SKC - 224 PCXR4 Low Flow Pump

SKC - 224 PCXR4 Low Flow Pump
Operating Instructions
Universal Sample Pump
Catalog No. 224-PCXR4
SKC Inc.
863 Valley View Road
Eighty Four, PA 15330 USA
Form #37712 Rev 9904
Table of Contents
Description ............................................................................................. 1
Specifications ......................................................................................... 2
Operation ............................................................................................... 4
High Flow Applications ............................................................. 4
Low Flow Applications .............................................................. 5
Preventive Maintenance ........................................................................ 7
Battery Pack .............................................................................. 7
Pump Inlet Filter ........................................................................ 7
Diagrams/Part Description ...................................................................... 9
Fig. 1 - Model 224-PCXR4 ......................................................... 10
Fig. 2 - Adjustable Flow Holder .............................................. 11
Fig. 3 - Replacement Parts ......................................................... 12
Optional Accessories .............................................................................. 13
Service Policy ...................................................................................... 14
Warranty .............................................................................................. 15
Description
The result of extensive research and development, the PCXR4 is a constant flow air
sampler suited for a broad range of applications. It is ideal for industrial hygiene
studies as well as environmental testing.
Specifically designed for “on worker” and “fenceline” applications, the PCXR4 is
typically used with collecting devices such as filters, impingers, sorbent sample tubes
and sample bags.
Durable RFI-Shielded Case
Low Flow Regulator
provides protection from
radio frequency interference
between 27 and 1000 MHz.
allows pump to be switched
from high to low flow.
Rechargeable NiCad Battery
External Exhaust Port
provides continuous 8-hour
operation on a single charge.
for bag sampling.
Digital LCD Display
Built-in Fluid/Particulate Trap
shows run time,
battery check and
fault indications.
in see-through housing
protects pump.
AIRCHEK
SAMPLER
UL
SAMPLE PERIOD — MINUTES
START
HOLD
FLOW
AND
BATTERY
CHECK
LISTED 124U
®
INTRINSICALLY SAFE
PORTABLE AIR SAMPLING PUMP
FOR USE IN HAZARDOUS LOCA
TIONS CLASS I, GROUPS A B C D
AND CLASS II, GROUPS E F G
AND CLASS III, TEMPERATURE
CODE T3C.
WARNING - SUBSTITUTION OF
COMPONENTS MAY IMPAIR INTRIN
SIC SAFETY. USE ONLY UL LISTED
PORTABLE AIR SAMPLING PUMP
BATTERY PACK MODEL P21661
FLOW
4
3
AIRCHEK SAMPLER
MODEL 224-PCXR4
ON
5
SERIAL NO.
SKC INC.
EIGHTY FOUR
PA
2
1
15330
ADJ
Anti-tamper Cover
Built-in Rotameter
prevents inadvertant
changes to settings.
provides a visible check of flow rate
during sampling, from 0 to 5 LPM.
Accessory Mounting Screws
Touch Keypad
allow sampling accessories such as
impinger holders to be secured to pump.
Recessed Flow Adjustment
adjusts flow rate between
750 and 5000 ml/min.
1
Specifications
Operating Range:
5-5000 ml/min
(5-500 ml/min requires adjustable low flow holder)
Weight:
34 oz (964 gm)
Dimensions:
1.9 x 4.7 x 5.1 inches; 46.5 cubic inches
(4.9 x 11.9 x 13 cm, 758 cubic cm)
Compensation
Range:
750 to 2500 ml/min—to 40 inches water back pressure
2500 to 4000 ml/min—to 20 in water back pressure
Flow Control:
±5% set point constant flow
Run Time:
8 hrs min at 4000 ml/min & 20 in water back pressure
Flow Indicator:
Built-in flow indicator with 250 ml division;
scale marked at 1, 2, 3, 4, & 5 LPM
Battery
Assembly:
Plug in battery pack, rechargeable NiCad 2.0 Ah,
6.0 V UL Listed.
Intrinsically
Safe:
UL Listed for: Class I, Groups A, B, C, D;
Class II, Groups E, F, G; and Class III.
Temp Code T3C.
Operating Temp: -20 C to +45 C (-4 F to +113 F)
Storage Temp:
-40 C to +45 C (-40 F to +113 F)
Charging Temp: +5 C to +45 C (+41 F to +113 F)
Operating
Humidity:
0 to 95% Relative
Multiple
Sampling:
Built-in constant pressure regulator allows user to take
up to four simultaneous samples at different flow rates up to
500 ml/min (maximum total combined flow
1350 ml/min) using optional low flow control.
