Gaymar TP 702 Localized Temperature Therapy System, TP 22B, TP 22C, TP 3E, TP 12E, TP 22E, TP 26E, TP 22G Mul•T•Pad Operator's Manual
Below you will find brief information for Localized Temperature Therapy System TP 702, Mul•T•Pad TP 22B, Mul•T•Pad TP 22C, Mul•T•Pad TP 3E, Mul•T•Pad TP 12E, Mul•T•Pad TP 22E, Mul•T•Pad TP 26E, Mul•T•Pad TP 22G. The TP 702 is a localized temperature therapy system that provides therapy by warming or cooling the enclosed water and circulating it through the Gaymar Mul•T•Pad. The pad is connected to the Gaymar T/Pump with easy-to-use Clik-Tite connectors. The Mul•T•Pad can be interconnected using Clik-Tite connectors to provide therapy to more than one body site at a time
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
T/PUMP
®
TP702 Localized Temperature
Therapy System
Operator's Manual [EN]
0086
www.gaymar.com
P/N 100963000 Rev A 8/08
BEFORE YOU BEGIN . . .
CONTENTS
SAFETY PRECAUTIONS ..............................1
INTRODUCTION............................................2
INDICATIONS FOR USE ...............................4
CLINICAL CONTRAINDICATIONS ............4
PUMP FEATURES ..........................................5
KEYPAD FEATURES TP702 .........................6
CLIK-TITE
®
CONNECTORS ...........................7
START-UP PROCEDURE ..............................8
SHUTDOWN PROCEDURE .........................8
THERAPY CYCLES,
PROFESSIONAL ONLY. .............................9
TROUBLESHOOTING.. .............................. 10
STORAGE / CLEANING ............................. 13
PREVENTIVE MAINTENANCE/
SERVICE....................................................15
FUNCTIONAL CHECK ............................... 16
SPECIFICATIONS ........................................ 19
WARRANTIES .............................................. 21
SYMBOLS USED WITHIN THIS
MANUAL AND ON THE PRODUCT
Warm Water Fill Line
Cold Water Fill Line
Attention, consult Operator's
Manual.
Water Levels
Water Flow
Dangerous Voltage
Type BF Applied Part
Do not penetrate with sharp object.
Length of cycle time
Continuous
OPERATOR'S MANUAL
TP702 T/PUMP
BEFORE YOU BEGIN . . .
Read and understand this T/PUMP OPERATOR'S
MANUAL and all SAFETY PRECAUTIONS
(see page 3) prior to using the T/Pump.
Only qualified medical service personnel should repair or perform function tests on the T/Pump. Contact your dealer or
Gaymar's Technical Service Department for assistance:
Telephone: 716 662-2551
RECEIVING INSPECTION
Check the shipping carton for damage immediately upon receipt. If package damage is discovered, the device should be unpacked with the carrier's agent present. Any claims for shortage or damage must be filed with the delivering carrier by the purchaser. Do not return pumps that were damaged in shipment to GAYMAR without contacting our Technical Service
Department for advice (see phone numbers below). If damaged goods are returned to GAYMAR without notifying the carrier,
GAYMAR will assume the repairs will be made at the customer's expense.
TO RETURN PUMPS TO FACTORY FOR REPAIR
OR EXCHANGE
Merchandise returned to GAYMAR must be accompanied by a
Return Goods Number (RG#), issued by GAYMAR, authorizing goods to be returned. Contact your local dealer or Gaymar's
International Department at:
International: +1 716-662-8636
Fax: +1 716-662-0730
Provide the model, serial number, and detailed nature of the problem. You will be given a Return Goods Number (RG#).
The serial number can be found on the bottom of the T/Pump.
T/PUMP, T/PAD, Mul•T•Pad, Clik-Tite, and Gaymar are registered trademarks of Gaymar Industries, Inc. U. S. PATENT 4,068,870
© 2008. Gaymar Industries, Inc. All rights reserved.
OPERATOR'S MANUAL
TP702 T/PUMP
SAFETY
PRECAUTIONS
www.gaymar.com
SAFETY PRECAUTIONS
DANGER
• Risk of explosion. Do not use in the presence of flammable anesthetics.
• Risk of electric shock. Disconnect power before servicing the
WARNING
• This device pumps temperature controlled water through a pad. Set the pad temperature only as directed by and under the guidance of the prescriber.
• Check the skin integrity of the body surface to which therapy is applied. Evaluate patient response to temperature application.
• Check patient's skin for adverse reactions every 30 minutes or as directed by the prescriber.
