DeVilbiss 525DS, 525DS-Q, 525KS, 525KS-LT Service Manual
Below you will find brief information for Oxygen Concentrator 525DS, Oxygen Concentrator 525DS-Q, Oxygen Concentrator 525KS. This manual provides instructions for installing, operating, and maintaining the DeVilbiss Oxygen Concentrator. It is designed to provide qualified service technicians and homecare providers with the proper maintenance, service, safety, and repair procedures for the DeVilbiss Oxygen Concentrator. The concentrator draws in room air, separates the nitrogen from the oxygen, and delivers concentrated oxygen to the patient through an oxygen port.
Advertisement
Advertisement
DeVilbiss® 5-Liter Oxygen Concentrator Service Manual DANGER – NO SMOKING MODEL 525DS MODEL 525DS-Q MODEL 525KS MODEL 525KS-LT MODEL 525PS CAUTION-Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. TA B L E O F C O N T E N T S GENERAL INFORMATION Introduction.............................................................................................................................................................................................................................. 3 Important Safeguards.............................................................................................................................................................................................................. 3 Safety Precautions and General Warnings............................................................................................................................................................................. 3 UNPACKING AND SETUP Initial Inspection....................................................................................................................................................................................................................... 5 Patient Setup........................................................................................................................................................................................................................... 5 Operating Instructions............................................................................................................................................................................................................. 5 MAINTENANCE Patient Alert System................................................................................................................................................................................................................ 7 Routine Patient Maintenance.................................................................................................................................................................................................. 7 Periodic Homecare Provider Preventative Maintenance......................................................................................................................................................... 7 Warranty.................................................................................................................................................................................................................................. 8 Provider's Notes...................................................................................................................................................................................................................... 8 Return and Disposal................................................................................................................................................................................................................ 8 Preventative Maintenance Summary...................................................................................................................................................................................... 8 TROUBLESHOOTING System Operation.................................................................................................................................................................................................................... 9 Normal Operating Sequence................................................................................................................................................................................................... 9 Simplified Troubleshooting.................................................................................................................................................................................................... 10 Troubleshooting Chart A........................................................................................................................................................................................................ 11 Troubleshooting Chart B....................................................................................................................................................................................................... 12 Troubleshooting Chart C....................................................................................................................................................................................................... 12 Troubleshooting Chart D....................................................................................................................................................................................................... 12 Troubleshooting Chart E....................................................................................................................................................................................................... 13 Troubleshooting Chart F........................................................................................................................................................................................................ 13 COMPONENT TESTING, REPAIR, AND REPLACEMENT Proper Repair Procedures..................................................................................................................................................................................................... 14 Cabinet Removal................................................................................................................................................................................................................... 14 Accumulator Pressure Test................................................................................................................................................................................................... 15 Auxiliary Oxygen Port............................................................................................................................................................................................................ 16 Capacitor............................................................................................................................................................................................................................... 16 Check Valves/Manifold.......................................................................................................................................................................................................... 16 Compressor........................................................................................................................................................................................................................... 16 Cooling Fan........................................................................................................................................................................................................................... 18 Flow Meter............................................................................................................................................................................................................................. 18 Hour Meter............................................................................................................................................................................................................................ 19 Manifold................................................................................................................................................................................................................................. 19 Molecular Sieve Beds............................................................................................................................................................................................................ 20 Power Cord........................................................................................................................................................................................................................... 20 Power Switch......................................................................................................................................................................................................................... 20 Pressure Regulator............................................................................................................................................................................................................... 21 Printed Circuit Board (PC Board).......................................................................................................................................................................................... 21 Rotary Valve.......................................................................................................................................................................................................................... 21 FIGURES, DIAGRAMS AND PARTS LIST 525DS Unit - Serial #F & 525DS / 525DS-Q Units - Serial #J (115V Units)......................................................................................................................... 23 525DS Unit - Serial #N & 525DS-Q Unit - Serial #R (115V Units with Auxiliary Oxygen Port)............................................................................................ 25 525KS Unit - Serial #F & 525KS / 525KS-LT Units - Serial #J (220/230/240V Units).......................................................................................................... 28 525KS-LT Unit - Serial #N & 525KS Unit - Serial #R (220/230/240V Units with Auxiliary Oxygen Port)............................................................................. 30 525PS Unit - Serial #F & 525PS Unit - Serial #J (220/230/240V Units)............................................................................................................................... 33 Accessories and Tools........................................................................................................................................................................................................... 35 Pneumatic and Wiring Diagrams........................................................................................................................................................................................... 36 ORDERING INFORMATION AND PARTS RETURN Ordering Information............................................................................................................................................................................................................. 38 Parts Return.......................................................................................................................................................................................................................... 38 SPECIFICATIONS................................................................................................................................................................................................................. 39 GUIDANCE AND MANUFACTURER’S DECLARATION..................................................................................................................................................... 40 LT-2023 2 G E N E R A L I N F O R M AT I O N INTRODUCTION • This service manual was designed to provide DeVilbiss Healthcare qualified service technicians and homecare providers with the proper maintenance, service, safety, and repair procedures for the DeVilbiss Oxygen Concentrator. • be used only after one or more settings have been individually determined or prescribed for you at your specific activity levels. be used with the specific combination of parts and accessories that are in line with the specification of the concentrator manufacturer and that were used while your settings were determined. Read and understand all the information contained in this service manual before attempting to operate or perform any maintenance on the concentrator. • For your safety, the oxygen concentrator must be used according to the prescription determined by your physician. An oxygen concentrator is a device that delivers highly concentrated oxygen for therapeutic applications. • Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. • This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device components. • Installation of 515LF-607 low output flow meter package will cause the low flow alarm to not work. • Use only water-based lotions or salves that are oxygen-compatible before and during oxygen therapy. Never use petroleum or oil-based lotions or salves to avoid the risk of fire and burns. • Do not lubricate fittings, connections, tubing, or other accessories of the oxygen concentrator to avoid the risk of fire and burns. • Use only spare parts recommended by the manufacturer to ensure proper function and to avoid the risk of fire and burns. • Use of this device at an altitude above 4000 meters or above a temperature of 95˚F (35˚C) or greater than 95% relative humidity may affect the flow rate and the percentage of oxygen and consequently the quality of the therapy. Refer to specifications for details regarding parameters tested. • Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is turned on but not in use, the oxygen will make the materials flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment. • If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately to avoid harm. • Geriatric, pediatric or any other patient unable to communicate discomfort can require additional monitoring and/or a distributed alarm system to convey the information about