RFI-Shielding
Performance:
Complies with requirements of EN 55022, FCC Part 15
Class B, EN 50082-1, Frequency range of the radiated
susceptibility test was 27 MHz to 1000 MHz.
CE approved.
2
Specifications (cont)
Flow Fault:
Fault shutdown with LCD indicator and time display
retention if flow is restricted.
Battery Test:
LCD indicator verifies battery condition prior to
sampling.
Time Display:
LCD indicator displays sampler run time in minutes.
Timing Accuracy: ±0.05%
Sampling Pause: Allows user to temporarily halt sampling without loss of
timing data. Restart does not require resetting time.
3
Operation
High Flow Applications (750-5000 ml/min)
Setup
Fully charge the battery by connecting the charger plug to the sampler charging jack
(Figure 1, #18). Use only an SKC charger designated for this model. CAUTION: DO
NOT CHARGE IN A HAZARDOUS ENVIRONMENT. Using flexible tubing,
connect the sampling media to the pump intake (Figure 1, #9). Make sure the pump is
set for high flow. (See “Return to High Flow” p. 4).
Setting the Flow Rate
Test the battery pack by turning the sampler on using the ON/OFF switch (Figure 1,
#4). Press the START/HOLD key then the FLOW AND BATTERY CHECK key.
Adjust the flow to 2 L/min using the FLOW ADJUST SCREW (Figure 1, #7). The LCD
should indicate “BATT OK” in the upper left corner, if not, recharge the battery. Press
the FLOW AND BATTERY CHECK key to place the pump in “HOLD” mode.
Connect a flowmeter to the intake of the sampling media using flexible tubing. [For
pressure applications, insert the exhaust port fitting into the exhaust port (Figure 1,
#15) and connect the sample bag to this fitting.] Press the FLOW AND BATTERY
CHECK key to start the pump and set the flow rate using the FLOW ADJUST SCREW
(Figure 1, #7). When the flow rate is set, press the FLOW AND BATTERY CHECK key
to place the pump in “HOLD”.
Caution: When using impingers, place an in-line trap between the pump and the
impinger to protect the sampler from liquid or vapors. FAILURE TO USE THE
IMPINGER TRAP VOIDS THE WARRANTY. The impinger and trap may be
mounted to the sampler using the accessory mounting screws (Figure 1, #8).
Sampling
For personal sampling, clip the sample collection media to the worker in the breathing zone. While the LCD shows “HOLD,” start the test cycle by pressing the START/
HOLD key at the beginning of the desired sampling period. “SAMPLE RUNNING”
will display. Record the start time. The time display will automatically track the
sampling period time elapsed.
User options during sampling:
Pause - pause (shutdown) by pressing the START/HOLD key. All timing data will
freeze. To resume sampling press the START/HOLD key, timing data will resume.
Fault shutdown - during restricted flow or low battery conditions the sampler will
shut down. “HOLD” will display on the LCD and timing functions will pause. “LO
BATT” or “FLOW FAULT” will display on the LCD depending on the cause of the
shutdown. To restart a pump in “FLOW FAULT,” correct the flow blockage and press
the START/HOLD key. A pump displaying “LO BATT” must be recharged before
sampling.
4
Low Flow Applications (5-500 ml/min)
Setup
Fully charge the battery by connecting the charger plug to the sampler charging jack
(Figure 1, #18). Use only an SKC charger designated for this model. CAUTION: DO
NOT CHARGE IN A HAZARDOUS ENVIRONMENT.
Test the battery pack by turning the sampler on using the ON/OFF switch (Figure 1, #4). Press
the START/HOLD key then the FLOW AND BATTERY CHECK key and adjust the flow to 1.5
L/min using the FLOW ADJUST SCREW (Figure 1, #7). If performing multiple sampling using
an adjustable flow tube holder (dual, tri, or quad), the flow rate of the pump must be greater
than the sum of the flow rates through the tubes; the flow rate through any one tube cannot
exceed 500 ml/min. The LCD should indicate “BATT OK” in the upper left corner. If not,
recharge the battery. Press the FLOW AND BATTERY CHECK key to place the pump in
“HOLD”.
Remove the screw cap (Figure 1, # 14) covering the regulator isolation valve. Turn the
exposed screw 4-5 turns counterclockwise. Replace the screw cap. The pump is now
set for low flow. Connect an adjustable low flow holder (Figure 2) to the pump intake
(Figure 1, #9) using flexible tubing. Insert an opened sorbent tube into the rubber sleeve of
the low flow holder so the arrow on the tube points toward the holder.
