Below you will find brief information for Capacitive level meters CLM-36 CLM-36N, Capacitive level meters CLM-36 CLM-36Xi, Capacitive level meters CLM-36 CLM-36NT, Capacitive level meters CLM-36 CLM-36XiT. The CLM-36 level meters are designed for continuous level measurement of liquids, powders and bulk-solid materials in vessels, tanks, sumps, containers, silos, etc. The level meters are available in different variants for non-explosive and explosive areas. It is important to note that the level meters are classified as intrinsically safe for hazardous areas.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Capacitive Level Meters CLM-36 Service Manual | Manualzz

SERVICE MANUAL

Capacitive level meters

CLM-36

Before level meter´s usage please read instructions in this service manual and keep it for further use. Service manual is for all types of level meters series CLM-36. Producer reserves the right to make changes without any previous announcement.

Content

Content .................................................................................................................................2

I.

II.

Description .................................................................................................................3

Technical specification ........................................................................................... 4-7

III. Types and performance of electrode..........................................................................8

IV. Linearity......................................................................................................................9

V. Mounting and dismounting................................................................................. 10-12

VI. Instalation of level meter and placing in operation...................................................13

VII. Preparing of level meter for measuring....................................................................14

VIII. Setting of level meter with unknown value of permittivity of measured medium......15

IX. Setting of level meter with known value of permittivity of measured medium.... 16-18

X. Maintainance and service of level meter..................................................................19

XI. Protection, safety, compatibility and explosion proof ...............................................20

XII. Marking of level meters............................................................................................21

XIII. Marking of labels................................................................................................ 22-23

XIV. Accessories..............................................................................................................24

XV. Table of relative permittivity (dielectric constants) of selected materials............ 25-26

XIV. Notes........................................................................................................................27

Contacts ..............................................................................................................................28

- 2 - 3 -

- 2 -

I. Description

Capacitive level meters CLM-36 are designed for continuous level measurement of liquids, powders and bulk-solid materials in vessels, tanks, sumps, containers, silos, etc.

Variants according working areas:

- for non-explosive area:

- for explosive area:

CLM-36N_-_ _-_

CLM-36Xi_-_ _-_

Available is high temperature performance for both working areas:

- high temperature performance for non-explosive area:

- high temperature performance for explosive area:

CLM-36NT-_ _-_

CLM-36XiT-_ _-_

Classification of areas:

Performance of level meter

CLM-36N

Part of the level meter with electrode

Part of the level meter with housing

non-explosive

CLM-36NT

CML-36Xi

CLM-36XiT non-explosive

II1GDT83°CEEXiaIIBT5 with ISIR

1) zone 0, zone 20

II1/2GDT83°CEEXiaIIBT5 with ISIR

1) zone 0, zone 20 zone 1, zone 21

The electrode and its surroundings (the wall of vasel, the reference tube, etc.) make capacitor. The electrode is designed according measured material:

- variant with rod electrode:

- variant with rod electrode and reference tube:

- variant with rope electrode:

Available is mechanical connection:

CLM-36_ _-10, 12

CLM-36_ _-20, 22

CLM-36_ _-30, 31, 32

- with thread:

- without thread:

CLM-36_ _-_ _-M

CLM-36_ _-_ _-G1

CLM-36_ _-_ _-Cl

M36x2

G1"

Triclamp

Inside of the level meter´s housing is placed removable electronic module with adjustment elements (see chapter VII).

1) ISIR - Intrinsically safe isolating repeater (example IRU-420)

- 3 -

II. Technical specification

Type of level meter

Supply voltage: - var. N, NT

-

var. Xi,XiT

Current output: - non-activated electrode

- full-activated electrode

Max. internal values of Xi version

Sensitivity ranges

Maximum capacity of flooded electrode (range)

Initial capacity regulation ratio

Nonlinearity

Temperature error

Voltage error

Internal resistance / electric strength

(electrode - housing)

Coupling capacity / electric strength

(housing - supply leads)

Allowed temperature range in zone 0 (EN 50284)

Allowed pressure range in zone 0 (EN 50284)

Protection class: - housing

- connector type GDM 2009

- connecor type GDM-K 2000

Connector socket type Hirschmann DIN 43 650

Recommended cable

Max. load (serial) resistance (U = 24 V)

