Keithley 2510, -AT SourceMeter Service Manual
Below you will find brief information for SourceMeter 2510, SourceMeter 2510-AT. The SourceMeter 2510 is a versatile, high-performance source measure unit. It can be used to test a wide variety of electronic components and circuits. This manual will help you troubleshoot the unit and replace parts if necessary.
PDF
Download
Document
Advertisement
Advertisement
Models 2510 and 2510-AT TEC SourceMeter® Service Manual A GREATER MEASURE OF CONFIDENCE WARRANTY Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment. Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation. During the warranty period, we will, at our option, either repair or replace any product that proves to be defective. To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days. LIMITATION OF WARRANTY This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY. Keithley Instruments, Inc. 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 1-888-KEITHLEY (534-8453) • www.keithley.com Sales Offices: Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02-363 00 40 • Fax: 02/363 00 64 Yuan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-6202-2886 • Fax: 8610-6202-2892 Tietäjäntie 2 • 02130 Espoo • Phone: 09-54 75 08 10 • Fax: 09-25 10 51 00 3, allée des Garays • 91127 Palaiseau Cédex • 01-64 53 20 20 • Fax: 01-60 11 77 26 Landsberger Strasse 65 • 82110 Germering • 089/84 93 07-40 • Fax: 089/84 93 07-34 Unit 2 Commerce Park, Brunel Road • Theale • Berkshire RG7 4AB • 0118 929 7500 • Fax: 0118 929 7519 Flat 2B, Willocrissa • 14, Rest House Crescent • Bangalore 560 001 • 91-80-509-1320/21 • Fax: 91-80-509-1322 Viale San Gimignano, 38 • 20146 Milano • 02-48 39 16 01 • Fax: 02-48 30 22 74 New Pier Takeshiba North Tower 13F • 11-1, Kaigan 1-chome • Minato-ku, Tokyo 105-0022 • 81-3-5733-7555 • Fax: 81-3-5733-7556 FL., URI Building • 2-14 Yangjae-Dong • Seocho-Gu, Seoul 137-130 • 82-2-574-7778 • Fax: 82-2-574-7838 Postbus 559 • 4200 AN Gorinchem • 0183-635333 • Fax: 0183-630821 c/o Regus Business Centre • Frosundaviks Allé 15, 4tr • 169 70 Solna • 08-509 04 679 • Fax: 08-655 26 10 Kriesbachstrasse 4 • 8600 Dübendorf • 01-821 94 44 • Fax: 01-820 30 81 1FL., 85 Po Ai Street • Hsinchu, Taiwan, R.O.C. • 886-3-572-9077• Fax: 886-3-572-9031 BELGIUM: CHINA: FINLAND: FRANCE: GERMANY: GREAT BRITAIN: INDIA: ITALY: JAPAN: KOREA: NETHERLANDS: SWEDEN: SWITZERLAND: TAIWAN: 2/02 Models 2510 and 2510-AT TEC SourceMeter® Service Manual All references in this manual to the Model 2510 also apply to the Model 2510-AT unless otherwise specified. ©2000, Keithley Instruments, Inc. All rights reserved. Cleveland, Ohio, U.S.A. Fourth Printing, February 2002 Document Number: 2510-902-01 Rev. D Manual Print History The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page. Revision A (Document Number 2510-902-01) ................................................................ April 2000 Revision B (Document Number 2510-902-01) ................................................................. June 2000 Revision C (Document Number 2510-902-01) ........................................................ December 2000 Revision D (Document Number 2510-902-01) .......................................................... February 2002 All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders. Safety Precautions The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present. This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications. If the product is used in a manner not specified, the protection provided by the product may be impaired. The types of product users are: Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained. Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits. Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel. Service personnel are trained to work on live circuits, and perform safe installations and repairs of products. Only properly trained service personnel may perform installation and service procedures. Keithley products are designed for use with electrical signals that are rated Installation Category I and Installation Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O signals are Installation Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Installation Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the Manual. Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring. Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be exposed. Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card. Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use. When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator. For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting ca2/02 bles or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers. Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured. The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired. Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card. When fuses are used in a product, replace with same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid interlock. If a The screw is present, connect it to safety earth ground using the wire recommended in the user documentation. ! symbol on an instrument indicates that the user should refer to the operating instructions located in the manual. The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages. The WARNING heading in a manual explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure. The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty. Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information. To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing. Table of Contents 1 Performance Verification Introduction ................................................................................ 1-2 Verification test requirements ..................................................... 1-2 Environmental conditions ................................................... 1-2 Warm-up period .................................................................. 1-2 Line power .......................................................................... 1-3 Recommended test equipment ................................................... 1-3 Resistor characterization ..................................................... 1-4 Verification limits ....................................................................... 1-4 Example limits calculation .................................................. 1-4 Resistance limits calculation ............................................... 1-4 Restoring factory defaults .......................................................... 1-5 Performing the verification test procedures ............................... 1-5 Test summary ...................................................................... 1-5 Test considerations .............................................................. 1-6 Adjusting setpoints .............................................................. 1-6 Voltage accuracy ........................................................................ 1-6 Output voltage accuracy ...................................................... 1-6 Voltage readback accuracy .................................................. 1-7 Voltage limit accuracy ......................................................... 1-8 Current accuracy ........................................................................ 1-8 Output current accuracy ...................................................... 1-8 Current readback accuracy .................................................. 1-9 Current limit accuracy ....................................................... 1-10 AC resistance accuracy ............................................................ 1-10 Sensor measurement accuracy ................................................. 1-11 2 Calibration Introduction ................................................................................ Environmental conditions .......................................................... Temperature and relative humidity ..................................... Warm-up period .................................................................. Line power .......................................................................... Calibration considerations .......................................................... Calibration cycle ................................................................. Recommended calibration equipment ................................. Resistor characterization ..................................................... Calibration menu ........................................................................ Unlocking calibration ................................................................. Unlocking calibration from the front panel ........................ Unlocking calibration by remote ........................................ 2-2 2-2 2-2 2-2 2-2 2-3 2-3 2-3 2-4 2-5 2-5 2-5 2-6 Changing the password ............................................................... 2-6 Changing the password from the front panel ...................... 2-6 Changing the password by remote ...................................... 2-6 Resetting the calibration password ............................................. 2-7 Viewing calibration dates and calibration count ......................... 2-7 Calibration errors ........................................................................ 2-7 Front panel error reporting .................................................. 2-7 Remote error reporting ........................................................ 2-7 Aborting calibration steps ........................................................... 2-8 Front panel calibration ................................................................ 2-8 Remote calibration .................................................................... 2-20 Remote calibration command summary ............................ 2-20 Remote calibration procedure ........................................... 2-21 3 Routine Maintenance Introduction ................................................................................ 3-2 Line fuse replacement ................................................................. 3-2 4 Troubleshooting Introduction ................................................................................ 4-2 Safety considerations .................................................................. 4-2 Repair considerations ................................................................. 4-2 Power-on self-test ....................................................................... 4-3 Front panel tests .......................................................................... 4-3 KEYS test ............................................................................ 4-3 DISPLAY PATTERNS test ................................................. 4-4 CHAR SET test ................................................................... 4-4 Principles of operation ................................................................ 4-5 Overall block diagram ......................................................... 4-5 Analog circuits .................................................................... 4-6 Power supply ..................................................................... 4-12 Digital circuitry ................................................................. 4-13 Display board circuits ........................................................ 4-15 Troubleshooting ........................................................................ 4-15 Display board checks ........................................................ 4-15 Power supply checks ......................................................... 4-16 Digital circuitry checks ..................................................... 4-16 Analog circuitry checks ..................................................... 4-17 Internal fuse replacement ......................................................... 4-17 Power supply module fuse replacement ............................ 4-17 Digital I/O +5V supply fuse replacement ......................... 4-18 No comm link error .................................................................. 4-18 5 Disassembly Introduction ................................................................................ Handling and cleaning ............................................................... Handling PC boards ............................................................ Solder repairs ...................................................................... Static sensitive devices ............................................................... Assembly drawings .................................................................... Case cover removal .................................................................... Mother board removal ................................................................ Front panel disassembly ............................................................. Removing power components .................................................... Power supply module removal ............................................ Power module removal ....................................................... Fan removal ......................................................................... Instrument reassembly ............................................................... 6 5-2 5-2 5-2 5-2 5-3 5-3 5-4 5-4 5-5 5-5 5-5 5-6 5-6 5-6 Replaceable Parts Introduction ................................................................................ Parts lists .................................................................................... Ordering information ................................................................. Factory service ........................................................................... Component layouts .................................................................... A Specifications B Calibration Reference 6-2 6-2 6-2 6-2 6-2 Introduction ............................................................................... B-2 Command summary .................................................................. B-2 Miscellaneous commands ......................................................... B-3 SENSe commands ..................................................................... B-6 SOURce commands .................................................................. B-8 DAC commands ........................................................................ B-9 RESistance commands .............................................................. B-9 Detecting calibration errors ..................................................... B-10 Reading the error queue ................................................... B-10 Error summary ................................................................. B-10 Status byte EAV (Error Available) bit .............................. B-10 Generating an SRQ on error ............................................ B-10 Detecting calibration step completion ..................................... B-12 Using the *OPC? query ................................................... B-12 Using the *OPC command ............................................... B-12 Generating an SRQ on calibration complete ................... B-13 C Calibration Program Introduction ............................................................................... Computer hardware requirements ............................................. Software requirements ............................................................... Calibration equipment ............................................................... General program instructions .................................................... Program C-1 Model 2510 calibration program .............. C-2 C-2 C-2 C-2 C-3 C-4 List of Illustrations 1 Performance Verification Figure 1-1 Figure 1-2 Figure 1-3 Figure 1-4 Connections for voltage verification tests .............................. 1-7 Connections for current verification tests .............................. 1-9 Connections for AC resistance verification test ................... 1-11 Connections for sensor resistance accuracy verification ..... 1-12 2 Calibration Figure 2-1 Figure 2-2 Figure 2-3 Figure 2-4 Figure 2-5 Figure 2-6 Voltage calibration connections ............................................. 2-9 Current and current protection calibration connections ....... 2-11 Thermistor and RTD calibration resistor connections ......... 2-12 Shorted input calibration connections .................................. 2-13 I-SS and V-SS calibration resistor connections ................... 2-17 AC ohms calibration connections ........................................ 2-18 3 Routine Maintenance Figure 3-1 Rear panel .............................................................................. 3-2 4 Troubleshooting Figure 4-1 Figure 4-2 Figure 4-3 Figure 4-4 Figure 4-5 Figure 4-6 Figure 4-7 Figure 4-8 Figure 4-9 Figure 4-10 Overall block diagram ........................................................... 4-5 Block diagram of analog circuits ........................................... 4-6 Simplified schematic of class D amplifier ............................. 4-7 Simplified schematic of H-bridge output stage ...................... 4-8 H-bridge switching and step-down regulator ......................... 4-9 Sensor conditioning for 10kΩ thermistor ............................ 4-10 PID hardware control loop ................................................... 4-11 Power supply block diagram ................................................ 4-12 Digital circuitry overall block diagram ................................ 4-13 Digital control circuits ......................................................... 4-14 List of Tables 1 Performance Verification Table 1-1 Table 1-2 Table 1-3 Recommended verification equipment ................................... 1-3 Characterized resistor values .................................................. 1-4 Sensor resistance measurement accuracy limits ................... 1-13 2 Calibration Table 2-1 Table 2-2 Table 2-3 Table 2-4 Table 2-5 Table 2-6 Recommended calibration equipment .................................... 2-4 Characterized resistor values .................................................. 2-4 Calibration menu .................................................................... 2-5 Calibration step summary ...................................................... 2-9 Remote calibration command summary ............................... 2-20 Remote calibration step summary ........................................ 2-21 3 Routine Maintenance Table 3-1 Power line fuse ....................................................................... 3-3 4 Troubleshooting Table 4-1 Table 4-2 Table 4-3 Table 4-4 Display board checks ........................................................... 4-15 Power supply checks ............................................................ 4-16 Digital circuitry checks ........................................................ 4-16 Analog circuitry checks ........................................................ 4-17 6 Replaceable Parts Table 6-1 Table 6-2 Table 6-3 Mother board parts list ........................................................... 6-3 Display board parts list ........................................................... 6-9 Miscellaneous parts list ........................................................ 6-10 B Calibration Reference Table B-1 Table B-2 Calibration commands .......................................................... B-2 Calibration errors ................................................................ B-11 1 Performance Verification Models 2510 and 2510-AT Performance Verification 1-2 Introduction Use the procedures in this section to verify that Model 2510 accuracy is within the limits stated in the instrument’s one-year accuracy specifications. You can perform these verification procedures: • • • • When you first receive the instrument to make sure that it was not damaged during shipment. To verify that the unit meets factory specifications. To determine if calibration is required. Following calibration to make sure it was performed properly. WARNING NOTE The information in this section is intended for qualified service personnel only. Do not attempt these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous voltages, which could cause personal injury or death if contacted. Use standard safety precautions when working with hazardous voltages. If the instrument is still under warranty and its performance is outside specified limits, contact your Keithley representative or the factory to determine the correct course of action. Verification test requirements Be sure that you perform the verification tests: • • • • • Under the proper environmental conditions. After the specified warm-up period. Using the correct line voltage. Using the proper test equipment. Using the specified output signals and reading limits. Environmental conditions Conduct your performance verification procedures in a test environment with: • • An ambient temperature of 18-28°C (65-82°F). A relative humidity of less than 70% unless otherwise noted. Warm-up period Allow the Model 2510 to warm up for at least one hour before conducting the verification procedures. 1-3 Performance Verification Models 2510 and 2510-AT If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range. Also, allow the test equipment to warm up for the minimum time specified by the manufacturer. Line power The Model 2510 requires a line voltage of 100 to 240V and a line frequency of 50 or 60Hz. Verification tests must be performed within this range. Recommended test equipment Table 1-1 summarizes recommended verification equipment and pertinent specifications. You can use alternate equipment as long as that equipment has specifications at least as good as those listed in Table 1-1. Keep in mind, however, that test equipment uncertainty will add to the uncertainty of each measurement. Generally, test equipment uncertainty should be at least four times better than corresponding Model 2510 specifications. Table 1-1 Recommended verification equipment Description Manufacturer/Model Specifications Digital Multimeter Keithley 2002 DC Voltage1 20V: ±6.8ppm Resistance1 20Ω: 200Ω: 2kΩ: 20kΩ 200kΩ: ±23ppm ±19ppm ±7.4ppm ±7.4ppm ±29.8ppm Resistors2 Isotec RUG-Z-1R00-0.1 INPUT/OUTPUT Mating Connector3 Keithley CS-846 1Ω, ±0.1%, 100W 100Ω, ±1% 1kΩ, ±1% 10kΩ, ±1% 100kΩ, ±1% 1. Ninety day, full-range accuracy specification of ranges required for various measurement points. 2. Characterize all resistors using 4-wire ohms function of recommended DMM before use. 3. One connector supplied with Model 2510. Models 2510 and 2510-AT Performance Verification 1-4 Resistor characterization The resistors listed in Table 1-1 must be characterized using the 4-wire ohms function of the recommended DMM before use. Be sure to use the lowest resistance range possible for each measurement for best accuracy. Characterized values can be recorded in Table 1-2. Table 1-2 Characterized resistor values Nominal resistance Characterized resistance* 1Ω _________Ω 100Ω _________Ω 1kΩ _________kΩ 10kΩ _________kΩ 100kΩ _________kΩ *Characterize value using DMM (see Table 1-1). Verification limits The verification limits stated in this section have been calculated using only the Model 2510 one-year accuracy specifications, and they do not include test equipment uncertainty. If a particular measurement falls outside the allowable range, recalculate new limits based on Model 2510 specifications and corresponding test equipment specifications. Example limits calculation As an example of how verification limits are calculated, assume you are testing the operating voltage function using a 10V value. Using the Model 2510 operating voltage one-year accuracy specification of ±(0.1% of reading + 4mV offset), the calculated output limits are: Limits = 10V ± [(10V × 0.1%) + 4mV] Limits = 10V ± (0.01 + 0.004) Limits = 10V ± 0.014V Limits = 9.986 to 10.014V Resistance limits calculation When verifying accuracy for tests using the precision resistors, it will be necessary to recalculate resistance limits based on the actual characterized resistance values. You can calculate resistance reading limits in the same manner described above, but be sure to use the actual characterized resistance values and the Model 2510 one-year accuracy specifications for your calculations (see Appendix A). 1-5 Performance Verification Models 2510 and 2510-AT Restoring factory defaults Before performing the verification procedures, restore the instrument to its factory front panel (BENCH) defaults as follows: 1. Press the MENU key. The instrument will display the following prompt: MAIN MENU 2. Select SAVESETUP, and then press ENTER. The unit then displays: SAVESETUP MENU SAVESETUP COMMUNICATION CAL SAVE RESTORE POWERON RESET 3. Select RESET, and then press ENTER. The unit displays: RESET ORIGINAL DFLTS 4. Select BENCH, and then press ENTER to restore BENCH defaults. BENCH GPIB Performing the verification test procedures Test summary • • • • Voltage accuracy Current accuracy AC resistance accuracy Sensor measurement accuracy If the Model 2510 is not within specifications and not under warranty, see the calibration procedures in Section 2 for information on calibrating the unit. Models 2510 and 2510-AT Performance Verification 1-6 Test considerations WARNING The maximum common-mode voltage (voltage between INPUT/OUTPUT - terminals and chassis ground) is 30V DC. Exceeding this value may cause a shock hazard. CAUTION The maximum voltage between INPUT/OUTPUT sense (S) terminals is 1V. Exceeding this voltage may result in instrument damage. When performing the verification procedures: • • • • • • Be sure to restore factory front panel defaults as previously outlined. Make sure that the test equipment is fully warmed up and properly connected to the Model 2510 INPUT or OUTPUT terminals as required. Be sure that the Model 2510 output is turned on before making measurements. Be sure the test equipment is set up for the proper function and range. Allow the Model 2510 output signal to settle before making a measurement. Do not connect test equipment to the Model 2510 through a scanner, multiplexer, or other switching equipment. Adjusting setpoints Before making many verification measurements, you must properly adjust the setpoint as covered below. 1. 2. 3. First select the function to be tested as indicated in the procedure. Press any one of the four EDIT keys (, , , ) to enter the setpoint edit mode. Using the EDIT keys, adjust the setpoint to the required value. Voltage accuracy Output voltage accuracy Follow the steps below to verify that Model 2510 output voltage accuracy is within specified limits. This test involves setting the output voltage to a specific value and accurately measuring the voltage with a DMM. 1. 2. 3. 4. With the power off, connect the digital multimeter to the Model 2510 OUTPUT terminals, as shown in Figure 1-1. Turn on the Model 2510 and DMM, and allow them to warm up for at least one hour. Select the multimeter DC volts measuring function, and choose the 20V range. Press the Model 2510 V key to select the voltage function, and make sure the source output is turned on. 1-7 Performance Verification 5. 6. 7. 8. Models 2510 and 2510-AT Using the EDIT keys, adjust the Model 2510 voltage setpoint to exactly +9.000V. Verify that the DMM reading is within 8.987 to 9.013V limits. Repeat steps 5 and 6 for a -9.000V output value. Turn off the output when the test is completed. Figure 1-1 Connections for voltage verification tests INPUT S- INPUT F+ INPUT S+ INPUT FWARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. SENSE Ω 4 WIRE INPUT HI INPUT HI 1100V PEAK ! LO DCV ACV DCI ACI Ω2 Ω4 FREQ NEXT RANGE REL TRIG STORE RECALL INFO LOCAL CHAN AUTO FILTER MATH CONFIG MENU F EXIT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) ENABLE-DIG I/O RS-232 TRIGGER LINK LINE FUSE SLOWBLOW R FRONT/REAR 2A 250V RANGE POWER SCAN 500V PEAK INPUTS TEMP DISPLAY INPUT F+ S+ S- F- F+ F- S+ SISOLATION FROM EARTH: 30V MAX. INPUT LO 2002 MULTIMETER PREV OUTPUT ! CAL AMPS ! 2.5A, 250V 120 350V PEAK CAT I LINE RATING 100-240VAC 50, 60 HZ 90VA MAX ENTER CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2002 DMM Model 2510 Voltage readback accuracy Follow the steps below to verify that Model 2510 voltage readback accuracy is within specified limits. This test involves setting the output voltage to a specific value as measured by a DMM and verifying that the Model 2510 voltage readback display reading is within limits. 1. 2. 3. 4. 5. 6. 7. 8. With the power off, connect the digital multimeter to the Model 2510 OUTPUT terminals, as shown in Figure 1-1. Turn on the Model 2510 and DMM, and allow them to warm up for at least one hour. Select the multimeter DC volts measuring function, and choose the 20V range. Press the Model 2510 V key to select the voltage function, and make sure the source output is turned on. Using the EDIT keys, adjust the Model 2510 output voltage as measured by the DMM to +9.000V (or closest possible value). Verify that the Model 2510 readback display (top display line) is within 8.987 to 9.013V limits. Repeat steps 5 and 6 for a -9.000V output value. Turn off the output when the test is completed. Models 2510 and 2510-AT Performance Verification 1-8 Voltage limit accuracy Follow the steps below to verify that Model 2510 voltage limit accuracy is within specified limits. The test involves setting the voltage limit to a specific value and making sure the output voltage is limited to the required value. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. With the power off, connect the digital multimeter to the Model 2510 OUTPUT terminals, as shown in Figure 1-1. Turn on the Model 2510 and the DMM, and allow them to warm up for one hour. Select the multimeter DC voltage measuring function, and set the unit to the 20V range. Press the Model 2510 V key to select the voltage function. Using the EDIT keys, adjust the Model 2510 voltage setpoint to +10.000V. Press CONFIG then V, select PROTECTION, then set the voltage limit to 9.00V. Turn the source output on. Verify that the “VLIM” message is displayed, then note the DMM voltage reading. Verify that the voltage measured in step 8 is within 8.55 to 9.45V limits. Turn off the output when the test is completed. Current accuracy Output current accuracy Follow the steps below to verify that Model 2510 output current accuracy is within specified limits. The test involves setting the output current to a specific value and measuring the current using a 1Ω resistor and a digital multimeter. 1. 2. 3. 4. 5. 6. 7. 8. 9. With the power off, connect the digital multimeter and 1Ω resistor to the Model 2510 OUTPUT terminals, as shown in Figure 1-2. Turn on the Model 2510 and the DMM, and allow them to warm up for one hour. Select the multimeter DC voltage measuring function, and set the unit to the 20V range. Press CONFIG then I, choose PROTECTION, then set the current limit to 5.00A. Press the Model 2510 I key to select the current function, and make sure the source output is turned on. Using the EDIT keys, adjust the Model 2510 setpoint to exactly +4.5000A. Compute the current as follows: I = V/R, where V is the DMM voltage reading, and R is the characterized value of the 1Ω resistor. Verify that the current calculated in step 6 is within 4.474 to 4.526A limits. Repeat steps 6 and 7 for an output current of -4.5000A. Turn off the output when the test is completed. 1-9 Performance Verification Models 2510 and 2510-AT Figure 1-2 Connections for current verification tests 1Ω Resistor OUTPUT S- OUTPUT F+ OUTPUT F- OUTPUT S+ Connect sense (S) leads as close as possible to resistor body. WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. SENSE Ω 4 WIRE INPUT HI 350V PEAK INPUT HI LO ACV DCI ACI Ω2 Ω4 FREQ NEXT REL TRIG STORE RECALL INFO LOCAL CHAN AUTO FILTER MATH CONFIG MENU F EXIT (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) ENABLE-DIG I/O TRIGGER LINK RS-232 LINE FUSE SLOWBLOW R FRONT/REAR 2A 250V RANGE POWER SCAN 500V PEAK INPUTS TEMP RANGE IEEE-488 MADE IN U.S.A. ISOLATION FROM EARTH: 30V MAX. CAL AMPS ENTER ! 2.5A, 250V 120 DCV INPUT INPUT LO 2002 MULTIMETER PREV DISPLAY ! F+ S+ S- F- F+ F- S+ S- 1100V PEAK ! CAT I OUTPUT LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2002 DMM Model 2510 Current readback accuracy Follow the steps below to verify that Model 2510 current readback accuracy is within specified limits. The test involves setting the output current to a specific value using a 1Ω resistor and a digital multimeter. 1. 2. 3. 4. 5. 6. 7. 8. With the power off, connect the digital multimeter and 1Ω resistor to the Model 2510 OUTPUT terminals, as shown in Figure 1-2. Turn on the Model 2510 and the DMM, and allow them to warm up for one hour. Select the multimeter DC voltage measuring function, and set the unit to the 20V range. Press CONFIG then I, choose PROTECTION, then set the current limit to 5.00A. Press the Model 2510 I key to select the current function, and make sure the source output is turned on. Using the EDIT keys, adjust the Model 2510 output current to +4.5000A, as determined from the DMM voltage reading and resistor value. Compute the current as follows: I = V/R, where V is the DMM voltage reading, and R is the characterized value of the 1Ω resistor. Verify that the current calculated in step 7 is within 4.474 to 4.526A limits. Repeat steps 6 and 7 for an output current of -4.5000A. Models 2510 and 2510-AT Performance Verification 1-10 Current limit accuracy Follow the steps below to verify that Model 2510 current limit accuracy is within specified limits. The test involves setting the current limit to a specific value and making sure the current is limited to the required value. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. With the power off, connect the digital multimeter and 1Ω resistor to the Model 2510 OUTPUT terminals, as shown in Figure 1-2. Turn on the Model 2510 and the DMM, and allow them to warm up for one hour. Select the multimeter DC voltage measuring function, and set the unit to the 20V range. Press the Model 2510 V key to select the voltage function. Press CONFIG then I, select PROTECTION, then set the current limit to 4.50A. Make sure the source output is turned on. Using the EDIT keys, adjust the Model 2510 output voltage to +10.000V. Verify that the “ILIM” message is displayed, then measure the output current, as determined from the DMM voltage reading and resistor value. Compute the current as follows: I = V/R, where V is the DMM voltage reading, and R is the characterized value of the 1Ω resistor. Verify that the current calculated in step 9 is within 4.275 to 4.725A limits. Turn off the output when the test is completed. AC resistance accuracy Follow the steps below to verify that Model 2510 AC resistance accuracy is within specified limits. The test involves connecting a 100Ω resistor to the OUTPUT terminals and verifying that the Model 2510 AC resistance reading is within limits. 1. 2. 3. 4. 5. 6. With the power off, connect the 100Ω resistor to the Model 2510 OUTPUT terminals, as shown in Figure 1-3. Connect the Model 2510 to the serial port or IEEE-488 interface of the computer. Turn on the Model 2510, and allow it to warm up for one hour before testing. Press CONFIG then R. Select AC-OHMS, then press ENTER. Note the reading, and verify that the resistance measurement is within required limits. (Recalculate limits using the characterized resistance value and Model 2510 AC resistance specifications.) For a 100Ω resistance, the limits are: 99.88 to 100.12Ω. 1-11 Performance Verification Figure 1-3 Connections for AC resistance verification test Models 2510 and 2510-AT 100Ω Resistor OUTPUT S- OUTPUT F+ OUTPUT F- OUTPUT S+ Connect sense (S) leads as close as possible to resistor body. WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 2.5A, 250V 120 ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 Sensor measurement accuracy Use the following steps to verify that Model 2510 temperature sensor measurement accuracy is within specified limits. This procedure involves connecting characterized resistors to the INPUT terminals and verifying sensor resistance readings are within required limits. NOTE The following procedure tests thermistor sensor accuracy and will verify sensor measurement accuracy for all sensor types. 1. With the power off, connect the 100Ω characterized resistor to the Model 2510 INPUT terminals, as shown in Figure 1-4. Turn on the Model 2510, and allow it to warm up for one hour before testing. Select the Model 2510 voltage function by pressing the V key. 2. 3. Models 2510 and 2510-AT Figure 1-4 Connections for sensor resistance accuracy verification Performance Verification 1-12 100Ω, 1kΩ, 10kΩ, or 100kΩ Resistor INPUT S- INPUT F+ Connect sense (S) leads as close as possible to resistor body. INPUT F- INPUT S+ WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 2.5A, 250V 120 ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 4. Configure the Model 2510 for thermistor sensor type, resistance range, and 4-wire sense mode as follows: • Press CONFIG then T. The instrument will display the following: CONFIG TEMPERATURE PROTECTION SENSOR-TYPE UNITS PID • Select SENSOR-TYPE, and then press ENTER. The following will be displayed: SENSOR TYPE THERMISTOR RTD I-SS V-SS • Select THERMISTOR, and then press ENTER. The following will be displayed: THERMISTOR SETTINGS • Select RANGE, then press ENTER. The unit displays: THERMISTOR RANGE RANGE A B C I-SRC 100 1K 10K 100K • • 5. 6. Select 100, then press ENTER. From the THERMISTOR SETTINGS menu, choose SENSE-MODE, then press ENTER. • From the SENSE MODE menu, choose 4-WIRE, then press ENTER. • Press EXIT as required to return to normal display. Repeatedly press the DISPLAY TOGGLE key until the sensor resistance (RT) is displayed on the lower display line on the right. For example: RT: +099.98Ω. Verify that the sensor reading for the 100Ω resistance in Table 1-3 is within stated limits. 1-13 Performance Verification NOTE 7. Models 2510 and 2510-AT The reading limits in Table 1-3 are based on nominal resistance values. Recalculate new limits based on characterized resistance values and Model 2510 one-year thermistor sensor accuracy specifications (see Appendix A). Repeat steps 1 through 5 for the remaining resistance values in Table 1-3. Be sure to select the correct resistance range from the THERMISTOR RANGE menu. Table 1-3 Sensor resistance measurement accuracy limits Sensor range Test resistance1 100Ω 1kΩ 10kΩ 100kΩ 100Ω 1kΩ 10kΩ 100kΩ Model 2510 sensor reading limits2 (1 year, 18°C-28°C) 99.89 to 100.11Ω 0.9992 to 1.0008kΩ 9.995 to 10.005kΩ 99.939 to 100.061kΩ 1. Nominal resistance values. 2. Reading limits based on Model 2510 1-year accuracy specifications and nominal resistance values. Recalculate reading limits using actual characterized resistance values and Model 2510 1-year accuracy specifications. See “Verification limits” earlier in this section for details. 