Luxtec 9300 XSP, XDP Xenon Light Source Operation and Service Manual

Luxtec 9300 XSP, XDP Xenon Light Source Operation and Service Manual

The 9300XSP/9300XDP 300 Watt Xenon Light Source is designed to supply high-intensity white light to a fiber optic cable for illumination of a surgical field or other area of examination or operation. It delivers 300-Watts of cool white IR filtered light and comes with a rotary turret for attaching light guide cables for instrument or headlight use. The lamp is housed in a lamp module that can be easily removed and replaced without special tools.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Luxtec 9300XSP/9300XDP Xenon Light Source Operation and Service Manual | Manualzz

Operation and Service Manual

9300XSP/9300XDP

300 Watt Xenon Light Source

ENGLISH

TABLE OF CONTENTS

Table of Symbols

General Warnings

9300XSP/9300XDP Overview

Indication For Use

Inspection Before Use

Optional Floor Stand Assembly

Installation to Floor Stand

Light Source Operation

Xenon Lamp Module Replacement

Turret Change

Troubleshooting

Maintenance and Cleaning

Light Source Specifications

Electromagnetic Compatibility (EMC)

User Information

Block Diagram

Ordering Equipment Parts List

Limited Warranty

Repair and Return

12

12

13

6

7 - 8

8 - 10

11

3

4

4 - 5

4 - 5

5

6

14 - 16

17

18

18

19

TABLE OF SYMBOLS

Found on Light Source

Attention: Consult Accompanying

Documents

Caution: High Voltage

Brightness

Equipotentiality

Protective Earth

ON

OFF

Fuse

Type CF Equipment

High Temperature

Intensity

Alternating Current

Meets requirements of the Medical Device

Directive

Listed by CSA

3

GENERAL WARNINGS

The user of this Luxtec product should be thoroughly familiar and trained in the assembly, use and care of this product.

The user should carefully study the Operation and Service Manual before making any attempt to use the equipment in the clinical environment. Instructions should be followed specifically, with special attention given to warnings, controls and user specifications. The

Manual should also be available to the appropriate personnel.

This Manual contains information about the proper procedures for inspecting and preparing this product before its use, and care and storage after use.

Follow the instructions and restrictions in the operating manuals of other manufacturer’s equipment when they are used in conjunction with this product.

Before every procedure, carefully inspect the Light Source to ensure it has been properly maintained and cleaned, and that it is fully functional. DO NOT use if inspection reveals any damage such as cable cuts or loose connectors.

To reduce the risk of fire and electric shock, do not expose electrical equipment to moisture. When cleaning, do not immerse any electrical device in liquid.

SAFETY PRECAUTIONS MUST ALWAYS BE EXERCISED WHEN USING ELECTRICAL

EQUIPMENT TO PREVENT OPERATOR/PATIENT SHOCK, FIRE HAZARD OR

EQUIPMENT DAMAGE.

All electrical equipment must be used with approved hospital grade power cords and power plugs inserted properly into grounded AC power outlets.

This product must be removed from the surgical field prior to the use of a defibrillator.

The Light Source should never be used in ocular surgery or in a surgical procedure requiring direct illumination of the eye.

The Light Source produces high intensity light. Thermal burns can be the result of improper use of the Light Source or the light output of the light guide cable.

Use only fiber optic light guide cables that are approved for high intensity light output.

OVERVIEW

Indication for Use

This device is designed to supply high-intensity white light to a fiber optic cable for illumination of a surgical field or other area of examination or operation.

The 9300XDP and 9300XSP deliver 300-Watts of cool white IR filtered light. The lamp is housed in a lamp module that can be easily removed and replaced by the user without special tools.

4

OVERVIEW

An illuminated push-on and push-off power switch is located on the front panel. See Figures 1 and 2.

The rotating turret wheel is designed to accommodate the user’s choice of 4 different light guide endfitting styles. The standard turret has ACMI, Olympus, Wolf and Storz adapter ports.

The intensity of the light output is controlled manually by the light intensity knob located to the right of the turret wheel.

The bulb life is indicated in hours of use on the front panel by a digital meter.

