F u n c t i o n G e n e r a t o r
H M 8 0 3 0 - 6
Service-Manual
G e n e r a l i n f o r m a t i o n r e g a r d i n g t h e C E m a r k i n g
Manufacturer
Fabricant
Industriestraße 6
D-63533 Mainhausen
Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt
The HAMEG Instruments GmbH herewith declares conformity of the product
HAMEG Instruments GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation:
Funktionsgenerator
Générateur de fonctions
Typ / Type / Type: HM8030-6 mit / with / avec: HM8001-2
Optionen / Options / Options: mit den folgenden Bestimmungen / with applicable regulations / avec les directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Angewendete harmonisierte Normen /
Harmonized standards applied /
Normes harmonisées utilisées
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
Sicherheit / Safety / Sécurité:
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994
EN 61010-1/A2: 1995 / IEC 1010-1/A2: 1995 / VDE 0411 Teil 1/A1: 1996-05
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility / Compatibilité
électromagnétique
EN 61326-1/A1
Störaussendung / Radiation / Emission: Tabelle / table / tableau 4, Klasse / Class
/ Classe B.
Störfestigkeit / Immunity / Imunitee: Tabelle / table / tableau A1.
EN 61000-3-2/A14
Oberschwingungsströme / Harmonic current emissions / Émissions de courant harmonique: Klasse / Class / Classe D.
EN 61000-3-3
Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker /
Fluctuations de tension et du fl icker.
Datum/Date/Date
22.07.2004
Unterschrift / Signature /Signatur
Manuel Roth
Manager
General information regarding the CE marking
HAMEG instruments fulfi ll the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and product standards. In cases where different limit values are applicable,
HAMEG applies the severer standard. For emission the limits for residential, commercial and light industry are applied. Regarding the immunity (susceptibility) the limits for industrial environment have been used.
The measuring- and data lines of the instrument have much infl uence on emmission and immunity and therefore on meeting the acceptance limits. For different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) suffi ciently screened cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than
3 meters and not be used outside buildings. If an interface has several connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the manual for a shorter length, signal lines must be less than 3 meters and not be used outside buildings.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened cables (RG223/U, RG214/U) must be used.
3. Infl uence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fi elds, even with careful setup of the measuring equipment an infl uence of such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments specifi cations may result from such conditions in individual cases.
HAMEG Instruments GmbH
2
Subject to change without notice
C o n t e n t
Declaration of Conformity
General information regarding the CE-marking
Function Generator HM8030-6
Specifi cations
Control elements 6
Adjustment 7
Circuit and layout diagrams 9
4
5
2
2
Subject to change without notice
3
H M 8 0 3 0 - 6
1 0 M H z F u n c t i o n G e n e r a t o r
H M 8 0 3 0 - 6
Option HO801
HZ33, HZ34
Test cable BNC/BNC
HZ20 Connector
BNC to 4mm socket
Frequency range 50 mHz to 10 MHz
High signal purity and amplitude stability
Distortion factor ‹ 0.5 % up to 1 MHz
Output voltage 20 V
pp
(10 V
pp
into 50 Ω)
Surge- and short-circuit-proof output
Rise and fall time typ. 15 ns
Internal and external sweep
Pulse width adjustment
Highly accurate digital frequency display
Mainframe HM8001-2 required for operation
4
Subject to change without notice
10 MHz Function Generator HM8030-6
Valid at 23 °C after a 30 minute warm-up period
Operating modes
Sine, square, triangle, pulse; free running, internal sweep or external frequency modulation, with or without DC offset
Frequency ranges
0.05 Hz to 10 MHz in 8 ranges, variable: x 0.09 to x 1.1 (12:1)
Frequency drift: ‹ 0.5 % / hr or 0.8 %/24 hrs. at constant ambient temperature
Waveform characteristics
Sine wave distortion
0.05 Hz to 1 MHz:
1 MHz to 10 MHz:
Square wave rise time:
Overshoot:
Triangle non-linearity: max. 0.5 % max. 5 % typ. 15 ns
‹ 5 % (for termination into 50 Ω)
‹ 1 % (to 100 kHz)
Displays
Frequency:
Accuracy: up to 5 Hz:
5 Hz to 10 MHz:
5-digit, 7-segment LED, each 8 x 5 mm
± (1 % + 3 digits)
± (5 x 10 -5 + 1 digit)
LED indicators for mHz, Hz, kHz and sec
Outputs
Signal output: protected against ext. voltage up to ± 45 V
Impedance: short-circuit proof, max. (30 sec.)
