Gaymar TP 702 Localized Temperature Therapy System, TP 22B, TP 22C, TP 3E, TP 12E, TP 22E, TP 26E, TP 22G Mul•T•Pad Operator's Manual

Gaymar TP 702 Localized Temperature Therapy System, TP 22B, TP 22C, TP 3E, TP 12E, TP 22E, TP 26E, TP 22G Mul•T•Pad Operator's Manual
Add to My manuals

Below you will find brief information for Localized Temperature Therapy System TP 702, Mul•T•Pad TP 22B, Mul•T•Pad TP 22C, Mul•T•Pad TP 3E, Mul•T•Pad TP 12E, Mul•T•Pad TP 22E, Mul•T•Pad TP 26E, Mul•T•Pad TP 22G. The TP 702 is a localized temperature therapy system that provides therapy by warming or cooling the enclosed water and circulating it through the Gaymar Mul•T•Pad. The pad is connected to the Gaymar T/Pump with easy-to-use Clik-Tite connectors. The Mul•T•Pad can be interconnected using Clik-Tite connectors to provide therapy to more than one body site at a time

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Localized Temperature Therapy System TP 702 Operator's Manual | Manualzz

T/PUMP

®

TP702 Localized Temperature

Therapy System

Operator's Manual [EN]

0086

www.gaymar.com

P/N 100963000 Rev A 8/08

BEFORE YOU BEGIN . . .

CONTENTS

SAFETY PRECAUTIONS ..............................1

INTRODUCTION............................................2

INDICATIONS FOR USE ...............................4

CLINICAL CONTRAINDICATIONS ............4

PUMP FEATURES ..........................................5

KEYPAD FEATURES TP702 .........................6

CLIK-TITE

®

CONNECTORS ...........................7

START-UP PROCEDURE ..............................8

SHUTDOWN PROCEDURE .........................8

THERAPY CYCLES,

PROFESSIONAL ONLY. .............................9

TROUBLESHOOTING.. .............................. 10

STORAGE / CLEANING ............................. 13

PREVENTIVE MAINTENANCE/

SERVICE....................................................15

FUNCTIONAL CHECK ............................... 16

SPECIFICATIONS ........................................ 19

WARRANTIES .............................................. 21

SYMBOLS USED WITHIN THIS

MANUAL AND ON THE PRODUCT

Warm Water Fill Line

Cold Water Fill Line

Attention, consult Operator's

Manual.

Water Levels

Water Flow

Dangerous Voltage

Type BF Applied Part

Do not penetrate with sharp object.

Length of cycle time

Continuous

OPERATOR'S MANUAL

TP702 T/PUMP

BEFORE YOU BEGIN . . .

Read and understand this T/PUMP OPERATOR'S

MANUAL and all SAFETY PRECAUTIONS

(see page 3) prior to using the T/Pump.

Only qualified medical service personnel should repair or perform function tests on the T/Pump. Contact your dealer or

Gaymar's Technical Service Department for assistance:

Telephone: 716 662-2551

RECEIVING INSPECTION

Check the shipping carton for damage immediately upon receipt. If package damage is discovered, the device should be unpacked with the carrier's agent present. Any claims for shortage or damage must be filed with the delivering carrier by the purchaser. Do not return pumps that were damaged in shipment to GAYMAR without contacting our Technical Service

Department for advice (see phone numbers below). If damaged goods are returned to GAYMAR without notifying the carrier,

GAYMAR will assume the repairs will be made at the customer's expense.

TO RETURN PUMPS TO FACTORY FOR REPAIR

OR EXCHANGE

Merchandise returned to GAYMAR must be accompanied by a

Return Goods Number (RG#), issued by GAYMAR, authorizing goods to be returned. Contact your local dealer or Gaymar's

International Department at:

International: +1 716-662-8636

Fax: +1 716-662-0730

Provide the model, serial number, and detailed nature of the problem. You will be given a Return Goods Number (RG#).

The serial number can be found on the bottom of the T/Pump.

T/PUMP, T/PAD, Mul•T•Pad, Clik-Tite, and Gaymar are registered trademarks of Gaymar Industries, Inc. U. S. PATENT 4,068,870

© 2008. Gaymar Industries, Inc. All rights reserved.

OPERATOR'S MANUAL

TP702 T/PUMP

SAFETY

PRECAUTIONS

www.gaymar.com

SAFETY PRECAUTIONS

DANGER

• Risk of explosion. Do not use in the presence of flammable anesthetics.

• Risk of electric shock. Disconnect power before servicing the

WARNING

• This device pumps temperature controlled water through a pad. Set the pad temperature only as directed by and under the guidance of the prescriber.