NOTE: You must use ONLY a Luxtec supplied lamp module assembly for replacement in this Light Source.

Use of other lamps or lamp assemblies will void all warranties for this product. Refer to Parts List later in this manual.

Figure 1

Digital Lamp Age Meter

Power Switch

Figure 2

Secondary Turret

Digital Lamp Age Meter

Power Switch

Turret Light Intensity Knob

9300XSP Front View

Primary Turret

Beamsplitter

Light Intensity Knobs

9300XDP Front View

Set Up and Inspection Before Use

The 9300XDP and 9300XSP Light Sources come with the power cord attached. Please verify that both components have been received undamaged.

All fiber optic light guides should be properly cleaned and sterilized prior to first time use. See light guide manufacturer’s manual for information.

Warnings:

Do not use or store liquids on or above the Light Source.

Electric shock hazard. If unit is not functioning properly, DO NOT OPEN. Please refer to the

Repair and Return Section of this Manual.

Explosion Hazard. Do not use in the presence of flammable anesthetics, liquids, vapors, gases or dusts.

Use care not to point any light guide directly at the eye while operating the Light Source.

Keep cooling vent and fans free of obstructions.

FIRE HAZARD: DO NOT DRAPE OR COVER THE LIGHT SOURCE WHILE IT IS OPERATING.

When Light Source is not in use, turn off the power, or rotate the turret 1/8 turn to block light output.

5

OPTIONAL FLOOR STAND ASSEMBLY

Figure 3

There are three (3) parts to the floor stand assembly:

See Figure 3.

Swivel-top base

Fiber optic headlight hanger

Power cord hanger

Interlocking adapter plate lever

Nylon bushing

Top section

1. Casters base

2. Two-piece column

3. Swivel top base with interlocking adapter plate with cord hangers

Two-piece column

To assemble the floor stand, remove the base from the box and stabilize it by placing a stop in front of one of the casters.

Bottom section

Casters base

Assemble the two column sections as shown in

Figure 3. One section of the column has an arrow that should point up as the stand is assembled. Slide the lower end of the top section over the upper end of the bottom section. Position the tapered end of the two piece column in the base. Be sure the nylon bushing in the upper section of the column is at the top.

Figure 4

Interlocking adapter plate lever

(“open”position)

Figure 5

Rim

Swivel-top base

Interlocking adapter plate plunger

Interlocking adapter plate

(rear)

Swivel-top base

Interlocking adapter plate lever

(“locked”position)

Place the swivel top base through the nylon bushing. The base should rotate freely within this bushing.

Open the interlocking adapter on the swivel top base by pulling down on the plunger and sliding the lever to the “open” position as labeled. See Figure 4. The swivel top base is now ready to receive the Light

Source.

Installation to Floor Stand

1. Place Light Source on the floor stand base so the front edge is up to the rim of the base. Verify that all rubber feet of the

Light Source are positioned on the base.

2. Slide the interlocking adapter plate to the lock position. The plate plunger will snap into the locked position as shown in

Figure 5.

3. Check the rear of the Light Source to be sure the floorstand plate captures the support plate underneath the Light

Source and is secure.

Setting Up

Before turning power on the Light Source, make sure the unit is plugged into any standard 100V to 240V

50/60Hz (as appropriate) three-conductor outlet.

Grounding reliability is guaranteed only when connected to a “hospital grade” receptacle.

6

LIGHT SOURCE OPERATION

Power on the Light Source

The power switch is located on the front panel. See

Figure 8. Ignition clicking sounds are normal. Some light will show through the turret wheel and the fan outlets. When the power switch is pressed, it will light up indicating power is on. The Light Source is now ready to use.

Rotate turret wheel 1⁄8 turn to block light output.

See Figure 6.

When finished using the Light Source, press the power switch again to turn off the unit.

Turret

The 9300XDP and 9300XSP Light Sources are equipped with a rotary turret. See Figures 6 and 7.

This turret is to attach light guide cables for instrument or headlight use to the Light Source.

Luxtec offers three turret options:

1. Kit “T”: ACMI, Wolf, Storz and Olympus (standard)

2. Kit “P”: ACMI, Wolf, Storz and Pilling

3. Kit “D”: Wolf, Storz, Designs for Vision and

Olympus.

