Soil Moisture Sensor
Operator’s
Manual
For Models EC-20, EC-10, and EC-5
Version 5
Decagon Devices, Inc.
2365 NE Hopkins Court
Pullman WA 99163 USA
“ECH
2
Trademarks:
O” is a registered trademark of Decagon, Devices, Inc All rights reserved..
©2006 Decagon Devices, Inc.
All rights reserved.
10371-05
ECH
2
O Probe User’s Manual
Contents
Contents
1. Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . 1
Welcome . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . 1
Contact Information . . . . . . . . . . . . . . . . . . . 2
Warranty Information . . . . . . . . . . . . . . . . . 2
Seller’s Liability . . . . . . . . . . . . . . . . . . . . . . . 2
2. About the ECH2O Probes
. . . . . . . . . . . 4
The EC-10 and EC-20 . . . . . . . . . . . . . . . . . . 4
The EC-5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Probe Features . . . . . . . . . . . . . . . . . . . . . . . 5
Wiring Diagrams . . . . . . . . . . . . . . . . . . . . . . 5
3.5mm plug wiring . . . . . . . . . . . . . . . . . . . 5
Wiring to Non-Decagon Dataloggers . . . 6
Extension cables . . . . . . . . . . . . . . . . . . . . . . 7
3. Installing the Probes
. . . . . . . . . . . . . . . . 8
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Orientation . . . . . . . . . . . . . . . . . . . . . . . . 10
Removing the Probe . . . . . . . . . . . . . . . . . 10
4. Collecting Data
. . . . . . . . . . . . . . . . . . . . . 11
Datalogger Requirements . . . . . . . . . . . . . . 11
Connecting to a Datalogger . . . . . . . . . . . 12
Sample Program . . . . . . . . . . . . . . . . . . . . . 12
SCWin (Short Cut) Directions . . . . . . . . . . 14
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Probe calibration values . . . . . . . . . . . . . 16 i
ECH
2
O Probe User’s Manual
Contents
Troubleshooting . . . . . . . . . . . . . . . . . . . . . 19
Declaration of Conformity
. . . . . . . . . . .20
Index
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 ii
ECH
2
O Probe User’s Manual
1. Introduction
1. Introduction
Welcome
Welcome to the ECH
2
O Dielectric Aquameter sensors for measuring soil water content. These innovative sensors will enable you to monitor soil moisture accurately and affordably.
Specifications
Measurement Time: 10ms (milliseconds)
Accuracy:
EC-10 and EC-20:
± .04 m
3
/m
3
(±4%) < 0.5 dS/m
With soil-specific calibration: ±.02 m
3
EC-5:
/m
3
(±2%) at least 0.003 m
3
/m
3
all soils, up to 8 dS/m
With soil-specific calibration: ±.02 m
3
/m
3
(±2%)
Resolution:
EC-10 and EC-20: 0.002m
3
/m
3
EC-5: 0.001 m
3 growing media
/m
3
VWC in mineral soils, 0.25% in
Power:
Requirements:
EC-10 and EC-20: 2.5VDC @ 2mA to 5VDC @ 7mA
EC-5: 3VDC @ 10mA
Output: 10-40% of excitation voltage (250-1000mV at
2500mV excitation)
Operating Environment:
EC-10 and EC-20: 0 to 50°C
1
ECH
2
2
O Probe User’s Manual
1. Introduction
EC-5: -40 to +60 °C
Range of Measurement:
EC-10 and EC-20: 0 to saturated VWC
EC-5: 0 to 100% VWC
Probe dimensions:
EC-20: 25.4cm x 3.17cm x .15cm
EC-10: 14.5cm x 3.17cm x .15cm
EC-5: 8.9cm x 1.8cm x 0.7cm
Cable length: 5m standard, extension cable available
Connector types: 3.5 mm plug or stripped and tinned
lead wires
Datalogger Compatibility (not exclusive):
Decagon: Em5, Em50 (and radio-enabled versions)
Campbell Scientific: CR10X, 21X, 23X, CR1000, CR3000, etc.
Contact Information
To contact Decagon for customer support or questions:
E-mail:
Fax:
(509) 332-5158
Telephone:
(US and Canada only) 1-800-755-2751, or 509-332-2756.