• Check the skin integrity of the body surface to which therapy is applied. Evaluate patient response to temperature application.

• Check patient's skin for adverse reactions every 30 minutes or as directed by the prescriber.

• Failure to adhere to these warnings could result in patient injury.

• The following Groups/Conditions require additional surveillance:

Group/Condition at risk

Pediatric patients: The portion of an infant’s skin surface in contact with a pad, in relationship to their body mass, can potentially affect their body temperature.

Potential Injury

Hyperthermia/

Hypothermia

Patients with impaired circulation Ischemia

Areas of application under pressure Ischemia

• Only qualified medical service personnel should repair the

T/Pump. Improper repair may result in death or serious injury, equipment damage, or malfunction.

• Use T/Pump TP702 controls with Mul•T•Pads. For catalog numbers and descriptions, see page 5.

• Do not place additional heat sources between the patient and pad.

Skin damage may result.

CAUTION

• Federal law restricts this device to sale by or on the order of a physician.

• Do not cover the control unit with blankets, pillows or other insulating materials. Air flow is required to maintain system performance.

1

2

INTRODUCTION

INTRODUCTION

Connecting the Pads

OPERATOR'S MANUAL

TP702 T/PUMP

Heat therapy is effective in the dilation of blood vessels, thereby increasing the blood flow to the heated area. Heat therapy has a variety of uses, the most common being treatment of aches and pains in joints and muscles.

Cooling therapy assists in vasoconstriction, decreasing blood flow and decreasing the metabolism in the affected area. Cooling therapy is applied in the acute phase of injury minimizing blood loss, inflammation of the tissue, and can be effective in pain management.

The GAYMAR T/Pump

®

Localized Temperature Therapy System provides therapy by warming or cooling the enclosed water, and circulating it through the Gaymar Mul•T•Pad. The pad is connected to the Gaymar

T/Pump with easy-to-use Clik-Tite

®

connectors.

The Mul•T•Pad provides the interface for delivering the temperature therapy. The unique button design allows water to flow and provides trouble-free operation when the pad is folded to form a customized fit.

This reduces the number of pads your facility must keep in inventory.

The pads are applied to the part of the body requiring therapy, and the circulating water maintains the pad at the setpoint temperature. The setpoint temperature can be locked to prevent tampering.

Figure 1A: Localized Temperature

Therapy System with Single Pad with the TP702.

The Mul•T•Pads can be interconnected using Clik-Tite

®

connectors to provide therapy to more than one body site at a time.

Figure 1B: Localized

Temperature Therapy

System with Multiple Pads.

www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

Catalog Descriptions

INTRODUCTION

Catalog # Product Name

TP702 Professional Control Unit

TP22B Mul•T•Pad: 15"w x 22"l (38cm x 56cm)

All Polymer. 10 per carton

TP22C Mul•T•Pad: 15"w x 22"l (38cm x 56cm)

Nonwoven fabric on one side, pliable polymer on the other side. 1 per carton

TP3E Mul•T•Pad: 3"w x 23"l (8cm x 58cm)

Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton

TP12E Mul•T•Pad: 13"w x 18"l (33cm x 46cm)

Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton

TP22E Mul•T•Pad: 15"w x 22"l (38cm x 56cm)

Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton

TP26E Mul•T•Pad: 18"w x 26"l (46cm x 66cm)

Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton

TP22G Mul•T•Pad: 15"w x 22"l (38cm x 56cm)

Heavy Polymer, Reuseable. 10 per carton www.gaymar.com

To order any of these products contact your dealer or Gaymar's

International Department at:

International:

Fax:

+1 716-662-8636

+1 716-662-0730

Or, visit our website at www.gaymar.com

3

4

INDICATIONS AND CONTRAINDICATIONS

INDICATIONS FOR USE

OPERATOR'S MANUAL

TP702 T/PUMP

Localized temperature therapy is recommended in treating the following applications:

Orthopedic Conditions such as:

acute injuries, chronic pain, lower back pain, muscle spasm and strains

Skin Trauma such as: abscesses, boils, bruises, burns and contusions

Cold Indications muscle spasm contusions tendonitis pain management

Other conditions such as:

chronic arthritis, neuritis, phlebitis, tendonitis and I.V. infiltration, infection and localized pain.

Other applications only as prescribed.

OK for Use With:

• Non-acute traumatized tissue.

• Impaired mental status.

• Insensate body surface.

• O

2

therapy. However, if oxygen tent is in use, do not use pump inside tent.

CLINICAL

CONTRAINDICATIONS

Heating contraindications are:

Application to a body surface with compromised blood flow

(Ischemia, area under pressure, arterial insufficiency).