• Failure to adhere to these warnings could result in patient injury.
• The following Groups/Conditions require additional surveillance:
Group/Condition at risk
Pediatric patients: The portion of an infant’s skin surface in contact with a pad, in relationship to their body mass, can potentially affect their body temperature.
Potential Injury
Hyperthermia/
Hypothermia
Patients with impaired circulation Ischemia
Areas of application under pressure Ischemia
• Only qualified medical service personnel should repair the
T/Pump. Improper repair may result in death or serious injury, equipment damage, or malfunction.
• Use T/Pump TP702 controls with Mul•T•Pads. For catalog numbers and descriptions, see page 5.
• Do not place additional heat sources between the patient and pad.
Skin damage may result.
CAUTION
• Federal law restricts this device to sale by or on the order of a physician.
• Do not cover the control unit with blankets, pillows or other insulating materials. Air flow is required to maintain system performance.
1
2
INTRODUCTION
INTRODUCTION
Connecting the Pads
OPERATOR'S MANUAL
TP702 T/PUMP
Heat therapy is effective in the dilation of blood vessels, thereby increasing the blood flow to the heated area. Heat therapy has a variety of uses, the most common being treatment of aches and pains in joints and muscles.
Cooling therapy assists in vasoconstriction, decreasing blood flow and decreasing the metabolism in the affected area. Cooling therapy is applied in the acute phase of injury minimizing blood loss, inflammation of the tissue, and can be effective in pain management.
The GAYMAR T/Pump
®
Localized Temperature Therapy System provides therapy by warming or cooling the enclosed water, and circulating it through the Gaymar Mul•T•Pad. The pad is connected to the Gaymar
T/Pump with easy-to-use Clik-Tite
®
connectors.
The Mul•T•Pad provides the interface for delivering the temperature therapy. The unique button design allows water to flow and provides trouble-free operation when the pad is folded to form a customized fit.
This reduces the number of pads your facility must keep in inventory.
The pads are applied to the part of the body requiring therapy, and the circulating water maintains the pad at the setpoint temperature. The setpoint temperature can be locked to prevent tampering.
Figure 1A: Localized Temperature
Therapy System with Single Pad with the TP702.
The Mul•T•Pads can be interconnected using Clik-Tite
®
connectors to provide therapy to more than one body site at a time.
Figure 1B: Localized
Temperature Therapy
System with Multiple Pads.
www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
Catalog Descriptions
INTRODUCTION
Catalog # Product Name
TP702 Professional Control Unit
TP22B Mul•T•Pad: 15"w x 22"l (38cm x 56cm)
All Polymer. 10 per carton
TP22C Mul•T•Pad: 15"w x 22"l (38cm x 56cm)
Nonwoven fabric on one side, pliable polymer on the other side. 1 per carton
TP3E Mul•T•Pad: 3"w x 23"l (8cm x 58cm)
Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton
TP12E Mul•T•Pad: 13"w x 18"l (33cm x 46cm)
Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton
TP22E Mul•T•Pad: 15"w x 22"l (38cm x 56cm)
Nonwoven fabric on one side, pliable polymer on the other side. 20 per carton
TP26E Mul•T•Pad: 18"w x 26"l (46cm x 66cm)
Nonwoven fabric on one side, pliable polymer on the other side. 10 per carton
TP22G Mul•T•Pad: 15"w x 22"l (38cm x 56cm)
Heavy Polymer, Reuseable. 10 per carton www.gaymar.com
To order any of these products contact your dealer or Gaymar's
International Department at:
International:
Fax:
+1 716-662-8636
+1 716-662-0730
Or, visit our website at www.gaymar.com
3
4
INDICATIONS AND CONTRAINDICATIONS
INDICATIONS FOR USE
OPERATOR'S MANUAL
TP702 T/PUMP
Localized temperature therapy is recommended in treating the following applications:
Orthopedic Conditions such as:
acute injuries, chronic pain, lower back pain, muscle spasm and strains
Skin Trauma such as: abscesses, boils, bruises, burns and contusions
Cold Indications muscle spasm contusions tendonitis pain management
Other conditions such as:
chronic arthritis, neuritis, phlebitis, tendonitis and I.V. infiltration, infection and localized pain.
Other applications only as prescribed.
OK for Use With:
• Non-acute traumatized tissue.
• Impaired mental status.
• Insensate body surface.
• O
2
therapy. However, if oxygen tent is in use, do not use pump inside tent.
