when the user forgot his password or finished the program. b. Doctor select z
Click the “Select” button after selecting a doctor, or double click after selection. c. Insertion of Password z
Enter the password when password window appears and press “OK”.
If you forget your password, contact AS center of our company for protection of personal information.
44
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
4. Daily Test
- DEXXUM T requires daily check, according to the intervals set up in configuration, using daily check phantom.
It is recommended to carry out daily test once a day for the more exact use of equipment. a. Click the “Daily Test” icon, the second icon from the upper end of main screen of user program. b. There are red/green indicators that show communication connection in right bottom of the screen.
In case of red, stand by for a while until it turns to green.
Red: communication between equipment and PC is in preparation.
Green: equipment and PC is communicating with each other.
45
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 c. Click the “Daily Test Start” button at the bottom when it turns to green. d. Carriage of equipment starts to move to default position and laser pointer is turned on.
1
2
Do not stare directly at the laser pointer, as it can cause serious damage to eyes.
46
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 e. Place the daily test phantom with reference to picture A and B.
Place the phantom on a cross mark of laser point of equipment. At this time, have the cross mark indicated in phantom face inwards as shown in picture A, and parallel inclined plane of phantom and a side of equipment.
평행
FIG A.
FIG B
47
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 f. If you intend to implement the daily check in the message window of “Start or Cancel”, click the “Start”button to implement the daily check, and click the
“Cancel”button to cancel it. g. Implement the daily check. h. If the massage of “Daily Phantom” does not pop up, place the phantom in the exact position and measure it again. i. When the check is completed, go to the initial position.
Daily check is implemented automatically until the equipment check is completed. Be careful not to be injured due to carelessness during equipment operation as equipment may move.
48
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
5. Personal Information Registration of Subjects
Insert the personal information of subjects. Exact values shall be entered as measurement mode varies according to the personal data of patients.
- Add: Register the subjects.
- Search: Search a subject satisfying the condition in the subject list.
- Modify: Modify the personal data of the selected subject.
- Delete: Delete the all the data of the selected subject. a. Click the “Measurement” of user program. b. Register the new patient by clicking “Add”button at the bottom of subject list. c. Register the information when “Patient New” window pops up.
Patient ID: Enter the chart number or subject ID.
Name: Enter the name of the subject.
Birth data: Enter the date of the subject’s birth.
Gender: Enter the gender of the subject.
Ethnicity: Enter the ethnicity of the subject.
Height: Enter the height of the subject. (unit: cm)
Weight: Enter the weight of the subject. (unit: kg)
Menopause: When the subject is a female, select if she has experienced menopause. d. Click “OK” when having finished inserting.
49
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
6. Modification of Subject Personal Information
- Select a subject and click the “Modify” button to modify the personal information of the subject.
7. Deletion of Subject Data and Personal Information
- Select a subject and click the “Delete” button to delete the subject data and personal information.
Data deleted by users on their own cannot be restored.
50
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
8. Search of Subject Data
- You can search a desired subject using the date of his birth, name, and ID as occasion demands and patient list according to search conditions is created.
51
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
9. Measurement of Bone Mineral Density
a. After entering subject information for bone mineral density measurement, double click the subject in subject list or click the “Measurement” on the right with subject selected and then “Measurement window” appears. b. When scanning, spine, left femur, and right femur can be selected, and it is possible to scan by area and there is also one scan with which two or three regions can be scanned at a time.
What is One Scan?
Unlike the conventional method implemented in the process of equipment settingÎpositioningÎone region measurementÎanalysisÎ equipment setting for other site ÎpositioningÎ other region measurementÎ analysis Îcompletion,
One Scan has benefits of reducing patient waiting time and convenience of user equipment use with the steps of equipment settingÎpositionÎ position of the second areaÎfirst region measurementÎ
second region measurementÎanalysis Îcompletion when measuring more than two sites.
52
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 c. Predict the measurement points of subject referring to the picture.
Right Femur
Right Forearm
Spine
Left Femur
Left Forearm
Spine: Place the light of laser pointer on the navel and implement the measurement.
