ASTEC MONAIR 10, 15, 20, 25, 30 fume cupboard OPERATING MANUAL
The Astec Monair 10, 15, 20, 25, 30 range of fume cupboards is designed to provide operator safety with an easy-to-use cabinet. The main work area has a face velocity of >0.5 m/sec to provide containment of fumes. The work surface is polypropylene and will contain spillages. The head unit contains the mains switch, low airflow alarms, carbon filter and optional lights and safety filter. The five units in the range vary in filter volume and size only. The Monair range of fume cupboards are suitable for use in a variety of laboratory settings, including schools, universities, and research facilities.
Advertisement
Advertisement
Caution:
Please read this Manual thoroughly.
If you do not understand or are unsure of any information contained in this Manual please contact your Agent or Astec Microflow Ltd. for clarification.
OPERATING MANUAL
ASTEC
MONAIR RANGE
(10, 15, 20, 25, 30)
Ref: CC/PM/06/00
Astec Microflow Limited
30/31 Lynx Crescent
Weston-super-Mare
North Somerset
BS24 9BP
Tel: 01934 410500
Fax: 01934 410542
http://www.astec-microflow.co.uk
Warranty Information
Introduction
Installation
Fitting and Removal of Filters
General Operating Instructions
Control Panel/Display Board
CONTENTS
7
8
5
6
2
3
4
1
2
3
4
5
6
7
8
Monitoring
Maintenance
In Case of Difficulty
9 Filter Types
10 Chemical Applications
11 Occupational Exposure Limits 2000
12 Specification
Appendices
1 Notes on COSHH Regulations (UK Only)
Low Airflow Alarm - Testing and Calibration
Airflow Adjustment
Filter Saturation Alarm - Testing and Calibration
Electrical Diagram
UK Approved Service Companies Telephone and Fax Numbers
Protocol for Testing, Monitoring, Replacement and Disposal of Filters
Care and Cleaning
Important, please read this manual before using this equipment.
Thank you for purchasing our Astec Microflow product. We are confident that you will be completely satisfied with its design, function and quality.
Export and Agent Sales Only
Your warranty arrangements will be with your Agent. Please confirm the warranty directly with your Agent.
UK Sales Only
To register the product for our warranty you can either copy this page and fax the details below to us on 01934 410542 or e-mail [email protected]. Full conditions of our warranty and how to arrange a warranty visit are on the next three pages.
The units are calibrated for airflow before leaving the factory however trained personnel can recalibrate them on site.
If you require any help, advice, or have any problems please contact us on our
Technical Help-Line 0800 716107. If a warranty visit is required please complete the
Warranty Visit Request Fax and send to us on 01934 410542.
WARRANTY REGISTRATION FAX
HOSPITAL / COMPANY
CONTACT
TELEPHONE NUMBER
AND EXTENSION
ADDRESS
EQUIPMENT
MODEL
SERIAL NUMBER
LOCATION
PLEASE TICK IF YOU REQUIRE DETAILS OF OUR
SERVICE RECOMMENDATIONS AND PRICES
Warranty and Liability
Astec Microflow produces products that are warranted under normal usage against defects in workmanship and materials for one-year parts and labour costs, from the date of manufacture. The Warranty is stated in the Standard Terms and Conditions of sale.
Export and Agent retailed products are warranted directly by the Agent.
Please confirm your warranty and liability status with the Agent.
In addition, the Warranty is void unless the following conditions are met:
(a) The Warranty Registration Fax has been returned to Astec Microflow UK.
The Warranty Registration Fax is located on the previous page.
(b) The product has been installed and used as stated within the Instruction
Manual.
(c) The warranty does NOT include servicing or maintenance. An approved service company who have attended our training courses for your product must carry out maintenance of product. Failure to maintain or service this product will invalidate the warranty. Maintenance must be carried out in accordance with the Service Manual and include tasks within stated periods.
Failure to use approved service companies or Astec Microflow trained personnel for maintenance also affects the CE Marking status of the product, removing Astec Microflow’s Duty of Care and responsibility under the following European Directives:
(i) Medical Device **
(ii) EMC
(iii) LVD
(d) Supplies used with the product, such as cleaning solutions **, disinfectants **, are not covered by this Warranty.
(e) Consumables such as carbon filters, pre-filters, HEPA filters, light bulbs and tubes, not warranted.
(f) This Warranty is void if faults are caused by accidental damage, mishandling, adjustment by unauthorised personnel or failure to follow the correct maintenance and safety precautions as stated in the Instruction Manual.
(g) The Warranty expressly provided for herein is the sole Warranty provided in connection with the product and no other Warranty, expressed or implied, is provided. Astec Microflow assumes no responsibility for any other claims, consequential (including lost time or profit) or other damage, whether based in contract, tort or otherwise, not specifically stated in this Warranty.
(h) Except in respect of death or personal injury caused by Seller’s negligence, or as expressly provided in these Conditions, Seller shall not be liable to Buyer by reason of any representation (unless fraudulent), or any implied warranty, condition or other term, or any duty at common law, or under the express terms of the Contract for any loss of profit or any indirect, special or consequential loss, damage, costs, expenses or other claims (whether caused by the negligence of Seller, its servants or agents or otherwise) which arise out of or in connection with the supply of the Goods or their use or resale by Buyer, and the entire liability of Seller under or in connection with the Contract shall not exceed the price of the Goods.
Note:
When requesting a Warranty visit, please have the following information available:
(i) Product model number and name.
(ii) Serial number.
(iii) Date of last service, and Service Company.
(iv) Nature of fault and any other comments likely to indicate cause of fault.
(v) A Purchase Order number to cover costs incurred if visit is outside the scope of the Warranty.
Astec Microflow, or our sister company Servicecare, or other nominated personnel will carry out warranty visits.
(j) In the event of any health and safety incidents please advise us in writing at the earliest opportunity.
(k) This warranty and all other contractual issues shall be governed by English law and the parties agree to submit to the nonexclusive jurisdiction of the courts of England.
Items marked ** are relevant to our Safescope models
Please copy before use.
WARRANTY VISIT REQUEST FAX
Customer name
…………………………………………………………………………………………….……
Location of product including address and contact telephone number
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
Product model number and name
………………………………………………………………………………………………….
Serial number (on front or right hand side on unit)
………………………………………………………………………………………………….
Date of last service, and Service Company
………………………………………………………………………………………………….
Nature of fault and any other comments likely to indicate cause of fault
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
…………………………………………………………………………………………………
Purchase Order number to cover costs incurred if visit is outside the scope of the
Warranty ………………………………………………………………………..…………….
Export Only UK Only
Please fax to your product Agent. Please fax to 01934 410542.
1. INTRODUCTION
The Monair fume cupboard has been designed to produce a unit that will provide operator safety with an easy to use cabinet.
The main work area has a face velocity of >0.5 m/sec to provide containment of fumes. The work surface is polypropylene and will contain spillages.
The head unit contains the mains switch, low airflow alarms, carbon filter and optional lights and safety filter.
The five units in the range vary in filter volume and size only.
MONAIR 10
MONAIR15
1 x 10 kg filter
2 x 10 kg filters
MONAIR 20 2 x 10 kg filters
MONAIR 25 2 x 10 kg filters
or 3 x 10 kg filters
MONAIR 30 3 x 10 kg filters
2.
