CTI-CRYOGENICS 8500 compressor Installation, Operation and Service Instructions
Below you will find brief information for 8500 compressor. The 8500 compressor is a vital component of a cryo-vacuum system, designed to compress helium gas, which is essential for cooling the cryopump. It ensures efficient and reliable operation of the cryopump system.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
8500 Compressor
(For Use with Cryo-Torr High-Vacuum Pumping System)
Installation, Operation and Service
Instructions
8040251
Rev. L (3/97)
View our inventory
HELIX TECHNOLOGY CORPORATION http://www.helixtechnology.com
The information in this document is believed to be accurate and reliable. However,
Helix Technology Corporation, cannot accept any financial or other responsibilities that may result from the use of this information. No warranties are granted or extended by this document.
Helix Technology Corporation reserves the right to change any or all information contained herein without prior written notice. Revisions may be issued at the time of such changes and/or deletions.
Any duplication of this manual or any of its parts without expressed written permission from
Helix Technology Corporation is strictly prohibited.
Any correspondence regarding this document should be forwarded to:
Helix Technology Corporation
Mansfield Corporate Center
Nine Hampshire Street
Mansfield, Massachusetts 02048-9171 U.S.A.
Telephone: (508) 337-5000
FAX: (508) 337-5464
The following Helix Technology Corporation trademarks and service marks may appear in this document:
CTI-Cryogenics
®
Cryo-Torr
®
Cryogem
®
On-Board
®
RetroFast
®
Value Line
™
FastRegen
™
TurboLink
™
ThinLine
™
GUTS
®
Cryodyne
®
Cryogenerator
®
Helix
®
RetroEase
TurboPlus
®
®
All other trademarks or registered trademarks are the property of their respective holders.
Copyright© 2000 Helix Technology Corporation Printed in U.S.A.
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping System)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table of Contents
Section 1 - Introduction
Section 2 - Inspection
Section 3 - Installation
Assembling the Compressor Input-Power Cable . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Electrical Preparation of Compressor Control Module . . . . . . . . . . . . . . . . . . . 3-5
Cooling Water: Flow and Pressure Requirements . . . . . . . . . . . . . . . . . . . . . . . 3-8
Section 4 - Maintenance Procedures
Suggested Unscheduled Maintenance Equipment . . . . . . . . . . . . . . . . . . . . . . . 4-4
Helium Circuit Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Cryopump Decontamination Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Compressor Decontamination Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Priming the Compressor Oil System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Appendix A - Customer Support Centers
Appendix B - Troubleshooting Procedures
P/N 8040251 iii
8500 Compressor (For Use With Cryo-Torr High-Vacuum Pumping System)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table of Contents (continued)
Appendix C - Electrical Schematic for 8500 Compressor
Appendix D - Components in the Electrical Control Module of
8500 Compressor
Appendix E - 8500 Compressor Flow Diagram
Appendix F - Multiple Cryopump Installation with Single 8500 Compressor
Appendix G - Remote Cryo-Torr High-Vacuum Pump Operation Using
8500 Compressor
iv
Figures
Figure 1-1: 8500 Compressor Front and Side Views . . . . . . . . . . . . . . . . . . . . 1-2
Figure 3-1: Block Diagram for 8500 Compressor Installation . . . . . . . . . . . . . 3-1
Figure 3-2: Removing Compressor Pump Shipping Supports . . . . . . . . . . . . . . 3-2
Figure 3-3: Stripping Compressor Input-Power Cable Insulation . . . . . . . . . . . 3-3
Figure 3-4: Input Power Cable Receptacle Plug . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Figure 3-5: Wiring Diagram for Standard and Grounded Delta Configurations 3-5
Figure 3-6: Preparing the 8500 Compressor Control Module- . . . . . . . . . . . . . 3-6
Figure 3-7: 8500 Compressor Cooling Water Flow and Pressure Requirements 3-8
Figure 3-8: 8500 Compressor Water Discharge Temperature Increase (°F) . . 3-10
Figure 4-1: Disconnecting the Adsorber Self-Sealing Couplings . . . . . . . . . . . 4-2
Figure C-1: Electrical Schematic for 8500 Compressor, P/N 8031351 Rev. D C-3
Figure D-1: Components in the Electrical Control Module of 8500 Compressor D-2
Figure E-1: Flow Diagram of 8500 Compressor . . . . . . . . . . . . . . . . . . . . . . . . E-2
P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping System)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table of Contents (continued)
Figure F-1: Typical Multiple Cryopump Installation with 8500 Compressor . . F-1
Figure G-1: Remote High-Vacuum Pump Switching Schematic . . . . . . . . . . . G-2
Tables
Table 1-1: 8500 Compressor—Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Table 1-2: 8500 Compressor Specifications—Power Requirements Steady-State
Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Table 1-3: 8500 Compressor Specifications—General Specification . . . . . . . . 1-3
Table 2-1: Cryo-Torr High-Vacuum Pump System . . . . . . . . . . . . . . . . . . . . . 2-1
Table 3-1: Compressor Power Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Table B-1: Compressor Troubleshooting Procedures . . . . . . . . . . . . . . . . . . . . B-2
Table C-1: 1FU Through 14FU—(All Slow-Blow Type) . . . . . . . . . . . . . . . . . C-1
P/N 8040251 v
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
Section 1 - Introduction
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
General
The manual provides instructions for installing, operating and servicing the
8500 Compressor. If you are installing or operating a Cryo-Torr High-
Vacuum System you should also have available the high-vacuum manual that applies to your particular system:
• 8040240 Cryo-Torr 100, 7, 8 and 8F High-Vacuum Pumps
• 8040252 Cryo-Torr 10 High-Vacuum Pump
• 8040219 Cryo-Torr 250F High-Vacuum Pump
• 8040259 Cryo-Torr 400 High-Vacuum Pump
• 8040266 Cryo-Torr 500 High-Vacuum Pump
The manuals for a system cover two basic components: the high-vacuum pump and the compressor. Each manual outlines the details necessary for installation, operation and servicing of that component. A manual is shipped with each system component (high-vacuum pump and compressor).
When you purchase a system, you will receive the two manuals necessary for system installation, plus a loose-leaf binder with index tab separators, allowing you to compile a complete indexed notebook.
Installation, Operation and Service Instructions
Installation, Operation and Service Instructions for your 8500 Compressor provide easily accessible information. All personnel with installation, operation, and servicing responsibilities should become familiar with the contents of these instructions to ensure high quality, safe, reliable performance.
Unit must be wired by an authorized electrician in accordance with the national electrical code, ANSI/NFPA 70-1987, as well as the local codes
P/N 8040251 1-1
Introduction
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
A
B
C
D
Figure 1-1: 8500 Compressor Front and Side Views
Table 1-1: 8500 Compressor—Dimensions
Location
A
B
C
D
Compressor Dimension
Inches (MM)
19.52 max. (496 mm)
21.50 max. (546 mm)
24.24 max. (616 mm)
2.30 (55 mm)
Table 1-2: 8500 Compressor Specifications—Power Requirements Steady-State Conditions
Compressor
8500, Part No.
8031348G001
8500, Part No.
8031348G002
Volts
208/230
200
460
380
Hz
60
50
60
50
Phase
3
3
3
3
Full Load
Current
(Amps)
16
14
8
8
Operating
Votage Range
(Volts)
198-250
190-230
395-506
342-457
Maximum
Phase Run
Current
(Amps)
23
22
13
13
1-2 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table 1-3: 8500 Compressor Specifications—General Specification
Specification Description
340 lbs (154 kg) approximate
Weight
Weight (shipping)
Power consumption
Compressor input-power cable (customer-supplied)
Helium pressure
445 lbs (202 kg) approximate
5 kw, nominal operating power, 5.75 kW maximum
Nominal power factor 0.8: 7.6 kVA max.
