Coviden Puritan Bennett 560 Ventilator Service Manual
Advertisement
Advertisement
Service Manual
Puritan Bennett
TM
560 Ventilator
Copyright information
COVIDIEN, COVIDIEN with logo, the Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. All other brands are trademarks of a Covidien company.
© 2012, 2014, 2015, 2017 Covidien.
The information contained in this manual is the sole property of Covidien and may not be duplicated without permission. This manual may be revised or replaced by Covidien at any time and without notice. You should ensure that you have the most current applicable version of this manual; if in doubt, contact Covidien Technical
Support department or visit the Covidien product manual web page at: http://covidien.com/rms/sales-support/product-manuals
While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals. Covidien’s sole responsibility with respect to the ventilator, and its use, is as stated in the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Covidien’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Covidien has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.
Contents
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
Service Manual i
Contents
Exhalation Valve (Internal Valve) Pressure Sensor. . . . . . . . . . . . . . . . . . . . . . . 3–7
Internal Battery Charging and Discharging . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–11
Power Management PCBA Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 3–14
Ventilator Compartment Temperature Sensor. . . . . . . . . . . . . . . . . . . . . . . . 3–15
for Various Oxygen and Ventilator Settings. . . . . . . . . . . . . . . . . . . . . . 3–17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–1
How to Pause the Audible Portion of an Alarm . . . . . . . . . . . . . . . . . . . . . . . . . 4–3
ii Service Manual
Contents
Troubleshooting Ventilation Alarm Messages. . . . . . . . . . . . . . . . . . . . . . . . . 4–12
Troubleshooting Miscellaneous Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–31
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
Adhesive Feet Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–13
Exhalation Block Removal, Cleaning, and Replacement . . . . . . . . . . . . . . . 5–14
Carrying Handle Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 5–19
Exhalation Solenoid Valve Removal and Replacement . . . . . . . . . . . . . . . . 5–20
Inspiratory Block Removal, Cleaning and Replacement. . . . . . . . . . . . . . . . 5–30
Connector and Cable Removal and Replacement . . . . . . . . . . . . . . . . 5–32
Buzzer PCBA Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–33
Exhalation Conical Fitting Removal and Replacement. . . . . . . . . . . . . . . . . 5–35
Captive Knob Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–36
Oxygen Solenoid Valve Removal and Replacement . . . . . . . . . . . . . . . . . . . 5–37
Turbine Assembly Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . . 5–39
Turbine Control PCBA Removal and Replacement . . . . . . . . . . . . . . . . . . . . 5–41
Turbine Foam Pads Removal and Replacement . . . . . . . . . . . . . . . . . . . . . . . 5–43
Power Management PCBA Removal and Replacement . . . . . . . . . . . . . . . . 5–45
Battery Connection PCBA Removal and Replacement . . . . . . . . . . . . . . . . . 5–47
Remote Alarm (Nurse Call) Cable Removal and Replacement . . . . . . . . . . 5–47
Oxygen Connector/Tube Assembly Removal and Replacement . . . . . . . 5–51
Service Manual iii
Contents
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1
Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–37
Sensor Detection Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–38
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–1
iv Service Manual
Figures
Service Manual
Ventilator with Single Limb Circuit and Test Lung. . . . . . . . . . . . . . . . . . . . . . 3–8
Ventilator with Dual Limb Patient Circuit and Test Lung. . . . . . . . . . . . . . . . 3–9
FiO2 for various Oxygen and Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . 3–18
Figure 4-1.
Alarm Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–3
Alarm Settings Menu During Audio Paused Period. . . . . . . . . . . . . . . . . . . . . 4–4
Alarm Settings Menu During Alarm Paused Condition . . . . . . . . . . . . . . . . . 4–5
Alarm Log Screen With and Without Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–6
Alarm Log Screen With and Without Alarm Information. . . . . . . . . . . . . . . . 4–7
Identification of Short and Long Ends of Spring. . . . . . . . . . . . . . . . . . . . . . . 5–11
Pneumatic Tubing Connections (top view) . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–22
CPU PCBA Tubing Connections (underside view) . . . . . . . . . . . . . . . . . . . . . 5–25
v
Figures vi
Removing the 3V Battery (LCD panel removed) . . . . . . . . . . . . . . . . . . . . . . 5–30
Cable Routing and Tape Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–34
Exhalation Conical Fitting and Baffle Conduit Fittings . . . . . . . . . . . . . . . . 5–37
Power Management/turbine Control Cable Detail . . . . . . . . . . . . . . . . . . . . 5–41
Figure 5-29.
Turbine Control PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–43
Power Management PCBA Mounting Screw Locations . . . . . . . . . . . . . . . 5–46
Remote Alarm Cable and Oxygen Tubing Assembly . . . . . . . . . . . . . . . . . . 5–49
Aligning Harness Connector Tab with Housing Notch . . . . . . . . . . . . . . . . 5–51
Orientation of the Connector and Tightening Locking Nut. . . . . . . . . . . . 5–51
Remote Alarm Harness Installed into Ferrite. . . . . . . . . . . . . . . . . . . . . . . . . . 5–52
Exhalation Flow Sensor Calibration Test Setup . . . . . . . . . . . . . . . . . . . . . . . 6–20
Pressure Calibration Points in Turbine Performance Test . . . . . . . . . . . . . 6–24
Test setup for Oxygen Solenoid Functional Test . . . . . . . . . . . . . . . . . . . . . . 6–26
Sensor Calibration Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–38
Sensor Cable Disconnected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–40
Remote Alarm Test Box Connected to Ventilator (rear view) . . . . . . . . . . 6–42
Pneumatic Components and Connection Diagram . . . . . . . . . . . . . . . . . . . 7–11
Service Manual
Figures
Power Management PCBA Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–3
Exhalation Solenoid Valve During Inspiration and Exhalation. . . . . . . . . . . A–4
Service Manual vii
Figures
This page intentionally blank viii Service Manual
Tables
Service Manual
Default Preferences (after software download) . . . . . . . . . . . . . . . . . . . . . . . 2–17
Table 2-10.
Electromagnetic Immunity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–21
Electromagnetic Immunity — Conducted and radiated RF . . . . . . . . . . . . 2–22
Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the Ventilator . . . . . . . . . . . . . . . . . . . . . . 2–23
On-screen symbols, abbreviations, and internal labels . . . . . . . . . . . . . . . . 2–40
Pneumatic Component Operating Ranges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–4
Table 6-1.
Technical Fault Error Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . 4–10
Ventilation Alarm Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–14
Miscellaneous Symptom Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . 4–31
Specifications for Measurement Check Verification . . . . . . . . . . . . . . . . . . . . 5–7
Testing and Calibration Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–2
ix
Tables
Major System Component Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–2
Electrical System Component Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–6
Pneumatic System Component Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . 7–9
Turbine Assembly Component Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . 7–12
x Service Manual
Service Manual
Preface
About this Manual
This manual provides information needed to service the Puritan Bennett™ 520 Ventilator. This manual is intended for use by factory-trained biomedical engineering technicians or personnel with equivalent experience and training in servicing this type of equipment. It is recommended that the user complete the Puritan Bennett™ 520 Ventilator training class.
Note:
This manual is written assuming English (US) language is selected in ventilator SETUP (
messages will be different depending upon the language selected.
While this manual covers the ventilator configurations currently supported by Covidien, it may not be all-inclusive and may not be applicable to your ventilator. You should ensure you have the most current applicable version of this manual; if in doubt, contact Covidien or visit the Puritan Bennett product manual web page at: http://covidien.com/rms/sales-support/product-manuals. Contact your local representative for questions regarding the applicability of the information.
Qualification of Personnel
Installation and maintenance of the device must be made by authorized and trained personnel. In particular, training for the handling of products sensitive to electrostatic discharges must include the use of Electrostatic
Discharge (ESD) protection devices and knowledge of the ESD symbol’s meaning, as well as using original spare parts and respecting quality assurance and traceability rules approved by Covidien.
Electromagnetic Susceptibility
The Puritan Bennett™ 520 Ventilator complies with the requirements of IEC 60601-1-2:2007 (EMC Collateral Standard), including the E-field susceptibility requirements at a level of 20 volts per meter at frequencies up to 1 GHz, and 10 volts per meter at frequencies between 1 GHz and 2.5 GHz, and the ESD requirements of this standard.
However, even at this level of device immunity, certain transmitting devices (cellular phones, walkie-talkies, cordless phones, paging transmitters, etc.) emit radio frequencies that could interrupt ventilator operation if located in a range too close to the ventilator. It is difficult to determine when the field strength of these devices becomes excessive.
Practitioners should be aware that radio frequency emissions are additive, and that the ventilator must be located a sufficient distance from transmitting devices to avoid interruption. Do not operate or store the ventilator in the presence of strong electromagnetic fields such as a magnetic resonance imaging (MRI) environment.
describes possible ventilator alarms and what to do if they occur. Contact an authorized service representative or you local durable medical equipment (DME) representative in case of interrupted ventilator operation and before
P-1
Preface relocating any life support equipment.
WARNING
Accessory equipment connected to the power receptacle, analog, and digital interfaces must be certified according to IEC 60601-1. Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Any person who connects additional equipment to the power receptacle, signal input part, or signal output part of the Puritan Bennett™ 520
Ventilator configures a medical system, and is therefore responsible for ensuring that the system complies with the requirements of the system standard IEC 60601-1-1. If in doubt, consult Puritan Bennett Technical Support at 1.800.255.6774 or your local representative.
Warranty
Information regarding your product warranty is available from your sales representative or Covidien.
Extended Service
The Puritan Bennett™ 560 Ventilator offers extended service contracts/warranties for purchase when the ventilator is purchased. Please contact your local Covidien sales or service representative for additional information.
Customer Assistance
For online technical support, visit the SolvIT
SM
Center Knowledge Base by clicking the link at www.medtronic.com/covidien/support/solvitcenter-knowledge-base/. Here, you will find answers to frequently asked shooting help or to obtain service for your ventilator, call your local
Puritan Bennett Technical Support center. If you require further assistance, contact your local Puritan Bennett representative.
questions about the Puritan Bennett™ 520 Ventilator, and other
Puritan Bennett products 24 hours a day, 7 days a week. For trouble-
P-2 Service Manual
Year of Manufacture
Year of Manufacture
The year of manufacture for ventilators is indicated on the bottom of the ventilator as shown here.
2007
Manufacturer
The manufacturer and authorized representative are shown below.
Covidien llc
15 Hampshire Street
Mansfield, MA 02408
Covidien Ireland Limited, IDA Business & Technology Park, Tullamore.
Service Manual P-3
Preface
Service Centers
The following table lists Covidien international service centers.
Table 1-1. Covidien Service Centers
Covidien Argentina
Vedia 3616
Buenos Aires
Argentina
Tel: 5411 4863 5300
Fax: 5411 4863 4142
Covidien Belgium BVBA/SPRL
Generaal De Wittelaan 9/5
2800 Mechelen
Belgium
Tel: +32 220 08260
Fax: +32 270 06690
Covidien China
1st Floor, Covidien Plaza,
Building 14
No. 99 Tian Zhou Rd
CAOHEJING High Tech Park
Xu Hui District
Shanghai 200233
P.R. China
Tel: +86 4008 21 6699
Fax: +86 2154 4511 18
Covidien Danmark A/S
Langebrogade 6E, 4. sal
DK-1411 København
Danmark
Tel: +45 43 68 21 71
Fax: +45 43 31 48 99
Covidien Asia
Singapore Regional Service
Centre
15 Pioneer Hub, #06-04
Singapore 627753
Tel: 65 6578 5288
Fax: 65 6515 5260
Covidien Australia
52A Huntingwood Drive
Huntingwood, NSW 2148
Australia
Tel: +61 1800 350702
Fax: +61 2967 18118
Covidien Deutschland GmbH
Technisches Service Center
Raffineriestr. 18
93333 Neustadt / Donau
Germany
Tel: + 49 69 51 709670
Fax: + 49 69 29 9571608
Covidien ECE s.r.o.
Galvahiho 7 / A
821 04 Bratislava Slovakia
Slovenska Republika
Tel: +421 2 4821 4573
Fax: +421 2 4821 4501
Covidien Austria GmbH
Campus21
Europaring F09402
A-2345 Brunn am Gebirge
Österreich
Tel: +43 1 20609 1143
Fax: +43 1 20609 2457
Covidien Brazil
Praça Agrícola La Paz Tristante,
121
Osasco – São Paulo / CEP
06276-035
São Paulo, SP
Brasil 04578-000
Tel: +55 11 2187 6543
Fax: 2187 6380
Covidien Canada
19600 Clark Graham
Baie d'Urfe, QC, H9X 3R8
Canada
Tel: 1 877 664 8926, option
2
Fax: 1 514 695 7534
Covidien Chile
Lo Boza 107
Pudahuel
Santiago
Chile
Tel: 562 2739 3000
Fax: 562 783 3149
Covidien Colombia
Edificio Prados de la Morea
Carretera Central Del Norte
(Cra 7a) Kilometro 18,
Chia-Cundinamarca
Bogota, Colombia
Tel: 571 668 3777
Fax: 571 668 3777,ext. 181
Covidien Costa Rica
Global Park, Parkway 50
La Auroa de Heredia
40104
Costa Rica
Tel: 506 2293 4854
Fax: 506 2239 9108
Covidien ECE s.r.o.
Organizacní Složka
Prosek, Prosecká 852/66
190 00 Praha 9
Tel: +420 241 095 735
Fax: +420 239 016 856
Covidien Finland Oy
Rahtitie 3
FI-01530 Vantaa
Finland
Tel: +358 9725 192 88
Fax: +358 9725 192 89
Covidien France SAS
Parc d’affaires Technopolis
Bat.Sigma, 3 Avenue du
Canada
LP 851 Les Ulis
91975 Courtaboeuf Cedex
France
Tel: +33 1 57 32 35 10
Fax: +33 1 57 32 70 10
Covidien Hong Kong
Unit 12-16, 18/F
BEA Tower
Millennium City 5
4187 Kwun Tong Road
Kwum Tong,
Kowloon, Hong Kong
Tel: + 852 2574 3251
Fax: + 852 2838 0749
Covidien India
10th Floor Building No 9B
DLF Cyber City Phase III
Gurgaon
Haryana - 122002
India
Tel: + 91 1244 709800
Fax: + 91 1244 206850
Covidien ECE s.r.o.
Magyarországi Fióktelepe
Mariassy u. 7
1095 Budapest
Hungary
Tel: + 36 1880 7975
Fax: + 36 1778 9459
P-4 Service Manual
Service Centers
Table 1-1. Covidien Service Centers (Continued)
Covidien Ireland
Block G, Ground Floor,
Cherrywood Business Park,
Loughlinstown
County Dublin, Ireland
Tel: +353 0 1.4073173
Fax: +353 0 1.9075668
Covidien Israel
5 Shacham St.
North Industrial Park
P.O. Box 3069
Caesarea, Israel 3088900
Tel: +972 4.6277388
Fax:+972 6.6370053
Covidien Korea
5F, Hibrand Living Gwan, #215
Yangjae-Dong,
Seocho-Gu
Seoul, Korea
Tel: +822 6196 5459
Fax: +822 6196 5498
Covidien Mexico
Av. Insurgentes Sur 863, Pisos
15 y 16
Col. Nápoles
Del. Benito Juarez
Mexico, D.F. 03810 Mexico
Tel: 5255 5804 1524
Fax: 5255 5536 1326
Covidien Italia S.p.A.
Via S. Bovio 3
20090 S. Felice di Segrate
(MI)
Italy
Tel: +39 02 91483320
(option 3)
Fax: +39 02 91294863
Covidien Nederland BV
Hogeweg 105
5301 LL Zaltbommel
Nederland
Tel: +31 202061470
Fax: +31 707 709229
Covidien Japan Inc.
Technical Support Center
83-1, Takashimadaira
1-Chome
Itabashi-ku, Tokyo 175-0082
Japan
Tel: +81 (0) 3 6859 0120
Fax: +81 (0) 3 6859 0142
Covidien New Zealand
Cnr Manu Tapu Dr & Joseph
Hammond Pl.
Auckland Airport
New Zealand
Tel: + 64 508 489 264
Covidien Norge AS
Postboks 343
N1372 Askerr
Norway
Tel: +47 2415 9887
Fax: +47 2302 4955
Covidien Panama
Parque Industrial Costa del
Esta
Calle Primera, Edifio #109
Panama City, Panama
Tel: 507 264 7337
Fax: 507 236 7408
Covidien Polska
Al. Jerozolimskie 182
02-222 Warszawa
Polska
Tel: +48 22 279 04 05
Fax: +48 22 279 04 03
Covidien Portugal Lda.
Estrada do Outeiro de Polima,
Lote 10-1° Piso Abóboda
2785-521 S.Domingos de
Rana
Portugal
Tel: +35 21 761 62 44
Fax: +35 800 781385
Covidien Puerto Rico
Palmas Industrial Park
Road 869 Km 2.0 Bdlg. #1
Cataño, PR 00962
Tel: 787 993 7250
Ext. 7222 & 7221
Fax: 787-993-7234
Covidien Russia
53 bld. 5 Dubininskaya Street
Moscow
Russia 119054
Tel: +70 495 933 64 69
Fax: +70 495 933 64 68
Covidien Spain S.L.
Servicio Técnico
WTC Almeda Park
Plaça de la Pau, S/N - Edif. 7, 3ª
Planta
08940 Cornellá de Llobregat
Barcelona, Spain
Tel: +34 91 275 48 54 (option
3)
Fax: +34 91 276 89 33
Covidien Sverige AB
Box 54
17174 Solna
Sweden
Tel: +46 8 517 615 73
Fax: +46 8 502 521 10
Covidien UK
Unit 2,
Talisman Business Park
London Road, Bicester
OX26 6HR, United Kingdom
Tel: +4420 3027 1757
Fax: +4420 3684 8869
Covidien USA
2101 Faraday Ave
Carlsbad, CA 92008
Tel: 1 800 255 6774 (option 4)
Email: [email protected]
Covidien Saglik A.S.
Maslak Mahallesi Bilim
Sokak No: 5, Sun Plaza Kat:
2-3
Sisli, Istanbul 34398
Turkey
Tel: +90 212 366 20 00
Fax: +90 212 276 35 25
Covidien South Africa
Corporate Park North
379 Roan Crescent
Randjespark Midrand,
South Africa
Tel: +27 115 429 500
Fax: +27 115 429 624
Covidien Switzerland
Roosstrasse 53
8832 Wollerau
Schweiz
Tel: +41 44 511 82 71
Fax: +4144 511 16 34
Covidien Thailand
99 Soi Rubia,
Sukhumvit 42 Road
13 - 14 Fl., Berli Jucker
Building
Prakanong, Klongtoey
Bangkok 10110, Thailand
Tel: +662 2073- 100
Fax: +662 657 6325
Service Manual P-5
Preface
This page intentionally blank
P-6 Service Manual
1 Safety Information
1.1
Definitions
This manual uses three indicators to highlight critical information: Warning, Caution, and
Note. They are defined as follows:
WARNING
Indicates a condition that can endanger the patient or the ventilator operator.
Caution
Indicates a condition that can damage the equipment.
Note:
Indicates points of particular emphasis, that make operation of the ventilator more efficient or convenient.
Please take the time to familiarize yourself with the following caveats as they cover safety considerations, special handling requirements, and regulations that govern the use of the Puritan Bennett™ 520 Ventilator.
1.2
Warnings
Fire Hazard Warnings
• To avoid a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition
(e.g., flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses.
• In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen supply, facility power, and battery.
• Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials such as grease or oils. (Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.)
• The ventilator must not be used with flammable anesthetic substances.
• Never expose any batteries to direct flame.
1-1
Safety Information
Warnings for General Repair Safety
• To ensure proper servicing and avoid the possibility of physical injury, only qualified personnel should attempt to service or make authorized modifications to the ventilator.
• To avoid an electrical shock hazard while servicing the ventilator, be sure to remove all power to the ventilator by disconnecting the power source and turning off all ventilator power switches before servicing. Follow accepted safety practices for electrical equipment when testing or making equipment adjustment, or repairs.
• When replacing the ventilator's internal battery, the ventilator must be disconnected from all external power supplies and turned off.
• When servicing the ventilator, be sure to familiarize yourself with, and adhere to all posted and stated safety warning and caution labels on the ventilator and its components, and on any service equipment and materials used. Failure to adhere to such warnings and cautions at all times may result in injury or property damage.
• To prevent personal injury or death, do not attempt any ventilator service while a patient, or other person, is connected to the ventilator.
• Never touch the ventilator's internal components, including the battery, and the patient simultaneously.
• To ensure proper performance of the ventilator, follow the preventive maintenance
.
• To prevent possible personal injury, always disconnect the oxygen source from the ventilator before service.
• Use all cleaning solutions and products with caution. Read and follow the instructions associated with the cleaning solutions you use to clean your ventilator. Use only those solutions listed in
• Do not clean any gas pathway with a liquid cleaner. Refer to
for allowable cleaning and disinfecting agents for use on other parts of the ventilator.
• Use personal protective equipment whenever exposure to toxic fumes, vapor, dust particles, blood pathogens, and other transmittable diseases and hazardous material can be expected. If in doubt, consult an environmental health and safety specialist or an
industrial hygienist before servicing the ventilator. Refer to Section 5.6.1
procedure.
• Chemicals from a broken LCD panel are toxic when ingested. Use caution when handling a ventilator with a broken display panel.
• To prevent possible damage to the power supply, remove the battery from the ventilator
(
) during disassembly and reassembly.
Warnings about Reducing Infection
• To reduce the risk of infection, wash your hands thoroughly before and after handling the ventilator or its accessories.
1-2 Service Manual
Service Manual
• Dirty or contaminated equipment is a potential source of infection. Clean the ventilator and its accessories regularly and systematically before and after each use and following any maintenance procedure to reduce the risks of infection. The use of a bacteria filter at the ventilator's outlet — or both ports if a double-limb circuit is used — is recommended. Refer to the “Cleaning” chapter in the Clinician’s manual.
• When handling any part of the ventilator, always follow appropriate infection control guidelines for handling infectious material.
• Covidien recognizes that cleaning, sterilization, sanitation, and disinfection practices vary widely among healthcare institutions. It is not possible for Covidien to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of cleaning, sterilization, and other practices carried out in the patient care setting.
• Covidien recommends that users of its products that require cleaning and sterilization/ disinfection consider the National Standards and Recommended Practices for Sterilization published by the Association for the Advancement of Medical Instrumentation (AAMI), as well as the following Center for Disease Control (CDC) publications: Guidelines for
Maintenance of In-use Respiratory Therapy Equipment and Guidelines for Prevention of
Nosocomial Pneumonia .
Warnings Before Using Equipment
• For a thorough understanding of ventilator operations, be sure to thoroughly read this manual and the Clinician’s manual before attempting to use the ventilator.
• The ventilator must be used only under the responsibility and on the prescription of a doctor.
• Check the ventilator periodically as outlined in this manual; do not use if defective.
Immediately replace parts that are broken, missing, obviously worn, distorted, or contaminated.
• Before activating any part of the ventilator, be sure to check the equipment for proper operation.
• To prevent patient injury, do not use a ventilator if it requires repair.
• If the ventilator is damaged or its external housing is not correctly closed or it behaves in a way that is not described in this manual (excessive noise, heat emission, unusual odor, alarms not triggered during the start-up procedure), the oxygen and power supplies should be disconnected and use of the device stopped immediately.
• Verify the functionality of the alarms before making ventilator available for patient use.
• Do not perform ventilator alarm tests while the patient is connected to the ventilator.
• If the ventilator fails any alarm tests, refer to
Chapter 5 to service appropriate
components. Refer to Table 4-3
to troubleshoot specific alarm messages.
• Do not start ventilation until you ensure the device is correctly assembled, the inlet air filter is properly installed and unobstructed, and there is proper clearance all around the unit. Also ensure the patient circuit is correctly connected to both the ventilator and the patient, and the patient circuit, including all hoses, is not damaged or obstructed.
Safety Information
1-3
Safety Information
• To minimize the risk of damage, you must use the ventilator's factory approved packaging to transport the ventilator.
• Do not use the ventilator without a properly installed inlet air filter or with a dirty inlet air filter.
• The inlet air filter is not reusable. Do not attempt to wash, clean, or reuse it.
• If you cannot determine the cause of a problem with the ventilator, contact Covidien
Technical Support or your local representative. Do not use the ventilator on a patient until the problem has been corrected.
• To avoid injury and/or possible damage to the ventilator: before using the ventilator, use a flow meter (flow regulator) to regulate the oxygen supply to specifications before connecting the ventilator to the oxygen supply.
General Warnings Regarding Use of Equipment
• The ventilator must be used only under the responsibility and on the prescription of a doctor.
• The Puritan Bennett 560 Ventilator is not intended to be a comprehensive monitoring device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment.
• Patients on life-support equipment should be appropriately monitored by competent medical personnel and suitable monitoring devices.
• An alternative source of ventilation should always be available when using the Puritan
Bennett 560 Ventilator.
• Handle the ventilator with care during and after use, particularly when ambient temperatures are high. Some ventilator surfaces may become hot, even if safety specifications are not exceeded.
• Do not use sharp objects to make selections on the keyboard.
• Do not clean any gas pathway with a liquid cleaner. Refer to
for allowable cleaning and disinfecting agents for use on other parts of the ventilator.
• Use all cleaning solutions and products with caution. Read and follow the instructions associated with the cleaning solutions you use to clean your ventilator. Use only those solutions listed in
• Never use a liquid cleaner inside the patient circuit, or on any component of a gas pathway.
• The SET UP menu is only accessible if ventilation is stopped and the ventilator is powered on while the ALARM CONTROL key is pressed.
1-4 Service Manual
Service Manual
• Even though the Puritan Bennett 560 Ventilator meets current safety standards, the internal lithium-ion battery of the device exceeds the 100Wh threshold and is therefore considered to be Dangerous Goods (DG) Class 9 – Miscellaneous, when transported in commerce. As such, the Puritan Bennett 560 Ventilator and/or the associated lithium-ion battery are subject to strict transport conditions under the Dangerous Goods Regulation for air transport (IATA: International Air Transport Association), International Maritime
Dangerous Goods code for sea and the European Agreement concerning the
International Carriage of Dangerous Goods by Road (ADR) for Europe. Private individuals who transport the device are excluded from these regulations although for air transport some requirements apply. For air transport; the Puritan Bennett 560
Ventilator is permitted as checked-in or carry-on baggage. Two spare batteries per person may be taken on board as carry-on luggage only, with the prior approval of the airline. This classification and regulatory requirements may vary depending upon the country and mode of transport. Therefore it is recommended that users verify with the carrier / airline as to which measures to take before the voyage.
• A continuous alarm condition will be activated if the ventilator power is switched off during ventilation. When the power is switched back on, ventilation will resume without having to press the ventilator’s ON/OFF key.
• To ensure correct and lasting operation of the device, ensure the ventilator is installed and operated in the environmental conditions recommended in Appendix B,
“Specifications” of the Clinician’s manual.
• To ensure correct and lasting operation of the ventilator, ensure its air circulation holes
(main inlet or cooling) are never obstructed. Place the device in an area where air can freely circulate around the ventilator and avoid installing it near floating fabrics, such as curtains.
• The ventilator should never be immersed in any liquid, and any liquid on the surface of the device should be wiped away immediately.
• If the ventilator has been transported or stored at a temperature that differs more than
20 °C (
68 °F) from the temperature in which it will be operating, the ventilator should be allowed to stabilize in its operating environment for at least two (2) hours prior to use.
• If the ambient temperature is above 104 °F (40 °C), the battery's thermal cut-off circuit may prevent the battery from recharging fully-even if the INTERNAL BAT charging indicator indicates the battery is completely recharged.
• The default setting for altitude compensation is YES. Altitude compensation should always be set to YES for accurate volume delivery calculations at all elevations.
• If the ventilator is damaged or its external housing is not correctly closed or it behaves in a way that is not described in this manual (excessive noise, heat emission, unusual odor, alarms not triggered during the start-up procedure), the oxygen and power supplies should be disconnected and use of the device stopped immediately.
• The Puritan Bennett 560 Ventilator offers a variety of breath delivery modes.
Throughout the patient’s treatment, the clinician should carefully select the ventilation mode or modes to use for that patient, based on clinical judgment, considering the condition and needs of the individual patient, as they change from time to time, and considering the benefits, limitations and operating characteristics of each mode.
Safety Information
1-5
Safety Information
Warnings Regarding the Environment
• For environmental protection, the ventilator and its components, whatever their respective conditions of operation, cannot be disposed of with household waste and must be submitted for suitable selective collection and possible recycling. Observe all applicable regulations when disposing of the ventilator and any of its components.
• Follow local governing ordinances and recycling regulations regarding disposal or recycling of batteries and other device components.
Warnings Regarding Hoses and Accessories
• The ventilator must not use, nor be connected to, any anti-static or electrically conductive hoses, tubing, or conduits.
• Resistance of the exhalation valve and accessories (water traps, filters, etc.) must be as low as possible.
• The exhalation valve must allow rapid discharge of the circuit pressure. Ensure the exhalation valve is always clean and its evacuation aperture (exhaust port) is never obstructed.
• Never use a ventilator or any components or accessories that appear to be damaged. If any signs of damage are evident, contact your equipment supplier or Covidien.
• After assembling, cleaning, or reassembling the patient circuit, inspect the hoses and other components to ensure there are no cracks or leaks and that all connections are secure.
• The hose connecting the ventilator to the oxygen source must be designed exclusively for use with medical-grade oxygen. Under no circumstances should the oxygen hose be modified by the user. In addition, the hose must be installed without the use of lubricants.
• The exhalation block is intended for single use by a single patient. It may periodically be cleaned, but it cannot be disinfected or sterilized.
• Ensure the exhalation block is completely dried after cleaning and prior to use.
• When an exhalation block is set up, each time it is removed, or after installing a new exhalation block on the machine, it is essential that the exhalation flow sensor be recalibrated before the exhalation block is used. Refer to
.
Warnings Regarding Electrical Power
• The maximum recommended shelf life of the internal battery is two (2) years. Do not use a battery that has been stored for two years prior to its first use.
• Periodic recharging is important to help maximize useful life of the battery. Do not store the internal battery for extended periods, without recharging, as this may reduce the maximum life.
• Due to typical voltage fluctuations that occur during normal power wheelchair use, the wheelchair mains battery should never be used to power the Puritan Bennett 560
Ventilator, nor should the ventilator’s battery be used to power the wheelchair. The ventilator should always be connected to an independent power source (e.g. AC power, extra batteries, or DC power source).
1-6 Service Manual
Service Manual
• For the AC (“mains”) power cable to be properly secured, the attachment located on the power cable must be fitted into the power cable holder incorporated in the battery access cover and located under the AC (mains) power socket.
• The power supply to which the ventilator is connected (both AC and DC) must comply with all applicable standards and provide electrical power corresponding to the voltage
characteristics inscribed on the rear of the ventilator and in Table 2-3 to ensure correct
operation.
• After replacing the ventilator's internal battery, push on the battery's cover from the rear towards the front to take pressure off the cover mounting brackets, thereby avoiding breakage when the screws are tightened.
• Do not leave power cables lying on the ground where they may pose a hazard.
• Before cleaning the ventilator, first disconnect the ventilator and the patient circuit.
• Ensure that the AC power cable is in perfect condition and not compressed. The device should not be turned on if the AC power cable is damaged.
• To connect the ventilator to an external power source, first ensure the ventilator’s power switch is off. Then, connect the desired power cable to the ventilator. Finally, connect the power cable to the external power source.
• To disconnect the ventilator from an external power source, first power-down the ventilator. Then, disconnect the power cable from the external power source and, finally, the ventilator.
Warnings Regarding Oxygen
• Strictly follow the instructions provided in the Clinician’s manual for connecting the oxygen supply, which include the use of a flow regulator and special coupler.
• The oxygen supply must be regulated using a flow meter connected to the source gas outlet.
• The oxygen supply must be shut off when ventilation is off. Before disconnecting the oxygen hose, allow the ventilator to continue for a few cycles without oxygen to flush the patient circuit of excess oxygen.
• Before connecting the oxygen supply, ensure the stud on the oxygen connector is protruding outward.
• Inspect the oxygen connector before use to ensure its o-ring is attached and in good condition. Do not use an oxygen connector with a missing, worn, or damaged o-ring.
• In the event of an oxygen leak, shut down the supply of oxygen at its source. In addition, remove and/or keep any incandescent source away from the device, which may be enriched with oxygen. Circulate fresh air into the affected room to bring the oxygen level down to normal.
• To ensure stability, when the Puritan Bennett 560 Ventilator is mounted on a cart, the weight of the oxygen bottle should not exceed 14 kg (30 lb).
• Ensure the oxygen supply pressure to the machine never exceeds 7 psi (50 kPa) or a flow of 15 lpm.
Safety Information
1-7
Safety Information
• The hose connecting the ventilator to the oxygen source must be designed exclusively for use with medical-grade oxygen. Under no circumstances should the oxygen hose be modified by the user. In addition, the hose must be installed without the use of lubricants.
• Ensure the only gas supplied to the ventilator through the dedicated oxygen supply connector is medical-grade oxygen.
Warnings Regarding USB Memory Device
• Always verify that the serial number of the ventilator is correctly associated with the correct patient before using a USB memory device to transfer data between the ventilator and a PC.
1-8 Service Manual
2 General Information
This manual provides information needed to service the Puritan Bennett™ 560 Ventilator.
It is intended for use by biomedical engineering technicians or personnel with equivalent experience who have successfully completed Covidien training on this product.
This chapter provides introductory information on the Puritan Bennett 560 Ventilator including:
• a description of the ventilator, its accessories, and its controls and indicators
• a preventive maintenance schedule
• detailed specifications and required and test equipment used for service and repair
• an outline of the service philosophy.
2.1
About this Manual
COVIDIEN recommends that you become familiar with this manual and the Puritan
Bennett 560 Ventilator Clinician’s Manual ( for patient circuit setup, operation, and breath delivery information) before attempting to operate or maintain the ventilator.
This manual is intended to be a comprehensive guide for servicing the Puritan Bennett
560 Ventilator.
provides safety information
Chapter 2 provides general information as described above.
describes the ventilator’s technical operating theory.
contains alarm, diagnostic, and troubleshooting information, and suggests which components to replace based upon troubleshooting results.
describes service prerequisites and provides step-by-step instructions for disassembling the ventilator, replacing components or assemblies, and reassembling the ventilator.
defines required performance verification tests.
contains parts lists and detailed assembly drawings for ordering replacement parts.
2-1
General Information
2.2
Product Description and Intended Use
The Puritan Bennett 560 Ventilator is indicated for the mechanical ventilatory support of patients weighing at least 5 kg (11 lb). It is not intended for ventilator-dependent patients. The servicing of the ventilator is intended to be performed by qualified, trained personnel. The ventilator is applicable for adult and pediatric patients who require the following general types of ventilatory support, as prescribed by an attending physician:
• Positive Pressure ventilation
• Assist/Control or CPAP modes of ventilation
• Pressure Control, and Pressure Support (ST) breath types.
The ventilator is suitable for use in institutional, home, and portable settings. It is not intended for use as an emergency transport ventilator.
The Puritan Bennett 560 Ventilator is a microprocessor-controlled ventilator consisting of a low-inertia micro turbine (blower) used for flow generation. A three-way solenoid valve is used to pilot an exhalation valve at the outlet of exhalation block. Flow and pressure sensors provide information to a CPU which uses control algorithms to provide closedloop feedback control for breath delivery. An internal fan provides controlled cooling for electronic and pneumatic ventilator components.
The ventilator can be powered via external AC power, an external DC power source, or an internal battery during situations where there is a loss or absence of external power. A
Power Management PCBA determines which power source (AC mains, external DC power source, or internal battery) is the primary supply depending on the availability of each source.
The servicing of the ventilator is intended to be performed by qualified, trained personnel and its use is applicable for adult and pediatric patients.
provides a list of accessories that are available for the Puritan Bennett 560
Ventilator.
To order parts or accessories, contact your equipment supplier or COVIDIEN representative.
WARNING
Use only accessories recommended by Covidien.
Note:
The Puritan Bennett 560 Ventilator is shipped with the following items: clinician’s manual on CD-
ROM (multi-language), one set of six (6) inlet air filters, carrying bag, one O
2
inlet connector,
European power cable, and an adult, double limb breathing circuit. Table 2-2
lists available breathing circuits and other consumable items .
