Agilent Cary 8454
UV-Visible Spectroscopy
System
Operator’s Manual
Notices
© Agilent Technologies, Inc. 2002,
2003-2008, 2011, 2013, 2014, 2016
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1115-90002
Edition
09/16
Agilent Technologies Australia (M) Pty Ltd
679 Springvale Road
Mulgrave, Victoria, Australia, 3170 www.agilent.com
Research Use Only
This product may be used as a component of an in vitro diagnostic system if the system is registered with the appropriate authorities and complies with the relevant regulations. Otherwise, it is intended only for general laboratory use.
Microsoft is a U.S. registered trademark of
Microsoft Corporation.
Software Revision
This handbook is for B.05.xx revisions of the
Agilent ChemStation software, where xx is a number from 00 through 99 and refers to minor revisions of the software that do not affect the technical accuracy of this handbook.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should
Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and
Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a
CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
In This Guide…
This book is an introduction to your Agilent Cary 8454 UV-Visible spectroscopy system. It includes step-wise procedures and examples for basic operations and tasks.
For more detailed information see the following manuals:
• Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
• Understanding Your UV-Visible Spectroscopy System
• Agilent Cary 8454 Service Manual
• UV-Visible ChemStation Help
1 Safety Information
Introduces your UV-Visible spectrophotometer and the Agilent ChemStation software.
2 Introduction to Your System
Introduces your UV-Visible spectrophotometer and the Agilent ChemStation software.
3 Installation and Start Up
Summarizes the system installation process, and explains how to start a measurement session.
4 Good Measurement Practices
Provides guidance for good measurement practices.
5 Using your UV-Visible Spectroscopy System
Provides examples of basic measurements and related tasks.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 3
4 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Agilent Cary 8454 Contents
1 Safety Information 9
Safety Information
10
Verifying Safe State
10
Warning Symbols 10
Precautions 12
Solvent Information
13
Flow Cells 13
Solvents 13
Other Precautions 15
Protective Earth 15
Disconnecting Device
15
Allowing Hot Instrument to Cool
16
Environmental Conditions and Power Requirements
17
Pollution level 18
Regulatory Information
18
Electromagnetic Compatibility 18
2 Introduction to Your System 21
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview 22
Optical System Overview
22
Spectrophotometer Description 26
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy —
Overview
30
User Interface Elements 31
Software Structure 35
Standard Mode Tasks
37
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 5
Contents
6
Standard Mode Data Processing
40
3 Installation and Start Up 47
Installation Summary for Your General Purpose UV-Visible System 48
General
48
Spectrophotometer
48
PC
49
Starting a Measurement Session 50
4 Good Measurement Practices 51
General Considerations
52
Spectrophotometer Design
52
Making Measurements
52
Sample Cell Material 53
Optical Specifications of Cells 54
Apertured Cells
55
Flow Cells 56
Handling and Maintaining Cells 57
Solvents 59
Sample Preparation 60
Photosensitive Samples 61
Stirring and Temperature Control 62
Checklist for Best Results
62
Inserting a Cell 65
5 Using your UV-Visible Spectroscopy System
67
Starting Your First Measurement Session
68
Starting Your UV-Visible Software 70
Measuring Caffeine Absorbance at 273 nm 71
Saving Your Parameters as a Method
74
Retrieving and Printing a Method
76
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Contents
Saving and Retrieving Data
79
Saving your Samples 79
Saving a Selected Spectrum
81
Retrieving Spectra 83
Deleting Current Spectra 84
Print Preview of Reports 85
Finding the Caffeine Absorbance Maximum 88
Entering your Cell’s Path Length
92
Controlling your Sipper System 93
Using your Multicell Transport
95
Quantitative Analysis using a Calibration with Standards
98
Setup 99
Calibration 101
Analysis
103
How Can I Be Sure That My Spectrophotometer Works Properly?
105
Spectrophotometer Self test
105
What Maintenance Does My Spectrophotometer Require?
107
Cleaning 107
Spare Parts
107
How Can I Get a Deeper Understanding of UV-Visible Spectroscopy?
108
When Do I Have to Measure a Blank?
110
Index 111
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 7
Contents
8 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Agilent Cary 8454 UV-Visible Spectroscopy System
Operator’s Manual
1
Safety Information
Environmental Conditions and Power Requirements 17
Agilent Technologies
9
1 Safety Information
Safety Information
Safety Information
Agilent products must only be used in the manner described in the Agilent product user guides. Any other use may result in damage to the product or personal injury. Agilent is not responsible for any damages caused, in whole or in part, by improper use of the products, unauthorized alterations, adjustments or modifications to the products, failure to comply with procedures in Agilent product user guides, or use of the products in violation of applicable laws, rules or regulations.
Verifying Safe State
The following general safety precautions must be observed during all phases of operation, maintenance and service of this instrument. To ensure continued safety of the instrument after maintenance or service procedures, verify the instrument is returned to a safe state for the customer and service personnel user. This includes running performance checks to verify the instruments safety systems are functioning correctly. Check the general condition of the instrument during operation, and for wear or signs of corrosion that are likely to inhibit function or safety.
Warning Symbols
To avoid personal injury, death, or damage to the instrument, the operator must refer to an explanation in the manual. This symbol is placed on the product where it is necessary for you to refer to the manual in order to understand a hazard.
10 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Safety Information 1
Safety Information
The following symbols appear on the instrument for your information. The hazard they describe is also shown.
Single phase alternating current.
When attached to the rear of the instrument, indicates that the product complies with the requirements of one or more
EU directives.
Indicates a protective ground terminal.
Eye damage may result from directly viewing light produced by deuterium lamps used in detectors and spectrophotometers. Always turn off the deuterium lamp before opening the lamp door on the instrument.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 11
1 Safety Information
Precautions
Precautions
Before applying power, comply with the installation section. Additionally the following must be observed.
If the instrument is used in any manner not specified by Agilent, the protection provided by the equipment may be impaired.
Do not remove instrument covers when operating. Before the instrument is switched on, all protective earth terminals, extension cords, auto-transformers, and devices connected to it must be connected to a protective earth via a ground socket. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in serious personal injury. Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any intended operation.
Make sure that only fuses with the required rated current and of the specified type (normal blow, time delay, and so on) are used for replacement. The use of repaired fuses and the short-circuiting of fuseholders must be avoided.
Some adjustments described in the manual, are made with power supplied to the instrument, and protective covers removed. Energy available at many points may, if contacted, result in personal injury.
Any adjustment, maintenance, and repair of the opened instrument under voltage should be avoided as much as possible. When inevitable, this should be carried out by a skilled person who is aware of the hazard involved. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present. Do not replace components with power cable connected.
Do not operate the instrument in the presence of flammable gases or fumes.
Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Do not install substitute parts or make any unauthorized modification to the instrument.
12 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Safety Information 1
Precautions
Capacitors inside the instrument may still be charged, even though the instrument has been disconnected from its source of supply. Dangerous voltages, capable of causing serious personal injury, are present in this instrument. Use extreme caution when handling, testing and adjusting.
Solvent Information
CAUTION
Organic and aqueous solvents can be hazardous if handled improperly. Always follow the precautions listed below to protect both operator and instrument.
Flow Cells
Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and thus impair the optical properties of flow cells.
Solvents
Always filter solvents, small particles can permanently block capillaries. Avoid the use of the following steel-corrosive solvents:
• Solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on).
• High concentrations of inorganic acids like nitric acid, sulfuric acid especially at higher temperatures (replace, if your analysis method allows, by phosphoric acid or phosphate buffer which are less corrosive against stainless steel).
• Halogenated solvents or mixtures which form radicals and/or acids, for example:
2CHCl3 + O2
2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 13
1 Safety Information
Precautions
• Analysis-grade ethers, which can contain peroxides (for example, THF, dioxane, di-isopropylether) such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides.
• Solutions of organic acids (acetic acid, formic acid, and so on) in organic solvents. For example, a 1-% solution of acetic acid in methanol will attack steel.
• Solutions containing strong complexing agents (for example, EDTA, ethylene diamine tetra-acetic acid).
• Mixtures of carbon tetrachloride with 2-propanol or THF.
14 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Safety Information 1
Other Precautions
Other Precautions
Use of the spectrometer system and accessories may involve materials, solvents and solutions which are flammable, corrosive, toxic or otherwise hazardous. Careless, improper or unskilled use of such materials, solvents and solutions can create explosion hazards, chemical burn hazards, fire hazards, toxicity and other hazards that can result in death, serious personal injury or damage to equipment. Apply all necessary precautions including use of lab coats, safety goggles and other appropriate forms of personal protection. All wastes should be disposed of in accordance with local regulatory requirements.
Protective Earth
WARNING
Shock Hazard
Connecting the Agilent Cary 8454 UV-Vis to a power source that is not equipped with a protective earth contact creates s shock hazard for the operator and can damage the instrument. Always operate your instrument from a power outlet which has a ground connection.