To operate the turret, rotate the wheel until the desired connection lines up with the light channel.

When the fiber optic cable is inserted, a click can be heard indicating the cable is properly seated.

Light Attenuation

The 9300XDP and 9300XSP have a manually operated light intensity knob to control the desired light output level. To increase the light output, rotate the knob clockwise until the desired level is reached.

See Figures 6 and 7.

NOTE: It is strongly recommended that the light intensity be adjusted to the minimum level of brightness for the procedure. There is a risk that prolonged exposure on one spot during surgery, at close proximity may cause an increase in tissue temperature. Minimum intensity levels and exposure will mitigate this occurrence.

Figure 6

Figure 7

Turret

9300XSP

Light intensity knob

Primary Turret

Secondary Turret

9300XDP

Light intensity knobs

7

LIGHT SOURCE OPERATION

Dual Port Mirror Mode (9300XDP only)

The 9300XDP has a knob on the front panel which indicates single or dual port capability. See

Figure 8. The knob rotates to allow the operator to use the Light Source in a single or dual port mode.

NOTE: The second port is NOT a back-up port for the first port.

From the front panel symbols:

• Two yellow bulbs indicate two port operation.

• One yellow bulb and one black bulb indicates single port operation.

Figure 8

Single or dual port indicator knob

9300XDP

XENON LAMP MODULE REPLACEMENT

Note: Please adhere to appropriate safety precautions when performing lamp replacement. Only qualified Personnel should service this device. Protective facemask and gloves should be worn when replacing lamp module. Before changing the lamp module, turn power off and allow the Light

Source to cool for at least fifteen (15) minutes.

Please read and comply with all precautions listed in the General Warnings section of this manual.

The Xenon Lamp Module should be replaced after 650 hours of use or if the lamp fails to start when the power is turned on.

NOTE: You must use ONLY a Luxtec supplied lamp module assembly for replacement in this

Light Source. Use of other lamps or lamp assemblies will void all warranties for this product.

Refer to Parts List later in this manual.

8

XENON LAMP MODULE REPLACEMENT

Figure 9

Figure 10A

Rear panel fasteners

To remove:

1. Make sure power switch is turned off.

Disconnect the power cord from the hospital grade receptacle.

2. Allow the unit to cool down for a minimum of

15 minutes before replacing the lamp module.

The lamp module may represent a burn hazard if not allowed to cool sufficiently prior to servicing the device.

3. Remove the two side screws and loosen the two quarter turn screws in the back of the unit. See

Figure 9. Pull back on the top cover handle as indicated in Figure 10A and then pull up on the top cover to open as indicated in Figure 10B.

4. Firmly grasp the lamp module by the protective cover with one hand while pressing the release button with the other. Pull up to remove the assembly. See Figure 11. Do not use excessive force or prying tools.

Figure 10B

9

XENON LAMP MODULE REPLACEMENT

To replace:

1. Replace the lamp module assembly only with a Luxtec provided lamp module assembly.

2. Firmly grasp the lamp module by the protective cover with one hand while pressing the release button with the other. Do not touch the glass surface of the lamp itself.

3. Line up the base with its mating piece and firmly press the lamp module into place.

See Figure 12.

Figure 11

Lamp module

4. Disengage the release button to secure lamp in its proper position. The lamp module should not tilt or jiggle while in place.

5. Pull the top cover down to close the Light

Source (reverse movements shown in

Figure 10) then push the cover toward the front of the unit to line up side screw holes. Replace side screws. Secure quarter turn screws in back of the unit by turning to the right.

Figure 12

After closing the unit, push the front panel power button to turn the unit on. Record the number of hours indicated by the digital lamp age meter on the front panel. Note: the Light Source must be operating to reset the digital lamp age meter. See

Figure 13. A maintenance log sticker is located on the rear panel of the unit for convenient recording and tracking of this information.

After recording the hours of operation, the digital lamp meter may be reset by momentarily depressing the rear panel reset switch as shown in Figure 13. Turn the Light Source off after completing this procedure. Do not leave the unit operating while unattended.