Femur: Lower it about 15cm ~ 20cm from the pelvis bones to leg direction and place the light of laser pointer in the middle of the thigh and scan femur.
Forearm: Place the laser pointer in the middle of wrist and scan forearm.
Forearm does not support One Scan.
Basic System
53
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 d. Place the subject on equipment and fix the measurement assistant according to scanning regions.
When scanning a point of spine
Outlines of Measurement Assistant
54
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
When scanning one point of femur or in case of One Scan
Fixation of feet when measuring a region of femur or in case of One Scan
55
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 e. When pressing the “Positioning” button in “measurement” window, it stands by in the initial position for 5 minutes.
The message said it will go to “Default position” appears. f. Check if subject has injury risk by equipment, and click the “Yes” button.
(“Yes” -> automatically go to Default position
“No” -> manual movement using op-control switch ) g. When having moved to Default position, laser pointer is turned “on”, and laser pointer LED of user operation switch is turned on. h. Go to the proper position where you can select spine, left femur, right femur of subject by pressing Top/Bottom/Left/Right keys and operating Arm of user operation switch. Press the “Select” switch of appropriate site. Once pressing the “Select” key, appropriate coordinates are stored in equipment. If you want
One Scan, go to another measurement points and set up the appropriate site once again. When coordinates are set up in each site, LED (green) is turned on.
If pressing Select switch of appropriate site one time, LED (green) is turned on and present position is stored, when pressing once more, LED (green) is turned off and the stored position information is deleted.
In case of One Scan, LED of selected SW is turned on while standing by the measurement, and it informs which site is being measured while flickering, when set up site is scanned.
56
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1
2
3
4
5
1. Scanning areas: It means the measurable areas with equipment.
2. Navel: When measuring spine, set laser pointer to a navel. Equipment automatically lowers about 4cm for measuring and scans it.
3. Right femur: Lower it about 15cm ~ 20cm from a right pelvis level and set laser pointer to the middle of femur to measure right femur.
4. Left femur: Lower it about 15cm ~ 20cm from a left pelvis level and set laser pointer to the middle of femur to measure left femur.
5. Measurement assistant :It bends inwards and fixes feet so that femur neck part may not overlap the surrounding textures and bones when measuring femur. i. Click the “Measure Start” button of user program and it starts scanning at the appointed position, after measurement points are set up.
57
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Picture A. Picture B. Picture C.
- Picture A: the state position is not set up
- Picture B: the state the positions of spine and left femur are set up.
- Picture C: spine is being scanned with two positions set up. j. Scanning k. When measurement completes, the screen automatically connects to analysis screen and Arm of equipment moves to the initial position
58
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Allow the subject to lower after equipment completely returns to the initial position for the safety of subject.
In case of One Scan, when finishing one site, it goes to the next site without messages. At this time, if the subject gets up while recognizing unconsciously the measurement has completed, he can get hurt by moving Arm.
59
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
10. Interruption of Bone Mineral Density Measurement
If there is a situation needs canceling measurement immediately during bone mineral density measurement, wrong measurements, or patients’ complaints of inconvenience, bone mineral density measurement shall be interrupted. In case of emergency, use the “emergency switch”, and if it is not emergency situation, normal termination is possible as follows. a. Click the “Stop” button while measuring. b. The following message appears.
1. Go to Next:
Store the data being measured and go to the next scanning site. .
2. All measurement Stop and File isn’t saved:
Stop the current measurement and do not save the data.
3. Cancel:
Continue to measure.
60
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
11. Re Scan
- Rescan is used to move scanning region into the region of interest without resetting during measurement when the subject’s bones that are being measured are not within the region of interest. a. Click the “ReScan” during measurement. b. Drag and click the middle of box within a red outline so that the regions to be measured may be exposed. c. The following display window appears.
1. Move and scan: Restart the scanning at the moved region.
2. Repositioning: Return to reset a region. d. Scan is restarted in the new region.
Picture A. Rescan during Measurement
61
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Picture B. Resetting of Regions
Picture C. Rescanning and Resetting of Regions
62
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
12. Description of Analysis Screen
- Subject’s data which was measured are analyzed by doctors.
