ITech IT6121, IT6151, IT6153, IT6154 User Manual
The Itech IT6153 is a versatile power supply that finds its application in laboratories, and in production and service areas wherever it is necessary to have a stabilized DC power supply. It is a digitally controlled, programmable, high resolution DC Power Supply. The model is single-channel, has a voltage range of 0 to 60 V and a maximum current of 5 A. Other specifications include minimum voltage resolution of 10 mV and a current resolution of 1 mA. The product combines compact size with a large color TFT-LCD display for easy programming and monitoring. Itech IT6153 is programmable through the front panel and remotely via USB, RS232, and LAN. Additional features include built-in OVP, OCP, OPP, OTP protections and a built-in cooling fan that makes the device suitable for continuous operation even in harsh ambient conditions.
Advertisement
Advertisement
High Resolution and High
Speed Programmable Power
Supply
Series IT6100 User’s Manual
Model: IT6121/IT6122/IT6123/IT6151/IT6152/
IT6153/IT6154
Version: V1.1
Notices
© Itech Electronics, Co., Ltd. 2015
No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into a foreign language) without prior permission and written consent from Itech Electronics,
Co., Ltd. as governed by international copyright laws.
Manual Part Number
IT6100-402196
Revision
1st Edition: May 26, 2015
Itech Electronics, Co., Ltd.
Trademarks
Pentium is U.S. registered trademarks of Intel Corporation.
Microsoft, Visual Studio, Windows and
MS Windows are registered trademarks of Microsoft Corporation in the United
States and/or other countries and regions.
Warranty Safety Notices
The materials contained in this document are provided “as is”, and is subject to change, without prior notice, in future editions. Further, to the maximum extent permitted by applicable laws, ITECH disclaims all warrants, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. ITECH shall not be held liable for errors or for incidental or indirect damages in connection with the furnishing, use or application of this document or of any information contained herein.
Should ITECH and the user enter into a separate written agreement with warranty terms covering the materials in this document that conflict with these terms, the warranty terms in the separate agreement shall prevail.
Technology Licenses
The hardware and/or software described herein are furnished under a license and may be used or copied only in accordance with the terms of such license.
A CAUTION sign denotes a hazard.
It calls attention to an operating procedure or practice that, if not correctly performed or adhered to, could result in damage to the product or loss of important data.
Do not proceed beyond a
CAUTION sign until the indicated conditions are fully understood and met.
A WARNING sign denotes a hazard. It calls attention to an operating procedure or practice that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
Restricted Rights Legend
Restricted permissions of the U.S. government. Permissions for software and technical data which are authorized to the U.S. Government only include those for custom provision to end users
ITECH provides this customary commercial license in software and technical data pursuant to FAR 12.211
(Technical Data) and 12.212 (Computer
Software) and, for the Department of
Defense, DFARS 252.227-7015
(Technical Data – Commercial Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or
Computer Software Documentation).
NOTE
A NOTE sign denotes important hint. It calls attention to tips or supplementary information that is essential for users to refer to.
IT6100 User’s Manual
Quality Certification and Assurance
We certify that IT6100 series power supply meets all the published specifications.
Warranty
ITECH warrants that the product will be free from defects in material and workmanship under normal use for a period of one (1) year from the date of delivery (except those described in the Limitation of Warranty below).
For warranty service or repair, the product must be returned to a service center designated by ITECH.
The product returned to ITECH for warranty service must be shipped
PREPAID. And ITECH will pay for return of the product to customer.
If the product is returned to ITECH for warranty service from overseas, all the freights, duties and other taxes shall be on the account of customer.
Limitation of Warranty
This Warranty will be rendered invalid if the product is:
Damage caused by circuit installed by customer or using customer own products or accessories;
Modified or repaired by customer without authorization;
Damage caused by circuit installed by customer or not operating our products under designated environment;
The product model or serial number is modified, deleted, removed or illegible;
Damaged as a result of accidents, including but not limited to lightning, moisture, fire, improper use or negligence.
Safety Symbols
ON (power on) Direct current
Alternating current OFF (power off)
Both direct and alternating current
Power-on state
Protective conductor terminal Power-off state
Earth (ground) terminal Reference terminal
Caution, risk of electric shock Positive terminal
Negative terminal Warning, risk of danger (refer to this manual for specific
Warning or Caution information)
Frame or chassis terminal - -
Copyright © ITECH Electronics Co., Ltd.
i
IT6100 User’s Manual
Safety Precautions
The following safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or specific warnings elsewhere in this manual will constitute a default under safety standards of design, manufacture and intended use of the instrument. ITECH assumes no liability for the customer’s failure to comply with these precautions.
Do not use the instrument if it is damaged. Before operation, check the casing to see whether it cracks. Do not operate the instrument in the presence of inflammable gasses, vapors or dusts.
The power supply is provided with a three-core power line during delivery and should be connected to a three-core junction box. Before operation, be sure that the power supply is well grounded. Make sure to use the power cord supplied by ITECH.
Check all marks on the instrument before connecting the instrument to power supply.
Use electric wires of appropriate load. All loading wires should be capable of bearing maximum short-circuit of electronic load without overheating. If there are multiple loads, each pair of the load power cord must be carry out the full rated short-circuit output current of the power securely.
Ensure the voltage fluctuation of mains supply is less than 10% of the working voltage range in order to reduce risks of fire and electric shock.
Do not install alternative parts on the instrument or perform any unauthorized modification.
Do not use the instrument if the detachable cover is removed or loosen.
To prevent the possibility of accidental injuries, be sure to use the power adapter supplied by the manufacturer only.
Never use the instrument with a life-support system or any other equipment subject to safety requirements.
Failure to use the instrument as directed by the manufacturer may render its protective features void.
Always clean the casing with a dry cloth. Do not clean the internals.
Make sure the vent hole is always unblocked.
Environmental Conditions
The instrument is designed for indoor use and an area with low condensation. The table below shows the general environmental requirements for the instrument. The speed of fan will change intelligently by the temperature of radiator. When the temperature is up to 40°C, the fan will be on and adjust intelligently when temperature changes.
Environmental Conditions
Operating temperature
Operating humidity
Storage temperature
Requirements
0°C to 40°C
20%-80% (non-condensation)
-20°C to 70 °C
Copyright © ITECH Electronics Co., Ltd.
ii
IT6100 User’s Manual
Altitude
Pollution degree
Installation category
≤2,000m
Pollution degree 2
Ⅱ
NOTE
To make accurate measurements, allow the instrument to warm up for 30 min before operation.