CLINICAL
CONTRAINDICATIONS
Heating contraindications are:
•
•
•
•
•
•
•
Application to a body surface with compromised blood flow
(Ischemia, area under pressure, arterial insufficiency).
Application to a patient with an increased tendency to bleeding
(aggravates potential for hemorrhage).
Application to a body surface with possibility of malignancy (tissue metabolism is increased and therefore, the growth potential of the malignant tissues).
Treatment of hematoma within first 24-48 hours (potential for re-bleeding and hemorrhage).
Recent sprain or fracture (acute inflammatory response).
In combination with topical solutions whose toxicity may be affected by the application of heat.
In combination with other heat sources.
Cooling contraindications are:
• Application to a body surface with compromised blood flow or hypersensitivity to cold. (Ischemia, Reynauds Syndrome).
www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
FEATURES
PUMP FEATURES
Easy to Use
Keypad
Attached hose
See KEYPAD FEATURES on page 8.
Flow indicator
Warm/Cool
Delivery
Therapy Cycles
10 ft (305 cm) dual hose. Connectors allow pads to be connected to the pump (Figures 1A and 1B, page 4).
Indicates no flow. Turns off heater if pump is tipped.
Four temperature setpoints on the TP702.
On/Standby
Button
Over Temp Safety
Thermostat
Self Check
Hose/cord
Management
Comfortable
Handle Design
Dual Micro
Processor
Tethered Easy-
Open Cap
Handle Vents
Easy to Use Keypad
Choose from 20-minute, 30-minute, or Continuous cycles.
Indicates power is supplied to the unit.
Limit thermostat shuts off pump and heater if the high temperature limit is exceeded.
Automatic system check at startup.
Convenient and easy storage areas for hose and cord.
Vents in Handle
Designed for a more comfortable grip when moving the pump.
Two electronic circuits, one over temperature sensing circuit.
Prevents misplacing the cap. Only 1/4 of a turn is needed to remove or secure the cap.
The vents in the handle allow air flow to keep the motor and heater inside the unit cool.
On/Standby Button
Comfortable Handle
Flow Indicator
Tethered Easy-Open Cap
Cord storage
Temperature Setpoints for warm or cool delivery
Therapy Time Cycles)
Hose storage
Hose Connections
Figure 2: T/Pump Features www.gaymar.com
5
6
FEATURES
KEYPAD FEATURES
TP702
30
20
OPERATOR'S MANUAL
TP702 T/PUMP
T/Pump
Professional
®
!
C
42
38
35
10
F
107
100
95
50
Figure 3: TP702 Keypad
Indicator/
Warning
Light
Control Panel
100269
See the Troubleshooting section on page 12.
T/Pump Professional CE
Water Flow
Check hoses or clamps for kinks or occlusions.
Setpoints
Water Level
Check water level.
Press the button at the bottom of the setpoint indicator to toggle through the four setpoints. Temperatures are identified in °C and °F.
Setpoint
Lock
Prevents tampering.
Press and hold for 2 seconds to lock or unlock the setpoint.
Therapy
Cycles Continuous cycle,
30-minute cycle
20-minute cycle.
Green indicates the unit is on.
Yellow indicates power is supplied to the unit but the unit is not on. www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
CLIK-TITE
®
CONNECTORS
Note: Refer to Figure 1B on page 4 when connecting multiple pads.
OPERATING INSTRUCTIONS
The TP702 T/Pump is supplied with Clik-Tite
®
connectors.
To connect and disconnect Clik-Tite
®
connectors from hose to pad:
1.
2.
Insert male fittings into female fittings with a twisting motion
(Figures 4A and 4B).
When fittings are fully inserted, snap locking ring into place
(Figures 4C and 4D).
3.
4.
To disconnect, reverse the procedure.
To open or close the hose pinch clamps:
• Open the clamp by pushing the serrated end (Figure 5B).
• Close the clamp by pressing the clamp together (Figure 5C).
www.gaymar.com
5B
5C
Figure 4: Clik-Tite
®
Connectors
7
OPERATING INSTRUCTIONS
START-UP PROCEDURE
OPERATOR'S MANUAL
TP702 T/PUMP
1. Before filling the pump, always attach a pad to the connector hose or close the clamps on the connector hose ends. Make sure that there are no kinks in the hose or pad. Open the hose clamps.
2. Open the fill cap on top of the pump.
3. To fill for cooling: a. Fill with cold water to the Cooling water line. b. Fill with ice to the full capacity of the reservoir.