Table 2-1. Ventilator Accessories
Description
Carrying bag
Oxygen inlet connector (x10)
Carrying belt
Part number
3829000
2962799
3819100
2-2 Service Manual
Service Manual
Table 2-1. Ventilator Accessories (continued)
Description
Suspension belt
Backpack, padded suspenders
Dual Bag, Pink
Dual Bag, includes:
•
Backpack padded suspenders, 2 ea. (3819300)
•
Suspension belt (3819200)
•
Carrying belt (3819100)
DC power source (external battery, Puritan Bennett™ 560
Ventilator Power Pack), includes interconnect cable
24 V DC cable for DC power source (300 mm)
AC (mains) power cable assembly, Australia
AC (mains) power cable assembly, Canada
AC (mains) power cable assembly, UK
AC (mains) power cable assembly, Japan
AC (mains) power cable assembly, China
AC (mains) power cable assembly, SAF/India
AC (mains) power cable assembly, EU
AC (mains) power cable assembly, Latin America
Part number
3819200
3819300
2967200P
2967299
4098100
3841000
2981499
2977099
2971899
2972099
2972399
2972499
2971799 various, contact your local representative
10094625 12 V cable (for connection to automobile DC power port)
RAII, with 1/4 inch jack connector:
-Remote alarm cable, NO (normally open)
-Remote alarm cable, NC (normally closed)
Remote alarm cable without jack connector:
-Remote alarm cable, RA
-Remote alarm cable, RAII
RS232 serial cable
USB cable, mini B
USB 2.0 serial converter
Utility Cart
10003919
10094353
10094626
10000914
4-074688-00 or local supplier
10094627 or equivalent
Local supplier
4096800
General Information
2-3
General Information
Table 2-2. Consumable Parts
Description
Exhalation block, single-patient use (blue)
Part number
10046892
Inlet air combi filter, fine (pack of 6)
Inspiratory bacteria filter:
Barrierbac S
Barrierbac S Angled
Hygrobac
Hygrobac S
Hygrobac S Angled
Hygroboy
Hygroster
Hygroster Mini
Sterivent
Sterivent S
Sterivent Mini*
10028771
Double-limb patient circuit with exhalation valve, 180 cm,
PVC, non-sterile, ADULT
Double-limb patient circuit with exhalation valve, 180 cm,
PVC, non-sterile, PEDIATRIC
Single-limb patient circuit with exhalation valve, 180 cm,
PVC, non-sterile, ADULT
Single-limb patient circuit with exhalation valve, 180 cm,
PVC, non-sterile, PEDIATRIC
5093900
5093600
5093500
Single-limb patient circuit without exhalation valve, 180 cm,
PVC, non-sterile, ADULT
Single-limb patient circuit without exhalation valve, 180 cm,
PVC, non-sterile, PEDIATRIC
5093300
5093100
*Do not use Sterivent Mini filters to test or calibrate the ventilator
350U5879
350U19006
352U5805
352U5877
352U5996
355U5430
354U5876
354U19028
351U5410
351U5878
351U5979
5094000
2-4 Service Manual
Service Manual
2.3
Specifications
Weight
Dimensions (excluding accessories)
Connectors
Temperature
Atmospheric pressure
Altitude
Humidity
Combination temperature and humidity
DC voltage supply
Turbine
Oxygen inlet
AC input
DC supply
Table 2-3. Ventilator Specifications
Physical characteristics
9.9 lb (4.5 kg)
9.3 in wide x 12.4 in deep x 6.0 in high
(235 mm wide x 315 mm deep x 154 mm high)
Inspiratory limb connector: ISO 22 mm (OD) 15 mm (ID) conical
Exhalation limb connector (on exhalation block): ISO 22 mm (ID) conical
Oxygen inlet: female connector with valve
Exhalation pilot port: accommodates 3.2 mm to 4 mm ID tubing
Proximal pressure port: accommodates 5 mm to 6 mm ID tubing
Environmental requirements
Operating: 5º to 40º C (41º to 104º F) 20 minutes after conditioning at 23 o
C
Storage: -40º to 70º C (-40º to 158º F)
Operating: 450 to 825 mmHg (600 to 1100 hPa; 8.7 to 15.9 psi)
Storage: 375 to 825 mmHg (500 to 1100 hPa; 7.3 to 15.9 psi)
Operating/storage: -500 ft to 13,000 ft (-152 m to 3962 m)
Operating/storage: 10% to 95% RH noncondensing
Operating: 113º F (45º C) and 75% RH
-10% to +10% of nominal (12-30 V DC)
Pneumatic specifications
Maximum flow: 240 lpm at 24 cmH
2
O ± 1%
Maximum pressure:
70 cmH
2
O at 0 lpm
Pressure: 7 psi max (50 kPa)
Flow: 15 lpm max
Electrical specifications
100 V AC – 240 V AC nominal (84 V AC – 264 V AC)
50 Hz – 60 Hz nominal (47 Hz – 63 Hz)
180 VA max
430 BTUs per hour
12 V DC ± 10%, 8.3 A to 30 V DC ± 10%, 3.3 A max
General Information
2-5
General Information
Table 2-3. Ventilator Specifications (continued)
Internal battery (Puritan Bennett
560 Ventilator) lithium ion 26 V DC (nominal), 4.8 Ah full load capacity
Charging current:
Standby ventilation: 1.5 A/h (duration: < 6 h)
During ventilation: 0.5 A/h (duration: <13 h)
Average operating time (- 10%) at 25 ºC (± 5 ºC) with a fully charged battery with less than 50 charge/discharge cycles at the following ventilator settings:
11 h when Vt = 200 mL (± 5 mL), Pi = 10 cmH
2
Rtot = 20 bpm
O (± 2 cmH
2
O),
9 h when Vt = 300 mL (± 5 mL), Pi = 20 cmH
2
Rtot = 15 bpm
O (± 2 cmH
2
O),
6.5 h when Vt = 500 mL (± 5 mL), Pi = 30 cmH
2
Rtot = 15 bpm
O (± 2 cmH
2
O),
4.5 h when Vt = 750 mL (± 5 mL), Pi = 45 cmH
2
Rtot = 20 bpm
O (± 2 cmH
2
O),
Shelf life:
Installed in ventilator (without recharging): 6 months
Not installed in ventilator (spare battery): 24 months
WARNING
Even though the Puritan Bennett 560 Ventilator meets current safety standards, the internal lithium-ion battery of the device exceeds the 100Wh threshold and is therefore considered to be Dangerous Goods (DG) Class 9
– Miscellaneous, when transported in commerce. As such, the Puritan Bennett™ 560 Ventilator and/or the associated lithium-ion battery are subject to strict transport conditions under the Dangerous Goods Regulation for air transport (IATA: International Air Transport Association),
International Maritime Dangerous Goods code for sea and the European Agreement concerning the International
Carriage of Dangerous Goods by Road (ADR) for Europe.
Private individuals who transport the device are excluded from these regulations although for air transport some requirements apply. For air transport; the Puritan
Bennett™ 560 Ventilator is permitted as checked-in or carry-on baggage. Two spare batteries per person may be taken on board as carry-on luggage only, with the prior approval of the airline. This classification and regulatory requirements may vary depending upon the country and mode of transport. Therefore it is recommended that users verify with the carrier / airline as to which measures to take before the voyage.
2-6 Service Manual
Service Manual
General Information
Table 2-3. Ventilator Specifications (continued)
Communication capabilities Remote alarm (nurse call) port. Allows alarm conditions to be annunciated at locations away from the ventilator (for example, when the ventilator is in an isolation room). The ventilator signals an alarm using a normally open or a normally closed signal.
The ventilator asserts a remote alarm when there is an active alarm condition, unless the audio paused function is active, and when the ventilator power switch is turned off. The remote alarm port is an8-pin female connector. Allowable current is 100 mA at
24 V DC (maximum).
2
3
1
4
8
5
6
7
Nurse call pin-out (view from back of ventilator)
4
5
6
Pin
1
2
3
7
8
Signal relay common normally open (NO) normally closed (NC) not used not used not used not used not used
Remote alarm cable wire color black brown orange
--
--
--
--
--
WARNING
To connect the ventilator to a Nurse Call device, contact Covidien to check the ventilator’s compatibility with the Nurse Call device and order a suitable connection cable. Before using the Nurse Call system, ensure that its connections are secure and it operates properly. Do not use Nurse Call devices that operate based on the closure of an electrical circuit, because the devices often do not take into account possible cable disconnection or a total loss of power. Ensure that the Nurse Call device is always connected to the ventilator.
PC communication port. Allows communication between the ventilator and a PC. Used for software uploads and data downloads.
2-7
General Information
Table 2-3. Ventilator Specifications (continued)
Alarm volume
Ventilator operating volume
Inspiratory resistance
Exhalation resistance
Internal ventilator compliance
Working pressure
Maximum pressure limit
Sound pressure level
USB Type A ports. Two Type A USB ports are included on the
Puritan Bennett 560 Ventilator for obtaining trended, event, and monitoring information.
65 to 80 dBA ± 10% (as measured per IEC 60601-1-8)
30 dBA + 10%
Ventilator airway resistance
1.0 cmH
2
O at 30 lpm
3.7 cmH
2
O at 60 lpm
0.5 cmH
2
O at 30 lpm
1.1 cmH
2
O at 60 lpm
Ventilator airway compliance
0.0001 L/cmH
2
O
Performance specifications
5 cmH
2
O - 55 cmH
2
O
60 cmH
2
O
30 dBA per ISO 17510-1 test conditions
Adult double branch circuit with exhalation valve
Patient circuit inspiratory resistance
2 cmH
2
O at 60 lpm
Pediatric double branch circuit with exhalation valve
2 cmH
2
O at 30 lpm
Filter type
Filtration efficiency
Flow resistance
Maximum allowable flow resistance (pressure drop)
Inlet air filter specifications mechanical; foam and fine particle filter
99.999982% at 30 lpm (filtering microbes 3.3 μm)
0.06 cmH
2
O at 24 lpm
Inspiratory bacteria filter specifications
4 cmH
2
O at 60 lpm
2.4
Range, Resolution, and Accuracy
Ranges, resolutions, and accuracies, for ventilator settings, alarm settings, and patient data are listed in
Note:
Accuracies only apply in monitored patient data values.
2-8 Service Manual
Service Manual
General Information
Table 2-4. Range, Resolution, and Accuracy
Setting or monitored parameter Range, resolution, and accuracy
Ventilator settings
Apnea Time (available in PSV ST, CPAP,
P SIMV, and V SIMV)
Range: AUTO or 1 - 60 s
Resolution: 1 s
Default value: AUTO
Dependencies: Backup R, R-Rate
In PSV ST, Apnea time: AUTO = 60/Backup R s
In V SIMV or P SIMV, Apnea Time: AUTO = 12 s
WARNING
•
In SIMV modes, no apnea alarm occurs if set Apnea Time > 60/R-Rate.
•
In CPAP mode, APNEA TIME only drives the alarm signal — there is no backup rate.
If the user has set Apnea Time to OFF in the SETUP menu, then the patient will receive backup ventilation during an APNEA condition, but will not sound an alarm.
Backup Rate
(Backup R) (available in PSV ST)
Range: 4 - 40 bpm or OFF
Resolution: 1 bpm
Default value: 13
Dependencies: Low I time
In P SIMV and V SIMV, Backup rate = Max (8, R-Rate) and is not adjustable
Expiratory Sensitivity
(Esens) (available in PSV ST, V SIMV,
P SIMV)
Flow pattern (available in V A/C)
Inspiratory Sensitivity
(Insp Sens) (available in V A/C, P A/C, PSV ST,
P SIMV, and V SIMV)
Range: 5% to 95% of peak flow or -95% to -5% of peak flow depending upon E Sens setting selected in SETUP mode; positive or negative)
Resolution: 5%
Default value: 75% (AUTO)
In CPAP Exh Sens is fixed at 25% and is not adjustable.
Range: Square, descending ramp, or sinusoidal
Resolution: N/A
Default value: Descending ramp
In V SIMV, flow pattern is set to square and is not adjustable
Range: OFF or 0P - 5 (0P indicates a pediatric setting.
Autotriggering may result if this setting is used while ventilating an adult patient.)
Resolution: 1
Default value: 2
In CPAP, Insp Sens is fixed at 2 and is not adjustable
2-9
General Information
Table 2-4. Range, Resolution, and Accuracy (continued)
Setting or monitored parameter
Inspiratory Time
(Insp Time) (available in P A/C, V A/C, P SIMV,
V SIMV)
Mode
PEEP
Inspiratory Pressure
(Pi)
Pressure Support
(P Support) (available in P SIMV, V SIMV, PSV
ST)
Respiratory Rate
(R-Rate) (available in V A/C, P A/C, V SIMV, P
SIMV)
Range, resolution, and accuracy
Range: 0.3 s to 6.0 s in P A/C and V A/C modes; 0.3 s to
2.4 s in P SIMV and V SIMV modes
Resolution: 0.1 s
Default value: 1.5 s
Dependencies: R-Rate in P SIMV mode; R-Rate and Vt in
V SIMV mode
Range: V A/C, P A/C, V SIMV, P SIMV, PS V ST, CPAP
Resolution: N/A
Default value: P A/C
Range: OFF or 1 to 20 cmH
2
O; OFF corresponds to a 0.5 cmH
2
O target
Resolution: 1 cmH
2
O
Default value: OFF
Dependencies: Pi, P Support, Low PIP, High PIP
Note: The minimum PEEP value in CPAP mode is 4
(pressure units chosen by user) and OFF is not available.
In non-invasive ventilation the minimum value of PEEP is 4 (pressure units chosen by user).
Range: 5 cmH
2
O to 55 cmH
2
0 in P A/C in valve configuration or 6 cmH
2
O to 30 cmH
2
0 in leak configuration (See
for the differences between valve and leak configurations)
Resolution: 1 cmH
2
O
Default value: 15 cmH
2
O
Dependencies: PEEP
Range: OFF or 5 cmH
2
O to 55 cmH
2
0 in valve configuration or 5 cmH
2
O to 30 cmH
2
0 in leak configuration (See
for the differences between valve and leak configurations)
Resolution: 1 cmH
2
O
Default value: 15 cmH
2
O
Dependencies: PEEP, Low PIP, High PIP
Range: 1 bpm - 60 bpm in V A/C and P A/C modes;1 bpm - 40 bpm in P SIMV and V SIMV modes with I:E not exceeding 1:2Resolution: 1 bpm
Default value: 13
Dependencies: Insp Time in P SIMV mode; Insp Time and Vt in V SIMV mode
2-10 Service Manual
Service Manual
General Information
Table 2-4. Range, Resolution, and Accuracy (continued)
Setting or monitored parameter
Rise Time (available in P A/C, PSV ST, P SIMV,
V SIMV)
Tidal Volume Vt
(available in V SIMV, V A/C)
I:E ratio (available in P A/C and V A/C)
I/T ratio (available in P A/C and V A/C)
Vt target (available in P A/C,PSV ST)
High Inspiratory Time
(High I time)
Low Inspiratory Time
(Low I time)
Range, resolution, and accuracy
Range: 1-4 (1 is fastest, 4 is slowest)
Resolution: 1
Default value: 2
Dependencies: Insp time
Range: 50 mL to 2000 mL
Resolution: 10 mL
Default value: 500 mL
Dependencies: Insp time, R-Rate, Low VTI, High VTI,
Low VTE, High VTE, Low Min Vol, High Min Vol
Range: 9.9:1 to 1:9.9
Resolution: 0.1
Default value: 1:2
Range: 20% to 50%
Default value: 33%
Resolution:1%
Range: 50 mL to 2000 mL; OFF
Resolution: 10 mL
Default value: OFF (100 mL)
Range: 0.8 to 3.0 s
Resolution: 0.1 s
Default value: AUTO (minimum of 3 s or 30/monitored rate)
Dependencies: Low I time, R-Rate
Range: AUTO or 0.1 to2.8 s
Resolution: 0.1 s
Default value: AUTO (Rise time + 300 ms)
Dependencies: High I time, Backup R, Rise time
Alarm settings (based on breath mode)
High Inspired Tidal Volume (High VTI)
High Exhaled Tidal Volume (High VTE)
Pressure A/C mode
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value: 2000 mL
Dependencies: Low VTI
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value:1000 mL
Dependencies: Vt, Low VTE
2-11
General Information
Setting or monitored parameter
High Respiratory Rate (High Rtot)
Low Exhaled Tidal Volume (Low VTE)
Low Inspired Tidal Volume (Low VTI)
High leak
High Exhaled Tidal Volume (High VTE)
High Inspired Tidal Volume
(High VTI)
High Respiratory Rate
(High Rtot)
Table 2-4. Range, Resolution, and Accuracy (continued)
Low Exhaled Tidal Volume (Low VTE)
Low Inspired Tidal Volume
(Low VTI)
Range, resolution, and accuracy
Range: 10 bpm to 70 bpm; OFF
Resolution: 1 bpm
Default value: OFF
Dependencies: R-Rate
Range: 30 mL to 1990 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: Vt, High VTE
Range: 30 mL to 2000 mL; OFF
Resolution: 10 mL
Default value: 300 mL
Dependencies: High VTI
Range: 5 lpm - 200 lpm; OFF
Resolution: 5 lpm
Default value: OFF
To be used in non-invasive ventilation without an exhalation valve
PSV ST mode
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value:1000 mL
Dependencies: Vt, Low VTE
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value: 2000 mL
Dependencies: Low VTI
Range: 10 bpm to 70 bpm; OFF
Resolution: 1 bpm
Default value: OFF
Dependencies: Backup R
Range: 30 mL to 1990 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: Vt, High VTE
Range: 30 mL to 2000 mL; OFF
Resolution: 10 mL
Default value: 300 mL
Dependencies: High VTI
2-12 Service Manual
Service Manual
Setting or monitored parameter
High leak
High Exhaled Tidal Volume
(High VTE)
High PIP
High Respiratory Rate
(High Rtot)
Low PIP
Table 2-4. Range, Resolution, and Accuracy (continued)
Low Exhaled Tidal Volume
(Low VTE)
High Exhaled Tidal Volume
(High VTE)
High Inspired Tidal Volume
(High VTI)
Range, resolution, and accuracy
5 - 200 lpm or OFF
To be used in non-invasive ventilation without an exhalation valve
V SIMV mode
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value:1000 mL
Dependencies: Vt, Low VTE
Range: 12 cmH
2
O to 60 cmH
2
O
Resolution: 1 cmH
2
O
Default value: 40 cmH
2
O
Dependencies: PEEP, P support, Low PIP
Range: 17 bpm to 70 bpm; OFF
Resolution: 1 bpm
Default value: OFF
Dependencies: R-Rate
Range: 2 cmH
2
O to 52 cmH
2
O
Resolution: 1 cmH
2
O
Default value: 2 cmH
2
O
Dependencies: PEEP, P support, High PIP
The Low PIP value will automatically be raised to set
PEEP + 2 if PEEP is increased to a level that surpasses
Low PIP - 2
Range: 30 mL to 1990 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: Vt, High VTE
P SIMV mode
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value: 1000
Dependencies: Low VTE
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value: 2000 mL
Dependencies: Low VTI
General Information
2-13
General Information
Setting or monitored parameter
High Respiratory Rate
(High Rtot)
Low Exhaled Tidal Volume
(Low VTE)
Low Inspired Tidal Volume
(Low VTI)
High Exhaled Tidal Volume
(High VTE)
High Peak Inspiratory Pressure
(High PIP)
High Respiratory Rate
(High Rtot)
Table 2-4. Range, Resolution, and Accuracy (continued)
Low Exhaled Tidal Volume
(Low VTE)
Low Peak Inspiratory Pressure
(Low PIP)
Range, resolution, and accuracy
Range: 17 bpm to 70 bpm; OFF
Resolution: 1 bpm
Default value: OFF
Dependencies: R-Rate
Range: 30 mL to 1990 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: High VTE
Range: 30 mL to 2000 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: High VTI
Volume A/C mode
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value:1000
Dependencies: Vt, Low VTE
Range: 12 cmH
2
O to 60 cmH
2
O
Resolution: 1 cmH
2
O
Default value: 40 cmH
2
O
Dependencies: PEEP, Low PIP
The High PIP value will automatically be raised to set
PEEP + 10 if PEEP is increased to a level that surpasses
High PIP - 10
Range: 10 bpm to 70 bpm; OFF
Resolution: 1 bpm
Default value: OFF
Dependencies: R-Rate
Range: 30 mL to 1990 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: Vt, High VTE
Range: 2 cmH
2
O to 52 cmH
2
O
Resolution: 1 cmH
2
O
Default value: 2 cmH
2
O
Dependencies: PEEP, High PIP
The Low PIP value will automatically be raised to set
PEEP + 2 if PEEP is increased to a level that surpasses
Low PIP - 2
2-14 Service Manual
Service Manual
General Information
Setting or monitored parameter
High Inspired Tidal Volume
(High VTI)
High leak
High Respiratory Rate
(High Rtot)
Low Inspired Tidal Volume
(Low VTI)
Exhaled Tidal Volume (VTE)
Expiratory Time (Etime)
I:E Ratio (I:E)
I/T Ratio
Table 2-4. Range, Resolution, and Accuracy (continued)
Minute Volume (M. VoI)
Range, resolution, and accuracy
CPAP
Range: 80 mL to 3000 mL; OFF
Resolution: 10 mL
Default value: 2000 mL
Dependencies: Low VTI
Range: 5 to 200 lpm; OFF
Resolution: 5 lpm
Default value: OFF
Range: 10 bpm to 70 bpm; OFF
Resolution: 1 bpm
Default value: OFF
Dependencies: R-Rate
Range: 30 mL to 2000 mL; OFF
Resolution: 10 mL
Default value: 300
Dependencies: High VTI
Monitored patient data
Resolution: 5 mL
Accuracy: ± (10 mL + 10%)
Resolution: 0.1 s
Accuracy: ± (100 ms)
Resolution: 0.1
Accuracy: Insp. time ±50 ms and Exh. time ±50 ms or I:E ratio ±10%, whichever is greater
Resolution: 1%
Accuracy: Insp. time ±50 ms and Exh. time ±50 ms or I/T ratio ±10%, whichever is greater
Resolution: 0.1 L
Accuracy: ± (10 mL + 10% VTI) x rate in valve configuration and ± (20 mL + 20% VTI) x rate in leakage configuration)
2-15
General Information
Table 2-4. Range, Resolution, and Accuracy (continued)
Setting or monitored parameter
Inspiratory Tidal Volume (VTI)
Range, resolution, and accuracy
In valve configuration:
Resolution: 5 mL
Accuracy: ± (10 mL + 10%) and ± (20 mL +20%) in CPAP mode above 200 mL
In leak configuration:
Resolution: 5 mL
Accuracy: ± (20 mL +20%)
Positive End Expiratory Pressure
(PEEP)
Total Respiratory Rate
(R tot)
Leak (displayed only in leak ventilation without exhalation valve)
% Spontaneous cycles (appears in the ventilation report menu)
Apnea index (AI)
Resolution: 1 cmH
2
O
Accuracy: ± (2 cmH
2
O + 8% of reading)
Resolution: 1 bpm
Accuracy: ± 1 bpm
Resolution: 1 lpm
Accuracy:± (3 lpm + 20%)
Resolution: 1%
Accuracy:± (1%)
Resolution: 1
Accuracy:± 1 ev/h
2-16 Service Manual
Service Manual
General Information
2.5
Default Settings
through
lists the default settings of the Puritan Bennett 560
Ventilator. Downloading a new software version to the ventilator is the only way to automatically reset the ventilator’s parameters to these settings. You should, however, manually enter these settings when you are finished servicing the ventilator if applicable.
Table 2-5. Default Settings
Parameter
Mode
Pi
PEEP
Rise Time
R - Rate
I:E
Insp Sens
Vt Target
Setting
P A/C
15 cmH
2
O
OFF
2
10
1:3.6 or 1.3s or (I/T of 22%)
2
OFF
Table 2-6. Default Alarm Settings
Alarm
VTI (ml)
VTE (ml)
Rtot (bpm)
Setting
Low 300 mL; High 2000 mL
Low 300 mL; High 1000 mL
Low -- High OFF
Table 2-7. Default Preferences (after software download)
Setting Preference
Backlight
Contrast
OFF
Alarm Volume
Key Sound
Apnea Alarm
Disconnection alarm
Display Waveforms
Pediatric Circuit
Ventilation Report
Accept tone
YES
15s
OFF
OFF
N/A
2-17
General Information
Note: Access the Preferences menu from the main ventilation screen by placing the cursor on the
Preferences position and pressing ENTER .
The ventilation report contains average values for a 24-hour period from 8:00 a. m. to 8:00 a.m. The report includes VTI, VTE, airway pressure, Rate, Leak, Apnea Index (apneic events per hour), Apnea
Time, % Spontaneous cycles, machine hours, and patient hours.
2.6
Compliance and Approvals
The Puritan Bennett 560 Ventilator was developed in accordance with pertinent North
American and International standards (
). The manufacturing facility for this product is certified to EN ISO 13485:2000 (ISO 13485:1996) Quality Systems - Medical
Devices - Particular Requirements for the Application of ISO 9001:1994.
The ventilator’s IEC 60601-1/EN 60601-1 classification is protection class II internally and mains powered (with detachable power cord), Type BF applied parts, IPX1 (protection from ingress of moisture, continuous operation. The device is not suitable for use with flammable anesthetics, and is not suitable for sterilization.
Table 2-8. Ventilator Compliance with Standards
Standards/certifications
IEC 60601-1 Medical electrical equipment, Part 1: General requirements for safety: 1988 and all its amendments up to 1995 and EN 60601-1:1990
UL 60601-1 Medical electrical equipment, Part 1: General requirements for safety: 2003
IEC/EN 60601-1-2 Medical electrical equipment, Part 1-2: General requirements for safety, Collateral Standard:
Electromagnetic compatibility - Requirements and tests: 2007
IEC 60601-1-4 Medical electrical equipment, Part 1: General requirements for safety-2- Collateral Standard:
Programmable Electrical Medical Systems: 2000
EN 60601-1-4 Medical electrical equipment, Part 1: General requirements for safety-2- Collateral Standard:
Programmable Electrical Medical Systems: 2004
IEC 60601-1-6 Medical electrical equipment, Part 1: General requirements for safety-2- Collateral Standard:
Usability: 2006
EN 60601-1-6 Medical electrical equipment, Part 1: General requirements for safety-2- Collateral Standard:
Usability: 2007
IEC 60601-1-8 General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems: 2006
EN 60601-1-8 General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems: 2007
EN ISO 10651-2 Lung ventilators for Medical Use - Particular Requirements for basic safety and Essential
Performance, Part 2: Home Care Ventilators for Ventilator-Dependent Patients: 2009 and Chinese version
YY0600.2-2007
CAN/CSA C22.2 No. 601-1-M90 Supplement 1-94 Medical Electrical Equipment, Part 1: General Requirements for Safety: Reaffirmed 2005
Medical Device Directive 93/42/EEC as amended by 2007/47/EC.
Directive 2006/66/EC: 2006 Batteries and Accumulators and Waste Batteries and Accumulators
Directive 2002/96/EC: 2003 Waste Electrical and Electronic Equipment
2-18 Service Manual
Service Manual
General Information
Table 2-8. Ventilator Compliance with Standards
Standards/certifications
EN ISO 5356-1 Anesthetic and respiratory equipment - Conical connectors - Part 1: Cones and sockets: Part 1 re-issued 2004
EN 50419 Marking of Electrical and Electronic Equipment: 2005
ISO 10651-3: 1997 Sections 21.6a (vibration sinusoidal) 21.6b (random vibration), and 21.6.c (bump test) of
Lung Ventilator for Medical Use - Part 3: Particular Requirements for Emergency Transport Ventilators
Compatible with drop test according to 10.1.3 d article of draft IEC 60601-1-11: 2008 Medical Electrical
Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard:
Requirements for Medical Electrical Equipment and Medical Electrical Systems used in the Home Healthcare
Environment (document IEC/SC62A 62A/624/CD)
RTCA DO-160E: 2007 section 21 (emission of Radio Frequency Energy) of Environmental Conditions and Test
Procedures for Airborne Equipment
EN ISO 10651-6: 2009 Lung Ventilators for Medical Use - Particular Requirements for Basic Safety and Essential
Performance Part 6: Home Care ventilatory support devices and Chinese version YY0600.1-2007
2.6.1 Manufacturer’s Declaration
The following tables contain the manufacturer’s declarations for the Puritan Bennett 560
Ventilator electromagnetic emissions, electromagnetic immunity, recommended separation distances between ventilator and portable and mobile RF communications equipment, and a list of compliant cables.
WARNING
•
Portable and mobile RF communications equipment can affect the performance of the
Puritan Bennett 560 Ventilator. Install and use this device according to the information contained in this manual and the Puritan Bennett 560 Ventilator Operator’s and
Clinician’s Manual.
• The Puritan Bennett 560 Ventilator should not be used adjacent to or stacked with other equipment, except as specified in this manual and the Puritan Bennett 560 Ventilator
Operator’s and Clinician’s manual. If adjacent or stacked use is necessary, the ventilators should be observed to verify normal operation in the configurations in which they will be used.
The Puritan Bennett 560 Ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should assure that it is used in such an environment.
2-19
General Information
Emissions test
Radiated RF emissions
CISPR 11
Conducted RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Table 2-9. Electromagnetic Emissions
Compliance level
Group 1
Class B
Group 1
Class B
Class A
Complies
Electromagnetic environment – guidance
The Puritan Bennett 560
Ventilator uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Puritan Bennett 560
Ventilator is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
2-20 Service Manual
Service Manual
General Information
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
Table 2-10. Electromagnetic Immunity
IEC 60601-1-2 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV lines/lines
± 2 kV lines/earth
< 5% U
T
(> 95% dip in U
T
for
0.5 cycle)
40% U
T
(60% dip in U
T
for
5 cycles)
70% U
T
(30% dip in U
T
for
25 cycles)
< 5% U
T
(> 95% dip in U
T
for 5 s)
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
Electromagnetic environment – guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
± 1 kV for input/output lines
± 1 kV lines/lines
± 2 kV lines/earth
< 5% U
T
(> 95% dip in U
T
for
0.5 cycle)
40% U
(60% dip in U
T
for
5 cycles)
70% U
T
T
(30% dip in U
T
for
25 cycles)
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Puritan Bennett 560
Ventilator requires continued operation during power mains interruptions, it is recommended that the
Puritan Bennett 560
Ventilator be powered from an uninterruptible power supply or a battery.
< 5% U
T
(> 95% dip in U
T
for 5 s)
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
2-21
General Information
Table 2-11. Electromagnetic Immunity — Conducted and radiated RF
Immunity test
Conducted RF
IEC 61000-4-6
IEC 60601-1-2 test level
3 Vrms
150 kHz to 80 MHz outside
ISM bands a
Compliance level
3 Vrms
150 kHz to 80 MHz outside
ISM bands
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the
Puritan Bennett 560
Ventilator, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance d = 0.35 P
Radiated RF
IEC 61000-4-3
10 Vrms inside ISM bands a
10 V/m
80 MHz to 2.5 GHz
10 Vrms inside ISM bands
10 V/m
80 MHz to 2.5 GHz d = 1.2 P d = 1.2 P to 800 MHz
80 M
Hz d = 2.3 P to 2.5 GHz
800
MHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m) b
. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey c
, should be less than the compliance level in each frequency range d
. Interference may occur in the vicinity of equipment marked with the following symbol:
2-22 Service Manual
Service Manual
General Information
Table 2-11. Electromagnetic Immunity — Conducted and radiated RF (continued)
Immunity test IEC 60601-1-2 test level Compliance level
Electromagnetic environment – guidance
Note:
•
At 80 MHz and 800 MHz, the higher frequency range applies.
•
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. a
The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz’; 13.553 MHz to 13.567
MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Puritan Bennett 560 Ventilator is used exceeds the applicable RF compliance level above, the Puritan
Bennett 560 Ventilator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Puritan Bennett 560 Ventilator.
d
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Table 2-12. Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the Ventilator
Rated maximum output power of transmitter (W)
0.01
0.1
1
10
100
Separation distance according to frequency of transmitter (m)
150 kHz to 80
MHz outside ISM bands d = 0.35 P
150 kHz to 80
MHz in ISM bands d = 1.2 P
80 MHz to 800
MHz d = 1.2 P
800 MHz to 2.5 d =
GHz
2.3 P
0.035
0.11
.35
1.1
3.5
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
•
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
•
The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to
40.70 MHz.
2-23
General Information
•
An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
WARNING
The use of accessories and cables other than those specified with the exception of parts sold by Covidien as replacements for internal components, may result in increased emissions or decreased immunity of the Puritan Bennett 560 Ventilator.
Table 2-13. Compliant Cables and Accessories
Cable or accessory
P/N 2981499
Mains power cord, Australia
P/N 2977099
Mains power cord, Canada
P/N 2971899
Mains power cord, UK
P/N2972099
Mains power cord, Japan
P/N 2972399
Mains power cord, China
P/N 2972499
Mains power cord, SAF/India
P/N 2971799
Mains power cord, EU
P/N various, contact your local representative
(PN 10003736, Brazil)
Mains power cord, Latin America
P/N 10094626
Remote alarm cable
P/N 10094625
12V cable (for connection to automobile DC power port)
5.74 ft (1.75 m)
Maximum length
5.74 ft (1.75 m)
5.74 ft (1.75 m)
5.74 ft (1.75 m)
5.74 ft (1.75 m)
5.74 ft (1.75 m)
5.74 ft (1.75 m)
5.74 ft (1.75 m)
16.40 ft (5 m)
4.26 ft (1.3 m)
2.7
Tools, Equipment, and Service Materials
Table 2-14 lists the tools, equipment, and service materials needed to service the Puritan
Bennett 560 Ventilator. All test equipment used by Covidien service personnel shall be calibrated and maintained in accordance with Field Service quality system procedures.
Note:
Part numbers are subject to change.
2-24 Service Manual
Service Manual
Tool
Bacteria filters (x2)
Table 2-14. Tools, equipment, and service materials
Barbed coupling, silicone,
22mm ID (quantity 3)
Calibration shell
Connector, 22 mm barbed to 22 mm conical
DC power supply adjustable, 0-18 VDC, 0-3 Amps
DC test cable
Digital multimeter (DMM) with 3 decimal place resolution
Disinfectant/cleaner
Electrical safety analyzer
Electrostatic shielding bags
8 x 5 in. (20.3 x 12.7cm)
11 x 14 in. (27.9 x 35.6 cm)
18 x 14 in. (45.7 x 35.6 cm
Manufacturer/model or COVIDIEN part number
See
for recommended part numbers
4-003443-00
S-231702-00B or equivalent
4-018075-00
Extech model 382202 or equivalent
10000021 modified to attach to 12 V power supply
Fluke 87 or equivalent
The following cleaners/disinfectants are suitable for use on the exterior surfaces of the Puritan Bennett 560 Ventilator:
Mild dishwashing detergent
70% isopropyl alcohol (rubbing alcohol)
10% chlorine bleach (90% tap water)
Glutaraldehyde
Hospital disinfectant cleaners (phenolicbased: o-Phenylphenol 10.5%, o-Benzyl-p-chlorophenol 5.0%; Amphyl or equivalent)
Hydrogen peroxide
15% ammonia (85% tap water)
Ammonia based household cleaners
Household cleaners (Alkyl Dimethyl
Benzyl Ammonium Chloride 0.3%, 409 or equivalent)
Fluke ESA 620 or equivalent analyzer capable of performing electrical safety tests for Class II, Type BF equipment.
Performance verification
Where used
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
(ventilator/safety tester connection)
Performance verification
General cleaning
General servicing of electronics
4-009803-00
4-009800-00
4-009801-00
General Information
2-25
General Information
Table 2-14. Tools, equipment, and service materials (continued)
Tool
FiO
2
sensor measurement kit
(attached to UUT)
Flex tube (standard tube), 22 mm x 21.0 in.
Ground test cable
Leak detector solution
Oxygen connector, male
Patient circuit tubing, dual limb, adult
Patient circuit tubing, dual limb, pediatric
Patient circuit tubing, single limb, adult
Patient circuit tubing, single limb, pediatric
Patient lead test cable
Manufacturer/model or COVIDIEN part number
3814199
4-018506-00
10000017
Where used
Performance verification
Performance verification
Performance verification
(ventilator/safety tester ground connection)
4-004489-00 Leak testing oxygen solenoid valve connections
2962700 Performance verification
5094000 Performance verification
5093900 Performance verification
5093600 Performance verification
5093500
Supplied with Fluke ESA 620
Performance verification
Performance verification (electrical safety test)
Personal computer with
Windows OS and also USB capability
Power cord assembly
PTS 2000 Pneumatic Calibration
Analyzer
Puritan Bennett™ 500 Series
Ventilator Test Software
Remote alarm test box
Local supplier
2977099
4-076185-00
10059515
10000019 (compatible with RAI remote alarm harness black plastic connector)
10005329 (compatible with RAII remote alarm harness gray metallic connector)
Software upload to ventilator
Event log download from ventilator
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
2-26 Service Manual
Service Manual
Table 2-14. Tools, equipment, and service materials (continued)
Tool
Silicone tubing, 7/32” ID (5 to
6 mm ID), various lengths
Standard exhalation block
Manufacturer/model or COVIDIEN part number
Local supplier
10046892
Local supplier
Where used
Performance verification
Performance verification
Service and repair Standard tool kit including:
Hexalobular (Torx®) drivers sizes
T8, T9, T10, T20
#1 Phillips screwdriver
Needle nose pliers
Small flat bladed screwdriver
Tie wraps
Cable cutter
Static-dissipative field service kit
(includes wrist strap, static dissipative mat, and ground cord)
Stopper, no. 1
4-018149-00 or equivalent
4-009523-00
General servicing of electronics
Stopper, no. 2
Test lung, Covidien
Test lung, 4L
Tubing connector, 3/16 to 1/8
Tubing junction, 22 mm OD, with pressure port
Tubing, silicone, 1/8 ID, 6 in.
Tubing, silicone, 3/16 ID, 3 ft.