50 Ω
DC
Output voltage:
Attenuation:
2 attenuators:
Variable:
10 V pp into 50 Ω load; 20 V max. 60 dB each 20 dB ± 0.2 dB
0 to 20 dB pp (open circuit)
Amplitude error:
0.05 Hz to 0.5 MHz:
0.5 MHz to 10 MHz:
DC offset: into 50 Ω load: in open circuit:
Trigger output:
(sine wave/triangle) max. 0.2 dB max. 0.5 dB variable (on/off, except pulse function) max. ±2.5 V max. ±5 V square wave synchronous to signal output, approx. +5 V/TTL
S p e c i f i c a t i o n s
FM input
(VCF, BNC connector on rear panel of HM8001-2 and option HO801)
Frequency deviation:
Input impedance:
Input voltage: approx. 1 : 100
6 kΩ II 25 pF max. ± 30 V
Internal sweep
Sweep speed:
Sweep range:
20 ms to 15 s approx. 1:100
Miscellaneous
Power supply
(from mainframe):
Weight:
+ 5 V/200 mA
+ 16 V/300 mA
- 16 V/250 mA
Operating temperature:
(∑= 9.8 W).
+10° C to + 40° C
Max. relative humidity: 80 % (without condensation)
Dimensions (W x H x D) (without 22-pole flat plug):
135 x 68 x 228 mm approx. 0.80 kg
Accessories supplied: Operator’s Manual
Optional accessories: HZ33/HZ34 BNC Test Cable, HZ22 50 Ω feed-through terminal, HZ10 Silicone test leads w w w . h a m e g . c o m
HM8030-6E/140705/ce · Subject to alterations · © HAMEG Instruments GmbH · ® Registered Trademark · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com · info
@hameg.com
A Rohde & Schwarz Company
Subject to change without notice
5
C o n t r o l e l e m e n t s
10 MHz FUNCTION GENERATOR HM8030-6 s kHz
Hz mHz
FREQUENCY
START
STOP
TIME
SWEEP
50 mHz 10 MHz
MADE IN GERMANY
Instruments
ON
10% 90%
-5 V +5V
OFF
OFFSET min. max.
AMPLITUDE
ATTENUATOR
-20 dB -20 dB
FUNCTION
TRIG.
OUTP.
(TTL)
!
50 Ω
OUTP.
max.
20Vpp
1 2 3 4 5 6 7 8 9 10 11 12
Control elements
DISPLAY (7 segment LED)
5-digit frequency meter. LED indicators for mHz, Hz, kHz and s
SWEEP (push button) and SWEEP- Indication (LEDs)
Button activates internal sweep generator. The LEDs indicate the function chosen with the SWEEP-Button. Settings are changed with or .
FREQUENCY (adjustment knob)
Continuous and linear frequency fi ne ad-justment, with the setting range from 0.09 to 1.1 (approx 0.045 to 1.1 in 10 MHzrange) overlapping the ranges selected with
FREQUENCY (2 pushbuttons)
Frequency range selection from 50 mHz to 10 MHz in 8 decade steps.
– (LED s)
Indication of selected function.
FUNCTION – (pushbutton)
Mode selection: Triangle, Sine, Square, Pulse and Off.
TRIGGER OUTPUT (BNC connector)
This short-circuit-proof output supplies a square signal in synchronism with the output signal. It is TTL compatible and has a duty-factor of approx. 50%.
OFFSET (adjustment knob)
Adjustment of the positive or negative offset voltage. This
DC voltage can be super-imposed on the output signal. The max. offset voltage is ±5V (o.c.) or ±2.5V respectively when terminated into 50Ω. The offset voltage is available to all functions except for pulse and activated by . In operation mode OFF (no function activated) it can be used separatey.
In pulse mode the pulse width is set with this control from
10% to 90%.
ON (pushbutton)
Activates the offset function except in pulse mode. If the
ON-button is pushed in pulse mode, pulse width is set with the control from 10% to 90%. In OFF-position the fi xed pulse width amounts to 50%.
50Ω OUTPUT (BNC connector)
Short-circuit proof signal output of the generator. The output impedance is 50Ω and the max. output amplitude is 20 V pp
(o.c.) or 10 V pp
respectively when terminated into
50Ω.
–20dB, –20bB (pushbutton)
Two parately. When both pushbuttons are activated, a total attenuation of 40dB results. Including the amplitude control
, the max. attenuation amounts to 60dB (factor 1000).
AMPLITUDE (adjustment knob)
Continuous adjustment of the output ampli-tude from 0 to
–20dB terminated into 50Ω.
6
Subject to change without notice
A d j u s t m e n t
Adjustment
Please observe the correct sequence of steps!
