IP 140 IMPAC-Pyrometer Operation Manual · Betriebsanleitung Proven Quality IMPAC pyrometer IP 140 © LumaSense Technologies 2009. All rights reserved. LumaSense Technologies 3033 Scott Blvd. Santa Clara, CA 95054-3316 LumaSense Technologies 16 Thornton Road Oakland, NJ 07436 Tel.: +1 408 727-1600 Fax: +1 408 727-1677 Tel.: +1 201 405-0900 Fax: +1 201 405-0090 Internet: www.lumasenseinc.com E-mail: [email protected] [email protected] Internet: www.mikroninfrared.com E-mail: [email protected] Please call 800-631-0176 for Factory Repair and Return • • • • There are no user-serviceable components in the instrument Disassembly of the instrument is not allowed, and the warranty is VOID if the instrument is disassembled, tampered with, altered or otherwise damaged, without prior written consent from LumaSense Technologies or if considered by LumaSense Technologies to be abused or used in abnormal conditions. No adjustments may be made to the targeting laser. It is fixed at the factory. No adjustments may be made to the targeting laser’s power level. General Information about the user manual Congratulations on choosing the high quality and highly efficient IMPAC pyrometer. Please read this manual carefully, step by step, including all notes to security, operation and maintenance before installing the pyrometer. For installation and operation of the instrument this manual is an important source of information and work of reference. To avoid handling errors keep this manual in a location where you always have access to. When operating the instrument it is necessary to follow the general safety instructions (see section 3, Safety). Additionally to this manual the manuals of the components used are valid. All notes – especially safety notes – are to be considered. Limit of liability and warranty All general information and notes for handling, maintenance and cleaning of this instrument are offered according to the best of our knowledge and experience. LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect. LumaSense Technologies reserves the right to revise this document and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes. All series 140 instruments from LumaSense Technologies have a warranty of two years from the invoice date. This warranty covers manufacturing defects and faults which arise during operation only if they are the result of defects caused by LumaSense Technologies. The Windows compatible software was thoroughly tested on a wide range of Windows operating systems and in several world languages. Nevertheless, there is always a possibility that a Windows or PC configuration or some other unforeseen condition exists that would cause the software not to run smoothly. The manufacturer assumes no responsibility or liability and will not guarantee the performance of the software. Liability regarding any direct, or indirect damage caused by this software is excluded. Copyright All copyrights reserved. This document may not be copied or published, in part or completely, without the prior written permission of LumaSense Technologies GmbH. Contraventions are liable to prosecution and compensation. All rights reserved. 2 IMPAC pyrometer IP 140 Contents 1 Technical data.......................................................................................................................................... 5 1.1 Dimensions ................................................................................................................................... 6 2 Overview................................................................................................................................................... 6 2.1 Appropriate use............................................................................................................................. 7 2.2 Scope of delivery........................................................................................................................... 7 3 Safety........................................................................................................................................................ 7 3.1 General.......................................................................................................................................... 7 3.2 Laser targeting light....................................................................................................................... 7 3.3 Electrical connection ..................................................................................................................... 7 3.4 Ambient temperature..................................................................................................................... 7 4 Electrical Installation .............................................................................................................................. 8 4.1 Pin assignment of the male socket on the back of the pyrometer ................................................ 8 4.1.1 Connector pin J.............................................................................................................. 8 4.2 Connecting the pyrometer to a PC................................................................................................ 9 4.2.1 Connecting to RS232 interface...................................................................................... 9 4.2.2 Connecting to RS485 interface...................................................................................... 9 4.3 Connection of additional analyzing devices ................................................................................ 10 5 Mechanical Installation ......................................................................................................................... 10 5.1 Accessories (optional)................................................................................................................. 10 6 Sighting .................................................................................................................................................. 11 6.1 Thru-lens view finder................................................................................................................... 11 6.2 Laser targeting light..................................................................................................................... 11 7 Optics ..................................................................................................................................................... 12 7.1 Adjusting the required measuring distance................................................................................. 12 8 Instrument settings ............................................................................................................................... 13 8.1 Settings at the instrument ........................................................................................................... 13 8.2 Key panel operation .................................................................................................................... 13 8.3 Selection of the serial interface................................................................................................... 13 8.4 Test function................................................................................................................................ 14 8.5 Factory settings........................................................................................................................... 14 9 Parameter descriptions / settings........................................................................................................ 14 9.1 Emissivity (Emi)........................................................................................................................... 14 9.2 Compensation of ambient temperature (ta) ................................................................................ 14 9.3 Exposure time (t90)..................................................................................................................... 15 9.4 Clear time of the maximum value storage (tClear) ..................................................................... 15 9.5 Analog output (mA) ..................................................................................................................... 16 9.6 Subrange (from / to) .................................................................................................................... 16 9.7 Address (Adr) .............................................................................................................................. 16 9.8 Baud rate (Baud)......................................................................................................................... 16 9.9 Temperature display (C / F) ........................................................................................................ 16 9.10 Wait time (tw) .............................................................................................................................. 16 9.11 Maximum internal temperature (MaxIntTemp)............................................................................ 17 9.12 Error status (Status) .................................................................................................................... 17 3 IMPAC pyrometer IP 140 10 Settings via interface and software ..................................................................................................... 17 10.1 Connecting the pyrometer to a PC.............................................................................................. 17 10.2 Installation ................................................................................................................................... 17 10.3 Program start .............................................................................................................................. 17 10.4 The start menu ............................................................................................................................ 17 10.5 Beginning .................................................................................................................................... 18 10.6 Number of devices ...................................................................................................................... 18 10.7 Basic settings .............................................................................................................................. 18 10.8 Measurement color bar ............................................................................................................... 19 10.9 Measurement online trend .......................................................................................................... 19 10.10 Listing (analyzing) ....................................................................................................................... 20 10.11 Output .TXT file (analyzing) ........................................................................................................ 20 10.12 Trend output (analyzing) ............................................................................................................. 20 10.13 PC sampling rate (time interval between two measurements) ................................................... 21 10.14 Spot size calculator ..................................................................................................................... 21 11 Transport, packaging, storage............................................................................................................. 21 12 Maintenance........................................................................................................................................... 21 12.1 Safety .......................................................................................................................................... 21 12.2 Service ........................................................................................................................................ 21 12.3 Optics replacement ..................................................................................................................... 21 13 Trouble shooting ................................................................................................................................... 22 14 Data format UPP® (Universal Pyrometer Protocol) ............................................................................ 23 15 Reference numbers ............................................................................................................................... 25 15.1 Reference numbers instruments................................................................................................. 25 15.2 Reference numbers accessories: ............................................................................................... 25 Index................................................................................................................................................................ 26 4 IMPAC pyrometer IP 140 Legend Note: The note symbol indicates tips and useful information in this manual. All notes should be read with regard to an effective operation of the instrument. Security note laser beam Indicates to the danger of a built-in laser targeting light. Attention: This sign indicates special information which is necessary for a correct temperature measurement Shortcut for Temperature range (in German: Messbereich) MB Terminology The used terminology corresponds to the VDI- / VDE-directives 3511, page 4. Disposal / decommissioning Inoperable IMPAC pyrometers have to be disposed corresponding to the local regulations of electro or electronic material. 1 Technical data Temperature ranges: Signal processing: 50 to 400°C (MB 4) 160 to 1200°C 75 to 550°C (MB 5.5) 200 to 1300°C 100 to 700°C (MB 7) any range adjustable within the temperature range minimum span 51°C photoelectric current, digitized Spectral range: IR detector: 2 to 2.8 µm PbS Power supply: Power consumption: Analog output: 24 V AC or DC (14 to 30 V AC or DC) (AC: 48 to 62 Hz) Max. 6 W 0 to 20 mA or 4 to 20 mA (linear), switchable; Test current 10 mA or 12 mA by pressing test key 0 to 500 Ω RS232 or RS485 addressable (half duplex), switchable; Baud rate 2400 up to 115200 Bd 0.1°C on interface and display; < 0.03% of temperature range at the analog output power supply, analog output and digital interface are galvanically isolated from each other green LED Illuminated LC display for temperature indication or parameter settings Sub range: Load: Digital Interface: Resolution: Isolation: Operation indication: LC display: Parameter: Emissivity: Exposure time t90: Maximum value storage: Switch contact: (MB 12) (MB 13) Adjustable at the device or via interface: Emissivity ε, exposure time t90, 0 to 20 or 4 to 20 mA switch for analog output, sub range, clear times for maximum value storage, automatically or external deletion of maximum value storage, address, baud rate, wait time tW Readable at the device or via interface: Measuring temperature, internal instrument temperature. 10 to 100% adjustable in the instrument or via interface in steps of 0.1% 1.5 ms (with dynamical adaptation at low signal levels); adjustable at 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s Built-in single or double storage. Clearing with adjusted time tclear (off; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the next measuring object max. 0.15 A (only active with automatic clear mode or tclear ≥ 0.25 s) 5 IMPAC pyrometer IP 140 Uncertainty: (with ε=1, t90=1 s, Tamb.=10 to 40°C) up to 400°C: above 400°C: 2°C 0.3% of measured value in °C + 1°C (Note: the pyrometer must be operate at least 30 min before these values are valid) Repeatability: Ambient temperature: Storage temperature: Protection class: Weight: 0.1% of measured value in °C + 1°C 0 to 53°C - 20 to 60°C IP65 (DIN 40050) approx. 550 g CE-label: Sighting: According to EU directives about electromagnetic immunity Laser targeting light (max. power level < 1 mW, λ = 630-680 nm, CDRH class II) or built-in optimized thru-lens view finder 1.1 (ε=1, t90=1 s, Tamb.=10 to 40°C) Dimensions Pyrometer with thru lens view finder: Pyrometer with laser targeting light: All dimensions in mm 2 Overview 7 8 1 9 2 10 3 4 1 2 3 4 5 6 6 5 Focusable optics (with label with optical data) 4 threads for fixing the pyrometer or accessory parts Mounting rail Type label LC display Extendable back cover 11 6 7 8 12 Screws for rear cover (3 mm allen screws) Laser targeting light on/off switch or parallax free view finder, dependent on the instrument’s type) 9 Operating status for targeting light / laser (on instruments with laser targeting light) 10 Male socket for electrical connections 11 LC display, extended 12 Setting keys IMPAC pyrometer IP 140 Caution: The radiation temperature in the spot area of the pyrometer is not allowed to be higher than 150°C above the end of the basic range because the temperature sensor can be damaged. 2.1 Appropriate use The IP 140 is a digital pyrometer especially designed for non-contact temperature measurement of metals or ceramics. 2.2 Scope of delivery Device with thru-lens sighting or laser targeting light, selectable optics, PC software „InfraWin”, allen key 3 mm, operation manual. Note: 3 A connection cable is not included with the instrument and has to be ordered separately (see section 15, Reference numbers). Safety This section offers an overview about important safety aspects. Additionally in the several sections there are concrete safety aspects to avert danger. These aspects are indicated with symbols. Labels and markings at the instrument have to be noticed and keep in a permanent readable condition. 3.1 General Each person working with the pyrometer must have read and understood the user manual before operation. Also this has to be done if the person already used similar instruments or was already trained by the manufacturer. The pyrometer has only to be used for the purpose described in the manual. It is recommended to use only accessories offered by the manufacturer. 3.2 Laser targeting light For easy alignment to the measuring object the pyrometers can be equipped with a laser targeting light. This is a visible red light with a wavelength between 630 and 680 nm and a maximum power of 1 mW. The laser is classified as product of laser class II. Warning: To reduce the risk of injury to the eyes, do not look directly into the targeting laser and do not point the targeting laser into anyone's eyes. The instrument is equipped with a class II laser that emits radiation. Safety regulations: • Never look directly into the laser beam. The beam and spot can be watched safely from side. • Make sure that the beam will not be reflected into eyes of people by mirrors or shiny surfaces. 