Caution! Long duration color detector tubes require a special tube cover which
accommodates an in-line trap tube. The trap tube protects the pump from caustic
fumes which are often released from detector tubes. FAILURE TO USE THE TRAP
TUBE VOIDS THE WARRANTY.
Setting the Flow Rate
Connect a flowmeter to the exposed end of the sorbent tube. Loosen the screw on the low
flow holder, for Tri and Quad models first rotate the anti-tamper cover (Figure 2, #1) to
expose the brass screw(s) (Figure 2, #2). Activate the pump by pressing the FLOW AND
BATTERY CHECK key and adjust the flow rate by turning the brass screw until the
flowmeter indicates the desired flow. Do not adjust the flow on the pump. Adjust the flow
only by using the brass screw (Figure 2, #2) on the low flow holder.
When the flow rate is set, place the pump in “HOLD” by pressing the FLOW AND
BATTERY CHECK key and disconnect the flowmeter. Replace the sorbent tube used
for setting the flow with a new sorbent tube for sample collection. Place the appropriate size tube cover (Figure 2, #5) over the tube, and screw it in place on the low flow
holder.
5
Sampling
For personal sampling, clip the sample collection media to the worker in the breathing zone. While the LCD shows “HOLD,” start the test cycle by pressing the START/
HOLD key at the beginning of the desired sampling period. “SAMPLE RUNNING”
will display. Record the start time. The time display will automatically track the
sampling period time elapsed.
User options during sampling:
Pause - pause (shutdown) by pressing the START/HOLD key. All timing data will
freeze. To resume sampling press the START/HOLD key, timing data will resume.
Fault shutdown - during restricted flow or low battery conditions the sampler will
shut down. “HOLD” will display on the LCD and timing functions will pause. “LO
BATT” or “FLOW FAULT” will display on the LCD depending on the cause of the
shutdown. To restart a pump in “FLOW FAULT,” correct the flow blockage and press
the START/HOLD key. A pump displaying “LO BATT” must be recharged before
sampling.
Return to High Flow
To return to High Flow, remove the screw cap (Figure 1, # 14) covering the regulator
isolation valve. Turn the exposed screw clockwise until it stops. (Do not over-tighten.)
Replace the screw cap. The pump is now set for high flow.
Bag Sampling by Positive Pressure
Using flexible tubing, connect the sampling media to the pump intake (Figure 1,#9).
[For sample bags using positive pressure filling, insert the exhaust fitting into the
exhaust port (Figure 1, #15). After setting the flow rate, you will connect the sample
bag to this fitting instead.]
6
Preventive Maintenance
Battery Pack Maintenance
Removal—Remove the two screws (Figure 1, #16) which secure the battery pack
(Figure 1, #17) and loosen the four case screws above and below the belt clip.
Carefully slide the battery pack out from under the belt clip (Figure 1, #19) being
careful to keep it straight.
Replacement—Slip the front edge of the battery pack (Figure 1, #17) under the belt
clip (Figure 1, #19) and rotate the battery pack so the rails engage the slots on the case
front. Push the battery pack until it is properly located. Reinstall battery screws and
tighten the case screws (Figure 1, #16).
Charge Maintenance
For proper maintenance of battery packs, SKC produces an optional cycling charger
(Catalog No. 223-426) which discharges and recharges the battery automatically to
protect against memory effects.
Rotate the use of any spare pack to avoid idle periods in excess of one month. Fully
charge packs before or after use or storage.
SKC UL listed battery packs (SKC Catalog No. P21661) contain a protective device to
eliminate potential short circuiting while the pump is in use. If the indicator light on
the charger will not light while charging, either the battery pack, charger, or wall
outlet is inoperative. If you are unable to determine which is inoperative, please
contact SKC Technical Support at 724-941-9701 or e-mail [email protected]
Caution: Do not charge in a hazardous environment.
Warning: Using a non-approved charger voids the SKC warranty.
Warning: Tampering with the battery pack voids the SKC warranty
and the UL Intrinsic Safety listing.
Pump Inlet Filter
The SKC sampler is fitted with a filter/trap inside the clear plastic intake port
housing. This prevents particulates from being drawn into the pump mechanism.
Occasionally, the filter should be visually checked to assure that it does not become
clogged. If maintenance is necessary:
1. Clean all dust and debris from around the filter housing.
2. Remove the four screws (Figure 1, #10) and the front filter housing.
3. Remove and discard the filter membrane (Figure 1, #12) and O-ring (Figure 1, #11).
4. Clean the filter housing.
7
5. Insert a new filter membrane and o-ring.
(Filter Replacement Kit, SKC Catalog No. P22409)
6. Reattach the front filter housing and cross-tighten the four screws.
Pump Service
Pumps under warranty should be sent to SKC Inc. for servicing (see Service p. 14). For
further information on pump maintenance, testing and replacing pump components,
and troubleshooting, request the Universal Pump Service Manual (SKC Publication
No. 1377).