Regulatory Markings
The CE mark is a registered trademark of the European
Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
The instrument complies with the
WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard the electrical/electronic product in domestic household waste.
This symbol indicates the time period during which no hazardous or toxic substances are expected to leak or deteriorate during normal use. The expected service life of the product is
10 years. The product can be used safely during the 10-year Environment
Friendly Use Period (EFUP). Upon expiration of the EFUP, the product must be immediately recycled.
Waste Electrical and Electronic Equipment (WEEE)
Directive
2002/96/EC Waste Electrical and Electronic Equipment
(WEEE) Directive
This product complies with the WEEE Directive (2002/96/EC) marking requirement. This affix product label indicates that you must not discard the electrical/electronic product in domestic household waste.
Product Category
With reference to the equipment classifications described in the
Annex I of the WEEE Directive, this instrument is classified as a
“Monitoring and Control Instrument” product.
To return this unwanted instrument, contact your nearest ITECH office.
Copyright © ITECH Electronics Co., Ltd.
iii
IT6100 User’s Manual
Compliance Information
Complies with the essential requirements of the following applicable European
Directives, and carries the CE marking accordingly:
Electromagnetic Compatibility (EMC) Directive 2014/30/EU
Low-Voltage Directive (Safety) 2014/35/EU
Conforms with the following product standards:
EMC Standard
IEC 61326-1:2012/ EN 61326-1:2013 ¹²³
Reference Standards
CISPR 11:2009+A1:2010/ EN 55011:2009+A1:2010 (Group 1, Class A)
IEC 61000-4-2:2008/ EN 61000-4-2:2009
IEC 61000-4-3:2006+A1:2007+A2:2010/ EN 61000-4-3:2006+A1:2008+A2:2010
IEC 61000-4-4:2004+A1:2010/ EN 61000-4-4:2004+A1:2010
IEC 61000-4-5:2005/ EN 61000-4-5:2006
IEC 61000-4-6:2008/ EN 61000-4-6:2009
IEC 61000-4-11:2004/ EN 61000-4-11:2004
1. The product is intended for use in non-residential/non-domestic environments. Use of the product in residential/domestic environments may cause electromagnetic interference.
2. Connection of the instrument to a test object may produce radiations beyond the specified limit.
3. Use high-performance shielded interface cable to ensure conformity with the EMC standards listed above.
Safety Standard
IEC 61010-1:2010/ EN 61010-1:2010
Copyright © ITECH Electronics Co., Ltd.
iv
IT6100 User’s Manual
CONTENTS
Quality Certification and Assurance ................................................................................................................ i
Warranty ........................................................................................................................................................... i
Limitation of Warranty ..................................................................................................................................... i
Safety Symbols ................................................................................................................................................. i
Safety Precautions ........................................................................................................................................... ii
Environmental Conditions .............................................................................................................................. ii
Regulatory Markings ...................................................................................................................................... iii
Waste Electrical and Electronic Equipment (WEEE) Directive .................................................................... iii
Compliance Information ................................................................................................................................ iv
Chapter1 Installing the Instrument .............................................................................................................. 1
1.1 Verifying the Shipment.............................................................................................................................. 1
1.2 Instrument Size Introduction ..................................................................................................................... 1
1.3
Adjusting the Carrying Handle ................................................................................................................. 3
1.4 Racking Mount the Instrument ................................................................................................................. 3
1.5 Connecting the Power Cord ...................................................................................................................... 3
Chapter2 Quick Start ..................................................................................................................................... 5
2.1 Brief Introduction ...................................................................................................................................... 5
2.2 Front Panel Introduction ........................................................................................................................... 5
2.3 Keyboard Introduction .............................................................................................................................. 7
2.4 Rear Panel Introduction ............................................................................................................................. 8
2.5 Power-on Selftest ...................................................................................................................................... 9
2.6 Output Checkout ..................................................................................................................................... 11
Chapter3 Functions and Characteristics .................................................................................................. 13
3.1 Front-panel Operation Overview ............................................................................................................ 13
3.2 Setting Voltage......................................................................................................................................... 13
3.3 Setting Current ........................................................................................................................................ 14
3.4 Saving and Recalling Operation ............................................................................................................. 14
3.5 Menu Operation ....................................................................................................................................... 15
3.6 Over Voltage Protection Function ........................................................................................................... 22
3.6 Output Operation ..................................................................................................................................... 23
3.7 Remote Sense .......................................................................................................................................... 23
3.8 Milliohm Meter ....................................................................................................................................... 24
3.9 Voltage Meter .......................................................................................................................................... 24
Chapter4 Technical Specification .............................................................................................................. 26
4.1 Specifications .......................................................................................................................................... 26
4.2 Supplemental Characteristics .................................................................................................................. 27
Chapter5 Remote Operation Mode ........................................................................................................... 29
5.1 Communication Cable ............................................................................................................................. 29
5.2 Communication between Power Supply and Computer ......................................................................... 30
Appendix ............................................................................................................................................................... 31
Specifications of Red and Black Test Lines .................................................................................................. 31
Copyright © ITECH Electronics Co., Ltd.
v
Installing the Instrument
Chapter1 Installing the Instrument
1.1 Verifying the Shipment
Unpack the box and check the contents before operating the instrument. If wrong items have been delivered, if items are missing, or if there is a defect with the appearance of the items, contact the dealer from which you purchased the instrument immediately. The package contents include:
Checklist of Package Contents
Item
IT6100 power supply
Power cord
USB cable
CD
Ex-factory Test
Report
Qty. Model x1 x1 x1 x1 x1
IT6100 series
IT-E171/IT-E172
/IT-E173/IT-E17
4
-
-
-
Remarks
The IT6100 series include:
IT6121/IT6122/IT6123/
IT6151/IT6152/IT6153/IT6154
User may select an appropriate power cord that matches the specifications of power socket used in the area. See the
Section Connecting the Power
Cord for details.
-
It contains IT6100 power supply
User’s Manual, Programming
Guide and other user documentations.
It is the test report of the instrument before delivery.
Note
Upon verification of the shipment, keep the package and relevant contents thereof in a safe place. When returning the instrument for warranty service or repair, the specified packing requirements shall be met. Refer to IT6100 User’s Manual for detailed requirements on returns.