Weight of the housing (without electrode)

CLM-36__-__-_

9 ÷ 36 V DC

9 ÷ 30 V DC

4 mA

20 mA

Ui=30VDC, Ii=132mA, Pi=0,99W,

Ci=370nF, Li=0,9mH

20, 30, 50, 100, 150,

300, 500, 1000 pF

70, 250, 600, 1200, 3000,

7000, 18000, 36000 pF min. 1:2 max. 1 % max. 0,05% / K max. 0,3 μA / V

1 MΩ / 250 V AC var. N 51 nF / 250 V AC var. Xi 26 nF / 500 V AC

-20 to +60°C

0,8 to 1,1 bar (0,08 to 0,11 MPa)

IP 67

IP 65

IP 67

GDM 2009 (standard)

GDM-K 2000 (optional)

2 x 0,75 mm

2

R max

=750 Ω var. N, Xi aprox. 0,5 kg var. NT, XiT aprox. 1 kg

- 4 - 5 -

- 4 -

II. Technical specification

Temperature resistance of CLM-36:

Variant of performance

CLM-36_ _-10, 20-_

CLM-36_ _-12, 22-_

CLM-36_ _-30, 31, 32-_

Temperature of medium

(at electrode)

-40 to +200°C

-40 to +120°C

-40 to +105°C

Ambient temperature range t

a

(at housing)

variant N

-40 to +85°C

variant Xi

-40 to +75°C

-40 to +85°C

-40 to +85°C

-40 to +75°C

-40 to +75°C

Pressure resistance of performance CLM-36N(Xi):

Variant of performance

CLM-36_-10, 20-_

CLM-36_-12, 22-_

CLM-36_-30, 31, 32-_

for t

a

Maximum operating overpressure

= -40 to +20°C for t

a

= -40 to +85°C

3 MPa 1 MPa

3 MPa

1 MPa

1 MPa

0,5 MPa

Pressure resistance of high temperature performance CLM-36NT(XiT):

Temperature in connection point

Maximum overpressure

+ 100°C +120°C +150°C +180°C

3,0 MPa 2,0 MPa 1,5 MPa 0,5 MPa

Used materials:

Part of the level meter (type)

housing (all) insulation bushing (all) electrode (10, 12, 20, 22) electrode (30, 31, 32) electrode coating (12, 22, 32) electrode coating (30, 31) weight insulation (32) weight / anchor mechanism (30, 31, 32) reference tube (20, 22)

standard

Material

st. steel W. Nr. 1.4301

optional

st. steel W. Nr. 1.4571,

Incoloy 825 (W. Nr.

2.4858)

PTFE st. steel W. Nr. 1.4301

zinc coated steel rope st. steel W. Nr. 1.4571

-

FEP polyolefin (modified PE)

PTFE st. steel W. Nr. 1.4301

st. steel W. Nr. 1.4306

-

PTFE

-

st. steel W. Nr. 1.4571

- 5 -

II. Technical specification

Dimensions of each variants: with rod electrode:

CLM-36_ _-10-_

CLM-36_ _-12-_ with rod electrode and reference tube:

CLM-36_ _-20-_

CLM-36_ _-22-_ with rope electrode:

CLM-36_ _-30-_

var. M, G - housing with

thread var. Cl - housing with

Triclamp coupling

Z* - var. M - thread M-36x2

- var. G - thread G1“

- 6 - 7 -

- 6 -

II. Technical specification

Dimensions of each variants: with rope electrode:

CLM-36_ _-32-_ with rope electrode with anchor:

CLM-36_ _-31-_

connector

high temperature performance:

CLM-36_T-_ _-_

steel anchor bushing or dusttight bushing wall of container pushing spring housing and type of electrode according specific version

- 7 -

III. Types and performance of electrodes

Characteristic of variants:

CLM-36_ _-10-_

With uncoated rod electrode - for level measurement of non-conductive liquids (oils, diesel, benzine) and bulk-solid materials (flour, sand, cement, plastic granulates, etc.). Max. electrode length 5 m.

CLM-36_ _-12-_

With coated rod electrode (FEP isolating) - for level measurement of water and other conductive liquids including waste liquids in metal vessels, concrete sumps, reservoirs, etc.