Recommended type SO-4 conductor, 600V, neoprene jacket and 10-gauge wire. Install per Figure B.1, electrical schematic diagram, ensuring compliance with all national, state and local standards.
Static: 195-205 psig (1345 - 1415 kPa) at 70 to 80° F (21 to
27° C) Return: normal operation: 75-85 psig (517 - 590 kPa) at operating temperature.
50 to 100°F (10 to 38°C) Ambient operating temperature range
Interface Cold-head power receptacles (3): Mates with plugs on cold head power cables.
Compressor input-power receptacle: Mates with Hubbel twist lock plug No. 2723 (for Model G001). For Model 8500 Model
G002, Part No. 2433
Gas-supply connector: 1/2-inch self-sealing coupling
Gas-return connector: 1/2-inch self-sealing coupling
Remote control receptacle (24 VAC): compressor remote current: 2.7A inductive, coldhead remote current: 0.15A mates with P5 connector P/N MS3106A-14, supplied by
CTI-CRYOGENICS.
Adsorber service schedule Replace every three years.
Cooling water requirements 70°F water supply, 2.5 gpm, 14 psid (pressure drop: water inlet to water outlet).
The pressure value does not consider discharge back-pressure conditions.
Maximum/minimum water requirements
Maximum outlet temperature 100°F (38°C)
Minimum inlet temperature 40°F (4°C)
Nominal inlet pressure, maximum 100 psig
P/N 8040251 1-3
Introduction
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
This compressor has been designed to operate with water having a pH value of 6.0 to 8.0 and a calcium carbonate concentration of less than 75 parts per million (typical municipal drinking water quality). For applications of lower pH or greater hardness, water conditioning may be necessary.
1-4 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
1 7
8 9
10
11
4
3
6
5
12
13
2
18
17
16
15 14
LEGEND
1. Power ON indicator lamp.
2. Oil sight glass
3. Cooling water output.
4. Cooling water input.
5. Helium gas-return connector self-sealing coupling.
6. Helium gas-supply connector self-sealing coupling.
7. Compressor overload reset.
8. Voltage selector switches (S3).
9. Voltage selector switches (S2).
10. Elapsed time meter (ETM).
11. Cold head power circuit breakers (CB2 and CB3).
12. Cold head power cable receptacles (J2, J3, and J4).
13. Input power cable receptacle (J1).
14. Helium gas charge fitting and valve.
15. Helium return pressure gauge.
16. Safety interlock.
17. System power ON/OFF switch circuit breaker (CB1).
18. System remote control receptacle (J5) (Jumper plug installed).
Figure 1-2: 8500 Compressor
P/N 8040251 1-5
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
Section 2 - Inspection
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Packaging of the System
The Cryo-Torr High-Vacuum Pump System is packaged in three separate cartons. Listed below are the contents of each carton, depending upon which cryopump is included in the system. Note that an Installation,
Operation and Service Manual is included in the carton for the highvacuum pump and for the compressor; each manual covers the component packaged in that carton.
When installing a Cryo-Torr High-Vacuum Pump System, CTI-
CRYOGENICS recommends that as you unpack a component, you perform an inspection and the necessary tasks for system installation for the component according to the manual included with the component. Final system installation and operation will be performed following procedures in the high-vacuum pump manual that applies to your particular system.
Table 2-1: Cryo-Torr High-Vacuum Pump System
Carton Label System Component
Cryo-Torr Cryo-Torr 100, 7, 8 and 8F Cryopump
Cryo-Torr Cryo-Torr 10 Cryopump
Cryo-Torr Cryo-Torr 250F Cryopump
Cryo-Torr Cryo-Torr 400 Cryopump
Cryo-Torr Cryo-Torr 500 Cryopump
Compressor 8500 Compressor
Accessories Installation and Scheduled Maintenance Tool Kit and Accessories, P/N 8032040G009 for G001 compressor or P/N 8032040G010 for G002 compressor.
Manual Included
804024
8040252
8040219
8040259
8040266
8040251
None
P/N 8040251 2-1
Inspection
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
The Compressor
On receipt, remove the 8500 Compressor from its shipping carton and inspect the compressor for evidence of damage as described in this Section.
1. Unpackage and remove the compressor from its shipping carton.
2. Check the carton contents. It should contain: a. 8500 Compressor.
b. Two barbed fittings for compressor cooling water connections.
c. Compressor remote start connector.
d. Input power connector.
e. Oil prime manifold, P/N 8018129.
f.
8500 Compressor Manual, P/N 8040251.
g. Spare fuses.
2-2 P/N 8040251
Installation
Section 3 - Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Removing Compressor Pump
Shipping Supports
(Refer to Page 3-2)
Setting Voltage
Tap Switches
(Refer to Page 3-5)
Cooling Water Connections to Compressor
Cooling Water PH Factors and Conservation
Cooling Water Flow and Pressure Requirements
Cooling Water Temperature Rise
(Refer to Page 3-7)
Oil Circuit Stabilization
(Refer to Page 3-10)
Helium Gas Return and
Supply Connections
Verifying Helium Static Pressure
(Refer to Page 3-10)
Making Electrical
Connections
(Refer to Page 3-11)
3-1
Figure 3-1: Block Diagram for 8500 Compressor Installation
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Compressor Installation
Installation of your compressor requires no special tools other than those supplied in the Installation and Scheduled Maintenance Tool Kit.
Preparing the Compressor
1. After removing the compressor front panel, refer to Figure 3-2 and remove the front two compressor pump shipping supports.
a. Remove the front two nuts and red flat washers that secure the compressor pump tightly in position.
NOTE: Retain the two red flat washers for use during shipment.
b. Remove the two rubber grommets that are in the package attached to the compressor: install them, flat side up, onto the front two mounting posts.
c. Reinstall the two nuts and screw them down flush with the tops of the mounting posts.
d. Replace the compressor front panel.
3-2
Compressor Base
Hex Head Nut
Red Flat Washer
Hex Head Nut
Red Flat Washer
Rubber Grommets
Figure 3-2: Removing Compressor Pump Shipping Supports
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Assembling the Compressor Input-Power Cable
CAUTION
Use round cable (diameter 0.595 inch to 1.50 inch) with the CTI-
CRYOGENICS supplied compressor receptacle plug. Do not tin the conductor leads.
Assemble the compressor input-power cable using the CTI-supplied compressor receptacle plug and a 600-volt power cable that has a SO-4 conductor, 600-volt rating neoprene jacket and 10-gauge wire as follows.
1. Prepare the input power cable by stripping wires nd jacket of the cable as shown in Figure 3-3.
3-3
Figure 3-3: Stripping Compressor Input-Power Cable Insulation
2. Disassemble the receptacle plug. See Figure 3-4.
a. Loosen the two assembly screws (1) securing the receptacle plug face (2) to the cover (3) until plug face can be removed from cover.
b. Open cord clamp (5) by loosening the two clamp screws (4) securing the clamp to the cover.
NOTE: If you are using cable with a diameter greater than 0.895 inch, remove the auxiliary insert (6) from the cover.
c. Insert the stripped end of the power cable into the cover opening by starting from clamp end (6) until the stripped conductor ends are exposed.
d. Loosen and back off all terminal screws on receptacle plug face (2).
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3-4
1.
Assembly Screw
2.
Plug Face
3.
Cover
4.
Clamp Screw (2)
3.
Cord Clamp
4.
Auxiliary Insert
5.
Key
6.
Terminal Screw (4)
Figure 3-4: Input Power Cable Receptacle Plug
a. Insert all conductors into their proper wire entrance holes on receptacle plug face. See Figure 3-5. Check that there are no bare conductors exposed. Tighten terminal screws to 9-12 in.-lbs. torque.
