ESCHWEILER Combi Blood Gas Electrolyte Analyser Service Manual
The Eschweiler combiline is a blood gas and electrolyte analyzer that is designed for rapid and accurate measurement of blood gas and electrolyte parameters. It is a compact and easy-to-use system that is suitable for use in a variety of clinical settings. The combiline is equipped with a range of features that make it a versatile and reliable analyzer, including a built-in thermometer, a printer, and a user-friendly interface.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
combi
n line
Blood-Gas-Electrolyte-Analyser
For In-Vitro-Diagnostic Use
!
Service Manual
Eschweiler GmbH & Co. KG
Development - Manufacturing - Sales - Technical Service
Made in Germany
Service Manual
Manual Version
Manual History
Software Copyrights
Manufacturer
The latest Manual version is: 4.01
Order No.: 60 8 90 10 Print: SMe COMBILINE 401.PDF
Copyright © 2010 by
Eschweiler GmbH & Co. KG, Holzkoppelweg 35, 24118 Kiel
No part of this Manual may be reproduced in any form without the written permission of Eschweiler GmbH & Co. KG.
Manual ANALYSER-Software
Version Version
4.01
2500
Revision
Date
April 2010
The software for the combiline Analyser is the User and Instrument Control Software for the combiline. The combiline Software is the intellectual property of Eschweiler GmbH & Co. KG hereafter Eschweiler.
Intellectual property rights shall remain with Eschweiler.
You are entitled to use the combiline Software and the printed accompanying material at your place of work only.
Any violations of property rights or copyright or trademark or conditions of use may be subject to legal action.
Eschweiler GmbH & Co. KG
Holzkoppelweg 35
24118 Kiel, Germany
Phone: -49- 431-54 65 80
FAX: -49- 431-54 94 23
Internet: http://www.eschweiler-kiel.de
E-Mail: [email protected]
!
Service Manual
Consumable solutions and spare parts from other manufacturers used in ESCHWEILER Analysers.
We point out specifically that fault-free operation of our analysers can be guaranteed only when original ESCHWEILER solutions and spare parts are used.
The name ESCHWEILER with its decades of experience in the field of blood gas analysis also stands for the quality of the spare parts and consumable solutions. ESCHWEILER sensors and solutions form a functional unit whose components are optimally adapted to each other. Measuring accuracy and long life of the sensors are the goal of this system.
Through intensive development work, the chemical compositions of our solutions have been adjusted optimally to the ESCHWEILER sensors and guarantee, besides precise measurements, the greatest possible protection of the sensitive sensor surfaces.
For example, special wetting solutions are used which have a decisive influence on the unconstrained contact between sensor and solution in the measuring capillary. At the same time suitable preservatives prevent the growth of micro-organisms in the consumable solutions and keep them from being introduced into the measuring capillaries of the ESCHWEILER analysers.
The characteristics and concentrations of both the wetting- and the preservative substances are selected according to the special requirements of our sensors.
Fault-free operation of our analysers with measuring accuracy and long sensor life are the main goals of our quality requirements. Only if
ESCHWEILER original parts and solutions are used we can meet these requirements.
Please support us in doing this!
Service Manual
Table of Contents
Safety Issues ................................................................................................................................................ 2-5
Technical Data. ................................................................................................................................................ 6
Menue Guide. .................................................................................................................................................. 7
System Description ...................................................................................................................................... 8-11
Tubing. ....................................................................................................................................................... 12-15
Specification of Valves. ....................................................................................................................................16
Blockdiagram. .................................................................................................................................................17
Power Supply ..................................................................................................................................................18
Power Supply ESW-MVTC. ........................................................................................................................ 19-24
Adjustment of the operating voltage for Rollerpumps RP1 and RP2. ................................................................21
Adjustment of Thermostate Temperature .........................................................................................................23
Analog-Board ESW-ANALOG..................................................................................................................... 25-35
Adjustment of O2 Polarization Voltage and O2 Zero Point. ..............................................................................27
Circuit diagram for Sample-light-Barrier ...........................................................................................................28
Circuit Diagram for Hb-unit ..............................................................................................................................28
Testing the Airpressure Indication. ...................................................................................................................29
Adjustment of Thermostat Temperature-Indication. ..........................................................................................30
Adjustment of the Sample Light-Barrier ...........................................................................................................31
Capillary exchange in Sample-Light-Barrier .....................................................................................................32
Hb-Measurement module ........................................................................................................................... 33-35
CPU / AD-Converter CL Z80ADC. .............................................................................................................. 36-40
Serial Interface with SD Card Reader RS232 SD INT1.4 ............................................................................. 41-47
Fill Level Control CL FÜLLSTAND. .............................................................................................................. 48-50
Keyboard. .......................................................................................................................................................51
Display / Contrast. ..........................................................................................................................................52
Thermoprinter GPT-4352-60 .............................................................................................................................54
Liquid-Valves .............................................................................................................................................. 55-56
Delay Circuit for Ventilation Valve ....................................................................................................................57
Suction-Pump ............................................................................................................................................ 58-59
Sensor Spare Parts and Lifetime Information. ............................................................................................. 60-62
Consumeables ........................................................................................................................................... 63-65
Spare Parts ................................................................................................................................................ 66-68
TÜV Certificate ................................................................................................................................................69
Declaration of Conformity ................................................................................................................................70
combiline
Service- Manual - Version 4.01
Safety Issues
Please read the Instruction Manual in its entirety prior to operating the line
. In order to ensure a high level of performance, all warnings and references to technical safety in this Instruction Manual have to be followed.
! 1 Hazards and Precautions
The cautions and safety regulations in this Instruction Manual meet international classifications:
Warns of a risk of injury or of a risk to life (for example by electrical shock).
DANGER Warns of a risk of injury or of the analyser being severely damaged.
CAUTION
Biohazard
NOTE
Electrical safety
DANGER
Warns of a risk of biological material like blood or other human or animal specimen. Risk of infection! Potential infectious area/material!
Introduces rules to be observed.
The following safety issues have to be observed at all times:
- Check that the operating voltage is set correctly before you connect the device to the main power supply.
- To connect the analyser to the main power supply, use only sockets which are grounded to avoid the risk of an electrical shock.
- Use only grounded extension power cords. The used power cord has to correspond to the country specific regulations.
- Never intentionally disconnect the grounding contacts.
There is the risk of electrical shock if the protective conductor is interrupted within or outside the analyser, and/or the grounded contacts have been disconnected from the line.
- Never remove protective guards or secured components since you could expose electrically live parts in this way.
- Electrical connection contacts (plugs, sockets, etc.) can be electrically live.
Service- Manual - Version 4.01
combiline
Safety Issues
- Even after an analyser has been switched off, components (e.g. capacitors) can be under voltage as the result of an electrical charge.
- All current carrying parts are sources of danger for an electrical shock.
- Surfaces (floors, work tables) have not to be moist when you are working with any electrical device.
- Carry out only the maintenance work and/or the replacement of parts described in these Instruction Manual.
- Unauthorized work on the analyser can lead to the guarantee obligation becoming null and void with necessary expensive service work to correct it.
- All work, which requires the analyser to be opened, have to be carried
out by an authorised technician who is familiar with the risks related thereto.
- Use only replacement fuses of the stated type and with the stated nominal current. Never use fuses, which have been “repaired”.
- There is a Lithium battery on MAIN BOARD that have to be replaced
by the authorised technical service every 5 years! In case of a battery fault, all memorised parameter might be lost.
Mechanical Safety (Analyser is operating)
- Never open or unscrew analyser's casing parts while it is switched ON.
There is a risk of injury due to moving parts as fans, pumps, stepper-motors or any other mechanical movements.
CAUTION
Biological Material - Risk of Infection
- Wear gloves in all cases if there is a risk of infection.
Biohazard
- Avoid any direct contact with samples that are potentially infectious or which may generate other risks to the human body (Aids, Hepatitis etc).
In case of direct body contact go and wash the contaminated area immediately. Use a suitable skin disinfection solution. Ask your doctor for aid.
- If sample material (blood) or reagent is spilled onto the analyser, wipe it off immediately and refer to the chapter ! 2 Maintenance and Hygiene.
- Do not open a waste bottle as long as the analyser is in process.
combiline
Service- Manual - Version 4.01
Safety Issues
Reagents and Controls
CAUTION
- Observe the suggestions in the package inserts for an exact use of the reagents and quality controls. Note that reagents and quality control material can be biological material!
- Don't use Reagents, Controls and other Liquids after expiration date!
- Avoid any liquid penetration into the analyser.
Restrictions for Samples, Reagents and Controls
- For consumables no guarantee can be provided for any resistance against organic solvents. For this reason, do not use any organic solvents unless such solvents are expressly indicated.
CAUTION
- Do not use any other rinsing or cleaning solutions as recommended by the manufacturer or representative.
- Waste liquid has to be disposed in compliance with the legislation.
Accuracy and Precision of the
Measured Results
CAUTION
- In order to ensure a flawless operation of the analyser measure control samples and watch the function of the analyser closely.
- Faulty measurement results may result in an incorrect diagnosis or range danger for patient.
Fire and Explosion Hazards
- Do not place any flammable or hazardous explosive material in the proximity of the device. Electrical sparks could cause fire or explosions.
CAUTION
Operator Qualification
- The analyser should only be operated by trained personnel. Improper operation of the analyser might cause inaccurate measuring results!