Resistant against aggressive medium, lower adhesion to some mediums. Max. electrode length 3 m.

CLM-36_ _-20-_

With uncoated rod electrode and reference tube - for level measurement of clean nonconductive liquids (oils, diesel, benzine). By means of reference tube the output signal does not depend on the dimension and shape of the vessel. Impossible to use for waste and high viscosity liquids and bulk-solid materials. Max. electrode length 3 m.

CLM-36_ _-22-_

With coated rod electrode (FEP isolating) and reference tube - for level measurement of clean conductive liquids. Main use is for measurement in plastic and glass vessels and for fine measuring. Not suitable for waste water and high viscosity liquids and bulk-solid materials. Max. electrode length 3 m.

CLM-36_ _-30-_

With coated rope electrode (polyolefine isolating) and uncoated weight - for level measurement of bulk-solid materials (grains, sand, flour, cement, etc.). Max. electrode length 12 m.

CLM-36_ _-31-_

With coated rope electrode with anchor (polyolefine isolation) and uncoated weight - as the type 30. For use for higher silos. Max. electrode length 20 m.

CLM-36_ _-32-_

With coated rope electrode (FEP isolation) and coated weight (isolation PTFE) - for level measurement of conductive and non-conductive liquids. Max. electrode length 20 m.

- 8 - 9 -

- 8 -

IV. Linearity

Influence of the tank shape on a linearity of measured capacity of nonconductive liquids:

1. In the tank with straight wall (for example vertical cylinder) and with the sensor placed

parallely with the wall capacity change is linear.

2. In a curved tanks (most frequently horizontal cylinder) capacity change during

measuring of electrically non-conductive material is non-linear. Linearity is done

by reference tube (CLM-36_ _-20, 22-_).

- 9 -

V. Mounting and dismounting

Level meters with coated electrode have protection cover at the end of electrode, which is necessary take down befere mounting.

Level meters mount in vertical position to upper lid of tank or reservoir by steel welding flange ON-36x2 (steel 11375), stainless steel welding flange NN-36x2 (stainless steel

1.4301), fixing nut UM-36x2 (stainless steel 1.4301) or Clamp flange.

For mounting level meter to the metal tank or reservoir is not necessary to ground housing again.

In case of installation in concrete sumps or silos it is properly to install level meter at metalic auxiliary construction (console, lid, etc.) and then connect with metal all the time submerged object, eventually with steel armouring.

For measuring of material in plastic and glass vessels by level meter without reference tube is necessary to connect grounding screw at housing with auxiliary electrode which is fixed by the appropriate method to the outer casing of vessels (or at inner wall). Material of auxiliary electrode is necessary to choose with consideration to the working environment and properities of measured material.

Orientation of proportion for mounting of level meter (including eventualoly auxiliary electrode) shows following pictures.

CLM-36_ _-10, 12-_

installation into the metal and non-metal vessels

auxiliary electrode

width = min. 30 mm

(necessary only for non-metal vessels)

E - the length of electrode [mm] - the end of the electrode has to be dipped min. 20 mm below the lowest measured

level

h - the distance from the bottom - min. 50 mm

a - the distance from the wall - min. approx. E/20

d - the diameter of auxiliary tube vessel - min. 40 + E/20 (smaller dimension has to be discussed with producer)

- 10 - 11 -

- 10 -

V. Mounting and dismounting

CLM-36_ _-20, 22-_

installation with reference tube

h - the distance from the bottom - min. 50 mm with respect on possibility of presence of heavy fraction (water) in oil

products

k - the distance from the wall - optional

CLM-36_ _-30, 32-_

installation into the metal tank and concrete containers connection to the armouring

E - the length of electrode [mm] - the end of the electrode has to be dipped min. 20 mm below the lowest measured

level

h - the distance from the bottom - min. 100 mm

a - the distance from the wall - approx. the same as the distance from the inlet

- 11 -

V. Mounting and dismounting

CLM-36_ _-31-_

installation of rope electrode with anchor

steel anchor welding cylinder

(mat. 11375) or

dust-tight bushing

to be welded into the conical bottom wall

E - the length of electrode [mm]

c - the distance from the centre line of the silo - has to be as minimal as possible

t - the length of the conduction rod - approx. 500 mm

p - the length of the press spring - approx. 200 mm

- 12 - 13 -

- 12 -

VI. Instalation of level meter and placing in operation

Mounting, installation, placing in operation, maintainance, using and service of level meter CLM-36 must be done in conformity with this service manual and has to be performed according the norms for installation of electrical devices.