Lamp module

Figure 13

Maintenance log sticker

Reset switch

10

TURRET CHANGE

Figure 14

Figure 15

9300XSP

9300XDP

How to change the turret

To remove current turret assembly:

1. Turn off Light Source. Unplug unit.

2. Rotate turret until Wolf port is lined up at

12 o’clock.

3. Through the Wolf port hole, remove the screw with a Phillips screwdriver.

4. Rotate turret so the Wolf large hole is over the opposite screw, at 6 o’clock.

5. Through this large hole, remove the second screw.

6. While pulling forward, gently remove the turret assembly.

To install a new turret assembly:

1. Push desired turret into hole in bezel. A small pin in the plate behind the cup will line up with a small hole in the turret cup. It will allow the cup to seat only one way.

2. Rotate turret until Wolf port is lined up at

12 o’clock.

3. Through the Wolf port, install one of the flat head screws (supplied with turret).

4. Rotate turret so the Wolf large hole is over the opposite screw, at 6 o’clock.

5. Through the Wolf port, install the second screw.

6. When secured, turret should not rock but should rotate freely.

Rotate turret to assure it is seated and secure. Light source is now ready to use.

11

TROUBLESHOOTING

Problem Cause

No light output Light source not turned on

Bad/no lamp

Attenuator closed

Turret mispositioned

Blown fuse

Reduced light output Cable mismatched to turret

Attenuator mispositioned

Bad lamp

No power Light source not plugged in

Top cover not closed

MAINTENANCE AND CLEANING

Action

Turn power on

Check lamp seating/replace lamp module

Check position of knob on front panel

Rotate turret to desired adapter fitting

Replace fuse as indicated in maintenance section

Rotate turret to matching adapter fitting

Check position of knob on front panel

Replace lamp module

Plug in Light Source

Close and secure top cover

Maintenance:

If the Light Source does not operate properly when connected to a grounded receptacle, check the fuse(s). Do not attempt to repair the unit if the lamp fails in use. Turn off the unit and allow it to cool for at least 15 minutes, then try to restart the unit. If the lamp still fails to illuminate, the lamp may be defective, the lamp has exceeded its useful life (see lamp replacement), or there is a power failure. Try replacing the lamp module. If you still experience difficulties, return the unit to Luxtec Corporation for evaluation.

Cleaning:

The Light Source can be cleaned and disinfected using 70% isopropyl alcohol.

Unplug the power cord before cleaning.

Allow 5 minutes for alcohol to evaporate before reconnecting to power.

To replace the fuse:

The fuse(s) for the Light Source is (are) located in the power entry module in the rear of the unit. Remove the power cord by unscrewing the cord lock from the back of the Light Source with a small screwdriver, then unplug the cord from the unit. Using the small flat screwdriver, pry out the red plastic back from the power entry module. Check to see if the fuse(s) is (are) blown, then replace with fuse(s) of the same rating (T6.3A 250V).

To replace the red block in the housing, snap the retaining ring in place. Plug cord back into Light Source and retest the unit.

If unit now works, replace the cord lock with screws previously removed to secure cord to unit.

12

SPECIFICATIONS

Lamp

Type

Wattage

Color Temperature

Bulb Warranty

Position

Light Source

Dimensions

Weight

Power Input

Circuit Protection

Power Consumption

AC Power Leakage

Classification

Electrical Safety

Electromagnetic Compatibility

Environment:

OperatingTemperature

Temperature Range

Humidity

Xenon Short Arc Lamp

300 Watts

>6000 Kelvin, IR Filtered

650 Hours, Prorated per Hour

Horizontal ±15°

15"L x 12"W x 5"H

(381mm L x 305mm W x 127mm H)

21 lbs. (9.5kg)

100~240V AC 50 – 60 Hz

Fuse 6.3A 250V

450 VA

Leakage current to chassis (with ground wie intact), less than 100 microamps

Leakage current to chassis (with ground wie interupted), less than 500 microamps

Type CF, Class 1

CSA Listed, Conforms to UL60601-1 and

CSA C22.2 NO. 601.1

EN60601-1-2:2001

10°C – 40°C

10°C – 40°C

95% Relative Humidity Non-Condensing

Maximum

13

Electromagnetic Compatibility (EMC) User Information

WARNING: Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying documents.