INSTALLATION
WARNING:
INSTALLATION SHOULD ONLY BE CARRIED OUT BY TRAINED AND
APPROVED ENGINEERS OR AGENTS.
ASTEC MICROFLOW OR ITS AGENTS CANNOT ACCEPT RESPONSIBILITY
FOR DAMAGE, LOSS OR INJURY CAUSED BY, OR RESULTING FROM,
INCORRECTLY INSTALLED EQUIPMENT.
The units are supplied with an open base and can be positioned over an existing utility such as a sink, or on a balance bench, or can be used with the removable spillage tray.
The unit is complete with an electrical plug and only requires the fitting of the filters.
The plug is fitted with a 5 amp fuse.
MONAIR 10
ASSEMBLY
Place the fume cupboard base unit on the bench in the required position. Place the head unit on top of the cupboard. Slide the spillage tray into the base of the cupboard section. If a drip cup and tap are supplied, the tray will be fixed in the base of the cupboard and is not removable. Connect the mains lead into the socket on the right hand side of the head unit and plug into a 13 amp supply.
FILTERS
The filters for the unit are packed separately and require fitting into the machine before operation.
Fitting of filters - the filters are located in the head unit.
MAIN FILTER
Fitting
1. Open the front access panel. Carefully lift off and remove the electric earthing connection.
2. Release the clamp mechanism by pulling the handle out and across the unit.
3. Slide the filter into the unit locating the end lugs in the guide rails, and push back until the back stop is reached.
4. Clamp the filter by moving the handle from left to right into the unit.
5. Replace the door, refit earthing strap and lock into place.
PRE-FILTER
The white pre-filter is supplied ready to fit. The supporting frame is located on four retaining clips on the underside of the head unit. The pre filter is laid on the retaining frame and the frame is then pushed up into the retaining clips. Access to insert the retaining frame and pre-filter are from inside the main cupboard unit.
MONAIR 15,20,25
ASSEMBLY
Place the fume cupboard base unit on the bench in the required position. Place the head units on top of the cupboard. Place the top control panel on top of the filter boxes and in front of the fan boxes. Connect the tubing for the low airflow alarm and filter saturation alarm (if fitted) as shown in the diagrams. Slide the spillage tray into the base of the cupboard section. If a drip cup and tap are supplied, the tray will be fixed in the base of the cupboard and is not removable. Connect the mains lead into the socket on the right hand side of the head unit and plug into a 13 amp supply.
Fitting of filters is identical to the Monair 10 unit. Two filters are fitted in the head units one in each clamp arrangement.
Two pre-filters are fitted in the units, one before each main carbon filter.
AIRFLOW
If on checking the airflow, the reading is below the required 0.5 m/s (with the door flaps closed) the fan can be regulated as per the instructions in Appendix 3.
MONAIR 30 (MONAIR 25, 3 FILTER VERSION)
ASSEMBLY
Place the fume cupboard base unit on the bench in the required position. Place the head units on top of the cupboard. Place the top control panel on top of the filter boxes and in front of the fan boxes. Connect the tubing for the low airflow alarm and filter saturation alarm (if fitted) as shown in the diagrams. Slide the spillage tray into the base of the cupboard section. If a drip cup and tap are supplied, the tray will be fixed in the base of the cupboard and is not removable. Connect the mains lead into the socket on the right hand side of the head unit and plug into a 13 amp supply.
Fitting of filters is identical to the Monair 10 unit. Three filters are fitted in the head units one in each clamp arrangement.
Three pre-filters are fitted in the units, one before each main carbon filter.
AIRFLOW
If on checking the airflow, the reading is below the required 0.5 m/s (with the door flaps closed) the fan can be regulated as per the instructions in Appendix 3.
Monair 20 Plastic Pipe Connections Stage 1 - Gastec Connection
LH
RH
One pipe direct to Gastec port red plastic hole
Should not involve plastic T joining pieces but tube increaser is used
Monair 20 Plastic Pipe Connections Stage 2 - Filter Saturation
LH
RH
Should not involve plastic T joining pieces
Filter saturation alarm to one chamber with rubber boot
Monair 20 Plastic Pipe Connections Stage 3 - Airflow Connections
LH
RH
Both chambers linked via plastic T to grey plastic pressure sensor
White or blue plastic T junction
Monair 30 Plastic Pipe Connections Stage 1 - Gastec Connection
LH RH
One pipe direct to Gastec port red plastic hole
Should not involve plastic T joining pieces but tube increaser is used
Monair 30 Plastic Pipe Connections Stage 2 - Filter Saturation
LH
Should not involve plastic T joining pieces
RH
Filter saturation alarm to one chamber with rubber boot
Monair 30 Plastic Pipe Connections Stage 3 - Airflow Connections
LH RH
All three chambers linked via plastic T to grey plastic pressure sensor
White or blue plastic T junction
3.
FITTING AND REMOVAL OF FILTERS
The protocol for testing, monitoring, replacement and disposal of filters is detailed in
Appendix 7.
Hazards associated with the removal and disposal of used filters will depend on the use to which the filtration fume cupboard has been put. If an activated carbon filter is used with hydrocarbon solvents, the filter will retain the solvents without loss, and can be removed in the open laboratory. A pair of plastic gloves and a plastic bag are supplied with each filter. The plastic gloves should be used when removing filters.
The used filters should be placed in the plastic bag and sealed prior to waste disposal, preferably by incineration.
If dangerous materials such as asbestos dust or radioactive chemicals have been contained by the filter, then operator protection which includes the use of respirators and protective clothing may be required. The used filter may require disposal by a specialist company.
As the conditions of use are outside ASTEC's control, it is the responsibility of the user to ensure that any personnel changing filters are advised of any potential hazards in handling the filters, and are provided with any necessary protective equipment or clothing. The safety officer in your organisation can normally advise.
PRE-FILTERS
The pre-filters are located inside the main cupboard area on a supporting frame located on four retaining clips directly under the head and main carbon filter. Open the front cupboard door and pull the frame downwards out of the two front retaining clips. Remove the dirty pre-filter and place in a plastic bag, seal the bag ready for disposal.
The clean pre filter is laid on the retaining frame and the frame is then pushed up into the retaining clips.
Return the door to normal operating position.
MAIN CARBON FILTERS
Changing the filters is identical for all the models. The Monair 10 contains 1 filter, the Monair 15, 20 and 25 have 2 filters and the Monair 25 (alternate version) and
Monair 30 have 3 filters. All filters are the same size and fit all models.
Carefully open the front panel on the head unit and disconnect the earthing strap.
Put on plastic gloves supplied with the new filter.
Release the clamping mechanism by pulling the handle forward and across the front of the unit.
Pull the filter forward by its handle until the clamp mechanism guide rails are fully extended.
Lower the front of the filter with the yellow lugs still locate in the fully extended guide rails allowing the filter to hang as shown.
Slide the plastic bag supplied up over the filter. Remove the filter from runners and seal the bag.
The filter is ready for disposal. The preferable method of disposal is incineration.
Slide the new filter into the unit locating the yellow lugs in the guide rails. Push back until located against the back stop. Move the filter clamping handle back across the unit and into the head. The filter is now evenly clamped in position. Replace the front door panel, ensuring the earthing strap is reconnected.