Notice: This operating instruction may not address all safety concerns (if any) associated with this product and its
use. The user is responsible for determining and following the appropriate safety and health practices and
regulatory limitations (if any) before using the product. The information contained in this document should not be
construed as legal advice, opinion, or as a final authority on legal or regulatory procedures.
8
Diagrams/Part Description for Figure 1
Model 224-PCXR4
No. Description
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
A
B
LCD: Indicators for all sampler functions.
FLOW AND BATTERY CHECK Key: Allows setting flow rate and testing
battery condition.
START/HOLD Key: Used when ready to begin the sampling cycle, pause the
sampling cycle and restart the cycle after pause.
ON/OFF Switch: Allows the pump to be shut down completely, clears time
display.
Anti-tamper Cover: Protects controls from incidental contact or tampering.
Cover Screw: Fastens anti-tamper cover.
Flow Adjustment Control: Adjusts flow from 750-5000 ml/min.
Accessory Mounting Screws (2): Secure accessories such as impinger and trap
holders.
Filter Housing (intake): Air intake port and trap.
Filter Housing Screws (4): Secure filter housing.
Filter O-ring: Leak seal for filter in housing.
Filter (10 micron nylon): Filters particulates before entering pump.
Built-in Flowmeter: Monitors flow changes.
Regulator Isolation Cap: Accesses regulator isolation valve.
Exhaust Port Cap: Accesses exhaust port.
Battery Pack Screws (2): Secures pack to pump.
Battery Pack Assembly: Provides power to pump.
Charging Jack: Connector for battery charger.
Belt Clip: Secures pump to worker.
Compensation Pot A: Adjusts pump compensation which is factory set. Access
screw guards against accidental contact or tampering.
Compensation Pot B: Adjusts pump compensation which is factory set. Access
screw guards against accidental contact or tampering.
9
Figure 1
8
9
AIRCHEK
SAMPLER
1
10
U® L
2
LISTED 124U
SAMPLE PERIOD — MINUTES
3
FLOW
AND
BATTERY
CHECK
START
HOLD
5
11
4
12
10
3
AIRCHEK SAMPLER
MODEL 224-PCXR4
A
2
1
13
B
4
ON
FLOW
ADJ
5
6
7
FRONT
16
17
18
16
19
14
TOP
9
15
10
BACK
Figure 2
2
;;;
;;;
;;;
2
1
y
@
€
À
;
@
€
À
;
y
@
€
À
;
y
3
4
➁
➂
5
Quad Adjustable Low Flow
Holder
Single Adjustable Low Flow Holder
Figure 2 - Adjustable Low Flow Holder
1. Anti-tamper Cover (tri and quad only)
2. Manifold Flow Adjustment
3. Rubber Sleeve
4. Sorbent Sample Tube
5. Protective Cover (not included)
11
Figure 3 — Replacement Parts for 224-PCXR4
P22433N
P22433R
P22433T
For further information on
testing and replacing pump
components, request the
Universal Pump Service
Manual (SKC Publication
No. 1377).
P21411
P21661
P22433RS2
P22417BC
P22417J
Part Description for Figure 3
Replacement Parts
P22409*
P22433R
P22417C*
P22417D
P22417E
P22417F
P22417G
P22417H
P22417J
P22417K
Replacement Filter Kit
Cap Screws (set of 2)
Exhaust Port Fitting
Filter Housing Assembly
Bottom Plate Assembly
Valve Plate Assembly
Pump Body
Diaphragm/Yoke Assembly
Regulator Assembly
Pulsation Dampener
Assembly (2)
P22417M Motor/Eccentric Assembly
P21661
Battery Pack Assembly
P21411
Case Parts
(excluding Battery Case)
P22433L
Flowmeter Assembly
P22433T
Control Board
P22433RS2 Replacement Stack
P22417BC Belt Clip with Screws
P22433N
Keyboard Assembly
P22433C* Anti-tamper Cover
P22433ES* External Screw Pack
P22417K
P22433L
P22417D
P22417F
P22417G
P22417H
P22417M
P22417E
Exploded view of stack # P22433RS2
* Not shown in Figure 1 or 2
12
Optional Accessories
Adjustable Low Flow Holders:
224-26-01 Single Holder
224-26-02 Dual Holder
224-26-03 Tri Holder
224-26-04 Quad Holder
Protective Sample Tube Covers:
for tubes up to:
224-29A 70 mm long
224-29B 110 mm long
224-29C 150 mm long
224-29D 220 mm long
224-29T 115 mm with tandem trap tube cover
Battery Chargers:
223-226 Single Battery Charger 115 V
223-227 Single Battery Charger 230 V
223-426 Deluxe 5 Station Battery Charger, Switchable for 115 or 230 V operation
Miscellaneous:
224-11
Sampler Tool Kit
224-95
Protective Nylon Pouch with belt and shoulder strap, brown
224-95A Protective Nylon Pouch with belt and shoulder strap, red
13
Service
Product to be serviced should be sent, freight prepaid, to:
SKC Inc.