USB cable, mini type B
USB memory device (128 MB minimum memory requirement and formatted in the 32-bit file format)
G-061574-00
LNG800P or equivalent
4-075578-00 or equivalent
4-006104-00 or equivalent
4-011521-00
4-008578-00
4-008577-00
10094627 or equivalent
Local supplier
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
Performance verification
Software upload to ventilator
Event log download from ventilator
Performance verification
General Information
2-27
General Information
2.8
Preventive Maintenance
Caution
•
To prevent component damage due to excessive wear, perform preventive maintenance and
replace components at recommended intervals, as indicated in Table 2-15 . You may find it
convenient to note anticipated replacement dates for all components based on typical usage rates or recommended intervals.
•
Covidien has determined the Puritan Bennett 560 Ventilator preventive maintenance schedule based on the life expectancies of ventilator parts. Be aware that these parts, listed below, could also require unscheduled corrective maintenance and that other parts, not listed, could also require corrective maintenance during your ventilator’s lifetime.
Table 2-15 lists the periodic maintenance activities required for the Puritan Bennett 560
Ventilator. You can find the total machine hours on the welcome screen that appears when you turn on the power switch, in the preferences screen in normal operation, and also when you enter Maintenance mode (refer to
for information on how to enter Maintenance mode.)
Table 2-15. Preventive Maintenance Schedule
Frequency
As needed
Part
Ventilator external surface
Ventilator dual bag
Inspiratory bacteria filter
Maintenance
Clean and disinfect.
Refer to
ventilator cleaning instructions.
Clean dual bag regularly (can be machine washed).
Replace.
According to institutional protocol and/or manufacturer recommendation
With each new patient
Exhalation bacteria filter
Patient circuit
Inspiratory bacteria filter
Exhalation block
Replace.
Replace.
Recalibrate exhalation flow sensor if replaced. See
for calibration instructions.
Exhalation bacteria filter
Patient circuit
Replace.
Replace.
2-28 Service Manual
Service Manual
General Information
Table 2-15. Preventive Maintenance Schedule (continued)
Frequency
Check or replace once per month or more often.
Every four months or with each new patient
Part
Inlet air filter
Exhalation block*
* See Clinician’s manual
Every 15,000 hours of use Oxygen solenoid valve
Turbine
Exhalation solenoid valve
Cooling fan
Maintenance
Replace.
Note: In particularly dusty environments, replace the inlet air filter more frequently to prevent clogging even if the preventive maintenance period has not elapsed.
Refer to
for inlet air filter replacement instructions.
Replace exhalation block and calibrate exhalation flow sensor after re-installation of exhalation block.
Refer to
for exhalation block replacement instructions.
Refer to
for calibration instructions.
Replace.
solenoid valve removal and replacement instructions.
removal and replacement instructions.
Replace.
Refer to
for exhalation solenoid valve removal and replacement instructions.
Replace.
fan removal and replacement instructions.
2-29
General Information
Table 2-15. Preventive Maintenance Schedule (continued)
Frequency
Every two years
Part
Clean and disinfect Insp block
Measurements check and calibration
Battery, lithium-ion 4.8 Ah memory
Maintenance
*Clean and disinfect the inspiratory block using one of the disinfectants listed in
Re-calibrate insp flow sensor
Reference
inspiratory flow sensor calibration instructions.
Refer to
and
for measurements check and calibration instructions.
Replace. Refer to
battery removal and replacement instructions.
Battery, lithium, 3V
Buzzer PCBA
Replace. Refer to
for 3V battery removal and replacement instructions.
Replace. Refer to
for buzzer PCBA removal and replacement instructions.
*To prevent cross contamination, both cleaning and disinfection of the inspiratory block and flow sensor calibration should be considered before new patient use in the event that filters were not used at the inspiratory port or proximal Y piece.
Buzzer PCBA
Cooling fan
Exhalation block
Exhalation solenoid valve
Inlet air filter (6 pk)
Oxygen solenoid valve
Turbine
Table 2-16. Preventive Maintenance Parts List
Part description
Battery, lithium, 3V
Battery, lithium-ion 4.8 Ah memory
Part number
10022884
10087047 (Requires minimum version power management software AL020002)
10108395
10028754
10046892
3818699
10028771
10028763
10028768
2-30 Service Manual
2.9
Controls, Indicators, and Interfaces
This section lists the Puritan Bennett 560 Ventilator controls, indicators, and interfaces.
Figure 2-1. Control Panel
General Information
Service Manual
Index
1
2
3
Table 2-17. Control Panel Keys and Indicators
Labeling
N/A
Function
LCD panel.
Displays ventilation modes and parameters, waveforms, alarms, and ventilator configuration information.
AC power indicator.
When lit, green
LED indicates ventilator is connected to AC power.
DC power indicator.
When lit, green
LED indicates ventilator is connected to an external DC power source (e.g., automobile cigarette lighter or power port).
2-31
General Information
Index
Figure 2-1
4
5
6
7
8
9
Table 2-17. Control Panel Keys and Indicators (continued)
Labeling Function
Internal battery indicator. When lit, green LED indicates ventilator is operating on internal battery power if no other power source is connected.
LED flashes when the internal battery is charging, and turns off when the battery is charged.
Ventilator status indicator.
When lit, blue LED indicates ventilator is in standby mode, and enables the user to locate the Ventilation ON/OFF key at night.
During ventilation, LED is off.
Ventilation ON/OFF key.
Press and release key to start ventilation.
Press and hold key for 3 seconds to stop ventilation. Confirm stop by pressing the Ventilation ON/OFF key again.
Menu key.
Use this key to switch between the ventilator parameters, alarm settings, and waveform screens.
When a USB memory device is inserted into the USB port, this key activates the
USB memory device screen.
Up/down arrow and unfreeze/freeze waveform keys.
Up arrow key moves the cursor upward in a menu, increases a setting value when a parameter is selected, or activates the waveform display (unfreeze) when viewing waveforms.
Down arrow key moves the cursor downward in a menu, decreases a setting value when a parameter is selected or freezes the waveform on the screen when viewing waveforms.
Enter key. Accepts a setting or enables selections within a submenu.
2-32 Service Manual
Service Manual
General Information
Index
Figure 2-1
10
11
Table 2-17. Control Panel Keys and Indicators (continued)
Labeling Function
Alarm control key.
Press and release once to pause the audio portion of the alarm for 60 s. Press and release twice to activate the alarm paused function.
When alarm paused is active, visual and audible alarm signals are suspended until a new alarm becomes active.
Alarm indicators.
Red LED indicates an active high priority alarm. Yellow
LED indicates active medium priority alarms.
Figure 2-2. Front View
2-33
General Information
2
3
4
5
Index
1
Table 2-18. Front Panel
Labeling
N/A
Function
Exhalation block from-patient port.
Gas return port to which exhalation limb of double-branch patient circuit connects for exhaled volume measurements.
Single patient use: Do not use the exhalation block on multiple patients.
Front air vent.
Allows air flow through the ventilator for electronics cooling.
Exhalation valve pilot port.
Port used to apply pressure to the exhalation valve diaphragm.
Patient proximal pressure port.
Port for measuring patient pressure.
To patient port. Gas output port to which inspiratory limb of doublebranch patient circuit connects.
2-34 Service Manual
Figure 2-3. Rear View
General Information
Service Manual
Index
1
2
3
4
Table 2-19. Rear Panel Labels and Markings
Labeling
N/A
Function
USB ports.
Two Type-A receptacles designed to use a 32-bit formatted USB memory stick.
Carrying handle. Handle is integrated into the top housing.
PC communication port.
One USB
Type B mini receptacle for connection to a personal computer (PC). Used for software uploads and data downloads.
DC power cord connector.
Keyed connector for DC power cord.
2-35
General Information
6
7
8
9
Index
Figure 2-3
5
Table 2-19. Rear Panel Labels and Markings (continued)
Labeling Function
AC power cord connector.
Connector for AC mains power.
I/O
N/A
N/A
ON/OFF switch.
Applies/removes power to the ventilator. Protective cover prevents accidental switch actuation.
Power cord retainer.
Prevents inadvertent power cord disconnection.
Internal battery cover.
Provides access to internal battery.
Oxygen connector. Connector for low pressure oxygen line.
10
11
Nurse call connector.
Connector for nurse call cable.
Air inlet label.
Identifies ventilator air inlet.
See
individual symbols.
2-36 Service Manual
Figure 2-4. Top View
General Information
Service Manual 2-37
General Information
Figure 2-5. Bottom View
2-38
Index
1 a b c
Table 2-20. Bottom Panel Labels
Labeling e d
Function
Product Identification label: a: Device name/model b: Serial number c: Storage temperature range d: Manufacturer name and address, country of origin e: Manufacture date (year)
Service Manual
Figure 2-6. Side View
General Information
Service Manual
Index
1
2
Table 2-21. Side Panel
Labeling
N/A
See
Function
Side air vents.
Allow air flow through the ventilator for electronics cooling.
Exhaled gas outlet port. Allows exhaled flow to exit the ventilator through the patient circuit exhalation valve (when dual limb circuit is used).
2.10 Symbols, Abbreviations, and Internal
Labels
The following table lists symbols and abbreviations found on the Puritan Bennett 560
Ventilator screens or on the device or its labels, and contains labels not referenced in
through
2-39
General Information
Table 2-22. On-screen symbols, abbreviations, and internal labels
Symbol Meaning
Symbols found on ventilator screens
Flow pattern.
In VAC mode only, you can select between square (SQ), descending ramp (D), or sinusoidal (S) flow patterns. In volume ventilation mode, you can select between square (SQ) and descending ramp (D) flow patterns.
Pressure rise times.
In pressure ventilation modes, you can select one of four rise times:
Setting 1 = fastest
Setting 2
Setting 3
Setting 4 = slowest
To avoid invalid settings, the rise time will automatically adjust to ensure the set inspiratory time is less than rise time setting + 300 ms.
Freeze indicator.
Appears on the waveforms screen when the waveform tracing has been frozen.
±
Battery capacity indicator.
Battery capacity appears as a percentage of full charge.
Software lock enabled.
Access to ventilator patient settings, preferences menu, alarm settings, and maintenance mode is not permitted. There is access, however, to the ventilation menu, alarm menu, waveforms menu, USB menu, alarm memory menu, and play/ pause of the breath delivery waveforms.
To lock or unlock the ventilator: With the ventilator in standby or ventilation mode, simultaneously press and hold the up and down arrow keys for 3 seconds.
Cursor.
Shows cursor’s current position.
Available parameter.
Parameter may be selected when device is not locked.
Unavailable parameter.
Parameter cannot be changed when device is locked.
Setting adjust indicator.
Appears when you select a parameter for adjustment.
Breath trigger indicator.
Indicates a patient-initiated inspiration.
Exhalation valve presence indicator.
Indicates an exhalation valve is present in the patient circuit (valve configuration). Displayed on title line of ventilator screen.
2-40 Service Manual
Service Manual
General Information
Table 2-22. On-screen symbols, abbreviations, and internal labels (continued)
Symbol Meaning
Exhalation valve absence indicator.
Indicates an exhalation valve is not present in the patient circuit (leakage configuration). Displayed on title line of ventilator screen.
Alarm off (Apnea off) indicator.
Indicates apnea alarm as been set to
OFF in the Preferences menu. Displayed on title line of ventilator screen.
Alarm paused indicator.
Indicates an alarm paused condition.
Displayed on title line of ventilator screen.
Audio paused indicator.
Indicates an audio paused condition.
Displayed on title line of ventilator screen.
Apnea time
CV (V A/C)
PC (P A/C)
PSV/ (ST)
V SIMV
P SIMV
CPAP
Time interval after which the ventilator detects an apnea condition.
Volume Assist/Control Ventilation mode
Pressure Assist/Control Ventilation mode
Pressure Support Ventilation
Volume Synchronized Intermittent Mandatory Ventilation mode
Pressure Synchronized Intermittent Mandatory Ventilation mode
Continuous Positive Airway Pressure mode
Symbols and markings found on ventilator surfaces or labels
Recycle electrical and electronic items with this marking according to local regulations (EN50419). Do not discard with ordinary household waste. This includes the ventilator and the battery.
Electrostatic sensitive devices label (IEC 60417-5134). Located on inside of lower housing near Power Management PCBA.
Observe necessary precautions when handling devices sensitive to electrostatic discharge (ESD).
Type BF applied part (IEC 60417-5333).
A regulatory standard classification for protection against electrical shock for the part of the device that contacts the patient.
Attention - consult accompanying documents (ISO 7000-0434A).
Consult instructions for use (ISO 7000-1641).
2-41
General Information
Table 2-22. On-screen symbols, abbreviations, and internal labels (continued)
Symbol Meaning
Insulation class II equipment (IEC 60417-5172).
A regulatory standard classification for protection against electric shock. Equipment in which protection against electric shock does not rely on basic insulation only, but in which additional safety precautions such as double insulation or reinforced insulation are provided, there being no provision for protective earthing or reliance upon installation conditions.
CSA certification mark that signifies the product has been evaluated to the applicable Underwriters Laboratories Inc. (UL) and CSA standards for use in the US and Canada.
Ingress protection classification. Protected against access to hazardous parts with a tool. Protected against solid foreign objects of 2.5 mm diameter and greater. Protected against vertically falling water drops.
Do not reuse. Single patient use (ISO 7000-1051).
Manufacturer (BS EN-980)
2009
Manufacture date (ISO 7000-2497).
Storage temperature range (ISO 7000-0632)
Mains warning label (located on Power Management PCBA power supply guard)
2-42 Service Manual
General Information
Table 2-22. On-screen symbols, abbreviations, and internal labels (continued)
Symbol Meaning
Pneumatic cable routing diagram label (located on inside top housing)
Battery installation label (located inside battery cover)
Service Manual
2.11 Serial Number and Software Versions
The serial number can be found on the label on the bottom of the ventilator or on the
Maintenance mode screen. The CPU and Power Management software versions appear on the welcome screen for approximately 10 seconds after turning the ventilator’s power switch on. When the ventilator is in Maintenance mode the CPU and Power Management software versions are displayed on the screen (see
for instructions on how to enter Maintenance mode).
2.12 Ventilator Configuration
The Setup screen allows you to configure the following parameters:
• Language: English, English (US), Spanish, French, Greek, Italian, Japanese, Korean,
Netherlands, Norwegian, Polish, Portuguese, Russian, Finnish, Swedish, Turkish,
Chinese, Danish, German.
• Date: DD MMM YYYY format
• Time: HH:MM:SS format
• Intentional Vent Stop: YES or OFF
• Pressure unit: cmH
2
O, mbar, hPa
• Alarm Tone: Compliant (softer) or Original (louder)
• Patient hours: XXXXX h XX min
• Reset hours: YES or OFF
2-43
General Information
• Restore Defaults: YES or OFF
This parameter allows the user to reset all settings back to the original manufacturer defaults, except for the Language, Date, and Time.
• Maintenance
• Next
Any settings changes are retained in the ventilator’s memory after the ventilator is turned off.
To access the Setup screen and Maintenance mode:
1. Ensure the ventilator’s power switch is in the OFF position.
2. Press and hold the ALARM CONTROL key while simultaneously switching the ventilator to the ON position. Continue holding the key until the Setup screen
Figure 2-7. Setup Screen
2-44
Note:
If the ventilator fails to enter SETUP, stop ventilation by pressing the Ventilation ON/OFF key for 3 seconds and pressing again to confirm stop. Power off the ventilator then repeat steps 1 and 2 above.
To change the language, date, time, or other settings
1. Use the UP or DOWN arrow keys to place the cursor at the desired position and press
ENTER .
2. Use the UP or DOWN arrow keys to change settings.
3. Press ENTER to confirm your selection.
Service Manual
Service Manual
General Information
Note:
•
When a parameter contains several setup fields (e.g., Date, Time) press ENTER to move from one field to the next.
•
If a change is not confirmed with ENTER after 7 seconds, the ventilator restores the previous value.
To change the alarm tone:
1. Use the UP or DOWN arrows to place the cursor on “Alarm Tone.”
2. Press ENTER .
3. Use the UP or DOWN arrows to select Compliant or Original .
The default setting is Compliant. The audible sound of Compliant is softer than the
Original tone, and meets the requirements of standard EN-60601-1-8. Original refers to the alarm tone that was shipped with the ventilator from initial release until the
LX010023/LX010101 software update.
4. Press ENTER to confirm the selection.
To reset patient hours:
1. Use the UP or DOWN arrow keys to place the cursor at the "Patient Hours" position.
2. Press ENTER . The cursor moves to the "Reset Hours “position.
3. Press ENTER . "OFF" flashes.
4. Press the UP or DOWN arrow key to change the message to “YES”.
5. Press ENTER . "YES" appears continuously, a long "beep" sounds, and the patient counter display resets to 00000 h 00 min and the display shows "Reset Hours: OFF".
To restore the default settings:
1. Use the UP or DOWN keys to place the cursor on “Restore Defaults.”
2. Press ENTER . “OFF” flashes.
3. Press UP or DOWN to change “OFF” to “YES.”
4. Press ENTER to reset all settings back to the manufacturer defaults except for
Language, Date, and Time. “OFF” will reappear.
To enter Maintenance mode:
1. Press the DOWN arrow key to place the cursor at the “Maintenance” position.
2. Press ENTER . The Maintenance screen appears.
3. To go back to the Setup screen, press the UP or DOWN arrow key to place the cursor at the “Back” position.
4. Press ENTER to return to the Setup screen.
2-45
General Information
Once the ventilator is in Maintenance Mode, a menu appears with the following
Figure 2-8. Maintenance Screen
1
2
Ventilator software version
Power management CPU software version
3
Ventilator serial number
• Faults check
• Measurement check
• Sensors calibration including a submenu with Patient pressure, Int valve pressure,
Inspiration Flow, and Exhalation Flow
• Turbine speed
• Back
To access the Setup 2 menu:
1. Press the UP or DOWN arrows to place the cursor beside “Next.”
2. Press ENTER . The Setup 2 menu is displayed.
2-46 Service Manual
Figure 2-9. Setup 2 Menu
General Information
Service Manual
The parameters in this menu are:
• Cycling Mode: I:E or I/T
I:E ratio is the ratio of inspiratory time to expiratory time. I/T is the inspiratory time divided by the total breath cycle expressed as a percentage. The mode that uses this value is P A/C.
• Relative Pressure: OFF or YES.
• E Sens Setting: Positive, Negative
• Back
The Back parameter allows the user to return to the Setup menu.
2.13 Service Philosophy
Field service of the ventilator is limited to the service activities described in this manual.
For field service, technical support, or information on technical training, contact your local Covidien representative.
Servicing the Puritan Bennett 560 Ventilator consists of a three-step process
(
).
2-47
General Information
Figure 2-10. Service Process
Troubleshoot ventilator
Repair device or perform preventive maintenance
Final performance verification
1.
Troubleshoot the ventilator (as required) — Determine the cause of the problem and what components or assemblies need to be replaced. Use the
information in Chapter 4 to diagnose ventilator malfunctions.
2.
Repair the ventilator or perform preventive maintenance — See the schedule in
of this chapter for preventive maintenance intervals. Chapter 5
contains step-by-step instructions for disassembly, component repair or replacement, and reassembly.
3.
Final performance verification — Calibrate the ventilator’s sensors and check the ventilator’s performance characteristics after servicing as described in
to ensure all problems have been corrected and no new problems have been introduced.
2-48 Service Manual
3 Theory of Operation
This chapter contains the theory of operation of the Puritan Bennett™ 560 Ventilator and includes the following information:
• overview of ventilator operation and the ventilator’s safety net
• description of the ventilator components and their interactions during breath delivery
• pneumatic system description
• electrical system description.
3.1
Operational Overview
When you turn the ventilator on, Power On Self Test (POST) automatically runs during the first 15 seconds. During this period, the ventilator checks the status of AC and internal battery power sources, tests the LEDs and alarms, and checks the integrity of memory.
The LCD panel displays device and patient hours, ventilator software version, date, time, and copyright information. If POST fails, the ventilator resorts to a safe state where ventilation is not allowed: the exhalation valve opens, the turbine stops, the alarm sounds continuously, and the ventilation stand-by LED flashes. When POST passes, the ventilator remains in Standby mode until you press the Ventilation ON/OFF key to start ventilation. When ventilation is started, the software automatically performs a turbine speed test.
The keyboard is used to enter breath delivery and patient alarm settings into the ventilator and to enter Maintenance mode — a non-ventilatory mode reserved for sensor calibrations and other performance checks. The keyboard is also used to set a software lock to prevent inadvertent changes to the prescribed settings. (Refer to
for instructions on how to enable and disable the software lock.) Ventilator and alarm settings are retained in non-volatile memory during periods when the ventilator is turned off.
Breath delivery is based on a closed loop control system that adjusts the speed of the turbine. The speed of the turbine is controlled to the patient pressure signal or the inspired flow signal depending on whether pressure ventilation or volume ventilation is selected. The turbine speed control algorithms vary according to the ventilation mode, ventilator settings and whether the ventilator is in an inspiratory or expiratory phase. The speed of the turbine adjusts under software control during the expiratory phase to maintain the set PEEP level. Setting the pressure rise time or flow pattern influences the level of turbine acceleration at the start of inspiration. The transition between the
3-1
Theory of Operation inspiratory and expiratory phase is controlled by a deceleration or braking action proportional to the pressure difference between the two phases.
Flow-triggering algorithms, using measurements from an inspiratory flow sensor detect when inspiration begins and also determine the end of the inspiration phase when the ventilator is in PSV ST mode. When altitude compensation is turned on (default setting), the flow measurement is automatically corrected as a function of the atmospheric pressure measured by the ventilator’s internal barometer, btps reference temperature, and standard pressure and temperature references.
The turbine spins continuously throughout the breath, creating flow in the patient circuit. The exhalation valve, located at the outlet of the exhalation block on Puritan
Bennett 560 Ventilator, is fully closed during the inspiratory phase. A three-way solenoid valve on the CPU PCBA controls the pressure to the exhalation valve during the expiratory phase to allow purge and bias flows (created by the spinning turbine) to clear the patient circuit of exhaled gas. When a dual limb patient circuit is used, exhaled flow is measured by an exhalation flow sensor on the CPU PCBA via a port located in the exhalation block. This flow measurement is used to calculate the expired tidal volume.
Inspiratory and exhalation volumes are reported in body temperature and pressure saturated (BTPS) units when altitude compensation is set to YES, and ambient temperature and pressure dry (ATPD) units when altitude compensation is set to OFF. It is recommended to always set altitude compensation to YES. BTPS reference temperature is 37°C, and BTPS reference pressure is 1013 hPa (1033 cmH
2
O).
The various measurement signals used in control and detection are electronically filtered in order to minimize electrical noise and prevent erroneous alarms.
Consult the Puritan Bennett 560 Ventilator Clinician’s Manual for a detailed description of breath delivery in each mode.
3.2
Safety Net
The ventilator uses software and hardware features to prevent and detect conditions that could adversely affect the patient’s safety or the safety of the device.
A multiple priority alarm system is designed into the ventilator that includes operatorselectable low and high limits for breath delivery parameters, and built-in, nonadjustable alarms that can detect and announce unsafe patient conditions and various potential ventilator malfunctions. Alarms are indicated by audible alarm tones, flashing
LEDs on the ventilator’s control panel, and current alarm message displays on the ventilator settings and alarm screens. A secondary audio alarm source is available in the event of failure of the primary alarm. The backup alarm can be tested in the
Measurements Check menu of Maintenance Mode. Past alarms and technical faults can be viewed from the LCD screen or downloaded to a computer using Puritan Bennett™
500 Series Ventilator Test Software. The alarm system is explained in detail in Chapter 4 .
Watchdog monitors exist to ensure correct operation for microprocessors, data acquisition channel integrity, and A/D reference voltage accuracy. Software checks these watchdogs and will trigger alarms if out-of-range values are found.
3-2 Service Manual
Service Manual
Theory of Operation
Sensors measure the internal ventilator temperature and turbine temperature and the
CPU monitors these temperatures within specified limits. Internal temperatures are also used to control the speed of the fan inside the ventilator. As temperatures rise, the fan’s speed increases. If temperatures go beyond safe limits, the ventilator stops ventilation and sounds an alarm.
To guard against potential high pressure conditions, software monitors the pressure transducers and will declare a high pressure alarm if any of the following conditions occur:
• inspiratory pressure is higher than the set high PIP limit
• inspiratory pressure is greater than the set inspiratory pressure + 5 cmH
2
O for pressures less than or equal to 29 cmH
2
O
• inspiratory pressure is greater than the set inspiratory pressure + 10 cmH
2
O for pressures from 30 to 60 cmH
2
O
• inspiratory pressure is greater than 60 cmH
2
O
• expiratory pressure is greater than the limits described above for longer than 100 ms.
An internal pressure sensor monitors the pressure at the ventilator outlet to provide a backup for pressure measurements if the proximal pressure tubing becomes disconnected.
If a high pressure condition occurs during inspiration, the ventilator immediately switches to exhalation. If the high pressure occurs during exhalation, the exhalation valve is opened fully and the turbine stops.
Supplemental oxygen is available via the oxygen inlet connector (maximum 7 psi or 15 lpm). The oxygen inlet connector uses an internal valve to prevent leaks when connecting and disconnecting an oxygen supply.
Ventilator parameters and alarm limits are linked to prevent incompatible settings adjustments.
lists the dependencies for each parameter.
In the event of a ventilator shut-down, the system is designed so that the exhalation valve remains open and that the patient can breathe room air through the turbine inlet and both limbs of the patient circuit.
A software lock is available to prevent inadvertent changes of ventilator settings. (refer to
for instructions on how to enable and disable the software lock.)
A power switch cover is provided to prevent the ventilator from accidentally being switched off, and an alarm is present that annunciates if the ventilator is switched off using the power switch instead of stopping ventilation with the keypad first, then turning off power.
3.3
Ventilator Components
The Puritan Bennett 560 Ventilator provides positive pressure ventilation using an electronically controlled micro-turbine to deliver air and supplemental oxygen to the patient. The two main systems that comprise the Puritan Bennett 560 Ventilator are the pneumatic system and the electronic system.
3-3
Theory of Operation
3.4
Pneumatic System
This section describes the components of the pneumatic system.
pneumatic block diagram of the Puritan Bennett 560 Ventilator, including the patient circuit.
lists the main pneumatic components and their operating ranges.
Table 3-1. Pneumatic Component Operating Ranges
Component
Inspiratory flow sensor
Exhalation flow sensor
Proximal pressure sensor
Inspiratory pressure sensor
Exhalation valve pressure sensor
Barometric pressure sensor
Turbine
Oxygen valve
Operating range
0 to 1000 sccm
0 to 1000 sccm
-10 to 100 cmH
2
O ± 1.5 cmH
2
O
-10 to 100 cmH
2
O ± 1.5 cmH
2
O
-10 to 100 cmH
2
O ± 1.5 cmH
2
O
600 to 1100 cmH
2
O ± 10 cmH
2
O
50,000 rpm max flow range: 0 to 240 lpm
Inlet pressure: 7 psi
Inlet flow: 15 lpm
3.4.1 Inlet Air Filter
The inlet air filter combines a fine particle filter with a foam backing. The filter traps particles at 3.3 microns with 99.999982% efficiency at 30 lpm flow.
3.4.2 Low Pressure Oxygen Inlet/valve
This female inlet connector includes an internal valve to prevent leakage when connecting and disconnecting an oxygen source. The maximum oxygen source pressure is approximately 7 psi (50 kPa) at 15 lpm flow. A special coupling is provided with the ventilator to connect the oxygen supply hose to the inlet connector.
3.4.3 Oxygen Solenoid Valve
The oxygen solenoid valve is a normally closed +24V DC two-way valve that is used to prevent oxygen from accumulating inside the ventilator. It opens when the ventilator is powered on and ventilating, and closes when the ventilator is in Standby or turned off.
The CPU PCBA also commands this valve closed if there is an active DEVICE FAULT 6
(turbine overheat) alarm condition.
3.4.4 Inlet Silencer
Part of the turbine assembly, the inlet silencer reduces noise from the air inlet side of the turbine.
3-4 Service Manual
Service Manual
Theory of Operation
3.4.5 Turbine Assembly
The turbine is a low-inertia blower driven by a brushless DC motor that delivers gas by responding to pressure and flow measurements that are processed on the CPU PCBA.
The turbine spins at a maximum of 50,000 rpm, and can deliver flows of up to 240 lpm and maximum pressures of 70 cmH
2
O to 90 cmH
2
O. A resistive-type temperature sensor resides in the turbine housing. The CPU monitors the temperature and will shut the turbine down if the temperature rises above 70° C while the turbine speed is less than
1000 rpm.
3.4.6 Outlet Silencer
Part of the turbine assembly, the outlet silencer reduces noise from the outlet side of the turbine.
3.4.7 Exhalation Solenoid Valve
The exhalation solenoid valve is an electromagnetic, three-way proportional valve
mounted on the CPU PCBA ( Figure 3-1 ). It is used to control the pressure on the patient
circuit exhalation valve diaphragm throughout the breathing cycle. Pressure supplied from a port at the turbine connects to solenoid valve port 1 (
). Solenoid valve port 2 directs pressure through the exhalation valve port on the inspiratory block to the patient circuit exhalation valve diaphragm, where the pressure is used to completely or partially close the exhalation valve.
The voltage on the exhalation solenoid valve is under closed-loop control on a breath-tobreath basis using measurements from the inspiratory flow sensor. During inspiration, turbine gas flows from port 1 to 2 to apply pressure to the diaphragm to completely close the exhalation valve. At the beginning of exhalation, flow from the turbine is vented from port 2 to 3, relieving pressure on the exhalation valve diaphragm. This allows an initial purge flow and accompanying bias flow to flush the patient’s exhaled gas from the
patient circuit for the duration of exhalation (
Figure A-3 ).During exhalation without PEEP,
the flow path from port 2 to 3 is completely open allowing purge and bias flows. During exhalation with PEEP, flow is directed form port 1 to port 2 to apply a partial pressure to the patient circuit exhalation valve, while port 2 and port 3 are partially open simultaneously allowing purge and bias flows. The maximum bias flow does not exceed
40 lpm after a 15 breath-cycle stabilization period.
3-5
Theory of Operation
Figure 3-1. Exhalation Solenoid Valve
3-6
3
4
1
From turbine
2
To tee intersection with inspiratory block, exhalation valve port, and exhalation pressure transducer
To atmosphere
Exhalation solenoid valve
3.4.8 Inspiratory Block
The inspiratory block provides the means for inspiratory flow and pressure measurements. It includes a screen which creates a small pressure drop used to divert a portion of the inspired gas to the inspiratory flow sensor. This gas flows through a port on the upstream side of the screen, through the flow sensor, back into the gas stream through a port on the downstream side of the screen, then out the ventilator’s To Patient port. Pressure ports for inspiratory, proximal, and exhalation pressure measurements are also located in the inspiratory block.
3.4.9 Inspiratory Flow Sensor
A mass air flow sensor located on the CPU PCBA measures inspiratory flow. A portion of the inspired flow is measured by the sensor which contains a heated sensing element and internal circuitry used to determine the flow rate. The CPU uses this flow measurement to control the turbine during volume ventilation, determine when an inspiratory flow trigger occurs (based on the inspiratory sensitivity setting), determine
Service Manual
Service Manual
Theory of Operation the beginning of exhalation (based on the expiratory sensitivity setting), and to calculate inspired tidal volume.
3.4.10 Proximal Pressure Sensor
The proximal pressure sensor, located on the CPU PCBA measures pressure at the patient wye via a tube that connects from the wye to the Proximal pressure port on the inspiratory block. This pressure measurement is used in the feedback loop to control the turbine speed during pressure ventilation modes.
3.4.11 Inspiratory Pressure Sensor
The inspiratory pressure sensor, located on the CPU PCBA, measures pressure of the gas as it exits the ventilator. The ventilator substitutes this pressure measurement for the proximal pressure measurement as a safety backup in the event that the proximal pressure tube becomes disconnected.
3.4.12 Exhalation Valve (Internal Valve) Pressure Sensor
The exhalation pressure sensor, located on the CPU PCBA, is used as a safety feature to detect the presence and proper operation of the patient circuit exhalation valve. The exhalation valve port on the inspiratory block provides the pathway for the pressure measurement.
3.4.13 Exhalation Block
On the Puritan Bennett 560 Ventilator (when using a dual limb patient circuit), the exhalation block provides the means to measure exhaled flow. A pressure drop is created as exhaled gas passes through an orifice in the exhalation block. Pressure taps, located on either side of the orifice, provide the connections to the exhalation flow sensor on the
CPU PCBA. The measured flow is used to calculate exhaled tidal volume.
3.4.14 Exhalation Flow Sensor
The exhalation flow sensor is a mass air flow sensor that measures exhalation flow upstream of an orifice in the exhalation block. A portion of the exhalation flow is diverted through the sensor which contains a heated sensing element that converts the mass of air flowing past the element into a voltage. The flow is returned to the exhalation block after it passes through the exhalation flow sensor. The CPU uses the flow measurement to calculate exhaled tidal volume.
3.4.15 Barometric Pressure Sensor
The barometric pressure sensor is used to correct flow measurements when the device’s
Altitude Compensation feature is enabled. Covidien recommends that Altitude
Compensation always be set to YES in the Measurements Check screen of Maintenance
Mode.
3-7
Theory of Operation
3.4.16 Patient Circuit
The Puritan Bennett 560 Ventilator can be used with a single- or double-limb patient circuit. Both circuits use an exhalation valve to allow ventilation with a set PEEP level. The double-limb circuit allows for exhaled tidal volume measurements by routing exhaled gas through the exhalation block. A special patient circuit having only an inspiratory limb with a proximal pressure line connected and no exhalation valve must be used when NIV ventilation in CPAP or pressure support mode is selected.
A humidifier and water traps can be used with either circuit as shown in Figure 3-2 and
. The patient must wear a vented NIV interface if a single-limb circuit without an exhalation valve is used.
Figure 3-2. Ventilator with Single-Limb Circuit and Test Lung
3-8
1 Inspiratory bacteria filter
2 Humidifier
3 Water trap
4 Single limb patient circuit with exhalation valve
Service Manual
Figure 3-3. Ventilator with Dual-Limb Patient Circuit and Test Lung
Theory of Operation
Service Manual
1 Inspiratory bacteria filter
2 Humidifier
3 Water traps
4 Dual limb patient circuit
5 Exhalation bacteria filter
6 Exhalation valve at outlet of exhalation block
3.5
Electrical System
This section describes the components of the electrical system and their functions.
3-9
Theory of Operation
3.5.1 Power Management PCBA
The Power Management PCBA performs the following functions in the Puritan
Bennett 560 Ventilator:
• converts AC voltage to DC voltage
• converts, regulates, and distributes DC voltages
• manages internal battery charging and discharging
• automatically adapts battery management to the battery capacity identified
• prioritizes the use of AC, external DC and internal battery power sources
• provides USB mini type B interface for Covidien service personnel to upload application software to the ventilator
• provides two Type A USB connectors used for obtaining trended information which can be transferred to a PC
• acquires and responds to voltage, current, and temperature measurements via a microcontroller.
shows a block diagram of the Power Management PCBA.
3.5.2 AC to DC Conversion
AC power enters the Power Management PCBA through the mains connector with nominal input voltage and frequency ranges of 90 to 250 V AC and 50 to 60 Hz, respectively. The AC power is converted to 33 V DC (nominal) and supplies up to 3 A current.
3.5.3 DC Voltage Conversion and Distribution
If AC power is unavailable, the ventilator may be connected through the DC input connector on the Power Management PCBA to an external DC power source (12 to 30 V
DC), such as an automobile power port, or will obtain its power directly from the internal battery (26 V, nominal). Whether the DC voltage is supplied from AC, external DC, or internal battery, it is converted to four regulated voltages:
• 24 V — Used to supply the turbine and fan directly from the Power Management
PCBA, and distributed to the CPU PCBA. The Power Management PCBA monitors this voltage to ensure it remains within the range of 22 V to 26 V. The current output is limited to 4 A.
• 5 V — Used by the Power Management PCBA electronics and distributed to the CPU
PCBA for event memory. The Power Management PCBA monitors this voltage to ensure it remains within the range of 4.5 to 5.5 V. The current output is limited to 1 A.
• 3.3 V — Used for RAM voltage supply and distributed to the CPU PCBA for monitored parameter memory. The Power Management PCBA monitors this voltage to ensure it remains within the range of 3.0 to 3.6 V. The output current is limited to 500 mA.
• 3 V ref — Reference for voltage measurements.
3-10 Service Manual
Service Manual
Theory of Operation
3.5.4 Internal Battery Charging and Discharging
The Power Management PCBA manages internal battery charging and discharging using a combination of voltage, current, and temperature measurements, and charging time.
Additional protections are in place to prohibit battery charging if charging current and voltage out-of-bounds conditions are detected.
Battery charging occurs in two phases, and only when the ventilator is connected to an AC power source and when battery voltage drops below 28.5 V. In the first phase, the battery is charged using 500 mA (when ventilator is operating) or 1.5 A (when ventilator is not operating) until the battery reaches 29.4 V ± 1%. The second phase maintains this voltage and monitors the current. When the current drops to below 150 mA, the charging circuit turns off. When the battery discharges to 28.5 V, the charging cycle starts again.