the discomfort and/or the medical urgency to the responsible caregiver to avoid harm. • Smoking during oxygen therapy is dangerous and is likely to result in facial burns or death. Do not allow smoking within the same room where the oxygen concentrator or any oxygen carrying accessories are located. If you intend to smoke, you must always turn the oxygen concentrator off, remove the cannula and leave the room where either the cannula or mask or the oxygen concentrator is located. If unable to leave the room, you must wait 10 minutes after you have turned off the oxygen concentrator before smoking. • Open flames during oxygen therapy are dangerous and are likely to result in fire or death. Do not allow open flames within 2 m of the oxygen concentrator or any oxygen carrying accessories.” • The Device is classified as IPX0 which means it does not provide protection from water ingress. Keep the device dry. • Do not service or clean this device while in use with a Patient. • Under certain circumstances, oxygen therapy can be hazardous. Seek medical advice before using an oxygen concentrator. • When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 145˚F (63˚C). Keep body parts a minimum of 32” away from this area. • Improper use of the power cord and plugs can cause a burn, fire, or other electric shock hazards. Do not use the unit if the power cord is damaged. Room air is a mixture of 78% nitrogen, 21% oxygen, 1% argon and other gases. The concentrator draws in room air, separates the nitrogen from the oxygen, and delivers concentrated oxygen to the patient through an oxygen port. For more in-depth classroom type training, contact the Respiratory Technical Service Department at 1-800-338-1988 (814-443-4881). NOTE–DeVilbiss reserves the right to alter or change the design of the DeVilbiss Oxygen Concentrator series. Hence, slight differences in construction or components may exist between the unit in hand and what is described in this manual. IMPORTANT SAFEGUARDS DANGER NO SMOKING Read all instructions before operating the oxygen concentrator. Important information is highlighted by these terms: WARNING: Safety information for hazards that might cause serious injury or death. CAUTION–Information for preventing damage to the product. NOTE–Information to which you should pay special attention. SAFETY PRECAUTIONS AND GENERAL WARNING: A warning indicates the possibility of injury to the user or the operator. • Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 5 feet (1.6m) from hot, sparking objects or naked sources of flame. • Electric shock hazard. Do not remove cabinet. The cabinet should only be removed by a qualified DeVilbiss homecare provider. • Disconnect the power cord from the wall outlet before attempting repairs on the unit. Extra care should be taken if it is necessary to operate the unit with the cabinet removed. • Do not use oils, greases, or any petroleum-based solvents/cleaners on or near the unit. Use only materials that are compatible with oxygen. • Electric Shock Hazard. When replacing the capacitor, do not touch the terminals or allow metal objects to come in contact with the terminals on the capacitor. The capacitor may hold a charge for several days after the unit is turned off. The capacitor is located in the base of the unit next to the cooling fan. • To ensure you receive the therapeutic amount of oxygen delivery according to your medical condition, 525 Oxygen Concentrator must: 3 LT-2023 G E N E R A L I N F O R M AT I O N • The DeVilbiss oxygen concentrators manufactured since July 1, 2012 are equipped with a fire mitigating outlet fitting that prevents propagation of fire into the unit. DeVilbiss oxygen concentrators manufactured prior to July 1, 2012 must be retrofitted by your provider with the fire mitigating outlet fitting. • In order to prevent a fire propagating from the patient through the cannula towards the unit, a means of protection should be located as close to the patient as practicable. Please contact your provider for this means of protection. • Before attempting any cleaning procedures, turn the unit “Off.” • Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid inadvertent damage. • Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying documents. • Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. • The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used. CAUTION: A Caution indicates the possibility of damage to the device. • Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. • It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician. • To prevent product damage, do not attempt to operate the unit without the air filter or while the filter is still damp. • Do not place a humidifier with an oxygen patient unless prescribed by a physician and then only a bubble-type humidifier should be used. • Do not connect the oxygen concentrator to an electrical outlet controlled by a wall switch; the outlet should be independent of other appliances. • Do not use an electrical adapter or extension cord with the oxygen concentrator. • Only operate the oxygen concentrator with all filters in place; do not operate if the air filter is wet. • Use only DeVilbiss concentrator replacement parts and accessories. • Do not use regenerated sieve material. LT-2023 4 U N PA C K I N G A N D S E T U P INITIAL INSPECTION Oxygen Tubing with Humidification Connection It is suggested that an initial inspection be performed upon receiving the oxygen concentrator. If the physician has prescribed an oxygen humidifier as part of the patient’s therapy, follow these steps (If using a prefill, go to Step 3.): 1.Fill the humidifier bottle with distilled, demineralized, or boiled water. Do not overfill. 2.Thread the wing nut located on the top of the humidifier bottle to the oxygen outlet port so that it is suspended. Make sure it is securely tightened. 3.Attach the 5/32" (4 mm) I.D. oxygen tubing, not to exceed 50 feet (15 meters), directly to the humidifier bottle outlet fitting. 1.After removing the DeVilbiss Oxygen Concentrator from the carton, examine it for any external damage. If shipping damage has occurred, contact the DeVilbiss Customer Service Department at 1-800-338-1988 (814-443-4881) for specific instructions. Save the carton for possible later return; note the position of the unit and placement of the packing material. 2.Open the filter door and record the number of hours on the hour meter. Check to make sure the air filter is in place. 3. Check to be sure the intake bacteria filter is in place. NOTE–For optimum performance, the DeVilbiss Oxygen Concentrator has a preset nominal output pressure of 8.5 psi (58.6 kPa). Use only “bubble-type” humidifiers. Do not use “jet-type” humidifiers. 4.Plug the unit into an electrical outlet, turn the unit “On,” and check the audible and visible alerts. 5.Set the flow meter to maximum recommended liter flow and let the unit run for at least 20 minutes. 6. Use an oxygen analyzer to check the concentration. 7. With unit still running, unplug to test the power fail alarm. NOTE–Condensation from the humidifier may occur in longer lengths of tubing or if the tubing is laying on a cold floor. When ready for operation NOTE–If the unit fails to operate properly (oxygen concentration not within specification) or if internal damage is found, contact the DeVilbiss Customer Service Department at 1-800-338-1988 (814-443-4881). 1.Attach the nasal cannula to the oxygen tubing (per the manufacturer’s directions). PATIENT SETUP OPERATING INSTRUCTIONS 1.Position the unit near an electrical outlet in the room where the patient spends most of his or her time. 1.Remove the power cord completely from the strap. Make sure the power switch is in the “Off” position. NOTE–Do not connect to an electrical outlet controlled by a wall switch. 2. 2. 2.Position the unit at least 6 inches (16 cm) from walls, draperies, or any other objects that might prevent the proper flow of air in and out of the oxygen concentrator. 3.Locate the unit a minimum of 5 feet (1.6 meters) from fireplaces, radiators, heaters, and hot-air registers. Follow the Operating Instructions. 1 15 Volt Units– Insert the plug into an electrical outlet. The DeVilbiss Oxygen Concentrator uses a two-prong polarized plug and is doubleinsulated to protect against electric shock. 220/230/240 Volt Units – Ensure cord is connected to the unit before inserting plug into an appropriate electrical outlet. WARNING The plug on the DeVilbiss 525DS concentrators has one blade wider than the other. To reduce the risk of electric shock, this plug is intended to fit in a wall outlet only one way. Do not attempt to defeat this safety feature. WARNING Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 5 feet (1.6m) from hot, sparking objects or naked sources of flame. Improper use of the power cord and plugs can cause a burn, fire, or other electric shock hazards. Do not use the unit if the power cord is damaged. Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 5 feet (1.6m) from hot, sparking objects or naked sources of flame. Electric Shock Hazard. Only qualified DeVilbiss Healthcare homecare providers may remove the cabinet. 4.Attach the appropriate oxygen accessories (oxygen tubing or humidifier) to the oxygen outlet port. WARNING NOTE–A maximum of 50 feet (15 meters) of tubing plus 7 feet (2.1 meters) of cannula plus a bubble humidifier is allowed between the concentrator and the patient. The DeVilbiss oxygen concentrators manufactured since July 1, 2012 are equipped with a fire mitigating outlet fitting that prevents propagation of fire into the unit. DeVilbiss oxygen concentrators manufactured prior to July 1, 2012 must be retrofitted by your dealer with the fire mitigating outlet fitting. NOTE–The oxygen supply accessory (patient tubing) shall be equipped with a means that in case of fire stops the delivery of oxygen to the patient. This means of protection should be located as close to the patient as practicable. In order to prevent a fire propagating from the patient through the cannula towards the unit, a means of protection should be located as close to the patient as practicable. Please contact your dealer for this means of protection. Oxygen Tubing Only Connection 1.Thread the cannula fitting onto the oxygen outlet port. 3.Press the power switch to the “On” position. When the unit is turned “On,” all four indicator lights (Power, Service Required, Low Oxygen, and Normal Oxygen) on the front panel will briefly illuminate and an audible signal will sound (the patient alert system) momentarily. After a few seconds, only the 2.Attach the 5/32” (4 mm) I.D. oxygen tubing. 5 LT-2023 U N PA C K I N G A N D S E T U P “Power” and “Normal Oxygen” lights will remain on. NOTE– DeVilbiss recommends for optimal service life that the DeVilbiss Oxygen Concentrator to be operated for at least 30 minutes after it is powered on. Shorter periods of operation may affect the long term reliable operation of the product. 4.Slowly turn the flow meter knob until the flow meter ball is centered on the line next to the appropriate flow rate. NOTE–When the flow meter knob is turned clockwise, the flow decreases (and eventually will shut off the oxygen flow). When the knob is turned counter-clockwise, the flow increases. NOTE–Use low output flow meter (part #515LF-607) for flow rates under 1 lpm. NOTE–The unit may require up to 20 minutes for the oxygen concentration and flow rate to stabilize. The flow rate should be monitored and readjusted if necessary. 5.The flow meter has a locking device. If it is necessary to preset and lock in the prescribed flow rate, tighten the set screw located on the hex nut just below the control knob using a 1/16" Allen bit. No adjustment can be made without loosening the set screw. 6. The DeVilbiss oxygen concentrator is now ready for use. LT-2023 6 MAINTENANCE PATIENT ALERT SYSTEM ROUTINE PATIENT MAINTENANCE The DeVilbiss Oxygen Concentrator patient alert system will detect unit component failure. This system is comprised of both visible and audible alerts which signal the patient if a malfunction should occur. DeVilbiss recommends using only original DeVilbiss parts and filters in order to guarantee a reliable operation of the product. DeVilbiss OSD® Operation Cannula, Tubing, and Humidifier Bottle The OSD is a device within DeVilbiss concentrators that monitors the oxygen produced by the unit. The OSD operates as follows: • Normal Oxygen (green light) - oxygen purity normal • Low Oxygen (yellow light) - oxygen purity low–requires servicing The patient should clean and replace the cannula, tubing, and humidifier bottle according to the manufacturer’s instructions. The oxygen patient should perform the following maintenance: Air Filter and Oxygen Outlet Connector The air filter and oxygen outlet connector should be cleaned at least once a week by the patient. To clean, these steps should be followed: NOTE–If the oxygen purity continues to fall, an audible signal will sound intermittently. If the oxygen purity continues to fall to a low enough level, the yellow “Low Oxygen” light will turn off and the red “Service” light will turn on. 1.Remove the air filter located in the door on the back of the unit. Remove the oxygen outlet connector (if used) from oxygen outlet port. NOTE–Refer to the Alerts section below for specific alert settings. 2. NOTE–After Power On, the OSD conducts a continuous diagnostic evaluation to check for a fault in the piezo electronics. If this condition is detected by the OSD electronics at any time during concentrator operation, the green “Normal Oxygen” OSD light will turn off and the beeping audible alert and blinking red “Service” light will activate. Otherwise for the first fifteen minutes of operation, the green “Normal Oxygen” light will remain illuminated during the oxygen stabilization process. After that time, the OSD will begin monitoring the oxygen purity every second. 