Application to a patient with an increased tendency to bleeding

(aggravates potential for hemorrhage).

Application to a body surface with possibility of malignancy (tissue metabolism is increased and therefore, the growth potential of the malignant tissues).

Treatment of hematoma within first 24-48 hours (potential for re-bleeding and hemorrhage).

Recent sprain or fracture (acute inflammatory response).

In combination with topical solutions whose toxicity may be affected by the application of heat.

In combination with other heat sources.

Cooling contraindications are:

• Application to a body surface with compromised blood flow or hypersensitivity to cold. (Ischemia, Reynauds Syndrome).

www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

FEATURES

PUMP FEATURES

Easy to Use

Keypad

Attached hose

See KEYPAD FEATURES on page 8.

Flow indicator

Warm/Cool

Delivery

Therapy Cycles

10 ft (305 cm) dual hose. Connectors allow pads to be connected to the pump (Figures 1A and 1B, page 4).

Indicates no flow. Turns off heater if pump is tipped.

Four temperature setpoints on the TP702.

On/Standby

Button

Over Temp Safety

Thermostat

Self Check

Hose/cord

Management

Comfortable

Handle Design

Dual Micro

Processor

Tethered Easy-

Open Cap

Handle Vents

Easy to Use Keypad

Choose from 20-minute, 30-minute, or Continuous cycles.

Indicates power is supplied to the unit.

Limit thermostat shuts off pump and heater if the high temperature limit is exceeded.

Automatic system check at startup.

Convenient and easy storage areas for hose and cord.

Vents in Handle

Designed for a more comfortable grip when moving the pump.

Two electronic circuits, one over temperature sensing circuit.

Prevents misplacing the cap. Only 1/4 of a turn is needed to remove or secure the cap.

The vents in the handle allow air flow to keep the motor and heater inside the unit cool.

On/Standby Button

Comfortable Handle

Flow Indicator

Tethered Easy-Open Cap

Cord storage

Temperature Setpoints for warm or cool delivery

Therapy Time Cycles)

Hose storage

Hose Connections

Figure 2: T/Pump Features www.gaymar.com

5

6

FEATURES

KEYPAD FEATURES

TP702

30

20

OPERATOR'S MANUAL

TP702 T/PUMP

T/Pump

Professional

®

!

C

42

38

35

10

F

107

100

95

50

Figure 3: TP702 Keypad

Indicator/

Warning

Light

Control Panel

100269

See the Troubleshooting section on page 12.

T/Pump Professional CE

Water Flow

Check hoses or clamps for kinks or occlusions.

Setpoints

Water Level

Check water level.

Press the button at the bottom of the setpoint indicator to toggle through the four setpoints. Temperatures are identified in °C and °F.

Setpoint

Lock

Prevents tampering.

Press and hold for 2 seconds to lock or unlock the setpoint.

Therapy

Cycles Continuous cycle,

30-minute cycle

20-minute cycle.

Green indicates the unit is on.

Yellow indicates power is supplied to the unit but the unit is not on. www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

CLIK-TITE

®

CONNECTORS

Note: Refer to Figure 1B on page 4 when connecting multiple pads.

OPERATING INSTRUCTIONS

The TP702 T/Pump is supplied with Clik-Tite

®

connectors.

To connect and disconnect Clik-Tite

®

connectors from hose to pad:

1.

2.

Insert male fittings into female fittings with a twisting motion

(Figures 4A and 4B).

When fittings are fully inserted, snap locking ring into place

(Figures 4C and 4D).

3.

4.

To disconnect, reverse the procedure.

To open or close the hose pinch clamps:

• Open the clamp by pushing the serrated end (Figure 5B).

• Close the clamp by pressing the clamp together (Figure 5C).

www.gaymar.com

5B

5C

Figure 4: Clik-Tite

®

Connectors

7

OPERATING INSTRUCTIONS

START-UP PROCEDURE

OPERATOR'S MANUAL

TP702 T/PUMP

1. Before filling the pump, always attach a pad to the connector hose or close the clamps on the connector hose ends. Make sure that there are no kinks in the hose or pad. Open the hose clamps.

2. Open the fill cap on top of the pump.

3. To fill for cooling: a. Fill with cold water to the Cooling water line. b. Fill with ice to the full capacity of the reservoir.

4. To fill for heating, fill with room temperature water to the Heating water line.

5. Plug the T/Pump into a properly grounded Hospital Grade receptacle.

6. Press the On/Standby button.

7.

The light next to the selected temperature begins to flash.

Use the keypad to set the temperature as directed by the prescriber.

After setting the temperature, press and hold the lock Temperature

Setpoint button for 2 seconds to lock the setpoint.

