METTLER TOLEDO ICS226 QA3F, QA6F, QA15F, QA3R, QA6R, QA15R, QA3FCL, QA6FCL, QA15FCL, QA3RCL, QA6RCL, QA15RCL Compact scale User Manual
Below you will find brief information for Compact scale ICS226 QA3F, Compact scale ICS226 QA6F, Compact scale ICS226 QA15F, Compact scale ICS226 QA3R, Compact scale ICS226 QA6R, Compact scale ICS226 QA15R. This manual provides a detailed overview of the ICS226 Compact Scale, including its features, operation, and maintenance. It also includes safety instructions and technical specifications. The manual is designed to help you get the most out of your scale.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
ICS226
Compact Scale
Congratulations on choosing the quality and precision of METTLER
TOLEDO. Proper use according to these instructions and regular calibration and maintenance by our factory-trained service team ensure dependable and accurate operation, protecting your investment.
Contact us about a ServiceXXL agreement tailored to your needs and budget.
We invite you to register your product at www.mt.com/productregistration
so we can contact you about enhancements, updates and important notifications concerning your METTLER TOLEDO contact
.
2
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
Contents
.................................................................................................. 5
Charging and discharging battery ................................................................ 7
3
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
1 Introduction
1.1 Safety instructions
Product safety is very important for METTLER TOLEDO.
Non-observance of the following instructions can lead to damage to the weighing terminal and/or injuries.
▲ Read this manual carefully before operating or servicing the equipment.
▲ Store these instructions for future use.
▲ Strictly observe these instructions.
▲ Only permit qualified personnel to make checks, tests and adjustments to be carried out with power on. Failing to observe these precautions can result in bodily harm.
▲ Always disconnect the device from the power supply before in- stalling, servicing, cleaning or performing maintenance.
▲ Remove the battery if the device will not be used for a long time.
▲ Replace the used batteries by the batteries of the same type. Re- placement by wrong battery type can result in battery explosion.
▲ Check the power cable regularly for damage. If it is damaged, immediately disconnect the device from the power supply.
▲ Do not open the scale. The warranty is void if this stipulation is ignored. The scale may only be opened by authorized persons.
4
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
1.2 Description
1.2.1 General
Versions of ICS226:
● ICS226-QAxF for front mounting compact version
● ICS226-QAxR for rear mounting compact version
● ICS226-QAxFCL Front version with Ni- MH rechargeable battery
● ICS226-QAxRCL rear version with Ni- MH rechargeable battery
Features of ICS226:
● Two weighing functions can be performed:
- Simple weighing (zero, tare)
- Check weighing
1.3 Display
5
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
1.3.1 Status indicators
Icon
>0<
~
B/G
NET lb/oz/kg
Meaning
Indicators for check weighing
Zero setting activated
Motion indicator
Brutto/gross weight
The displayed weight value is net weight value
Condition of the battery
Currently chosen weight unit
1.3.2 Keys
Key Operating mode Menu
● Scrolling back
Zeroing
● Press and hold : abort current setting
Tare
Function key
Enter
● Scrolling forward
● When editing numeric values: increase the value
● Enter to edit value
● When editing numeric values:move the cursor to left
● Activating menu item
● Accepting selected setting
6
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
2 Putting into operation
2.1 Unpacking
1. Inspect the package immediately upon receipt.
– If the package is damaged, check for internal damage and file a freight claim with the carrier, if necessary.
– If the container is undamaged, open the package.
2. Remove the scale from the package and place it on a solid, flat surface.
3. Check all parts with the checklist. Make sure no part is damaged or missing.
Keep the packing material and inserts, in case you need to return the scale to a METTLER TOLEDO representative.
2.2 Charging and discharging battery
ICS226 provides a Ni-MH rechargeable battery as an option. Please contact METTLER TOLEDO service organization if the battery needs to be exchanged.
WARNING
Risk of explosion due to wrong operations of the battery or improper connection!
▲Please don’t use battery for other applications.
▲Please don’t manipulate on the battery pack.
