MCube Europe BioCon-500 T-BS-S04 bladder scanner Troubleshooting Guide
BioCon-500 Cubescan is a device used for assessing bladder volume, it offers several features like: standard and advanced modes, storage options, and printing options.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
MCube Europe version: 01-2006
T-BS-S04
Procedure
Goal
User
Indication
TROUBLESHOOTING
- Evaluate problem and solve it
- Sales Representatives / End users
- After questions / complains
- Send in unit for repair / maintenance Possible actions
Needed items
BioCon-500 Cubescan
Calibration phantom
Protocol
User’s Manual
Service Manual
: BC500-1
: BC5C-0008
: L-BS-S03 Procedure calibration check
: P-BS-P01
: P-BS-S01
Service request form. : L-BS-03 A
__________________________________________________________________
Table of Contents
Page
1. System does not start up
2. System can’t be turned off
2.
2.
3. Initial screen is not displayed when turning on the system 2-3.
4. The green LED is not turned on when the power adapter is connected 3.
5. No Scan head message is displayed during / before scanning 3.
6. No Paper message is displayed during printing / no printed image 3-4.
7. The system turns off while in use
8. No image is displayed after scanning
9. Error in printing (No feeding)
4.
4-5.
5.
10. System starts up with white and black squares 5.
11. Measurements (scanned volumes) are not accurate 5-6.
12. Error Messages 6.
A. How can I change the Time & Date
B. How can I put in the Hospital Name
C. How can I change the print options
7.
8.
10.
D. How can I change thermal paper in the printer
E. How can I change the brightness of the LCD screen
11.
12.
F. How can I review all 12 scanning images 13.
G. How can I only store the last scanned value in stead of highest value 14.
H. How can I display the scan status 17.
1
MCube Europe version: 01-2006
T-BS-S04
1. System does not start up
1
Is POWER key pressed for more than 1 second?
The system operates when pressing the POWER key for more than 1 second.
P-BS-P01 Users manual
P-BS-P02 Users protocol
2
Measure the voltage of battery module.
3 Is battery module connected?
More than +7V
If not connected, connect the battery module.
Medical Technical
Departm.
Renew Battery
Article: BC5C-0005
P-BS-S01
P. 22
Back side main unit
In compartment
1
2
2. System can’t be turned off
The system operates
Is POWER key pressed more than
1 second? when pressing POWER key for more than 1 second.
Is the system not turned off even when pressing the Power key for more than 1 second?
Turn the power off by removing the battery module from the system.
And exchange control boards.
P-BS-P01 Users manual
P-BS-P02 Users protocol
Medical Technical
Departm.
P-MC-C01 Catalogue
Article: BC5J-0001 ~ 0007
1
3. Initial screen is not displayed when turning on the system
The system is to cold.
Is the backlight flickering on the
Wait until the unit has
LCD screen ?
Should be between
(10 º ~ 40 º C) warmed up. (18 º C)
2 Is the screen very dark or bright. ? Adjust the brightness
See point E.
P-BS-P01
P.26
3
System starts up with white and black squares
System is in the
Maintenance Mode
4 Is backlight on in LCD?
5
Is the FPC connector to the LCD module well connected?
Check if the connector for backlight of LCD is connected properly.
Check if the connector is connected properly.
6
7
Check the voltage between the
Control board ground and 12th pin of JP5.
Check the voltage between the
Control board ground and 13th pin of JP5.
Check if -22V (+/-1V).
Check if -21V (+/-1V).
8 Is the system turned off?
Try again after replacing the LCD module.
Change system to patient mode
See procedure L-BS-S03
Medical Technical
Departm.
P-BS-S01
Medical Technical
Departm .
P-BS-S01
Medical Technical
Departm .
P-BS-S01
Medical Technical
Departm
.
P-BS-S01
Medical Technical
Departm
.
P-BS-S01
2
MCube Europe version: 01-2006
T-BS-S04
9 The system is not turned off. Test again after replacing the Control board.
Medical Technical
Departm.
P-BS-S01
1
2
3
4
5
4. The green LED is not turned on when the power adapter is connected
Is green LED not turned on even when the system is OFF?
Yes ->Refer to item 2.
No -> Refer to item 4
Medical Technical
Departm.
P-BS-S01
See procedure
Green LED is not turned on when the system (power adapter) is connected to power
Check if connection of the charger in rolling cart or extension cord in the case is OK
P-BS-S03
Rolling cart assembling
Is the 4 pin connector connecting the Keyboard and Charging board connected?
Check the connection status and if not connected, connect 4 pin connector.
Medical Technical
Departm.
P-BS-S01
Green LED is not turned on even when the system is OFF?
Replace charging board.
Medical Technical
Departm.
P-BS-S01
P-BS-P01
Users manual
Green LED is on when the system is OFF, but it is turned off when the system is ON.