Note: If you toggle past the desired setpoint, keep toggling to start at the beginning of the setpoint column.

If warming, the selected water temperature will be reached in approximately 11 minutes and the light next to the selected temperature becomes steady.

Check the water level. If it drops below the operating level, add water. 8.

9. Apply the Mul•T•Pad to the patient as prescribed. Follow the

Mul•T•Pad instructions.

10. Position the pump at or above the level of the pad.

Note: If the pump is placed below the pad(s), water will drain into the pump when it is shut off. If the pump has been overfilled or if multiple pads are connected, excess water can leak.

SHUTDOWN PROCEDURE

1.

2.

3.

4.

5.

6.

7.

Press the On/Standby button so that the Standby light is lit.

Unplug the T/Pump.

Close all hose clamps.

Disconnect pad(s) from pump.

To prevent water spillage, always disconnect pad from pump with connectors raised above the level of the pad and pump.

Coil the hose, and attach the Clik-Tite

®

connectors together on the hose (See Figure 4, Page 9), where applicable.

Secure the hose to the T/Pump using the tube set strap.

Wrap the power cord around the unit.

8 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

THERAPY CYCLES,

PROFESSIONAL ONLY

OPERATING INSTRUCTIONS

Therapy Cycles:

1.

2.

Turn the unit on, and place it into 20 (or 30) minute therapy cycle. The unit will start to heat up. Select warming period for

20 (or 30) minute cycle, LED should be solid, and the selected temperature LED is flashing.

Once the T/Pump reaches the desired temperature, the Warming

LED (Red) flashes with a short audible beep. This is to let the operator/user know the unit is at temperature and the timed therapy is starting. The warm-up period will take about 5 to 10 minutes depending on the selected Setpoint.

a. Note: If this is not the first time the unit has cycled (the

unit is already warmed up) the warm up period is about 2

minutes.

3.

4.

Once the 20 (or 30) minute period is up, the heater and pump shuts off. This stops any flow to the pad. The Setpoint and Therapy time LED's will be blinking during the off Time period. This is to inform the operator that the unit is not just in standby. During this time the pad temperature will slowly drop to the patients body temperature.

Once the Off Time period is up, the pump and heater will restart.

The patient may feel a cooling in the pad due to the drop in temperature in the hoses. Time Therapy LED goes solid, and the

Setpoint LED will flash until the system temperature reaches the

Setpoint. Once the Setpoint is reached the Red LED flashes with a short audible beep to let the operator/user know the next therapy cycle has started.

www.gaymar.com

9

TROUBLESHOOTING

TROUBLESHOOTING

Problem

T/Pump will not turn on.

T/Pump will not pump.

Flow indicator light is on.

Warning indicator & Audible alarm (Flash / Beep).

OPERATOR'S MANUAL

TP702 T/PUMP

Possible Cause

The electrical cord is not plugged into a properly grounded Hospital

Grade receptacle.

Remedy

Insert the plug fully into the properly grounded Hospital Grade receptacle.

Water level is low or reservoir is empty.

Refill with room temperature water to proper level.

Water flow to pad or hose is restricted.

Clamp is closed.

Water level is low or reservoir is empty.

T/Pump is filled with water that is too hot.

Straighten the hose.

Open the clamp.

Refill with room temperature water to proper level.

Refill with room temperature water to proper level.

A High Heat (107°F / 42

°

C) or

Cooling Setpoint was selected

(50

°

F / 10

°

C).

Loss of power while unit was in a

Therapy mode. (Possible Power

Fail.)

Indication only: A Setpoint outside body temperature range is selected.

Insert the plug fully into the receptacle, place the unit into

Standby mode, then unplug the

T/Pump. If power is removed while unit is in On-Mode, the Power Fail alarm will beep for approximately 10 minutes.

Unit is running after a 20- or 30- minute “Off “therapy cycle period, has reached the desired Setpoint, and is now timing the 20- or 30- minute

On cycle period.

Indication only to indicate an “On”

Therapy cycle period is timing.

The unit just went into, or came out of Lock mode.

Indication only.

Safety Circuit Function Test has started.

Note: This would be followed with the unit starting, while still in Standby mode.

See the Functional Check section on page 18 for details.

Note: This should only be performed by medical service personnel.

If this mode was started in error, press the On/Standby button to stop the test and go back to Standby mode with the pump off.

10 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

TROUBLESHOOTING

Problem

T/Pump running with the

Standby light on.

Warning indicator on with unit in Standby mode.

Flow indicator and Standby indicator are on with

T/Pump not pumping.

Temperature Setpoint light blinking.

Possible Cause

Safety Circuit Function Test has started.