63
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
- Measured data are divided into BMD value and T-score value in screen left according to the measurement date by region, images appear in the middle of screen when clicking data.
- Right top shows normal/osteopenia/osteoporosis by age-matched T-score distribution of analyzed subjects.
- Right bottom shows bone mineral density value, T-score, and Z-core by ROI of subjects.
Data deletion button
Subject data
T-score graph
64
Measurement value
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
- Icons next to an image indicate the following.
1. ROI SET
2. Brush
3. ROI Line Sort
4. Auto ROI
5. L1 on/off
6. L2 on/off
7. L3 on/off
8. L4 on/off
1. ROI set: It fixes or moves ROI.
Î
In case of fixation, fixed region value appears, when moving, the value does not appear.
2. Brush: It is used to erase the factors causing trouble to subject’s image or to include bone parts that were not captured within ROI.
3. ROI sort: It does not detect ROI automatically but implements basic sort.
4. Auto ROI: It is a function detecting ROI automatically by algorithm.
5. L1 on/off: It removes L1 region ROI from total results or adds when measuring spine. Spine.
6. L2 on/off: It removes L2 region ROI from total results or adds when measuring spine.
7. L3 on/off: It removes L3 region ROI from total results or adds when measuring spine.
8. L4 on/off: It removes L4 region ROI from total results or adds when measuring spine.
65
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
- Icons at the bottom of images are as follows.
1
2
1. It determines image colors. Black-and-white images or color images can be selected.
2. Comments about subject’s selection image can be inserted.
66
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
13. Image Analysis (SPINE)
- It analyzes the subject’s measured image of spine. a. Click the subject’s measured spine images in the list on the left of the images. b. Press the “ROI set” button when images appear. c. Press the “ROI set” button again when ROI is captured normally, as automatic
ROI is set up. d. When ROI has not been captured normally, capture the exact ROI by dragging
3 points in red line on images. ROI needs resetting for exact examination of subjects whose spine is bent or pressed.
Middle point: It is used to adjust top and bottom line.
Points on both sides :It is used to adjust the inclination of lines. e. When it is difficult to adjust ROI with bare eyes, use the histogram on the right of the images. Histogram indicates image density in graph from above, in case of high density, the right of histogram is marked high, and marked low for low density. f. Once clicking “ROI set” button, ROI is fixed and calculated value appears as measurement value.
67
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
14. Image Analysis (Femur)
- It analyzes the subject’s measured images of femur.
Refer to 13.image analysis (spine) as the analysis methods are basically the same as those. a. If ROI of femur has not been captured automatically, practice the following.
Press the “ROI set” button. b. Drag the red box to capture ROI of femur and place the pointer in the middle of the box in the “Neck”. c. Have the inclinations of Neck and rectangle box be perpendicular by adjusting inclinations with pointer outside the square.
68
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01 d. Place the pointer that can be dragged out of square on the same line with bottom-right of lessor-Trochanter, have the point placed on the position meeting perpendicular line from outside the greater-Trochanter.
Greater-Trochanter
Lesser-Trochanter e. When clicking “ROI set” button, ROI is fixed and a triangle region is created. f. The results are shown.
69
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
15. Use of Brush
- Brush is used to eliminate factors affecting the measurement values of subject images or include removed parts on one’s own into measurement values. It is possible to adjust the regions of bone, tissue, neutral, and artifact with appeared images. a. Click a subject’s images in the list. b. Click the ROI set button when an image appears. c. Click the “Brush”. d. Select the check box of function to be used among Bone, tissue, neutral, and artifact on the right menu. e. Select the Brush size. f. Click a certain area and paint it while rubbing with mouse clicked. g. Click the “Result” button after setting regions. h. Check the image if the painted regions are well reflected in the results. i. Click the “Reset” button if the painting fails. It returns to original image of patients. j. Click the “Cancel” button if you want to cancel the region setting during region setting.
When Brush function is applied to original images, BMD, T-score values are affected directly. Therefore, be sure to consider carefully when using Brush function.
70
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
- Bone Region Labeling
It is used when a bone to be calculated has not been captured in ROI regions.