! 2 Maintenance and Hygiene
- No organic acid based cleaning substances should be applied. Instead use cleaner designed for cleaning and disinfecting laboratory analysers.
Only use a dampened cloth to clean the analyser.
CAUTION
- Never spray or pour cleaning solution directly onto the analyser, that may negatively impact the analyser’s functions significantly.
- Keep the analyser clean and do not spill liquids onto it.
- In case liquids were spilled onto the analyser, immediately absorb liquid with a suitable cloth.
- Contact your distributor if your control measurements do not produce the expected results.
Service- Manual - Version 4.01
combiline
Safety Issues
Recommended Disinfectant
Solutions.
A suitable disinfectant is available by BODE Chemie, Hamburg:
Bacillol ® , is recommended for areas at risk of infection, where rapid contact times and drying times are necessary.
BODE Chemie GmbH & Co
Melachtonstr. 27 - 22525 Hamburg, Germany
Phone: +49- (0)40-54 00 60 - Internet: www. bode-chemie.de
Another suitable disinfectant is available by S&M Schülke & Mayr:
TERRALIN ® Liquid
Schülke & Mayr GmbH
Robert-Koch-Str. 2 - 22851 Norderstedt, Germany
Phone: + 49- (0)40-521 00-0 - Internet: www.schuelke-mayr.com
See the website for your nearest representation. Ask for an EC Material
Safety Data Sheet according to EC 91/155.
! 3 First-Aid Measures
If a person swallowed any chemical solution may be while servicing, wash out mouth with water.
If provided person is unconscious, than call a physician.
If inhaled, remove to fresh air. If breathing becomes difficult, than call a phy-
sician.
In case of contact, immediately wash skin with soap and copious amount of water.
In case of contact with eyes, flush with copious amounts of water for at least
15 minutes.
Assure adequate flushing by separating the eyelids with fingers, than call a physician.
! 4 Repairs
CAUTION
- Repairs to the analyser may only be carried out by trained personnel, and replacement parts have to comply with the analyser specifications.
- In case of analyser problems contact your representative.
Useful hint
NOTE
In this Manual a bullet "•" draws your attention to an instruction.
Example:
• Press enter-key to confirm ...
combiline
Service- Manual - Version 4.01
Technical Data
Service- Manual - Version 4.01
combiline
Menue-Guide
POWER UP COMBILINE
Rev. 2.500
WARM UP PHASE
OF 1 HOUR press 1 x
OPERATION key to select Menu
ECONOMY
PROGRAM
AUTO-CALIBRATION
OF SENSORS
STAND BY
DIALOGUE
DISPLAY
MEASURED
MEASURING DIALOGUES
Sample introduction by selection of:
MEASURE
(SYRINGE or CAPILLARY) specify:
- BGA and meta values only
- ISE and meta values only
QC specify:
- LEVEL 1
- LEVEL 2
- LEVEL 3
- LEVEL 4
ASPIRATION GAS
Attention!
Sofware functions marked with * are not accessible during Warming Up!
Software functions marked with ** are not available in BGA types!
Software functions marked with *** are not available in ISE types!
OPERATION
1 = RINSE
2 = CALIBRATION *
3 = CALIBRATION BGA * **
4 = CALIBRATION ISE * **
5 = ACTIVATION HB *
6 = SENSOR PARAMETER *
7= ECONOMY MODE *
8 = DISPLAY MEASURED press 2 x
OPERATION key to select Menu
SERVICE - TEST
1 = BGA 1 / 3 ***
2 = BGA 2 / 4 ***
3 = CAL 3 **
4 = CAL 4 **
5 = GAS ***
6 = AIR / TEMP
7 = LIGHT BARRIER
8 = PROTEIN *
9 = REAGENT TEST press 3 x
OPERATION key to select Menu
STANDARD DATA
1 = CALIBRATION DATA ***
2 = DATE / TIME
3 = CORRELATION
4 = SETTINGS
5 = SENSOR SELECTION
6 = AB-PARAMETER ***
combiline
Service- Manual - Version 4.01
System Description
Opened front cover
1 - LCD-Display
2 - Function keys
3 - Numerical keys
Order no.
F100.4.2
50 8 20 70 (keyboard compl.)
50 8 20 70 (keyboard compl.)
4 - Screw joint of Reagentwall
5 - Display-Contrast
6 - Peristaltic Pump meta 50 7 40 00
7 - Peristaltic Pump
8 - WASH-Valve
50 7 40 00
F200.25
9 - Sampleport-Insert 1-8.106
10- Liquid-Valves for Cal.Solutions 50 8 30 20
11- Suction-Valve
12- Sampleport-Valve
13- Input-Valve
14- Sample Light-Barrier
Order no.
F200.25
F200.28
F200.25
40 7 10 20
15- Thermostat with Sensors
16- Hb-Sensor type II
17- Sensorcatch
50 2 10 01
50 7 20 10
18- Panel for therm. illumination 50 7 20 21
19- Metabolite valve
20-Metabolite unit
F200.25
Service- Manual - Version 4.01
combiline
System Description
Opened Front
1 - Switching power supply
2 - Thermoprinter GPT-4352-60
3 - Intermediate Vessel for suction
4 - Dampness Absorber
5 - Interface RS232 with SD-Card Reader
6 - Pinch Valve VE1 and VE2
7 - Filling Level Control board CL-FÜLLSTAND
8 - Suction Pump
Order no.
60 8 10 90
60 8 20 90
40 7 10 50
40 7 10 50
60 8 20 40
F200.25
40 8 20 00
F10.19
combiline
Service- Manual - Version 4.01
System Description
Opened Rear Wall
11 - Switching Power Supply
12 - Intermediate Vessel
13 - Thermoprinter GPT-4352-60
14 - Analogboard ESW-ANALOG
Order no.
60 8 10 90
40 7 10 50
60 8 20 90
60 8 10 xx / 60 8 11 xx
15 - Filling Level Control board CL-FÜLLSTAND 40 8 20 00
16 - Rollerpump Motor with gear
17 - Suction Pump
50 7 40 10
F10.19
Service- Manual - Version 4.01
combiline
System Description
Opened Rearwall, inside
CPU Board Z80ADC 1.2 and Power Supply ESW_MVTC
ESW_MVTC 2.0
60 8 10 80
1 - Potentiometer Rollerpump 2
2 - Potentiometer Rollerpump 1
3 - Buzzer
4 - Voltage input from Switching Power Supply
CPU CL Z80ADC 1.2
60 8 10 70
5 - L4 Connector for Analogboard ESW ANALOG
6 - Connector for DC-Supply (+5V)
7 - Driver ICs for Valve and Motor Control
8 - Printer Interface 9600 Baud
9 - Connector for RS232 Interface, 9600 Baud
10- Lithium Battery 3V Type BR2325
11- L11 LCD-Display
12- L10 Keyboard
13- Eprom with Analyser Software
14- L12 Filling Level board CL FÜLLSTAND
15 -RAM
16 - SD-Card Reader with RS232 Interface
combiline
Service- Manual - Version 4.01
Reagent-Transport
combiline BGA
n n pe ffe
o s o ma om nor str r rbe r bso A tef ess ch pn
Feu
Dam el ess
äß e-V gef
8 .2
R20 ed hen erm
Zw
Int
1 0.2
10 Rm
Service- Manual - Version 4.01
combiline
Reagent-Transport
combiline ISE
n n pe ffe
o s o ally rm om no str r rbe bso ter s A efil cht nes mp
Feu
Da el ess
äß
-V gef
8 0.2
ate hen
R2 edi erm
Zw
Int
1 0.2
10 Rm
1 0.2
10 Rm
1 0.2
10 Rm
combiline
Service- Manual - Version 4.01
Reagent-Transport
combiline BGA+E
n n pe ffe
o s o ally rm om no str er rb r bso te
A efil ess cht pn
Feu
Dam el
ß ess fä
-V
.28
ate nge he
R20 edi erm
Zw
Int
1 0.3
10 Rm
1 0.3
10 Rm
0 0.3
10 Rm
Service- Manual - Version 4.01
combiline
Reagent-Transport
combiline meta
n n pe ffe
o s o ma om nor str r rbe er bso filt
A te ess uch pn
Fe
Dam se
ß
Ves fä
8 ge
.2
tehen
R20 ed erm
Zw
Int
1 0.3
10 Rm
1 0.3
10 Rm
0 0.3
10 Rm
combiline
Service- Manual - Version 4.01
Valves
Determination of Valves
EV
BGA1
BGA2
Input-valve, module holder (one stage pinch valve, normally open)
Liquid-valve for calibration solution BGA 1 (normally closed) 1
Liquid-valve for calibration solution BGA 2 (normally closed) 1
LV
ABS
META
VE1
VE2
PV
BGA 3 Liquid-valve for calibration solution BGA 3 (normally closed) 2
BGA 4 Liquid-valve for calibration solution BGA 4 (normally closed) 2
CAL 3 Liquid-valve for calibration solution CAL 3 (normally closed) 2
CAL 4 Liquid-valve for calibration solution CAL 4 (normally closed) 2
Liquid-valve for ventilation (normally closed)
Suction-valve (one stage pinch valve, normally open)
Valve for triggering of the metabolite sensors 3
Ventilation-valve, intermediate vessel (one stage pinch valve, norm. open)
Ventilation-valve, suction pump (one stage pinch valve, normally open)
Sample port valve (one stage pinch valve, normally open)
SP Rinsing solution valve (one stage pinch valve, normally closed)
1
2
3 not available in combiline with electrolytes or metabolites not available in combiline BGA available in combiline meta only
Service- Manual - Version 4.01
combiline
Blockdiagram rs er ve ib dri
Tre
D LC T IN PR Z80 OL CO R IE LT PE
OG AL AN
combiline ly pp su erow il
A te
5-2 g p etz
0-4 hin altn
11 itc sw
Sch
PM
Service- Manual - Version 4.01
Switching Power-Supply
Type: PM110-45-2A
+5V
±12V
+24V
Supply for Logic (with Protection Diode1N5630A): Z80 ADC, Level Control board CL Füllstand,
LCD Converter
Analogboard ESW- ANALOG, CL Z80 ADC Multiplexer
Printer, Valves, Suction-Pump, Thermostat-Heating
Power Supply für Pumps, Valves and Heating
Type: ESW-MVTC
Rollerpump 1
+6,2V
+11,5V
(E slow) Rollerpump 1 - slow speed (Testpoint MOTOR)
(E fast) Rollerpump 1 - fast speed (Testpoint MOTOR)
Rollerpump 2 (available in meta version only)
+6,0V (E slow) Rollerpump 2 - slow speed (Testpoint MOTOR_META)
+8,0V (E fast) Rollerpump 2 - fast speed (Testpoint MOTOR_META)
Service- Manual - Version 4.01
combiline
Power-Supply ESW-MVTC
Voltage Supply for Pumps and Valves
combiline
Service- Manual - Version 4.01
Power-Supply ESW-MVTC
Voltage Supply for Pumps and Valves
Service- Manual - Version 4.01
combiline
Power-Supply ESW-MVTC
Adjustment of the operating voltage for Roller pump RP
Preparations:
1. Switch off the analyser.
2. Open the rear wall.
3. Have a voltmeter ready
4. The board ESW-MVTC is located at the inner side of the rear wall
Adjustment of the roller pump 1:
1. Switch the equipment on again.
2. Press Program key and Select menu SERVICE-TEST.
3. Activate the program 9 REAGENT TEST.
4. Connect the voltmeter to the test-points MOTOR (+) and GND1 (ground).
5. Press key „3“ (calibration solution) to run the roller pump 1 with slow speed.
6. Set the voltage at test-point MOTOR to 6.2V using potentiometer TR_SLOW.
7. Press key „2“ (WASH solution) to run the roller pump 1 with high speed.
8. Set the voltage at test-point MOTOR to 11.5V using potentiometer TR_FAST.
Continue with point 14 if second roller-pump is not available.
Adjustment of the roller pump 2:
9. Connect the voltmeter to the test-points MOTOR_META (+) and GND1 (ground).
10. Press key „8“ (Meta) to run the roller pump 2 with slow speed.
11. Set the voltage at test-point MOTOR_META to 6.2V ±0.3V using potentiometer TRM_SLOW.
12. Press key „7“ (Gas) to run the roller pump 2 with high speed.
13. Set the voltage at test-point MOTOR_META to 8.0V ±0.3V using potentiometer TRM_FAST.
14. Switch off the instrument, disconnect Voltmeter
15. Close the rear wall
16. Switch analyser on again
Remark:
The operating voltage for the roller pump has to be checked and adjusted only when an intact, good-as-new roller pump tube is used.
combiline
Service- Manual - Version 4.01
Power-Supply ESW-MVTC
Circuit Diagram Temperature-Regulation
Service- Manual - Version 4.01
combiline
Power-Supply ESW-MVTC
Temperature-Regulation
Description:
The Thermostate-Temperature Regulation is located on the power supply board ESW-MVTC
Via the plug connection V1 the low voltage 24 V~/4.5A is supplied. Furthermore the external elements for heavy-duty heating resistors as well as the temperature sensor (both in thermostats) are connected.
Exchange of Temperature regulation board (ESW-MVTC) or the NTC
After exchange of one of these components, the temperature, indicated on the display, is adjusted to 37°C with potentiometer T-CAL. The PIC IC must not be exchanged. Otherwise a re-adjustment of the circuit is required. For this adjustment special equipment is required.
Thermostat temperature correction on-site:
There should be made no adjustments without the required auxiliaries. Adjustments should not be done without contacting Eschweiler Service department!
LED for Temperature control indication:
A yellow LED for Temperature indication is located at the analyser’s front, below the reagent-wall. This LED is an indicator for temperature regulation status.
The more the actual thermostat temperature reaches to the nominal value (37°C), the longer the LED is
“ON”.
The LED is permanently “ON” when the temperature is in a range of ±0.5°C from 37°C.
The LED is blinking when the temperature deviation from 37°C is more than ±0.5°C.
The higher the blinking frequency, the higher is the temperature difference (e.g. Warm-up Phase).
In case of defective NTC-sensor or released temperature safety switch, the LED just blinks very short.
The LED is permanently on when the heating system is functioning.
combiline
Service- Manual - Version 4.01
Power-Supply ESW-MVTC
Metabolite Temperature Regulation, Circuit Diagram
Service- Manual - Version 4.01
combiline
Analog-Board ESW ANALOG 1_01
Layout
combiline
Service- Manual - Version 4.01
Analog-Board ESW ANALOG 1_01
Circuit-Diagram for Sensor Amplifiers
Service- Manual - Version 4.01
combiline
Analog-Board ESW ANALOG 1_01
Adjustment of polarization voltage
Preparations:
1. Switch off the equipment
2. Open the rear cover of the instrument
3. The board ESW ANALOG is located at the rear-side of the reagent wall
Adjustment of polarization voltage:
1. Switch the equipment on again
2. Connect a voltmeter to the testpoint UPOL ant pos.B and Ground
3. Adjust the indicated voltage to 0.700V with the potentiometer POL at pos. A
4. Switch the equipment off
5. Close the analogue box, the reagent-wall and the frot cover again
Remark:
The polarization voltage is connected to the platinum cathode of the pO2 sensor. It is set to 0.700V at the factory
Adjustment of O
2
zero point
1. Activate Testprogram 5, GAS, in the SERVICE menue and remove the plug from the pO
2
-sensor
2. Adjust the voltage, indicated on the screen, to -1.700V with potentimeter ZERO at pos. C
3. Connect sensor cable to pO2 sensor, stop the Testprogram and start a calibration
4. Repeat this adjustment after approx. 30 min. operation time!
combiline
Service- Manual - Version 4.01
Analog-Board ESW ANALOG 1_01
Circuit-Diagram for BP, TEMP, Sample-Light-Barrier, Hb
Service- Manual - Version 4.01
combiline
Analog-Board ESW ANALOG 1_01
Adjustment and Testing of the Barometric Pressure indication
1. Find out the actual atmospheric pressure (Barometric Pressure)
2.
Press PROGRAM key and select the menu SERVICE-TEST
3. Select Program 6 AIR / TEMP
4. The atmospheric pressure measured by the internal air-pressure sensor is indicated as follows:
AIRPRESSURE 760 MMHG (example!)
5. Adjust the indicated value to the actual Barometric Pressure with the potentiometer T PS at pos. D
Remark:
A testpoint for measurement of the output voltage of the Barometric Pressure sensor is located at pos. E
combiline
Service- Manual - Version 4.01
Analog-Board ESW ANALOG 1_01
Adjustment of thermostate-temperature indication-circuit
Preparation:
1. Switch OFF the analyser
2. Open the front cover of the instrument
3. The board ESW ANALOG is located at the rear side of the reagentwall
The adjustment can be made with a suitable PT100-Transmitter or alternatively with 2 calibrated resistors. The adjustment is made for the value +20°C with a resistance of 107.80 Ohm and for +40ºC with a resistance of
115.54 Ohm. At the transmitter and the resistors the same lengths of cable must be used as on the built-in temperature sensor.
Adjustment procedure:
1. Switch the analyser ON again
2. Press PROGRAM key, and Select menu SERVICE-TEST.
3. Then activate the program 6 - AIR / TEMP
4. Disconnect the temperature sensor at plug PT100 at pos. H
5. Set +20ºC at the PT100-Transmitter and connect it at pin1 and pin3 of plug M2. Alternatively, connect a precision resistance* of 107.8 Ohm.
6. Set the temperature value in the display to 20.0°C with potentiometer T_ZERO at pos. G
7. Set +40ºC at the transmitter or connect a precision resistance* of 115.54 Ohm.
8. Set the temperature value in the display to 40.0ºC with Potentiometer T_GAIN at pos F
* Corresponding resistors with suitable plugs are available at ESCHWEILER
It may be necessary to repeat this procedure several times until both values are reached.
Service- Manual - Version 4.01
combiline
Analog-Board ESW ANALOG 1_01
Adjustment of the Sample Light-Barrier
The Sample Light Barrier can be adjusted at potentiometer TR_LB at pos. K on the analog board ESW
ANALOG
For Adjustment:
1. Open the Reagentwall and move it to the front
2. Press OPERATION key, and select menue SERVICE-TEST
3. Then activate the program 7, LIGHTBARRIER
4. Adjust the voltage, indicated for air, with potentiometer TR_LB at pos. K to a value between 2,000 V -
2.300V
4. Press key TEST (WASH solution is filled into the measuring cell). The indicated voltage must be between
0.300V and 0.500V!