WARNING
For installation where one of the 3-phase legs is at or near ground potential, connect that leg to terminal X on the compressor as shown in Figure 3-5.
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
PHASE A
PHASE B
PHASE C
GROUND
COMPRESSOR-RECEPTACLE PLUG
SUPPLIED BY CTI-CRYOGENICS
Z
G
X
Y
STANDARD CONFIGURATION
Figure 3-5: Wiring Diagram for Standard and Grounded Delta Configurations
3. Assemble the receptacle plug by sliding the cover over the wire assembly noting key (7). Align the key on the cover to the keyway on the plug face nd tighten the two assembly screws (1) to secure the cover and plug face. Alternately tighten the clamp screws (4) to
8-10 in.-lbs. torque.
WARNING
Insure that the ground wire is returned to the equipment grounding conductor of the circuit supplying the receptacle.
4. Refer to Multipump Installation, for correct phasing checkout procedure.
Do not connect the compressor to the power source at this time. All of the preparation must be completed and all the panels reinstalled before electrically connecting the compressor.
Electrical Preparation of Compressor Control Module
1. Refer to Table 1-2, for electrical power requirements. Then, using a voltmeter, measure the phase-to-phase voltage from the power source.
2. Prepare the compressor voltage selector switch S3 as required in accordance with Figure 3-6.
3-5 P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3-6
Voltage
Selector Switch
(S2)
Voltage
Selector Switch
(S3)
Elapsed Time
Meter
Cold Head Power
Circuit Breakers
(CB2, CB3)
System Power
ON/OFF Switch Circuit
Breaker (CB1)
RESET
Cold Head Power
Cable Receptacles
(J2, J3, J4)
System Remote
Control Receptacle
(J5)
Input Power Cable
Receptacle
(J1)
Figure 3-6: Preparing the 8500 Compressor Control Module
-
Compressor
Configuration
P/N 8031348G001
208/230 VAC, 50/60Hz
P/N 8031348G002
380 VAC, 50 Hz
460 VAC, 60 Hz
Table 3-1: Compressor Power Information
Frequency (Hz)
50
60
50
60
Line
Volatge
190-210
210-230
198-230
230-250
342-400
400-457
395-460
460-506
2 -
2*
2 -
2*
2 -
2*
2 -
2*
Settings
S2 S3
3 -
4*
3 -
4*
3 -
4*
3 -
4*
* Factory Settings
Cold-head
Voltage
137-153
131-144
145-169
144-158
126-147
125-143
145-169
144-158
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Cooling Water: Preparation
If flexible water hose connections are used, install the barbed fittings supplied with the compressor on the input and output connections:
1. Apply a light coating of standard plumbing thread sealant on the barbed fitting threads.
2. Tighten fittings on 1/2-inch FPT input and output connections. DO
NOT OVERTIGHTEN.
3. Connect flexible hoses to the fittings and secure with hose clamps.
If hard piping is desired, install the water lines directly onto the compressor
1/2-inch FPT input and output connections. DO NOT OVERTIGHTEN.
CAUTION
Check water connections for leaks.
Cooling Water: General Considerations
NOTE: Adjust your water flow to maintain an optimum discharge water temperature of 80°F with a minimum input pressure of 7 psi. For detailed water requirements, see below.
1. Cooling water must meet flow and pressure requirements as indicated in the following subsections.
2. Cooling water should have a pH value of 6.0 to 8.0 and a calciumcarbonate concentration of less than 77 ppm. The quality of typical municipal drinking water is generally acceptable. If the cooling water has a pH value lower than 6.0 or a calcium-carbonate concentration higher than 77 ppm, water conditioning may be required.
3. To conserve water, the cooling water should be shut off when the compressor is not running.
CAUTION
If cooling water below 45°F (7° is allowed to run through the compressor while the compressor is not operating, the compressor oil will change viscosity and thicken, causing the compressor to overheat and shut off at startup. in this event, repeatedly restart the compressor and allow it to run until it has shut off several times. The oil temperature will rise and thereby allow continuous compressor operation.
3-7 P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
4. Drain and purge water from the compressor before shipping it back to the factory or subjecting it to freezing conditions. Purge water from the compressor by blowing compressed air, regulated to 30 to
40 psig (200 to 275 kPa) into the compressor output connection and allowing water to exit from the water input connection.
Cooling Water: Flow and Pressure Requirements
CAUTION
If your water supply pressure falls below 7 psig due to back pressure, the compressor will overheat and shut down.
Use the two graphs in Figure 3-7 to determine the minimum acceptable cooling water supply pressure at different flow rates and temperatures.
Find the minimum pressure:
NOTE: Without consideration of back pressure.
3-8
Figure 3-7: 8500 Compressor Cooling Water Flow and Pressure Requirements
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
1. Determine the temperature of the cooling water. Allow a ±10°F to the present water temperature if a variation cannot be ascertained. Plot the high and low temperatures on the vertical axis of the lower graph.
The example describes cooling water that varies between 45° and 67° F.
2. Determine the optimum water flow rate by drawing a horizontal line from the upper temperature variation figure on the lower graph to the upper curve of the allowable operating range indicated by cross-hatching. Draw a line from this intersecting point straight down to the horizontal axis to find the optimal flow rate.
The example shows a solid arrow extending from 67°F and intersecting the allowable operating range. Dashed arrows pointing downward indicate a water flow rate of 2.5 gallons per minute.
3. Determine the cooling water supply pressure drop by drawing a straight line up from the flow rate in the lower graph to the upper graph. At the point at which this line intersects the upper graph, draw a line leftward to the vertical axis and find the water supply pressure drop.
The example shows dashed arrows extending from the lower to the upper graph. On the upper graph the dashed arrows intersect the graph curve at approximately 15 psig.
4. Determine the total cooling water supply pressure drop by adding the pressure drop determined from the graph to the water back pressure at the installation site.
Cooling Water: Temperature Rise
CAUTION
The temperature of the cooling water as it leaves the compressor should not exceed 100°F.
Use the graph in Figure 3-8 to determine the rise in cooling water temperature as it passes through the compressor. The recommended discharge temperature for the compressor is 80° F. This information is used by plant engineering personnel to determine cooling water requirements.
3-9 P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3-10
NOTE: Used for clarification of example presented in text.
Figure 3-8: 8500 Compressor Water Discharge Temperature Increase (°F)
Oil Circuit Stabilization
1. Check to insure that the cover and all panels are properly installed on the compressor.
2. With the system power ON/OFF switch in the OFF position, connect the input-power cable to the power source. If necessary, refer to Table 1-2, for electrical requirements.
3. Place the system power switch to the ON position and allow the compressor to run 15 minutes to stabilize the oil circuit. If the compressor shuts down, oil system priming may be necessary. Refer to
Section 4 - Maintenance Procedures for oil priming procedures.
Connecting the Compressor to the Cryopump
NOTE: If your application requires the use of multiple compressors on a single gas manifold, you must install an external check valve on each compressor. Order CTI-CRYOGENICS Kit P/N 8080275K004 and contact the Product Service Department for installation guidance on manifolding compressors.
Make the connections between the cryopump and compressor. See Figure
3-9.
1. Remove all dust plugs and caps from the supply fittings and return lines, compressor, and cryopump cold head. Check all fittings.
2. Connect the helium-gas return line from the gas-return connector on the rear of the compressor to the gas-return connector on the cryopump cold head.
P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3. Connect the helium-gas supply line from the gas-supply connector on the rear of the compressor to the gas-supply connector on the cryopump cold head.
4. Attach the supply and return line identification decals (CTI supplied) to their respective connecting piping ends.
Verify proper helium supply static pressure by con- firming that the helium pressure gauge reads 195 - 205 psig (1345 - 1415 kPa), in an ambient temperature range of 50 to 10° F (10 to 38° C).