Note: This would be followed with the unit starting, while still in Standby mode.

See the Functional Check section on page 18 for details.

Remedy

Important: This should only be performed by medical service personnel.

If this mode was started in error, press the On/Standby button to stop the test and go back to Standby mode with the pump off.

Unit shut down in an over temperature condition.

Safety Circuit Function Test completed successfully.

Empty the reservoir and refill with room temperature water.

Make sure all clamps are open.

Press the On/Standby button.

Verify flow through the pad.

The Warning light will turn off within 5 minutes.

Allow the system to cool for 60 minutes or until the Warning light turns off.

Unit detected a Flow warning for more than 5 minutes, thus goes to standby.

Reference “Flow indicator light is on” above.

Correct the problem, and press the

On/Standby to put the unit back into Run mode.

Unit is warming up to the selected setpoint.

Indication only.

Unit is in Cooling mode, for longer than 40 minutes.

Follow the shutdown procedure.

Drain the water in reservoir to ice fill level, and refill with ice. Follow the start-up procedure.

www.gaymar.com

11

TROUBLESHOOTING

OPERATOR'S MANUAL

TP702 T/PUMP

Problem

Both the Temperature and

Therapy Cycle Setpoint lights are blinking.

T/Pump will not heat.

T/Pump will not cool.

Temperature or Therapy Time buttons do not work.

Water leaks from hose connectors.

Possible Cause

Unit is in “Off” Therapy cycle time.

Remedy

Indication only.

Reservoir is empty.

Flow is blocked.

Reservoir is empty.

Flow is blocked.

Ice is depleted.

Refill with room temperature water to proper level.

Reference “Flow indicator light is on” above.

Refill with room temperature water to proper level.

Reference “Flow indicator light is on” above.

Drain excess water to Cooling water line and fill remainder of reservoir with ice.

The buttons have been locked.

Damaged O-ring.

Press and hold the lock button for two seconds.

Replace Clik–Tite

®

connector.

Locking ring on Clik–Tite

®

connector is not snapped into place (See

Figure 4 on page 9.)

Snap Clik–Tite

®

connector shut.

12 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

STORAGE / CLEANING

Storage (Short term)

Less than 1 day

Storage (Long term)

Draining

STORAGE / CLEANING

1.

2.

3.

4.

5.

6.

Close the hose clamps.

Disconnect the pad.

Connect ends of the connector hoses together, where applicable.

Open the hose clamps.

Leave water in the reservoir.

Coil and fasten the hose using the tube set strap and wrap the power cord around the unit.

1.

2.

3.

Drain the pump. (See instructions below.)

Coil the hose, rather than folding it, to prevent hose kinks.

Fasten the hose using the tube set strap and wrap the power cord around the unit.

1.

2.

3.

4.

5.

6.

7.

Disconnect the T/Pump from AC power.

Clamp the hose clamps.

Disconnect the pad or hoses from one another, keeping hoses at or above the level of the T/Pump.

Open the hose clamps.

Remove the fill cap and invert the T/Pump over a sink.

When all fluid has drained from the hoses and reservoir, replace the fill cap.

Connect the hoses together, where applicable.

www.gaymar.com

13

STORAGE / CLEANING

Institutional

(Hospital) Cleaning

Instructions

OPERATOR'S MANUAL

TP702 T/PUMP

Note: Clean and change the water monthly or more often depending on use.

5.

6.

7.

2.

3.

4.

Clean the outer surfaces of the T/Pump with one of the following:

• A damp cloth and soapy water.

1.

• A spray cleaner such as Fantastik

• A mild abrasive cleanser without bleach.

Prepare a germicidal solution according to the manufacturer's instructions. Use AirKem A-33, available from Ecolabs, Inc., 370

Wabasha, St. Paul, MN 35102 (phone: 1 800 247-5362), or from

GAYMAR, product catalog MTA33.

Drain the pump.

Connect hose set together.

Fill the reservoir to the Heating water line on the back of the reservoir.

Select the 95°F (35°C) temperature setpoint on the keypad.

Start the T/Pump, and circulate the solution for one hour.

Drain the solution and refill the pump with clean water.

NOTE: In a home environment, perform only step 2 and the

refill instructions in step 7.

Pads / Accessories

Only use Mul•T•Pads

®

. The unique button design allows optimal water flow and provides trouble-free operation when the pad is folded. This reduces the number of different sizes of pads your facility must keep in inventory.

The Mul•T•Pads with Clik-Tite

®

connectors can be interconnected to provide therapy to more than one body site at a time (Figure 1B, page 4). Refer to Catalog Descriptions on page 5 for a list of various pads and ordering information.