- Tissue Region Labeling
It is used when setting tissues on both sides of bone.
71
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
- Neutral Region Labeling
It is used when setting the regions excluded from BMD calculation.
- Artifact Region Labeling
It is used when removing foreign substances on images.
72
\
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
16. Print of Results and PACS Linkage
- Print out the acquired results. a. Print out the information of spines and femurs by site through “Information
Check” in analysis window after analyzing. Also, when clicking once the data in the list of subject measurement data, the data of list is selected, and when clicking the “Print” button, the selected data is printed. b. It is possible to transmit subject’s data easily using PACS if PACS linkage system is installed in the hospital.
Select the data in “Information check” and then click the “Send PACS” button.
Trend
PACS linkage (Send PACS)
73
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1
2
3
4
5
6
1: hospital logo
2: address and telephone numbers of hospital
3: subject information
4: results of multi-site No.1
5: results of multi-site No.2
6: comments
74
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1
2
3
4
5
1: hospital name
2: hospital address
3: subject’s information
4: data information of No. 1 site measurement
5: data information of No. 2 site measurement
75
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
17. Print of Trend
- Trend shows the trend of subject’s osteoporosis as a graph and BMD values.
Values by ROI region of each site as well as total BMD values can be printed. a. Select the site of interest in the first window of “search condition”. b. Select the region of interest in the second window of “search condition” c. BMD values are indicated after sorted by date, graph also indicates the data of measurement date as a diagram.
1
2
3
4
5
6
1: Selection of Site
2: Selection of ROI Region
3: Print of BMD values of selected regions by date
4: Go to analysis window
5: trend graph
6: trend print
76
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1
2
1: subject information
2: trend graph
77
Basic System
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
DEXXUM T Related Documents
1.
Documents Related to Radiation of DEXXUM T ................................................................. 79
1) Aluminum Equivalent .................................................................................................................... 79
2) Radiation Amount of DEXXUM T Leaked Outside ........................................................... 80
3) Patient Dose ....................................................................................................................................... 81
2.
Label ....................................................................................................................................................... 84
3.
Detailed Specifications of DEXXUM T ...................................................................................... 89
4.
List of Installation Items of DEXXUM T ............................................................................... 91
5.
Warranty ............................................................................................................................................. 92
6.
Remarks ................................................................................................................................................ 93
Contents of Related Documents
78
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1. Documents Related to Radiation of DEXXUM T
1) Aluminum Equivalent
Aluminum Equivalent of DEXXUM T:
Samarium filter
X-ray tube inner filtration
Total Filtration Equivalence
2.0mmAL
0.8mmAL
2.8mmAL
79
Related Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2) Radiation Amount Leaked Outside
The amount of leaked radiation of DEXXUM T is as follows.
80
Related Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3) Patient Dose Amount Table(mR)
Measured
Site
Spine
Femur
Site
Spine
Femur
Scan speed
Fast
Normal
Fast
Normal
Scan speed
Fast
Normal
Fast
Normal
Scan
Mode
Thin
Measured
Time
85 sec tube current
0.4 mA
Standard 85 sec 1.5 mA
Thick 136 sec 1.5 mA
Thin 187 sec 0.4 mA
Standard 187 sec 1.5 mA
Thick 344 sec 1.5 mA
Thin 65 sec 0.4 mA
Standard 65 sec 1.5 mA
Thin 119 sec 0.4 mA
Standard 119 sec 1.5 mA
Scan
Thin
Measured
Time
85 sec tube current
0.4 mA
Standard 85 sec 1.5 mA
Thick 136 sec 1.5 mA
Thin 187 sec 0.4 mA
Standard 187 sec 1.5 mA
Thick 344 sec 1.5 mA
Thin 65 sec 0.4 mA
Standard 65 sec
Thin 119 sec
1.5 mA
0.4 mA
Standard 119 sec 1.5 mA
Patient Dose Amount in Normal Operation tube voltage
83 kV tube voltage
83 kV
Dose Ratio
(mR)
0.766
3.74
5.95
1.675
8.2
14.94
0.589
2.885
1.066
5.14
Dose Ratio
(uSv)
6.71016
32.7624
52.122
14.673
71.832
130.8744
5.15964
25.2726
9.33816
45.0264 mAs
34
127.5
204
74.8
280.5
516
26
97.5
47.6
178.5 mAs
34
127.5
204
74.8
280.5
516
26
97.5
47.6
178.5
81
Related Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
A. Patient Dose Amount Graph (83KV, 1.5mA Unit : mR)
B. Patient Dose Amount Graph (83KV, 1.5mA Unit : uSv)
82
Related Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
C. Patient Dose Amount Graph (83KV, 0.4mA Unit : mR)
D. Patient Dose Amount Graph (83KV, 0.4mA Unit : uSv)
83
Related Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
2. Label
- Label positions of equipment are as follows.
