SERVICE MANUAL

Valor

5000 Scales

SERVICE MANUAL

Valor

5000

Scales

The information contained in this manual is believed to be accurate at the time of publication, but Ohaus

Corporation assumes no liability arising from the use or misuse of this material. Reproduction of this material is strictly prohibited.

Material in this manual is subject to change.

© Copyright 2008 Ohaus Corporation, all rights reserved.

TM Registered trademark of Ohaus Corporation.

TABLE OF CONTENTS

Page No.

CHAPTER 1 GETTING STARTED

1.1 Introduction ................................................................................................................1-1

1.2 Service Facilities........................................................................................................1-1

1.3 Tools and Test Equipment Required .........................................................................1-1

1.4 Specifications.............................................................................................................1-2

1.5.1 Overview of the Controls ....................................................................................1-4

1.5.2 Overview of the Display Indicators .....................................................................1-5

1.6 Legal For Trade (LFT) ...............................................................................................1-6

1.6.1 LFT Setup Procedure .........................................................................................1-6

CHAPTER 2 TROUBLESHOOTING

2.1 Troubleshooting .........................................................................................................2-1

2.2 Diagnostic Guide .......................................................................................................2-1

2.2.1 Diagnosis............................................................................................................2-1

CHAPTER 3 MAINTENANCE PROCEDURES

3.1 Preventive Maintenance ............................................................................................3-1

3.2 Opening the Scale .....................................................................................................3-1

3.2.1 Opening Valor 5000 Standard and Washable Models .......................................3-2

3.3 Replacing the Load Cell.............................................................................................3-3

3.4 Replacing the Printed Circuit Board and Display.......................................................3-4

3.5 Replacing the Housing Seals.....................................................................................3-5

3.6 Replacing the Function Label ....................................................................................3-6

3.6 Replacing the Plastic Insert .......................................................................................3-6

CHAPTER 4 TESTING

4.1. Testing .......................................................................................................................4-1

4.1.1 Test Masses Required .......................................................................................4-1

4.3 Segment Display Test................................................................................................4-1

4.4. Load Cell Test Using RAMP ......................................................................................4-1

4.5.4 Off-Center Load Test..........................................................................................4-5

4.5.5 Adjusting Off Center Load ..................................................................................4-6

CHAPTER 5 DRAWINGS AND PARTS LISTS

5-1 Valor 5000 V31X3 Scale: Housing & Internal Parts...................................................5-2

5-2 Valor 5000 V31X6 Scale: Housing & Internal Parts...................................................5-4

5-3 Valor 5000 V31XH2 Scale: Housing & Internal Parts ................................................5-6

5.4 Valor 5000 V31XH202 Scale: Housing & Internal Parts ............................................5-8

5.5 Valor 5000 V31XW3 Scale: Housing & Internal Parts .............................................5-10

TABLE OF CONTENTS

Page No.

5.6 Valor 5000 V31XW3-M Scale: Housing & Internal Parts .........................................5-12

5.7 Valor 5000 V31XW6 Scale: Housing & Internal Parts .............................................5-14

5.8 Valor 5000 V31XW6-M Scale: Housing & Internal Parts .........................................5-16

5.9 Valor 5000 V31XW301 Scale: Housing & Internal Parts .........................................5-18

5.10 Valor 5000 V31XW301-M Scale: Housing & Internal Parts .....................................5-20

Appendix A STANDARD CALIBRATION

A.1 Calibration................................................................................................................. A-1

Appendix B SERVICE CALIBRATION

B.1 Entering the Service Menu ....................................................................................... B-1

B.2 Ramp ........................................................................................................................ B-1

B.5 Geographical Adjustment Factor (GEO)................................................................... B-4

B.6 Expanded Readability (E.PAnd) ............................................................................... B-5

APPENDIX C SOFTWARE SERVICE TOOL INSTRUCTIONS

C.1 Hardware and Software Setup.................................................................................. C-1

C.2 Configuring the Scale (after replacing the PCB or Load Cell) .................................. C-3

APPENDIX D

GEOGRAPHICAL ADJUSTMENT VALUES

TABLE D-1. Geographical Adjustment Values................................................................... D-1

LIST OF TABLES

TABLE NO. TITLE Page No.

1-2

1-3

Capacity and Readability.......................................................................................1-3

Functions of Display Controls................................................................................1-4

1-4 Display

1-5 Valor 5000 Menu Structure ...................................................................................1-7

4-1 Calibration Mass Values........................................................................................1-2

4-2 Tolerances ............................................................................................................4-2

4-3 Repeatability

4-4 Linearity Test Masses ...........................................................................................4-5

5-1

5-2

5-3

5.4

Valor 5000 V31X3 Scale: Housing & Internal Parts ..............................................5-3

Valor 5000 V31X6 Scale: Housing & Internal Parts ..............................................5-5

Valor 5000 V31XH2 Scale: Housing & Internal Parts............................................5-7

Valor 5000 V31XH202 Scale: Housing & Internal Parts........................................5-9

5.5

5.6

5.7

5.8

Valor 5000 V31XW3 Scale: Housing & Internal Parts.........................................5-11

Valor 5000 V31XW3-M Scale: Housing & Internal Parts.....................................5-13

Valor 5000 V31XW6 Scale: Housing & Internal Parts.........................................5-15

Valor 5000 V31XW6-M Scale: Housing & Internal Parts.....................................5-17

TABLE OF CONTENTS

TABLE NO.

5.9

TITLE Page No.

Valor 5000 V31XW301 Scale: Housing & Internal Parts.....................................5-19

5.10 Valor 5000 V31XW301-M Scale: Housing & Internal Parts.................................5-21

D-1. Geographical Adjustment Values ......................................................................... D-1

LIST OF ILLUSTRATIONS

FIGURE NO.

1-1

TITLE

Power Connection and Seal..................................................................................1-4

1-2

1-3

Valor 5000 Display ................................................................................................1-4

Valor 5000 Display Indicators................................................................................1-5

1-4 LFT/Cal Lock Switch .............................................................................................1-6

3-1 Scale

3-2

3-3

3-4

3-5

Technique for separating Top and Bottom Housing on Valor 5000 ......................3-2

Red lock between battery compartments ..............................................................3-3

Printed Circuit Board and Display .........................................................................3-4

Bottom of Valor 5000 Xtreme W, with Battery Cover on .......................................3-5

3-6 Plastic Insert and Function Label ..........................................................................3-6

3-7 Plastic Insert, viewed from inside Top Housing.....................................................3-6

4-1 Segment

4-2

5-1

5-2

5-3

5.4

5.5

5.6

5.7

Scale drawing of Valor 5000 Load Cell and Weighing Pan...................................4-6

Valor 5000 V31X3 Scale: Housing & Internal Parts ..............................................5-2

Valor 5000 V31X6 Scale: Housing & Internal Parts ..............................................5-4

Valor 5000 V31XH2 Scale: Housing & Internal Parts............................................5-6

Valor 5000 V31XH202 Scale: Housing & Internal Parts........................................5-8

Valor 5000 V31XW3 Scale: Housing & Internal Parts.........................................5-10

Valor 5000 V31XW3-M Scale: Housing & Internal Parts.....................................5-12

Valor 5000 V31XW6 Scale: Housing & Internal Parts.........................................5-14

5.8

5.9

Valor 5000 V31XW6-M Scale: Housing & Internal Parts.....................................5-16

Valor 5000 V31XW301 Scale: Housing & Internal Parts.....................................5-18

5.10 Valor 5000 V31XW301-M Scale: Housing & Internal Parts.................................5-20

C-1 RS232 Interface connected to Service Adapter, connected to PCB .................... C-1

C-2

C-3

C-4

Com Port Configuration screen with correct settings ........................................... C-2

Config Tool’s Test Command screen ................................................................... C-2

The Config Tool’s Configuration Screen .............................................................. C-3

TABLE OF CONTENTS

CHAPTER 1 GETTING STARTED

1.1 INTRODUCTION

This service manual contains the information needed to perform routine maintenance and service on the Ohaus Valor 5000 Series scales. Familiarity with the scale’s Instruction Manual is assumed. The contents of this manual are contained in five chapters:

Chapter 1 Getting Started

– Contains information on service facilities, tools and test equipment, specifications, and the mechanical and electronic functions of the scale.

Chapter 2 Troubleshooting

– Contains a diagnostic guide and error code table.

Chapter 3 Maintenance Procedures

– Contains preventive maintenance procedures and disassembly, repair and replacement procedures.

Chapter 4 Testing – Contains a list of required test masses, an operational test, segment

display test, performance tests and adjustments.

Chapter 5 Drawings and Parts

Lists

– Contains exploded views of Valor 3000 scales identifying all serviceable components.

Appendix A Standard Calibration – Explains procedures for Standard Calibration, performed

prior to using a scale, and after service.

Appendix B Service Calibration – Describes the Service Menu and sub-menus, which allow

authorized service personnel to perform factory Linearity and Span calibrations (no pre-set limits), take Ramp readings, adjust the GEO Factor, and use E.PAnd to set the readability to

1/10 th

of the standard readability.