3.Rinse thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling. WARNING Do not attempt to operate the unit without the air filter or while the filter is still damp. NOTE–The air filter should be monitored more closely in environments with abnormal amounts of dust and lint. Alerts: CAUTION–Operation of the DeVilbiss Oxygen Concentrator in extreme environments or without the air filter will prematurely occlude the intake bacteria filter and cause a decrease in the unit performance. There are two visible service alerts located on the front panel. For serial numbers starting with J, N or R Wash in a solution of warm water and dishwashing detergent. For serial numbers starting with F Exterior Cabinet The patient should clean the concentrator exterior cabinet by using a damp cloth or sponge with a mild household cleaner and wiping it dry. Low O2% Service Low O2% WARNING Service Do not apply liquids directly to the cabinet or utilize any petroleum-based solvents or cleaning agents. The audible alert system is internally powered; no batteries are required. If the indicator lights illuminate or the audible alert sounds other than during start-up, a problem has occurred • Power Failure (Blinking red “Service” light and pulsing audible alert) • Low Flow (Below 0.5 lpm) (Continuous red “Service” light and audible alert) • Below Normal Oxygen (84% to 75%, yellow “Low Oxygen” light. 75% to 60%, yellow “Low Oxygen” light and beeping audible alert. Less than 60%, red “Service” light and beeping audible alert.) PERIODIC HOMECARE PROVIDER PREVENTATIVE MAINTENANCE Use only DeVilbiss concentrator replacement parts and accessories. Every DeVilbiss Oxygen Concentrator is tested at the factory. To assure continued trouble-free performance, the following preventative maintenance should be performed by the homecare provider during periodic oxygen patient visits. Failure to properly maintain the unit will void the warranty. The visible and audible alerts will activate for approximately 15 minutes in a no power situation. If the unit is turned “On” without power or power is removed later, no alert will sound for the first 10 seconds. After that time, the alert will produce an audible pulse every few seconds while the visible alert blinks. Power for this alert is provided by a capacitor on the PC board. 1.Check the oxygen concentration with an oxygen analyzer (part #R217P62 or R218P12)—every 3 years. a.Calibrate the oxygen analyzer prior to checking the oxygen concentration. The analyzer should be properly calibrated using the manufacturer’s recommended procedure. NOTE–If the concentrator has been unused for an extended period, the unit must run several minutes before the power fail alert will activate. NOTE–Changes in temperature, altitude, or humidity may affect the analyzer’s oxygen concentration reading. The analyzer should be calibrated in similar conditions to the location of the concentrator. The PC (printed circuit) board is responsible for controlling the system and alerts. NOTE–A high pressure condition is indicated by the audible (a “popping” sound) release of pressure from a pressure relief valve located on the compressor head. b.Power the unit, set the flow meter to 5 LPM and connect the analyzer to the unit’s oxygen outlet port. c.Wait 20 minutes for the display to stabilize. The concentrator must operate for a minimum of 20 minutes before checking the oxygen concentration. NOTE–Installation of 515LF-607 low output flow meter package will cause the low flow alarm not to work. d. 7 Record the reading. LT-2023 MAINTENANCE manufacturer or by a qualified DeVilbiss provider/service technician. 2.Check the audible alert and indicator lights during every service. When the power switch is turned “On,” listen for the audible alert and check to see if the front panel indicator lights are operating. NOTE–If the following described complete processing of the concentrator by a qualified DeVilbiss provider/technician is not possible, the device must not be used by another patient! 3.Inspect intake filter (part #MC44D-605) during every PM Check. Replace as necessary. DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients. a.Open the filter door and replace filter as required. 4.Inspect the final bacteria filter (part #PV5LD-651) during every compressor service. Replace as necessary. NOTE–If preventative maintenance is due at this time, these procedures should be carried out in addition to the servicing procedures. a.Use the Cabinet Removal instructions found under CABINET REMOVAL in this manual to remove and attach the cabinets. 1.Dispose of all accessory components that are not suitable for reuse, i.e., particularly the oxygen tubing, the nasal cannula/mask, oxygen outlet connector and humidifier bottle. b.Remove the hose from each end of the filter and discard the filter. c.Install the new final bacteria filter with the “IN” fitting toward the flow meter. 2. 5.Inspect the compressor filter during every compressor service. Replace as necessary. AUTION–the concentrator must be disconnected from the power C supply for this step: Open the concentrator and remove all dust deposits inside the cabinet with an appropriate vacuum cleaner. 3.Clean and disinfect all parts of the cabinet inside and outside and the power cord with a suitable disinfecting agent, e.g., Microbac Forte, Terralin®, Aldahol, Cidex OPA or Peract. NOTE–A change was made to the type of compressor filter used in the 525 series concentrator. The unit may have one of two different types of compressor filters depending on when the concentrator was manufactured. The original one is a white plastic filter (part #525D-622) that is located near the compressor. The newer style is a sintered bronze filter (525DD-626) that is installed in one end of the compressor exhaust hose near the rotary valve. All models of the 525 series are now being manufactured with the sintered bronze compressor filter. See figure below. 4.Check the cord, the plug on the back of the device, the power switch, the fuse holder and the indicator light for possible damage. 5. Replace all damaged or worn components. 6. Replace the cabinet air filter on the back of the device. 7.Check the oxygen concentration. If the device is within specifications, the intake bacteria filter does not need to be replaced between patients. If the concentration is not within specifications, the provider should refer to the service manual section on Troubleshooting. Sintered Bronze Compressor Filter (525DD-626) NOTE–This PM Schedule reflects: • 5000 hour usage equal to one year • a normal, clean operating environment. RETURN AND DISPOSAL This device may not be disposed of with household waste. After use of the device, please return the device to the dealer for disposal. This device contains electrical and/or electronic components that must be recycled per EU Directive 2012/19/EU-Waste Electrical and Electronic Equipment (WEEE). Non-infectious used accessories (e.g. nasal cannula) can be disposed of as residential waste. The disposal of infectious accessories (e.g. nasal cannula from an infected user) must be made via an approved waste disposal company. Names and addresses can be obtained from the local municipality. The homecare provider is responsible for: • determining the condition of the concentrator operating environment. • determining a preventative maintenance interval frequency* which takes into consideration the specific operating environment. * S tandard intervals are noted below. Service interval may be more or less frequent than stated below provided that the Home Care Provider establishes and documents appropriate protocols. PREVENTATIVE MAINTENANCE SUMMARY WARRANTY Patient / Caregiver DeVilbiss provides warranty for this device. Use only DeVilbiss approved original spare parts for maintenance and repair to retain warranty. Using unapproved or non-original spare parts would void warranty. Clean and replace oxygen tubing, cannula / mask, and humidifier bottle (if used) according to manufacturer’s instructions. PROVIDER’S NOTES - Cleaning and Disinfection When There is a Patient Change During each inspection Homecare Provider Wash/Replace cabinet filter. Check audible alert and indicator lights. During each PM check – every 3 years for the 525 series When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary), and the next patient, user or third party should be protected by appropriate handling and preparation. Inspect/Replace intake bacteria filter as necessary. Check oxygen purity. During compressor service Inspect/ Replace final bacteria filter. Inspect/ Replace the compressor filter Therefore, when there is a patient change, people must be protected during the transport and handling of the device, and the device must be fully processed, i.e., cleaned and disinfected, by suitably trained personnel before reuse to protect the next patient. The complete processing may only be done by the LT-2023 8 TROUBLESHOOTING SYSTEM OPERATION NORMAL OPERATING SEQUENCE The DeVilbiss Oxygen Concentrator uses a pressure swing adsorption system. The air is drawn into the unit through air filters and into a double-head compressor. When the concentrator is turned “On,” the following cycling sequence can be observed by attaching a pressure gauge to the manifold or accumulator tank test point. 1.The rotary valve is quickly cycled several times to relieve residual bed pressure preventing a static condition in the compressor. This rapid cycling only happens on start-up and is clearly heard as pressure is being quickly exhausted several times. The pressure exhausts through an exhaust muffler that is connected to the valve. 2.The PC board applies a short DC voltage signal to the valve. The valve will stop for several seconds causing the right bed to pressurize first while the left bed depressurizes. 3.Voltage is again applied to the valve for a short time. The valve will stop for approximately a second. During this time the sieve bed pressures are equalized. 4.A short DC voltage signal is again applied to the valve. The valve will stop for several seconds causing the left bed to pressurize while the right bed depressurizes. 5.A short DC voltage signal is again applied to the valve. The valve will stop for approximately a second. During this time, the sieve bed pressures are equalized. 6. The cycle then repeats with step 2. A pneumatic diagram of the system is shown on Page 36. The compressed air passes through a rotary valve, which is cycled at a predetermined rate, and is directed into one of two sieve beds. The sieve beds contain molecular sieve material which is a synthetically-produced inorganic silicate. It is very porous and has the unique ability to selectively adsorb nitrogen from the air as it passes through the sieve bed. As one bed is being pressurized, the other bed is quickly depressurized. This allows the nitrogen that was adsorbed during its pressurization cycle to be exhausted from the sieve material. The nitrogen is released through exhaust ports located on the rotary valve assembly. The ports are connected to a single piece of hose running from the valve to the exhaust muffler. Also during each bed pressurization, a small amount of oxygen flows through an orifice from the pressurized bed into the depressurizing bed. This helps purge the nitrogen from the depressurizing bed. The beds will continue to be alternately pressurized and depressurized as the unit operates. NOTE–In the “turn-down” mode, the fixed cycle time is decreased to less than 2 seconds. Oxygen leaving the sieve beds is directed through a check valve to the accumulator tank. A pressure regulator on the tank controls the oxygen pressure as it leaves the accumulator and enters the flow meter. The flow meter allows the oxygen flow to be controlled and adjusted to the level prescribed by the patient’s physician. From the flow meter the oxygen passes through the final bacteria filter and finally the oxygen outlet port to the patient. The DeVilbiss Oxygen Concentrator operates on a timed cycle (3.6 sec. @ 5LPM) that is controlled by the PC board. The PC board will send voltage to the valve causing it to shift and alternately pressurize the sieve beds. The PC board also activates the electronic alert system. Low flow, system abnormality, and power failure are indicated by audible and visible alerts. A high pressure condition will be indicated with a “popping” type sound produced by release of pressure from a pressure relief valve on the compressor head. The 525 operating system incorporates “turn-down” technology. The PC board constantly monitors the flow rate and will decrease the cycle time whenever the flow rate is less than 1.2 LPM. Therefore it “turns-down” the cycle based on lower oxygen demand. As a result, the unit runs cooler and less power is consumed. 9 LT-2023 TROUBLESHOOTING Type II—Operation Issues (Service Indicator SIMPLIFIED TROUBLESHOOTING Light and/or an audible alert is activated or the Pressure Relief Valve releases pressure with a ‘popping’ sound) The key to simple troubleshooting is to recognize which type of problem exists and select the most effective approach to solving the problem. The different types of problems and the approaches for solutions are as follows: WARNING Type I— Purity Issues (Low Purity Indicator Light and/or an audible alert is activated or the Pressure Relief Valve releases pressure with a ‘popping’ sound). Electric Shock Hazard. Extra care should be taken if it is necessary to operate the unit with the cabinet removed. WARNING WARNING Mechanical Hazard. Keep fingers, loose clothing, etc. away when working on compressor. NOTE–Check for leaks using a certified leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). Apply leak test solution to all fittings and hose connections with unit running. If an air leak is present, the solution will bubble. All leaks should be repaired before putting the unit back in service. CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. Electric Shock Hazard. Extra care should be taken if it is necessary to operate the unit with the cabinet removed. WARNING Mechanical Hazard. Keep fingers, loose clothing, etc. away when working on compressor. Observe the pressure cycle at the oxygen tank. 1.Connect a calibrated pressure gauge to the unit’s oxygen tank test point or manifold. 2. Set the unit’s flow meter to 5 LPM. 3.Power the unit and allow it to operate for a minimum of 5 minutes before observing the pressure cycle. 