Warranty Information
The ECH
2
O probes have a 30-day satisfaction guarantee and a one-year warranty.
Seller’s Liability
Seller warrants new equipment of its own manufacture against defective workmanship and materials for a period
2
ECH
2
O Probe User’s Manual
1. Introduction
of one year from date of receipt of equipment (the results of ordinary wear and tear, neglect, misuse, accident and excessive deterioration due to corrosion from any cause are not to be considered a defect); but Seller’s liability for defective parts shall in no event exceed the furnishing of replacement parts F.O.B. the factory where originally manufactured. Material and equipment covered hereby which is not manufactured by Seller shall be covered only by the warranty of its manufacturer. Seller shall not be liable to Buyer for loss, damage or injuries to persons
(including death), or to property or things of whatsoever kind (including, but not without limitation, loss of anticipated profits), occasioned by or arising out of the installation, operation, use, misuse, nonuse, repair, or replacement of said material and equipment, or out of the use of any method or process for which the same may be employed. The use of this equipment constitutes Buyer’s acceptance of the terms set forth in this warranty. There are no understandings, representations, or warranties of any kind, express, implied, statutory or otherwise (including, but without limitation, the implied warranties of merchantability and fitness for a particular purpose), not expressly set forth herein.
3
ECH
2
2
O Probe User’s Manual
2. About the ECH2O Probes
2. About the ECH
2
O Probes
The EC-10 and EC-20
The ECH
2
O probe measures the dielectric constant of the soil in order to find its volumetric water content. Since the dielectric constant of water is much higher than that of air or soil minerals, the dielectric constant of the soil is a sensitive measure of water content. The ECH
2
O probe has a very low power requirement and high resolution. This gives you the ability to make as many measurements as you want (even hourly) over a long period of time (like a growing season, for example), with minimal battery usage.
The EC-5
The EC-5 varies from its EC-10 and EC-20 cousins.
Although the principles of measurement are the same, its two-prong design and higher measurement frequency allows the EC-5 to measure VWC from 0 to 100%, and allows accurate measurement of all soil types and a much wider range of salinities.
4
Probe Features
ECH
2
O Probe User’s Manual
2. About the ECH2O Probes
Circuitry
Sensor
Fig.1:
ECH
2
O
probe diagram Fig. 2: EC-5 probe diagram
Wiring Diagrams
3.5mm plug wiring
The ECH
2
O probe comes with a 3.5mm “stereo plug” connector. This allows for rapid connection directly to
Decagon’s Em50 and Em5 logger to the ECH
2
O Check.
5
ECH
2
2
O Probe User’s Manual
2. About the ECH2O Probes
Below is a diagram showing the wiring configuration for this connector.
Analog out
Ground Excitation
Fig. 2: ECH
2
O plug wiring configuration
Wiring to Non-Decagon Dataloggers
Models with stripped and tinned leads are pre-configured for connecting to non-Decagon dataloggers. Simply wire the lead into the datalogger as described in “Connecting to a Datalogger” in Chapter 4.
If your model uses the standard 3.5mm plug, you have two choices when attaching ECH
2
O probes to non-Decagon dataloggers. First, you can clip off the plug on the probe cable, strip and tin the wires, and wire it directly into the datalogger. This has the advantage of creating a direct connection with no chance of the probe becoming un-plugged; however, it then cannot be used in the future with a Decagon Em50 or Em5 logger. The other choice is to obtain an ECH
2
O adapter cable from Decagon. The 3wire probe adapter cable has a connector for the ECH
2
O probe jack on one end, and three wires on the other end for connection to a datalogger (this type of wire is often referred to as a “pigtail” adapter). Both the probe wire and adapter cable wire have the same wire output (shown in
6
ECH
2
O Probe User’s Manual
2. About the ECH2O Probes
Fig. 3); the white wire is excitation, red is output, and the bare wire is ground.
Analog out (Red)
Ground (Bare)
Sensor cable
Excitation (White)
Fig. 3: 3-wire cable wiring configuration
Extension cables
Decagon supplies 50-foot (15.25m) and 10-foot (3m) extension cables for use with the ECH
2
O probe. You can safely connect up to 4 of the 50-foot cables without signal attenuation. For most applications, you will want to seal the connections from the elements to maintain a good connection and to prevent corrosion.