7
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
3 Operation
3.1 Switching on and off
3.1.1 Switching on
Press and hold until the display lights go on.
At first, the scale performs a self-test. Then the display lights up and shows the software number.
3.1.2 Switching off
In simple weighing mode, press and hold until -OFF- is displayed.
3.2 Zero
Zeroing corrects the influence of slight changes on the load plate.
3.3 Tare
Place an empty container on the scale and press button.
The zero display and the NET indicator appears.
3.4 Units Switch
F2.1.1 = ON is set in the operator menu.
Press to switch between the main and second unit.
3.5 Check weighing
F2.2.1 = ON is set in the operator menu.
Press to enter into check weighing mode.
The scale starts to check weighing as soon net weight is greater than 10d.
8
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
3.5.1 Setting target weight
There are two ways to set new target values.
If F2.2.2 = MANUAL is set in the operator menu.
1. Press and hold until last target value appears and blinks.
2. Enter the new target weight. Using to move the cursor one place to left, to increase the displayed digit.
3. Press to save the new target weight. Or press to abort new value and quit.
If F2.2.2 = WEIGHT is set in the operator menu.
1. Press and hold until the blinking current weight value is displayed.
2. Put the target weight to the plate.
3. Press to save the new target weight. Or press „Z“ to abort new value and quit.
3.6 Selecting the scale location
The correct location is important to the accuracy of the weighing results.
Select a stable, vibration-free and horizontal location for the weighing platform.
Observe the following environmental conditions:
- No direct sunlight
- No strong drafts
- No excessive temperature fluctuations
3.7 Leveling the scale
Weighing scales should be leveled precisely horizontally to provide accurate weighing results.
1. Turn the adjustable feet of the weighing platform until the air bubble is inside the inner circle.
2. Tighten the lock nuts of the adjustable feet.
9
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
3.8 Notes on battery operation
ICS226 provides an optional available Ni- MH rechargeable battery, which could continuous work for 90 hours w/o backlight.
The indicator shows the state of the battery:
indicator appears => Battery operation
indicator doesn’t appears => Adapter operation
indicator blinks => the battery has to be charged immediately.
Connect the adapter to the scale and power on the scale, if the indicator blinks, that means batteries are being charged.
Take the following measures to prolong the batteries operation time:
Set F3.1 to the required parameter for turning the backlight off or shorten the backlight time.
Set auto power-off/sleep time (F3.2).
Remove the battery if the device will not be used for a long time.
3.9 Cleaning
DANGER!
Electric shock hazard due to ingress of moisture!
▲ Before cleaning the scale, pull off the power plug to disconnect the unit from the power supply.
Further notes on cleaning
Use a damp cloth.
Do not use any acids, alkalis or strong solvents.
Follow all the relevant instructions regarding cleaning intervals and permissible cleaning agents.
10
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
4 Settings in the menu
The menu consists of the following 4 main blocks which contains various submenus on several levels.
F1 - Scale menu setting
F2 - Application menu setting
F3 - Instrument menu setting
F4 - Maintenance menu setting
End - Exit menu
4.1 Operating the menu
The menu has 2 deifferent operating levels: Operator and Supervisor. The operator level can only access F2 and F3 menu settings. The supervior level can set all menu settings.
4.1.1 Calling up the menu and entering the password
Press and hold until PSD is displayed to prompt user for a password.
Operator menu
Supervisor menu
If F1.1 = CN.
The following settings are only accessible in Chinese: F1,1, F1.2, F1.3,
F1.4, F1.7, F2.1 and F4.3.
4.1.2 Numeric entry
1. Press to start editing the displayed value.
2. Increase the displayed digit using the key.
3. When entering multi-digit numbers, use the key to move the cursor one place to left.
4. Change the digit as described in step2.
5. Repeat setps 2, 3 and 4 if necessary.
6. When all digits are entered, use the key to confirm the entry.
11
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
4.2 F1 – Scale menu
Factory settings are printed in bold characters.
4.2.1 F1.1 – Approval no
OINL no approval
Approval for China
4.2.2 F1.2 – Full capacity
Possible settings 3kg, 6kg, 15kg
4.2.3 F1.3 – Calibration
F1.3.1 – Geo value
Adaptation of the scale to the geographical location.