3.3 Electrical connection Follow common safety regulations for mains voltage (230 or 115 V AC) connecting additional devices operating with this mains voltage (e.g. transformers). Touching mains voltage can be mortal. A non expert connection and mounting can cause serious health or material damages. Only qualified specialists are allowed to connect such devices to the mains voltage. 3.4 Ambient temperature The pyrometer is designed for ambient temperatures of 0 to 53°C with non-condensing conditions. An operation out of these conditions can damage or malfunction the pyrometer. 7 IMPAC pyrometer IP 140 4 Electrical Installation The IP 140 is powered by a voltage of 24 V (possible range 14 to 30 V) DC or AC (48 to 62 Hz). The instrument needs some time to warm up. This time depends of the ambient temperature and connected power supply. During this time the display indicates “wait”. The analog output is 0 mA. After the warm up time the display shows the measuring temperature. For switching off the instrument, interrupt the power supply Green control light or unplug the electrical connector. Directly after connecting the power supply the display shows Internal Firmthe internal firmware version for approximately 1 s. The ware version for approx. 1 s green control light on the rear cover is switched on permanently. This control light is blinking as long as the laser tar- “wait“ indication geting light is switched on (only for instruments equipped during warm-up period with a targeting light). Note: At object temperatures below 150°C and emissivity smaller than 90% the pyrometer needs another half an hour until the final accuracy is reached. Otherwise a sufficient accuracy is reached after approx. 10 min. To meet the electromagnetic requirements (EMV), a shielded connecting cable must be used. The shield of the connecting cable has to be connected only on the pyrometer’s side. On side of the power supply (switch board) the shield must be open to avoid ground loops. LumaSense offers connecting cables, they are not part of standard scope of delivery. The connecting cable has wires for power supply, interface, analog output, external laser switch and external clear of maximum value storage via contact (see section 15, Reference numbers) and 12 pin connector. The cable includes a short RS232 adapter cable with a 9 pin SUB-D connector for direct PC communication. This adapter is not used in combination with RS485 interface. 4.1 Pin assignment of the male socket on the back of the pyrometer Pin K A L B H J G F C D E M 4.1.1 Color white brown green yellow gray Indication + 24 V power supply (or 24 V AC) 0 V power supply + Ioutp. analog output – Ioutp. analog output external switch for targeting light (bridge to K) see 4.1.1: operating mode contact or external clearing of pink maximum value storage or hold function red DGND (Ground for interface) black RxD (RS232) or B1 (RS485) violet TxD (RS232) or A1 (RS485) gray/pink B2 (RS485) (bridge to F) red/blue A2 (RS485) (bridge to C) Screen only for cable extension, orange don’t connect at the switchboard Male socket E F C G M D L B H J A K Pin-assignment (side of male inserts) Connector pin J The connector pin J can be used for 3 different functions: 1) Operating mode contact: During the warm-up of the pyrometer (after connection to the power supply, the LED display on the converter indicates “wait”) pin J is connected to the power supply voltage. This voltage is connected by a relays switch (max. 0.2 A at 50 V) to pin K (power supply voltage). This relays switch opens when the warm-up is finished and the pyrometer is in operating mode. For that reason this pin can be used as operating mode contact. 8 IMPAC pyrometer IP 140 2) External clearing of the maximum value storage: When the pyrometer is in operating mode, pin J can be used for external clearing of maximum value storage (see 9.4). To clear the maximum value storage, connect pin J for a short time to pin K (power supply voltage). The function “external clearing” is activated with the following conditions: • Warm-up is finished (no “wait“ indication on the display) • The clear time is set to “extern“ (see 9.4 and 10.7). 3) Hold function: when the hold function mode is activated the current temperature reading is frozen as long as J and pin K are connected (see 9.4 clear time for the maximum value storage). 4.2 Connecting the pyrometer to a PC The pyrometers are equipped with a serial interface RS232 or RS485 (switchable at the pyrometer). Standard on a PC is the RS232 interface. At this interface one pyrometer can be connected if the interface is set to RS232. Only short distances can be transmitted with RS232 and electromagnetic interferences can affect the transmission. With RS485 the transmission is to a large extend free of problems, long transmission distances can be realized and several pyrometers can be connected in a bus system. If RS485 is not available at the PC, it can be realized with an external converter which converts the RS485 in RS232 for a standard connection to a PC. When using a converter RS485 Ù RS232 take care, that the converter is fast enough to receive the pyrometer’s answer to an instruction of the master. Most of the commonly used converters are too slow for fast measuring equipment. So it is recommended to use the LumaSense-converter Ι-7520 (order no. 3 852 430). With a slow RS485 connection it is also possible to set a wait time at the pyrometer which delay the response of a command to the pyrometer (see 9.10 Wait time tw). Connecting to RS232 interface The transmission rate (in baud) of the serial interface is dependent on the length of the cable. Values between 2400 and 115200 Bd may be set. TxD (violet) RxD (black) DGND (red) E The baud rate has to be reduced by 50% when the transmission distance is doubled (see also 9.8 Baud rate) F M D C L A 1 K 2 6 3 7 4 5 8 9 PC’s side (soldering side female cable connector) Pyrometer’s side (soldering side female cable connector) Connecting to RS485 interface C L K Pyrometer 1 e.g. address 00 C L A K Pyrometer 2 e.g. address 01 F C DGND B1 E G M D H J A2 G M B B2 F A1 B1 E D H J A B2 G M B A1 B1 F D A2 E DGND A2 Terminator 120 Ohm B2 Master Half-duplex mode: A1 and A2 as well as B1 and B2 are A bridged in the 12-pin round connector of the connecting cable, to prevent B reflections due to long stubs. It also S safeguards against the interruption of the RS485 bus system should a connecting plug be pulled out. The master labels mark the connections on the RS485 converter. The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. Values between 2400 and 115200 Bd may be set. The baud rate is reduced by 50% when the transmission distance is doubled (see 9.8 Baud rate). Typical cable length for 19200 Bd is 2 km. A1 4.2.2 H J B Typical cable length for RS232 at 19200 Bd is 7 m. cable G DGND 4.2.1 L B H J A K Pyrometer 32 e.g. address 31 9 IMPAC pyrometer IP 140 4.3 Connection of additional analyzing devices Additional analyzing instruments, for example a LED digital display instrument only needs to be connected to a power supply and the analog outputs from the pyrometer. Another Instruments like a controller or printer can be connected to the display in series as shown above (total load of resistance max. 500 Ohm). white 230V ~ 24 V DC Power supply brown green °C LED digital display Controller Writer yellow 5 Mechanical Installation For mounting the pyrometer is equipped with a mounting rail on the bottom. This rail allows easy fixing of an adjustable mounting angle or a ball and socket mounting. Another possibility for fixing the pyrometer are 4 thread holes M 5 on the front of the instrument. 5.1 Accessories (optional) Numerous accessories guarantee easy installation of the pyrometer. The following overview shows a selection of suitable accessories. You can find the entire accessory program with all reference numbers on section 15, Reference numbers. Mounting: For mounting and aligning the pyrometer to the measured object a mounting angle or a ball and socket mounting is available. The ball and socket mounting is an easy way to align the pyrometer to the measured object. The quick-clampingscrews of the ball and socket mounting enable an easy and fast adjustment of the pyrometer in all directions. Mounting angle Ball and socket mounting Cooling: The pyrometer can be used in ambient temperatures outside of the specifications if preventive maintenance is taken. The cooling plate is used to protect the pyrometer from heat coming from the front. The completely covered water cooling jacket made from stainless steel protects the pyrometer if exposed to a hot environment. It is designed for ambient temperatures up to 180°C. Cooling plate Water cooling jacket Displays: Additionally to the built-in temperature indicator of the pyrometer LumaSense offers several digital displays which can also be used for remote parametrizing of the pyrometer. Digital display DA 6000 10 LED large display IMPAC pyrometer IP 140 Miscellaneous: The air purge protects the lens from contamination with dust and moisture. It has to be supplied with dry and oilfree pressurized air and generates an air stream shaped like a cone. The scanning attachment SCA 140 moves the measuring beam of the pyrometer from 0 to 12°. This angle is adjustable to smaller values. The scanning frequency is as well adjustable from 1 to 5 Hz. In most cases the scanning attachment SCA 140 is used as a peak picker for measuring smaller objects like thin wires which may be moving. Air purge Scanning attachment SCA 140 The 90°-mirror enables the capture of objects at an angle of 90° to the pyrometer axis. 90° mirror The emissivity enhancer can be used for objects with extremely low emissivity values e.g. objects with polished metallic surfaces. Emissivity enhancer 6 Sighting For exact aiming to the object the pyrometers are equipped with a thru-lens view finder or with an laser targeting light. 6.1 Thru-lens view finder The view finder can be used to align the measured object through direct observation. The view finder is true-sided and parallax-free; a circle marks the position of the measuring area, but not the exact size. 6.2 Thru-lens view finder Laser targeting light The laser targeting light is a red laser beam used to align the pyrometer at a target. The laser marks the center of the measuring spot. The laser targeting light can be used during operation without effecting the measurement. When the laser targeting light is switched on, the green control light on the rear cover is blinking and the display shows “PILT”. Laser targeting light push button (on/off) PILT display Blinking control light Warning: To reduce the risk of injury to the eyes, do not look directly into the targeting laser and do not point the targeting laser into anyone's eyes. The instrument is equipped with a class II laser that emits radiation. 11 IMPAC pyrometer IP 140 7 Optics According to requirements the instrument will be delivered with one of the below mentioned focusable optics. This allows the adjustment to the needed measuring distance to offer the smallest possible spot sizes. Optics 3-P Optics Optics Optics 2-P 1-P 0-P Meas. distance a [mm] a = 70 mm a = 78 mm a = 90 mm a = 105 mm a = 120 mm a = 150 mm a = 200 mm a = 260 mm a = 440 mm a = 345 mm a = 580 mm a = 1000 mm a = 4300 mm Aperture D [mm] *) MB 4 1.7 2.0 2.5 2.3 2.8 4.0 4.1 5.6 11.2 6.7 13 23 105 Spot size M90 [mm] MB 5.5 MB 7 MB 12 0.9 0.7 0.4 0.9 0.7 0.4 1.0 0.8 0.4 1.0 0.8 0.4 1.2 1.0 0.5 1.6 1.3 0.6 1.8 1.4 0.7 2.4 1.8 0.8 4.3 3.3 1.3 2.7 2.0 0.9 5 3.9 1.6 9 6.8 2.6 41 31 11 14 to 17 12 to 14 Objective length S [mm] MB 13 0.3 0.3 0.3 0.3 0.4 0.5 0.5 0.7 1.0 0.7 1.5 2.4 10 8 to 10 26 13 0 26 13 0 26 13 0 26 13 6.5 0 ) * The aperture is the effective lens diameter. It is depending on the objective length. The biggest aperture value belongs to the fully extended objective (S = 26), the smallest aperture value for objective length S = 0. Intermediate values have to be interpolated. The spot sizes, mentioned in the table above, will be only achieved at the adjusted distance. Decreasing or increasing the measuring a1 distance enlarges the spot a a2 size. Spot sizes for intermediate distances, that are Ø M1 Ø M2 ØM not shown on the optical profiles, may be calculated a2 a M1 = 1 (M + D) − D M2 = (M − D) + D using the following formula: Aperture D a a Note: Note: 7.1 The InfraWin program includes a Spot size calculator that roughly estimates the unknown values. The measuring object has to be bigger or at least as big as the spot size of the pyrometer. Adjusting the required measuring distance The required measuring distance must be adjusted to achieve the spot sizes mentioned in the tables above. This can be done between the smallest and the biggest limit value. For releasing the optics has to be turned anticlockwise. Then it can be pushed or pulled to find the correct measuring distance. For fixing the optics has to be turned clockwise. Adjusting the measuring distance with help of the table: The table mentions the minimum and maximum measuring distance for each optics. This corresponds to the longest or the shortest objective length. As an example a further value between max. and min. is shown. The objective length “S” can be measured with a caliper. fix push / pull release Objective length S The following methods can be used for measuring distances: Adjusting the measuring distance with help of the thru-lens view finder: The focusable optics is correctly adjusted to the required distance, if the measuring object is shown as a sharp image in the view finder. A circle marks the position of the measuring spot. 12 IMPAC pyrometer IP 140 Adjusting the measuring distance with help of the laser targeting light: On the focused measuring distance the laser has its smallest spot size and is illustrated exactly. 8 Instrument settings The IP 140 is equipped with a wide range of settings for optimal adaption to the required measuring condition and for getting the correct measuring temperature (description of all available parameters see section 9, Parameter descriptions / settings). All instrument settings can be done directly at the instrument or via serial interface and software InfraWin (see section 10, Settings via interface and software, user of an own communication software find all interface commands on section 14, Data format UPP®) 8.1 Settings at the instrument The LC-display as well as the push buttons for displaying and setting of the parameters are found inside the unit. The pyrometer is opened by 4 allen screws. If unscrewed, the rear cover can be pulled out along with the attached display and push buttons. The pullout is limited by the lengths of the screws. The backlight of the display is always powered in either status, opened or closed pyrometer. Note: Screw connection 3 mm allen screw Illuminated Display Setting keys Please make sure that the pyrometer is not contaminated while open. Max. pull-out 8.2 1 2 3 8.3 Rear cover Key panel operation PAR: With the PAR button all available parameters Temperature disare displayed in the following description play or parameter short form (section 10). Pushing the button again changes the display to the next parameter Push button and on the display a corresponding short short description form is displayed (see section 9, in brackets behind the parameter names). : With the arrow keys and all parameter 1 2 3 settings can be displayed. Pushing the button longer changes the settings in fast mode. ESC / ENT: Pushing the ESC button changes the pyrometer to measuring mode. If a parameter is changed with the arrow keys the indication of the ESC button changes to ENT. Pressing the button again confirms the value into the pyrometer. Changing the parameters again by pushing the PAR button doesn’t confirm this value in the pyrometer. If no button is pressed for 30 s the pyrometer changes to the temperature indication without accepting the changed value. Selection of the serial interface Opposite to the display there is a switch to select a serial interface RS232 or RS485. The LC display shows as chosen either RS232 or RS485. Interface switch Display of the selected interface Test button 13 IMPAC pyrometer IP 140 8.4 Test function The diagnostic push button „test“ generates a current on function active the analog output which is used to check if a connected Test (here the analog output external indicator shows the correct temperature value. is set to 0 -20 mA) The test current output is centered to the chosen analog output span, consequently 10 mA is supplied if the analog Display which also has output is adjusted to 0 to 20 mA and 12 mA is supplied if to be on an external the analog output span is set from 4 to 20 mA. The LC indication instrument display indicates the respective current along with the (here temp. range corresponding temperature. For example if a measuring 75 to 550°C) range of 75°C to 550 °C is selected the temperature shown in the display is 312.5°C). This temperature must be reflected exactly by the indicator which is supplied by the respective current. If this is not the case the selected analog input current span of the indicator is not equivalent to the chosen current output span of the pyrometer and one of the current spans or temperature range have to be modified. By pressing the „test“ push button once again or by pressing any push button of the LC-display the test current is switched off. Also after 30 seconds idle time the “test” current is switched off. The unit will be in the measurement mode again. 8.5 Factory settings Emissivity (Emi) = 100% Exposure time (t90) = min Clear time (tClear) = off Analog output (mA) = 0 ... 