4.Compare the high pressures and low pressures to those expected for the current elevation and use the following chart to find the appropriate action. All consecutive high pressures should be within 1 psi of each other and all consecutive low pressures should be within 1 psi of each other. Operation Alarm Chart If this is happening It’s because of this Pressure relief valve has a ‘popping’ noise with Rotary valve not cycling possible uneven tank properly. pressure. Continuous red light with a Low flow alert – flow setting pulsing audible alert while below 0.5 lpm. compressor is operating. NOTE–For normal system pressures refer to Specifications. NOTE–Check for leaks using a certified leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). Apply leak test solution to all fittings and hose connections with unit running. If an air leak is present, the solution will bubble. All leaks should be repaired before putting the unit back in service. Continuous red light with a Internal temperature is too continuous audible alert high. and unit is warm to touch. CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. Pressure Diagnostic Chart If Pressure Reading Is Do This To This Higher than expected and/ or pressure relief valve Replace Sieve beds (2) has a ‘popping’ noise Inspect and replace, as Filters needed Inspect and correct or replace, Tubing connections Lower than expected as needed Compressor Troubleshoot See Component Testing, Repair and Replacement Inspect and correct, as Leaks from sieve beds to As expected needed oxygen outlet port Dropping too low during Replace Check valves or manifold cycle change Inspect and replace, as Rotary valve wire harness Uneven - high pressures needed & low pressures are not Rotary valve or PC board. consistent during cycle Troubleshoot or replace See Component Testing, Repair and Replacement Continuous red light with a pulsing audible alert and Internal power failure to fan is operating but compressor. compressor is not. Do this to resolve Test rotary valve. Refer to page 21. Correct setting and educate user and / or install low flow meter, if prescription is below 1 lpm. See section on the flow meter. Move unit to cooler location. Ensure unit vents are clear. Ensure unit filters are clean. Ensure cooling fan is operating, replace as needed. Ensure there is proper voltage to the unit. If voltage is correct, replace capacitor or compressor. Test voltage at compressor connector. If no voltage, replace main wire harness. If voltage present, replace capacitor or compressor. Ensure voltage is correct and / or the part is functional for the following: AC outlet, power cord, IEC connector or cord connection, main wire harness, circuit breaker; correct as needed. If issue persists, replace power switch. Blinking red light with a pulsing audible alert and compressor and fan not operating. External OR Internal power failure. Blinking red light with a pulsing audible alert and compressor and fan are operating. Pressure relief valve has a ‘popping’ noise. Inspect wire harness and Internal power failure at valve replace as needed. wire harness or PC board. Inspect fuse, if blown, replace fuse or PC board. Type III—The concentrator runs and continues to cycle but has low oxygen concentrations and no alarms are activated. This problem is similar to Type I in the way it is diagnosed. Observe pressure cycle at oxygen tank as recommended in Type I. LT-2023 10 TROUBLESHOOTING TROUBLESHOOTING CHART A Visible Alarm Audible Alarm Compressor Power Light OFF OFF ON ON OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Pulsating air noise Intake filter not in place or defective Check filter and replace if necessary Excessive noise Fluctuating oxygen flow Little or no oxygen flow Low oxygen concentration Audible alarm does not sound during power failure Compressor intake hose disconnected Reconnect hose Loose or defective motor mounts Replace motor mounts Defective compressor Replace compressor Defective cooling fan Replace cooling fan Occluded humidifier Clean or replace humidifier Use of improper humidifier Use only a bubble-type humidifier Occluded filters Clean or replace filters Occluded or defective cannula and tubing Detach cannula from oxygen delivery tubing. If proper flow is not attained, check tubing for kinks or other obstructions. Clean or straighten as required or replace tubing if necessary Use of excess oxygen tubing The unit is designed to deliver 5 lpm with a cannula on 50 feet (15 meters) of approximately 5/32” (4 mm) inside diameter tubing. Smaller diameter tubing or the addition of any other flow restriction may prevent obtaining the desired flow rate. Defective flow meter Replace flow meter Leak in system Check for leaks in all hoses and fittings Defective compressor Replace compressor Defective compressor reed valve Replace compressor reed valve Defective check valve Replace check valve or manifold Pressure regulator not adjusted properly or defective Adjust or replace pressure regulator Flow meter not adjusted properly Adjust flow meter Hose disconnected to flow meter Reconnect hose Oxygen delivery tubing is kinked or blocked Straighten tubing or remove obstruction Occluded humidifier Clean or replace humidifier Leak in system Check for leaks in all hoses and fittings Defective sieve bed check valve Replace check valve or manifold Defective compressor reed valve Replace compressor reed valve Defective compressor Replace compressor Rotary valve not operating correctly Replace valve Occluded filters Clean or replace filters Contaminated sieve beds Replace sieve beds Unit has not been used for an extended period of time. NOTE–If the concentrator has been unused for an extended period, the unit must run several minutes before the power fail alarm will activate. Allow unit to run for 20 minutes and retry Defective PC board Replace PC board Defective power switch Replace power switch Defective wire harness Replace wire harness Audible alarm does not sound when unit is turned “On” Defective PC board Replace PC board Pressure relief valve activated “popping” sound PC board connectors not properly latched Be sure tabs are pushed completely into place Defective PC board Replace PC board Defective rotary valve Replace valve PC board connectors not properly latched Be sure tabs are pushed completely into place Defective PC board Replace PC board Defective light panel Replace light panel Service Required light does not illuminate when unit is turned “On” 11 LT-2023 TROUBLESHOOTING TROUBLESHOOTING CHART B Visible Alarm Audible Alarm Blinking Pulsing Compressor Power Light OFF OFF OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Fan off Line cord not properly installed or defective Insert plug in receptacle or replace line cord. On 220/230/240 Volt units, check that the IEC connector on the back of the unit is attached. No power at receptacle Check building circuit breaker or fuse, or have house wiring checked by qualified electrician Circuit may be fully loaded with other appliances and another receptacle may be required. Wall outlet should be independent of other appliances Oxygen concentrator circuit breaker activated Press the circuit breaker reset button. If unit circuit breaker opens again, check internal wiring. Line cord quick-connect terminal inside unit is disconnected Reconnect quick-connect terminal Defective power switch Replace power switch Defective circuit breaker Replace circuit breaker TROUBLESHOOTING CHART C Visible Alarm Audible Alarm Compressor Power Light Blinking Pulsing ON ON/OFF OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Fan and compressor operating Pressure relief valve activated – “popping” sound Connector on PC board not connected Connect connector Blown fuse on PC board Replace fuse Defective PC board Replace PC board TROUBLESHOOTING CHART D Visible Alarm Audible Alarm Compressor Power Light OFF ON ON ON OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Fan operating Main wiring harness disconnected/defective Reconnect/replace wiring harness Loose compressor wire Tighten or attach wire Defective capacitor Replace capacitor Defective compressor Compressor overheated due to: Replace compressor 1. Occluded filters 1. Clean or replace filters Defective cooling fan Replace cooling fan Defective compressor Replace compressor Unit warm to the touch and cannot be restarted for several minutes 2. Restricted input or output air passage 3. Low or high line voltage LT-2023 2. Remove obstruction 3. Check line voltage; use alternate circuit independent of other appliances 12 TROUBLESHOOTING TROUBLESHOOTING CHART E Visible Alarm Audible Alarm ON ON OTHER SYMPTOMS POSSIBLE CAUSE Fluctuating or no flow System pressure below 9 psi (62.1 kPa) due to: Compressor Power Light ON ON POSSIBLE REMEDY 1. Leak in system 1. Check for leaks in all hoses and fittings 2. Defective compressor 2. Replace compressor TROUBLESHOOTING CHART F Visible Alarm Audible Alarm Compressor Power Light See Below See Below ON ON OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY No OSD lights are illuminated. Defective OSD. Check concentration with an oxygen analyzer. If the concentration is within specification, replace the PC board. No OSD lights are illuminated, but red “Service Required” light is illuminated accompanied by a beeping audible alarm Oxygen level is low* Check concentration with an oxygen analyzer. If the concentration is within specification, replace the PC board. If the concentration is low, refer to low oxygen concentration symptom in Troubleshooting Chart A. Both OSD lights are illuminated. Defective OSD Replace PC board. Yellow Low Oxygen light is illuminated. Oxygen level is low* Check concentration with an oxygen analyzer. If the concentration is within specification, replace the PC board. If the concentration is low, refer to low oxygen concentration symptom in Troubleshooting Chart A. Yellow Low Oxygen light is illuminated and an intermittent audible alarm sounds every few seconds. Oxygen level is low* Check concentration with an oxygen analyzer. If the concentration is within specification, replace the PC board. If the concentration is low, refer to low oxygen concentration symptom in Troubleshooting Chart A. *Refer to Alerts page for oxygen purity levels. NOTE-If unit does not run or perform to specification, verify the proper AC input voltage and frequency are being used. 13 LT-2023 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T PROPER REPAIR PROCEDURES CABINET REMOVAL WARNING For units with Serial Numbers starting with J, N or R When servicing the DeVilbiss Oxygen Concentrator, be absolutely certain that the correct tools are used and that the parts are free of oil and grease or any material not compatible with oxygen. Teflon® tape is recommended and must be applied to the male threads omitting the first thread to eliminate the possibility of tape particles entering the oxygen system. To remove back cabinet: The majority of all the servicing and repairs can be done without removing the front cabinet completely. However, it may be loosened or removed to gain access to the components behind it. 1. Unplug the unit from the wall outlet. 2. Remove the six screws that secure the back cabinet – 2 behind handle, 2 near the bottom, and 1 on each side. Electric shock hazard. Do not remove cabinet. The cabinet should only be removed by a qualified DeVilbiss homecare provider. Disconnect the power cord from the wall outlet before attempting repairs on the unit. Extra care should be taken if it is necessary to operate the unit with the cabinet removed. NOTE–All six screws are the same size. NOTE–Be sure to read all of the steps involved before beginning any of the procedures in this manual. 3. Remove the back cabinet by sliding it toward the rear until clear. 4. To reassemble, reverse steps 2-3. NOTE– After repairing or replacing a component, run the unit for 20 minutes, check the oxygen concentration and test for leaks. To loosen the front cabinet: 5. Remove the two screws (located directly above the hour meter) that hold the front cabinet to the unit’s internal structure. 6. Remove the screw located near the bottom of the recessed humidifier compartment on the front of the unit. The top of the front cabinet can now be tilted forward to allow access to the components behind it. The DeVilbiss Oxygen Concentrator is designed for ease of service. To aid service personnel a Service Kit (part #444-501) is available which contains the necessary gauges, tools, and testing instruments to properly service the oxygen concentrator. See list below. In addition, you will also need an oxygen analyzer (part #R217P62 or R218P12) to periodically check oxygen concentration levels and leak test solution. To remove the front cabinet completely: 7. Tilt cabinet forward and disconnect the ribbon connector from the PC board. 8. Before disconnecting the wires from the power switch and circuit breaker note their positions in order to reconnect them properly; then disconnect the wires. 9. Disconnect the hose at the bottom of the flow meter and remove cabinet. 10. To reassemble reverse steps 5 – 9 making sure bottom of cabinet is inserted securely in base of unit. WARNING A certified leak detection solution, such as SWAGELOK #MS-Snoop® or equivalent, is needed to test for system leaks. The solution must not contain ethylene glycol. CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. NOTE–Apply leak test solution to all fittings and hose connections with the unit running. If an air leak is present, the solution will bubble. All leaks should be repaired before putting the concentrator back in service. NOTE–All 525 models are now being built using Plastite cabinet screws. These are special thread forming screws; so do not over tighten. They should be torqued to 18 – 20 in-lbs. The original cabinet screw has much finer threads and is not a thread forming screw. The following parts are included in the Service Kit: 1 1 1 1 1 1 2 1 2 1 1 1 1 1 1 LT-2023 Slotted bit #1 Phillips bit #2 Phillips bit Crescent wrench 8” Duckbill pliers Voltmeter Pressure/Vacuum gauge Tool box Test Fittings Torx screwdriver w/bits Channel Lock Pliers 1/4” Ratchet wrench 10mm Socket l/4” Drive 1/4” Drive extension Plastic storage case These screws are not interchangeable, so be sure to order the correct part number. See figure below. Plastite Cabinet Screw Original Cabinet Screw (525DD-628) For units with Serial Numbers starting with F To remove front and back cabinets: 1. Ensure the unit is unplugged from the wall outlet. 2. Remove the six screws that secure the back cabinet – 2 behind handle, 2 near the bottom, and 1 on each side. NOTE–All six screws are the same size. 3. Remove the back cabinet by sliding it toward the rear until clear. 4. Remove the front cabinet by pushing the top shoulders toward the back of the unit, then outward away from behind the bib. Tilt the top of the front cabinet forward until it can be pulled out of the base of the unit. 5. To reassemble, reverse steps 2-4. The majority of all the servicing and repairs can be done without removing the front bib. However, it may be loosened or removed to gain access to the components behind it. 14 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T To loosen the bib: ee Figures below showing pressures being checked at accumulator tank "T" S fitting and manifold. 