RS232 Interface Kit
INSTRUCTION MANUAL
This Interface Kit is for use only with the balance it was intended for.
This accessory has been tested and complies with the approvals listed in the Instruction Manual of the balance.
1. INTERFACE INSTALLATION
1
.
2
.
3
.
4
.
Note: The interface module must be slid fully forward until it snaps into place.
2. RS232 CONNECTION
The RS232 Interface Kit cable terminates with a 9 pin Sub-D female connector.
The active pins on the connector end are:
Pin 2 = TXD
Pin 3 = RXD
Pin 4 = Connected to pin 6 for external hardware handshake
Pin 5 = Ground
Pin 6 = Connected to pin 4 for external hardware handshake
Pin 7 = CTS
Pin 8 = RTS
The Interface cable allows direct connection to a computer or compatible printer.
Some applications may require a 9 to 25 pin adapter which may be purchased from your local dealer.
2
RS232 Input
The balance will respond to various commands sent via the interface adapter.
The balance will respond only to the following commands when terminated with a CR or CRLF.
? Prints current weighing unit
0A Turns Auto-Print off
(n)A
C
L
0M
(n)M
T
V
P
Auto-Print on 1 to 3600 second intervals (n = 1 to 3600)
Initiates Span calibration
Initiates Linearity calibration
Switch to gram weighing unit
Switch to another weighing unit (n = sequence number of unit)*
Same as pressing ON-ZERO
Print software version
Same function as pressing the PRINT key
LE
0S
Prints last error code (e.g. Err 2)
Print unstable data
1S Print stable data only
*Some balance models have a limited number of available weighing units. Refer to the user manual of the balance. Attempting to select a weighing unit that is disabled or unavailable may return an Error message.
6
3. BALANCE SETUP
The balance’s interface is preset to communicate using the following settings:
2400 baud, 7 bit, no parity, no handshake
If different settings are desired, it will be necessary to change the settings in the balance, computer, or printer to match.
Menu Structure
Upon installation and power up the balance will recognize the RS232 interface and add the PRINT and RS232 headings to the menu structure. Configure the balance to operate as desired. Refer to the User Manual of the balance for assistance in using the menus.
The two supplemental menus are shown below. Some balances may have additional menu items. Refer to the User Manual of the balance for further information. Default menu settings are shown in bold.
PRINT RS232
- Cont
- On.stbl
- 15sec
- 30sec
- 60sec
- Yes
- No
- 600
- 1200
- 4800
- 9600
- 19200
- 7-even
- 7-odd
Handshake
End
- No
3
Setup Definitions
PRINT menu:
Stable - On .................................Only stable data is output
Auto Print - Continuous ..............Data is output continuously
Auto Print - On Stable ................Data is automatically output upon stability
Auto Print - (xx) sec....................Data is output every (xx) seconds
Auto Print - Off............................Data is only output when PRINT button on balance is pressed
RS232 menu:
USB - Off ....................................Turns interface off
Baud, Parity, Handshake ...........Set parameters to match printer or computer
4. OPERATION
RS232 Output
With the interface installed the balance will operate in one of the following three ways:
• RS232 = on, Auto Print = off, Stable = on or off
Pressing PRINT will send the display data. If Stable is set to On, the balance will wait for a stable reading before sending data.
• RS232 = on, Auto Print = on, Stable = on or off
The balance will automatically send data based on the settings in the Menu.
If Stable is set to On, only stable will be sent.
=
The interface is turned off. During battery operation, turning the RS232
Interface off saves significant power.
Auto-Print (A-Print) Operation
Once Auto-Print is enabled the balance will continue to send data as required.
To temporarily stop Auto-Printing, press the PRINT key. A second press will resume Auto-Printing.
4
Data sent from the balance is in standard ASCII format terminated with a CRLF
(carriage return-line feed). The Output string is as follows:
[weight] 11 characters (right justified, includes – sign if needed)
[unit]
[stability ind.]
[legend]
5 characters max (left justified)
1 character: “?” when unstable, blank when stable
10 characters max: hh:mm:ss (when interval printing).
Output examples: (Note: * and _ indicate spaces)
*****192.21_ g - Manual, Continuous or On-Stable printing
*******0.01_ g _? - Unstable reading
*******0.01_ g ___00:00:00 - Interval Printing (15 sec intervals)
*****176.30_ g _?_00:00:15 - Unstable reading at 15 seconds
*****192.08_ g ___00:00:30 - Stable reading at 30 seconds