20 mA Sub range (from / to) same as temperature range 9 Parameter descriptions / settings 9.1 Emissivity (Emi) Address (Adr) = 00 Baud rate (Baud) = 19200 Bd Temperature display (C / F) = °C Wait time (tw) (for RS485) = 10 Interface (RS485 / RS232) = RS232 For a correct measurement it is necessary to adjust the emissivity. This emissivity is the relationship between the emission of an real object and the emission of a Display blackbody radiation source (this is an object which absorbs all incoming rays and of the has an emissivity of 100%) at the same temperature. Different materials have dif- adjusted ferent emissivities ranging between 0% and 100% (settings at the pyrometer be- emissivity tween 10 and 100%, the set value is indicated on the display). Additionally the emissivity is depending on the surface condition of the material, the spectral range of the pyrometer and the measuring temperature. The emissivity setting of the pyrometer has to be adjusted accordingly. Typical emissivity values of various common materials for the two spectral ranges of the instruments are listed below. The tolerance of the emissivity values for each material is mainly dependent on the surface conditions. Rough surfaces have higher emissivities. Measuring object “Black body furnace“ Extruded Aluminum Brass Brass oxidized (tarnished) Copper Copper, oxidized Inconel Inconel, oxidized Oxidized Iron Steel rolling scale 9.2 Emissivity [%] (at 2.4 µm) 100 13 18 65 to 70 5 70 to 80 30 85 85 to 90 80 to 88 Measuring object Steel, molten Nickel Titanium, non-oxidized Titanium, oxidized Molybdenum Molybdenum, oxidized Black Carbon Graphite Stoneware, glazed Porcelain rough Emissivity [%] (at 2.4 µm) 20 to 30 15 to 20 50 75 to 80 10 75 to 80 95 80 to 90 60 80 to 90 Compensation of ambient temperature (ta) The compensation of the ambient temperature can be used at low measuring temperatures (below 100°C). This compensation is used for a very few special applications only. The standard setting of this parameter is “auto”, because the temperature of the air around the pyrometer is normally the ambient temperature of the 14 IMPAC pyrometer IP 140 measured object. Should the measured object be placed in an area with a higher wall temperature (e.g. inside a furnace), the measurement might be falsified (probably too high temperature indication). This influence can be compensated by presetting of the ambient temperature of the object (presetting within the measuring range of the instrument). Settings: auto 00°C (32°F) . .. 53°C (127°F) It has to be considered, that this method only improves the results if the ambient temperature at the place of the measured object is always constant and the emissivity is well known. 9.3 Exposure time (t90) The exposure time is the time interval when the measured temperature has to be preSettings: sent after an abrupt change so that the output value of the pyrometer reaches a given min measurement value. The time taken is to reach 90 % of the recorded temperature dif0.01 s ference. In the “min” position, the device operates using its time constant. 0.05 s .. . The dynamic exposure time adjustment prolongs the exposure time at the lower range 10.00 s limit, also it t90 is set to a lower value. The following table shows the dependence of the exposure time from the measured temperature. Switch points on increasing temperatures: Temperature range 50 to 75 to 100 to 160 to 200 to 400°C 550°C 700°C 1200°C 1300°C (MB 4) (MB 5.5) (MB 7) (MB 12) (MB 13) switch points for ε = 1 35 ms → 5 ms 5 ms → 1.5 ms 102°C 129°C 137°C 180°C 172°C 224°C 255°C 329°C 272°C 352°C switch points for ε = 0,5 35 ms → 5 ms 5 ms → 1.5 ms 114°C 147°C 157°C 205°C 196°C 254°C 291°C 373°C 310°C 400°C If the maximum value storage is selected and the measuring object has a low emissivity and the measuring temperature is at the beginning of the range, a higher exposure time must be chosen to prevent measurement errors. Longer exposure times can be used for the measurement of objects which have rapidly fluctuating temperatures to achieve constant temperature reading. 9.4 Clear time of the maximum value storage (tClear) If the maximum value storage is switched on always the highest last temperature value will be displayed and stored. The storage has to be cleared at regular intervals for exchanging by a new and actual value. This feature is particularly useful when fluctuating object temperatures cause the display or the analog outputs to change too rapidly, or the pyrometer is not constantly viewing an object to be measured. In addition, it may also be beneficial to periodically delete and reset the stored maximum values. Settings: off 0.01 s . .. 25 s extern auto Hold The following settings are possible: off: The max. value storage is switched off and only momentary values are measured. 0.01...25 s: If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and held in double storage mode. After the entered time the storage will be deleted. extern: The external clearing can be activated and used within an own software (see section 14, Data format UPP®) or via an external contact (for connection see 4.1 Pin assignment of the male socket on the back of the pyrometer). In this case, the storage operates only in single storage, because only a single deletion mechanism is used. auto: The “auto” mode is used for discontinuous measuring tasks. For example objects are transported on a conveyer belt and pass the measuring beam of the pyrometer only for a few seconds. Here the maximum value for each object has to be indicated. In this mode the maximum value is stored until a new hot object appears in the measuring beam. The temperature which has to be recognized as “hot“ is defined by the low limit of the adjusted sub range. The stored maximum value will be deleted when the temperature of the new hot object exceeds the low limit “from“ of the sub range by 1% or at least 2°C. If a lower limit is not entered, the maximum value storage will be deleted whenever the lower level of the full measuring range has been exceeded. Hold: The function “hold” enables to freeze the current temperature reading at any moment. For this an external push button or switch has to be connected (see 4.1.1 connector pin J) which holds the temperature reading as long as the contacts are closed. 15 IMPAC pyrometer IP 140 Operation note: dependent on the settings the maximum value storage either works in single storage mode or in double storage mode: Single storage: the single storage is used when you want to reset the stored value using an external impulse via one contact closure from an external relay (i.e. between two measured objects). The relay contact is connected directly to the pyrometer between pins J and K. This mode allows a new value to be established, after each impulse from the reset signal. double storage: when entering the reset intervals via push buttons or PC interface the double storage is automatically selected. This mode utilizes two memories in which the highest measured value is held and is deleted alternately in the time interval set (clear time). The other memory retains the maximum value throughout the next time interval. The disadvantages of fluctuations in the display with the clock frequency are thereby eliminated. Note: 9.5 The maximum value storage follows the function of adjustment of exposure time. This results in: • clear time ≤ the adjusted response time is useless • clear times must be at least 3 times longer than the response time • only maxima with full maximum value can be recorded, which appear at least 3 times longer than the response time. Analog output (mA) The analog output has to be selected according to the signal input of the connected instrument (controller, PLC, etc.). 9.6 Settings: 0 ... 20 mA 4 ... 20 mA Subrange (from / to) You have the opportunity to choose a subrange (minimum 51°C) within the basic measuring range of the pyrometer. This subrange corresponds to the analog output. “from“ describes the beginning of this measuring range, “to“ the end of the range. Additionally with the setting of a subrange it is possible to fulfill the requirements of the “auto” clear mode of the maximum value storage (see above). 9.7 Address (Adr) For the connecting of several pyrometers with RS485 with one serial interface it is necSettings: essary to give each instrument an individual address for communication. First it is nec00 .. . essary to connect each single instrument to give it an address. After that all instruments 97 can be connected and addressed individually. If parameters may be changed simultaneously on all pyrometers, the global Address 98 can be used. This allows you to program all pyrometers at the same time, regardless of the addresses that have already been assigned. If the address of a pyrometer is unknown, it is possible to communicate with it using the global Address 99 (connect only one pyrometer). 9.8 Baud rate (Baud) The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. A standard cable length with RS232 for 19200 Bd is 7 m, with RS485 2 km. The baud rate is reduced by 50% if the transmission distance is doubled. 9.9 Temperature display (C / F) The temperature can be displayed in °C or °F. 9.10 .. 115.2 kBd Settings: °C °F Wait time (tw) Using a pyrometer with RS485 it is possible that the connection is not fast enough to receive the pyrometer’s answer to an instruction of the master. In this case a wait time can be set to slow down the data transfer (e.g.: tw = 02 at a baud rate 9600 means a wait time of 2/9600 sec). 16 Settings: 2.4 .kBd Settings: 00 .Bit .. 99 Bit IMPAC pyrometer IP 140 9.11 Maximum internal temperature (MaxIntTemp) Shows the maximum internal temperature the device ever reached. 9.12 Error status (Status) In case of a device error the pyrometer displays a hex code which identifies this error to LumaSense service. The standard display at this point is “ok”. 10 Settings via interface and software The operating and analyzing software InfraWin is included in delivery of the pyrometer. With this software all pyrometer functions also can be used on the PC (except changing the interface or using the test current function). This section gives an overview about the functions of the software. Additionally there is a software description in the program’s help menu. The following descriptions refer to the program version 4.0. The latest version is available for free as download from the homepage www.lumasenseinc.com. 10.1 Connecting the pyrometer to a PC The program InfraWin can operate up to two devices. For two devices using the RS232 interface, two PC interfaces must be used. Two devices using RS485 may be operated simultaneously by the same interface, if two different addresses have been properly entered (see 9.7 Address). 10.2 Installation For installation select the setup program “setup.exe“ from the InfraWin-CD or from the downloaded and unpacked zip file from the internet and follow the installation instructions. 10.3 Program start After installation and the first program start a language must be chosen (German, English, French, Italian, Spanish. The language also can be changed in the program). On the start page the screen shows the following icons: 10.4 The start menu Opens a saved file Storage of measured values for further processing Online measurement with color bar display Online measurement with graphic display Setting of the parameters of the instrument Setting of interface, baud rate and pyrometer addresses (RS485) Time interval between two measurements Number of connected instruments (max. 2) Listing of measured or stored values in tabular form Processing of measured (stored) readings in graph form Processing of measured (stored) readings in a text file Calculation of spot sizes in various measuring distances Only if available: controls the programmable controller PI 6000 17 IMPAC pyrometer IP 140 10.5 Beginning Before using the software, the serial interface connected to the pyrometer has to be selected under the Computer icon. For two devices using the RS232 interface, two PC interfaces must be used. 10.6 Number of devices With a click on “number of devices” InfraWin changes to the display of 1 or 2 devices. If 2 devices are selected, always 2 windows are displayed for settings or evaluation. 10.7 Basic settings Under pyrometer parameters all preset values can be displayed and modified if necessary. The window pyrometer parameter contains all parameter settings described in 9, Parameter descriptions / settings. Choose the correct setting for your application, the actual setting is displayed. Notes: • “Basic range” displays the total range of the pyrometer automatically and can not be changed. • Under “Material“ you have the possibility to store the names of different measuring objects with their emissivity values and to recall them from the list. • Choose whether the temperature should be displayed in °C (Celsius) or °F (Fahrenheit). • If the pyrometer is equipped with a laser targeting light the laser icon ( ) is visible. A click on the laser targeting light icon turns the laser targeting light on or off at this point. After approx. two minutes the laser targeting light is switched off automatically. • After switching over to “man”, you can decide if the off-set compensation for the ambient temperature should be activated (also see 9.2 Compensation of ambient temperature). The corresponding data field T(amb) is then activated so that the ambient temperature value can be entered. In the “auto“ mode position, the device assumes that the ambient temperature of the measured object is equal to that of the pyrometer. The open / save button enable to store and recall own pyrometer configurations. „1 meas.“ shows the current measuring temperature in the pyrometer parameters window for approx. 1 second. A click on the “Test” icon opens a window that allows the direct communication with the pyrometer via the interface commands (see section 14, Data format UPP®). After entering an interface command (00 is the adjusted address ex works, “ms“ is the command “reading temperature value“) and a click on “Send“ the following window is opened: This window already shows the answer of the pyrometer in 1/10°. The actual temperature reading is 325.7°C. “Len“ indicates the length of the answered data string, incl. Carriage Return (Chr(13)). In the lower part of the window the connection with the preset baud rate can be checked. Here the command was send 500 times with 19200 baud. It has taken 4.56 seconds without transmission errors. Note: 18 If the pyrometer parameters window displayed, changing of settings on the pyrometer is blocked. IMPAC pyrometer IP 140 10.8 Measurement color bar This window displays: • current temperature, graphically as color bar and numerically • temperature range or adjusted sub range • file size and quantity of the measured values of the current measurement • emissivity ε • the internal temperature of the instrument (Tint) • minimum (Tmin) and maximum values (Tmax) • temperature of the limit contacts The color bar display shows the span of the temperature range or the adjusted sub-range. Entering temperature values in the white fields on the left and right side of the color bar, limits for the color change of the color bar can be set. These limits can also be changed by moving the small bar with the PC mouse. The color bar displays temperatures within the two limits in green color, outside the limits in red color. The targeting light ( ) can be switched on or off at this point is the pyrometer is equipped with it. In addition, there is an input field ε for the emissivity in the window. If the emissivity is changed, the temperature change connected with this can be read off directly. Emi: AutoFind: If the true temperature of the measured object is known, you can calculate the emissivity of the measured object using the "Emi: AutoFind“ function: • A measured temperature is displayed with the current set emissivity (in this example 100%) (here: 824°C). • If you press “Emi: Autofind“ a window will open which allows you to enter the "true" temperature. • Once the temperature entry has been entered and confirmed with "OK", InfraWin will then calculate the emissivity which occurs with the new temperature. This is displayed immediately and can be used for further temperature measurement. 10.9 Measurement online trend This window displays: • temperature as graphical diagram • current temperature • quantity of the measured values and file size of the current measurement The example shows a sample reading over the period of approx. 10 seconds with a temperature range between 75 and 550°C. The final temperature (at the end of the reading) is 244.3°C. If the pyrometer is equipped with a laser targeting light ( ) this can be switched on or off. 19 IMPAC pyrometer IP 140 • • • With “Mark zone“ a temperature range can color marked for easier recognition. Setting a temperature under “Threshold“ prevents the recording of values above or below this temperature to keep the file size small. With “Scaling trend“ the view of the temperature range can be limited. Note: 10.10 The measuring values of “measurement color bar” or “measurement online trend” are automatically saved as "standard.i12". Should you need to edit the data later, you need to save the file as another .i12-file because old values are over-written when a new measurement is taken. Files from older program versions (.i10-files) can be opened and saved as .i12. Listing (analyzing) For analyzing the measured values in this field all measured data appears in a numeric list. The date beside the time gives more exactly values to see what happened on time units smaller 1 s. The value specifies the time in seconds after midnight (0:00 h). The amount of data depends on the frequency that readings were taken (settings at 10.13 PC sampling rates). As the amount of data increases, so does the amount of storage space required to save it. In order to save room, all .i12 data files are stored by a binary code. 10.11 Output .TXT file (analyzing) The same file as under „Output listing” may be converted into a text file and can be easily opened, for example with EXCEL. With the standard import settings EXCEL automatically formats the columns accordingly (tabulator as separators). 10.12 Trend output (analyzing) The graph’s curve depicts the temperature change over time within the specified temperature range. Additionally, other information appears in this window; such as recorded time (x-axis) and temperature in degrees (y-axis) as well as the time and temperature at the vertical cursor line which can be dragged with the mouse. Selecting the Trend output initially causes all the saved data to be displayed. If the data exceeds an amount that can be represented reasonably, you may “Zoom“ in on a partial segment using the mouse (such as the segment represented in the example). Under “Total” you can return to the representation of the entire curve. Note: 20 The last reading is saved in the standard.i12 file and automatically appears in this form upon opening Listing or Trend output. If file open was loaded using another file, the previous file will be overwritten and replaced by the standard.i12 file. IMPAC pyrometer IP 140 10.13 PC sampling rate (time interval between two measurements) This function sets a time interval. After each interval one measured value is stored on the PC. The bigger the time interval the smaller will be the stored file. This function is mainly used for long term measurements. 10.14 Spot size calculator The spot size calculator is only used for pyrometers with fixed optics. After input of the aperture and the main spot size, the input of interim values it calculates spot sizes in different measuring distances of the fixed optics. 