Measuring equipment required:
– Digital Multimeter ( HM8012 or similar)
– Frequency counter (HM8021 or similar)
– Distortion meter (HM8027 or similar)
– Oscilloscope
1) Turning service mode on a) For activating service mode push both buttons „50mHz“ and „10MHz“ simultaneously and switch on the
HM8030.
b) If the EPROM is empty or if the HM8030 is new, the display will show „E“, otherwise „F 3.1“. Push button „FUNC-
TION“.
c) If the display shows „XXX3“ (X = any fi gure, fi gure ‚3’ blinking), turn off the instrument.
2) Switches and controls
These settings must be kept constant during the whole adjustment procedure!
a) Open the top cover of the instrument.
b) Set OFFSET control to center position.
c) Button OFFSET deactivated.
d) Set AMPLITUDE control the right stop.
e) Buttons ATTENUATOR deactivated.
f) Push button SWEEP as often as needed to extinguish the START, STOP and TIME LEDs.
3) Internal triangle generator a) Set HM8030 to triangle function with f = 5250 Hz.
b) Connect HM8012 (measuring range: 500 mV
DC
) with a probe (with the low pass fi lter shown) at PT100.
c) Adjust 0 mV ±1 mV with VR103.
d) Connect HM8012 (measuring range: 500 mV
AC
) with a probe (without low pass fi lter) at PT100.
e) Adjust 577,3 mV ±1 mV with VR102.
f) As the settings are interacting, repeat steps b) to e) as often as is necessary for an optimum result.
4) Duty cycle 1 a) Connect HM8021 (function: pulse) to OUT.
b) Set HM8030 to square function with f = 52,50 kHz.
c) Adjust VR101 so that the positive pulse width is equal to the negative pulse width.
d) Set HM8030 to square function with f = 4,75 kHz .
e) Adjust VR104 so that the positive pulse width is equal to the negative pulse width.
f) As the settings are interacting, repeat steps b) to e) as often as is necessary for an optimum result.
5) Duty cycle 2 a) Connect HM8021 (function: pulse) to OUT.
b) Set HM8030 to square function with f = 475 Hz.
c) Adjust VR106 so that the positive pulse width is equal to the negative pulse width.
6) OUTPUT Offset a) Connect HM8012 (measuring range: 500 mV
DC
) with low pass fi lter to OUT.
b) Set HM8030 to triangle function with f = 5250 Hz.
c) Adjust 0 mV ±1 mV with VR111.
7) Distortion a) Connect HM8027 (20k Frequency Range) to OUT.
b) Set HM8030 to sine function with f = 5250 Hz.
c) Set HM8027 (settings: 100% CAL) with the blue control to 100, than switch to 10%.
d) Adjust the distortion factor to less than 0.3% by alternately adjusting VR108 and VR109.
e) If this adjustment is not possible, return to step 4) because the duty cycle was not correctly adjusted.
8) Square 1 a) Connect oscilloscope HM1507 (settings: 1V/div, 100μs/ div) with 50Ω to OUT.
b) Set HM8030 to triangle function with f = 5250 Hz.
c) Adjust VAR on the oscilloscope so that the amplitude of the signal equals to 6 div.
d) Set HM8030 to square function with f = 5250 Hz.
e) Adjust VR105, VR107 and VR110 so that the amplitude of the square signal amounts to 6 div.
f) Repeat step e) with f = 52,5 kHz.
9) Square 2 a) Connect oscilloscope HM1507 (settings: 1V/div, 50ns/div) with a 50Ω load at OUT.
b) Set HM8030 to square function with f = 5250kHz.
c) Adjust VC102 and VC104 so that the amplitude of the square signal equals to 6 div.
10) Triangle a) Connect oscilloscope HM1507 (settings: 1V/div, 50ns/div) with a 50Ω load at OUT.
b) Set HM8030 to triangle function with f = 7000 kHz.
c) Adjust VC101 so that the amplitude of the triangle signal equals to 6 div.
11) Duty cycle of pulse a) Connect HM8021 (function: pulse) to OUT.
b) Set HM8030 to pulse function with f = 5250 Hz.
c) Adjust VR1 so that the positive pulse width is equal to the negative pulse width.
Subject to change without notice
7
A d j u s t m e n t
Adjustment HM8030-6
30. November 2004
1) Turning service mode on
.
a)
If the EPROM is empty or if the HM8030 is new, the display will show „E“, otherwise „F 3.1“. Push button „FUNC-
TION“. If the display shows „XXX3“ (X = any figure, figure ‚3’ blinking), turn off the instrument.
2) Switches and controls - These settings must be kept constant during the whole adjustment! a) Open the top cover of the instrument.
b) Set OFFSET control to center position.
c) Button OFFSET deactivated. d) Set AMPLITUDE control to the right stop.
e)
Buttons ATTENUATOR deactivated.
f) Push button SWEEP as often as is needed to extinguish the START, STOP and TIME LEDs.
