Ohio Medical PTS-CVR Vacuum Regulator Service Manual

Ohio Medical PTS-CVR Vacuum Regulator Service Manual
Add to My manuals

The Ohio Medical PTS-CVR is a lightweight, compact unit used throughout the hospital primarily for pharyngeal/tracheal suctioning (airway management). Various models provide regulated or full-line vacuum for hospital suction procedures. Each has a positive pressure safety relief valve to prevent pressurization by either failed injector vacuum (Venturi) units or inadvertent cross connection to pressurized gasses. In addition, the Low PTS-CVR models include a vacuum relief valve to limit maximum suction.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Ohio Medical PTS-CVR Service Manual | Manualzz

®

Vacuum Regulator, Push-To-Set

TM

Continuous (PTS-CVR)

Service Manual

Adult, Pediatric, Neonatal, C.A.S.S.

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

600

400

300

500

60

50

40

80

70

High Flow

High Vacuum

30

20

200

700

90

100

10

100

- kPa

-mm

Hg

®

Two-Mode

ANSI

Analog

Adult (Standard) - (1225)

Adult (High) - (1330)

C.A.S.S. - (1227)

Neonatal - (1231)

Pediatric - (1234)

Digital

Adult (Standard) - (1325)

Adult (High) - (1430)

C.A.S.S. - (1327)

Neonatal - (1331)

Pediatric - (1334)

ISO

Analog

Adult (High) - (1230)

Neonatal - (1232)

Pediatric - (1235)

Digital

Adult (High) - (1337)

Neonatal - (1332)

Pediatric - (1335)

MAX

600

400

300

500

80

70

60

50

40

High Flow

High Vacuum

30

20

700

90

100

10

200

100

- kPa

-mm

Hg

®

MAX

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

®

MAX

Three-Mode

ANSI

Analog

Adult (Standard) - (1224)

Adult (High) - (1226)

Digital

Adult (Standard) - (1324)

Adult (High) - (1326)

ISO

Analog

Adult (High) - (1229)

Digital

Adult (High) - (1329)

8700-0006-000 (Rev.8) 10/2014

User Responsibility

This Product will perform as described in th is operating manua l and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective product should not be used. Parts that are broken, missing, worn, distorted or contaminated should be replaced immediately.

For service advice, Ohio Medical recommends that a telephone request be made to the nearest Ohio Medical Regional Service Center. This product and any of its parts should only be repaired using written instructions provided by Ohio Medical or by Ohio Medical trained personnel. The

Product must not be altered without the prior written approval of Ohio Medical’s Quality Assurance Department.

The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Ohio Medical.

AAA A 12345 This alpha character indicates the year of product manufacture and when the serial number was assigned; “L” = 2007, “M” = 2008, “N” = 2009, etc. “I” and “O” are not used.

Safety Instructions

This manual provides you with important information about the Push-To-Set

TM

Continuous Vacuum Regulator

(PTS-CVR). To ensure the safe and proper use of this device, READ and UNDERSTAND all of the safety and operating instructions. IF YOU DO NOT UNDERSTAND THESE INSTRUCTIONS, OR HAVE ANY QUESTIONS,

CONTACT YOUR SUPERVISOR, DEALER OR THE MANUFACTURER BEFORE ATTEMPTING TO USE THE

DEVICE.

Intended Use

The vacuum regulator is intended to be used in the medical facility as a means to evacuate media (i.e. fluids) from the body. DO NOT use this vacuum regulator for anything other than its intended use.

2 8700-0006-000 Rev.8 (10/2014)

1/Precautions

1.1 Definitions .............................................................. 4

1.2 Warnings ................................................................ 5

1.3 Cautions ................................................................. 5

2/Scope

2.1 ANSI Vacuum Regulators (North American) .......... 6

2.2 ISO Vacuum Regulators (International) ................. 6

3/Description and Specifications

3.1 Description ............................................................. 7

3.2 Specifications ......................................................... 8

4/Cleaning, Disinfection and Sterilization

4.1 Cleaning and Disinfection ...................................... 9

4.1.1 Routine Exterior Cleaning ............................. 9

4.1.1.1 Approved Cleaning Solutions ........... 9

4.2 Sterilization ............................................................ 9

5/Troubleshooting

................................................ 10

6/Service - Disassembly and Assembly

6.1 Service Tools and Equipment .............................. 12

6.2 Continuous Vacuum Regulators (All Models) ...... 12

6.2.1 Disassembly ............................................... 12

6.2.2 Assembly .................................................... 12

6.3 Suction Level Limit Setting (PED/NEO/C.A.S.S.

Models ONLY) ............................................. 13

6.3.1 Vacuum Relief Valve Adjustment ................ 13

6.3.2 Vacuum Limiting Set Screw Adjustment

(PED and NEO Units ONLY) .............. 13

6.3.2.1 Pediatric Unit ................................. 13

6.3.2.2 Neonatal Unit ................................. 13

6.4 Regulator Module ................................................ 13

6.4.1 Disassembly ............................................... 14

6.4.2 Assembly .................................................... 14

6.5 Digital Gauge ....................................................... 14

6.5.1 Removal of Batteries .................................. 14

6.5.2 Inserting Batteries ....................................... 14

Table of Contents

7/Service Checkout Procedure

7.1 Set-up .................................................................. 15

7.2 Push-To-Set

TM

Test............................................... 15

7.3 Gauge Test ........................................................... 15

7.3.1 High Vacuum Gauges ONLY ...................... 15

7.3.2 Standard and PED/NEO/C.A.S.S. Vacuum

Gauges ONLY ............................................. 15

7.4 Regulation Test .................................................... 16

7.5 Maximum Vacuum Test (PED/NEO/C.A.S.S.

Models ONLY) ............................................. 16

7.6 Bleed Test ............................................................ 16

7.7 Leak Test ............................................................. 16

7.7.1 Supply Side ................................................ 16

7.7.2 Patient Side ............................................... 17

7.8 Flow Test .............................................................. 17

8/Maintenance

8.1 General Maintenance of Suction Equipment ....... 18

8.2 Recommended Maintenance Schedule ............... 18

8.2.1 Maintenance Schedule ............................... 18

8.3 Repair Policy ........................................................ 18

8.4 Technical Assistance ............................................ 19

8.5 Return Instructions ............................................... 19

8.6 Installation Procedure for Adapters/Probes and Fittings .......................................................... 19

9/Ordering Information

9.1 Illustrated Parts .................................................... 20

8700-0006-000 Rev.8 (10/2014) 3

1/Precautions

1.1 Definitions

Note: A Note provides additional information to clarify a point in the text.

Important: An Important statement is similar to a note but of greater emphasis.

CAUTION: A CAUTION statement is used when the possibility of damage to the equipment exists.

WARNING: A WARNING statement is used when the possibility of injury to the patient or the operator exists.