Appendix C

Geographical Adjustment Values

The chart of scale settings for every

geographical latitude away from the equator (in degrees and minutes) and every elevation above sea level (in meters or feet).

To service a scale, the service area should meet the following requirements:

• Should be temperature controlled and meet scale specifications for temperature environmental requirements.

• Must be free of vibrations such as fork lift trucks close by, large motors, air currents or drafts from air conditioning/heating ducts, open windows, people walking by, fans, etc.

• Area must be clean and free of excessive dust.

• Work surface must be stable and level.

• Scale must not be exposed to direct sunlight or radiating heat sources.

• Use an approved Electro-Static Device.

CHAPTER 1 GETTING STARTED

1.3 TOOLS AND TEST EQUIPMENT REQUIRED

The service shop should contain the following equipment:

1. Standard hand tools.

2. Digital Voltmeter (DVM).

3. Standard Electronics tool kit.

4. Grounding mat and clip.

5. Razor

The repair method for the Valor 5000 Series Scales is direct substitution of major assemblies.

The available repair parts are listed and illustrated in Chapter 5.

There are four major replaceable assemblies:

– Cover Assembly with Membrane Switch

– Load Cell Assembly with Frame

– Lead Acid Rechargeable Battery and

– Main PC Board with LCD.

The Load Cell, LCD and Membrane Switch may also be ordered and replaced separately.

This service manual contains sufficient information to isolate the problem, replace the component, test and restore the scale to original factory specifications.

CHAPTER 1 GETTING STARTED

1.5 SPECIFICATIONS

Complete specifications for the Ohaus Valor 5000 Scales are listed in Tables 1-1 and 1-2.

When a scale has been serviced, it must meet the specifications listed in the table. Before servicing the scale, determine what specifications are not met.

TABLE 1-1. SPECIFICATIONS, STANDARD DENSITY MODELS

Model

Capacity x Readability

Max x d (nonapproved)

V51P3 V51P6 V51P15 V51P30

3 kg x 0.0005 kg

3000 g x 0.5 g

6 lb x 0.001 lb

96 oz x 0.02 oz

96 oz x 1/4 oz

6 lb 0 oz x 0.1 oz

6 lb 0 oz x 1/4 oz

6 kg x 0.001 kg

6000 g x 1 g

15 lb x 0.002 lb

240 oz x .05 oz

240 oz x 1/4 oz

15 lb x 0.1 oz

15 lb x 1/4 oz

15 kg x 0.002 kg

15000 g x 2 g

30 lb x 0.005 lb

480 oz x 0.1 oz

480 oz x 1/4 oz

30 lb 0 oz x 0.1 oz

30 lb 0 oz x 1/4 oz

30 kg x 0.005 kg

30000 g x 5 g

60 lb x 0.01 lb

960 oz x 0.2 oz

960 oz x 1/4 oz

60 lb 0 oz x 0.2 oz

60 lb 0 oz x 1/4 oz

Maximum Displayed

Resolution

Capacity x Readability

Max x d (approved)

6000 7500 7500 6000

Approved Resolution

Repeatability (kg)

Linearity (kg) (+/-)

Weighing Units

Application Modes

Checkweigh

Indication

3 kg x 0.001 kg

3000 g x 1 g

6 lb x 0.002 lb

96 oz x 0.05 oz

96 oz x 1/4 oz

6 kg x 0.002 kg

6000 g x 2 g

15 lb x 0.005 lb

240 oz x .1 oz

240 oz x 1/4 oz

15 kg x 0.005 kg

15000 g x 5 g

30 lb x 0.01 lb

480 oz x 0.2 oz

480 oz x 1/4 oz

30 kg x 0.01 kg

30000 g x 10 g

60 lb x 0.02 lb

960 oz x 0.5 oz

960 oz x 1/4 oz

3000 3000 3000 3000

0.001 0.002 0.005 0.01

0.001 0.002 0.005 0.01 g, kg, lb, oz (decimal/fractional), lb:oz (decimal/fractional oz)

Weighing, Checkweighing, Percentage

3 LEDs (yellow, green, red) with configurable alert beeper

Power

9-12 VDC 0.5A-0.8A AC adaptor; internal rechargeable sealed lead-acid battery

Battery Operating

Time

Tare range

Stabilization Time

Auto zero tracking

Operating

Temperature Range

Storage

Temperature Range

Approved

Temperature Range

Pan Size (D X W)

Scale dimensions

120 hours

To capacity by subtraction

Within 2 seconds

Off, 0.5, 1 or 3 divisions

–10° to 40°C / 14° to 104°F

– 40° to 70°C / – 40° to 104°F

0° to 40°C / 32° to 104°F

209 x 209 mm / 8.2 x 8.2 in

300 x 223 x 71 mm / 9 x 12 x 2.8 in.

CHAPTER 1 GETTING STARTED

1.4 SPECIFICATIONS

TABLE 1-2. SPECIFICATIONS, HIGH DENSITY MODELS

Model

Capacity x Readability

Max x d (nonapproved)

V51PH3 V51PH6 V51PH15 V51PH30

3 kg x 0.0001 kg

3000 g x 0.1 g

6 lb x 0.0002 lb

96 oz x 0.005 oz

96 oz x 1/4 oz

6 lb 0 oz x 0.1 oz

6 lb 0 oz x 1/4 oz

6 kg x 0.0002 kg

6000 g x 0.2 g

15 lb x 0.001 lb

240 oz x .01 oz

240 oz x 1/4 oz

15 lb x 0.1 oz

15 lb x 1/4 oz

15 kg x 0.001 kg

15000 g x 1 g

30 lb x 0.001 lb

480 oz x 0.02 oz

480 oz x 1/4 oz

30 lb 0 oz x 0.1 oz

30 lb 0 oz x 1/4 oz

30 kg x 0.001 kg

30000 g x 1 g

60 lb x 0.002 lb

960 oz x 0.05 oz

960 oz x 1/4 oz

60 lb 0 oz x 0.1 oz

60 lb 0 oz x 1/4 oz

Maximum Displayed

Resolution

Repeatability (kg)

Linearity (kg) (+/-)

Weighing Units

Application Modes

Checkweigh

Indication

30000 30000 30000 30000

0.0002

0.0002

0.0004

0.0004

0.002

0.002

0.002

0.002 g, kg, lb, oz (decimal/fractional), lb:oz (decimal/fractional oz)

Weighing, Checkweighing, Percentage

3 LEDs (yellow, green, red) with configurable alert beeper

Power

9-12 VDC 0.5A-0.8A AC adaptor; internal rechargeable sealed lead-acid battery

Battery Operating

Time

Tare range

Stabilization Time

Auto zero tracking

Operating

Temperature Range

Storage

Temperature Range

Approved

Temperature Range

Pan Size (D X W)

Scale dimensions

120 hours

To capacity by subtraction

Within 2 seconds

Off, 0.5, 1 or 3 divisions

–10° to 40°C / 14° to 104°F

– 40° to 70°C / – 40° to 104°F

0° to 40°C / 32° to 104°F

209 x 209 mm / 8.2 x 8.2 in

300 x 223 x 71 mm / 9 x 12 x 2.8 in.

CHAPTER 1 GETTING STARTED

1.5 SCALE OPERATION

This section contains information on the basic operation of the scale.

1.5.1 OVERVIEW OF THE CONTROLS

Figure 1-1. Valor 5000 Display.

TABLE 1-3. FUNCTIONS OF DISPLAY CONTROLS

Button/

Function

Primary

Function

(short press)

Secondary

Function

(long press)

Menu

Function

Target Setting

Function

ON / ZERO

If scale is off, turns scale on.

If scale is on, sets zero.

Off

If scale is on, turns scale off.

G/N/T

Momentarily

Displays Gross

Net and Tare values

Units

Changes weighing unit

TARGET

Initiates target setting mode.

Menu

Enters menu mode.

TARE

Sets tare.

Clears tare, if pan is empty.

None

Yes

Accepts current setting and advances to the next menu item

Shifts to next digit

No

Rejects current setting or menu item and advances to the next available item

Increments value of the active digit

Back

Reverts back to previous menu item.

Accepts current displayed value.

Exit

• Immediately exits menu mode

• Aborts calibration in progress

Changes sign

(+ / –) of displayed value.

CHAPTER 1 GETTING STARTED

1.5.2 OVERVIEW OF THE DISPLAY INDICATORS

8

Not used

1

2

3

Figure 1-2. Display Indicators.

4

5

TABLE 1-4. DISPLAY INDICATORS

No Function

1 Indicates that the measured value has become stable

2 Center of Zero indicator

3 Checkweigher mode indicator

7

6

6 Symbols for weighing modes: kg – Kilograms, g – grams, lb:oz – Pounds & Ounces,

lb – Pounds, oz – Ounces,

3

/

8

– Fractional ounces (P in % mode)

7 Battery charging indicator

1.5.3 Power

Check that the AC adapter supports the voltage in use locally. Then plug the adapter into the receptacle marked “Power In” located on the underside of the scale.

The first time the scale is powered on, the internal rechargeable battery should be fully charged. The scale can be operated during the charging process. The battery is protected against over charging and the scale can remain connected to the AC power line. The display’s battery symbol will stay lit until the battery is fully charged, and then it will turn off.