Over-charge voltage protection ensures that battery charging is canceled if battery voltage exceeds approximately 30 V. Over-discharge voltage protection prevents the battery from discharging below approximately 14 V. Battery usage is not allowed until the battery is charged to greater than approximately 19 V.
Battery temperature also affects the ability of the battery to charge and discharge.
Software controls are in place to discontinue battery charging if the battery temperature is greater than 45° C or less than 0° C. Software also prevents the battery from being used as the power source if the battery temperature is greater than 60° C or less than -10° C.
Software estimates the battery capacity (mAh) delivered during charging and compares it to a maximum threshold value. If the estimated capacity is greater than the maximum value, the software determines that the charging time is too long, and a new battery charge cycle is not allowed.
Battery discharging cycles are calculated every 5% of battery discharge, and are a function of the battery capacity at the start of a discharge, the remaining battery capacity, and the number of previous discharges (including partial discharges). The number of discharging cycles can be read from the Internal Battery Menu with the ventilator in Maintenance mode. The battery should be replaced when battery capacity is less than 3840 mAh.
3.5.5 Power Source Priority and Switching
The Puritan Bennett 560 Ventilator can operate using Mains AC, DC power supplied by an external DC power source (e.g., automobile power port), or DC power supplied by an internal battery. The Power Management PCBA drives the LEDs on the ventilator’s control panel that indicate which power source is in use.
When the ventilator is running using an external DC power source, the battery charging circuit does not function. When the ventilator is connected to more than one power source, the Power Management PCBA ensures that the resultant power source is available according to the priority shown in
. In the table, presence means that there is at least the minimum power requirement available from the indicated source.
During ventilator operation, if AC mains power is lost, the ventilator will automatically switch to external DC power, if available, or internal battery. When the ventilator is running on internal battery, an audible signal is annunciated every hour as a reminder
3-11
Theory of Operation that the ventilator is running on battery power. A battery “fuel gauge” appears on the
LCD panel letting you know the approximate capacity left in% or time units.
Table 3-2. Power Source Switching Priority
AC present
–
External DC power source present
–
Internal battery present
Resultant power source used for ventilator operation
AC
External DC power source
Internal battery –
–
–
–
–
–
–
–
–
–
AC
AC
AC
External DC power source
No ventilation
3.5.6 USB Interface
Two Type A and one mini B USB ports are provided on the Power Management PCBA. The type A ports are used for obtaining trended, event, and monitoring information. Only one USB port may be used at a time, or an alarm message will annunciate. Up to 12 months of trended ventilator data may be stored on a USB key (128 MB minimum memory and 32-bit file format required). This information may be transferred to a PC. The mini B port is used for performance verification with the Puritan Bennett™ 500 Series Ventilator Test
Software and for uploading application software to the ventilator. The mini B port is used as a communication interface between a PC and the Power Management PCBA. A 115 K Baud rate is supported.
WARNING
Always verify that the serial number of the ventilator is correctly associated with the correct patient before using a USB memory device to transfer data between the ventilator and a PC.
3.5.7 Microcontroller Functions
A PIC microcontroller on the Power Management PCBA processes data from the battery’s memory and temperature sensor, the internal ventilator temperature sensor (CT1), and voltage regulators. These data are communicated to the CPU PCBA via a Serial Peripheral
Interface (SPI) bus and are monitored to determine the state of the ventilator.
The microcontroller reads the internal ventilator temperature sensor and the battery temperature sensor and uses these readings to control the speed of the cooling fan, which maintains proper internal operating temperatures for the electronics.
3-12 Service Manual
Service Manual
Theory of Operation
During events such as ventilator operating software uploads or ventilator event log downloads, the Power Management PCBA microcontroller transmits PC communication data (at a 256 K Baud rate) to the CPU PCBA.
At ventilator startup, during POST, the Power Management PCBA is responsible for temporarily disabling alarms. If an error is detected, the three power source LEDs flash, alarm functions are restored and a Very High Priority alarm sounds. The ventilator is prevented from entering Standby mode and ventilation is disallowed.
3.5.8 CPU PCBA
The CPU PCBA performs the following functions:
• controls breath delivery functions of the ventilator
• displays information to the user through the LCD panel
• communicates with the Power Management PCBA
• provides continuous status and monitoring for errors
• stores ventilator settings, patient data, and events in memory
• interfaces with external devices.
3.5.9 Breath Delivery Functions
Software running in the microcontroller on the CPU PCBA processes ventilator and alarm setting inputs from the keyboard. A turbine interface circuit uses the ventilator settings to generate a pulse-width-modulated (PWM) signal that controls the turbine speed appropriately for inspiration and exhalation. The microcontroller detects an inspiratory trigger when a combination of the current and previous inspiratory flow sensor measurements are greater than the inspiratory sensitivity setting in all modes except for
CPAP, in which case the inspiratory flow readings are compared with an average flow.
Exhalation occurs after the rise time has elapsed and when inspiratory flow has decreased from peak flow to a predetermined percentage of peak flow (or the set expiratory sensitivity in PSV ST). Signals from the on-board flow transducers provide feedback upon which the turbine adjusts its speed to maintain the set volume or pressure and PEEP level.
A PWM signal drives the exhalation solenoid valve on the PCBA that controls the patient circuit exhalation valve. The voltage to the exhalation solenoid valve varies depending upon whether the breath cycle is in the inspiratory or exhalation phase. During the inspiratory phase, 24 V is applied to the valve which allows all flow to pass from port 1 through port 2, keeping the patient circuit exhalation valve closed. During exhalation, the voltage is modulated to allow flow through ports 2 and 3 (if no PEEP is set), or
simultaneously through ports 2 and 3 and ports 1 and 2 (if PEEP is set). (See Figure A-3 .)
During ventilation, a PWM signal opens a 2-way solenoid valve to allow the use of supplemental oxygen. The valve remains open as long as the device is in ventilation mode. It closes when the ventilator is turned off or in standby mode, or during a TURBINE
OVERLOAD alarm condition which can occur if there is a combination of high turbine temperature and low turbine speed.
3-13
Theory of Operation
3.5.10 LCD Panel Interface
The LCD panel connects to the CPU PCBA through a parallel bus and provides a visual display of ventilator and alarm settings, breath waveforms, and service information.
3.5.11 Power Management PCBA Communication
Communication between the CPU PCBA and the Power Management PCBA occurs through an SPI bus. The CPU PCBA receives AC, DC, and battery supply voltage status from the Power Management microcontroller and relays this information to the keyboard power LED indicators.
A UART bus enables communication between the CPU and a PC through the Power
Management PCBA. A 115.2 kBaud data transfer rate for event log downloads is supported, while software updates can occur at up to 256 kBaud. An analog signal is supplied by the Power Management PCBA for transmitting the battery voltage to the
CPU PCBA.
3.5.12 Status and Error Monitoring
The CPU PCBA continuously monitors signals and voltages to determine the status of the ventilator. A watchdog circuit detects proper execution of processor instructions by receiving a pulse from the microcontroller at specified time intervals. If the watchdog does not receive the pulse at the appropriate time, a reset signal is generated, after which
POST runs. If POST passes, ventilation continues as previously set. During POST, the ventilator is in its safe state, with turbine stopped and exhalation valve open. If POST fails, a Very High Priority alarm occurs and ventilation is not allowed.
Reference voltage monitors on the CPU PCBA detect the status of 5 V, and 10 V reference voltages. If a drop in either of these voltages beyond 0.5 V, or 1.1 V, respectively, occurs, a digital signal is generated and an alarm is annunciated. A monitor for the 3.3 V supply voltage will generate a digital signal and alarm if a drop of greater than 0.3 V is detected.
3.5.13 Memory
There are four types of memory on the CPU PCBA: ventilator settings memory, event memory, monitoring memory, and random access memory (RAM).
Ventilator parameter and alarm settings are stored in a minimum of 4 kB of non-volatile memory to ensure settings are retained when the ventilator’s power is turned off or during a loss-of-power condition.
A minimum of 512 kB of non-volatile event memory stores the event log. This log records information such as ventilator stops and starts, confirmed ventilator and alarm settings, and alarm and technical fault history including any associated audio paused or alarm paused key-presses, alarm resets, and acknowledgment key presses. This information is also retained when the ventilator is turned off and during power losses.
A minimum of 128 MB of non-volatile monitoring memory capacity is available for recording monitored patient data. These data are linked to the time of occurrence of
3-14 Service Manual
Service Manual
Theory of Operation associated events. A minimum of 64 KB of RAM is available for storage of operating software program variables. The data stored in RAM are not saved during periods when the ventilator is turned off.
3.5.14 Turbine Control PCBA
The Turbine Control PCBA controls turbine speed and braking and transmits turbine speed and temperature feedback signals to the CPU PCBA. The Turbine Control PCBA also monitors the current in the turbine motor windings and the position of the motor via a position sensor input signal from the motor. In the event of a turbine malfunction, fault detection signals are generated from the Turbine Control PCBA and transmitted to the CPU PCBA. A 24 V input from the CPU PCBA to the Turbine Control PCBA produces a
15 V output to drive the motor controller logic devices.
3.5.15 Buzzer PCBA
Two alarm buzzers are located on the buzzer PCBA. The buzzer PCBA is capable of producing 65 - 80 dBA alarm tones based on the alarm priority. A 4.8 V Ni-MH battery is mounted on the buzzer PCBA to power the ventilator’s POWER SUPPLY LOSS alarm. The battery is charged to a maximum voltage of 5.6 V. At this point the charger switches to a trickle-charge mode which maintains the battery voltage.
3.5.16 Battery Connection PCBA
The Battery Connection PCBA provides the physical connection between the battery and the Power Management PCBA. A connector with spring-loaded “fingers” on this PCBA touches the battery contacts providing DC power to the Power Management PCBA and allowing the battery temperature and memory to be read.
3.5.17 Internal Battery
The rechargeable internal battery is a 14-cell lithium-ion battery that includes a temperature sensor and a memory device which stores information, such as battery identification, capacity, number of discharge cycles, manufacture date and time and battery cell type. The ventilator displays this information in the Internal Battery menu, accessible through Maintenance mode and Measurement Check. The battery identification information is checked by the Power Management PCBA and if the stored information is incorrect, or if the data are corrupt, a BATT FAULT 3 alarm will occur.
3.5.18 Ventilator Compartment Temperature Sensor
This sensor, located on the Power Management PCBA, measures the air temperature inside the ventilator. The temperature signal is used to control the speed of the fan in the ventilator compartment.
3-15
Theory of Operation
3.5.19 Fan
A fan, located in the main ventilator compartment and controlled via the Power
Management PCBA, produces air flow through the ventilator and battery compartments to cool the electronics and to expel any oxygen vented through the exhalation solenoid valve to the outside of the ventilator.Room air enters the ventilator through the vents in the battery cover at the rear of the device and exits through the cooling vents in the front and sides where the upper and lower housings meet (see
and
).
The fan runs continuously when the ventilator is connected to AC power regardless of whether the ventilator is powered off, powered on but in Standby mode, or ventilating. If the external or internal battery is used, the fan runs any time the power switch is turned on. The fan’s speed adjusts to maintain an internal ventilator temperature of no greater than 60° C (measured on the Power Management PCBA), and a battery temperature of no greater than 40° C. Additionally, the fan will run at maximum speed if the internal ventilator temperature sensor should fail.
3.6
Ventilation Features
3.6.1 Vt target
The Puritan Bennett 560 Ventilator has the Vt target (Target Volume) feature as a part of pressure-based ventilation modes in non-invasive ventilation without an exhalation valve. This allows a range of pressures to be used to reach a set volume. If the mode is PSV
ST or P A/C, you may select Vt Target within the range of 50 to 2000 mL (or OFF) which controls the inspired tidal volume to the target value you specify. An increasing or decreasing series of small pressure adjustments are applied and the inspired volume is measured. If the inspired volume is lower than the target volume, the pressure of the next breath increases a small amount and the inspired volume is measured again. If the inspired volume is higher than the target volume, the pressure of the next breath is decreased slightly, until the inspired volume matches the target volume. The pressure increases stop if the maximum inspiratory pressure is reached.
3.6.2 Leak Compensation
Leak compensation is available on the Puritan Bennett 560 Ventilator in pressure-based ventilation modes. Leak compensation is used with non-invasive ventilation with a vented mask. It is normal for vented masks to have a known leak associated with them.The leak from the mask is estimated at the end of exhalation and this leak is taken into account in the patient’s inspired volume measurements.
3.6.3 Circuit Detection and Management
The Puritan Bennett 560 Ventilator can check the configuration of the patient circuit and high priority alarms will sound when the ventilator settings are incompatible with the circuit used. Some examples are:
• inspiratory pressure greater than 30 without an exhalation valve in the patient circuit
3-16 Service Manual
Service Manual
Theory of Operation
• exhalation pressure less than 4 without an exhalation valve in the patient circuit
• using CPAP mode with an exhalation valve in the patient circuit
• using volume ventilation without an exhalation valve in the patient circuit
• inspiratory or exhalation pressure difference less than 5 without a valve in the patient circuit.
These alarms will sound if a mode change is made that violates any of the aforementioned conditions. Upon detection of the incompatible circuit, the ventilator will not start ventilation, but will immediately detect when the circuit has been changed.
3.6.4 Relative or Absolute Pressure
You can determine how you want to view pressure settings — in the relative or absolute mode. For example, if you configure the ventilator for relative mode, and pressure support or pressure control is set to 10 cmH
2
O and PEEP is set to 5 cmH
2
O then the inspiratory pressure delivered will read 15 cmH
2
O. If absolute pressure is selected with the same settings, the inspiratory pressure is 10 cmH
2
O. This also applies to P Max in target Vt. The pressure convention is selected in the SETUP menu. Refer to
open the SETUP menu and configure the ventilator for the desired pressure display.
Relative or absolute pressure is different than the pressure units the ventilator uses for displayed pressure. The pressure units can be selected in the SETUP menu with cmH
2
O, hPa, or mbar available. Peak Inspiratory pressure and End Inspiratory Pressure are calculated as follows:
Peak Inspiratory Pressure during pressure rise shall not be more than (P SUPPORT or Pi +
PEEP) if Relative pressure is YES or (P SUPPORT or Pi) if Relative pressure is OFF.
End inspiratory pressure shall be within the setting (P SUPPORT or Pi + PEEP) ± 1 cmH
2
O
+ 10%) if Relative pressure is YES or (P SUPPORT or Pi) ± 1 cmH
2
O + 10%) if Relative pressure is OFF. The pressure overshoot during the pressure fall phase shall not be less than PEEP x 0.8 - 1 cmH
2
O.
3.6.5 Invasive or Non-invasive Ventilation
The ventilator can be used invasively or non-invasively. If used invasively, the ventilator is said to be used in valve configuration. Valve configuration means that the ventilator has an exhalation valve attached to the ventilator at the exhalation port.The inspiratory volume should be greater than or equal to 150 mL with no leak present. If used noninvasively, leak configuration means that there is no exhalation valve used — there is a known leak in the patient system.
3.6.6 FiO
2
for Various Oxygen and Ventilator Settings
contains a chart showing the relationship between oxygen flow and inhaled flow for various oxygen concentrations and PEEP settings for the Puritan Bennett 560
Ventilator.
3-17
Theory of Operation
Figure 3-4. FiO
2
for various Oxygen and Ventilator Settings x y
O
2
flow (l/min)
Inhalation flow (l/min)
Note:
Tests conducted in a valve configuration. Results can vary according to whether the circuit is configured with or without a valve and patient lung characteristics.
WARNING
The Puritan Bennett 560 Ventilator can be used with an oxygen analyzer with minimum and maximum concentration alarms. Always measure the delivered oxygen with a calibrated oxygen analyzer that features a minimum and maximum concentration alarm in order to ensure that the prescribed oxygen concentration is delivered to the patient.
3-18 Service Manual
4 Alarms and Troubleshooting
This chapter provides information on how to view and respond to alarms and technical faults and suggests ways to troubleshoot the Puritan Bennett™ 560 Ventilator
The ventilator responds to fault conditions with two types of alarms:
• technical faults
• ventilation alarms.
Technical faults indicate conditions that do not immediately affect ventilator operation, but need correcting to restore proper ventilator function. Technical faults do not produce audible alarms but are recorded in the ventilator’s Technical Fault log which is accessed through the Faults Check menu in Maintenance mode. A service technician is required to correct technical faults.
Ventilation alarms are those that audibly and visually alert the user to degraded patient conditions or a ventilator malfunction and require immediate attention. When this type of alarm occurs, a message appears on the LCD panel and is recorded in the ventilator’s alarm log. Many ventilation alarm limits are adjustable and are set after entering the patient’s ventilation parameters. (Refer to the Alarm settings section of
of adjustable alarms.) Other ventilation alarms, such as AC power disconnection and low battery cannot be adjusted. Most ventilation alarm conditions can be responded to and corrected by the clinician or caregiver. In some cases, however, ventilation alarms can
indicate problems with the ventilator hardware or patient circuit. See Section 4.4.2
information on troubleshooting these types of ventilation alarms.
4.1
Alarm Classification
Alarms are classified according to how their causes affect ventilator operation.
lists alarm classifications and associated ventilator responses.
4-1
Alarms and Troubleshooting
Classification
Very High Priority
(VHP)
High Priority
(HP)
Medium Priority
Table 4-1. Alarm Priority Classification
Effect on ventilation and ventilator response
No ventilation. Depending upon the cause, the ventilator always generates a continuous audible alarm at maximum volume and the following indicators may or may not be displayed:
• red continuously lit LED
• alarm message displayed on ventilator screen
• illuminated backlight.
Compromised ventilation. Ventilator responds with:
• a repeating series of 3 beeps then 2 beeps at the user-selected volume
• red flashing LED illumination
• alarm message
• illuminated backlight.
Note: If the cause of an alarm is not corrected, and if the Alarm
Control key is not pressed within 60 seconds of an alarm trigger, high priority alarms will sound at the maximum volume.
Ventilation not immediately affected, but could harm patient if left uncorrected. Ventilator responds with:
• a repeating series of 3 beeps at the user-selected volume
• yellow flashing LED illumination
• alarm message
• illuminated backlight.
There are currently no Low Priority alarms Low Priority
4.2
How to Respond to Alarms
When an alarm condition is detected, visual and audible indicators occur according to the alarm classification described in
Table 4-1 . The alarm display message appears only
on the ventilator parameters or alarm settings menus. If the waveforms screen is being viewed when an alarm occurs, the display will switch to either the ventilator parameters or alarm settings menu (
4-2 Service Manual
Service Manual
Ventilator parameters menu
Figure 4-1. Alarm Display Messages
Alarm
Alarm settings menu
Alarms and Troubleshooting
Alarm message area
If multiple alarms occur at the same time, the LED indicator and alarm sound of the highest priority alarm takes effect, but messages for all active alarms scroll in succession.
4.2.1 How to Pause the Audible Portion of an Alarm
To pause the audible portion of an alarm (audio paused):
Press the ALARM CONTROL key once.
This pauses the audio portion of the alarm for all active alarms for up to 60 seconds and
displays the Audio Paused symbol on the screen ( Figure 4-2 ). The
ALARM CONTROL key also pauses the audible portion of a remote alarm (nurse call) for 60 seconds if connected. Alarm messages and LED indicators remain active during this 60-second period. The Audio Paused period expires after 60 seconds or immediately, when a new alarm condition is detected.
4-3
Alarms and Troubleshooting
Figure 4-2. Alarm Settings Menu During Audio Paused Period
Audio paused symbol
4.2.2 How to Pause an Alarm
Most alarms are automatically cleared as soon as the conditions causing them no longer exist. The exceptions are the Apnea alarm, which clears after two patient breaths, and the
High Pressure alarm, whose LED indicator and alarm message remain active after the high pressure condition is corrected.
Some alarms, however, can be paused manually even if the cause of their activation remains (these alarms are identified in
To manually pause an alarm (alarm paused):
Press the ALARM CONTROL key twice.
The alarm sound and LED indicator are suspended and the Alarm Paused symbol appears at the top right of the Ventilator Parameters, Alarm Settings, and Waveforms screens
(
Figure 4-3 ). The alarm message disappears from the Ventilator Parameters screen.
When no other alarms are currently active, the last alarm paused appears continuously in the alarm message window in the Alarms Settings menu, along with the date and time of its activation.
4-4 Service Manual
Service Manual
Figure 4-3. Alarm Settings Menu During Alarm Paused Condition
Alarms and Troubleshooting
Alarm paused symbol
4.2.3 How to Reactivate a Paused Alarm
Alarms whose causes continue to exist can be reactivated so that audible and visual indicators are restored.
To reactivate a paused alarm:
1. If not currently displayed, press MENU
to access the Alarm Settings menu.
2. Press the UP or DOWN arrow key to place the cursor at the “Alarm Log” position.
3. Press ENTER
, to confirm your selection. The alarm log screen appears ( Figure 4-4
).
4-5
Alarms and Troubleshooting
Figure 4-4. Alarm Log Screen With and Without Alarms
4-6
4. Press the UP arrow key to place the cursor at the “REACTIVATE PAUSED ALARM” position.
5. Press and hold ENTER for at least three (3) seconds. The following events occur:
• a beep tone sounds
• the alarm sounds
• the alarm LED indicator flashes
• the messages of all active alarms are displayed in succession in the Ventilation
Parameters and Alarm Settings menus
• the Audio Paused symbol disappears (if it was displayed)
• the Alarm Paused symbol disappears.
6. Exit the alarm log screen by pressing the down arrow key to place the cursor at the
“Back” position.
7. Press ENTER to confirm your selection. The display returns to the Alarms Settings menu.
4.2.4 Alarm Reset Function
The Alarm Reset function applies only to the High Pressure alarm. A High Pressure alarm occurs when the inspiratory pressure exceeds the high pressure alarm threshold set in the Alarm Settings menu.
During a high pressure condition, the audible High Priority alarm sounds, the red LED flashes, and an alarm message displays on the screen, flashing in inverse video. If the high pressure condition corrects itself, the audio portion of the alarm resets (becomes silent), but the red LED and the alarm message remain flashing.
To manually reset the High Pressure alarm, which turns off the illuminated LED and clears the flashing alarm message, press the ALARM CONTROL key twice.
Service Manual
Alarms and Troubleshooting
4.3
How to Access Stored Ventilator Diagnostic
Logs
The Puritan Bennett 560 Ventilator stores information about its performance that can be used to troubleshoot and correct ventilator malfunctions. The Alarm, Technical Fault, and
Event logs provide this information.
The Alarm and Technical Fault logs are viewed directly from the ventilator screen.
Additionally, the Event log can be downloaded to a PC. Contact your Covidien service representative for more information.
The Alarm, Technical Fault, and Event logs are stored in non-volatile memory on the CPU
PCBA, ensuring that the information is retained when the ventilator is powered off and during power loss conditions.
4.3.1 Alarm Log
The Alarm log can be viewed from the ventilator screen when the ventilator is in standby mode or ventilation mode. This log displays up to eight (8) of the most recent ventilation
alarms, including the date and time of occurrence ().
To view the Alarm Log screen:
1. Follow steps 1-3 of Section 4.2.3
. The alarms appearing in the log are described in
.
Figure 4-5. Alarm Log Screen With and Without Alarm Information
Service Manual
To exit the Alarm Log screen, use the UP or DOWN arrow key to place the cursor at the
“Back” position and press ENTER . The Alarm Log screen is dismissed automatically when a new alarm occurs, or after 15 seconds if no key is pressed.
4-7
Alarms and Troubleshooting
4.3.2 Technical Fault Log
The Technical Fault log can be viewed when the ventilator is in Maintenance mode, and displays up to nine (9) of the most recent technical faults including date and time of
occurrence, and machine hours. A list of technical faults is provided in Table 4-2 .
If the software lock is enabled (a key symbol appears in the upper left corner of ventilator screen), you must disable it before you can view the technical fault log.
To unlock the ventilator software:
1. Simultaneously press and hold the UP or DOWN arrow keys for 3 seconds. The key symbol in the upper left corner of the display disappears, indicating that the software is unlocked.
To enter Maintenance mode and view the technical fault log:
1. If the ventilator is ventilating, stop ventilation via the Ventilation ON/OFF key prior to turning the ventilator power switch OFF.
2. Press and hold the ALARM CONTROL key while simultaneously switching the ventilator to the ON position. Continue holding the key until the Setup screen appears.
3. Use the UP or DOWN key to place the cursor at the Maintenance position.
4. Press ENTER . The maintenance screen appears (
).
The Maintenance screen displays the CPU and Power management CPU software versions, number of ventilator hours, serial number, and a menu of checks and calibrations designed to troubleshoot and verify the performance of the ventilator.
Figure 4-6. Maintenance Screen
4-8 Service Manual
Alarms and Troubleshooting
5. After entering Maintenance mode, the cursor’s default position is Faults Check. Press
ENTER
to view the list of technical faults ( Figure 4-7 ). If there are no technical faults in
memory, the message “NO DATA” appears on the screen.
Figure 4-7. Faults Screen With no Faults Present
Service Manual
4.3.3 Event Log
An event log, capable of storing up to 5000 entries, is maintained in the ventilator’s nonvolatile memory. After 5000 entries are stored in the event log, the ventilator overwrites the data using a first-in-first-out (FIFO) strategy.
The event log stores the following information:
• ventilation starts and stops
• confirmed ventilator parameter settings
• confirmed alarm settings
• alarm and technical fault occurrences with associated actions such as audio paused, alarm reset, and alarm paused actions, and acknowledgment key presses.
Each event is identified by its name, the type of event (technical fault or alarm event), a description of the event, and a date and time stamp.
The event log records any changes to the system's real-time clock by logging the current date/time followed by the new date/time and a unique event code indicating the change.
4.4
Troubleshooting
This section provides guidance on how to troubleshoot and correct technical faults and ventilation alarm conditions that may have been caused by hardware or patient circuit malfunctions.
Table 4-2 lists technical fault error codes and
4-9
Alarms and Troubleshooting ventilation alarm messages along with suggested troubleshooting steps.
WARNING
Some troubleshooting steps require working with exposed electronics connected to AC or battery power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools (except for electrical measuring devices such as a DMM) from touching electrical components.
4.4.1 Troubleshooting Technical Faults
Access the technical fault log using the information provided in Section 4.3
Table 4-2. Technical Fault Error Troubleshooting Guide
Error code n01 n02 n03
Description
Constant flow for 1 minute; faulty inspiratory flow sensor
Inspiratory flow sensor calibration fault
Exhalation flow sensor calibration fault
Troubleshooting steps
1. Calibrate inspiratory flow sensor
(see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Calibrate the inspiratory flow sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Verify correct patient circuit is used.
2. Re-run exhalation flow sensor calibration.
3. If running exhalation flow sensor calibration, try a new external exhalation valve and/ or a new patient circuit.
4. Calibrate inspiratory flow sensor
(see
5. Re-calibrate the exhalation flow sensor.
6. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
7. Replace exhalation block (see
8. Replace CPU PCBA (see
).
4-10 Service Manual
Service Manual
Alarms and Troubleshooting
Error code n04 n05 n06 n07
Table 4-2. Technical Fault Error Troubleshooting Guide (continued)
Description
Internal valve (exhalation solenoid valve) pressure sensor calibration fault
Internal pressure sensor calibration fault
Abnormal turbine speed measurement-
Defect sensor speed
Loss of clock parameters.
Troubleshooting steps
1. Calibrate internal valve pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Calibrate internal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Perform Flow Sensor Capacity test (see
2. Check that the turbine motor wiring harness is connected to the turbine control PCBA (J2), and that the cable between the turbine control PCBA (J22) and the CPU PCBA (J4) is connected.
3. Verify 24V is present on the turbine control PCBA connector
J1 pin1 (-) and pin 2 (+).
4. Verify 24V output on Power
Management PCBA [connector
J9, pin 1 (-) and pin 2 (+)]
5. Check fuse F1 on turbine control PCBA. Replace if necessary.
6. Replace turbine control PCBA/
CPU cable.
7. Replace turbine (see
).
8. Replace turbine control PCBA
(see
).
9. Replace CPU PCBA (see
).
1. Reset the date and time (see
2. Replace the 3V battery on the
CPU PCBA (see
).
3. Replace CPU PCBA (see
).
4-11
Alarms and Troubleshooting
Error code n08 n09 n10 n11 n12 n13
Table 4-2. Technical Fault Error Troubleshooting Guide (continued)
Description Troubleshooting steps
Proximal pressure sensor calibration fault.
1. Calibrate proximal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Verify proper tubing connections.
4. Replace CPU PCBA (see
).
Negative proximal pressure measurement for 15 seconds.
1. Calibrate proximal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
Negative internal pressure measurement for 15 seconds.
Negative internal valve (exhalation solenoid valve) pressure measurement for
15 seconds.
Loss of memory event pointers.
FiO
2
sensor offset fault
1. Calibrate internal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Calibrate internal valve (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Check the voltage across the exhalation solenoid valve (J19) connector pins on CPU PCBA. If
24 V is not present, replace exhalation solenoid valve, otherwise replace CPU PCBA
(see
).
1. Replace CPU PCBA (see
).
1. Calibrate FiO
2
sensor (see
4.4.2 Troubleshooting Ventilation Alarm Messages
lists possible ventilation alarm messages (in alphabetical order), alarm detection method and ventilator response, priority and whether they respond to the audio paused and/or alarm paused functions and suggests troubleshooting steps for possible hardware or patient circuit causes. See the Puritan Bennett 560 Ventilator
Clinician’s Manual for appropriate responses to alarms when the ventilator is being used
4-12 Service Manual
Service Manual
Alarms and Troubleshooting on a patient. The alarm message text shown appears when you have configured the ventilator for English in the SETUP menu.
If medium, high, or very high priority is listed for an alarm, the priority refers to the alarm
LEDs and tones. Very high priority means the red alarm LED is continuously illuminated with a continuous alarm tone, and ventilation stops. High priority means the red alarm
LED flashes with five alarm tones, and medium priority means an orange LED flashes with three alarm tones. If the ventilator gives a visual priority, then no LED is illuminated nor do alarm tones annunciate. Only an alarm message appears on the screen.
Some alarm messages appear with an asterisk (*) after the first part of the message and before the second part of the message. This indicates that an alarm message contains more information than can be displayed on a single screen. These messages scroll to display the additional information.
Access the alarm log using the information provided in Section 4.3
.
Note:
All alarm messages listed are in US English language as set in Preferences Menu.
4-13
Alarms and Troubleshooting
Alarm message
AC POWER
DISCONNECTION
APNEA
Table 4-3. Ventilation Alarm Troubleshooting Guide
How detected/ventilator response
Loss of AC power
•
If CHECK SUPPLY alarm is
OFF, immediately detected. If CHECK SUPPLY alarm is ON, detected after
5 seconds if ventilation is stopped or
•
Detected at the start of inspiration when ventilation is in progress.
Ventilator response: Switch to external DC power supply if present, otherwise to internal battery. Automatically clears when AC power returns.
No inspiratory trigger detected by the ventilator and
Apnea setting in Preferences menu is set to YES
Apnea occurs in PSV ST; V
SIMV; P SIMV; CPAP
Priority/
Audio paused/
Alarm paused
Medium
Priority/
Yes/Yes
Medium
Priority/
Yes/
Yes
Troubleshooting steps
1. Verify the presence of AC power at the wall outlet.
2. Check the continuity of the AC power cord.
3. If AC LED is illuminated, check the voltage across the 24V fuse
(FU2) on the CPU PCBA. If voltage is correct, consider replacing CPU PCBA (see
). If voltage is incorrect, consider replacing
Power Management CPU PCBA
(see
1. Check ventilator operation using a known good patient circuit and test lung.
2. Replace the inlet air filter (see
).
3. Verify flow and pressure sensor calibration (see
).
Calibrate as necessary.
4. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
4-14 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
BATTERY FAULT1
RESTART/SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Detection of a fault in the internal battery.
Ventilator response: Internal battery is disabled from use. A
-
symbol appears beside the battery symbol instead of remaining internal battery capacity.
Priority/
Audio paused/
Alarm paused
Medium
Priority/
Yes/Yes
Troubleshooting steps
1. Try to operate the ventilator on internal battery only (unplug from AC power). If the ventilator does not run, replace battery. If ventilator runs proceed to step
2. Check the battery voltage using the Internal Battery Menu in the
Measurements Check screen in
Maintenance mode (see
).
•
If displayed battery voltage is greater than upper battery voltage spec limit (29.4 V
+1%) remove battery (see
) and measure voltage. If measured voltage is less than displayed voltage, check the 24 V output (J7) pin 1 (-) and pin 3
(+) of the power management PCBA.
•
If 24 V output is out of range, consider replacing
CPU PCBA (see
) or Power management PCBA (see
). If displayed battery voltage is below 25V, remove the battery and measure voltage. Replace battery if voltage is less than
25 V (see
).
•
If displayed battery voltage is below 25V but measured battery voltage is > 28 V, check the voltage at connector J10, pins 1 (-) and
9 (+) on the Power
Management PCBA. If this measurement matches the measured battery voltage, consider replacing Power
Management PCBA (see
), otherwise replace Battery Connection
PCBA (see
).
4-15
Alarms and Troubleshooting
Alarm message
BATTERY FAULT2
RESTART/SRVC
BATTERY FAULT3
RESTART/SRVC
BUZZER FAULT1
RESTART/SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
No internal battery detected.
Ventilator response: A
-
symbol appears beside the battery symbol instead of remaining internal battery capacity.
Immediately detected if ventilator does not recognize the battery as a
Puritan Bennett supplied battery
Priority/
Audio paused/
Alarm paused
Medium
Priority/
Yes/No
Troubleshooting steps
1. Install a battery.
2. Replace Battery Connection
PCBA (see
).
3. Replace Power Management
PCBA (see
).
Detected when there is a hardware fault on the buzzer
PCBA
Medium
Priority/
Yes/No
Medium
Priority/
Yes/No
1. Inspect battery. If
Puritan Bennett battery is installed, replace with new
Puritan Bennett battery. If non-
Puritan Bennett battery is installed, replace with a
Puritan Bennett battery (see
).
2. Replace Power Management
PCBA (see
).
3. Replace CPU PCBA (see
).
1. Switch the ventilator off and switch it on again. If alarm persists go to Step 2. If alarm clears, the device is safe for use.
2. Switch the ventilator on (no need to ventilate) for a minimum of 15 minutes and up to two hours. Switch the ventilator off and switch it on again. If the alarm does not trigger again, the device is safe for use. If alarm persists go to
Step 3.
3. Switch the ventilator off and switch it on again in Setup mode and go to Maintenance menu. Go to the Measurements
Check menu, place the cursor on Buzzer Voltage check, and press ENTER . Buzzer should buzz and voltage should be
1.7 V – 2.1 V. If the voltage is out of range low or reading 0 while sounding, go to Step 4. If voltage is within range, the device is safe for use.
4-16 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
BUZZER FAULT1
RESTART/SRVC
(cont.)
BUZZER FAULT2
RESTART/SRVC
BUZZER FAULT3
RESTART/SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Priority/
Audio paused/
Alarm paused
Watchdog (Power Fail
Interrupt) voltage > 30 V.
Failure detected in the watchdog circuitry which would prevent the Very High
Priority alarm from audibly annunciating in the event of a loss of power during ventilation.
Ventilator response: No audio signaling for Very High
Priority alarms
Buzzer battery charging failure (voltage < 4.8 V).
Occurs when a hardware fault is detected on battery charger of the buzzer PCBA.
Medium
Priority/
Yes/Yes
High
Priority/
Yes/No
Troubleshooting steps
4. Open device and verify and/or reseat the Buzzer PCBA cable connection to the CPU PCBA.
Retest the Buzzer Voltage in
Maintenance Mode /
Measurement Check (Per step 3, check buzzer voltage). If vent continues to fail go to Step 5.
Replace Buzzer PCBA (see
).
5. Replace CPU PCBA (see
).
Check watchdog voltage in
Maintenance mode (see
). If watchdog voltage
>30 V, replace Power Management
PCBA (see
). If watchdog voltage < 30 V, replace
CPU PCBA (see
1. Ensure the alarm switch is properly set. See
to properly set the switch.
2. Consider replacing Buzzer PCBA
(see
persists, replace ventilator.
4-17
Alarms and Troubleshooting
Alarm message
BUZZER LOW
BATTERY
CALIBRATION
FAIL
CHECK BATTERY
CHARGE
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
Priority/
Audio paused/
Alarm paused
How detected/ventilator response
Buzzer battery voltage <4.8 V with no failure of alarm battery and alarm battery charging circuit. When this alarm occurs, the Power
Supply Loss (VHP) alarm cannot annunciate in the event that the ventilator’s power switch is turned off during ventilation.
Re-sets when voltage > 4.9 V.
Failure of an exhalation flow sensor calibration point.
Ventilator response: the failed calibration point is replaced by the default value and ventilator continues the calibration process at the next flow point.
Internal battery charging circuit fault detected.
Ventilator response: Internal battery does not charge.
Medium
Priority/
Yes/Yes
Medium
Priority/
Yes/Yes
Medium
Priority/
Yes/No
Troubleshooting steps
1. Switch the ventilator off and switch it on again. If alarm persists go to Step 2. If the alarm clears, the device is safe for use.
