ART-L3 CURING LIGHT Pro Fiber Type User Manual

ART-L3 CURING LIGHT Pro Fiber Type User Manual
ART-L3 CURING LIGHT Pro
Fiber Type
User Manual
BONART CO. LTD.
Tel: 909-594-8066
Fax:909-594-7266
Website:http://www.bonartmed.com
Email: [email protected]
Jorgen Kruuse A/S
Marslev Byvej 35, DK-5290, Marslev,
Denmark
COPYRIGHT© 2008-12 BONART CO. LTD.
NOTICE
Please keep the contact points
clean to avoid bad connection.
Power saving & charging.
The ART-L3 LED cordless Curing Light System is
designed to automatically turn to sleep mode after the
unit has been idle for a few minutes to be energy efficient.
Insert the handpiece into the charging unit after use.
-1-
Do not operate.
Do not operate 20-second curing cycle for
more than 3 times continuously to avoid
LED from overheating.
Keep clean
Check & clean LED area after each use.
-2-
NOTE:
As an alternative option for user to
use the system while in low battery,
Is to plug the power cord adapter to
the system.
-3-
BONART Co., Ltd. owns all rights to this publication. This publication should be used solely for the purpose of
reference, operation, maintenance, and repair of BONART equipment. Not any part of this document may be
reproduced or disseminated in any forms or any means, for any other purpose. In the event of inadvertent or
deliberate publication, BONART intends to enforce its rights to this manual under the copyright law. No one
may copy the materials presented in this manual without the express written permission of BONART Co., Ltd.
Product specifications, as well as information presented in this manual, are subject to change without prior
notice.
PROPERTY OF BONART CO., LTD.
ALL RIGHTS RESERVED
ART is a registered trademark of BONART Co., Ltd. & its affiliates.
Document No. L3F1001 Rev.1.0 (12/08)
Printed in Taiwan, R.O.C.
-4-
Operator Safety:
Please read carefully and completely before operating unit.
ENGLISH VERSION
ART-L3 LED CURING LIGHT Pro
Only properly trained personnel are designated to use this system as described in this manual.
Only qualified personnel shall carry out any kinds of adjustment, maintenance and repair of the
equipment.

WARNING!
WARNING:
The intended use of ART-L3 is for light curing only; it is not suitable for work
in full power more then 2 minutes continuously.
WARNING:
For continuous operation, please wait for at least 30 seconds between each
curing cycle.
WARNING:
Use only specified accessories.
-5-
Table of Contents
NOTICE
1
Copyright
4
Operator Safety
5
Contents
6
Installation and Operations
7
System Maintenance
8
Specifications
11
-6-
Installation and Operations:
1.
2.
3.
4.
5.
6.
7.
8.
9.
Plug the optic fiber into the nozzle of the handpiece.
Attach the adaptor into the wall plug.
Attach the eye protecting cap to the tip of the optic fiber or put on blue light resistant goggles
as a protection of user’s eyes.
Adjust the curing time by pressing the key “T”. Two timing, 10, 20 second, can be selected,
LED indicates the selected curing time.
Select curing mode by pressing the key “M”. There are 3 modes: Fast mode, Pulse mode and
Ramp mode can be choose. Mode LED indicator indicates which curing mode was selected.
Adjust the optic fiber direction to suit different restoration position in mouth. The optic fiber
can be rotated by 360°.
Start curing: Press the key “S” to activate the LED output.
Curing will be ended automatically or press key “S” again to end.
Press the key “S” to terminate the sleeping mode.
-7-
System Maintenance:
1. Prevention of cross infection, cleaning, disinfection and storage
It is mandatory that doctors need to wear sterilized gloves during these procedures at all time to
avoid any possibilities of incomplete sterilization and/or getting infected.
After use, remove the optic fiber and eye protecting cap. Clean and disinfect body with
commercial alcohol based surface disinfecting solution (Dürr FD 322, Henkel Incidin Liquid,
S&M Mikrozid Liquid, B. Braun Meliseptol rapid). Keep other solvents or flammable liquids as
well as intense sources of heat away from the unit as they may damage its plastic housing.
Always safeguard charger, Body against moisture as this may cause electrical short-circuit or
dangerous malfunction.
After each usage, there will be saliva or other debris left on the light guide fiber and eye
protecting cap, it is necessary to clean the light guide fiber and eye protecting cap with a
supersonic cleaner first. This can be done manually by scrubbing with a brush or by use of an
ultrasonic cleaner with a solution of detergent and water. After scrubbing. The optic fiber and eye
protecting cap should be rinsed thoroughly with water to remove all detergent and dried the rinsed
light guide fiber and eye protecting cap, and finally, put the light guide fiber and eye protecting
cap in a bag and then put it into a medical-equipment autoclave and sterilize the fiber at
260℉(127℃) for 30 minutes or as recommended by the manufacturer of the particular sterilizer
used.
-8-
2. Light output control
Make sure that the optic fiber is clean and scratch-free, otherwise light output will be reduced and
may be insufficient for proper curing of the material.
Control of light efficiency is recommended (e.g. by using CureRite™ light intensity testing
device).
The light intensity measured by CureRite testing device should exceed 1000 mW/cm2.
3. Battery maintenance
The battery needs approximately 2 hours to be fully recharged.
While the battery is recharging, the indicator light will flash.
If fully charged, the battery has a capacity of approximately 180 ten-second cures (approximately
30 minutes operating time). Under most conditions of use, the device only needs to be recharged
1.5 hours.
When battery is low, the indicator light will come on and flash. There should be approximately
80 ten-second cures left in the reserved circuit to complete of the light curing treatment. Light
output is not reduced during this period. After these 80 ten-second cures, the device will
automatically switched off.
Caution: To gain better performance of the battery, the new battery must be fully charged for
2.5 hours.
4.Charging method
A. (1)Plug in charging Stand. (2)Insert the Handpiece into the charging stand refer Fig.3 .
B. Plug in Handpiece refer Fig.2 .
-9-
Components
1. Light guide
2. Handpiece
3. Eyes protection cap
4. Power supply
5. Charger
Fig.1
Fig.2
Fig.3
- 10 -
Fig.4
Specifications:
Power:
LED Curing Light:
Optical specifications: wavelength 450~470nm
Flux intensity: 1000mW/cm² (Min.)
LED power consumption: 3watt
Equipment class: Class II
Protection from electric shock: Type BF
Operation environment
Temperature:0℃~40℃
Relative Humidity:
10% ~ 90% (non-condensing)
Power Adaptor:
Input: 100~240VAC, 50~60Hz,0.6A
Output: 5.0VDC/1.0A, 5W Max.
Battery:
Type: Li-ion
Voltage:DC.3.6V
Capacity: 700mAh
Handpiece will operate approximately 30 minutes
time starting upon complete recharging.
Time for battery recharge: approximately 2 hours
Autoclavable optic fiber light guide : 8mm
Transport and storage conditions
Temperature:0℃~60℃
Relative Humidity:
10% ~ 90% (non-condensing)
Atmospheric pressure:860~1060 hPa
- 11 -