“Power In”

Receptacle

Foot Switch

Receptacle

Figure 1-3. Scale Bottom

The scale will automatically switch to battery operation if the power cord is removed or there is a power failure. A fully charged battery can operate the scale for up to 120 hours (with the backlight off). During battery operation, the battery symbol will light to show the current status of the battery (1 segment = 25% capacity). A flashing symbol indicates that the battery requires recharging. The scale will automatically shut off when the battery is fully discharged.

CHAPTER 1 GETTING STARTED

With power applied, press ON/ZERO/Off. Allow 20 minutes warm-up time. The scale should be calibrated before use. (See Appendix A.)

With the scale on, press and hold the ON/ZERO Off button until OFF is displayed.

Programmable features of the Valor 5000 scales are contained in menus which are accessed by pressing TARGET/Menu on the Display Panel. The menu structure consists of three levels:

Main, Sub-menu, and Menu item. The Main menu contains the Sub-menus and each Sub-menu contains several Menu items. Each Menu item contains two or more settings.

Use the four buttons marked with Yes, No, Back and Exit to navigate through the menus.

(See Figure 1-4.) For more detail on using the menus, see the Instruction Manual.

Yes Enters the displayed sub-menu or menu item and accepts the displayed setting.

No Skips to the next sub-menu or menu item and changes the displayed setting.

Back Goes back to the previous sub-menu or menu item.

Exit Immediately exits the Menu mode.

Figure 1-4. Valor 5000 Menu Structure

CHAPTER 1 GETTING STARTED

1.6 Legal for Trade Settings

Consult local weights and measures officials for regulations before placing the scale in Legal for

Trade mode. When the LFT menu setting is On, the following conditions apply:

– The Calibration (CAL) sub-menu is hidden.

– The Calibration Lock (LCAL) menu item is hidden.

– The Auto Zero Tracking (AZt) menu item is set to 0.5 d.

lb:oz and P units are disabled.

To regain access to the locked menu settings, break the seal holding the Security Slide on the bottom of the scale closed, and shift the slide open. (See

Figure 1-6.) This turns LFT off by moving a menulock switch on the PCB to the Off position. (See

Figure 1-5.)

After the LFT/CAL Lock setting has been turned off, the scale must be inspected and sealed in accordance with local weights and measures regulations before it can be used in LFT mode again.

Figure 1-5. Menu-lock switch on PCB.

1.6.1 Sealing the Scale

When used in conjunction with the Lockout Menu and the Security Switch, the scale may be sealed to prevent or detect unauthorized changes to scale settings. For Legal for Trade applications, set the LFT menu item to ON, set the Security Slide to the On position by sliding the switch to cover the screw, and then seal the scale to prevent access to the metrological parameters. Seal the scale according to the local weights and measures regulations.

To regain access to the scale settings, break the seal and set the security slide to the OFF position.

ON

OFF

Security

Slide

Figure 1-5. Security Switch Settings.

CHAPTER 2 DIAGNOSTIC GUIDE

2.1 TROUBLESHOOTING

This section of the manual contains troubleshooting information. Information is contained to isolate specific problems using Table 2-1, Diagnostic Guide. Follow all directions step by step.

Make certain that the work area is clean. Handle scale components with care. Use appropriate

Electro-Static Device.

Table 2-1 is a Diagnostic Guide designed to help locate the problem area quickly and easily.

The probable causes are listed with the most common cause first. If the first remedy does not fix the problem, proceed to the next remedy. Before attempting to repair the scale, read all chapters of this manual to be familiar with the scale components and operation.

2.2.1 Diagnosis

1. Isolate and identify the symptom

2. Refer to Table 2-1, Diagnostic Guide and locate the symptom.

3. Follow the suggested remedies in the order they appear.

4. Perform the indicated checks, or see the appropriate section of the manual.

5. Repair or replace the defective section of the scale.

Note: If more than one symptom is observed, approach one area at a time,

and remember that the symptoms may be interrelated.

If there is a problem that is not covered in this manual, contact Ohaus for further information.

2.2.2 Testing the Membrane Switch

If the scale does not turn on, or menus cannot be accessed, the Membrane Switch may be defective. It can be tested after removing the Top Housing. (See Section 3.1.)

1. Using an Ohmmeter, measure the resistance between pin 5 and each other pin on the Membrane Switch Cable. (See Figure 2-2, next page.) The readings should be infinite resistance.

2. Press each button and check that the resistance is under 50 ohms for each button.

3. Check that the Membrane Switch is not shorted to the Top Housing case.

Check pins 1 through 5 to the case, which should be infinite resistance.

4. If any test fails, the Membrane Switch is defective. Replace it. (See Section 3.5.)

Figure 2-1. Membrane Switch Cable connections.

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.3 Checking Load Cells for Trouble

This section contains three methods of testing the Load cell.

1. Perform a Ramp Test. (See Appendix B.) If the Ramp Test passes, continue with the rest of this procedure.

2. Perform a Resistance Test, to determine if the Load Cell is severely damaged or a short circuit to the frame has occurred.

Note: The Load Cell must be completely disconnected from the Printed Circuit

Board and at no load when the resistance readings are taken.

Remove the Load Cell cable connector from the PC Board. Using an Ohmmeter, measure Load Cell Resistance Readings on the cable. (See Table 2-1.) If the resistance readings vary more than the table readings allow, replace the Load Cell.

(See Chapter 3.)

TABLE 2-1. LOAD CELL RESISTANCE READINGS (in Ohms)

MEASUREMENT POINTS* READING

Pins 1& 4 (Exe+ to Exe–)

Pins 2& 3 (S+ to S–)

*

See Figure 2-2.

408 ± 10

350 ± 4

Figure 2-2. Main Printed Circuit Board Connections

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.2 Checking Load Cells for Trouble

2. Perform an Output Voltage Test: Measure the no load, 50% load and full load output.

The reading should meet the Load Cell specifications. The Load Cell output should be very close to linear over its capacity range.

NOTE: The following steps involve power applied to the scale. Load Cell solder

contacts can be used as measuring points. (See Figure 2-2.)

TABLE 2-2. COLOR CODE FOR LOAD CELL WIRING*

BLACK RED GREEN WHITE

EXE+ SIG+ EXE– SIG–

*Color codes may vary.

– Insert the Pan Support Caps into the Load Cell Frame, place the Platform on top, and turn on power to the scale.

– Using a voltmeter, measure and record the excitation voltage supplied to the

PCB: with no load on the Platform, measure the voltage across points 1 and 4 of the Load Cell connection on the PCB: +EXE and –EXE. (See Figure 2-2.) This voltage must be approximately 5.0 Volts dc with the Load cell connected. If the voltage is lower, disconnect the Load Cell cable from the PCB and measure again. If the voltage is 5 Volts dc, the Load Cell is defective and must be replaced. If the voltage remains low, the PCB is defective and must be replaced.

CAUTION:

IN THE NEXT STEP, DO NOT OVERLOAD THE SCALE BEYOND FULL

CAPACITY RATING.

– Measure the voltages on +SIG and –SIG wires, disconnected from the PCB

(points 2 and 3 in Figure 2-2). These measurements represent the output of the

Load Cell. Record measurements at Zero Load, 50% and full scale capacities.

See Table 2-3 for typical readings.

NOTE: Table 2-3 indicates typical readings. Actual values can vary, but should

remain linear throughout the range. If readings are out of tolerance, replace the

Load cell. (See Section 3.6.)

TABLE 2-3. LOAD CELL OUTPUT READINGS (in mV with 5V Excitation)

Capacity

3kg

Zero Load

1.2 ± 1.5

50% Load

Zero + 2.2

100% Load

Zero + 4.4

6kg

15kg

30kg

0.4 ± 1.5

0.2 ± 1.5

0.2 ± 1.5

Zero + 2.0

Zero ± 2.5

Zero ± 3.0

Zero + 4.0

Zero ± 5.0

Zero ± 6.0

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.3 Testing the Printed Circuit Board (PCB)

The PCB can be tested by measuring voltages and by using a simulator. The simulator replaces the Load Cell during testing and is a useful tool for diagnosing problems.

PCB Voltage Measurements

Note: Prior to the voltage measurements, the battery should be fully charged and tested.

1. Disconnect power from the scale, and remove the Top housing. (See Section 3.1.)

Leave the Membrane Switch connected to the scale.

2. Disconnect the Battery from the PCB.

3. Connect the AC Adapter to the scale and turn the scale on.

4. Using a DVM, measure the excitation voltage across pins 1 and 4 of J1. (See Figure

2-2.) The reading should be 5 volts dc. This is the excitation voltage for the Load

Cell and is regulated. If the voltage is lower, replace the PCB. (See Section 3.4.)

Then perform Operational Tests. (See Chapter 4.)

5. Measure incoming power from the AC Adapter connector shown in Figure 2-2. The voltage should read between 9 and 14 volts dc with power off and above 9 Volts dc with power on.

6. With the battery disconnected from the PCB, measure voltage across pins 1 and 3 on the battery connector on the PCB. (See Figure 2-2.) The voltage should read

1 to 3 Volts dc.

