Dinel CLM-36 N, Xi, NT, XiT Capacitive level meters Service manual
Below you will find brief information for Capacitive level meters CLM-36 CLM-36N, Capacitive level meters CLM-36 CLM-36Xi, Capacitive level meters CLM-36 CLM-36NT, Capacitive level meters CLM-36 CLM-36XiT. The CLM-36 level meters are designed for continuous level measurement of liquids, powders and bulk-solid materials in vessels, tanks, sumps, containers, silos, etc. The level meters are available in different variants for non-explosive and explosive areas. It is important to note that the level meters are classified as intrinsically safe for hazardous areas.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
SERVICE MANUAL
Capacitive level meters
CLM-36
Before level meter´s usage please read instructions in this service manual and keep it for further use. Service manual is for all types of level meters series CLM-36. Producer reserves the right to make changes without any previous announcement.
Content
Content .................................................................................................................................2
I.
II.
Description .................................................................................................................3
Technical specification ........................................................................................... 4-7
III. Types and performance of electrode..........................................................................8
IV. Linearity......................................................................................................................9
V. Mounting and dismounting................................................................................. 10-12
VI. Instalation of level meter and placing in operation...................................................13
VII. Preparing of level meter for measuring....................................................................14
VIII. Setting of level meter with unknown value of permittivity of measured medium......15
IX. Setting of level meter with known value of permittivity of measured medium.... 16-18
X. Maintainance and service of level meter..................................................................19
XI. Protection, safety, compatibility and explosion proof ...............................................20
XII. Marking of level meters............................................................................................21
XIII. Marking of labels................................................................................................ 22-23
XIV. Accessories..............................................................................................................24
XV. Table of relative permittivity (dielectric constants) of selected materials............ 25-26
XIV. Notes........................................................................................................................27
Contacts ..............................................................................................................................28
- 2 - 3 -
- 2 -
I. Description
Capacitive level meters CLM-36 are designed for continuous level measurement of liquids, powders and bulk-solid materials in vessels, tanks, sumps, containers, silos, etc.
Variants according working areas:
- for non-explosive area:
- for explosive area:
CLM-36N_-_ _-_
CLM-36Xi_-_ _-_
Available is high temperature performance for both working areas:
- high temperature performance for non-explosive area:
- high temperature performance for explosive area:
CLM-36NT-_ _-_
CLM-36XiT-_ _-_
Classification of areas:
Performance of level meter
CLM-36N
Part of the level meter with electrode
Part of the level meter with housing
non-explosive
CLM-36NT
CML-36Xi
CLM-36XiT non-explosive
II1GDT83°CEEXiaIIBT5 with ISIR
1) zone 0, zone 20
II1/2GDT83°CEEXiaIIBT5 with ISIR
1) zone 0, zone 20 zone 1, zone 21
The electrode and its surroundings (the wall of vasel, the reference tube, etc.) make capacitor. The electrode is designed according measured material:
- variant with rod electrode:
- variant with rod electrode and reference tube:
- variant with rope electrode:
Available is mechanical connection:
CLM-36_ _-10, 12
CLM-36_ _-20, 22
CLM-36_ _-30, 31, 32
- with thread:
- without thread:
CLM-36_ _-_ _-M
CLM-36_ _-_ _-G1
CLM-36_ _-_ _-Cl
M36x2
G1"
Triclamp
Inside of the level meter´s housing is placed removable electronic module with adjustment elements (see chapter VII).