Disconnecting Device
WARNING
Shock Hazard
To disconnect the instrument from line power, disconnect the power cord from the supply. The power supply still uses some power and a shock hazard remains inside the instrument, even if the power switch on the front panel is turned off.
For safety reason do not position the equipment so that it is difficult to operate the disconnecting device.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 15
1 Safety Information
Other Precautions
Allowing Hot Instrument to Cool
WARNING
Hot Surface
The lamps and its surroundings remain hot for up to five minutes after the lamp is turned off. To avoid burns, allow the lamp compartment to cool before carrying our any work in this area.
UV Light Source
WARNING
Eye Hazard
Eye damage may result from directly viewing light produced by deuterium lamps.
Always turn off the deuterium lamps before opening the lamp door on the instrument. Direct viewing of ultraviolet rays without protection causes permanent blindness or vision impairment.
Power Cord
NOTE
Do not replace the mains power cord with one that is inadequately rated. When replacing the mains power cord, use only a mains power cord with a rating equivalent to the one supplied with the instrument.
Lithium Batteries Information
WARNING
Explosion Hazard
The Lithium battery can be found on the Spectrometer Processor Main Board (SPM).
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the equipment manufacturer. Do not dispose of lithium batteries in domestic waste.
16 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Safety Information 1
Environmental Conditions and Power Requirements
Environmental Conditions and Power Requirements
NOTE
Table 1
Suitable environmental conditions for the 8454 UV-Vis
Condition Altitude Temperature Humidity (%RH) non-condensing
90% Non-operating
(Storage or shipping)
Operating within specifications
Up to 4,600 m
(14,950 ft.)
Up to 2,000 m
(6,500 ft.)
- 40 to 70 °C
(-4 to 158 °F)
5-40 °C
(41-104 °F)
<95%
For optimum analytical performance, it is recommended that the ambient temperature of the laboratory be between 20-25 °C and be held +/- 2 °C throughout the entire working day.
WARNING
Hot Surface
If you use the instrument at environmental temperatures higher than 50 °C (122 F), the rear of the instrument may get hot.
CAUTION
Do not store, ship or use your instrument under conditions where temperature fluctuations could cause condensation within the instrument. Condensation will damage the electronics. If your instrument was shipped in cold weather, leave it in its box and allow it to warm up slowly to room temperature to avoid condensation.
Avoid air circulation of varying temperatures (like from refrigerators or air conditioners) and exposure to direct sun light to ensure the instrument operates within performance specifications.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 17
1 Safety Information
Environmental Conditions and Power Requirements
Pollution level
The pollution level is 2, based on IEC61010-1. Pollution level describes the degree to which a solid, liquid or gas that deteriorates dielectric strength is adhering. '2' applies to a normal indoor atmosphere, where only non-conductive pollution occurs.
Regulatory Information
CE Compliance
Your Agilent instrument has been designed to comply with the requirements of the Electromagnetic Compatibility (EMC) Directive and the Low Voltage
(electrical safety) Directive (commonly referred to as the LVD) of the
European Union. Agilent has confirmed that each product complies with the relevant Directives by testing a prototype against the prescribed EN
(European Norm) standards.
Proof that a product complies with these directives is indicated by:
• the CE Marking appearing on the rear of the product, and the documentation package that accompanies the product containing a copy of the Declaration of Conformity. The Declaration of Conformity is the legal declaration by Agilent that the product complies with the directives listed above, and shows the EN standards to which the product was tested to demonstrate compliance.
Electromagnetic Compatibility
EN55011/CISPR11
Group 1 ISM equipment: group 1 contains all Industrial, Scientific and
Medical (ISM) equipment in which there is intentionally generated and/or used conductively coupled radio- frequency energy which is necessary for the internal functioning of the equipment itself.
18 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Safety Information 1
Environmental Conditions and Power Requirements
Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
This device complies with the requirements of CISPR11, Group 1, Class A as radiation professional equipment. Therefore, there may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted as well as radiated disturbances.
Operation is subject to the following two conditions:
1 This device may not cause harmful interference.
2 This device must accept any interference received, including interference that may cause undesired operation.
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try one or more of the following measures:
1 Relocate the radio or antenna.
2 Move the device away from the radio or television.
3 Plug the device into a different electrical outlet, so that the device and the radio or television are on different electrical circuits.
4 Make sure that all peripheral devices are also certified.
5 Make sure that appropriate cables are used to connect the device to peripheral equipment.
6 Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance.
Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 19
1 Safety Information
Environmental Conditions and Power Requirements
South Korean Class A EMC declaration
A 급 기기 ( 업무용 방송통신기자재 )
This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
이 기기는 업무용 (A 급 ) 전자파적합기기로서 판 매자 또는 사용자는 이 점을 주
의하시기 바라 며 , 가정외의 지역에서 사용하는 것을 목적으 로 합니다 .
20 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Agilent Cary 8454 UV-Visible Spectroscopy System
Operator’s Manual
2
Introduction to Your System
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview 22
General Purpose Agilent ChemStation Software for UV-Visible
This chapter introduces the principles of data acquisition, data evaluation and data handling to help you to run your system successfully.
Your spectroscopy system is based on a UV-Visible spectrophotometer with general purpose Agilent ChemStation software for UV-Visible spectroscopy, running on a PC with a supported Microsoft Windows operating system. The spectrophotometer and PC are linked by a network connection. This type of link is very flexible. It can be used for a direct connection between the spectrophotometer and the PC, as well for integration in a enterprise network with network access from the PC to the spectrophotometer.
The spectrophotometer acquires and provides absorbance data which is handled by the PC’s application software. All of the data display, evaluation and long term storage is done under software control on the PC.
Agilent Technologies
21
2 Introduction to Your System
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
This section gives an overview of the optical system and explains the spectrophotometer front and back panels. It also explains the layout and construction of the spectrophotometer including the electronic and mechanical assemblies inside the spectrophotometer.
Optical System Overview
Optical System
The optical system of the spectrophotometer is shown in Figure 1. Its radiation source is a combination of a deuterium-discharge lamp for the ultraviolet (UV) wavelength range, and a tungsten lamp for the visible and short wave near-infrared (SWNIR) wavelength range. The image of the filament of the tungsten lamp is focused on the discharge aperture of the deuterium lamp by means of a special rear-access lamp design which allows both light sources to be optically combined and to share a common axis to the source lens. The source lens forms a single, collimated beam of light. The beam passes through the shutter/stray-light correction filter area then through the sample to the spectrograph lens and slit. In the spectrograph, light is dispersed onto the diode array by a holographic grating. This allows simultaneous access to all wavelength information. The result is a fundamental increase in the rate at which spectra can be acquired.
22 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
Cuvette
Spectrograph lens
Slit
Grating
Shutter/stray light filter
Tungsten lamp
Deuterium lamp
Source lens
Photo diode array
Figure 1 Optical system of spectrophotometer
• Lamps
The light source for the UV wavelength range is a deuterium lamp with a shine-through aperture. As a result of plasma discharge in a low pressure deuterium gas, the lamp emits light over the 190 nm to approximately
800 nm wavelength range. The light source for the visible and SWNIR wavelength range is a low-noise tungsten lamp. This lamp emits light over the 370 nm to 1100 nm wavelength range.
• Source Lens
The source lens receives the light from both lamps and collimates it. The collimated beam passes through the sample (if one is present) in the sample compartment.
• Shutter
The shutter is electromechanically actuated. It opens and allows light to pass through the sample for measurements. Between sample measurements it closes, to limit exposure of sample to light. If the measurement rate is very fast, you can force the shutter to remain open (using
Agilent ChemStation software) or it stays open automatically (handheld controller software).
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 23
2 Introduction to Your System
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
• Stray-Light Correction Filter
In a standard measurement sequence, reference or sample intensity spectra are measured without and then with the stray-light filter in the light beam.
Without the filter, the intensity spectrum over the whole wavelength range from 190–1100 nm is measured.
The stray-light filter is a blocking filter with 50 % blocking at 420 nm. With this filter in place, any light measured below 400 nm is stray light only. This stray-light intensity is then subtracted from the first spectrum to give a stray-light corrected spectrum. Depending on the software, you can switch off the stray light correction (using Agilent ChemStation software) if you want to do very fast repetitive scans, or it is switched off automatically.
• Sample Compartment
The spectrophotometer has an open sample compartment for easier access to sample cells. Because of the optical design a cover for the sample area is not required. The spectrophotometer is supplied with a single-cell holder already installed in the sample compartment. This can be replaced with the
Peltier temperature control accessory, the thermostattable cell holder, the long path cell holder or the multicell transport. All of these optional cell holders mount in the sample compartment using the same simple mounting system. An optical filter wheel is also available for use with the spectrophotometer and most of the accessories.