4. To fill for heating, fill with room temperature water to the Heating water line.
5. Plug the T/Pump into a properly grounded Hospital Grade receptacle.
6. Press the On/Standby button.
7.
The light next to the selected temperature begins to flash.
Use the keypad to set the temperature as directed by the prescriber.
After setting the temperature, press and hold the lock Temperature
Setpoint button for 2 seconds to lock the setpoint.
Note: If you toggle past the desired setpoint, keep toggling to start at the beginning of the setpoint column.
If warming, the selected water temperature will be reached in approximately 11 minutes and the light next to the selected temperature becomes steady.
Check the water level. If it drops below the operating level, add water. 8.
9. Apply the Mul•T•Pad to the patient as prescribed. Follow the
Mul•T•Pad instructions.
10. Position the pump at or above the level of the pad.
Note: If the pump is placed below the pad(s), water will drain into the pump when it is shut off. If the pump has been overfilled or if multiple pads are connected, excess water can leak.
SHUTDOWN PROCEDURE
1.
2.
3.
4.
5.
6.
7.
Press the On/Standby button so that the Standby light is lit.
Unplug the T/Pump.
Close all hose clamps.
Disconnect pad(s) from pump.
To prevent water spillage, always disconnect pad from pump with connectors raised above the level of the pad and pump.
Coil the hose, and attach the Clik-Tite
®
connectors together on the hose (See Figure 4, Page 9), where applicable.
Secure the hose to the T/Pump using the tube set strap.
Wrap the power cord around the unit.
8 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
THERAPY CYCLES,
PROFESSIONAL ONLY
OPERATING INSTRUCTIONS
Therapy Cycles:
1.
2.
Turn the unit on, and place it into 20 (or 30) minute therapy cycle. The unit will start to heat up. Select warming period for
20 (or 30) minute cycle, LED should be solid, and the selected temperature LED is flashing.
Once the T/Pump reaches the desired temperature, the Warming
LED (Red) flashes with a short audible beep. This is to let the operator/user know the unit is at temperature and the timed therapy is starting. The warm-up period will take about 5 to 10 minutes depending on the selected Setpoint.
a. Note: If this is not the first time the unit has cycled (the
unit is already warmed up) the warm up period is about 2
minutes.
3.
4.
Once the 20 (or 30) minute period is up, the heater and pump shuts off. This stops any flow to the pad. The Setpoint and Therapy time LED's will be blinking during the off Time period. This is to inform the operator that the unit is not just in standby. During this time the pad temperature will slowly drop to the patients body temperature.
Once the Off Time period is up, the pump and heater will restart.
The patient may feel a cooling in the pad due to the drop in temperature in the hoses. Time Therapy LED goes solid, and the
Setpoint LED will flash until the system temperature reaches the
Setpoint. Once the Setpoint is reached the Red LED flashes with a short audible beep to let the operator/user know the next therapy cycle has started.
www.gaymar.com
9
TROUBLESHOOTING
TROUBLESHOOTING
Problem
T/Pump will not turn on.
T/Pump will not pump.
Flow indicator light is on.
Warning indicator & Audible alarm (Flash / Beep).
OPERATOR'S MANUAL
TP702 T/PUMP
Possible Cause
The electrical cord is not plugged into a properly grounded Hospital
Grade receptacle.
Remedy
Insert the plug fully into the properly grounded Hospital Grade receptacle.
Water level is low or reservoir is empty.
Refill with room temperature water to proper level.
Water flow to pad or hose is restricted.
Clamp is closed.
Water level is low or reservoir is empty.
T/Pump is filled with water that is too hot.
Straighten the hose.
Open the clamp.
Refill with room temperature water to proper level.
Refill with room temperature water to proper level.
A High Heat (107°F / 42
°
C) or
Cooling Setpoint was selected
(50
°
F / 10
°
C).
Loss of power while unit was in a
Therapy mode. (Possible Power
Fail.)
Indication only: A Setpoint outside body temperature range is selected.
Insert the plug fully into the receptacle, place the unit into
Standby mode, then unplug the
T/Pump. If power is removed while unit is in On-Mode, the Power Fail alarm will beep for approximately 10 minutes.
Unit is running after a 20- or 30- minute “Off “therapy cycle period, has reached the desired Setpoint, and is now timing the 20- or 30- minute
On cycle period.
Indication only to indicate an “On”
Therapy cycle period is timing.
The unit just went into, or came out of Lock mode.
Indication only.
Safety Circuit Function Test has started.
Note: This would be followed with the unit starting, while still in Standby mode.
See the Functional Check section on page 18 for details.