1) ISIR - Intrinsically safe isolating repeater (example IRU-420)
- 3 -
II. Technical specification
Type of level meter
Supply voltage: - var. N, NT
-
var. Xi,XiT
Current output: - non-activated electrode
- full-activated electrode
Max. internal values of Xi version
Sensitivity ranges
Maximum capacity of flooded electrode (range)
Initial capacity regulation ratio
Nonlinearity
Temperature error
Voltage error
Internal resistance / electric strength
(electrode - housing)
Coupling capacity / electric strength
(housing - supply leads)
Allowed temperature range in zone 0 (EN 50284)
Allowed pressure range in zone 0 (EN 50284)
Protection class: - housing
- connector type GDM 2009
- connecor type GDM-K 2000
Connector socket type Hirschmann DIN 43 650
Recommended cable
Max. load (serial) resistance (U = 24 V)
Weight of the housing (without electrode)
CLM-36__-__-_
9 ÷ 36 V DC
9 ÷ 30 V DC
4 mA
20 mA
Ui=30VDC, Ii=132mA, Pi=0,99W,
Ci=370nF, Li=0,9mH
20, 30, 50, 100, 150,
300, 500, 1000 pF
70, 250, 600, 1200, 3000,
7000, 18000, 36000 pF min. 1:2 max. 1 % max. 0,05% / K max. 0,3 μA / V
1 MΩ / 250 V AC var. N 51 nF / 250 V AC var. Xi 26 nF / 500 V AC
-20 to +60°C
0,8 to 1,1 bar (0,08 to 0,11 MPa)
IP 67
IP 65
IP 67
GDM 2009 (standard)
GDM-K 2000 (optional)
2 x 0,75 mm
2
R max
=750 Ω var. N, Xi aprox. 0,5 kg var. NT, XiT aprox. 1 kg
- 4 - 5 -
- 4 -
II. Technical specification
Temperature resistance of CLM-36:
Variant of performance
CLM-36_ _-10, 20-_
CLM-36_ _-12, 22-_
CLM-36_ _-30, 31, 32-_
Temperature of medium
(at electrode)
-40 to +200°C
-40 to +120°C
-40 to +105°C
Ambient temperature range t
a
(at housing)
variant N
-40 to +85°C
variant Xi
-40 to +75°C
-40 to +85°C
-40 to +85°C
-40 to +75°C
-40 to +75°C
Pressure resistance of performance CLM-36N(Xi):
Variant of performance
CLM-36_-10, 20-_
CLM-36_-12, 22-_
CLM-36_-30, 31, 32-_
for t
a
Maximum operating overpressure
= -40 to +20°C for t
a
= -40 to +85°C
3 MPa 1 MPa
3 MPa
1 MPa
1 MPa
0,5 MPa
Pressure resistance of high temperature performance CLM-36NT(XiT):
Temperature in connection point
Maximum overpressure
+ 100°C +120°C +150°C +180°C
3,0 MPa 2,0 MPa 1,5 MPa 0,5 MPa
Used materials:
Part of the level meter (type)
housing (all) insulation bushing (all) electrode (10, 12, 20, 22) electrode (30, 31, 32) electrode coating (12, 22, 32) electrode coating (30, 31) weight insulation (32) weight / anchor mechanism (30, 31, 32) reference tube (20, 22)
standard
Material
st. steel W. Nr. 1.4301
optional
st. steel W. Nr. 1.4571,
Incoloy 825 (W. Nr.
2.4858)
PTFE st. steel W. Nr. 1.4301
zinc coated steel rope st. steel W. Nr. 1.4571
-
FEP polyolefin (modified PE)
PTFE st. steel W. Nr. 1.4301
st. steel W. Nr. 1.4306
-
PTFE
-
st. steel W. Nr. 1.4571
- 5 -
II. Technical specification
Dimensions of each variants: with rod electrode:
CLM-36_ _-10-_
CLM-36_ _-12-_ with rod electrode and reference tube:
CLM-36_ _-20-_
CLM-36_ _-22-_ with rope electrode:
CLM-36_ _-30-_
var. M, G - housing with
thread var. Cl - housing with
Triclamp coupling
Z* - var. M - thread M-36x2
- var. G - thread G1“
- 6 - 7 -
- 6 -
II. Technical specification
Dimensions of each variants: with rope electrode:
CLM-36_ _-32-_ with rope electrode with anchor:
CLM-36_ _-31-_
connector
high temperature performance:
CLM-36_T-_ _-_
steel anchor bushing or dusttight bushing wall of container pushing spring housing and type of electrode according specific version
- 7 -
III. Types and performance of electrodes
Characteristic of variants:
CLM-36_ _-10-_
With uncoated rod electrode - for level measurement of non-conductive liquids (oils, diesel, benzine) and bulk-solid materials (flour, sand, cement, plastic granulates, etc.). Max. electrode length 5 m.
CLM-36_ _-12-_
With coated rod electrode (FEP isolating) - for level measurement of water and other conductive liquids including waste liquids in metal vessels, concrete sumps, reservoirs, etc.