6. Remove the two screws (located directly above the hour meter) that hold the bib to the unit’s internal structure. The bib can now be tilted forward to allow access to the components behind it. To remove the bib completely: 7. Disconnect the ribbon connector from the PC board. 8. Before disconnecting the wires from the power switch and circuit breaker note their positions in order to reconnect them properly; then disconnect the wires. 9. Tilt the top of the bib forward to release it from the slot in the body of the concentrator. 10. Disconnect the hose at the bottom of the flow meter and remove bib. 11. To reassemble reverse steps 6 – 10 making sure bottom of front cabinet is inserted securely in base of unit. Accumulator "T" Fitting 5.Turn the unit “On” with the flow rate set to maximum recommended flow, which is usually 5 lpm. Allow the unit to run for 5 minutes before observing the pressures. During each timed cycle, the average pressure in the oxygen accumulator will rise and fall. The high pressures should be consistent and the low pressures should be consistent. NOTE–The bib tab must be inserted into the slot above the rotary valve before securing bib. NOTE–All 525 models are now being built using Plastite cabinet screws. These are special thread forming screws; so do not over tighten. They should be torqued to 18 – 20 in-lbs. The original cabinet screw has much finer threads and is not a thread forming screw. NOTE–Pressure drop will be 4-5 psi each cycle. NOTE–Expected normal pressures observed depend on altitude and flow rate. See the Expected Accumulator Tank Pressure Range chart below. • Increases in altitude and flow rate will slightly decrease accumulator pressures. • Lower altitudes and flow rates will slightly increase accumulator pressures. These screws are not interchangeable, so be sure to order the correct part number. See figure below. Plastite Cabinet Screw Manifold NOTE–A defective check valve, in either the purge harness or the manifold, may cause a rapid drop in accumulator pressure below the minimum value. Original Cabinet Screw (525DD-628) EXPECTED ACCUMULATOR TANK PRESSURE RANGE @ 5LPM Altitude Psi kPa ACCUMULATOR PRESSURE TEST 0 to 457 m 0 to 1500 ft. 23-32 159-221 To check accumulator pressures: 457 to 914 m 1500 to 3000 ft. 21-29 145-200 914 to 1524 m 3000 to 5000 ft. 20-27 138-186 1. Make sure the unit is “Off.” 2.Use the Cabinet Removal instructions listed previously to open the unit for testing. 3.Use the pressure gauge (part #PVO2D-601) and pressure test assembly (part #303DZ-637) included in the Service Kit. Includes a pressure drop of 4-5 psi and 27-34 kPa during the pressure cycle. 6. Refer to the Type 1 – Purity Issues, found under Simplified Troubleshooting, to determine the appropriate action to take in resolving abnormal pressure cycles. NOTE–A defective compressor will be indicated by slowly rising pressure. Pressure may only reach a certain level and then stop. Pressure Gauge Low oxygen concentration levels and accumulator pressures higher than normal may indicate defective sieve beds. Severely contaminated beds may also cause the pressure relief valve on the compressor to open. Pressure Test Assembly NOTE–A malfunctioning rotary valve may also cause high accumulator tank pressure and activation of the pressure relief valve. In this case it should be determined whether the problem is with the sieve beds, valve, or both. 4.Remove the tubing cap from the accumulator tank fitting or from the manifold attached to the tank, and attach the 1/16" (1.6 mm) diameter tubing from the gauge to the accumulator tank fitting just vacated above. 15 LT-2023 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T 7. Install the new capacitor and secure with a new cable tie. 8. Replace the compressor and reassemble cabinet. 9.Replace the back cabinet and secure with the six screws. AUXILIARY OXYGEN PORT All 525 series concentrators are now being manufactured with an auxiliary oxygen port located on the back of the unit. Concentrators with the auxiliary port have serial numbers starting with “N” or “R”. CHECK VALVES/MANIFOLD To fill oxygen cylinders: This external port can be used to fill oxygen cylinders with an FDA-cleared cylinder filling device that is designed to use oxygen from a concentrator to fill a cylinder. The port is only for use with FDA-cleared filling devices with compatible oxygen input specifications.The flow meter should be set at 3 LPM or less when the concentrator is being used during cylinder fill. The port does not affect concentrator performance if properly used. See figures below. Auxiliary Port Output Specifications: Outlet Pressure ................................... 8.5±0.5 psi Outlet Flow....................................... 2.5±0.5 LPM Outlet Oxygen...............................................>90% Operation Time ....................................Continuous Refer to the cylinder filling device instruction guide for the oxygent input/output specifications, connection and operating instructions. Manifold And Sieve Bed Check Valves The manifold check valves, located in the manifold attached to the accumulator tank, and the sieve bed check valves, located in the purge harness, between the outlet of each sieve bed and the accumulator tank, allow oxygen to pass from the sieve beds to the accumulator tank, when the bed pressure is greater than the accumulator tank pressure. These valves also prevent reverse flow of oxygen from the accumulator tank to the sieve beds. The manifold also directs a small amount of pressurized oxygen into the discharging sieve bed to aid the nitrogen exhaust process. The purge harness includes a fixed orifice that performs the same function. A defective manifold or sieve bed check valve will result in lower oxygen concentrations and accumulator pressures. Use the Accumulator Pressure Test described in this section to troubleshoot manifold or sieve bed check valves. • To replace a defective sieve bed check valve, remove the tubing on either side of it and install a new valve making sure the outlet end is toward the accumulator tank. • To replace a defective manifold check valve, replace the manifold. NOTE– Also see manifold section on page 19. Final Check Valve Transfill Hose (PF1100TUB) Auxiliary Port The final check valve is located, on older units, between the final bacteria filter and the oxygen outlet fitting. Transfiller Caddy (525DD-650) To replace the final check valve: 1. Make sure the unit is unplugged from the wall outlet. 2. Use the Cabinet Removal instructions listed previously to open the unit. 3. Remove the hose from the outlet side of the final bacteria filter. 4. Remove the two screws from the back of the oxygen outlet fitting assembly and remove the assembly. 5. Remove the hose from each end of the final check valve. 6. Attach the hoses to a new check valve. Make sure that the flat side of the check valve is directed toward the oxygen outlet fitting. 7. Replace the outlet fitting assembly and connect the hose to the filter. 8. Reassemble the cabinet. CAPACITOR The capacitor enables the compressor to start and run by supplying voltage to the windings of the compressor motor. A defective capacitor will result in the compressor running slower or not starting. CAUTION–The 525DS concentrators use a GSE compressor with a 22 mfd capacitor. If replacement is necessary, be sure the correct capacitor is installed. CAUTION–The 525KS/525PS concentrators use a GSE compressor with a 10 mfd capacitor. If replacement is necessary, be sure the correct capacitor is installed. WARNING Electric Shock Hazard. When replacing the capacitor, do not touch the terminals or allow metal objects to come in contact with the terminals on the capacitor. The capacitor may hold a charge for several days after the unit is turned off. COMPRESSOR The DeVilbiss Oxygen Concentrator uses a double-head, oil-free compressor. The compressor is secured to the base with four motor mounts. If a defective capacitor is suspected, a new one must be installed. The capacitor is strapped into a well molded into the bottom of the unit next to the cooling fan. Pressure Relief Valve Intake Fitting To replace the capacitor: 1. Make sure the unit is unplugged from the wall outlet. 2.Remove the back cabinet. Use the Cabinet Removal instructions listed previously. 3.Remove the compressor. Use the To Remove the Compressor instructions listed under Compressor. 4.Disconnect the two wires from the terminals on the capacitor. 5.Cut the nylon cable tie holding the capacitor in place and remove the capacitor. 6. Reconnect the wires to the new capacitor. Exhaust Fitting Electrical Connector LT-2023 16 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T A compressor that is worn or defective may: • cause pressure to rise slowly. • cause excessive noise and/or vibration. • cause lower oxygen concentrations. To test the compressor operating voltage: The compressor requires line voltage to operate. If the compressor does not start when the unit is turned on, the voltage input must be tested: 1.This voltage can be checked at the compressor connector using a voltmeter or test light connected to the brown and blue wires. The voltmeter is the best way to test. A worn or defective compressor can be caused by a defective internal component such as: • reed valve • o-ring • gasket • cup seal Voltmeter Leads These components are included in the Compressor Rebuild Kit (525D-643 and 525K-643). Cup Seal Rod Screw Spacer Rod (beneath cup retainer) Sleeve Cup Retainer Testing Compressor Voltage 2.If no voltage is detected, disconnect power and check for loose or broken wires between the compressor connector and switch or wire harness. 3.If there is voltage at the compressor connector, then either the capacitor or the compressor itself is defective. To test the compressor for proper output: Valve Plate NOTE–If the compressor is not providing a high enough output the patient alert system may be activated. Retainer 1. Use the Cabinet Removal instructions listed previously to open the unit for testing. 2. Use the Accumulator Pressure Test listed previously to observe the high pressures, low pressures and the pressure drop. 3. Refer to the Type 1 – Purity Issues, found under Simplified Troubleshooting, to determine the appropriate action to take in resolving abnormal pressure cycles. O-Ring Reed Valve or Flapper Screw Sleeve PR Valve Exhaust Fitting NOTE–A compressor, which slowly builds pressure that remains below 25 psi, indicates worn cup seals and/ or reed valves. If these conditions are observed then: Compressor Head Head Gasket Connecting Tubes •The unit filter(s) may be occluded—check the air filter, compressor filter, and intake filter for occlusions. •There may be a severe leak in the system—check for air leaks using a leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). Intake Fitting CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. •The compressor reed valves, cup seal, or the compressor itself may be defective. If the filters are not occluded and no leaks are found, the compressor must then be removed and repaired or replaced. NOTE–A built-in thermal cutoff switch will shut the compressor off if it becomes overheated. This protects the compressor from damage caused by heat build-up. (Some models have an auxiliary thermostat mounted within the compressor compartment.) Should this condition occur, the compressor will require several minutes for the thermo-protective device to reset. NOTE–A pressure relief (PR) valve is located on the pressure head to prevent high pressure build up in the system should a component malfunction occur. To remove the compressor: 1. Make sure the unit is unplugged from the wall outlet. 2. Use the Cabinet Removal instructions listed previously to open the unit. 3.Disconnect the compressor wires by disconnecting the compressor electrical connector. 4.Remove the ladder clamp and hose from the exhaust fittings on the compressor head and compressor filter if applicable. 5.Carefully place the concentrator on its front side. From the bottom of the unit, remove the four 10 mm hex nuts that secure the motor mounts. 17 LT-2023 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T 5.Secure motor mounts with hex nuts. 6. Reconnect hoses to the fittings at compressor exhaust and filter. 7. Reconnect the compressor electrical connector. 6.Lift compressor partially out of the compressor housing area and loosen ladder clamp securing hose to the intake fitting, then remove hose. 7.Compressor may now be removed from the compressor housing area. CAUTION–If the unit has been running recently, the compressor may be hot. COOLING FAN 8.Unscrew motor mounts from compressor feet by hand. The cooling fan provides a constant air flow to cool the compressor. The cooling fan is located in the bottom of the unit below the compressor. To inspect and/or replace internal components: A defective cooling fan may cause the compressor’s internal thermo-protective (thermal cut off) device to activate and shut the compressor off. Should this condition occur, the compressor will require several minutes for the thermoprotective device to reset. 1.Remove the eight screws that hold the compressor heads in place. When removing the heads, be sure to keep each head and its components with the correct compressor side. 2.Check for proper placement of or damage to the gaskets on the bottom of the compressor heads. Replace if damaged. 3.Remove reed valve plates. A reed valve is located on each side of the valve plate. 4.The compressor reed valves should be flush with the valve plate. If the valve is broken or not flush with the valve plate, or foreign matter is detected inside the head, clean or replace the compressor reed valves. If the cooling fan is defective, it must be replaced: 1. Make sure the unit is unplugged from the wall outlet. 2. Use the Cabinet Removal instructions listed previously to open the unit. 3. Use the To Remove the Compressor instructions listed under Compressor. 4. Disconnect the cooling fan terminals. 5.Note the position of the fan and fan guard before removing the four retaining screws that secure the fan to the base of the unit. 6.Remove the defective fan and secure the replacement fan in position with the four retaining screws. To replace the compressor reed valves: a.Remove the screw holding the compressor reed valves in position on the valve plate and discard the used reed valves. b.Position the new reed valves so that they are centered and completely cover the holes in the valve plate. c.Place the metal retainer on the reed valves and secure with the reed valve screw. 