No problem.
Because of charging control in Control board sometime Green LED is off when system is on.
1
2
3
5. No Scan head message is displayed during / before scanning
Is the ultrasonic probe connected well to the main body?
Is open between the the 3rd and
6th pin of the probe connector?
If not, NO SCANHEAD message is displayed when the scan button is pushed.
Test again after replacing with other ultrasonic probe
(Re-)Connect the Scan head / Probe
P-BS-P01 Users manual
Medical Technical
Departm .
P-BS-S01
The message is displayed when the 3rd and 6th pin of Probe connector is short
Request service from the headquarter
Medical Technical
Departm.
P-BS-S01
1
2
3
6. No Paper message is displayed during printing / no printed image
Does printer module have thermal paper?
If not, insert thermal paper.
P-BS-P01 Users manual
Page 27
No image printed on the thermal paper
Paper roll does not fit in the printer
Check if the right side up of the thermal paper is placed in the printer
Check dimensions of the paper roll
Turn the paper roll and try again or adjust Print
Density : Medical
Technical Departm.
P-BS-S01
P-MC-C01 Catalogue
Article:BC5C-0002
3
MCube Europe version: 01-2006
T-BS-S04
4
Does message still appears even when there is paper?
Try again after replacing the control board.
Medical Technical
Departm.
P-BS-S01
1
7. The system turns off while in use
Is auto power-off function active
(on) in the setup menu of the
Maintenance mode?
When auto power-off function is active, system will be turned off if no key input is made during the fixed time.
Medical Technical
Departm.
P-BS-S01
2
Is the system turned off with the message of “BATTERY LOW,
SYSTEM WILL BE TURNED
OFF”?
Turn off the system to protect the battery module if the power is insufficient. Try again after charging.
If the system turns off after a few scans and the battery has been charged: renew the battery
P-MC-C01
Catalogue
Article:BC5C-0005
8. No image is displayed after scanning
Image Description
1
In this case, the mode showing the scan result is set as
CONTOUR.
This can be changed into
B-mode image
See : P-BS-S01
Page 41
2
3
Normal Image
Scanned image from open object(in air)
No bladder volume
(empty bladder 0 ml)
Make sure that the probe
(dome) is faced downwards during scanning in air (as when you scan a patient)
Image ,when the transducer connection is open.
Check the resistance from the ultrasonic probe connector to the 18th and 20th pin
Medical Technical
Department.
P-BS-S01
4
MCube Europe version: 01-2006
T-BS-S04
4
Error in transducer connection.
Retry after replacing the probe
Same result:
Medical Technical
Department .
P-BS-S01
5
9. Error in printing (No feeding)
Check if ‘SCAN
RESULT’ is set as
‘CONTOUR’.
Yes, retry after setting as ‘B-
MODE’.
No, request service from the headquarter.
Changing into B-mode image
See : P-BS-S01
Page 42
Same result:
Medical Technical
Department .
P-BS-S01
1 Thermal Paper
No paper.
Properly inserted.
P-BS-P01
Users manual
Page 27
2 Motor sound?
Yes, go to 1
No, check the printer connector
P-BS-P01
Users manual
Page 27
3 Not solved?
Replace control board or printer
Medical Technical
Department.
P-BS-S01
10. System starts up with white and black squares
1
System starts up with white and black squares
System is in the
Maintenance Mode
Change system to patient mode
See procedure L-BS-S03
1
2
3
11. Measurements (scanned volumes) are not accurate
Re-angle the probe or
One displayed image (bladder) is
See procedures : not seen on the screen move the probe away from the pubic bone
P-BS-P01
Training : T-BS-03
One displayed image (bladder) is not seen on the screen
Scanned volume is different from catheterized volume
Check image with point 8.
Accuracy of the system is
± 20 % (0~699 ml)
Make sure that the images on the screen are correct and that the bladder is shown in the centre of the scanning
Medical Technical
Department.
P-BS-S01
See procedures :
P-BS-P01
Training : T-BS-03
- P-BS-P02
Users protocol
5
MCube Europe version: 01-2006
T-BS-S04
3 Scanned volume is different from catheterized volume area (targeting circle)
The probe should be placed 4 cm above the pubis on the midline
(lower abdomen)
Apply ultrasonic gel on probe cap evenly
(no air trapped in Gel)
Obese patients:
Probe below belly
The scan button on the probe should be on the right side of the patient
The gender should be selected
Perform a Calibration
Check
(between 121 ~ 139 ml) result Negative ►
Send unit to Medical
Technical Department for
Calibration
- P-BS-P03
Protocol for ward
- P-BS-P04
Quick reference
- P-BS-P05 ~ 14
Protocols per ward
- T-BS-07
Learning package
- L-BS-S03
Calibration check
Medical Technical
Department.