Resistant against aggressive medium, lower adhesion to some mediums. Max. electrode length 3 m.
CLM-36_ _-20-_
With uncoated rod electrode and reference tube - for level measurement of clean nonconductive liquids (oils, diesel, benzine). By means of reference tube the output signal does not depend on the dimension and shape of the vessel. Impossible to use for waste and high viscosity liquids and bulk-solid materials. Max. electrode length 3 m.
CLM-36_ _-22-_
With coated rod electrode (FEP isolating) and reference tube - for level measurement of clean conductive liquids. Main use is for measurement in plastic and glass vessels and for fine measuring. Not suitable for waste water and high viscosity liquids and bulk-solid materials. Max. electrode length 3 m.
CLM-36_ _-30-_
With coated rope electrode (polyolefine isolating) and uncoated weight - for level measurement of bulk-solid materials (grains, sand, flour, cement, etc.). Max. electrode length 12 m.
CLM-36_ _-31-_
With coated rope electrode with anchor (polyolefine isolation) and uncoated weight - as the type 30. For use for higher silos. Max. electrode length 20 m.
CLM-36_ _-32-_
With coated rope electrode (FEP isolation) and coated weight (isolation PTFE) - for level measurement of conductive and non-conductive liquids. Max. electrode length 20 m.
- 8 - 9 -
- 8 -
IV. Linearity
Influence of the tank shape on a linearity of measured capacity of nonconductive liquids:
1. In the tank with straight wall (for example vertical cylinder) and with the sensor placed
parallely with the wall capacity change is linear.
2. In a curved tanks (most frequently horizontal cylinder) capacity change during
measuring of electrically non-conductive material is non-linear. Linearity is done
by reference tube (CLM-36_ _-20, 22-_).
- 9 -
V. Mounting and dismounting
Level meters with coated electrode have protection cover at the end of electrode, which is necessary take down befere mounting.
Level meters mount in vertical position to upper lid of tank or reservoir by steel welding flange ON-36x2 (steel 11375), stainless steel welding flange NN-36x2 (stainless steel
1.4301), fixing nut UM-36x2 (stainless steel 1.4301) or Clamp flange.
For mounting level meter to the metal tank or reservoir is not necessary to ground housing again.
In case of installation in concrete sumps or silos it is properly to install level meter at metalic auxiliary construction (console, lid, etc.) and then connect with metal all the time submerged object, eventually with steel armouring.
For measuring of material in plastic and glass vessels by level meter without reference tube is necessary to connect grounding screw at housing with auxiliary electrode which is fixed by the appropriate method to the outer casing of vessels (or at inner wall). Material of auxiliary electrode is necessary to choose with consideration to the working environment and properities of measured material.
Orientation of proportion for mounting of level meter (including eventualoly auxiliary electrode) shows following pictures.
CLM-36_ _-10, 12-_
installation into the metal and non-metal vessels
auxiliary electrode
width = min. 30 mm
(necessary only for non-metal vessels)
E - the length of electrode [mm] - the end of the electrode has to be dipped min. 20 mm below the lowest measured
level
h - the distance from the bottom - min. 50 mm
a - the distance from the wall - min. approx. E/20
d - the diameter of auxiliary tube vessel - min. 40 + E/20 (smaller dimension has to be discussed with producer)
- 10 - 11 -
- 10 -
V. Mounting and dismounting
CLM-36_ _-20, 22-_
installation with reference tube
h - the distance from the bottom - min. 50 mm with respect on possibility of presence of heavy fraction (water) in oil
products
k - the distance from the wall - optional
CLM-36_ _-30, 32-_
installation into the metal tank and concrete containers connection to the armouring
E - the length of electrode [mm] - the end of the electrode has to be dipped min. 20 mm below the lowest measured
level
h - the distance from the bottom - min. 100 mm
a - the distance from the wall - approx. the same as the distance from the inlet
- 11 -
V. Mounting and dismounting
CLM-36_ _-31-_
installation of rope electrode with anchor
steel anchor welding cylinder
(mat. 11375) or
dust-tight bushing
to be welded into the conical bottom wall
E - the length of electrode [mm]
c - the distance from the centre line of the silo - has to be as minimal as possible
t - the length of the conduction rod - approx. 500 mm
p - the length of the press spring - approx. 200 mm
- 12 - 13 -
- 12 -
VI. Instalation of level meter and placing in operation
Mounting, installation, placing in operation, maintainance, using and service of level meter CLM-36 must be done in conformity with this service manual and has to be performed according the norms for installation of electrical devices.