Should it not be posssible to reach the a.m. voltage values, an exchange of the Sample Light Barrier or its capillary might be required. For instructions for LB-exchange please note next page.
combiline
Service- Manual - Version 4.01
Sample-Light-Barrier LB
Capillary exchange in Sample Light-Barrier LB
By the time, the sensitivity of the sample-light-barrier can decrease. That means samples as well as calibration solutions are not recognized by the system. This leads to filling level indication for calibration solutions or even to permanent wash-cycles. Samples can be washed out without beeing measured.
The loss of sensitivity is mainly caused by deposits on the inner surface of the light-barrier-capillary. If an adjustment as described on previous page is not possible, the capillary needs to be replaced. For replacement proceed as follows:
- Activate Reagenttransport Program in menu SERVICE
- Remove sensors on left side (figure 1).
- Remove tube from LB-capillary (a)
- Put the exchange tool on the LB-capillary as shown in figure 2.
- Push the capillary out of the light-barrier as shown in figure 3.
- Take a new LB-capillary, place it on the tip of the exchange tool and push it left into the light-barrier as shown in figure 4.
- Reinstall sensors, connect tube to new LB-capillary and Quit Reagenttransport Program.
After replacement, check and adjust light-barrier voltage as described on previous page.
Service- Manual - Version 4.01
a figure 1 figure 2 figure 3 figure 4
combiline
Analog-Board ESW ANALOG 1_01 Hb
Measure method
The Hb-sensor is based on the combination of reflexion (scattered) and Transmission measurement of infrared light.
The data processing of Hb-values in the range of 6 - 30g/dl is done by the Reflexion-measurement method.
The data processing of Hb-values in the range of 3 - 6g/dl is done by the Transmission-measurement method.
The Hb-measure unit is located on the analog amplifier board ESW-ANALOG 1_01.
The following has to be noticed:
- The use of Hb-calibration is not required, only the input of two parameters forrefexion and Transmission
(HB-R and HB-T) is necessary.
- After setting the Combiline into operation, sensor related parameters have to be entered by keyboard. After that the Hb-sensor is active for measurement. These parameters (HB-R and HB-T) are printed on a white flag at the Hb-sensor cable.
- The sample volume required for a fully equipped Combiline BGA+E+HB is min. 210 - 240µl. The sample volume required for a combiline BGA+HB is approx. 160µl. So it is of particular relevance to use capillaries with adequate volume.
- Please note that this measure method is suitable for measurement of whole-blood only! Quality controls are not usable!
- The measure range of the Hb-sensor is 3 - 30g/dl
- Hb will be measured if the sample light-barrier detects a whole-blood sample. Otherwise the programmed standard Hb-value is displayed (e.g. 15g/dl) and used for the calculation of the acid-base parameters.
- In case of too less sample volume, a “Transmission-voltage” of -4.000V is detected. The displayed Hbvalue will be 0.0g/dl then. This can be observed on a sensor-parameter printout which can be made after a sample measurement (see attached example of printout)
combiline
Service- Manual - Version 4.01
Analog-Board ESW ANALOG 1_01 Hb
Activation of Hb-Sensor type II
Carry out of an Hb-Sensor activation in STAND-BY mode
In order to maintain the measuring accuracy for determination of Hb and to activate the Hb-sensor, an input of two factors HB-R and HB-T has to be carried out in STAND-BY mode
In case the COMBILINE was switched off for less than 24 hours the Hb-factors are retained in the memory and the sensor remains ACTIVE.
In case the COMBILINE was switched off for more than 24 hours the Hb-factors have to be entered as described below, the Hb-sensor is marked with OFF.
The mentioned Hb-factors are printed on a white flag at the Hb-sensor cable (see attached photo)
Carry out Hb-factor input to activate the Hb-sensor:
• Press the 5-key in OPERATION MENU to perform the Hb-activation process, the following dialogue appears:
HB-SENSOR
HB R : 0.000
HB T :
• Enter the HB-R value given on the white ensign at the Hb-sensor by keypad e.g. 0.970 and press the
Enter-key to confirm.
The following dialogue appears:
HB-SENSOR
HB R : 0.970
HB T : 0.000
• Enter the HB-T value given on the white ensign at the Hb-sensor by keypad e.g. 0.517 and press the
Enter-key to confirm.
• The input range for HB-R and HB-T is 0.300 to 1.800
For quality control of the Hb-Sensor, whole blood specimens whose Hb-concentration have already been determined with the help of suitable measuring methods might be used (e.g. oximeter).
Service- Manual - Version 4.01
combiline
Analog-Board ESW ANALOG 1_01
HB - SENSOR
HBR / HBT 0.970 0.517
MESS. [mV] -1182 -320 printout of Hb-sensor parameter
HB/R = Reflexion factor
HB/T = Transmission factor
MESS = Hb-sensor voltages for Reflexion and Transmission measurement of the last sample measurement (in millivolt)
Hb
Hb-factors HB-R and HB-T
Hb-sensor type II
Hb-sensor type II installed in combiline (optional)
combiline
Service- Manual - Version 4.01
CPU Board CL Z80 ADC12
Layout
Service- Manual - Version 4.01
combiline
CPU Board CL Z80 ADC12
Circuitdiagram 1/3
a a
F
F
T
0n
0n uF
F
F
F
T
F
F
F
0n
0n
0n uF
0n
0n
0n
F 0n
10
10
10
10
10
10
10
10
10
10
10
1 C1
C9
C8
3
0
4
C1
C1
C1
C7
C6
C5
C4
C3
combiline
E AC RF TE IN HB ste ce nitt rfa ch te
S se se ria rie r r nte cke pri
Dru
Service- Manual - Version 4.01
CPU Board CL Z80 ADC12
Circuitdiagram 2/3
Service- Manual - Version 4.01
combiline
CPU Board CL Z80 ADC12
Circuitdiagram 3/3
22
20
18
16
26
24
14
12
10
8
23
21
25
19
17
13
11
15
4
6
2
5
3
1
9
7
7
6
5
4
9
8
3
2
1
IN
IN
IN
IN
IN
IN
IN
IN
IN
10
IN
IN
IN
IN
IN
IN
IN
16
15
14
13
12
11
8
9
7
10
11
26
25
5
6
4
23
22
24
21
20
19
combiline
Service- Manual - Version 4.01
CPU Board CL Z80 ADC12
Specification
AD-conversion: (2 x 8 Bit parallel) 12 Bit AD-converter, ICL7109 int., range -4.096V to +4.096V, resolution
1mV. 16 channel multiplexer, HI506
Driver Output (24 pieces) 3 x 8 Bit outputs, with D-Latch (74HC273), Optocoupler (galvanic insulation) and driver circuits for loads up to 1A (ULN2803)
Display:
Level control:
Graphic-display NLC240x128BTGLE at L11 via ECODISP
(1 x 8 Bit) 8 Input-Ports for checking the filling levels of reagents via L12 CL FÜLLSTAND
Serial interface: 1. External interface RS-232 SD INT at L2
2. Thermo printer at L3
EPROM: 64KB Program (27C512)
RAM: 8KB (8K x 8)
Keyboard: Two parallel Ports at L10
Reset-component: battery operation, automatic reset generation, switch over to battery operation
4 Timer and counter channels: Baudrate generation (2 channels) Timer (multitasking software modul)
Remark:
An automatic reset is generated if the operation voltage of the CPU board becomes lower than 4.75V or higher than 5.25V.
Service- Manual - Version 4.01
combiline
Serial Interface RS-232 SD INT 1.4
Serial interface with SD-Card Reader RS-232 SD INT 1.4
Technical Data:
Baudrate: 9600 bit/sec
Databit: 8
Parity:
Startbit: no
1 (STX = 02HEX))
Stopbit: 1 (ETX = 03HEX) no handshake (unidirectional) connector: terminal RS232 at the rear panel of the instrument
Description:
All data which is sent to the printer after completition of a sample measurement, will also be transmitted to the RS232 interface at the same time together with CR (carriage return) and LF (line feed) at the end of the line. The transmission is unidirectional, there is no dialog between COMBILINE and PC. The data has to be received in a polling mode after starting a measurement. The data is given out in standard ASCII code.
Data recalled from the memory (SD-Card) will not be sent via RS232 !
The data transmission is unidirectional only.