Electric connection of feeder cable is realized in state without voltage through the use of thread terminal in dismountable connector GDM at level meter’s housing.

Positive pole of supply (+U) will be connected at terminal (1), negative pole of supply

(0V) will be connected at terminal (2) - see following pictures.

Connection scheme of level meter

The inside scheme of the GDM connector socket

+U shielding

0V

Cable from level meter to the further unit is by two-wire cable with outside diameter of

6 ÷ 8 mm (recommended cross section of cores 0,5 ÷ 0,75 mm

2

).

A shielded cable must be only used if the lead is longer than 30 m or in the event that a power line must be used simultaneously with the lead. Connect shielding to the socket , shielding do not connect to the power panel.

Power supply for variants CLM-36N and CLM-36NT can be by stabilized safety voltages power supply 9 ÷ 36 V DC (SELV), example PSU-2400 or DSU-2400.

Or following evaluation device, for example programmable display unit PDU-4xx. For variants CLM-36Xi and CLM-36XiT must be connected with intrinsically safe isolating repeater 9 ÷ 30 V DC, for example IRU-420.

Activation is by turning on of the power supply (for example PSU-2400 či DSU-2400 for variants CLM-36N and CLM-36NT) or intrinsically safe isolating repeater (for example IRU-

420 for variants CLM-36Xi and CLM-36XiT).

- 13 -

VII. Preparing of level meter for measuring

Preparing of level meter

1. For access to the adjustment elements disconnect a connector and unscrew a nut

(attention for inside connecting wires). Connector connect again.

2. Level meter connect to the supply unit through miliampermeter (controler, etc.).

3. Trimmer

2)

20 mA bring into the basic position (this position is set by producer):

a) Turn out trimmer totally rightwards (in clockwise).

b) Return back for 3 turns leftwards.

Adjustment elements of level meter

trimmer 4 mA

(for compensation of initial capacity)

blue wire

(-)

green-yellow wire

( )

brown wire

(+)

trimmer 20 mA

(for sensitive span adjustment)

brass tie

(for pull out of elecronic module)

(the top view on the internal electronic module)

DIP switches

(for range select)

2) The trimmers are without backstops - approx. 15 turns.

3) Sensitivity - minimal capacity change of electrode to reach a current range 4 ÷ 20 mA.

4) Range - maximum capacity of housing and activated electrode to reach current range 4 ÷ 20 mA.

5) In case of short of electrode on housing or setting of a very sensitive range can caused to current restriction of level meter on value max. 30 mA.

- 14 - 15 -

- 14 -

VIII. Setting with unknown value of permittivity

Setting of level meter with unknown value of permittivity of measured medium

1. Empty tank to the minimum level.

2. By DIP switches on level meter set range

4)

No. 2 (250 pF).

3. Turn by screwdriver in trimmer 4 mA and set quiescent current of the level meter to

4 mA. Turning rightwards (in clockwise) increase the current, turning leftwards

decrease the current. When is impossible to set required current to 4 mA, then switch

DIP switches to the nearest higher range and set up of current 4 mA perform at this

range.

4. Fill the tank to the maximum level. When it is impossible fill the tank to the maximum

state then we use some known (measurable) level and current for next setting

calculate according to formula:

I out = 4 + (0,16 x level in %) [mA]

5. By DIP switches stepwise switch to higher ranges till the current is greather then

20 mA (or calculated value Iout), thru level meter then flows current e.g. 21 mA

Trimmer 20 mA is still in basic position.

5)

.

6. Then set by trimmer 20 mA current of level meter to 20 mA (or to calculated value I out

- see bellow).

7. To reach a maximum accuracy is useful to check over the setting of 4 mA

(compensation of initial capacity - quiescent current).

Table of ranges:

No.