7. Connect the battery to the PCB and measure the voltage across pins 1 and 3 on the battery connector. The full battery voltage should read a minimum of 6.0 Volts dc.

If the voltage is lower, the battery may require charging or it may be defective. Test the battery. (See Section 2.5.)

8. Perform simulator testing.

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.3 Testing the Printed Circuit Board (PCB)

Simulator Testing

To perform these tests, the use of a Simulator is required. The function of a Simulator is to simulate the output of a full bridge Load Cell, allowing the scale to be separated from the Load

Cell for the purposes of troubleshooting and calibration. The Load Cell used in the scale is rated at 2mV/V output with a 5 Volt excitation voltage applied.

General Load Test

This test checks the Main PC Board circuitry by simulating accurate Load Cell voltages at zero load, 50% and 100% load capacities.

1. Disconnect power from the scale, and remove the Top housing. (See Section 3.1.)

Leave the Membrane Switch connected to the scale. Disconnect the battery.

2. Disconnect the Load Cell cable from connector J1 on the main PC Board.

(See Figure 2-2.)

3. With the Simulator set to zero, connect it to connector J1 using Alligator clips.

4. Connect a known good AC Adapter to the scale and connect to a power source.

5. Turn on the scale. CLEAR PAN, ERROR 14 may appear. This is normal.

6. Set the scale to indicate weight in kilograms (kg) and set the calibration value to maximum span value.

7.

Adjust the Simulator to simulate 0% load, 50% load and 100% load for the capacity that the scale is rated for. (See Table 2.3 for values to use.) If the resulting readings are unstable, the Main PC Board is defective.

8.

Use the Simulator to calibrate the scale in the next procedure to verify if the Main PC

Board is good or bad.

Calibration Test

This test calibrates the scale using the simulator and can verify that the Main PC Board is functioning properly or improperly.

1. With the scale on and the Membrane Switch connected, enter the scale menu and perform a span calibration. (See Appendix A.)

2. Follow the scale prompts. When the scale indicates a given weight to be placed on the scale, set the simulator to an equivalent value based on Table 2-3.

(For example, when a span value of 6 kg for a 6 kg scale is shown, the simulator should be set to .8 mV/V.)

3. Upon completion of calibration, the PCB can be further checked using the Simulator to simulate various weight values. If simulator settings and weight reading on the scale agree, the PCB is functional. If the scale readings vary, or do not agree with readings in Table 2-3, the Main PC Board is defective and should be replaced.

(See Chapter 3.)

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.4 Testing the Battery

The Valor 5000 contains a rechargeable, 6.0 Volt, 4.5 Ampere rated, Lead-Acid Battery.

WARNING! DEATH OR SERIOUS INJURY CAN OCCUR

– Charge the battery only with the scale’s charger. Charging the battery under any other conditions may cause the battery to overheat, emit hydrogen gas, leak, ignite or burst.

– Operate the battery at normal temperature range as specified for the scale.

– Do NOT short the battery terminals under any conditions.

– Do NOT dispose of the battery in incinerators, or crush or try to open.

Dispose of the defective battery in accordance with local regulations for lead-acid type batteries.

– Check the battery for any irregularities in appearance. If there is any damage to the case such as cracks, deformation or leakage, replace the battery.

– Do NOT charge the battery with the charger terminals reversed.

A new battery when installed in a Valor 5000 scale will provide many years of useful service.

The length of service depends upon the type of use of the scale. If left connected to the AC

Adapter all of the time, the battery receives a small charge when the battery voltage is low, and service life can be many years.

Valor 5000 Scales can be operated for up to 120 hours per charge. Battery life can be maximized by frequent charging.

1. Upon receipt of the scale, plug in the AC Adapter to a power source and the scale.

Allow the battery to charge. If the battery required a charge, the battery indicator on the front of the LCD display will indicate charging.

2.

Remove the Top Housing. (See Section 3.1.) Leave the Membrane Switch on the cover connected to the scale.

3. With the AC Adapter connected, measure the battery voltage at the battery terminals. The voltage should be approximately 6 to 7 Volts dc when charging. It should be approximately 6.4 Volts dc ten minutes after charging stops. The scale will operate until the battery voltage drops to approximately 5.6 Volts dc.

NOTE: If the scale operates less than 120 hours, charge the battery and turn on the

scale and test how long the scale operates. If the scale operates less than 60 hours with no backlight, the battery should be replaced. This is 50% capacity.

4.

Table 2-4 indicates the capacity of the battery in percentage after a full charge.

Allow a charged battery to stand overnight without being charged and then measure the terminal voltage to see capacity of the battery.

5.

If the battery requires replacement, see Section 3.6.

ÅCHECK

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.4 Testing the Battery

TABLE 2-4. BATTERY CAPACITY.

7.00

6.75

6.50

(Temperature 25° C)

6.25

6.00

5.75

5.50

5.25

5.00

0

20

40

60

Residual Capacity (%)

80

100

CHAPTER 2 DIAGNOSTIC GUIDE

2.2.5 Diagnosis

Symptom

TABLE 2-5. DIAGNOSTIC GUIDE

Possible Cause Remedy

Cannot turn on with

AC adapter supplied.

No power to scale Verify connections and voltage.

DC input connector at the bottom of scale is defective.

Reconnect the ac adapter to the scale. Check dc voltage on the PCB or test points as shown in Figure 2-2.

Voltage should read between 9 and

14 Volts (power off), and above 9

Volts dc (power on). If voltage is not present, check and replace the

Harness.

Membrane Switch defective.

Main PC Board is defective.

Test the Membrane Switch. (See

Section 2.2.) Replace if necessary.

(See Chapter 3.)

If the scale fails to turn on with a new

Membrane Switch, test the PCB.

If defective it should be replaced.

(See Section 3.4.)

Scale will not turn on using battery power.

Battery discharged or defective.

Wiring harness defective or battery clips connection broken.

Membrane Switch defective.

PCB is defective.

Test the battery. Replace if necessary.

Check voltage at battery contacts first. Check dc voltage at pins 1 and

3 of battery connector on the PCB.

(See Figure 2-2.) Voltage should read approximately 6 Volts dc minimum. If voltage is not present at the connector, disconnect the leads from the battery and make a continuity test between the connectors on the Harness to connector pins 1 and 3 on the PCB.

Resistance should be 0 ohms for the red lead and 0 ohms for the black lead. If an open condition exists, replace wiring as necessary.

Test the Membrane Switch.

Replace if necessary.

If the scale fails to turn on with a new

Membrane Switch, test the PCB.

If defective it should be replaced.

(See Sections 2.3 and 3.4.)

CHAPTER 2 DIAGNOSTIC GUIDE

Symptom

Scale does not respond to front panel controls.

Scale reads incorrectly.

Scale reading unstable.

Cannot calibrate

Error 1

Error 2

Possible Cause

Membrane Switch is defective.

Display is not on or partial characters are displayed.

Main PC Board is defective or

LCD may be defective.

Needs calibration.

Load Cell is damaged.

Main PC Board defective.

Incorrect weights.

LFT lock switch is in locked position; software lock menu is set to ON.

Load Cell is defective.

PCB is defective.

Remedy

Test the Membrane switch. Replace if necessary.

If Membrane Switch is OK, test the

PCB. If defective, replace it.

Test the PCB. Replace PCB or LCD.

NOTE: If the LCD shows signs of

damage (cracked, dim or partial characters), it should be replaced

Perform Service Calibration. (See

Appendix B.) If calibration fails, see next symptom.

Test the Load Cell. (See Section

2.2.3.) If defective replace.

(See Section 3.3.)

Test the PCB. (See Section 2.2.3.)

If defective replace it.

(See Section 3.4.)

Verify that proper weights are used.

Set LFT lock switch to unlocked position and/or set the software lock menu to OFF. (See Section 1.6.)

Test the Load Cell. (See Section

2.2.3.) If defective replace it.

(See Section 3.3.)

Test the PCB. (See Section 2.2.3.)

If defective replace it.

(See Section 3.4.)

The load on the Pan exceeds the rated capacity of the scale

(Overload condition).

Remove excess load. Turn scale off then on.

Needs calibration. Perform service calibration.

(See Appendix A.)

Damaged Load Cell.

Test the Load Cell. (See Section

2.2.3.) If defective replace it.

(See Section 3.3.)

Pan missing. (Underload)

Needs calibration.

Damaged load cell.

Replace the Pan.

Perform Calibration. (Appendix A.)

Test the Load Cell. (See Section

2.2.3.) If defective replace it.

(See Section 3.3.)

CHAPTER 2 DIAGNOSTIC GUIDE

Symptom

Error 9

Error 14

Error 21

CAL E

Possible Cause

Checksum error.

(Configuration data.)

Needs calibration.

PCB defective.

Check capacity. (See Appendix B.)

Perform Service Calibration.

(See Appendix B.)

Replace PCB.

At power up, the load cell signal is outside the allowable limit.

(Zero range error.)

Remove load from pan. If problem persists, t

est the Load Cell. (See

Section 2.2.3.) If defective replace it.

(See Section 3.3.)