• Spectrograph
The spectrograph housing material is ceramic, to reduce thermal effects to a minimum. It main components of the spectrograph are the lens, the slit, the grating and the photo diode array with front-end electronics. The mean sampling interval of the diode array is about 0.9 nm over the wavelength range 190 nm to 1100 nm. The nominal spectral slit width is 1 nm.
• Spectrograph Lens
The spectrograph lens is the first of the parts which are collectively known as the spectrograph. It is mounted on the housing of the spectrograph. The spectrograph lens refocuses the collimated light beam after it has passed through the sample.
24 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
• Slit
The slit is a narrow aperture in a plate located at the focus of the spectrograph lens. It is exactly the size of one of the photo diodes in the photo diode array. By limiting the size of the incoming light beam, the slit makes sure that each band of wavelengths is projected onto only the appropriate photodiode.
• Grating
The combination of dispersion and spectral imaging is accomplished by using a concave holographic grating. The grating disperses the light onto the diode array at an angle linear proportional to the wavelength.
• Diode Array
The photodiode array is the heart of the spectrograph. It is a series of 1024 individual photodiodes and control circuits etched onto a semiconductor chip. With a wavelength range from 190 nm to 1100 nm the sampling interval is about 0.9 nm.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 25
2 Introduction to Your System
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
Spectrophotometer Description
Your spectrophotometer is very easy to use. It has a line power indicator, a status indicator and some push buttons. All electrical connections are made at the rear of the spectrophotometer.
Front View
The front view of the spectrophotometer is shown in Figure 2. Notice that the sample compartment is open. Unlike conventional spectrophotometers the
Agilent Cary 8454 UV-Visible spectrophotometer does not suffer from ambient false light. The open sample area makes it easier to access for cuvette handling and to connect tubing to a flow cell or thermostatable cell holder. The spectrophotometer is shipped with the standard single-cell holder. Standard and accessory cell holders can be removed and replaced in seconds with few or no tools.
Line power switch with green light
Pushbuttons
Indicator
Figure 2 Front view of spectrophotometer
The line power switch is located at the lower-left part of the spectrophotometer. Pressing it in turns on the spectrophotometer. It stays pressed in and shows a green light when the spectrophotometer is turned on.
When the line power switch stands out and the green light is off, the spectrophotometer is turned off.
26 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
On the front panel of the spectrophotometer is a status indicator which will display different colors depending of the actual condition of the spectrophotometer.
• Green—the spectrophotometer is ready to measure.
• Green, blinking—the spectrophotometer is measuring.
• Yellow—the spectrophotometer is in not-ready state, for example, turning one of the lamps on or if both lamps are switched off.
• Red—error condition, that is, the spectrophotometer does not pass one of the self-tests which are run when the spectrophotometer is turned on or an error occurred during operation. In this case the UV-Visible operating software gives a detailed error message and possible explanations are in the online help system and in your Service Manual Chapter 3 “Diagnostics and Troubleshooting”.
• Red, blinking—error condition of the spectrophotometer processor system.
Because in this case there is no communication with the computer there will be no error message. The online help system and your Service Manual
Chapter 3 “Diagnostics and Troubleshooting” give more information about troubleshooting.
The four measure push buttons on the front panel cause the following actions to be performed and the resulting data being sent to the computer. The push button functionality is controller by the UV-Visible ChemStation software and adjusted to the current measurement task.
• Blank—the spectrophotometer takes a blank measurement. This comprises a reference measurement that is used in all subsequent sample measurements until a new blank measurement is taken. Following the reference measurement an additional baseline spectrum is measured and displayed on the PC.
• Sample—the spectrophotometer takes a sample measurement or starts a series of measurements. This depends on the parameters set in your software.
• Standard—the spectrophotometer takes a measurement of a standard.
Additional information such as concentration and so on, have to be entered in the operating software.
• Stop—the spectrophotometer and/or software aborts any ongoing activity and returns to a to ready state.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 27
2 Introduction to Your System
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
Rear View
All connections are made at the rear of the spectrophotometer, see Figure 3.
Remote connector
Multicell transport connector
GPIB connector and switches
MIO slot for
LAN interface
GP-IO connector
RS 232 connector
Line connector
Figure 3 Rear view of spectrophotometer
• The multicell connector allows you to connect a multicell transport.
• The GPIO (general-purpose input/output) connector allow you to control a sipper and autosampler or other accessories depending on the software you are using.
• The remote connector is currently not used by Agilent instrument control software. It can be used e.g. to synchronize instruments.
• The RS232C connector may be used to control the spectrophotometer from a computer through RS232 connection, using appropriate software. The
Multiple Instrument Firmware Tool provided can be used with RS232 communication. The UV-Visible ChemStation software currently supports
LAN communication only.
• The MIO board slot is reserved for a LAN interface board.
• The accessory board slot is reserved for future use.
28 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
Agilent Cary 8454 UV-Visible Spectrophotometer — Overview
• The power input socket does not have a voltage selector because the power supply has wide-ranging capability, for more information see your Service
Manual Chapter 1 “Specifications”. There are no externally accessible fuses, because automatic electronic fuses are implemented in the power supply. The security lever at the power input socket prevents that the spectrophotometer cover is taken off when line power is still connected.
Side of the Spectrophotometer
On the right side of the spectrophotometer is a double door for access to the lamps. For exchanging lamps both doors, the plastic door and the sheet-metal door must be opened. Safety light switches are automatically turning off the lamps when the inner sheet metal door is opened.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 29
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
General Purpose Agilent ChemStation Software for UV-Visible
Spectroscopy — Overview
This section gives an overview of the elements of the user interface implemented with your Agilent ChemStation software and the data analysis concept behind. It explains how data are processed and what the advantages of this processing are from a practical point of view.
30 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
User Interface Elements
Your general purpose Agilent ChemStation software for UV-Visible spectroscopy facilitates operation of your diode-array-based UV-Visible spectrophotometer in daily routine operation. The focus of this software is on ease-of-use and ease-of-learning. A graphical user interface visualizes the spectrophotometer operation and usage. This user interface consists of a number of elements as described in the following sections.
Method name
Tool bar
Menu bar
Analysis panel
Data files
Navigation bar
Modes
Instrument panel
Tabular results
Spectra
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 31
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Menu
The more traditional menu interface at the top of the Agilent ChemStation window provides access to all operations. When you choose an item from the menu bar, a list of commands and submenus is displayed. An operation is performed by choosing a command (mouse click or Enter key).
Toolbar
The tool bar below the menu bar provides direct access to basic operations such as printing result reports, loading methods, and saving methods and data.
Navigation Bar
The navigation bar consists of mode switching bars and a file browsing tree view for saved files. The navigation bar is dockable, and can be hidden or resized. The navigation bar window provides fast, direct access to file system and mode switch functionality. The size and position of the analysis and instrument panels are fixed, but are a function of your display’s resolution.
Files can be loaded from the navigation bar tree view by double-clicking.
Clicking on the "Browse" section allows access to any available data directory.
Analysis Panel
The analysis panel gives you a graphical visualization of the current working context. In addition, it provides access to the setup dialog of your actual task by means of the Setup button.
32 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Instrument Panel
The instrument panel is below the analysis panel. It displays and controls your sampling devices and spectrophotometer. Some of the graphical elements on this panel are active items which are used, for example, to switch lamps on or off, or run a peristaltic pump.
When you move the mouse across an active area, the cursor changes. Click in an active area to bring up a small menu with selections, or to perform an operation.
In addition you can select a sampling system from the list of available systems.
Its parameters can be adjusted by means of the Setup button.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 33
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
View
The area on the right side of the side panels provides a view on a certain aspect of your current task. A view consists of one or more separate windows.
These windows provide information such as a graph showing your measured sample spectra, or a table with the calculated results.
Views are usually handled automatically by the operation you performed, but you may also use the view menu’s commands to select the view you prefer.
34 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Software Structure
The software is divided into specific applications, called modes. In addition levels of operation are available, and there is support for data evaluation sessions without spectrophotometer control.
Agilent ChemStation Sessions
Your Agilent ChemStation belongs to the Agilent’s ChemStation family of instrument control software. An installation of Agilent ChemStation software can control up to four different UV-Visible instruments on a single PC. Each of these instruments has its own session
If you start a session, its name is indicated in the main application’s window title bar, for example, 845x UV-Visible System[1].
Each instrument session is available for data analysis only, or for instrument control. An instrument control session has the suffix (Online) and can be started as the first instance only on your PC. In addition a data analysis session, which has the suffix (Offline), can be launched. The offline session allows recalculations based on stored data, and is useful in the development of analytical methods.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 35
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Operation Levels
The operation levels manager level and operator level apply to all modes and allow managing an application and running an application only. In the managing level of an application, methods can be developed and stored permanently to disk. The manager level of operation is password protected.
This assures the integrity of predefined methods and operation sequences.
In the operator level, a reduced set of functions is available. Functions which may affect the integrity of an analytical procedure are not available. However, an operator may use their own settings. This is flagged on the tool bar and is indicated on printed reports.