1

6)

2

3

4

5

6

7

8

Sensitivity

(S)

3)

20 pF

30 pF

50 pF

100 pF

150 pF

300 pF

500 pF

1 000 pF

Range

4)

(R)

70 pF

250 pF

600 pF

1 200 pF

3 000 pF

7 000 pF

18 000 pF

36 000 pF

Position of

DIP switch

6) Range No. 1 (70 pF) is very sensitive and we recommend to use it only at level meters with short electrode

(approx. to 500 mm) and measured media with low permittivity.

- 15 -

IX. Setting with known value of permittivity

Setting of level meter with known value of permittivity of measured medium

1. Calculation of measuring range:

a) Count sum of capacity of housing (CH) and capacity of non-activated

electrode (CNE).

C1 = CH + CNE

Capacity of housing

(typical values)

Capacity of housing - normal performance

CH

21 pF

Capacity of housing - high temperature performance 28 pF

Capacity of non-activated electrodes

(electrode length 1000 mm)

all types without reference tube in distance 250 mm from conducting wall uncoated rod electrode with reference tube (type 20) coated rod electrode with reference tube (type 22)

CNE

(air)

12 ÷ 15 pF approx. 50 pF approx. 55 pF

b) Count sum of capacity of housing (CH) and capacity of activated electrode

(CAE) - selected mediums see the table.

C2 = CH + CAE

Capacity of activated electrodes

(electrode length 1000 mm)

all types without reference tube in distance 250 mm from conducting wall uncoated rod electrode with reference tube (type 20) coated rod electrode with reference tube

(type 22)

CAE

(benzine)

25 ÷ 30 pF approx. 100 pF approx. 110 pF

CAE

(water)

500 pF

(coated electorde)

-

500 pF

- 16 - 17 -

- 16 -

IX. Setting with known value of permittivity

c) Set the possible ranges -

capacity of housing and activated electrode

(C2) will not be greather than range value (R).

C2 < R

d) Sensitivity check - difference of capacity activated (CAE) and non-activated

(CNE) electrode must be greather than sensitivity of calculated range.

C2 - C1 > S (CAE - CNE > S)

e) Choose the range - both previous conditions must be used together c) and d).

2. Empty the tank to the minimum level.

3. By DIP switches on level meter set up pre-counted range.

4. Turn by screwdriver in trimmer 4 mA and set quiscent current of the level meter to

4 mA. Turning rightwards (in clockwise) increase the current, turning leftwards

decrease the current. When is impossible to set required current to 4 mA, then switch

DIP switches to the nearest higher range and set up of current 4 mA perform at this

range.

5. Fill the tank to the maximum level. When it is impossible fill the tank to the maximum

state then we use some known (measurable) level and current for next setting

calculate according to formula:

I out = 4 + (0,16 x level in %) [mA]

6. On preliminary range of level meter set by trimmer 20 mA current of level meter to

20 mA (or to calculated value I out - see bellow). Turning rightwards (in clockwise)

increase the current, turning leftwards decrease the current.

7. To reach a maximum accuracy is useful to check over the seting of 4 mA (compensation

of initial capacity - quiscent current).

Example of calculation:

Level meter CLM-36N-20-M E1000, measured medium is benzine (ε r

= 2,2), distance from wall to level meter with uncoated rod electrode with reference tube is optional.

1.a) Sum of capacity of housing and non-activated electrode:

C1 = CH + CNE = 21 + 51 =

72 pF

1.b) Sum of capacity of housing and activated electrode:

C2 = CH + CAE = 21 + 100 =

121 pF

- 17 -

IX. Setting with known value of permittivity

1.c) Possible ranges:

C2 < R

121 pF < (2) 250 pF, (3) 600 pF, (4) 1200 pF, (5) 3000 pF, (6) 7000 pF,

(7) 18000 pF, (8) 36000 pF

1.d) Check the sensitivity:

C2 - C1 > S

C2 - C1 = 121 - 72 = 49 pF

1.e) Recommended accurate range:

range (2) 250 pF

49 pF > (1) 20 pF,

(2) 30 pF

Level meter CLM-36N-20 with electrode E = 1000 mm by flowage of benzine has capacity

120 ÷ 135 pF - typical measuring range is range (2) 250 pF.