High Flow

High Vacuum = high flow, high vacuum

High Flow

Low Vacuum

| (ON)

= high flow, low vacuum

= on

O (OFF) = off

Consult Instructions for Use

Manufacturer

Abbreviations Used In This Manual

CCW

CW

CVR

MAX inHg kPa

LPM mmHg

Counter-clockwise (decrease)

Clockwise (increase)

Continuous Vacuum Regulator

Full Line Vacuum

Inches of mercury

Kilopascals (kPa x 7.50 = mmHg)

Liters per minute

Millimeters of mercury (mmHg x 0.133 = kPa) mL Milliliters

°C

°F

N-m ft-lb

Degrees Celsius

Degrees Fahrenheit

Newton-Meter (N-m x 0.737 = ft-lb)

Foot-Pound Force (ft-lb x 1.356 = N-m) oz Ounces

DISS

OES

NCG

NPT

Diameter Index Safety System

Oxequip

®

Suction

National Compressed Gases (Chemetron)

National Pipe Thread (USA)

NPTF

MPTS

National Pipe Thread Female (USA)

Multi-Purpose Therapy Stand gal Gallon

PTFE Polytetrafluoroethylene

PTS Push-To-Set

TM

ETO

IFU

Ethylene Oxide

Instructions for Use

4 8700-0006-000 Rev.8 (10/2014)

1/Precautions

1.2 Warnings

This device should be repaired only by qualified Ohio

Medical or Ohio Medical-trained, qualified personnel, using only Ohio Medical recommended parts. There are risks associated with using anything other than Ohio Medical parts. Ohio Medical will assume no responsibility for incidents which may occur if the product was not repaired in accordance with procedures authorized by Ohio Medical.

If the vacuum regulator is repaired or disassembled in any manner, the service checkout procedure must be performed before using the equipment on a patient.

After patient use, if regulator is contaminated then handle in accordance with you hospital’s infection control policy.

To reduce transportation personnel and/or service personnel exposure to hazardous contamination,

DO NOT ship any suction equipment that has been contaminated.

Do not use this device in the presence of flammable anesthetics. Static charges may not dissipate and a possible explosion hazard exists in the presence of these agents.

Connection to positive pressure sources such as oxygen and medical air, even momentarily, could injure the patient or operator.

Ohio Medical will assume no responsibility for incidents which may occur if the product is not used in accordance with product labeling.

To help prevent aspirate from entering the device, wall outlet and pipeline equipment, a safety trap should be attached prior to its use. Aspirate in the regulator, wall outlet and pipeline equipment may impair its operations. The use of the safety trap and suction filter will help prevent this and extend the life of suction equipment.

1.3 CAUTION

Do not use any Loctite® products or any products which contain Methacrylate Ester as an active ingredient to seal the threads on the adapters/probes and fittings.

Use of lubricants other than recommended may degrade plastic or rubber components.

Do not steam autoclave or liquid sterilize the regulator. Severe impairment to the operation of the

8700-0006-000 Rev.8 (10/2014)

regulator will result. The only acceptable method of sterilization is with gas (ethylene oxide).

Sterilization with ethylene oxide mixtures may cause crazing (minute superficial cracking) of some plastic parts. Crazing will be more pronounced when mixtures containing Freon® are used.

Following sterilization with ethylene oxide, parts should be quarantined in a well ventilated area to allow dissipation of residual ethylene oxide gas absorbed by the material. Aerate parts for 8 hours at 54°C (130°F). Follow your hospital sterilization procedure.

Do not use harsh chemical or cleaning solution. Do not spray cleaners directly onto suction regulators.

Only use chemical recommended in this manual.

If any evidence of damage is found, repair as necessary or contact your authorized service provider.

Connection to positive pressure sources such as

oxygen and medical air, even momentarily, could damage the equipment.

The suction control knob must be completely pushed in to adjust the vacuum level. Failure to do so may damage the vacuum regulator.

Not for transport use: The categories of field and transport user are specifically defined in ISO 10079-

3. “Field” means use at accidents or emergencies outside a hospital. “Transport” means use in ambulances, cars and airplanes. These situations may expose the equipment to uneven support, water, dirt, and mechanical shock and temperature extremes. Ohio Medical Suction equipment has not been tested to comply with the specific requirements of these categories.

Note: Ohio Medical requests that parties acquiring this device:

Report the device’s purchase, receipt in trade, return after sale, loss, destruction, or retirement.

Contact your Ohio Medical customer service representative to obtain manual updates.

Authorized Service Center / Customer Service

Call 1-866-549-6446 or +1 847-855-0500 for service and repair information.

5

2/Scope

This service manual contains service, maintenance and parts information on all models of the Push-To-Set

TM

Continuous

Vacuum Regulator.

2.1 ANSI Vacuum Regulators (North American)

®

®

F

H

M

L

MEDICAL VACUUM

F

H

M

L

®

®

MAX MAX

®

H

F

M

L

®

MEDICAL VACUUM

F

H

M

L

MAX

200

300

400

-mm Hg

500

600

100

700

®

H

F

M

L

MEDICAL VACUUM

F

H

M

L

®

LO

W

40

60

20

MED

100 mm Hg

140

160

C.A.S.S.

®

H

F

M

L

MEDICAL VACUUM

F

H

M

L

®

C.A.S.S.

Analog (1225) Digital (1325)

Standard Two Mode

Continuous

Adult

MAX

MAX

200

300

400

-mm Hg

500

600

100

700

MAX

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

Analog (1224) Digital (1324)

Standard Three Mode

Continuous

Adult

MED

W

LO

40

60

20

80

100

120

Low Vacuum

140 mm Hg

160

®

H

F

M

L

®

MEDICAL VACUUM

F

H

M

L

Analog (1330) Digital (1430)

High Two Mode

Continuous

Adult

MED

LO

W

40

20

60

80

100

120

Low Vacuum

140 mm Hg

160

HIG

H

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

Analog (1227) Digital (1327)

Standard Two Mode

Continuous

C.A.S.S.

Analog (1226) Digital (1326)

High Three Mode

Continuous

Adult

Analog (1231) Digital (1331)

Low Two Mode

Continuous

Neonatal

Analog (1234) Digital (1334)

Low Two Mode

Continuous

Pediatric

2.2 ISO Vacuum Regulators (International)

600

400

300

500

80

70

60

50

40

High Flow

High Vacuum

30

20

90 10

200

100

700 100

- kPa

-mm

Hg

®

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

MAX

600

700

500

80

400

70

300

60

50

40

High Flow

High Vacuum

30

20

200

90

100

10

100

- kPa

-mm

Hg

®

MAX

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

100

120

140

20

18

16

14

80

12

10

8

6

4

2

60

40

20

High Flow

Low Vacuum

- k

Pa

-m m

Hg

®

H

F

M

L

®

MEDICAL VACUUM

F

H

M

L

140

120

100

20

18

16

14 12

10

80

8

6

2

4

60

40

20

High Flow

Low Vacuum

- k

Pa

-m m

Hg

®

F

H

M

L

®

MEDICAL VACUUM

F

H

M

L

6

Analog (1230) Digital (1337)

High Two Mode

Continuous

Adult

Analog (1229) Digital (1329)

High Three Mode

Continuous

Adult

Analog (1232) Digital (1332)

Low Two Mode

Continuous

Neonatal

Analog (1235) Digital (1335)

Low Two Mode

Continuous

Pediatric

8700-0006-000 Rev.8 (10/2014)

3/Description and Specifications

3.1 Description

The PTS-CVR is a lightweight, compact unit used throughout the hospital primarily for pharyngeal/tracheal suctioning

(airway management). Various models provide regulated or full-line vacuum for hospital suction procedures.