IT6100 series power supply has the following optional accessory:
IT-E151:19 inch rack mount
IT-E121:RS232 communication cable
IT-E122:USB communication cable
IT-E135:GPIB connector
IT-E123:RS485 communication cable
1.2 Instrument Size Introduction
The instrument should be installed at well-ventilated and rational-sized space.
Please select appropriate space for installation based on the power supply size.
Model IT6120 Series
Copyright © ITECH Electronics Co., Ltd.
1
Installing the Instrument
Overall dimensions
Width: 214.5 mm
Height: 88.2 mm
Depth: 354.6 mm
Detailed dimensional drawings
Model IT6150 Series
(Unit: mm)
Overall dimensions
Width: 429 mm
Height: 88.2 mm
Depth: 354.6 mm
Detailed dimensional drawings
Copyright © ITECH Electronics Co., Ltd.
2
Installing the Instrument
1.3
Adjusting the Carrying Handle
To adjust the position, grasp the handle by the sides and pull outward. Then, rotate the handle to the desired position.
Bench-top viewing positions Carrying position
1.4 Racking Mount the Instrument
You can mount the IT6120 and IT6150 series power supply in a standard 19-inch rack cabinet. For details, please see the corresponding rack mounting installation instruction.
1.5 Connecting the Power Cord
Before connecting the power cord, please ensure the power switch of the instrument is turned OFF. Only use the power cord supplied as a standard accessory. A summary of connection procedures is given below.
AC power input level
IT6120/IT6150 series power supply working voltage is 110V and 220V. Please pay
Copyright © ITECH Electronics Co., Ltd.
3
Installing the Instrument attention to the AC power input level as below.
AC power input level:
Option Opt.01: 220VAC ± 10%, 47 to 63 Hz
Option Opt.02: 110 VAC ± 10%, 47 to 63 Hz
Categories of power cords
IT6100 series power supply provides the standard power cords as below.
Please select appropriate power cords appropriate to local voltage based on the specifications of power cords below. If purchased model fails to meet local voltage requirements, please contact distributor or factory for change.
E
N
E
L N
E
L
China
IT-E171
E
L N
America,
Canada,Japan
IT-E172
Europe
IT-E173
L N
Britain
IT-E174
Copyright © ITECH Electronics Co., Ltd.
4
Quick Start
Chapter2 Quick Start
2.1 Brief Introduction
IT6100 Series power supplies are high resolution and high speed programmable
DC power supplies. This series have very small size of 2U, and they are equipped with 5 2
1
digital voltage meter and m
Ω
meter. That means you don’t need to invest extra money to purchase digital meter. This series offer flexible solution to general laboratory and workshop requirement.
VFD display
Liner programmable power supply with low ripple and noise
Very high accuracy and resolution of 0.1mV/0.1mA
High rise speed<30mS
Equipped with 5 2
1
digital voltage meter and m
Ω
meter
Small size can be installed in standard 19’’ rack or bench top
Remote control via USB/RS232/GPIB
Limit voltage protection
Over current/temperature protection
Free software for control the machine on computer
Model
IT6121
IT6122
IT6123
IT6151
IT6152
IT6153
IT6154
Voltage Current Power
20V 5A 100W
32V
72V
5.2V
20V
30V
60V
3A
1.2A
60A
27A
18A
9A
96W
86W
312W
540W
540W
540W
2.2 Front Panel Introduction
IT6120 Series’ Front Panel as shown below.
Copyright © ITECH Electronics Co., Ltd.
5
Quick Start
Fig 1.1 IT6120 series’ front panel
1
VFD display
2
Rotary knob
3
Power switch
4
Numeric keys and ESC key
5
Function keys
6
Up , down key and ENTER key
7
Output terminals
IT6150 Series’ Front Panel as shown below.
1 2
3 4 5 6 7
Fig 1.2 IT6150 series’ front panel
1
V FD display
2
Rotary knob
3
Power switch
Copyright © ITECH Electronics Co., Ltd.
8
6
Quick Start
4
Numeric keys and ESC key
5
Function keys
6
Up, down key and ENTER key
7
Input terminals
8
Output terminals
2.3 Keyboard Introduction
Detailed description of keys: keys
Power
0 ~ 9
Name and functions
Power on the instrument.
Numeric button
Dot mark.
Esc key. Exit under any conditions.
Set the voltage value
Set the current value
Save
Recall
On/Off
Shift
Save the currently data of the power supply to internal register
Recall the data from the internal register
Set the output state of power supply
Use it with multiple key
Enter
Up key. Select menu items or increase output voltage in menu operation.
Down key. Select menu items or decrease output voltage in menu operation.
Enter key.
Combination keys. Press (Shift) button first and then other keys to achieve all kinds functions in the following table.
Shift
Shift
+
+
( Menu)
( List)
Combine with
Shift
key to realize menu operation and set the corresponding parameters of power supply.
Set LIST operation parameters.
Copyright © ITECH Electronics Co., Ltd.
7
Quick Start
Shift
Shift
+ ( Trigger)
+ (
V/mΩ
)
Cause an immediate trigger.
Voltmeter/Milliohm meter
Shift Milliohm meter 0.1W range.
+ (0.1W)
Shift Milliohm meter 1W range.
+ (1W)
Shift
Shift
+ (10W) Milliohm meter 10W range.
Local key. Remote mode switch to local mode.
+ ( Local)
2.4 Rear Panel Introduction
IT6120 Series’ Rear Panel as shown below.
8-Pin Trigger and Remote sensing connectors
9-Pin COM port interface connector
Power switch (110V / 220V)
Cooling window
Voltage measurement terminals
Power socket
IT6150 Series’ Rear Panel as shown below.
Copyright © ITECH Electronics Co., Ltd.
8
Quick Start
1 9-Pin COM port interface connector
2 4-Pin Trigger and Remote sensing connectors
3 Power socket
4 Power switch (110V / 220V)
5
Cooling window
2.5 Power-on Selftest
A successful selftest indicates that the purchased power product meets delivery standards and is available for normal usage.
Before operation, please confirm that you have fully understood the safety instructions.
To avoid burning out, be sure to confirm that power voltage matches with supply voltage.
Be sure to connect the main power socket to the power outlet of protective grounding. Do not use terminal board without protective grounding. Before operation, be sure that the power supply is well grounded.
To avoid burning out, pay attention to marks of positive and negative polarities before wiring.