National Service Center
863 Valley View Road
Eighty Four, PA 15330
Care should be taken in packaging to prevent damage in transit. Please include a contact name and phone
number, shipping address, and a brief description of the problem. For nonwarranty repairs, a purchase
order number and billing address is also required. The Service Center will contact nonwarranty
customers with an estimate before proceeding with repairs.
SKC QualityCare
QualityCare is a cost-effective preventive maintenance program that assures that pumps are tested,
repaired, and calibrated on an annual basis. Participants will receive certificates of compliance for each
pump, each year, to demonstrate adherence to Occupational Health and Safety Management Systems or
company quality programs.
For more information on QualityCare call our SKC Customer Service Team at 724 941-9701.
Note: SKC Inc. will accept for repair any SKC product which is not contaminated with hazardous materials.
Products determined to be contaminated will be returned unserviced.
Universal Pump Service Manual
Customers who wish to self-service their out-of-warranty pumps should request the Universal Pump
Service Manual (SKC Publication No. 1377).
14
SKC INC. LIMITED ONE YEAR WARRANTY
1.
SKC warrants that its instruments provided for industrial hygiene, air pollution, gas analysis, and safety and
health applications are free from defects in workmanship and materials under normal use for a period of one (1) year.
This warranty does not cover claims due to abuse, misuse, neglect, alteration, or accident, or use in application for
which the instrument was either not designed or not approved by SKC Inc., or due to the buyer’s failure to maintain normal
maintenance, improper selection or misapplication. This warranty shall further be void if changes or adjustments to the
instrument are made by other than an employee of the seller, or if the operating instructions furnished at the time of
installation are not complied with.
2.
SKC Inc. hereby expressly disclaims all warranties either expressed or implied, including any implied
warranties of merchantability or fitness for a particular purpose, and neither assumes nor authorizes any other
person to assume for it any liability in connection with the sale of these instruments. No description of the goods
being sold has been made a part of the basis of the bargain or has created or amounted to an express warranty that the
goods will conform to any such description. Buyer shall not be entitled to recover from SKC Inc. any consequential
damages, damages to property, damages for loss of use, loss of time, loss of profits, or income or other incidental
damages. Nor shall buyer be entitled to recover from SKC Inc. any consequential damages resulting from defect of
the instrument including, but not limited to, any recovery under section 402A of the Restatement, Second of Torts.
3.
This warranty extends only to the original purchaser of the warranted instrument during the term of the
warranty.
4.
Completion and return by the buyer of the instrument registration card is a condition precedent to warranty
coverage and performance. In addition, the buyer may be required to present proof of purchase in the form of a paid
receipt for the instrument.
5.
This warranty covers the instrument purchased and each of its component parts.
6.
In the event of a defect, malfunction, or other failure of the instrument not caused by any misuse or damage to the
instrument while in possession of the buyer, SKC Inc. will remedy the failure or defect, without charge to the buyer. The
remedy will consist of service or replacement of the instrument, or refund of the purchase price, at the option of SKC Inc.
However, SKC Inc. will not elect refund unless it is unable to provide replacement, and repair is not commercially
practicable.
7.
The term of this warranty begins on the date the instrument is delivered to the buyer, and continues for a period
of one (1) year,
8.
(a) To obtain performance of any obligation under this warranty, the buyer shall return the instrument, freight
prepaid, to SKC Inc., at the following address:
SKC Inc., National Service Center, 863 Valley View Road, Eighty Four, PA 15330 USA
(b) To obtain further information on the warranty performance you may telephone (724) 941-9701 at the address
above.
This warranty shall be construed under the laws of the Commonwealth of Pennsylvania which shall be deemed to be
9.
the situs of the contract for purchase of SKC Inc. instruments.