2 Calibration 2-2 Calibration Models 2510 and 2510-AT Introduction Use the procedures in this section to calibrate the Model 2510. These procedures require accurate test equipment to measure precise DC voltages and resistances. Calibration can be performed either from the front panel or by sending SCPI calibration commands over the IEEE-488 bus or RS-232 port with the aid of a computer. WARNING The information in this section is intended for qualified service personnel only. Do not attempt these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous voltages. Environmental conditions Temperature and relative humidity Conduct the calibration procedures at an ambient temperature of 18-28°C (65-82°F) with relative humidity of less than 70% unless otherwise noted. Warm-up period Allow the Model 2510 to warm up for at least one hour before performing calibration. If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range. Allow the test equipment to warm up for the minimum time specified by the manufacturer. Line power The Model 2510 requires a line voltage of 100 to 240V at line frequency of 50 or 60Hz. The instrument must be calibrated while operating from a line voltage within this range. Models 2510 and 2510-AT Calibration 2-3 Calibration considerations WARNING The maximum common-mode voltage (voltage between INPUT/OUTPUT - terminals and chassis ground) is 30V DC. Exceeding this value may cause a shock hazard. CAUTION The maximum voltage between INPUT/OUTPUT sense (S) terminals is 1V. Exceeding this voltage may result in instrument damage. When performing the calibration procedures: • • • • Make sure that the test equipment is properly warmed up and connected to the Model 2510 INPUT or OUTPUT terminals as required. Always allow the source signal to settle before calibrating each point. Do not connect test equipment to the Model 2510 through a scanner or other switching equipment. If an error occurs during calibration, the Model 2510 will generate an appropriate error message. See Appendix B for more information. Calibration cycle Perform calibration at least once a year to ensure the unit meets or exceeds its specifications. Recommended calibration equipment Table 2-1 lists the recommended equipment for the calibration procedures. You can use alternate equipment as long that equipment has specifications at least as good as those listed in the table. For optimum calibration accuracy, test equipment specifications should be at least four times better than corresponding Model 2510 specifications. 2-4 Calibration Models 2510 and 2510-AT Table 2-1 Recommended calibration equipment Description Manufacturer/Model Specifications Digital Multimeter Keithley 2002 DC Voltage1 20V: ±6.8ppm Resistance1 20Ω: 200Ω: 2kΩ: 20kΩ 200kΩ: ±23ppm ±19ppm ±7.4ppm ±7.4ppm ±29.8ppm Resistors2 Isotec RUG-Z-1R00-0.1 INPUT/OUTPUT Mating Connector3 Keithley CS-846 Clean Copper Wire 1Ω, ±0.1%, 100W 100Ω, ±1% 1kΩ, ±1% 10kΩ, ±1% 100kΩ, ±1% #18-22 AWG 3. Ninety day, full-range accuracy specification of ranges required for various measurement points. 4. Characterize all resistors using 4-wire ohms function of recommended DMM before use. 5. One connector supplied with Model 2510. Resistor characterization The resistors listed in Table 2-1 must be characterized using the 4-wire ohms function of the recommended DMM before use. Be sure to use the lowest resistance range possible for each measurement for best accuracy. Record the characterized values in Table 2-2. Table 2-2 Characterized resistor values Nominal resistance Characterized resistance* 1Ω _________Ω 100Ω _________Ω 1kΩ _________kΩ 10kΩ _________kΩ 100kΩ _________kΩ *Characterize value using DMM (see Table 2-1). Models 2510 and 2510-AT Calibration 2-5 Calibration menu Table 2-3 summarizes the main calibration menu selections. To enter the calibration menu, press the MENU key, select CAL, then press ENTER. Use the EDIT keys to move the cursor and scroll through menu selections. Press ENTER to select a MENU item. Table 2-3 Calibration menu Menu selection Description UNLOCK Unlock calibration using password (default: 002510). EXECUTE VOLTAGE CURRENT I-PROTECTION TEMPERATURE AC-OHMS V-SOURCE Execute calibration steps. Voltage calibration. Current calibration. Current protection calibration. Temperature calibration. AC resistance calibration. Voltage source calibration. VIEW-DATES View calibration dates. SAVE Save calibration constants. LOCK Lock out calibration. CHANGE-PASSWORD Change calibration password. Unlocking calibration Before performing calibration, you must first unlock calibration by entering or sending the calibration password as explained in the following paragraphs. Unlocking calibration from the front panel 1. Press the MENU key, choose CAL, and press ENTER. The instrument will display the following: CALIBRATION UNLOCK EXECUTE VIEW-DATES SAVE LOCK CHANGE-PASSWORD 2. Select UNLOCK, and then press ENTER. The instrument will display the following: PASSWORD: Use , , , , ENTER or EXIT. 3. 4. Use the and keys to select the letter or number, and use the and arrow keys to choose the position. (Press for letters; for numbers.) Enter the present password on the display. (Front panel default: 002510.) Once the correct password is displayed, press the ENTER key. You can then proceed with the calibration procedure. 2-6 Calibration Models 2510 and 2510-AT Unlocking calibration by remote To unlock calibration via remote, send the following command: :CAL:PROT:CODE '<password>' For example, the following command uses the default password: :CAL:PROT:CODE 'KI002510' Changing the password The default password (002510) may be changed from the front panel or via remote as discussed. Changing the password from the front panel Follow the steps below to change the password from the front panel: 1. Press the MENU key, choose CAL, and press ENTER. The instrument will display the following: CALIBRATION UNLOCK EXECUTE VIEW-DATES SAVE LOCK CHANGE-PASSWORD 2. 3. Select UNLOCK, then enter the password. (Default: 002510.) Select CHANGE-PASSWORD, and then press ENTER. The instrument will display the following: New Pwd: 002510 Use , , , , ENTER or EXIT. 4. 5. Using the EDIT keys, enter the new password on the display. Once the desired password is displayed, press the ENTER key to store the new password. Changing the password by remote To change the calibration password by remote, first send the present password, and then send the new password. For example, the following command sequence changes the password from the 'KI002510' remote default to 'KI_CAL': :CAL:PROT:CODE :CAL:PROT:CODE 'KI002510' 'KI_CAL' You can use any combination of letters and numbers up to a maximum of eight characters. NOTE If you change the first two characters of the password to something other than “KI”, you will not be able to unlock calibration from the front panel. Models 2510 and 2510-AT Calibration 2-7 Resetting the calibration password If you lose the calibration password, you can unlock calibration by shorting together the CAL pads, which are located on the display board. Doing so will reset the password to the factory default (002510, front panel; KI002510, remote). See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the display board component layout drawing at the end of Section 6 for the location of the CAL pads. Viewing calibration dates and calibration count When calibration is locked, only the UNLOCK and VIEW-DATES selections will be accessible in the calibration menu. To view calibration dates and calibration count at any time: 1. From normal display, press MENU, select CAL, then press ENTER. The unit will display the following: CALIBRATION UNLOCK EXECUTE VIEW-DATES 2. Select VIEW-DATES, and then press ENTER. The Model 2510 will display the next and last calibration dates and the calibration count as in the following example: NEXT CAL: 02/15/2001 Last calibration: 02/15/2000 Count: 0001 Calibration errors The Model 2510 checks for errors after each calibration step, minimizing the possibility that improper calibration may occur due to operator error. Front panel error reporting If an error is detected during comprehensive calibration, the instrument will display an appropriate error message (see Appendix B). The unit will then prompt you to repeat the calibration step that caused the error. Remote error reporting You can detect errors while in remote by testing the state of EAV (Error Available) bit (bit 2) in the status byte. (Use the *STB? query to request the status byte.) Query the instrument for the type of error by using the :SYST:ERR? query. The Model 2510 will respond with the error number and a text message describing the nature of the error. See Appendix B for details. 2-8 Calibration Models 2510 and 2510-AT Aborting calibration steps To abort a calibration step from the front panel, press the EXIT key. To abort a calibration step via remote, send the :ABORt command. Front panel calibration The front panel calibration procedure described below calibrates all functions. Note that each function is separately calibrated, and the procedure must be performed in the order shown. Step 1: Prepare the Model 2510 for calibration 1. 2. 3. Turn on the Model 2510 and the digital multimeter, and allow them to warm up for at least one hour before performing calibration. Configure the INPUT terminals for 4-wire sensing as follows: • Press CONFIG then T. • Select SENSOR-TYPE, then press ENTER. • Choose THERMISTOR, then press ENTER. • Select SENSE-MODE, then press ENTER. • Choose 4-WIRE, then press ENTER. • Press EXIT as required to back out of the menu structure and return to normal display. Press the MENU key, choose CAL, and press ENTER. Select UNLOCK, and then press ENTER. The instrument will display the following: PASSWORD: Use , , , , ENTER or EXIT. 4. 5. 6. Use the and keys to select the letter or number, and use the and arrow keys to choose the position. (Press for letters; for numbers.) Enter the present password on the display. (Front panel default: 002510.) Press ENTER to complete the process. Select EXECUTE, then press ENTER to enter the CAL EXECUTE menu. Table 2-4 summarizes the various menu selections and calibration steps, which are covered in more detail throughout the procedure. Models 2510 and 2510-AT Calibration 2-9 Table 2-4 Calibration step summary Function* Calibration step Test connections VOLTAGE Output +8.5V Output 0V Output -8.5V DMM to OUTPUT terminals (Figure 2-1) DMM to OUTPUT terminals (Figure 2-1) DMM to OUTPUT terminals (Figure 2-1) CURRENT 1Ω resistor 1Ω resistor to OUTPUT terminals (Figure 2-2) I-PROTECTION 1Ω resistor 1Ω resistor to OUTPUT terminals (Figure 2-2) 100Ω resistor, short 1kΩ resistor, short 10kΩ resistor, short 100kΩ resistor, short 100Ω resistor, short 1kΩ resistor, short 1kΩ resistor 1kΩ resistor, short 100Ω resistor then short to INPUT terminals (Figure 2-3 and Figure 2-4) 1kΩ resistor then short to INPUT terminals (Figure 2-3 and Figure 2-4) 10kΩ resistor then short to INPUT terminals (Figure 2-3 and Figure 2-4) 100kΩ resistor then short to INPUT terminals (Figure 2-3 and Figure 2-4) 100Ω resistor then short to INPUT terminals (Figure 2-3 and Figure 2-4) 1kΩ resistor then short to INPUT terminals (Figure 2-3 and Figure 2-4) DMM and 1kΩ resistor to INPUT terminals (Figure 2-5) DMM, 1kΩ resistor; short to INPUT terminals (Figure 2-4 and Figure 2-5) AC-OHMS 100Ω resistor 100Ω resistor to OUTPUT terminals (Figure 2-6) V-SOURCE Open leads None (open OUTPUT terminals) TEMPERATURE THERMISTOR RTD I-SS V-SS * CAL EXECUTE menu selections. Step 2: Voltage calibration 1. From the CAL EXECUTE menu, select VOLTAGE, then press ENTER. The instrument will display the following message: VOLTAGE CAL Connect only a DMM to Peltier. Then press ENTER. 2. Connect the DMM to the Model 2510 OUTPUT terminals, as shown in Figure 2-1. Figure 2-1 Voltage calibration connections OUTPUT F+ OUTPUT S+ WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. INPUT HI INPUT HI CAT I 350V PEAK 1100V PEAK ! INPUT LO LO PREV NEXT DCV ACV DCI ACI Ω2 Ω4 FREQ TRIG STORE RECALL INFO LOCAL CHAN AUTO FILTER MATH RANGE SCAN CONFIG MENU F OUTPUT S- R FRONT/REAR 2A 250V POWER EXIT ENTER IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) ENABLE-DIG I/O 500V PEAK INPUTS TEMP RANGE REL INPUT F+ S+ S- F- F+ F- S+ SISOLATION FROM EARTH: 30V MAX. 2002 MULTIMETER DISPLAY OUTPUT ! CAL AMPS RS-232 TRIGGER LINK OUTPUT F- LINE FUSE SLOWBLOW ! Model 2002 DMM 2.5A, 250V 120 SENSE Ω 4 WIRE LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 2-10 Calibration Models 2510 and 2510-AT 3. 4. Select the DMM DC voltage function, and enable auto-range. Press ENTER. The unit will then prompt for the +8.5V step: VOLTAGE CAL Press ENTER to output +8.5V NOTE The output will turn on automatically when you perform the next step. 5. Press ENTER. The Model 2510 will source +8.5V and simultaneously display the following: DMM RDG: +08.50000 V 6. Note and record the DMM reading, and then use the EDIT keys to adjust the Model 2510 display to agree exactly with the actual DMM reading. After adjusting the display to agree with the DMM reading, press ENTER. The instrument will then display the following: Use , , , , ENTER or EXIT. 7. VOLTAGE CAL Press ENTER to output 0.0V 8. Press ENTER. The Model 2510 will source 0V and at the same time display the following: DMM RDG: +00.00000 V 9. Note and record the DMM reading, and then adjust the Model 2510 display to agree with that reading. After adjusting the display value to agree with the DMM reading, press ENTER. The unit will then display the following: VOLTAGE CAL Use , , , , ENTER or EXIT. 10. Press ENTER to output -8.5V 11. Press ENTER. The Model 2510 will source -8.5V and display the following: DMM RDG: -08.50000 V Use , , , , ENTER or EXIT. 12. 13. Note and record the DMM reading, adjust the Model 2510 display to agree with the DMM reading, and press ENTER. Disconnect the DMM from the OUTPUT terminals. Models 2510 and 2510-AT Calibration 2-11 Step 3: Current and current protection calibration 1. From the CAL EXECUTE menu, select CURRENT, then press ENTER. The instrument will display the following message: CURRENT CAL Connect a 1Ω load to Peltier. Then press ENTER. 2. Figure 2-2 Current and current protection calibration connections Connect the 1Ω resistor to the Model 2510 OUTPUT terminals, as shown in Figure 2-2. 1Ω Resistor OUTPUT S- OUTPUT F+ OUTPUT F- OUTPUT S+ Connect sense (S) leads as close as possible to resistor body. WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 2.5A, 250V 120 ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 3. Press ENTER. The Model 2510 will display the following: RES VALUE: 1.00000Ω Use , , , , ENTER or EXIT. 4. 5. Adjust the Model 2510 display to agree exactly with the characterized value of the 1Ω resistor, then press ENTER. Select I-PROTECTION, then press ENTER. The instrument will then display the following: CURRENT PROTECTION Connect 1Ω load to Peltier. Then press ENTER. 6. 7. With the 1Ω resistor still connected to the OUTPUT terminals, press ENTER. The Model 2510 will automatically complete the current protection calibration process. Disconnect the 1Ω resistor from the OUTPUT terminals. 2-12 Calibration Models 2510 and 2510-AT Step 4: Temperature calibration 1. 2. Connect the 100Ω resistor to the Model 2510 INPUT terminals, as shown in Figure 2-3. From the CAL EXECUTE menu, select TEMPERATURE, then press ENTER. The instrument will display the following message: TEMPERATURE CAL THERMISTOR RTD I-SS V-SS Figure 2-3 Thermistor and RTD calibration resistor connections 100Ω, 1kΩ, 10kΩ, or 100kΩ Resistor INPUT S- INPUT F+ Connect sense (S) leads as close as possible to resistor body. INPUT F- INPUT S+ WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 120 2.5A, 250V ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 3. Select THERMISTOR, then press ENTER. The instrument will display the following: THERMISTOR CAL 100Ω 1kΩ 10kΩ 100kΩ 4. Select 100Ω, then press ENTER. The unit will display the following: THERMISTOR CAL Connect 100Ω to temperature sensor and press ENTER. 5. Make sure the 100Ω resistor is properly connected, then press ENTER. The Model 2510 will display the following: RES VALUE: 0.100000kΩ 6. Use the EDIT keys to adjust the Model 2510 display to agree exactly with the characterized 100Ω resistance value, then press ENTER. The unit will display the following: THERMISTOR CAL Use , , , , ENTER or EXIT. Connect a Short to temperature sensor and press ENTER. 7. Short all four INPUT terminals together using clean copper wire, as shown in Figure 2-4. Allow one minute for thermal equilibrium. Models 2510 and 2510-AT Calibration Figure 2-4 Shorted input calibration connections 2-13 Short INPUT Terminals with Clean Copper Wire WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 2.5A, 250V 120 ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 The characterized 1Ω resistor can be used in place of the shorted INPUT terminals, if desired. Enter the characterized value of the 1Ω resistor at the RES VALUE prompt. NOTE 8. Press ENTER. The Model 2510 will display the following: RES VALUE: 00.00000Ω Use , , , , ENTER or EXIT. 9. 10. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. From the THERMISTOR CAL menu, select 1kΩ, then press ENTER. The unit will display the following: THERMISTOR CAL Connect 1kΩ to temperature sensor and press ENTER. 11. 12. Connect the 1kΩ resistor to the Model 2510 INPUT terminals (see Figure 2-3). Press ENTER. The Model 2510 will display the following: RES VALUE: 01.00000kΩ Use , , , , ENTER or EXIT. 13. Use the EDIT keys to adjust the Model 2510 display to agree exactly with the characterized 1kΩ resistance value, then press ENTER. The unit will display the following: THERMISTOR CAL Connect a Short to temperature sensor and press ENTER. 14. 15. Short all four INPUT terminals together using clean copper wire (see Figure 2-4). Allow one minute for thermal equilibrium. Press ENTER. The Model 2510 will display the following: RES VALUE: 00.00000Ω Use , , , , ENTER or EXIT. 16. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. 2-14 Calibration 17. Models 2510 and 2510-AT From the THERMISTOR CAL menu, select 10kΩ, then press ENTER. The unit will display the following: THERMISTOR CAL Connect 10kΩ to temperature sensor and press ENTER. 18. 19. Connect the 10kΩ resistor to the Model 2510 INPUT terminals (Figure 2-3). Press ENTER. The Model 2510 will display the following: RES VALUE: 010.0000kΩ Use , , , , ENTER or EXIT. 20. Use the EDIT keys to adjust the Model 2510 display to agree exactly with the characterized 10kΩ resistance value, then press ENTER. The unit will display the following: THERMISTOR CAL Connect a Short to temperature sensor and press ENTER. 21. 22. Short all four INPUT terminals together using clean copper wire (see Figure 2-4). Allow one minute for thermal equilibrium. Press ENTER. The Model 2510 will display the following: RES VALUE: 00.00000Ω Use , , , , ENTER or EXIT. 23. 24. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. From the THERMISTOR CAL menu, select 100kΩ, then press ENTER. The unit will display the following: THERMISTOR CAL Connect 100kΩ to temperature sensor and press ENTER. 25. 26. Connect the 100kΩ resistor to the Model 2510 INPUT terminals (Figure 2-3). Press ENTER. The Model 2510 will display the following: RES VALUE: 0100.000kΩ Use , , , , ENTER or EXIT. 27. Use the EDIT keys to adjust the Model 2510 display to agree exactly with the characterized 100kΩ resistance value, then press ENTER. The unit will display the following: THERMISTOR CAL Connect a Short to temperature sensor and press ENTER. 28. 29. Short all four INPUT terminals together using clean copper wire (see Figure 2-4). Allow one minute for thermal equilibrium. Press ENTER. The Model 2510 will display the following: RES VALUE: 00.00000Ω Use , , , , ENTER or EXIT. Models 2510 and 2510-AT Calibration 2-15 30. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. 31. Press EXIT to return to the TEMPERATURE CAL menu. 32. Select RTD, then press ENTER. The unit displays the following: RTD CAL Connect 100Ω to temperature sensor and press ENTER. 33. Connect the 100Ω resistor to the Model 2510 INPUT terminals (Figure 2-3). 34. Press ENTER. The Model 2510 will display the following: RES VALUE: 0.100000kΩ Use , , , , ENTER or EXIT. 35. Adjust the display to agree with the value of the 100Ω resistor, then press ENTER. The unit will display the following: RTD CAL Connect a Short to temperature sensor and press ENTER. 36. Short all four INPUT terminals together using clean copper wire (see Figure 2-4). Allow one minute for thermal equilibrium. 37. Press ENTER. The Model 2510 will display the following: RES VALUE: 00.00000Ω Use , , , , ENTER or EXIT. 38. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. 39. Connect the 1kΩ resistor to the Model 2510 INPUT terminals (Figure 2-3). 40. Press ENTER. The Model 2510 will display the following: RES VALUE: 01.00000kΩ Use , , , , ENTER or EXIT. 41. Adjust the display to agree with the value of the 1kΩ resistor, then press ENTER. The unit will display the following: RTD CAL Connect a Short to temperature sensor and press ENTER. 