P-BS-S01
12. Error Messages
NO DATA AVAILABLE
No measured data or viewing B-mode image
See point 8.
See point 7. BATTERY LOW.
RECHAGE BEFORE NEXT USE
SYSTEM WILL BE TURNED OFF
NO PAPER
NO SCANHEAD
Power shortage
No printer paper
No connecting probe
See point 6 & 9
See point 5.
6
MCube Europe version: 01-2006
T-BS-S04
A. How can I change the Time & Date ( in Standard Mode)
Reference: P-BS-P01 / Users manual / Page 13
To enter the SETUP mode, push UP KEY ( red arrow ) a. Move the cursor at Set Date/Time by up or down key b.
Push the (green arrow) Enter key c. Move the cursor to the position by left or right arrow key, push up or down key for selecting digits. d. Push the Enter key (green Arrow) after changing date / time. e. Move the cursor to Exit by
down arrow key. Push the Enter key (green
Arrow) to exit setup page
7
MCube Europe version: 01-2006
T-BS-S04
B. How can I put in the Hospital Name (in Advanced Mode only)
Reference: P-BS-P01 / Users manual / Page 16
To put in the hospital name the system should be in the advanced mode
To enter the SETUP mode, push UP KEY ( red arrow ) a. Move the cursor to System Mode by up or down key b.
Push the Enter key (inside arrow keys) c. Push the left or right arrow key to select a desired mode
- Standard
- Advanced d. Push the Enter key after changing system mode.
e. Move the cursor to Exit by down arrow key. Push the
Enter key to exit setup page
8
MCube Europe version: 01-2006
T-BS-S04
To enter the SETUP mode / Advanced, push UP KEY ( red arrow ) a. Move the cursor at Hospital Name by up or down key b.
Push the Enter key (inside arrow keys) c. Move the cursor to a desired position and push the
Enter key.
(e.g. M) d. After finish input the name, move the cursor to DONE.
Push the Enter key.
e. Move the cursor to Exit by down arrow key. Push the
Enter key to exit setup page.
9
MCube Europe version: 01-2006
T-BS-S04
C. How can I change the print options (in Standard Mode)
Reference: P-BS-P01 / Users manual / Page 15
To enter the SETUP mode, push UP KEY ( red arrow )
Fig. 3.17
a. Move the cursor at Print Option by up or down key b.
Push the Enter key (inside arrow keys) c. Push the left or right arrow key to select a desired mode.
- Value Only
- Raw Image
- Walls
- All Planes
. d. Push the Enter key to finish print option change.
e. Move the cursor to Exit by down arrow key. Push the Enter key to exit setup page.
10
MCube Europe version: 01-2006
T-BS-S04
D. How can I change thermal paper in the printer
Reference: P-BS-P01 / Users manual / Page 27 a. Open the printer cover as the left image. b. Grab a printer paper in one hand as the left image and slightly pull out the paper to insert in the paper cart by using the other hand.
Make sure that the printable side is facing the pint head
(see left image) c. Close the printer cover after inserting papers in printer cart.
11
MCube Europe version: 01-2006
T-BS-S04
E. How can I change the brightness of the LCD screen
Reference: P-BS-P01 / Users manual / Page 26
Regulating the LCD brightness
Press the enter button in the centre (see blue arrow) and at the same time
the arrow button pointed upwards (=brighter)
or
the arrow button pointed downwards (=darker) ( see red arrows )
Short key functions
Short key Function
Enter + →
Turns over another images of 12 images
Remark
Displays 2 images on LCD
Enter + ←
Enter + ↑
Enter + ↓
Regulates contrast of images on 4 steps
Regulates LCD brightness to brighter
Regulates LCD brightness to darker
Rotates 4 steps
12
MCube Europe version: 01-2006
T-BS-S04
F. How can I review all 12 scanning images
Reference: P-BS-P01 / Users manual / Page 26
After having a scan result on the LCD screen (see below)
SCAN result (B-Mode)
Review other images, displayed per 2 images on LCD Screen
Short key functions
Short key Function
Enter + →
Turns over another images of 12 images
Enter + ←
Regulates contrast of images on 4 steps
Remark
Displays 2 images on LCD
Rotates 4 steps
13
MCube Europe version: 01-2006
T-BS-S04
G. 1 . How can I only store the last scanned value in stead of highest value
(see also G. 2 . below)
(in Advanced Mode only)
ON
1. Stores the maximum SCAN RESULT of one session in Flash memory
2. SCAN RESULT will not be deleted if the power is put off and back on
OFF
1. Flash memory does not store any data
To enter the SETUP mode / Advanced, push UP KEY ( red arrow ) a. Move the cursor at Flash Store by up or down key b.