Selftest steps
Normal selftest procedures:
1. Correctly connect the power cord. Press Power key to start up.
2. After selftest, VFD display the output voltage and current status as below:
0.000V 0.00000A OFF
0.000V 3.000V
Copyright © ITECH Electronics Co., Ltd.
9
Quick Start
Error Information References
The following error information may occur when an error occurs during Power On self-test:
If the EEPROM was damaged, the VFD will display “EEPROM Error”.
If the lastest operation data in EEPROM is lost, then VFD will display “Error
Config Data”.
If the calibration data in EEPROM is lost, then VFD will display “Error
Calibration Data”.
If the factory calibration data in EEPROM is lost, and then the VFD will display
“Error Factory Cal.Data”.
Exception handling
If the power supply cannot start normally, please check and take measures by reference to steps below.
1. Check whether the power cord is correctly connected and confirm whether the power supply is powered.
Correct wiring of power cord => 2
Incorrect wiring of power cord => Re-connect the power cord and check whether the exception is removed.
2. Check whether the power in On. Power key is under “ ” On status.
Yes => 3
No => Please check the Power key to start power and check whether the exception is removed.
3. Check whether the fuse of power supply is burned out.
If yes, change fuse. Detailed steps:
1. Use a screwdriver to push and turn the fuse box on the rear panel of the power supply, refer to the below picture.
Fuse specification list
Model Fuse Sepcification
(110VAC)
IT6121
IT6122
IT6123
IT6151
T2.5A 250V
T2.5A 250V
T2.5A 250V
T10A 250V
IT6152
IT6153
IT6154
T10A 250V
T10A 250V
T10A 250V
Copyright © ITECH Electronics Co., Ltd.
Fuse Sepcification
(220VAC)
T1.25A 250V
T1.25A 250V
T1.25A 250V
T5A 250V
T5A 250V
T5A 250V
T5A 250V
10
Quick Start
2. After the fuse box is opened, you can see the fuse in it. Please replace with a fuse of the same specification..
2.6 Output Checkout
The following procedures check to ensure that the power supply develops its rated outputs and properly responds to operation from the front panel.
2.6.1 Voltage Output Checkout
The following steps verify basic voltage functions without load.
1. Turn on the power supply.
2. Enable the outputs
3. Set the voltage value
Set some different voltage values, then wait till the Meter mode to check if the
VFD displayed voltage value is the same as the set voltage value, and to check if the VFD displayed current value is nearly zero.
4. Ensure that the voltage can be adjusted from zero to the full rated value.
2.6.2 Current Output Checkout
The following steps check basic current functions with a short across the power supply’s output.
1. Turn on the power supply.
2. Disable the output
Press
On/Off key to ensure that the output is disabled.
3. Connect a short across (+) and (-) output terminals with an insulated test lead.
Use a wire size sufficient to handle the maximum current.
4. Enable the output
5. Adjust the voltage value to 1.0 volt
Adjust the voltage to 1.0 volt to ensure the power supply is in CC operation mode.
The CC announciator will turn on.
6. Adjust the current value
Set some different voltage values, then wait till the Meter mode to check if the
VFD displayed current value is the same as the set voltage value, and to check if the VFD displayed voltage value is nearly zero
Copyright © ITECH Electronics Co., Ltd.
11
Quick Start
7. Ensure that the current can be adjusted from zero to the full rated value
8.
Turn off the power supply and remove the short wire from the output terminals
Copyright © ITECH Electronics Co., Ltd.
12
Functions and Characteristics
Chapter3 Functions and Characteristics
This chapter elaborates on the functions and characteristics of IT6100 series power supply. Contents following sections:
Front-Panel Operation Overview
Setting Voltage
Setting Current
Saving and Recalling Operation
Menu Operation
Output Operation
Remote Sense
Milliohm Meter
Voltage Meter
3.1 Front-panel Operation Overview
The following section describes an overview of the front-panel keys before operating your power supply.
1. The power supply is shipped from the factory configured in the front-panel operation mode. At power-on, the power supply is automatically set to operate in the front-panel operation mode. When in this mode, the front panel keys can be used.
2. When the power supply is in remote operation mode, you cannot use the front-panel. A change between front-panel and remote operation modes will not result in any change in the output parameters. You can change the front-panel and remote operation modes by computer.
3. The output of the power supply can be enabled or disabled from the front panel by pressing
On/Off key.
4. The VFD display shows the present operating status of the power supply with announciators. Turn on the power supply, VFD displays two lines data. The first line shows the actual output voltage value, current value and the state of the power supply. The second line shows the voltage value tested by the voltage meter and the output value of the power supply.
3.2 Setting Voltage
The constant voltage range is from 0V to the maximum voltage value of each model. It is very easy for you to set the constant voltage output. You have 2 solutions to set the constant voltage value.
Solution1:
Step1: Power on the IT6100 series instrument.
Step2: Press the
▲
and
▼
keys to change the value.
Solution2:
Step1: Power on the IT6100 series instrument.
Copyright © ITECH Electronics Co., Ltd.
13
Functions and Characteristics
Step2: Press .
0
▲
and
▼
keys to change the voltage value.
Step4: Press
Solution3:
Enter to confirm.
Step1: Power on the IT6100 series instrument.
Step2: Press .
Step3: Use rotary knob to adjust the voltage value.
Setp4: Press
Enter to confirm.
3.3 Setting Current
The constant current output range is from 0A to the maximum current value of each type. It is very easy for you to set the constant current output.
Solution1:
Step1: Power on the IT6100 series instrument.
9
▲ and
▼ keys to change the current value.
Solution2:
Step1: Power on the IT6100 series instrument.
Step3: Use rotary knob to adjust the voltage value.
3.4 Saving and Recalling Operation
You can use
Save or
Recall
or use the SCPI order *SAV
、
*RCL to store up to
50 different output states in storage register locations(1 to 50). If the fast recall function is turned on, you can recall the saved settings by pressing numerical key
( 0~9 ) . You can store the followings: the maximum output voltage value, the voltage setting value, the current setting value and the step value of voltage.
Saving Operation: You can use
Save
and 0~50 numeric key, press and store the parameter of IT6100 series into the register.
Enter
Recalling Operation : You can use
Recall
and 0~50 numeric key, press and recall the parameter stored before from the register.
Enter
,
,
Copyright © ITECH Electronics Co., Ltd.