5.Check for proper placement of or damage to the rubber o-ring on the bottom of the valve plate. Replace if damaged. 6.Remove piston sleeves by pulling upward and inspect cup seal on pistons. Replace if badly worn or damaged. NOTE–When installing the fan, be sure the air flow directional arrow on the side of the fan is directed away from the compressor and fan guard is reinstalled properly. 7. 8. FLOW METER To replace cup seal: There are two flow meters available for the DeVilbiss 525 series oxygen concentrators: a standard 5 lpm flow meter (505DZ-607) and a low output flow meter (515LF-607). The low output flow meter should be used when the prescribed flow rate is at or under 1 lpm. a. Remove rod screw from top of piston. b. Remove the cup retainer plate. c. Discard defective cup seal. d. Place new cup seal into position. e. Replace cup retainer plate. f. Secure with screws. 7.Discard old sleeve and position new sleeve at 45 degree angle over the piston. Carefully push it down as you rotate it slightly around the top of the piston until it is in place. 8.Place valve plates on the compressor so that heads of reed valve screws are aligned with the indentation in top of pistons. 9.Install the compressor heads so that the holes in the heads are aligned with the holes in the compressor housing. 10. Secure compressor heads with the screws. NOTE–Installation of a low output flow meter eliminates the low flow alert for flow delivery under 0.5 lpm. Both pressure-compensated flow meters have an accuracy level of ±5% at full scale (exception: +0% / -5% at 5 lpm) and are designed for use at 8.5 psi (58.6 kPa). Both flow meters can be locked using a 1/16th inch Allen wrench and tightening the locking screw behind the flow meter knob. Locking Screw To check for leaks in the flow meter tubing: 1.Check for leaks using a certified leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). 2.Apply leak test solution to all fittings and hose connections with the unit running. To replace the compressor: CAUTION–The 525DS concentrators use a GSE compressor with a 22 mfd capacitor. The 525KS/525PS concentrators use a GSE compressor with a 10 mfd capacitor. If replacement is necessary, be sure the correct capacitor is installed. CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. 1.Inspect the motor mounts. Replace if damaged. Attach mounts to compressor feet. 2.Inspect the capacitor to determine if replacement is necessary. If capacitor is wrong value for compressor or replacement is desired, refer to Capacitor section. 3. Reconnect hose to the compressor intake fitting. 4.Position compressor on the base of the unit so that the studs on the motor mounts protrude through the holes in the base of the unit. LT-2023 Reconnect the electrical connector. Reinstall the compressor. 3. If an air leak is present, the solution will bubble. All leaks should be repaired before putting the concentrator back in service. WARNING Electric Shock Hazard. Use caution when leak testing near electrical connections. 18 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T 6. To replace the flow meter: 1. Make sure the unit is unplugged from the wall outlet. 2.Use the Cabinet Removal instructions listed previously to open the unit. 3. From behind the bib/front cabinet, remove the 2 hoses from the flow meter. 4.While squeezing tabs on flow meter brackets, push the flow meter through the bib/front cabinet or just unscrew flow meter fittings. 5. Install new flow meter and reconnect hoses. Replace the flow meter. a. Remove all tubing from the current flow meter. b. Unscrew white plastic connectors or fittings. To install or replace the low output flow meter: 1. 2. 3. Ensure that the unit is unplugged from the wall outlet. Use the Cabinet Removal instructions listed previously to open the unit. Disconnect: a. All wires, terminals and connectors attached to the PC board. b.The 1/8 ID blue tubing from the Oxygen Sensing Device (OSD) and from the flow meter. c. The screw that secures the PC board to the unit. 4.Build the T-fitting assembly. (Skip this step, if replacing an existing low output flow meter.) a.Cut the existing 1/8 inch blue tubing that connected the OSD and the flow meter into two equal pieces. 7. c.Remove the metal brackets and push the meter through the case. Save the meter and metal brackets for later use, if desired. d.Align the new metal brackets on the flat surfaces of the new meter. e. Push the flow meter into the case with the highest flow at the top. Reassemble the unit. a.Connect one end of the 1/8 inch blue tubing to the T-fitting assembly to the bottom of the flow meter and the other end to the bottom of the OSD. b.Reconnect the PC board wires, tubing, terminals and connectors. HOUR METER To replace the hour meter: 5. 1. Make sure the unit is unplugged from the wall outlet. 2.Use the Cabinet Removal instructions listed previously to open the unit. 3. Remove PC board. 4.Squeeze locking tabs on both sides of hour meter while pushing upward to remove meter. 5.Install a new hour meter by applying downward pressure until it snaps into position. 6. Replace PC board and reconnect all wires, connectors and tubing. b.Attach each piece of the cut tubing to the large ends of the T-fitting, included in the kit. c.Attach one end of the 1/16 inch ID blue tubing, included in the kit, to the small end of the T-fitting. Position the bleed-off orifice and connect tubing. 1 CAUTION–Do not apply any force or flex the PC Board when connecting or disconnecting electronic or pneumatic components. Damage to the electronic assembly is possible. 2 MANIFOLD The manifold is attached to the accumulator tank and performs the same function as a purge harness. It directs a small amount of pressurized oxygen into the discharging sieve bed to aid the nitrogen exhaust process while it ensures that the majority of pressurized oxygen is directed into the accumulator tank. The manifold also prevents reverse flow of oxygen from the accumulator to the sieve beds. Insert an opened paperclip through the hole in the case, behind where the PC board was installed (Fig. 1), until it appears in the filter cavity (Fig. 2). 3 4 Place the orifice fitting, included in the kit, onto the end of the paperclip (Fig. 3). Push the fitting through the foam until it’s inside the case. Remove the paperclip and attach the 1/16 inch ID tubing (Fig. 4). See CHECK VALVES: MANIFOLD and SIEVE in this manual for additional information. 19 LT-2023 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T 1. Ensure any contamination problem has been corrected. 2. Ensure the unit is unplugged from the wall outlet. 3.Cut the plastic cable ties that secure the sieve beds to the internal structure of the unit. 4. Remove the tubing from the fittings at the top of each sieve bed. 5.Remove the hose clamps and hose from the bottom of the sieve beds. 6.Install new sieve beds in reverse order using new plastic cable ties. Position the new beds so that the bed serial number label is at the top of the unit. 7.Leak test all connections with a certified leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). Apply leak test solution to all fittings and hose connections with unit running. If an air leak is present, the solution will bubble. All leaks should be repaired before putting the unit back in service. MOLECULAR SIEVE BEDS The build and release of pressure in the sieve beds indicates the health of the sieve material and the operation of contributing components. Determine ‘good’ or expected pressures for your altitude by testing bed pressures on multiple oxygen concentrators that are producing at least 93% purity. Check sieve bed pressures at the accumulator tank using the manifold test point or the accumulator tank test point. See Figures below. CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. Accumulator "T" Fitting Manifold POWER CORD NOTE–The pressure will rise and fall as it cycles through the rotary valve. To replace the power cord - 115 volt units only: Also consider the acceptable pressure range for various altitudes as shown below. 1. Make sure the unit is unplugged from the wall outlet. 2. Use the Cabinet Removal instructions listed previously to open the unit. 3. Disconnect the power cord connector. 4. Note wire colors and socket locations before removing wires. 5.Using a pair of duckbill pliers, squeeze the power cord strain relief and pull it out of the base of the unit. 6.Insert a new power cord through the hole in the base of the unit and secure with strain relief. 7.Insert sockets into connector housing and then reconnect the power cord connector. 8. Replace back cabinet and secure with the six screws. EXPECTED ACCUMULATOR TANK PRESSURE RANGE @ 5 LPM Altitude Psi kPa 0 to 457 m 23-32 159-221 0 to 1500 ft. 457 to 914 m 21-29 145-200 1500 to 3000 ft. 914 to 1524 m 20-27 138-186 3000 to 5000 ft. Includes a pressure drop of 4-5 psi and 27-34 kPa during the pressure cycle. To test sieve bed pressures: POWER SWITCH To replace the power switch: 1.Remove the plugged piece of 1/16" tubing from the manifold or the accumulator tank. 2.Connect the pressure gauge to the test point and observe the cycling of pressures. Refer to the section on Normal Operating Sequence and also Accumulator Pressure Test. a.If it is determined that the valve is not cycling the pressure, refer to the section on Rotary Valve testing. b. If the unit is cycling properly, allow it to operate for 20 minutes. 3.After 20 minutes of operation, observe the high and low pressures and check the oxygen concentration level. a.If the pressures are within the expected range and the oxygen concentration is within specification, the sieve material is effective. b.If pressures are not within the expected range and/or the oxygen concentration is not within specification, refer to the Simplified Troubleshooting Table in this manual to determine the proper corrective action. 1. Make sure the unit is unplugged from the wall outlet. 2.Use the Cabinet Removal instructions listed previously to open the unit. 3.Note the position of the wires and switch before removing the wires from the switch terminals. 4.While squeezing the locking tabs on the sides of the switch, push the switch out of the front of the unit. 5.Install the new switch in the correct orientation making sure that it locks into position. 6. Reconnect the wires to the switch terminals. Switch Detail NOTE–If the molecular sieve material is found to be no longer effective, first search for the source of a malfunction in the system; then, for a cause for contamination (such as leaks) and take corrective action. Brown Wire (or 2 Black Wires) *Red Wire 2 Black Wires (or 1 Brown Wire) *Either red wire can be connected to either left hand terminal of the switch To replace the molecular sieve beds: NOTE–Make sure that the sealing caps remain on the new sieve beds until just prior to connecting hoses and tubing. LT-2023 *Red Wire 20 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T NOTE–If the concentrator has been unused for an extended period, the unit must run 20 minutes before the power fail alert will be enabled. This alert is powered by a capacitor on the PC board. PRESSURE REGULATOR The pressure regulator stabilizes the flow of oxygen to the patient and establishes back pressure on the system. It is preset at 8.5 psi (58.6 kPa) and should not have to be adjusted in the field. It is not necessary to test the pressure regulator unless there is a problem with flow rate accuracy. The following test is only needed when troubleshooting flow rate issues. CAUTION–Do not apply any force or flex to the PC Board when connecting or disconnecting electronic or pneumatic components. Damage to the electronic assembly is possible. To test the pressure regulator: To remove and replace the PC board: 1. Turn the unit “On.” 2. Set the flow meter at 2 lpm. 3.Attach a pressure gauge (part #PVO2D-601) to the oxygen outlet to obtain a reference pressure. Use this reference pressure to determine if further testing is needed. 4. If the reference pressure varies from the expected pressure by more than ± 0.5 psi or ± 5.9 kPa, connect a ‘T’ fitting directly between the pressure regulator fitting and the pressure regulator tubing and attach the pressure gauge to the 3rd leg of the ‘T’ fitting. If the pressure reading is not within 8.5 ± .5 psi (58.6± 5.9 kPa), adjustment to the pressure regulator is required. 1. Make sure the unit is unplugged from the wall outlet. 2. Use the Cabinet Removal instructions listed previously to open the unit. 3. Disconnect all wires, terminals and connectors. 4. Remove the 1/8" (3.2mm) tubing attached to the sensor. 5.Remove the screw that secures the board to the unit and remove the PC board. 6. Install the new PC board and secure it using the screw. 7.Reconnect all electrical wires, terminals, connectors and the tubing to the sensor. To adjust the pressure regulator: The timed rotary valve alternately distributes pressure supplied by the compressor to the sieve beds. While one bed is being pressurized the other bed is being exhausted through the valve exhaust port. ROTARY VALVE 1. If necessary, the pressure regulator can be adjusted by turning the allen screw on top of the regulator until the pressure is within specification. Turn clockwise to increase the pressure, and counterclockwise to decrease the pressure. The valve contains two revolving discs powered by a stepper motor to cycle the pressure between the beds. DC voltage is supplied by the PC board to the motor windings causing the internal discs to turn and direct pressure to the proper sieve bed. NOTE–Before adjusting the pressure regulator, make sure no leaks exist by using a certified leak detection solution such as Snoop® or equivalent (must not contain ethylene glycol). Stepper Motor CAUTION–Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly. Valve Wire Harness A malfunction in the pressure regulator will cause either a loss or fluctuation in the oxygen flow which will be seen on the flow meter or a decrease in oxygen concentration. To replace the pressure regulator: 1. Make sure the unit is unplugged from the wall outlet. 2.Use the Cabinet Removal instructions listed previously to open the unit. 3. Remove the tubing clamp and tubing from the pressure regulator. 