Agilent ChemStation Modes
The Agilent ChemStation modes are application oriented. Each mode has its own mode-specific menu, panels, operations and set of views. Your general purpose software for UV-Visible spectroscopy is the platform for all modes. It is split into a Standard mode, an Execute Advanced Method mode and a
Verification and Diagnostics mode.
36
Depending on your needs, modes are available for Advanced operation,
Dissolution Testing runs, Multibath Dissolution Testing runs, Combined
Reports evaluation, Kinetics measurements, Thermal Denaturation studies and Color Calculations .
These modes of operation can be switched within a running
Agilent ChemStation session. All current raw data will be preserved during such a switch. This allows you to view your data from different perspectives.
Most of the modes offer you the ability to define your analytical task by a set of parameters and, if necessary, data. A set of parameters and data can be saved as a method. This allows you to repeat your analysis task under defined conditions simply by loading a method and running your samples.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Standard Mode Tasks
The Agilent ChemStation modes also offer the ability to focus on specific tasks.
The standard mode of your general purpose software for UV-Visible spectroscopy is oriented towards the most common tasks performed in a analytical laboratory that employs UV-Visible spectroscopy. Four tasks are available:
• Fixed Wavelength
• Spectrum/Peaks
• Ratio/Equation
• Quantification
A task is selected and activated from the analysis panel’s selection box.
The task orientation lets you quickly adjust the software to give you the right view and answers based on your data. These tasks have been derived from a survey of the most common UV-Visible spectroscopic tasks done on a routine basis in analytical laboratories. Historically, these tasks have been developed on filter photometers or scanning spectrophotometers.
With your spectrophotometer you have the advantage that, by default, the entire UV-Visible spectrum of your samples is available. So these four tasks offer different views of your acquired data.
A task switch within the Standard mode is much faster than an entire mode switch. An additional benefit of these tasks is that the definition of a method is done within a single dialog.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 37
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Fixed Wavelength
The Fixed Wavelength task is used to look at measured sample data at up to six different wavelength. This data is available as absorbance, transmittance and first to fourth derivative.
Due to the spectral acquisition, additional techniques such as internal reference or three-point, drop-line background corrections can be applied.
Spectrum/Peaks
38
In the Spectrum/Peaks task, you can view absorbance minima and maxima.
The focus here is more on the wavelength scale, but the relevant absorbance readings are also provided.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Ratio/Equation
The Ratio/Equation task is used to evaluate a user-definable equation based on measured data and sample information. An equation can be set up using sample data at up to six wavelengths. Weight and volume data is entered with the samples measured. By means of an equation, for example, analysis results based on chemical test kits can be automatically calculated and reported.
Another application for this task is to use a ratio of data values to check for the identity or the purity of a sample.
Quantification
The Quantification task allows you to do single component analysis based on four different types of calibration curves and a set of standards. Due to the spectral acquisition, background corrections can also be applied to your data.
The calibration can be optimized for your concentration range of interest by changing the wavelength used for calibration. A new calibration and a new analysis is automatically performed, based upon your current standard data.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 39
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Standard Mode Data Processing
General Data Processing
Although knowledge of the internal design of the data flow and processing is not required to use Agilent ChemStation software, it can be useful to understand how Agilent ChemStation processes your data and how this processing is controlled by your method parameters.
The data processing can be described using a model of data containers and operations visualized in a flow diagram.
Spectral Processing
Raw Data
Data Access
Processed
Spectra
Data Evaluation
Used
Wavelength
Results
All basic data go into a raw data container. This container is empty when you start your Agilent ChemStation session and it is filled by measuring data or loading data from a file.
The raw data container held the originally acquired data as specified with your acquisition parameters and stamped with, for example, acquisition date and time and the acquisition operator’s name.
40 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Spectral
Processing
Used Wavelength
Your method defines how this data is analyzed. The first processing step is spectral processing. The processed raw data spectra are transferred automatically into a second container for processed spectra . This allows you to examine the results of this processing step by viewing the processed spectra container’s content. If, for example, you specified the first derivative data type, the first derivative spectra of all your raw data are available in the processed spectra container after analysis. The type of spectral operation is defined by your method settings.
Results
The next step in the data analysis process is the provision of data for further evaluation specified in terms of wavelength and background correction operations such as an internal reference calculation or three-point, drop-line calculation.
This data is stored in the used wavelength container. In the Fixed
Wavelength task, for example, a tabular view of this data is available with the Sample/Results Table window.
An additional step is the evaluation of the accessed data.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 41
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
In the Ratio/Equation task, for example, this used wavelength data is processed by an evaluation of the equation specified which generates the calculation results. These calculated result values are stored in the results container.
The results are available with the Sample/Results Table.
Summary The generalized basic data processing is divided into three steps.
1 spectral processing
2 data access
3 data evaluation
These steps are always performed in the above order identically for all spectra in the raw data container. The results are placed in the results container. The previous content of the processed spectra , used wavelength and results container is replaced.
42 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Processing with Standards
In the fourth task, standards are used in addition to your sample data. This requires an extension to the above concept to handle standards. Two independent sets of containers are implemented, one for standards, and another for samples. All processing containers are also duplicated. As with the sample-only processing, all evaluation steps are done in parallel on both the samples and the standards.
Samples Standards
Spectral Processing
Raw Data Raw Data
Processed
Spectra
Used
Wavelength
Data Access
Processed
Spectra
Data Evaluation
Used
Wavelength
Calibration
Results
Calibration
Results
The evaluation in quantification is now a calibration using standards. The coefficients are calculated based on the method’s settings and the current standards in the Agilent ChemStation memory. This means that results now become a function of a measured set of standards and of the standard concentrations specified.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 43
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
44
These coefficients are then used to calculate concentration results for the samples currently in memory. The same processing steps, applied to sample and standard data, lead to the most precise results.
Advantages
A diode-array-based spectrophotometer, in combination with powerful data evaluation based on Agilent ChemStation, offers many advantages over traditional spectrophotometer systems. Some of these advantages are:
• Virtually unlimited number of standards
Measure your standards before or after your samples and, as well as the minimum required number of standards, use as many standards as you wish with your calibration.
• Easy optimization
The availability of all raw data—samples and standards—means you can easily optimize your method’s settings by choosing a different calibration wavelength and re-calibrating your system. Also, by simply removing this standard from your standard data set, you can eliminate outliers in your calibration.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
• Calibrated method
When you save your method, the standards currently in memory are always saved with the method. After loading a method, you can directly analyze your samples.
• Optimization for a particular sample
You can optimize your wavelength settings for a sample outside the linear range of your actual calibration. Due to the excellent wavelength reproducibility of your spectrophotometer, you can switch to a wavelength with a lower extinction coefficient for precise analysis of such a sample.
In summary, the availability of spectral raw data offers you many additional opportunities to optimize your calibration and analysis for best results. This optimization can be done quickly, just by setting new method parameters. You get new answers almost instantaneously with consistent and reliable results.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 45
2 Introduction to Your System
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
46 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Agilent Cary 8454 UV-Visible Spectroscopy System
Operator’s Manual
3
Installation and Start Up
Installation Summary for Your General Purpose UV-Visible System 48
Starting a Measurement Session 50
This chapter does not replace the information available with the Agilent Cary
8454 UV-Visible Spectroscopy System Installation Guide. It is intended as a reminder of the key steps of the installation and system startup.
Agilent Technologies
47
3 Installation and Start Up
Installation Summary for Your General Purpose UV-Visible System
Installation Summary for Your General Purpose UV-Visible
System
48
General
A detailed description of your UV-Visible system is given with the manual
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide. This summary reminds you of the key points of installation.
Spectrophotometer
CAUTION
✔ Make sure that your spectrophotometer has the LAN interface card installed.
✔ Check that your spectrophotometer is either connected to your PC directly using a crossover LAN cable, or to your LAN using a direct connection.
The Agilent Cary 8454 UV-Visible spectrophotometer does not have a CAN interface.
Do not connect the LAN adapter of your PC to the CAN interface of the Agilent 8454 spectrophotometer, otherwise the LAN adapter of the PC will be seriously damaged, because the operating voltage of the CAN interface (12 V) is higher than the operating voltage of the LAN adapter (5 V).
WARNING
✔ Check that your spectrophotometer is connected to a power outlet.
Always operate your instrument from a power outlet which has a ground connection. Always use the power cord designed for your region.
✔ Before you switch on your spectrophotometer, make sure that the Agilent
Bootp Service is installed on your PC or your network administrator has assigned an IP address to your spectrophotometer. For details see the chapter “LAN Communication, Installation, Connection and Configuration” in your Agilent Cary 8454 UV-Visible Spectroscopy System Installation
Guide.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Installation and Start Up 3
Installation Summary for Your General Purpose UV-Visible System
PC
✔ Make sure that all of your PC components are connected to line power.