11 Transport, packaging, storage With faulty shipping the instrument can be damaged or destroyed. To transport or store the instrument, please use the original box or a box padded with sufficient shock-absorbing material. For storage in humid areas or shipment overseas, the device should be placed in welded foil (ideally along with silika gel) to protect it from humidity. The pyrometer is designed for a storage temperature of -20 to 70°C with non-condensing conditions. A storing out of these conditions can damage or malfunction the pyrometer. 12 Maintenance 12.1 Safety Attention during pyrometer services: Should the pyrometer be integrated in a running machine process the machine should be switched off and secured against restart before servicing the pyrometer. 12.2 Service The pyrometer does not have any parts which require regular service, only the lens has to be kept clean. The lens can be cleaned with a soft cloth in combination with alcohol (do not use acid solutions or dilution). Also standard cloths for cleaning glasses or photo objectives can be used. 12.3 Optics replacement The pyrometers of series 140 are equipped with a focusable optics. This optics can be changed against another. For replacement optics for different measuring distances can be used without recalibration of the instrument. Replacement can be necessary if the lens is scratched or the pyrometer will be used for other measuring distances. Replacement: Only the lens will be replaced for changing the focusable optics. The fixing ring has to be removed with a suitable objective wrench. After removing the old lens, put in the new one with the convex side to the front. Fix the lens with a new fixing ring. On the inside of this ring is the sticker with the optics data. Lens Fixing ring Sticker with optics data For differentiation they are marked with a color mark at the border of the lens: Focusable optics 0-P (measuring distance 70 ... 90 mm) (without color mark) Focusable optics 1-P (measuring distance 100 ... 133 mm) (pink) Focusable optics 2-P (measuring distance 190 ... 340 mm) (black) Focusable optics 3-P (measuring distance 350 ... 2600 mm) (blue) 21 IMPAC pyrometer IP 140 13 Trouble shooting Before sending the pyrometer for repair, try to find the error and to solve the problem with the help of the following list. Temperature indication too low • Incorrect alignment of the pyrometer to the object ⇒ New correct alignment to achieve the max. temperature signal (see 6) • Measuring object smaller than spot size ⇒ check measuring distance, smallest spot size is at nominal measuring distance (see 7) • Measuring object is not always in the measuring spot of the pyrometer ⇒ Use max. value storage (see 9.4). • Emissivity set too high ⇒ Set lower correct emissivity corresponding to the material (see 9.1) • Lens contaminated ⇒ Clean lens carefully (see 12.2) Temperature indication too high • Emissivity set too low ⇒ Set lower correct emissivity corresponding to the material (see 9.1) • The measurement is influenced by reflections of hot machine parts ⇒ Use mechanical construction to avoid the influence of the interfering radiation Measuring errors • Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens ⇒ Clean lens. Recommendation: use of air purge (see 12.2, 5.1) • Indicated temperature is decreasing during the use of the pyrometer, although the air purge unit is used. Probably compressed air is not clean or air failed ⇒ Clean the lens and use clean, dry and oil free compressed air • Air contamination in the sighting path between pyrometer and object ⇒ Change position of the pyrometer with a clean sighting path (if necessary use a ratio pyrometer) • HF-interferences ⇒ Correct the connection of the cable shield (see 4) • Instrument overheated ⇒ Use cooling jacket with air or water cooling (see 5.1) • Temperature Indication is fluctuating, probably caused by changing emissivity ⇒ Wrong pyrometer type, use of ratio pyrometer recommended Laser targeting light • Laser targeting light fails ⇒ Instruments max. temperature is exceeded. Use cooling jacket (see 5.1) Note: 22 The wavelength band of the IP 140 reacts at low measuring temperatures to incandescent lamps or very bright daylight (not valid for fluorescent tube). For a correct measurement strong external light to the measured object should be avoided. IMPAC pyrometer IP 140 14 Data format UPP® (Universal Pyrometer Protocol) Via interface and a suitable communication software or via “Test” function of the InfraWin software (see 10.7 Basic settings Æ Test) commands can be exchanged directly with the pyrometer. The data exchange occurs in ASCII format with the following transmission parameters: The data format is: 8 data bits, 1 stop bit, even parity (8,1,e) The device responds to the entry of a command with: output (e.g. the measuring value) + CR (Carriage Return, ASCII 13), to pure entry commands with "ok" + CR. Every command starts with the 2-digit device address AA (e.g. "00"). This is followed by 2 small command letters (e.g. "em" for level of emissivity ε), finished with CR This is followed, if necessary for that command, by the ASCII parameter "X". If this parameter "X" is omitted, then the device resets with the current parameter. A „?“ after the small command letters answers with the respective settings (only at setting commands, not at enquiry commands). Example: Entry: “00em“ + <CR> The emissivity setting (ε) of the device with the address 00 is returned Answer: “0970“ + <CR> means Emissivity = 0.97 or 97.0% Description Reading temperature value: Command AAms Reading temperature value repeated: Emissivity: Ambient temperature compensation: AAmsXXX Exposure time t90: AAezX Clear times of the maximum value storage: AAlzX External clearing: Analog output: Reading basic temperature range: AAlx AAasX AAmb Reading temperature sub range: AAme Setting of temperature sub range: Address: AAm1XXXXYYYY AAemXXXX AAutXXXX AAgaXX Parameters Output: XXXXX (dec., in 1/10 °C or °F) (77770 = Instrument’s temperature too high or chopper wheel rotates with wrong speed 88880 = temperature overflow) XXX = 000...999 (XXX = number of measuring values) XXXX = (0010 ... 1000‰) (decimal) XXXX = value of ambient temperature, 4-digit, hex XXXX e.g. FFEC corresponds to -20 degrees - 99dez = FF9Dhex means: automatic, no manual compensation X = 0 ... 6 (decimal) 0 = intrinsic time constant of the device 1 = 0.01 s 4 = 1.00 s 2 = 0.05 s 5 = 3.00 s 3 = 0.25 s 6 = 10.00 s X = 0 ... 9 (dec.) 0 = Maximum value storage off 1 = 0.01 s 4 = 1.00 s 7 = external deletion 2 = 0.05 s 5 = 5.00 s 8 = automatically deletion 3 = 0.25 s 6 = 25.00 s 9 = Hold Clearing the max. value storage (only with lz = 7, external) X = 0 or 1 0 = 0 - 20 mA; 1 = 4 - 20 mA Output: XXXXYYYY (hex 8-digit, °C or °F) XXXX = beginning of temp. range YYYY = end of temp. range Output: XXXXYYYY (hex 8-digit, °C or °F) XXXX = beginning of temp. range YYYY = end of temp. range XXXX (hex 4-digit) beginning of temp. range (°C or °F) YYYY (hex 4-digit) end of temp. range (°C or °F) XX = (00 ... 97) 00 ... 97 = regular device addresses 99 = Global address with response 98 = Global address without response (only setting commands!) 23 IMPAC pyrometer IP 140 Baud rate: AAbrX Temp. display °C / °F Wait time: Internal temperature: Max. internal temperature: Error status: AAfhX AAtwXX AAgt AAtm Laser targeting light: AAlaX Reading interface: Lock keyboard: AAin AAlkX Reading parameters: AApa Device type: Serial number: Device type / software version: AAna AAsn AAve Detailed Software version: Reference number: AAvs AAfs AAbn X=1...6 or 8 (dec.) 1 = 2400 baud 5 = 38400 baud 2 = 4800 baud 6 = 57600 baud 3 = 9600 baud (7 is not available) 4 = 19200 baud 8 = 115200 baud Output: X = 0: display in °C; X = 1: display in °F XX = 00 ... 99 (decimal, in bit time of the current baud rate) Output: XXX (dec. 000 ... 099°C, 032 ... 210°F Output: XXX (dec. 000 ... 099, always in °C) Output: XX; XX=00…FF (00 = no error) (01…FF: error code for LumaSense service) X = 0 switch off laser X = 1 switch on laser Output: 1 or 2 (1 = RS232, 2 = RS485) X = 0 ... 3 1 = lock lk1, removal with command lk0 or power off-on 0 = removal of lock lk1 3 = continuous lock lk3, removal only with command lk2 2 = removal of lock lk3 Output 11-digit, decimal: Digit 1 und 2 (10...99 or 00): Emissivity Digit 3 (0 ... 6): Exposure time Digit 4 (0 ... 8): Clear time max. storage Digit 5 (0 ... 1): Analog output Digit 6 und 7: (00 ... 98): Internal temperature Digit 8 und 9 (00 ... 97): Address Digit 10 (0 ... 6 or 8): Baud rate Digit 11 (0,1, 3): Keyboard lock Output: “IP 140” (16 ASCII-characters) Output: XXXXX (dec. 5-digit) Output: XXYYZZ (6-digit decimal) XX = 63 (IP 140) YY = Month of software version ZZ = Year of software version tt.mm.yy XX.YY tt = day; mm = month; yy = year; XX.YY = software version Output: XXXXXX (hex 6-digit) Note: the letter “l” means the lower case letter of “L”. Additional instruction for the RS 485 interface: Requirements to the master system during half-duplex operation: 1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some older interfaces are not fast enough for this). 2. The pyrometer's response will follow after 5 ms at latest. 3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated. 24 IMPAC pyrometer IP 140 15 Reference numbers 15.1 Reference numbers instruments Temperature range MB 4: 50 ... 400°C MB 5.5: 75 ... 550°C MB 7: 100 ... 700°C MB 12: 160 ... 1200°C MB 13: 200 ... 1300°C With laser pilot light 3 875 500 3 875 520 3 875 540 3 875 560 3 875 580 With thru-lens view finder 3 875 510 3 875 530 3 875 550 3 875 570 3 875 590 Ordering note: When ordering please select one focusable optics (0, 1, 2 or 3) (see section 7, Optics). A connection cable is not included in scope of delivery and has o be ordered separately. 15.2 Reference numbers accessories: 3 848 210 3 848 250 3 848 260 3 848 270 Optics 0-P (70 ... 90 mm) Optics 1-P (100 ... 133 mm) Optics 2-P (190 ... 340 mm) Optics 3-P (350 ... 2600 mm) 3 820 340 3 820 530 3 820 540 3 820 830 3 820 840 3 820 550 3 820 330 3 820 500 3 820 510 3 820 810 3 820 820 3 820 520 3 820 740 3 820 750 Connection cable, length 5 m, 90° connector Connection cable, length 10 m, 90° connector Connection cable, length 15 m, 90° connector Connection cable, length 20 m, 90° connector Connection cable, length 25 m, 90° connector Connection cable, length 30 m, 90° connector Connection cable, length 5 m, straight connector Connection cable, length 10 m, straight connector Connection cable, length 15 m, straight connector Connection cable, length 20 m, straight connector Connection cable, length 25 m, straight connector Connection cable, length 30 m, straight connector Connection cable, length 5 m, straight connector, temperature resistant up to 200°C Connection cable, length 5 m, 90° connector, temperature resistant up to 200°C 3 834 280 3 834 270 3 835 230 3 837 290 3 835 060 3 834 140 3 837 240 3 835 450 3 843 520 Adjustable mounting angle Ball and socket mounting Air purge Cooling jacket, stainless steel Air purge for cooling jacket Heavy ball and socket mounting for cooling jacket Cooling plate 90° mirror with quartz glass window Scanner SCA 140, (scanning angle 0 ... 12°, scanning frequency 1 ... 5 Hz), with quartz glass window Air purge for scanner SCA 140 3 835 290 3 826 500 Power supply for DIN rail mounting NG DC (100 to 240 V AC ⇒ 24 V DC, 1 A) Power supply NG 2D, as NG 0D: additionally with 2 limit switches (not for US sale) LED digital display DA 4000-N LED digital display DA 4000: with 2 limit switches LED digital display DA 6000-N: with possibility for pyrometer parameter settings for digital IMPAC pyrometers; RS232 interface LED digital display DA 6000-N: with possibility for pyrometer parameter settings for digital IMPAC pyrometers; RS485 interface LED digital display DA 6000; DA 6000-N additional with 2 limit switches and analog input and output, RS232 interface LED digital display DA 6000; DA 6000-N additional with 2 limit switches and analog input and output, RS485 interface HT 6000, portable battery driven indicator and instrument for pyrometer parameter setting 3 826 510 PI 6000: PID programmable controller, very fast, for digital IMPAC pyrometers 3 825 430 Ι-7520, RS232 ⇔ RS485 converter 3 852 290 3 852 550 3 890 640 3 890 650 3 890 560 3 890 570 3 890 520 3 890 530 25 IMPAC pyrometer IP 140 Index A L Accessories (optional) ........................................10 Address...............................................................17 Ambient temperature ............................................7 Ambient temperature compensation...................15 Analog output......................................................17 Analyzing devices, additional .............................10 Appropriate use ....................................................7 Laser targeting light ............................................. 7 Listing (analyzing) .............................................. 21 B Basic settings......................................................19 Baud rate ............................................................17 C Clear time ...........................................................16 Color mark ..........................................................23 Connection cable ..................................................7 Converter ..............................................................9 D Data format UPP® ...............................................24 Dimensions ...........................................................6 E Electromagnetic requirements..............................8 Emi: AutoFind .....................................................20 Emissivity............................................................15 Error Status.........................................................18 Exposure time.....................................................16 M Maintenance ...................................................... 22 Maximum value storage....................................... 9 Measurement online trend ................................. 20 Measuring distance............................................ 12 O Online trend ....................................................... 20 Operating mode contact ...................................... 8 Optics................................................................. 12 Optics replacement ............................................ 22 P Parameter descriptions / settings ...................... 15 Pin assignment of the male socket ...................... 8 Pyrometer parameters ....................................... 19 R Reference numbers ........................................... 26 S Focusable optics.................................................12 Safety................................................................... 7 Scope of delivery ................................................. 7 Serial interface ................................................... 13 Shield ................................................................... 8 Sighting .............................................................. 11 Software............................................................. 18 Subrange ........................................................... 17 H T Hold function.........................................................9 Temperature display .......................................... 17 Temperature range ............................................ 20 Test .................................................................... 19 Test function ...................................................... 15 Thru-lens view finder.......................................... 11 Time interval ...................................................... 22 Transport, packaging, storage ........................... 22 Trend output (analyzing).................................... 21 Trouble shooting ................................................ 23 TXT file............................................................... 21 F I i12 .................................................................21, 22 InfraWin ..............................................................18 Installation, electrical ............................................8 Installation, mechanical ......................................10 Instrument settings .............................................13 Interface................................................................9 Interface commands ...........................................19 Interface settings ................................................18 Internal temperature ...........................................18 K Key panel operation............................................13 U UPP® data format............................................... 24 W Wait time ............................................................ 17 26 IMPAC-Pyrometer IP 140 Inhaltsverzeichnis Allgemeines ........................................................................................................................................... 29 Informationen zur Betriebsanleitung........................................................................................................ 29 Haftung und Gewährleistung................................................................................................................... 29 Symbolerklärung / Bezeichnungen.......................................................................................................... 29 Terminologie............................................................................................................................................ 29 Urheberschutz ......................................................................................................................................... 29 Entsorgung / Außerbetriebnahme ........................................................................................................... 29 1 Technische Daten.................................................................................................................................. 30 1.1 Abmessungen ............................................................................................................................. 31 2 Übersicht ................................................................................................................................................ 31 2.1 Bestimmungsgemäße Verwendung............................................................................................ 32 2.2 Lieferumfang ............................................................................................................................... 32 3 Sicherheit ............................................................................................................................................... 32 3.1 Allgemeines................................................................................................................................. 