14
Functions and Characteristics
3.5 Menu Operation
Menu Description
Press MENU to indicate operation mode. View the menu in VFD, and use
▲
and
▼ to scroll through the complete menu list as following. If press ENTER , you could get the selected menu function. Press page.
ESC back to the previous menu selection
MENU
Config
Config Init.
Out Recall
On
Off<Default>
PWR-ON Recall
On
Off<Default>
Key Sound Set
On<Default>
Off
Knob Lock Set
Return to the factory default setup value.
Setting Power-on state of power supply.
When users turn on the power supply; its setup value will keep the state of last time as users turned off the power supply.
Disable this function.
Setting power-on parameter of power supply
When users turn on the power supply; its setup value will keep the state of last time as users turned off the power supply.
Disable this function.
Keypad sound setting.
Enable key sound.
Disable key sound
Setup Rotary knob lock state.
On
Off< Default >
Remote Sense
Lock Rotary knob.
Unlock Rotary knob.
Setup voltage measurement Mode.
On
Off< Default >
ShortCut Recall
The electronic load will measure input voltage from the remote sense connector.
The power supply will measure input value from the front panel connector.
Quickly recall the data stored before
On Enable this function
Off<Default> Disable this function
OVP Set
On
Off< Default >
Setting over voltage protection
Enable this function
Disable this function
Meter Rate Set
High
Setting the speed of power supply
High speed
Low< Default >
Baudrate set
Low speed
Setting baud rate
Baudrate 4800 <
Default >
Baudrate 9600
Baudrate 19200
Baudrate 38400
Comm. Parity Communication parity bit set.
None< Default >
Copyright © ITECH Electronics Co., Ltd.
15
Functions and Characteristics
Even
Odd
Address Set
Address=
**
Port Mode
Trigger< Def >
Setting communication address(range from 0to 254)
Select function of port
RI/DIF
DIGITAL I/O
Trig Source
Immediat<Def>
External
Bus
Setting trigger mode
Trigger signals from
Shift
+
Trigger key
Trigger signals from the TRIG connector in the rear panel.
Communication command trigger mode.
RI Mode
Off< Default >
Control the output mode
Disable this function
Latching
Live
DFI Source
System Set
Discrete Fault Indicator
Off< Default >
QUES
OPER
ESB
RQS
Question Bit
Operation Bit
Event State Bit
Require Bit
Key Lock Set Setting keypad password.
Press ENTER directly to disable the key lock function.
Password= ﹡﹡﹡﹡
Exit
Max Volt. set Setup the Maximum Voltage.
Max= ﹡﹡﹡﹡
Step Volt Set Set voltage step
Exit
List Set
Step= ﹡﹡﹡﹡
Call ListFile
Edit ListFile
Recall list operation file.
Continuous
Once
Edit list operation file.
Repeat
Step
Once
Repeat
Save Mode Set
8 X 25 Steps
4 X 50 Steps
2 X 100 Steps
Users can choose 4 kinds of memory space to save the list file.
Copyright © ITECH Electronics Co., Ltd.
16
Functions and Characteristics
Exit
1 X200 Steps
Timer time you set.
Timer State Setting POWER ON timer state
On When users choose the timer state ON, and then turn on the power supply’s output, the POWER ON TIMER will start working, and when the POWER ON TIMER is reach the setup time, the power supply’s output will turn off automatically.
Off< Default >
Timer Set Setting time of POWER ON timer.
Exit
Exit
Timer= ﹡﹡ S
Note:
Press
Shift
+
1
(Menu) to enter menu operation, then press
Esc
to exit. During any function operations, press
Esc key to exit the current operation state.
Output Recall
This function can help you set the output state when the power supply is powered on. If you select On , the power supply will keep the state of last time as it is turned off. If you select Off , this function is disabled. Default is On.
Key Sound
This instruction can switch on/off the buzzing sound when you press any key, if you select On , the buzzer will sound when any key was pressed. If you select Off , the buzzer will not sound when the keys were pressed. Default is On .
ShortCut Recall
This function can help you recall the data stored before.
Option:
1. Press press
Shift
ENTER
+
MENU to confirm. into menu operation, VFD displays
2. Press
▲
and
▼to select ShortCut Recall
, press ENTER to confirm.
3. Press
▲
and
▼to select
On, press
ENTER to confirm.
Config ,
4. Press
ESC two times to escape menu operation.
5. Press press
6. Press
ENTER or to confirm.
Save
, VFD displays Store 1
9
9
Copyright © ITECH Electronics Co., Ltd.
17
Functions and Characteristics number
( range from 1 to 50
)
, press ENTER to confirm.
7. You can recall the data stored before by pressing the register number.
Setting Communication Baud Rate (>BAUDRATE)
This function can help you set communication baud rate. There are 4 kinds of baud rate, 4800HZ, 9600HZ, 19200HZ and 38400HZ. Make sure that power’s baud rate is as the same as your computer before communication. Default baud rate is 4800HZ.
Setting Communication Parity Bit (>Comm.Parity)
This function can set communication parity bit. There are three kinds of communication parity bit, None, Even and Odd. Make sure that the communication parity bit set to None during remote communication operation. Default communication parity bit is None .
Setting Address (>ADDRESS)
This instruction can set the communication address for each power supply. The address range is from 0 to 30. Before the communication, you must make sure that there is same address between the power supply and the computer. Default address is 0.
Option:
1. Press press
Shift
ENTER
+
MENU to confirm. into menu operation, VFD displays
2. VFD displays Address=
* *
9 press ENTER to confirm.
Config ,
3. Press
ESC two times to escape menu operation.
Setting Port Mode
The level of the port on the rear panel is TTL, and the port has 3 kinds of function as follows:
TRIGGER : pin1 and pin 2 can be used as the external trigger source for power supply, and can control the list operation.
RI/DFI : Inhibit Input can control the output state of power supply, Fault Output can find fault of power supply.
DIGITAL I/O : It can read and control output and input state. mode Trigger RI/DFI DIGITAL I/O
Pin
1
2
3
4
Trigger in
GND
No Use
No Use
Inhibit Input
GND
Fault Output
GND
Digital Input
GND
Digital Output
GND
Copyright © ITECH Electronics Co., Ltd.
18
Functions and Characteristics
Setting Trigger Operation (Shift+Trigger)
The power supply has 3 kinds of trigger mode. You must choose trigger source before trigger operating.
Shift
Trigger Key: If this function is enabled, press will start trigger operation once.