Checksum error. (User data menu settings.)

Corrupt checksum error

(calibration).

Incorrect weight used

(Calibration error).

Load cell damaged.

PC board damaged.

Remedy

Check menu settings.

Perform Service Calibration.

(See Appendix B.)

Calibrate using correct calibration weights and in the correct order.

Test the Load Cell. (See Section

2.2.3.) If defective replace it. (See

Section 3.3.)

Test the PCB. (See Section 2.2.3.)

If defective replace it.

(See Section 3.4.)

CHAPTER 3 MAINTENANCE PROCEDURES

3.1 PREVENTIVE MAINTENANCE

Ohaus scales are precision instruments and should be carefully handled, stored in a clean, dry, dust-free area, and cleaned periodically. Follow these precautionary steps:

– When a scale has had chemicals or liquids spilled on it, all exterior surfaces should be cleaned as soon as possible with warm water on a damp cloth.

– Do not leave a mass on the scale when the scale is not in use.

– Allow time for the scale to stabilize after moving it from an area which is at a different temperature than the area where it is to be operated. Allow one hour for each 5°F

(2.7°C) temperature change before using the scale. After temperature stabilization, allow an additional 20 minutes after turning the scale on, for the scale electronics to stabilize.

3.1.1 Preventive Maintenance Checklist

The scale should be inspected and checked regularly, as follows:

1. Remove the Pan and Sub Pan to inspect and clean the area beneath the Pan.

2. Clean the outside of the scale using a damp cloth with warm water.

CAUTION

DO NOT USE CHEMICAL CLEANERS OR SOLVENTS OF ANY TYPE.

SOME CLEANERS ARE ABRASIVE AND MAY AFFECT THE SCALE’S FINISH.

3. Check the Power Cord for broken or damaged insulation.

4. Make a visual inspection for faulty connectors, wiring, and loose hardware.

CHAPTER 3 MAINTENANCE PROCEDURES

3.2 REMOVING THE TOP HOUSING

Common hand tools are sufficient to disassemble the Valor 5000 scales. Turn the scale off and unplug the power cord before you begin.

1. Carefully lift and remove the Pan from the scale.

2. Unscrew and remove the four Pan Support Caps from the top of the scale.

3.

Turn the scale over and remove the five screws which hold the Top Housing. (See

Figure 3-1.) Note that one of these is smaller, and is beneath the Security Slide.

Battery screws

Load Cell screws

Load Cell screws

Housing screws

LFT

Security

Figure 3-1. Scale Bottom.

CAUTION

Use care in the next step as the Membrane Switch wiring is attached to the PC Board.

4. Place the scale upright and carefully lift the Top Housing from the Base Housing slightly, and tilt forward. Reach under the Top Housing from the front and carefully disconnect the flexible Membrane Switch Cable from PC Board.

5.

Set the Top Housing aside.

CHAPTER 3 MAINTENANCE PROCEDURES

3.3 REPLACING THE LOAD CELL (Either as Assembly or as Component)

The Load Cell may need to be replaced because of scale instability, or because the scale does not calibrate or repeat, or because it is physically broken or displays an error code.

The entire Load Cell Assembly, with the Frame, can be replaced, or only the Load Cell itself, as an individual component. (See Chapter 5 for parts information.) If it is replaced as a component, the Overload Protection Stops must be adjusted. (See Section 3.3.2.)

3.3.1 Replacing the Load Cell Assembly (with Frame)

The Load Cell is factory assembled and the Overload Protection Stops are pre-adjusted.

There are no adjustments to be made on this assembly after installation.

1. Remove the Top Housing, and disconnect the flexible Membrane Switch Cable from the PC Board. (See Section 3.2.)

2. Disconnect the Load Cell Cable from the PCB. (See Figure 3-2.)

3. Remove the four screws holding the Load Cell Assembly to the Bottom Housing.

(See Figure 3-1.)

4. Lift the Load Cell Assembly out of the bottom housing and replace it with a new one. Install the screws in the Bottom Housing that were removed in Step 2.

5. Connect the Load Cell Cable to the PCB.

6. Replace the Top Housing and connect the Membrane Switch cable to the PCB.

Screws holding Load

Cell Frame

Load Cell Cable, disconnected from PCB

7. Make sure the LFT Security Slide is in the OFF position.

8. Turn on the scale. The display may show an error code. This is normal.

9. Calibrate the scale according to the instructions in Appendix A.

10. Turn the scale off then on. It will go through all the self-testing.

11. Perform a Service Calibration.

(See Appendix B.)

12. Run Performance Tests. (See Chapter 4.)

Figure 3-2. Valor 5000 with Top Housing off.

CHAPTER 3 MAINTENANCE PROCEDURES

3.3.1 Replacing the Load Cell Component

Use this procedure to keep the existing Load Cell Frame and just replace the Load cell.

Note: When the new Load Cell is mounted to the existing Frame, the Overload

Protection Stops must be adjusted.

1. Remove the Top Housing, and disconnect the flexible Membrane Switch Cable from the PC Board. (See Section 3.2.)

2. Disconnect the Load Cell Cable from the PCB. (See Figure 3-2.)

3. Remove the four screws holding the Load Cell Assembly to the Bottom Housing.

(See Figure 3-1.)

4. Remove the screws holding the Load Cell Frame to the Load Cell, and lift it off.

5. Remove the cable clamp holding the Load Cell cable to the Frame.

7. Install the new Load Cell: insert the four screws that hold it to the Frame. Torque the screws as indicated in Table 4-1. Be sure the Load cell is properly aligned.

TABLE 3-1. LOAD CELL TORQUE SETTINGS

CAPACITY TORQUE

3 kg

6 kg

15 kg

30 kg

10 N.m / 88 in. lb

10 N.m / 88 in. lb

10 N.m / 88 in. lb

10 N.m / 88 in. lb

8. Re-attach the cable clamp holding the Load Cell Cable to the Frame assembly

9. Re-install the four screws holding the Load Cell Assembly to the Bottom Housing.

(See Figure 3-1.)

10. Re-connect the Load Cell Cable to the PC Board.

11. Replace the Top Housing and connect the Membrane Switch cable to the PCB.

12. Make sure the LFT Security Slide is in the OFF position.

13. Turn on the scale. The display may indicate an error code. This is normal.

14. Perform a Service Calibration. (See Appendix B.)

15. Run Performance tests. (See Chapter 4.)

CHAPTER 3 MAINTENANCE PROCEDURES

3.3.2 Overload Protection Stop Adjustment

When a Load Cell has been replaced in the existing Load Cell Frame, the Overload Protection

Stops must be set before calibrating or testing the scale. This requires mm feeler gauges ranging from 0.35 mm to 2.1 mm or inch feeler gauges 0.013 in. to 0.082 in. (See Table 3-2.)

1. At each Overload Stop Protection location, insert the proper feeler gauge as specified in

Table 3-2.

2. With the correct feeler gauge positioned in the first Overload Protection Stop, press down on the Frame and check for movement of the Frame. Adjust the Overload

Protection Stop until there is no movement with the feeler gauge inserted. The resulting gap should be the thickness of the feeler gauge.

3. Repeat this procedure for all of the Overload Protection Stops, using the appropriate feeler gauges.

4. When the initial gap settings have been made, recheck all gap settings and adjust as required.

5. Use a thread locking solution to coat all down stop screws so they will not move.

TABLE 3-2. LOAD CELL OVERLOAD PROTECTION STOP GAP SETTINGS

CAPACITY

3 kg

6 kg

15 kg

30 kg

LONG ARM

0.7 mm / 0.028 in.

1.0 mm / 0.039 in.

1.4 mm / 0.055 in

2.5 mm / 0.098 in

MID

0.5 mm / 0.02 in.

0.5 mm / 0.02 in.

0.7 mm / 0.028 in.

0.9 mm / 0.035 in

SHORT ARM

0.5 mm / 0.019 in.

0.75 mm / 0.029 in.

1.2 mm / 0.047 in.

1.4 mm / 0.055 in

6. Reassemble the scale.

7. Perform a Service Calibration. (See Appendix B.)

8. Run Performance tests. (See Chapter 4.)

CHAPTER 3 MAINTENANCE PROCEDURES

3.4 Replacing the Printed Circuit Board and LCD Display

The Printed Circuit Board (PCB) is located inside the scale towards the front. To replace the

PCB it is necessary to disassemble the scale. The LCD Display may also be replaced as a separate item if it is cracked or partial displays are shown.

1. Remove the Top Housing, and disconnect the flexible Membrane Switch Cable from the

PC Board. (See Section 3.2.)

2. Disconnect the Load Cell Cable from the PCB. (See Figure 3-2.)

Screws holding

PCB in place

Figure 3-3. PCB Connector Locations

3. Remove the connector plugs from the Battery and AC Adapter. Remove the four screws holding the PCB. (See Figure 3-3.)

4. Lift the PCB up from the Housing. Insert the replacement PCB.

5. Follow these steps in reverse order to install the new PCB.

6.

Replace the Top Housing and connect the Membrane Switch cable to the PCB.