42. Short all four INPUT terminals together using clean copper wire (see Figure 2-4). Allow one minute for thermal equilibrium. 43. Press ENTER. The Model 2510 will display the following: RES VALUE: 00.00000Ω Use , , , , ENTER or EXIT. 2-16 Calibration Models 2510 and 2510-AT 44. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. 45. From the TEMPERATURE CAL menu, select I-SS, then press ENTER. The unit displays the following: I-SS CAL Connect 1kΩ to temperature sensor and press ENTER. 46. Connect the DMM and 1kΩ resistor to the Model 2510 (Figure 2-5). Press ENTER. The Model 2510 will display the following: RES VALUE: 01.00000kΩ Use , , , , ENTER or EXIT. 47. Use the EDIT keys to adjust the Model 2510 display to agree exactly with the characterized 1kΩ resistance value, then press ENTER. The unit will prompt for the DMM voltage reading: DMM RDG: +0.833330 V Use , , , , ENTER or EXIT. 48. Note the DMM reading, then adjust the Model 2510 display to agree with that value. Press ENTER. 49. From the TEMPERATURE CAL menu, select V-SS, then press ENTER. The unit displays the following: V-SS CAL Connect 1kΩ to temperature sensor and press ENTER. 50. Make sure the 1kΩ resistor and the DMM are still connected to the Model 2510 INPUT terminals (Figure 2-5). 51. Press ENTER. The Model 2510 will display the following: DMM RDG: +2.500000V Use , , , , ENTER or EXIT. 52. Note the DMM reading, use the EDIT keys to adjust the Model 2510 display to agree exactly with that value, then press ENTER. The unit will display the following: V-SS CAL Connect a Short to temperature sensor and press ENTER. 53. Short all four INPUT terminals together using clean copper wire (see Figure 2-4). Allow one minute for thermal equilibrium. 54. Press ENTER. The Model 2510 will display the following: DMM RDG: +0.000000V Use , , , , ENTER or EXIT. Models 2510 and 2510-AT Calibration 55. Press ENTER to complete the shorted calibration step, then remove the shorting wires from the INPUT terminals. 56. Press EXIT to return to the CAL EXECUTE menu. Figure 2-5 I-SS and V-SS calibration resistor connections 1kΩ Resistor INPUT S- INPUT F+ INPUT S+ Connect sense (S) leads as close as possible to resistor body. INPUT FWARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. SENSE Ω 4 WIRE INPUT HI INPUT HI 1100V PEAK ! LO ACV DCI ACI Ω2 Ω4 FREQ NEXT REL TRIG STORE RECALL INFO LOCAL CHAN AUTO FILTER MATH CONFIG MENU F EXIT INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW R FRONT/REAR 2A 250V RANGE POWER SCAN 500V PEAK INPUTS TEMP RANGE OUTPUT ! INPUT LO 2002 MULTIMETER DCV DISPLAY CAT I CAL AMPS ! 2.5A, 250V 120 350V PEAK PREV 2-17 LINE RATING 100-240VAC 50, 60 HZ 90VA MAX ENTER CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2002 DMM Model 2510 2-18 Calibration Models 2510 and 2510-AT Step 5: AC ohms calibration 1. From the CAL EXECUTE menu, select AC-OHMS, then press ENTER. The instrument will display the following message: AC-OHMS CAL Connect a 100Ω load to Peltier. Then press ENTER. 2. 3. Connect the 100Ω resistor to the OUTPUT terminals (see Figure 2-6). Press ENTER. The instrument will display the following: RES VALUE: 0.100000kΩ 4. Use the EDIT keys to adjust the Model 2510 display to agree exactly with the characterized 100Ω resistance value, then press ENTER. Use , , , , ENTER or EXIT. Figure 2-6 AC ohms calibration connections 100Ω Resistor OUTPUT F+ OUTPUT S- OUTPUT S+ OUTPUT F- Connect sense (S) leads as close as possible to resistor body. WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! IEEE-488 MADE IN U.S.A. INPUT (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 2.5A, 250V 120 ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Model 2510 Step 6: Voltage source calibration 1. From the CAL EXECUTE menu, select V-SOURCE, then press ENTER. The instrument will display the following message: V-SRC CAL Open Peltier leads then press ENTER. 2. 3. 4. Disconnect all resistors and test leads from the Model 2510 OUTPUT terminals. Press ENTER to complete voltage source calibration. Press EXIT to return to the CALIBRATION menu. Models 2510 and 2510-AT Calibration 2-19 Step 7: Enter calibration dates and save calibration NOTE 1. For temporary calibration without saving new calibration constants, proceed to Step 8: Lock out calibration. From the CALIBRATION menu, select SAVE, then press ENTER. The unit will prompt you for the calibration date: CAL DATE: 02/15/2000 Use , , , , ENTER or EXIT. 2. 3. Using the EDIT keys, change the displayed date to today's date, then press the ENTER key. Press ENTER again to confirm the date. The unit will then prompt for the calibration due date: NEXT CAL: 02/15/2001 Use , , , , ENTER or EXIT. 4. 5. Set the calibration due date to the desired value, then press ENTER. Press ENTER again to confirm the date. Once the calibration dates are entered, calibration is complete, and the following message will be displayed: CALIBRATION COMPLETE Press ENTER to save; EXIT to abort 6. Press ENTER to save the calibration data (or press EXIT to abort without saving calibration data). Step 8: Lock out calibration From the CAL EXECUTE menu, select LOCK, then press ENTER to lock out calibration. Press EXIT to return to normal display. 2-20 Calibration Models 2510 and 2510-AT Remote calibration Use the following procedure to perform remote calibration by sending SCPI commands over the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are separately summarized for each step. Remote calibration command summary Table 2-5 summarizes remote calibration commands used in this section. NOTE For a detailed description of all calibration commands and queries, refer to Appendix B. Table 2-5 Remote calibration command summary Command Description :CALibration :PROTected :CODE '<password>' :SENSe :VOLTage <NRf> :CURRent <NRf> :TEMPerature <NRf> :SOURce <NRf> :AUTO :IPDac :RESistance [:AC] <NRf> :DATE <y>,<m>,<d> :NDUE <y>,<m>,<d> :SAVE :LOCK Calibration subsystem. Calibration commands protected by code/password. Unlock calibration. (Default code/password: KI002510.) Calibrate Peltier voltage measurement. Calibrate current measurement. Calibrate temperature measurement, Calibrate Peltier source/sense voltage. Calibrate voltage source. Calibrate current protection DAC. Calibrate AC ohms. Program calibration year, month, day. Program calibration due year, month, day. Save calibration data in EEPROM. Lock out calibration. Models 2510 and 2510-AT Calibration 2-21 Remote calibration procedure Step 1: Prepare the Model 2510 for calibration 1. 2. 3. 4. 5. With the power off, connect the Model 2510 to the controller IEEE-488 interface or RS-232 port using a shielded interface cable. Turn on the Model 2510 and the test equipment, and allow them to warm up for at least one hour before performing calibration. If you are using the IEEE-488 interface, make sure the primary address of the Model 2510 is the same as the address specified in the program you will be using to send commands. (Use the MENU key and the COMMUNICATION menu to access the IEEE-488 address.) Send the following command to unlock calibration: :CAL:PROT:CODE ‘KI002510’ Table 2-6 summarizes the various calibration steps and associated commands, which are covered in more detail throughout the procedure. Table 2-6 Remote calibration step summary Calibrated function Calibration command Voltage :OUTP ON :SOUR:VOLT 8.5 :CAL:PROT:SENS:VOLT <DMM_Reading> :CAL:PROT:SOUR <DMM_Reading> :SOUR:VOLT 0 :CAL:PROT:SENS:VOLT <DMM_Reading> :CAL:PROT:SOUR <DMM_Reading> :SOUR:VOLT -8.5 :CAL:PROT:SENS :VOLT <DMM_Reading> :CAL:PROT:SOUR <DMM_Reading> :OUTP OFF Current Current protection :OUTP ON :CAL:PROT:SENS:CURR <Resistance_Value> :CAL:PROT:IPD :OUTP OFF Test connections DMM to OUTPUT (Figure 2-1) DMM to OUTPUT (Figure 2-1) DMM to OUTPUT (Figure 2-1) DMM to OUTPUT (Figure 2-1) DMM to OUTPUT (Figure 2-1) DMM to OUTPUT (Figure 2-1) 1Ω to OUTPUT (Figure 2-2) 1Ω to OUTPUT (Figure 2-2) 2-22 Calibration Models 2510 and 2510-AT Table 2-6 (cont.) Remote calibration step summary Calibrated function Calibration command Temperature Thermistor sensor :SYST:RSEN ON :SENS:TEMP:TRAN THER :SENS:TEMP:THER:RANG 100 :CAL:PROT:TEMP <Resistance_Value> :CAL:PROT:TEMP 0 :SENS:TEMP:THER:RANG 1e3 :CAL:PROT:TEMP <Resistance_Value> :CAL:PROT:TEMP 0 :SENS:TEMP:THER:RANG 1e4 :CAL:PROT:TEMP <Resistance_Value> :CAL:PROT:TEMP 0 :SENS:TEMP:THER:RANG 1e5 :CAL:PROT:TEMP <Resistance_Value> :CAL:PROT:TEMP 0 :SENS:TEMP:TRAN RTD :SENS:TEMP:RTD:RANG 100 :CAL:PROT:TEMP <Resistance_Value> :CAL:PROT:TEMP 0 :SENS:TEMP:RTD:RANG 1000 :CAL:PROT:TEMP<Resistance_Value> :CAL:PROT:TEMP 0 :SENS:TEMP:TRAN ISS :CAL:PROT:TEMP <Resistance_Value> :CAL:PROT:TEMP <DMM_Reading> :SENS:TEMP:TRAN VSS :CAL:PROT:TEMP <DMM_Reading> :CAL:PROT:TEMP 0 RTD sensor Current sensor Voltage sensor Test connections 100Ω to INPUT (Figure 2-3) Short to INPUT (Figure 2-4) 1kΩ to INPUT (Figure 2-3) Short to INPUT (Figure 2-4) 10kΩ to INPUT (Figure 2-3) Short to INPUT (Figure 2-4) 100kΩ to INPUT (Figure 2-3) Short to INPUT (Figure 2-4) 100Ω to INPUT (Figure 2-3) Short to INPUT (Figure 2-4) 1000Ω to INPUT (Figure 2-3) Short to INPUT (Figure 2-4) 1kΩ/DMM to INPUT (Figure 2-5) 1kΩ/DMM to INPUT (Figure 2-5) Short to INPUT (Figure 2-4) AC ohms :CAL:PROT:RES <Resistance_Value> 100Ω to OUTPUT (Figure 2-6) Voltage source :OUTP ON :CAL:PROT:SOUR:AUTO :OUTP OFF None (open OUTPUT) Models 2510 and 2510-AT Calibration 2-23 Step 2: Voltage calibration 1. 2. 3. Connect the DMM to the Model 2510 OUTPUT terminals, as shown in Figure 2-1. Select the DMM DC voltage function, and enable auto-range. Send this command to turn on the output: :OUTP ON 4. Send the following command to output +8.5V: :SOUR:VOLT 8.5 5. Note and record the DMM reading, and then send that value as the parameter for the following commands: :CAL:PROT:SENS:VOLT <DMM_Reading> :CAL:PROT:SOUR <DMM_Reading> For example, if the DMM reading is 8.51V, the correct commands are: :CAL:PROT:SENS:VOLT 8.51 :CAL:PROT:SOUR 8.51 6. Send the following command to output 0V: :SOUR:VOLT 0 7. Note and record the DMM reading, then send that value as the parameter for the following commands: :CAL:PROT:SENS:VOLT <DMM_Reading> :CAL:PROT:SOUR <DMM_Reading> 8. Send the following command to output -8.5V: :SOUR:VOLT -8.5 9. Note and record the DMM reading, then send that value as the parameter for the following commands: :CAL:PROT:SENS:VOLT <DMM_Reading> :CAL:PROT:SOUR <DMM_Reading> 10. Send this command to turn off the output: :OUTP OFF 11. Disconnect the DMM from the OUTPUT terminals. Step 3: Current and current protection calibration 1. 2. Connect the 1Ω resistor to the Model 2510 OUTPUT terminals, as shown in Figure 2-2. Send this command to turn on the output: :OUTP ON 3. Send the following command with the characterized 1Ω resistor value as the command parameter: :CAL:PROT:SENS:CURR <Resistance_Value> For example, if the actual resistor value is 1.01Ω, the correct command is: :CAL:PROT:SENS:CURR 1.01 2-24 Calibration Models 2510 and 2510-AT 4. With the 1Ω resistor still connected, send the following command to calibrate the current protection DAC: :CAL:PROT:IPD 5. Send this command to turn off the output: :OUTP OFF 6. Disconnect the 1Ω resistor from the OUTPUT terminals. Step 4: Temperature calibration 1. 2. Connect the 100Ω resistor to the Model 2510 INPUT terminals, as shown in Figure 2-3. Send the following command to select the thermistor type temperature sensor: :SENS:TEMP:TRAN THER 3. Send the following command to enable 4-wire sensing: :SYST:RSEN ON 4. Send this command to select the 100Ω resistance range: :SENS:TEMP:THER:RANG 100 5. Send the following command with the characterized 100Ω resistance value as the parameter to calibrate the 100Ω range: :CAL:PROT:SENS:TEMP <Resistance_Value> 6. Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 7. 8. Connect the 1kΩ resistor to the Model 2510 INPUT terminals (Figure 2-3). Send this command to select the 1kΩ resistance range: :SENS:TEMP:THER:RANG 1e3 9. Send the following command with the characterized 1kΩ resistance value as the parameter to calibrate the 1kΩ range: :CAL:PROT:SENS:TEMP <Resistance_Value> 10. Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 11. 12. Connect the 10kΩ resistor to the Model 2510 INPUT terminals (Figure 2-3). Send this command to select the 10kΩ resistance range: :SENS:TEMP:THER:RANG 1e4 13. Send the following command with the characterized 10kΩ resistance value as the parameter to calibrate the 10kΩ range: :CAL:PROT:SENS:TEMP <Resistance_Value> 14. Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 15. Connect the 100kΩ resistor to the Model 2510 INPUT terminals (Figure 2-3). Models 2510 and 2510-AT Calibration 2-25 16. Send this command to select the 100kΩ resistance range: :SENS:TEMP:THER:RANG 1e5 17. Send the following command with the characterized 100kΩ resistance value as the parameter to calibrate the 100kΩ range: :CAL:PROT:SENS:TEMP <Resistance_Value> 18. Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 19. 20. Connect the 100Ω resistor to the INPUT terminals (Figure 2-3). Send the following command to select the RTD type temperature sensor: :SENS:TEMP:TRAN RTD 21. Send this command to select the 100Ω range: :SENS:TEMP:RTD 100 Send the following command with the characterized 100Ω resistance value as the parameter to calibrate the RTD measurement function: :CAL:PROT:SENS:TEMP <Resistance_Value> 22. 23. Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 24. 25. Connect the 1000Ω resistor to the INPUT terminals (Figure 2-5). Send this command to select the 1000Ω range: :SENS:TEMP:RTD:RANG 1000 Send the following command with the characterized 1000Ω resistance value as the parameter to calibrate the RTD measurement function: :CAL:PROT:SENS:TEMP <Resistance_Value> Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 Connect the DMM and the 1kΩ resistor to the INPUT terminals (Figure 2-5). Send the following command to select a solid-state, current-type temperature sensor: :SENS:TEMP:TRAN ISS 26. 27. 28. 29. 30. Send the following command with the characterized 1kΩ resistance value as the parameter: :CAL:PROT:SENS:TEMP <Resistance_Value> 31. Note the DMM reading, then send that value as the parameter for the following command: :CAL:PROT:SENS:TEMP <DMM_Reading> 32. Make sure the 1kΩ resistor and DMM are still connected to the Model 2510 INPUT terminals (Figure 2-5). Send the following command to select a solid-state, voltage-type temperature sensor: :SENS:TEMP:TRAN VSS 33. 2-26 Calibration Models 2510 and 2510-AT 34. Note the DMM reading, then send the following command with the DMM reading value as the parameter to calibrate the current type solid-state sensor: :CAL:PROT:SENS:TEMP <DMM_Reading> 35. Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for thermal equilibrium, then send this command: :CAL:PROT:SENS:TEMP 0 Step 5: AC ohms calibration 1. 2. Connect the 100Ω resistor to the OUTPUT terminals (see Figure 2-6). Note that the output will be turned off for the AC ohms calibration step. Send the following command with the characterized 100Ω resistance value as the command parameter to calibrate the AC ohms function: :CAL:PROT:RES <Resistance_Value> Step 6: Voltage source calibration 1. 2. Disconnect all resistors and test leads from the Model 2510 OUTPUT terminals. Turn on the output by sending: :OUTP ON 3. Send this command to complete voltage source calibration: :CAL:PROT:SOUR:AUTO 4. Turn off the output by sending: :OUTP OFF Step 7: Program calibration dates Use following commands to set the calibration date and calibration due date: :CAL:PROT:DATE :CAL:PROT:NDUE <yyyy>, <mm>, <dd> <yyyy>, <mm>, <dd> (Calibration date) (Next calibration due date) Note that the year, month, and date must be separated by commas. Step 8: Save calibration constants Calibration is now complete, so you can store the calibration constants in EEROM by sending the following command: :CAL:PROT:SAVE NOTE Calibration will be temporary unless you send the SAVE command. Step 9: Lock out calibration To lock out further calibration, send the following command after completing the calibration procedure: :CAL:PROT:LOCK 3 Routine Maintenance 3-2 Routine Maintenance Models 2510 and 2510-AT Introduction The information in this section deals with routine type maintenance that can be performed by the operator. Line fuse replacement WARNING Disconnect the line cord at the rear panel, and remove all test leads connected to the instrument (front and rear) before replacing the line fuse. The power line fuse is accessible from the rear panel and is integral with the AC power module (see Figure 3-1). Figure 3-1 Rear panel Model 2510 WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. CAT I OUTPUT ! INPUT IEEE-488 MADE IN U.S.A. (ENTER IEEE ADDRESS WITH FRONT PANEL MENU) F+ S+ S- F- F+ F- S+ S- ENABLE-DIG I/O ISOLATION FROM EARTH: 30V MAX. RS-232 TRIGGER LINK LINE FUSE SLOWBLOW 2.5A, 250V 120 ! LINE RATING 100-240VAC 50, 60 HZ 90VA MAX CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. Line Fuse Models 2510 and 2510-AT Routine Maintenance 3-3 Perform the following steps to replace the line fuse: 1. 2. Using a small flat-blade screwdriver, carefully release the locking tab that secures the fuse carrier to the power module. Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1. CAUTION 3. NOTE To prevent instrument damage, use only the fuse type specified in Table 3-1. Reinstall the fuse carrier, pushing it in firmly until it locks into place. If the power line fuse continues to blow, a circuit malfunction exists and must be corrected. Refer to the troubleshooting information in Section 4 of this manual for additional information. Table 3-1 Power line fuse Line voltage Rating Keithley part no. 100-240V 250V, 2.5A, Slow Blow 5 × 20mm FU-106-2.5 4 Troubleshooting 4-2 Troubleshooting Models 2510 and 2510-AT Introduction This section of the manual will assist you in troubleshooting and repairing the Model 2510. Included are self-tests, test procedures, troubleshooting tables, and circuit descriptions. Note that disassembly instructions are located in Section 5, and component layout drawings are at the end of Section 6. Safety considerations WARNING The information in this section is intended for qualified service personnel only. Do not perform these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous voltages that could cause personal injury or death. Use caution when working with hazardous voltages. Repair considerations Before making any repairs to the Model 2510, be sure to read the following considerations. CAUTION • • • • • The PC boards are built using surface mount techniques and require specialized equipment and skills for repair. If you are not equipped and/or qualified, it is strongly recommended that you send the unit back to the factory for repairs or limit repairs to the PC board replacement level. Without proper equipment and training, you could damage a PC board beyond repair. Repairs will require various degrees of disassembly; however, it is recommended that the Front Panel Tests be performed prior to any disassembly. The disassembly instructions for the Model 2510 are contained in Section 5 of this manual. Do not make repairs to surface mount PC boards unless equipped and qualified to do so (see previous CAUTION). When working inside the unit and replacing parts, be sure to adhere to the handling precautions and cleaning procedures explained in Section 5. Many CMOS devices are installed in the Model 2510. These static-sensitive devices require special handling as explained in Section 5. Whenever a circuit board is removed or a component is replaced, the Model 2510 must be recalibrated. See Section 2 for details on calibrating the unit. Models 2510 and 2510-AT Troubleshooting 4-3 Power-on self-test During the power-on sequence, the Model 2510 will perform a checksum test on its EPROM and test its RAM. If one of these tests fails, the instrument will lock up. Front panel tests There are three front panel tests: one to test the functionality of the front panel keys and two to test the display. In the event of a test failure, refer to Display board checks for details on troubleshooting the display board. KEYS test The KEYS test lets you check the functionality of each front panel key. Perform the following steps to run the KEYS test. 1. 