There are several models of the PTS-CVR. All models contain a vacuum gauge which indicates suction supplied by the regulator. Each has a positive pressure safety relief valve to prevent pressurization by either failed injector vacuum

(Venturi

®

) units or inadvertent cross connection to pressurized gasses. In addition, the Low PTS-CVR models include a vacuum relief valve to limit maximum suction.

Some models operate in a regulated or non-regulated MAX mode. Others operate only in the regulated mode.

In the non-regulated MAX mode, the vacuum source is connected directly to the fitting/patient port. The regulator module is bypassed and full-line vacuum is provided.

In the regulated mode, the vacuum source is connected through the regulator module which functions as an automatic valve. Turning the suction control knob adjusts the position of the regulator module and allows selection of a predetermined level of suction when set according to instructions.

During use, as the flow requirement increases, the valve automatically opens to maintain suction at the preset level.

When the flow requirement decreases, the valve automatically closes to maintain suction at the preset level. The same mechanism compensates for changes in supply vacuum and automatically maintains the pre-set suction level when set according to instructions.

1. Suction Control Knob - Allows easy adjustment of suction to the patient.

2. Mode Selector Switch - Allows quick and easy mode changes.

a. | (ON) - Suction can be adjusted with the suction control knob.

b. O (OFF) - No suction is supplied to the patient.

c. MAX - Maximum full-line vacuum is supplied to the patient.

3. Vacuum Gauge - The suction level to the patient is displayed during use.

Mode

Selector

Switch

MAX

Vacuum

Gauge

Adapter/Probe

Port

®

Suction

Control

Knob

Fitting/Patient Port

8700-0006-000 Rev.8 (10/2014) 7

3/Description and Specifications

3.2 Specifications

Performance

Flow rate

1

Positive Pressure

Safety Relief Valve:

Gauge Accuracy

Standard

High

Pediatric

Neonatal

Standard

80 LPM

High

80 LPM

Pediatric

80 LPM

Neonatal

80 LPM

C.A.S.S.

80 LPM

Located in the vacuum supply line to prevent pressurization of the patient connection by failed injector vacuum (Venturi) units, or inadvertent cross connection to pressured gases.

Gauge Range

0-200 mmHg (0-26.7 KPA)

Analog

2

±5% Full-scale

Digital

3

±1% Full-scale

0-760 mmHg (0-101.3 KPA) ±5% Full-scale ±1% Full-scale

0-160 mmHg (0-21.3 KPA)

0-160 mmHg (0-21.3 KPA)

0-100 mmHg (0-13.3 KPA)

0-160 mmHg (0-21.3 KPA)

±5% Full-scale

±5% Full-scale

±5% Full-scale

±5% Full-scale

±1% Full-scale

±1% Full-scale

±1% Full-scale

±1% Full-scale

C.A.S.S.

Physical

Dimensions

6.7”H x 2.8”W x 4.8”D

(17.02 cm x 7.1 cm x 12.2 cm)

1 lb 4 oz (0.57 kg)

Weight

Battery

Lithium

Two 2/3 AA batteries, 3.6V, 1.6 Ah, lithium

1 without fittings at full increase setting depending on supply vacuum and open air flow.

2 full scale deflection

3 full range at 22°C

ENVIRONMENTAL SPECIFICATIONS

Operating Temperature Range:

Storage Temperature Range:

50 to 104ºF (10 to 40ºC)

-13 to 158ºF (-25 to 70ºC)

Operating and storage Relative Humidity: 5 to 95% RH (Non-condensing)

Ingress Protection Ratings: IP 20

8 8700-0006-000 Rev.8 (10/2014)

4/Cleaning, Disinfection and Sterilization

4.1 Cleaning and Disinfection

WARNING: After patient use, regulators may be contaminated. Handle in accordance with your hospital’s infection control policy.

CAUTION: Suctioned fluids drawn into a vacuum regulator do not stop in the regulator. They proceed through it into the wall outlet and pipeline system. Failure to clean and disinfect the wall outlet and pipeline system may result in damage to this equipment.

4.1.1 Routine Exterior Cleaning

Routine cleaning of the regulator is recommended as a standard procedure after each use. Wipe all exterior surfaces with a solution of water and mild detergent and/or an approved cleaning solution.

4.1.1.1 Approved Cleaning Solutions

• Sodium Hypochlorite 0.5% (Bleach): Mixture of 13 fl. oz. of bleach to 1 gallon (128 fl. oz.) tap water

• Isopropyl Alcohol 70%

• Hydrogen Peroxide 3%

• Cavicide

®

Ready to use full strength

4.2 Sterilization

Should misuse occur, resulting in accidental flooding of the regulator, the regulator may be sterilized using Ethylene oxide

(ETO). After sterilization, follow the service checkout procedures (Section 7 Service Checkout Procedure).

WARNING: Following sterilization with ethylene oxide, parts should be quarantined in a well ventilated area to

allow dissipation of residual ethylene oxide gas absorbed by the material. Aerate parts for 8 hours at 130°F

(54°C). Follow your hospital sterilization procedure.

CAUTION: Do not steam autoclave or liquid sterilize the intermittent vacuum regulator. Severe impairment to the operation of the regulator will result. The only acceptable method of sterilization is with gas (ethylene oxide).

CAUTION: Sterilization with ethylene oxide mixtures may cause crazing (minute superficial cracking) of some plastic parts. Crazing will be more pronounced when mixtures containing Freon

®

are used.

Note: The vacuum regulator should only be sterilized if it is contaminated or maintenance is to be performed.

8700-0006-000 Rev.8 (10/2014) 9

5/Troubleshooting

SYMPTOM

Unit fails to provide vacuum at patient port in all modes.

POSSIBLE CAUSES POSSIBLE SOLUTION

Mode Selector Switch is at O(OFF) position Move mode selector switch [34] to desired mode.

No Supply Vacuum

Damaged/Clogged Supply side or Patient side fittings

Check Hospital vacuum level (should be at least 500 mmHg (66.7 KPA).

Replace fittings if damaged/clogged

Retighten/Torque if found loose

Knob fails to return when pushed

Knob stuck at OFF position (all the way counter clock wise direction)

Push and rotate the knob [2] if stuck

Replace actuator [6], drive gear [5],

Regulator Case [3], OTS assembly [10].

Push and Rotate Knob [2] to clock wise direction

Damaged Regulator Module

Clogged external filter

Regulator Module Assembly [12]

Replace external filter

Overflow protection device is shut off

Loose or incorrect set-up

Aspirant draw into unit

Gauge assembly is not properly aligned.