+
Trigger
, the power supply
External trigger signal (TTL): There is a trigger input port on the rear panel.
When this function is enabled, please give this trigger input port a pulse about 5 mS, and the power supply will start trigger operation once.
Bus: When this function is enabled, and the power supply receives order *TRG or
TRIgger , the power supply will start trigger operation once.
Setting Remote Inhibit
RI input has 3 modes: LATCHING , L IVE , OFF
LATCHING : When the level of RI port changes from high to low, the output of power supply is off.
LIVE : The output state of power supply changes along with the level of RI port. If the level of RI is high, the output is on; and the level of RI is low, the output of power supply is off.
OFF : The level state of RI do not affect the output state of power supply.
Setting Discrete Fault Indicator
DFI source contains QUES 、 OPER 、 ESB 、 RQS 、 OFF.
QUES: The output level of DFI changes along with the state of QUES bit. When
QUES bit is 1, DFI outputs low level. When QUES bit is 0, DFI outputs high level.
OPER: The output level of DFI changes along with the state of OPER bit.
ESB: The output level of DFI changes along with the state of ESB bit.
RQS: The output level of DFI changes along with the state of RQS bit.
OFF: The output level of DFI remains high.
Setting KEY LOCK
This instruction can set a password (1 through 4 digits) to lock the function keys operation. After setting the password, all the function keys on the front panel will be locked except the OUT on/off key. You must enter the correct password to unlock them, then you can continue to do the function key operation. If you don’t want to lock the function keys, please don’t press any number key when you enter the >KEY LOCK instruction, just press ENTER key to unlock it.
Note: When shipped from factory, there is no password and function keys are unlocked. The start bit of your desired password shouldn’t be 0.
Setting Maximum Output Voltage (>Max Volt. Set)
The range of maximum voltage value is from 0V to full rated output voltage, the steps are as follows:
1. Press
Shift
+
1
(Menu) to enter menu function. VFD displays > Config ,
Copyright © ITECH Electronics Co., Ltd.
19
Functions and Characteristics
List Set
Press and to select > System Set , then press ENTER to confirm.
2. Press and to select > Max Volt. Set , then press ENTER to confirm.
3. VFD displays Max=73.000V
, select press ENTER to confirm.
0
-
9
to set maximum voltage value,
4. Press
Esc
twice time to exit menu operation.
Default value of Max Volt. Set is full rated output voltage.
Before you edit the list file, please set trigger source as “immediate” in the menu.
Action:
1. Press
Shift
+
Menu into the menu.
2. Press up or down key to select “Config” , press
Enter to confirm.
3. Press up or down key to select “Trig Source” , press
4. Press up or down key to select “Immediat” , press
Enter
Enter to confirm. to confirm.
You can make the input change order by editing every step value of list operation.
The parameter of list operation includes the name of list file, input step (no more than 400 steps), step time (the minimum is 1mS) and every step value. The list file can be stored in ROM whose capacity is 4K, and it can be fast recalled. This store area is divided into A, B, C and D four areas. In area A, there is only 1 group whose capacity is 4Kb. In area B, there are 2 groups and each group’s capacity is 2Kb. In area C, there 4 groups and each group’s capacity is 1Kb. In area D, there are 8 groups and each group’s capacity is 512b.
If the list operation mode is CONTINIOUS, when you receiving a trigger signal, the power supply will start list operation until the list operation is over or a trigger is received again.
Operation:
1. Press
Shift
+
Menu into menu operation.
2. VFD displays Config , press
▼to select List Set
, press
Enter to confirm.
3. VFD displays confirm.
Call ListFile , press
▼to select Edit ListFile
, press
Enter to
4. VFD displays Continuous , press
Enter to confirm.
5. VFD displays Once , press
▼ to select Repeat , press select circle operation.
Enter to confirm,
6. VFD displays Second , press
▼ to select MilliSecond , press
Enter to confirm, select time unit.
7. VFD displays List Count= _ , press numeric key or move the rotary knob, set circle times ( in this example, count is 2),press
Enter to confirm.
8. VFD displays 1th= * .
**** V , press numeric key or move the rotary knob, set the maximum voltage, press
Enter to confirm.
9. VFD displays 1th= * .
**** A , press numeric key or move the rotary knob,
Copyright © ITECH Electronics Co., Ltd.
20
Functions and Characteristics set the maximum current, press
Enter to confirm.
10. VFD displays 1th= * mS , press numeric key or move the rotary knob, set delay time, press
Enter to confirm.
11. VFD displays 2th= * .
**** V , press numeric key or move the rotary knob, set the maximum voltage, press
Enter to confirm.
12. VFD displays 2th= * .
**** A , press numeric key or move the rotary knob, set the maximum current, press
Enter to confirm.
13. VFD displays 2th= * mS , press numeric key or move the rotary knob, set delay time, press
Enter to confirm.
14. VFD display Store File_ , press numeric key or move the rotary knob, set the register number ( 1 to 8 ) , press
Enter to confirm.
15. Press
Esc
two times to escape menu operation.
16. Press
Shift
+ the list file. Press
List
Shift to set the list file, then press
+
List
to stop.
Shift
+
Trigger to run
If you have edited several list files, you can call the list file that you need by “Call
ListFile” function in the menu. Press then press
Shift
+
List
Esc
two times to exit menu operation. And to set it, press
Shift
+
Trigger
to run the file that you call. Press
Shift
+
List to stop running.
Trigger Trigger
When the list operation mode is STEP , the power supply will go to the next step when receive a trigger signal.
Copyright © ITECH Electronics Co., Ltd.
21
Functions and Characteristics
Trigger
Trigger
Trigger Trigger
Trigger
Trigger
Trigger Trigger
Setting Output Timer Function (>Out On Timer)
The loading time of IT6100 power supply can be set. When the timer state is ON, the power supply output opens and starts with load. When the loading time reaches the set time, the output will be turned off automatically. When the timer state is OFF, the timer function will be disabled. Default value is OFF.
The detailed operations are as follows:
Set timer state:
1. Press press
Shift
and
+
1
(Menu) to enter menu function, VFD displays > Config , to select > Out On Timer , then press
Enter to confirm.
2. Press and to select > Timer State , then press
Enter to confirm.
3. VFD displays Off<Default> (or On), press and to select On or Off, then press
Enter to confirm.
4. Press
Esc
twice time to exit the operation.