SAFETY FILTERS (IF FITTED)
To remove the safety filter undo the screws on the retaining frame on the top of the head unit. Put on plastic gloves supplied. Lift the filter out and place in the plastic bag provided. Seal the bag and the filter is ready for disposal. The preferred method of disposal is incineration.
Place the new filter on the top of the exhaust with the foam seal downwards. Screw the retaining frame in place to seal the filter securely.
4.
GENERAL OPERATING INSTRUCTIONS
1. The fume cupboard may only be operated with the correct filter installed for the application. Refer to Sections 9 and 10 of this manual for further information. If in doubt, call ASTEC directly, or contact your local dealer.
2. To start the units, switch the power switch on. The fan will automatically run to give 0.5 m/s.
3. Check the airflow and filter saturation on a regular basis. Section 5 of this manual describes monitoring systems in greater detail and discusses COSHH requirements.
4. The face velocity at the working aperture (and therefore containment of fumes) is at a maximum with the front lower flap in the closed position. The unit should be used in the closed position whenever practical.
5. A Bunsen burner or other heat source should not be placed too close (<150 mm) to the side or back panels. It is not recommended that an open flame or heat source be positioned closer than 300 mm from the bottom of the filter bed.
6. All units are provided with a polypropylene spillage tray that has good chemical resistance.
7. It should be noted that filterblocks do not absorb carbon monoxide or hydrogen. However small quantities (such as used in schools) will not present a hazard because of the large dilution factor from the airflow through the fume cupboard, and retardation of the chemical in the filter matrix.
8. ASTEC fume cupboards are designed to handle fumes and vapours given off during normal laboratory procedures. These will be at the ppm level in the airstream entering the filter block. It is NOT recommended that large quantities of solvents or acids should be boiled off in the fume cupboard.
9. Always keep a spare set of filters available.
10. High concentrations of fumes entering the filterblock may temporarily reduce the filtration efficiency. For this reason any major spillage within the fume cupboard should be cleared up immediately, preferably using spillage absorption granules (available from BDH or other laboratory supply houses), rather than tissue paper which may aggravate the evaporation of toxic fumes from the spillage area.
11. Following a major spillage, the main filters must be changed, as the heat of wetting may reduce filter efficiency. After stabilisation, the old filters can normally be re-used, provided saturation has not been reached.
12. The electrical equipment in the cabinet including light fittings and control equipment are in separate enclosures, on the clean side of the filter. The equipment should not be used in a flammable room atmosphere. Specially modified equipment can be provided for use in solvent rooms. Contact
ASTEC or your local dealer/distributor for further information.
5.
CONTROL PANEL/DISPLAY BOARD
LOW AIRFLOW
MAINS ON/OFF
FILTER
ALARM
(OPTIONAL) HOUR COUNTER ON/OFF
Press the switch down to switch on. A green indicator lamp will indicate that power is connected to the unit. The fan will run at full speed, ready for normal operation of the fume cupboard and the lights, if fitted, will come on. The cupboard can be used immediately.
HOUR COUNTER
This indicates the total number of hours running time of the fan.
LOW AIRFLOW ALARM
The alarm is a red light located at the front of the unit. In normal operation, the red light will not come on.
If the red indicator lamp starts to flashes on intermittently or stays on, the pre-filter is starting to block with dust and an airflow reading should be taken at the aperture to determine if the pre-filter requires changing. We advise changing the pre-filter if an airflow of 0.3 m/sec cannot be maintained.
The correct operation of the alarm may be tested as described in Appendix 2, instructions on replacing filters are given in Section 3 of this manual.
FILTER ALARM/STATUS (IF FITTED)
The green LED is constantly displayed while the filter is actively absorbing the chemical fumes. The light goes out every six seconds as the mechanism samples the air stream. When a chemical breakthrough is detected the red light comes on intermittently and an audible alarm bleeps 4 times every minute showing the filter requires changing. See Appendix 4 for calibration and testing.
ACID DETECTION SYSTEM (IF FITTED)
A simple optional detector system for inorganic acid gases can be fitted to the fume cupboard.
The green LED will indicate the absence of acid gases, and a red LED will alert the user to acid fumes in the return air to the laboratory, an audible alarm will also sound.
TESTING THE ACID DETECTOR ALARM
The alarm can be simply checked by removing the indicator paper from the holder as follows:
1) Switch on the unit, the green LED should be on.
2) Open right-hand cupboard door.
3) Slide the cover off the detector.
4) Remove the indicator paper. The green LED should go out, the red LED will come on and the alarm sound.
Insert the detector paper, green light should come back on.
6.
MONITORING
GENERAL
Under the "Control of Substances Hazardous to Health" (COSHH) regulations, it is mandatory to check safety equipment at "suitable intervals" for correct operation. A suggested maintenance schedule is given in Section 7 of this manual. This section of the manual reviews the manual methods of checking the unit.
The purpose of monitoring is to detect when the pre-filters or main filters cease to operate effectively. If the pre-filters are blocked, the airflow will be reduced at the fume cupboard aperture. If the main filters are saturated, they will cease to remove the fumes effectively.
The Monair units are fitted with a low airflow alarm. Details for testing and calibration are in Appendix 2.
The Monair units do not have any filter saturation detectors. You should carry out a manual check of filter saturation once every three months, as described below.
If an odour is noticed, it is sensible to check the fume cupboard. However, it must be remembered that the sense of smell is very sensitive for some chemicals (e.g. ammonia or hydrogen sulphide) and a slight smell does not mean that the exhaust levels of chemical have approached the maximum acceptable concentration.
MANUAL MONITORING
Manual monitoring should be carried out yearly to check the integral systems on the unit.
AIRFLOW MEASUREMENTS
An anemometer should be used to check the airflow (face velocity) at the working aperture, with the lower panel down. Any suitable anemometer may be used, including hot wire, propeller or vane anemometer. A minimum of six readings should be taken across the working aperture. Note the result in a record book kept for this purpose. This is mandatory under COSHH regulations. We advise changing the pre-filter if the average airflow drops below 0.5 m/s.
The pre-filters will normally need to be changed more frequently than the main filters.
A blocked HEPA filter (where fitted) is also indicated by reduced airflow which is not restored after a pre-filter change.
MANUAL FILTER SATURATION DETECTION
1. Select a suitable test chemical and matching Gastec sampling tube.
Examples include alcohols, toluene, trichloroethylene, or any suitable chemical in routine use in the fume cupboard, provided it is well adsorbed and is not dangerously toxic.
2. Place 6 ml of chemical in a beaker on a hotplate. Adjust the hotplate to boil off the chemical in about two minutes. This gives a concentration of about
100 - 200 ppm to challenge the filter. Use 3 ml of chemical for Monair 10.
3. For testing ACI filters (acid adsorbing), or multi-filters with acid adsorbing layers, use sulphur dioxide gas (SO
2
) at 2 bubbles per second through water
(1 bubble per second for Sensair 10).
4. Using the Gastec tube, sample the outlet airstream from the unit, following the instructions given with the Gastec tube (one pump stroke for trichloroethylene, eight pump strokes for sulphur dioxide for example).