Missing/Damaged O-Rings

Gauge doesn’t respond to changes in suction

Gauge assembly is damaged

Check(Reset float)/Replace Overflow protection device

Set-up unit correctly without any leak.

Follow your facilities procedures for handling contaminated products. (DO NOT SEND

UNIT BACK TO THE MANUFACTURER)

Ensure gauge assembly [6] is properly aligned.

Replace O-Rings [9]

Replace gauge assembly [6]

Digital Unit Only:

Depleted batteries

Unable to push and rotate knob

Suction level cannot be adjusted

Damaged OTS assembly

Regulator module is stuck in full OFF or full

ON position

Snap cap separated from regulator module housing.

Damaged/Missing spring inside regulator module assembly

Damaged/Missing O-ring, Quad-ring on regulator module

Damaged Diaphragm inside regulator module

Damaged Regulator Module Assembly

Digital Unit Only:

Replaced Gauge [6]/Batteries

Push and rotate knob [2] if stuck

Replace actuator [6], drive gear [5],

Regulator Case [3], OTS assembly [10].

Replace OTS assembly [10]

Rotate the knob [2] to free the Regulator

Module Assembly [12]

Reassemble snap cap [28] onto regulator module housing [34]

Replace snap cap [28] and/or regulator module housing [34]

Replace Regulator Module Assembly [12]

Replace Regulator module assembly [12]

Replace O-ring [35] and/or Quad-ring [36]

Replace Diaphragm [30] or regulator module assembly [12]

Replace Regulator module assembly [12]

10 8700-0006-000 Rev.8 (10/2014)

SYMPTOM POSSIBLE CAUSES

Damaged Diaphragm inside regulator module

Erratic gauge movement resulting from regulator adjustment

Damaged/Missing O-ring, Quad-ring on regulator module

Gauge assembly is damaged

Knob stuck in full OFF or full ON position

Excessive force applied

Damaged drive gear

Suction level can be adjusted without pushing the knob in

Damaged actuator

Damaged Knob

5/Troubleshooting

POSSIBLE SOLUTION

Replace Diaphragm [30] or regulator module assembly [12]

Replace O-ring [35] and/or Quad-ring [36]

Replace gauge assembly [6]

Rotate the knob [2] to free the Regulator

Module Assembly [12]

Replace drive gear [5]

Replace actuator occlude-to-set [6]

Replace knob [2]

Damaged gauge assembly Replace gauge assembly [6]

Gauge won’t zero

Whistling/buzzing noise from the unit

Need to apply grease on stem on regulator module

Blocked bleed hole orifice

Apply grease on stem [31]

Clean bleed hole orifice [32]

Gauge responds slowly to changes in suction/ failed bleed down test

Damaged gauge assembly

Damaged/Clogged Supply side or Patient side fittings

Damaged regulator case

Replace gauge assembly [6]

Replace fittings if damaged/clogged

Retighten/Torque if found loose

Replace regulator case [3]

Vacuum relief valve activates below specified range

{PED and NEO ONLY}

Damaged diaphragm

Damaged/loose set screw

Damaged vacuum relief assembly

Vacuum relief valve does not activate at specified range

{PED and NEO ONLY}

Damaged vacuum relief assembly

Replace diaphragm [30]

Replace/Adjust set screw 8700-0007-400.

Replace vacuum relief assembly

6700-0045-700.

Replace vacuum relief assembly and follow

Section 6.3 suction level limit setting.

8700-0006-000 Rev.8 (10/2014) 11

6/Service - Disassembly and Assembly

6.1 Service Tools and Equipment

CAUTION: Use of lubricants other than recommended may degrade plastic or rubber components.

The following items should be on hand during any service procedure.

• Supply vacuum: 500 mmHg ± 10 mmHg (67 kPa ±

1.3 kPa) & 80 LPM open air flow minimum

• Supply Vacuum Regulator with Gauge, 760 mmHg

(101.3 kPa) Full Scale.

• Low Vacuum Calibration Gauge, 225 mmHg (30 kPa)

Full Scale* (Ohio Medical P/N 6700-0353-800)

• High Vacuum Calibration Gauge, 760 mmHg (101.3 kPa) Full Scale*(Ohio Medical P/N 6700-0352-800)

• 80 LPM Flowmeter

• Phillips Head Screwdriver, No. 2

• Flat Head Screwdriver, 1/4 inch

• Tubing Clamp

• Bubble Leak Tester

• Tweezers (Filter Remover)

• Wooden Tooth Pick (O-ring Remover)

• Phillips head screwdriver No.1

• Dow Corning

®

111 grease (Ohio Medical P/N 6700-

0074-200)

• Loctite

®

242 removable thread locker (Ohio Medical

P/N 0220-5016-300)

• Stop watch or vacutimer (Ohio Medical P/N 6700-

0438-800)

• Open end adjustable wrench

• 10 LPM flowmeter

(*) Accuracy: ±1% of full scale deflection

6.2 Continuous Vacuum

Regulators (All Models)

6.2.1 Disassembly

WARNING: If the Vacuum Regulator is repaired or disassembled in any manner, the Service

Checkout Procedure must be performed before using the equipment on a patient.

WARNING: Clean and sterilize all suction equipment if contaminated before disassembly to ensure service personnel are not exposed to hazardous contamination.

WARNING: When servicing a Low Vacuum Regulator, perform the Vacuum Relief Valve Adjustment and Low Vacuum Limit Setting Procedure.

CAUTION: The gauge assembly must be handled with utmost care to retain its precision. If the lens is removed, do not rest the gauge on its face.

Note: See Section 9.1 Illustrated Parts

1. Remove the four cover screws from the back of the regulator.

2. With regulator facing up, carefully pull the cover assembly off the back body.

3. To remove the gauge assembly, grasp the assembly and pull straight out. The snap-fit lens can also be removed (if applicable) for replacement.

4. Remove actuator and PTS assembly.

5. To remove the regulator module from the manifold assembly, rotate the regulator module clockwise.

6. Using tweezers, remove the filter. Replace with a new filter.

7. Grasp the positive pressure safety relief valve, and pull it from the back body.

8. Remove the mounting screws from the actuator support bracket and remove the support bracket, detent plate and selector switch.

9. Low models ONLY: Grasp the vacuum relief valve and pull it from the back body.

6.2.2 Assembly

CAUTION: To prevent stripping the plastic threads, place the screw in the hole and turn counter-clockwise until it drops into the original thread, then tighten the screws.

1. Place the positive pressure safety relief valve in position and push onto the back body.

2. Low models ONLY: Lubricate the vacuum relief valve

O-ring and install it in the valve housing.

3. Lubricate switch with a thin coat of Dow Corning

111 on the overmolded area. Place in manifold with grease side down.

4. Add detent plate making sure the fingers are on either side of the switch shaft and pointing down.

5. Place O-ring on post coming up through center of detent plate and set actuator support bracket over

O-ring. Tighten down with 4 screws.