2. Switch the ventilator on (no need to ventilate) for a minimum of 15 minutes and up to two hours. Switch the ventilator off and switch it on again. If the alarm does not trigger again, the device is safe for use. If alarm persists go to
Step 3.
3. Check the orientation of the switch on the Buzzer PCBA (see
Figure 5-23 ) and be sure that it
is set to the left as shown. If ventilator continues to fail, go to Step 4.
4. Replace Buzzer PCBA (see
). If alarm persists, replace ventilator.
1. If running exhalation flow sensor calibration, try a new external exhalation valve and/ or a new patient circuit.
2. Calibrate inspiratory flow sensor
(see
3. Re-calibrate the exhalation flow sensor.
4. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
5. Replace exhalation block (see
).
6. Replace CPU PCBA (see
).
1. Check internal battery capacity by running the measurement check (see
).
2. Replace Power Management
PCBA (see
).
4-18 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
CHECK
EXH VALVE
CHECK EXH VALVE
PRESSURE
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
Priority/
Audio paused/
Alarm paused
How detected/ventilator response
Exhaled tidal volume < 40% of inspired tidal volume, and inspired tidal volume > 20 mL.
Activation: after 15 s in V/ AC;
P/AC; PSV ST; V SIMV; and P
SIMV
Occurs when the software detects an abnormal signal from the valve pressure sensor on the CPU PCBA
High
Priority/
Yes/No
High
Priority/
Yes/No
Troubleshooting steps
1. Check ventilator operation with a known good patient circuit.
2. Re-calibrate exhalation flow sensor using known good exhalation block (see
3. Re-calibrate internal valve pressure sensor (see
).
4. Re-calibrate inspiratory flow sensor (see
5. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
6. Verify integrity of exhalation block. Replace if necessary (see
).
7. Verify integrity of inspiratory block. Replace if necessary (see
).
8. Replace CPU PCBA (see
), otherwise replace CPU PCBA (see
).
1. Recalibrate internal valve pressure sensor (see
).
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Check the voltage across the exhalation solenoid valve (J19) connector pins on CPU PCBA. If
24 V not present, replace exhalation solenoid valve (see
).
4-19
Alarms and Troubleshooting
Alarm message
CHK EXH VALVE
PRESSURE
CHECK
PROXIMAL LINE 1*
CHECK
REMOTE ALARM
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
Priority/
Audio paused/
Alarm paused
•
How detected/ventilator response
Valve pressure < 1.5 cmH
2
O at ventilation start.
Activation: At the first breath in V A/C; P A/C; PSV ST;V SIMV;
P SIMV
Occurs when proximal line is not connected at start of ventilation. Loss of signal from the proximal pressure sensor after one ventilation cycle or
•
No signal detection from the proximal pressure sensor after 17 seconds in
P A/C and V A/C modes or after Apnea Time + 4 seconds in PSV ST,P SIMV, and V SIMV modes.
Ventilator response: Switch to internal pressure sensor for pressure measurement.
Failure of remote alarm relay circuit.
High
Priority/
Yes/No
Medium
Priority/
Yes/No
Medium
Priority/
Yes/Yes
Troubleshooting steps
1. Re-calibrate internal valve pressure sensor (see
).
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Check the voltage across the exhalation solenoid valve (J19) connector pins on CPU PCBA. If
24 V is not present, replace exhalation solenoid valve (see
), otherwise replace CPU PCBA (see
).
1. Re-calibrate proximal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Perform remote alarm test
(see
2. If test fails, replace CPU PCBA
(see
4-20 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
CHECK SETTINGS
CONNECT VALVE
CONNECT VALVE
OR CHANGE PRESS
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Immediately detected upon loss of memorized parameters or after installation of a new ventilator software version or
Setting out of range or loss of SETUP menu settings
Ventilator response:
Deactivates software lock; out-of-range settings replaced by default values
Occurs at ventilation start in
VC, PSIMV, or VSIMV modes when the exhalation valve is not detected.
Occurs at ventilation start in
PC or PSV ST when pressure settings match operation with exhalation valve configuration but the exhalation valve is not detected.
Ventilator cooling fan operating speed is not appropriate for the internal device temperature
Priority/
Audio paused/
Alarm paused
Medium
Priority/
Yes/Yes
High
Priority/
Yes/No
High
Priority/
Yes/No
Troubleshooting steps
1. Re-enter ventilation parameters.
2. Turn ventilator off and back on again to verify memory. If settings are not retained, replace CPU PCBA (see
).
Connect an exhalation valve
COOLING FAN
RESTART / SRVC
Medium
Priority/
Yes/Yes
1. Install or replace exhalation valve.
2. Calibrate internal valve pressure transducer (see
3. Replace CPU PCBA (see
).
1. Verify fan spins freely. Replace if necessary (see
2. Verify 24 V output at fan connector (J6). If voltage is incorrect, replace Power
Management PCBA (see
).
3. Replace CPU PCBA (see
).
Remove exhalation valve.
CPAP MODE
Occurs when CPAP ventilation is started and an exhalation valve is used.
High
Priority/
Yes/No
4-21
Alarms and Troubleshooting
Alarm message
DC POWER
DISCONNECTION
DEVICE FAULT1
RESTART/SRVC
DEVICE FAULT2
RESTART/SRVC
DEVICE FAULT3
RESTART / SRVC
DEVICE FAULT4
RESTART/SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
Priority/
Audio paused/
Alarm paused
How detected/ventilator response
Loss of external DC power.
Activation: After 5 seconds if ventilation is stopped or at the start of inspiration when ventilation is in progress.
Ventilator response: Switch to internal battery.
Internal pressure sensor signal constant (± 0.5 cmH
2
O) for
15 seconds.
Proximal pressure sensor signal constant (± 0.5 cmH
2
O) for 15 seconds.
24 V failure
Ventilator cooling fan operating speed is not appropriate for the internal device temperature.
Medium
Priority/
Yes/Yes
High
Priority/
Yes/No
Medium
Priority/
Yes/No
High
Priority/
Yes/No
Medium
Priority/
Yes/Yes
Troubleshooting steps
1. Verify the voltage at the external power source and recharge as necessary.
2. If DC source has power available, but DC LED is not illuminated, replace Power
Management PCBA (see
).
3. If DC LED is illuminated, remove the fuse and check continuity
(FU2 on the CPU PCBA). If continuity is good, replace CPU
PCBA (see
).
4. If 24 V not present, replace
Power Management CPU PCBA
(see
1. Re-calibrate internal pressure sensor (see
).
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Re-calibrate proximal pressure sensor (see
).
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Check 24V output at connector
J9 pin 1 (-) and pin 2 (+) on
Power Management PCBA. If
24 V not present, replace Power
Management PCBA (see
).
1. Verify fan spins freely. Replace if necessary (see
2. Verify 24 V output at fan connector (J6). If voltage is incorrect, replace Power
Management PCBA (see
).
3. Replace CPU PCBA (see
).
4-22 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
DEVICE FAULT5
RESTART / SRVC
DEVICE FAULT6
RESTART / SRVC
DEVICE FAULT7
RESTART / SRVC
DEVICE FAULT8
RESTART/SRVC
DEVICE FAULT9
RESTART/SRVC
DEVICE FAULT10
RESTART/SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Detection of a fault in the electrical power supply system.
Ventilator response: A
-
symbol appears beside the battery symbol instead of remaining internal battery capacity
Turbine speed too low coupled with high internal temperature. Ventilator response: ventilation stops within 5 seconds; oxygen solenoid valve closes
Priority/
Audio paused/
Alarm paused
Medium
Priority/
Yes/Yes
Troubleshooting steps
Replace Power Management PCBA
(see
CPU 5 V reference voltage <
4.5 V or 10 V reference voltage
< 9 V
Inspiratory flow is constant
(+1 lpm) with normal turbine temperature and speed conditions
RAM error during POST
Flash memory checksum error during POST
High
Priority/
No/No
High
Priority/
Yes/No
High
Priority/
Yes/No
Very High
Priority/
No/No
Very High
Priority/
No/No
1. Verify electrical connection at J1 on Turbine PCBA and that 24 V is present. If not present consider replacing:
•
Power Management PCBA
(see
•
Turbine control PCBA (see
)
2. Check the fuse, F1, on the turbine control PCBA. If fuse is blown, replace turbine control
PCBA (see
).
3. Check connection and cable between turbine control PCBA and CPU PCBA. Replace cable if necessary.
4. Replace turbine (see
).
5. Replace CPU PCBA (see
).
1. Consider replacing:
•
Power Management PCBA
(see
•
CPU PCBA (see
).
1. Re-calibrate inspiratory flow sensor (see
).
2. Consider replacing CPU PCBA
(see
1. Consider replacing CPU PCBA
(see
1. Consider replacing CPU PCBA
(see
4-23
Alarms and Troubleshooting
Alarm message
DEVICE FAULT11
RESTART/SRVC
DEVICE FAULT12
RESTART/SRVC
DEVICE FAULT13
RESTART/SRVC
EMPTY BATTERY
E SENS FAULT
OR CIRC LEAK
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
EEPROM error during POST
Reference voltage error during POST
Priority/
Audio paused/
Alarm paused
Very High
Priority/
No/No
Very High
Priority/
No/No
Very High
Priority/
No/No
Troubleshooting steps
1. Consider replacing CPU PCBA
(see
1. Consider replacing CPU PCBA
(see
1. Contact your local Covidien service representative.
Software does not carry the
LX xxxxxx designation.
Activation: At ventilator startup during POST.
Ventilator response:
Ventilation prohibited.
Internal battery capacity
< 10 min. or 3% (battery voltage < 22.5 V)
Ventilator response:
Ventilation stops immediately
At least 4 of the last 6 breaths within the last minute are terminated by time in P SIMV or V SIMV modes.
High
Priority/
No/No
Medium
Priority/
Yes/No
1. Charge the battery by connecting the device to AC power.
2. Replace the battery (see
).
3. Replace battery connection
PCBA (see
).
4. Replace Power Management
PCBA (see
).
1. Check for leaks at all tubing connections inside the ventilator.
2. Replace CPU PCBA (see
).
4-24 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
EXH VALVE
LEAKAGE
ERASE ERROR
TECHNICAL
PROBLEM
HIGH INT TEMP
COOL VENT*
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
Priority/
Audio paused/
Alarm paused
How detected/ventilator response
Exhaled flow detected during the inspiratory phase for three consecutive breaths (using double-branch patient circuit) in V A/C; P A/C; PSV ST;
V SIMV; P SIMV
Medium
Priority/
Yes/No
Troubleshooting steps
1. Check ventilator operation with a known good patient circuit.
2. Re-calibrate exhalation flow sensor using known good exhalation block (see
3. Re-calibrate internal valve pressure sensor (see
4. Re-calibrate inspiratory flow sensor (see
).
5. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
6. Verify integrity of exhalation block. Replace if necessary (see
).
7. Verify integrity of inspiratory block. Replace if necessary (see
).
8. Replace CPU PCBA (see
).
Try another USB memory device.
Occurs if the memory erase operation on USB memory device fails.
Device internal temperature out of tolerance range. (> 60 o
C)
Visual priority/
No/No
Medium
Priority/
Yes/Yes
1. Check device internal temperature on the
Measurements Check menu in
Maintenance mode (see
). If displayed temperature > 140 °F (60° C), move device to cooler location.
2. Replace Power Management
PCBA (see
).
3. Replace CPU PCBA (see
).
4-25
Alarms and Troubleshooting
Alarm message
HIGH/LOW
BATTERY TEMP*
INCOMPATIBLE
SOFTWARE
VERSION
INSP FLOW
RESTART / SRVC
INTENTIONAL
VENT STOP
KEYBOARD FAULT
RESTART / SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Battery temperature out of tolerance. Ventilator response:
Battery charging stops.
Priority/
Audio paused/
Alarm paused
Medium
Priority/
Yes/Yes
Troubleshooting steps
1. Check battery temperature displayed on Internal Battery menu in Measurements Check menu of Maintenance mode
(see
temperature > 104
F (40
C) or
< 41
F (5
C), move device to cooler or warmer location as appropriate.
2. Verify internal fan is operating.
3. Consider replacing:
•
Power Management PCBA
(see
•
CPU PCBA (see
).
•
Battery (see
).
1. Contact your local Covidien representative.
Occurs at startup when a hardware problem is detected on CPU PCBA memories or critical CPU functions. Also occurs on start up after software download with corrupted software or with software that is not suitable for PB560 ventilator.
Inspiratory flow is constant
(+1 lpm) with normal turbine temperature and speed conditions
High
Priority/
No/No
High
Priority/
Yes/No
1. Re-calibrate inspiratory flow sensor (see
).
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Consider replacing CPU PCBA
(see
Ensure that ventilator was shut down intentionally.
Detected when ventilator is shut down with the power switch while ventilating.
Alarm lasts until the next power-on.
Key pressed for more than
45 seconds (stuck key).
High
Priority/
Yes/Yes
High
Priority/
No/No
1. Replace keypad (see
).
2. Replace CPU PCBA (see
).
4-26 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
LOW BATTERY
NO BATTERY
DETECTED
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Internal battery capacity < 30 min. or 8%
(battery voltage < 25 V)
Priority/
Audio paused/
Alarm paused
Immediately detected when no battery is installed
High
Priority/
Yes/No
Medium
Priority/
Yes/No
Troubleshooting steps
1. Charge the battery by connecting the device to AC power.
2. Replace battery (see
).
3. Replace battery connection
PCBA (see
).
4. Replace Power Management
PCBA (see
).
Install Covidien approved battery.
1. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
2. Re-calibrate proximal pressure sensor (see
).
3. Replace CPU PCBA (see
).
NO
PROXIMAL LINE2*
Proximal pressure < 0.6 cmH
2
O for 100 ms during inspiration phase of 2nd breath cycle
Ventilator response: Switch to internal pressure sensor for pressure measurement.
•
Occurs when proximal line is not connected during ventilation. Loss of signal from the proximal pressure sensor after one ventilation cycle or
• No signal detection from the proximal pressure sensor after
17 seconds in P A/C and V A/C modes or after Apnea Time +
4 seconds in PSV ST,P SIMV, and V SIMV modes.
Ventilator response: Switch to internal pressure sensor for pressure measurement.
Medium
Priority/
Yes/No
4-27
Alarms and Troubleshooting
Alarm message
OCCLUSION
CHECK CIRCUIT*
PATIENT
DISCONNECTION*
POWER FAULT
RESTART/SRVC
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
Priority/
Audio paused/
Alarm paused
How detected/ventilator response
Less than 20 mL of inspired tidal volume and P Support or
Pi -20% < Pi < P Support or
Pi+20% measured for three consecutive breaths in PSV ST,
P A/C, and P SIMV modes.
Note:
A high respiratory rate or backup rate may not sufficiently flush out CO
2
in some vented pediatric masks.
Activates if one of the following conditions remains for 15 s or Apnea Time + 2 seconds, whichever is greater:
Inspiratory flow > 130 lpm during the inspiratory phase or
•
In V A/C and V SIMV modes, patient pressure ≤ Low PIP or
•
In PSV ST, CPAP, P A/C, and P SIMV modes patient pressure <
(inspiratory pressure set point) - 20%.
A detection of a fault in the electrical power supply system. A- symbol displayed instead of battery capacity.
Occurs when communication between CPU PCBA and
Power Management PCBA is lost.
High
Priority/
Yes/No
High
Priority/
Yes/No
Medium
Priority/
Yes/Yes
Troubleshooting steps
1. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator or in patient circuit.
2. Ensure the patient circuit matches the set ventilator preference (Use pediatric circuit when pediatric circuit preference set to YES; use adult circuit when Pediatric circuit preference set to OFF).
3. Switch to a vented system with a larger leak configuration.
4. Re-calibrate inspiratory flow sensor (see
5. Replace CPU PCBA (see
).
1. Re-calibrate the patient and internal pressure sensors (see
2. Verify all internal tubes are connected properly to their respective sensors.
3. Check for n06 technical fault. If present, replace turbine (see
).
4. Replace CPU PCBA (see
).
1. Replace Power Management
CPU PCBA (see
2. 2. Replace CPU PCBA (see
).
4-28 Service Manual
Service Manual
Alarms and Troubleshooting
Alarm message
POWER LOSS
(no message)
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Electrical power interrupted when ventilation is in progress or
•
Ventilator operating on internal battery and battery fully discharged or
• stopping the ventilator without using the power switch
•
Ventilator response:
Immediate loss of ventilation. Ventilation resumes after AC or DC power is restored or when power switch is turned on and vent is connected to
AC power
Priority/
Audio paused/
Alarm paused
VHP/No—
Alarm
Cancel
Only/No—
Alarm
Cancel
Only
Troubleshooting steps
1. Press the Alarm Control key to pause the audible portion of the alarm.
2. Turn the ventilator power switch back on to immediately resume ventilation.
3. Press and hold the Ventilation
ON/OFF key for 3 seconds to stop ventilation. Confirm stop by pressing the Ventilation ON/
OFF key again.
PRES SENS FLT1
RESTART/SRVC
Internal pressure sensor signal constant
(± 0.5 cmH
2
O) for 15 seconds.
PROX SENS FLT2
RESTART/SRVC
REMOTE ALARM
FAULT
REMOVE VALVE
Proximal pressure sensor signal constant
(± 0.5 cmH
2
O) for 15 seconds
Failure of remote alarm relay circuit.
Occurs in CPAP if ventilator detects an exhalation valve is being used.
High
Priority/
Yes/No
Medium
Priority/
Yes/Yes
1. Re-calibrate internal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Re-calibrate proximal pressure sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
1. Replace CPU PCBA (see
).
Medium
Priority/
Yes/Yes
High
Priority/
Yes/No
1. Remove exhalation valve.
4-29
Alarms and Troubleshooting
Alarm message
REMOVE VALVE OR
CHANGE PRESS
TRANSFER ERROR
USB
DISCONNECTION
TRANSFER ERROR
USB FULL
TRANSFER ERROR
TECHNICAL
PROBLEM
Table 4-3. Ventilation Alarm Troubleshooting Guide (continued)
How detected/ventilator response
Occurs in P A/C;PSV ST when pressure settings match operation with calibrated leak configuration but an exhalation valve is detected
Occurs when USB memory device is disconnected during data transfer.
Occurs when USB memory device is full during data transfer
Occurs if a data transmission failure occurs during transfer
Priority/
Audio paused/
Alarm paused
High
Priority/
Yes/No
Visual priority/
No/No
Visual priority/
No/No
Visual priority/
No/No
Troubleshooting steps
1. Remove exhalation valve.
1. Reinsert USB device and retry.
1. Delete USB device contents and retry.
2. Try a new USB device.
1. Try a new USB device.
1. Ensure the USB device has been configured for FAT 32 format.
TRANSFER NOT
ALLOWED
INCOMPATIBLE
SETTINGS
TRANSFER NOT
ALLOWED FILE
NOT FOUND OR
WRONG
TRANSFER NOT
ALLOWED
REMOVE A KEY
VALVE MISSING
Occurs on data transfer request from the USB memory device to the ventilator when the setting file to be imported is not suitable for PB560
Occurs on data transfer request from the USB memory device to the ventilator when there is no setting file on the
USB memory device or if file's data are corrupted.
Occurs when more than one
USB memory device is connected to the ventilator.
Occurs at ventilation start in
V/AC; PSIMV; or VSIMV modes when the exhalation valve is not detected
VTI NOT
REACHED*
Inspired tidal volume does not match Vt setting (± 10%) for six consecutive breaths in
V A/C and V SIMV modes.
Visual priority/
No/No
Visual priority/
No/No
Visual priority/
No/No
High
Priority/
Yes/No
High
Priority/
Yes/No
1. Ensure the USB device has been configured for FAT 32 format.
2. Try a new USB device.
1. If two USB keys are connected to the ventilator, remove one key.
1. Verify exhalation valve is connected to exhalation valve pilot port.
2. Check for disconnected tubing inside ventilator.
1. Re-calibrate inspiratory flow sensor (see
2. Check that there are no leaks or obstructions in the pneumatic tubes inside the ventilator.
3. Replace CPU PCBA (see
).
4-30 Service Manual
Service Manual
Alarms and Troubleshooting
4.4.3 Troubleshooting Miscellaneous Symptoms
provides information to troubleshoot faults that may or may not cause alarms or technical faults.
Table 4-4. Miscellaneous Symptom Troubleshooting Guide
Symptom
AC, DC, and internal battery LEDs lit in the presence of a connected AC or DC power source.
LED indicating power source is not lit, but device functions.
Troubleshooting steps
1. Consider replacing CPU PCBA (see
Excessive heat from the ventilator
Whistling air sounds
LCD panel does not update or respond to key presses
Dark display or no display
2. Check that the keypad cable is correctly connected to connector (J1) on the CPU PCBA. Check for broken LEDs, damaged keypad cable. If LEDs are broken or cable damaged, replace keypad (see
).
3. Replace CPU PCBA (see Section 5.7.10
1. Check the air inlet for obstructions and clear if necessary.
2. Replace the inlet air filter (see
3. Perform the Flow Sensor Capacity test (See
).
4. Replace turbine (see
).
1. Check that the inlet air filter is present.
2. Check for leaks in the oxygen connector and delivery tube. Replace tie wraps or connector or tube if necessary.
3. Replace turbine (see
).
1. Verify the software lock is unlocked (see
in this chapter for instructions on how to unlock).
2. Ensure the LCD panel cable is connected properly to the CPU PCBA.
3. Replace keypad (see
4. Verify the voltage across pins 21 (-) and 22 (+) of connector J17 on the CPU PCBA is 5 V.
5. Replace LCD panel (see Section 5.7.11
).
6. Replace CPU PCBA (see Section 5.7.10
1. Adjust display contrast and brightness.
2. If device is below specified operating temperature
(See
Table 2-4 for operating temperature
specification), place in warmer environment.
3. Ensure the keypad cable is connected properly to the CPU PCBA.
4. Verify the voltage across pins 21 (-) and 22 (+) of connector J17 on the CPU PCBA is 5 V.
5. Replace LCD panel (see Section 5.7.11
).
6. Replace CPU PCBA (see Section 5.7.10
4-31
Alarms and Troubleshooting
Table 4-4. Miscellaneous Symptom Troubleshooting Guide (continued)
Symptom
No access to waveforms
Troubleshooting steps
Set Display Waveforms to YES in Preferences menu
(only available when ventilating.)
Set Backlight to OFF in Preferences menu.
Display backlight does not switch off during ventilation
Alarm sound level too low or too high
Poor display visibility
Corrupted characters on display
Condensation appears on screen
Adjust alarm volume level in Preferences menu.
Adjust contrast in Preferences menu.
Replace LCD panel (see Section 5.7.11
).
1. Open case, let air dry, and check for failures.
2. Replace keypad (see
3. Replace LCD panel (see Section 5.7.11
).
Replace CPU PCBA (see Section 5.7.10
Displayed barometric pressure does not fall within tolerance when compared with a calibrated standard barometer (fails barometric pressure test in Performance
Verification).
Frequent inspiratory or expiratory tidal volume alarms
Ensure circuit type used (adult or pediatric) matches circuit type set in Preferences Menu
4.4.4 PCBA Test Points
Some troubleshooting steps given in
voltages at various test points. The following sections provide illustrations for each PCBA showing the referenced connectors and location of pin 1 on each connector, and list the nominal voltages expected at these test points.
4-32 Service Manual
Service Manual
4.4.4.1 CPU PCBA
Top view
Top View
Bottom View
Bottom view
Alarms and Troubleshooting
4-33
Alarms and Troubleshooting
3
6
7
4
5
1
2
Reference Designator
FU2
J19
J17
J1
J11
J5
Comments
Circled item number is the referenced pin of connector
Circled item number is the referenced pin of connector
J17 located beneath CPU LCD interface PCBA under LCD panel
8
9
10
11
12
13
Reference Designator
J21
J2
J18
J3
J4
J7
Connector
J7
J19
FU2
Pin (-)
1
2
N/A
Pin (+)
13
1
N/A
Nominal
Voltage (v)
5
24
24
4-34 Service Manual
4.4.4.2 Power Management PCBA
Alarms and Troubleshooting
Service Manual 4-35
Alarms and Troubleshooting
1
2
3
4
5
Reference
Designator
J2
J7
F1
6
Comments
Power management PCBA
Power Supply PCBA
Circled item number is the referenced pin of connector
Connector
J6
J7
J9
J10
J7
Pin(-)
2
1
1
1
1
7
8
9
10
11
Reference
Designator
PROG EEP
TURBINE
Comments
Power supply PCBA connector
VENTIL Fan connector
Battery connection PCBA connector
Pin (+)
1
3
2
9
13
Nominal Voltage (V)
24
24
24
27
5
4-36 Service Manual
4.4.4.3 Turbine Control PCBA
Alarms and Troubleshooting
Service Manual
1
2
3
Reference
Designator
J3
J6
F1
4
5
Reference
Designator
J1
Comments
Circled item number is the referenced pin of connector (for
CPU PCBA harness)
Connector
J1
F1
Pin (-)
1
N/A
Pin (+)
2
N/A
Nominal
Voltage (V)
24
24
4-37
Alarms and Troubleshooting
This page intentionally blank
4-38 Service Manual
5 Service and Repair
This chapter describes how to repair the major components of the Puritan Bennett™ 560
Ventilator and includes removal, replacement, and adjustment procedures. For a complete parts breakdown, including assembly drawings and part numbers, refer to
5.1
Warnings
WARNING
When servicing the ventilator, be sure to familiarize yourself with, and adhere to all posted and stated safety warning and caution labels on the ventilator and its components, and on any service equipment and materials used. Failure to adhere to such warnings and cautions at all times may result in injury or property damage.
To prevent patient injury, do not use a ventilator if it requires repair.
To prevent personal injury or death, do not attempt any ventilator service while a patient, or other person, is connected to the ventilator.
Use personal protective equipment whenever exposure to toxic fumes, vapor, dust particles, blood pathogens, and other transmittable diseases and hazardous material can be expected. If in doubt, consult an environmental health and safety specialist or an industrial hygienist before servicing the ventilator. Refer to
To prevent electrical shock hazard and possible personal injury, always disconnect electrical power sources before servicing the ventilator. Follow accepted safety practices for electrical equipment when testing or making equipment, adjustment, or repairs.
To prevent possible personal injury, always disconnect the oxygen source from the ventilator before service.
Chemicals from a broken LCD panel are toxic when ingested. Use caution when handling a ventilator with a broken display panel.
To prevent possible damage to the power supply, remove the battery from the ventilator
(
) during disassembly and reassembly.
To prevent damage to electrostatic discharge (ESD) sensitive components, always follow
ESD guidelines when servicing the ventilator. Adhere to ESD control techniques when repairing ESD sensitive components.
Damage to components may occur due to over-tightening of screws. Care should be taken during reassembly not to over-tighten screws where instructed.
5-1
Service and Repair
Use only recommended tools, test equipment, and service materials (see
servicing the ventilator.
As you repair the ventilator, perform any applicable cleaning and inspection procedures listed below.
Visually inspect any removed ventilator parts, including those removed to gain access to a suspected faulty part. Inspect the exposed area behind the removed parts as well. Clean removed parts to facilitate further inspection as necessary.
Investigate and determine the cause of any detected abnormality. Repair the unit or contact
Covidien Technical Support for help in diagnosing unresolved symptoms.
Replace or repair all parts that are worn, missing, damaged, cracked, corroded, burnt, warped, bent, disfigured, or broken. Consult
for parts availability.
Covidien recommends that customers or technical services personnel follow local governing ordinances and recycling regulations regarding disposal or recycling of the ventilator, battery, or other device components.
The repair instructions assume the patient system, humidifier, and oxygen source are already removed from the ventilator.
In the repair procedures, PCBA reference designators for connectors and sensors are given in parentheses. For example, the reference designator for the Remote Alarm connector on the CPU PCBA is (J2); the reference designator for the exhalation valve pilot pressure sensor fitting is (1 VALVE).
Do not start ventilation until you ensure the device is correctly assembled, that the air inlet filter is properly installed and unobstructed, and that there is proper clearance all around the unit. Also ensure the patient circuit is correctly connected to both the ventilator and the patient, and that the patient circuit, including all hoses, is not damaged or obstructed.
5.2
Electrical Cables and Pneumatic Tubing
• To ensure proper reassembly, note or label wire and tube positions before disconnecting parts. Refer to the pneumatic cabling diagram label on the inside of
the ventilator top housing or to correctly connect pneumatic tubes.
• To avoid shredding a silicone tube when removing from a fitting, gently pull the tube while turning.
• Make sure all tubes, and harnesses or cables, are installed using tie wraps where required.
• Make sure wiring does not interfere with, and cannot be damaged by, hinged or moving ventilator parts.
5.3
Electrostatic Discharge Control
It is important to follow ESD control procedures whenever the ventilator is repaired.
Electrostatic discharge can permanently damage ESD sensitive microelectronic components or assemblies when they are handled, and even when no direct contact is
5-2 Service Manual
Service Manual made with the component or assembly. ESD damage may not be immediately detectable; however, ESD damage will show up at a later time. It can manifest as a premature catastrophic failure of a component or assembly, or as an intermittent failure, all of which can be difficult and costly to locate.
5.3.1 ESD Procedures and Precautions
Follow these procedures and precautions to prevent ESD damage to the ESD-sensitive microelectronic components and assemblies in the Puritan Bennett 560 Ventilator.
• Use a personnel-grounding system. Before opening the ventilator’s housings, ensure a personnel grounding system such as that listed in
Table 2-14 (wrist strap, static-
dissipative mat, and ground cord) is worn correctly and is properly connected to a reliable ground.
• Follow correct procedures when using an antistatic mat. Place tools, test equipment, and the ESD sensitive device on the mat before starting repairs. Conduct all work on the mat.
• Handle ESD sensitive components properly. Do not handle ESD sensitive component connection points, connector pins, leads, or terminals.
• Keep nonconducting materials away from the work area. Static charges from nonconducting material, (i.e. plastic containers, foam cups, synthetic clothing, cellophane tape, etc.) cannot be removed by grounding. These items must be kept away from the work area when handling ESD sensitive devices.
• Follow correct procedures for use of static-shielding bags. Store and transport all ESD sensitive devices in static-shielding bags at all times, except when being worked on.
Never place more than one ESD sensitive device in a static-shielding bag. Never place static generating nonconducting material inside a static-shielding bag with an ESD sensitive device. Place any faulty ESD-sensitive devices in a static-shielding bag immediately after removal, to prevent additional damage. Close the bag to ensure the shield is effective.
5.4
Replacement Part Ordering
Ordering correct parts requires that you properly identify the ventilator version and part.
To replace a part that is not stocked or that is unavailable, order the next higher assembly. Retain the part to be replaced until the replacement part is obtained, and compare the two for compatibility, if possible.
provides lists of orderable parts and assemblies.
5.5
Patient System and Accessories
To service the patient system and accessories, refer to directions for use supplied with the patient system.
Service and Repair
5-3
Service and Repair
5.6
Service Prerequisites
Before repairing the ventilator, perform the tasks in this section.
5.6.1 Cleaning
The following cleaning guidelines pertain to parts that require cleaning while servicing the ventilator. Patient breathing circuits and parts listed in
, are single patient use only, and should be disposed of when dirty. Do not clean or reuse these parts.
Replace all parts that cannot be cleaned. Refer to
for recommended cleaning solutions.
Note:
•
When cleaning reusable components, do not use hard brushes or other implements that could damage surfaces.
•
After you clean the components, inspect them for damage, such as cracks and crazing. Replace any damaged components.
Note:
Discard all parts removed from the ventilator during the maintenance procedures in accordance with your institution's protocol. Sterilize or disinfect parts before nondestructive disposal. Follow local governing ordinances and recycling plans regarding disposal or recycling of device components.
To clean the ventilator
1. Clean ventilator exterior surfaces before disassembly. Table 2-14 of this manual lists
acceptable cleaning and disinfecting agents.
• Use a clean, lint-free cloth, and squeeze excess liquid from the cloth. Do not allow excess cleaning liquids to enter the ventilator.
• Allow cleaned ventilator parts and surfaces to air-dry.
2. If required upon ventilator disassembly, vacuum ventilator interior using ESD safe equipment. Do not clean the ventilator interior or exterior surface with high-pressure air.
5.6.2 Labeling Checks
Verify the presence of the following labels and replace if necessary:
• inlet air filter label
• nurse call alarm label
• O
2
inlet label
• identification label on the bottom of the ventilator
• inspiratory block labels (exhalation pilot pressure and proximal pressure label
• exhalation block, single patient use label
• exhalation block, valve connection label.
5-4 Service Manual
Service Manual
5.6.3 External Inspection
1. Inspect the external surfaces of the ventilator for damage. Replace if necessary.
screen
2. Check that the handle fasteners are tight. Tighten if required (if missing, refer to
3. Verify the presence of the following items:
• inlet air filter (if dirty, refer to
for replacement)
•
power switch cover (if missing, refer to Section 5.7.2
• four (4) feet on the bottom of the ventilator (if missing, refer to
for replacement)
• all fasteners on the housing and battery compartment (replace as necessary).
5.6.4 Measurements Check
This series of steps verifies various internal voltages, including the internal battery, checks the barometric pressure sensor, and tests the alarm buzzers.
Note:
Before running the Measurement check, warm up the ventilator for at least ten minutes in normal operation (ventilating) with alarms turned off and default parameters set. This is particularly important so that the blower will be at the correct temperature during the Measurement check.
To perform the measurements check:
1. Disconnect any external DC power source, if present, and connect the ventilator to
AC power.
2. Enter Maintenance mode (refer to Section 2.12
for instructions on how to enter
Maintenance mode).
3. With the Maintenance screen displayed, use the UP or DOWN arrow keys to place the cursor at the Measurement Check position and press ENTER (
Measurements Check screen appears (
).
Service and Repair
5-5
Service and Repair
Figure 5-1. Selecting Measurement Check
Measurement check
Figure 5-2. Measurement Check Screen
4. Ensure the values are within the test limits in
.
5-6 Service Manual
Service Manual
Table 5-1. Specifications for Measurement Check Verification
Parameter
24 V check 24 V ± 0.5 V
Specification
Watchdog
Barometric pressure
Internal temperature
Blower temperature
Buzzer
Back-up buzzer
Altitude
Compensation
24 V ± 0.5 V
Ventilator barometer reading within
± 11 mmHg of reference barometer
Internal temp 30 °C - 55 °C
(86°F - 131 °F)
Blower temp 35 °C - 65 °C
(95 °F - 149 °F)
Long beep sounds
Buzzer voltage: 1.7V - 2.1V
Long beep sounds
Covidien recommends altitude compensation is always used (YES)
Troubleshooting steps
1. Verify 24 V on connector J7
2. Ensure Power Management cable is connected properly and is not damaged.
3. Replace Power Management
PCBA (see
.
4. Replace CPU PCBA (see
1. Replace CPU PCBA (see
1. Ensure no tubing is connected to the barometric pressure transducer.
2. Replace CPU PCBA (see
1. Move ventilator to a cooler location.
2. Ensure side and front cooling vents are not obstructed.
3. Ensure fan is operating properly.
4. Replace Power Management
PCBA (see
.)
1. Warm up the ventilator for at least ten minutes in normal operation (ventilating) with alarms turned off and default parameters set. This is particularly important so that the blower will be at the correct temperature during the Measurements check.
2. Ensure turbine control cable is connected properly and not damaged.
3. Replace turbine.
1. Refer to Buzzer Fault alarms in
.
1. Replace CPU PCBA (see
N/A
Service and Repair
5-7
Service and Repair
Table 5-1. Specifications for Measurement Check Verification (continued)
Parameter Specification
Internal battery menu
Supplier Suppliers displayed
Troubleshooting steps
First Use Date
Theoretical capacity
Capacity
Record value
4800 mAh
3840 mAh minimum
If supplier cannot be determined,
replace the battery (see Section
.)
N/A
If not 4800 mAh, Replace battery.
Cycles done
Battery voltage
Battery temperature
Record value
23.5 V to 29.7 V following battery charge
0°C - 40°C (32°F - 104°F)
If battery capacity is not at least
3840 mAh, replace battery (see
.).
N/A
1. Charge battery.
2. Replace battery.
3. Verify contacts and proper connection of Battery
Connection PCBA and Power
Management PCBA.
4. Replace Battery Connection
PCBA (see
).
5. Replace Power Management
PCBA (see
)
1. Check for proper operation of fan.
2. Verify cooling vents in battery cover are not obstructed.
3. Replace battery (see
.).
5. With the cursor at the Buzzer position, press ENTER .
a. Verify alarm emits a long beep.
b. Verify buzzer voltage is 1.7 V to 2.1 V.
6. Place the cursor at the Back-up buzzer position and press ENTER .
7. Verify alarm emits a long beep.
8. Place the cursor at the Internal Battery Menu position and press ENTER .
9. Verify the displayed information as shown in Table 5-1 . If the measurements check
parameters are not as displayed in the table, refer to the Troubleshooting steps contained in that table.
10. When you have finished, place the cursor at the Back position and press ENTER to return to the Maintenance screen.
5-8 Service Manual
Service Manual
Service and Repair
5.6.5 Review Technical Fault Log
To review the technical fault log:
1. Enter Maintenance mode (refer to Section 2.12
for instructions on how to enter
Maintenance mode).