If the indicated pressure is higher than 205 psig (1415 kPa), reduce the pressure as follows:
1. Remove the flare cap from the gas charge fitting located on the rear of the compressor.
2. Open the gas charge valve very slowly. Allow a slight amount of helium gas to escape until the helium pressure gauge reads 200 psig
(1380 kPa).
3. Close the gas charge valve and reinstall the flare cap.
If the indicated pressure is lower than 195 psig (1345 kPa), add helium gas as described in Section 4 - Maintenance Procedures.
Electrical Connection of Compressor
WARNING
The system power switch on the rear of the compressor must be in the
OFF position before making any and all electrical connections.
1. Connect the power cable to the rear panel of the compressor and the other end to the electrical power connector on the cryopump.
2. Plug the compressor input power cable into the power source.
3. Turn the system power switch on the rear of the compressor to the
ON position.
4. If the compressor does not start, the most likely cause is that the main power input phasing is incorrect. To correct the phasing, interchange any two of the three phase wires at the compressor input power receptacle plug. See Figure 3-5.
3-11 P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Multipump Installation
Your 8500 Compressor can be connected to more than one high-vacuum pump at a time. For example, three Cryo-Torr 8 vacuum pumps can be connected to a single 8500 Compressor.
CAUTION
For detailed instructions on multiple cryopump installation, refer to
Appendix F.
If a remote switching system is desired, refer to Appendix G for fabrication of a remote switching circuit. Remove the supplied jumper plug connected to the compressor remote connector J5 and connect the switching circuit to J5 connector. If the remote switching circuit is to be discontinued, then simply disconnect the remote switching circuit from the
J5 connector and reinstall the jumper plug.
NOTE: The compressor will not operate if the jumper plug is not reinstalled.
3-12 P/N 8040251
Installation
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
User’s
Vacuum
Chamber
Cryo-Torr 8
Helium Supply Line
Helium Return Line
Power Cable
3-13
Figure 3-9: 8500 Compressor Installation
P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Section 4 - Maintenance Procedures
WARNING
Always disconnect the compressor from all sources of electrical power before performing any maintenance procedures.
Unit must be wired by an authorized electrician in accordance with the national Electrical Code, ANSI/NFPA 70-1996, as well as the local codes. This shall include installation of a readily accessible disconnect device into the fixed wiring supplying power.
An insulated safety grounding conductor that is identical in size, insulation material and thickness to the circuit supply conductors, except that it is green with or without one or more yellow stripes is to be installed as part of the branch circuit which supplies the unit or system. The grounding conductor described is to be connected to the equipment grounding conductor of the supply circuit.
Scheduled Maintenance
The only scheduled maintenance required on the 8500 Compressor is replacement of the compressor adsorber (P/N 8080275K001) every three years.
P/N 8040251 4-1
Maintenance Procedures
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Compressor
Rear
Panel
4-2
This 1 1/8 in. wrench holds the coupling in a stationary position.
This 1 3/16 in. wrench is used to loosen the self sealing coupling connector. Note the direction of the large arrow.
This 1 1/8 in. wrench holds the coupling in a stationary position.
This 1 3/16 in. wrench is used to tighten the self sealing coupling connector. Note the direction of the large arrow.
To Disconnect The Coupling To Connect The Coupling
Figure 4-1: Disconnecting the Adsorber Self-Sealing Couplings
Removing the Compressor Adsorber
To remove the compressor adsorber see Figure 4-1 and proceed as follows:
1. Shut down the compressor.
2. Disconnect the compressor input power cable from its electrical power source and disconnect the water lines.
3. Disconnect the flex lines from the gas-return and gas-supply connectors at the rear of the compressor.
NOTE: Use the two wrenches supplied in the Installation and Scheduled
Maintenance Tool Kit to avoid loosening the body of the coupling from its adapter. Hold one wrench fast on the coupling half attached to the rear side of the compressor. Use the other wrench to loosen the coupling on the helium supply or return line.
4. Loosen the four screws that hold the compressor rear panel and remove the panel.
5. Disconnect the adsorber-inlet self-sealing coupling. See Figure 4-1.
6. Remove the bolts, nuts, and washers that secure the adsorber to the base of the compressor. Save all hardware.
7. Carefully slide the adsorber outward until it clears the compressor and remove the adsorber.
P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
WARNING
Depressurize the adsorber before disposing of it. Attach the depressurization fitting (included in the Installation and Scheduled Maintenance
Tool Kit) to the coupling half at either end of the adsorber and tighten it slowly.
Installing the Compressor Adsorber
1. Install the replacement adsorber as follows: a. Remove the dust caps from the self-sealing coupling halves at each end of the replacement adsorber.
b. Install the replacement adsorber following the steps for compressor adsorber removal in reverse order. Use the hardware saved from the previous procedure.
2. Connect the adsorber to the compressor internal piping as follows: a. Check the self-sealing connector flat rubber gasket to make sure that it is clean and properly positioned.
b. Make the first turns by hand and then firmly seal the connection using the two wrenches until the fittings “bottom”.
CAUTION
Make sure to hold fast on the left coupling nut while tightening the right coupling nut, as shown in Figure 4-1.
c. Using the two wrenches supplied in the Installation and
Scheduled Maintenance Tool Kit, make the connection quickly to minimize minor gas leakage. See Figure 4-1.
3. Replace the panel and secure it by tightening the four screws.
4. Ensure that the pressure gauge reads 195-205 psig (1345-1415 kPa). If additional gas pressure is required, follow the instructions under, Adding Helium Gas. Record the date that the adsorber was replaced, and also make a note that the next adsorber replacement is in three years.
5. Connect the flex lines to the gas-return and gas-supply connectors at the rear of the compressor.
P/N 8040251 4-3
Maintenance Procedures
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
6. Connect the water lines to the compressor fittings and secure with hose clamps. If hard piping is used, apply a light coating of plumbing thread sealant on threads and install water lines onto the compressor. DO NOT OVERTIGHTEN.
CAUTION
Check water connections for leaks.
7. Connect the cold head power cables to the compressor connectors.
8. Plug the compressor input power cable into the power source.
Unscheduled Maintenance
Suggested Unscheduled Maintenance Equipment
It is advisable to keep on hand the unscheduled maintenance equipment and disposable supplies listed below.
1. Helium, 99.999% pure.
2. Pressure regulator (0-3000/0-400 psig).
3. Maintenance manifold, P/N 8080250K003*.
4. Helium charging line terminating in a 1/4-inch female flare fitting,
P/N 7021002P001.
5. Installation and Scheduled Maintenance Tool Kit, P/N
8032040G014.
*Available from stock; consult the factory or your sales representative.
Adding Helium Gas
Use only 99.999% pure helium gas.
CAUTION
If the compressor helium pressure gauge reads 0, decontamination is required. Refer to Helium Circuit Decontamination, or contact the
Product Service Department.
1. A User-supplied helium charging line terminating in a 1/4-inch female flare fitting, and a two-stage pressure regulator rated at
0-3000/0-400 psig is required for this operation.
4-4 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
2. If you need to add helium more than once every several months, check for leaks caused by improperly connected self-sealing connections or any mechanical joint within the compressor.
There are two conditions that require the addition of helium gas:
1. Compressor not operating; helium pressure gauge reads 195 psig (1345 kPa), or below.
2. Compressor operating; helium pressure reads 75 psig (517 kPa), or below.
To add helium gas:
1. Attach a two-stage regulator (0-3000/0-400 psig) and charging line to a helium gas (99.999% pure). DO NOT OPEN THE
BOTTLE AT THIS TIME. Purge the regulator and charging lines as instructed in steps a through e below. Do not use helium gas that is less than 99.999% pure.
a. Open the regulator a small amount by turning the adjusting knob clockwise until it contacts the diaphragm, then turn approximately 1/8 to 1/4 turn more, so that the regulator is barely open.
b. Slowly open the bottle valve, and purge the regulator for 10 to 15 seconds. Turn the regulator knob counterclockwise until the helium stops flowing.
c. Connect the charge line to the helium pressure regulator.
d. Remove the flare cap of the gas charge fitting on the rear of the compressor. Loosely connect the charge line to the charge fitting.
e. Set the helium pressure regulator to 10 to 25 psig (70-125 kPa). Allow helium gas to flow through the charging line and around the loosened flare fitting for 30 seconds to purge the charging line of air. Then tighten the flarenut at the end of the charge line.