7. Make sure the LFT Security Slide is in the OFF position.

8. Turn on the scale. When self-testing is completed, the display may indicate an error code. This is normal.

9. Configure the Scale. (See Appendix C.)

10.

Perform a Service Calibration. (See Appendix B.)

11.

Run Performance tests. (See Chapter 4.)

Note: The PCB and the LCD Display are supplied as a single unit. However, if only the

LCD Display needs replacement, it can be separated from the PCB by removing the two screws in the bottom of the PCB, and unsoldering the 24 fine lead-wires connecting it to the PCB. When installing the new LCD Display, carefully feed the lead-wires through their holes, then insert the two screws, and finally solder the 24 lead-wires.

CHAPTER 3 MAINTENANCE PROCEDURES

3.5 REPLACING THE MEMBRANE SWITCH

The membrane switch is affixed to the Top Housing of the scale. To replace the Membrane

Switch, the scale must be disassembled to gain access to the switch connections.

1. Remove Top Housing. (See Section 3.2.)

CAUTION

Use care in the next step as the Membrane Switch wiring is attached to the PC Board.

2. Place the scale upright and carefully lift the Top Housing from the Base Housing slightly.

Reach under the Top Housing from the front and carefully disconnect the flexible

Membrane Switch Cable from the PCB.

3. On the Top Housing, lift up the defective Membrane Switch (if necessary carefully pry it up with a knife) and gently peel it off the Upper Housing.

4. Carefully clean the Top Housing Membrane Switch area, removing all traces of adhesive. The best method is to use a flat razor blade.

5. Insert the cable from the new Membrane Switch through the hole in the Top Housing.

Peel off the protective film from the new Membrane Switch and carefully align and affix to the Top Housing.

6. Press the Membrane Switch down uniformly. Using your fingers with a cloth, roll from the center of the Membrane Switch outward towards the edges to remove any air bubbles that may be trapped.

CAUTION

DO NOT PRESS ON THE CLEAR WINDOW AREA OR ON THE SWITCHES.

7. Position the Top Housing over the Bottom Housing and connect the flexible Membrane

Switch Cable to the PC Board.

8. Reassemble the scale.

9.

Replace the Pan Support Caps and Pan on top of the scale.

10.

Run Performance tests. (See Chapter 4.)

CHAPTER 3 MAINTENANCE PROCEDURES

This procedure is used when it is required to change the Battery.

WARNING! DEATH OR SERIOUS INJURY CAN OCCUR

– Charge the battery only with the scale’s charger. Charging the battery under any other conditions may cause the battery to overheat, emit hydrogen gas, leak, ignite or burst.

– Operate the battery at normal temperature range as specified for the scale.

– Do NOT short the battery terminals under any conditions.

– Do NOT dispose of the battery in incinerators, or crush or try to open.

Dispose of the defective battery in accordance with local regulations for lead-acid type batteries.

– Check the battery for any irregularities in appearance. If there is any damage to the case such as cracks, deformation or leakage, replace the battery.

– Do NOT charge the battery with the charger terminals reversed.

1. Remove the Top Housing. (See Section 3.2.)

2. Disconnect the wire Harness from the Battery.

3. Remove the two screws holding the Battery Holder on top of the Battery. These screws are reached from underneath the scale on the Bottom Housing.

4. Lift the Battery out of the Bottom Housing.

5. Position the new Battery in the Bottom Housing.

6.

Reinstall the Battery holder using the two screws.

7. Place the Battery Holder over the Battery and secure it with the two screws previously removed.

8. Attach the two battery wires with quick-clips to the new Battery posts. Attach the Red wire to the positive terminal on the Battery and the Black wire to the negative terminal.

CAUTION:

Dispose of the Battery according to local regulations for hazardous materials.

Do not incinerate, crush or throw out in regular garbage containers.

Battery contains an acid solution.

CHAPTER 4 TESTING

4.1 TESTING

Before and after servicing a Valor 5000 scale, various performance tests should be made to confirm that the scale meets specifications. Turn the scale on and allow it to warm up for at least 20 minutes before performing these tests.

NOTE:

Make sure the test area is free from drafts and that the scale rests on a level and vibration-free surface.

4.1.1 Test Masses Required

The masses required to test the Ohaus Valor 5000 scales must meet the requirements of ASTM

Class 4 or OIML F2 Tolerance. The mass values are listed in Table 4-1.

TABLE 4-1. CALIBRATION MASS VALUES

MODEL Span Cal Choices Linearity Cal (fixed)

V51P3

1, 2, 3 kg

V51P6

2, 4, 6 kg

V51P15

5 10, 15 kg

V51P30

10, 20, 30 kg

4.2 PERFORMANCE TESTS

1 or 2, 3 kg

2, 6 kg

10, 15 kg

15, 30 kg

Service Cal.

1, 3 kg

2, 6 kg

5, 15 kg

10, 30 kg

Accurate performance of the Valor 5000 scales is determined by a series of performance tests.

The displayed readings are compared with the tolerances listed in Tables 1-1 and 1-2.

Tolerance values are expressed in counts. A one-count difference is shown in the last digit on the scale display.

NOTE:

The following performance tests are used to evaluate scale operation before and after repairs. The scale must meet the requirements specified in each test as well as the specifications listed in Tables 1-1 and 1-2. Before proceeding with the following tests, the scale should be calibrated. (See Appendix A.)

4.2.1 Segment Display Test

Turn the scale on, and ensure that all segments are enabled and displayed briefly. This is a Segment Display Test. (See Figure 4-1.)

Figure 4-1. Segment Display

4.2.2 Load Cell Test Using RAMP

To test the Load Cell using RAMP, see Appendix B.

CHAPTER 4 TESTING

4.2.3 Precision Test

The Precision Test measures the Standard Deviation of a set of similar weight readings, which should match the specification for each model, listed in Tables 1-1 and 1-2.

1. Power on the balance. The reading on the display should be 0g.

2. Select a mass weighing near the maximum capacity of the balance, and place it on the center of the Pan. Observe and record the reading.

3. Remove the mass. The reading should return to 0g ±1 count.

4. Repeat this test three times. The reading should be within ±1 count of the reading recorded. If so, the balance passes the Precision Test.

5. If the deviation for any set of readings (using the same mass placed on the center of the Pan) is greater than the tolerance listed in Tables 1-1 and 1-2, the balance does not meet the precision specification. Inspect and correct the following areas:

– Check for mechanical obstructions. Any foreign object touching any part of the moving assemblies will cause a balance to fail the Precision Test. Inspect and correct as necessary.

– If the scale does not meet specifications, move it to a suitable location, ensure that it is level, and try again. If it still does not meet specifications, perform a service calibration, and try again. (See Appendix B for Service Calibration.)

CHAPTER 4 TESTING

Repeatability is the Standard Deviation of a set of similar weight readings.

Requirements:

– To perform this test a single mass must be used for all readings.

– The test mass should be approximately ½ of the capacity of the instrument.

– Wear gloves when handling the mass.

Before starting a repeatability test, set up the instrument as follows:

Set Up:

Enter the menu and adjust and record the following settings:

A. Set the Stability setting to its lowest setting.

B. Set the Filter level to medium or the center of its range.

C. Set the AZT (Auto Zero Tracking) to .5d or its lowest setting. Do not turn it off.

D. Set the instrument to display the same units as the performance specifications.

(Usually kg, g, or mg)

Record Settings:

Stability Setting =

Filter Level Setting =

____________

____________

Auto Zero Tracking Setting = ____________

Displayed Units = ____________

Mass Used = ____________

TEST PROCEDURE:

1. Zero the instrument, if it does not read zero.

2. Using a test mass approximately half the capacity of the instrument, place the mass on the center of platform. Record the reading on the worksheet provided.

3. Remove the mass from the platform.

4. Repeat this test starting at Step 1 until you record a total of ten readings

Fill in the worksheet (Table 4-2) with the ten (10) readings.

CHAPTER 4 TESTING

TABLE 4-2. REPEATABILITY WORKSHEET

n Reading

1

Delta = Reading – Mean

2

3

4

5

6

7

8

9

10

Delta x Delta

n = number of Reading Mean = Sum of readings / 10 Delta = Reading – Mean

Standard Deviation = Square Root of (sum of (Delta x Delta) / 9)

5. Add the ten readings and divide the total by 10 to find the Mean (average).

6. Mean = (Reading 1 + Reading 2 + Reading 3 + Reading 4 + Reading 5

7. + Reading 6 + Reading 7+ Reading 8 + Reading 9 + Reading 10) / 10

Mean =________

6. Calculate the Delta for each reading and record in the work sheet.

Delta = Reading – Mean

7. Calculate the Delta x Delta for each reading and record in worksheet.

8. Add the ten Delta x Delta values and divide by 9

9.

Calculate the Standard Deviation by applying the square root of the result from step 8.

Standard Deviation

=___________

10. Note: If the balance does not meet specifications, move it to a suitable location, ensure that it is level, and try again. If it still does not meet specifications, perform a service calibration, and try again. (See Appendix B for Service Calibration.) If the Scale continues failing to meet specifications, the Load Cell may need to be replaced.

CHAPTER 4 TESTING

This test is used to determine the linearity of the unit throughout its operating range.