Electric connection of feeder cable is realized in state without voltage through the use of thread terminal in dismountable connector GDM at level meter’s housing.
Positive pole of supply (+U) will be connected at terminal (1), negative pole of supply
(0V) will be connected at terminal (2) - see following pictures.
Connection scheme of level meter
The inside scheme of the GDM connector socket
+U shielding
0V
Cable from level meter to the further unit is by two-wire cable with outside diameter of
6 ÷ 8 mm (recommended cross section of cores 0,5 ÷ 0,75 mm
2
).
A shielded cable must be only used if the lead is longer than 30 m or in the event that a power line must be used simultaneously with the lead. Connect shielding to the socket , shielding do not connect to the power panel.
Power supply for variants CLM-36N and CLM-36NT can be by stabilized safety voltages power supply 9 ÷ 36 V DC (SELV), example PSU-2400 or DSU-2400.
Or following evaluation device, for example programmable display unit PDU-4xx. For variants CLM-36Xi and CLM-36XiT must be connected with intrinsically safe isolating repeater 9 ÷ 30 V DC, for example IRU-420.
Activation is by turning on of the power supply (for example PSU-2400 či DSU-2400 for variants CLM-36N and CLM-36NT) or intrinsically safe isolating repeater (for example IRU-
420 for variants CLM-36Xi and CLM-36XiT).
- 13 -
VII. Preparing of level meter for measuring
Preparing of level meter
1. For access to the adjustment elements disconnect a connector and unscrew a nut
(attention for inside connecting wires). Connector connect again.
2. Level meter connect to the supply unit through miliampermeter (controler, etc.).
3. Trimmer
2)
20 mA bring into the basic position (this position is set by producer):
a) Turn out trimmer totally rightwards (in clockwise).
b) Return back for 3 turns leftwards.
Adjustment elements of level meter
trimmer 4 mA
(for compensation of initial capacity)
blue wire
(-)
green-yellow wire
( )
brown wire
(+)
trimmer 20 mA
(for sensitive span adjustment)
brass tie
(for pull out of elecronic module)
(the top view on the internal electronic module)
DIP switches
(for range select)
2) The trimmers are without backstops - approx. 15 turns.
3) Sensitivity - minimal capacity change of electrode to reach a current range 4 ÷ 20 mA.
4) Range - maximum capacity of housing and activated electrode to reach current range 4 ÷ 20 mA.
5) In case of short of electrode on housing or setting of a very sensitive range can caused to current restriction of level meter on value max. 30 mA.
- 14 - 15 -
- 14 -
VIII. Setting with unknown value of permittivity
Setting of level meter with unknown value of permittivity of measured medium
1. Empty tank to the minimum level.
2. By DIP switches on level meter set range
4)
No. 2 (250 pF).
3. Turn by screwdriver in trimmer 4 mA and set quiescent current of the level meter to
4 mA. Turning rightwards (in clockwise) increase the current, turning leftwards
decrease the current. When is impossible to set required current to 4 mA, then switch
DIP switches to the nearest higher range and set up of current 4 mA perform at this
range.
4. Fill the tank to the maximum level. When it is impossible fill the tank to the maximum
state then we use some known (measurable) level and current for next setting
calculate according to formula:
I out = 4 + (0,16 x level in %) [mA]
5. By DIP switches stepwise switch to higher ranges till the current is greather then
20 mA (or calculated value Iout), thru level meter then flows current e.g. 21 mA
Trimmer 20 mA is still in basic position.
5)
.
6. Then set by trimmer 20 mA current of level meter to 20 mA (or to calculated value I out
- see bellow).
7. To reach a maximum accuracy is useful to check over the setting of 4 mA
(compensation of initial capacity - quiescent current).
Table of ranges:
No.
1
6)
2
3
4
5
6
7
8
Sensitivity
(S)
3)
20 pF
30 pF
50 pF
100 pF
150 pF
300 pF
500 pF
1 000 pF
Range
4)
(R)
70 pF
250 pF
600 pF
1 200 pF
3 000 pF
7 000 pF
18 000 pF
36 000 pF
Position of
DIP switch
6) Range No. 1 (70 pF) is very sensitive and we recommend to use it only at level meters with short electrode
(approx. to 500 mm) and measured media with low permittivity.