7
ECH
2
2
O Probe User’s Manual
3. Installing the Probes
3. Installing the Probes
When selecting a site for installation, it is important to remember that the soil adjacent to the probe surface has the strongest influence on the probe reading and that the probe measures the volumetric water content. Therefore any air gaps or excessive soil compaction around the probe can profoundly influence the readings. Also, do not install the probes adjacent to large metal objects such as metal poles or stakes. This can attenuate the probe’s electromagnetic field and adversely affect output readings. Because the EC-5 has gaps between its prongs, it is also important to consider the size of the media you are inserting the probe into. It is possible to get sticks, bark, roots or other material stuck between the probe prongs, which will adversely affect readings. Finally, be careful when inserting the probes into dense soil, as the prongs will break if excessive sideways force is used when pushing them in.
Procedure
When installing the ECH
2
O probe, it is best to maximize contact between the probe and the soil. There are two methods to accomplish this.
For the EC-10/EC-20:
1. (Recommended) Use Decagon’s Probe Installation
Kit to install the probe. This kit has a custom-shaped blade to make the insertion in the soil, then another tool to place the probe into the insertion. For deeper installations, use an augur to reach the desired depth,
8
ECH
2
O Probe User’s Manual
3. Installing the Probes
then use the Installation kit with extension rods to install the probe.
2. Use a thin implement like a trenching shovel, gardening spade, or flat bar to make a pilot hole in the soil.
Then insert the probe into the hole, making sure the entire length of the probe is covered. Finally, insert the shovel again into the soil a few inches away from the probe, and gently force soil toward the probe to provide good contact between the probe and the soil. For deeper installation, excavate down to the level you wish to measure, then install the probe as described.
For the EC-5:
1. Insert the probes into the soil, making sure that the prongs are buried completely up to the black overmolding, as shown below.
The tip of each prong has been sharpened to make it easier to push the probe in - be careful with the sharp tips!
The probe may be difficult to insert into extremely
9
ECH
2
2
O Probe User’s Manual
3. Installing the Probes
compact or dry soil. If you have difficulty inserting the probe, try loosening the soil somewhat or wetting the soil.
Never pound it in!
Orientation
The probe can be oriented in any direction. However, orienting the flat side perpendicular to the surface of the soil will minimize effects on downward water movement.
Removing the Probe
When removing the probe from the soil,
do not pull it out of the soil by the cable!
Doing so may break internal connections and make the probe unusable.
10
ECH
2
O Probe User’s Manual
4. Collecting Data
4. Collecting Data
Datalogger Requirements
The ECH
2
O sensors are designed to work most efficiently with Decagon’s 5-channel Em5, Em50 or our ECH
2
O
Check handheld readout. They can, however, be adapted for use with other dataloggers, such as those from Campbell Scientific, Inc., for example. The ECH
2
O requires an excitation voltage in the range of 2 to 5 volts. It produces an output voltage that depends on the dielectric constant of the medium surrounding the probe, and ranges between 10 and 50% of the excitation voltage. Any datalogger which can produce a 2.5 to 5V excitation with approximately 10 millisecond duration and read a voltlevel signal with 12-bit or better resolution should be compatible with the ECH
2
O probes. For the EC-10 and EC-
20 probes, the current requirement at 2.5V is around
2mA, and at 5V it is 7-8mA. For the EC-5, it is 10mA at
2.5V.
NOTE: ECH
2
O probes are intended only for use with dataloggers and readout devices which can provide short excitation pulses, leaving the probes turned off most of the time. Continuous excitation not only wastes battery power, but may, under certain circumstances, cause the probe to exceed government specified limits on electromagnetic emissions.
11
ECH
2
2
O Probe User’s Manual
4. Collecting Data
NOTE: All Decagon readout devices use either a2.5V or
3V excitation, so the remainder of the discussion assumes this as the excitation. If other voltages are used, adjustments to the calibration equations provided will be needed.