32 3.2 Laserpilotlicht .............................................................................................................................. 32 3.3 Elektrischer Anschluss ................................................................................................................ 32 4 Elektrische Installation ......................................................................................................................... 33 4.1 Pin-Belegung des Flanschsteckers auf der Rückseite des Pyrometers ..................................... 33 4.1.1 Stecker-Pin J ............................................................................................................... 33 4.2 Allgemeinhinweise zum Anschluss des Pyrometers an einen Rechner ..................................... 34 4.2.1 Anschluss an Schnittstelle RS232............................................................................... 34 4.2.2 Anschluss an Schnittstelle RS485............................................................................... 34 4.3 Anschluss zusätzlicher Auswertegeräte ..................................................................................... 35 5 Mechanische Installation...................................................................................................................... 35 5.1 Zubehör (optional)....................................................................................................................... 35 6 Visiereinrichtungen ............................................................................................................................... 36 6.1 Durchblickvisier ........................................................................................................................... 36 6.2 Laserpilotlicht .............................................................................................................................. 36 7 Optik ....................................................................................................................................................... 36 7.1 Einstellen auf den benötigten Messabstand ............................................................................... 37 8 Geräteeinstellungen .............................................................................................................................. 38 8.1 Einstellungen am Gerät............................................................................................................... 38 8.2 Anzeigen und Einstellen der Parameter ..................................................................................... 38 8.3 Wahl der seriellen Schnittstelle ................................................................................................... 38 8.4 Testfunktion................................................................................................................................. 39 8.5 Werkseinstellungen..................................................................................................................... 39 9 Parameterbeschreibung / Einstellungen............................................................................................. 39 9.1 Emissionsgrad ε (Emi) ................................................................................................................ 39 9.2 Umgebungstemperaturkompensation (ta) .................................................................................. 40 9.3 Erfassungszeit (t90) .................................................................................................................... 40 9.4 Löschzeit des Maximalwertspeichers (tClear) ............................................................................ 40 9.5 Analogausgang (mA) .................................................................................................................. 41 9.6 Teilmessbereich (from / to) ......................................................................................................... 41 9.7 Adresse (Adr) .............................................................................................................................. 41 9.8 Baudrate (kBaud) ........................................................................................................................ 42 9.9 Temperaturanzeige (C / F).......................................................................................................... 42 9.10 Wartezeit (tw) .............................................................................................................................. 42 9.11 Maximale Innentemperatur (MaxIntTemp).................................................................................. 42 9.12 Fehler-Status (Status) ................................................................................................................. 42 27 IMPAC-Pyrometer IP 140 10 Einstellungen über Schnittstelle / Software ....................................................................................... 42 10.1 Anschluss des Pyrometers an einen PC..................................................................................... 42 10.2 Installation ................................................................................................................................... 42 10.3 Programmstart ............................................................................................................................ 42 10.4 Das Menü .................................................................................................................................... 43 10.5 Vorbereitung................................................................................................................................ 43 10.6 Anzahl Pyrometer........................................................................................................................ 43 10.7 Grundeinstellungen ..................................................................................................................... 43 10.8 Messung (Farb-Balken)............................................................................................................... 44 10.9 Messung (Online-Grafik) ............................................................................................................. 45 10.10 Tabelle (Auswertung) .................................................................................................................. 45 10.11 Ausgabe .TXT-Datei (Auswertung) ............................................................................................. 46 10.12 Ausgabe Grafik (Auswertung) ..................................................................................................... 46 10.13 PC-Aufnahmerate (Zeitintervall zwischen zwei Messungen)...................................................... 46 10.14 Messfeld-Rechner ....................................................................................................................... 46 11 Transport, Verpackung, Lagerung....................................................................................................... 46 12 Wartung .................................................................................................................................................. 47 12.1 Sicherheit .................................................................................................................................... 47 12.2 Allgemeines................................................................................................................................. 47 12.3 Austausch der Optik.................................................................................................................... 47 13 Fehlerdiagnose ...................................................................................................................................... 47 14 Datenformat UPP® (Universelles Pyrometer-Protokoll) .................................................................... 48 15 Bestellnummern .................................................................................................................................... 50 15.1 Bestellnummern Geräte .............................................................................................................. 50 15.2 Bestellnummern Zubehör............................................................................................................ 50 Stichwortverzeichnis..................................................................................................................................... 51 28 IMPAC-Pyrometer IP 140 Allgemeines Informationen zur Betriebsanleitung Wir beglückwünschen Sie zum Kauf dieses hochwertigen und leistungsfähigen IMPAC-Pyrometers. Lesen Sie diese Betriebsanleitung mit allen Hinweisen zu Sicherheit, Bedienung und Wartung bitte sorgfältig Schritt für Schritt durch. Sie dient als wichtige Informationsquelle und Nachschlagewerk für Installation und Betrieb des Gerätes. Zur Vermeidung von Bedienungsfehlern muss diese Anleitung so aufbewahrt werden, dass jederzeit darauf zugegriffen werden kann. Die allgemeinen Sicherheitsbestimmungen (siehe Kap. 3, Sicherheit) müssen bei Betrieb des Gerätes unbedingt eingehalten werden. Neben dieser Betriebsanleitung gelten die Betriebsanleitungen der mitbenutzten Komponenten. Die darin enthaltenen Hinweise – insbesondere Sicherheitshinweise – sind zu beachten. Sollten weitergehende Fragen auftreten, steht Ihnen unser technischer Kundendienst unter der Rufnummer +49 (0)69 973 73-0 in D-60326 Frankfurt telefonisch gerne zur Verfügung. Haftung und Gewährleistung Alle Angaben und Hinweise für die Bedienung, Wartung und Reinigung dieses Gerätes erfolgen unter Berücksichtigung unserer bisherigen Erfahrung nach bestem Wissen. LumaSense Technologies übernimmt keine Haftung für die in diesem Handbuch aufgeführten Beispiele und Verfahren oder für Schäden, die daraus eventuell entstehen könnten oder für den Fall, dass der Inhalt dieses Dokuments möglicherweise unvollständig oder fehlerhaft ist. LumaSense Technologies behält sich das Recht vor, Änderungen an diesem Dokument und den darin beschriebenen Produkten vorzunehmen, ohne die Verpflichtung einzugehen, irgendeine Person über solche Änderungen zu informieren. LumaSense Technologies gibt auf die Pyrometer der Serie 140 eine Gewährleistung von zwei Jahren ab Datum der Lieferung. Diese bezieht sich auf Fabrikationsfehler sowie Fehler, die sich während des Betriebes einstellen und auf einen Fehler der Firma LumaSense Technologies hinweisen. Die Gewährleistung erlischt, wenn das Gerät ohne vorherige schriftliche Zustimmung von LumaSense zerlegt oder modifiziert wurde. Die Windows-Software wurde unter diversen Windows-Betriebssystemen in mehreren Sprachen nach bestem Wissen getestet. Es kann jedoch nicht grundsätzlich ausgeschlossen werden, dass es eine Konfiguration aus PC und Windows-Betriebssystem oder andere Umstände gibt, in denen sie nicht einwandfrei arbeitet. Auf den Einsatz der PC-Software können keine Haftungs- oder Gewährleistungsansprüche hergeleitet werden. Jede Haftung für direkte, indirekte, verursachte oder gefolgerte Schäden, die durch die Verwendung dieses Programms entstehen könnten, ist ausgeschlossen. Symbolerklärung / Bezeichnungen Hinweis: Das Hinweissymbol kennzeichnet Tipps und besondere nützliche Informationen dieser Betriebsanleitung. Alle Hinweise sollten im Interesse einer effektiven Bedienung des Gerätes beachtet werden. Sicherheitshinweis Laserstrahlung: Weist auf die Gefahren eines eingebauten Laserpilotlichts hin. Achtung: Das Achtung-Symbol kennzeichnet besondere Informationen, die für eine korrekte Temperaturmessung nötig sind MB Abkürzung für Messbereich Terminologie Die verwendete Terminologie bezieht sich auf die VDI- / VDE-Richtlinie 3511, Blatt 4. Urheberschutz Alle Unterlagen sind im Sinne des Urheberrechtgesetzes geschützt. Weitergabe sowie Vervielfältigung von Unterlagen, auch auszugsweise, Verwertung und Mitteilung ihres Inhaltes sind nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwiderhandlungen sind strafbar und verpflichten zu Schadenersatz. Alle Rechte der Ausübung von gewerblichen Schutzrechten behalten wir uns vor. Entsorgung / Außerbetriebnahme Nicht mehr funktionsfähige IMPAC-Pyrometer sind gemäß den örtlichen Bestimmungen für Elektro- / Elektronikmaterial zu entsorgen. 29 IMPAC-Pyrometer IP 140 1 Technische Daten Grundmessbereiche: Interne Messwertverarbeitung: 50 ... 400°C (MB 4) 160 ... 1200°C (MB 12) 75 ... 550°C (MB 5.5) 200 ... 1300°C (MB 13) 100 ... 700°C (MB 7) Beliebig innerhalb des Grundmessbereichs einstellbar mit Mindestmessbereichsumfang 51°C Wechsellichtsignal, wird digitalisiert Spektralbereich: IR-Detektor: 2 ... 2,8 µm PbS Spannungsversorgung: Leistungsaufnahme: Analogausgang: 24 V AC oder DC (14 ... 30 V AC oder DC) (AC: 48 ... 62 Hz) Max. 6 W 0 ... 20 mA oder 4 ... 20 mA (linear), umschaltbar; Teststrom 10 mA bzw. 12 mA auf Tastendruck 0 ... 500 Ω RS232 oder RS485 adressierbar (halbduplex), umschaltbar; Baudrate 2400 bis 115200 Bd 0,1°C an Schnittstelle und Display; < 0,03% des eingestellten Teilmessbereichs am Analogausgang Versorgung, Analogausgang und digitale Schnittstelle sind gegeneinander galvanisch getrennt Grüne LED Beleuchtetes LC-Display zur Temperaturanzeige oder Parametereinstellung Teilmessbereich: Bürde: Digital-Schnittstelle: Auflösung: Isolation: Betriebsanzeige: LC-Anzeige: Parameter: Emissionsgrad: Erfassungszeit t90: Maximalwertspeicher: Bereitschaftskontakt: Messunsicherheit: (bei ε=1, t90=1 s, TUmg.=10...40°C) Am Gerät oder über Schnittstelle einstellbar bzw. ablesbar: Emissionsgrad ε, Erfassungszeit t90, 0 ... 20 oder 4 ... 20 mAUmschaltung für Analogausgang, Teilmessbereich, Löschzeiten für Maximalwertspeicher, automatisches oder externes Löschen des Maximalwertspeichers, Adresse, Baudrate, Wartezeit tW Am Gerät oder über Schnittstelle ablesbar: Messtemperatur, Geräteinnentemperatur. 10 ... 100% einstellbar im Gerät oder über Schnittstelle in Stufen von 0,1% 1,5 ms (mit dynamischer Anpassung bei kleinen Signalen); einstellbar auf 0,01 s; 0,05 s; 0,25 s; 1 s; 3 s; 10 s Eingebauter Einfach- bzw. Doppelspeicher. Löschen durch eingestellte Zeit tclear (off; 0,01 s; 0,05 s; 0,25 s; 1 s; 5 s; 25 s), extern, über Schnittstelle oder auch automatisch bei neuem Messgut max. 0,15 A (nur aktiv bei automatischem Löschen oder tclear ≥ 0,25 s) Bis 400°C: 2°C Über 400°C: 0,3% vom Messwert in °C + 1°C (Hinweis: das Pyrometer muss 30 min. in Betrieb sein, bevor diese Werte gelten) Wiederholbarkeit: 0,1% vom Messwert in °C + 1°C (bei ε=1, t90=1 s, TUmg.=10...40°C) Zul. Umgebungstemperatur: Zul. Lagertemperatur: Schutzart: Gewicht: Gehäuseabmessung: Bestandene EMV-Prüfungen: Visiereinrichtung: 30 0 ... 53°C - 20 ... 60°C IP65 nach DIN 40050 ca. 550 g siehe Kap. 1.1 Entsprechend den EU-Richtlinien für elektromagnetische Verträglichkeit Laserpilotlicht (Laserklasse 2, max. Laserleistung < 1 mW, 630-660 nm) oder eingebautes parallaxenfreies Durchblickvisier IMPAC-Pyrometer IP 140 1.1 Abmessungen Pyrometer mit Durchblickvisier: Pyrometer mit Laserpilotlicht: Alle Maße in mm 2 Übersicht 7 8 1 9 2 10 3 4 1 2 3 4 5 6 5 Vario-Optik (mit Aufkleber mit Optikangaben) 4 Gewinde zum Befestigen des Pyrometers oder von Zubehörteilen Montageschiene Typenschild LC-Display Ausziehbarer Rückwanddeckel 11 6 12 7 Schrauben für Rückwanddeckel (3 mm Innensechskant) 8 Laserpilotlicht-Ein- / Ausschalter (oder parallaxefreies Durchblickvisier, je nach Gerätetyp) 9 Kontrollleuchte für Betriebszustand oder Laser (bei Geräten mit Laserpilotlicht) 10 Flanschstecker für elektrische Anschlüsse 11 LC-Display, ausgezogen 12 Einstelltasten Achtung: Die Strahlungstemperatur im Messfeld des Pyrometers darf nicht mehr als 150°C über dem Ende des Grundmessbereichs liegen, da sonst der Temperatursensor Schaden nehmen kann. 31 IMPAC-Pyrometer IP 140 2.1 Bestimmungsgemäße Verwendung Das IP 140 ist ein digitales Infrarot-Messgerät zur berührungslosen Temperaturmessung an Metallen, Keramik, Graphit etc. Es stehen 5 Temperaturbereiche zwischen 50 und 1300°C zur Verfügung. 2.2 Lieferumfang Gerät mit Durchblickvisier oder Laserpilotlicht, Optik nach Wahl, PC-Software InfraWin, Innensechskantschlüssel 3 mm, Betriebsanleitung. Hinweis: Ein Anschlusskabel ist nicht im Lieferumfang enthalten und muss separat bestellt werden (siehe Kap. 15, Bestellnummern). 3 Sicherheit Dieser Abschnitt bietet einen Überblick über wichtige Sicherheitsaspekte. Zusätzlich sind in den einzelnen Kapiteln konkrete Sicherheitsaspekte zur Abwendung von Gefahren gegeben und mit Symbolen gekennzeichnet. Darüber hinaus sind am Gerät befindliche Schilder und Beschriftungen zu beachten und in ständig lesbarem Zustand zu halten. 3.1 Allgemeines Jede Person, die damit beauftragt ist, Arbeiten am oder mit dem Gerät auszuführen, muss die Betriebsanleitung vor Beginn gelesen und verstanden haben. Dies gilt auch, wenn die betreffende Person mit einem solchen oder ähnlichen Gerät bereits gearbeitet hat oder durch den Hersteller bereits geschult wurde. Das Pyrometer darf nur zu dem in der Anleitung beschriebenen Zweck benutzt werden. Es wird empfohlen, nur das vom Hersteller angebotene Zubehör zu verwenden. 3.2 Laserpilotlicht Die Pyrometer können zum leichteren Ausrichten auf das Messobjekt mit einem Laserpilotlicht ausgestattet sein. Dieses strahlt ein sichtbares rotes Licht mit einer Wellenlänge zwischen 630 und 660 nm aus und hat eine maximale Leistung von unter 1 mW. Der Laser ist eingestuft als Produkt der Laserklasse 2. Achtung: Nicht in den Laserstrahl blicken! Laserklasse 2 nach IEC 60825-1-3-4 Hinweis: Die am Gerät angebrachten Warnschilder sollten möglichst auch nach der Montage des Gerätes gut sichtbar sein. Sicherheitsregeln: • Niemals direkt in den Laserstrahl schauen. Der Strahl kann sicher von der Seite angesehen werden. • Es ist sicherzustellen, dass der Strahl nicht in die Augen einer Person reflektiert wird (durch einen Spiegel oder eine glänzende Oberfläche). 3.3 Elektrischer Anschluss Beim Anschluss zusätzlicher Geräte, die unter Netzspannung stehen (z.B. Transformatoren), sind die allgemeinen Sicherheitsrichtlinien beim Anschluss an die 230 V-Versorgung zu beachten. Netzspannung kann beim Berühren tödlich wirken. Eine nicht fachgerechte Montage kann schwerste gesundheitliche oder materielle Schäden verursachen. Der Anschluss solcher Netzgeräte an die Netzspannung darf nur von qualifiziertem Personal durchführt werden. 32 IMPAC-Pyrometer IP 140 4 Elektrische Installation Zum Betrieb des IP 140 wird wahlweise eine Spannung von 24 V DC oder 24 V AC (48 ... 62 Hz) benötigt (möglicher Bereich: 14 ... 