- 15 -
IX. Setting with known value of permittivity
Setting of level meter with known value of permittivity of measured medium
1. Calculation of measuring range:
a) Count sum of capacity of housing (CH) and capacity of non-activated
electrode (CNE).
C1 = CH + CNE
Capacity of housing
(typical values)
Capacity of housing - normal performance
CH
21 pF
Capacity of housing - high temperature performance 28 pF
Capacity of non-activated electrodes
(electrode length 1000 mm)
all types without reference tube in distance 250 mm from conducting wall uncoated rod electrode with reference tube (type 20) coated rod electrode with reference tube (type 22)
CNE
(air)
12 ÷ 15 pF approx. 50 pF approx. 55 pF
b) Count sum of capacity of housing (CH) and capacity of activated electrode
(CAE) - selected mediums see the table.
C2 = CH + CAE
Capacity of activated electrodes
(electrode length 1000 mm)
all types without reference tube in distance 250 mm from conducting wall uncoated rod electrode with reference tube (type 20) coated rod electrode with reference tube
(type 22)
CAE
(benzine)
25 ÷ 30 pF approx. 100 pF approx. 110 pF
CAE
(water)
500 pF
(coated electorde)
-
500 pF
- 16 - 17 -
- 16 -
IX. Setting with known value of permittivity
c) Set the possible ranges -
capacity of housing and activated electrode
(C2) will not be greather than range value (R).
C2 < R
d) Sensitivity check - difference of capacity activated (CAE) and non-activated
(CNE) electrode must be greather than sensitivity of calculated range.
C2 - C1 > S (CAE - CNE > S)
e) Choose the range - both previous conditions must be used together c) and d).
2. Empty the tank to the minimum level.
3. By DIP switches on level meter set up pre-counted range.
4. Turn by screwdriver in trimmer 4 mA and set quiscent current of the level meter to
4 mA. Turning rightwards (in clockwise) increase the current, turning leftwards
decrease the current. When is impossible to set required current to 4 mA, then switch
DIP switches to the nearest higher range and set up of current 4 mA perform at this
range.
5. Fill the tank to the maximum level. When it is impossible fill the tank to the maximum
state then we use some known (measurable) level and current for next setting
calculate according to formula:
I out = 4 + (0,16 x level in %) [mA]
6. On preliminary range of level meter set by trimmer 20 mA current of level meter to
20 mA (or to calculated value I out - see bellow). Turning rightwards (in clockwise)
increase the current, turning leftwards decrease the current.
7. To reach a maximum accuracy is useful to check over the seting of 4 mA (compensation
of initial capacity - quiscent current).
Example of calculation:
Level meter CLM-36N-20-M E1000, measured medium is benzine (ε r
= 2,2), distance from wall to level meter with uncoated rod electrode with reference tube is optional.
1.a) Sum of capacity of housing and non-activated electrode:
C1 = CH + CNE = 21 + 51 =
72 pF
1.b) Sum of capacity of housing and activated electrode:
C2 = CH + CAE = 21 + 100 =
121 pF
- 17 -
IX. Setting with known value of permittivity
1.c) Possible ranges:
C2 < R
121 pF < (2) 250 pF, (3) 600 pF, (4) 1200 pF, (5) 3000 pF, (6) 7000 pF,
(7) 18000 pF, (8) 36000 pF
1.d) Check the sensitivity:
C2 - C1 > S
C2 - C1 = 121 - 72 = 49 pF
1.e) Recommended accurate range:
range (2) 250 pF
49 pF > (1) 20 pF,
(2) 30 pF
Level meter CLM-36N-20 with electrode E = 1000 mm by flowage of benzine has capacity
120 ÷ 135 pF - typical measuring range is range (2) 250 pF.
Typical measuring ranges of CLM-36 with electrode 1 m, 2 m and 5 m length for selected medias:
Medium Sand, grains
Type 1 m 5 m
10 2 3
12
30
31
Medium
Type
10
12
20
22
32
2
2
2
1 m
Water
2 m
-
3
-
4
-
3
3
4
4
3
3
3
2
2
2
Cement
1 m 5 m
2 4
4
4
4
1
1
1
1 m
Flour
5 m
1 3
3
3
3
Benzine, petroleum, diesel, oils
1 m 2 m
1 2
1
2
2
3
2
1
3
2
Note:
- 18 -
Types 10, 12, 30, 31, 32 in distance 250 mm from conducting wall.