✔ Make sure that your general purpose software for UV-Visible spectroscopy is installed.
✔ A printer must be configured on your PC.
• Adjust paper size (for example, Letter, A4)
• Set Orientation to Portrait
✔ The TCP/IP protocol must be installed and configured on your PC.
✔ Your spectrophotometer must be configured with its IP address.
✔ If you connected your spectrophotometer directly to your PC be sure that the Agilent Bootp Service is installed and configured. If you connect to your spectrophotometer using a LAN make sure that your network administrator assigns the IP address to your spectrophotometer.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 49
3 Installation and Start Up
Starting a Measurement Session
Starting a Measurement Session
NOTE
With a network connection to your spectrophotometer, it is important that your spectrophotometer is recognized by the software. This requires the assignment of a unique IP address to your spectrophotometer when it is powered on. The assignment is done either by the Agilent BootP Service application running on your PC with a direct connection to the spectrophotometer, or by a server application on your LAN. Therefore it is important that either one of these applications is up and running before you switch your spectrophotometer on.
✔ Switch on your PC and boot your PC operating system. If a printer is connected to your system, switch the printer on.
✔ Make sure that the Agilent BootP Service is running or you are logged onto your LAN.
✔ Switch your spectrophotometer on and wait until the spectrophotometer’s indicator light turns to green. This process includes the spectrophotometer’s self test and takes about one minute. For details on the startup sequence see chapter “Installation and Start Up” in your Agilent
Cary 8454 UV-Visible Spectroscopy System Installation Guide.
✔ Launch your measurement session by pressing Start > All Programs or All
Apps > UV-Visible ChemStations > Instrument 1 online.
✔ You are ready to use your system, if the blue busy status display on the system’s bottom message line turns off.
✔ The first measurement you have to perform is a reference measurement.
After this alignment you are ready to measure absorbance data and spectra.
It takes about 15 minutes for the lamps to reach stable state conditions. For best results, do not perform measurements before this period of time has elapsed.
50 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Agilent Cary 8454 UV-Visible Spectroscopy System
Operator’s Manual
4
Good Measurement Practices
This chapter describes
• making measurements
• selecting material, optical specification and type of cell
• handling and maintaining cells
• checklist for good results
• solvents selection
• sample preparation
• use of filters
• stirring and temperature control of sample
• how to insert cells into the cell holder.
Agilent Technologies
51
4 Good Measurement Practices
General Considerations
General Considerations
There are many factors that can affect the results of your measurements. This section provides brief discussions of some of the more important ones.
Spectrophotometer Design
The sample compartment of the spectrophotometer is open. Unlike conventional instruments the UV-Visible spectrophotometer does not suffer from ambient false light. The open sample area makes it easier to access it generally and to connect tubing to a flow cell or thermostattable cell holder.
Making Measurements
Blank (Reference) and Sample Measurement
Your spectrophotometer is a single beam instrument so you must measure a blank before you measure a sample. For the high accuracy measurements, the blank and the sample measurement should closely follow each other.
In general, a blank measurement should be repeated as often as is practical.
Even in a thermally stable environment, a blank measurement should be taken at least every half hour to ensure accurate results.
Chemically, the only difference between the blank and the sample should be the presence of the analyte(s). For measurements with liquid samples, the blank should be a sample cell filled with the solvent you plan to use.
52 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
General Considerations
Sample Cell Material
Quartz sample cells or sample cells with quartz face plates are required if you want to use the full 190 to 1100 nm wavelength range of your spectrophotometer.
If you plan on working only in the visible and/or short-wave near-infrared range of 350 to 1100 nm, you can use good quality glass cells.
Disposable plastic sample cells , for measurements in the range 400–1100 nm, are also available. The quality of these cells varies and they are generally not recommended.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 53
4 Good Measurement Practices
General Considerations
Optical Specifications of Cells
The accuracy of the readings of a diode-array spectrophotometer is very sensitive to spatial shifts of the measurement light beam. Cells having non parallel opposite faces, or so called wedge shaped cells, lead to a spatial shift of the light beam (see Figure 4). Therefore, the opposite cell walls illuminated by the analysis light beam have to be very parallel. The parallelism is measured in terms of the angle between the two opposite cell walls . We recommend that you use 10 mm path length cells with an angle that is below
0.1 degrees of an arc .
Parallel
Cell Walls
Light Beam In
Light Beam in
Figure 4 Shift of the spectrophotometer light beam due to non-parallel cell walls
54 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
General Considerations
Apertured Cells
In applications where sample volume is limited, apertured or microcells are used. The width of these cells is reduced to reduce the volume and the blank part of the cell must be blackened to avoid unwanted transmission and reflection through the side walls. If the side walls are not blackened the result will be poor photometric accuracy and, if different concentrations are measured, poor linearity.
The disadvantage of apertured and microcells is that part of the light beam is blocked. Not all the light passes through the sample and there can be some loss of sensitivity. See Figure 5 for recommended and Figure 6 for cells you should not use with the instrument.
Figure 5
Quartz cells
Recommended cells
Quartz cells with black
CAUTION
* Quartz cells with black apertures smaller than 2 mm, when used with a multicell transport, can lead to measurements with poor reproducibility.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 55
4 Good Measurement Practices
General Considerations
Quartz cells with transparent apertures, fluorescence cells, plastic cells
Figure 6 Cells you should not use with the instrument
Flow Cells
We recommend a sipper system with a flow cell for obtaining high precision measurements. Using a flow cell eliminates the necessity of moving the cell between blank measurement and sample measurement. Also, the cell can be rinsed thoroughly with the solution to be measured.
The design of the flow cell should minimize entrapment of bubbles and flow channeling to provide the most reliable results.
56 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
General Considerations
Handling and Maintaining Cells
Passivating New Cells
When filling a non-passivated new cell with your sample, you will observe that air bubbles stick on the windows of your cell. To prevent the formation of sticky bubbles, rinse the cell with cleaning and passivating fluid (part number
5062-8529). The cleaning procedure is described on the label of the cleaning fluid container.
Cleaning Cells
The fats in fingerprints are significant absorbers in the UV region and, if left on optical surfaces, can cause erroneous results. Wipe off all fingerprints and contaminants before using a sample cell.
Use only high quality lens tissues (part number 9300-0761) and never dry the inside of a cell with lens tissues . Dry the inside of the cell with pressurized, oil free air, that prevents the cell from contamination with tissue particles, or rinse the cell with blank or sample solution. Floating particles in the cell will deflect the light beam and result in very poor quality of the measured spectrum.
Floating particles will deflect and scatter the light beam
Figure 7 Floating particles in a cell
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 57
4 Good Measurement Practices
General Considerations
58
CAUTION
Figure 8 Spectrum taken with floating particles in the light path
Lens tissues for glasses or other uses often contain detergents or lubricants which can affect your measurements. If possible, avoid cleaning the faces of your cell between blank and sample measurements.
Handling Cells
Always install a cell so that it faces the same direction to minimize problems with cell non-uniformity. For best results with microcells, leave your sample cell clamped in position throughout the measurement sequence. Solutions should be removed and replaced by pipette or use flow cells.
If glass pasteur pipettes are used, make sure that the optical windows of the cell are not touched or scratched by the pipette.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
General Considerations
Solvents
Your choice of solvents should be based primarily on the solvent’s absorbance characteristics over the wavelengths of interest, its suitability as a solvent for the analyte, and on experimental conditions. Table 2 lists common solvents and the lower limit of their useful wavelength range.
Table 2
Lower Limit
180–195 nm
200–210 nm
210–220 nm
245–260 nm
265–275 nm
280–290 nm
Above 300 nm
Lower limit of UV transmission for some common solvents
Solvent
Sulfuric acid (96%)
Water
Acetonitrile
Cyclopentane n-Hexane
Glycerol
2,2,4-Trimethylpentane
Methanol n-Butyl alcohol
Isopropyl alcohol
Cyclohexane
Ethyl ether
Chloroform
Ethyl acetate
Methyl formate
Carbon tetrachloride
Dimethyl sulfoxide
Dimethyl formamide
Acetic acid
Benzene
Toluene m-Xylene
Pyridine
Acetone
Carbon disulfide
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 59
4 Good Measurement Practices
General Considerations
WARNING
Many of the solvents in Table 2 are hazardous. Be sure you fully understand their properties before using them.
When using volatile solvents such as acetone or methylene chloride, make sure that the sample cell is stoppered. Evaporation of a solvent does change the solute concentration and can cause solution noise due to solute convection currents. Both of these will affect the accuracy of your measurements. We also recommend stirring and temperature control when you use volatile solvents.
When using water as solvent we recommend using UV grade or HPLC grade water to reduce unwanted absorbance from impurities in the water. If you are using the sipper/sampler system, the water should be degassed to avoid bubble formation in the flow cell, especially if the water comes from a pressurized water supply.