6. Place cylindrical white filter in corresponding cavity.

7. Screw regulator module counter-clockwise into manifold.

8. Place PTS assembly in manifold directly below regulator module.

12 8700-0006-000 Rev.8 (10/2014)

9. Set actuator in brackets of the actuator support bracket and rest the loop on the top of the PTS assembly.

10. Place gear over regulator post and rest it on the actuator.

11. Place lens on gauge with single tab at the top and the two tabs at the lower portion of the gauge and set gauge into support bracket.

12. Place cover strip in curved slot inside case.

13. Mate the cover with the manifold. With both thumbs positioned just below the lens, press firmly to snap the lens in.

14. Install the 4 cover screws.

6.3 Suction Level Limit Setting

(PED/NEO/C.A.S.S. Models ONLY)

WARNING: This Low Vacuum Limit Setting Procedure must be followed when repairing Low Vacuum

Regulators. Failure to do this may result in suction exceeding the prescribed level.

6.3.1 Vacuum Relief Valve Adjustment

1. Remove the cover (see Section 6.2.1 Disassembly).

2. Connect the supply vacuum to the adapter/probe port.

3. Connect an external vacuum test gauge to the patient port.

4. Move the mode selector switch to | (ON).

5. Slowly rotate the regulator module to increase the vacuum level. Note the suction level at which the relief valve opens. The relief valve should open at

140 mmHg ± 5 mmHg (18.7 kPa ± 0.7 kPa) on the external test gauge.

6. If adjustment is required, grip the vacuum relief valve firmly with pliers and rotate the screw about 1/8 turn with a screwdriver. Repeat the previous step and check the suction level at which the valve opens.

Clockwise rotation will increase the suction level at which the relief valve opens. Counterclockwise rotation will decrease it.

7. Lock the adjusting screw with a drop of removable thread locker such as Loctite 242. Repeat step 5 to verify the vacuum relief valve setting.

8. Re-attach the cover (see Section 6.2.2 Assembly).

WARNING: Excess Loctite

®

may seal the steel ball to the seat. This will disable the vacuum relief

6/Service - Disassembly and Assembly

valve and may allow suction to exceed the preset limit.

CAUTION: When Loctite is used on the vacuum relief safety valve, ensure that it only contacts the metal parts. Loctite causes many plastic parts to deteriorate.

6.3.2 Vacuum Limiting Set Screw

Adjustment (PED and NEO Units

ONLY)

6.3.2.1 Pediatric Unit

1. Adjust the regulator until the unit’s gauge reads 135 mmHg (18.0 kPa).

2. Rotate the set screw located above the regulator knob clockwise until it stops against the regulator module and then place one drop of Loctite 242 onto the set screw.

3. Turn the regulator counter-clockwise then clockwise until it stops against the set screw. Ensure that the gauge reads 135 mmHg ± 5 mmHg (18.0 kPa ± 0.7 kPa) and that the relief valve does not activate. For the digital unit, ensure that the gauge reads 135 mmHg ± 2 mmHg (18.0 kPa ± 0.3 kPa).

6.3.2.2 Neonatal Unit

1. Adjust the regulator until the unit’s gauge reads 100 mmHg (13.3 kPa).

2. Rotate the set screw located above the regulator knob clockwise until it stops against the regulator module and then place one drop of Loctite 242 onto the set screw.

3. Turn the regulator counter-clockwise then clockwise until it stops against the set screw. Ensure that the gauge reads 100 mmHg ± 5 mmHg (13.3 kPa ± 0.7 kPa) and that the relief valve does not activate. For the digital unit, ensure that the gauge reads between

99 mmHg and FULL (13.2 kPa and FULL).

6.4 Regulator Module

CAUTION: Do not lubricate any internal components of the regulator module.

CAUTION: The cap screws can strip the regulator module housing threads if they are screwed in too tight.

8700-0006-000 Rev.8 (10/2014) 13

6/Service - Disassembly and Assembly

6.4.1 Disassembly

1. Remove the quad-ring from the piston stem and the O-ring from the housing. Lubricate or replace if needed.

2. Lift and pull tabs to pry off and remove the cap.

3. Peel the diaphragm from the piston head.

4. With the piston head facing down, lift the housing to separate.

5. Remove springs from the piston stem.

6.4.2 Assembly

1. With the piston head facing down, place the springs on the piston stem.

2. Place the housing over the piston and slide the piston stem into the housing.

3. Place the quad-ring on the tip of the piston stem.

4. Place the O-ring on the housing.

5. Fold the diaphragm and place it on the piston head.

6. Place the cap on the housing and press until it snaps into place.

6.5 Digital Gauge

Note: Both batteries are positioned with the positive side facing up relative to the display.

6.5.1 Removal of Batteries

1. Remove digital gauge as described in Section 6.2.1

Disassembly.

2. Using a small flat head screwdriver, push batteries out at a 45 o

angle.

6.5.2 Inserting Batteries

1. Place negative side of battery into battery slot first

(battery will be at an angle).

2. Then using a flat head screwdriver, gently push in positive battery contact while sliding battery in place.

WARNING: If the Vacuum Regulator is repaired or disassembled in any manner, the Service

Checkout Procedure must be performed before using the equipment on a patient.

14 8700-0006-000 Rev.8 (10/2014)

Important: This entire Service Checkout Procedure must be performed in numerical order.

7.1 Set-up

1. Verify that there is 500 mmHg ± 10 mmHg (66.7 kPa

± 1.3 kPa) vacuum on the supply gauge.

2. The supply open flow must be 80 LPM minimum.

3. Connect the supply vacuum to the adapter/ probe port.

Mode

Selector

Switch

Supply Vacuum Gauge

0-760 mmHg (0-101.3 kPa)

Supply Vacuum Reg. set to 500 mmHg +

10 mmHg

(64 kPa + 1.3 kPa)

Supply

Vacuum

500 mmHg

(67 kPa) minimum and open flow of 80 LPM minimum

Flowmeter

50 LPM

High Vacuum

Calibration Gauge

760 mmHg/101.3 kPa

Occluded

Low Vacuum

Calibration Gauge

225 mmHg/30 kPa

7.2 Push-To-Set

TM

Test

1. Connect the 10 LPM flowmeter to the patient port.

2. Move the mode selector switch to | (ON).

3. Set 200 mmHg (26.6 kPa) on the gauge and keep the knob pushed in. For the pediatric unit set at 135 mmHg (18.0 kPa), for the neonatal unit set at 100 mmHg (13.3 kPa) and for the C.A.S.S. unit set at 140 mmHg (18.7 kPa)

4. Make certain leakage at the patient port connection is no more than 1 LPM.

5. While observing the flowmeter with the knob still pushed in, rotate the knob. Leakage should be no more than 1 LPM. Release the knob and move the mode selector switch to O (OFF).

7/Service Checkout Procedure

7.3 Gauge Test

Note: Analog gauges are supplied with an accuracy of ±5% of full scale deflection throughout their range. See step 4 of Section 7.4 Regulation Test.