Push the Enter key (inside arrow keys) c. Move the cursor by left or right key to on or off and push the Enter key. d. Move the cursor to Exit by down arrow key. Push the Enter key to exit setup page.
14
MCube Europe version: 01-2006
T-BS-S04
G. 2 . How can I only store the last scanned value in stead of highest value
(in Maintenance Mode only)
* For this input ask your Medical Technical Department
Maintenance Mode
Maintenance mode can be used to adjust the system variables. Set-up of those variables follows the procedure below.
1
2
3
Check ID OK!!!
Erase Sector OK!!!
Blank Check OK!!!
Program OK!!!
Verify OK!!!
Programming is completed!!!
<<MAINTENANCE MODE>>
Maintenance mode should be accessed by only qualified person.
1. To go to maintenance mode press ENTER key.
2. To return top menu press DOWN key.
To enter the Maintenance mode, push LEFT key with ENTER key being pushed and then push
PRINT key after the system is turned on and the initial screen is displayed. At this time, the information as left is displayed on the screen.
Turn off the system and again turn it on to have the checkers as shown left on LCD.
The initial screen of the system is displayed after approximately 10 seconds of the appearance of checkers. At this time, push RIGHT key with LEFT key being pushed together to display the screen as left.
Push ENTER key to display maintenance setup menu, DOWN key to return to the previous page.
15
MCube Europe version: 01-2006
T-BS-S04
4
5
▶ Calibrat.(M2) Return
Min. Contrast 2
Max. Contrast 8
Store Mode Current
Auto Power 5 minutes
Print Density 8
Scan Status on
Exit
Check ID OK!!!
Erase Sector OK!!!
Blank Check OK!!!
Program OK!!!
Verify OK!!!
Programming is completed!!!
Push ENTER key from the screen 3 above to display maintenance set-up menu as left.
*) Key manipulation
Use up/down key to move menu items from the menu.
▶ appears beside the current menu item. To replace the item value, push ENTER key while the ▶ is still displayed for the corresponding item. To change the item value, scroll until the desired value appears by using left/right value and push ENTER key.
*) Refer to the table below for the details.
When the set-up is done, push LEFT key with
ENTER key being pushed and then push LEFT key to display the initial screen without having the checkers during rebooting. At this time, the information as left is displayed on the screen. And then turn the system off.
Store Mode
Appoints which one to store from the current session. In case of current, the last scanned value is stored. And in case of maximum, the image data with maximum volume is stored.
Inapplicable at present but an upgrade is scheduled in the near future.
16
MCube Europe version: 01-2006
T-BS-S04
H. How can I display the scan status (in Maintenance Mode only)
* For this input ask your Medical Technical Department
Maintenance Mode
Maintenance mode can be used to adjust the system variables. Set-up of those variables follows the procedure below.
1
2
3
Check ID OK!!!
Erase Sector OK!!!
Blank Check OK!!!
Program OK!!!
Verify OK!!!
Programming is completed!!!
<<MAINTENANCE MODE>>
Maintenance mode should be accessed by only qualified person.
1. To go to maintenance mode press ENTER key.
2. To return top menu press DOWN key.
To enter the Maintenance mode, push LEFT key with ENTER key being pushed and then push
PRINT key after the system is turned on and the initial screen is displayed. At this time, the information as left is displayed on the screen.
Turn off the system and again turn it on to have the checkers as shown left on LCD.
The initial screen of the system is displayed after approximately 10 seconds of the appearance of checkers. At this time, push RIGHT key with LEFT key being pushed together to display the screen as left.
Push ENTER key to display maintenance setup menu, DOWN key to return to the previous page.
17
MCube Europe version: 01-2006
T-BS-S04
4
5
▶ Calibrat.(M2) Return
Min. Contrast 2
Max. Contrast 8
Store Mode Current
Auto Power 5 minutes
Print Density 8
Scan Status on
Exit
Check ID OK!!!
Erase Sector OK!!!
Blank Check OK!!!
Program OK!!!
Verify OK!!!
Programming is completed!!!
Push ENTER key from the screen 3 above to display maintenance set-up menu as left.
*) Key manipulation
Use up/down key to move menu items from the menu.
▶ appears beside the current menu item. To replace the item value, push ENTER key while the ▶ is still displayed for the corresponding item. To change the item value, scroll until the desired value appears by using left/right value and push ENTER key.
*) Refer to the table below for the details.
When the set-up is done, push LEFT key with
ENTER key being pushed and then push LEFT key to display the initial screen without having the checkers during rebooting. At this time, the information as left is displayed on the screen. And then turn the system off.
Scan Status
On/off control scan status information which shows how much current scan is centered.
18
advertisement
Key Features
- Standard and Advanced Modes
- Storage of scan data
- Print options
- LCD brightness control
- Calibration Check
- Troubleshooting guide