(This procedure is required to ensure that both the regulator and the charging line will be purged of air and that the air trapped in the regulator will not diffuse back into the helium bottle. For best results, CTI-
CRYOGENICS suggests a dedicated helium bottle, regulator, and line, which are never separated, for adding helium.)
2. Set the helium pressure regulator to 300 psig (2070 kPa). Depending on the compressor operating state, add helium gas:
P/N 8040251 4-5
Maintenance Procedures
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION a. If the compressor is running (approximately 2 hours operating time) under normal operating conditions, slowly open the helium charge valve on the rear of the compressor.
When the helium pressure gauge rises to 75-85 psig (517-
590 kPa) tightly close the charge valve.
b. If the compressor is not running, slowly open the helium charge valve. When the helium pressure gauge rises to 195-
205 psig (1345-1415 kPa), tightly close the charge valve.
CAUTION
Add helium gas slowly to prevent relief valve blow-off.
3. Ensure that the helium charge valve on the compressor is tightly closed. Shut off the helium pressure regulator on the helium bottle and remove the charging line from the male flare fitting. Shut off the helium gas bottle valve. Reinstall the flare cap.
Helium Circuit Decontamination
Contamination of the helium-gas circuit is indicated by sluggish or intermittent operation (ratchetting) of the cold head drive mechanism. With severe contamination the cold head drive may seize and fail to operate. One of the major sources of contamination is using helium gas of less than the required purity. When performing the decontamination process, use only
99.999% pure-helium gas, and the regulator and charging line must be properly connected and purged. This contamination procedure will remove contaminants from the cold head and/or compressor, thereby restoring system performance. The cold-trapping of contaminants inside the cold head during this procedure will also decontaminate the compressor if the contamination of the system is not severe. Separate decontamination of the compressor is required whenever the compressor has been opened to atmosphere, or the pressure dropped to zero.
Cryopump Decontamination Procedures
1. Cool down the cryopump and operate it for one to three hours.
If the system will not cool down, proceed to step 2. Operating the cryopump will isolate the contaminants by “freezing” them in the cold head. The contaminants in the helium-gas circuit of the cryopump tend to become frozen inside the cold head. The longer the cryopump is operated beyond the one-hour period, the greater is the amount of contamination that becomes isolated inside the cold head.
4-6 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
2. Shut down the compressor as follows: a. Close the Hi-Vac valve in your vacuum system.
b. Turn off the system power ON/OFF switch.
CAUTION
Exposing the cryopump to atmosphere during warm-up will cause excessive water vapor adsorption by the charcoal of the 15K array.
3. Immediately disconnect the helium-gas supply and helium-gas return lines from the gas-supply and gas-return connectors at the rear of the compressor. Leave them attached to the cold head.
4. Attach the maintenance manifold (P/N 8080250K003) to the disconnected ends of the helium-gas return and helium-gas supply lines.
5. Reduce the pressure in the cold head to a level of 45 psig by using the maintenance manifold.
6. Allow the second stage of the cold head to warm up to room temperature. Warm-up time can be reduced by purging the cryopump with warm dry argon or nitrogen gas. Using the gas heater, CTI P/N
8080250K020, will reduce warm-up time about 50 percent, and will maintain the gas temperature below the 150°F (66°C) limit.
7. Once the cryopump has reached room temperature, attach a twostage regulator (0-3000/0-400 psig) and charging line to a helium bottle (99.999% pure). DO NOT OPEN THE BOTTLE VALVE AT
THIS TIME. Purge the regulator and charging line as instructed in steps a through e under, Adding Helium Gas. Do not use helium gas that is less than 99.999% pure.
8. Perform in sequence: a. Backfill the cold head and helium-gas return and supply lines with helium to a static charge pressure of 195-205 psig
(1345-1415 kPa) by adjusting the regulator to the required pressure, and opening the valve on the manifold. Close the valve when the pressure is correct.
b. Depressurize the cold head by slowly opening the ball valve and allowing the helium to bleed out slowly. Do not reduce the pressure to less than 30 psig or the cold head may be further contaminated.
c. Perform flushing steps a and b three more times.
P/N 8040251 4-7
Maintenance Procedures
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION d. Pressurize the cold head to the static charge pressure of 195-
205 psig (1345-1415 kPa) and run the cold head drive motor for 10 to 30 seconds by actuating the controller ON/OFF switch to on.
e. Perform steps b through d three more times for a total of 16 flushes and a total of 4 drive-motor runs.
9. Verify that the cold head is pressurized to the static charge pressure of 195-205 psig (1345-1415 kPa).
10. Disconnect the maintenance manifold from the helium-gas return and helium-gas supply lines.
11. Reconnect the helium-gas return and helium-gas supply lines to the return and supply connectors at the rear of the compressor. The cryopump is now ready for operation.
Compressor Decontamination Procedures
The procedure to decontaminate a compressor is similar to the above procedure with certain exceptions.
• There is no need to operate the cryopump before decontaminating the compressor.
• The maintenance manifold and helium-gas supply and helium-gas return lines will be connected to the supply and return fittings on the compressor.
1. Open the ball valve slightly on the maintenance manifold and allow the helium to bleed out and depressurize the compressor
(if pressurized) to 30 psig.
2. Charge the compressor slowly to approximately 200 psig (1380 kPa) by opening the 1/8-inch valve on the maintenance manifold.
3. Run the compressor for about 30 seconds.
4. Repeat steps 1 and 2, one time.
5. Disconnect the maintenance manifold from the helium-gas return and helium-gas supply lines.
6. Reconnect the helium-gas return and helium-gas supply lines to the return and supply connectors on the cold head. The compressor is now ready for operation.
NOTE: After connecting the compressor to the cryopump, and operating the system for a period of time, it may be necessary to decontaminate the cryopump as some residual contamination from the compressor may become trapped in the cold head. If the entire system was reduced to zero
4-8 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
psig (a broken flex line, for example), then the cryopump and compressor would have to be decontaminated according to the Cryopump
Decontamination Procedures section.
Priming the Compressor Oil System
The oil-prime manifold (P/N 8018129) supplied with the 8500 Compressor is used to prime the compressor oil system when any of the following conditions exist:
1. The compressor has been inoperative for a period of three months or longer.
2. The compressor is not operating and the supply pressure gauge reads 0 psig. If this condition exists contact the Product Service
Department for corrective action before priming the compressor.
3. The compressor has automatically shut off because of a loss of helium supply pressure.
Priming procedures are:
1. Disconnect the helium return and supply lines from the rear of the compressor.
2. Install the oil-prime manifold on the compressor supply and return connectors.
3. Ensure that the pressure gauge reads 195-205 psig (1345-1415 kPa). If additional gas pressure is required, follow the instructions under, Adding Helium Gas.
4. Turn the compressor switch to the ON position and allow the compressor to run between 15 and 30 minutes.
5. After completing the oil-system priming turn off the compressor and remove the oil-prime manifold.
6. Reconnect the helium return and supply lines on the rear of the compressor.
P/N 8040251 4-9
Appendix A - Customer Support Centers
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix A - Customer Support Centers
Introduction
Refer to Table A-1 for the nearest Customer Support Center for technical assistance or service. North American customers may call 800-FOR-
GUTS (800-367-4887) 24 hours a day, seven days a week. All other customers must call their local Customer Support Center.