Note: Digital gauges are supplied with an accuracy of

±1% of full range at 22

°

C.

Note: All gauge needles should come to rest within the zero range bracket or return to the stop pin when no suction is being supplied. Gauges which do not comply may be out of calibration. Digital gauge should read “0” when no suction is applied.

No calibration required on digital gauges.

Note: When checking the accuracy of the gauge on the analog unit, be sure that the calibration gauge has an accuracy of 1% of full scale deflection or better.

7.3.1 High Vacuum Gauges ONLY

1. Connect the regulator’s fitting/patient port to the high calibration gauge with tubing.

2. Move the mode selector switch to | (ON).

3. Ensure that the gauge is in agreement with the high vacuum calibration gauge within the ± 38 mmHg (±

5.1 kPa) tolerance. Recommended test points are

100, 300, and 450 mmHg (13.3, 40.0 and 60.0 kPa).

4. Push and rotate the suction control knob fully clockwise and ensure that the vacuum calibration gauge reads at least 450 mmHg (60.0 kPa).

5. Three mode models ONLY: Move the mode selector switch to MAX and ensure that the vacuum calibration gauge reads at least 450 mmHg (60.0 kPa).

6. Push and rotate the suction control knob fully counter-clockwise (decrease) and verify the gauge reading decreases to zero.

7.3.2 Standard and PED/NEO/C.A.S.S.

Vacuum Gauges ONLY

1. Connect the regulator’s fitting/patient port to the low calibration gauge with tubing.

2. Move the mode selector switch to | (ON).

3. Ensure that the gauge is in agreement with the low vacuum calibration gauge within ± 10 mmHg

(± 1.3 kPa) for the standard model. For the PED/

NEO/C.A.S.S. models, ensure that the calibration gauge is within ±8 mmHg (± 1.1 kPa). Recommended

8700-0006-000 Rev.8 (10/2014) 15

7/Service Checkout Procedure

test points are 40 and 80 mmHg (5.3 and 10.7 kPa) for the standard/PED/NEO/C.A.S.S. models and 140 mmHg (18.7 kPa) for standard model only.

4. For the standard model ONLY: Push and rotate the suction control knob fully clockwise. Ensure that the vacuum calibration gauge reads at least 450 mmHg

(60.0 kPa) and that the gauge on the unit is in the

FULL VAC range.

5. Push and rotate the suction control knob fully counter-clockwise (decrease) and verify the gauge reading decreases to zero.

7.4 Regulation Test

1. Remove any equipment attached to the patient port.

2. If using a high vacuum regulator, set its gauge to

300 mmHg (40.0 kPa). If using a standard/PED/

NEO/C.A.S.S. vacuum regulator, set its gauge to 100 mmHg (13.3 kPa).

3. Open and close the fitting/patient port several times to create various flow rates through the regulator.

4. With the fitting/patient port occluded, the gauge should return to the setting listed in step 2 within the following tolerances:

• For analog units, the gauge should return to ±

10 mmHg (± 1.3 kPa) for the standard model, ±

8 mmHg (± 1.1 kPa) for the PED/NEO/C.A.S.S. models, and ± 38 mmHg (± 5.1 kPa) for the high model.

• For digital units, the gauge should return to ± 2 mmHg (± 0.3 kPa) for the standard/PED/C.A.S.S. models, ± 1 mmHg (± 0.1 kPa) for the NEO model, and ± 8 mmHg (± 1.0 kPa) for the high model.

7.5 Maximum Vacuum Test

(PED/NEO/C.A.S.S. Models ONLY)

1. Occlude the fitting/patient port.

2. Rotate the suction control knob fully clockwise

(increase).

3. Verify that the suction delivered does not exceed 100 mmHg ± 5 mmHg (13.3 kPa ± 0.7 kPa) for the NEO model, 135 mmHg ± 5 mmHg (18.0 kPa ± 0.7 kPa) for the PED model, and 140 mmHg ± 5 mmHg (18.7 kPa ± 0,.7 kPa) for the C.A.S.S. model.

Note: For setting the suction level limit, refer to

Section 6.3 Suction Level Limit Setting (PED/NEO/

C.A.S.S. Models ONLY).

16

7.6 Bleed Test

1. Occlude the fitting/patient port and set the vacuum level to 100 mmHg (13.3 kPa).

2. Move the selector switch to O (OFF) and observe the gauge needle. It must return to the zero range bracket or stop pin within 10 seconds (digital gauge will return to “0”).

7.7 Leak Test

7.7.1. Supply Side

1. Connect the adapter/probe port to port “A” of the bubble leak tester. Allow the fitting/patient port to be open to air.

2. Move the mode selector switch to O (OFF).

3. Connect port “B” of the bubble leak tester to the supply vacuum regulated to 500 mmHg (66.7 kPa).

4. Wait 20 seconds. No more than 6 bubbles should appear in the next ten seconds.

Important: Prior to venting port “A” of the bubble leak tester to atmosphere (i.e. turning the supply regulator off), ensure the tubing from port “B” has been disconnected from the adapter/ probe port of the vacuum regulator.

Test Gauge

Adaptor/

Probe

Port

Port

A

Port

B

Bubble Leak Test - Supply Side

8700-0006-000 Rev.8 (10/2014)

7.7.2 Patient Side

1. Connect the supply vacuum tubing to the adapter/ probe port.

2. Connect the fitting/patient port to port “A” of the bubble leak tester with tubing. Allow port “B” of the bubble leak tester to be open to air.

3. Rotate the suction control knob a minimum of one full turn clockwise (increase). No bubbles should appear in the next ten seconds.

4. Rotate the suction control knob fully counter-clockwise

(decrease).

5. Move the mode selector switch to | (ON). No bubbles should appear in the next ten seconds.

7/Service Checkout Procedure

• The unit is set to 100 mmHg ± 5 mmHg (12.7 kPa

± 0.7 kPa) in | (ON) mode.

• Adult/High models ONLY: the unit should produce a minimum of 40 LPM at maximum vacuum in | (ON) mode.

1. Connect the fitting/patient port to the 50 LPM flowmeter.

2. Move the mode selector switch to | (ON).

3. Set the unit to 100 mmHg ± 5 mmHg (12.7 kPa ± 0.7 kPa).

4. Release the suction control knob and verify that the flow rate exceeds 25 LPM.

5. Adult models ONLY: Push and rotate the suction control knob fully clockwise.

6. Release the suction control knob and verify that the flow rate exceeds 40 LPM.

MAX

Port

A

Port

B

Fitting/

Patient

Port

Bubble Leak Test - Patient Side

7.8 Flow Test

Note: All PTS-CVR models should produce a minimum of 25 LPM from the patient port given the following conditions:

• The supply open flow is 80 LPM minimum.

• The supply vacuum pressure is 500 mmHg ± 10 mmHg.

• The unit has fittings on the fitting/patient and adapter/probe ports with minimum inner diameters of 0.140”.