WARNING: Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.

WARNING: The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used.

NOTE: The EMC tables and other guidelines that are included in the Instruction Manual provide information to the customer or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic

Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing other Equipment and Systems or non-medical electrical equipment.

Table 201

Guidance and Manufacturer’s Declaration – Emissions All Equipment and Systems

The XSP/XDP is intended for use in the electromagnetic environment specified below. The customer or user of the XSP/XDP should assure that it is used in such an environment.

Emissions Test Electromagnetic Enforcement – guidance

RF Emissions

CISPR 11

Compliance

Group 1

RF Emissions

CISPR 11

Class B

Radiated

And

Conducted

Emissions

The XSP/XDP uses RF energy only for its internal function.

Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The XSP/XDP is suitable for use in all establishments including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

N/A

Conducted Emissions Tests Performed at both 240VAC, 50Hz

And 120VAC, 60Hz

Equipment intended for Professional Use Only Harmonics

IEC 61000-3-2

Flicker

IEC 61000-3-3

N/A Equipment intended for Professional Use Only

14

Electromagnetic Compatibility (EMC) User Information

Table 202

Guidance and Manufacturer’s Declaration—Immunity All Equipment and Systems

The XSP/XDP is intended for use in the electromagnetic environment specified below. The customer or user of the XSP/XDP should assure that it is used in such an environment.

Immunity Test

Electrostatic

Discharge (ESD)

IEC 61000-4-2

Electrical Fast

Transient/burst

IEC 61000-4-4

IEC 60601

Test Level

±6kV contact

±8kV air

±2kV on AC

Mains

Compliance

Level

Electromagnetic Environment - Guidance

±6kV contact

±8kV air

Floors should be wood, concrete or ceramic tile. If floors are synthetic, the relative humidity should be at least 30%.

±2kV on AC Mains Mains power quality should be that of a typical commercial or hospital environment.

Surge

IEC 61000-4-5

±1kV

Differen tial

±2kV Common

±1kV Differential

±2kV Common

Note - Tests Performed at both 240VAC, 50Hz and 120VAC, 60Hz

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips,

Short interruptions and

Voltage variations on power supply input lines

IEC 61000-4-11

>95% Dip for

0.5 Cycle

60% Dip for

5 Cycles

30% Dip for

25 Cycles

>95% Dip for

5 Seconds

>95% Dip for

0.5 Cycle

60% Dip for

5 Cycles

30% Dip for

25 Cycles

>95% Dip for

5 Seconds

Note - Tests Performed at both 240VAC,

50Hz and 120VAC, 60Hz

Mains power quality should be that of a typical commercial or hospital environment. If the user of the XSP/XDP requires continued operation during power mains interruptions, it is recommended that the XSP/XDP be powered from an uninterruptible power supply or battery.

Power

Frequency

50/60Hz

Magnetic Field

IEC 61000-4-8

3A/m 3A/m

Note - Tests Performed at both 240VAC, 50Hz and 120VAC, 60Hz

Power frequency magnetic fields should be that of a typical location in a typical commercial or hospital environment.

Note - Tests Performed at both 50Hz and 60Hz

15

Electromagnetic Compatibility (EMC) User Information

Table 204

Guidance and Manufacturer’s Declaration – Emissions Equipment and Systems that are NOT Life-

Supporting

The XSP/XDP is intended for use in the electromagnetic environment specified below. The customer or user of the XSP/XDP should ensure that it is used in such an environment.

Immunity

Test

Conducted RF

IEC 61000-4-6

IEC 60601

Test Level

3 Vrms from

150 kHz to

80 MHz

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to

2.5 GHz

Compliance

Level

Electromagnetic Environment – Guidance

V1 = 3 Vrms Portable and mobile RF communications equipment should be separated from the XSP/XDP by no less than the recommended separation distances calculated/listed below:

E1 = 3V/m

D

=

( 3 .

5 /

V

1 )

P

D

=

( 3 .