4. Unscrew the regulator from the accumulator tank. 5.Install a new regulator on the accumulator tank and attach the tubing and tubing clamp. Front View PRINTED CIRCUIT BOARD Rear View If the rotary valve is not shifting properly, one bed may pressurize continuously; the pressure building until the pressure relief valve, on the compressor, releases the excess with a sputtering sound. The printed circuit (PC) board is responsible for monitoring and controlling the DeVilbiss Oxygen Concentrator. The PC board has preset alerts for low flow and power failure. Should any of the alert values be exceeded, the patient alert system will activate. There are several reasons why the rotary valve could malfunction; therefore the cause of failure must be determined before corrective action can be taken. To test the rotary valve: 1.Use the Cabinet Removal instructions listed previously to open the unit for testing. 2.Connect a pressure gauge to the test point on the manifold or accumulator tank to observe unit cycling and the high and low bed pressures. Refer to the section on Normal Operating Sequence and also Accumulator Pressure Test. a.If it is determined that the valve is shifting properly and the high and low pressures are consistent, the valve is operating correctly. OSD Valve Wire Harness Fuse 21 LT-2023 C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T b.If it is determined that the valve did not shift properly or the high pressures or low pressures are uneven, continue testing with step 3. 3.The stepper motor on the rotary valve has multiple windings so there are several voltage readings that need to be checked in order to determine if the problem is being caused by the PC board or the valve itself. Testing for proper voltage is done at the wire harness connectors on the valve or on the PC Board using a voltmeter. Voltmeter Leads Testing Valve Voltage at PC Board elow are the valve voltage test positions and voltages that should be present at B each one: NOTE–The 1-3 Volt reading will go on and off as the valve cycles. 525 Series have a fixed cycle time of 3.6 seconds @5 LPM. However, when the concentrator goes into its “turn-down” mode, the cycle time is shortened. Negative Lead Positive Lead Voltage Reading Orange Yellow or White 12 VDC Orange Gray 5 VDC Orange Brown 5 VDC (intermittent) Black or Green or Red or Blue Yellow or White 1-3 VDC (intermittent) a.If any of these readings are not obtained, check for loose or broken wires in the harness. If wire harness is okay, replace the main PC board. b. If proper voltages are obtained, replace valve. To replace the rotary valve: 1. Make sure the unit is unplugged from the wall outlet. 2.Use the Cabinet Removal instructions listed previously to open the unit. 3. Unplug valve wire harness from the valve. 4.Loosen clamps and remove bed hoses from each side of valve. 5.Loosen ladder clamps and remove pressure intake and exhaust hoses from backside of valve. 6.Install the new rotary valve by reversing the above procedure. LT-2023 22 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "F" (115V Units) 525DS & 525DS-Q Units - Serial # "J" (115V Units) Front View 10-11 Rear View 48 40 29-30 43 23 5 7 17 9 46 8 14 45 3 Front Inside 33 53 36 1 44-32 47 Behind Front Cover/Bib Rear Inside 49 4 38 28 or 41 52 24 27 53 25 39 16 (Final Check Valve not visible - behind oxygen outlet port) Fan & Capacitor 34 15 12 50 35 6 20-21 54 42 19 18 13 Compressor Box 51 55 22 31 37 23 LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "F" (115V Units) 525DS & 525DS-Q Units - Serial # "J" (115V Units) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 LT-2023 PART DESCRIPTION Accumulator Tank Auxiliary Oxygen Port Cabinet Parts: Base Bib Communication Port Door Compressor Box Filter Door Front Cover Rear Cover Cabinet Screw (Original) Cabinet Screw (plastite) Cable Tie, (Sieve Bed) Capacitor (Motor Start/Run) Caster, Non-locking Check Valves: Bed Check Valve Final Check Valve Circuit Breaker Cooling Fan Cooling Fan Guard Compressor Compressor Rebuild Kit Exhaust Muffler Filters: Cabinet Air Filter Intake Bacteria Filter Compressor Filter (Plastic) Compressor Filter (Sintered Bronze) Final Bacteria Filter Fittings: Accumulator Tank "Tee" Flow Meters: Standard Flow Meter Low Output Meter Foam Kit Fuse (For PC Board) Hoses: 1/8" ID (Blue) (4' Lg) 1/2" ID PVC Braided (4' Lg) 1/2" ID Silicone Braided (2' Lg) Hose Clamps: Plastic (1/4" ID Hose) Ladder Clamp (1/2" ID Hoses) Hour Meter Intake Canister Light Panel / Ribbon Cable Purge Manifold Assembly Motor Mounts Oxygen Outlet Port PC Board Power Cord w/Strain Relief Power Cord Strap Power Cord Strain Relief Power Switch Pressure Regulator Pressure Relief Valve Rotary Valve Sieve Bed Wire Harness: Communication Harness/Port Main Harness Rotary Valve Harness 525DS "F" SERIAL # 525DS "J" SERIAL # 525DS-Q "J" SERIAL # 525D-603 525D-604 525D-640 525D-605 525D-639 525D-601 525D-602 525DD-628 N/A 505DZ-617 525D-616 501DZ-603 525DD-603 N/A 525DD-640 525DD-605 525DD-639 525DD-601 525DD-602 525DD-628 N/A 505DZ-617 525D-616 501DZ-603 525DD-603 N/A 525DD-640 525DD-605 525DD-639 525DD-601 525DD-602 525DD-628 N/A 505DZ-617 525D-616 501DZ-603 PVO2D-607 PVO2D-607 525D-613 525DD-634 N/A 525D-625 525D-643 515A-705 PVO2D-607 PVO2D-607 525D-613 525DD-634 N/A 525D-625 525D-643 515A-705 PVO2D-607 PVO2D-607 525D-613 525DD-634 N/A 525D-625 525D-643 515A-705 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 444-582 444-582 444-582 505DZ-607 515LF-607 525D-600 515ADZ-627 505DZ-607 515LF-607 N/A 515ADZ-627 505DZ-607 515LF-607 N/A 515ADZ-627 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 N/A 444-566 PV5LD-617 525DD-614 525D-615 N/A 505IZ-609 525DD-606 525DS-622 PV5LD-618 MC29D-657 505DZ-645 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525DS-622 PV5LD-618 MC29D-657 505DZ-645 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525DS-622 PV5LD-618 MC29D-657 505DZ-645 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 525D-608 515DZ-623 525D-621 525D-608 515DZ-623 525D-621 525D-608 515DZ-623 525D-621 525D-610 N/A 24 525D-610 N/A 525D-610 N/A F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "N" (115V Units with Auxiliary Oxygen Port) 525DS-Q Unit - Serial # "R" (115V Units with Auxiliary Oxygen Port) Front View Rear View 40 11 48 7 23 43 2 29-30 8 17 14 46 9 45 47 3 Front Inside 1 44 49 36 (Plastic Clamp Behind Front Cover 27 Compressor Box - Serial # "N" not shown secures hose to regulator) 33 41 54 32 26 52 55 (Sintered filter located in hose) Fan & Capacitor 12 19 18 13 Compressor Box - Serial # "R" 22 31 37 51 25 LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "N" (115V Units with Auxiliary Oxygen Port) 525DS-Q Unit - Serial # "R" (115V Units with Auxiliary Oxygen Port) Rear Inside - Serial # "N" Rear Inside - Serial # "R" 38 24 34 39 25 50 35 6 20-21 42 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 LT-2023 PART DESCRIPTION Accumulator Tank Auxiliary Oxygen Port Cabinet Parts: Base Bib Communication Port Door Compressor Box Filter Door Front Cover Rear Cover Cabinet Screw (Original) Cabinet Screw (plastite) Cable Tie, (Sieve Bed) Capacitor (Motor Start/Run) Caster, Non-locking Check Valves: Bed Check Valve Final Check Valve Circuit Breaker Cooling Fan Cooling Fan Guard Compressor Compressor Rebuild Kit Exhaust Muffler Filters: Cabinet Air Filter Intake Bacteria Filter Compressor Filter (Plastic) Compressor Filter (Sintered Bronze) Final Bacteria Filter Fittings: Accumulator Tank "Tee" 525DS "N" SERIAL # 525DS-Q "R" SERIAL # 525DD-603 N/A N/A 525DD-615 525DD-639 525DD-611 525DD-612 N/A N/A 505DZ-617 525D-616 501DZ-603 525DD-603 N/A N/A 525DD-615 525DD-639 525DD-611 525DD-612 N/A N/A 505DZ-617 525D-616 501DZ-603 N/A N/A 525D-613 525DD-634 N/A 525D-625 525D-643 515A-705 N/A N/A 525D-613 525DD-634 N/A 525D-625 525D-643 515A-705 303DZ-605 MC44D-605 525D-622 525DD-626 PV5LD-651 303DZ-605 MC44D-605 525D-622 525DD-626 PV5LD-651 N/A N/A 525D-610 N/A 26 525D-610 N/A F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "N" (115V Units with Auxiliary Oxygen Port) 525DS-Q Unit - Serial # "R" (115V Units with Auxiliary Oxygen Port) PART DESCRIPTION 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 Flow Meters: Standard Flow Meter Low Output Meter Foam Kit Fuse (For PC Board) Hoses: 1/8" ID (Blue) (4' Lg) 1/2" ID PVC Braided (4' Lg) 1/2" ID Silicone Braided (2' Lg) Hose Clamps: Plastic (1/4" ID Hose) Ladder Clamp (1/2" ID Hoses) Hour Meter Intake Canister Light Panel / Ribbon Cable Purge Manifold Assembly Motor Mounts Oxygen Outlet Port PC Board Power Cord w/Strain Relief Power Cord Strap Power Cord Strain Relief Power Switch Pressure Regulator Pressure Relief Valve Rotary Valve Sieve Bed Wire Harness: Communication Harness/Port Main Harness Rotary Valve Harness 27 525DS "N" SERIAL # 525DS-Q "R" SERIAL # 505DZ-607 515LF-607 N/A 515ADZ-627 505DZ-607 515LF-607 N/A 515ADZ-627 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 N/A 444-566 PV5LD-617 N/A 525D-615 525DD-617 505IZ-609 525DD-606 525DD-622 PV5LD-618 MC29D-657 505DZ-645 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525DD-622 PV5LD-618 MC29D-657 505DZ-645 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 525DD-623 525D-621 N/A 525DD-623 525D-621 LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS Unit - Serial # "F" (220/230/240V Units) 525KS & 525KS-LT Units - Serial # "J" (220/230/240V Units) Front View 10-11 Rear View 48 40 29-30 43 23 7 5 17 9 46 8 56 14 45 Behind Front Cover/Bib 3 Front Inside 33 53 36 1 44-32 Rear Inside 49 38 28 or 41 24 52 27 53 19 18 35 50 31 6 20-21 51 13 Compressor Box 55 22 42 54 LT-2023 not visible - behind oxygen outlet port) 25 34 15 16 (Final Check Valve Fan & Capacitor 39 12 4 28 57 37 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS Unit - Serial # "F" (220/230/240V Units) 525KS & 525KS-LT Units - Serial # "J" (220/230/240V Units) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 PART DESCRIPTION Accumulator Tank Auxiliary Oxygen Port Cabinet Parts: Base Bib Communication Port Door Compressor Box Filter Door Front Cover Rear Cover Cabinet Screw (Original) Cabinet Screw (plastite) Cable Tie, (Sieve Bed) Capacitor (Motor Start/Run) Caster, Non-locking Check Valves: Bed Check Valve Final Check Valve Circuit Breaker Cooling Fan Cooling Fan Guard Compressor Compressor Rebuild Kit Exhaust Muffler Filters: Cabinet Air Filter Intake Bacteria Filter Compressor Filter (Plastic) Compressor Filter (Sintered Bronze) Final Bacteria Filter Fittings: Accumulator Tank "Tee" Flow Meters: Standard Flow Meter Low Output Meter Foam Kit Fuse (For PC Board) Hoses: 1/8" ID (Blue) (4' Lg) 1/2" ID PVC Braided (4' Lg) 1/2" ID Silicone Braided (2' Lg) Hose Clamps: Plastic (1/4" ID Hose) Ladder Clamp (1/2" ID Hoses) Hour Meter Intake Canister Light Panel / Ribbon Cable Purge Manifold Assembly Motor Mounts Oxygen Outlet Port PC Board Power Cord w/Strain Relief Power Cord Strap Power Cord Strain Relief Power Switch Pressure Regulator Pressure Relief Valve Rotary Valve Sieve Bed Wire Harness: Communication Harness/Port Main Harness Rotary Valve Harness IEC Connector Thermostat 525KS "F" SERIAL # 525KS "J" SERIAL # 525KS-LT "J" SERIAL # 525K-603 525D-604 525D-640 525D-605 525D-639 525D-601 525D-602 525DD-628 N/A 505DZ-617 525K-616 501DZ-603 525KD-603 N/A 525DD-640 525DD-605 525DD-639 525DD-601 525DD-602 525DD-628 N/A 505DZ-617 525K-616 501DZ-603 525KD-603 N/A 525DD-640 525DD-605 525DD-639 525DD-601 525DD-602 525DD-628 N/A 505DZ-617 525K-616 501DZ-603 PVO2D-607 PVO2D-607 515KZ-615 515UK-634 N/A 525K-625 525K-643 515A-705 PVO2D-607 PVO2D-607 515KZ-615 515UK-634 N/A 525K-625 525K-643 515A-705 PVO2D-607 PVO2D-607 515KZ-615 515UK-634 N/A 525K-625 525K-643 515A-705 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 444-582 444-582 444-582 505DZ-607 515LF-607 525D-600 515AKZ-610 505DZ-607 515LF-607 N/A 515AKZ-610 505DZ-607 515LF-607 N/A 515AKZ-610 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 N/A 444-566 PV5LD-617 525DD-614 525D-615 N/A 505IZ-609 525DD-606 525KS-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525KS-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525KS-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 525D-608 515AKS-623 525D-621 N/A N/A 525D-608 515AKS-623 525D-621 N/A N/A 525D-608 515AKS-623 525D-621 N/A N/A 525D-610 N/A 29 525D-610 N/A 525D-610 N/A LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS-LT Unit - Serial # "N" (220/230/240V Units with Auxiliary Oxygen Port) 525KS Unit - Serial # "R" (220/230/240V Units with Auxiliary Oxygen Port) Front View 11 Rear View 40 48 7 23 43 2 29-30 8 17 46 9 45 14 56 3 Front Inside 1 44 49 Behind Front Cover 36 (Plastic Clamp not shown - secures hose to regulator) 33 27 41 32 52 Fan & Capacitor 19 18 55 12 13 Compressor Box - Serial # "N" 51 Compressor Box - Serial # "R" 37 57 54 26 LT-2023 30 (Sintered filter located in hose) 22 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS-LT Unit - Serial # "N" (220/230/240V Units with Auxiliary Oxygen Port) 525KS Unit - Serial # "R" (220/230/240V Units with Auxiliary Oxygen Port) Rear Inside - Serial # "N" Rear Inside - Serial # "R" 38 24 34 39 25 50 35 6 20-21 42 PART DESCRIPTION 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 Accumulator Tank Auxiliary Oxygen Port Cabinet Parts: Base Bib Communication Port Door Compressor Box Filter Door Front Cover Rear Cover Cabinet Screw (Original) Cabinet Screw (plastite) Cable Tie, (Sieve Bed) Capacitor (Motor Start/Run) Caster, Non-locking Check Valves: Bed Check Valve Final Check Valve Circuit Breaker Cooling Fan Cooling Fan Guard Compressor Compressor Rebuild Kit Exhaust Muffler Filters: Cabinet Air Filter Intake Bacteria Filter Compressor Filter (Plastic) Compressor Filter (Sintered Bronze) Final Bacteria Filter 31 525KS-LT "N" SERIAL # 525KS "R" SERIAL # 525D-610 N/A 525D-610 N/A 525KD-603 N/A N/A 525DD-615 525DD-639 525DD-611 525DD-612 N/A N/A 505DZ-617 525K-616 501DZ-603 525KD-603 N/A N/A 525DD-615 525DD-639 525DD-611 525DD-612 N/A N/A 505DZ-617 525K-616 501DZ-603 N/A N/A 515KZ-615 515UK-634 N/A 525K-625 525K-643 515A-705 N/A N/A 515KZ-615 515UK-634 N/A 525K-625 525K-643 515A-705 303DZ-605 MC44D-605 525D-622 525DD-626 PV5LD-651 303DZ-605 MC44D-605 525D-622 525DD-626 PV5LD-651 LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS-LT Unit - Serial # "N" (220/230/240V Units with Auxiliary Oxygen Port) 525KS Unit - Serial # "R" (220/230/240V Units with Auxiliary Oxygen Port) PART DESCRIPTION 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 LT-2023 Fittings: Accumulator Tank "Tee" Flow Meters: Standard Flow Meter Low Output Meter Foam Kit Fuse (For PC Board) Hoses: 1/8" ID (Blue) (4' Lg) 1/2" ID PVC Braided (4' Lg) 1/2" ID Silicone Braided (2' Lg) Hose Clamps: Plastic (1/4" ID Hose) Ladder Clamp (1/2" ID Hoses) Hour Meter Intake Canister Light Panel / Ribbon Cable Purge Manifold Assembly Motor Mounts Oxygen Outlet Port PC Board Power Cord w/Strain Relief Power