• The unit is either directly connected to the vacuum source, or connected with a tube no more than 4” in length with a minimum inner diameter of 0.25”.

• The unit is connected to a flowmeter with a tube no more than 4” in length with a minimum inner diameter of 0.25”

8700-0006-000 Rev.8 (10/2014) 17

8/Maintenance

8.1 General Maintenance of Suction Equipment

WARNING: The Pre-use Checkout Procedure must be performed before using this equipment on each patient.

If the regulator fails any part of the Pre-use Checkout Procedure, it must be removed from service and repaired by qualified service personnel.

WARNING: If the unit has been contaminated, clean and disinfect all suction equipment before disassembly.

PTS-CVR should be kept in use or used on a rotating basis. Internal parts of unused equipment may tend to deteriorate.

Maintenance of the vacuum piping system is as important as maintenance of the suction equipment. The use of collection canisters with reliable shut-off valves, overflow safety trap assemblies and disposable suction filters will protect the regulator, wall outlet, and piping system. The flow rate at the wall outlet should be checked on a yearly basis and suitable cleaning of the outlets should be performed as needed. The flow rate measurement should meet NFPA

®

99 and/or ISO

® standards.

Routine maintenance and inspection are important to the performance of suction equipment. The following is a recommended list for care of suction equipment after each patient use.

1. Keep connecting tubing, fittings and adapters/probes clean.

2. Wipe all exterior surfaces with a solution of water and mild detergent or approved cleaning solutions.

3. Perform a careful visual inspection of the vacuum regulator.

4. Check that the high flow disposable suction filter is clean and in good condition.

5. Check that all tubing is in good condition and connected securely to the correct ports.

6. Check the floats in the Overflow Protection Device and collection canister for correct operation.

7. Perform the Pre-use Checkout Procedure.

8.2 Recommended Maintenance Schedule

We recommend the following to determine the maintenance schedule appropriate for each facility:

1. Periodically (as required, but no less than annually) inspect the overall condition of the vacuum regulator. Test gauge accuracy (Section 7.3 Gauge Test) and perform the pre-use checkout procedure (See PTS-CVR Instructions for Use for the Pre-use Checkout Procedure). If the regulator does not pass, refer to troubleshooting (Section 5

Troubleshooting).

2. Determine a maintenance schedule based on data from your periodic inspections. Follow the guidelines below.

8.2.1 Maintenance Schedule

Item

Service Checkout

Procedure

Elastomeric parts, O-rings, gaskets, diaphragms, internal filters

Minimum Frequency Comments

Every 24 months or as required

As required

If the regulator does not pass, refer to troubleshooting (Section

5 Troubleshooting). Repair as needed

Cleaning, lubrication and replacement interval depends on hours of usage and environmental conditions. Replace, lubricate, and repair as needed

8.3 Repair Policy

WARNING: To reduce transportation personnel and/or service personnel exposure to hazardous contamination, clean and disinfect all suction equipment before shipping for service.

CAUTION: Do not steam autoclave or liquid sterilize the PTS-CVR. Severe impairment to the operation of the regulator will result. The only acceptable method of sterilization is with gas (ethylene oxide).

CAUTION: Only competent individuals trained in the repair of this equipment should attempt service.

Do not use malfunctioning equipment. Make all necessary repairs. Have the equipment repaired by qualified service personnel or by Ohio Medical. After repair, perform the Service Checkout Procedure to ensure that the product is functioning properly, and complies with the manufacturer’s published specifications.

18 8700-0006-000 Rev.8 (10/2014)

8/Maintenance

8.4 Technical Assistance

If technical assistance is required, contact Ohio Medical technical support at +1 857-855-0500 or toll free at

866-549-6446.

8.5 Return Instructions

1. Clean and disinfect the vacuum regulator.

2. Package securely for protection, preferably in the original container.

3. Include a letter describing in detail any difficulties experienced with the product. Include the person, title, and telephone number to contact for functional questions.

4. If the vacuum regulator is covered under warranty, include the warranty information that came with the device and a copy of the invoice.

5. Call toll free 866-549-6446 or +1 847-855-0500 and ask customer service for an RMA number to include with your shipment.

6. Ship the vacuum regulator prepaid. Write your return address and billing address information on the package or letter that comes with the package.

For Warranty and non-warranty repairs, mail the package to

Ohio Medical Corporation

1111 Lakeside Drive

Gurnee, IL 60031 USA

RMA #_________________

In other locations contact your nearest Ohio Medical office or authorized Ohio Medical distributor.

8.6 Installation Procedure for Adapters/Probes and Fittings

CAUTIONS:

Do not use any Loctite® products or any products which contain

Methacrylate Ester as an active ingredient to seal the threads on the adapters/probes and fittings.

All Ohio Medical regulators are supplied with 1/8” NPT female ports. The standard port facilitates simple adaptation to any quick disconnect system. Fittings are available from Ohio Medical to adapt to your quick connect system. We recommend you purchase the appropriate fittings with your regulator at the time of purchase. The fittings will then be factory installed prior to shipping.

Wall Adapter

Port (Rear)

In the event that you must assemble or disassemble fittings, please follow these instructions:

1. Prior to installing the fittings wrap the thread with Teflon tape

Ohio Medical P/N 6700-1987-800 or equivalent.

2. Apply appropriate torque 4.0 ft-lb (5.4 N-m) minimum to 10.0 ft-lb (13.6 N-m) maximum.

Patient Connection

Port (Bottom)

3. Adapters/Probe and fittings which are not keyed for specific orientation, should be torque approximately 6.0 ft-lb (8.1

N-m).

4. Adapters/Probes and fittings that are keyed to specific orientation, must be torque initially to 4.0 ft-lbs (5.4 N-m).

Additional torque is applied only until orientation is correct.

5. Make sure wall adapter are installed to specific orientation so that they are mount straight on wall.

6. The regulator is now ready to be place in service on your suction system.

8700-0006-000 Rev.8 (10/2014) 19

9/Ordering Information

9.1 Illustrated Parts

12.

13.

14.

15.

8.

9.

10.

11.

6.

7.

4.

5.

Item Part Number

1.

6700-0078-400

Description

Screw, PHH PNH, 4-20 Plastite

2.

3.

8700-0017-500

See Table

Knob

Case, Regulator

8700-0041-500

8700-0021-500

Gauge Lens

Drive Gear, Occlude-To-Set

See table below Gauge, Vacuum

8700-0038-500 Cover Strip, Switch

8700-0015-500

6700-0175-500

8700-0002-700

8700-0043-500

Actuator, Occlude-To-Set

O-Ring, 2-013 Nitrile

OTS Assembly

Support Bracket, Actuator

8700-0001-700 Assembly, Reg Mod

8700-0040-500 Plate, Detent

8700-0016-500 Selector Switch, Overmolded

16.

See table below Plug, Pull Tab

17.

See table below Vacuum Relief Assembly

18.

See table below O-Ring, 2-014, Nitrile, Buna-N

19.