Factory setting
16
Possible settings 0...31
Note Re-calibrate after changing Geo value
F1.3.2 – Calibration
1. E SCL appears, unload scale.
Perform calibration
Note
2. Start caalibration by pressing .
The scale determines the zero point.
The calibration weight to be placed on the scaleis shown on the display.
3. Place the calibration weight on the scale and confirm with .
4. donE appears, calibration completed.
In step 2, if you want to change the calibration weight, press the key repeatedly until the desired calibration weight appears on the display.
Press and hold to abort calibration process
12
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
F1.3.3 – Linearity Calibration
Perform calibration
1. E SCL appears, unload scale.
2. Start calibration with .
After capturing zero point, first weight is shown on the display.
3. Place first weight on the scale and press
.
4. First weight confirmed. Second weight is shown on the display.
5. Remove first weight and place second weight on the scale and press
.
6. donE appears, calibration completed.
In step 2 and 4, if necessary, change the weight displayed value with .
Note
Press and hold to abort calibration process
4.2.4 F1.4 – Zero
F1.4.1 – Automatic zero setting
Factory setting
0.5d
Possible settings OFF, 1d, 2d, 5d, 10d
F1.4.2 – Power up zero
2%
20%
Power up zero within +/-2%
Power up zero from -2% ~ %18
* For approved mode, the setting is always 2%.
F1.4.3 – Pushbutton zero
Factory setting
2%
Possible settings 2%, 20%,Off
13
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
4.2.5 F1.5 – Tare
F1.5.1 – Auto Tare
Factory setting
OFF
Possible settings ON,OFF
F1.5.2 – Auto clear tare
Factory setting
OFF
Possible settings ON, 9d
F1.5.3 – Pushbutton tare
Factory setting
OFF
Possible settings ON
F1.5.4 – Chain tare
Factory setting
OFF
Possible settings ON
F1.5.5 – Auto tare threshold
F1.5.1=ON
Factory setting 1d
Possible settings 0 ... maximun load
4.2.6 F1.6 – Digital filter
F1.6.1 – Digital filter
Factory setting
Middle
Possible settings Low, HIgH
F1.6.2 – Stability
Factory setting
Stand
Possible settings Fast, Exact
14
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
4.2.7 F1.7 – Display
F1.7.1 – Weighing unit
Factory setting
kg
Possible settings g, lb, oz
F1.7.2 – Resolution
Capacity 3 kg
Factory setting
0.001 kg
Possible settings 0.0002 kg,
0.0005 kg
6 kg
0.002 kg
0.0005 kg,
0.001 kg
15 kg
0.005 kg
0.001 kg,
0.002 kg
4.2.8 F1.10 – Reset F1 parameters to factory setting
Reset all F1 parameters to factory setting, except the calibration and full capacity values.
4.3 F2 – Application menu
Factory settings are printed in bold characters.
4.3.1 F2.1 – Units switch
F2.1.1 – Unit switch
Factory setting
OFF
Possible settings ON
F2.1.2 – Second unit
F2.1.1=ON
Factory setting
g
Possible settings kg, lb, oz
4.3.2 F2.2 – Check weighing function
F2.2.1 – Check weighing
Factory setting
OFF
Possible settings ON
15
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
F2.2.2 – Setting the target weigt
F2.2.1=ON
WEIGHt
ManUAL
By weighing in an actual sample weight
By numeric entry
F2.2.3 – Upper tolerance
F2.2.1=ON
Factory setting
1d
Possible settings 0 ... maximum load
F2.2.4 – Lower tolerance
F2.2.1=ON
Factory setting
1d
Possible settings 0 ... maximum load
4.3.3 F2.10 – Reset F2 parameters to factory setting
Reset all F2 parameters to factory setting.
4.4 F3 - Terminal menu
Factory settings are printed in bold characters.