Set the time of output timer:
1. Press
Shift
+
1
(Menu) to enter menu function, VFD displays > Config , press and to select > Out On Timer , then press
2. Press and to select > Timer Set , then press
Enter
Enter to confirm. to confirm.
3. VFD displays confirm.
Timer=20S , press
0
-
9
to set time, then press
Enter to
4. Press
Esc
twice time to exit the operation.
Digital I/O
When the digital port of the power supply mode is DIGITALI/O and power supply is under remote sense, you could send SCPI order (DIGital:INPut[:STATe?] and
DIGital:OUTPut[:STATe?]) to read and set the state of output and input port.
3.6 Over Voltage Protection Function
When the output voltage exceeds the constant voltage by 5%, or the connection is unsuccessful under remote sense mode, the power supply will protect itself automatically. The power supply must be turned on again. When the power is in overvoltage protection, the output is off, and buzzer is mooing. VFD displays Over
Voltage.
Copyright © ITECH Electronics Co., Ltd.
22
Functions and Characteristics
Over Voltage
When the power supply with other power supply is in parallel connection, please disable this function.
3.6 Output Operation
If you control the power supply by front panel, you can press
On/Off
and switch the state of output. If power supply is under remote sense, you can send SCPI order (OUTPut ON|OFF) to change the state of output.
3.7 Remote Sense
The wire connected the power supply and electronic load will produce voltage difference when the current of the power supply is so big. In order to make the measurement accuracy much accurate, there are 4 pin trigger and remote sensing connectors on the rear panel and you can use it to measure the output voltage of the tested instrument. You must set remote sense mode before you start remote test function.
IT6120 8 pin connector in rear panel
+ +S
-
S INH GND FLT GND
+,: Output port, they are as the same as the output in front panel
+S,-S : Sense port
INH : this port hat multiple function
When “Port Mode” in the menu is set with “Trigger”, INH is trigger port.
When “Port Mode” in the menu is set with “RI/DIF”, INH can set the output status on/off. Here INH port has 3 working mode to choose.
LATCHING : When the level of INH port changes from high to low, the output of power supply is off.
LIVE : The output state of power supply changes along with the level of INH port. If the level of INH is 1, the output is on; and the level of INH is 0, the output of power supply is off.
OFF : The level state of INH do not affect the output state of power supply.
When “Port Mode” in the menu is set with “DIGITAL I/O”, INH port becomes the digital input port to read input status by communication order.
FLT: This port hat multiple function
When “Port Mode” in the menu is set with “Trigger”, FLT is an invalid port.
When “Port Mode” in the menu is set with “RI/DIF”, FLT port can find fault of power supply.
FLT source contains QUES, OPER, ESB, RQS, OFF .
QUES : The output level of FLT changes along with the state of QUES bit. When
Copyright © ITECH Electronics Co., Ltd.
23
Functions and Characteristics
QUES bit is 1, FLT outputs low level. When QUES bit is 0, FLT outputs high level.
OPER : The output level of FLT changes along with the state of OPER bit.
ESB : The output level of FLT changes along with the state of ESB bit.
RQS : The output level of FLT changes along with the state of RQS bit.
OFF : The output level of FLT remains high.
When “Port Mode” in menu is set with “DIGITAL I/O”, FLT port becomes output port to control the output status by communication order.
IT6150 4 pin connector in rear panel
GND TRIN
+
GND, TRIN : Trigger port
-, + : Remote sense port
3.8 Milliohm Meter
SOURCE METER can measure resistant value accurately and the biggest resistance it can measure is 1k
Ω.
In order to protect the resistor, please select the power range of the resistor before you measure it. Wiring diagram as follows:
There are 3 ranges to be chosen: 0.1W
、 1W 、 10W.
The accuracy of the measurement is <1%, the higher range you select, the more accurate result you can get.
Option:
1. Press
Shift
+V/m
Ω
( if VFD displays **.** V , press
Shift
+V/m
Ω
) m : ** , you can measure resistance value.
2. Press
Shift
+ 0.1W /1W /10W , you can set different range of milliohm meter.
3.9 Voltage Meter
Wiring diagram as follows:
Copyright © ITECH Electronics Co., Ltd.
24
Functions and Characteristics
Press press
Shift
Shift
+V/m
+V/m
Ω
( if VFD displays Range: * * ,please
Ω
) , VFD displays **.** V , you can begin to test the voltage value.
The accuracy is <0.1%.
Copyright © ITECH Electronics Co., Ltd.
25
Technical Specification
Chapter4 Technical Specification
4.1 Specifications
IT6120 Series DC Power Supply
Parameter
Output Ratings
( 0 °C~40 °C)
Load Regulation
±
(
% of output+offset)
Line Regulation
±
(
% of output+offset)
Setup resolution
Readback resolution
Setup accuracy, 12months,
(25 °C ± 5 °C)
±
(
% of output+offset)
Readback accuracy
12months,
(25 °C ± 5 °C)
±
(
% of output+offset)
Ripple
(20Hz ~20MHz)
Temperature coefficient,
(0 °C~40 °C)
±
(
% of output+offset)
Readback temperature, coefficient,
±
(
% of output+offset)
Dimension(mm)
Weight
IT6121 IT6122
Voltage 0 ~20V
Current 0~5A
LVP
Voltage
Current
Voltage
0 ~32V
0~3A
0 ~21V 0 ~33V
≤0.01
%+2mV
≤0.05
%+1mA
≤0.01
%+1mV
Current
≤0.05
%+0.1mA
Voltage 0.5mV
Current 0.1mA
Voltage
Current 0.05mA
0.1mV
0.01mA
Voltage
≤0.03%+3mV
Current
Voltage
≤0.05
%+2mA
≤0.02
%+3mV
IT6123
0~72V
0~1.2A
0 ~73V
≤0.01
%+2mV
≤0.05
%+0.3mA
≤0.01
%+2mV
≤0.05
%+0.05m
A
2mV
0.02mA
0.5mV
0.01mA
≤0.03
%+6mV
≤0.05
%+1mA
≤0.02
%+5mV
Current
Voltage
≤0.05
%+2mA
Voltage
≤
3mVp-p
≤
4mVp-p
Current
≤3mArms ≤
3mArms
≤0.02%+3mV
Current
≤0.05
%+2mA
Voltage
Current
≤0.02%+3mV
≤0.05
%+2mA
≤
0.05%+1mA
≤
5mVp-p
≤3mArms
≤0.02%+5mV
≤0.05
≤0.05
%+0.5mA
≤0.02%+5mV
%+0.5mA
214.5mmW×88.2mmH×354.6mmD
8 kg
IT6150 Series High Resolution Power Supply
Parameter
Output Ratings
( 0 °C~40 °C)
Voltage
Current
IT6151
0~5.2V
0~60A
IT6152
0~20V
0~27A
The model of remote sense±(%ofoutput
LVP 0~5.5V
Voltage <0.05%+30mV
Current <0.1%+10mA
IT6153
0~30V
0~18A
0~21V 0~31V
<0.05%+20mV<0.05%+15mV
<0.1%+5mA
IT6154
0~60V
0~9A
0~61V
<0.05%+10mV
<0.1%+2mA
Copyright © ITECH Electronics Co., Ltd.