2. Place the cursor at the Faults Check position and press ENTER . The Faults screen appears. If there are no faults in the fault log, the ventilator displays NO FAULTS.
3. To return to the Maintenance screen, place the cursor at the Back position and press ENTER .
5.7
Ventilator Repair
This section describes how to remove major ventilator components.
5.7.1 Inlet Air Filter Removal and Replacement
Caution
•
Check the cleanliness of the inlet air filter located on the rear of the ventilator and replace it if necessary before the recommended replacement period expires. Pay particular attention when the ventilator is installed on a wheelchair, as the environmental conditions may cause the filter to become dirty more rapidly.
•
Failure to replace a dirty inlet air filter or operating the ventilator without a filter may damage the ventilator.
•
The inlet air filter is not reusable. Do not attempt to wash, clean, or reuse it.
To remove the inlet air filter:
Grasp the center of the filter and pull outward (
5-9
Service and Repair
Figure 5-3. Removing the Inlet Air Filter
To replace the inlet air filter:
With the foam surface of the filter towards the ventilator, push the filter into position, tucking the corners into the housing opening. Ensure the filter is not distorted when installed.
5.7.2 Power Switch Cover Replacement
WARNING
To avoid eye injury when working with the cover spring, wear safety glasses when performing this procedure.
To remove the power switch cover:
1. Grasp the power switch cover.
2. Gently pull the cover, spring, and pin assembly away from the ventilator.
To assemble the power switch cover retainer:
1. Remove the pin, spring, and switch cover from the existing switch cover assembly.
5-10 Service Manual
Service Manual
Service and Repair
2. Assemble the spring and switch cover, ensuring proper orientation.
a. Identify the long end of the spring and the short end of the spring. Older versions of the spring do not have a formed end, newer versions of the spring do.
Figure 5-4. Identification of Short and Long Ends of Spring
2
1
2
Long leg of spring to inside of switch cover
Short leg of spring to outside of switch cover b. With the front of the switch cover out and ensuring the short end of the spring is outside the cover, slide the pin through right hole at the top of the switch cover and through the right side of the spring and the center hinge of the switch cover.
c. Twist the long end of the spring, placing the spring under tension, tucking it behind the switch cover.
d. Push the pin out the left hole at the top of the switch cover. e. Ensure the longer leg of the spring is to the inside of the cover and the shorter leg of the spring is to the outside of the cover.
Figure 5-5. Switch cover, spring, and pin detail
5-11
Service and Repair
To replace the power switch cover:
2. Grasp the power switch cover, sliding an index finger under the supporting leaf.
3. Bring supporting leaf to a 90-degree position from the switch cover.
Figure 5-6. Positioning and Inserting the Switch Cover
4. Insert into the rear of the lower housing, ensuring the pin and spring remain centered within the slots on the lower housing.
5. Secure the power switch cover with the provided tie-wrap.
a. Insert tie wrap through the slots of the supporting leaf.
b. Firmly bend the tie-wrap on both sides to prevent slippage.
c. Secure the tie-wrap, allowing some room between the housing and the tie-wrap.
d. Verify the switch cover is correctly attached.
Figure 5-7. Securing Switch Cover with Tie-wrap
6. Lift and lower the cover to check that it operates correctly.
5-12 Service Manual
Service Manual
5.7.3 Adhesive Feet Removal and Replacement
To remove the adhesive feet:
1. Turn the ventilator bottom side up.
2. Use pliers to pull each foot away from the housing.
To replace the adhesive feet:
1. Clean the recess of the housing with isopropyl alcohol to remove old adhesive.
2. Peel the adhesive backing from the foot and press it into its recess (4 pl).
5.7.4 Battery Removal and Replacement
WARNING
• Ensure the ventilator is off and AC power is disconnected when removing or installing the battery.
• Do not open the battery for any reason.
• Use care to apply proper pressure on the battery cover when installing screws.
• To prevent possible damage to the power supply, remove the battery from the ventilator during disassembly and reassembly.
• To prevent possible damage following removal, cover battery terminals in nonconductive packaging to avoid shorting and to store the battery so that it cannot be damaged.
To remove the battery:
1. Remove the three screws from the battery cover ( Figure 5-8 ) using a T10 Torx® driver.
2. Pull the cover out of the lower housing.
3. Remove the battery from the cover.
Service and Repair
5-13
Service and Repair
Figure 5-8. Removing/replacing the Battery
5-14
1 Battery contacts toward this corner
To replace the battery:
1. Install the battery so that the terminals will make contact with the Battery
Connection PCBA terminals when the cover and battery are installed into the lower housing. (
Figure 5-8 shows the correct orientation of the battery.)
2. Slide the battery and cover together into the lower housing. Apply pressure to the battery cover to hold in place. Use a T10 Torx® driver to install the two screws at the rear cover (do not over-tighten).
3. Install the screw at the bottom of the cover.
Note:
Dispose of used batteries according to your local governing ordinances and recycling regulations.
5.7.5 Exhalation Block Removal, Cleaning, and
Replacement
The exhalation block is for single patient use only and must not be reused with a new patient. It may, however, be removed from the ventilator, cleaned, and re-installed for use on the same patient.
Service Manual
Service Manual
Service and Repair
To remove the exhalation block:
Refer to
Figure 5-9 and perform the following steps:
1. Tilt the ventilator upward and unscrew the captive knob holding the exhalation block in place.
2. Grasp the “From patient” fitting and slide the exhalation block away from the ventilator.
Figure 5-9. Removing the Exhalation Block
To clean the exhalation block:
1. Wash the exhalation block with warm, soapy water.
2. Remove excess water with a towel and air dry or blow dry with compressed air.
Caution
To prevent damage to the ventilator electronics, ensure the exhalation block is completely dry before reinserting it back into the ventilator.
To replace the exhalation block:
1. Slide the exhalation block into the exhalation conical fitting and ensure it is fully seated on the two connectors.
2. Tighten the captive screw to lock the exhalation block into place.
Note:
After cleaning or replacing the exhalation block, perform exhalation flow sensor calibration.
5-15
Service and Repair
5.7.6 Housing Disassembly and Assembly
WARNING
To prevent possible injury, always remove the battery before servicing the ventilator.
To disassemble the top housing:
1. Place the ventilator bottom side up on the work surface, and remove the 5 screws that hold the housings together using a T20 Torx® driver (
). Be careful not to scratch the keypad window.
2. Turn the ventilator right side up and gently lift the top housing off the base and turn it right side down on the work surface. Be careful not to pull on the keypad flex circuit.
3. Disconnect the keypad flex circuit by lifting up on the keypad connector locking tab
(J1) (located on the CPU PCBA) and sliding the circuit out of the connector ().
4. Check for the presence of the following internal ventilator labels and replace if damaged or illegible:
• Pneumatic schematic label on the inside of the upper housing
• Attention Caution label on the Power Management PCBA guard
• ESD label inside the device at rear of lower housing.
5-16 Service Manual
Figure 5-10. Removing Top Cover
Service and Repair
Service Manual
Housing screws (5 pl)
5-17
Service and Repair
Figure 5-11. Disconnecting the Keypad Flex Circuit
5-18
To reassemble the upper housing:
1. Slip the keypad flex circuit fully into connector J1 on the CPU PCBA and lock the connector.
Caution
To prevent possible damage, do not fold or crease the flex circuit.
2. Route the flex circuit so that it does not become pinched between the housings, and place the top housing onto the ventilator base assembly, lining up the case halves.
Ensure the inspiratory block is properly seated in the lower and upper housings, and that the inlet air filter is properly positioned in its slot.
Caution
When reassembling the upper housing, avoid damage to the flex circuit by routing it inside the
Buzzer PCBA and inspiratory block.
3. Hold the housings together and turn the ventilator bottom side up and install the 5 screws. Do not over-tighten.
To replace the upper housing:
When replacing the upper housing, you must also install a new keypad and pneumatic diagram label if your housing did not come with these parts pre-assembled. Use the old housing as a reference for the label location (place label on curved surface on inside of upper housing).
Service Manual
Service Manual
Service and Repair
1. To install the pneumatic diagram label, clean the inside housing label surface with isopropyl alcohol using a lint-free wipe, then peel off the adhesive backing and position the label on the new housing.
2. Follow the steps in Section 5.7.8
to install the keypad, then reassemble the upper housing according to the instructions above.
5.7.7 Carrying Handle Removal and Replacement
To remove the carrying handle:
1. Unscrew the two screws on the inside of the upper housing using a T20 Torx® driver.
2. Turn the upper housing over and remove the handle.
To replace the carrying handle:
1. Place the handle in position on the upper housing.
2. While holding the handle in place, turn the upper housing right side down. Using a
T20 Torx® driver, fasten the two screws previously removed. Do not over-tighten.
5.7.8 Keypad Removal and Replacement
Caution
To prevent possible damage, do not fold or crease the flex circuit.
To remove the keypad:
1. Slide a small flat-blade screwdriver or other flat tool under one edge of the keypad and gently lift to break the adhesive bond with the upper housing.
2. Lift the keypad away from the housing.
3. Clean the housing with alcohol to remove any adhesive residue.
To replace the keypad:
1. Ensure the window does not have finger prints or dust prior to assembly. If necessary, remove protective film from window.
2. Peel away the protective paper backing from the new keypad to expose the adhesive.
Use care during handling to ensure no fingerprints or contaminants are deposited on the adhesive surface of the keypad.
3. Route the flex circuit through the slot in the upper housing (
5-19
Service and Repair
4. Align the keypad within the recess in the upper housing and press into place. Use care when re-installing the upper housing so the flex circuit is not pinched.
Caution
Use care when installing the keypad, as any attempt to reposition it can damage the keys.
Figure 5-12. Replacing the Keypad
5-20
5.7.9 Exhalation Solenoid Valve Removal and Replacement
The exhalation solenoid valve is mounted to the CPU PCBA and can be removed without removing the PCBA from the ventilator.
To remove the exhalation solenoid valve:
Refer to
and do the following:
1. Disconnect the exhalation solenoid valve electrical cable from the CPU PCBA connector (J19).
2. Disconnect the two pneumatic tubes from the valve.
3. Remove the two screws that fasten the solenoid valve to the CPU PCBA using a T9
Torx® driver.
Service Manual
Figure 5-13. Pneumatic Tubing Connections (top view)
Service and Repair
Service Manual
1
From inspiratory block port
2
3
From inspiratory block proximal pressure port
Pressure sensors
5
6
4
Connects with tee to exhalation valve pilot pressure port
Solenoid valve mounting screws
Exhalation solenoid valve connector J19
7
8
9
10
Connects with tee to exh. valve pilot pressure port and valve pressure sensor
From turbine
11
12
Exhalation solenoid valve
Do not connect
Inspiratory gas from turbine
FiO
2
connector
To replace the exhalation solenoid valve:
1. Fasten the valve to the CPU PCBA with the two screws previously removed using a T9
Torx® driver. Do not over-tighten.
2. Connect the pneumatic tubes to the valve ensuring correct placement as shown in
Figure 5-13 . The tube from the turbine connects to the valve “IN” port (from turbine)
and the tube to the inspiratory block exhalation valve port connects to the valve
“OUT” port (to inspiratory block). Do not connect any tube to the barbed fitting closest to the upper edge of the PCBA (the “IN” and “OUT” ports are silk-screened on the CPU PCBA (
to unlock
5-21
Service and Repair
3. Connect the valve’s electrical cable to CPU PCBA connector (J19).
5.7.10 CPU PCBA Removal and Replacement
To remove the CPU PCBA:
1. Disconnect the following cables and pneumatic tubes from the CPU PCBA:
• power management PCBA cable (J7)
• turbine control PCBA cable (J4)
• exhalation valve pressure sensor tube (1 VALVE)
• inspiratory pressure sensor tube (2 INTERNE)
• proximal pressure sensor tube (3 PROXIMALE)
• O
2
valve cable (VALVE O
2
)
• remote alarm cable (J2)
• alarm cable (J18).
5-22 Service Manual
Figure 5-14. CPU PCBA Cables and Tubes (top view)
2
Service and Repair
Service Manual
1
2
Power management cable
Turbine control PCBA cable
3
4
PCBA mounting screw
Exh. valve pilot pressure tube
5
Insp. pressure tube
6
7
Prox. pressure tube
O
2
valve cable
PCBA mounting screw
8
9
Remote alarm cable
10
PCBA mounting screw
5-23
Service and Repair
2. Remove the three CPU PCBA mounting screws using a T10 Torx® screwdriver and turn the PCBA over to disconnect the following pneumatic tubes from the sensors on the
PCBA:
• inspiratory flow sensor (IN) and (OUT) tubes
• exhalation flow sensor (IN) and (OUT) tubes
• exhalation solenoid valve (IN) and (OUT) tubes.
Note:
Observe tube routing to ensure tubes are not kinked or pinched upon reassembly.
Figure 5-15. CPU PCBA Tubing Connections (underside view)
5-24
2
3
1
Exh flow sensor (OUT)
Insp flow sensor (IN)
Exh solenoid valve (IN)
5
6
4
Exh solenoid valve (OUT)
Insp flow sensor (OUT)
Exh flow sensor (IN)
To remove the turbine tube assembly:
1. Gently pull the tube from the turbine and inspiratory block connectors.
To replace the turbine tube assembly:
1. Push the tube onto the inspiratory block and turbine connectors, as shown in
. Ensure the tube is inserted fully on the connectors.
To replace the CPU PCBA:
1. Reconnect the pneumatic tubing to the flow sensors and exhalation solenoid valve on the underside of the CPU PCBA according to the diagram shown in (
for tubing lengths and diameters.
Service Manual
Figure 5-16. Pneumatic Tubing Diagram
Service and Repair
Service Manual
1 O
2
connector (rear of ventilator)
2 Turbine
3 Insp flow sensor
10 To exh valve port on insp block
11 Proximal pressure sensor (PROXIMALE)
12 Inspiratory pressure sensor (INTERNE)
4 Exh flow sensor
5 Exhalation block
13 Exh pressure sensor (VALVE)
14 Exh solenoid valve (IN)
6 Exhalation conical fitting support 15 Exh solenoid valve (OUT)
7 To prox port on insp block 16 Do not connect
8 Insp block 17 See Note in step 3 below
9 To insp pressure port on insp block 18 O
2
solenoid valve
5-25
Service and Repair
2. Place the CPU PCBA on the standoffs and install 3 mounting screws using a T10 Torx® screwdriver. Ensure the tubing is routed underneath the PCBA so that it is not kinked or pinched.
3. Connect the tubing to the pressure sensor ports on top of the CPU PCBA according to the diagram shown in
.
Note:
Ensure the wider end of the straight connector is attached to the larger diameter tube and that the narrower end of the connector is attached to the smaller diameter tube. See
numbers. The tubes must be fully seated to the connector.
4. Make the following cable connections:
• power management PCBA cable (J7)
• turbine control PCBA cable (J4)
• O
2
valve cable (VALVE O
2
)
• remote alarm cable (J2)
• alarm cable (J18)
5.7.11 LCD Panel Removal and Replacement
To remove the LCD panel:
1. Using a T10 Torx® screwdriver, remove the four screws that fasten the LCD panel to the CPU PCBA.
5-26 Service Manual
Figure 5-17. Removing the LCD Panel
Service and Repair
Service Manual
LCD panel mounting screws (4 pl)
2. Gently lift the LCD panel away from the PCBA and lay it bottom side up on the work surface, being careful not to pull or damage the flex circuit.
3. Release the locking tabs on the LCD panel connector and slide the flex circuit away from the panel.
5-27
Service and Repair
1
Figure 5-18. LCD Flex Circuit Detail
5-28
1 LCD panel
2 Pull connector tabs outward to unlock flex circuit, push inward to lock
3 CPU PCBA
To replace the LCD panel:
1. Slide the flex circuit into the unlocked connector on the LCD panel and push the tabs inward to lock the circuit in place.
2. Turn the LCD panel over and mount to the standoffs on the CPU PCBA using four screws previously removed.
3. Peel back the protective cover from the LCD panel.
5.7.12 3 V Battery Removal and Replacement
The 3 V battery is located on the CPU PCBA under the LCD panel (G1).
Service Manual
To remove the 3 V battery from the CPU PCBA:
Avoid causing a short-circuit by using a small wood or plastic tool (the bare end of a cotton swab is shown in
) to push the battery out of its retainer on the
CPU PCBA.
Figure 5-19. Removing the 3V Battery (LCD panel removed)
Service and Repair
Service Manual
To replace the 3 V battery on the CPU PCBA:
Slide the new battery into the retainer (G1), making sure that the positive side of the
battery is facing upward ( Figure 5-20
).
Note:
Verify time and date are correct after replacing battery.
5-29
Service and Repair
Figure 5-20. Replacing the 3 V Battery
5-30
5.7.13 Inspiratory Block Removal, Cleaning and
Replacement
To remove the inspiratory block:
1. Detach the large tube connecting the inspiratory block (if not already removed) to the turbine assembly.
2. Remove the inspiratory block and connected pneumatic tubes from the ventilator lower housing.
To clean the inspiratory block:
1. Remove the pneumatic tubes from the assembly and replace if dirty. Inspect the inspiratory block for signs of occlusion or damage. Replace if necessary.
2. Dampen a soft cloth with a cleaning solution (see
for a list of approved cleaners and disinfectants) and wipe the inspiratory block thoroughly.
3. Ensure the inspiratory block is completely dry and in good condition before reinstalling into the ventilator.
To replace the inspiratory block:
1. Replace pneumatic tubes prior to installing the inspiratory block (
for tubing part numbers and a detailed assembly drawing.
2. Push the large tube onto the turbine outlet fitting, making sure that the inspiratory block fitting rests on the support bracket.
Service Manual
Note:
Perform inspiratory flow sensor calibration after cleaning or replacing inspiratory block.
Figure 5-21. Installing the Inspiratory Block
Service and Repair
Service Manual
1 Turbine outlet fitting
2 Support bracket
3 Inspiratory block fitting
4 Not all tubes shown
5-31
Service and Repair
5.7.14 FiO
2
Connector and Cable Removal and
Replacement
To remove the FiO
2
connector and cable:
The FiO
2
cable must be replaced together with the connector, as it is soldered in place.
1. If not already disconnected, disconnect the FiO
2
cable connector from (J11 FiO the CPU PCBA.
2
) on
2. Remove the 10-mm hex nut holding the cable assembly to the inspiratory block, and slide the nut away from the connector.
3. Pull the ferrite away from the double-sided tape securing it to the lower housing.
4. Open the ferrite using a flat-bladed screwdriver and remove the cable assembly.
5. Remove the FiO
2
cable from the ferrite, and retain the 10 mm hex nut for reassembly, or use a new one if supplied with the connector.
6. Remove the FiO
2
Cable and Connector from the Inspiratory Block.
To replace the FiO
2
connector and cable:
1. If installed, remove the 10-mm hex nut (and lock washer if present) from the FiO
2 connector/cable assembly.
2. Route the cable wires through the inspiratory block and install the 10 mm hex nut to hold the connector in place in the inspiratory block. Tighten hand-tight. The lock washer is not used during re-assembly.
3. Orient the arrow on the connector so it points toward the lung symbol on the inspiratory block label (straight up).
4. Tighten the 10-mm hex nut to secure the connector. Do not over-tighten.
5. Route the cable wires as shown in
Figure 5-22 , ensuring the wires are routed around
the rear of the PCBA mounting post and are not inside the turbine tubing support area.
6. Ensure the PCBA connector exits the ferrite toward the front of the ventilator.
7. Close the ferrite, ensuring the cable is not pinched.
8. Replace the double-sided tape to ensure good adhesion and secure the ferrite to the tape (see
for tape part number). Tape and ferrite should be placed next to cable guide 3 (
).
9. Re-connect FiO
2 components.
cable to (J 11 FiO
2
) on CPU PCBA after reassembling internal
5-32 Service Manual
Figure 5-22. FiO
2
Cable Routing and Tape Placement
Service and Repair
Service Manual
1 Double sided tape
5.7.15 Buzzer PCBA Removal and Replacement
To remove the buzzer PCBA:
1. If not already disconnected, remove buzzer PCBA connector from J18 on the CPU
PCBA.
2. Slide the PCBA upward out of its slot in the lower housing.
5-33
Service and Repair
Figure 5-23. Buzzer PCBA and Cable Detail
5-34
1 Buzzer PCBA connector J 18
2 Buzzer battery
3 Buzzer PCBA
Service Manual
To replace the buzzer PCBA:
Note:
To ensure the buzzer PCBA switch is set properly, slide the switch to the left (
Figure 5-24. Buzzer PCBA Switch (underside shown)
Service and Repair
Service Manual
1 Slide switch to the left
1. Insert the PCBA into its slot in the lower housing. Ensure the PCBA seats fully into the slot.
2. Re-connect the connector at J18 on the CPU PCBA.
5.7.16 Exhalation Conical Fitting Removal and
Replacement
Note:
The exhalation block may remain in the conical fitting during this procedure and the assembly may be removed as a unit.
To remove the exhalation conical fitting:
1. Tilt the ventilator upward and unscrew the captive knob holding the exhalation block and housing in place.
2. Remove the cover from the ventilator.
3. Lift the exhalation conical fitting and connected pneumatic tubes out of the lower housing.
4. Inspect the pneumatic tubing and exhalation connectors. If dirty, remove the lock nuts fastening the connectors using a 16-mm wrench and socket. If necessary, replace pneumatic tubes and tubing connectors. If the connectors and pneumatic tubes are clean, remove them and set aside for reassembly.
5-35
Service and Repair
To replace the exhalation conical fitting:
1. Install new tubing connectors (if necessary) into the new exhalation conical fitting using the lock nuts previously removed. Ensure the tubing connectors seat properly into the hex-shaped recesses in the inside of the exhalation conical fitting
(
). Do not over-tighten.
Figure 5-25. Exhalation Conical Fitting and Baffle Conduit Fittings
5-36
2. Replace pneumatic tubes (if necessary) prior to installing the exhalation conical
for part numbers and a detailed assembly
drawing.
3. Place the exhalation conical fitting into the lower housing and hand-tighten the captive knob to hold it in place.
4. Route the pneumatic tubes in front of the CPU PCBA mounting standoff.
5.7.17 Captive Knob Removal and Replacement
To remove the captive knob:
Using needle-nose pliers, pull the e-clip away from the threaded portion of the captive knob and remove the knob from the bottom of the lower housing.
Service Manual
Figure 5-26. E-clip Detail
Service and Repair
Service Manual
1
2
Approx. 1/4-in or 8 threads
Pull e-clip with pliers to remove
To replace the captive knob:
1. Push the threaded end of the knob through the opening in the lower housing, and rest the housing on the work surface right side up.
2. Install the e-clip so that it is flush with the surface of the lower housing ( Figure 5-26
).
There should be approximately 1/4 inch (6mm) or 8 threads protruding above the housing surface.
5.7.18 Oxygen Solenoid Valve Removal and Replacement
Refer to
for oxygen solenoid valve removal and replacement.
5-37
Service and Repair
To remove the oxygen solenoid valve:
1. Cut the tie wrap and disconnect the silicone pneumatic tubing from the valve outlet fitting. Be careful not to cut the tubing.
2. Open the tabs on the cable guide 3 (
Figure 5-35 ) to release the oxygen solenoid valve
harness.
3. Remove the screws that hold the valve to its mounting bracket using a T9 Torx® driver.
4. Lift the valve assembly up, unscrew the knurled retainer ring, and pull the oxygen tubing from the barbed inlet fitting. A small flat-bladed screwdriver may be useful for loosening the tubing from the fitting.
Note:
If you need to replace the oxygen inlet tube and connector, you will have to remove the turbine assembly and power management PCBA first (see
Figure 5-27. Oxygen Valve Detail
5-38
1 Oxygen tubing
2 Knurled retaining ring
3 Oxygen valve
4 Tie wrap
5 Valve outlet
6 Valve wiring harness
7 Cable guide
8 Mounting screw
9 Valve mounting bracket
10 Valve inlet
Service Manual
Service Manual
Service and Repair
To replace the oxygen solenoid valve:
1. Fasten the new oxygen solenoid valve to the mounting bracket using the two screws previously removed.
2. Connect the oxygen tubing to the valve inlet fitting, and tighten the knurled retainer ring.
WARNING
Ensure the oxygen tubing is connected tightly to the inlet fitting to prevent oxygen from leaking into the ventilator compartment.
3. Place the oxygen solenoid valve/bracket assembly into the lower housing and fasten in place with the two screws previously removed. Ensure the wiring harness is not caught underneath the mounting bracket.
4. Connect the silicone tube from the turbine assembly to the valve outlet fitting and fasten with a tie wrap using a tie wrap gun to ensure a tight grip on the tubing.
5. Route the valve’s wiring harness through cable guide 3 and bend the tabs closed.
To test the connections for leaks:
1. Connect an O
2
supply to the O
2
inlet connector at the rear of the ventilator. Adjust the source pressure to less than 7 psi or flow less than15 lpm.
2. Apply leak detector solution to each end of the tube where it connects with the oxygen inlet fitting at the rear of the ventilator and the oxygen solenoid valve inlet, and to the tube connecting the oxygen solenoid valve outlet to the turbine.
• Verify no bubbles form at the connections
• If a leak is detected, tighten connection and repeat test until no leaks are found
3. Wipe any remaining leak detector solution from the connections.
5.7.19 Turbine Assembly Removal and Replacement
To remove the turbine assembly:
1. Disconnect the power management/turbine control cable from the turbine control
PCBA connector (J1) by pushing in the tab and pulling the connector.
5-39
Service and Repair
Figure 5-28. Power Management/turbine Control Cable Detail
5-40
1 Cable clamp
2 Toroid
3 Connector J 1
4 Turbine control PCBA
5 Connector to CPU PCBA (J 4)
2. Open the cable clamp and remove the toroid and cable.
3. Cut the tie-wrap that secures the tube connected to the oxygen solenoid valve, and remove the tube from the turbine. Be careful not to cut the tube.
4. Disconnect the large tube from the turbine outlet and inspiratory block.
5. Disconnect the tube between the turbine and the exhalation solenoid valve.
6. Lift the turbine assembly out of the housing.
To replace the turbine assembly:
1. Position the turbine assembly so that it rests on the foam pads in the lower housing.
2. Attach the tube connected to the oxygen solenoid valve to the turbine with a tiewrap as shown in
Figure 5-28 . Tighten the tie-wrap with a tie-wrap applicator gun.
Service Manual
Service Manual
Service and Repair
3. Connect the tube between the exhalation solenoid valve and the turbine.
4. Connect the large silicone tube between the turbine outlet and inspiratory block.
5. Route the toroid/cable through the cable clamp and close the clamp tightly around
the toroid as shown in Figure 5-28 .
6. Connect the cable to connector J1 on the Turbine Control PCBA.
5.7.20 Turbine Control PCBA Removal and Replacement
To remove the Turbine Control PCBA:
1. Disconnect the motor and temperature sensor cables from the Turbine Control PCBA
connector (J6) ( Figure 5-29 ).
2. Open the cable clamp and remove the toroid and cable.
3. Remove the four screws fastening the PCBA and mounting bracket to the turbine housing, using a T10 Torx® driver.
4. Lift the PCBA and bracket from the turbine assembly.
5. Disconnect the toroid/cable assembly from the PCBA by pushing in the tab on the connector at J1 and pulling the connector out of the PCBA. This harness should remain connected to the Power Management PCBA.
5-41
Service and Repair
Figure 5-29. Turbine Control PCBA
5-42
1 Motor cable (J 6)
2 Mounting screw (4 pl)
3 Temp sensor cable
4 Do not pinch cables when reinstalling bracket
5 Tie wrap; Do not use tie wrap gun to tighten
To replace the Turbine Control PCBA:
1. Place the Turbine control PCBA on the turbine assembly standoffs aligning the holes in the PCBA with the standoffs.
2. Place the bracket over the PCBA, and route the cables under the bracket as shown in
Figure 5-29 . Make sure that the bracket captures the two cable assemblies, but does
not pinch them against the turbine assembly body.
3. Evenly tighten the screws using a T10 Torx® driver. Do not over-tighten.
4. Connect the motor cable and temperature sensor cables as shown in
to connector J6 on the PCBA.
5. Place a tie-wrap around the harnesses as shown in Figure 5-29 and hand-tighten.Do
not tighten the tie-wrap with a tie-wrap gun. Cut tail of tie-wrap with cable cutter.
Service Manual
6. If not already assembled, adhere the cable clamp to the corner of the turbine as shown in
7. Connect the toroid/cable assembly to J1 on the turbine control PCBA (
).
8. Secure the toroid in the cable clamp as shown in
.
5.7.21 Turbine Foam Pads Removal and Replacement
To remove and replace lower turbine support pads:
Lift the turbine support pads out of the lower housing and replace with new ones.
There is no adhesive on these pads.
Figure 5-30. Turbine Support Pad Detail
Service and Repair
Service Manual
1 Foam pads
To remove the upper turbine pads:
1. Using a flat tool or scraper, remove the pads from the top of the turbine.
2. Clean the residual adhesive from the turbine housing using isopropyl alcohol. Wipe dry.
5-43
Service and Repair
Figure 5-31. Upper Turbine Pad Detail
5-44
1
Front of ventilator
2
Adhere pads to metal part of turbine housing as shown
To replace the upper turbine pads:
Peel the adhesive backing from the turbine pads and apply to the turbine housing in
the positions shown in Figure 5-31 .
Service Manual
5.7.22 Power Management PCBA Removal and
Replacement
The power management PCBA consists of the power management PCBA and the power supply PCBA. The power supply PCBA is mechanically fastened to the power management PCBA via four standoffs.
To remove the power management PCBA:
1. Remove the battery to prevent damage from an accidental short-circuit.
2. Disconnect the fan from the power management PCBA connector (J6).
Figure 5-32. Power Management PCBA Mounting Screw Locations
Service and Repair
Service Manual
1 Mounting screws
2 Mounting screw
3 Fan cable routed under PCBA prior to connecting
4 Mounting screw
3. Remove the four screws fastening the power management PCBA to the lower housing using a T10 Torx® driver (
5-45
Service and Repair
4. Lift the front of the Power Management PCBA to disconnect it from the Battery
Connection PCBA, then remove from the ventilator housing taking care not to damage the connectors at the rear of the assembly.
To replace the power management PCBA:
1. Place the power management PCBA onto the standoffs in the lower housing so that the connectors fit through the holes at the rear. Make sure that the connector on the power management PCBA is seated properly into Battery Connection PCBA
connector (J1), and that the fan cable is routed underneath the PCBA ( Figure 5-32 ).
2. Fasten the PCBA in place using the four screws previously removed. Do not overtighten.
3. Connect the fan cable to the PCBA connector (J6).
4. Reinstall battery.
5.7.23 Fan Removal and Replacement
To remove the fan:
1. Remove the two screws fastening the fan to the lower housing using a T10 Torx®
Figure 5-33. Fan Detail
5-46
1 Wires routed through retainer in fan housing
To replace the fan:
1. Ensure the fan is oriented as shown in
Figure 5-33 with air flow pointing UP and route
the wiring harness through the retainer on the fan housing.
2. Fasten the fan to the lower housing using two screws placed diagonally as shown in
Figure 5-33 . Do not over-tighten.
Service Manual
Service Manual
5.7.24 Battery Connection PCBA Removal and
Replacement
To remove the battery connection PCBA:
1. Remove the screw fastening the battery connection PCBA stop to the housing using a T10 Torx® driver.
2. Lift the battery connection PCBA out of its slot (
).
Figure 5-34. Battery Connection PCBA
Service and Repair
To replace the battery connection PCBA:
1. Insert the battery connection PCBA into the slot in the lower housing.
2. Install the bracket with the screw previously removed so that it retains the PCBA. Do not over-tighten.
5.7.25 Remote Alarm (Nurse Call) Cable Removal and
Replacement
To remove the remote alarm cable:
1. Bend back the tabs on cable guide 1 to release the cable from the lower housing.
2. Remove the connector lock nut from the inside of the lower housing using an 11-mm open-end wrench or needle-nose pliers (
3. Remove the cable from the back of the lower housing.
5-47
Service and Repair
Figure 5-35. Remote Alarm Cable and Oxygen Tubing Assembly
2
8
1 Cable guide 3 (for oxygen valve harness) 5 Oxygen connector
2 Ferrite 6 Cable guide 1
3 Remote alarm harness
4 Remote alarm cable connector
7
8
Cable guide 2
Tubing shown disconnected from turbine for clarity
To replace the remote alarm harness
1. Remove nut and washer prior to pushing connector through slot in rear of enclosure.
(
)
5-48 Service Manual
Service Manual
Service and Repair
2. Form cable and connector as shown and push through nut. Be careful not to damage wire during this operation.
Figure 5-36. Remote Alarm Harness
1 Location for harness
3. Thread washer and nut over connector as shown in ( Figure 5-38 ) and line up the tab
with the notch on the housing. (
4. Ensure connector socket is orientated so that pin 1 is at the top (
)
5. Hand tighten when connector is correctly aligned. (
6. Tighten locking nut (
) using 12mm wrench. Once complete ensure the socket is orientated so that pin 1 is at the top. Do not over-tighten.
5-49
Service and Repair
Figure 5-37. Aligning Harness Connector Tab with Housing Notch
1 Notch on housing
2 Tab on harness connector
Figure 5-38. Orientation of the Connector and Tightening Locking Nut
5-50
1 Connector
2 Locking nut
7. Route the cable through cable guide 1 (
Figure 5-35 ) and secure it by bending the tabs
over the cable.
8. Open the ferrite.
9. Twist remote alarm cable so that it sits inside the ferrite and clamp the ferrite closed
(
).
Service Manual
10. Secure ferrite to cable guide 3 with a 25 mm length of double-sided tape (See
for part number).
11. Refer to
for replacement label kit.
Figure 5-39. Remote Alarm Harness Installed into Ferrite
Service and Repair
Service Manual
1 Cable guide 3
2 Ferrite
5.7.26 Oxygen Connector/Tube Assembly Removal and
Replacement
To remove the oxygen connector/ tube assembly:
1. Bend back the tabs on cable guide 2 (
Figure 5-35 ) to release the tubing from the
lower housing.
2. Remove the connector lock nut from the inside of the lower housing using a 16-mm open-end wrench or a needle-nose plier (
3. Remove the assembly from the back of the lower housing.
5-51
Service and Repair
To replace the oxygen connector/ tube assembly:
1. Insert the tube assembly through the opening in the back of the lower housing.
Ensure the connector is oriented as shown in
.
Figure 5-40. Oxygen Connector Properly Oriented
5-52
2. Install the lock nut on the connector to secure it to the housing. Do not over-tighten.
3. Route the tube through cable guide 2 ( Figure 5-35 ) and secure it by bending the tabs
over the tube being careful not to pinch the tube.
4. Connect the tubing from the oxygen solenoid valve to the turbine assembly, and secure it with a tie-wrap using a tie-wrap gun.
5.7.27 Lower Housing Replacement
When replacing the lower housing, you must also transfer the old serial number to the new housing. Use care when removing the serial number. If the serial number is damaged during removal, call your local tech support representative for serial number replacement.
If the serial number label (with window) is damaged during removal, order a new label kit
(part number is provided in
).
To replace the lower housing:
1. Follow the removal steps listed above to remove all components from the lower housing.
2. Reverse the disassembly steps to reassemble ventilator.
Service Manual
Service Manual
5.8
Post-ventilator Repair Information
If you have made any repairs that have required opening the ventilator housing, you must run the Performance Verification procedures described in
describes what testing and calibration is required following ventilator service
Table 5-2. Testing and Calibration Requirements
Type of service
Ventilator installation
Software download
Biennial checkout (every two years)
Required test or calibration
POST
1. Software download
2. POST
3. Visual inspection
1. POST
2. Visual inspection
3. Electrical safety
4. Performance verification testing, including calibrations
15K preventive maintenance
1. POST
2. Visual inspection
3. Electrical safety
4. Performance verification testing, including calibrations
Battery replacement
1. POST
2. Visual inspection
3. Measurements check
Keypad and/or upper housing assembly replacement
1. POST
2. Visual inspection
3. Electrical safety
LCD replacement
1. POST
2. Electrical safety
3. Calibrations
4. Performance verification testing, including calibrations
Service and Repair
5-53
Service and Repair
Table 5-2. Testing and Calibration Requirements (continued)
Type of service
CPU PCBA or power management PCBA replacement
Required test or calibration
1. POST
2. Visual inspection
3. Software download
4. Electrical safety
5. Performance verification testing, including calibrations
Replacement of:
•
Battery connection PCBA
•
Buzzer PCBA
•
Exhalation conical fitting
•
Exhalation valve solenoid
•
Inspiratory block
•
Lower housing
•
Oxygen connector
•
Oxygen solenoid
•
Turbine
•
Turbine controller PCBA
1. POST
2. Visual inspection
3. Electrical safety
4. Performance verification testing, including calibrations
5-54 Service Manual
6 Performance Verification
This chapter describes how to verify the performance of the Puritan Bennett™ 560
Ventilator after it has been repaired or has undergone preventive maintenance. It includes calibration and final functional tests. Refer to
Table 2-14
for required tools and service materials.