14 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

PREVENTIVE

MAINTENANCE / SERVICE

MAINTENANCE

DANGER

Risk of electric shock. Disconnect power before servicing.

WARNING

Only qualified medical service personnel should repair the

T/Pump in accordance with the Service Manual. Improper repair may result in death or serious injury, equipment damage, or malfunction.

Refer to functional check. To obtain a copy of the Service Manual, contact your local dealder or Gaymar's International Department or visit our web site www.gaymar.com.

www.gaymar.com

15

FUNCTIONAL CHECK

FUNCTIONAL CHECK

Operating

Temperatures &

Over-Temperature

Safety Circuit Check

16

Interval

Required Tools

Test Setup

OPERATOR'S MANUAL

TP702 T/PUMP

This section provides a check of the T/Pump’s temperature control, and over-temperature safety circuit. Follow the steps in this section carefully, paying particular attention to each step, and its expected result. If at any time the expected result can not be verified, press the On/Standby button to stop the test, then restart the procedure. If after a second attempt the expected result can not be verified, press the On/Standby button to stop the test, then unplug the T/Pump and call your dealer or

Gaymar’s Technical Service Department for assistance.

WARNING

Only qualified service personnel should perform this Functional

Check. Improperly following the test procedure may result in equipment damage.

Do not perform this Functional Check with an empty reservoir.

Damage to the T/Pump may result.

Read through and understand each step before performing the test.

To assure optimum performace, dependability and safety, the following

Functional Check should be performed once per year (or as specified in the facility’s preventive maintenance program).

TFC1 or equivalent (Thermometer with 30

°

52

°

C) range with 2

°

F (1

°

C) accuracy.

F to 125

°

F (-2

°

C to

TPT9 is used to measure the temperature and the flow of the water entering the pad.

Stop Watch for testing Over-Temperature Safety Circuit

Mul•T•Pad

Note: To order a TFC1, TPT9, and Mul•T•Pad, contact your dealer or

Gaymar's International Department.

Connect the T/Pump to the TPT9 and Mul•T•Pad as shown in Figure 6.

Thermometer

Mul•T•Pad

TPT9

Flow Indicator

Figure 6: Functional Check Setup www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

Test Procedure

Note: If at any time the Function test needs to be stopped, press the

On/Standby button.

FUNCTIONAL CHECK

Note: Stay present for the full duration of the test to observe all readings and to prevent damage to the T/Pump.

1.

2.

Attach a pad to the connector hose (See Figure 6 on page 18). Then, unkink the pad and hose. Open the hose clamps.

Open the fill cap on top of the pump.

3.

4.

5.

6.

7.

8.

9.

Fill with room temperature water to the Heating water line.

Plug the T/Pump into a properly grounded Hospital Grade receptacle.

Press the On/Standby button.

The T/Pump performs its self test of the lights and audible alarm, with light pattern and short audible beep.

The T/Pump goes to Standby mode with only the Standby light on.

Press and hold the Temperature Setpoint button.

While still holding the Temperature Setpoint button, press the On/

Standby button. Hold both buttons simultaneously for 3 seconds and then release.

The T/Pump gives a long audible beep, and the Warning light flashes.

The T/Pump goes into Operating Temperature Function Test mode, and the:

T/Pump starts pumping with the Standby light on with a flow of

9 GPH minimum.

107°F (42°C) setting is selected.

Continuous light is on.

• System disables the Wave feature to maintain a steady 107°F supply of water to the pad.

The T/Pump controls to 107

°

F (42

°

C) at the inlet to the pad for

15 minutes, the duration of the test.

After 12 minutes, verify with the TPT9 that the Temperature is at 107

°

+/- 2

°

F (42ºC+/-1°C) and flow is 9 GPH minimum.

F

Note: If the T/Pump does not reach the temperature and flow, press the On/Standby button, unplug the unit, check the pad and hoses for kinks and start over.

After 15 minutes, observe that the T/Pump starts the Over

Temperature Safety Circuit Test.

The T/Pump gives a long beep with the Warning light on. This signals the start of the Over Temperature Safety Circuit Test.

www.gaymar.com

17

FUNCTIONAL CHECK

Note: If the system does not shut down in an over temperature condition in 30 seconds, press the

On/Standby button, then unplug the T/Pump. Contact your dealer or Gaymar’s Technical Service

Department for assistance

Telephone: 716 662-2551

OPERATOR'S MANUAL

TP702 T/PUMP

10. Start the stop watch to make sure the Over Temperature condition is detected in 30 seconds.

WARNING: If this step is not followed, damage to the T/Pump may result.

The T/Pump gives a long beep and flash of the Warning light.

The pump stops. No water will be flowing to the pad.