Showned ranges are orientation. Exact range for existing configuration electrode / tank must be set directly in application on mearured medium.
- 19 -
- 18 -
X. Maintainance and service of level meter
Maintainance of level meter
Maintainance of device is to control the housing of level meter and lead-in calbe. In case of any visible damages please contact producer or supplying company.
At level meter is prohibited to perform any changes or repair without producer confirmation.
Service of level meter
Eventual reparation of mechanical damage of level meter must be done by producer or by producer authorized service organization.
Replacement of defective electronic
In case of detected electronic failure it is possible to replace the electronic module directly in process place of level meter by following procedure:
1. Disconnect connector and unscrew a nut (attention for inside connecting wires).
2. Defective electronic module pick up by pliers onto brass tie and remove by tensile
from housing of level meter.
3. Check placing of white seal O-ring in housing and on pressure ring.
4. Repaired or new electronic module return back into housing of level meter and press
down by clamping ring (attention that contact pin will be a little open before insertion
and contact springs were not pressed under level of electronic casing).
5. Check placing of black seal O-ring on connector.
6. Screw a nut back (attention for inside connecting wires) and connect the connector.
trimmers brass tie contact springs electronic module cover contact pin
- 19 -
XI. Protection, safety, compatibility and explosion proof
Protection
Level meter is equipped with protection against reverse polarity of supply voltage and against overvoltage.
Safety
Safety against dangerous contact is ensure by safe power supply according norm EN 33
2000-4-41 (SELV).
Compatibility
Electromagnetic compatibility is provided by conformity with norms EN 55022/B, EN
61326-1, EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6.
Explosion proof
Explosion proof of CLM-36Xi and CLM-36XiT is examined by FTZU-AO 210 Ostrava
- Radvanice, certificate No. 02 ATEX 0235X.
- 20 - 21 -
- 20 -
XII. Marking of level meters
Method of marking of level meters CLM-36
CLM-36 E
Basic thread dimension
Performance:
N
- normal (for non-explosive areas)
Xi - explosion proof - intrinsically safe for hazardous
(explosive) areas (clasification zone 0)
NT, XiT - high temperature
Electrode length in mm
Process connection:
M - thread M36x2
G - thread G1"
Cl - Triclamp
Electrode performance:
10 - rod uncoated electrode (max. 5 m)
12 - rod coated (FEP isolation) electrode (max. 3 m)
20 - rod uncoated electrode with reference tube (max. 3 m)
22 - rod coated (FEP isolation) electrode with reference tube (max. 3 m)
30 - rope electrode with coated (polyolefine isolation) rope and uncoated weight (max. 12 m)
31 - rope electrode with anchor with coated rope (polyolefine isolation) and uncoated weight
(max. 20 m)
32 - rope electrode with coated rope (FEP isolation) and coated weight (isolation PTFE) (max.