30 V). Das Gerät benötigt eine gewisse Aufwärmzeit, abhängig von Umgebungstemperatur und angelegter Versorgungsspannung, bis es betriebsbereit ist. Während dieser Zeit erscheint auf dem Display die Anzeige „wait“. Der Analogausgang gibt in dieser Zeit 0 mA aus. Ist das Gerät Betriebsbereit, erscheint auf dem Display die aktuelle MesstemperaGrüne Kontrolltur. Zum Ausschalten des Pyrometers ist die Spannungsversor- leuchte gung zu unterbrechen, z.B. indem der Anschlussstecker Interne Firmwareversion für abgezogen wird. Mit Anlegen der Spannungsversorgung wird für etwa eine etwa 1 s Sekunde die interne Firmware-Version im Display ange- “wait“-Anzeige zeigt. An der Geräterückseite leuchtet die grüne Kontroll- während der leuchte. Ist das Pyrometer mit einem Laserpilotlicht ausge- Aufwärmzeit stattet, so blinkt die Kontrollleuchte, wenn dieses eingeschaltet ist. Hinweis: Bei Objekttemperaturen unter 150°C und einem Emissionsgrad kleiner als 90% benötigt das Pyrometer etwa eine halbe Stunde, bis die volle Messgenauigkeit erreicht ist. Ansonsten wird eine ausreichende Genauigkeit nach etwa 10 Minuten erreicht. Um die Anforderungen der elektromagnetische Verträglichkeit zu erfüllen (EMV), dürfen nur abgeschirmte Anschlusskabel verwendet werden. Die Abschirmung des Anschlusskabels wird nur auf der Pyrometerseite angeschlossen. Auf der Seite der Spannungsquelle (Schaltschrank) bleibt die Abschirmung offen, um Masseschleifen zu verhindern. LumaSense bietet Anschlusskabel als Zubehör an, sie sind nicht im Standard-Lieferumfang enthalten. Das Anschlusskabel hat Leitungen für Spannungsversorgung, Schnittstelle, Analogausgang, externe Pilotlichtsteuerung und externes Löschen über Kontakt inklusive 12-poligem Steckverbinder (siehe Kap. 15, Bestellnummern). Es ist mit einem kurzen RS232-Verlängerungsstück mit 9-poligem D-Sub-Stecker für direkten PC-Anschluss (RS232) versehen, das bei RS485 nicht verwendet wird. 4.1 Pin-Belegung des Flanschsteckers auf der Rückseite des Pyrometers Stecker-Pin K A L B H J G F C D E M 4.1.1 Farbe weiß braun grün gelb grau Bedeutung + 24 V Versorgungsspannung (oder 24 V AC) 0 V Versorgungsspannung + Iausg. Analogausgang – Iausg. Analogausgang Pilotlicht extern einschalten (Brücke zu K) siehe 4.1.1: Maximalwert extern löschen, Bereitrosa schaftskontakt oder extern hold rot DGND (Masse für Schnittstelle) schwarz RxD (RS232) bzw. B1 (RS485) violett TxD (RS232) bzw. A1 (RS485) grau/rosa B2 (RS485) (gebrückt mit F) rot/blau A2 (RS485) (gebrückt mit C) Abschirmung, nur zur Kabelverlängerung verbinden, orange im Schaltschrank nicht auflegen Flanschstecker E F M D C G L B H J A K Pin-Belegung (Stiftseite) Stecker-Pin J Der Stecker-Pin J kann für 3 verschiedene Funktionen verwendet werden: 1) Bereitschaftskontakt: Während das Pyrometer sich in der Warmlaufphase befindet (nach Anlegen der Spannungsversorgung, auf dem Display wird „wait“ angezeigt), liegt am Pin J die Versorgungsspannung an. Diese Spannung wird über einen internen Relaiskontakt (max. 0,2 A bei 50 V) von Pin K (Spannungsversorgung) geschaltet und öffnet sich, wenn das Gerät betriebsbereit ist. Der Pin kann damit als Bereitschaftskontakt verwendet werden. 33 IMPAC-Pyrometer IP 140 2) Externes Löschen des Maximalwertspeichers : Ist das Pyrometers betriebsbereit, dient der Pin J als Eingang für das externe Löschen des Maximalwertspeichers (siehe auch 9.4). Um den Maximalwertspeicher zu löschen, muss Pin J kurz an die Versorgungsspannung (Pin K) gelegt werden. Die Funktion „externes Löschen“ funktioniert nur, wenn folgende Voraussetzungen erfüllt sind: • Die Warmlaufphase muss abgeschlossen sein (keine Anzeige „wait“) • Die Löschzeit muss auf „extern“ eingestellt sein (Einstellungen siehe 9.4 bzw. 10.7). 3) Hold-Funktion: Wird bei aktivierter Hold-Funktion (siehe 9.4 Löschzeit des Maximalwertspeichers) der Stecker-Pin J mit Pin K verbunden, wird die Temperaturanzeige während dieser Zeit auf dem aktuellen Wert gehalten. 4.2 Allgemeinhinweise zum Anschluss des Pyrometers an einen Rechner Das Pyrometer verfügt über eine Schnittstelle RS232 oder RS485 (am Pyrometer umschaltbar). Die Übertragung mit RS232 ist nur über relativ kurze Distanzen möglich und elektromagnetische Störungen können die Übertragung beeinträchtigen. Die Übertragung mit RS485 ist weitestgehend störunanfällig, es lassen sich sehr lange Übertragungsstrecken realisieren und es können mehrere Pyrometer in einem Bussystem an eine Schnittstelle angeschlossen werden. Steht keine RS485 am Rechner zur Verfügung, kann diese mit einem externen Konverter realisiert werden. Dieser wandelt die RS485 in RS232 und lässt sich damit an die Standardschnittstelle anschließen. Bei der Verwendung von RS485 Ù RS232-Konvertern ist zu beachten, dass der Konverter schnell genug sein muss, um die Antwort des Pyrometers auf einen Befehl des Masters rechtzeitig zu erfassen. Die meisten handelsüblichen Konverter sind für schnelle Messgeräte nicht geeignet. Daher wird dringend empfohlen, den LumaSense-Konverter Ι-7520 (Best.-Nr. 3 852 430) zu benutzen. Weiterhin besteht bei einer zu langsamen RS485-Verbindung auch die Möglichkeit, über Schnittstelle eine Wartezeit einzugeben, die die Antwort des Pyrometers auf einen Befehl verzögert (siehe auch 9.10 Wartezeit tw). Anschluss an Schnittstelle RS232 Die Übertragungsgeschwindigkeit (in Baud) der seriellen Schnittstelle ist von der Leitungslänge abhängig. Einstellbar sind Werte zwischen 2400 und 115200 Bd. Die zu verwendende Baudrate halbiert sich jeweils mit der Verdoppelung der Übertragungsstrecke (siehe auch 9.8 Baudrate (kBaud)). Ein Richtwert bei RS232 für 19200 Bd ist: 7 m Leitungslänge. TxD (violett) RxD (schwarz) DGND (rot) E F G M D C L B Kabelverbindung H J A 1 K 4 8 5 9 Anschluss an Schnittstelle RS485 M D C L B A K Pyrometer 1 z.B. Adresse 00 C M L B A K Pyrometer 2 z.B. Adresse 01 DGND E F G M D H J B1 G A2 F A1 B1 DGND E D H J A2 G B2 F A1 B1 E DGND A2 B2 Abschlusswiderstand 120 Ohm A1 Halb-Duplex-Betrieb: A1 und A2 sowie B1 Master und B2 sind im 12-poligen RundsteckA verbinder des Anschlusskabels gebrückt, um lange Stichleitungen und damit Reflektio- B nen zu vermeiden und beim Abziehen eines S Anschlusssteckers den RS485-Datenbus nicht zu unterbrechen. Die Master-Bezeichnungen kennzeichnen die Anschlüsse am RS485-Konverter. Die Übertragungsgeschwindigkeit der Schnittstelle (in Baud) ist von der Leitungslänge abhängig. Einstellbar sind Werte zwischen 2400 und 115200 Bd. Die zu verwendende Baudrate halbiert sich jeweils mit der Verdoppelung der Übertragungsstrecke (siehe auch 9.8 Baudrate (kBaud)). Ein Richtwert für 19200 Bd sind 2 km Gesamtleitungslänge. 34 3 PC-Seite (Lötseite Kabeldose) Pyrometerseite (Lötseite Kabeldose) 4.2.2 2 6 7 B2 4.2.1 C L B H J A K Pyrometer 32 z.B. Adresse 31 IMPAC-Pyrometer IP 140 4.3 Anschluss zusätzlicher Auswertegeräte Zusätzliche Auswertegeräte wie z.B. eine LED-Digitalanzeige benötigen lediglich den Anschluss an eine Spannungsversorgung sowie die analoge Verbindung mit dem Pyrometer. Weitere Geräte wie z.B. ein Regler oder Drucker werden dem Anschlussbild gemäß in Reihe zu dem Anzeigegerät geschaltet (GesamtLastwiderstand max. 500 Ω). weiß 230V ~ Netzteil 24 V DC braun grün °C LED-Digitalanzeige Regler Drucker gelb 5 Mechanische Installation Zur Befestigung des Pyrometers ist an der Unterseite eine Montageschiene vorhanden. Dort lässt sich ein Haltewinkel befestigen oder eine Kugelgelenkhalterung einschieben, die mit einer Schnellspannschraube fixiert werden kann. Weiter befinden sich an der Stirnseite des Gerätes 4 Gewindebohrungen M 5. 5.1 Zubehör (optional) Umfangreiches Zubehör garantiert Ihnen problemlosen Anschluss sowie Montage des Pyrometers. Einen Überblick geben die folgenden Bilder bzw. die Zeichnungen / Beschreibungen sowie die Artikelbezeichnung (siehe auch 15, Bestellnummern). Befestigung: Zur Befestigung und Ausrichtung der Pyrometer auf das Messobjekt steht ein Montagewinkel oder eine Kugelgelenkhalterungen zur Verfügung. Die Kugelgelenkhalterung ist eine schnelle und einfache Möglichkeit, um das Pyrometer auf das Messobjekt auszurichten. Die Schnellspannschraube am Kugelgelenk ermöglicht ein sehr schnelles und einfaches Justieren des Pyrometers in allen Richtungen. Montagewinkel Kugelgelenkhalterung Kühlung: Das Pyrometer kann in Umgebungstemperaturen betrieben werden, die außerhalb der des Pyrometers liegt, wenn entsprechende Schutzmaßnahmen getroffen werden. Für den Schutz vor Fronthitze steht die Kühlplatte zur Verfügung. Das komplett geschlossenen Wasserkühlgehäuse aus Edelstahl dient dem Schutz des Pyrometers in extrem heißen Umgebungen. Es ist für eine Umgebungstemperatur von bis zu 180°C ausgelegt. Kühlplatte Wasserkühlgehäuse Anzeigegeräte: Zusätzlich zur integrierten Temperaturanzeige am Pyrometer gibt es passende Einbau-Anzeigegeräte, die u.a. auch zur Fern-Parametrierung des Pyrometers verwendet werden können. Digitalanzeige DA 6000 LED-Großanzeige 35 IMPAC-Pyrometer IP 140 Verschiedenes: Der Blasvorsatz schützt die Linse vor Verschmutzungen durch Staub, Feuchtigkeit oder Schwebstoffe. Er muss mit trockener, ölfreier Druckluft betrieben werden und erzeugt einen kegelförmigen Luftstrahl. Der Schwenkvorsatz SCA 140 bewegt den Messstrahl des Pyrometers in einem Winkel von 0 ... 12° (einstellbar). Die Schwenkfrequenz ist einstellbar von 1 ... 5 Hz. Der Schwenkvorsatz wird hauptsächlich zum Erfassen des Maximalwertes von kleineren Messobjekten, z.B. schwingender, dünner Drähte eingesetzt. Blasvorsatz Der 90°-Umlenkspiegel ermöglicht die Erfassung des Messobjektes in einem Winkel von 90° zur Pyrometerachse. Schwenkvorsatz SCA 140 90°-Umlenkspiegel Der Emissionsgradverstärker kann eingesetzt werden, wenn der Emissionsgrad des zu messenden Materials zu niedrig für eine exakte Messung ist, z.B. bei Objekten mit polierter Oberfläche. Emissionsgradverstärker 6 Visiereinrichtungen Zum exakten Anvisieren des Messobjektes sind die Pyrometer der Serie 140 mit einem Durchblickvisier oder mit einem Laserpilotlicht ausgestattet. 6.1 Durchblickvisier Mit dem Durchblickvisier kann das Messobjekt durch direkte Beobachtung anvisiert werden. Das Durchblickvisier ist seitenrichtig und parallaxenfrei; ein Kreis markiert die Lage des Messfeldes, nicht jedoch die exakte Größe. 6.2 Durchblickvisier Laserpilotlicht Das Laserpilotlicht zeigt einen roten Punkt an der Position des Messobjekts an. Der Laserpunkt markiert das Zentrum des Messfeldes, aber nicht dessen Größe. Das Laserpilotlicht kann ohne Beeinflussung der Messung betrieben werden. Ist das Laserpilotlicht eingeschaltet, so blinkt die grüne Kontrollleuchte an der Pyrometer-Rückseite und auf dem Display wird „PILT“ angezeigt. Laserpilotlicht (Ein- / AusTaster) PILT-Anzeige Blinkende Kontrollleuchte Achtung: Nicht in den Laserstrahl blicken! Laserklasse 2 nach IEC 60825-1-3-4 7 Optik Je nach Bedarf wird das Gerät mit einer der hier vorgestellten Vario-Optiken geliefert. Diese lässt sich auf die benötigte Messentfernung einstellen und ermöglicht so sehr kleine Messfelder bei unterschiedlichen Entfernungen zum Messobjekt. 36 IMPAC-Pyrometer IP 140 Optik 3-P Optik 2-P Optik 1-P Optik 0-P Messabstand a [mm] a = 70 mm a = 78 mm a = 90 mm a = 105 mm a = 120 mm a = 150 mm a = 200 mm a = 260 mm a = 440 mm a = 345 mm a = 580 mm a = 1000 mm a = 4300 mm Apertur D [mm] *) MB 4 1,7 2,0 2,5 2,3 2,8 4,0 4,1 5,6 11,2 6,7 13 23 105 Messfelddurchmesser M90 [mm] MB 5.5 MB 7 MB 12 0,9 0,7 0,4 0,9 0,7 0,4 1,0 0,8 0,4 1,0 0,8 0,4 1,2 1,0 0,5 1,6 1,3 0,6 1,8 1,4 0,7 2,4 1,8 0,8 4,3 3,3 1,3 2,7 2,0 0,9 5 3,9 1,6 9 6,8 2,6 41 31 11 14 … 17 12 … 14 MB 13 0,3 0,3 0,3 0,3 0,4 0,5 0,5 0,7 1,0 0,7 1,5 2,4 10 8 … 10 Auszug S [mm] 26 13 0 26 13 0 26 13 0 26 13 6,5 0 ) * Die Apertur bezeichnet den wirksamen Durchmesser der Optik. Sie ist abhängig vom Optikauszug. Der größte Wert gilt bei voll ausgezogenem Objektiv (S = 26), der kleinste Wert gilt bei S = 0. Dazwischenliegende Werte müssen bei Bedarf interpoliert werden. Die Tabellen geben Werte bei verschiedenen Optikeinstellungen an. Wird der fest eingestellte Abstand zum Messobjekt verkleinert oder a1 vergrößert, ergibt sich eine a a2 Unschärfe, die sich in einer Vergrößerung des MessfelØ M1 Ø M2 ØM des äußert. Berechnungen zu Zwischenwerten können a2 a M1 = 1 (M + D) − D M2 = (M − D) + D mit der folgenden Formel Apertur D a a bestimmt werden: Hinweis: Das PC-Programm InfraWin beinhaltet einen Rechner („IR-Rechner“), der bei Eingabe der Daten a, M und D die benötigten Zwischenwerte liefert. Hinweis: Das Messobjekt muss mindestens so groß sein wie das Messfeld. 7.1 Einstellen auf den benötigten Messabstand Bei einer Vario-Optik muss der benötigte Messabstand eingestellt werden, um die angegebenen Messfelder zu erreichen. Das lässt sich innerhalb des kleinsten und größten Wertes der jeweiligen Optik vornehmen. Die Optik muss zunächst durch eine Drehung gegen den Uhrzeigersinn gelöst werden. Anschließend kann sie vor- und zurückbewegt werden, bis der korrekte Messabstand gefunden ist. Danach muss sie durch Drehung im Uhrzeigersinn wieder arretiert werden. Einstellen des Messabstandes mit Hilfe der Tabelle Die Tabelle gibt für jede Vario-Optik die minimal und maximal mögliche Messentfernung an. Dies entspricht einer voll ausgezogenen oder eingeschobenen Optik. Als Beispiel ist ein weiterer Zwischenwert angegeben. Der Optikauszug „S“ für diesen Wert lässt sich mit einem Messschieber exakt nachmessen. Für alle anderen Messabstände finden die folgenden Methoden Anwendung: arretieren ziehen / drücken lösen Auszug S Einstellen des Messabstandes mit Hilfe des Durchblickvisiers: Die Vario-Optik ist korrekt auf die gewünschte Entfernung eingestellt, wenn das Messobjekt beim Anvisieren mit dem Durchblickvisier scharf dargestellt wird. Ein Zielkreis im Okular markiert dabei die Lage des Messfeldes. Einstellen des Messabstandes mit Hilfe des Laserpilotlichtes: Der Laserstrahl hat seinen kleinsten Durchmesser und damit die schärfste Abbildung auf dem fokussierten Messabstand. 37 IMPAC-Pyrometer IP 140 8 Geräteeinstellungen Das IP 140 ist mit umfangreichen Einstellmöglichkeiten ausgestattet, um es an die jeweilige Messbedingung optimal anzupassen (Beschreibung aller verfügbaren Parameter siehe Kap. 9, Parameter) und die Temperatur des Messobjekts korrekt zu erfassen. Sämtliche Einstellungen lassen sich direkt am Pyrometer oder über serielle Schnittstelle und Software InfraWin vornehmen (siehe Kap. 10, Einstellungen über Schnittstelle / Software. Benutzer eines eigenen Kommunikationsprogramms finden alle Schnittstellenbefehle in Kap. 14, Datenformat UPP®). 8.1 Einstellungen am Gerät Das LC-Display sowie Taster zum Einstellen und Ablesen der Parameter befinden sich im Inneren des Gerätes und sind nach Aufschrauben und Abziehen des Rückwanddeckels erreichbar. Dazu sind die vier Innensechskantschrauben zu lösen und der Deckel ist zusammen mit der Anzeige / Einstelleinheit gerade nach hinten abzuziehen. Der Auszug ist durch die Länge der Schrauben beschränkt. Die Anzeige ist sowohl im eingebauten als auch im ausgezogenen Zustand immer beleuchtet. Hinweis: Achten Sie darauf, dass beim Öffnen keine Verunreinigungen in das Gerät gelangen. 8.2 1 2 3 8.3 Beleuchtetes Display Einstelltasten Max. Auszug Rückwanddeckel Anzeigen und Einstellen der Parameter PAR: Das Anzeigen der verfügbaren Parameter Temperaturanzeige erfolgt mit der PAR-Taste in der nachfolgend oder Parameterbeschriebenen Reihenfolge (Kap. 10). Mit kurzform jedem erneuten Drücken wechselt die AnTasten-Kurzzeige zum nächsten Parameter und auf der beschreibung Anzeige erscheint eine entsprechende Kurzform (in Kap. 9, Parameter in Klammern hinter dem jeweiligen Parameternamen). 1 2 3 : Mit den Pfeiltasten und lassen sich für jeden Parameter die entsprechenden Einstellwerte anzeigen. Längeres Drücken der Taster verändert die Einstellungen im Schnellmodus. ESC / ENT: Mit Betätigen der ESC-Taste kehrt das Pyrometer zurück in den Messmodus. Wurde ein Parameter mit den Pfeiltasten verändert, so ändert sich die Bezeichnung der ESC-Taste in ENT. Der neue Wert muss damit zur Übernahme in das Pyrometer bestätigt werden. Ein Wechsel auf einen anderen Parameter (mit der PAR-Taste) übernimmt einen mit den Pfeiltasten veränderten Wert nicht. Erfolgt ca. 30 s lang keine Eingabe, wechselt das Pyrometer ohne Übernahme des evtl. geänderten Parameters in den Messmodus zurück. Wahl der seriellen Schnittstelle Gegenüber dem Display befindet sich ein Umschalter, mit dem die Schnittstelle RS232 oder RS485 gewählt werden kann. Die Auswahl wird auf dem LC-Display angezeigt. 38 Verschraubung 3 mm-Innensechskant Schnittstellenumschalter TestTaster Anzeige der gewählten Schnittstelle IMPAC-Pyrometer IP 140 8.4 Testfunktion Der „Test“-Taster schaltet einen Teststrom auf den AnaTestfunktion aktiv logausgang, mit dem sich die korrekte Anzeige eines (hier ist der Analogexternen Anzeigegerätes überprüfen lässt. Der ausgegeausgang auf 0 … 20 mA eingestellt) bene Teststrom liegt in der Mitte des eingestellten Analogausgangs, also 10 mA bei einem eingestellten AnaAnzeige, die auch auf logausgang von 0 ... 20 mA und 12 mA bei einem Anaeinem externen Anzeigegerät zu sehen logausgang von 4 ... 20 mA. Auf dem LC-Display wird sein muss (hier Messdieser Wert angezeigt, zusammen mit der entsprechenbereich 75 ... 550°C) den Temperatur, also der Mitte des eingestellten (Teil-) Messbereichs (z.B. 312,5°C bei einem Messbereich von 75 ... 550°C). Diese Temperatur muss sich auch auf dem externen Anzeigegerät wiederfinden. Ist das nicht der Fall, so ist das Anzeigegerät auf einen anderen Strom oder Messbereich eingestellt, als das Pyrometer. Nach erneutem Druck auf den „Test“-Taster, einen Taster an der LC-Anzeige oder nach einer Zeit von ca. 30 Sekunden wird das Signal wieder ausgeschaltet und der Messmodus ist aktiv. 8.5 Werkseinstellungen Emissionsgrad (Emi) = 100% Erfassungszeit (t90) = min Löschzeit (tClear) = off Analogausgang (mA) = 0 ... 20 mA Teilmessbereich (from / to) entspricht Grundmessbereich Adresse (Adr) = 00 Baudrate (Baud) = 19200 Bd Temperaturanzeige (C / F) = °C Wartezeit (tw) (für RS485) = 10 Schnittstelle (RS485 / RS232) = RS232 9 Parameterbeschreibung / Einstellungen 9.1 Emissionsgrad ε (Emi) Unter dem Emissionsgrad ε versteht man das Verhältnis der abgestrahlten LeisAnzeige tung eines beliebigen Objekts zur abgestrahlten Leistung eines „Schwarzen Strahdes eingelers“ gleicher Temperatur (ein „Schwarzer Strahler“ ist ein Körper, der alle einfalstellten lenden Strahlen absorbiert mit einem Emissionsgrad von 100%). Der EmissionsEmissionsgrad ist materialabhängig und liegt zwischen 0% und 100% (Einstellmöglichkeiten grades am Pyrometer: 10 ... 