Connecting to a Datalogger
Connect the wires to the datalogger as shown, with the supply wire connected to the excitation, the analog out wire to an analog input, and the bare ground wire to ground:
Supply
Analog out
Ground
Exc.
H
Analog
In
L
Datalogger
G
Fig. 5: Datalogger configuration
Sample Program
The following program is an example that can be used with Campbell Scientific’s CR10X datalogger and our
ECH
2
O probes at a 2500mv excitation:
Important Note: Multiplier and offset in instruction 2 (P4) convert mV reading to % water content. Instruction 7 (P37) converts %
volumetric water content to m
3
m
-3
. Also, “--” (such as the one in
12
ECH
2
O Probe User’s Manual
4. Collecting Data
Instruction 4 (P89), line 1) increment the input location for the next time through the loop. A “--” is applied to the input location by
clicking on the input location # and typing the letter “c”.
;{CR10X}
; Example ECH2O Datalogger Program for CR10X
; Wiring:
; White: Excitation Channel 1
; Red: Input Single Ended Channel 1
; Black: Ground
*Table 1 Program
01: 1 Execution Interval (seconds)
; Factory calibration equations for ECH2O
; probes convert mV output of ECH2O to
; volumetric water content (VWC, m3/m3)
; EC-20: VWC = 0.00069 * mV - 0.29
; EC-10: VWC = 0.00093 * mV - 0.376
; EC-5: VWC = 0.00119 * mV - 0.400
; The multiplier and offset in this
; example are for the EC-5
1: Excite-Delay (SE) (P4)
1: 1 Reps
2: 5 2500 mV Slow Range
3: 1 SE Channel
4: 1 Excite all reps w/Exchan 1
5: 1 Delay (0.01 sec units)
6: 2500 mV Excitation
7: 1 Loc [ Probe_VWC ]
8: .00119 Multiplier
9: -.4 Offset
*Table 2 Program
02: 0.0000 Execution Interval (seconds)
*Table 3 Subroutines
End Program
1 [ Probe_VWC ] -W-- 0 1 ----- ------ ---
2 [ _________ ] ---- 0 0 ----- ------ ---
3 [ _________ ] ---- 0 0 ----- ------ ---
4 [ _________ ] ---- 0 0 ----- ------ ---
5 [ _________ ] ---- 0 0 ----- ------ ---
6 [ _________ ] ---- 0 0 ----- ------ ---
7 [ _________ ] ---- 0 0 ----- ------ ---
8 [ _________ ] ---- 0 0 ----- ------ ---
9 [ _________ ] ---- 0 0 ----- ------ ---
10 [ _________ ] ---- 0 0 ----- ------ ---
11 [ _________ ] ---- 0 0 ----- ------ ---
13
ECH
2
2
O Probe User’s Manual
4. Collecting Data
12 [ _________ ] ---- 0 0 ----- ------ ---
13 [ _________ ] ---- 0 0 ----- ------ ---
14 [ _________ ] ---- 0 0 ----- ------ ---
15 [ _________ ] ---- 0 0 ----- ------ ---
16 [ _________ ] ---- 0 0 ----- ------ ---
17 [ _________ ] ---- 0 0 ----- ------ ---
18 [ _________ ] ---- 0 0 ----- ------ ---
19 [ _________ ] ---- 0 0 ----- ------ ---
20 [ _________ ] ---- 0 0 ----- ------ ---
21 [ _________ ] ---- 0 0 ----- ------ ---
22 [ _________ ] ---- 0 0 ----- ------ ---
23 [ _________ ] ---- 0 0 ----- ------ ---
24 [ _________ ] ---- 0 0 ----- ------ ---
25 [ _________ ] ---- 0 0 ----- ------ ---
26 [ _________ ] ---- 0 0 ----- ------ ---
27 [ _________ ] ---- 0 0 ----- ------ ---
28 [ _________ ] ---- 0 0 ----- ------ ---
SCWin (Short Cut) Directions
Following are instructions for using our SCWin (Short
Cut) program to read EC-5, EC-10 and EC-20 ECH
2 probes.