Cord Strap Power Cord Strain Relief Power Switch Pressure Regulator Pressure Relief Valve Rotary Valve Sieve Bed Wire Harness: Communication Harness/Port Main Harness Rotary Valve Harness IEC Connector Thermostat 32 525KS-LT "N" SERIAL # 525KS "R" SERIAL # N/A N/A 505DZ-607 515LF-607 N/A 515AKZ-610 505DZ-607 515LF-607 N/A 515AKZ-610 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 N/A 444-566 PV5LD-617 N/A 525D-615 525DD-617 505IZ-609 525DD-606 525KD-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525KD-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 525KD-623 525D-621 N/A N/A N/A 525KD-623 525D-621 N/A N/A F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525PS Unit - Serial # "F" (220/230/240V Units) 525PS Unit - Serial # "J" (220/230/240V Units) Front View 10-11 Rear View 48 40 29-30 43 23 7 5 17 9 46 8 56 14 45 Behind Front Cover/Bib 3 Front Inside 33 53 36 1 44-32 Rear Inside 49 38 28 or 41 24 52 27 53 not visible - behind oxygen outlet port) 19 25 18 34 15 16 (Final Check Valve Fan & Capacitor 39 12 4 35 50 31 6 13 20-21 22 Compressor Box 51 42 55 57 54 26 37 33 (Sintered filter located in hose) LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525PS Unit - Serial # "F" (220/230/240V Units) 525PS Unit - Serial # "J" (220/230/240V Units) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 LT-2023 PART DESCRIPTION Accumulator Tank Auxiliary Oxygen Port Cabinet Parts: Base Bib Communication Port Door Compressor Box Filter Door Front Cover Rear Cover Cabinet Screw (Original) Cabinet Screw (plastite) Cable Tie, (Sieve Bed) Capacitor (Motor Start/Run) Caster, Non-locking Check Valves: Bed Check Valve Final Check Valve Circuit Breaker Cooling Fan Cooling Fan Guard Compressor Compressor Rebuild Kit Exhaust Muffler Filters: Cabinet Air Filter Intake Bacteria Filter Compressor Filter (Plastic) Compressor Filter (Sintered Bronze) Final Bacteria Filter Fittings: Accumulator Tank "Tee" Flow Meters: Standard Flow Meter Low Output Meter Foam Kit Fuse (For PC Board) Hoses: 1/8" ID (Blue) (4' Lg) 1/2" ID PVC Braided (4' Lg) 1/2" ID Silicone Braided (2' Lg) Hose Clamps: Plastic (1/4" ID Hose) Ladder Clamp (1/2" ID Hoses) Hour Meter Intake Canister Light Panel / Ribbon Cable Purge Manifold Assembly Motor Mounts Oxygen Outlet Port PC Board Power Cord w/Strain Relief Power Cord Strap Power Cord Strain Relief Power Switch Pressure Regulator Pressure Relief Valve Rotary Valve Sieve Bed Wire Harness: Communication Harness/Port Main Harness Rotary Valve Harness IEC Connector Thermostat 525PS "F" SERIAL # 525PS "J" SERIAL # 525K-603 525D-604 525D-640 525D-605 525D-639 525D-601 525D-602 525DD-628 N/A 505DZ-617 525K-616 501DZ-603 525KD-603 N/A 525DD-640 525DD-605 525DD-639 525DD-601 525DD-602 525DD-628 N/A 505DZ-617 525K-616 501DZ-603 PVO2D-607 PVO2D-607 515KZ-615 515UK-634 N/A 525PS-625 525K-643 515A-705 PVO2D-607 PVO2D-607 515KZ-615 515UK-634 N/A 525PS-625 525K-643 515A-705 303DZ-605 MC44D-605 525D-622 N/A PV5LD-651 303DZ-605 MC44D-605 525D-622 525DD-626 PV5LD-651 444-582 444-582 505DZ-607 515LF-607 525D-600 515AKZ-610 505DZ-607 515LF-607 N/A 515AKZ-610 444-554 444-549 505DZ-634 444-554 444-549 505DZ-634 N/A 444-566 PV5LD-617 525DD-614 525D-615 N/A 505IZ-609 525DD-606 525KS-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 N/A 444-566 PV5LD-617 525DD-614 525D-615 525DD-617 505IZ-609 525DD-606 525KS-622 VARIOUS MC29D-657 N/A 505DZ-508 MC29D-612 515ADZ-614 515ADZ-702 525D-619 525D-608 515AKS-623 525D-621 N/A N/A 525D-608 515AKS-623 525D-621 N/A N/A 525D-610 N/A 34 525D-610 N/A F I G U R E S , D I A G R A M S A N D PA R T S L I S T ACCESSORIES Carton with Shipping Inserts 525D-606 Oxygen Outlet Connector - Plastic XM-1 Casters, Locking 525DS-603 Low Output Flow Meter 515LF-607 Transfill Caddy 525DD-650 Transfill Hose PF1100TUB TOOLS Service Kit 444-501 Pressure Gauge PVO2D-601 Pressure Test Assembly 303DZ-637 Oxygen Analyzer (Max O2) Oxygen Analyzer (Handi+) R217P62 R218P12 35 LT-2023 F I G U R E S , D I A G R A M S A N D PA R T S L I S T Pneumatic Diagram LT-2023 36 F I G U R E S , D I A G R A M S A N D PA R T S L I S T Wiring Diagram 37 LT-2023 O R D E R I N G I N F O R M AT I O N A N D PA R T S R E T U R N ORDERING INFORMATION When ordering components, instruction guides, or service manuals the following must be provided: • Unit Catalog Number • Unit Serial Number • Part Number • Quantity Required DeVilbiss Concentrator 5 Liter Series Instruction Guide: • A-525D3 525DS English, Spanish, French • PS-525P3 525PS, 525KS, 525DS English, Spanish, Arabic • SE-525K3 525DS, 525KS, 525PS English, Spanish, French, German, Italian, Dutch, Portuguese, Polish, Turkish DeVilbiss Concentrator 5 Liter Series Service Manual: • LT-2023 525DS, 525KS & 525PS English • LT-2023-DE 525DS, 525KS & 525PS German • LT-2023-ES 525DS, 525KS & 525PS Spanish • LT-2023-FR 525DS, 525KS & 525PS French • LT-2023-PT 525DS, 525KS & 525PS Portuguese DeVilbiss Oxygen Concentrator Service Log: • A-1007AllEnglish Orders may be placed by calling: • Customer Service 800-338-1988 • Warranty parts - U.S.A. 800-338-1988 • International Department 814-443-4881 • Europe +49 (0) 621-178-98-0 PARTS RETURN Before returning parts or units to the factory, call the DeVilbiss Healthcare Customer Service Department (800-338-1988) or (814-443-4881) to obtain a return authorization number. Include in the package a note indicating the return authorization number along with your company name, address, phone number, and account number. The return authorization number should also be written on the outside of the package. To expedite your order for warranty or non-warranty parts, the following information should be given to the representative: • Catalog number • Serial number • Hour meter reading for each concentrator • Account number • Company name and address • Description of problem LT-2023 38 S P E C I F I C AT I O N S DEVILBISS 5-LITER SERIES Catalog Number Delivery Rate (Lower delivery rates available for low flow applications) Maximum Recommended Flow(@ nominal outlet pressures of zero and 7 kPa) Outlet Pressure 525DS / 525DS-Q 525KS / 525KS-LT 525PS 0.5 to 5 LPM 0.5 to 5 LPM 0.5 to 5 LPM 5 LPM 5 LPM 5 LPM 8.5±0.5 psig (58.6±3.5 kPa) Outlet Pressure: 8.5±0.5 psig (58.6±3.5 kPa) Outlet Flow: 2.5±0.5 LPM Continuous 8.5±0.5 psig (58.6±3.5 kPa) Operating Voltage Range 97-127 V~, 60 Hz Oxygen Percentage perating Altitude O (tested at 21°C only) 0-1500 M (0-4921 ft) 1-5 LPM=87%-96% 8.5±0.5 psig (58.6±3.5 kPa) Outlet Pressure: 8.5±0.5 psig (58.6±3.5 kPa) Outlet Flow: 2.5±0.5 LPM Continuous 220-230 V~, 50 Hz, 1.55 Amp 230 V~, 60 Hz, 1.9 Amp 187-253 V~, 50 Hz 195-255 V~, 60 Hz 1-5 LPM=87%-96% Across the voltage range: No degradation of performance Tested at nominal voltage only: No degradation of performance Across the voltage range: No degradation of performance Tested at 230V/50Hz only: No degradation of performance Auxiliary Oxygen Port**** Electrical Rating 1500-4000 M (4921-13123 ft) perating Environment Range* O 5°C to 40°C, humidity range of 10% to 95% Power Consumption Weight Sound Level (ISO 8359:1996 from front) Dimensions Pressure Relief Valve Operating System he visible “low oxygen” indicator will T activate at the following level Storage Conditions 115 V~, 60 Hz, 3.3 Amp 220-230 V~, 60 Hz, 1.68 Amp 187-253 V~, 60 Hz 1-5 LPM=87%-96% Across the voltage range: No degradation of performance Tested at 230V/60Hz only: No degradation of performance No degradation in performance across the operating voltage range. No degradation in performance across the No degradation in performance across the operating voltage range. operating voltage range. 230V / 50 Hz - 312 Watts Average 230V / 50 Hz - 296 Watts Average @ 230V / 60 Hz - 334 Watts Average 310 Watts Average 1.2 LPM & below 230V / 60 Hz - 297 Watts Average @ 275 Watts @ 1.2 LPM & below 230V / 60 Hz - 387 Watts Average 2.5 LPM & below 230V / 60 Hz - 369 Watts Average @ 1.2 LPM & below 36 lbs. (16.3 Kilograms) 36 lbs. (16.3 Kilograms) 36 lbs. (16.3 Kilograms) 48 dBA (525DS) <40 dBA (50Hz) (525KS) 40 dBA (50Hz) 46 dBA (525DS-Q) 48 dBA (525KS-LT) 24.5”H x 13.5”W x 12”D 24.5”H x 13.5”W x 12”D 24.5”H x 13.5”W x 12”D (62.2 x 34.2 x 30.4 cm) (62.2 x 34.2 x 30.4 cm) (62.2 x 34.2 x 30.4 cm) 45 psig±5 psig (310 kPa±34.5 kPa) 45 psig±5 psig (310 kPa±34.5 kPa) 45 psig±5 psig (310 kPa±34.5 kPa) Time Cycle / Pressure Swing Time Cycle / Pressure Swing Time Cycle / Pressure Swing 3.6 sec. at 5 LPM 3.6 sec. at 5 LPM 3.6 sec. at 5 LPM 84% ± 2% 84% ± 2% 84% ± 2% (The audible alarm will alert at approximately (The audible alarm will alert at approximately (The audible alarm will alert at approximately 75%. At less than 60%, the red “service 75%. At less than 60%, the red “service 75%. At less than 60%, the red “service required” light will activate.) required” light will activate.) required” light will activate.) -40°C to 70°C, humidity range of 10% to -40°C to 70°C, humidity range of 10% to -40°C to 70°C, humidity range of 10% to 100%, including condensation 100%, including condensation 100%, including condensation Equipment Class and Type Class II Equipment Double Insulated; Type B Applied Part Ordinary Equipment, IPX0 Approval Body and Safety Standard CSA CAN/CSA-C22.2 No. 601.1-M90 EMC Compliance To** Typical Accumulator Pressure at Maximum Recommended Flow*** 0-457 M (0-1,500 Ft) 457-914 M (1,500 - 3,000 Ft) 914-1524 M (3,000-5,000 Ft) N/A Class II Equipment Double Insulated; Type B Applied Part Ordinary Equipment, IPX0 TUV approved for 50 Hz only to EN 60601+A1+A2 ISO8359: 1996/Amd.1:2012 EN60601-1-2 EN60601-1-2 23-32 psig (159-221 kPa) 21-29 psig (145-200 kPa) 20-27 psig (138-186 kPa) 23-32 psig (159-221 kPa) 21-29 psig (145-200 kPa) 20-27 psig (138-186 kPa) Class II Equipment Double Insulated; Type B Applied Part Ordinary Equipment, IPX0 TUV EN 60601+A1+A2 ISO8359: 1996/Amd.1:2012 EN60601-1-2 23-32 psig (159-221 kPa) 21-29 psig (145-200 kPa) 20-27 psig (138-186 kPa) *NOTE: The OSD performance at 5°C to 40°C, 95% R.H. through voltage range on the 525DS verified at 670m. **NOTE: This oxygen concentrator has been tested to and meets the EMC requirements of EN 60601-1-2. Do not place the concentrator near other equipment or devices that create or attract electromagnetic fields. Examples of such equipment are defibrillators, diathermy equipment, CB radios, microwave ovens, etc. ***NOTE–The minimum and maximum pressures are peak accumulator pressures at 5 LPM. This includes a 4-5 psig drop during pressure cycle. ****NOTE-The maximum recommended flow is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port. NOTE–Turn-down mode may cease at 0.5 LPM if a low output flow meter is not being used. Specifications subject to change without notice 39 LT-2023 D E V I L B I S S G U I D A N C E A N D M A N U F A C T U R E R ’ S D E C L A R AT I O N WARNING Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying documents. Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used. NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing other Equipment and Systems or non-medical electrical equipment. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment – Guidance RF Emissions CISPR 11 Group 1 This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF Emissions CISPR 11 Class B Harmonic emissions IEC 61000-3-2 Class B Voltage fluctuations / flicker emissions Complies This device is suitable for use in all establishments including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment. Compliance Level Immunity Test IEC 60601 Test Level Electrostatic discharge (ESD) IEC 61000-4-2 ±6kV contact ±8kV air Complies Radiated RF IEC 61000-4-3 3 V/m 80MHz to 2.5GHz Complies Conducted RF IEC 61000-4-6 3 Vrms 150kHz to 80MHz Complies Electrical fast transient IEC 61000-4-4 Surge IEC 61000-4-5 ±2kV power line ±1kV I/O lines ±1kV differential ±2kV common Power frequency magnetic field IEC 61000-4-8 3 A/m Voltage dips, short interrupts and voltage variations on power supply input lines IEC 61000-4-11 >95% dip 0.5 cycle 60% dip 5 cycles 70% dip 25 cycles 95% dip 5 secs. LT-2023 Electromagnetic Environment - Guidance Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30% Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3 V/m. Interference may occur in the vicinity of equipment marked with the following symbol: Complies Mains power quality should be that of a typical commercial or hospital environment. Complies Complies Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Complies Mains power quality should be that of a typical commercial or hospital environment. If the user of this device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or battery. 40 DeVilbiss Healthcare LLC 100 DeVilbiss Drive Somerset, PA 15501-2125 USA 800-338-1988 • 814-443-4881 DeVilbiss Healthcare Ltd Unit 3, Bloomfield Park Bloomfield Road Tipton, West Midlands DY4 9AP UNITED KINGDOM +44 (0) 121 521 3140 DeVilbiss Healthcare Pty. Limited 15 Carrington Road, Unit 8 Castle Hill NSW 2154 AUSTRALIA +61-2-9899-3144 EC REP DeVilbiss Healthcare SAS 13/17, Rue Joseph Priestley 37100 Tours FRANCE +33 (0) 2 47 42 99 42 0044 DeVilbiss Healthcare GmbH Kamenzer Straße 3 68309 Mannheim GERMANY +49 (0) 621-178-98-0 D e Vi l b i ss H eal t hcare LLC • 100 De Vilb iss Dr ive • S o me r set, PA 15 5 0 1 • USA 800-338-1988 • 814-44 3 - 4 8 8 1 • www.DeVilbissHea lthcar e .co m DeVilbiss® and OSD® are registered trademarks of DeVilbiss Healthcare. Teflon® is a registered trademark of DuPont. Snoop® is a registered trademark of SWAGELOK © 2017 DeVilbiss Healthcare LLC. 05.17 All Rights Reserved. LT-2023 Rev. F ">
Download
Advertisement
Key features
Provides concentrated oxygen for therapy
Pressure swing adsorption system
Patient alert system
Audible and visible alerts
Easy to use
Portable and lightweight
Quiet operation
Durable construction
Frequently asked questions
When the concentrator is turned “On,” the rotary valve is quickly cycled several times to relieve residual bed pressure preventing a static condition in the compressor. The PC board applies a short DC voltage signal to the valve. The valve will stop for several seconds causing the right bed to pressurize first while the left bed depressurizes. Then, the valve will stop for approximately a second. During this time the sieve bed pressures are equalized. After that, the left bed will pressurize while the right bed depressurizes.
Remove the air filter located in the door on the back of the unit and rinse it thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling.
If the low oxygen light comes on, it means that the oxygen purity is low and you should contact your homecare provider for service.