100%, der eingestellte Wert wird auf dem Display angezeigt). Zusätzlich ist der Emissionsgrad von der Oberflächenbeschaffenheit des Materials, dem Spektralbereich des Pyrometers und der Messtemperatur abhängig. Der Emissionsgrad muss am Pyrometer entsprechend eingestellt werden. Typische Emissionsgrade für die Spektralbereiche der Geräte liefert folgende Tabelle. Die angegebenen Toleranzen bei den einzelnen Materialien sind hauptsächlich von der Oberflächenbeschaffenheit abhängig. Raue Oberflächen haben höhere Emissionsgrade. Messobjekt „Schwarzer Strahler“ Aluminium, stranggepresst Messing Messing oxidiert Kupfer Kupfer oxidiert Inconel Inconel oxidiert Eisen verzundert Stahl-Walzhaut Emissionsgrad [%] (bei 2,4 µm) 100 13 18 65 ... 70 5 70 ... 80 30 85 85 ... 90 80 ... 88 Messobjekt Flüssigstahl Nickel Titan blank Titan oxidiert Molybdän Molybdän oxidiert Ruß Graphit Porzellan, glasiert Porzellan, rau Emissionsgrad [%] (bei 2,4 µm) 20 ... 30 15 ... 20 50 75 ... 80 10 75 ... 80 95 80 ... 90 60 80 ... 90 39 IMPAC-Pyrometer IP 140 9.2 Umgebungstemperaturkompensation (ta) Die Umgebungstemperaturkompensation kann bei niedrigen Messtemperaturen (unter Einstellungen: 100°C) benutzt werden. Sie bleibt im Normalfall auf der Standardeinstellung „auto“ auto stehen, da die das Pyrometer umgebende Luft in der Regel die gleiche Temperatur wie 00°C (32°F) .. . die Umgebung des Messobjekts hat. Befindet sich das Messobjekt jedoch in einem 53°C (127°F) Bereich mit höherer Wandtemperatur (z.B. in einem Ofen), kann das Messergebnis verfälscht werden (zu hohe Temperaturanzeige). Zur Kompensation dieses Einflusses kann mit Hilfe dieser Funktion die Umgebungstemperatur (innerhalb des Pyrometermessbereichs) des Messobjektes eingegeben werden. Es ist allerdings zu bedenken, dass diese Methode nur dann zum korrekten Messwert führt, wenn dieser Umgebungstemperatureinfluss am Messobjekt immer konstant ist und der Emissionsgrad gut bekannt ist. 9.3 Erfassungszeit (t90) Die Erfassungszeit ist die Zeitspanne, in der die Messtemperatur bei sprunghafter ÄnEinstellungen: derung mindestens im Messfeld anstehen muss, damit der Ausgangswert des Pyromemin ters einen vorgegebenen Messwert erreicht. Die Zeiten beziehen sich dabei auf 90% 0,01 s des gemessenen Temperatursprungs. Bei „min“ arbeitet das Gerät mit seiner Eigenzeit0,05 s .. . konstanten. 10,00 s Die dynamische Anpassung bei niedrigen Signalpegeln sorgt für eine automatische Verlängerung der Erfassungszeit am Messbereichsanfang, auch wenn t90 auf einen kleineren Wert gestellt ist. Die folgende Tabelle zeigt die Abhängigkeit der Erfassungszeit von der Messtemperatur. Umschaltpunkte bei steigender Temperatur: Messbereich 50 ... 75 ... 100 ... 160 ... 200 ... 400°C 550°C 700°C 1200°C 1300°C (MB 4) (MB 5.5) (MB 7) (MB 12) (MB 13) Umschaltpunkte für ε = 1 35 ms → 5 ms 5 ms → 1,5 ms 102°C 129°C 137°C 180°C 172°C 224°C 255°C 329°C 272°C 352°C Umschaltpunkte für ε = 0,5 35 ms → 5 ms 5 ms → 1,5 ms 114°C 147°C 157°C 205°C 196°C 254°C 291°C 373°C 310°C 400°C Am Messbereichsanfang und bei kleinem Emissionsgrad muss bei eingestelltem Maximalwertspeicher eine längere Erfassungszeit gewählt werden, um Messfehler durch Rauschen zu unterdrücken. Langsamere Zeiten können sinnvoll sein, um über schnelle Schwankungen der Objekttemperatur zu mitteln. 9.4 Löschzeit des Maximalwertspeichers (tClear) Bei eingeschaltetem Maximalwertspeicher wird immer der höchste, letzte Messwert angezeigt und gespeichert. Der Speicher muss regelmäßig zurückgesetzt werden, damit er durch einen neuen, aktuellen Wert ersetzt werden kann. Angewendet wird ein solcher Speicher z.B. bei schwankenden Temperaturen, wo die Anzeige sehr „unruhig“ ist oder das Messobjekt nur kurz am Messstrahl vorbeigeht. Damit dieser Wert für jedes Messobjekt neu ermittelt werden kann, ist es sinnvoll, den Speicher regelmäßig oder vor der Messung eines neuen Messobjekts zu löschen. Einstellungen: off 0,01 s . .. 25 s extern auto Hold Folgende Einstellungen stehen zur Verfügung: off: Bei Löschzeit „off“ ist der Maximalwertspeicher abgeschaltet und der Momentanwert wird gemessen. 0,01...25 s: Wird eine Löschzeit zwischen 0,01 und 25 s gewählt, wird der Maximalwert ermittelt und im Doppelspeicher festgehalten. Nach der gewählten Zeit wird er wieder gelöscht. extern: Ein externes Löschen lässt sich über eine eigene Software aktivieren und verwenden (siehe auch Kap. 14, Datenformat UPP®) oder auch über einen externen Löschkontakt (zum Anschluss siehe 4.1 Pin-Belegung des Flanschsteckers auf der Rückseite des Pyrometers). In diesem Fall wirkt der Speicher nur als Einfachspeicher, da nur ein Löschmechanismus zur Verfügung steht auto: Der Modus „auto“ wird für diskontinuierliche Messaufgaben verwendet. Es werden z.B. Objekte auf einem Förderband transportiert und passieren das Pyrometer nur für einige Sekunden. Da40 IMPAC-Pyrometer IP 140 Hold: bei soll die Maximaltemperatur von jedem Teil erfasst werden. Im „auto“-Modus wird der Maximalwert so lange gehalten, bis ein neues heißes Objekt in den Messstrahl kommt. Die Temperatur, die als „heiß“ erkannt werden soll, ist dabei durch die untere Grenze des Messbereichs definiert. Der gespeicherte Maximalwert wird dann gelöscht, wenn die Temperatur eines neuen heißen Objektes die untere Grenze des Messbereichs oder eingestellten Teilmessbereichs um 1% oder mindestens 2°C überschreitet. Die Funktion „Hold“ ermöglicht das festhalten des aktuellen Messwertes zu einem beliebigen Zeitpunkt. Dazu muss ein externer Taster oder Schalter angeschlossen werden (siehe 4.1.1 Stecker-Pin J), welcher bei Betätigung den aktuellen Messwert solange hält und speichert, bis er wieder gelöst wird. Funktions-Hinweis: Je nach gewählter Einstellung arbeitet der Maximalwertspeicher entweder als Einfachspeicher oder als Doppelspeicher: Einfachspeicher: Der Einfachspeicher kommt zum tragen, wenn Sie zum Löschen des Speichers einen externen Kontakt (gibt einen Löschimpuls) angeschlossen haben (beispielsweise zwischen zwei Messobjekten). Dieser Kontakt ist direkt am Pyrometer zwischen Stecker PIN J und K anschließbar. Hierbei nimmt nach jedem Löschimpuls das Gerät immer erst den jeweiligen neuen, aktuellen Messwert an, um sich dann schrittweise dem neuen Maximalwert zu nähern. Doppelspeicher: Geben Sie die Löschzeiten über die Tasten am Pyrometer bzw. über Schnittstelle oder PC ein, wird automatisch der Doppelspeicher benutzt. Es handelt sich dabei um zwei Speicher, auf die der jeweils höchste Wert der Messspannung geleitet wird und die immer abwechselnd mit der eingegebenen Taktzeit gelöscht werden, so dass der andere Speicher den Maximalwert noch für eine Zykluszeit behält. Damit wird verhindert, dass die Temperaturanzeige mit der Taktfrequenz einbricht. Hinweis: Der Maximalwertspeicher ist der Erfassungszeitfunktion nachgestellt. Dies hat zur Konsequenz, dass: • Löschzeiten ≤ der eingestellten Erfassungszeit sinnlos sind • die Löschzeit mind. 3 x größer als die Erfassungszeit sein muss • nur Maxima mit vollem Maximalwert erfasst werden können, die länger als 3 x Erfassungszeit anliegen. 9.5 Analogausgang (mA) Der Analogausgang muss so gewählt werden, dass er mit dem Signaleingang Ihres Auswertegerätes (z.B. Regler, SPS, ...) übereinstimmt. 9.6 Einstellungen: 0 ... 20 mA 4 ... 20 mA Teilmessbereich (from / to) Es besteht die Möglichkeit, einen Teilmessbereich (Mindestumfang 51°C) innerhalb des Gesamtmessbereichs auszuwählen. Dieser Teilmessbereich entspricht dem Analogausgang. „from“ stellt den Teilmessbereichsanfang dar, „to“ das Teilmessbereichsende. Mit Hilfe des Teilmessbereichs ist es außerdem möglich, die Anforderung an den „Auto“-Löschmodus des Maximalwertspeichers zu konfigurieren (s.o.). 9.7 Adresse (Adr) Zum Betrieb mehrerer Geräte mit RS485-Schnittstellen ist es nötig, jedem Gerät eine Einstellungen: eigene Adresse zuzuweisen, unter der es angesprochen werden kann. Dazu muss 00 .. . zunächst jedes Gerät einzeln mit einer Adresse versehen werden. Danach können alle 97 Geräte angeschlossen werden. Sollen bestimmte Parameter bei allen Geräten gleichzeitig verändert werden, so ist das mit der globalen Adresse 98 möglich (es erfolgt keine Antwort der Geräte). Sollte die Adresse eines Gerätes unbekannt sein, so haben Sie die Möglichkeit, jedes Gerät unabhängig von der eingestellten Adresse mit der globalen Adresse 99 anzusprechen (nur ein Gerät anschließen). 41 IMPAC-Pyrometer IP 140 9.8 Baudrate (kBaud) Die Übertragungsgeschwindigkeit der seriellen Schnittstelle (in Baud) ist von der Leitungslänge abhängig. Ein Richtwert bei RS232 für 19200 Bd sind: 7 m Leitungslänge, bei RS485: 2 km. Die Baudrate halbiert sich jeweils mit der Verdoppelung der Übertragungsstrecke. Einstellungen: 2,4 .kBd 9.9 Einstellungen: °C °F Temperaturanzeige (C / F) Die Anzeige der Temperatur kann wahlweise in °C oder °F erfolgen. 9.10 115,2 kBd Wartezeit (tw) Beim Betrieb eines Pyrometers über RS485 kann es vorkommen, dass die Verbindung nicht schnell genug ist, um die Antwort des Pyrometers auf einen Befehl des Masters rechtzeitig zu erfassen. In diesem Fall kann eine Wartezeit eingegeben werden, die den Datentransfer verzögert (z.B.: tw = 02 bei einer Baudrate von 9600 bedeutet eine Wartezeit von 2/9600 sec). 9.11 .. Einstellungen: 00 .Bit .. 99 Bit Maximale Innentemperatur (MaxIntTemp) Die maximale Innentemperatur, die das Gerät jemals erreicht hat, kann hier ausgelesen werden. 9.12 Fehler-Status (Status) Im Falle eines Gerätefehlers gibt das Gerät einen Hex-Code aus, mit dem der LumaSense-Service diesen Fehler lokalisieren kann. Als Standard zeigt diese Anzeige „ok“ an. 10 Einstellungen über Schnittstelle / Software Im Lieferumfang der Pyrometer ist die Bearbeitungs- und Auswertesoftware InfraWin enthalten. Damit haben Sie die Möglichkeit, sämtliche Funktionen, die das Pyrometer bietet, am PC abzulesen und einzustellen. Zusätzlich lassen sich die Messdaten grafisch und numerisch anzeigen und auswerten. Dieses Kapitel gibt einen Überblick über die einzelnen Funktionen der Software. Eine Beschreibung findet sich auch direkt im Hilfemenü von InfraWin. Die hier beschriebenen Möglichkeiten beziehen sich auf die Programmversion 4.0. Die jeweils aktuellste Version ist kostenlos als Download von der Homepage www.lumasenseinc.com erhältlich. 10.1 Anschluss des Pyrometers an einen PC Das Programm InfraWin kann ein oder zwei Geräte betreuen. Bei zwei Geräten mit RS232-Schnittstelle müssen zwei PC-Schnittstellen benutzt werden. Zwei Geräte mit RS485 können an der gleichen Schnittstelle parallel betrieben werden, wenn ihre Adressen unterschiedlich eingestellt wurden (siehe 9.7 Adresse). 10.2 Installation Zum Installieren wählen Sie das Installations-Programm „setup.exe“ von der InfraWin-CD oder aus dem heruntergeladenen und entpackten Zip-Archiv und folgen Sie den Anweisungen. 10.3 Programmstart Nach der Installation und dem ersten Programmstart können Sie eine Sprache wählen (deutsch, englisch, italienisch, spanisch. Die Sprache kann auch später noch geändert werden). Anschließend ist die Startseite mit dem folgenden Startmenü zu sehen: 42 IMPAC-Pyrometer IP 140 10.4 Das Menü Öffnen einer gespeicherten Datei Speichern der erfassten Messwerte zur späteren Nachbearbeitung Messung mit Farbbalkendarstellung Messung mit graphischer Darstellung Einstellung der Geräteparameter Wahl der Schnittstelle, Baudrate und der Pyrometer-Adresse (bei RS485) Zeitwerte zwischen den Messwertabfragen Anzahl der angeschlossenen Geräte (maximal zwei) Auswertung der gemessenen oder gespeicherten Werte in Tabellenform Auswertung gemessener oder gespeicherter Werte in einer Grafik Auswertung gemessener oder gespeicherter Werte in einer Text-Datei Berechnet Messfelddurchmesser bei verschiedenen Messabständen Nur wenn verfügbar: Steuerung des Programmreglers PI 6000 10.5 Vorbereitung Bevor das Programm benutzt wird, ist zunächst unter Computer die Schnittstelle auszuwählen, mit der das Pyrometer verbunden ist. Beim Anschluss von 2 Pyrometer mit RS232-Schnittstelle müssen auch 2 Schnittstellen ausgewählt werden. 10.6 Anzahl Pyrometer Durch Anklicken von „Anzahl der Geräte“ wechselt InfraWin auf die Anzeige von 1 oder 2 Geräten. Sind 2 Geräte ausgewählt, so stellt InfraWin jeweils 2 Fenster zur Eingabe oder Auswertung dar. 10.7 Grundeinstellungen Unter Pyrometer-Parameter können alle voreingestellten Werte ausgelesen oder ggf. verändert werden. Beschreibungen zu den Eingabemöglichkeiten sind unter Kap. 9, Parameterbeschreibung / Einstellungen zu finden. Wählen Sie in dem jeweiligen Listenfeld den für Sie in Frage kommenden Parameter aus, die aktuelle Einstellung des Gerätes wird angezeigt. Hinweise: • Der Grundmessbereich gibt den Gesamtmessbereich des Pyrometers an, wird automatisch angezeigt und kann nicht geändert werden. • Unter „Material haben Sie die Möglichkeit, verschiedene Materialien mit den dazugehörigen Emissionsgraden einzugeben und aus der Liste jederzeit wieder aufzurufen. • Wählen Sie, ob die Temperaturangaben in °C (Celsius) oder °F (Fahrenheit) angezeigt werden sollen. • Das Laserpilotlicht-Symbol ( ) ist nur zu sehen, wenn das Gerät damit ausgestattet ist. Ein Klick auf das Laserpilotlicht-Symbol schaltet das Laserpilotlicht ein, nach erneutem Klick oder automatisch nach ca. 2 min wird es wieder ausgeschaltet. • Zur Aktivierung der Umgebungstemperatur-Kompensation muss auf „man.“ umgeschaltet werden. Das Eingabe-Feld T(Umg) wird dann aktiviert, in dem die Umgebungstemperatur eingegeben werden kann. In der Stellung „auto“ geht das Gerät davon aus, dass die Umgebung des Messobjekts die gleiche Temperatur wie das Pyrometer hat (Standardeinstellung). 43 IMPAC-Pyrometer IP 140 Mit den Öffnen- / Speichern-Feldern lassen sich eigene Pyrometer-Konfigurationen aufrufen sowie abspeichern. „1 Messung“ zeigt für etwa eine Sekunde im Fenster der Pyrometer-Parameter die aktuelle Messtemperatur an. „Test“ öffnet ein Fenster, das die direkte Kommunikation mit dem Pyrometer über die Schnittstellenbefehle ermöglicht (siehe Kap. 14, Datenformat UPP® (Universelles Pyrometer-Protokoll)). Nach Eingabe eines Schnittstellenbefehls (00 ist die voreingestellte Geräteadresse, „ms“ ist z.B. der Befehl „Messwert abfragen“) und einem Klick auf „Senden“ öffnet sich das hier abgebildete Fenster. Hier ist bereits die Antwort des Pyrometers in 1/10 °C zu sehen. Die aktuelle Messtemperatur beträgt in diesem Fall 323,2°C. „Len“ bezeichnet die Länge des zurückgegebenen Datenstrings inklusive des Carriage Return (Chr(13)). Im unteren Teil des Fensters besteht die Möglichkeit, die Verbindung mit der zuvor eingestellten Baudrate zu überprüfen. Der Befehl wurde 500 x mit 19200 Baud gesendet, hat 4,5 sec dafür benötigt und keine Übertragungsfehler gemeldet. Hinweis: Ist das Fenster Pyrometer-Parameter geöffnet, so ist das Verändern der Parameter am Pyrometer blockiert. 10.8 Messung (Farb-Balken) Dieses Fenster stellt dar: • Aktuelle Messtemperatur, graphisch als Farb-Balken-Darstellung und numerisch • Messbereich bzw. eingestellter Teilmessbereich • Dateigröße und Anzahl der gemessenen Werte der aktuellen Messung • Emissionsgrad ε • Die aktuelle Geräteinnentemperatur (Tint) • Minimal- (Tmin) und Maximalwerte (Tmax) Der Farbbalken zeigt den Messbereich oder eingegebenen Teilmessbereich an. Durch Eingabe von Temperaturwerten in den weißen Feldern rechts und links vom Farbbalken oder durch Verschieben der danebenliegenden Striche mit der Maus können Grenzen für den Farbwechsel des Balkens eingestellt werden. Bei Temperaturen innerhalb der beiden Grenzen wird der Balken grün dargestellt, außerhalb rot. Zusätzlich befindet sich in dem Fenster ein Eingabefeld für den Emissionsgrad ε. Wird der Emissionsgrad verändert, so kann eine damit verbundene Temperaturänderung direkt abgelesen werden. Das Laserpilotlicht ( ) kann ein- oder ausgeschaltet werden, wenn das Pyrometer damit ausgestattet ist. Emi: AutoFind: Zusätzlich befindet sich in dem Fenster ein Eingabefeld für den Emissionsgrad ε. Wird der Emissionsgrad verändert, so kann eine damit verbundene Temperaturänderung direkt abgelesen werden. Für den Fall, dass die wahre Temperatur des Messobjekts bekannt sein sollte, kann mit der Funktion „Emi: AutoFind“ der Emissionsgrad des Messobjekts berechnet werden: 44 IMPAC-Pyrometer IP 140 • • • Mit dem aktuell eingestellten Emissionsgrad (in diesem Bsp. 100%) wird eine Messtemperatur angezeigt (hier: 824°C). Durch Drücken von „Emi: Autofind“ wird ein Fenster geöffnet, das die Eingabe der „wahren“ Temperatur ermöglicht. Nach Eingabe und Bestätigung der Temperatureingabe mit „OK“ berechnet InfraWin den Emissionsgrad, der sich mit der neuen Temperatur ergibt. Dieser wird sofort angezeigt und direkt für die weitere Temperaturmessung verwendet. 10.9 Messung (Online-Grafik) Dieses Fenster stellt dar: • Temperatur als grafische Darstellung • Aktuelle Messtemperatur • Anzahl der gemessenen Werte sowie die Dateigröße der aktuellen Messung Das hier dargestellte Beispiel zeigt den Ausschnitt einer Messung über den Zeitraum von etwa 10 s bei einem Messbereich von 75 550°C und der aktuellen Temperatur von 244,3°C. Das Laserpilotlicht kann ein- oder ausgeschaltet werden, wenn das Pyrometer damit ausgestattet ist. • Mit „Zone markieren“ kann ein Temperaturbereich zum leichteren Erkennen farbig markiert werden. • Mit „Schwellwert“ kann eine Temperatur eingegeben werden, ober- oder unterhalb der keine Messwerte mehr aufgezeichnet werden. Die Größe der gespeicherten Datei lässt sich so kleiner halten. • „Graphik-Grenzen“ grenzt die Darstellung des Temperaturbereichs auf den benötigten Bereich ein. Hinweis: Bei Aufruf von einer der Messungen Online-Grafik oder Farb-Balken werden die Messdaten automatisch gespeichert unter der Bezeichnung standard.i12. Sollen die Daten später zur Nachbearbeitung zur Verfügung stehen, bietet es sich an, die Datei in einer anderen .i12-Datei zu speichern, da der erneute Beginn einer Messung die Werte der alten Messung überschreibt. Dateien aus älteren Programmversionen mit der Endung .i10 lassen sich öffnen und als .i12 abspeichern. 10.10 Tabelle (Auswertung) Hier werden die gemessenen Temperaturwerte zur nachträglichen Auswertung oder Analyse numerisch aufgelistet. Da während der kleinsten Zeiteinheit von 1 s mehrere Daten anfallen können, gibt es noch eine zweite Zeitangabe, die die Zeit in sec. nach Mitternacht (0:00 h) angibt. Die Menge der Daten hängt davon ab, wie häufig eine Messung durchgeführt wird (Eingabe unter 10.13 Zeitintervall zwischen zwei Messungen). Mit der Menge der Daten wächst auch der Speicherbedarf, der nötig ist, um die Datei zu speichern. Um Platz zu sparen, sind die Daten in .i12Dateien binär codiert abgelegt. 45 IMPAC-Pyrometer IP 140 10.11 Ausgabe .TXT-Datei (Auswertung) Die gleiche Datei, wie unter „Ausgabe Tabelle“ lässt sich umwandeln in eine Text-Datei, die sich z.B. unter EXCEL einfach öffnen lässt. EXCEL formatiert die Spalten mit den Standard-Importeinstellungen (Tabulator als Trennzeichen) automatisch richtig. 