1)
84
2)
3)
4)
5)
6)
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
1) Warning label against pinched fingers.
2) Label for daily check phantom
3) Warning label against laser pointer
85
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
4) Equipment label
(Korean)
기재사항
(1) 제품명 : X선 골밀도 측정기
(2) 형명(모델명) : DEXXUM T
(3) 제조업자 : ㈜오스테오시스
서울시 구로구 구로동 212-13 벽산디지털벨리 3차 308,309,310
(4) 제조업허가번호 : 제 1125 호
(5) 사용목적 : 뼈또는 인접조직에 X선을 투과시켜 뼈의 밀도와 미네랄 함량을 측정하는 기구
(6) 제조번호 : 추후기재
(7) 제조년월일 : 추후기재
(8) 중량 :130.5Kg
(9) 포장단위: 1set
(10) 제품의 유효기간 (반영구적)
(11) 성능 및 사용방법 : 제품설명서 참고
(12) 사용상 주의사항 : 제품설명서 참고
(13) 정격전원의 상수(단상)
(14) 주파수 : 50/60Hz
(15) 표준전압 : 100-240Vac
(16) 소비전력 : 400VA
(17) X선관의 형식명 : DF-151SBR
(18) 초점의 크기 : 0.5 mm
(20) 고유여과 : Inherent Filtration 2.8mm Al
(21) 전격에 대한 보호형식 : 1급기기
(22) 전격에 대한 보호등급 : B형기기
86
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
(English)
X-ray Bone Densitometer
MODEL: DEXXUM T
S/N:
50/60Hz,
X-ray Tube Voltage : 70-99kV
X-ray Tube Current: 0.2-1.5mA
CLASS I Equipment
Filtration: Total filtration: 2.8mmAl please refer to the user manual
Tube: 0.8mmAl, Samarium: 2.0mmAl
Take adequate precaution to prevent the possibility of any persons
Carelessly exposing themselves or other to radiation
2008
5) Label for subject table
DEXXUM T Patient Table
H2AY-002A
2008
87
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
6) Label for X-ray generator
X-ray Generator
Model: EMB-0165, Manufacture: Ecotron Co., Ltd
Serial No.: , Mfg Date:
Line input Power: 100-240Vac, Single Phase: 50/60Hz
Max Output Power: 165W, Inherent Filtration: 2.0 mm Al
Rating: 70-99kV, 0.2-1.5mA, Continuous
X-ray TUBE HEAD
Model: DF-151SBR, Manufacture: TOSHIBA
Focal spot: Large: 1.5mm, Small: 0.5mm
Operating Tube Voltage: 40-110kV
Anode Heat Storage Capacity: 35.5 kJ
Target Angle: 16 degrees. Inherent Filtration: At least 0.8mm Al
88
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
3. Detailed Specifications of DEXXUM T
Contents
Manufacturer
Model name
Grade and type
Product size
Weight
Descriptions
Equipment
OsteoSys
DEXXUM T
1 st grade B type instrument
2083(W) x 1066(D) x 1230H)
130.5Kg
Input voltage and frequency 100VAC ~ 240VAC / 50Hz ~ 60Hz
Power consumption 400VA
Valid period Semi-permanent
Manufacturer
Address
X-ray generator
ECOTRON
11F, 1002 Hansin IT Tower 2 nd , 60-18,
Gasan-dong, Geumcheon-gu, Seoul,
KOREA
EMB-0165 Model name
Maximum tube voltage
Maximum tube current
Power consumption
Capacity
110KV
1.5mA
50W
165W
Input voltage and frequency 110VAC/220VAC ± 10%
Lowest Current Time product 1mAs (0.2mA, 5seconds)
Aluminum equivalent 2.8mmAL (samarium + X-ray tube)
X-ray tube
Manufacturer
Model name
Maximum tube voltage
Maximum tube current
Focal spot size
Cathode
Anode
Anode input
TOSHIBA
DF-151SBR
121KV
3.1A
0.5mm
Tungsten filament
14 fixed-type tungsten target
248VDC / 1.6A
Remarks
89
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Manufacturer
Model name
Image sensor
Detector
EV products
EV-381
CdZnTe
Image area 11mm diameter
Remarks
Manufacturer
Model name
Size
Image size (spine)
Image size (Femur)
Reproducibility