Typical measuring ranges of CLM-36 with electrode 1 m, 2 m and 5 m length for selected medias:

Medium Sand, grains

Type 1 m 5 m

10 2 3

12

30

31

Medium

Type

10

12

20

22

32

2

2

2

1 m

Water

2 m

-

3

-

4

-

3

3

4

4

3

3

3

2

2

2

Cement

1 m 5 m

2 4

4

4

4

1

1

1

1 m

Flour

5 m

1 3

3

3

3

Benzine, petroleum, diesel, oils

1 m 2 m

1 2

1

2

2

3

2

1

3

2

Note:

- 18 -

Types 10, 12, 30, 31, 32 in distance 250 mm from conducting wall.

Showned ranges are orientation. Exact range for existing configuration electrode / tank must be set directly in application on mearured medium.

- 19 -

- 18 -

X. Maintainance and service of level meter

Maintainance of level meter

Maintainance of device is to control the housing of level meter and lead-in calbe. In case of any visible damages please contact producer or supplying company.

At level meter is prohibited to perform any changes or repair without producer confirmation.

Service of level meter

Eventual reparation of mechanical damage of level meter must be done by producer or by producer authorized service organization.

Replacement of defective electronic

In case of detected electronic failure it is possible to replace the electronic module directly in process place of level meter by following procedure:

1. Disconnect connector and unscrew a nut (attention for inside connecting wires).

2. Defective electronic module pick up by pliers onto brass tie and remove by tensile

from housing of level meter.

3. Check placing of white seal O-ring in housing and on pressure ring.

4. Repaired or new electronic module return back into housing of level meter and press

down by clamping ring (attention that contact pin will be a little open before insertion

and contact springs were not pressed under level of electronic casing).

5. Check placing of black seal O-ring on connector.

6. Screw a nut back (attention for inside connecting wires) and connect the connector.

trimmers brass tie contact springs electronic module cover contact pin

- 19 -

XI. Protection, safety, compatibility and explosion proof

Protection

Level meter is equipped with protection against reverse polarity of supply voltage and against overvoltage.

Safety

Safety against dangerous contact is ensure by safe power supply according norm EN 33

2000-4-41 (SELV).

Compatibility

Electromagnetic compatibility is provided by conformity with norms EN 55022/B, EN

61326-1, EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6.

Explosion proof

Explosion proof of CLM-36Xi and CLM-36XiT is examined by FTZU-AO 210 Ostrava

- Radvanice, certificate No. 02 ATEX 0235X.

- 20 - 21 -

- 20 -

XII. Marking of level meters

Method of marking of level meters CLM-36

CLM-36 E

Basic thread dimension

Performance:

N

- normal (for non-explosive areas)

Xi - explosion proof - intrinsically safe for hazardous

(explosive) areas (clasification zone 0)

NT, XiT - high temperature

Electrode length in mm

Process connection:

M - thread M36x2

G - thread G1"

Cl - Triclamp

Electrode performance:

10 - rod uncoated electrode (max. 5 m)

12 - rod coated (FEP isolation) electrode (max. 3 m)

20 - rod uncoated electrode with reference tube (max. 3 m)

22 - rod coated (FEP isolation) electrode with reference tube (max. 3 m)

30 - rope electrode with coated (polyolefine isolation) rope and uncoated weight (max. 12 m)

31 - rope electrode with anchor with coated rope (polyolefine isolation) and uncoated weight

(max. 20 m)

32 - rope electrode with coated rope (FEP isolation) and coated weight (isolation PTFE) (max.

20 m)

Examples of correct marking

CLM-36N-10-G E1100

CLM-36XiT-20-M E900

CLM-36N-12-Cl E2000

CLM-36NT-12-M E1500

CLM-36Xi-30-M E12000

CLM-36N-32-G E6000

- 21 -

XIII. Marking of labels

On label of level meter type CLM-36N(T) are following datas:

- symbol of producer: logo Dinel

®

- contact: Dinel, s.r.o., Zlin, Czech Republic, www.dinel.cz, [email protected]

- connection scheme and labelling of wires: +U, 0V, GND

- type of level meter: CLM-36-N_-__-_, include length of electrode: E in mm

- serial number: Ser. No.: xxxxx - production year (first two position) and sequence

factory number (next three position)

- power supply voltage: U i

= 9 ÷ 36 V DC

- output current range: I = 4 ÷ 20 mA

- ambient temperature range: t a

= -40 ... +85 °C

- protection degree: IP 65 / IP 67

- label of shody: CE

- symbol of electro-equipment for back collection and separate collection of electro-

waste

On label of level meter type CLM-36Xi are following datas:

- symbol of producer: logo Dinel

®

- contact: Dinel, s.r.o., Zlin, Czech Republic, www.dinel.cz, [email protected]

- connection scheme and labelling of wires: +U, 0V, GND

- type of level meter: CLM-36-Xi-__-_, include length of electrode: E in mm

- serial number: Ser. No.: xxxxx - production year (first two position) and sequence

factory number (next three position)

- label of non-explosive device: Ex in hexagon

- performance: II1GDT83°CEExiaIIBT5

- number of authorized person examining control of system quality: 1026

- number of certificate of intrinsically safety: FTZU 02 ATEX 0235X

- limiting operating parameters: U i

= 30 V DC, I i

= 132 mA, P i

= 0,99 W, C i

= 370 nF,

L i

= 0,9 mH

- ambient temperature range: t a

= -40 ... +75 °C, t a

= -20 ... +60 °C (zone 0)

- protection degree: IP 65 / IP 67

- label of shody: CE

- symbol of electro-equipment for back collection and separate collection of electro-

waste

- 22 - 23 -

- 22 -

XIII. Marking of labels

On label of level meter type CLM-36XiT are following datas:

- symbol of producer: logo Dinel

®

- contact: Dinel, s.r.o., Zlin, Czech Republic, www.dinel.cz, [email protected]

- connection scheme and labelling of wires: +U, 0V, GND

- type of level meter: CLM-36-XiT-__-_, include length of electrode: E in mm

- serial number: Ser. No.: xxxxx - production year (first two position) and sequence

factory number (next three position)

- label of non-explosive device: Ex in hexagon

- performance: II1/2GDT83°CEExiaIIBT5

- number of authorized person examining control of system quality: 1026

- number of certificate of intrinsically safety: FTZU 02 ATEX 0235X

- limiting operating parameters: U i

= 30 V DC, I i

= 132 mA, P i

= 0,99 W, C i

= 370 nF,

L i

= 0,9 mH

- ambient temperature range: t a

= -40 ... +75 °C, t a

= -20 ... +60 °C (zone 0)

- protection degree: IP 65 / IP 67

- label of shody: CE

- symbol of electro-equipment for back collection and separate collection of electro-

waste

- 23 -

XIV. Accessories

Standard accessories:

For each level meter:

- 1 pc of connector socket GDM 2009 (IP 65)

- 1 pc of seal (material AFM)

For each delivery (each 5 pcs):

- 1 pc of screwdriver for adjustment

Optional accessories:

For mounting of level meter:

- stainless fixing nut UM-36x2 (material stainless steel 1.4301)

- steel welding flange ON-36x2 (material steel 11375)

- steel welding flange with lock and Al seal ON-36x2-Z (material steel 11375)

- stainless welding flange NN-36x2 (material stainless steel 1.4301)

Other:

- connector socket GDM-K 2000 with 5 m of calbe (IP 67)

- seal (on request - material PTFE, Al, etc.)

- 24 - 25 -

- 24 -

XV. Table of relative permittivity

Table of relative permittivity (dielectric constants) of selected materials

Material

Acetone

Acetylentetrachloride

Aminoplasts

Ammonia

Aniline

Ash

Benzaldehyde

Benzene

Benzol

Celluloide

Cement

Dioxane

Ebonit

Ethanol

Ethylacetate

Ethyleneglycol

Ethylenechloride

Flint crystal

Flour

Food salt (NaCl)

Freon R22

Fused quartz

Gasoline (benzin)