Please have the following information available when calling so that we may assist you:
• Product Part Number
• Product Serial Number
• Product Application
• Specific Problem Area
• Hours of Operation
• Equipment Type
• Vacuum System Brand/Model/Date of Manufacture
For your convenience, you may also e-mail us at:
Visit us at our corporate website:
www.helixtechnology.com
A-1
Appendix A - Customer Support Centers
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
United States and Canada
Table A-1: Customer Support Centers
United States and Canada
Helix Technology Corporation
Customer Response Center
Nine Hampshire Street
Mansfield, Massachusetts 02048 U.S.A.
Tel: 508-337-5000
Tel: 800-379-7224 (within USA)
Fax: 508-337-5169
Dial: 800-FOR-GUTS (800-367-4887)
24 hours a day, seven days a week
United States and Canada
Helix Technology Corporation
3350 Montgomery Drive
Santa Clara, California 95054 U.S.A.
Tel: 408-592-5940
Tel: 800-324-6449 (within USA)
Fax: 408-988-6630
Dial: 800-FOR-GUTS (800-367-4887)
24 hours a day, seven days a week
United States and Canada
Helix Technology Corporation
4120 Freidrich Lane
Suite 600
Austin, Texas 78744 U.S.A.
Tel: 512-912-2800
Tel: 800-324-6445 (within USA)
Fax: 512-912-2888
Dial: 800-FOR-GUTS (800-367-4887)
24 hours a day, seven days a week
Helix Technology Corporation
Colorado Operations
6450 Dry Creek Parkway
Longmont, Colorado 80503 U.S.A.
Tel: 303-443-7660
Tel: 800-776-6543 (within USA)
Fax: 303-443-2546
Dial: 800-FOR-GUTS (800-367-4887)
24 hours a day, seven days a week
A-2
Appendix A - Customer Support Centers
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table A-1: Customer Support Centers (Continued)
Germany, Italy, Denmark, Switzerland,
Holland, Norway, The Netherlands
France, Spain, Portugal, Greece
Helix Technology Corporation
Haasstrasse 15
D-64293 Darmstadt
Germany
Tel: +(49) 6151-959-55
Fax: +(49) 6151-959-57
Helix Technology
Domaine Technologique de Saclay
4, rue Rene Razel, Bât Apollo
F-91892 Orsay Cedex
France
Tel: +(33) 1-6935-2600
Fax: +(33) 1-6985-3725
Dial: +(33) 1-6935-2600
24 hours a day, seven days a week
Japan
Dial: +(49) 6151-959-55
24 hours a day, seven days a week
United Kingdom, Ireland, Sweden,
N. Ireland
Helix Technology Corporation
Fleming Road
Kirkton Campus
Livingston, West Lothian
Scotland EH54 7BN
Tel: +(44) 1-506-460017
Fax: +(44) 1-506-411122
Helix Technology Corporation
Queens Tower A 14F
3-1, Minatomirai 2-chome
Nishi-ku, Yokohama
Japan 220-6014
Tel: +(81) 45-682-5470
Fax: +(81) 45-682-5475
Dial: +(81) 45-682-5470
24 hours a day, seven days a week Dial: +(44) 1-506-460017
24 hours a day, seven days a week
Korea Taiwan
Helix Vacuum Technology Limited
4F-1 No. 81 Shui Lee Road
Hsin-Chu 300, Taiwan
Helix Technology
Zeus Company, Ltd.
Zeus Building
3-16, Yangjae-Dong, Socho-ku
Seoul 137-130, Korea
Tel: +(82) 2-577-3181/3186
Fax: +(82) 2-576-3199
Dial: +(82) 2-577-3181
24 hours a day, seven days a week
Tel: +(886) (3) 572-3999
Fax: +(886) (3) 571-7670
Dial: +(886) (3) 516-9022
24 hours a day, seven days a week
A-3
Appendix A - Customer Support Centers
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table A-1: Customer Support Centers (Continued)
Australia, New Zealand, Tasmania China
Helix Technology Corporation
Tritek International Company
Da An Square, 9A Middle Bldg
829 Yan An Middle Road
Shanghai, 200040
China
Tel: +(86) 21-6279-1389
Fax: +(86) 21-6247-1539
Dial +(86) 21-6279-1389
24 hours a day, seven days a week
Helix Technology Corporation
CTI-Cryogenics Products
AVT Services Pte. Ltd.
Unit 1, 12 Pioneer Avenue
Thornleigh NSW 2120
Sydney, Australia
Tel: +(612) 9-4810748
Fax: +(612) 9-4810910
Dial: +(612) 9-4810748
24 hours a day, seven days a week
Australia, New Zealand, Tasmania
Helix Technology Corporation
Granville-Phillips Products
Stanton Scientific
2/6 Jonson St.
Byron Bay, NSW 2481
Australia
Tel: +(61) 66-856-902
Fax: +(61) 66-858-530
Singapore, Malaysia, Philippines,
Indonesia
Helix Technology Corporation
APP Systems Services Pte Ltd.
2 Corporation Road
#06-14 Corporation Place
Singapore 2261
Tel: +(65) 268-2024
Fax: +(65) 268-6621
Dial: +(65) 268-2024
24 hours a day, seven days a week
A-4
Appendix A - Customer Support Centers
Table A-1: Customer Support Centers (Continued)
India India
Helix Technology Corporation
CTI-Cryogenics Products
Goodwill Cryogenics Enterprise
213 Nirman, Sec. 17, Vashi
New Bombay, 400703
India
Tel: +(91) 22-7632906
Fax: +(91) 22-7822769
Dial: +(91) 22-7632906
24 hours a day, seven days a week
Helix Technology Corporation
Granville-Phillips Products
Ashwini Enterprises
C-2/13/03, Sector-16, Vashi
New Bombay 400 705
India
Tel: +(91) 22-766 1107
Fax: +(91) 22-766 2631
Israel
Helix Technology Corporation
CTI-Cryogenics Products
Vacuum System & Technology Ltd.
P.O. Box 4137
Petach Tikva 49130
Israel
Tel: +(972) 3-9247710
Fax: +(972) 3-9247711
Dial: +(972) 3-9247710
24 hours a day, seven days a week
Israel
Helix Technology Corporation
Granville-Phillips Products
El-Tan Technologies
P.O. Box 1020
Hod-Hasharon 45110
Israel
Tel: +(972) 9-740-5551
Fax: +(972) 9-740-6690
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
A-5
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix B - Troubleshooting Procedures
Troubleshooting the Compressor
The compressor troubleshooting procedures are summarized in Table B-1.
Technical Inquiries
Please refer to Appendix A of this manual for a complete list of the
CTI-CRYOGENICS’ world wide customer support centers.
WARNING
Disconnect the compressor before performing any troubleshooting procedures.
The compressor pump is hot after operating. Wait for the pump to cool down before working on the inside of the compressor
Do not change or modify any compressor internal wiring circuits, this may cause failure of the compressor and cold head due to improper phasing.
P/N 8040251 B-1
Appendix B - Troubleshooting Procedures
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table B-1: Compressor Troubleshooting Procedures
Problem Possible Cause Corrective Action
1) System power ON/
OFF switch (CB1) remains in the ON position when switched on but the compressor will not run. Refer to Figure
C-1 and D-1, for identification of all electrical components.
1) Incorrect phasing of input power causing phase monitor switch to remain open.
2) No power coming from the source.
3) Insufficient voltage at power source causing phase monitor switch to remain open.
1) Check phasing of input power from the source against its schematic, Figure 3-5.
2) Check source fuses, circuit breakers, and wiring associated with the power source, and repair as needed.
3) Verify adequate phase-tophase input voltage. Refer to
Table 1-2.