O
1. Download EchoCSI.zip from http://www.decagon.com/appnotes/EchoCSIappnote.pdf.
2. Unzip the folder EchoCSI.zip.
3. Locate the file containing SCWin.exe. It should be in
C:\Program Files\Campbellsci\SCWin. Place the following files from the unzipped EchoCSI.zip folder into the folder with SCWin.exe:
AM1632Z.MUX
AM416Z.MUX
EC10.SCS
EC101632.SCS
EC10416.SCS
EC20.SCS
EC201632.SCS
EC20416.SCS
EC5.SCS
EC5632.SCS
14
ECH
2
O Probe User’s Manual
4. Collecting Data
EC5416.SCS
SCWIN-DECAGON.CNT
SCWIN-DECAGON.HLP
Note: If you are not able to find this directory path,
search for the folder that contains SCWIN.exe and place the files into that folder.
4. Open up SCWin.exe (Short Cut). If you are using a
V.3 copy of LoggerNet, there is a tab for SCWin
(Short Cut) on the tool bar.
5. Select “New” to start a new program to read the
ECH
2
O probes.
a. Select the datalogger you will be using to read the probes.
b. Select the measurement interval (a shorter measurement interval, i.e. 1 sec., is sometimes desirable when testing the probes).
6. Click on Sensors (this should open a new page with a file tree on it).
7. Under the “Sensors” file tree, double-click on “Meteorological” and then select “Soil Moisture.”
8. ECH
2
O Probe EC-10 and ECH
2
O Probe EC-20 should appear on the tree along with other soil moisture probes.
a. If they don’t appear, check to make sure you have pasted the files above into the correct location.
Calibration
The ECH
2
O probes come pre-calibrated for most soil types. However, this general calibration may not be appli-
15
ECH
2
2
O Probe User’s Manual
4. Collecting Data
cable for all probes and all soil types. For example, the standard calibrations for the EC-20 and EC-10 probes do not perform in soils with high sand or salt content. Therefore, for added accuracy we encourage our customers to perform soil-specific calibrations.
The calibration equation that you will use depends on where you will be using it. If you will be using it with probes connected to a non-Decagon datalogger OR using the "SEN" value given from the ECH
2
O Check, you will need to use the 2500 mV calibration. If you use any Decagon software (DataTrac, ECH
2 calibration menu in the ECH
2
O Utility, etc.) or the user
O Check, you will need to use the RAW calibration. The difference between the two is the slope constant. To increase the resolution of the sensor output, Decagon uses all available increments of the 12-bit number (value of 4096) where the measurement is stored. Thus, the output of the sensors read by the
ECH
2
O Check and Decagon loggers must be multiplied by 0.61 AND the 2500 mV slope to give the right value.
Probe calibration values
Following is a list of the both the millivolt and RAW calibration values for the ECH
2
O probes, where θ is the volumetric water content, mV is the millivolt output of the probe, and where x is the RAW probe output.
EC-20
2500mV: θ (m
3
/m
3
) = 0.000695mV - 0.29
RAW: θ (m
3
/m
3
) =0.000424 x - 0.29
EC-10:
2500mV: θ (m
3
/m
3
) = 0.000936mv - 0.376
16
ECH
2
O Probe User’s Manual
4. Collecting Data
RAW: θ (m
3
/m
3
) = 0.000571 x - 0.376
EC-5:
The EC-5 is much less sensitive to variation in texture and electrical conductivity because it runs at a much higher measurement frequency. Therefore, its general calibration equation should apply for all mineral soils up to 8 dS/m.
Its calibration equations are shown below for mineral soil, potting soil, and rockwool growing media:
Mineral Soils
According to our tests, a single calibration equation will generally suffice for all mineral soil types with electrical conductivities from 0.1 dS/m to 10 dS/m. Volumetric water content (q) is given by q =
8.5 * 10
-4
* Raw - 0.48 (1) where Raw is the output from the Decagon datalogger. If you are using a non-Decagon datalogger, VWC is given by q =
11.9 * 10
-4
* mV - 0.401
(2) where mV is the output of the probe when excited at 2500 mV. Please note that the equation will reach a maximum at ~60% volumetric water content (VWC) in pure water.
To display data on a scale from 0 to 100%, VWC should be modeled with a quadratic equation (which would result in a 100% VWC in water), but a linear equation fits the mineral soil VWC range as well as the quadratic, and linear equations are easier to deal with, especially since mineral soil typically saturates at ~40 - 50% VWC.