6700-0115-400 Relief Valve, Positive Pressure

1

1

1

1

1

0210-0525-300 O-Ring, 9.32 OD x 5.32 IS x 1/16 W 1

1

1

1

1

1

1

1

1

1

1

Qty

5

1

1

23.

24.

25.

26.

Item Part Number

20.

0206-5159-300 Plug, Filter

21.

22.

6700-0121-400

8700-0009-700

Description

Orifice, Continuous Vacuum Regulator

Manifold Assembly

Qty

1

1

1

6700-0151-400

8700-0016-100

8700-0005-100

8700-0056-500

Screw, Self Tap, #6 Pan Head

Serial Number Label

Knob, Label

Plug

1

1

4

1

31.

32.

33.

34.

27.

See table below Label

28.

8700-0006-500 Cap

29.

30.

8700-0046-500

8700-0008-500

Cap Plug

Diaphragm

35.

36.

8700-0007-500

8700-0011-400

8700-0012-400

8700-0005-500

8700-0023-500

8700-0024-500

Piston with stem

Spring, 0.475" ID

Spring, 0.246" ID

Housing

O-Ring, -018 Nitrile

Quad-Ring, -006 Nitrile

8700-0006-000 PTS-CVR Service Manual

1

1

1

1

1

1

1

1

1

1

1

Model #

1224/1324

1225/1325

1226/1326/1229

1231/1331/1235

1234/1334/1232

1330/1430/1230

Item 3

Case, Regulator

8700-0012-500

8700-0022-500

8700-0012-500

8700-0053-500

8700-0053-500

8700-0022-500

Model #

1224/1324

1225/1325

1226/1326

1231/1331

1234/1334

1330/1430

Color Case for

Digital Regulators Only

(Item 3)

Purple

Red

Blue

Yellow

Lt. Blue

Pink

Item 16

Plug, Pull Tab

8700-0039-500

8700-0039-500

8700-0039-500

N/A

N/A

8700-0039-500

2-Mode and C.A.S.S.

8700-0065-500

8700-0066-500

8700-0067-500

8700-0068-500

8700-0069-500

8700-0070-500

Item 17

Vacuum Relief

Assembly

N/A

N/A

N/A

6700-0045-700

6700-0045-700

N/A

3-Mode

8700-0092-500

8700-0093-500

8700-0094-500

8700-0095-500

8700-0096-500

8700-0097-500

Item 6

Gauge, Vacuum

Analog (ANSI) / Digital

8700-0002-400 / VR-DGP-200MM

8700-0002-400 / VR-DGP-200MM

8700-0022-400 / VR-DGP-760MM

8700-0024-400 / VR-DGP-160MM

8700-0024-400 / VR-DGP-160MM

8700-0022-400 / VR-DGP-760MM

Item 18

O-Ring, 2-014,

Nitrile, Buna-N

N/A/

N/A

N/A

6700-0135-500

6700-0135-500

N/A/

Pediatric

8700-0108-500

8700-0109-500

8700-0110-500

8700-0111-500

8700-0112-500

8700-0113-500

Item 6

Gauge, Vacuum

Analog (ISO)

8700-0023-400

8700-0025-400

8700-0025-400

8700-0023-400

Item 27

Label

N/A

N/A

N/A

8700-0036-102

N/A

N/A

Neonatal

8700-0124-500

8700-0125-500

8700-0126-500

8700-0127-500

8700-0128-500

8700-0129-500

20 8700-0006-000 Rev.8 (10/2014)

9/Ordering Information

8700-0006-000 Rev.8 (10/2014) 21

Notes:

22 8700-0006-000 Rev.8 (10/2014)

Notes:

8700-0006-000 Rev.8 (10/2014) 23

North America

United States

Customer Service, Distribution Center

Technical Support, Sales and Service

Equipment Service Center

Ohio Medical Corporation

1111 Lakeside Drive

Gurnee, IL 60031 USA

P: +1 866 549 6446

P: +1 847 855 0800

F: +1 847 855 6218

Ohio Medical Corporation Authorized

Representative

EC REP

OxygenCare Ltd.

2 Holfeld Business Park

Kilmacanogue Co Wicklow

Ireland

Tel: +353 1 2769700

Fax: +353 1 2764970

© 2014 Ohio Medical Corporation.

This document contains information that is proprietary and confidential to Ohio Medical Corporation.

Use of this information is under license from Ohio Medical Corporation.

Any use other than that authorized by Ohio Medical Corporation is prohibited.

Ohio Medical Corporation is a registered trademark and Push-to-Set is a trademark of Ohio Medical Corporation.

ANSI is a registered trademark of American National Standards Institute.

ISO is a registered trademark of International Organization for Standardization.

NFPA is a registered trademark of National Fire Protection Association Inc.

Lexan is a registered trademark of Sabic Innovative Plastics IP B.V. Company.

Virex is a registered trademark of S.C. Johnson & Son, Inc.

Kleenaseptic is a registered trademark of Metrex Research Corporation.

Betco is a registered trademark of Betco Corporation.

Cavicide is a registered trademark of Metrex Research Corporation.

Vac Kote is a registered trademark of the Ball Aerospace & Technology Corp.

Loctite is a registered trademark of Henkel Corporation.

Teflon is a registered trademark of E.I. duPont de Demurs and Company.

Oxequip is a registered trademark of Allied Healthcare Products, Inc.

Dow Corning is a registered trademark of Dow Corning Corporation.

Venturi is a registered trademark of Talley Group, Ltd

8700-0006-000 (Rev.8)

10/2014

advertisement

Key Features

  • Lightweight, compact design
  • Regulated or full-line vacuum
  • Positive pressure safety relief valve
  • Vacuum relief valve (Low models)
  • Easy suction adjustment
  • Quick and easy mode changes
  • Accurate vacuum gauge

Frequently Answers and Questions

What is the intended use of the PTS-CVR vacuum regulator?
The vacuum regulator is intended to be used in the medical facility as a means to evacuate media (i.e. fluids) from the body. DO NOT use this vacuum regulator for anything other than its intended use.
What is the purpose of the positive pressure safety relief valve?
The positive pressure safety relief valve is located in the vacuum supply line to prevent pressurization of the patient connection by failed injector vacuum (Venturi) units, or inadvertent cross connection to pressured gases.
How do I adjust the suction level on the PTS-CVR?
Turn the suction control knob to adjust the position of the regulator module and allows selection of a predetermined level of suction when set according to instructions.
What are the different modes of operation for the PTS-CVR?
Some models operate in a regulated or non-regulated MAX mode. Others operate only in the regulated mode. In the non-regulated MAX mode, the vacuum source is connected directly to the fitting/patient port. The regulator module is bypassed and full-line vacuum is provided. In the regulated mode, the vacuum source is connected through the regulator module which functions as an automatic valve.
How do I sterilize the PTS-CVR?
The regulator may be sterilized using Ethylene oxide (ETO). After sterilization, follow the service checkout procedures (Section 7 Service Checkout Procedure).