4) Control circuit fuses missing or blown.
4) Check control fuses 1FU,
2FU, and 3FU. Also check control fuse 4FU. If any of these fuses are blown, contact the Product Service Department.
5) Improperly wired external remote control circuit. This will apply only if the remote control feature if being utilized.
5) Verify correct installation of remote control feature per
Appendix G.
6) Remote control jumper plug not in place. This will apply only if remote circuit is not being used.
6) Check to insure that remote jumper plug is fully seated.
See Figure C-1, for jumper plug location on J5 remote connector.
7) Incorrect control circuit voltage.
7) Check the voltage setting on switch S2/S3 per Figure 3-6.
B-2 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table B-1: Compressor Troubleshooting Procedures (Continued)
Problem Possible Cause Corrective Action
8) Excessive current drawn by the compressor pump has caused the overload relay in the motor contactor (MI) to open.
1) Loss or degradation of power coming from power source.
8) a. Reset the overload relay in the motor contactor. Refer to
Figure C-1, for its location.
b. Contact the Product Service
Department.
1) Ensure a constant supply of power per Table 1-3
2) System power ON/
OFF switch (CB1) remains in the ON position, and the compressor stops after several minutes of operation and remains off.
2) Thermal protective switches are open.
2) Check for inadequate water cooling, see Table 1-3.
3) System power ON/
OFF switch (CB1) remains in the ON position, and the compressor stops after several minutes of operation and remains off.
3) Very cold cooling water has caused a restriction of oil flow through the oil injection orifice during start-up.
1) a. Loss of oil injection has caused the oil flow thermal switch (TS2) to open.
b. The solenoid valve (SOL
1) is not opening.
c. Insufficient oil in the compressor pump.
d. The oil injection orifice is blocked.
3) Recheck for proper cooling water temperature per Table
1-3. Restart the compressor repeatedly until continuous operation is achieved.
1) a. Check the thermal switch
(TS2) for continuity.
b. Check the solenoid coil for continuity.
c. Confirm that oil is visible in the compressor sight glass.
d. Contact the Product Service
Department for assistance.
2) Excessive current drawn by the compressor pump or the solenoid valve (SOL 1) has caused the overload relay in the motor contactor
(M1) to open.
2) a. Reset the overload relay in the motor contactor, Refer to
Figure 1-1, for its location.
b. Contact the Product Service Department for assistance.
3) Control circuit fuses blown.
3) Check condition of control circuit fuses (F1, F2, F3, and F4).
Refer to Figure D-1, for location of fuses.
P/N 8040251 B-3
Appendix B - Troubleshooting Procedures
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table B-1: Compressor Troubleshooting Procedures (Continued)
Problem Possible Cause Corrective Action
4) The remote control system switch has opened. This will apply only if the remote control feature is incorporated per Appendix G, Figure G-1.
1) Intermittent loss or degradation of input power.
4) Close the remote control system switch.
1) Confirm that there is a constant supply of power per
Table 1-2.
4) System power ON/
OFF switch (CB1) remains in the on position, the compressor and all cold heads stop and recycle.
5) Cold head circuit breakers (CB2, CB3) shut off, or will not remain on.
1) Damaged cold head drive cable, connectors or drive motor.
1) Check for compressor operation with each cold-head connector (P2, P3 or P4) disconnected. Contact the
Product Service Department if the compressor operates improperly.
6) Cold head hunts or ratchets during operation.
2) Damaged components in the cold head control circuit
(K3, K4, K5, K6, K7, K8,
K9, ETM, TS1, TS2, TS3,
M1 (Coil), CB1, T1, T2,
T3.)
1) Improper voltage select settings.
2) Refer to Figure C-1, and check for damaged components and contact the Product Service
Department.
1) Low control voltage, remove jumper plug from J5 remote connector and check control voltage between pins F and G on the remote connector for correct voltage readings. If incorrect reading check setting of voltage selector switch(s) S2 or S3, refer to Figure 3-6.
B-4 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix C - Electrical Schematic for 8500
Compressor
CB1
CB2, CB3
ETM
J1
J2
J3
J4
J5
J6, J9
K3
K4
K5
K6
K7
K8
K9
1LT
M1
M1OL
PM1
S2
S3
SOL
SOL2
T1
Table C-1: 1FU Through 14FU—(All Slow-Blow Type)
Fuse Designation
1FU
2FU
3FU
4FU
14FU
G001
Amp Rating
G002
3.2
3.2
3.2
4.0
0.125
1.5
1.5
1.5
4.0
0.125
Table C-2: Legend for Figure C-1
System Power Switch and Circuit Breaker
Circuit Breakers, 3A, Cryopump Power
Elapsed Time Meter
Input Power Connector
Number One Cold-Head Cable Connector — 6 Pins
Number Two Cold-head Cable Connector — 6 Pins
Number Three Cold Head Cable Connector — 6 Pins
System Remote Connector — 14 Pins
Chassis Connectors
Voltage Select Remote Relay
System Power Answerback Relay
Thermal Lockout Relay
Coldhead Interlock Relay
Number One Cold-Head Remote Start Relay
Number Two Cold-Head Remote Start Relay
Number Three Cold-Head Remote Start Relay
Compressor Status Indicator Lamp
Motor Contactor
Motor Overload Protector
Phase Monitor
Voltage Selector Switch
Voltage Selector Switch
Solenoid Valve, Normally Closed
Solenoid Valve, Normally Open
Cold-Head Drive Transformer
P/N 8040251 C-1
Appendix C - Electrical Schematic for 8500 Compressor
T2
T3
TS1
TS2
TS3
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table C-2: Legend for Figure C-1 (Continued)
Cold-Head Drive Transformer
Control Voltage Transformer
Thermal Protective Switch — Oil Flow/Gas Discharge Temperature
Thermal Protective Switch — Oil Flow/Motor Temperature
Thermal Protective Switch — Oil /Water Temperature
C-2 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
P/N 8040251
Figure C-1: Electrical Schematic for 8500 Compressor, P/N 8031351 Rev. D
C-3
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix D - Components in the Electrical
Control Module of 8500 Compressor
P/N 8040251 D-1
Appendix D - Components in the Electrical Control Module of 8500 Compressorr
4
5
6
7 8
9
10-16
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3
2
1
17
18
19
20
21
D-2
Figure D-1: Components in the Electrical Control Module of 8500 Compressor
Table D-1: Component Callouts
Item Number
7
8
5
6
3
4
1
2
9
10
11
12
Description
Cold-Head Drive Transformer
Cold-Head Drive Transformer
Compressor Status Indicator Lamp
ETM Fuse
Control Circuit Fuse
Control Voltage Transformer
Control Circuit Fuse
Control Circuit Fuse
Control Circuit Fuse
Number Three Cold-Head Remote Start Relay
Number Two Cold-Head Remote Start Relay
Number One Cold-Head Remote Start Relay
Symbol
Designation
T2
T1
1LT
14FU
4FU
T3
3FU
2FU
1FU
K9
K8
K7
P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
Table D-1: Component Callouts (Continued)
Item Number
17
18
19
20
21
13
14
15
16
Description
Compressor Start Relay
System Power Remote Relay
Voltage Select Relay
Interlock Relay
Phase Monitor
Motor Overload
Motor Contactor
Chassis Connector
Chassis Connector
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Symbol
Designation
K5
K4
K3
K6
PM1
M1OL
M1
J9
J6
P/N 8040251 D-3
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix E - 8500 Compressor
Flow Diagram
P/N 8040251 E-1
Appendix E - 8500 Compressor Flow Diagram
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
E-2
Figure E-1: Flow Diagram of 8500 Compressor
P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix F - Multiple Cryopump
Installation with Single 8500 Compressor
Your 8500 Compressor may be used to drive a number of multiple Cryo-
Torr cryopump combinations.