20 m)
Examples of correct marking
CLM-36N-10-G E1100
CLM-36XiT-20-M E900
CLM-36N-12-Cl E2000
CLM-36NT-12-M E1500
CLM-36Xi-30-M E12000
CLM-36N-32-G E6000
- 21 -
XIII. Marking of labels
On label of level meter type CLM-36N(T) are following datas:
- symbol of producer: logo Dinel
®
- contact: Dinel, s.r.o., Zlin, Czech Republic, www.dinel.cz, [email protected]
- connection scheme and labelling of wires: +U, 0V, GND
- type of level meter: CLM-36-N_-__-_, include length of electrode: E in mm
- serial number: Ser. No.: xxxxx - production year (first two position) and sequence
factory number (next three position)
- power supply voltage: U i
= 9 ÷ 36 V DC
- output current range: I = 4 ÷ 20 mA
- ambient temperature range: t a
= -40 ... +85 °C
- protection degree: IP 65 / IP 67
- label of shody: CE
- symbol of electro-equipment for back collection and separate collection of electro-
waste
On label of level meter type CLM-36Xi are following datas:
- symbol of producer: logo Dinel
®
- contact: Dinel, s.r.o., Zlin, Czech Republic, www.dinel.cz, [email protected]
- connection scheme and labelling of wires: +U, 0V, GND
- type of level meter: CLM-36-Xi-__-_, include length of electrode: E in mm
- serial number: Ser. No.: xxxxx - production year (first two position) and sequence
factory number (next three position)
- label of non-explosive device: Ex in hexagon
- performance: II1GDT83°CEExiaIIBT5
- number of authorized person examining control of system quality: 1026
- number of certificate of intrinsically safety: FTZU 02 ATEX 0235X
- limiting operating parameters: U i
= 30 V DC, I i
= 132 mA, P i
= 0,99 W, C i
= 370 nF,
L i
= 0,9 mH
- ambient temperature range: t a
= -40 ... +75 °C, t a
= -20 ... +60 °C (zone 0)
- protection degree: IP 65 / IP 67
- label of shody: CE
- symbol of electro-equipment for back collection and separate collection of electro-
waste
- 22 - 23 -
- 22 -
XIII. Marking of labels
On label of level meter type CLM-36XiT are following datas:
- symbol of producer: logo Dinel
®
- contact: Dinel, s.r.o., Zlin, Czech Republic, www.dinel.cz, [email protected]
- connection scheme and labelling of wires: +U, 0V, GND
- type of level meter: CLM-36-XiT-__-_, include length of electrode: E in mm
- serial number: Ser. No.: xxxxx - production year (first two position) and sequence
factory number (next three position)
- label of non-explosive device: Ex in hexagon
- performance: II1/2GDT83°CEExiaIIBT5
- number of authorized person examining control of system quality: 1026
- number of certificate of intrinsically safety: FTZU 02 ATEX 0235X
- limiting operating parameters: U i
= 30 V DC, I i
= 132 mA, P i
= 0,99 W, C i
= 370 nF,
L i
= 0,9 mH
- ambient temperature range: t a
= -40 ... +75 °C, t a
= -20 ... +60 °C (zone 0)
- protection degree: IP 65 / IP 67
- label of shody: CE
- symbol of electro-equipment for back collection and separate collection of electro-
waste
- 23 -
XIV. Accessories
Standard accessories:
For each level meter:
- 1 pc of connector socket GDM 2009 (IP 65)
- 1 pc of seal (material AFM)
For each delivery (each 5 pcs):
- 1 pc of screwdriver for adjustment
Optional accessories:
For mounting of level meter:
- stainless fixing nut UM-36x2 (material stainless steel 1.4301)
- steel welding flange ON-36x2 (material steel 11375)
- steel welding flange with lock and Al seal ON-36x2-Z (material steel 11375)
- stainless welding flange NN-36x2 (material stainless steel 1.4301)
Other:
- connector socket GDM-K 2000 with 5 m of calbe (IP 67)
- seal (on request - material PTFE, Al, etc.)
- 24 - 25 -
- 24 -
XV. Table of relative permittivity
Table of relative permittivity (dielectric constants) of selected materials