Remark:
The data given by the interface can be detected by any terminal program for example TERMINAL.
combiline
Service- Manual - Version 4.01
Serial Interface RS-232 SD INT 1.4
Printout:
Example of print out of measurering results with corresponding data output
* see next page for description
Data output
27,-24.20
28,-17.1
29,4.4
30,171.1
31,22.81
32,1.8
33,0.25
34,7.373
35,0.00
36,0.00
37,760
38,101.3
65
0
11,45.0
12,42.4
13,1.36
14,126
15,37.0
16,20.9
17,0.85
18,1.0
19,13.3
20,-19.0
21,-21.5
22,1.1
23,29.0
24,99.4
25,20.5
26,27.67
(position no.,Value)
STX
80,CL3000
81,0945250808
82,09:45,25.08.08
0,171.1
1,1.8
2,7.373
3,5.9
4,104
5,1.38
6,141
7,0.92
8,0.0
9,0.0
10,15.0
255
ETX
Service- Manual - Version 4.01
Meaning
(Parameter
Start of Text
Type of analyser
Sample number
Time and Date pO
2 pCO
2 pH
K +
Na
Ca
Cl
Li +
GLU
LAC
Hb
-
+
++
(QC) unit)
(02HEX) mmHg or kPa mmHg or kPa mmol/l mmol/l mmol/l mmol/l mmol/l mmol/l mmol/l g/dl
HCT
H +
% nmol/l
Ca ++ 7.4
mmol/l
Cl - (Blood / Serum) mmol/l
Temperature
FIO
2
RQ
°C
%
HCO3A
HCO3S
BE
SBE mmol/l mmol/l mmol/l mmol/l
TCO
2
BB
O
2
O
2
SAT
-CT
P50 mmol/l mmol/l
%
%
AADO
A-GAP
2
SHUNT mmHg mmHg mmol/l
% pO
2 pO
2
, temp. correct. mmHg
, temp. correct. kPa pCO
2 pCO
2
, temp. correct. mmHg
, temp. correct. kPa pH, temp corrected
GLU
LAC mg/dl mg/dl
Barometric Pressure mmHg
Barometric Pressure kPa
Sample type
Timeouts
(MESART)*
(ADCMSK)*
Active Sensors (MESMSK)*
End of Text (03HEX)
combiline
Serial Interface RS-232 SD INT 1.4
Meaning of the last three Values of a data output:
1. Value - MESART= kind of measurement
2. Value - ADCMSK = sensors which exceeded the permissible time for measurement (measure-value with ?)
3. Value - MESMSK = sensors which were ACTIVE during measurement
All values are available in ASCII-Format as hexadecimal value. The value range is from 0 to 32768.
Explanation:
Bit Valence
00 1
01 2
02 4
03 8
04 16
05 32
06 64
07 128
08 256
09 512
10 1024
11 2048
12 4096
13 8192
14 16384
15 32768 -
-
-
-
-
-
MESART
Syringe
Capillary
BGA only
ISE only
QC
K
NA
Respiratory Gas CA
Serum, Plasma CL
ADCMSK, MESMSK
PO2
PCO2
PH
-
-
Blood
-
LI
HB
-
-
-
-
-
-
combiline
Service- Manual - Version 4.01
Serial Interface RS-232 SD INT 1.4
Example for MESART:
65 64 + 1
82 64 + 16 + 2
133 128 + 4 + 1
Example for ADCMSK:
0 0
3 2 + 1
Examples for MESMSK:
255 128+64+32+16+8+4+2+1
7 4 + 2 + 1
Serum, Plasma ; Syringe
Serum, Plasma ; QC ; Capillary
Blood ; BGA only ; Syringe
All Sensors comply with permissible measure times pO
2
+ pCO
2
did not comply with permissible measure times pO
2
, pCO
2
, pH, K + , Na + , Ca ++ , Cl , Li + were ACTIVE during measurement pO
2
+ pCO
2
+ PH were ACTIVE during measurement
Service- Manual - Version 4.01
combiline
Serial Interface RS-232 SD INT 1.4
Layout
RS 232 Interface with SD Card-Reader
combiline
Service- Manual - Version 4.01
Serial Interface RS-232 SD INT 1.4
Circuit Diagram
Service- Manual - Version 4.01
combiline
Serial Interface RS-232 SD INT 1.4
Interface Cable
combiline
Service- Manual - Version 4.01
Fill Level Control CL FÜLLSTAND
Description:
Via L2 the waste bottle sensor is connected. The control of the Calibrants is made by the Sample Light
Barrier in the Sensorholder (Thermostat).
Via L1 the voltage supply (+5V / GND digital) and the outputs of the comparators are connected. The switch thresholds of the light barrier LS1 for WASH is adjusted with potentiometer TR1.
The light barriers must fulfil the following conditions:
1. Rinsing solution/calibration solution in tube Umax = 0.300 V
2. Air in tube Umin = 1.000 V
The outputs of the comparators supply the following signals: empty full rinsing solution T8 0 V +5 V
Push bar BGA 1/3 T4
Push bar BGA 2/4 T3 waste bottle T5
0 V
0 V
+5
+5 V (released / pressed)
+5 V (released / pressed)
0 V
Check of the waste bottle control
1. Reference voltage at
2. Sensors of waste bottle: in air, in tap water,
T6 = 2.500 V
T7 approx. 3.500 V
T7 approx. 2.000 V in distilled water T7 approx. 2.400 V
For testpoints (T...) see circuit diagram CL FÜLLSTAND
Transport of calibration solutions CAL3, CAL4, BGA1, BGA2, BGA3 and BGA4 is controled by the Sample
Light Barrier during calibration procedure!
Service- Manual - Version 4.01
combiline
Fill Level Control CL FÜLLSTAND
Layout
Filling-Level Control Board CL-FÜLLSTAND
combiline
Service- Manual - Version 4.01
Fill Level Control CL FÜLLSTAND
Circuit Diagram
Service- Manual - Version 4.01
combiline
Keyboard
Specification
The keyboard is operated directly at a parallel 8-bit port. Debouncing of the contacts takes place by by a progranmable Controller (PIC)
The PIC scans the keyboard-matrix permanently. If a key is pressed, the corresponding value is sent to the port. By an interrupt the information is interpreted by the analyser software. The corresponding action will be performed then.
Cleaning
Use a suitable cleaning solution as described in chapter !2 Maintenance and Hygiene and pay attention to the notes.
Keyboard Adaptor CL-TASTATUR
TASTATUR
ENTER
F1
1
2
3
F2
4
5
6 .
F4
C
0
F3
7
8
9
3
2
1
5
4
L2
7
6
9
8
AR1
4*10K
Q1
4.0 MHz
+5V
C1
100nF
1
TOCKI
2
+U
6
RA0
7
RA1
8
RA2
9
RA3
IC1
28
MCLR
PIC 16C57C
10
RB0
11
RB1
12
RB2
13
14
RB3
RB4
RC0
18
19
RC1
20
RC2
21
RC3
22
RC4
23
RC5
24
RC6
25
RC7
27
OSC1
26
OSC2
GND
4
13
11
9
7
25
26
17
15
5
3
1,2
L1
L10
Z80ADC
ESCHWEILER CombiLine
Tastaturadapter CL TASTATUR Rev. 1.2
CL TASTATUR 12.tcd / 01.02.2010
Ident-No. 3031
TASTATURADAPTER CL TASTATUR Rev. 1.2
SCHALTBILD UND BESTÜCKUNGSPLAN / CIRCUIT DIAGRAM AND LAYOUT
combiline
ESCHWEILER GmbH & CO. KG KIEL GERMANY
DATUM / DATE : 01.02.2010
BEZ.: CL TASTATUR 12.TCD
IDENT Nr.: 3031
BLATT / PAGE : 1 / 1
Keyboard Adaptor, Circuit Diagram and Layout
Service- Manual - Version 4.01
LCD-Display
Specification
Type: NLC240x128BTGLE
15 lines LCD-Display 30 characters / line
Contrast Adjustment
The trim potentiometer for contrast adjustment is located above the roller pump, as shown in figure 42a.
• Turn the potentiometer knob until the contrast of the display is pleasant for the operator.
Contrast Potentiometer for LCD-Display
Display contrast adjustment
Service- Manual - Version 4.01
combiline
LCD-Converter
Circuit-Diagram and Layout
combiline
Service- Manual - Version 4.01
Thermoprinter GPT-4352-60
Pin configuration for Printer GPT-4352-60
Voltage supply plug:
Pin 1: GND
Pin 2:GND
Pin 3:GND
Pin 4:+24V
Pin 5:CTS (output)
Paper specifications:
Figure 44 printer board
Interface (TTL/ 4,5V Pegel ):
Pin 1:GND
Pin 2:TXD (input)
Pin 3:RXD (output)
Pin 4:RTS (input)
Pin 5:CTS (output)
Roll diameter:
Roll width: up to 60mm
57mm
Paper thickness: 0.04mm
Key-Functions
FEED / ENTER key (1)
Printer self test:
Press key 1 then switch analyser ON.
Keep key 1 pressed for at least 3 seconds.
The printer software version and the character set will be printed
Paper feed will be performed during normal operation
1 LED
Service- Manual - Version 4.01
combiline
Liquid-Valve SMC
Description
Liquids will be sucked in by the control of Liquid Valves. 5 of them are sampled together to an array back beside of the Sample Port inside the housing.
The Liquid Valves are controlled by the NT-Move board.
Following liquids are controlled by the Liquid Valves (depending onconfiguration)
BGA 1 for Calibration Solution BGA 1
BGA 2 for Calibration Solution BGA 2
BGA 3 for Calibration Solution BGA 3
BGA 4 for Calibration Solution BGA 4
CAL 3 for Calibration solution CAL 3
CAL 4 for Calibration solution CAL 4 (+M)
LV for Air
Sedimentation
After a longer period of use, or the COMBILINELayout has been parked for a longer time, it can be possible that a sedimentation on the seal can cause a leakage.
Error evidence:
Valve does not respond of commands: Check magnetic coil with a digital multimeter. In case of trouble, replace valve.
In case of sedimentations inside of the valve, the liquid cannot be transported properly. The valve must be cleaned inside.