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
6.1
Test Prerequisites
Note:
•
This manual assumes English (US) language is selected in the ventilator SETUP (
Ventilator messages will differ, depending upon the language selected.
•
Test the device under ambient conditions. If the device has been in use recently, allow it to reach ambient temperature before testing.
•
All tests must be run in series, and the device must pass all tests before returning the unit to service.
•
Follow your institution’s policy for documenting the Performance Verification.
•
Set Apnea alarm to OFF in Preferences menu.
•
Turn all alarms OFF or set to their minimum and maximum values.
To troubleshoot any performance verification or calibration:
For any performance verification test or calibration that fails, use the following steps to troubleshoot the failure:
1. Verify PTS 2000, BreathLab PTS RPTS application, and ventilator settings and tubing setups are correct.
2. Repeat the test or calibration steps.
3. Replace associated component(s).
4. Check for specific alarm messages. Reference Chapter 4 to troubleshoot calibration
failures.
6-1
Performance Verification
6.2
When to Run Performance Verification
Run performance verification tests following ventilator service as summarized in
.
Type of service
Software download
Table 6-1. Testing and Calibration Requirements
Biennial checkout (every two years) or
15000 hour preventive maintenance
Battery, lithium, 3V and any internal harness replacement
Required test or calibration
1. POST
2. Visual inspection
1. POST
2. Visual inspection
3. Electrical safety
4. All calibrations
5. All performance verification testing
1. POST
2. Visual inspection
3. Electrical safety
4. Measurements check.
5. All calibrations.
6. All alarms tests
7. Remote alarm test
8. Power off while ventilating test
Battery, lithium-ion 4.8 Ah memory
Keypad or upper housing assembly replacement
LCD replacement
CPU or power management PCBA replacement
1. POST
2. Visual inspection
3. Measurements check
4. Power management software download if not version AL020002 or later
1. POST
2. Visual inspection
3. Electrical safety
1. POST
2. Electrical safety
3. Calibrations
1. POST
2. Visual inspection
3. Software download, as applicable
4. Electrical safety
5. All calibrations
6. All performance verification testing
6-2 Service Manual
Service Manual
Table 6-1. Testing and Calibration Requirements (continued)
Type of service
Replacement of:
•
Battery connection PCBA
•
Buzzer PCBA
•
Fan
Replacement of:
•
Exhalation conical fitting support
•
Exhalation valve solenoid
•
Lower housing assembly
•
Inspiratory conical fitting
•
Oxygen female connector
•
Oxygen solenoid valve
•
Any internal pneumatic tubing
•
Turbine/blower
•
Turbine controller PCBA
Required test or calibration
1. POST
2. Visual inspection
3. Electrical safety
4. Measurements check
5. All alarms tests except remote alarm test
6. Power off while ventilating test
1. POST
2. Visual inspection
3. Electrical safety
4. All calibrations
5. All performance verification tests
Performance Verification
6-3
Performance Verification
6.3
Required Materials and Equipment
• PC operating on Windows XP or 7
• Serial cable, RS232, DB9 male to DB9 female, 4-074688-00
• USB 2.0 serial converter, local supplier
• PTS 2000 pneumatic analyzer, 4-076185-00
• BreathLab PTS RPTS application software, 4-075356-00
• Patient circuit, pediatric dual-limb, with exhalation valve, 5093900
• Patient circuit, adult dual-limb, with exhalation valve, 5094000
• Test Lung, 1.0-liter, Deluxe Adult, White with Restrictor, P/N LNG800P or equivalent
• Test Lung, 4-liter, 4-075578-00
• Gold standard tube, 22-mm ID X 21.0 inch flex tube, 4-018506-00
• Connector, tubing junction, 22-mm OD barbed with pressure port, 4-011521-00
• Tee connector, 1/8-inch ID tubing, local supplier
• Silicone tubing, 1/8-inch ID, various lengths, 4-008578-00 or equivalent
• Silicone tubing, 7/32-inch ID, various lengths, local supplier
• Silicone tubing, 3/16-inch ID, various lengths, 4-008577-00 or equivalent
• Connector, tubing, 3/16-inch ID to 1/8 inch ID, 4-006104-00 or equivalent
• Silicone barbed coupling, 22-mm ID, 4-003443-00 (3 required)
• Calibration shell, S-231702-00B
• Adapter, 15-mm ID/22-mm OD x barb, 4-002902-00 or equivalent
• Stopper, no. 2, G-061574-00 or equivalent
• Stopper, no. 1, 4-009523-00 or equivalent
• ReFlex filter, barbed, 4-074644-00
• DC power supply, adjustable, minimum 2 A output, Extech model 382202 or equivalent
• DC power, test cable, 10000021
• Digital multimeter, Fluke 87 or equivalent
• Electrical Safety Analyzer, Fluke ESA620 or equivalent
• Fluke ESA620 accessory kit (test leads, etc.)
• Test cable, ground, 10000017
• Remote alarm test box, RA I (black plastic connector), 10000019
• Remote alarm test box, RA II (gray metallic connector), 10005329
• FiO
2
sensor kit, 3814199
• Oxygen connector, male, 2962700
• Oxygen supply, regulated to <7 psi at < 15 lpm, local supplier
6-4 Service Manual
Performance Verification
• USB memory device, 128 MB to 4 GB, formatted in 32-bit file format, local supplier
• Static-dissipative Field Service Kit, 4-018149-00
6.3.1 PTS 2000 Setup
1. Set the PTS 2000 up as shown ( Figure 6-1 ) and described below.
Figure 6-1. PTS 2000 Setup
Service Manual
2. Attach a silicone barbed coupling to the high flow inlet port of the PTS 2000.
3. Attach the 22-mm tubing connector with pressure port to the barbed coupling.
4. Attach another barbed coupling to the 22-mm tubing connector.
5. Attach the barbed Re/Flex filter to the barbed coupling, making sure that the flow direction arrow on the filter housing points towards the PTS 2000.
6. Connect one end of a short piece of 3/16-inch ID tubing to the pressure port on the
22-mm tubing junction.
7. Connect the 3/16-inch x 1/8-inch tubing connector to other end of the 3/16 ID tubing.
6-5
Performance Verification
8. Connect the 3/16-inch x 1/8-inch tubing connector to on end of short piece of 1/8inch ID tubing.
9. Connect the other end of 1/8-inch ID tubing to (+) low pressure port on the front of the PTS 2000.
10. Connect the USB 2.0 serial converter to a USB port on the PC.
11. Connect the serial cable to the USB 2.0 serial converter.
12. Connect the other end of the USB serial converter to a serial port on the rear of the
PTS 2000.
13. The USB port must be set to COM1 on the PC for the USB serial converter to communicate with the PTS 2000.
14. Power on the PTS 2000 and allow a ten (10) minute warm-up.
15. Perform oxygen sensor calibration per the PTS 2000 User's Manual.
16. Set the PTS 2000 to REMOTE OPERATION mode before launching BreathLab PTS RPTS application software.
6.4
Electrical Safety Test
This test verifies that the Puritan Bennett 560 Ventilator meets the leakage current requirements of IEC 60601-1:1990 or 1988 for Class II, Type BF devices. You may use any electrical safety analyzer capable of performing the required tests for this device classification. Refer to
Table 6-2 for appropriate test limits. Refer to the analyzer operating
instructions for details on test setup and test steps.
Required equipment:
• electrical safety analyzer
• electrical safety test ground cable
• patient lead test cable
.
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
To perform an electrical safety test:
1. Test the ventilator according to the electrical safety analyzer’s instructions.
2. Verify the test results are within the listed limits.
6-6 Service Manual
Service Manual
Performance Verification
Test
Enclosure Leakage Current
Enclosure Leakage Current
Enclosure Leakage Current
Enclosure Leakage Current
Mains on Applied Part
Mains on Applied Part
Patient Leakage Current, AC
Patient Leakage Current, AC
Patient Leakage Current, AC
Patient Leakage Current, AC
Patient Leakage Current, DC
Patient Leakage Current, DC
Patient Leakage Current, DC
Patient Leakage Current, DC
Table 6-2. Electrical Safety Test Limits
Line configuration
Normal Condition (NC)
Normal Condition/Reversed Mains (NCRM)
Open Supply (OS)
Open Supply/Reversed Mains (OSRM)
Single Fault Condition (SFC)
Single Fault Condition, Reversed Mains
Open Supply (OS)
Normal Condition (NC)
Open Supply/Reversed Mains (OSRM)
Normal Condition/Reversed Mains (NCRM)
Open Supply (OS)
Normal Condition (NC)
Open Supply/Reversed Mains (OSRM)
Normal Condition/Reversed Mains (NCRM)
6.5
Power-on Indicators Check
WARNING
•
Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
To perform a check of the power-on indicators:
1. Connect the ventilator to AC power.
2. Switch the ventilator power switch to the ON position.
3. Verify the red, yellow, and blue LEDs flash briefly and two alarm tones sound.
4. Verify the blue LED lights and remains lit.
5. Verify the AC LED lights and remains lit.
Test limit
100
A
100
A
500
A
500
A
5000
A
5000
A
500
A
100
A
500
A
100
A
500
A
100
A
500
A
100
A
6-7
Performance Verification
6.6
Ventilator Configuration
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
The Performance Verification process requires service technicians access the Setup screen to configure the ventilator, if not already configured. Configuration includes setting the language, date, time, resetting patient hours to zero, and turning on altitude compensation from the Maintenance Mode > Measurement Check menu. Setting the current time and date provides a frame of reference for any technical faults or errors that may occur during calibration or verification steps.
6.6.1 SETUP Menu Access
To access the SETUP menu:
1. Ensure the ventilator power switch is in the OFF position.
2. Press and hold the ALARM CONTROL key while simultaneously switching the ventilator power switch to the ON position. Continue holding the ALARM CONTROL key until the Setup screen appears (
Figure 6-2. Setup Screen
6-8
Information displayed on this screen is for illustrative purposes only
Service Manual
Service Manual
Performance Verification
Note:
If the ventilator fails to enter SETUP, stop ventilation by pressing the Ventilation ON/OFF key for 3 seconds and pressing again to confirm stop. Power off the ventilator then repeat steps 1 and 2 above.
6.6.2 Ventilator Settings
To configure ventilator settings:
1. Access the SETUP menu. Reference
.
2. Use the UP or DOWN arrow keys to move the cursor as desired.
3. Place the cursor on LANGUAGE.
4. Press ENTER .
5. Use the UP or DOWN arrow keys to select ENGLISH (US) .
6. Press ENTER .
7. Place the cursor on DATE .
8. Press ENTER .
9. Use the UP or DOWN arrow keys to set the current date.
10. Press ENTER .
11. Place the cursor on TIME .
12. Press ENTER .
13. Use the UP or DOWN arrow keys to set the current time.
14. Press ENTER .
6.7
Maintenance Mode
6.7.1 Faults Check
To review the Faults Log:
1. Access the SETUP menu. Reference
2. Use UP or DOWN arrow keys to move cursor as desired.
3. Select MAINTENANCE .
4. Press ENTER .
5. Select FAULTS CHECK .
6. Press ENTER to access the Faults Log.
7. Review the Faults Log and determine any corrective action necessary from the displayed fault codes.
6-9
Performance Verification
Figure 6-3. Maintenance Screen
6-10
Values shown are for reference only
6.7.2 Measurements Check
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
To perform the measurements check:
Before running Measurements Check , warm up the ventilator for at least ten minutes by running turbine at full speed with calibration shell attached to inspiratory port. See
“To run turbine at full speed:” on page 6-22. This is particularly important so that the blower
will be at the correct temperature during the Measurements Check .
1. Disconnect the ventilator from any external DC power source.
2. Connect the ventilator to AC power.
3. Power on ventilator in normal mode and set default parameters per the tables in
4. Ensure that Key Sound is set to Accept Tone in PREFERENCES .
5. Power on ventilator in Maintenance mode (
6. Access the SETUP menu .
Service Manual
7. Use UP or DOWN arrow key to select MAINTENANCE .
8. Press ENTER to access the MAINTENANCE menu .
9. Select Measurements Check (
).
Figure 6-4. Measurements Check Screen
Performance Verification
Service Manual
10. Press ENTER.
11. After the Measurements Check screen appears, verify the displayed values are within
the listed specifications summarized in Table 6-3
.
Table 6-3. Measurements Check Specifications
Parameter
24 V check
Watchdog
Barometric pressure
Specification
23.5 V- 24.5 V
23.5 V - 24.5 V
Ventilator reading is + 11 mmHg of PTS 2000 barometric pressure measurement
Troubleshooting steps
1. Ensure Power
Management cable is connected properly and is not damaged.
2. Verify 24 V on connector
J7 (see
3. Replace Power
Management PCBA
(see
4. Replace CPU PCBA (see
).
Replace CPU PCBA (see
).
1. Ensure no tubing is connected to the barometric pressure transducer.
2. Replace CPU PCBA (see
).
6-11
Performance Verification
Table 6-3. Measurements Check Specifications (continued)
Parameter
Internal temperature
Blower temperature
Buzzer
Specification
30 o
C - 55 o
C (86 o
F - 131 o
F)
35 o
C - 65 o
C (95 o
F - 149 o
F)
Long beep sounds and
Buzzer voltage = 1.7 V - 2.1 V
Troubleshooting steps
1. Relocate ventilator to a cooler location.
2. Ensure side and front cooling vents are not obstructed.
3. Ensure fan is operating properly.
4. Replace Power
Management PCBA (see
).
1. Warm up the ventilator for ten minutes in maintenance mode with calibration shell attached to inspiratory port and turbine speed set to maximum.
2. Ensure turbine control cable is connected properly and not damaged.
3. Replace turbine (see
).
1. Verify default parameters are set as listed in
particularly important that Key Sound is set to
Accept tone in
Preferences menu for correct Buzzer voltage result.
2. Charge the buzzer battery by allowing ventilator to remain powered on and connected to AC power for at least 30 minutes and up to 2 hours.
3. Verify that PCBAmounted switch on buzzer PCBA is set to
OFF.
4. Replace Buzzer PCBA
(see
5. Replace CPU PCBA (see
).
6-12 Service Manual
Service Manual
Performance Verification
Parameter
Backup buzzer
Table 6-3. Measurements Check Specifications (continued)
Specification
Long beep sounds
Altitude Compensation YES
Troubleshooting steps
Replace CPU PCBA (see
).
Set to YES in Measurement
Check menu.
Supplier
Internal battery menu
Approved supplier
First use date
Theoretical capacity
Capacity
Cycles done
Record value
4800 mAh
3840 mAh minimum
N/A
If supplier is not displayed:
1. Power management software download if not version AL020002 or later.
2. Replace lithium-ion battery (see
).
N/A
Replace lithium-ion battery
(see
).
Replace lithium-ion battery
(see
).
N/A
6-13
Performance Verification
Parameter
Battery voltage
Battery temperature
Table 6-3. Measurements Check Specifications (continued)
Specification
23.5 V to 29.7 V
0 o
C - 40 o
C (32 o
F - 104 o
F)
Troubleshooting steps
1. Connect ventilator to AC power source to charge battery.
2. Replace lithium-ion battery (see
).
3. Inspect electrical contacts and proper connection of Battery
Connection PCBA and battery for damage or oxidation.
4. Verify proper connection of Battery Connection
PCBA and Power
Management PCBA.
5. Replace Battery
Connection PCBA (see
).
6. Replace Power
Management PCBA (see
).
7. Verify airflow vents in battery cover are not obstructed.
8. Check for proper operation of fan.
9. Replace lithium-ion battery (see
).
12. Select Buzzer.
13. Press ENTER .
14. Verify a long beep sounds and voltage displayed is 1.7 V to 2.1 V .
Note:
For Buzzer voltage above 2 .
1 V, allow buzzer battery to charge by leaving ventilator powered on in maintenance mode for 30 minutes and up to two hours
.
For Buzzer voltage below 1.7 V
,
verify Key
Sound is set to Accept tone in PREFERENCES menu.
15. Select Back-up Buzzer .
16. Press ENTER .
17. Verify a long beep sounds.
18. Select Internal Battery Menu .
19. Press ENTER .
20. Verify the displayed values are within the listed specifications in
.
6-14 Service Manual
Service Manual
Performance Verification
21. Select Altitude Comp .
22. Press ENTER .
WARNING
The default setting for altitude compensation is YES . Altitude compensation should always be set to YES for accurate volume delivery calculations at all elevations.
23. Press the UP or DOWN arrow keys to select YES if it is not already set to YES.
24. Press ENTER.
Note:
The default setting for altitude compensation is YES .
Altitude compensation should always be set to
YES for accurate volume delivery calculations at all elevations.
25. Select BACK .
26. Press ENTER to return to the MAINTENANCE menu.
6.7.3 Sensors Calibration
This series of steps calibrates the patient pressure sensors and exhalation valve (internal valve) pressure sensor.
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
6.7.3.1 Patient Pressure Sensors Calibration
The Patient Pressure test provides pressure offsets at 0 and 40 cmH
2
O to calibrate the internal and proximal pressure sensors.
Required equipment:
• PTS 2000 pneumatic calibration analyzer
• calibration shell
• tee connector
• 1/8” ID silicone tubing
• 7/32” ID silicone tubing
• silicone coupling.
6-15
Performance Verification
To set up Patient Pressure Sensor and Exhalation Valve (Internal Valve)
Pressure Sensor calibration:
1. Connect the ventilator and the PTS 2000 ( Figure 6-5 ):
2. Block the exhalation valve pilot port using a section of 1/8 inch tubing blocked at one end.
3. Use a 22-mm ID barbed coupling to connect the calibration shell assembly to the ventilator inspiratory port.
Figure 6-5. Pressure Sensors Calibration Test Setup
PTS 2000
7/32” ID tubing silicone coupling calibration shell
1/8” ID tubing tee connector
4. Connect the low pressure tube from the calibration assembly to the positive (+) low pressure port on the front of the PTS ventilator tester.
5. Set the PTS 2000 display to Low Pressure cmH
2
0.
6. Verify PTS 2000 Low Pressure reads 0.0 + 0.1 cmH
2
O. If the PTS 2000 low pressure is not 0.0 ± 0.1 cmH
2
O, zero the PTS 2000 low pressure transducer per the PTS 2000
User's Manual before proceeding.
To calibrate the Patient Pressure Sensor:
1. Enter the ventilator MAINTENANCE Menu.
2. Select Sensors Calibration.
3. Press ENTER .
4. Select Patient Pressure .
6-16 Service Manual
Service Manual
Performance Verification
5. Press ENTER .
6. Press and hold ENTER to accept the 0 (zero) reading.
7. Adjust the Turbine Speed using the UP or DOWN arrow keys to read 39.80 to 40.20 cmH
2
O on the PTS 2000 low pressure measurement.
8. Press and hold ENTER until a short beep sounds. A long beep indicates the calibration failed.
9. Press MENU twice to verify 39.60 cmH
2
O to 40.40 0 cmH
2
O is measured on the PTS
2000 for low pressure.
10. Press MENU to complete the calibration check.
To calibrate Exhalation Valve (Internal Valve) Pressure Sensor:
1. Use the down arrow key to place the cursor at the Int. Valve pressure position.
2. Press ENTER .
3. Confirm successful calibration, indicated by a short beep. If a long beep sounds, the calibration failed.
6.7.3.2 Inspiratory flow Sensor Calibration
Required equipment:
• PTS 2000 pneumatic calibration analyzer
• gold standard tube, 22-mm dia. flex tube.
To calibrate the inspiratory flow sensor:
1. Connect a gold standard (22-mm) tube to the ventilator inspiratory port and the high flow inlet of PTS 2000 as shown in
6-17
Performance Verification
Figure 6-6. Inspiration Flow Calibration Test Setup
6-18
2. Set the PTS 2000 to measure HIGH FLOW, SLPM, air.
3. Ensure the cap on the PTS 2000 High Flow exhaust port is removed.
4. Using the UP or DOWN arrow keys, place the cursor at the Inspiration Flow position.
5. Press ENTER .
6. Verify the display flashes 0 and the PTS 2000 reads 0.00 ± 0.10 slpm.
7. Press and hold ENTER to accept the calibration and proceed to the next calibration point.
8. For each of the flows listed below, use the UP or DOWN arrow keys to adjust the turbine speed to obtain the specified flow measurement after each reading stabilizes.
Service Manual
Service Manual
Performance Verification
9. Press and hold ENTER until a short beep sounds.
0 slpm ± 0.10 slpm
4.90 slpm to 5.10 slpm
11.76 slpm to 12.24 slpm
19.6 slpm to 20.4 slpm
36.26 slpm to 37.74 slpm
58.8 slpm to 61.2 slpm
88.2 slpm to 91.8 slpm
127.4 slpm to 132.6 slpm
If a long beep sounds after any calibration step, the calibration failed and the value is discarded. The previous value is automatically restored.
To verify the inspiration flow calibration:
1. Place the cursor at the Inspiration Flow position.
2. Press MENU .
3. Verify the PTS 2000 readings are within the listed ranges.
0 slpm ± 0.10 slpm
4.50 slpm to 5.50 slpm
11.1 slpm to 12.9 slpm
19.0 slpm to 21.0 slpm
35.1 slpm to 38.9 slpm
57.0 slpm to 63.0 slpm
85.5 slpm to 94.5 slpm
123.5 slpm to 136.5 slpm
4. Press MENU to complete the check.
To troubleshoot inspiratory flow calibration failures:
1. Check vent and PTS 2000 setup.
2. Verify the PTS 2000 and ventilator have warmed up for ten (10) minutes.
3. Recalibrate inspiratory flow.
4. Check for kinked or disconnected internal pneumatic tubing.
6.7.3.3 Exhalation Flow Calibration
Each time the exhalation block is replaced, the exhalation flow sensor calibration must be completed. There is no need to use the PTS 2000 for this calibration. Appropriately set the preferences for patient circuit based on the patient circuit being used. This procedure
6-19
Performance Verification describes how to perform the exhalation flow sensor calibration using an adult dual-limb patient circuit, but tests may also use a pediatric dual-limb patient circuit.
Required equipment:
• adult dual-limb patient circuit or pediatric dual-limb patient circuit
• 22-mm barbed to 22-mm conical connector
• No. 1 stopper.
To calibrate the exhalation flow sensor:
1. Remove the gold standard tube from the PTS 2000 and the ventilator.
2. Connect an adult dual-limb patient circuit to the ventilator.
Note:
Ensure the ventilator Preferences have been set to “OFF” for pediatric dual-limb patient circuit before calibration when using an adult dual-limb patient circuit.
3. Insert a no. 1 stopper into the patient wye.
Figure 6-7. Exhalation Flow Sensor Calibration Test Setup
.
6-20
4. Use the UP or DOWN arrow keys to place the cursor at the Exhalation Flow position.
5. Press ENTER . The ventilator automatically performs the calibration at eight (8) flow points: 0 slpm, 5 slpm, 12 slpm, 20 slpm, 37 slpm, 60 slpm, 90 slpm, 130 slpm.
6. Verify the ventilator emits a short beep after successfully calibrating each flow rate. A short beep indicates pass while a long beep indicates failure.
7. Verify all of the flow points pass calibration as indicated by a short beep.
Service Manual
Service Manual
Performance Verification
Note:
In the event of a calibration error, the ventilator emits a long beep. The ventilator uses the previously saved value as the default and automatically switches to the next calibration point.
Verify proper exhalation valve and patient circuit connections before attempting recalibration.
8. Perform exhalation flow sensor calibration verification by pressing MENU to successively step through each of these flow measurements: 0 slpm, 5 slpm, 12 slpm,
20 slpm, 37 slpm, 60 slpm, 90 slpm, and 130 slpm, verifying that the Inspiration Flow reading is within test limits as specified.
0 slpm ± 0.10 slpm
4.50 slpm to 5.50 slpm
11.1 slpm to 12.9 slpm
19.0 slpm to 21.0 slpm
35.1 slpm to 38.9 slpm
57.0 slpm to 63.0 slpm
85.5 slpm to 94.5 slpm
123.5 slpm to 136.5 slpm
Note:
If any exhalation flow verification is not within test limits, perform Exhalation Flow Sensor calibration again and verify each calibration point passes as indicated by a short beep.
9. Press MENU once more to complete the calibration check.
6.8
Functional Tests
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
6.8.1 Flow Sensor Capacity Test
This test checks the maximum flow capability of the turbine when there is no pressure in the circuit. An external measurement device (PTS 2000) is not used during this test.
To perform the flow sensor capacity test:
1. Remove all tubing from the ventilator.
2. If not already in Maintenance mode, access the Maintenance menu. Reference
3. Use the UP and DOWN arrow keys to place the cursor at the Turbine Speed position.
6-21
Performance Verification
4. Press MENU .
5. Verify the inspiration flow reading is > 145 slpm on the ventilator display.
6. Use the UP and DOWN arrow key to stop the turbine.
6.8.2 Turbine Performance Test
This test verifies the turbine can develop sufficient pressure during a no-flow condition.
Required equipment:
• PTS 2000 pneumatic calibration analyzer
• calibration shell
• tee connector
• 1/8” ID silicone tubing
• 7/32” ID silicone tubing
• silicone coupling.
To run turbine at full speed:
1. Access the SETUP
2. Use UP or DOWN arrow keys to move cursor as desired.
3. Select MAINTENANCE .
4. Press ENTER .
5. Select TURBINE SPEED .
6. Press MENU .
To run the turbine performance test:
1. With the ventilator and PTS 2000 in the same setup from the previous test, connect the calibration shell to the To Patient port on the ventilator using the 22-mm ID barbed silicone coupling (
6-22 Service Manual
Figure 6-8. Test Setup for Turbine Performance Test
Performance Verification
Service Manual block hole in calibration shell
Block hole with your thumb during test
2. If necessary, set the PTS 2000 to measure LOW PRESSURE, cmH
2
O.
3. Verify that low pressure reads 0.0 + 0.1 on PTS 2000. If PTS 2000 low pressure measurement is not 0.0 + 0.1 cmH
2
O, zero the PTS 2000 low pressure transducer per the PTS 2000 User's Manual before proceeding.
4. Use the UP or DOWN arrow keys to position the cursor at the Turbine Speed position.
5. Press MENU to command the turbine to full speed.
6. Block the hole in the calibration shell for no more than three seconds while verifying the PTS 2000 indicates pressure reading > 70 cmH
2
O, and the ventilator displays patient pressure > 70 cmH
2
O.
Caution
Do not block hole for more than 3 seconds or turbine damage could result.
7. Unblock the hole in the calibration shell.
8. Remove tubing from patient pressure port.
9. Verify the patient pressure reading (in the right-hand column of the ventilator display) is 0 ± 0.5 cmH
2
O. If the right-hand column reads > 70 cmH
2
O and the lefthand column reads 0 ± 0.5 cmH
2
O, the internal tubes may be reversed (
6-23
Performance Verification
Figure 6-9. Pressure Calibration Points in Turbine Performance Test
6-24
1
1
2
Reading should be > 70 cmH
2
O
Reading should be 0 ± 0.5cmH
2
O
10. Press the UP or DOWN arrow key to stop the turbine.
6.8.3 Oxygen Solenoid Functional Test
This test ensures the proper function of the oxygen solenoid valve.
WARNING
Ensure the oxygen supply pressure to the machine never exceeds 7 psi (50 kPa) or a flow of
15 lpm.
Required equipment:
• PTS 2000 pneumatic calibration analyzer
• standard tube, 22-mm dia. flex tube
• external oxygen source with 3/16-in. inlet tubing
• oxygen connector, male
• exhalation valve assembly.
To perform the oxygen solenoid functional test:
1. Set the PTS 2000 to display Oxygen% measurement.
2. Connect an oxygen supply to the O
2
inlet connector at the rear of the ventilator.
3. Adjust the source pressure to < 7 psi (at no more than 15 lpm).
4. Ensure the oxygen supply is turned off.
Service Manual
Service Manual
Performance Verification
5. Connect a gold standard tube from the ventilator To Patient port to the High Flow
Inlet on the PTS 2000.
6. Block the exhalation valve pilot pressure port or attach exhalation valve pilot pressure line to the pilot pressure port.
7. With ventilator in normal ventilation mode on standby, set ventilation parameters to default settings.
• Ventilation mode: P A/C
• Pi: 15 cmH
2
0
• PEEP: OFF
• Rise Time: 2
• R-Rate: 10
• Insp Time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 2
• Vt Target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
8. Start ventilation by pressing Ventilation On/Off key.
9. After approximately 30 seconds, verify the PTS 2000 oxygen% measures: 21.0% ±
3.0%.
10. Stop ventilation.
11. Turn on the oxygen supply (< 7 psi, < 15 lpm).
12. Start ventilation.
13. After approximately 30 seconds, verify the PTS 2000 oxygen% reading: > 25.0%.
14. Turn off the oxygen supply and disconnect the oxygen supply hose and connector from the ventilator while continuing ventilation.
15. Allow the PTS 2000 oxygen% reading to return to 21%, to ensure the system is purged of oxygen.
16. Stop ventilation by placing the ventilator into standby mode.
17. Disconnect the gold standard tube from the ventilator.
6-25
Performance Verification
Figure 6-10. Test setup for Oxygen Solenoid Functional Test rear view front view
6.9
Breath Delivery Accuracy Tests
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
Note:
This test is designed to verify ventilator output (volume accuracy + (10 mL + 10%) of setting; pressure accuracy ± (2 cmH
2
O + 8%) of setting using the PTS 2000 in combination with BreathLab
RPTS software application.
6-26 Service Manual
Service Manual
Performance Verification
Required equipment:
• PTS 2000 pneumatic calibration analyzer
• pediatric dual-limb patient circuit
• adult dual-limb patient circuit
• 1.0-L test lung
• 4-L test lung
• gold standard tube, 22-mm dia. flex tube
• tee connector
• 1/8-in. ID silicone tubing
• 7/32-in. ID silicone tubing
• silicone coupling
• calibration shell.
Preconditions:
• PTS 2000 and ventilator must have been powered on for a minimum of ten (10) minutes for proper warm-up.
• Turn off all settable alarms.
• Access the SETUP menu and set Pressure Unit to cmH
2
O and relative pressure to
YES.
• Due to the test setup, the ventilator does not display a VTE reading during volume tests.
6.9.1 Pediatric Volume Accuracy
To verify pediatric volume accuracy:
1. Power on the ventilator in normal ventilation mode on standby.
2. Enter Preferences menu.
3. Set pediatric circuit to YES .
4. Connect pediatric dual-limb patient circuit and exhalation valve to ventilator.
5. Block pediatric dual-limb patient circuit wye using no. 1 stopper.
6. Press MENU to access the Alarms menu.
7. Select VTE .
8. Press ENTER twice so that OFF is flashing in middle (Current) column.
9. While OFF is flashing, use the UP or DOWN arrow key to select YES .
10. Press ENTER to perform exhalation flow calibration.
6-27
Performance Verification
11. Connect the pediatric patient circuit tubing wye to the PTS 2000 high flow inlet filter
(
).
Figure 6-11. Pediatric Volume Accuracy Test Setup
6-28
12. Connect the gold standard (22-mm) tube to the PTS 2000 High Flow outlet.
13. Connect the calibration shell to the other end of the gold standard tube.
14. Set the PTS 2000 to Volume, Air, mL, BTPS, threshold 15 slpm, and Flow-by Tare 4 slpm.
15. With ventilator in normal ventilation mode on standby, set ventilation to the listed parameters.
Service Manual
Service Manual
Performance Verification
16. Select the following settings:
• Mode: V A/C
• VT: 150ml
• PEEP: OFF
• Flow pattern: SQ
• R-Rate: 30 BPM
• Insp Time: 0.5s (I:E ratio: 1:3.0)
• Insp Sens: 3
• Sigh: OFF
• PIP Min: 2
• PIP Max: 60
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm Max: OFF
17. Start ventilation by pressing the Ventilation On/Off key.
18. After the ventilator delivers approximately 15 breaths, compare the VTI reading displayed in the ventilator’s Alarms menu screen is 125 mL to 175 mL.
19. Verify that the delivered volume measured by the PTS 2000 is within calculated test limits (VtI + (8% + 10 mL)).
20. Stop ventilation by using the Ventilation On/Off key.
6-29
Performance Verification
6.9.2 Pediatric Pressure Accuracy
To verify pediatric pressure accuracy:
1. Connect the 4-L test lung to the other end of the standard tube (
Figure 6-12. Pediatric Pressure Accuracy Test Setup
2. Set the PTS 2000 to Low Pressure, cmH
2
O.
6-30 Service Manual
Service Manual
Performance Verification
3. With the ventilator in normal ventilation mode on standby, set the listed parameters.
• Mode: P A/C
• Pi: 10 cmH
2
O
• PEEP: OFF
• Rise time: 1
• R-Rate: 10 BPM
• Insp Time: 1.3 sec (I:E = 1:3.6 or I/T = 22%)
• Insp Sens: 5
• Vt Target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
Note:
Turn OFF all settable alarms.
4. Start ventilation by pressing the Ventilation On/Off key.
5. After the ventilator delivers approximately 15 breaths, verify the Pi reading displayed in the alarm menu screen is 8 cmH
2
O to 12 cmH
2
O.
6. Verify the Pi reading displayed in the ventilator's alarms menu screen is within calculated test limits (PTS 2000 peak low pressure measurement ± (8.75% + 2.04 cmH
2
O)).
7. Stop ventilation by using the Ventilation On/Off key.
6.9.3 Adult Volume Accuracy
To verify adult volume accuracy:
1. Power on the ventilator in normal ventilation mode on standby.
2. Set the pediatric circuit to OFF in the preferences menu.
3. Connect the adult dual-limb patient circuit and exhalation valve to the ventilator.
4. Block the patient circuit wye with a no. 1 stopper.
5. Press MENU to access the Alarms menu.
6. Select VTE .
7. Press ENTER twice so that OFF is flashing in the middle column.
8. While OFF is flashing, use the UP or DOWN arrow key to select YES .
9. Press ENTER to perform exhalation flow calibration.
6-31
Performance Verification
10. Connect the adult dual-limb patient circuit tubing wye to the PTS 2000 high flow inlet
(
).
Figure 6-13. Adult Volume Accuracy Test Setup
11. Connect the gold standard tube to the output of the PTS 2000 High Flow outlet.
12. Connect the calibration shell to the other end of the gold standard tube.
13. Set the PTS 2000 to Volume, Air, mL, BTPS, threshold 15 slpm, and Flow-by Tare 3 slpm.
6-32 Service Manual
Service Manual
Performance Verification
14. With the ventilator in normal ventilation mode on standby, set the listed parameters.
• Mode: V A/C
• Vt: 500 mL
• PEEP: OFF
• Flow pattern: SQ
• R-Rate: 10 BPM
• Insp Time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 5
• Sigh: OFF
• PIP Min: 2
• PIP Max: 60
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm Max: OFF
15. Start ventilation by pressing the ventilation On/Off key.
16. After the ventilator delivers approximately 15 breaths, compare the VtI reading displayed in the alarm menu screen is 440 mL to 560 mL.
17. Verify that the VtI reading displayed on the ventilator's Alarms menu screen is within calculated test limits (PTS 2000 volume measurement ± (8% + 10 mL)).
18. Stop ventilation by using the ventilation On/Off key.
6.9.4 Adult Pressure Accuracy
To verify adult pressure accuracy:
1. Connect the 15-mm ID port of 22-mm OD/15-mm ID adapter to the open end of the gold standard tube.
6-33
Performance Verification
2. Connect the 1.0-liter hard-sided test lung to the 15-mm ID port of the 22-mm/15-mm adapter (
).
Figure 6-14. Adult Pressure Accuracy Test Setup
6-34
3. Set the PTS 2000 set to Low Pressure, cmH
2
O.
4. With the ventilator in normal ventilation mode on standby, set the listed parameters.
• Mode: P A/C
• Pi: 20 cmH
2
O
• PEEP: OFF
• Rise time: 1
• R-Rate: 10 bpm
• Insp Time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 5
• Vt target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
Service Manual
Service Manual
Performance Verification
5. After the ventilator delivers approximately 15 breaths, verify the Pi reading displayed in the alarm menu screen is 17 cmH
2
O to 23 cmH
2
O.
6. Verify that the Pi reading displayed in the ventilator's Alarms menu screen is within calculated test limits (PTS 2000 Low Pressure measurement ± (8.75% + 2.04 cmH
2
0)).
7. Stop ventilation by using the ventilation On/Off key.
6.10 AC/DC/Battery Power Switching Test
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
Required equipment:
• external DC power supply capable of at least 2 A output
• digital multimeter
• DC test cable.
To verify AC/DC/Battery power switching:
1. Disconnect AC power from the ventilator.
2. If not already connected, connect the DC test cable to the DC power supply’s positive and negative outputs and External Battery connector on the ventilator’s rear panel
6-35
Performance Verification
(
Figure 6-16 , rear view, Figure 6-15 , front view).
Figure 6-15. Power Switching Test Setup (front view)
Figure 6-16. Power Switching Test Setup (rear view)
6-36
3. Set the external DC power supply voltage to 10.0 V.
4. Use a multimeter to verify the voltage output.
5. Power on the ventilator in normal ventilation mode on standby.
6. Verify the Internal Battery indicator LED lights.
Service Manual
Service Manual
Performance Verification
7. Slowly increase the external DC power supply voltage until the ventilator’s DC power indicator LED lights.
a. Verify the ventilator’s DC power indicator LED lights when DC power supply voltage measures 10.5 V ± 0.3 V on the digital multimeter.
b. Verify the ventilator’s AC power indicator LED is not lit.