Note: This should only be performed by medical service personnel.
If this mode was started in error, press the On/Standby button to stop the test and go back to Standby mode with the pump off.
10 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
TROUBLESHOOTING
Problem
T/Pump running with the
Standby light on.
Warning indicator on with unit in Standby mode.
Flow indicator and Standby indicator are on with
T/Pump not pumping.
Temperature Setpoint light blinking.
Possible Cause
Safety Circuit Function Test has started.
Note: This would be followed with the unit starting, while still in Standby mode.
See the Functional Check section on page 18 for details.
Remedy
Important: This should only be performed by medical service personnel.
If this mode was started in error, press the On/Standby button to stop the test and go back to Standby mode with the pump off.
Unit shut down in an over temperature condition.
Safety Circuit Function Test completed successfully.
Empty the reservoir and refill with room temperature water.
Make sure all clamps are open.
Press the On/Standby button.
Verify flow through the pad.
The Warning light will turn off within 5 minutes.
Allow the system to cool for 60 minutes or until the Warning light turns off.
Unit detected a Flow warning for more than 5 minutes, thus goes to standby.
Reference “Flow indicator light is on” above.
Correct the problem, and press the
On/Standby to put the unit back into Run mode.
Unit is warming up to the selected setpoint.
Indication only.
Unit is in Cooling mode, for longer than 40 minutes.
Follow the shutdown procedure.
Drain the water in reservoir to ice fill level, and refill with ice. Follow the start-up procedure.
www.gaymar.com
11
TROUBLESHOOTING
OPERATOR'S MANUAL
TP702 T/PUMP
Problem
Both the Temperature and
Therapy Cycle Setpoint lights are blinking.
T/Pump will not heat.
T/Pump will not cool.
Temperature or Therapy Time buttons do not work.
Water leaks from hose connectors.
Possible Cause
Unit is in “Off” Therapy cycle time.
Remedy
Indication only.
Reservoir is empty.
Flow is blocked.
Reservoir is empty.
Flow is blocked.
Ice is depleted.
Refill with room temperature water to proper level.
Reference “Flow indicator light is on” above.
Refill with room temperature water to proper level.
Reference “Flow indicator light is on” above.
Drain excess water to Cooling water line and fill remainder of reservoir with ice.
The buttons have been locked.
Damaged O-ring.
Press and hold the lock button for two seconds.
Replace Clik–Tite
®
connector.
Locking ring on Clik–Tite
®
connector is not snapped into place (See
Figure 4 on page 9.)
Snap Clik–Tite
®
connector shut.
12 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
STORAGE / CLEANING
Storage (Short term)
Less than 1 day
Storage (Long term)
Draining
STORAGE / CLEANING
1.
2.
3.
4.
5.
6.
Close the hose clamps.
Disconnect the pad.
Connect ends of the connector hoses together, where applicable.
Open the hose clamps.
Leave water in the reservoir.
Coil and fasten the hose using the tube set strap and wrap the power cord around the unit.
1.
2.
3.
Drain the pump. (See instructions below.)
Coil the hose, rather than folding it, to prevent hose kinks.
Fasten the hose using the tube set strap and wrap the power cord around the unit.
1.
2.
3.
4.
5.
6.
7.
Disconnect the T/Pump from AC power.
Clamp the hose clamps.
Disconnect the pad or hoses from one another, keeping hoses at or above the level of the T/Pump.
Open the hose clamps.
Remove the fill cap and invert the T/Pump over a sink.
When all fluid has drained from the hoses and reservoir, replace the fill cap.
Connect the hoses together, where applicable.
www.gaymar.com
13
STORAGE / CLEANING
Institutional
(Hospital) Cleaning
Instructions
OPERATOR'S MANUAL
TP702 T/PUMP
Note: Clean and change the water monthly or more often depending on use.
5.
6.
7.
2.
3.
4.
Clean the outer surfaces of the T/Pump with one of the following:
• A damp cloth and soapy water.
1.
• A spray cleaner such as Fantastik
• A mild abrasive cleanser without bleach.
Prepare a germicidal solution according to the manufacturer's instructions. Use AirKem A-33, available from Ecolabs, Inc., 370
Wabasha, St. Paul, MN 35102 (phone: 1 800 247-5362), or from
GAYMAR, product catalog MTA33.
Drain the pump.
Connect hose set together.
Fill the reservoir to the Heating water line on the back of the reservoir.
Select the 95°F (35°C) temperature setpoint on the keypad.
Start the T/Pump, and circulate the solution for one hour.