5. The reading should be below the Occupational Exposure Limit (see Section
11). Enter the result in a record book kept for this purpose. This is mandatory under COSHH regulations.
6. If a significant level of chemical is noted in the exhaust air, the main filter should be changed.
7.
MAINTENANCE
IMPORTANT NOTICE
Under the "Control of Substances Hazardous to Health" (COSHH) regulations, effective from 1st October 1989, it is mandatory to maintain written records of checks, tests and repairs carried out on safety equipment, and these records must be kept for 5 years. A summary of COSHH regulations are provided in
Appendix 1.
Regular maintenance will reduce the possibility of hazard to the operator and prolong the life of the fume cupboard.
WARNING! Before attempting any inspection or replacement of electrical components in the head assembly, always isolate the fume cupboard from the
mains electricity supply.
AIRFLOW
The Monair units are fitted with a low airflow alarm. With this model you should test the alarm once a year as described in Appendix 2 of this manual, to ensure it is operating correctly.
FILTER SATURATION
The Monair units do not have any filter saturation detection systems. The filter condition should be monitored regularly, i.e. every 3 months or more frequently if required.
CLEANING AND INSPECTION
You should carry out the following procedures at six monthly intervals:
1. Remove the spillage tray (where fitted) and wash in dilute detergent solution.
2. Wash the interior surface of the cupboard with dilute detergent solution.
3. Inspect the cupboard frame and panels for mechanical damage.
4. Inspect the hinged door for corrosion and correct adjustment of friction hinges
(where fitted).
The following checks should be made every 12 months.
1. Check the condition of services to the cupboard, including water supply, drip cup, waste drain (where fitted), gas supply (where fitted) and electric cable and plug.
2. The electrical earthing and insulation should be inspected by a qualified electrician.
3. Check the fan for correct running, and the tension on the butterfly nut holding the regulating plate in place to ensure it is tight.
8.
IN CASE OF DIFFICULTY
WARNING! BEFORE ATTEMPTING ANY INSPECTION OR REPLACEMENT OF
ELECTRICAL COMPONENTS IN THE HEAD ASSEMBLY, ALWAYS ISOLATE
THE FUME CUPBOARD FROM THE MAINS ELECTRICITY SUPPLY.
Electrical components are mounted behind the control panel. To access the electrical components, remove the two retaining screws which secure the control head to the fan box(es), lift upwards then, if required, disconnect the electrical/tubing connections.
Some possible problems and their causes are shown below:
1. Unit will not operate, no lights or airflow: a) Check that unit is plugged in and switched on. b) Check fuse in mains supply or plug (where fitted). c) Check fuses in electrical input socket.
2. Unit operates, but one or both fluorescent lights do not come on (if fitted).
Open the front door of the head unit to gain access to the lights. Disconnect the electrical connection and remove the two screws. The light box can now slide forward for examination. a) Ensure light tube correctly seated. b) Replace starter lamp. c) Replace fluorescent tube.
3. Fan does not operate, a) Change motor start capacitor on fan. Access to fan is by removing the panel or safety filter (if fitted) on top of the unit. b) Motor failure - contact ASTEC or your local distributor for advice.
4. Fan operates initially, but then cuts out.
Motors are fitted with a thermal cut-out device, which will operate if the motor temperature rise exceeds 95°C. The most likely cause of overheating is a blockage of the airflow, either at the filters or at the exhaust outlet at the top of the unit. Low airflow alarm will operate.
5. Low airflow light - the red indicator lamp comes on at switch on and stays on. a) The pre-filter is blocked with dust. b) If replacement of the pre-filters does not cause the red light to stay off, then the low airflow alarm required recalibration, as described in
Appendix 2.
FOR FURTHER ADVICE PLEASE CONTACT THE ASTEC HELPLINE
ON 0800 716107 OR YOUR LOCAL DISTRIBUTOR.
9.
FILTER TYPES
PRE-FILTERS
Filtrete pre-filter. This is a high performance pre-filter, designed to remove particulates from the airstream. The filter material is based on electrets, which are permanently charged di-electrics. They remove particulates from polluted air by strong electrostatic forces generated by the fibres from which they are made.
The combination of strong electric charge and open structure provides a filter with high efficiency, low airflow resistance and high loading capacity. Measured efficiency figures for particles in the 0.5 - 2.0 micron range is 99%, with loading capacities up to 113 g/m². Filtrete will remove fine particles, aerosols and mists.
MAIN FILTERS
Fourteen different types of filter media are available for the main filter bed. Most of these are impregnated activated carbon, to provide a higher filter capacity for lower molecular weight organic compounds and inorganic gases and vapours. A number of filter efficiency studies have been carried out, and all results using single bed filters show efficiencies very close to 100%.
GP Filter. The most widely used filter in the range, primarily for solvent fume removal. It is manufactured from coconut-shell based activated carbon of 5-10 mesh size and surface area up to 1300 m²/gm. Filtration is achieved by the physical adsorption of molecules in the pores of the activated carbon by Van de Waals forces.
This filter will remove any chemical with a molecular weight greater than 30 and a boiling point greater than 60°C.
ACR Filter. This filter is impregnated with halide salts, and is used for the high efficiency removal of radioactive iodine and methyl iodide. It is frequently used for iodination reactions with low-level radioactive iodine, and efficiencies in excess of
99.99% have been measured.
ACM Filter. This filter is impregnated with iodine compounds for removal of mercury vapour. The filter has been demonstrated to remove mercury from a saturated airstream at ambient temperature to below 5 parts per billion (ppb) in the exhaust airstream.
AMM Filter. This filter is impregnated with copper compounds to efficiently remove vapour from dilute ammonia solutions, and to remove low molecular weight amines.
ACI Filter. This alkali impregnated filter will neutralise volatile inorganic acid vapours such as hydrochloric and hydrofluoric acids, and acid gases such as sulphur and nitrogen dioxides.
SUL Filter. A potassium iodide impregnated filter, designed to remove hydrogen sulphide and low molecular weight mercaptans.
CYN Filter. A multi-impregnated filter to a military specification, for removal of hydrogen cyanide gas. Many cyanide compounds will evolve HCN gas if acidified, so this filter is normally specified if working with any cyanide compound.
FOR Filter. This filter is impregnated with an oxidising agent to oxidise formaldehyde to formate salts. It is widely used in hospital pathology laboratories.
ETH Filter. Diethyl ether is adsorbed on activated carbon, but because of its low boiling point, the local heat of adsorption can reduce the capacity of the filter.
Special impregnation allows a chemical reaction which increases filter capacity.
MIL Filter. As the name implies, this filter is multi-impregnated to handle a wide range of chemicals defined by the Ministry of Defence. It will efficiently remove solvent fumes, acid fumes, hydrogen sulphide, cyanides and arsine at high efficiency
(>99%) but has a somewhat reduced capacity compared to single impregnated filters. It contains noble metals, and is thus rather expensive.
OAL Filter. This odour filter is normally used in air purifiers to deal with odours from excreta, urine and other materials essentially alkaline in nature.
OAC Filter. Air purifier odour filters for human and animal acid type smells caused by bacterial decay, such as cadaverine and putrescine, and other odours which are acidic in nature.