The masses used to perform this test can be utility masses.

NOTE:

The scale must pass the Precision and Repeatability Tests, and be calibrated before the Linearity Test may be performed.

TABLE 4-4. LINEARITY TEST MASSES

Capacity:

Test Mass:

3 kg

1 kg

6 kg

1 kg

15 kg

1 kg

30 kg

1 kg

Load 1:

Load 2:

0.5 kg

1 kg

2 kg

3 kg

5 kg

10 kg

10 kg

15 kg

Load 3:

2 kg 5 kg 14 kg 25 kg

NOTE:

All masses are nominal values. Be certain to use the same reference mass throughout the procedure.

1. Place the test mass on the Scale, record the weight and remove.

2. Place Load 1 on the Scale and press TARE.

3. Place the test mass on the Scale, record the weight and remove.

4. Place Load 2 on the Scale and press TARE.

5. Place the test mass on the Scale, record the weight and remove.

6. Place Load 3 on the Scale and press TARE.

7. Place the test mass on the Scale and record the weight.

8. The difference in the weights of the test mass should be within the tolerance in

Tables 1-1 and 1-2. If not, calibrate (see Appendix A.1) and repeat the test.

9. If the Scale remains out of tolerance, the Load Cell may need to be replaced.

CHAPTER 4 TESTING

The Off-Center Load Test is used to determine whether displayed weight values are affected by moving the sample to different areas of the Pan.

1. Place half of the scale’s capacity in the center of the Pan.

2. Note the reading.

3. Move the mass halfway (between the center and the edge) to the front of the Pan.

Note any differences in the displayed weight reading.

4. Repeat the test for the back, left, and right position of the Pan.

5. Maximum allowable change in displayed weight readings for each of the four positions can be found in Specifications Tables (Chapter 1). If this maximum is exceeded, follow procedures in Section 4.2.7, Adjusting Off Center Load.

Figure 4-2. Off-Center Load Test Weight Locations.

CHAPTER 4 TESTING

4.2.7 Adjusting Off Center Load

If the Off Center Load (OCL) is excessive, perform adjustment as follows:

Top view of Load Cell

Side view of Load Cell

2

3

Weighing Pan’s center, with points

A – D indicated

1

4

Figure 4-3. Scale drawing of Valor 5000 Load Cell and Weighing Pan.

1. Place the test weight in the center of the Weighing Pan.

2. Tare the balance.

3. Move the weight to point A and record the reading.

4. Move the weight to point B and record the reading.

5. Move the weight to point C and record the reading.

6. Move the weight to point D and record the reading.

7. If the reading at point A is negative, file at points 1 and 4 AT AN ANGLE.

8. If the reading at point B is negative, file at points 1 and 2 STRAIGHT ACROSS.

9. If the reading at point C is negative, file at points 2 and 3 AT AN ANGLE.

10. If the reading at point D is negative, file at points 3 and 4 STRAIGHT ACROSS.

11. Repeat 1 to 10 until within specifications.

Note: It is not recommended that you try to adjust more than –5 counts if the beam has

been filed already. If the beam has not been filed previously, you can adjust –10 counts.

Remember, when filing you are weakening the beam. File a little at a time.

CHAPTER 4 TESTING

CHAPTER 5 PARTS LISTS & DIAGRAMS

This section of the manual contains exploded views for the Valor 5000 Series scales. The exploded view drawings are designed to identify the parts which can be serviced on the scale in the field.

NOTE:

In all cases where a part is replaced, the scale must be thoroughly checked after the replacement is made. The scale MUST meet the parameters of all applicable specifications in this manual.

If further technical information is needed, please contact your local Ohaus distributor, or:

Ohaus Corporation, www.ohaus.com

19A Chapin Road

P.O. Box 2033

Pine Brook, NJ 07058-2033 USA

Tel: 973-377-9000

Fax: 973-593-0359

In the United States call toll free, 800-526-0659 between 8:00 a.m. and 6:00 p.m. EST.

CHAPTER 5 PARTS LISTS & DIAGRAMS

5.1 VALOR 5000 SERIES SCALES: HOUSING & INTERNAL PARTS

9

10

11

1

2

3

4

5

6

7

8

12

13

80

14

Figure 5-1. Valor 5000 Series Scales: Housing & Internal Parts.

CHAPTER 5 PARTS LISTS & DIAGRAMS

5.1 VALOR 5000 SERIES SCALES: HOUSING & INTERNAL PARTS

TABLE 5-1. VALOR 5000 SERIES SCALES: HOUSING & INTERNAL PARTS

Drawing Item

1

Description

Pan Support Caps (4)

4

6

Load Cell Assembly (includes Frame)

Main Printed Circuit Board, with LCD

9 Pan

10 Rechargeable Battery Pack

11 Harness for Rechargeable Battery

13

14

Seal, Adapter & Foot Switch Socket

Adjustable Feet, Set of 4

80 Adapter

CHAPTER 5 PARTS LISTS & DIAGRAMS

APPENDIX A STANDARD CALIBRATION

APPENDIX A.

STANDARD CALIBRATION

A.1 CALIBRATION

Standard calibration should be performed prior to using a scale, and after service. Calibration can be adjusted in two ways: Span and Linearity. Span calibration uses two calibration points to adjust the sensitivity of the scale. Linearity calibration uses three calibration points to correct for non-linear weighing results. See Section 4.2 for Calibration Masses required for each model.

Calibration is not available when the scale has been set to Legal for Trade operation. To disable LFT, and then restore it if necessary, see Section 1.6.

Note:

Be careful not to touch the scale or the table while calibration is in progress, as it will cause the process to fail.

Span Calibration Procedure

Press TARGET/Menu until CAL appears.

Press Yes.

Span appears. Press Yes to begin Span

calibration. Clear Pan flashes.

Clear the pan, then press Yes.

--C-- appears, followed by the span

calibration weight value.

Press No to change to an alternate span calibration weight value, or place the indicated calibration weight on the scale and press Yes.

--C-- appears then done, and then the

scale reverts to weighing mode.

Remove the calibration weight.

Figure A-1. Span Calibration Menu readings.

Note: The calibration procedure can be canceled at any time by pressing Exit

or turning the scale off.

APPENDIX A STANDARD CALIBRATION

A.1 CALIBRATION

Linearity Calibration Procedure

Press TARGET/Menu until CAL appears.

Press Yes. Span appears. Press No.

Lin appears. Press Yes to begin Linear

calibration. Clear Pan flashes.

Clear the pan if necessary, then press Yes.

--C-- appears, followed by the first

calibration weight value. Press No to change to an alternate first calibration weight value, or• place the indicated calibration weight on the scale and press

Yes.

--C-- appears, followed by the second

calibration weight value. Place the indicated calibration weight on the scale and press Yes.

--C-- appears, then “done,” and then the

scale reverts to weighing mode.

Remove the calibration weight.

Note: The calibration procedure can be

canceled at any time by pressing Exit or turning the scale off.

Figure A-2. Linear Calibration Menu readings.

APPENDIX B THE SERVICE MENU

APPENDIX B.

THE SERVICE MENU

This section describes the Service menu and sub-menus, which allow authorized service personnel to perform factory settings and calibration (no pre-set limits), as well as setting the

Geo code. There is also a Ramp display used to indicate the condition of the load cell. Please refer to the Service Menu diagram

Figure A-1. Service Menu Flow Chart

B.1 ENTERING THE SERVICE MENU

1. The following conditions must be met before entering the Service menu.

– The Security Slide on the bottom of the scale must be in the OFF position.

(See Figure 1-5, page 1-8)

NOTE: This requirement is waived when there is an EEPROM checksum error.

2. Turn the scale on.

3. To enter the Service Menu, press and hold TARE and G/N/T together for at least ten seconds, until SERVE appears briefly, followed by CAP.

B.2 NAVIGATION

A blinking item on the display indicates a choice to make:

Press Yes to accept the displayed choice.

Press No to advance to the next choice or the next menu.

Press Exit to exit the service menu and return to displaying the weight.

Press Back to back up one step in the menu.

APPENDIX B THE SERVICE MENU

B.3 CAPACITY

The capacity function allows you to set the capacity to match the installed Load Cell.

1. Press and hold TARE and G/N/T together for at least 10 seconds.

SERVE appears.

2. When you release the buttons, CAP appears.

3. To check the scale’s capacity, press ON/ZERO

Off

. 3 kg flashes.

SErVE

CAP

3

kg

4. To accept the displayed capacity, press Yes.

To change the capacity to match the Load Cell, press No until the correct capacity appears, then press Yes to accept.

Then GEO appears.

GEO

This menu item is used to compensate for slight variations in gravity at different geographical locations around the world. It allows entry of values from 0 to 31. (Complete Geographical

Adjustment Factors are listed in Appendix C.) The default Geo factor is 19 for Europe and 16 for the USA.

This feature allows authorized personnel to accurately calibrate the scale at a location other than the location where the scale is to be used. Prior to calibration, set the Geo factor to correspond to the geographical location where the calibration is being performed.

Following calibration

, change the Geo factor to match the location where the scale is to be used.

If required, the scale may also be sealed according to the required approval regulations.