Drain the solution and refill the pump with clean water.
NOTE: In a home environment, perform only step 2 and the
refill instructions in step 7.
Pads / Accessories
Only use Mul•T•Pads
®
. The unique button design allows optimal water flow and provides trouble-free operation when the pad is folded. This reduces the number of different sizes of pads your facility must keep in inventory.
The Mul•T•Pads with Clik-Tite
®
connectors can be interconnected to provide therapy to more than one body site at a time (Figure 1B, page 4). Refer to Catalog Descriptions on page 5 for a list of various pads and ordering information.
14 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
PREVENTIVE
MAINTENANCE / SERVICE
MAINTENANCE
DANGER
Risk of electric shock. Disconnect power before servicing.
WARNING
Only qualified medical service personnel should repair the
T/Pump in accordance with the Service Manual. Improper repair may result in death or serious injury, equipment damage, or malfunction.
Refer to functional check. To obtain a copy of the Service Manual, contact your local dealder or Gaymar's International Department or visit our web site www.gaymar.com.
www.gaymar.com
15
FUNCTIONAL CHECK
FUNCTIONAL CHECK
Operating
Temperatures &
Over-Temperature
Safety Circuit Check
16
Interval
Required Tools
Test Setup
OPERATOR'S MANUAL
TP702 T/PUMP
This section provides a check of the T/Pump’s temperature control, and over-temperature safety circuit. Follow the steps in this section carefully, paying particular attention to each step, and its expected result. If at any time the expected result can not be verified, press the On/Standby button to stop the test, then restart the procedure. If after a second attempt the expected result can not be verified, press the On/Standby button to stop the test, then unplug the T/Pump and call your dealer or
Gaymar’s Technical Service Department for assistance.
•
•
•
WARNING
Only qualified service personnel should perform this Functional
Check. Improperly following the test procedure may result in equipment damage.
Do not perform this Functional Check with an empty reservoir.
Damage to the T/Pump may result.
Read through and understand each step before performing the test.
To assure optimum performace, dependability and safety, the following
Functional Check should be performed once per year (or as specified in the facility’s preventive maintenance program).
•
•
•
•
TFC1 or equivalent (Thermometer with 30
°
52
°
C) range with 2
°
F (1
°
C) accuracy.
F to 125
°
F (-2
°
C to
TPT9 is used to measure the temperature and the flow of the water entering the pad.
Stop Watch for testing Over-Temperature Safety Circuit
Mul•T•Pad
Note: To order a TFC1, TPT9, and Mul•T•Pad, contact your dealer or
Gaymar's International Department.
Connect the T/Pump to the TPT9 and Mul•T•Pad as shown in Figure 6.
Thermometer
Mul•T•Pad
TPT9
Flow Indicator
Figure 6: Functional Check Setup www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
Test Procedure
Note: If at any time the Function test needs to be stopped, press the
On/Standby button.
FUNCTIONAL CHECK
Note: Stay present for the full duration of the test to observe all readings and to prevent damage to the T/Pump.
1.
2.
Attach a pad to the connector hose (See Figure 6 on page 18). Then, unkink the pad and hose. Open the hose clamps.
Open the fill cap on top of the pump.
3.
4.
5.
6.
7.
8.
9.
Fill with room temperature water to the Heating water line.
Plug the T/Pump into a properly grounded Hospital Grade receptacle.
Press the On/Standby button.
The T/Pump performs its self test of the lights and audible alarm, with light pattern and short audible beep.
The T/Pump goes to Standby mode with only the Standby light on.
Press and hold the Temperature Setpoint button.
While still holding the Temperature Setpoint button, press the On/
Standby button. Hold both buttons simultaneously for 3 seconds and then release.
The T/Pump gives a long audible beep, and the Warning light flashes.
The T/Pump goes into Operating Temperature Function Test mode, and the:
•
•
•
T/Pump starts pumping with the Standby light on with a flow of
9 GPH minimum.
107°F (42°C) setting is selected.
Continuous light is on.
• System disables the Wave feature to maintain a steady 107°F supply of water to the pad.
The T/Pump controls to 107
°
F (42
°
C) at the inlet to the pad for
15 minutes, the duration of the test.
After 12 minutes, verify with the TPT9 that the Temperature is at 107
°
+/- 2
°
F (42ºC+/-1°C) and flow is 9 GPH minimum.
F
Note: If the T/Pump does not reach the temperature and flow, press the On/Standby button, unplug the unit, check the pad and hoses for kinks and start over.