Accuracy
X-ray radiation time
Radiation dose
Descriptions
Daily Check Phantom
OsteoSys
D3MC-500A
170(W) x 50(D) x 150(H)
Performances
160mm X 192mm
120mm X 180mm
±
±
1.5% CV
1.5% CV
1min 25sec – 3min 50sec below 10m Rem per examination
(5m distance)
Below 1 mRem (5m distance) Residual radiation
Operating temperature
Suitable humidity
Suitable air pressure
Impact
Storage temperature range
Storage humidity range
Use environment above 15℃ below 40℃
20% ~ 80%
700 ~ 1060 Hpa
Within 2G in 6 X 10
-3 sec
Above -5℃ below 50℃
0% ~ 90%
90
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
4. Installation Items List of DEXXUM T
No. Product
1 Main frame of
Equipment
2 Computer
Qty’ Model e DEXXUM T
1
Vendor
Osteosys
Samsung
CHECK
3 Monitor 1
1
1
Minimum requirements:
P4 Pentium processor
512MB
Win XP
1GB hard drive capacity
(excluding DB)
17” LCD Monitor
(above 1280 X 1024 )
Color Printer
D3MC-500A
Samsung
Samsung
Osteosys
4 Printer
5 Daily check phantom
6 Protective wall
7 Power cable
8 Communication cable
9 A4 paper
10 Measurement assistant
11 Test report
12 User manual
13 Installation confirmation
1
1
1
1
1
1
1
1
-
-
-
G2AY-003A
220VAC 10A
5mm cross cable
A4
G3MJ-002A
Osteosys
Osteosys
-
-
Osteosys
Osteosys
Osteosys
Osteosys
91
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Warranty
Thank you very much for using OsteoSys’ bone densitometer.
We notify that OsteoSys’ equipment has been developed and manufactured conforming to the law related to medical equipment.
OsteoSys will provide a repair of the product at OsteoSys’ own expense for any such failure and defects arising within a year of the original date of purchase or installation, provided that such a defect which result from normal transportation and use and cannot be otherwise repaired. z
Repairs performed by OsteoSys are subject to charge to the purchaser at established service fee, only in the following cases, and at normal repair rates when received service after the warranty period.
1. When product failures occur after the warranty period
2. Failures caused by natural calamities such as fire, earthquake, a stroke of lightning, and flood
3. When product failures occur due to inadequate movement after installation and careless use
4. When product failures occur due to services by third parties other than OsteoSys, appointed service agencies, or OsteoSys’ designated engineers
5. When product failures occur due to disassembly and assembly of products on your own z
If you require assistance, please fill out the following form and send a fax to
+82-2-2025-2299 or contact the nearby agencies, stores, or head office by calling the service line of +82-2-2025-1650. Please fill out the following and send it for prompt after sales service or tell us abnormal symptoms of equipment.
Model name / serial No. : DEXXUM T /
Name of Installed hospital / City / Country :
Contact of Installed hospital :
Install date :
Abnormal symptoms :
OsteoSys Co., Ltd.
92
Document
User Manual (Rev 1.01) – English Version
OsteoSys Co., Ltd.
Model Designation: DEXXUM T DOC. No. : OT08-2F7430-MUL-01
Remarks:
93