HEPA Filter. The High Efficiency Particulate Air filter is a pleated glass-fibre material sealed with epoxy resin in an aluminium frame. Each filter is tested to ensure a stated particle filtration efficiency (usually 99.997%) according to US
Standard 209B or BS 3928. A HEPA filter is sometimes known as an absolute filter.
EDU Filter. The EDU filter has been specially formulated to adsorb the normal range of chemical fumes generated in Schools during GCSE and 'A' level classes.
MULTIPLE LAYER FILTERS
Fume cupboards may be fitted with filters that have up to four layers of filter material in one filter block. In addition, filters which are 50% HEPA and 50% carbon based may be supplied for particulate and chemical uses.
Advice is available from ASTEC on the choice of filterblock for specific applications.
Contact our Chemist on 0800 716107.
10.
CHEMICAL APPLICATIONS
INTRODUCTION
The table overleaf gives a list of common laboratory chemicals, together with the filter recommended and an estimate of filter capacity.
The recommended filter is the main single-layer filter used for the application. It may be possible to use other filters, such as the MIL filter or a Multi Layer filter. In all cases the filter efficiency for a single layer filter will be close to 100%. Chemicals marked * are poorly absorbed by all filters and should be used in small quantities only. Their exhaust concentration may however be quite low due to dilution with air and retardation in the filter matrix.
The filter capacity is given as the equilibrium saturation capacity, a standard test procedure for activated carbon (ASTM-D3467). The capacity is expressed as the final filter weight at saturation as a percentage of initial filter weight. In many cases these values have been determined by experiment, in other cases estimated values are used based on experience with activated carbon material and a knowledge of the chemical structure. The actual capacity obtained will depend on the conditions of use, but the breakthrough point to reach the OEL of the chemical may be up to 70% of this value for normal input loadings.
Ethylene *
N-heptane
Heptylene
Hexane
Hexylene
Methane *
N-nonane
GP
GP
GP
GP
GP
GP
GP
N-octane
N-octylene
Pentane
Propane *
GP
GP
GP
GP
GP Propylene
Aromatic Hydrocarbons
Benzene GP
Napthalene GP
Ninhydrin GP
Styrene monomer GP
Toluene
Toluidine
Xylene
GP
GP
GP
Esters
Butyl acetate GP
Cellosolve acetate GP
Ethyl acetate GP
Ethyl acrylate
Ethyl formate
GP
GP
Isopropyl acetate GP
Methyl acetate GP
Methyl acrylate GP
Methyl formate GP
Methyl methacrylate GP
Chemical Filter Equilibrium
Type Saturation
Capacity %
Acids
Acetic
Acetic anhydride GP
Acrylic
Butyric
GP
GP
Caprylic
Carbolic
Formic
Lactic
GP
GP
ACI
GP
Osmium tetroxide GP
Palmitic GP
Phenol
Proprionic
GP
GP
Valeric
Alcohols
Ethyl
GP
GP
GP
33
33
40
40
40
40
20
40
40
40
40
40
40
Amyl
Butyl
Cyclohexanol
Isopropyl
GP
GP
GP
GP
Methyl (methanol) GP
Propyl
Aliphatic Hydrocarbons
Acetylene GP
Iso-butane
Butylene
GP
GP
Butadiene
Cyclohexane
N-decane
Ethane *
GP
GP
GP
GP
GP
32
40
40
45
40
32
40
20
10
10
-
35
-
-
-
-
-
35
-
-
-
-
-
26
-
10
40
47
47
47
47
47
40
40
45
40
45
40
45
40
45
40
45
Aldehydes and Ketones
Acetone GP
Acetaldehyde
Acrolein
Benzaldehyde
Butyraldehyde
FOR
GP
GP
GP
32
10
32
40
32
Caproaldehyde GP
Crotonaldehyde GP
Cyclohexanol GP
40
40
40
Diethyl ketone GP
Dipropyl ketone GP
32
40
Formaldehyde FOR 10
Gluteraldehyde FOR 10
Mesityl oxide GP
Methyl butyl ketone GP
Methyl ethyl ketone GP
Methyl isobutylketoneGP
40
40
32
40
Propionaldehyde GP
Valeraldehyde GP
Valeric aldehyde GP
32
40
40
Ethers
Amyl
Butyl
Cellosolve
Dioxan
Diethyl (ethyl)
Ethylene oxide
GP
GP
GP
GP
35
35
40
45
ETH 10
GP 20
Isopropyl GP
Methyl cellosolve GP
Methyl *
Propyl
25
45
ETH 10
GP 30
Halogens
Bromine
Butyl chloride
GP
GP
53
40
Carbon tetrachloride GP
Chlorine
Chlorobenzene GP
Chlorobutadiene GP
Chloroform
Chloro picrin
GP
GP
GP
Chloro nitropropane GP
Dibromoethane GP
Dichlorobenzene GP
Dichlorodifluoro
methane GP
Dichlorodifluoro
ethane GP
Dichloroethylether GP
Dichloromethane GP
Dichloromonofluoro
methane GP
Dichloropropane GP
Dichlorotetrfluoro
ethane
Ethyl bromide
Ethyl chloride
GP
GP
GP
Ethylenechlorohydrin GP
Ethylene dichloride GP
Fluorotrichloro
methane GP
Freon (BP > -20°C) GP
Hydrogen bromide ACI
Hydrogen chloride ACI
Hydrogen iodide ACI
Iodine
Iodoform
Methyl bromide
GP
GP
GP
Methyl chloride GP
Methyl chloroform GP
Methylene chloride GP
Monochlorobenzene GP
Fluorotrichloro
methane GP
65
20
53
40
60
65
60
60
60
20
40
53
53
20
53
45
20
20
20
40
53
55
53
25
20
45
45
45
50
45
5
5
7
Paradichlorobenzene GP
Perchloroethylene GP
Phosgene
Propyl chloride
MIL
GP
Tetrachloroethane GP
Tetrachlorotheylene GP
Vinyl chloride GP
Sulphur Compounds
20
Carbon disulphide GP
Dimethyl sulphate GP
20
50
Ethyl mercaptan SUL 40
Hydrogen sulphide SUL 20
Mercaptans-high MW SUL 40
Sulphur dioxide ACI 10
Sulphur trioxide ACI
Sulphuric acid ACI
20
40
Tetrahydrothiapene GP 40
Nitrogen Compounds
Acetonitrile CYN 20
45
45
20
40
53
53
Ammonia AMM 10
Amines - low MW AMM 10
Amines - high MW AMM 40
Aniline
Diethyl amine
Diethyl aniline
Dimethyl amine
GP
AMM
GP
AMM
40
20
53
20
Ethyl amine AMM 20
Hydrogen cyanide CYN 20
Indole
Nicotine
GP
GP
53
40
Nitric acid fumes ACI
Nitrobenzene GP
Nitroethane GP
Nitrogen dioxide * ACI
10
53
53
-
Nitroglycerine
Nitromethane
Nitropropane
Nitrotoluene
Pyridine
Urea
Uric acid
GP
GP
GP
GP
53
40
40
53
AMM 53
GP
GP
53
53
Miscellaneous
Adhesives
Animal Odours
GP 40
OAL 30
Camphor GP
Carbon monoxide * GP
Carbon dioxide * GP
Citrus fruits GP
40
-
-
40
Cooking odours GP
Deodorisers
Detergents
GP
GP
40
20
40
Hospital odours OAC 30
Human odours OAC 30
Leather
Ozone
GP
GP
30
30
Nicotine
Perfumes
Petrol
Putrescine
Resins
Toilet odours
GP
GP
30
30
GP 40
OAC 30
GP 30
OAL 30
11.