17
ECH
2
2
O Probe User’s Manual
4. Collecting Data
Potting soil
The following equations can be used to convert EC-5 output to water content in potting soil. We tested several types of potting soil (Sunshine mix, Miracle Grow Potting
Mix, and Custom Nursery soil) at several salinities and found that VWC is given by q =
7.2 * 10
-4
* Raw - 0.393
(3) for a Decagon datalogger or q =
10.3 * 10
-4
* mV - 0.334
(4) for a datalogger with 2500mV excitation.
Rockwool
The EC-5 was calibrated in Grodan Master rockwool with solution electrical conductivities of 0.2, 1.0, 1.5, 2.0, and
4.5 dS/m. Volumetric water content can be calculated using q =
6.28 * 10
-7
* Raw
2
+ 1.37 x 10
-4
* Raw - 0.183
(5) for a Decagon datalogger or q =
2.63 * 10
-6
* mV
2
+ 5.07 x 10
-4
* mV - 0.0394
(6) for a datalogger with 2500 mV excitation.
NOTE: These calibration constants only apply to
2500mV excitations; use of these numbers with any other excitation voltage will result in erroneous readings!
18
ECH
2
O Probe User’s Manual
4. Collecting Data
Troubleshooting
If you encounter problems with the ECH
2
O probes, they most likely will manifest themselves in the form of incorrect or erroneous readings. Before contacting Decagon about the sensor, do the following:
• Check to make sure the connections to the datalogger are both
correct
and
secure
.
• Ensure that your datalogger’s batteries are not dead or weakened.
If you encounter problems that aren’t due to the datalogger, please contact Decagon at (509) 332-2756 and at
[email protected].
19
ECH
2
2
O Probe User’s Manual
Declaration of Conformity
Declaration of Conformity
Application of Council Directive:
Standards to which conformity is declared:
Manufacturer’s Name:
89/336/EE6
EN61326 : 1998
EN51022 : 1998
Decagon Devices, Inc.
2365 NE Hopkins Court
Pullman, WA 99163 USA
Type of Equipment:
ECH
2
O dielectric soil moisture probe.
Model Number:
Year of First Manufacture:
EC-10, EC-20, EC-5
2001
This is to certify that the ECH
2
O dielectric soil moisture probe, manufactured by Decagon Devices, Inc., a corporation based in Pullman, Washington, USA meets or exceeds the standards for CE compliance as per the Council Directives noted above. All instruments are built at the factory at Decagon and pertinent testing documentation is freely available for verification.
20
ECH
2
O Probe User’s Manual
Index
Index
A
accuracy
1
C
calibration for EC-10 16 for EC-20 16 for EC-5 17 for mineral soil 17 for potting soil 18 for rockwool 18 contact information
2
D
datalogger
2 connecting to 12 requirements 11 sample program 12 declaration of conformity
20
E
EC-10
EC-20
EC-5
4
4
4 calibration 17 email
2 e-mail address
2 extension cables
7
21
ECH
Index
2
2
O Probe User’s Manual
I
installation tips 8
8 procedure 8
M
measurement range mineral soil calibration 17
2
O
orientation of probe
10
P
plug wiring configuration 5 potting soil power
18 requirements 1 probe calibration 15 installation instructions 8 orientation 10 specifications 1 probe orientation program
12
10
R
range
2 removing the probe rockwool
18
10
22
S
seller’s liability
Short Cut directions specifications
1
2
14
T
troubleshooting
19
W
warranty
2 wiring diagram
5 wiring diagrams
5
ECH
2
O Probe User’s Manual
Index
23
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 Welcome
- 5 Specifications
- 6 Contact Information
- 6 Warranty Information
- 6 Seller’s Liability
- 8 The EC-10 and EC
- 8 The EC
- 9 Probe Features
- 9 Wiring Diagrams
- 11 Extension cables
- 12 Procedure
- 15 Datalogger Requirements
- 16 Connecting to a Datalogger
- 16 Sample Program
- 18 SCWin (Short Cut) Directions
- 19 Calibration