8. Slowly decrease the external DC power supply voltage until the ventilator’s internal battery LED lights.
a. Verify the ventilator’s internal battery indicator LED lights when DC power supply voltage is 9.8 V ± 0.3 V (on the digital multimeter).
b. Verify the ventilator’s DC power indicator LED is not lit.
c. Verify the ventilator shows the message AC Power Disconnection and on Internal
Battery .
9. Press the ALARM CONTROL key to pause the alarm.
10. Very slowly increase the external DC power supply voltage until the ventilator’s DC power indicator LED lights.
a. Verify the ventilator’s DC power indicator LED lights when DC power supply voltage is 12.8 V ± 0.3 V on the digital multimeter.
b. Verify the ventilator’s internal battery power indicator LED is not lit.
11. Reconnect the AC mains cord to the ventilator.
a. Verify the ventilator’s AC power indicator LED lights.
b. Verify the ventilator’s DC power indicator LED is not lit.
6.11 FiO
2
Sensor Calibration
Required equipment:
• FiO
2
sensor
• silicone coupling
• 2” length of 1/8” ID silicone tubing, plugged at one end
• 1” length of 1/8” ID silicone tubing
• 5” length of 3/16” ID silicone tubing
• calibration shell.
To calibrate FiO
2
sensor:
1. Set up the ventilator with FiO
2
sensor kit, adult patient circuit with exhalation valve
and 1.0-liter hard-sided test lung as shown in Figure 6-17 .
2.
Power on the ventilator in normal ventilation mode on standby.
3. Press MENU to enter the Alarms menu.
4. Turn off all settable alarms except FiO
2
%.
6-37
Performance Verification
5. Place the cursor at the FiO
2
position.
6. Press ENTER twice to move to the “Current” column.
7. Press the UP arrow key to select YES.
8. Press ENTER to perform the calibration. If no calibration errors are displayed, then the calibration is successful.
Figure 6-17. FiO
2
Sensor Calibration Setup
6-38
6.12 FiO
2
Sensor Detection Test
1. Set up the ventilator with FiO
2
sensor kit, adult patient circuit with exhalation valve
and 1.0-liter hard-sided test lung as shown in Figure 6-17 .
2.
Power on the ventilator in normal ventilation mode on standby.
Service Manual
Service Manual
3. Set the listed parameters.
• Ventilation mode: P A/C
• Pi: 15 cmH
2
O
• PEEP: OFF
• Rise time: 2
• R-Rate: 10 bpm
• Insp time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 5
• Vt Target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
4. Start ventilation by pressing ventilation On/Off key.
5. Verify the FiO
2
reading is 19% to 23%.
6. Disconnect the FiO
2
sensor cable from the FiO
2 electrical connector on the ventilator’s inspiratory
Figure 6-18
).
Performance Verification
6-39
Performance Verification
Figure 6-18. O
2
Sensor Cable Disconnected
6-40
7. Verify the FiO2 sensor missing message appears.
8. Set FiO
2
Min and Max alarms to OFF in Alarms menu.
9. Stop ventilation by using the ventilation On/Off key.
6.13 Remote Alarm Test
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
Required equipment:
• Adult patient circuit and test lung
• Remote alarm test box
• Multimeter
Service Manual
Performance Verification
To perform remote alarm test:
1. Set up ventilator with an adult patient circuit with exhalation valve and 1.0-liter test lung.
2. Connect the remote alarm test box cable to the ventilator (
Figure 6-19. Remote Alarm Test Setup (front view)
Service Manual 6-41
Performance Verification
Figure 6-20. Remote Alarm Test Box Connected to Ventilator (rear view)
6-42
3. With the ventilator in normal ventilation mode on standby, set the listed parameters:
• Mode: P A/C
• Pi: 15 cmH
2
O
• PEEP: OFF
• Rise time: 2
• R-Rate: 10 bpm
• Insp time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 5
• Vt Target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
4. Set the Alarm volume in the preferences menu to medium volume.
5. Access the Alarms menu and turn off all settable alarms.
6. Set the multimeter to read ohms.
7. Connect the multimeter negative lead to the common connection on the test box and the positive lead to the normally closed connection.
Service Manual
Service Manual
Performance Verification
8. Verify the multimeter reads less than one (1) ohm (closed).
9. Connect the positive lead to the normally open connection and verify the multimeter reads open.
10. Move the multimeter positive lead to the normally closed pin (pin 3), and verify the multimeter reads less than one (1) ohm (closed).
Figure 6-21. Remote Alarm Pin Diagram
4
3
2
1
8
7
5
6
1 relay common 5 not used
2 normally open (NO) 6 not used
3 normally closed (NC) 7 not used
4 not used 8 not used
11. Start ventilation by pressing ventilation On/Off key.
12. Disconnect the AC power cord from the ventilator.
13. Verify the multimeter reads open.
14. Verify the ventilator AC power indicator LED is OFF and the internal battery indicator
LED is lit.
15. Verify the ventilator's medium priority alarm sounds, the yellow alarm LED is flashing, and the AC POWER DISCONNECTION alarm is active.
16. Connect AC power cord.
17. Connect multimeter positive to normally open connection on the test box and the negative multimeter lead to the common connection on the test box.
18. Verify the multimeter reads open.
19. Disconnect AC power cord.
20. Verify the AC power indicator LED is off and internal battery indicator LED is lit.
21. Verify the multimeter reads less than one (1) ohm (closed).
22. Verify a medium-priority alarm sounds, yellow alarm LED flashes, and AC POWER
DISCONNECTION alarm is active.
23. Reconnect AC power.
24. Verify the multimeter reads open.
25. Stop ventilation by using the ventilation On/Off key.
6-43
Performance Verification
6-44
6.14 Alarms Tests
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
Required equipment:
• dual limb patient circuit
• 1.0-liter test lung.
6.14.1 AC Power Disconnection Alarm Test
To test the AC power disconnection alarm:
1. Setup ventilator with adult patient circuit with exhalation valve and 1.0-liter test lung.
2. With ventilator in normal ventilation mode on standby, set the listed parameters.
• Mode: P A/C
• Pi: 15 cmH
2
O
• PEEP: OFF
• Rise time: 2
• R-Rate: 10 bpm
• Insp time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 5
• Vt Target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
3. Access Alarms menu and turn off all settable alarms.
4. Start ventilation by pressing ventilation On/Off key.
5. Unplug the AC power cord from the ventilator.
6. Verify a medium priority alarm sounds.
7. Verify an amber alarm LED flashes continuously.
8. Verify AC POWER DISCONNECTION message appears in the alarm display area.
9. Reconnect the AC power cord to the ventilator.
10. Verify AC POWER DISCONNECTION alarm automatically resets.
Service Manual
Service Manual
Performance Verification
6.14.2 Patient Disconnect Alarm Test
To test the patient disconnect alarm:
1. With the device still in ventilation mode from the previous test, disconnect the inspiratory filter at the To Patient port.
a. Verify the High Priority alarm sounds within 15 seconds.
b. Verify the red High Priority LED flashes.
c. Verify the LOW PRESSURE DISCONNECT* *IF PERSISTS RESTART/SRVC messages appear in the alarm display area.
2. Reconnect the inspiratory filter to the To Patient port.
3. Verify the alarm auto-resets.
4. Stop ventilation by using the ventilation On/Off key.
6.14.3 Power Off While Ventilating Test
To power off while ventilating test:
1. Setup the ventilator with adult patient circuit, exhalation valve, and 1.0-liter hardsided test lung.
2. With ventilator in normal ventilation mode on standby, set the listed parameters.
• Ventilation mode: P A/C
• Pi: 15 cmH
2
O
• PEEP: OFF
• Rise time: 2
• R-Rate: 10 bpm
• Insp time: 1.3s (I:E ratio: 1:3.6 or I/T= 22%)
• Insp Sens: 5
• Vt Target: OFF
• VTI mL Min: OFF
• VTI mL Max: OFF
• VTE mL Min: OFF
• VTE mL Max: OFF
• R tot bpm High: OFF
3. Start ventilation by using the ventilation On/Off key.
4. While ventilating, power off the ventilator using the power switch.
a. Verify the VHP alarm sounds continuously for a minimum of 120 seconds.
b. Verify ventilation stops (no LED or alarm message appears on the display)
6-45
Performance Verification
5. After 120 seconds has elapsed, press the ALARM CONTROL key once to pause the alarm.
6. Power on the ventilator using the power switch.
7. Verify the ventilator immediately resumes ventilation.
8. Stop ventilation by using the ventilation On/Off key.
6.14.4 USB Ports Test
Required equipment:
USB memory device (128 MB minimum memory requirement and formatted in the
32-bit file format)
To test the USB ports:
1. With ventilator in normal ventilation mode on standby , connect a USB memory device to one of the USB type A ports at the rear of the ventilator.
2. Verify the ventilator annunciates two short beeps when the USB device is connected.
3. If the ventilator does not recognize the USB memory device, try another USB memory device.
4. Verify the ventilator screen displays the USB Key Management menu.
5. Verify the Wait message appears.
6. After the Wait message disappears, remove the USB memory device from the USB port.
7. Verify the ventilator annunciates two short beeps with disconnection of the USB device.
8. Insert the USB device into the other USB port and repeat steps 2 -7.
6.15 Verify Default Settings
WARNING
• Never perform any of the tests described in this chapter while a patient is connected to the ventilator.
• This test requires working with exposed electronics connected to AC power. In order to prevent injury or damage to the ventilator, use extreme caution in keeping hands and tools from touching electrical components.
To verify default ventilator settings
1. With ventilator in normal ventilation mode on standby, set the ventilator parameters, alarms, and preferences according to
, and
.
2. Verify the time and date are current.
6-46 Service Manual
Service Manual
Performance Verification
Table 6-4. Default Ventilator Settings
Mode
Parameter
Pi
PEEP
Rise Time
Respiratory rate
Insp Time or I:E
Insp Sens
Vt Target
Setting
P A/C
15 cmH
2
O
OFF
2
10
1.3s (I:E ratio: 1:3.6 or
I/T= 22%)
2
OFF
Table 6-5. Default alarm settings
Alarm
VTI
VTE
Leak (if applicable)
Rtot
Setting
Min 300 mL; Max 2000 mL
Min 300 mL; Max1000 mL
High OFF
High OFF
Table 6-6. Default preferences
Preference
Backlight
Contrast
Setting
YES
Alarm Volume
Key Sound
Waveforms Display
Pediatric Circuit
Accept tone
YES
NO
6.16 Clear Faults
Note:
Clearing the fault log also clears the alarm log and the event log. Any faults must be addressed prior to completing Performance Verification.
1. Press and hold the ALARM CONTROL key while simultaneously using the main power switch to power on the ventilator to enter SETUP mode.
2. Use the UP and DOWN arrow keys to place the cursor at the Maintenance position.
6-47
Performance Verification
3. Press ENTER . The cursor moves to the Faults Check position.
4. Press and hold the ALARM CONTROL key until a long beep sounds.
5. With the cursor at the Faults Check position, press ENTER to display the fault log.
6. Verify NO DATA appears on the FAULTS screen.
7. Press ENTER to exit the Faults Check function.
6.17 Reset Patient Hours
Note:
Resetting the patient hours will also reset the trend data stored in the device memory in preparation for a new patient.
1. Press and hold the ALARM CONTROL key while simultaneously using the main power switch to power on the ventilator to enter SETUP mode.
2. Use the UP and DOWN arrow keys to place the cursor at the Patient Hours position.
3. Press ENTER . The cursor moves to the Reset Hours position.
4. Press ENTER so OFF flashes.
5. Press the UP and DOWN arrow keys to change the message to YES .
6. Press ENTER .
7. Verify a long beep sounds and the patient counter display resets to 00000 h and
00 min .
8. Press the UP and DOWN arrow keys to exit the Reset Patient Hours function.
6-48 Service Manual
7 Parts List
This chapter identifies the repair parts available for the Puritan Bennett™ 560 Ventilator.
7.1
How to Use the Parts Lists
This chapter shows various views of the entire ventilator system and its components. To use the parts lists, refer to the appropriate drawing for the part(s) you need to order. The following assembly drawings are provided:
• major system components
• electrical system components
• pneumatic system components
• turbine assembly components
• labels.
The corresponding table for each drawing lists the item number, part number, quantity required, and part description. If a part number does not appear, it is because it cannot be ordered individually. You must order the assembly, instead. If part numbers are listed, even though they may be included in an assembly, then these may be ordered individually.
Abbreviations used in this chapter are listed in
.
Table 7-1. Abbreviations
Abbreviation
HEXALOBE
ID
OD
PCBA
Meaning
Hexalobular fastener (Torx®)
Inside Diameter
Outside Diameter
Printed Circuit Board Assembly
7-1
Parts List
5
6
10
11
12
13
7
8
9
14
15
16
7.2
Major System Components
Table 7-2. Major System Component Part Numbers
Item no.
1
2
3
--
4
Part no.
10046895
3823200
10049509
not available separately order upper housing assembly
10046894 (label kit)
Qty
1
1
1
--
1
Description
Keypad (part of upper housing assembly)
Handle (part of upper housing assembly)
Upper housing assembly (includes, keypad, handle with screws, and labels)
Upper housing (part of upper housing assembly)
20
21
22
17
18
19
2960600
2966100
3812500
10053276
10028768
3812400
2970800
10028753
10025114
10023294
10025115
2957500
10053275
3818699
2699400
10046894 (label kit)
10028771
2399700
2
4
4
4
3
1
4
4
2
1
1
4
1
1
1
1
1
Label, pneumatic cable diagram (part of upper housing assembly)
Screw, HEXALOBE, M4 x 10, galvanized steel countersunk (2 required for handle; 5 required for housings)
Galvanized steel TCB screw HEXALOBE M3 x
5
Shielding plate motor harness
Turbine control PCBA
Turbine
Hexagonal spacer M/F M3 LG5
Exhalation solenoid valve mounting screw
LCD panel
Screw, M3 x 8 (LCD panel mounting screws)
LCD panel spacer
LCD panel washer
Galvanized steel TCB screw HEXALOBE D3.5 x 8
CPU PCBA
Exhalation solenoid valve
Truarc retaining ring (holds captive knob)
(part of lower housing assembly)
ESD label (part of lower housing assembly)
Filter, inlet air (pack of 6)
Adhesive collar (cable guides) (part of lower housing assembly)
7-2 Service Manual
Service Manual
Item no.
Figure 7-1
23
26
27
28
--
29
30
31
24
25
32
33
34
35
Table 7-2. Major System Component Part Numbers (continued)
Part no.
10049508
10005243
2961500
2960600
10080325
3809700
10046892
10046803
10029945
10028747
See
country-specific part numbers not available separately: order power cord assembly not available separately: order power cord assembly
-not available separately: order switch cover assembly 10028765 not available separately: order switch cover assembly10028765
Qty
1
1
4
5
1
1
1
--
2
1
1
1
1
1
1
Description
Lower housing assembly (compatible with original remote alarm harness 10028752)
Lower housing assembly (compatible with updated remote alarm harness 10005246)
Adhesive feet 12.7 x 12.7 (part of lower housing assembly)
Screw, HEXALOBE, M4 x 10, galvanized steel countersunk (2 required for handle; 5 required for housings)
Inspiratory conical fitting assembly (with labels
Knurled captive knob (part of lower housing assembly
Exhalation block assembly (with labels)
Exhalation block assembly (with labels, 10pack)
Baffle conduit, fluted (exhalation connectors)
Exhalation conical fitting support
Power cable
Cable grip, 6.6 x 4.3 (part of power cord assembly)
Clip, cable grip, 6.6 x 4.3 (part of power cord assembly)
Cylindrical pin 2 x 28 stainless steel (part of switch cover assembly)
Switch cover (part of switch cover assembly)
Parts List
7-3
Parts List
Table 7-2. Major System Component Part Numbers (continued)
--
--
--
--
--
--
Item no.
Figure 7-1
--
Part no.
not available separately: order switch cover assembly 10028765
10046894 (label kit)
10046894 (label kit)
10028765
10046894 (label kit)
10046894 (label kit
10046894 (label kit
Qty
1
1
1
1
1
1
1
Description
Cover spring (part of switch cover assembly)
Label, valve connection (see Figure 7-7 , item
1)
Label, Exhalation single patient use (see
, item 2)
Switch cover assembly
Label O
2
Label, FiO
2
Label, Exh valve/proximal pressure see
, item 2
7-4 Service Manual
Figure 7-1. Major System Components
Parts List
Service Manual 7-5
Parts List
7.3
Electrical System Components
Table 7-3. Electrical System Component Part Numbers
Item no.
6
7
8
9
1
2
3
4
5
10
11
12
13
14
15
16
20
21
22
23
17
18
19
24
25
Part no.
10108395
10053275
10028757
3810200
2957500
10028754
10000265
10053277
2957500
2966300
10087047
2973900
2960100
10054177
1006894 (label kit)
10028752
10005246
10000914
10028763
3827900
10053276
10028751
10053277
10053278
1006894 (label kit)
10053275
--
1
1
1
1
1
1
1
1
1
Quantity
1
1
1
1
1
1
1
1
2
2
3
1
4
6
1
1
Description
Buzzer PCBA
CPU PCBA (includes LCD)
Battery connection PCBA
Stop, Battery connection PCBA
Zinc plated steel TCB screw HEXALOBE D3.5 x
8 (fastener for battery connection PCBA)
Fan with cable
Zinc plated steel TCB screw HEXALOBE D3.0 x
25
Power management PCBA
Zinc plated steel TCB screw HEXALOBE D3.5 x
8
Adhesive stop, diameter 8 mm, thickness 2.2 mm (part of battery cover assembly)
Li- ion battery, assembly 4.8 Ah capacity
(with labels) Requires minimum revision power management software AL020002
Silicone stop (part of battery cover assembly)
Screw, HEXALOBE D3 x 8, galvanized steel
TCB (part of battery cover assembly)
Battery cover assembly
Battery assembly label (part of battery cover assembly)
Original remote alarm harness
Updated remote alarm harness
Remote alarm cable (RAII)
2-way solenoid valve, O
2
Square support bracket, O
2
valve
Turbine control PCBA
Power supply/turbine harness
Power management PCBA
CPU/Power supply harness
Mains warning label
CPU PCBA (includes LCD)
CPU/turbine harness (part of 10053276)
7-6 Service Manual
Item no.
Figure 7-2
-
-
-
-
Table 7-3. Electrical System Component Part Numbers (continued)
Part no.
10046894 (label kit)
Quantity
1
10025113
2980100
10022884
1
1
1
Description
Label O
2
/RA (Two-piece label; O
2
) (part of battery cover assembly) See
, Items
5 & 6
Double sided tape (not shown)
Wire ferrite (not shown)
Battery, 3V, 35 mAh
Parts List
Service Manual 7-7
Parts List
Figure 7-2. Major Electrical Assemblies
POWER MANAGEMENT PCBA
7-8
TURBINE CONTROL PCBA
C CPU PCBA
(Bottom view)
Service Manual
Service Manual
7.4
Pneumatic System Components
Table 7-4. Pneumatic System Component Part Numbers
Item no.
1
Part no.
2969600
2
3
4
5
6
7
8
12
13
14
15
16
9
10
11
2970400
2969600
2969600
2970300
2969500
2969500
2963800
3827900
2100500
2970800
10028763
2962799
2962600
10038115
2966800
Qty see
for specific lengths
1 see
for specific lengths see
for specific lengths
1 see
for specific lengths see
for specific lengths
2
Description
Silicone tubing, 1.6 mm ID x 4.8 mm OD
Tee connector, 2.4, plastic
Silicone tubing, 1.6 mm ID x 4.8 mm OD
Silicone tubing, 1.6 mm ID x 4.8 mm OD
Straight connector, 2.4 x 3.2 plastic
Silicone tubing diameter 3.2 mm ID x 6.4 mm
OD
Silicone tubing diameter 3.2 mm ID x 6.4 mm
OD
1
2
2
1
1
1
1 see
for specific length
Galvanized steel countersunk screw
HEXALOBE M3 x 8 (O
2
valve bracket to lower housing)
Square support bracket, O
2
valve
Tie-wrap (for O
2
sol valve outlet, turbine inlet
Galvanized steel TCB screw HEXALOBE D3 x 6
(bracket screws)
2-way solenoid valve (O
2
)
Oxygen inlet connector
O
2
female connector with valve
Turbine/inspiratory port connection tube
Polyamide tubing 6 mm x 4 mm (0.2)
Parts List
7-9
Parts List
Item no.
Figure 7-3
17-20
21-22
Table 7-4. Pneumatic System Component Part Numbers (continued)
Part no.
2969500
2969600
Qty see
for specific lengths see
for specific lengths
Description
Silicone tubing diameter 3.2 mm ID x 6.4 mm
OD
Silicone tubing, 1.6 mm ID x 4.8 mm OD
7-10 Service Manual
Figure 7-3. Pneumatic Components and Connection Diagram
Parts List
Service Manual 7-11
Parts List
7.5
Turbine Assembly Components
Table 7-5. Turbine Assembly Component Part Numbers
Item no.
1
2
3
4
Part no.
2391200
10028768
2877500
2971600
Qty
2
1
2
1
Description
Black foam - width 40 mm; thickness 10 mm
Turbine
Neoprene adhesive strip
Adjustable adhesive collar (toroid clamp)
7-12 Service Manual
Figure 7-4. Turbine Assembly Drawing
Parts List
Service Manual 7-13
Parts List
7.6
Label Kit
lists the contents of the complete label kit and the general locations of each label. The subsequent tables break the kit down by location and correspond to the illustrations on their respective facing pages.
Table 7-6. Label Kit, P/N 10046894
Description
Label, Air inlet
Label, Battery assembly
Label, ESD
Label O
2
/RA (remote alarm portion)
Label O
2
/RA(O
2
portion)
Label, Pneumatic cabling
Label, exhalation valve proximal pressure
Label, exhalation single use
Label, identification
Label, valve connection
Label, exhalation, single patient use
Label, polyester blank (serial number window)
Location
Lower housing, back of ventilator
Inside battery cover (see Item 15)
Inside ventilator, lower housing (see Item 35)
Lower housing, back of ventilator
Battery cover
Inside ventilator, upper housing (see Item 4)
Front of ventilator, on inspiratory block (See
Item 3)
On exhalation block (See
Bottom of ventilator (See Figure 7-6 Item 1)
Exhalation block (See Figure 7-7 Item 1)
Exhalation block (see Figure 7-7 Item 2)
Under product ID label
7-14 Service Manual
Figure 7-5. Front and Back Panel Labels
Parts List
Service Manual 7-15
Parts List
Item no.
--
1
2
5
6
3
4
Part no.
10046894
--
--
--
--
--
--
Table 7-7. Front and Back Labels
1
1
1
1
Qty
--
1
1
Description
Label kit
Label, Exhalation single patient use
Label, FiO
2
Label, exhalation valve/proximal pressure
Label, Air inlet
Label O
2
/RA (remote alarm portion)
Label O
2
/RA (O
2
portion)
7-16 Service Manual
Service Manual
Figure 7-6. Bottom Panel Label
Parts List
1
Item no.
Part no.
label kit
Table 7-8. Bottom Panel Label
Qty.
1
Description
Product identification label
7-17
Parts List
Figure 7-7. Left Side Panel Labels
Item no.
Item no.
1
2
Part no.
Part no.
--
--
Table 7-9. Left Side Panel Labels
Qty
Qty
1
1
Description
Description
Label, valve connection
Label, exhalation, single patient use
7-18 Service Manual
A Appendix A
A.1
Overview
This appendix contains diagrams for the Puritan Bennett™ 560 Ventilator referenced in
.
Figure A-1. Pneumatic Block Diagram
A-1
Appendix A
1
Turbine control PCBA
2
3
19
Inlet air filter
Turbine
4
5
17
Exhalation solenoid valve
Exhalation valve pressure sensor
6
7
15
Inspiratory block
Inspiratory flow sensor
8
9
Inspiratory bacteria filter
Not shown with humidifier, nebulizer, or additional water traps
Inspiratory tubing
10
11
Proximal pressure tube
12
Patient wye
13
Inspiratory pressure sensor
14
Proximal pressure sensor
Exhalation valve pilot tube
16
Buzzer PCBA
Exhalation flow sensor
18
Exhalation tubing
Water trap
20
Exhalation bacteria filter
21
Exhalation block
22
Exhalation valve
23
Display
24
CPU PCBA
25
Battery connection PCBA
26
Internal battery
27
Cooling fan
28
29
Power supply (located above power management PCBA)
Power switch
30
AC input
31
DC input
32
PC port
33
Type A USB ports (2)
34
35
SpO
2
port (not used)
Nurse call port
36
37
Low pressure O
2 inlet
O
2 solenoid valve
A-2 Service Manual
Figure A-2. Power Management PCBA Block Diagram
Appendix A
Service Manual A-3
Appendix A
Figure A-3. Exhalation Solenoid Valve During Inspiration and Exhalation
1
2
3
Turbine flow applied to port 1
To patient circuit exhalation valve and exhalation pressure transducer
Port 3 vents to atmosphere
A-4 Service Manual
Index
A
Accessories, list of . . . . . . . . . . . . . . . . . . . . . . . .2-2
Adhesive feet, removal and replacement 5-13
CPAP mode . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
High inspired tidal volume (High VTI) . .
High leak. . . . . . . . . . . . . . . . . . . . . . . . . 2-15
High respiratory rate (high Rtot) . . 2-15
Low inspired tidal volume (Low VTI). . . .
P SIMV mode . . . . . . . . . . . . . . . . . . . . . . . . 2-13
High exhaled tidal volume (High VTE) . .
High inspired tidal volume (High VTI) . .
High respiratory rate (High Rtot) . . 2-14
Low exhaled tidal volume (Low VTE) . . .
Low inspired tidal volume (Low VTI). . . .
Pressure A/C mode . . . . . . . . . . . . . . . . . . 2-11 high exhaled tidal volume high (VTE). . .
2-11 high inspired tidal volume (high VTI) . . .
high inspiratory time (high I time). 2-11
high inspired tidal volume (high VTI) . . .
high leak . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
minimum inspiratory time (low I time) .
V AC mode
high exhaled tidal volume high VTE) . . .
2-13 high peak inspiratory pressure (high
Volume A/C mode . . . . . . . . . . . . . . . . . . . 2-14
Alarms
AC POWER DISCONNECTION . . . . . . . . . 4-14
APNEA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
BATTERY FAULT 1 . . . . . . . . . . . . . . . . . . . . 4-15
BATTERY FAULT 2 . . . . . . . . . . . . . . . . . . . . 4-16
BATTERY FAULT 3 . . . . . . . . . . . . . . . . . . . . 4-16
BUZZER FAULT 1 . . . . . . . . . . . . . . . 4-16, 4-17
BUZZER FAULT 2 . . . . . . . . . . . . . . . . . . . . . 4-17
BUZZER FAULT 3 . . . . . . . . . . . . . . . . . . . . . 4-17
CALIBRATION FAIL . . . . . . . . . . . . . . . . . . . 4-18
CHECK BATTERY CHARGE . . . . . . . . . . . . 4-18
CHECK EXH VALVE . . . . . . . . . . . . . . . . . . . 4-19
CHECK EXH VALVE PRESSURE. . . . . . . . . 4-19
CHECK PROXIMAL LINE 1 . . . . . . . . . . . . . 4-20
CHECK SETTINGS. . . . . . . . . . . . . . . . . . . . . 4-20
CONNECT VALVE OR CHANGE PRESS . 4-21
COOLING FAN . . . . . . . . . . . . . . . . . . . . . . . 4-21
DEVICE FAULT 1. . . . . . . . . . . . . . . . . . . . . . 4-22
DEVICE FAULT 10 . . . . . . . . . . . . . . . . . . . . 4-23
DEVICE FAULT 11 . . . . . . . . . . . . . . . . . . . . 4-24
DEVICE FAULT 12 . . . . . . . . . . . . . . . . . . . . 4-24
DEVICE FAULT 13 . . . . . . . . . . . . . . . . . . . . 4-24
DEVICE FAULT 2. . . . . . . . . . . . . . . . . . . . . . 4-22
DEVICE FAULT 3. . . . . . . . . . . . . . . . . . . . . . 4-22
DEVICE FAULT 4. . . . . . . . . . . . . . . . . . . . . . 4-22
DEVICE FAULT 5. . . . . . . . . . . . . . . . . . . . . . 4-23
DEVICE FAULT 6. . . . . . . . . . . . . . . . . . . . . . 4-23
DEVICE FAULT 7. . . . . . . . . . . . . . . . . . . . . . 4-23
DEVICE FAULT 8. . . . . . . . . . . . . . . . . . . . . . 4-23
DEVICE FAULT 9. . . . . . . . . . . . . . . . . . . . . . 4-23
E SENS FAULT. . . . . . . . . . . . . . . . . . . . . . . . 4-24
EMPTY BATTERY . . . . . . . . . . . . . . . . . . . . . 4-24
INCOMPATIBLE SOFTWARE VERSION . 4-26
INSP FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
INTENTIONAL VENT STOP . . . . . . . . . . . . 4-26
KEYBOARD FAULT . . . . . . . . . . . . . . . . . . . 4-26
Index-1
Index-2
LOW BATTERY. . . . . . . . . . . . . . . . . . . . . . . 4-27
OCCLUSION . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Power supply loss (no message) . . . . . 4-29
REMOTE ALARM FAULT . . . . . . . . . . . . . . 4-29
Atmospheric pressure sensor
See Barometric pressure sensor
B
Barometric pressure sensor . . . . . . . . . . . . . . .3-7
Battery
description. . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
removal and replacement . . . . 5-13 to 5-14
Battery (3V), removal and replacement. . 5-28
Battery Connection PCBA
description. . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
removal and replacement . . . . . . . . . . . 5-47
Buzzer PCBA
description. . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
removal and replacement . . . . . . . . . . . 5-33
C
Calibration
exhalation flow sensor. . . . . . . . . . . . . . . 6-19
inspiratory flow sensor . . . . . . . 6-17 to 6-19
Patient (proximal) pressure sensor . . . 6-15
Captive knob, removal and replacement . . . . .
Cleaning solutions . . . . . . . . . . . . . . . . . . . . . 2-25
Consumable parts, list of . . . . . . . . . . . . . . . . .2-4
Control panel
CPU PCBA
description. . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
removal and replacement . . . . . . . . . . . 5-22
D
Default settings . . . . . . . . . . . . . . . . . . 2-17, 6-46
Disinfectants . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Disposable parts
E
Electromagnetic emissions and immunity, manufacturer’s declaration. . 2-19 to 2-24
Exhalation block
description. . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Exhalation block housing (conical fitting), removal and replacement . . 5-35 to 5-36
Exhalation flow sensor
calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Exhalation solenoid valve
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
removal and replacement . . . . 5-20 to 5-22
Exhalation solenoid valve diagram . . . . . . . A-4
Exhalation valve (internal valve) pressure sensor
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
External inspection . . . . . . . . . . . . . . . . . . . . . . 5-5
F
Fan
description . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
removal and replacement . . . . . . . . . . . . 5-46
Faults, technical
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
troubleshooting table. . . . . . . . . . . . . . . . 4-10
H
Handle, removal and replacement. . . . . . . 5-19
High priority alarm indicator . . . . . . . . . . . . 2-33
Hours, how to reset . . . . . . . . . . . . . . . . . . . . . 2-45
Housing, disassembly and assembly . . . . . . . . .
I
Indicators
AC power. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
DC power. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
Inlet air filter
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
removal and replacement . . . . . . . . . . . . . 5-9
Inlet silencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Inspiratory block
cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
removal and replacement . . . . 5-30 to 5-31
Inspiratory flow sensor
calibration . . . . . . . . . . . . . . . . . . . 6-17 to 6-19
description . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Index
Inspiratory pressure sensor . . . . . . . . . . . . . . .3-7
Intended use, product description . . . . . . . .2-2
Internal battery
description. . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
removal and replacement . . . . 5-13 to 5-14
K
Keypad
controls and indicators . . . . . . . 2-31 to 2-33
removal and replacement . . . . 5-19 to 5-20
Keys
alarm control. . . . . . . . . . . . . . . . . . . . . . . . 2-33
Ventilation ON/OFF . . . . . . . . . . . . . . . . . 2-32
L
Labeling checks . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
Labels
front panel . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
parts lists. . . . . . . . . . . . . . . . . . . . . 7-14 to 7-17
rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35
side panel. . . . . . . . . . . . . . . . . . . . . . . . . . . 2-39
LCD panel
description. . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
removal and replacement . . . . 5-26 to 5-28
M
Maintenance mode, how to access . . . . . . . . . .
Manufacturer’s declaration . . . . . . 2-19 to 2-24
Medium priority alarm . . . . . . . . . . . . . . . . . 2-33
Apnea index . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Exhaled tidal volume (VTE). . . . . . . . . . . 2-15
Expiratory time (Etime) . . . . . . . . . . . . . . 2-15
I:E ratio (I:E) . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
I:T ratio (I/T) . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Inspiratory minute volume (M. Vol) . . 2-15
Inspiratory tidal volume (VTI) . . . . . . . . 2-15
Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Positive end expiratory pressure (PEEP) . . .
Spontaneous cycles . . . . . . . . . . . . . . . . . . 2-16
Total respiratory rate (R tot) . . . . . . . . . . 2-16
N
Nurse call port
O
ON/OFF switch . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
Outlet silencer . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Oxygen inlet valve . . . . . . . . . . . . . . . . . . . . . . . 3-4
Oxygen solenoid valve
functional test . . . . 6-24 to 6-26, ?? to 6-26
removal and replacement . . . . 5-37 to 5-39
P
Patient (proximal) pressure sensor, calibration
Patient circuit. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Patient hours, how to reset. . . . . . . . . . . . . . 2-45
Performance verification tests
AC/DC/battery power switching test. . 6-35
electrical safety test . . . . . . . . . . . . . . . . . . . 6-6
exhalation flow sensor calibration . . . . 6-19
flow sensor capacity test . . . . . . . . . . . . . 6-21
inspiratory flow sensor calibration . . . . . . . .
internal 24 V verification . . . . . . . . . . . . . . 5-7
internal battery information verification . .
internal temperature check . . . . . . . . . . . 5-7
oxygen solenoid valve functional test . . . .
patient disconnect alarm test . . . . . . . . 6-45
patient pressure sensors calibration . . 6-15
power off while ventilating. . . . . . . . . . . 6-45
remote alarm test . . . . . . . . . . . . . . . . . . . . 6-40
set default parameters/time & date verification . . . . . . . . . . . . . . . . . . . . . . 6-46
turbine performance test. . . . . . . . . . . . . 6-22
watchdog voltage verification . . . . . . . . . 5-7
Power Management PCBA . . . . . . . . . . . . . . 3-10
Power management PCBA block diagram A-3
Index-3
Index-4
Preventive maintenance
parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
schedule . . . . . . . . . . . . . . . . . . . . . 2-28 to 2-29
Proximal pressure sensor . . . . . . . . . . . . . . . . .3-7
R
Remote alarm port . . . . . . . . . . . . . . . . . . . . . . .2-7
S
Serial number, how to find . . . . . . . . . . . . . 2-43
Service philosophy . . . . . . . . . . . . . . . . . . . . . 2-47
Settings, default. . . . . . . . . . . . . . . . . . 2-17, 6-46
Setup screen, how to access . . . . . 2-44 to 2-45
Software lock . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40
Software version, how to find. . . . . . . . . . . 2-43
Specifications
performance . . . . . . . . . . . . . . . . . . 2-8 to 2-16
Symbols, onscreen and device label . . . . . . . . .
T
Technical faults
description. . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
how to troubleshoot . . . . . . . . . 4-10 to 4-12
Temperature sensor
turbine. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
ventilator compartment . . . . . . . . . . . . . 3-15
Test Preconditions . . . . . . . . . . . . . . . . . . . . . . .6-1
Tools, required for service . . . . . . . 2-24 to 2-25
Troubleshooting
miscellaneous symptoms . . . . . 4-31 to 4-32
technical faults . . . . . . . . . . . . . . . 4-10 to 4-12
ventilation alarms . . . . . . . . . . . . 4-12 to 4-18
Turbine assembly
description. . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
removal and replacement . . . . 5-39 to 5-41
Turbine Control PCBA
description . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
removal and replacement . . . . . . . . . . . . 5-41
V
Ventilator compartment temperature sensor
Ventilator pneumatic block diagram. . . . . . A-1
Ventilator settings
I to E ratio. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
I/T ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
inspiratory sensitivity (Insp Sens) . . . . . . 2-9
Vt target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Ventilator settings, backup rate . . . . . . . . . . 2-9
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1, 5-1
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Electrical Power . . . . . . . . . . . . . . . . . . . . . . . 1-6
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Fire Hazard . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
General Repair Safety . . . . . . . . . . . . . . . . . 1-2
Infection Reduction . . . . . . . . . . . . . . . . . . . 1-2
Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Performance Verification . . . . .2-2, 6-1, 6-6
Prior to Equipment Usage . . . . . . . . . . . . . 1-3
USB Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Use of Equipment. . . . . . . . . . . . . . . . . . . . . 1-4