2. 3. Display the MAIN MENU by pressing the MENU key. Using the EDIT keys, select TEST, and press ENTER to display the SELF-TEST MENU. Select DISPLAY-TESTS, and press ENTER to display the following menu: FRONT PANEL TESTS KEYS DISPLAY-PATTERNS CHAR-SET 4. 5. Select KEYS, and press ENTER to start the test. When a key is pressed, the label name for that key will be displayed to indicate that it is functioning properly. When the key is released, the message “No keys pressed” is displayed. Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT aborts the test and returns the instrument to the SELF-TEST MENU. Continue pressing EXIT to back out of the menu structure. 4-4 Troubleshooting Models 2510 and 2510-AT DISPLAY PATTERNS test The display test lets you verify that each pixel and annunciator in the vacuum fluorescent display is working properly. Perform the following steps to run the display test: 1. 2. 3. Display the MAIN MENU by pressing the MENU key. Select TEST, and press ENTER to display the SELF-TEST MENU. Select DISPLAY-TESTS, and press ENTER to display the following menu: FRONT PANEL TESTS 4. Select DISPLAY-PATTERNS, and press ENTER to start the display test. There are five parts to the display test. Each time a front panel key (except EXIT) is pressed, the next part of the test sequence is selected. The five parts of the test sequence are as follows: • Checkerboard pattern (alternate pixels on) and all annunciators. • Checkerboard pattern and the annunciators that are on during normal operation. • Horizontal lines (pixels) of the first digit are sequenced. • Vertical lines (pixels) of the first digit are sequenced. • Each digit (and adjacent annunciator) is sequenced. All the pixels of the selected digit are on. When finished, abort the display test by pressing EXIT. The instrument returns to the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu structure. KEYS DISPLAY-PATTERNS CHAR-SET 5. CHAR SET test The character set test lets you display all characters. Perform the following steps to run the character set test: 1. 2. 3. Display the MAIN MENU by pressing the MENU key. Select TEST, and press ENTER to display the SELF-TEST MENU. Select DISPLAY-TESTS, and press ENTER to display the following menu: FRONT PANEL TESTS 4. Select CHAR-SET, and press ENTER to start the character set test. Press any key except EXIT to cycle through all displayable characters. When finished, abort the character set test by pressing EXIT. The instrument returns to the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu structure. KEYS DISPLAY-PATTERNS CHAR-SET 5. Models 2510 and 2510-AT Troubleshooting 4-5 Principles of operation The following information is provided to support the troubleshooting tests and procedures covered in this section of the manual. Overall block diagram Figure 4-1 shows an overall block diagram of the Model 2510. Circuitry may be divided into three general areas: • • • Figure 4-1 Overall block diagram Analog circuits — includes source circuits such as the pulse-width modulator, H-drive, and class D output stages, as well as measurement circuits such as the sensor signal conditioning and A/D converter circuits. Digital circuits — includes the microcomputer that controls the analog section, front panel, and GPIB and RS-232 ports, as well as associated interfacing circuits. Power supplies — converts the AC line voltage into DC voltages that supply the power for the digital and analog circuits, and the class D power amplifier. Analog Section Pulse Width Modulator DAC Control Register H Drive Class D Output Stage Over I Limit TEC I/V Sensing Output to TEC Mux A/D Converter Display, Keyboard Front Panel Controller Sensor Signal Conditioning RS-232 Input from Temperature Sensor To Analog Circuits Trigger Link Trigger, Digital I/O GPIB Interface Digital Section To Digital Circuits ±15V +5V +15V +5V Analog Power Supply Output Stage Power Supply Digital Power Supply RS-232 I/O Microcomputer Digital I/O To Output Stage GPIB I/O Power Supply Line In 4-6 Troubleshooting Models 2510 and 2510-AT Analog circuits Figure 4-2 shows a block diagram of the analog circuits. These circuits are discussed in more detail in the following paragraphs. Figure 4-2 Block diagram of analog circuits DAC U127 Pulse Width Modulator U110 H Drive U107 Class D Output Stage Output Filter Q107Q110 L106 L107 L115 Source Control Register Over Current Limit TEC I/V Sensing U143 R102 U105 U132 U147 U151 Mux Sensor Signal Conditioning Measure Control Register U134U136 Digital Control From MPU U142 U124 U149 OUTPUT Terminals INPUT Terminals Measure Signal To A/D Converter Source circuits The Model 2510 source is a digitally controlled, class D power amplifier with the output stage in an H-bridge configuration. U127 is a 16-bit DAC that controls the pulse-width modulator U110. The H-drive IC, U107, switches the four output FETs, Q107 - Q110. Output filtering is provided by L106, L107, L115, and associated components. Control information for the DAC and H-drive circuit is fed through serial-to-parallel converter U143. The source amplitude voltage is controlled by varying the duty cycle of the pulse-width modulator and, ultimately, the output stage with the digital control information fed to the DAC and the serial-to-parallel converter. For example, with a duty cycle of 50%, the output is 2.5V. The hardware current limit circuit consists of R102, U105, U132, and associated components. The output stage current that flows through R102 develops a voltage that is amplified by U105 and compared with a programmed current limit in 10-bit DAC U132. If the current value exceeds the pre-programmed limit, the H-drive IC is disabled, turning off the output stage. Class D Amplifier Figure 4-3 shows a simplified schematic of the class D amplifier output stage. The H_DAC converts control information from the processor into an analog signal that controls the pulsewidth modulator (PWM) through the H-bridge control. The IP_DAC converts current limit control information from the processor and forms the BRIDGE OFF signal that turns off the output stage in an over-current condition. I_LIM 10K PWM_CLK F 10K + - 10K F 3.01K 0.1UF + BRIDGEOFF 2 470PF 1 0.1µf 30.1K 20K HIP4081 BRIDGEOFF PWM I_LIM 3 1.7K 500 F 0.02 +15V F F 1Ω Figure 4-3 Simplified schematic of class D amplifier IP_DAC H_DAC U118 F 1K Models 2510 and 2510-AT Troubleshooting 4-7 4-8 Troubleshooting Models 2510 and 2510-AT H-bridge Figure 4-4 shows a simplified schematic of the H-bridge output stage switching as well as the pulse-width modulator waveforms. The four output stage transistors function as simple switches with the switching phases and duty cycles controlled by the pulse-width modulator as shown. Figure 4-4 Simplified schematic of H-bridge output stage +15V AHI BHI DUT ALO BLO A. H-Bridge Switching ALO PWM Input Voltage Pulse Width Modulator BHI AHI BLO B. Pulse-Width Modulator Waveforms Models 2510 and 2510-AT Troubleshooting 4-9 Figure 4-5 shows alternate switching phases of the H-bridge output stage as well as a simplified schematic of the step-down “buck” regulator. Note that the bridge switches between alternate configurations as shown, with the duration and duty cycles controlled by the pulse-width modulator. The step-down “buck” regulator smooths the H-bridge switching waveform into a constant DC level. Figure 4-5 H-bridge switching and step-down regulator +15V +15V DUT DUT A. H-Bridge I B. Step-Down “Buck” Regulator 4-10 Troubleshooting Models 2510 and 2510-AT Measurement circuits Signal conditioning for the external temperature sensor is provided by U124, U149, and associated components. Voltage and current feedback signals from the TEC output terminals are developed by U147 and U151. The conditioned temperature sensor, voltage, and current, feedback signals are fed to the multiplexer IC, U142, that switches among the signals during the various phases of the measurement cycle. The output of the mux is fed to the A/D converter for conversion to digital data. Additional signals switched by the mux during the measurement cycle include a reference voltage, reference temperature, internal circuit board temperature signal, and an internal +15V measurement. Sensor conditioning Figure 4-6 shows a simplified schematic of the sensor conditioning circuits. In this example, measurement using a 10kΩ thermistor is assumed. Sensor conditioning for other thermistor ranges and sensor types is similar. Figure 4-6 Sensor conditioning for 10kΩ thermistor +15V 2 Wire THPFBK GND + DUT GND - 2 Wire SHUNT_10K 10K 100K MEAS_10K Models 2510 and 2510-AT Troubleshooting 4-11 A constant current is forced through the DUT (thermistor) as well as a reference resistor (in this example, 10kΩ for the 10kΩ range). As various phases of the measurement cycle, the voltages across the DUT and 10kΩ reference resistor are measured, and the actual DUT resistance is computed from the ratio of the two voltages. Since the resistance of the 10kΩ reference resistance is accurately known (as determined during instrument calibration), the DUT resistance and a given temperature can be accurately determined, and internal software converts the resistance into an equivalent temperature reading. A/D converter The Model 2510 unit uses a multi-slope charge balance A/D converter with a single-slope run-down. The converter is controlled by a gate array. Commands are issued by the MPU to the gate array, and the gate array IC sends A/D reading data back to the MPU for calibration and processing. PID hardware loop control Figure 4-7 shows an overall diagram of the hardware aspects of the PID control loop. Figure 4-7 PID hardware control loop HDAC “D” Amplifier U127 0.1Ω 68332 MUX ADC U142 ITHP VTHP REF ZERO THP_FBK PC_TEMP +15V MEAS TEMP_REF U147 U151 2Ω 4-12 Troubleshooting Models 2510 and 2510-AT As previously discussed, source control information from the 68332 processor is fed to the HDAC where it is converted into an analog signal to control the class D amplifier output. Power supply Figure 4-8 shows a block diagram of the Model 2510 power supply system. Figure 4-8 Power supply block diagram Analog Circuits +15VA +5VA Class D Power Amplifier A -15VA +15VPWR Digital Circuits Constant Frequency Low Noise Floating Switching Supply +15V DC D +5VD DC Regulator +5VF F Line In 100-240V AC PFC Power Supply Module DC Regulator The integrated switching power supply module provides all power for the instrument while providing universal inputs and power factor correction for the 120/240V line. The +15VPWR voltage from the module powers the output stage directly, and this supply voltage is further regulated to develop a +5VF (floating) supply voltage and the +5VD (digital) voltage that powers the digital circuits. A constant-frequency switching supply also runs off the +15V power module supply and generates the supply voltages for the analog circuits: +5VA and ±15FA. Models 2510 and 2510-AT Troubleshooting 4-13 Digital circuitry Refer to Figure 4-9 for the following discussion on digital circuitry. Figure 4-9 Digital circuitry overall block diagram ROM RAM U168 U160 Serial Interface U166 Reset RS-232 Interface E 2 PROM U141 A/D Control/Data GPIB Microprocessor U163 U167, U169 U170 IEEE-488 Interface A/D Interface U171 To Display Board Controller 16.78MHz Voltage Source Control Trigger Link U165 Digital I/O U164 Trigger Link Digital I/O The core digital circuitry uses a Motorola 68332 microcontroller (U163) running at 16.78MHz. The memory configuration includes a flash EEPROM (U168) and a RAM (U160). Flash ROM support allows internal firmware upgrades using either the serial or GPIB port for downloading new firmware. All calibration constants and the saved setups are stored in a separate serial EEPROM (U141). External communication is provided via GPIB and serial interfaces. A 9914 GPIB IEEE-488 standard interface IC (U167) is used for the GPIB, and a U166 provides the voltage conversion for the RS-232 port. U164 provides interfacing for the Digital I/O port. 4-14 Troubleshooting Models 2510 and 2510-AT Figure 4-10 shows a diagram for digital control that includes serial-to-parallel converters U134 to U136 and associated control signal nomenclatures. Figure 4-10 Digital control circuits A/D IN U136 MUXC MUXB MUXA 10_MA Range* N.C. SHUNT_100* SHUNT_1K SHUNT_10K U135 MEAS_DUT_HI* MEAS_DUT_LO* GND+* GND-* GAIN_4* GAIN_16* A2DGND* N.C. SDAT MUXSTB ADTS U163 68332 U145 ADTX U134 RNGSTB ADCLK ADRXB SCLKOUTN PWM_OUT PWM_CLK U116 OFFSTB U143 HCON IPDAC U132 HDAC U127 <8> <16> <16> MEAS_100K* MEAS_10K* MEAS_1K* MEAS_100* 2_WIRE* 1MA* 100UA* I_DIV_3 Models 2510 and 2510-AT Troubleshooting 4-15 Display board circuits U902 is the display microcontroller that controls the VFD (vacuum fluorescent display) and interprets key data. The microcontroller has four peripheral I/O ports that are used for the various control and read functions. Display data is serially transmitted to the microcontroller from the digital board via the TXB line to the microcontroller PD0 terminal. In a similar manner, key data is serially sent back to the mother board through the RXB line via PD1. The 4MHz clock for the microcontroller is generated on the digital board. DS901 is the VFD (vacuum fluorescent display) module, which can display up to 49 characters. Each character is organized as a 5 × 7 matrix of dots or pixels and includes a long underbar segment to act as a cursor. The display uses a common multiplexing scheme with each character refreshed in sequence. U903 and U904 are the grid drivers, and U901 and U905 are the dot drivers. Note that dot driver and grid driver data is serially transmitted from the microcontroller (PD3 and PC1). Troubleshooting Troubleshooting information for the various circuits is summarized below. Refer to the component layout drawings at the end of Section 6 for component locations. Display board checks If the front panel display tests indicate that there is a problem on the display board, use Table 4-1. See “Principles of operation” for display circuit theory. Table 4-1 Display board checks Step Item/component Required condition 1 2 3 4 5 6 Front panel test J1033 U902, pin 1 U902, pin 43 U902, pin32 U902, pin 33 Verify that all segments operate. +5V, ±5% Goes low briefly on power up, and then goes high. 4MHz square wave. Pulse train every 1 ms. Brief pulse train when front panel key is pressed. Remarks Use front panel display test. Digital +5V supply. Microcontroller RESET. Controller 4MHz clock. Control from main processor. Key down data sent to main processor. 4-16 Troubleshooting Models 2510 and 2510-AT Power supply checks Power supply problems can be checked out using Table 4-2. See “Principles of operation” for circuit theory on the power supply. Table 4-2 Power supply checks Step Item/component Required condition 1 2 3 4 5 6 7 8 9 Line fuse Line power J1003, pin 3 TP102 TP112 TP101 TP105 TP106 TP107 Check continuity. Plugged into live receptacle, power on. +15V, ±10% (+15VPWR) +5V, ±5% (+5VF) +5V, ±5% (+5VD) +15V, ±5% (+15VA) -15V, ±5% (-15VA) +5V, ±5% (+5VA) +5V, ±5% (+5VRF) Remarks Remove to check. Check for correct power-up sequence. Referenced to TP121. Referenced to TP121. Referenced to U163, pin 19. Referenced to TP122. Referenced to TP122. Referenced to U104, pin 2. Referenced to U104, pin 2. Digital circuitry checks Digital circuit problems can be checked out using Table 4-3. See “Principles of operation” for a digital circuit description. Table 4-3 Digital circuitry checks Step Item/component Required condition 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 Power-on test U163 pin 19 U163 pin 7 U163 pin 68 U163, A0-A19 U163, D0-D15 U163 pin 66 U166 pin 7 U166 pin 8 U167 pins 34-42 U167 pins 26-31 U167 pin 24 U167 pin 25 U163 pin 43 U163 pin 44 U163 pin 45 U163 pin 47 RAM OK, ROM OK. Digital common. common. +5V Low on power-up, then goes high. Check for stuck bits. Check for stuck bits. 16.78MHz. Pulse train during RS-232 I/O. Pulse train during RS-232 I/O. Pulse train during IEEE-488 I/O. Pulses during IEEE-488 I/O. Low with remote enabled. Low during interface clear. Pulse train. Pulse train. Pulse train. Pulse train Remarks Verify that RAM and ROM are functional. All signals referenced to digital common. Digital logic supply. MPU RESET line. MPU address bus. MPU data bus. MPU clock. RS-232 TX line. RS-232 RX line. IEEE-488 data bus. IEEE-488 command lines. IEEE-488 REN line. IEEE-488 IFC line. D_ADDATA D_DATA D_CLK D_STB Models 2510 and 2510-AT Troubleshooting 4-17 Analog circuitry checks Table 4-4 summarizes analog circuitry checks. NOTE All measurements are with the V function selected, +5V output voltage, 5A current limit, and 1Ω and 100Ω resistors connected to the OUTPUT and INPUT terminals respectively. Turn OUTPUT ON while measuring. Table 4-4 Analog circuitry checks Step Item/component1 Required condition 1 2 3 4 5 6 7 8 9 TP1031 TP1041 TP1082 TP1092 TP1102 TP1142 TP1182 TP1192 TP1202 +10.4V +4V +6.3V -12.5V +12.5V 0V Pulse train 0.5V 5V Remarks H drive signal H drive signal A/D reference signal -14V nominal supply +14V nominal supply Sensor feedback signal Mux output to A/D converter OUTPUT current sense signal OUTPUT voltage sense signal 6. Measured with respect to F common (TP121). 7. Measured with respect to A2 common (TP122). Internal fuse replacement WARNING Disconnect the line cord and all cables and test leads from the instrument before replacing internal fuses. CAUTION Do not install a fuse with a higher current rating than specified, or instrument damage may occur. Power supply module fuse replacement An internal fuse protects the power supply module from over-current conditions. Replace this fuse as follows: 1. 2. 3. 4. Turn off the power, and disconnect the line cord and all other test leads and cables from the instrument. Remove the case cover and mother board, as covered in Section 5. Locate the fuse on the power supply module circuit board. The module is mounted on the chassis bottom. Replace the fuse with one with the same current and voltage rating marked on the power supply module circuit board. 4-18 Troubleshooting Models 2510 and 2510-AT Digital I/O +5V supply fuse replacement An internal fuse protects the +5V supply line on the ENABLE-DIG I/O connector from over-current conditions. Replace this fuse as follows: 1. 2. 3. 4. 5. 6. Turn off the power, and disconnect the line cord and all other test leads and cables from the instrument. Remove the case cover, as covered in Section 5. Locate fuse F101 on the mother board. This fuse is located near the ENABLE-DIG I/O/ RS-232 connector assembly (J1007) at the back of the board. See the 2510-100 component layout drawing at the end of Section 6 for location. Carefully unsolder the blown fuse. Replace the fuse with one with the following rating: 0.6A, Keithley part number FU-103, then solder it in place. After soldering, clean the mother board as covered in Section 5, “Handling and cleaning precautions.” No comm link error A “No Comm Link” error indicates that the front panel processor has stopped communicating with the main processor, which is located on the mother board. This error indicates that the main processor ROM (U168) may require reseating in its socket. The ROM may be reseated as follows: 1. 2. 3. 4. Turn off the power, and disconnect the line cord and all other test leads and cables from the instrument. Remove the case cover as outlined in Section 5. Locate the firmware ROM, U168, located on the digital board. This ROM is the only IC installed in a socket. (Refer to the 2510-100 component layout drawing at the end of Section 6 for exact location.) Carefully push down on the ROM IC to make sure it is properly seated in its socket. CAUTION 5. Be careful not to push down excessively, or you might crack the mother board. Connect the line cord, and turn on the power. If the problem persists, additional troubleshooting will be required. 