Sample Preparation
The sample cell should be rinsed three to five times with your intended solvent before you fill it with the pure solvent that will be used in the measurement.
Turning the cell upside down on a small stack of absorbent tissues will help remove any residual solvent. This treatment will minimize contamination from previous experiments.
Samples which contain colloidal dispersions, dust or other particulate matter should be filtered, centrifuged or allowed to settle. If not, the overall attenuation-of-transmittance spectrum due to light scattering and/or reflection will hide the spectral information from the analyte.
60 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
General Considerations
Photosensitive Samples
A few substances are very photosensitive. They degrade or undergo photochemical reactions if exposed to light. This can be easily seen by a decrease of sample absorbance over time.
Use of Filters
The shorter wavelength, higher-energy UV light is most likely to degrade photosensitive samples. If this is a problem, you can selectively block portions of the UV spectrum with a UV cut-off filter. An optical filter wheel assembly with three cut-off filters is available for the spectrophotometer. The cut-off wavelength of the filter you choose should be low enough that it does not eliminate important spectral information, but high enough that it blocks the light that could degrade your sample. If you use a filter with your samples, you must use the same filter when you make your blank measurement.
Turning the D 2 -Lamp off
The short wavelength radiation leading to photodegradation comes from the light of the D2-lamp. For application where readings are taken at wavelengths above 400 nm, the D2-lamp can be turned off. The light intensity supplied by the Tungsten lamp is sufficient for a good signal to noise ratio over the wavelength range 400–1100 nm. When using cells with small apertures, you should check the signal to noise ratio by making sample measurements under the conditions of your application.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 61
4 Good Measurement Practices
General Considerations
Stirring and Temperature Control
Solution homogeneity can be a problem, especially for viscous solutions. There are cases where, due to convection induced gradients, rapid absorbance changes may give irreproducible data. These changes can be observed spectroscopically by taking measurements with short integration times. To minimize convection effects keep the temperature of your sample the same as the cell holder or environmental temperature. Problems like these can also be minimized by using a thermostattable cell holder and/or a stirring module.
A similar effect can occur in cases of incomplete mixing. This is especially true where the specific gravities or miscibilities of the solvent and analyte are quite different. Again, stirring is a way to prevent this kind of problem.
In an unstirred cell, it is sometimes possible to observe local photodegradation of sensitive analytes. Because the actual volume of the sample in the light path is very small, stirring the sample will reduce the time any given analyte molecule is in the light path. This minimizes the photodegradation and increases homogeneity. Using a flow cell with continuous flow can yield similar results.
62
Checklist for Best Results
Cell:
✔ Cell is made of quartz or glass
✔ Apertured cell has black sides
✔ Apertured cell has an aperture greater than or equal to 3 mm
✔ Cell windows are free of fingerprints and other contamination
✔ Flow cell used instead of an apertured standard cell
Measurements:
✔ Solution in cell is free of floating particles
✔ Solution in cell, and cell walls are free of bubbles
✔ Solution in cell is mixed homogeneously
✔ Blank measured on same solvent as sample
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
General Considerations
✔ Blank measurement shows a flat baseline (Figure 9 and Figure 10 on page 64 show a good and a poor baseline)
✔ Cell orientation of blank and sample measurements is the same
✔ Ideally the cell is not removed between the measurements, which means the cell is filled/rinsed using a pipette or a flow cell is used
✔ Time between blank and sample measurement should be short
Figure 9 Example of a blank on water showing a good baseline
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 63
4 Good Measurement Practices
General Considerations
NOTE
Figure 10 Example of a blank on water with bubbles causing a poor baseline
64 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Good Measurement Practices 4
Inserting a Cell
Inserting a Cell
Your spectrophotometer is shipped with the standard single-cell holder which you must install in the sample compartment. This cell holder accommodates standard cells or flow cells. To insert a sample cell in the cell holder:
1 Move the locking lever to its up position.
2 Insert the sample cell, making sure you orient it correctly. The frosted (non-clear) sides of the sample cell should not be in the path of the light beam.
3 Lock the sample cell in place by pushing the locking lever back down.
Utmost care should be taken when using small volume flow cells and particularly any cells with less than a 2 mm aperture. Here it is important that the cells are properly centered in the light path. The cell holder is designed for cells with 15 mm center height. If applicable, these cells should not be removed between the respective reference (blank) measurement and the sample measurement.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 65
4 Good Measurement Practices
Inserting a Cell
66 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Agilent Cary 8454 UV-Visible Spectroscopy System
Operator’s Manual
5
Using your UV-Visible Spectroscopy
System
Starting Your First Measurement Session 68
Starting Your UV-Visible Software 70
Measuring Caffeine Absorbance at 273 nm 71
Saving Your Parameters as a Method 74
Retrieving and Printing a Method 76
Finding the Caffeine Absorbance Maximum 88
Entering your Cell’s Path Length 92
Controlling your Sipper System 93
Using your Multicell Transport 95
Quantitative Analysis using a Calibration with Standards 98
How Can I Be Sure That My Spectrophotometer Works Properly? 105
What Maintenance Does My Spectrophotometer Require? 107
How Can I Get a Deeper Understanding of UV-Visible Spectroscopy? 108
When Do I Have to Measure a Blank? 110
Agilent Technologies
67
5 Using your UV-Visible Spectroscopy System
Starting Your First Measurement Session
Starting Your First Measurement Session
1 Make sure that your spectrophotometer has been installed correctly.
For details of installation see your Agilent Cary 8454 UV-Visible
Spectroscopy System Installation Guide.
2 Switch on your PC, monitor and printer.
3 Log on to your PC operating system.
Confirm that your Agilent BootP Service is running using the Control
Panel’s Administrative Tools Services dialog, or make sure that your network administrator has integrated your spectrophotometer into your local network.
68
4 Switch on your spectrophotometer.
A running BootP service will now assign the configured IP address to your spectrophotometer. In the standard installation, the Agilent BootP Service performs this task.
5 Start your measurement session by selecting Instrument 1 online from the menu.
Your Instrument Panel shows you the current state of the spectrophotometer and the Blank button is enabled.
6 The first task is to measure a reference. Typically the cell containing the solvent used with your samples is put in the measurement position and a
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
NOTE
NOTE
Using your UV-Visible Spectroscopy System 5
Starting Your First Measurement Session blank measurement performed. To start this measurement, click the Blank button on your Instrument Panel or press the spectrophotometer’s Blank button.
A blank measurement is a reference measurement combined with the measurement of a baseline spectrum. A baseline spectrum gives you additional hints on the absorbance of the cell windows and the solvent.
Areas with high noise indirectly indicate high absorbance.
For high precision measurements, wait until the spectrophotometer and the lamps have reached thermal equilibrium. The time required is a function of the environmental conditions. Your spectrophotometer should be ready after 45 minutes.
7 The next measurement is your sample measurement. To get the most precise results, use the same cell in the same orientation to the measurement beam. Flush your cell about three times with your sample solution and start the measurement by clicking the Instrument Panel’s
Sample button, or by pressing the spectrophotometer’s Sample button.
For details on how to mount your cell, see “Inserting a Cell" on page 65.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 69
5 Using your UV-Visible Spectroscopy System
Starting Your UV-Visible Software
Starting Your UV-Visible Software
This section describes how to start an Agilent ChemStation session on your
PC. If you want to perform measurements, you can start an online session, or you can start an offline session for optimizing the analytical parameters of a method, recalculating results or printing reports.
An online session can be started only as the first instance on your PC. An offline session can be started in parallel. This allows you to optimize your method settings by direct comparison based on identical sets of data.
1 Switch on your PC, monitor and printer.
2 Log on to your PC’s operating system
3 Start your Agilent ChemStation session by selecting Instrument 1 online, for a measurement session, or Instrument 1 offline, for method optimization and data evaluation.
4 Enter your name to log on to your Agilent ChemStation session. If you protected your manager level by password, you must enter the correct password. The system will then come up with the last-used mode and method.
70 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Measuring Caffeine Absorbance at 273 nm
Measuring Caffeine Absorbance at 273 nm
This section describes how to measure the caffeine sample that was shipped with your spectrophotometer. Measurement of this caffeine sample is also used for the IQ (installation qualification) of your spectrophotometer.
1 Make sure that you are in Standard mode. The mode is indicated on the toolbar of your Agilent ChemStation session.
2 Select the Fixed Wavelength task in the analysis panel’s selection box.
3 Click Setup in the analysis panel to open the parameter dialog.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 71
5 Using your UV-Visible Spectroscopy System
Measuring Caffeine Absorbance at 273 nm
4 Type the wavelength of interest in the Wavelengths section of the Fixed
Wavelength(s) Parameters dialog. Adjust your spectral display to a wavelength range from 190 nm to 400 nm in the Display spectrum section.
Click OK to set your parameters.