26
Technical Specification
+offset)
Line Regulation
±(
% of output+offset)
Voltage
Current
<0.02%+1mV
<0.1%+1mA
Setup resolution
Readback resolution
Setup accuracy,12 months±(%of output
+offset)
Readback accuracy
±(%of output+offset)
Voltage
Current
Voltage
Current
Voltage
Current
Voltage
Current
0.1mV
1mA
0.1mV
1mA
0.02%+2mV
<0.1%+30mA
0.02%+1.5mV
<0.05%+15mA
Ripple
(20Hz~20MHz)
Temperature
Coefficient
( 0 °C~40 °C)
±(%of output+offset)
Voltage
Current
Voltage
Current
≤
4mVp-p
<15mArms
<0.02%+2mV
<0.1%+30mA
Readback temperature coefficient±(%of output+offset)
Voltage
Current
Dimension(mm)
<0.02%+2mV
<0.1%+30mA
Weight
<0.02%+1mV
<0.01%+1mA
0.5mV
1mA
0.1mV
0.1mA
<0.02%+6mV
<0.1%+15mA
0.02%+3mV
<0.05%+10mA
≤
4mVp-p
<5mArms
<0.02%+5mV
<0.1%+15mA
<0.02%+5mV
≤0.05%+10mA
<0.02%+2mV
<0.01%+0.1mA
0.5mV
1mA
0.5mV
0.1mA
≤0.02%+12mV
<0.05%+10mA
0.02%+6mV
<0.05%+5mA
≤
5mVp-p
<3mArms
<0.02%+10mV
<0.05%+5mA
<0.02%+10mV
≤0.05%+5mA
429mmW×88.2mmH×458.9mmD
29 kg
* The above specifications may be subject to change without prior notice.
4.2 Supplemental Characteristics
State Storage Memory : Fifty (50) user-configurable stored states
Recommended Calibration Interval: Once a year
AC Input Ratings (selectable via switch on the rear panel):
Option Opt.01: 220VAC ± 10%, 47 to 63 Hz
Option Opt.02: 110 VAC ± 10%, 47 to 63 H
Maximum power:
IT6120 Series
300VA
Cooling: Fan cooled
IT6150 Series
1000VA
Operating Temperature : 0 to 40 °C
Storage Temperature : -20 to 70 °C
Copyright © ITECH Electronics Co., Ltd.
27
Technical Specification
Environmental Conditions
Designed for indoor use in an installation category II, pollution degree 2 environment. Designed to operate at maximum relative humidity of 95% and at altitudes of up to 2000 meters.
Copyright © ITECH Electronics Co., Ltd.
28
Remote Operation Mode
Chapter5 Remote Operation Mode
The DB9 interface connector on the rear panel of the power supply can be transferred to RS-232 interface, the following information will tell you how to use the computer to control the output of the power supply.
5.1 Communication Cable
IT-E131 RS232 Communication Cable
The DB9 interface connector on the rear panel of power supply is TTL voltage level; you can use the communication cable (IT-E131) to connect the DB9 interface connector of the power supply and the RS-232 interface connector of computer for the communication.
Computer side TTL -E131) PS side
IT-E131 communication cable
Power
IT-E132 USB Communication Cable
The DB9 interface connector on the rear panel of power supply is TTL voltage level; you can use the communication cable (IT-E132) to connect the DB9 interface connector of the power supply and the USB interface connector of computer for the communication.
Power
IT-E135 GPIB Communication Cable
The DB9 interface connector on the rear panel of power supply is TTL voltage level; you can use the GPIB communication cable (IT-E135) to connect the DB9 interface
Copyright © ITECH Electronics Co., Ltd.
29
Remote Operation Mode connector of the power supply, and then connect the GPIB interface of the IT-E135 and computer with GPIB/IEEE 488 line for the communication.
IT-E135 outer communication adapter
COM interface of
Power supply GPIB line
IT-E135 ISOLATED
Serial /IEEE 488 Controller
Copyright © ITECH Electronics Co., Ltd.
30
Appendix
Appendix
Specifications of Red and Black Test Lines
ITECH provides you with optional red and black test lines, which individual sales and you can select for test. For specifications of ITECH test lines and maximum current values, refer to the table below.
Model
IT-E301/10A
IT-E301/30A
IT-E301/30A
IT-E301/60A
IT-E301/120A
IT-E301/240A
IT-E301/240A
IT-E301/360A
Specification Cross section Length
10A
30A
30A
60A
120A
240A
240A
360A
-
6mm
2
6mm
2
20mm
2
50mm
70mm
2
2
70mm
2
95mm
2
1m
1.2m
2m
1.5m
2m
1m
2m
2m
For maximum current of AWG copper wire, refer to table blow.
AWG
The
Maximum current value( A)
10
40
12 14
25 20
16
13
18
10
20 22
7 5
24 26 28
3.5 2.5 1.7
Note: AWG ( American Wire Gage), it means X wire ( marked on the wire). The table above lists current capacity of single wire at working temperature of 30°C.
For reference only.
Copyright © ITECH Electronics Co., Ltd.
31
Contact US
Thank you for purchasing ITECH products. If you have any doubt about this product, please contact us as follow.
1. Please refer to the CD-ROM of related user’s manual in package.
2. Visit ITECH website www.itechate.com
.
3. Select the most convenient contact for further consultation.
Advertisement
Key features
- High Resolution-10mV Voltage & 1mA Current Resolution
- Rich Interface Options- USB, RS-232, and LAN
- Multiple Protection Functions
- Compact Size & Large Color Display
- Suitable for Continuous Operation, Built-in Cooling Fan