5 /

E

1 )

D

=

( 7 /

E

1 )

P

P

80 to 800 MHz

800 MHz to 2.5 GHz

Where P is the maximum power rating in watts and D is the recommended separation distance in meters.

Field strengths from fixed transmitters, as determined by an electromagnetic site survey, should be less that the compliance levels (V1 and E1).

Interference may occur in the vicinity of equipment containing a transmitter.

Table 206

Recommended Separation Distances Between Portable and Mobile RF Communications

Equipment and the XSP/XDP Equipment and Systems that are NOT Life-Supporting

The XSP/XDP is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the XSP/XDP can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the XSP/XDP as recommended below, according to the maximum output power of the communications equipment.

Maximum Output Power

(Watts)

Recommended Separation Distances for the XSP/XDP (meters)

150 kHz to 80 MHz 80 to 800MHz 800 MHz to 2.5 GHz

d

=

1 .

1667

P d

=

1 .

1667

P d

=

2 .

3333

P

0.01

0.1

1

10

100

0.11667

0.36894

1.1667

3.6894

11.667

0.11667

0.36894

1.1667

3.6894

11.667

0.23333

0.73785

2.3333

7.3785

23.333

16

BLOCK DIAGRAM

17

ORDERING EQUIPMENT

9300XSP/9300XDP Parts List

To place an order, contact your local Luxtec distributor or call Luxtec Customer Service at

1-800-325-8966 (USA & Canada) or at +1-508-835-9700.

Light Source

Note: Dual Turrets are identified as a two letter suffix. The first letter indicates the primary turret option and the second letter indicates the secondary turret option.

009300XDPCFDD 300 Watt Xenon Dual Turret Light Source with 2 D Turrets

009300XDPCFDP

009300XDPCFDT

009300XDPCFPD

009300XDPCFPP

009300XDPCFPT

009300XDPCFTD

009300XDPCFTP

009300XDPCFTT

300 Watt Xenon Dual Turret Light Source with 1 D and 1 P Turret

300 Watt Xenon Dual Turret Light Source with 1 D and 1 T Turret

300 Watt Xenon Dual Turret Light Source with 1 P and 1 D Turret

300 Watt Xenon Dual Turret Light Source with 2 P Turrets

300 Watt Xenon Dual Turret Light Source with 1 P and 1 T Turret

300 Watt Xenon Dual Turret Light Source with 1 T and 1 D Turret

300 Watt Xenon Dual Turret Light Source with 1 T and 1 P Turret

300 Watt Xenon Dual Turret Light Source with 2 T Turrets

009300XSPT

009300XSPP

009300XSPD

300 Watt Xenon Single Turret Light Source with A–W–O–S Turret

300 Watt Xenon Single Turret Light Source with A–W–P–S Turret

300 Watt Xenon Single Turret Light Source with W–D–O–S Turret

For any 9300XDP purchased prior to November 2000, call Luxtec Customer Service for parts.

T, D, and P Turret Kits

000662

000663

000664

Kit, “T” CF Turret, ACMI, Wolf, Storz and Olympus

Kit, “P” CF Turret, ACMI, Wolf, Storz and Pilling

Kit, “D” CF Turret, Wolf, Storz, Designs for Vision and Olympus

Lamp Assembly

(same for both XSP and XDP product lines)

000090

0025001

Lamp Assembly*

Lamp only*

* For warranty compliance, a Luxtec lamp assembly or lamp must be used in these Luxtec products. The use of non-Luxtec bulb or lamp assembly will void the warranty for that product.