5 Fill your 1 cm path length quartz cell with distilled water. Lift the lever at the left side of your cell holder. Put the cell in the cell holder and make sure the transparent windows face towards the font and the back of the spectrophotometer. Push the lever down to secure your cell in the cell holder.
72 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Measuring Caffeine Absorbance at 273 nm
6 Press the Blank button on the front of the spectrophotometer or click Blank on the Instrument Panel to start the measurement.
7 Remember the orientation of the cell in the cell holder. Lift the level to release the cell, remove it and flush it three times with about 1 ml of your caffeine sample. Then fill the cell with about 3 ml of the caffeine sample.
Make sure that the cell windows are clean and reposition the cell in the same orientation as for the reference measurement. Close the cell holder’s level.
8 Press the Sample button on the front of your spectrophotometer, or click
Sample on the Instrument Panel to start the measurement.
9 The view shows you the spectrum of your caffeine sample with a vertical line indicating your wavelength of interest. Below the spectrum is the
Sample/Result Table which shows the absorbance reading at 273 nm.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 73
5 Using your UV-Visible Spectroscopy System
Saving Your Parameters as a Method
Saving Your Parameters as a Method
This section describes how to save your settings for a future
Agilent ChemStation session. Simply by loading your method you adjust your
Agilent ChemStation to repeat your measurement. A library of methods facilitates routine laboratory work.
Let’s assume that your current settings are defining your method to analyze a caffeine sample. To be able to repeat such an analysis all set parameters can be stored permanently to disk. This allows you to load such a method on your system or even transfer such a method to your colleague with a
Agilent ChemStation system.
1 Choose Save Method As… from the File menu or click the icon.
74 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Saving Your Parameters as a Method
2 This displays the Save Method As… dialog box. Type the method name in the File name field, for example, Caffeine.m. Click OK to save your method.
NOTE
If you generate many methods, you can use Options & Infos… from the Method menu to add text for documentation purposes and to simplify method access. This displays the
Method Options & Information dialog box. In the Method Information section you can enter a descriptive text which becomes part of your method.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 75
5 Using your UV-Visible Spectroscopy System
Retrieving and Printing a Method
Retrieving and Printing a Method
This section describes how to access methods and print a method report.
1 Choose Load Method… from the File menu or click the icon in the toolbar.
NOTE
If your current method has been modified, a dialog box will ask you whether you want to save or ignore these changes.
76 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Retrieving and Printing a Method
2 The Load Method… dialog is displayed. The selected method information is shown in the File Information section of the dialog box.
NOTE
3 If you want to load this method, click OK.
Whenever you change a parameter of the current method, an indication is displayed in the toolbar’s modification field.
This triggers the reminder dialog mentioned above.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 77
5 Using your UV-Visible Spectroscopy System
Retrieving and Printing a Method
4 To print a method, choose Print, Method from the File menu.
NOTE
To print your method report, your printer must be properly configured and online. If your printer is currently not online, you can view the print preview on the screen.
78 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Saving and Retrieving Data
Saving and Retrieving Data
This section explains how to save and retrieve measured data. This data can be used for archiving, for method development at a later stage, or for exchange with other Agilent ChemStations.
Your Agilent ChemStation can store and retrieve your data using a binary, checksum-protected data format (extension *.sd,*.std). All current spectra
-samples or standards- can be saved to disk for permanent storage. Saving and loading of data is possible using local and network drives. In addition a single spectrum can be selected for storage.
Saving your Samples
1 Choose Save, Samples As… from the File menu or click the toolbar icon.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 79
5 Using your UV-Visible Spectroscopy System
Saving and Retrieving Data
2 Select a data files in the File name selection box of the Save Spectra As… dialog or type a valid file name into the File name edit box.
80
NOTE
A valid file name consist of eight alphanumeric characters and the file extension .sd or .std.
Usually the extension .std is used for standards only.
If the file name exists, a message box is displayed allowing you to abort or continue with the operation.
3 Click OK to start the operation.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Saving and Retrieving Data
Saving a Selected Spectrum
1 Select the spectrum of interest in the graphic window.
Or, in the tabular Sample/Results Table window.
2 Choose Save, Selected Spectra As… from the File menu.
NOTE
If you did not select either the Overlaid Sample Spectra window or the Sample/Results
Table window. you may get the warning, Select/activate a window!, on the message line.
3 Select a data files in the File name selector box of the Save Spectra As… dialog or enter a valid file name into the File name edit box.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 81
5 Using your UV-Visible Spectroscopy System
Saving and Retrieving Data
4 Press OK to start the operation.
82 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Saving and Retrieving Data
Retrieving Spectra
1 Choose Load, Samples… from the File menu.
2 Select the data file in the File name selector box of the Load Spectra to
SAMPLES dialog box.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 83
5 Using your UV-Visible Spectroscopy System
Saving and Retrieving Data
Select a file in the file selector box to view file information for that file.
NOTE
3 Click OK to start the operation. The spectra available with your data file are added to the files currently in the Agilent ChemStation sample container.
Deleting Current Spectra
1 Choose Clear, Samples from the Edit menu or click the toolbar icon.
NOTE
Clear, Standards and Clear, Math. Results can be used to delete the current standards and the current mathematical results spectra from the Agilent ChemStation memory.
84 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Print Preview of Reports
Print Preview of Reports
Print preview allows you to view the report in a separate Agilent ChemStation window based on the currently configured printer. All types of available reports can be checked page by page in the preview window. The number of pages generated and the layout also can be checked. In addition the currently displayed report can be printed.
Print Preview of a Results Report
The print previews work similarly for all available types of reports. The example below shows you how to preview your results report.
1 Choose Print Preview, Results from the File menu.
NOTE
Printing a results report requires you to have set all parameters properly and have data available for evaluation. If you do not have data, you may get the message No results
present! on the message line.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 85
5 Using your UV-Visible Spectroscopy System
Print Preview of Reports
2 The report generated will be displayed in the preview window.
86
This window allows you to look at your report page by page. Scroll bars are available if a page does not fit into the actual preview window.
In addition, you may use a different size for your preview display. Three sizes are available with the size selection box. Depending on your display resolution select the one which best fits your needs.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Print Preview of Reports
The following functions are available with the print preview window:
• Browse through the report pages with the Prev and Next buttons.
• Jump to a page directly using the selection box.
• The Print button sends the report to the printer, which is displayed in the lower left corner of the print preview window.
• The Close button closes your print preview window and discards the report.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 87
5 Using your UV-Visible Spectroscopy System
Finding the Caffeine Absorbance Maximum
Finding the Caffeine Absorbance Maximum
This section describes how you find the absorbance maximum for your caffeine IQ sample.
1 Make sure that you are in the Standard mode. The mode is indicated on the tool bar of your Agilent ChemStation session.
2 Select the Spectrum/Peaks task in the analysis panel’s selection box.
3 Use the Setup button of the analysis panel to open the parameter dialog.
88 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Finding the Caffeine Absorbance Maximum
4 Type 2 for the number of peaks to find and uncheck the Valley find option.
Set your data type to absorbance, and adjust your spectral display to a wavelength range from 190 nm to 400 nm in the dialog’s Display spectrum section. Click OK to set your parameters.
5 Fill your 1-cm path-length quartz cell with distilled water. Lift the lever at the left side of your cell holder. Put the cell in the cell holder and make sure the transparent windows face towards the front and the back of the spectrophotometer. Push the lever down to secure your cell in the cell holder.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 89
5 Using your UV-Visible Spectroscopy System
Finding the Caffeine Absorbance Maximum
6 Press the Blank button on the front of the spectrophotometer or click Blank on the Instrument Panel to start the measurement.
7 Remember the orientation of the cell in the cell holder. Lift the level to release the cell, remove it and flush it three times with about 1 ml of your caffeine sample. Then fill the cell with about 3 ml of the caffeine sample.
Make sure that the cell windows are clean, and reposition the cell in the same orientation as for the reference measurement. Close the cell holder’s level.
8 Press the Sample button on the front of your spectrophotometer or click
Sample on the Instrument Panel to start the measurement.
90 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Finding the Caffeine Absorbance Maximum
9 The view shows you the spectrum of your caffeine sample. Two peaks were found and are annotated with the wavelength. Below the spectrum graph the Sample/Result Table shows the wavelength of the peaks found and the measured absorbance values.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 91
5 Using your UV-Visible Spectroscopy System
Entering your Cell’s Path Length
Entering your Cell’s Path Length
The cells used for your measurements are specified with the sampling system parameters. In quantitative calculations, these parameters are used in results calculations. You must provide the correct value for the path length setting.
Usually you get this information from the supplier of your cells. You set the path length in manual mode cell handling as follows.
1 Click Setup on the Instrument Panel.
2 Type the path length in cm in the Setup Manual dialog box.
92
3 Click OK to set the specified path length.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Controlling your Sipper System
Controlling your Sipper System
A sipper system transfers your sample by means of a peristaltic pump into a flow cell for the measurement. To control your sipper system through the
Agilent ChemStation software, you must adjust your current sampling system for sipper introduction.