5 Disassembly 5-2 Disassembly Models 2510 and 2510-AT Introduction This section explains how to handle, clean, and disassemble the Model 2510. Disassembly drawings are located at the end of this section. Handling and cleaning To avoid contaminating PC board traces with body oil or other foreign matter, avoid touching the PC board traces while you are repairing the instrument. Mother board areas covered by the shield have high-impedance devices or sensitive circuitry where contamination could cause degraded performance. Handling PC boards Observe the following precautions when handling PC boards: • • • • • Wear cotton gloves. Only handle PC boards by the edges and shields. Do not touch any board traces or components not associated with repair. Do not touch areas adjacent to electrical contacts. Use dry nitrogen gas to clean dust off PC boards. Solder repairs Observe the following precautions when you must solder a circuit board: • • • • Use an OA-based (organic activated) flux, and take care not to spread the flux to other areas of the circuit board. Remove the flux from the work area when you have finished the repair by using pure water with clean, foam-tipped swabs or a clean, soft brush. Once you have removed the flux, swab only the repair area with methanol, then blowdry the board with dry nitrogen gas. After cleaning, allow the board to dry in a 50°C, low-humidity environment for several hours. Models 2510 and 2510-AT Disassembly 5-3 Static sensitive devices CMOS devices operate at very high impedance levels. Therefore, any static that builds up on you or your clothing may be sufficient to destroy these devices if they are not handled properly. Use the following precautions to avoid damaging them: CAUTION • • • • • • • Many CMOS devices are installed in the Model 2510. Handle all semiconductor devices as being static sensitive. Transport and handle ICs only in containers specially designed to prevent static buildup. Typically, you will receive these parts in anti-static containers made of plastic or foam. Keep these devices in their original containers until ready for installation. Remove the devices from their protective containers only at a properly grounded work station. Ground yourself with a suitable wrist strap. Handle the devices only by the body; do not touch the pins. Ground any printed circuit board into which a semiconductor device is to be inserted to the bench or table. Use only anti-static type desoldering tools. Use only grounded-tip solder irons. Once the device is installed in the PC board, it is normally adequately protected, and you can handle the boards normally. Assembly drawings Use the assembly drawings located at the end of this section to assist you as you disassemble and reassemble the Model 2510. Refer to these drawings for information about the Keithley part numbers of most mechanical parts in the unit. Assembly drawings include: • • • • • Front panel assembly — 2510-040 Chassis/power module assembly — 2510-050 Front panel/chassis assembly — 2510-051 Chassis assembly — 2510-052 Final inspection — 2510-080 5-4 Disassembly Models 2510 and 2510-AT Case cover removal Follow the steps below to remove the case cover to gain access to internal parts. WARNING 1. 2. NOTE 3. 4. 5. Before removing the case cover, disconnect the line cord and any test leads from the instrument. Remove handle — The handle serves as an adjustable tilt-bail. Adjust its position by gently pulling it away from the sides of the instrument case and swinging it up or down. To remove the handle, swing the handle below the bottom surface of the case and back until the orientation arrows on the handles line up with the orientation arrows on the mounting ears. With the arrows lined up, pull the ends of the handle away from the case. Remove mounting ears — Remove the screw that secures each mounting ear. Pull down and out on each mounting ear. When reinstalling the mounting ears, make sure to mount the right ear to the right side of the chassis, and the left ear to the left side of the chassis. Each ear is marked “RIGHT” or “LEFT” on its inside surface. Remove rear bezel — To remove the rear bezel, loosen the two screws that secure the rear bezel to the chassis, then pull the bezel away from the case. Remove bottom screws — Remove the four screws that secure the case to the chassis. They are located on the bottom of the case. Remove chassis — To remove the case, grasp the front bezel of the instrument, and carefully slide the chassis forward. Slide the chassis out of the metal case. Mother board removal Perform the following steps to remove the mother board. This procedure assumes that the case cover is already removed. 1. 2. 3. Remove the IEEE-488, ENABLE-DIG I/O, and RS-232 fasteners. The IEEE-488, ENABLE-DIG I/O, and RS-232 connectors each have two hex-head screws that secure the connectors to the rear panel. Remove these screws. Remove mother board mounting screws. Remove the two mounting screws that secure the mother board to the chassis. Unplug cables: • Unplug the display board ribbon cable from J1014. • Unplug the cables going to the power supply from J1003. • Unplug the cable going to the OUTPUT indicator from J102. • Unplug the fan cable from J1015. Models 2510 and 2510-AT 4. Disassembly 5-5 Remove mother board. Slide the mother board forward until the slots line up with the guide pins, then remove the board. During reassembly, replace the mother board, and start the IEEE-488, ENABLE-DIG I/O, and RS-232 connector screws and the board mounting screws. Tighten all the fasteners once they are all in place and the board is correctly aligned. Be sure to plug in all cables. Front panel disassembly Use the following procedure to remove the display board and/or the pushbutton switch pad. 1. 2. 3. 4. 5. Remove the power switch rod. Carefully disconnect the power switch rod from the power switch mounted on the rear panel power module. Slide the rod toward the rear until it clears the access hole in the front panel, then remove the rod. Remove the front panel assembly. This assembly has four retaining clips that snap onto the chassis over four pem nut studs. Two retaining clips are located on each side of the front panel. Pull the retaining clips outward and, at the same time, pull the front panel assembly forward until it separates from the chassis. Unplug the display board ribbon cables. Using a thin-bladed screw driver, pry the plastic PC board stop (located at the bottom of the display board) until the bar separates from the casing. Pull the display board from the front panel. Remove the switch pad by pulling it from the front panel. Removing power components The following procedures to remove the power supply and/or power module require that the case cover and mother board be removed, as previously explained. Power supply module removal Perform the following steps to remove the power supply module: 1. 2. 3. Remove the wires that connect the power supply module to the rear panel power module. Remove any cable clamps or cable ties that secure the power supply module wires. Remove the screws that secure the power supply to the chassis bottom, then remove the module. 5-6 Disassembly Models 2510 and 2510-AT Power module removal Perform the following steps to remove the rear panel power module: 1. 2. Disconnect the power module's ground wire. This green and yellow wire connects to a threaded stud on the chassis with a kep nut. Squeeze the latches on either side of the power module while pushing the module from the access hole. Fan removal 1. 2. Remove the two nuts that secure the fan to the rear of the chassis. Remove the fan from the chassis. Instrument reassembly Reassemble the instrument by reversing the previous disassembly procedures. Make sure that all parts are properly seated and secured, and that all connections are properly made. WARNING To ensure continued protection against electrical shock, verify that power line ground (green and yellow wire attached to the power module) is connected to the chassis. Also make sure the four bottom case screws are properly installed to secure and ground the case cover to the chassis. 6 Replaceable Parts 6-2 Replaceable Parts Models 2510 and 2510-AT Introduction This section contains replacement parts information and component layout drawings for the Model 2510. Parts lists The electrical parts lists for the Model 2510 are shown in the tables at the end of this section. For part numbers to the various mechanical parts and assemblies, use the Miscellaneous parts list and the assembly drawings provided at the end of Section 5. Ordering information To place an order, or to obtain information concerning replacement parts, contact your Keithley representative or the factory (see inside front cover for addresses). When ordering parts, be sure to include the following information: • • • • • Instrument model number (Model 2510) Instrument serial number Part description Component designation (if applicable) Keithley part number Factory service If the instrument is to be returned to Keithley Instruments for repair, perform the following: • • • • Call the Repair Department at 1-800-552-1115 for a Return Material Authorization (RMA) number. Complete the service form at the back of this manual, and include it with the instrument. Carefully pack the instrument in the original packing carton. Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label. Component layouts The component layouts for the circuit boards are provided on the following pages. Drawings include: • • Mother board — 2510-100 Display board — 2400-110 Models 2510 and 2510-AT Replaceable Parts 6-3 Table 6-1 Mother board parts list Circuit designation Description Keithley part no. C101,C102,C204,C205,C206,C207 CAP, 680U, 20%, 50V ALUM ELEC C-578-680 C103,C106 CAP, 2.2U, 10%, 50V TANTALUM C-563-2.2 C104,105,118,121,122,125,126,127,157,158 CAP, .1UF, 20%, 50V CERAMIC C-418-.1 C107,109,112,114,135,136,143,152,155,161 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 C110,C113 CAP, .033U, 10%, 50V CERAMIC C-491-.033 C111,C128,C188,C189,C192,C194,C290 CAP, 1UF, 20%, 35V, TANTALUM C-494-1 C115,C175 CAP, 2200P, 1%, 50V CERAMIC C-532-2200P C116,C174 CAP, 220PF, 10%, 100V, CERAMIC C-451-220P C119,C172,C185 CAP, 100PF, 5%, 100V CERAMIC C-465-100P C120 CAP, 470P, 10%, 100V, CERAMIC C-451-470P C123,131,132,133,134,137,138,141,142,146 CAP, 22UF, 20%, 25V, TANTALUM C-440-22 C130 CAP, 100UF, 20%, 10V, ALUM ELEC C-483-100 C139,C140,C144,C145 CAP, 47U, 20%, 50V, ALUM C-579-47 C151,156,178,187,199,210,236,150 CAP, 22UF, 20%, 25V, TANTALUM C-440-22 C162,163,167,170,171,181,182,196,198,200 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 C164,C184 CAP, 2200P, 10%, 100V CERAMIC C-430-2200P C165,166,186,193,220,225,226,229,230,231 CAP, .01UF, 10%, 50V CERAMIC C-491-.01 C168,169,183,195,197,250,268 CAP, 47P, 5%, 100V, CERAMIC C-465-47P C173,251,266,117,147,297-305 CAP, 1000PF,10%, 50V, MONO CERAMIC C-452-1000P C176,177,180,208,209,214,215,216,253,271 CAP, .1UF, 20%, 50V CERAMIC C-418-.1 C179,C247,C260,C265,C275,C292 CAP, 100P, 10%, 100V CERAMIC C-451-100P C201,202,203,211,212,213,219,221,222 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 C223 CAP, 47PF, 10%, 100V, CERAMIC C-451-47P C224,C228 CAP, 15P, 1%, 100V CERAMIC C-512-15P C227,233,234,237,238,239,240,241,243 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 C232,C235,C242,C245,C282 CAP, .01UF, 10%, 50V CERAMIC C-491-.01 C246,248,249,252,254,255,256,261,267,269 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 C270,C295,C218,C296 CAP, .1UF, 20%, 50V CERAMIC C-418-.1 C272,C273 CAP, 1UF, 20%, 100V, CERAMIC C-487-1 C277,278,280,281,283-289,291,293,294 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 C279 CAP, .1UF, 20%,100V, CERAMIC C-436-.1 CR101,CR113 DIODE, MBRS140T3 RF-110 6-4 Replaceable Parts Models 2510 and 2510-AT Table 6-1 (cont.) Mother board parts list Circuit designation Description Keithley part no. CR102 ULTRA FAST BRIDGE RECTIFIER, EDF1BM RF-123 CR103,CR104,CR105,CR106 ULTRAFAST POWER RECTIFIER RF-107 CR107,CR108,CR130 DIODE, SWITCHING, MMBD914 RF-83 CR109,CR110,CR111,CR112,CR133 DIODE, SWITCHING, MMSD914T19 RF-112 CR114 DIODE, DUAL HSM-2822T31 RF-95 CR115,CR116,CR122-CR129,CR131,CR132 DIODE, DUAL SWITCHING, BAV99L RF-82 CR117,CR118,CR119,CR120,CR121 DIODE, DUAL COMMON ANODE BAW56LT2 RF-98 F101 POLYSWITCH, SMD030-2 FU-103 HS101,HS105-HS110 HEAT SINK HS-55 J1001 CONNECTOR, MODULES CS-834 J1003 CONNECTOR, HEADER CS-784-4 J1006 CONN, MICRODIN W/GND FINGERS CS-792 J1007 CONN, D-SUB DUAL STACK M-F CS-1072-1 J1008 CONN, RIGHT ANGLE, 24 PIN CS-501 J101 CONN,BERG CS-339 J1014 CONN, HEADER STRAIGHT SOLDER PIN CS-368-16 J102,J1015 LATCHING HEADER,FRICTON, SGL ROW CS-724-3 K101 N.C RELAY, 1 FORMB, AQV214 RL-176 K102 MINI POWER RELAY, DK1A-5V RL-249 L101,L105 CHOKE, 22UH CH-66-22 L102,L103,L104 FERRITE BEAD CH-91 L106,L107,L116 HIGH CURRENT TOROID CH-103-1 L108,L109,L110,L111,L112,L113,L114,L118 FERRITE CHIP 600 OHM BLM32A07 CH-62 L115 EMI SUPPRESSION INDUCTOR CH-84 L117,L120 CHOKE, EMI CH-57 L119,L121,L122,L123,L124 FERRITE CHIP 600 OHM BLM32A07 CH-62 L125 CHOKE, 100MHZ CH-50 PS101 POWER SUPPLY DC-DC CONVERTER PS-76-1A Q101,Q102,Q104,Q119 TRANS, NPN, MMBT3904 TG-238 Q103 TRANS, PNP, MMBT3906L TG-244 Q105,Q106 TRANS, N CHANNEL MOSFET, BUZ11 TG-211 Q107,Q108,Q109,Q110 HEXFET POWER MOSFET, IRFZ44N TG-354 Models 2510 and 2510-AT Replaceable Parts 6-5 Table 6-1 (cont.) Mother board parts list Circuit designation Description Keithley part no. Q111,Q116,Q117 TRANS, P-CHAN, MOSFET, TP0610T TG-259 Q112,Q113,Q114,Q115,Q120,Q121,Q122,Q123 TRANS, N-MOSFET, VN0605T TG-243 Q124,Q125,Q126,Q127 TRANS, N-MOSFET, VN0605T TG-243 R101,R103,R104,R105,R188 RES, 10, 5%, 125MW METAL FILM R-375-10 R102 RES, .02, 1.5W, 1% R-468-.02 R107,R198,R200 RES, 20K, 1%, 100MW, THICK FILM R-418-20K R112,R113 RES, 200K, 1%, 125MW, METAL FILM R-391-200K R114,115,106,123,208,135,136,130,157,143 RES, 1K, 1%, 100MW, THICK FILM R-418-1K R116,R117,R150,R161,R162,R199,R209,R221 RES, 475, 1%, 100MW, THICK FILM R-418-475 R119,R118,R141 RES, 30.1K, 1%, 100MW, THICK FILM R-418-30.1K R120,R142 RES, 24.3, 1%, 100MW, THICK FILM R-418-24.3 R121 RES, 3.01K, 1%, 100MW, THICK FILM R-418-3.01K R122 RES, 1.74K, 1%, 100MW, THICK FILM R-418-1.74K R124,R126,R307 RES, .0499, 1%, 100MW, THICK FILM R-418-.0499 R125,127,131,165,151,152,168,169,171 RES, 10K, 1%, 100MW, THICK FILM R-418-10K R132,R311 RES, 7.5K, 1%, 100MW, THICK FILM R-418-7.5K R133 RES, 51, 5%, 125MW, METAL FILM R-375-51 R134,R137,R201,R306 RES, 200, 1%, 100MW, THICK FILM R-418-200 R138,R139,R316 RES, 5.11K, 1%, 100MW, THICK FILM R-418-5.11K R146,147,167,177,178,202,216,217,218,219 RES, 1K, 1%, 100MW, THICK FILM R-418-1K R148,R149,R255,R256,R257,R279,R284,R140 RES, 2.21K, 1%, 100MW, THICK FILM R-418-2.21K R153 RES, 49.9K, 1%, 100MW, THICK FILM R-418-49.9K R154 RES, 1.28M, .1%, 1/8W, METAL FILM R-176-1.28M R159 RES, 4.99K, 1%, 125mW, METAL FILM R-391-4.99K R160 RES, 100K, 1%, 100MW, THICK FILM R-418-100K R163 RES, 2.49K, 1%, 125MW, METAL FILM R-391-2.49K R164 RES, 20K, .1%, 1/10W, METAL FILM R-263-20K R170 RES, 2.49K, .1%, .125MW, THIN FILM R-456-2.49K R172 RES, 10K, .1%, .125W, THIN FILM R-456-10K R173 RES, 249K, .1%, .125W, THIN FILM R-456-249K R174 RES, 27.7K, .1%, .125W, THIN FILM R-456-27.7K R175 RES, 10K, .1%, 1/10W, METAL FILM R-263-10K 6-6 Replaceable Parts Models 2510 and 2510-AT Table 6-1 (cont.) Mother board parts list Circuit designation Description Keithley part no. R176 RES, 1K, .1%, 1/10W, METAL FILM R-263-1K R179 RES, 100K, .1%, 1/10W, METAL FILM R-263-100K R180 RES, 34K, 1%, 100MW, THICK FILM R-418-34K R181,R248,R254,R270 RES, 4.75K, 1%, 100MW, THICK FILM R-418-4.75K R182 RES, 82.5, 1%, 100MW, THICK FILM R-418-82.5 R183,R293,R295,R297,R299,R301,R303 RES, 5.11K, 1%, 100MW, THICK FILM R-418-5.11K R184 RES, 470,5%, 125MW, METAL FILM R-375-470 R185,189,239,294,296,298,300,302,304 RES, 100, 1%, 100MW, THICK FILM R-418-100 R186 RES, 6.65K, .1%, 1/10W, METAL FILM R-263-6.65K R187 RES NET TF-245 R190,R191,R192,R193,R194,R195,R196,R197 RES, 5.11K, 1%, 100MW, THICK FILM R-418-5.11K R203 RES, .1, 1%, 3W R-475-.1 R210,R305,R313,R314,R315,R320 RES, 10K, 1%, 100MW, THICK FILM R-418-10K R211,R215,R252 RES, 10M, 1%, 125MW, THICK FILM R-418-10M R212 RES, 18K, .1%, 1/10W, METAL FILM R-263-18K R213,235,236,241-247,249-251,253,263-266 RES, 10K, 1%, 100MW, THICK FILM R-418-10K R214 RES, 3.2K, .1%, 1/10W, METAL FILM R-263-3.2K R220,222,230,231,258-262,267,272,280,281 RES, 1K, 1%, 100MW, THICK FILM R-418-1K R223,R224,R225,R226,R227,R228,R229,R232 RES, 475, 1%, 100MW, THICK FILM R-418-475 R233,R234,R237 RES, 475, 1%, 100MW, THICK FILM R-418-475 R238 RES, 15k, 1%, 100MW, THICK FILM R-418-15K R240 RES, 332K, 1%, 100MW, THICK FILM R-418-332K R268,R269,R271,R274-277,282,283,285-290 RES, 10K, 1%, 100MW, THICK FILM R-418-10K R273 RES, 100, 5%, 250MW, METAL FILM R-376-100 R278 RES, 100, .1%, 1/10W, METAL FILM R-263-100 R291,292,312,R158 RES, 1K, 1%, 100MW, THICK FILM R-418-1K R308 RES, .1%, 1/10W, METAL FILM R-263-229 R309 RES, 56.2, .1%, 1/10W, METAL FILM R-263-56.2 RV101 TRANSIENT VOLTAGE SUPPRESSOR VR-25 SO-168 SOCKET PLCC-032-T-A SO-143-32 T101 TRANSFORMER TR-315A TP103,TP104,TP111-TP113,TP118-TP120 SURFACE MOUNT PCB TEST POINT CS-1026 Models 2510 and 2510-AT Replaceable Parts 6-7 Table 6-1 (cont.) Mother board parts list Circuit designation Description Keithley part no. TP121,TP122 SURFACE MOUNT PCB TEST POINT CS-1026 U101 IC, +5V VOLTAGE REGULATOR, LM2940CT IC-576 U102 IC, VOLTAGE REGULATOR IC-1133 U103 IC, -15V VOLTAGE REGULATOR IC-1135 U104 IC, VOLTAGE REGULATOR IC-1132 U105,U119 IC, DUAL FET OPAMP IC-1128 U107 IC, 80V 2.5A FULL BRIDGE FET DRIVER IC-1139 U108,U109,U123 IC, 2 - INPUT AND GATE IC-1140 U110 IC, HIGH SPEED PWM CONTROLLER IC-1119 U113 IC, SUPPLY VOLT SUPERVISOR,TL7705A IC-860 U114 IC, AJD SHUNT REGULATOR, TL431CLP IC-677 U115,U150 IC, SINGLE TRANSISTOR OPTO COUPLER IC-1182 U116 IC, HIGH SPEED PWN CONTROLLER IC-1120 U118 IC, VOLT COMPARATOR LM393D IC-775 U120,U124 IC, OPA177GS IC-960 U121 IC, PRECISION BIFET OPAMP IC-1194 U122 IC, DUAL BIPOLAR OP-AMP, LT1124CS8 IC-955 U125,U172 IC, VOLTAGE REFERENCE IC-1065 U126,U130,U131,U152,U153 IC, CMOS ANAL SWITCH, DG444DY IC-866 U127 IC, +5V 16 BIT DAC, MAX542ACSD IC-1176 U128 IC, VOLT. COMPARATOR, LM311M IC-776 U129 IC, OP-AMP, NE5534D IC-802 U132 IC, +5V 10 BIT DAC, MAX515CSA IC-1331 U133 IC, NCHAN LAT DMOS QUADFET, SD5400CY IC-893 U134,U135,U136,U143 IC, 8 STAGE SHIFT C074HC409AM IC-1026 U138 IC, QUAD D FLIP FLOP W/CLK, RESET 74HC175 IC-923 U139,U140 IC, QUAD 2 IN NOR, 74HCT02 IC-809 U141 IC, SERIAL EPROM 24LC16B LSI-153 U142 IC, 8-CHAN ANA MULTIPLEXER, DG408DY IC-844 U145 PROGRAM 2000-802A02 U146 IC, DUAL D-TYPE F/F, 74HC74 IC-773 6-8 Replaceable Parts Models 2510 and 2510-AT Table 6-1 (cont.) Mother board parts list Circuit designation Description Keithley part no. U147,U151 IC, DIFFERENTIAL AMP, INA117P IC-889 U148 IC, OP-AMP, LOW NOISE LT1007CS8 IC-949 U149 IC, OPAMP IC-1357 U155,U158,U154 IC, HCPL0631, PACK IC-1153 U156,U157 IC, HIGH SPEED OPTO COUPLER IC-1225 U159,U165 IC, POS NAND GATES/INVERT, 74HCT14 IC-656 U160 IC, 64KX16 BIT CMOS STATIC RAM LSI-215-1 U161 IC, 2 INPUT EXCLUSIVE OR GATE, NC7SZ86 IC-1180 U163 IC, MICROCONTROLLER MC68332-FC LSI-161 U164 IC, PROTECTED QUAD POWER DRIVERS IC-1212 U166 IC, +5V RS-232 TRANSCEIVER, MAX202 IC-952 U167 IC, GPIB ADAPTER, 9914A LSI-123 U168 PROGRAMMED ROM 2510-800* U169 IC, OCTAL INTER BUS TRANS,75161 IC-647 U170 IC, OCTAL INTERFACE BUS, 75160 IC-646 U171,U137 IC, 2-INPUT OR GATE IC-1206 U173 IC, CURRENT SOURCE, LM334H IC-421 U174 IC, CENTIGRADE TEMP SENSOR, LM35DZ IC-933 VR103,VR104 DIODE, ZENER, 6.2V MMSZ6V2 DZ-97 VR105 DIODE DZ-127 VR106 DIODE, ZENER 30V BZX84C30 DZ-106-30 W101,W102 JUMPER, .300 J-7-1 Y101 OSCILLATOR HIGH SPEED CMOS 12MHZ CR-37 Y102 CRYSTAL, FSM327 CR-41 *Order current firmware revision, for example 2510-800A01. Models 2510 and 2510-AT Replaceable Parts Table 6-2 Display board parts list Circuit designation Description Keithley part no. C901 CAP, 22UF, 20%, 6.3 TANTALUM C-417-22 C902,C904,C907,C908,C910 CAP, .1UF, 20%,100V, CERAMIC C-436-.1 C903,C905,C906,C909,C911 CAP, .1UF, 20%, 50V, CERAMIC C-418-.1 C912 CAP, 2.2UF, 20%,100V, ALUM ELEC C-503-2.2 C913,C914 CAP, 100UF, 20%, 16V, TANTALUM C-504-100 C915,C916 CAP, 33PF, 10%, 100V, CERAMIC C-451-33P CR901,CR902,CR903,CR904 DIODE, SWITCHING, 250MA,BAV103 RF-89 CR905,CR906 DIODE, SWITCHING, MMBD914 RF-83 DS901 VACUUM FLUORESCENT DISPLAY DD-51C J1032 CONN, BERG CS-339 J1033 CONN, HEADER STRAIGHT SOLDER PIN CS-368-16 Q901,Q902 TRANS, NPN GEN PURPOSE BC868 TG-293 R901 RES NET, 15K, 2%, 1.875W TF-219-15K R902 RES, 13K, 5%,125MW, METAL FILM R-375-13K R903,R904 RES, 4.7K, 5%, 250MW, METAL FILM R-376-4.7K R905 RES, 1M, 5%, 125MW, METAL FILM R-375-1M R906 RES, 1K, 5%, 250MW, METAL FILM R-376-1K R907 RES, 240, 5%, 250MW, METAL FILM R-376-240 R908 RES, 10M, 5%, 125MW, METAL FILM R-375-10M T901 TRANSFORMER, TDK, ER14.5 SERIES TR-300 U901,U904,U905 IC, LATCHED DRIVERS, UCN-5812EPF-1 IC-732 U902 PROGRAMMED ROM 7001-800* U903 IC, 32-BIT, SERIAL UCN5818EPF-1 IC-830 VR901 DIODE, ZENER, 8.2V, MMBZ5237 DZ-92 Y901 CRYSTAL, 4MHZ CR-36-4M *Order current revision level, for example 7001-800A02. 