Glass

Glass organic

Glass silicate

Glazed carton

Glycerine

Grains

Hardened textile

Hexane

ε

r

3,7

2,0 ÷ 2,2

3,7 ÷ 10

3 ÷ 3,6

16

3,5

47

3,0 ÷ 5,0

2 ÷ 6

1,9

24

6,0

38,7

10,5

4,5

2,5 ÷ 3,0

6,0

6,1

19,5 ÷ 21,4

8,1

5 ÷ 8

15 ÷ 24

6,9

1,5 ÷ 1,7

18,3

2,28 ÷ 2,3

2,3

3,5 ÷ 6,2

4,0

2,235

2,5 ÷ 5

Material

Chloroforme

Ice

Liquefied air

Liquefied CO2

Liquefied chlore

Liquefied propan

Marble

Methanol

Mica

Mikanite

Milk powder

Monochlorbenzene

Nitrobenzene

Ortonitrotoluene

Paper

Paraffin

Paraffin oil

Paraffinic oil

Petroleum

Polyacetal

Polyamide - PA

Polydichlorstyrole

Polyetheretherketone -

Polyetherimide - PEI

Polyethylene - PE

Polyfenylenesulfide - PPS

Polymethylmetakrylate

Polypropylene - PP

Polystyrole

Polysulphone - PSU

Polytetrafluorethylene

ε

r

3,0

2,16

3,3

2,56

2,0 ÷ 2,2

2,39

3,0

2,0 ÷ 2,1

1,9 ÷ 2,5

1,9 ÷ 2,5

2,2

2,0 ÷ 2,2

3,6 ÷ 3,7

4,0 ÷ 5,0

2,7

3,2

4,81

3,1

1,5

1,6

2,0

1,6 ÷ 1,9

9,3 ÷ 11,6

32,7 ÷ 33,8

5 ÷ 8

4,5 ÷ 6

3,5 ÷ 4

4,0

35,7

27,4

1,6 ÷ 2,6

- 25 -

XV. Table of relative permittivity

Table of relative permittivity (dielectric constants) of selected materials

Material

Polyvinylacetate

Polyvinylchloride - PVC

Polyvinylidenefluoride

Porcelain

Pyridine

Rape-seed (colza) oil

Resin - acrylic

Resin - epoxy

Resin - melamine

Resin - phenolic

Resin - polyester

Resin - styrene

Resin - ureal

Rubber

Sand

Silicon caoutchouc

Succinite

Sugar

Sulphur

Toluene

Trichlorethylene

Trolitule

Vaseline

Water

Water solution

White beeswax

Wood - damp

Wood - dry

ε

r

2,8 ÷ 3,3

2,9

3,0

3,4 ÷ 3,6

2,3 ÷ 2,38

3,43

2 ÷ 2,6

2,2 ÷ 2,9

81

50 ÷ 80

2,0 ÷ 2,9

10 ÷ 30

2 ÷ 6

2,7

2,51 ÷ 3,1

6,0 ÷ 7,4

4,5 ÷ 7

13,6

2,8

2,4 ÷ 4,5

2,5 ÷ 8,0

4,7 ÷ 10,2

4,0 ÷ 12,0

2,8 ÷ 8,1

2,3 ÷ 3,4

5,0 ÷ 8,0

2,0 ÷ 6,0

3,0 ÷ 5,0

- 26 - 27 -

XVI. Notes

- 26 - 27 -

e l e c t r o n i c l e v e l m e a s u r e m e n t s y s t e m s

industrial electronics

Dinel, s.r.o.

U Tescomy 249

760 01 Zlin

Czech Republic phone: (+420) 577 002 003 fax: (+420) 577 002 007 e-mail: [email protected]

www.dinel.cz

www.levelmeasurement.eu

The latest version of service manual can be found at www.dinel.cz.

advertisement

Key Features

  • Continuous level measurement
  • Suitable for liquids, powders and bulk-solids
  • Variants for non-explosive and explosive areas
  • Intrinsically safe for hazardous areas
  • Available with different types of electrodes
  • High temperature performance available
  • Adjustable sensitivity and range
  • Easy to install and maintain
  • Protection against reverse polarity and overvoltage

Frequently Answers and Questions

How do I install the CLM-36 level meter?
The level meter can be mounted in a vertical position to the upper lid of the tank or reservoir using a steel welding flange, stainless steel welding flange, fixing nut, or clamp flange. Refer to the service manual for specific installation instructions.
What is the difference between the CLM-36N and CLM-36Xi variants?
The CLM-36N variant is designed for non-explosive areas, while the CLM-36Xi variant is explosion proof and intrinsically safe for hazardous areas. The CLM-36NT and CLM-36XiT variants offer high-temperature performance.
How do I set up the CLM-36 level meter?
The level meter can be set up with known or unknown permittivity of the measured medium. Refer to the service manual for detailed setup instructions.