Figure F-1 depicts a typical multi-cryopump installation with a 8500
Compressor. As shown in this figure, a power cable is connected from the compressor to each cold head; also, the components are helium connected in parallel (all supply fittings piped together).
If you are considering a multiple cryopump installation please contact CTI-
CRYOGENICS, U.S.A. Application Engineering Department (1-800-447-
5007) for technical assistance in the selection and sizing of the manifolds and interconnecting lines for your particular installation.
1
2
3
Cryo-Torr
Cryopump
Cryo-Torr
Cryopump
Cryo-Torr
Cryopump
Helium Return Line
Helium Supply Line
Input Power
P/N 8040251
Figure F-1: Typical Multiple Cryopump Installation with 8500 Compressor
F-1
Appendix F - Multiple Cryopump Installation with Single 8500 Compressor
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Preliminary System Requirements
Installation of the 8500 Compressor for use with a single cryopump is covered in the Installation section. Installation for multiple cryopump usage is similar, but requires that system charge pressure be established for a given high-vacuum pump system. Additionally, the following parameters must be considered before the system can be assembled.
• All system components must be assembled and be operational for system static charge pressure determination.
• All system components must have static charge pressures of 195 -
205 psig at 70°F to 80°F (21° to 27°C).
Determining System Charge Pressure
To establish the helium gas charge pressure of a multiple cryopump installation proceed as follows:
1. Assemble your multiple cryopump system components.
2. Check the compressor pressure gauge to insure static pressure does not exceed 195 - 205 psig (1345-1415 kPa).
3. If it is necessary to reduce helium gas pressure then perform the following: a. Disconnect the charging line from gas charge fitting on rear of compressor.
b. Open the gas charge valve very slowly. Allow helium gas to escape until pressure gauge reads 50-100 psig (345-690 kPa).
c. Close the gas charge valve and reinstall the charging line to the gas charge fitting, refer to step 4 for procedure.
4. If the pressure is low, attach a helium bottle, regulator, and charging line to the compressor and perform the following: a. Remove the flare cap of the gas charge fitting on the rear of the compressor.
b. Loosely attach a charging line from the helium pressure regulator on the helium bottle to the 1/4-inch male flare fitting installed on the helium charge fitting of the compressor.
NOTE: Use only 99.999% pure helium.
c. Set the helium pressure regulator to 10 to 25 psig (70-125 kPa). Allow helium gas to flow through the charging line and around the loosened flare fitting for 30 seconds to purge
F-2 P/N 8040251
8500 Compressor (For Use with Cryo-Torr High-Vacuum Pumping Systems)
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION the charging line of air. Then tighten the flare nut at the end of the charge line.
d. Slowly add helium gas until compressor pressure gauge reads 195-205 psig (1345-1415 kPa).
5. Turn on the system power ON/OFF switch.
6. Note helium pressure gauge reading immediately after startup. It should read 50-100 psig (345-690 kPa). If necessary add additional helium gas by slowly opening the helium charge valve on the rear of the compressor until the helium pressure gauge rises to 50-100 psig (345-690 kPa).
7. Allow the cryopump to operate until a cooldown temperature of
20K or less is reached.
Adjust the helium pressure if necessary as described in step 6 until the helium pressure gauge reads 80-100 psig (550-690 kPa) while the system is operating.
8. Allow the system to reach steady state. Recheck the helium pressure.
9. When steady state is achieved, shut the system off and allow the system time to reach steady state conditions at room temperature.
10. When the system reaches room temperature, the pressure reading on the compressor gauge is the system charge pressure.
NOTE: Record the compressor static pressure in your operating log. This is the static pressure for your. This is the static pressure for your particular installation and should be used for checking compressor performance or when troubleshooting the installation.
11. Ensure that the helium charge valve on the compressor is tightly closed. Then shut off the helium pressure regulator or the helium bottle. Remove the charging line from the male flare fitting and reinstall the flare cap.
P/N 8040251 F-3
8500 Compressor for Use with Cryo-Torr High-Vacuum Pumping Systems
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Appendix G - Remote Cryo-Torr
High-Vacuum Pump Operation Using
8500 Compressor
Remote Operation
An electrical schematic is provided, see Figure G-1, for fabrication of an external switching circuit. This feature provides the capability for remote
ON/OFF switching of the complete system (compressor and all cold heads) and for individual switching of electrical power to each of the cold head connectors J2, J3 and J4. It is not mandatory that all three remote switching features be utilized simultaneously. The switching circuit is connected into the control module of the compressor by means of the 14-pin system remote connector J5, which is located at the rear of the control module.
Note that a J5 jumper connector is installed on the compressor as shipped.
Refer to Figure 1-1, for location of the connector.
Assembling the Switching Circuit
Assemble the remote switching circuit to suit your installation requirements as follows:
1. Uncouple the J5 jumper connector from the compressor. Wire the remote connector using 18-gauge stranded copper wire and a remote switch or relay contact. Cryopump remote contacts must be 24 VAC, 150 ma. Minimum rating of compressor remote contacts must be 24V AC, 2.7 A inductive. Follow the schematic in Figure G-1, ensuring compliance with all national, state and local standards.
2. After fabricating and checking the remote switching circuit, connect it to the system remote connector J5.
3. Ensure that the system power ON/OFF switch on the compressor is off, and that all the remote switches are off before connecting the input power cable to the power source. The system power ON/OFF switch and circuit breaker safety features on the compressor will remain operational even with the remote switching circuit being utilized.
P/N 8040251 G-1
Appendix G - Remote Cryo-Torr High-Vacuum Pump Using 8500 Compressor
C TI
-
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
4. Turn the system power ON/OFF switch on the compressor module to the ON position and check for proper operation of the remote switching circuit.
5. If the use of the remote control switching circuit is to be discontinued, then simply disconnect the remote control switching circuit from J5 and reinstall the J5 jumper connector.
NOTE: The cryopump system will not operate if the jumper connector is not reinstalled.
G-2
Master (system)
24 VAC, 2.7 A Inductive
Coldhead No. 3
24 VAC, 150 mA
Coldhead No. 2
24 VAC, 150 mA
Coldhead No. 1
24 VAC, 150 mA
Figure G-1: Remote High-Vacuum Pump Switching Schematic
P/N 8040251
advertisement
Key Features
- Helium gas compression
- Cooling water requirements
- Compressor oil system
- Adsorber replacement
- Helium circuit decontamination
- Multipump installation
- Remote operation capabilities
Frequently Answers and Questions
What is the purpose of the 8500 compressor?
How often does the compressor adsorber need replacement?
What are the cooling water requirements for the 8500 compressor?
Related manuals
advertisement
Table of contents
- 26 General
- 26 Installation, Operation and Service Instructions
- 27 Packaging of the System
- 27 The Compressor
- 28 Compressor Installation
- 28 Preparing the Compressor
- 28 Assembling the Compressor Input-Power Cable
- 28 Electrical Preparation of Compressor Control Module
- 28 Cooling Water: Preparation
- 28 Cooling Water: General Considerations
- 28 Cooling Water: Flow and Pressure Requirements
- 28 Cooling Water: Temperature Rise
- 28 Oil Circuit Stabilization
- 28 Connecting the Compressor to the Cryopump
- 28 Electrical Connection of Compressor
- 28 Multipump Installation
- 29 Scheduled Maintenance
- 29 Removing the Compressor Adsorber
- 29 Installing the Compressor Adsorber
- 29 Unscheduled Maintenance
- 29 Suggested Unscheduled Maintenance Equipment
- 29 Adding Helium Gas
- 29 Helium Circuit Decontamination
- 29 Cryopump Decontamination Procedures
- 29 Compressor Decontamination Procedures
- 29 Priming the Compressor Oil System