6-9 6-10 Replaceable Parts Models 2510 and 2510-AT Table 6-3 Miscellaneous parts list Qty. Description Keithley part no. 1 BEZEL, REAR 428-303D 1 CHASSIS 2510-301A 1 CONDUCTIVE RUBBER SWITCH 2510-315A 1 COVER 2510-307A 1 DISPLAY LENS 2510-311B 1 FAN FN-39-1 2 FOOT 428-319A 2 FOOT, EXTRUDED FE-22A 2 FOOT,RUBBER FE-6 1 FUSE HOLDER FH-39 1 FUSE, EXP. 2.5/5X20MM FU-106-2.5 1 HANDLE 428-329F 1 LENS, LED 6517-309B 1 LINE CORD CO-7 1 LINE MODULE PM-1-1B 1 MATING INPUT/OUTPUT CONNECTOR CS-846 1 MEMBRANE SWITCH, FRONT PANEL 6430-313A 1 MODIFIED FRONT PANEL 2510-308A 1 MOUNTING EAR, LEFT 428-338B 1 MOUNTING EAR, RIGHT 428-328E 1 OVERLAY FRONT PANEL 2510-309A 2 PC BOARD STOP 2001-371A 1 POWER ROD 2001-320A 1 POWER SUPPLY PS-77-1A 1 REAR PANEL 2510-303A A Specifications PURCHASED ITEM 2510 TEC SourceMeter The Model 2510 Thermoelectric Cooler Controller is designed to: • control the power to the TEC to maintain a constant temperature, current, voltage, or thermistor resistance • measure the resistance of the TEC • software PID loop TEC OUTPUT SPECIFICATIONS OUTPUT RANGE: ±10 VDC at up to ±5 ADC. OUTPUT RIPPLE: <5mV rms.5 AC RESISTANCE EXCITATION: ±(9.6mA + 190µA).10, 11 CONTROL SYSTEM SPECIFICATIONS TEC MEASUREMENT SPECIFICATIONS SET: Constant Peltier Temperature Constant Peltier Voltage Constant Peltier Current Constant Thermistor Resistance CONTROL METHOD: Programmable software PID loop. Proportional, Integral, and Derivative gains independently programmable. SETPOINT SHORT TERM STABILITY: ±0.005°C rms.2, 3 SETPOINT LONG TERM STABILITY: ±0.01°C.2, 4 SETPOINT RANGE: –50°C to 225°C. OVER TEMPERATURE LIMIT: 250°C max. UNDER TEMPERATURE LIMIT: –50°C max. SETPOINT RESOLUTION: 0.001°C, 1mV, 100µA, 0.01% of nominal (25°C) thermistor resistance. HARDWARE CURRENT LIMIT: 1.0A to 5.25A ±5%. SOFTWARE VOLTAGE LIMIT:±0.5 to 10.5V ±5%. FUNCTION Operating Resistance 1, 6, 7, 8 Operating Voltage 1, 6 Operating Current 6 AC Resistance 1, 13 1 Year, 23°C ±5°C ±(2.0% of rdg + 0.1Ω) ±(0.1% of rdg + 4mV) ±(0.4% of rdg + 8mA) ±(0.10% of rdg + 0.02Ω) OPEN SHORTED THERMOELECTRIC DETECTION LOAD IMPEDANCE: Stable into 1µF typical. COMMON MODE VOLTAGE: 30VDC maximum. COMMON MODE ISOLATION: >109Ω, <1500pF. MAX. SENSE LEAD RESISTANCE: 1Ω for rated accuracy. MAX. FORCE LEAD RESISTANCE: 0.1Ω. THERMAL FEEDBACK ELEMENT SPECIFICATIONS (1 Year, 23°C ±5°C) Sensor Type RTD Excitation 9 Compliance Nominal Resistance Range Excitation Accuracy Nominal Sensor Temperature Range Sensor Coefficients Measurement Accuracy ±(% rdg + offset) Thermistor 100 Ω 2.50 mA 1 kΩ 833 µA 100 Ω 2.5 mA 1 kΩ 833 µA 10 kΩ 100 µA 0–250 Ω ±2.9% 0-2.50 kΩ ±2.9% 0–1 kΩ ±2.9% 0–10 kΩ ±2.9% 0–80 kΩ ±2.9% 100 kΩ 33 µA 833 µA max 0–200 kΩ ±2.9% –50° to +250°C A, B, C –50° to +250°C A, B, C 0.04 + 0.4 Ω1 0.02 + 3 Ω1 –50° to +250°C α, β, δ –50° to +250°C –50° to +250°C α, β, δ A, B, C 0.04 + 0.07 Ω 0.04 + 0.4 Ω 0.04 + 0.07 Ω1 THERMISTOR MEASUREMENT ACCURACY14 Nominal Thermistor Resistance 100 Ω 1 kΩ 10 kΩ 100 kΩ 25°C 0.035°C 0.023°C 0.012°C 0.014°C 50°C 0.070°C 0.045°C 0.026°C 0.026°C BRUNING 40-21 62198-SBG A B REVISIONS 24052 REL 27123 REV –50° to +250°C A, B, C –40° to +100°C Slope & offset –40° to +100°C Slope & offset 0.04 + 21 Ω 0.03 + 100 nA 0.03 + 500 µV NOTES CMRR12 90 dB SOURCE OUTPUT MODES: Fixed DC level. PROGRAMMABILITY: IEEE-488 (SCPI-1995.0), RS-232, 3 user-definable power-up states plus factory default and *RST. POWER SUPPLY: Nominal 100 to 240VAC rms, 50–60Hz, 90VA. WARRANTY: 1 year. EMC: Conforms to European Union Directive 89/336/EEC, EN 61326-1. SAFETY: Conforms to European Union Directive 73/23/EEC, EN 61010-1. VIBRATION: MIL-PRF-28800F Class 3 Random Vibration. WARM-UP: 1 hour to rated accuracies. DIMENSIONS, WEIGHT: 89mm high × 213 mm wide × 370mm deep (31⁄2 in × 83⁄8 in × 149⁄16 in). Bench configuration (with handle & feet): 104mm high × 238mm wide × 370mm deep (41⁄8 in × 93⁄8 in × 149⁄16 in). Net Weight: 3.8kg (8.38 lbs). ENVIRONMENT: Operating: 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C. Storage: –25° to 65°C LTR ±2.9% COMMON MODE VOLTAGE: 30VDC. COMMON MODE ISOLATION: >109Ω, <1000pF MAX. VOLTAGE DROP IN INPUT FORCE LEADS: 1 volt. MAX. SENSE LEAD RESISTANCE: 100Ω for rated accuracy. SENSE INPUT IMPEDANCE: > 1·108Ω. 100°C 0.27°C 0.18°C 0.15°C 0.13°C GENERAL NOISE REJECTION: SPEED NPLC Normal 1.00 ±12% OPEN/SHORTED ELEMENT DETECTION SOFTWARE LINEARIZATION FOR THERMISTOR AND RTD Accuracy vs. Temperature 0°C 0.021°C 0.015°C 0.006°C 0.009°C Solid State Current Voltage Output (ISS) Output (VSS) +13.5V 2.5 mA 833 µA 15.75 V max APP. DATE DRN. SZ 4/12/00 CKD. SZ 4/10/02 APP. HW SK DATE DATE DATE 1 With remote voltage sense. 2 With 10kΩ thermistor as sensor. 3 Short term stability is defined as 24 hours with Peltier and Model 2510 at 25°C ±0.5°C. 4 Long term stability is defined as 30 days with Peltier and Model 2510 at 25°C ±0.5°C. 5 10Hz to 10MHz measured at 5A output into a 2Ω load. 6 Common mode voltage = 0V (meter connect enabled, connects Peltier low output to thermistor measure circuit ground). ±(0.1% of rdg + 0.1Ω) with meter connect disabled. 7 Resistance range 0Ω to 20Ω for rated accuracy. 8 Current through Peltier > 0.2A. 9 Default values shown, selectable values of 3µA, 10µA, 33µA, 100µA, 833µA, 2.5mA. Note that temperature control performance will degrade at lower currents. 10 AC Ohms is a dual pulsed measurement using current reversals available over bus only. 11 @23°C ±5°C. 12 For 1kΩ unbalance in LO lead. Minimum amplifier specification. 13 Resistance range 0Ω to 100Ω for rated accuracy. 14 Accuracy figures represent the uncertainty that the Model 2510 may add to the temperature measurement, not including thermistor uncertainty. These accuracy figures are for thermistors with typical A, B, C constants. HW 3/13/02 Rev. B 4/2/02 4/2/02 Keithley Instruments, Inc. Cleveland, Ohio 44139 PART NUMBER SPECIFICATIONS FORM 28777A-SBG SPEC-2510 Models 2510 and 2510-AT Specifications A-3 Accuracy calculations The information below discusses how to calculate accuracy for both TEC measurement and thermal feedback element specifications. Accuracy specifications are stated as follows: Accuracy = ±(% of reading + offset) As an example of how to calculate the actual limits, assume an output voltage of 5V. You can compute the limits from one-year operating voltage accuracy specifications as follows: Accuracy = ±(% of reading + offset) ±[(0.1% X 5V) + 4mV] ±(5mV + 4mV) ±9mV Thus, the actual operating voltage range is: 5V± 9mV, or from 4.991 to 5.009V. B Calibration Reference B-2 Calibration Reference Models 2510 and 2510-AT Introduction This appendix contains detailed information on the various Model 2510 remote calibration commands, calibration error messages, and methods to detect the end of each calibration step. Section 2 of this manual covers detailed calibration procedures. Command summary Table B-1 summarizes Model 2510 calibration commands. These commands are covered in detail in the following paragraphs. Table B-1 Calibration commands Command Description :CALibration :PROTected :CODE '<password>' :CODE? :SENSe :VOLTage <NRf> :DATA? :CURRent <NRf> :DATA? :TEMPerature <NRf> :DATA? :SOURce <NRf> :DATA? :AUTO :IPDac :DATA? :RESistance [:AC] <NRf> :DATA? :DATE <yyyy>,<mm>,<dd> :DATE? :NDUE <yyyy>,<mm>,<dd> :NDUE? :SAVE :LOCK :LOCK? :COUNt? Calibration subsystem. Calibration commands protected by code/password. Unlock calibration. (Default code/password: KI002510.) Query calibration code/password. Calibrate Peltier voltage measurement. Query Peltier voltage calibration constants. Calibrate current measurement. Query current calibration constants. Calibrate temperature measurement. Query temperature calibration constants. Calibrate Peltier source/sense voltage. Query Peltier source/sense calibration constants. Calibrate voltage source. Calibrate current protection DAC. Query current protection DAC calibration constants. Calibrate AC ohms. Query AC ohms calibration constants Program calibration year, month, day. Query calibration date. Program calibration due year, month, day. Query calibration due date. Save calibration data in EEPROM. Lock out calibration. Query if calibration is locked. (1 = locked; 0 = unlocked). Query number of times Model 2510 has been calibrated. Models 2510 and 2510-AT Calibration Reference B-3 Miscellaneous commands Miscellaneous commands are those commands that perform such functions as unlocking calibration, saving calibration constants, locking out calibration, and programming date parameters. :CODE (:CALibration:PROTected:CODE) Purpose To unlock calibration so that you can perform the calibration procedures. Format :cal:prot:code '<password>' Parameters Up to a 8-character string including letters and numbers. Description The :CODE command sends the password/code and enables calibration when performing these procedures via remote. The correct password must be sent to the unit before sending any other calibration command. The default remote password is KI002510. NOTE • The :CODE command should be sent only once before performing calibration. Do not send :CODE before each calibration step. • To change the code, first send the present code, then send the new code. • The password parameter must be enclosed in single quotes. • If you change the first two characters of the password to something other than “KI”, you will not be able to unlock calibration from the front panel. Example :CAL:PROT:CODE 'KI002510' Send default code of KI002510. :COUNT? (:CALibration:PROTected:COUNT?) Purpose To request the number of times the Model 2510 has been calibrated. Format :cal:prot:count? Response Number of times calibrated. Description The :COUNT? query may be used to determine the total number of times the Model 2510 has been calibrated. Example :CAL:PROT:COUNT? Request calibration count. B-4 Calibration Reference Models 2510 and 2510-AT :LOCK (:CALibration:PROTected:LOCK) Purpose To lock out calibration. Format Query :cal:prot:lock Response 0 1 Description The :LOCK command allows you to lock out calibration after completing the procedure. Thus, :LOCK performs the opposite of sending the password with the :CODE command. The :LOCK? query returns calibration lock status. NOTE :cal:prot:lock? Calibration unlocked Calibration locked To unlock calibration, send the :CODE command with the appropriate password. Example :CAL:PROT:LOCK Lock out calibration :SAVE (:CALibration:PROTected:SAVE) Purpose To save calibration constants in EEROM after the calibration procedure. Format :cal:prot:save Description The :SAVE command stores internally calculated calibration constants derived during comprehensive in EEROM. EEROM is non-volatile memory, and calibration constants will be retained indefinitely once saved. Generally, :SAVE is sent after all other calibration steps (except for :LOCK). NOTE Calibration will be only temporary unless the :SAVE command is sent to permanently store calibration constants. Calibration data will not be saved if calibration was not unlocked by sending the :CODE command or if invalid calibration data exists. Example :CAL:PROT:SAVE Save calibration constants Models 2510 and 2510-AT Calibration Reference B-5 :DATE (:CALibration:PROTected:DATE) Purpose To program the calibration date. Format :cal:prot:date <year>, <month>, <day> Parameters <year> = 2000 to 2099 <month> = 1 to 12 <day> = 1 to 31 Query :cal:prot:date? Response <year> , <month> , <day> Description The :DATE command allows you to store the calibration date in instrument EEROM for future reference. You can read back the date from the instrument by using the :DATE? query, or by using the front panel CAL menu. NOTE The year, month, and day parameters must be delimited by commas. Example :CAL:PROT:DATE 2000,11,20 Send cal date (11/20/2000). :NDUE (:CALibration:PROTected:NDUE) Purpose To send the next calibration due date to the instrument. Format :cal:prot:ndue <year>, <month>, <day> Parameters <year> = 2000 to 2099 <month> = 1 to 12 <day> = 1 to 31 Query :cal:prot:ndue? Response <year>, <month>, <day> Description The :NDUE command allows you to store the date when calibration is next due in instrument memory. You can read back the next due date by using the :NDUE? query, or by using the front panel CAL menu. NOTE The next due date parameters must be delimited by commas. Example :CAL:PROT:NDUE 2001,11,20 Send due date (11/20/2001). B-6 Calibration Reference Models 2510 and 2510-AT SENSe commands :VOLTage (:CALibration:PROTected:SENSe:VOLTage) (:CALibration:PROTected:SENSe:VOLTage:DATA?) Purpose To calibrate the Peltier voltage measurement. Format :cal:prot:sens:volt <DMM_Reading> Parameters <DMM_Reading> = Query :cal:prot:sens:volt:data? Description The :CAL:PROT:SENS:VOLT command calibrates the Model 2510 Peltier voltage measurement. During the calibration process, this command is sent three times; once each with parameters of approximately +F.S, 0, and -F.S. The actual parameters are voltage readings taken with a precision DMM. 3.5 to 12.5[V] = cal at + F.S. -2.5 to 2.5[V] = zero cal -3.5 to -12.5[V] = cal at - F.S. The :CAL:PROT:SENS:VOLT:DATA? query allows you to request the Peltier voltage calibration parameters. Example :CAL:PROT:SENS:VOLT 8.51 Calibrate positive voltage. :CURRent (:CALibration:PROTected:SENSe:CURRent) (:CALibration:PROTected:SENSe:CURRent:DATA?) Purpose To calibrate the current measurement. Format :cal:prot:sens:curr <Resistor_Value> Parameters <Resistor_Value> = 0.5 to 1.5[Ω] Query :cal:prot:sens:curr:data? Description The :CAL:PROT:SENS:CURR command calibrates the Model 2510 current measurement. During the calibration process, a 1Ω (nominal) resistor must be connected to the OUTPUT terminals, and the source is automatically set to +4.5V, 0V, and -4.5V. The command parameter is the characterized value of the 1Ω resistor. The :CAL:PROT:SENS:CURR:DATA? query allows you request the current calibration parameters. Example :CAL:PROT:SENS:CURR 1.01 Calibrate current. Models 2510 and 2510-AT Calibration Reference B-7 :TEMPerature (:CALibration:PROTected:SENSe:TEMPerature) (:CALibration:PROTected:SENSe:TEMPerature:DATA?) Purpose To calibrate the temperature sensor measurements. Format :cal:prot:sens:temp <Value> Parameters <Resistor_Value> = 100kΩ thermistor: 0 to 125e3[Ω] 10kΩ thermistor: 0 to 12.5e3[Ω] 1kΩ thermistor: 0 to 1.25e3[Ω] 100Ω thermistor: 0 to 125[Ω] 100Ω RTD: 0 to 125[Ω] 1000Ω RTD: 0 to 1250 [Ω] VSS: 0V to 5[V] ISS: 350 to 1.25e3[Ω] then 0.01 to 1.5[V] Query :cal:prot:sens:temp:data? Description The :CAL:PROT:SENS:TEMP command calibrates the Model 2510 temperature sensor measurements. During the calibration process, a 100Ω, 1kΩ, 10kΩ, or 100kΩ (nominal) resistor or a short must be connected to the INPUT terminals, and the characterized value, a zero value, or a voltage reading is sent as the command parameter, depending on the calibration step (see Section 2 for details). The appropriate sensor and range must first be selected with the following commands: • Enable 4-wire sensing: :SYST:RSEN ON • Thermistor sensor: :SENS:TEMP:TRAN THER • Thermistor range: :SENS:TEMP:THER:RANG <Range> • RTD sensor: :SENS:TEMP:TRAN RTD • RTD range: :SENS:TEMP:RTD:RANG <Range> • ISS sensor: :SENS:TEMP:TRAN ISS • VSS sensor: :SENS:TEMP:TRAN VSS The :CAL:PROT:SENS:TEMP:DATA? query allows you request the temperature sensor calibration parameters. Example :CAL:PROT:SENS:TEMP 100.5 Calibrate temperature sensor. B-8 Calibration Reference Models 2510 and 2510-AT SOURce commands :SOURCE (:CALibration:PROTected:SOURce) (:CALibration:PROTected:SOURce:DATA?) Purpose To calibrate the Peltier source/sense voltage. Format :cal:prot:sour <DMM_Reading> Parameters <DMM_Reading> = Query :cal:prot:sour:data? Description The :CAL:PROT:SOUR command calibrates the Model 2510 Peltier source/sense voltage. During the calibration process, this command is sent three times; once each with parameters of approximately +F.S (full scale), 0, and -F.S. The actual parameters are voltage readings taken with a precision DMM. 3.5 to 12.5[V] = cal at + F.S. -2.5 to 2.5[V] = zero cal -3.5 to -12.5[V] = cal at - F.S. The :CAL:PROT:SOUR:DATA? query allows you request the voltage calibration parameters. Example :CAL:PROT:SOUR 8.51 Calibrate with positive voltage. :AUTO (:CALibration:PROTected:SOURce:AUTO) Purpose To calibrate the voltage source. Format :cal:prot:sour:auto Description The :CAL:PROT:SOUR:AUTO command calibrates the Model 2510 voltage source. Disconnect all test leads or equipment from the OUTPUT terminals before sending this command. Example :CAL:PROT:SOUR:AUTO Calibrate voltage source. Models 2510 and 2510-AT Calibration Reference B-9 DAC commands :IPDac (:CALibration:PROTected:IPDac) (:CALibration:PROTected:IPDac:DATA?) Purpose To calibrate the current protection DAC. Format Query :cal:prot:ipd Description The :CAL:PROT:IPD command calibrates the current protection DAC. A resistor of less than 1.5Ω must be connected to the OUTPUT terminals before this command is sent. :cal:prot:ipd:data? The :CAL:PROT:IPD:DATA? query allows you request the current protection DAC calibration parameters. Example :CAL:PROT:IPD Calibrate current protection DAC. RESistance commands :RESistance (:CALibration:PROTected:RESistance[:AC]) (:CALibration:PROTected:RESistance[:AC]:DATA?) Purpose To calibrate the AC ohms circuit. Format :cal:prot:res <Resistance_Value> Parameters <Resistance_Value> = 10 to 1e3 [Ω] (100 default) Query :cal:prot:res:data? Description The :CAL:PROT:RES command calibrates the Model 2510 AC ohms circuit. A 10Ω to 1kΩ (100Ω preferred) resistor must be connected to the OUTPUT terminals, and the characterized value of that resistor is used as the :CAL:PROT:RES command parameter. The output is off. The :CAL:PROT:RES:DATA? query requests the AC ohms calibration constants. Example :CAL:PROT:RES 99.5 Calibrate with 100Ω nominal resistor. B-10 Calibration Reference Models 2510 and 2510-AT Detecting calibration errors If an error occurs during any calibration step, the Model 2510 will generate an appropriate error message. Several methods to detect calibration errors are discussed below. Reading the error queue As with other Model 2510 errors, any calibration errors will be reported in the error queue. (You can read the error queue by using the :SYST:ERR? query.) Error summary Table B-2 summarizes calibration errors. If errors occur, and :CAL:PROT:SAVE is sent, calibration constants without errors will be saved while those with errors will not. Status byte EAV (Error Available) bit Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the status byte will be set. Use the *STB? query to obtain the status byte, then test bit 2 to see if it is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query to read the error and at the same time clear the EAV bit in the status byte. Generating an SRQ on error To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an error occurs, send the following command: *SRE 4. This command will enable SRQ when the EAV bit is set. You can then read the status byte and error queue as outlined above to check for errors and to determine the exact nature of the error. Models 2510 and 2510-AT Calibration Reference Table B-2 Calibration errors Error number +500 +501 +502 +509 +510 +520 +521 +522 +523 +524 +525 +530 +531 +532 +533 +540 +541 +542 +543 +550 +551 +552 +553 +560 +561 +562 +563 +564 +565 +566 +570 Error message “Date of calibration not set” “Next date of calibration not set” “Calibration data invalid” “Not permitted with cal locked” “Not permitted with cal un-locked” “Source + gain data invalid” “Source + offset data invalid” “Source - gain data invalid” “Source - offset data invalid” “Source DAC Overflow” “Source DAC Underflow” “I Protection gain data invalid” “I Protection offset data invalid” “I Protection DAC Overflow” “I Protection DAC Underflow” “V Meas + gain data invalid” “V Meas + offset data invalid” “V Meas - gain data invalid” “V Meas - offset data invalid” “I Meas + gain data invalid” “I Meas + offset data invalid” “I Meas - gain data invalid” “I Meas - offset data invalid” “100 Ohm Thermistor data invalid” “1kOhm Thermistor data invalid” “10kOhm Thermistor data invalid” “100kOhm Thermistor data invalid” “RTD data invalid” “VSS data invalid” “ISS data invalid” “AC Ohms data invalid” B-11 B-12 Calibration Reference Models 2510 and 2510-AT Detecting calibration step completion When sending remote calibration commands, you must wait until the instrument completes the current operation before sending another command. You can use either *OPC? or *OPC to help determine when each calibration step is completed. Using the *OPC? query With the *OPC? (operation complete) query, the instrument will place an ASCII 1 in the output queue when it has completed each step. To determine when the OPC response is ready, do the following: 1. 