OCCUPATIONAL EXPOSURE LIMITS 2000
Occupational Exposure Limits are listed overleaf for some common laboratory chemicals. A full list may be obtained from:
The Health and Safety Executive
HSE Books
PO Box 1999
Sudbury
Suffolk
CO10 6FS
Tel: 01787 881165
Fax: 01787 313995
The exposure limits are expressed in parts per million by volume (ppm) under the following headings:
(a) LTEL - Long Term Exposure Limit (8 hour time weighted average reference period).
(b) STEL - Short Term Exposure Limit (15 minute reference period).
Gastec sampling tubes are available for * marked items.
Items under review are marked UR.
These Occupational Exposure Limits are supplied for guidance only. Current limits should be confirmed in the EH40 document available from the Health and Safety Executive.
Note:
These limits are correct at the time of publication of EH40/2000. As these limits can change annually, it is best practice to carry out an annual review for the materials in use.
100
3
0.2
2
2
1
2
25
1
10
2
100
25
5
5
5
---
5
0.2
6
4
2
---
35
---
25
6
150
---
---
10
10
LTEL STEL
--- 50
100
50
1000
150
---
---
400
5
200
200
500
---
250
250
0.1
2
0.5
50
0.3
---
1
---
10
2
0.1
---
---
0.3
25 50
1000 1250
200 400
200
100
10
250
300
---
1000 1250
UR UR
1000 1250
0.5
5
1.5
---
1000 1250
---
1
3
5
---
---
0.6
3
0.1
1
5
50
200
15
100
400
1000 1250
50 100
0.02 0.06
50
200
100
7
100
400
150
---
25
400
5
250
10
0.05
0.5
10
0.5
2
2
30
3
200
5000 15000
--- 0.2
100
500
---
310
---
---
2
15
---
5
Alcohols
Butan-1-ol *
Butan-2-ol *
Cyclohexanol
Ethanol *
Isopropanol *
2-Methoxyethanol
Methanol *
Propan-1-ol *
Halogens
Bromine *
Carbon Tetrachloride *
Chlorine *
Chlorobenzene *
Chlorobutadiene
Chloroform *
Chloropicrin *
1,4-Dichlorobenzene
Dichlorodifluoromethane
1,1-Dichloroethane
1,2-Dichloroethylene *
Dichloromethane
Dichlorofluoromethane
Dichloropropene
Dichlorotetrafluoroethane
Ethyl bromide *
Ethyl chloride
Epichlorohydrin
Ethylene dichloride
Fluorotrichloromethane
Hydrogen bromide
Hydrogen chloride *
Hydrogen fluoride *
Iodine *
Iodoform
Methyl bromide *
Methyl chloride
Methyl chloroform *
Trichlorofluoromethane
Perchloroethylene *
Phosgene *
Tetrachloroethane *
Tetrachloroethylene
Trichloroethane
Trichloroethylene *
Vinyl chloride *
Ethers
1,4-Dioxane *
Diethyl ether*
Ethylene oxide *
Isopropyl
Sulphur Compounds
Carbon disulphide *
Dimethyl sulphate
Ethyl mercaptan *
Hydrogen sulphide *
Methyl mercaptan *
Sulphur dioxide *
Miscellaneous
Camphor
Carbon monoxide *
Carbon dioxide *
Ozone *
Petrol *
12.
SPECIFICATION
AIRFLOW
Volume of air treated
Average face velocity (m/s)
10 15 20 25 30
225 450 450 450/675 675
>0.5 >0.5 >0.5 >0.5 >0.5
Lighting (optional)
15 W fluorescent tubes
FILTERS
Main
565 x 350 x 110 mm – 10 kg
Safety
450 x 495 x 40 mm – 4 kg (optional)
SIZE
Width (mm)
Depth (mm)
Overall height (mm)
Internal working height (mm)
ALL UNITS
2
1
1
3
2
2
3
2
2
2/3
2/3
CONSTRUCTION
Head and base unit
Colour
Back Panel
Side Panels
Front doors
Spillage Tray
Fan
Epoxy coated mild steel frame and head
Grey
Clear acrylic
Clear acrylic
Clear acrylic
Polypropylene
Centrifugal type
ELECTRICAL
Voltage Supply
Switches
Fuses
Power rating
230 V, 50 Hz, Single phase
Mains ON/OFF
10 and 20 - 2 x 3 Amp
10 and 20 - Max 690 Watts
OPTIONS
1. Internal lights
2. Trolley or stand
3. Electronic filter saturation alarm
3 4
3
3
750 1000 1250 1500 1750
695 695 695 695 695
1210 1210 1210 1210 1210
800 800 800 800 800
ENVIRONMENTAL CONDITIONS
Indoor use only
Altitude up to 2000 metres
Temperature 10°C to 40°C
Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% at 40°C.
Main supply voltage fluctuations not to exceed ± 10% of nominal.
Pollution degree 1.
APPENDIX 1
NOTES ON COSHH REGULATIONS (UK ONLY)
1. The "Control of Substances Hazardous to Health" (COSHH) regulations, effective from 1st October 1989.
2. The regulations are the UK implementation of an EEC Council
Directive 80/1107/EEC.
3. The regulations require an employer to protect his employees and any other people (whether working for him or not) from hazardous substances.
4. A hazardous substance is defined as:
A. A substance which is on the list of hazardous substances as defined by the Classification, Packaging and Labelling Regulations 1984 (b).
B. A substance for which an Occupational Exposure Limit
(OEL) value exists. This list is similar to US Threshold
Limit Value levels (TLV).
C. A micro-organism which creates a health hazard.
D. Dust at a substantial concentration in air.
E. Any substance which creates a hazard to health, similar to the hazards created by the substances in A - D.
NOTE paragraph 4E is a "catch-all" section.
5. The employer is responsible for assessing the risk to an employee.
6. The employer must prevent or control the exposure of an employee to hazardous substances.
7. The control of exposure "shall be secured by measures other than the provision of personal protective equipment". This means the fumes must be contained, rather than providing protective suits and masks to staff.
8. OEL values must not be exceeded.
9. The employer must ensure that safety equipment is properly used.
10. The employee must use safety equipment provided correctly.
11. The employer must maintain safety equipment in good working order.
In particular,
A. Exhaust ventilation equipment must be examined every
14 months.
B. Other safety equipment must be examined at "suitable intervals".
C. Records of checks, tests and repairs must be kept for
5 years.
12. Monitoring of exposure to hazardous substances must occur "in accordance with a suitable procedure". Records of results must be kept for 5 years for general monitoring and for 30 years when they relate to a specific employee.
13. Regular medical checks are required when working with certain listed substances, or where an identifiable disease is associated with a certain substance.
14. An employer must provide suitable instruction and training to employees regarding risks of substances and precautions to be taken.