Accessories

400456

400681

410116LX

600972

600987

601058

601059

601178

601317

601318

601949EUR6M

601949UK6M

601949US6P1

604035

850159

Power Switch Assembly

Digital Age Meter Assembly

Fan Assembly

Power Supply, 300 Watt Xenon

Fuse 6.3 Amp, Slow Blow

Intensity Knob

Actuator Knob

Attenuator O-Ring

Handle

Rubber Bumper Feet

European Power Cord 20’ (6m)

Hospital Grade UK Power Cord 20’ (6m)

Hospital Grade USA Power Cord 20’ (6m)

Interlock Switch

Operating and Service Manual 9300XSP/XDP

18

LIMITED WARRANTY

Luxtec warrants that the equipment manufactured by Luxtec shall be free from defects in material and workmanship under normal use and service for a period of three (3) years from the date of shipment from Luxtec. Luxtec’s sole and exclusive liability under the warranty shall be, at Luxtec’s option, either to repair or to replace any component which fails during the warranty period due to any defect in workmanship or material F.O.B. factory if:

1. Customer promptly reports such defect to Luxtec in writing,

2. If requested by Luxtec, customer returns equipment to Luxtec with shipping charges and,

3. Upon inspection, Luxtec finds the equipment to be defective.

This warranty is contingent upon normal and proper use of the equipment. It does not cover equipment modified without the written approval of Luxtec, subjected to unusual physical or electrical stress, or damaged during shipment. This warranty is non-transferable unless authorized in writing by Luxtec.

Luxtec reserves the right to make design changes on its products without liability to incorporate said change in Luxtec products previously designed or sold.

Upon receipt of the product, it should be carefully inspected. If any defect is discovered, notification must immediately given to the manufacturer.

REPAIR AND RETURN

This device must be clean when returned to Luxtec. Luxtec reserves the right to return unrepaired any equipment that is contaminated with blood or other organic material.

Warranty Service and Repair:

To obtain service under warranty or return product for repair, the customer should contact your local Luxtec distributor or call Luxtec Customer Service at 800-325-8966

(USA & Canada) or +1-508-835-9700.

Information contained in this manual is subject to change without notice.

99 Hartwell Street • West Boylston, MA 01583• USA

Telephone: (800) 325-8966 (USA & Canada) or (508) 835-9700

Facsimile: (508) 835-9976 E-mail address: [email protected]

P/N 850159H

advertisement

Key Features

  • 300 Watts of cool white IR filtered light
  • Rotary turret for attaching light guide cables
  • Easy to replace lamp module
  • Digital lamp age meter
  • Light intensity control
  • Dual port mirror mode (9300XDP only)

Frequently Answers and Questions

How do I replace the lamp module?
The lamp module should be replaced after 650 hours of use or if the lamp fails to start. Make sure the power is off and the unit has cooled down for at least 15 minutes before replacing the lamp module. Remove the two side screws and loosen the two quarter turn screws in the back of the unit. Pull back on the top cover handle and then pull up on the top cover to open. Firmly grasp the lamp module by the protective cover with one hand while pressing the release button with the other. Pull up to remove the assembly. To replace, line up the base with its mating piece and firmly press the lamp module into place. Disengage the release button to secure the lamp in its proper position. Close the Light Source and push the front panel power button to turn the unit on. Record the number of hours indicated by the digital lamp age meter on the front panel.
How do I change the turret?
To remove the current turret assembly, turn off the Light Source and unplug the unit. Rotate the turret until the Wolf port is lined up at 12 o’clock. Through the Wolf port hole, remove the screw with a Phillips screwdriver. Rotate the turret so the Wolf large hole is over the opposite screw, at 6 o’clock. Through this large hole, remove the second screw. While pulling forward, gently remove the turret assembly. To install a new turret assembly, push the desired turret into the hole in the bezel. Rotate the turret until the Wolf port is lined up at 12 o’clock. Through the Wolf port, install one of the flat head screws (supplied with the turret). Rotate the turret so the Wolf large hole is over the opposite screw, at 6 o’clock. Through the Wolf port, install the second screw. When secured, the turret should not rock but should rotate freely.
What are the safety precautions when using the Light Source?
Do not use or store liquids on or above the Light Source. Electric shock hazard. If the unit is not functioning properly, DO NOT OPEN. Explosion Hazard. Do not use in the presence of flammable anesthetics, liquids, vapors, gases or dusts. Use care not to point any light guide directly at the eye while operating the Light Source. Keep cooling vent and fans free of obstructions. FIRE HAZARD: DO NOT DRAPE OR COVER THE LIGHT SOURCE WHILE IT IS OPERATING. When the Light Source is not in use, turn off the power, or rotate the turret 1/8 turn to block light output.