The Heater is turned on at 100% power with no flow.

11.

The 107

°

F (42

°

C) Setpoint light starts flashing.

The 20-minute Cycle light starts flashing.

• When the T/Pump detects an Over Temperature Condition

(temperature trip point of 115

°

F

±

5°F [46

°

C

±

2°C]

)

, the T/Pump gives an audible beep and flashes the Warning light. The system then cycles and goes into Standby mode with the Warning light on.

When the Warning light turns on, stop the Stop-Watch.

12. Restart the Stop-Watch to make sure the T/Pump cools down to a point where the thermostat resets and the Warning light turns off in

60 minutes.

Or, to speed up the cool down go to step 13.

13. To speed up the cool down: a. Unplug the T/Pump.

b. Refill the reservoir with room temperature tap water.

c. Restart the unit in Cooling mode.

The system starts cooling down in Standby mode. The thermostat resets within 5 minutes.

The test is completed.

14. Shut down the T/Pump as described on page 10.

18 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

SPECIFICATIONS

SPECIFICATIONS

Classification

Size (approx.)

Weight

TP702

Class I equipment with Type BF applied part suitable for continuous operation. Not classified for protection against ingress of liquid.

Not classified for use in the presence of flammable anesthetics.

11.5” x 8” x 8”

29.2cm x 20.3cm x 20.3cm

6.5 lbs (2.9 kg) when empty

9 lbs (4.0 kg) with unit filled with water to heating level

Reservoir capacity

Flow rate

93 oz (2.75 l) maximum

9 gph (34 lph) minimum with pad attached

Ambient operating temperature

Environmental conditions for transport and storage

Temperature setpoints

60°F to 90°F (15.6°C to 32.2°C )

-20°F to 120°F (-28°C to 48°C )

At uncontrolled RH

TP702 Professional

107°F (42°C)

100°F (38°C)

95°F (35°C)

50°F (10°C)

Average temperature accuracy

±2°F at 107°F (±1°C at 42°C)

Maximum Contact Surface

Temperature

107°F (42°C)

High Limit Safety Temperature 111°F to 118°F

(44°C to 48°C)

Power cord

Current leakage

Modular: adaptable to Country of use

100 microamperes maximum

Ground resistance

Electrical requirements

Voltage (VAC)

Frequency (Hz)

Current (amps)

0.5 ohm max

230+10%

50 Hz

2.0 amperes

Certifications MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO

ELECTRICAL SHOCK, FIRE, AND MECHANICAL HAZARDS

ONLY IN ACCORDANCE WITH UL 60601-1 AND ASTM F 2196-

2002

IEC 60601-1-2 www.gaymar.com

19

WARRANTIES

WARRANTIES

PUMP

PAD, SINGLE PATIENT

USE

PAD, REUSABLE

PARTS

OPERATOR'S MANUAL

TP702 T/PUMP

GAYMAR equipment and products are warranted against defects in material and workmanship under normal use, and operation from the date of purchase, for the time periods listed below for the respective equipment and products. Except for such warranty, GAYMAR disclaims all other expressed and/or implied warranties including, but not limited to, the implied warranties of merchantability and of fitness for a particular purpose.

All labor performed and parts provided free of charge for a period of one (1) full year from the date of purchase, provided the equipment is returned with prior authorization prepaid to GAYMAR Industries.

Free replacement of product where defects in materials and/or workmanship are evident at time of delivery, provided the product is returned with prior authorization prepaid to GAYMAR Industries.

Free replacement of product where defects in materials and/or workmanship occur within 90 days from date of delivery, provided the product is returned with prior authorization prepaid to GAYMAR

Industries.

Defective parts will be exchanged free of charge where defects in materials and/or workmanship occur within 90 days from date of delivery, provided the parts are returned with prior authorization prepaid to

GAYMAR Industries.

20 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

ENGLISH

The TP702 complies with EN60601-1-2: 2001 Second Edition

(CISPR Classified as Class B, Group 1 ISM equipment)

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following pages.

Portable and mobile RF communications equipment can effect

Medical Electrical Equipment.

CAUTION !

The TP702 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the

TP702 should be observed to verify normal operation in the configuration in which it will be used.

EMC INFORMATION

www.gaymar.com

21

EMC INFORMATION

OPERATOR'S MANUAL

TP702 T/PUMP

EMC INFORMATION

Guidance and manufacturer’s declaration – electromagnetic immunity

The TP702 is suitable for use in the electromagnetic environment specified below.