15. Certain other regulations take precedence, such as Control of Lead at
Work, Control of Asbestos at Work, radioactive, explosive or flammable regulations, Mines and Quarries Act, and medical treatment regulations.
APPENDIX 2
LOW AIRFLOW ALARM – TESTING AND CALIBRATION
TESTING THE LOW AIRFLOW ALARM
CALIBRATION WARNING!!
CALIBRATION OF THE PRESSURE SWITCH INVOLVES CONTACT
WITH THE MAIN BOARD WHICH HAS ELECTRICAL CONTACTS AT
MAINS VOLTAGE WHICH IS ALSO MAINS OPERATED. CARE
SHOULD THEREFORE BE TAKEN WHEN CARRYING OUT THE
CALIBRATION PROCEDURE.
1. Ensure that new pre-filters are fitted in your filtration fume cupboard.
Switch on the unit, indicator lamp should be out.
2. Switch off unit. Block the pre-filters to give an airflow of <0.3 m/s.
Switch on. The red indicator lamp will start to flash.
3. If a malfunction is indicated by steps 1 or 2, then a pressure switch calibration procedure may be required. Details are provided below.
Remember to remove the plastic bags from the pre-filters before returning the unit to routine use!
CALIBRATION PROCEDURE
The low airflow alarm operates using a differential pressure switch detects a
"high vacuum" situation caused by a blocked pre-filter. The pressure switch is carefully calibrated before leaving our factory, and in the majority of cases calibration will not be required.
The factory calibration is carried out using a standard GP filter at 240, 220 or 110 V 50 Hz mains supply, depending on the market area. A standard
Filtrete pre-filter is used. If the end-user is known (e.g. school) then the calibration will be performed using the correct main filter in place (e.g. EDU filter).
In some cases it may be necessary to recalibrate the pressure switch in the customer's laboratory if the test procedure described above fails. The main reason for a test failure are:
1. A severe knock during transport
2. A change in the mains voltage or frequency
3. Use of a different main filter from that used during factory calibration.
The pressure switch is located behind the control panel.
The pressure switch is located by removing the grommet in the front control panel, and carefully insert a screwdriver to locate the control screw.
The filtration fume cupboard should be fitted with main filter and new Filtrete pre-filter. Switch on, adjustable pressure. Wait 30 seconds to warm up.
1. Unscrew (anticlockwise) switch by 2 turns. The red warning light should now be off.
2. Screw switch clockwise until the red warning light JUST comes on.
3. Unscrew switch by 15 degrees (half hour on the Hour hand of a clock) so the red warning light goes out.
4. Check calibration by switching the unit off, and wrap the pre-filter in a plastic bag and replace in the unit. Switch on. The warning light should come on.
It should be noted that the alarm is NOT an ON/OFF device, but will start to
flicker as the pre-filter progressively blocks with dust.
APPENDIX 3
AIRFLOW ADJUSTMENT
1. The required air flow of 0.5 m/s is set at the factory, however the air flow may need to be adjusted if alternative types of filters are fitted or a safety filter is added.
2. To adjust the airflow, disconnect the unit from the mains power supply and wait for the fan to stop spinning.
3. Remove the front filter door and disconnect the earth wire.
4. Remove the main filter
5. Insert a thin screw driver or similar item through the grill and rotate the butterfly to open or close the fan inlet. Opening the inlet will increase the air flow at the working aperture, closing the inlet will decrease the airflow.
6. Replace the filter, close the door and turn on the unit, check the airflow and repeat the process until the airflow is satisfactory.
7. If the butterfly feels loose remove the fan grill by undoing the four wing nuts securing it position and using a M6 spanner tighten the lock nut until the butterfly is stiff to move.
8. The butterfly nut tension should be checked annually as part of the twelve month service.
APPENDIX 4
FILTER SATURATION ALARM - TESTING AND CALIBRATION
TESTING THE FILTER SATURATION ALARM
1. Switch on the unit, the green indicator lamp comes on.
2. Release the filter clamp.
3. Place a small amount of alcohol (isopropanal) on a tissue and hold above the filter to introduce the fume into the airstream. The red indicator lamp should come on and the audible alarm sound.
4. If a malfunction is indicated by steps 1 or 2, then a calibration procedure may be required. Details are provided below. Remember to remove the paper before returning to routine use!!
CALIBRATION - FILTER SATURATION BOARD
The detector has been set up in the factory before delivery. However the unit may require a slight adjustment in some environments.
1. Remove the Control Head.
2. The Gas Sensor Board is located adjacent to the Control Panel inside the Control Head.
3. Using a multimeter, monitor the voltage at the test point between the gas sensor and the potentiometer and adjust to give 2.3 V.
5 V PSU
APPENDIX 5
ELECTRICAL DIAGRAM
LIGHT PL1 SK1
5 V PSU (WHEN FITTED) PROVIDES
POWER TO OPTIONAL FILTER SATURATION
AND ACID SENSOR ALARMS
LOW
AIRFLOW
ALARM
LED
PRESSURE
SENSOR
BR
CAPACITOR
BR BL BK
BL
FAN
BL
BR
G/Y
G/Y
SK3 PL3 LIGHT
HOUR
COUNTER
ON/OFF
N
L
E
POWER
IN
APPENDIX 6
UK APPROVED SERVICE COMPANIES TELEPHONE AND FAX NUMBERS
We only recommend BIOQUELL Service, as this company has been fully assessed for competency, trained engineers and insurance cover for our complete product range.
COMPANY CONTACT TELEPHONE FAX EMAIL
BIOQUELL Service Service Desk 01264 835 800 01264 835 801 [email protected]
BIOQUELL Service is our sister company which we use for installation, commissioning and warranty work. BIOQUELL
Service is the only approved trained company for Genesis models. There are however other companies who have trained engineers.
APPENDIX 7
PROTOCOL FOR TESTING, MONITORING, REPLACEMENT AND DISPOSAL OF
FILTERS
THIS INSTRUCTION SUPERSEDES ALL MANUAL INSTRUCTIONS REGARDING
FILTER CHANGE PERIODS AND METHODS OF TESTING.
1 BEST PRACTICE
In order to clarify best practice for the use of carbon filters Astec Microflow would like to recommend the following guidance. This guidance is primarily for Carbon filters but High Efficiency Particulate (HEPA) filters are referred to when appropriate.
2
3
FILTRATION
Carbon filter technology has been safely applied for the absorption of harmful vapours and fumes for many years. However, there still exists some doubt with users as to the best practice with regard to filter monitoring and changing. Astec Microflow has detailed best practice in the following product group Tables 1, 2 and 3. Each of these tables express recommended test monitoring and change periods in months, (assuming a 40 hour working week).
TEST PROTOCOL - TERMS
The test period is the maximum period between the filter installation test and the next test for fume bypass of the filter, in some applications where heavier loads of chemicals are applied or the chemicals are dangerous then filter testing should be completed weekly or monthly.
The monitoring period is the maximum period between the filter installation test and the regular weekly/monthly filter monitoring for fume bypass.
The change period is the maximum period between the filter installation and the time to change the filter.
4
All the test, monitor and change periods are expressed in months.
Therefore, in units of higher usage the month recommendation will be superseded by the earlier number of hours.
Note:
The test, monitor or change should be completed at the earliest point whether this be, for example, 6 months or 1000 hours which ever is sooner.