Material
Acetone
Acetylentetrachloride
Aminoplasts
Ammonia
Aniline
Ash
Benzaldehyde
Benzene
Benzol
Celluloide
Cement
Dioxane
Ebonit
Ethanol
Ethylacetate
Ethyleneglycol
Ethylenechloride
Flint crystal
Flour
Food salt (NaCl)
Freon R22
Fused quartz
Gasoline (benzin)
Glass
Glass organic
Glass silicate
Glazed carton
Glycerine
Grains
Hardened textile
Hexane
ε
r
3,7
2,0 ÷ 2,2
3,7 ÷ 10
3 ÷ 3,6
16
3,5
47
3,0 ÷ 5,0
2 ÷ 6
1,9
24
6,0
38,7
10,5
4,5
2,5 ÷ 3,0
6,0
6,1
19,5 ÷ 21,4
8,1
5 ÷ 8
15 ÷ 24
6,9
1,5 ÷ 1,7
18,3
2,28 ÷ 2,3
2,3
3,5 ÷ 6,2
4,0
2,235
2,5 ÷ 5
Material
Chloroforme
Ice
Liquefied air
Liquefied CO2
Liquefied chlore
Liquefied propan
Marble
Methanol
Mica
Mikanite
Milk powder
Monochlorbenzene
Nitrobenzene
Ortonitrotoluene
Paper
Paraffin
Paraffin oil
Paraffinic oil
Petroleum
Polyacetal
Polyamide - PA
Polydichlorstyrole
Polyetheretherketone -
Polyetherimide - PEI
Polyethylene - PE
Polyfenylenesulfide - PPS
Polymethylmetakrylate
Polypropylene - PP
Polystyrole
Polysulphone - PSU
Polytetrafluorethylene
ε
r
3,0
2,16
3,3
2,56
2,0 ÷ 2,2
2,39
3,0
2,0 ÷ 2,1
1,9 ÷ 2,5
1,9 ÷ 2,5
2,2
2,0 ÷ 2,2
3,6 ÷ 3,7
4,0 ÷ 5,0
2,7
3,2
4,81
3,1
1,5
1,6
2,0
1,6 ÷ 1,9
9,3 ÷ 11,6
32,7 ÷ 33,8
5 ÷ 8
4,5 ÷ 6
3,5 ÷ 4
4,0
35,7
27,4
1,6 ÷ 2,6
- 25 -
XV. Table of relative permittivity
Table of relative permittivity (dielectric constants) of selected materials
Material
Polyvinylacetate
Polyvinylchloride - PVC
Polyvinylidenefluoride
Porcelain
Pyridine
Rape-seed (colza) oil
Resin - acrylic
Resin - epoxy
Resin - melamine
Resin - phenolic
Resin - polyester
Resin - styrene
Resin - ureal
Rubber
Sand
Silicon caoutchouc
Succinite
Sugar
Sulphur
Toluene
Trichlorethylene
Trolitule
Vaseline
Water
Water solution
White beeswax
Wood - damp
Wood - dry
ε
r
2,8 ÷ 3,3
2,9
3,0
3,4 ÷ 3,6
2,3 ÷ 2,38
3,43
2 ÷ 2,6
2,2 ÷ 2,9
81
50 ÷ 80
2,0 ÷ 2,9
10 ÷ 30
2 ÷ 6
2,7
2,51 ÷ 3,1
6,0 ÷ 7,4
4,5 ÷ 7
13,6
2,8
2,4 ÷ 4,5
2,5 ÷ 8,0
4,7 ÷ 10,2
4,0 ÷ 12,0
2,8 ÷ 8,1
2,3 ÷ 3,4
5,0 ÷ 8,0
2,0 ÷ 6,0
3,0 ÷ 5,0
- 26 - 27 -
XVI. Notes
- 26 - 27 -
e l e c t r o n i c l e v e l m e a s u r e m e n t s y s t e m s
industrial electronics
Dinel, s.r.o.
U Tescomy 249
760 01 Zlin
Czech Republic phone: (+420) 577 002 003 fax: (+420) 577 002 007 e-mail: [email protected]
www.dinel.cz
www.levelmeasurement.eu
The latest version of service manual can be found at www.dinel.cz.
advertisement
Key Features
- Continuous level measurement
- Suitable for liquids, powders and bulk-solids
- Variants for non-explosive and explosive areas
- Intrinsically safe for hazardous areas
- Available with different types of electrodes
- High temperature performance available
- Adjustable sensitivity and range
- Easy to install and maintain
- Protection against reverse polarity and overvoltage
Frequently Answers and Questions
How do I install the CLM-36 level meter?
What is the difference between the CLM-36N and CLM-36Xi variants?
How do I set up the CLM-36 level meter?
Related manuals
advertisement
Table of contents
- 2 Content
- 3 Description
- 4 Technical specification
- 8 Types and performance of electrode
- 9 Linearity
- 10 Mounting and dismounting
- 13 Instalation of level meter and placing in operation
- 14 Preparing of level meter for measuring
- 15 VIII. Setting of level meter with unknown value of permittivity of measured medium
- 16 Setting of level meter with known value of permittivity of measured medium
- 19 Maintainance and service of level meter
- 20 Protection, safety, compatibility and explosion proof
- 21 Marking of level meters
- 22 XIII. Marking of labels
- 24 XIV. Accessories
- 25 Table of relative permittivity (dielectric constants) of selected materials
- 27 XIV. Notes
- 28 Contacts