2. 3. Repeatedly test the MAV (Message Available) bit (bit 4) in the status byte and wait until it is set. (You can request the status byte by using the *STB? query.) When MAV is set, a message is available in the output queue, and you can read the output queue and test for an ASCII 1. After reading the output queue, repeatedly test MAV again until it clears. At this point, the calibration step is completed. Using the *OPC command The *OPC (operation complete) command can be used to detect the completion of each calibration step. In order to use *OPC to detect the end of each calibration step, do the following: 1. 2. 3. 4. Enable operation complete by sending *ESE 1. This command sets the OPC (operation complete bit) in the standard event enable register, allowing operation complete status from the standard event status register to set the ESB (event summary bit) in the status byte when operation complete is detected. Send the *OPC command immediately following each calibration command. For example: :CAL:PROT:SENS:VOLT 8;*OPC Note that you must include the semicolon (;) to separate the two commands, and that the *OPC command must appear on the same line as the calibration command. After sending a calibration command, repeatedly test the ESB (Event Summary) bit (bit 5) in the status byte until it is set. (Use *STB? to request the status byte.) Once operation complete has been detected, clear OPC status using one of two methods: (1) use the *ESR? query, then read the response to clear the standard event status register, or (2) send the *CLS command to clear the status registers. Note that sending *CLS will also clear the error queue and operation complete status. Models 2510 and 2510-AT Calibration Reference B-13 Generating an SRQ on calibration complete An IEEE-488 bus SRQ (service request) can be used to detect operation complete instead of repeatedly polling the Model 2510. To use this method, send both *ESE 1 and *SRE 32 to the instrument, then include the *OPC command at the end of each calibration command line, as covered above. Clear the SRQ by querying the ESR (using the *ESR? query) to clear OPC status, then request the status byte with the *STB? query. Refer to your controller's documentation for information on detecting and servicing SRQs. C Calibration Program C-2 Calibration Program Models 2510 and 2510-AT Introduction This appendix includes a calibration program written in BASIC to help you to calibrate the Model 2510. Refer to Section 2 for more details on calibration procedures, equipment, and connections. Appendix B covers calibration commands in detail. Computer hardware requirements The following computer hardware is required to run the calibration programs: • • • IBM PC compatible computer. Keithley KPC-488.2, KPS-488.2, or KPC-488.2AT, or CEC PC-488 IEEE-488 interface for the computer. Two shielded IEEE-488 connecting cables (Keithley Model 7007). Software requirements In order to use the calibration programs, you will need the following computer software: • • • Microsoft QBasic (supplied with MS-DOS 5.0 or later) or Quick Basic. MS-DOS version 5.0 or later or Windows 95/98/Me. HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and CEC interface cards listed above). Calibration equipment The following calibration equipment is required: • • • Keithley Model 2002 DMM. 1Ω, 100Ω, 1kΩ, 10kΩ, and 100kΩ resistors. Clean copper wire. See Section 2 for detailed equipment specifications. Models 2510 and 2510-AT Calibration Program C-3 General program instructions 1. 2. 3. 4. 5. 6. With the power off, connect the Model 2510 and the digital multimeter to the IEEE-488 interface of the computer. Be sure to use shielded IEEE-488 cables for bus connections. Turn on the computer, the Model 2510, and the digital multimeter. Allow the Model 2510 and the multimeter to warm up for at least one hour before performing calibration. Make sure the Model 2510 is set for a primary address of 15. (Use the front panel MENU/COMMUNICATIONS/GPIB selection to check or change the address.) Make sure the digital multimeter primary address is set to 16. Make sure that the computer bus driver software (CECHP.EXE) is properly initialized. Enter the QBasic editor, and type in the program below. NOTE Be sure to change the resistor parameters (R$, R$(1) - R$(4)) to the characterized values of the 1Ω, 100Ω, 1kΩ, 10kΩ, and 100kΩ resistors. (See Section 2 for details.) 7. 8. Check thoroughly for errors, then save it using a convenient filename. Run the program, and follow the prompts on the screen to perform calibration. For test connections, refer to the following figures in Section 2: • • • • • • Voltage calibration DMM connections: Figure 2-1. Current and current protection calibration 1Ω resistor connections: Figure 2-2. Thermistor and RTD calibration 100Ω to 100kΩ resistor connections: Figure 2-3. Shorted INPUT connections: Figure 2-4. I-SS and V-SS calibration resistor/DMM connections: Figure 2-5. AC ohms calibration 100Ω resistor connections: Figure 2-6. C-4 Calibration Program Program C-1 Models 2510 and 2510-AT Model 2510 calibration program ‘ Model 2510 calibration program for use with Keithley 2002 DMM. ‘ 2510 primary address = 15. 2002 primary address = 16. OPEN “IEEE” FOR OUTPUT AS #1 ‘ Open IEEE-488 output path. OPEN “IEEE” FOR INPUT AS #2 ‘ Open IEEE-488 input path. PRINT #1, “INTERM CRLF” ‘ Set input terminator. PRINT #1, “OUTTERM LF” ‘ Set output terminator. PRINT #1, “REMOTE 15 16” ‘ Put 2510, 2002 in remote. PRINT #1, “OUTPUT 15;*RST” ‘ Initialize 2510. PRINT #1, “OUTPUT 15;*CLS” ‘ Clear 2510 status. PRINT #1, “OUTPUT 15;*ESE 1;*SRE 32” ‘ Enable OPC and SRQ. PRINT #1, “OUTPUT 16;:SYST:PRES” ‘ Initialize 2002. PRINT #1, “OUTPUT 16;:FORM:ELEM READ” ‘ 2002 reading data only. ‘ ‘ Replace nominal resistor values below with characterized values. R$ = “1” ‘ 1 ohm value. R$(1) = “100” ‘ 100 ohm value. R$(2) = “1e3” ‘ 1k ohm value R$(3) = “1e4” ‘ 10k ohm value. R$(4) = “1e5” ‘ 100k ohm value. ‘ CLS PRINT “Model 2510 Calibration Program” GOSUB KeyCheck PRINT #1, “OUTPUT 15;:CAL:PROT:CODE ‘KI002510’” J = 1 FOR I = 1 TO 46 READ Cmd$ SELECT CASE I CASE 1 PRINT “Connect DMM INPUT to 2510 OUTPUT terminals.” GOSUB KeyCheck CASE 3, 4, 6, 7, 9, 10, 39, 41 GOSUB ReadDMM Cmd$ = Cmd$ + Reading$ CASE 12 PRINT “Connect 1 ohm resistor to OUTPUT terminals.” GOSUB KeyCheck CASE 13 Cmd$ = Cmd$ + R$ CASE 19, 22, 25, 28, 32, 35 IF I = 32 THEN J = 1 PRINT “Connect “; R$(J); “ ohm resistor to INPUT terminals.” GOSUB KeyCheck Cmd$ = Cmd$ + R$(J) J = J + 1 Models 2510 and 2510-AT Calibration Program CASE 20, 23, 26, 29, 33, 36, 42 PRINT “Short INPUT terminals with clean copper wire.” PRINT “Wait 1 minute.” GOSUB KeyCheck CASE 38, 41 PRINT “Connect 1k ohm resistor and DMM to INPUT terminals.” IF I = 38 THEN Cmd$ = Cmd$ + R$(2) GOSUB KeyCheck CASE 43 PRINT “Connect 100 ohm resistor to OUTPUT terminals.” GOSUB KeyCheck Cmd$ = Cmd$ + R$(1) CASE 44 PRINT “Disconnect all equipment from OUTPUT terminals.” GOSUB KeyCheck END SELECT PRINT #1, “OUTPUT 15;”; Cmd$; “;*OPC” GOSUB CalEnd GOSUB ErrCheck NEXT I LINE INPUT “Enter calibration date (yyyy,mm,dd): “; D$ PRINT #1, “OUTPUT 15;:CAL:PROT:DATE “; D$ LINE INPUT “Enter calibration due date (yyyy,mm,dd): “; D$ PRINT #1, “OUTPUT 15;:CAL:PROT:NDUE “; D$ PRINT #1, “OUTPUT 15;:CAL:PROT:SAVE” ‘ Save calibration constants. GOSUB ErrCheck PRINT #1, “OUTPUT 15;:CAL:PROT:LOCK” ‘ Lock out calibration. PRINT “Calibration completed.” PRINT #1, “LOCAL 15 16” CLOSE END ‘ KeyCheck: ‘ Check for key press routine. WHILE INKEY$ <> ““: WEND ‘ Flush keyboard buffer. PRINT : PRINT “Press any key to continue (ESC to abort program).” DO: I$ = INKEY$: LOOP WHILE I$ = ““ IF I$ = CHR$(27) THEN GOTO EndProg ‘ Abort if ESC is pressed. RETURN ‘ CalEnd: ‘ Check for cal step completion. DO: PRINT #1, “SRQ?” ‘ Request SRQ status. INPUT #2, S ‘ Input SRQ status byte. LOOP UNTIL S ‘ Wait for operation complete. PRINT #1, “OUTPUT 15;*ESR?” ‘ Clear OPC. PRINT #1, “ENTER 15” INPUT #2, S PRINT #1, “SPOLL 15” ‘ Clear SRQ. INPUT #2, S RETURN C-5 C-6 Calibration Program Models 2510 and 2510-AT ‘ ErrCheck: ‘ Error check routine. PRINT #1, “OUTPUT 15;:SYST:ERR?” ‘ Query error queue. PRINT #1, “ENTER 15” INPUT #2, E, Err$ IF E <> 0 THEN PRINT Err$: GOTO ErrCheck’ Display error. RETURN ‘ ReadDMM: ‘ Get reading from DMM. SLEEP 3 PRINT #1, “OUTPUT 16;:READ?” PRINT #1, “ENTER 16” INPUT #2, Reading$ RETURN ‘ EndProg: ‘ Close files, end program. BEEP: PRINT “Calibration aborted.” PRINT #1, “OUTPUT 15;:CAL:PROT:LOCK” PRINT #1, “OUTPUT 15;*RST” PRINT #1, “LOCAL 15 16” CLOSE END ‘ CmdList: ‘ Calibration command list. DATA “:OUTP ON”,”:SOUR:VOLT 8.5”,”CAL:PROT:SENS:VOLT “,”CAL:PROT:SOUR “ DATA “:SOUR:VOLT 0”,”:CAL:PROT:SENS:VOLT “,”:CAL:PROT:SOUR “ DATA “:SOUR:VOLT -8.5”,”:CAL:PROT:SENS:VOLT “,”:CAL:PROT:SOUR “,”:OUTP OFF” DATA “:OUTP ON”, “CAL:PROT:SENS:CURR “,”CAL:PROT:IPD”,”:OUTP OFF” DATA “:SYST:RSEN ON”,”:SENS:TEMP:TRAN THER” DATA “:SENS:TEMP:THER:RANG 100”,”CAL:PROT:SENS:TEMP “,”CAL:PROT:SENS:TEMP 0” DATA “:SENS:TEMP:THER:RANG 1E3”,”:CAL:PROT:SENS:TEMP “,”CAL:PROT:SENS:TEMP 0” DATA “:SENS:TEMP:THER:RANG 1E4”,”:CAL:PROT:SENS:TEMP “,”CAL:PROT:SENS:TEMP 0” DATA “:SENS:TEMP:THER:RANG 1E5”,”:CAL:PROT:SENS:TEMP “,”CAL:PROT:SENS:TEMP 0” DATA “:SENS:TEMP:TRAN RTD”,”:SENS:TEMP:RTD:RANG 100” DATA “:CAL:PROT:SENS:TEMP “,”:CAL:PROT:SENS:TEMP 0” DATA “:SENS:TEMP:RTD:RANG 1E3”,”:CAL:PROT:SENS:TEMP “,”CAL:PROT:SENS:TEMP 0” DATA “:SENS:TEMP:TRAN ISS”,”:CAL:PROT:SENS:TEMP “,”:CAL:PROT:SENS:TEMP “ DATA “:SENS:TEMP:TRAN VSS”,”:CAL:PROT:SENS:TEMP “,”CAL:PROT:SENS:TEMP 0” DATA “:CAL:PROT:RES “,”:OUTP ON” DATA “:CAL:PROT:SOUR:AUTO”,”:OUTP OFF” Index Connections AC ohms calibration 2-18 Current and current protection calibration 2-11 for AC resistance verification test 1-11 for current verification tests 1-9 for sensor resistance accuracy verification 1-12 for voltage verification tests 1-7 I-SS and V-SS calibration resistor 2-15 Shorted input calibration 2-13 Thermistor and RTD calibration resistor 2-12 voltage calibration 2-9 Current accuracy 1-8 limit 1-10 Output 1-8 readback 1-9 A Aborting calibration steps 2-8 AC resistance accuracy 1-10 Accuracy calculations A-5 Analog circuitry checks 4-17 Assembly drawings 5-3 B Block diagram Analog circuitry 4-6 Digital circuitry 4-13 Overall 4-5 Power supply 4-12 C Calibration 2-1 Calibration commands B-2 Calibration considerations 2-3 Calibration cycle 2-3 Recommended calibration equipment 2-3 Resistor characterization 2-4 Calibration equipment C-2 Calibration errors 2-7, B-11 Front panel error reporting 2-7 Remote error reporting 2-7 Calibration menu 2-5 Calibration Program C-1 Calibration Reference B-1 Calibration step summary 2-9 Case cover removal 5-4 Changing the password 2-6 by remote 2-6 from the front panel 2-6 Characterized resistor values 1-4, 2-4 Class D Amplifier 4-6 Command summary B-2 Component layouts 6-2 Computer hardware requirements C-2 D DAC commands B-9 :IPDac B-9 Detecting calibration errors B-10 Error summary B-10 Generating an SRQ on error B-10 Reading the error queue B-10 Status byte EAV (Error Available) bit B-10 Detecting calibration step completion B-12 Generating an SRQ on calibration complete B-13 Using the *OPC command B-12 Using the *OPC? query B-12 Digital circuitry checks 4-16 Disassembly 5-1 Display board checks 4-15 E Environmental conditions 2-2 Line power 2-2 Temperature and relative humidity 2-2 Warm-up period 2-2 F P Factory service 6-2 Front panel calibration 2-8 Front panel disassembly 5-5 Front panel tests 4-3 CHAR SET test 4-4 DISPLAY PATTERNS test 4-4 KEYS test 4-3 Fuse replacement Digital I/O +5V supply 4-18 Power supply module 4-17 Parts list Display board 6-9 Miscellaneous 6-10 Mother board 6-3 Parts lists 6-2 Performance Verification 1-1 Performing the verification test procedures 1-5 Adjusting setpoints 1-6 Test considerations 1-6 Test summary 1-5 PID hardware loop control 4-11 Power line fuse 3-3 Power supply checks 4-16 Power-on self-test 4-3 Principles of operation 4-5 Analog circuits 4-6 A/D converter 4-11 Measurement circuits 4-10 Source circuits 4-6 Digital circuitry 4-13 Display board circuits 4-15 Overall block diagram 4-5 Power supply 4-12 G General program instructions C-3 Program C-1 Model 2510 calibration program C-4 H Handling and cleaning 5-2 Handling PC boards 5-2 Solder repairs 5-2 H-Bridge 4-8 I Instrument reassembly 5-6 Internal fuse replacement 4-17 Introduction 1-2, 2-2, 3-2, 4-2, 5-2, 6-2, B-2, C-2 L Line fuse replacement 3-2 M Miscellaneous commands B-3 :CODE B-3 :COUNT? B-3 :DATE B-5 :LOCK B-4 :NDUE B-5 :SAVE B-4 Mother board removal 5-4 N No comm link error 4-18 O Ordering information 6-2 R Rear panel 3-2 Recommended calibration equipment 2-4 Recommended test equipment 1-3 Resistor characterization 1-4 Recommended verification equipment 1-3 Remote calibration 2-20 command summary 2-20 procedure 2-21 Remote calibration command summary 2-20 Removing power components 5-5 Fan removal 5-6 Power module removal 5-6 Power supply module removal 5-5 Repair considerations 4-2 Replaceable Parts 6-1 Resetting the calibration password 2-7 RESistance commands B-9 :RESistance B-9 Restoring factory defaults 1-5 Routine Maintenance 3-1 S U Safety considerations 4-2 SENSe commands B-6 :CURRent B-6 :TEMPerature B-7 :VOLTage B-6 Sensor conditioning 4-10 Sensor measurement accuracy 1-11 Sensor resistance measurement accuracy limits 1-13 Software requirements C-2 SOURce commands B-8 :AUTO B-8 :SOURCE B-8 Specifications A-1 Static sensitive devices 5-3 T Troubleshooting 4-1, 4-15 Analog circuitry checks 4-17 Digital circuitry checks 4-16 Display board checks 4-15 Power supply checks 4-16 Unlocking calibration 2-5 by remote 2-6 from the front panel 2-5 V Verification limits 1-4 Example limits calculation 1-4 Resistance limits calculation 1-4 Verification test requirements 1-2 Environmental conditions 1-2 Line power 1-3 Warm-up period 1-2 Viewing calibration dates and calibration count 2-7 Voltage accuracy 1-6 limit 1-8 Output 1-6 readback 1-7 Service Form Model No. ______________ Serial No.___________________Date________________ Name and Telephone No. _________________________________________________ Company ______________________________________________________________ List all control settings, describe problem and check boxes that apply to problem. _________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ ❑ Intermittent ❑ Analog output follows display ❑ Particular range or function bad; specify _______________________________ ❑ IEEE failure ❑ Obvious problem on power-up ❑ Front panel operational ❑ All ranges or functions are bad ❑ Batteries and fuses are OK ❑ Checked all cables Display or output (check one) ❑ Drifts ❑ Overload ❑ Unable to zero ❑ Will not read applied input ❑ Calibration only ❑ Certificate of calibration required (attach any additional sheets as necessary) ❑ Unstable ❑ Data required Show a block diagram of your measurement including all instruments connected (whether power is turned on or not). Also, describe signal source. Where is the measurement being performed? (factory, controlled laboratory, out-of-doors, etc.)_______________ __________________________________________________________________________________________ What power line voltage is used? ___________________ Ambient temperature? ________________________ °F Relative humidity? ___________________________________________Other?___________________________ Any additional information. (If special modifications have been made by the user, please describe.) __________________________________________________________________________________________ __________________________________________________________________________________________ Be sure to include your name and phone number on this service form. Specifications are subject to change without notice. All Keithley trademarks and trade names are the property of Keithley Instruments, Inc. All other trademarks and trade names are the property of their respective companies. Keithley Instruments, Inc. 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 1-888-KEITHLEY (534-8453) • www.keithley.com Sales Offices: BELGIUM: CHINA: FINLAND: FRANCE: GERMANY: GREAT BRITAIN: INDIA: ITALY: JAPAN: KOREA: NETHERLANDS: SWEDEN: SWITZERLAND: TAIWAN: Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02-363 00 40 • Fax: 02/363 00 64 Yuan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-6202-2886 • Fax: 8610-6202-2892 Tietäjäntie 2 • 02130 Espoo • Phone: 09-54 75 08 10 • Fax: 09-25 10 51 00 3, allée des Garays • 91127 Palaiseau Cédex • 01-64 53 20 20 • Fax: 01-60 11 77 26 Landsberger Strasse 65 • 82110 Germering • 089/84 93 07-40 • Fax: 089/84 93 07-34 Unit 2 Commerce Park, Brunel Road • Theale • Berkshire RG7 4AB • 0118 929 7500 • Fax: 0118 929 7519 Flat 2B, Willocrissa • 14, Rest House Crescent • Bangalore 560 001 • 91-80-509-1320/21 • Fax: 91-80-509-1322 Viale San Gimignano, 38 • 20146 Milano • 02-48 39 16 01 • Fax: 02-48 30 22 74 New Pier Takeshiba North Tower 13F • 11-1, Kaigan 1-chome • Minato-ku, Tokyo 105-0022 • 81-3-5733-7555 • Fax: 81-3-5733-7556 2FL., URI Building • 2-14 Yangjae-Dong • Seocho-Gu, Seoul 137-888 • 82-2-574-7778 • Fax: 82-2-574-7838 Postbus 559 • 4200 AN Gorinchem • 0183-635333 • Fax: 0183-630821 c/o Regus Business Centre • Frosundaviks Allé 15, 4tr • 169 70 Solna • 08-509 04 679 • Fax: 08-655 26 10 Kriesbachstrasse 4 • 8600 Dübendorf • 01-821 94 44 • Fax: 01-820 30 81 1FL., 85 Po Ai Street • Hsinchu, Taiwan, R.O.C. • 886-3-572-9077• Fax: 886-3-572-9031 © Copyright 2001 Keithley Instruments, Inc. Printed in the U.S.A. 4/02 ">
Advertisement
Key features
- High output voltage and current
- Precision measurement capabilities
- Versatile configuration options
- Easy-to-use interface
- Comprehensive documentation
Frequently asked questions
The SourceMeter 2510 is a versatile, high-performance source measure unit that can be used to test a wide variety of electronic components and circuits.
The SourceMeter 2510 can source up to 250V and 2.5A.
The SourceMeter 2510 can measure voltage, current, resistance, and capacitance.
The accuracy specifications of the SourceMeter 2510 are detailed in the service manual. Contact the manufacturer for specific information about the specifications.