10) Triangle a) Connect oscilloscope HM1507 (settings: 1V/div,
50ns/div) with a 50 load at OUT. b) HM8030 - traingle function - f = 7000kHz. c) Adjust VC101 so that the amplitude of the triangle signal equals to 6 div.
4) Duty cycle 1 a) Connect HM8021 (function: pulse) to OUT. b) HM8030 - square function - f = 52,50kHz. c) Adjust the pos. pulse width is equal to neg. pulse width.
d) HM8030 - square function - f = 4,75kHz . e) Adjust VR104 so that the positive pulse width is equa to the negative pulse width.
f)
As the settings are interacting, repeat steps b) to e)
5) Duty cycle 2 a) Connect HM8021 (function: pulse) to OUT. b) Set HM8030 to square function with f = 475Hz.
c) Adjust VR106 so that the positive pulse width is equal to the negative pulse width.
11) Duty cycle of pulse a) Connect HM8021 (function: pulse) to
OUT.
b)
HM8030 - pulse function - f = 5250Hz.
c) Adjust VR1 so that the positive pulse width is equal to the negative pulse width.
8) Square 1 a) Connect oscilloscope HM1507 (settings: 1V/div, 100μs/div) with 50 to OUT. b) Set HM8030 to triangle function with f = 5250Hz. c) Adjust VAR on the oscilloscope so that the amplitude of the signal equals to 6 div.
d) Set HM8030 to square function with f = 5250Hz. e) Adjust VR105, VR107 and VR110 so that the amplitude of the square signal equals to 6 div.
f)
Repeat step e) with f = 52,5kHz.
9) Square 2 a) Connect oscilloscope HM1507 (settings: 1V/div, 50ns/div) with a 50 load at OUT. b) Set HM8030 to square function with f = 5250kHz. c)
Adjust VC102 and VC104 so that the amplitude of the square signal equals to 6 div.
VR106
VC101
PT100
VR102
VR106
VR1
VR109
VR111
VR105 VR110
VC104
3) Internal triangle generator a) HM8030 - triangle function - f = 5250Hz. b) HM 8012 (500mV
DC
) with a probe (with lowpass filter) at PT100. c) Adjust r 1mV with VR103.
d) Connect HM8012 (500mV
AC
) with a probe
(without lowpass filter) at PT100. e)
Adjust 577,3mV r 1mV with VR102.
f) As settings are interacting, repeat steps b) - e) as often as is necessary for an optimum result.
6) OUTPUT Offset a) Connect HM8012 (500mV
DC
) with lowpass filter to OUT. b) HM8030 - triangle function -
f = 5250Hz. c) Adjust 0mV r 1mV with VR111.
VC102
VR107
VR104
VR103
VR101
7) Distortion a) Connect HM8027 (20k Frequency
Range) to OUT. b) HM8030 - sine function - f = 5250Hz. c) Set HM8027 (settings: 100% CAL) with the blue control to 100, than switch to
10%.
d) Adjust the distortion factor to less than
0.3% by alternately adjusting VR108 and VR109.
e) If this adjustment is not possible, return to step 4) because the duty cycle was not correctly adjusted.
8
Subject to change without notice
C i r c u i t a n d l a y o u t d i a g r a m s
Subject to change without notice
9
C i r c u i t a n d l a y o u t d i a g r a m s
10
Subject to change without notice
C i r c u i t a n d l a y o u t d i a g r a m s
Subject to change without notice
11
C i r c u i t a n d l a y o u t d i a g r a m s
12
Subject to change without notice
C i r c u i t a n d l a y o u t d i a g r a m s
Subject to change without notice
13
C i r c u i t a n d l a y o u t d i a g r a m s
14
Subject to change without notice
C i r c u i t a n d l a y o u t d i a g r a m s
Subject to change without notice
15
C i r c u i t a n d l a y o u t d i a g r a m s
16
Subject to change without notice
C i r c u i t a n d l a y o u t d i a g r a m s
Subject to change without notice
17
C i r c u i t a n d l a y o u t d i a g r a m s
18
Subject to change without notice
C i r c u i t a n d l a y o u t d i a g r a m s
Subject to change without notice
19
Oscilloscopes
Spectrum Analyzer
Power Supplies
Modular System
8000 Series
Programmable Instruments
8100 Series authorized dealer
w w w . h a m e g . d e
Subject to change without notice
4S-8030-06E0 / 07-11-2005-gw
© HAMEG Instruments GmbH
A Rohde & Schwarz Company
® registered trademark
DQS-Certifi cation: DIN EN ISO 9001:2000
Reg.-Nr.: 071040 QM
HAMEG Instruments GmbH
Industriestraße 6
D-63533 Mainhausen
Tel +49 (0) 61 82 800-0
Fax +49 (0) 61 82 800-100 [email protected]