CARBON FILTER LIFE
It is important to note that once the sealed bags containing carbon filters have been opened they will absorb water from the atmosphere. Therefore after 24 months the filters potential life cannot be assured and should be replaced.
5
The filter test protocol should be applied during all test, monitoring, and filter change stages. The protocol is to confirm that the filter is correctly fitted and is absorbing the fumes intended.
It is best practice to regard our guidance time scales as Maximums.
Tables 1, 2 and 3 detail each product and the relevant test, monitor, and change periods for the main and safety filters.
Months to Hours conversion
1 month = 175 hours
3 months = 500 hours
6 months = 1000 hours
9 months = 1500 hours
12 months = 2000 hours
Note:
Use the shortest time scale that applies.
FILTER TEST PROCEDURE
After installation or when test or monitoring is required the following procedure should be completed:
(a) Ensure the unit is switched on and confirm the airflow is correct by measurement or examination of the low airflow alarm.
(b) Introduce the chemical challenge that is normally in use within the unit or a less harmful equivalent for example Iso Propyl Alcohol (IPA).
(c) Test the exhaust or the filter test port to confirm there is no evidence of the chemical.
If the chemical has been detected then the following checks should be completed.
This test can be simply carried out using Gastec or Draeger detector tubes and the appropriate hand held pump. Test pumps and tubes can be supplied by Astec Microflow Ltd.
(d) Note for some chemicals the chemical tube technique is not the best method for measurement. For example, Glutaraldehyde should be tested using OSHA 64 (American standard) method for accuracy.
Notes:
1 Ensure the filter is correctly fitted with the seal seated and compressed.
2 Ensure that there are no gaps in the filter seal.
3 Change the filter and retest.
4 If the chemical by pass continues contact your agent, service provider or Astec Microflow, Tel +44 (0) 1934 410500.
6 FILTER DISPOSAL
The only fully safe method of disposal of carbon and HEPA filters is incineration at the temperatures used for medical waste. However it is possible to dispose in some cases safely to land fill with other laboratory and medical wastes. Astec-Microflow recommends incineration as best practice.
Incineration is available from several companies including P & R Disposal
Services Ltd. and Shanks Ltd.
Table 1 - Airflow Fume Cupboards
MAIN FILTER SAFETY FILTER CARBON ‡
MODEL
A3000
A500
A5000
Astec EVE
Condor
Cytolab **
Downflow
Ducted
EDU 96 †
Exair †
Forensic
Mobile evacuation **
Monair 05
Monair 08
Monair 10
Monair 15
Monair 20
TEST MONITOR CHANGE TEST MONITOR CHANGE
6
6
6
6
6
6
9
9
6
6
9
9
Monitor monthly
8 uses or 24 months
Monitor monthly 12
Not applicable
6
6
6
9
9
9
12
12
12
6
8 uses or 24 months
9 12
9
9
9
9
12
12
12
12
12
12
12
12
12
6
6
6
6
6
6
6
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
12
Not applicable
12
12
12
Not applicable
Not applicable
Not applicable
12
12
12
24
24
24
24
24
24
24
Monair 25
Monair 30
Monair special
Omni Cytotoxic
Omni Lam flow
Omni PCR
Omni TC
Sensair 10
Sensair 15
Sensair 20
Sensair 25
6
6
6
9
9
9
Not applicable
Not applicable
Not applicable
Not applicable
6
6
6
6
9
9
9
9
12
12
12
HEPA
HEPA
HEPA
HEPA
12
12
12
12
6
6
6
6
6
6
6
12
12
12
24
24
24
Not applicable HEPA
Not applicable
Not applicable HEPA
Not applicable HEPA
12
12
12
12
Sensair 30 6 9
‡ Applies only when carbon safety filter is fitted.
12 6 12
† In addition apply CLEAPPS test protocols (UK standard).
** Not used continuously therefore calculated life based on absorption capability.
24
24
24
24
24
Table 2 - Endoscopy Range
Astec Microflow recommend that Endoscopy HEPA exhaust filters are changed every 24 months.
MODEL
MD
MDV
Mini F
Mini MD (f)
MP
Multi F
Safescope Mini PR
Safescope PR
Safescope Mini R
Safescope R
Water system
MAIN FILTER SAFETY FILTER (CARBON) ‡
TEST MONITOR CHANGE TEST MONITOR CHANGE
6
6
1
6
6
1
1
6
6
6
9
9
1
1
9
1
9
9
9
9
Not applicable
‡ Applies only when carbon safety filter is fitted.
12
12
3
3
12
3
12
12
12
12
1
1
Not applicable HEPA
Not applicable HEPA
1
1
Not applicable HEPA
6
6
1 1 6
Not applicable HEPA
Not applicable HEPA
Not applicable HEPA
Not applicable HEPA
Not applicable
Table 3 - Biological and Laminar Cabinets
HEPA exhaust and safety filters to be changed when airflow is reduced or breakthrough is detected.
MODEL
Laminar Flow
M10158
M25674
M25676
M40004
M50480
M50545
M50546
M50548
MAIN FILTER CARBON
TEST MONITOR CHANGE
Not
Applicable
HEPA
OPTION K**
CHANGE
Not
Applicable
HEPA
**
M50549
M50713
M50715
M51423
M51424
M51426
M70085
Microaire
Safety Cabinet
ABS 1000
ABS 1200
ABS 1500
ABS 1800
Special SK
Tissue Culture
6
6
6
6
6
9
9
9
9
9
6
6
6
6 9 12
Not applicable HEPA
6 9 12
9
9
9
12
12
12
12
12
12
12
12
6
6
9
9
12
12
Not applicable HEPA
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
Not Applicable
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
10 uses or 24 months
Not used continuously therefore calculated life based on absorption capability.
APPENDIX 8
CARE AND CLEANING
1
2
STAINLESS STEEL COMPONENTS
Considerable care has been taken in the selection and processing of the stainless steel components used in the construction of this equipment, however even stainless steel can be damaged by chemical attack.
It is therefore important to ensure that any cleaning or disinfecting procedures used will not cause a chemical attack that may damage the surface of the stainless steel. Work surfaces should be kept clean and free of chemical liquids, particularly those containing Chlorine.
Liquids or vapours containing Chlorine are known to cause gradual staining of stainless steel. To avoid this, when chemicals are used containing
Chlorine all surfaces should be thoroughly dried and all traces of the solution removed, the area should then be treated with a neutralising agent. Open vessels left inside the equipment which are likely to cause Chlorine vapours may also result in staining of the metal surfaces.
Should the surface of the stainless steel become stained, it may be cleaned by mechanical polishing and special treatment of the affected area. Our service department can help with this process.
PLASTIC COMPONENTS
Cleaning of the plastic items should only be carried out with mild detergent or chlorine based cleaning solutions. Care must be taken not to apply chlorine solutions to any stainless steel metalwork, as this will potentially cause staining.
3
Do not apply heat, abrasive materials, solvents or solvent wipes to the surface as this can cause irreparable damage.
PAINTED AND OTHER COMPONENTS
Caution:
Do not use water around electrical connections, switches etc.
To clean all painted and other components use a damp cloth and, when required, a liquid detergent.
Note:
Do not use abrasive cleaners.