Replacement of a Liquid Valve
Order no.:
40 8 20 10 Liquid-valve block SMC complete, for COMBILINE BGA+E and meta versions
40 8 20 15 Liquid-valve block SMC complete, for COMBILINE BGA
40 8 20 20 Liquid-valve block SMC complete, for COMBILINE ISE
Order no. of a single valve:
50 8 30 20 Liquid-valve SMC
• Proceed as follows:
• Power OFF COMBILINE and disconnect it from Mains.
• Unscrew the three slotted-head retaining screws at the right side of the reagent wall.
• Remove all tubes from the valves.
Each of the valves is assembled together with two screws.
• Unscrew the two two for dismantling.
• Replace the valve to be replaced.
• Plug in the cables at the new valve (don't misplace cable!)
• Assemble the valve group in reverse order.
continued next page
• Power ON COMBILINE and open REAGENT TRANSPORT in SERVICE menu
• Ventilate the replaced Valve by pressing the corresponding function-key
combiline
Service- Manual - Version 4.01
Liquid-Valve SMC
Cleaning of a Liquid Valve
Proceed as described as follows:
• Dismantle the Valve Group.
• Dismantle the Valve as described above
• Clean the internal parts with suitable cleaning solution.
• For assembling of the valve proceed in reverse order.
• For reassembling of the Valve Group proceed as described b
Liquid Valve - SMC Type
Liquid Valve SMC Type - Fluidics
Service- Manual - Version 4.01
air
CAL3
CAL4
BGA3/1
BGA4/2 outlet
combiline
Ventilation-Valve VE1
Delay-circuit, Layout and Circuit-Diagram
The delay-circuit CL ASV is mounted on the ventilation valve VE2
combiline
Service- Manual - Version 4.01
Suction-Pump
Replacement of the Suction Pump
The Suction Pump is mounted with an aluminium angle beside the transformer and is controlled by the Main Board M1.
For a replacement proceed as follows:
• Switch OFF COMBILINE and disconnect it from Mains.
• Unscrew the two retaining screws on the aluminium angle.
• Remove angle from pump.
• Exchange the two cables to the new motor (don't misplace cables!).
• Exchange the two tubes (don't misplace in/out).
• Assemble the new pump with aluminium angle to its place.
• Switch ON COMBILINE.
• Open the dialogue REAGENTTRANSPORT in SERVICE-TEST menu
• Press 1-key for SUCTION to check its functioning.
Cleaning of Pump's Head
The internal parts of the pump could properly be contaminated with biological material. There is a risk of infection!
• Wear gloves!
• Uninstall the Suction Pump as described in the chapter above.
• Remove lid (B) from pump head, see figure 49 on next page.Liquid-Valve
• Remove the two screws (A) and the front panel (don't loose centre ring for motor axis).
• Loosen the allen screw (D) at motor axis.
• Remove both retaining screws for motor and pull pump head from motor axis.
• Clean the corresponding internal parts with a suitable solution, see chapter
!2 Hygiene.
Attention: There is no spare part available!
• Proceed in reverse order to install the pump head.
• Check the functioning as described in the above.
Service- Manual - Version 4.01
combiline
Suction-Pump
C
G
D
ASF
E
A B
Figure 49 Suction Pump explored
F
combiline
Service- Manual - Version 4.01
ESCHWEILER
automatic analysing systems
Sensor Spare Parts and average lifetime information
Ref.
1
Haltbarkeit
(Monate) durability
(month)
12 - 36
pO
2
Haltbarkeit
(Monate) durability
(month)
pCO
2
Ersatzteil-Liste für BGA-Sensoren spare part list for BGA sensors
Haltbarkeit
(Monate) durability
(month)
pH
Haltbarkeit
(Monate) durability
(month)
1 1
2
3
4
12 - 36
2
3
4
3 - 6
12 - 48
2
3
3 - 6 4
12 - 24
1
3
2
12 - 24
5
6
12 - 48
5
6
12 - 48
Referenzsensor kompl.
reference sensor compl.
50 5 10 00
1
2
3
4
5
6
Ref. Ableitsystem ref. conduct. system
50 5 10 10
Schraubkopf screw cap
50 1 10 21
Ref.-Aufsatz ref.- extension
50 5 10 50
Hülse mit Membran membrane shell
50 5 10 30
Sensorgehäuse sensor casing
50 5 10 20
O-Ring 1.8 x 1.0
o-ring 1.8 x 1.0
50 1 10 25
1 pO
2 pO
2
Sensor kompl.
sensor compl.
50 1 10 00
2
3
4
5
6
Schraubkopf screw cap
50 1 10 21
Sensoreinheit sensor unit
50 1 10 10
O-Ring 8.0 x 1.0
o-ring 8.0 x 1.0
50 1 10 50
Hülse mit Membran membrane shell
50 1 10 30
Sensorgehäuse sensor casing
50 1 10 20
O-Ring 1.8 x 1.0
o-ring 1.8 x 1.0
50 1 10 25
5
6
12 - 48
1 pCO
2 pCO
2
Sensor kompl.
sensor compl.
50 1 20 00
2
3
4
5
6
Schraubkopf screw cap
50 1 10 21
Sensoreinheit sensor unit
50 1 20 10
O-Ring 8.0 x 1.0
o-ring 8.0 x 1.0
50 1 10 15
Hülse mit Membran membrane shell
50 1 20 30
Sensorgehäuse sensor casing
50 1 20 20
O-Ring 1.8 x 1.0
o-ring 1.8 x 1.0
50 1 10 25
3
4
12 - 24
1 pH Sensor kompl.
pH sensor compl.
50 1 30 00
2
3
4
Schraubkopf screw cap
50 1 10 21
Sensoreinheit sensor unit
50 1 30 10
Sensorgehäuse sensor casing
50 1 30 20
O-Ring 1.8 x 1.0
o-ring 1.8 x 1.0
50 1 10 25
BGA_combiline/15012009HEU
Service- Manual - Version 4.01
combiline
Sensor Spare Parts
Sensor Spare Parts and average lifetime information
combiline
Na
Haltbarkeit
(Monate) durability
(month)
ISE
Haltbarkeit
(Monate) durability
(month)
1
12 - 36
1
2
12 - 36 2
K
+
3 - 6
Ca
++
Li
+
3 - 6
3 - 6
Cl- 3 - 6
Na
+
12-24
3 3
4
12 - 24
4
Na
Na
+
+
Sensor kompl.
sensor compl.
50 3 20 00
ISE Sensoren:
ISE sensors:
K
+
/ Ca
++
/ Cl- / Li
+
1
2
3
Ableitsystem conducting system
50 3 20 10
O-Ring 3.2 x 1.8
o-ring 3.2 x 1.8
50 3 20 15
Na
+
Na
+
Sensorgehäuse
sensor casing
50 3 20 20
K
+
K
+
Sensor kompl.
sensor compl.
50 3 10 00
Ca
Ca
++
++
Sensor kompl.
sensor compl.
50 3 30 00
Cl- Sensor kompl.
Cl- sensor compl.
50 3 40 00
4
Ersatzteile ISE Sensoren / spare parts ISE sensors K
+
/ Ca
++
/ Cl- / Li
+
1 Ableitsystem conducting system
K
+
50 3 10 10
Ca
++
50 3 30 10
Cl-
Li
+
50 3 40 10
50 3 50 10
3 Sensorgehäuse sensor casing
K
+
50 3 10 20
Ca
++
50 3 30 20
Cl- 50 3 40 20
Li
+
50 3 50 20
2
O-Ring 1.8 x 1.0
o-ring 1.8 x 1.0
50 1 10 25
Hülse mit Membran membrane shell
K
+
50 3 10 30
Ca
++
50 3 30 30
Cl-
Li
+
50 3 40 30
50 3 50 30
Li
Li
+
Sensor kompl.
+
sensor compl.
50 3 50 00
4 O-Ring 1.8 x 1.0
o-ring 1.8 x 1.0
50 1 10 25
Service- Manual - Version 4.01
Lifetime of Sensors
Ref.
pO
2 pCO
2 pH
Lebensdauer von Sensoren / Lifetime of sensors
Referenzsensor reference sensor
50 5 10 00
Ableitsystem conducting system
12 - 36 Monate
12 - 36 months
Membran membrane
6 - 9 Monate
6 - 9 months pO
2 pO
2
Sensor
sensor
50 1 10 00 pCO
2
Sensor pCO
2
sensor
50 1 20 00 pH Sensor pH sensor
50 1 30 00
Sensor-Einheit sensor unit
12 - 48 Monate
12 - 48 months
12 - 24 Monate
12 - 24 months
12 - 24 Monate
12 - 24 months
Membran membrane
3 - 6 Monate
3 - 6 months
3 - 6 Monate
3 - 6 months
K
Cl
Li
+
Na
Ca
+
-
+
++
K+ Sensor
K+ sensor
50 3 10 00
Na+ Sensor
Na+ sensor
50 3 20 00
Ca++ Sensor
Ca++ sensor
50 3 30 00
Cl- Sensor
Cl- sensor
50 3 40 00
Li+ Sensor
Li+ sensor
50 3 50 00
Ableitsystem 1 conducting system
12 - 36 Monate
12 - 36 months
1
12 - 36 Monate
12 - 36 months
12 - 36 Monate
12 - 36 months
12 - 36 Monate
12 - 36 months
12 - 36 Monate
12 - 36 months
Membran membrane
3 - 6 Monate
3 - 6 months
12 - 24 Monate
12 - 24 months
3 - 6 Monate
3 - 6 months
3 - 6 Monate
3 - 6 months
3 - 6 Monate
3 - 6 months
GLU
LAC
GLU Sensor
GLU sensor
50 4 20 00
LAC Sensor
LAC sensor
50 4 10 00
30 Tage oder ca. 1000 Proben
30 days or approx. 1000 samples
15 Tage oder ca. 400 Proben
15 days or approx. 400 samples
1
Das Ableitsystem der ISE-Sensoren kann bei Beschädigung der Chloridschicht werksseitig regeneriert werden.