4.4.1 F3.1 – Backlight
Factory setting
ON
Possible settings 5s, 10s, 30s, 60s, OFF
4.4.2 F3.2 – Auto power off/sleep time
Factory setting
30
Possible settings 5, 10, 30, 60 (minutes)
4.4.3 F3.10 – Reset F3 parameters to factory setting
Reset all F3 parameters to factory setting.
16
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
4.5 F4 - Maintenance menu
Factory settings are printed in bold characters.
4.5.1 F4.1 – Keyboard test
The salce shows PrESS
Press all the keys to go through the test.
Note: If keypad test fails, please switch off the scale to quit test.
4.5.2 F4.2 – Display test
All the icons and segments light up.
4.5.3 F4.3 – Display expand resolution
Expand display resolution to 100,000.
Factory setting
OFF
Possible settings ON
4.5.4 F4.10 – Reset F4 parameters to factory setting
Reset all F4 parameters to factory setting.
4.6 End - Ending menu
1. Press .
StoreE? appears.
2. Press again to save changes.
- or –
Press to reject changes.
AbOrt appears.
Press .
17
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
5 Error messages
Error code
Error
┌----┐
Overload
Remedy
Unload scale
Reduce preload
└----┘
Underload
Zeroing the weighing platform
┌-no-┐
Zero/Tare setting outside setting range
└-no-┘
--no--
Unload scale
Empty platter before power up
Recalibrate zero point
Cannot perform the key function
Back to gross mode
------
scale is in motion
Err 53
EEPROM checksum error
Err 35
Calibration weight too small
Ensure the weighing platform is stable
Unplug power plug and plug in again
If this doesn‘t help, contact your authorized METTLER
TOLEDO service organization
Check calibration weight
Err 6
Scale not calibrated yet
Recalibrate scale
Err 8
Err 10
Illegal operation
Illegal target value
Switch off the scale and switch on again
Please set the correct target and tolerance values
18
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
6 Technical Data
6.1 Scale
Capacity
Resolution
Power Supply
IP Protection
Environment
Display
Material
Load cell
Platform Size
3kg/6kg/15kg
3000d/6000d/7500d
● 100~240 VAC
● 6V Ni-MH rechargable battery 3800mAh
(option)
● Terminal:IP67
● Scale:IP65
● Temperature range: -10
C ~ +40
C
8-seg x 6 LCD Display, LED white backlight
AISI304 stainless steel
Stainless steel, IP67
240 x 240mm
6.2 Product versions
Model
ICS226-QA3F
ICS226-QA6F
ICS226-QA15F
ICS226-QA3R
ICS226-QA6R
ICS226-QA15R
AC
Capacity Readability Installation Product Size
3kg
6kg
15kg
3kg
6kg
15kg
1g
2g
5g
1g
2g
5g
Front
Front
Front
Rear
Rear
Rear
375 x 331 x 86 mm
Internal Rechargeable Battery, AC/DC
Model Capacity Readability Installation Product Size
ICS226-QA3FCL
ICS226-QA6FCL
3kg
6kg
ICS226-QA15FCL 15kg
ICS226-QA3RCL 3kg
ICS226-QA6RCL 6kg
ICS226-QA15RCL 15kg
1g
2g
5g
1g
2g
5g
Front
Front
Front
Rear
Rear
Rear
359 x 331 x 158 mm
19
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
6.3 Dimensions
Front mounting (375X331X86mm) Rear mounting(359X331X158mm)
20
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
7 Appendix
7.1 Disposal
In conformance with the European Directive 2002/96/EC on Waste
Electrical and Electronic Equipment (WEEE), this device must not be disposed of in domestic waste. This also applies to countries outside the EU as per their specific regulations.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
21
METTLET TOLEDO
Service Manual ICS226 Compact Scale Order number 30032410A 03/01
Mettler Toledo
Industry/Retail
Add: 111 West Tai Hu Road, Xinbei District,
Changzhou, Jiangsu
Tel: 0519-86642040
Fax: 0519-86641991
E-mail: [email protected]
Printed in P. R. China 2012/11
Order number 30032410B
www.mt.com/service
For more information
* 3 0 0 3 2 4 1 0 B *