Geographical location flashes, for example, GEO 19.

GEO 19

6. Locate the Geographical Factor in Appendix C. Select a number from 0 to 31.

7. To change the Geographical factor, press No until the desired GEO

Factor appears, then press Yes.

CAL is now displayed.

GEO 16

CAL

APPENDIX B THE SERVICE MENU

Service calibration is used when span or linearity calibration fails or whenever a Load Cell or PC

Board has been changed. Using this procedure establishes a new set of parameters for the scale and new calibration data.

2. Remove any material from the Pan and press Yes. - - C- - appears.

3. After a few seconds, PLACE (n) kg flashes, where (n) is the first calibration weight.

4. Place the requested weight on the Pan and press Yes. - - C- - appears briefly, followed by PLACE (n) kg where (n) is the second calibration weight.

CLEAr PAN

--C--

PLACE 1

PLACE 3

kg

kg

5. Place the requested weights on the Pan and press Yes.

- - C- - appears briefly, followed by DONE.

DONE

Within a few seconds, the scale returns to weighing mode and displays the current weight on the Pan.

6.

Remove the weights from the Pan.

Note:

Be careful not to touch the scale or the table while calibration is in progress, as it will cause the process to fail.

APPENDIX B THE SERVICE MENU

B.6 Ramp

Ramp is used primarily to determine if the Load Cell is damaged from causes including abuse, heavy weights dropped on the Pan, or dropping the scale. The ramp function displays the duty cycle of the A/D converter output to within one tenth of a percent.

When in the Ramp Menu, placing a small weight on the Pan will cause the reading to increment slightly. Placing additional weights of the same value will also increment to a greater displayed value. This is normal operation. A Load Cell that is damaged will be non-linear as additional weights are added and might also have rapidly changing numbers. In this case, the Load Cell must be replaced.

1. When RAMP appears, press Yes to display the scale’s ramp values.

2. With an empty Pan, record the no load ramp value.

3. Place half load on the pan and record the ramp value.

4. Place a full load on the Pan and record the ramp value.

5. Subtract the no load ramp value from the half load ramp value.

6. Subtract the half load ramp value from the full load ramp value.

The two calculated values should be equal.

The ramp value should be between 20% to 60% throughout testing.

Values outside of this range indicate the load cell should be replaced.

7. At the end of the testing, press Yes.

E.PAND appears. Disregard this display. It is not used.

9.

Press Yes to exit and return to the weighing mode.

APPENDIX C GEOGRAPHICAL ADJUSTMENT VALUES

APPENDIX C.

GEOGRAPHICAL ADJUSTMENT VALUES

TABLE D-1. GEOGRAPHICAL ADJUSTMENT VALUES.

Elevation above sea level in meters

Geographical latitude away from the equator (North or South), in degrees and minutes.

0°00

′ - 5°46′

5°46

′ - 9°52′

9°52

′ - 12°44′

12°44

′ - 15°06

15°06

′ - 17°10′

17°10

′ - 19°02′

19°02

′ - 20°45′

20°45

′ - 22°22′

22°22

′ - 23°54′

23°54

′ - 25°21′

25°21

′ - 26°45′

26°45

′ - 28°06′

28°06

′ - 29°25′

29°25

′ - 30°41′

30°41

′ - 31°56′

31°56

′ - 33°09′

33°09

′ - 34°21′

34°21

′ - 35°31′

35°31

′ - 36°41′

36°41

′ - 37°50′

37°50

′ - 38°58′

38°58

′ - 40°05′

40°05

′ - 41°12′

41°12

′ - 42°19′

42°19

′ - 43°26′

43°26

′ - 44°32′

44°32

′ - 45°38′

45°38

′ - 46°45′

46°45

′ - 47°51′

47°51

′ - 48°58′

48°58

′ - 50°06′

50°06

′ - 51°13′

51°13

′ - 52°22′

52°22

′ - 53°31′

53°31

′ - 54°41′

54°41

′ - 55°52′

00 0

325

325

650

650

975

0 975

1300

1300

1625

Elevation above sea level in feet

1625

1950

000 0

1060

1060

2130

2130

3200

4260

5330

5330

6400

16

17

17

18

14

15

15

16

12

13

13

14

10

11

11

12

18

19

19

20

20

0 8

0 8

0 9

0 9

10

0 6

0 6

0 7

0 7

0 3

0 3

0 4

0 4

0 5

0 5

16

17

17

18

14

15

15

16

18

19

19

20

12

13

13

14

10

10

11

11

12

0 8

0 8

0 9

0 9

0 6

0 6

0 7

0 7

0 2

0 3

0 3

0 4

0 4

0 5

0 5

3200

4260

16

17

17

18

18

14

15

15

16

12

13

13

14

10

11

11

12

19

19

20

20

21

0 8

0 9

0 9

10

0 6

0 7

0 7

0 8

0 3

0 4

0 4

0 5

0 5

0 6

18

18

19

19

16

16

17

17

14

14

15

15

12

12

13

13

20

20

21

21

22

22

10

10

11

11

0 8

0 8

0 9

0 9

0 5

0 5

0 6

0 6

0 7

0 7

18

18

19

19

16

16

17

17

14

14

15

15

12

12

13

13

20

20

21

21

22

0 9

10

10

11

11

0 7

0 8

0 8

0 9

0 4

0 5

0 5

0 6

0 6

0 7

15

16

16

17

17

18

18

13

14

14

15

11

12

12

13

19

19

20

20

21

21

0 9

10

10

11

0 7

0 8

0 8

0 9

0 4

0 4

0 5

0 5

0 6

0 6

0 7

1950

2275

2275

2600

2600

2925

2925

3250

3250

3575

6400

7460

7460

8530

17

17

18

18

15

15

16

16

19

19

13

13

14

14

11

11

12

12

0 8

0 8

0 9

0 9

0 6

0 6

0 7

0 7

10

10

0 4

0 4

0 5

0 5

0 2

0 2

0 3

0 3

15

15

16

16

12

13

13

14

14

10

11

11

12

0 8

0 9

0 9

10

17

17

18

18

19

0 6

0 7

0 7

0 8

0 4

0 5

0 5

0 6

0 1

0 2

0 2

0 3

0 3

0 4

8530

9600

0 9600

10660

14

15

15

16

12

13

13

14

10

11

11

12

0 8

0 9

0 9

10

16

17

17

18

18

0 6

0 7

0 7

0 8

0 4

0 4

0 5

0 5

0 6

0 1

0 1

0 2

0 2

0 3

0 3

14

15

15

16

12

13

13

14

16

17

17

18

10

11

11

12

0 8

0 9

0 9

10

0 6

0 7

0 7

0 8

0 4

0 4

0 5

0 5

0 6

0 0

0 1

0 1

0 2

0 2

0 3

0 3

10660

11730

13

13

14

14

11

11

12

12

0 9

0 9

10

10

0 7

0 7

0 8

0 8

15

15

16

16

17

17

0 3

0 3

0 4

0 4

0 5

0 5

0 6

0 6

0 0

0 0

0 1

0 1

0 2

0 2

APPENDIX C GEOGRAPHICAL ADJUSTMENT VALUES

TABLE D-1. GEOGRAPHICAL ADJUSTMENT VALUES.

Elevation above sea level in meters

Geographical latitude away from the equator (North or South), in degrees and minutes.

55°52

′ - 57°04′

57°04

′ - 58°17′

58°17

′ - 59°32′

59°32

′ - 60°49′

60°49

′ - 62°09′

62°09

′ - 63°30′

63°30

′ - 64°55′

64°55

′ - 66°24′

66°24

′ - 67°57′

67°57

′ - 69°35′

69°35

′ - 71°21′

71°21

′ - 73°16′

73°16

′ - 75°24′

75°24

′ - 77°52′

77°52

′ - 80°56′

80°56

′ - 85°45′

85°45

′ - 90°00′

00 0

325

325

650

650

975

0 975

1300

1300

1625

Elevation above sea level in feet

1625

1950

000 0

1060

23

23

24

24

25

25

26

26

27

27

28

28

29

29

30

30

31

1060

2130

22

23

23

24

24

25

25

26

26

27

27

28

28

29

29

30

30

2130

3200

22

22

23

23

24

24

25

25

26

26

27

27

28

28

29

29

30

3200

4260

21

22

22

23

23

24

24

25

25

26

26

27

27

28

28

29

29

4260

5330

26

27

27

28

24

25

25

26

28

21

21

22

22

23

23

24

29

5330

6400

26

26

27

27

24

24

25

25

28

20

21

21

22

22

23

23

28

1950

2275

6400

7460

7460

8530

25

26

26

27

23

24

24

25

27

20

20

21

21

22

22

23

28

2275

2600

24

25

25

26

22

23

23

24

26

27

19

20

20

21

21

22

27

2600

2925

8530

9600

24

24

25

25

22

22

23

23

26

26

19

19

20

20

21

21

27

2925

3250

0 9600

10660

3250

3575

10660

11730

23

24

24

25

21

22

22

23

25

26

18

19

19

20

20

21

26

23

23

24

24

21

21

22

22

25

25

18

18

19

19

20

20

26

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com