After 15 minutes, observe that the T/Pump starts the Over
Temperature Safety Circuit Test.
The T/Pump gives a long beep with the Warning light on. This signals the start of the Over Temperature Safety Circuit Test.
www.gaymar.com
17
FUNCTIONAL CHECK
Note: If the system does not shut down in an over temperature condition in 30 seconds, press the
On/Standby button, then unplug the T/Pump. Contact your dealer or Gaymar’s Technical Service
Department for assistance
Telephone: 716 662-2551
OPERATOR'S MANUAL
TP702 T/PUMP
10. Start the stop watch to make sure the Over Temperature condition is detected in 30 seconds.
WARNING: If this step is not followed, damage to the T/Pump may result.
•
•
•
The T/Pump gives a long beep and flash of the Warning light.
The pump stops. No water will be flowing to the pad.
The Heater is turned on at 100% power with no flow.
11.
•
•
The 107
°
F (42
°
C) Setpoint light starts flashing.
The 20-minute Cycle light starts flashing.
• When the T/Pump detects an Over Temperature Condition
(temperature trip point of 115
°
F
±
5°F [46
°
C
±
2°C]
)
, the T/Pump gives an audible beep and flashes the Warning light. The system then cycles and goes into Standby mode with the Warning light on.
When the Warning light turns on, stop the Stop-Watch.
12. Restart the Stop-Watch to make sure the T/Pump cools down to a point where the thermostat resets and the Warning light turns off in
60 minutes.
Or, to speed up the cool down go to step 13.
13. To speed up the cool down: a. Unplug the T/Pump.
b. Refill the reservoir with room temperature tap water.
c. Restart the unit in Cooling mode.
The system starts cooling down in Standby mode. The thermostat resets within 5 minutes.
The test is completed.
14. Shut down the T/Pump as described on page 10.
18 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
SPECIFICATIONS
SPECIFICATIONS
Classification
Size (approx.)
Weight
TP702
Class I equipment with Type BF applied part suitable for continuous operation. Not classified for protection against ingress of liquid.
Not classified for use in the presence of flammable anesthetics.
11.5” x 8” x 8”
29.2cm x 20.3cm x 20.3cm
6.5 lbs (2.9 kg) when empty
9 lbs (4.0 kg) with unit filled with water to heating level
Reservoir capacity
Flow rate
93 oz (2.75 l) maximum
9 gph (34 lph) minimum with pad attached
Ambient operating temperature
Environmental conditions for transport and storage
Temperature setpoints
60°F to 90°F (15.6°C to 32.2°C )
-20°F to 120°F (-28°C to 48°C )
At uncontrolled RH
TP702 Professional
107°F (42°C)
100°F (38°C)
95°F (35°C)
50°F (10°C)
Average temperature accuracy
±2°F at 107°F (±1°C at 42°C)
Maximum Contact Surface
Temperature
107°F (42°C)
High Limit Safety Temperature 111°F to 118°F
(44°C to 48°C)
Power cord
Current leakage
Modular: adaptable to Country of use
100 microamperes maximum
Ground resistance
Electrical requirements
Voltage (VAC)
Frequency (Hz)
Current (amps)
0.5 ohm max
230+10%
50 Hz
2.0 amperes
Certifications MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO
ELECTRICAL SHOCK, FIRE, AND MECHANICAL HAZARDS
ONLY IN ACCORDANCE WITH UL 60601-1 AND ASTM F 2196-
2002
IEC 60601-1-2 www.gaymar.com
19
WARRANTIES
WARRANTIES
PUMP
PAD, SINGLE PATIENT
USE
PAD, REUSABLE
PARTS
OPERATOR'S MANUAL
TP702 T/PUMP
GAYMAR equipment and products are warranted against defects in material and workmanship under normal use, and operation from the date of purchase, for the time periods listed below for the respective equipment and products. Except for such warranty, GAYMAR disclaims all other expressed and/or implied warranties including, but not limited to, the implied warranties of merchantability and of fitness for a particular purpose.
All labor performed and parts provided free of charge for a period of one (1) full year from the date of purchase, provided the equipment is returned with prior authorization prepaid to GAYMAR Industries.
Free replacement of product where defects in materials and/or workmanship are evident at time of delivery, provided the product is returned with prior authorization prepaid to GAYMAR Industries.
Free replacement of product where defects in materials and/or workmanship occur within 90 days from date of delivery, provided the product is returned with prior authorization prepaid to GAYMAR
Industries.