In case of damaged chlorid layer the conducting system of ISE sensors can be regenerated by manufacturer.
The lifetimes given in the above table are average lifetimes
Service- Manual - Version 4.01
combiline
List of Consumeables
Article no.: description
Calibration- and Rinsing solutions
for combiline BGA
3-4.100
40 6 10 20
40 6 10 30
WASH 1 (Rinsing solution for BGA)
BGA 1 (Calibration solution for BGA / low)
BGA 2 (Calibration solution for BGA / high) packing/unit
6x330 ml
12x130 ml
12x130 ml
for combiline BGA+E
40 6 10 10
40 6 10 40
40 6 10 50
40 6 10 45
40 6 10 55
WASH 2 (Rinsing solution for BGA plus E)
BGA 3 (Calibration solution BGA plus E/low)
BGA 4 (Calibration solution BGA plus E/high)
CAL 3 (Calibration solution for BGA plus E/low)
CAL 4 (Calibration solution for BGA plus E/high) for COMBILINE meta
40 6 10 11
40 6 10 40
40 6 10 50
40 6 10 45
40 6 10 56
WASH 2+M Rinsing solution
BGA 3 Calibration solution BGA plus E/low
BGA 4 Calibration solution BGA plus E/high
CAL 3 Calibration solution for BGA plus E/low
CAL 4+M Calibration solution for BGA plus E/high
for combiline ISE
3-4.110
3-4.400
3-4.500
WASH 2 (Rinsing solution for ISE)
CAL 3 (Calibration solution for ISE/low)
CAL 4 (Calibration solution for ISE/high)
6x250 ml
12x130 ml
12x130 ml
12x150 ml
12x150 ml
8x250 ml
12x130 ml
12x130 ml
12x150 ml
12x150 ml
6x330 ml
6x330 ml
6x330 ml
Special Solutions
50 6 10 80
50 6 10 84
50 6 10 88
Protein remover
Electrode cleaner
Activating solution for pH-sensor
Quality Controls
Control Solutions for BGA
50 6 20 10
50 6 20 20
50 6 20 30
Level I BGA (acidosis)
Level II BGA (normal)
Level III BGA (alkalosis)
1 x 100 ml
50 ml
20 ml
30 ampoules
30 ampoules
30 ampoules
Control Solutions for BGA plus E
50 6 20 40
50 6 20 50
50 6 20 60
50 6 20 70
Level I BGA+E (acidosis)
Level II BGA+E (normal)
Level III BGA+E (alkalosis)
Level IV (high pO
2
)
Control Solutions for BGA plus E+Li+
50 6 20 45
50 6 20 55
50 6 20 65
Level I EGL (acidosis)
Level II EGL (normal)
Level III EGL (alkalosis)
30 ampoules
30 ampoules
30 ampoules
30 ampoules
30 ampoules
30 ampoules
30 ampoules
Sampling Systems
Capillaries
50 6 30 08
50 6 30 10
50 6 30 12
50 6 30 14
50 6 30 16
50 6 30 30
50 6 30 35
50 6 30 40 hep. capillaries 8 cm/lg hep. capillaries 10 cm/lg hep. capillaries 12 cm/lg hep. capillaries 14 cm/lg for instruments with Hb sensor
250 pcs.
250 pcs.
250 pcs.
250 pcs.
hep. capillaries 16 cm/lg for instr. with meta and Hb sensor 250 pcs.
metal stirrers for capillaries 250 pcs.
magnet for stirrer end-caps for capillaries 500 pcs.
combiline
Service- Manual - Version 4.01
List of Consumeables
Article no.: description
Aspiration Pipes
3-7.200
50 6 30 55 pipe for control solution, activating solution etc.
ampoule holder for aspiration pipe
Printer Paper
50 6 50 00
Cleaning Paste
50 1 10 41 printer paper cleaning paste (silicon carbide) packing/unit
10 pcs.
1 pc.
pack.=2 rolls
Accessories for Maintenance
40 6 40 00 service-set complete for COMBILINE without Metabolites consisting of:
40 6 40 10
40 6 40 15
40 6 40 20
40 6 40 25 set tubings for roller-pump (reagents) tubing for magnetic valve VE1 tubing for magnetic valve VE2 tubing for magnetic valve ABS
40 6 40 01 consisting of:
40 6 40 11
40 6 40 15
40 6 40 20
40 6 40 25 service-set complete for COMBILINE with Metabolites set tubings for roller-pump (reagents) tubing for magnetic valve VE1 tubing for magnetic valve VE2 tubing for magnetic valve ABS sample port insert (red) 1-8.106
Sensors pO
2
-Sensor
50 1 10 00
50 1 10 10
50 1 10 15
50 1 10 20
50 1 10 21
50 1 10 25
50 1 10 30
50 1 10 35
50 1 10 50 pO pO pO
2
2 gasket for Sensor-unit (8.0 x 1.0)
2
-Sensor complete
-Sensor-unit with membrane shell
-Sensor-casing with screw-cap screw cap for Sensor-casing gasket for Sensor-casing (1.8 x 1.0) pO pO
2
2
-membrane shells
-fill solution polishing kit for pO2 sensors
2 pcs.
1 pc.
1 pc.
1 pc.
2 pcs.
1 pc.
1 pc.
1 pc.
1 pc.
pack.= 5 pcs.
bot.=20 ml pCO
2
-Sensor
50 1 20 00
50 1 20 10
50 1 10 15
50 1 20 20
50 1 10 21
50 1 10 25
50 1 20 30
50 1 20 35 pH-Sensor
50 1 30 00
50 1 30 10
50 1 30 20
50 1 10 21
50 1 10 25 pCO pCO
2
2
-Sensor complete
-Sensor-unit with membrane shell gasket for Sensor-unit (8.0 x 1.0) pCO2-Sensor-casing screw cap for Sensor-casing gasket for Sensor-casing (1.8 x 1.0) pCO pCO
2
2
-membrane shells
-fillsolution pH-Sensor complete pH-Sensor-unit pH-Sensor-casing screw cap for Sensor-casing gasket for Sensor-casing (1.8 x 1.0) pack.=5 pcs.