The customer or the user of the TP702 should assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level

Electrostatic Discharge (ESD)

IEC 61000-4-2

Electrical fast Transient/burst

IEC61000-4-4

Surge

IEC 61000-4-5

±6 kV contact

±8 kV air

±2 kV for power supply lines

±1 kV for input/output lines

±1 kV differential mode

±2 kV common mode

±6 kV contact

±8 kV air

±2 kV for power supply lines

±1 kV for input/output lines

±1 kV differential mode

±2 kV common mode

Electromagnetic environment guidance

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.

Mains power quality should be that of a typical commercial or hospital environment

Mains power quality is that of a typical commercial and/or hospital environment

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC 61000-4-11

<5%Ut(95% dipUt) for

0,5 cycle

40%Ut (60% dop in Ut) for

5 cycles

70% Ut (30% dip in Ut) for

25 cycles

<5% Ut (>95% dip in Ut) for 5 sec.

<5%Ut(95% dipUt) for 0,5 cycle

40%Ut (60% dop in Ut)for

5 cycles

70% Ut (30% dip in Ut) for

25 cycles

<5% Ut (>95% dip in Ut) for 5 sec.

Mains power quality should be that of a typical commercial or hospital environment. If the user of the

TP702

requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.

Power frequency (50/60 Hz) magnetic field

IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial and/or hospital environment.

Note: U

T

is the a.c. mains voltage prior to application of the test level.

Recommended separation distances between portable and mobile RF communications equipment and the TP702.

The TP702 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the TP702 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the TP702 as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter

W

150 kHz to 80 MHz d = 1,2

P

Separation distance according to frequency of transmitter

80 MHz to 800 MHz d = 1,2

P

800 MHz to 2,5 GHz d = 2,3

P

0,01

0,1

1

0,12

0,38

1,2

0,12

0,38

1,2

0,23

0,73

2,3

10

100

3,8

12

3,8

12

7,3

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

22 www.gaymar.com

OPERATOR'S MANUAL

TP702 T/PUMP

EMC INFORMATION

TP702 EMC Information

The TP702 is suitable for use in the electromagnetic environment specified below.

The customer or the user of the TP702 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to

80 MHz

3 V/m

80 MHz to 2,5

GHz

3 Vrms

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the TP702 including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter.

Recommended Separation Distance d = 1,2

P

d = 1,2

P

80 MHz to 800 MHz d = 2,3

P

800 MHz to 2,5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a

Should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 Mhz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radiobroadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the TP702 is used exceeds the applicable RF compliance level above, the TP702 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the TP702.

b Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m

Guidance and manufacturer’s declaration – electromagnetic emissions

The TP702 is intended for use in the electromagnetic environment specified below.

The customer or the user of the TP702 should assure that it is used in such an environment

Emissions Test

RF emissions CISPR 11

Compliance

Group 1

Electromagnetic Environment

The TP702 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11 Class B The TP702 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Voltage fluctuations/Flicker emissions

IEC 61000-3-3

Class A

Complies www.gaymar.com

23

OPERATOR'S MANUAL

TP702 T/PUMP

T/PUMPS

®

AND T/PADS

®

ARE MADE IN THE USA

24

GAYMAR INDUSTRIES, INC.

10 Centre Drive

Orchard Park, NY

14127-2295

Phone:

716-662-8636

Fax:

716-662-0730 www.gaymar.com

advertisement

Key Features

  • Localized Temperature Therapy System
  • Warming or cooling water circulation
  • Mul•T•Pad interface
  • Clik-Tite connectors
  • Multiple pad connection
  • Temperature Setpoint lock
  • Therapy cycles (20, 30 min, Continuous)
  • Flow indicator
  • Over-temperature safety thermostat
  • Self check at startup

Frequently Answers and Questions

What are the temperature setpoints available on the TP702?
The TP702 Professional offers setpoints of 107°F (42°C), 100°F (38°C), 95°F (35°C), and 50°F (10°C).
How do I connect multiple pads to the TP702?
You can interconnect Mul•T•Pads using Clik-Tite connectors to provide therapy to more than one body site at a time. See Figure 1B on page 4 for guidance.
What are the different types of Mul•T•Pads available?
There are various Mul•T•Pads available, including TP22B, TP22C, TP3E, TP12E, TP22E, TP26E, and TP22G. Each pad has different features and dimensions. Refer to Catalog Descriptions on page 5 for ordering information.
How do I perform the Functional Check on the TP702?
Refer to the Functional Check section on page 18 for detailed instructions. This check should be performed once per year by qualified service personnel.
What are the warranty terms for the TP702 and Mul•T•Pads?
The pump is warranted for one year, the single-patient use pads are warranted for replacement if defective at time of delivery, reusable pads are warranted for 90 days, and parts are warranted for 90 days.

Related manuals