Defective parts will be exchanged free of charge where defects in materials and/or workmanship occur within 90 days from date of delivery, provided the parts are returned with prior authorization prepaid to
GAYMAR Industries.
20 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
ENGLISH
The TP702 complies with EN60601-1-2: 2001 Second Edition
(CISPR Classified as Class B, Group 1 ISM equipment)
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following pages.
Portable and mobile RF communications equipment can effect
Medical Electrical Equipment.
CAUTION !
The TP702 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the
TP702 should be observed to verify normal operation in the configuration in which it will be used.
EMC INFORMATION
www.gaymar.com
21
EMC INFORMATION
OPERATOR'S MANUAL
TP702 T/PUMP
EMC INFORMATION
Guidance and manufacturer’s declaration – electromagnetic immunity
The TP702 is suitable for use in the electromagnetic environment specified below.
The customer or the user of the TP702 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical fast Transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
Electromagnetic environment guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment
Mains power quality is that of a typical commercial and/or hospital environment
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
<5%Ut(95% dipUt) for
0,5 cycle
40%Ut (60% dop in Ut) for
5 cycles
70% Ut (30% dip in Ut) for
25 cycles
<5% Ut (>95% dip in Ut) for 5 sec.
<5%Ut(95% dipUt) for 0,5 cycle
40%Ut (60% dop in Ut)for
5 cycles
70% Ut (30% dip in Ut) for
25 cycles
<5% Ut (>95% dip in Ut) for 5 sec.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the
TP702
requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial and/or hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications equipment and the TP702.
The TP702 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the TP702 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the TP702 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
150 kHz to 80 MHz d = 1,2
P
Separation distance according to frequency of transmitter
80 MHz to 800 MHz d = 1,2
P
800 MHz to 2,5 GHz d = 2,3
P
0,01
0,1
1
0,12
0,38
1,2
0,12
0,38
1,2
0,23
0,73
2,3
10
100
3,8
12
3,8
12
7,3
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
22 www.gaymar.com
OPERATOR'S MANUAL
TP702 T/PUMP
EMC INFORMATION
TP702 EMC Information
The TP702 is suitable for use in the electromagnetic environment specified below.
The customer or the user of the TP702 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to 2,5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the TP702 including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter.
Recommended Separation Distance d = 1,2
P
d = 1,2
P
80 MHz to 800 MHz d = 2,3
P
800 MHz to 2,5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a
Should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 Mhz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radiobroadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the TP702 is used exceeds the applicable RF compliance level above, the TP702 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the TP702.
b Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m
Guidance and manufacturer’s declaration – electromagnetic emissions
The TP702 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TP702 should assure that it is used in such an environment
Emissions Test
RF emissions CISPR 11
Compliance
Group 1
Electromagnetic Environment
The TP702 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The TP702 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker emissions
IEC 61000-3-3
Class A
Complies www.gaymar.com
23
OPERATOR'S MANUAL
TP702 T/PUMP
T/PUMPS
®
AND T/PADS
®
ARE MADE IN THE USA
24
GAYMAR INDUSTRIES, INC.
10 Centre Drive
Orchard Park, NY
14127-2295
Phone:
716-662-8636
Fax:
716-662-0730 www.gaymar.com
advertisement
Key Features
- Localized Temperature Therapy System
- Warming or cooling water circulation
- Mul•T•Pad interface
- Clik-Tite connectors
- Multiple pad connection
- Temperature Setpoint lock
- Therapy cycles (20, 30 min, Continuous)
- Flow indicator
- Over-temperature safety thermostat
- Self check at startup
Frequently Answers and Questions
What are the temperature setpoints available on the TP702?
How do I connect multiple pads to the TP702?
What are the different types of Mul•T•Pads available?
How do I perform the Functional Check on the TP702?
What are the warranty terms for the TP702 and Mul•T•Pads?
Related manuals
advertisement
Table of contents
- 3 SAFETY PRECAUTIONS
- 4 INTRODUCTION
- 6 INDICATIONS FOR USE
- 6 CLINICAL CONTRAINDICATIONS
- 7 PUMP FEATURES
- 8 KEYPAD FEATURES TP
- 9 CONNECTORS
- 10 START-UP PROCEDURE
- 10 SHUTDOWN PROCEDURE
- 11 PROFESSIONAL ONLY
- 12 TROUBLESHOOTING
- 15 STORAGE / CLEANING
- 17 SERVICE
- 18 FUNCTIONAL CHECK
- 21 SPECIFICATIONS
- 23 WARRANTIES
- 20 Telephone