bot.=20 ml
Service- Manual - Version 4.01
combiline
List of Consumeables
Article no.: description
Hb-Sensor
50 2 10 01 Hb-Sensor Type II complete
K + -Sensor
50 3 10 00
50 3 10 10
50 3 10 20
50 1 10 25
50 3 10 30
50 3 10 35
Na + -Sensor
50 3 20 00
50 3 20 10
50 3 20 15
50 3 20 20
50 1 10 25
50 3 20 35
Ca ++ -Sensor
50 3 30 00
50 3 30 10
50 3 30 20
50 1 10 25
50 3 30 30
50 3 30 35
Cl -Sensor
50 3 40 00
50 3 40 10
50 3 40 20
50 1 10 25
50 3 40 30
50 3 40 35
K + -Sensor complete
K + -conducting system
K + -Sensor casing gasket for Sensor-casing 1.8 x 1.0)
K + -membrane shell with membrane
K + -fillsolution (organic tenside, buffer substances)
Na + -Sensor complete
Na + -conducting system gasket for conducting system (3.2 x 1.8)
Na + -Sensor-casing gasket for Sensor-casing 1.8 x 1.0)
Na + -fillsolution
Ca ++ -Sensor complete
Ca ++ -conducting system
Ca ++ -Sensor casing gasket for Sensor-casing 1.8 x 1.0)
Ca ++ -membrane shell with membrane
Ca ++ - fillsolution
Cl -Sensor complete
Cl - conducting system
Cl -Sensor casing gasket for Sensor-casing 1.8 x 1.0)
Cl -membrane shell with membrane
Cl -fillsolution
Li + -Sensor
50 3 50 00
50 3 50 10
50 3 50 20
50 1 10 25
50 3 50 30
50 3 50 35
Li + -Sensor complete
Li + -conducting system
Li + -Sensor casing gasket for Sensor-casing 1.8 x 1.0)
Li + -membrane shell with membrane
Li + -fillsolution
Reference-Sensor
40 5 10 00
50 5 10 10
50 3 20 15
50 5 10 20
50 1 10 25
40 5 10 50
50 5 10 30
50 5 10 35 reference-sensor complete reference-conducting system gasket for conducting system (3.2 x 1.8) reference-Sensor casing gasket for Sensor-casing 1.8 x 1.0) reservoir for reference-Sensor casing ref.-membrane shell with membrane reference-fillsolution packing/unit bot.=20 ml btl.= bottle bot.= 20 ml bot.=20 ml bot.=20 ml bot.=20 ml bot.=50 ml
combiline
Service- Manual - Version 4.01
List of Spare Parts
Article no.:
System casing
60 7 10 00
60 7 10 10 description
system casing, complete
front cover complete, painted, with print
Sensor fixing device / Thermostate
50 7 20 00
40 7 10 30
50 7 20 10
40 7 10 20
40 7 10 25
50 7 20 20
50 7 20 21
50 7 20 05
50 7 20 90 sensor fixing device, complete device for specimen input closing mechanism sample Light Barrier COMBILINE capillary for sample light barrier board for thermostate illumination, (LED) panel for thermostate illumination (LED), complete temperature safety switch temperature regulation sensor NTC (for Thermostat or Meta block)
Sensor fixing device for Metabolites
50 7 20 60
50 7 20 70
50 7 20 75
50 7 20 80
50 7 20 85
50 7 20 95 thermostate block, for Metabolite sensors, complete fixing screw, white, for GLU-Sensor fixing screw, black, for LAC-Sensor holding device for GLU-Sensor holding device for LAC-Sensor peltier-element for Meta-block
Calibration-bag adapter
40 7 10 35
G 40.1
40 7 10 40
G 41.3
G 40.4
G 40.5
40 7 10 36
Peristaltic pump
50 7 40 00
50 7 40 10
50 7 40 20
50 7 40 30
50 7 40 40
50 7 40 50
Reagent-wall
50 7 40 60
50 7 40 61
50 7 40 62
50 7 40 70
40 7 40 70
40 7 10 50
40 7 10 55 bag-adapter COMBILINE, complete consisting of: metal-casing slide-bar holder COMBILINE slide-bar COMBI 1 set spring for slide-bar 1 set cannula 1set
Suspension for Calibration-Bags peristaltic pump, complete micromotor with gear 97.3:1 bush for bearing bearing roller device for peristaltic pump clutch tube connector, black tube connector, white, reagent-wall tube connector, white, bottle compartment holder for roller pump tubing, reagents (low) holder for roller pump tubing WASH (high) intermediate vessel holder for intermediate vessel
Service- Manual - Version 4.01
combiline
List of Spare Parts
Article no.: description
Bottles for COMBILINE BGA+E
Rm 100.24
Rm 100.30
Rm 100.31
R 100.21
60 8 20 80 waste-bottle 250ml, empty suction tube for reagent bottle WASH (250ml) suction tube for reagent bottle CAL3/4(+M) (150ml) screw cap for reagent- WASH- or waste-bottle (150 / 250ml) waste bottle device, complete
Bottles for COMBILINE BGA and COMBILINE ISE
Rm 100.23
Rm 100.21
R 100.20
60 8 20 70 waste-bottle 330ml (empty) suction tube for reagent bottle WASH / CAL (330ml) screw cap for CAL- WASH or waste-bottle (330ml) waste bottle device, complete for 330ml bottle
Sample port
Rm 20.80
R 20.81 (1-8.105)
R 20.83 (1-8.106)
G 20.51
G 20.52
G 20.53
Wiring
SEm 100.21
SEm 200.21
SEm 300.21
SEm 400.21
SEm 500.21
SEm 600.21
SEm 700.21
SEm 800.21
SEm 900.22
Valves
F 200.25
F 200.28
50 8 30 20
40 8 20 11
40 8 20 16
40 8 20 21
40 8 20 25 sample port device, complete with insert consisting of: sample port insert with integr. Steel canula (white) sample port insert (red) sample port holder sample port cover sample port support sensor-cable pO2 sensor-cable pCO2 sensor-cable pH sensor-cable Na+ sensor-cable K+ sensor-cable Ca++ sensor-cable Li+ sensor-cable Clsensor-cable reference tube valve (normally open) sample port valve PV (normally open) liquid Valve SMC liquid-valve block SMC complete for COMBILINE BGA+E liquid-valve block SMC complete for COMBILINE BGA liquid-valve block SMC complete for COMBILINE ISE multilayer valve carrier SMC
combiline
Service- Manual - Version 4.01
List of Spare Parts
Article no.:
Electronic boards
(ESW- ANALOG)
60 8 10 10
60 8 10 11
60 8 10 15
60 8 10 20
60 8 10 25
60 8 10 26
60 8 10 30
60 8 10 35
60 8 11 10
60 8 11 11
60 8 11 15
60 8 11 20
60 8 11 25
60 8 11 26
60 8 11 30
60 8 11 35
60 8 10 40
60 8 10 45
60 8 10 50
60 8 10 55
40 8 10 60
60 8 10 70
60 8 10 80
60 8 10 90
40 8 20 00
E 210.5
60 8 20 40 description analog-board for pO2 -pCO2-pH-baro.pressure, temperature analog-board for pO2 -pCO2-pH-baro-pressure-temperature-Hb analog-board for BGA, BP, K-temperature analog-board for BGA, BP, K-Na-temperature analog-board for BGA, BP, K-Na-Ca-temperature analog-board for BGA, BP, K-Na-Cl-temperature analog-board for BGA, BP, K-Na-Ca-Cl-temperature analog-board for BGA, BP, K-Na-Ca-Cl-Li-temperature analog-board for pO2 -pCO2-pH-Meta-baro.pressure, temperature analog-board for pO2 -pCO2-pH-Meta-baro-pressure-temperature-Hb analog-board for BGA, Meta, BP, K-temperature analog-board for BGA, Meta, BP, K-Na-temperature analog-board for BGA, Meta, BP, K-Na-Ca-temperature analog-board for BGA, Meta, BP, K-Na-Cl-temperature analog-board for BGA, Meta, BP, K-Na-Ca-Cl-temperature analog-board for BGA, Meta, BP, K-Na-Ca-Cl-Li-temperature analog-board for K-Na-temperature analog-board for K-Na-Ca-temperature analog-board for K-Na-Ca-Cl-temperature analog-board for K-Na-Ca-Cl-Li-temperature air pressure sensor 1240-015A-3N
CPU board (Z80 ADC) power supply board for valves, motors and temperature regulation (CL TEMP NETZ) switching power supply PM110 45-2A for +5V, -12V, +12V, +24V compl.
board for filling level control (CL FÜLLSTAND) displayadapter (ECODISP) interface RS232 for data output, incl. SD-Card Reader
Various electronical and electrical devices
50 8 20 65
50 8 20 70
50 8 20 71
F 100.4.2
60 8 20 50
60 8 20 90
F 10.5.1
60 8 20 60
F 10.19
temperature measuring sensor keyboard, complete keyboard adapter CL TASTATUR
LCD Display 240x128 dpi
LCD Converter thermal printer GBT-4352 heating resistor for thermostate heating device fan for power supply 24V, 1W suction pump
Service- Manual - Version 4.01
combiline
combiline
Service- Manual - Version 4.01
ESCHWEILER
automatic analysing systems
Eschweiler GmbH & Co. KG z
Holzkoppelweg 35 z
D - 24118 Kiel phone: ..49-(0)431-546580 e-mail: [email protected] fax: ..49-(0)431-549423 homepage: www.eschweiler-kiel.de
Declaration of EC-Conformity
as defined by EC In-Vitro Diagnostica (IVD) directive 98/79/EG
Eschweiler GmbH & Co. KG
Holzkoppelweg 35 / D-24118 Kiel / Germany declares that the product
Description: blood gas-/ electrolyte- and metabolite - analysers
Model:
ESCHWEILER combi line
to which this declaration relates is in accordance with the provisions of the following directives
Applied harmonised standards:
EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use:
EN 61010-2-08/A1
EN 61010-2-101
Emission:
EN 55011 A2
EN 61000-3-2
EN 61000-3-3
Immunity:
EN 61326-1
Safety requirements for electrical equipment for measurement, control, and laboratory use:
Safety requirements for electrical equipment for measurement, control, and laboratory use:
Industrial, scientific and medical radio-frequency equipment – Emission: Class B
Harmonic current emissions
Voltage fluctuation and flicker
Electrical equipment for measurement, control and laboratory use – EMC requirements:
EN 61000-4-2
EN 61000-4-3 +A1
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-11
Electrostatic discharge immunity test
Radiated, radio-frequency electromagnetic field – immunity test
Electrical fast transient/burst immunity test
Surge immunity tests
Immunity to conducted disturbances, induced by radio frequency fields
Voltage dips, short interruptions and voltage variations immunity test
ESCHWEILER GmbH & Co. KG
Kiel, 01.04.2010
..................................... i.A Dr. H.-J. Heupke
(QCM)
EC declaration combilineV2 IVD engl/01012010
ESCHWEILER
GmbH & Co. KG
Holzkoppelweg 35
D-24118 Kiel
Telefon: 0431 / 546580 Deutsche Bank AG, Kiel Commerzbank, Kiel Postbank
Telefax: 0431 / 549423 [email protected] www.eschweiler-kiel.de
Kto.-Nr. 0324 426
(BLZ 210 700 20)
SWIFT: DEUTDEHH 210
IBAN: DE46 2107 0020 0032 4426 00
Kto.-Nr. 7224488
(BLZ 210 400 10)
SWIFT: COBADEFFXXX
IBAN: DE66 2104 0010 0722 4488 00
Hamburg 24676 - 204
(BLZ 200 100 20)
PBNKDEFF
IBAN: DE56 2001 0020 0024 6762 04
Gerichtsstand / Place of Jurisdiction: Kiel USt.-ID Nr.: DE 188048227
Handelsregister Kiel HRA 3736 KI Kompl.: Eschweiler Verwaltungs GmbH Kiel HRB 4610 KI Geschäftsführer: Horst H. Kock
Service- Manual - Version 4.01
combiline