10.12 Ausgabe Grafik (Auswertung) In der Grafik-Ausgabe wird die Kurve des Temperaturverlaufs über der Zeit im relevanten Messbereich dargestellt. Zusätzlich sind auf der rechten Seite des Fensters die der Messung zugrundeliegenden Daten, sowie die Uhrzeit und Temperatur an der Stelle der senkrechten, mit der Maus verschiebbaren Cursor-Linie zu sehen. Bei Aufruf der Grafik-Ausgabe werden zunächst alle gespeicherten Daten im Grafikfenster angezeigt. Überschreitet die Datenmenge eine vernünftig darzustellende Größe, so haben Sie die Möglichkeit, nach Drücken der Taste „Zoom“ mit der Maus einen Teilausschnitt zu wählen (wie der dargestellte Ausschnitt im Beispiel). Unter „Gesamt“ können Sie dann wieder die gesamte Kurve der Messung darstellen. Hinweis: Die jeweils letzte Messung wird in der Datei standard.i12 gespeichert und beim Öffnen von Tabelle oder Grafik-Ausgabe automatisch in diese hineingeladen. Wurde zuvor mit Datei öffnen eine andere Datei geladen, so wird diese geöffnet und die bisherige standard.i12 überschrieben. 10.13 PC-Aufnahmerate (Zeitintervall zwischen zwei Messungen) Mit dieser Eingabe wird ein Zeitintervall festegelegt, nach dem jeweils ein Messwert auf dem Rechner gespeichert wird. Je größer das Zeitintervall ist, desto kleiner bleibt die gespeicherte Datei. Diese Funktion wird hauptsächlich für Langzeitversuche eingesetzt. 10.14 Messfeld-Rechner Der Messfeld-Rechner wird nur für Pyrometer mit Festoptik verwendet! Nach Eingabe der Apertur und des Nenn-Messfelddurchmessers lassen sich durch einfache Eingabe Zwischenwerte des Messfelddurchmessers bei verschiedenen Messabständen einer Festoptik berechnen. 11 Transport, Verpackung, Lagerung Das Gerät kann durch unsachgemäßen Transport beschädigt oder zerstört werden. Steht die Originalverpackung nicht mehr zur Verfügung, ist zum Transport des Gerätes ein mit stoßdämpfendem PE-Material ausgelegter Karton zu verwenden. Bei Überseeversand oder längerer Lagerung in hoher Luftfeuchtigkeit sollte das Gerät durch eine verschweißte Folie gegen Feuchtigkeit geschützt werden (evtl. Silicagel beilegen). Die Pyrometer sind für eine Lagertemperatur von -20 ... 70°C ausgelegt. Die Lagerung des Pyrometers über oder unter dieser Temperatur kann zu Beschädigung oder Fehlfunktionen führen. 46 IMPAC-Pyrometer IP 140 12 Wartung 12.1 Sicherheit Vorsicht bei Wartungsarbeiten am Pyrometer. Ist das Pyrometer in laufende Prozesse einer Anlage integriert, so sollte diese gegebenenfalls ausgeschaltet und gegen Wiedereinschalten gesichert werden. Danach kann die Wartungsarbeit am Pyrometer durchgeführt werden. 12.2 Allgemeines Das Gerät besitzt keine Teile, die einer Wartung unterliegen, nur die Linse muss zur einwandfreien Messung in sauberem Zustand gehalten werden. Bei Verschmutzung kann die Linse mit einem weichen Tuch in Verbindung mit Spiritus gereinigt werden. Es können auch handelübliche Brillen- oder Foto-Objektiv-Reinigungstücher verwendet werden (keine säurehaltigen Mittel oder Lösungsmittel verwenden). 12.3 Austausch der Optik Es stehen 4 verschiedene Optiken zur Auswahl. Die Linsen dieser Optiken sind gegeneinander austauschbar, ohne dass das Gerät nachkalibriert werden muss. Zum Austauschen der Linse ist der innere Befestigungsring abzuschrauben, danach kann eine andere Linse mit der gewölbten Seite nach außen eingesetzt und mit einem neuen, passenden Ring verschraubt werden. Zur Unterscheidung sind sie mit einer Farbkennzeichnung am Linsenrand versehen: Vario-Optik 0-P: Vario-Optik 1-P: Vario-Optik 2-P: Vario-Optik 3-P: 13 (Messabstand 70 ... 90 mm) (Messabstand 105 ... 150 mm) (Messabstand 200 ... 440 mm) (Messabstand 345 ... 4300 mm) (ohne) (rosa) (schwarz) (blau) Linse innerer Befestigungsring Aufkleber mit Optikdaten Fehlerdiagnose Bevor das Pyrometer zur Reparatur eingesendet werden muss, können Sie versuchen, zunächst den Fehler anhand der nachfolgenden Liste zu erkennen und zu beheben. Temperaturanzeige zu niedrig • Pyrometer falsch auf das Messobjekt ausgerichtet ⇒ Neu ausrichten, um maximales Temperatursignal zu erreichen (siehe 6) • Messobjekt ist kleiner, als Messfeld ⇒ Messabstand überprüfen, kleinstes Messfeld ist bei Nennmessabstand (siehe 7) • Messobjekt befindet sich nicht ständig im Messfeld ⇒ Aktivieren des Maximalwertspeichers (siehe 9.4) • Emissionsgrad ist zu hoch eingestellt. ⇒ Emissionsgrad auf niedrigeren Wert entsprechend des Materials korrigieren (siehe 9.1) • Optik verschmutzt ⇒ Optik reinigen (siehe 12.2) Temperaturanzeige zu hoch • Emissionsgrad ist zu niedrig eingestellt. ⇒ Emissionsgrad auf höheren Wert entsprechend des Materials korrigieren (siehe 9.1) • Die Messung wird durch Reflektionen von heißen Anlagenteilen beeinflusst ⇒ Mit mechanischer Vorrichtung Störstrahlung abschirmen Messfehler • Angezeigte Temperatur wird im Laufe der Zeit niedriger, vermutlich Verschmutzung der Optik ⇒ Optik reinigen. Verwendung des Luftspülvorsatzes empfohlen (siehe 12.2, 5.1) • Angezeigte Temperatur wird trotz Luftspülvorsatz im Laufe der Zeit niedriger, vermutlich schmutzige Druckluft oder Druckluftausfall ⇒ Optik reinigen und saubere, ölfreie und trockene Luft verwenden • Sicht auf Messobjekt ist durch Staub oder Wasserdampf getrübt 47 IMPAC-Pyrometer IP 140 • • • ⇒ Pyrometerposition ändern, mit freier Sicht zum Messobjekt (ggf. Quotienten-Pyrometer verwenden) Messfehler infolge HF-Störungen. ⇒ Abschirmung falsch angeschlossen, gemäß Kapitel 4 anschließen Gerät überhitzt ⇒ Kühlvorrichtung mit Luft- oder Wasserkühlung verwenden (siehe 5.1) Schwankende Temperaturanzeige, wahrscheinlicht durch Änderung des Emissionsgrades ⇒ Falscher Pyrometertyp, Quotientenpyrometer verwenden. Laserpilotlicht • Laserpilotlicht funktioniert nicht mehr, Maximale Geräteinnentemperatur überschritten ⇒ Kühlvorrichtung mit Luft- oder Wasserkühlung verwenden (siehe 5.1) Hinweis: Der Spektralbereich des IP 140 reagiert bei niedrigen Messtemperaturen auf Glühlampenlicht oder sehr helles Tageslicht (gilt nicht für Leuchtstoffröhren). Um das Messergebnis nicht zu verfälschen, sollte starke Fremdlichteinwirkung auf das Messobjekt daher vermieden werden. 14 Datenformat UPP® (Universelles Pyrometer-Protokoll) Über Schnittstelle lassen sich mit einem geeignetem Kommunikationsprogramm oder über das TestEingabefeld in der Software InfraWin (siehe 10.7 Grundeinstellungen Æ Test) Befehle direkt mit dem Pyrometer austauschen. Der Datenaustausch erfolgt im ASCII-Format mit folgenden Übertragungsparametern: Das Datenformat ist: 8 Datenbit, 1 Stopbit, gerade Parität (8,1,e) Das Gerät antwortet bei Befehlseingabe mit: Ausgabe (z.B. dem Messwert) + CR (Carriage Return, ASCII 13), bei reinen Eingabebefehlen mit „ok“ + CR. Jeder Befehl beginnt mit der 2-stelligen Geräte-Adresse AA (z.B. „00“). Darauf folgen 2 kleine Buchstaben (z.B. „em“ für Emissionsgrad) gefolgt von ggf. erforderlichen ASCIIParametern „X“ und CR als Abschluss. Wird dieser Parameter „X“ weggelassen, so gibt das Gerät den momentan eingestellten Parameter zurück. Ein „?“ nach den 2 kleinen Buchstaben gibt die jeweiligen Grenzen aus (nur bei Parametrierbefehlen, nicht bei Abfragebefehlen). Bsp: Eingabe: “00em“ + <CR> Es wird der eingestellte Emissionsgrad des Gerätes mit der Adresse 00 zurückgegeben Antwort: “0970“ + <CR> bedeutet Emissionsgrad = 0,970 oder 97,0% Beschreibung Messwert lesen: Befehl AAms Messwert mehrf. lesen. AAmsXXX Emissionsgrad: AAemXXXX UmgebungstemperaAAutXXXX turkompensation: Erfassungszeit t90: AAezX Löschzeiten Maximalwertspeicher: AAlzX Externes Löschen: AAlx 48 Parameter Ausgabe: XXXXX (dez., in 1/10 °C oder °F) (77770 = Gerätetemperatur zu hoch oder Chopperrad dreht mit falscher Drehzahl 88880 = Temp.-Overflow) XXX = 000...999 (XXX = Anzahl Messwerte) XXXX = (0010 ... 1000‰) (dezimal) XXXX = Wert der Umgebungstemperatur, 4-stellig, hex XXXX z.B. FFEC entsprechen -20 Grad - 99dez = FF9Dhex bedeutet: Automatik, keine manuelle Kompensation X = 0 ... 6 (dezimal) 0 = Eigenzeitkonstante des Geräts 1 = 0,01 s 4 = 1,00 s 2 = 0,05 s 5 = 3,00 s 3 = 0,25 s 6 = 10,00 s X = 0 ... 9 (dez.) 0 = Maximalwertspeicher aus 1 = 0,01 s 4 = 1,00 s 7 = extern Löschen 2 = 0,05 s 5 = 5,00 s 8 = autom. Löschen 3 = 0,25 s 6 = 25,00 s 9 = Hold Maximalwertspeicher löschen (nur bei lz = 7, extern) IMPAC-Pyrometer IP 140 Analogausgang: Grundmessbereich lesen: AAasX AAmb Teilmessbereich lesen: AAme Teilmessbereich setzen: Geräteadresse: AAm1XXXXYYYY Baudrate: AAbrX Umschaltung °C / °F Wartezeit: Geräteinnentemperatur: Maximale GeräteInnentemperatur: Fehlerstatus: AAfhX AAtwXX AAgt AAtm Laserpilotlicht: AAlaX Schnittstelle lesen: Tastatur sperren: AAin AAlkX Parameter lesen: AApa Gerätename lesen: Seriennummer lesen: Gerätetyp / Softwareversion: AAna AAsn AAve Softwareversion ausführlich: Bestellnummer lesen: AAvs Hinweis: AAgaXX AAfs AAbn X = 0...1 0 = 0...20 mA 1 = 4...20 mA Ausgabe: XXXXYYYY (hex 8-stellig, °C oder °F) XXXX = Messbereichsanfang YYYY = Messbereichsende Ausgabe: XXXXYYYY (hex 8-stellig, °C oder °F) XXXX = Messbereichsanfang YYYY = Messbereichsende XXXX (hex 4-stellig) Messbereichsanfang (°C oder °F) YYYY (hex 4-stellig) Messbereichsende (°C oder °F) XX = (00 ... 97) 00 ... 97 = einstellbare Geräteadressen 99 = Globale Adresse mit Antwort 98 = Globale Adresse ohne Antwort (nur Einstellbefehle!) X=1...6 oder 8 (dez.) 1 = 2400 Baud 5 = 38400 Baud 2 = 4800 Baud 6 = 57600 Baud 3 = 9600 Baud (7 nicht vorhanden) 4 = 19200 Baud 8 = 115200 Baud Ausgabe: X = 0: Anzeige in °C; X = 1: Anzeige in °F XX = 00 ... 99 (dezimal, in Bitzeit der aktuellen Baudrate) Ausgabe: XXX (dez. 000 ... 099°C, 032 ... 210°F Ausgabe: XXX (dez. 000 ... 099, immer in °C) Ausgabe: XX; XX=00…FF (00 = kein Fehler) (01…FF: Fehlercode für LumaSense-Service) X = 0 Pilotlicht ausschalten X = 1 Pilotlicht einschalten Ausgabe: 1 oder 2 (1 = RS232, 2 = RS485) X = 0 ... 3 1 = Sperre lk1, aufhebbar mit Befehl lk0 oder power off-on 0 = Aufheben von Sperre lk1 3 = Dauerhafte Sperre lk3, nur aufhebbar mit Befehl lk2 2 = Aufheben von Sperre lk3 Ausgabe 11-stellig, dezimal: Stellen 1 und 2 (10...99 oder 00): Emissionsgrad Stelle 3 (0 ... 6): Erfassungszeit Stelle 4 (0 ... 8): Löschzeit Max.wertspeicher Stelle 5 (0 ... 1): Analogausgang Stellen 6 und 7: (00 ... 98): Gerätetemperatur Stellen 8 und 9 (00 ... 97): Geräteadresse Stelle 10 (0 ... 6 oder 8): Geräte-Baudrate Stelle 11 (0,1, 3): Tastensperre Ausgabe: „IP 140“ (16 ASCII-Zeichen) Ausgabe: XXXXX (dez. 5-stellig) Ausgabe: XXYYZZ (6-stellig dezimal) XX = 63 (IP 140) YY = Monat der Softwareversion ZZ = Jahr der Softwareversion tt.mm.yy XX.YY tt = Tag; mm = Monat; yy = Jahr; XX.YY = Softwareversion Ausgabe: XXXXXX (hex 6-stellig) Mit dem Buchstaben „l“ ist das kleine „L“ gemeint. Ergänzender Hinweis zur RS485-Schnittstelle: Anforderung an das Master-System bei Halb-Duplex-Betrieb: 1. Nach einer Anfrage ist der Bus innerhalb einer Übertragungszeit von 3 Bits freizuschalten (einige ältere Interfaces sind dafür nicht schnell genug). 2. Die Antwort des Pyrometers erfolgt spätestens nach 5 ms. 3. Erfolgt keine Antwort, so liegt ein Parity- oder Syntaxfehler vor und die Anfrage muss wiederholt werden. 49 IMPAC-Pyrometer IP 140 15 Bestellnummern 15.1 Bestellnummern Geräte Messbereich MB 4: 50 MB 5.5: 75 MB 7: 100 MB 12: 160 MB 13: 200 ... 400°C ... 550°C ... 700°C ... 1200°C ... 1300°C Mit Laserpilotlicht 3 875 500 3 875 520 3 875 540 3 875 560 3 875 580 Mit Durchblickvisier 3 875 510 3 875 530 3 875 550 3 875 570 3 875 590 Bestellhinweis: Bei Bestellung bitte eine Vario-Optik (0, 1, 2 oder 3) Ihrer Wahl mit angeben (siehe auch Kap. 7, Optik). Ein Anschlusskabel ist im Lieferumfang nicht enthalten und muss separat bestellt werden. 15.2 Bestellnummern Zubehör 3 848 210 3 848 250 3 848 260 3 848 270 Optik 0-P (70 ... 90 mm) Optik 1-P (105 ... 150 mm) Optik 2-P (200 ... 440 mm) Optik 3-P (345 ... 4300 mm) 3 820 340 3 820 530 3 820 540 3 820 830 3 820 840 3 820 550 3 820 330 3 820 500 3 820 510 3 820 810 3 820 820 3 820 520 3 820 740 3 820 750 Anschlusskabel, 5 m lang, Winkelsteckverbinder Anschlusskabel, 10 m lang, Winkelsteckverbinder Anschlusskabel, 15 m lang, Winkelsteckverbinder Anschlusskabel, 20 m lang, Winkelsteckverbinder Anschlusskabel, 25 m lang, Winkelsteckverbinder Anschlusskabel, 30 m lang, Winkelsteckverbinder Anschlusskabel, 5 m lang, gerader Steckverbinder Anschlusskabel, 10 m lang, gerader Steckverbinder Anschlusskabel, 15 m lang, gerader Steckverbinder Anschlusskabel, 20 m lang, gerader Steckverbinder Anschlusskabel, 25 m lang, gerader Steckverbinder Anschlusskabel, 30 m lang, gerader Steckverbinder Anschlusskabel, 5 m lang, gerader Stecker, temperaturbeständig bis 200°C Anschlusskabel, 5 m lang, Winkelsteckverbinder, temperaturbeständig bis 200°C 3 834 280 3 834 270 3 835 230 3 837 290 3 835 060 3 834 140 3 837 240 3 835 450 3 843 520 Justierbarer Montagewinkel Kugelgelenk-Halterung Blasaufsatz Vollmantel-Kühlgehäuse, Edelstahl Blasaufsatz für Kühlgehäuse Schwere Kugelgelenkhalterung für Vollmantel-Kühlgehäuse Kühlplatte 90°-Umlenkspiegel mit Quarzglasfenster Robuster Schwenkaufsatz SCA 140, (Abtastwinkel 0 ... 12°, Scanfrequenz 1 ... 5 Hz), mit Quarzglas-Fenster Blasaufsatz für Schwenkaufsatz SCA 140 3 835 290 3 852 290 3 852 550 3 890 640 3 890 650 3 890 560 3 890 570 3 890 520 3 890 530 3 826 500 Netzteil für C/Z-Schienenmontage NG DC (100...240 V AC ⇒ 24 V DC, 1 A) Netzgerät NG 2D, wie NG 0D: zusätzlich mit 2 Grenzkontakten LED-Digital-Anzeige DA 4000-N LED-Digital-Anzeige DA 4000: mit zwei Grenzkontakten LED-Digitalanzeige DA 6000-N: mit Parametrierfunktion für digitale IMPAC-Pyrometer; RS232-Schnittstelle LED-Digitalanzeige DA 6000-N: mit Parametrierfunktion für digitale IMPAC-Pyrometer; RS4852-Schnittstelle LED-Digitalanzeige DA 6000; wie DA 6000-N, zusätzlich mit zwei Grenzkontakten und analogem Ein- und Ausgang, RS232-Schnittstelle LED-Digitalanzeige DA 6000; wie DA 6000-N, zusätzlich mit zwei Grenzkontakten und analogem Ein- und Ausgang, RS485-Schnittstelle HT 6000, tragbares Handterminal zum Parametrieren von stationären Pyrometern 3 826 510 PI 6000: PID-Programmregler, extrem schnell, für digitale IMPAC-Pyrometer 3 825 430 Ι-7520, Schnittstellenwandler RS232 ⇔ RS485 50 IMPAC-Pyrometer IP 140 Stichwortverzeichnis A Abmessungen.....................................................31 Abschirmung.......................................................33 Adresse...............................................................41 Analogausgang...................................................41 Anschlusskabel...................................................32 Austausch der Optiken .......................................47 Auswertegeräte, zusätzliche...............................35 B Baudrate .............................................................42 Bereitschaftskontakt ...........................................33 Bestellnummern..................................................50 Bestimmungsgemäße Verwendung ...................32 D Datenformat UPP® ..............................................48 Durchblickvisier...................................................36 E Elektromagnetische Verträglichkeit ....................33 Emi: AutoFind .....................................................44 Emissionsgrad ....................................................39 Erfassungszeit ....................................................40 Error-Status ........................................................42 F Farb-Balken-Messung ........................................44 Farbkennzeichnung ............................................47 Fehlerdiagnose ...................................................47 G Lieferumfang ...................................................... 32 Löschzeit............................................................ 40 M Maximalwertspeicher ......................................... 34 Messabstand...................................................... 37 Messung (Farb-Balken) ..................................... 44 Messung (Online-Grafik).................................... 45 O Online-Grafik-Messung...................................... 45 Optik................................................................... 36 P Parameterbeschreibung / Einstellungen............ 39 Pin-Belegung des Flanschsteckers ................... 33 Pyrometer-Parameter ........................................ 43 S Schnittstelle............................................ 34, 38, 42 Schnittstellenbefehle.......................................... 44 Sicherheit ........................................................... 32 Software............................................................. 42 T Tabelle (Auswertung)......................................... 45 Technische Daten .............................................. 30 Teilmessbereich................................................. 41 Temperaturanzeige............................................ 42 Test .................................................................... 44 Testfunktion ....................................................... 39 Transport, Verpackung, Lagerung ..................... 46 TXT-Datei, Ausgabe .......................................... 46 Geräteeinstellungen............................................38 Glühlampenlicht ..................................................48 Grafik-Ausgabe (Auswertung) ............................46 Grundeinstellungen.............................................43 U H V Hold-Funktion .....................................................34 Vario-Optik ......................................................... 36 I W i12 .................................................................45, 46 InfraWin ..............................................................42 Innentemperatur, maximale................................42 Installation, elektrische .......................................33 Installation, mechanische ...................................35 IR-Rechner ...................................................37, 46 Warnschilder ...................................................... 32 Wartezeit............................................................ 42 Wartung ............................................................. 47 Werkseinstellungen............................................ 39 L Umgebungstemperaturkompensation................ 40 UPP®-Datenformat............................................. 48 Z Zeitintervall......................................................... 46 Zubehör (optional).............................................. 35 Laserpilotlicht................................................32, 36 51 LumaSense Technologies 3033 Scott Blvd. Santa Clara, CA 95054-3316 Tel.: +1 408 727-1600 Fax: +1 408 727-1677 Internet: www.lumasenseinc.com E-mail: [email protected] [email protected] LumaSense Technologies Inc. 16 Thornton Road Oakland, NJ 07436 Tel.: +1 201 405-0900 Fax: +1 201 405-0090 Internet: www.mikroninfrared.com E-mail: [email protected] LumaSense Technologies GmbH Kleyerstr. 90 D-60326 Frankfurt/Main Tel.: +49 (0)69 973 73-0 Fax: +49 (0)69 973 73-167 Internet: www.lumasenseinc.com E-Mail: [email protected] 3 875 227-c ">
Advertisement