In addition, due to the length of tubing, the dead volume of your flow cell and the flow rate of your pump, you must adjust your sipper system parameters.
For details see the Installing and Operating Your Sipper System manual.
1 Select the Sipper in the Instrument Panel selection box.
2 Click Setup on the Instrument Panel. Type the path length of your flow cell in cm and click OK.
3 Click Setup again, and access the Sipper Parameter dialog by clicking
Parameter. The parameters required can be determined using the Flow Test task of your Verification and Diagnostics mode.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 93
5 Using your UV-Visible Spectroscopy System
Controlling your Sipper System
NOTE
4 Click OK in the Sipper Parameter dialog and click OK in the Sipper dialog to set the parameters.
Every measurement you start by clicking one of the Instrument Panel measurement buttons or by pressing the spectrophotometer buttons uses the sipper for sample introduction.
Sipper introduction is also used by an automated sequence.
94 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Using your Multicell Transport
Using your Multicell Transport
NOTE
The multicell transport is a cell changer which allows you to position up to 8 cells in the measurement position automatically. You may use different cells in each measurement position. The path length can be specified individually for each of the cell positions.
For details of your multicell transport see the Installing and Operating Your Multicell
Transport manual.
1 Select Multicell (8-cell) in the Instrument Panel selection box.
2 Click Setup on the Instrument Panel. Type the path lengths of all cells used in cm and click OK.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 95
5 Using your UV-Visible Spectroscopy System
Using your Multicell Transport
3 To move your cell for the next measurement into the measurement position, click the cell on the Instrument Panel or choose Multicell Transport
Position from the Spectrophotometer menu to access the Multicell transport Control dialog. In the Multicell Transport Control, press one of the position buttons to move the multicell transport.
96 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Using your Multicell Transport
NOTE
In an automated sequence, the multicell transport can be used for automatic sample introduction. A maximum of 8 samples can be introduced. If you specify more than 8 measurements this warning is displayed:
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 97
5 Using your UV-Visible Spectroscopy System
Quantitative Analysis using a Calibration with Standards
Quantitative Analysis using a Calibration with Standards
Your quantitative analysis task is based on a calibration with standards. After a successful calibration the measured standards can become part of your method. Such a method can be used directly for quantitative analysis of samples.
After having set up your method, calibrated samples can be analyzed. Several views of both your standards and the calibration, as well as of your samples and results are available.
As a quick introduction, a calibration using Beer’s law with a single standard and the analysis of a sample is described. The only limitation on the number of samples and standards is the memory capacity of your Agilent ChemStation.
For the practical experiment we use the IQ caffeine sample as standard and a
1:1 dilution with distilled water as sample. For the calibration we use absorbance data at 273 nm.
98 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Quantitative Analysis using a Calibration with Standards
Setup
1 Make sure that you are in Standard mode. The mode is indicated on the tool bar of your Agilent ChemStation session.
2 Select the ‘Quantification’ task in the analysis panel’s selection box.
3 A new task panel is displayed and the Quantification Parameters dialog is opened automatically.
NOTE
If you are already in the Quantification task, use the Setup button on the analysis panel to open the parameter dialog.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 99
5 Using your UV-Visible Spectroscopy System
Quantitative Analysis using a Calibration with Standards
4 Setup your analysis wavelength at 273 nm (Use wavelength), type Caffeine for the Analyte name, set the Calibration curve type to Linear, select
Concentration entry and use mg/l as Unit. Check the Prompt for standard information and the Prompt for sample information. Select Absorbance as
Data type and set Display spectrum From 190 nm To 340 nm.
NOTE
5 Click OK to set your parameters.
Now you are ready to run your measurements.
100 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Quantitative Analysis using a Calibration with Standards
Calibration
1 Fill your 1-cm path-length quartz cell with distilled water. Lift the lever at the left side of your cell holder. Put the cell in the cell holder and make sure the transparent windows face towards the front and the back of the spectrophotometer. Push the lever down to secure your cell in the cell holder.
2 Press the Blank button on the front of the spectrophotometer or click Blank on the Instrument Panel to start the measurement.
3 Remember the orientation of the cell in the cell holder. Lift the level to release the cell, remove it and flush it three times with about 1 ml of your caffeine sample. Then fill the cell with about 3 ml of the caffeine sample.
Make sure that the cell windows are clean and reposition the cell in the same orientation as for the reference measurement. Close the cell holder’s level.
4 Press the Standard button on the front of your spectrophotometer, or click
Standard on the Instrument Panel to start the measurement.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 101
5 Using your UV-Visible Spectroscopy System
Quantitative Analysis using a Calibration with Standards
5 Enter your standard’s information in the Standard Information dialog box and click OK.
Your Agilent ChemStation software automatically calibrates and displays the calibration results. After a successful calibration the task panel’s calibration curve shows green. This indicates that your method is ready for analysis.
102
NOTE
The Task panel’s Samples and Standards buttons can be used to switch your current view to the samples or standards view.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
Quantitative Analysis using a Calibration with Standards
Analysis
1 Remember the orientation of the cell in the cell holder. Lift the level to release the cell, remove it and flush it three times with about 1 ml of your caffeine sample (1:1 dilution with distilled water). Then fill the cell with about 3 ml of the caffeine sample. Make sure that the cell windows are clean and reposition the cell in the same orientation as for the standard measurement. Close the cell holder’s level.
2 Press the Sample button on the front of your spectrophotometer or click
Sample on the Instrument Panel to start the measurement.
3 Enter your sample’s information in the Sample Information dialog box and click OK. The view switches to the samples and your quantitative results will be displayed with the Sample/Result table.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 103
5 Using your UV-Visible Spectroscopy System
Quantitative Analysis using a Calibration with Standards
NOTE
To save your data for future use or for documentation purposes, see “Saving and Retrieving
Data" on page 79 for more information.
104 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
How Can I Be Sure That My Spectrophotometer Works Properly?
How Can I Be Sure That My Spectrophotometer Works
Properly?
The quality of your measurement data is dependent on the performance of your spectrophotometer. For a full performance verification external standards are required. Standard kits are available for operational qualification and performance verification (OQ/PV). OQ/PV kits for the
Agilent Cary 8454 UV-Visible spectrophotometer are available as spare parts.
A quick check without the need of standards can be performed by the
Verification and Diagnostics mode’s self test. This test can be performed always after starting the spectrophotometer.
Spectrophotometer Self test
1 Make sure that you are in the Verification and Diagnostics mode. The mode is indicated on the tool bar of your Agilent ChemStation session.
2 Select the Self-Test task in the analysis panel’s selection box.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 105
5 Using your UV-Visible Spectroscopy System
How Can I Be Sure That My Spectrophotometer Works Properly?
3 Choose Self-Test > Start from the Task menu or click Start to start the self-test.
NOTE
The spectrophotometer should be in stable working conditions before you initiate the test.
If these conditions are not met, you may get a warning message.
106
NOTE
4 The self-test results are displayed with pass/fail criteria.
Self-test results can be stored with the spectrophotometer. Storing the self-test results with the spectrophotometer allows you to monitor the performance over time. Graphical representations of the self-test histories can be generated.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
What Maintenance Does My Spectrophotometer Require?
What Maintenance Does My Spectrophotometer Require?
Cleaning
Any spills should be wiped up immediately.
The exterior surfaces of the spectrophotometer should be kept clean. All cleaning should be done with a soft cloth. If necessary, this cloth can be dampened with water or with a mild detergent. Do not use organic solvents or abrasive cleaning products.
Spare Parts
For spare parts and consumables ordering information, refer to the Agilent
Technologies website at www.agilent.com.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 107
5 Using your UV-Visible Spectroscopy System
How Can I Get a Deeper Understanding of UV-Visible Spectroscopy?
How Can I Get a Deeper Understanding of UV-Visible
Spectroscopy?
The basic principles of UV-Visible spectroscopy are covered with your help system. This covers Basic Principles through to details of Derivative
Spectroscopy.
108
In addition, solutions to selected topics in UV-Visible analysis are described.
There you can find help on topics such as Enhancing Sensitivity or
Determining Purity.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Using your UV-Visible Spectroscopy System 5
How Can I Get a Deeper Understanding of UV-Visible Spectroscopy?
More specific examples, including data, are available in the Examples section of the Standard Mode help. These can be used to run your
Agilent ChemStation software for the special purpose described.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 109
5 Using your UV-Visible Spectroscopy System
When Do I Have to Measure a Blank?
When Do I Have to Measure a Blank?
The measurement data acquired by your spectrophotometer is instrument independent. To achieve this independence, a reference measurement must be performed. All consecutive measurements refer to the last measured reference.
In the Agilent ChemStation software, the reference measurement is combined with a baseline measurement. The baseline displayed is information about the quality of the current reference. In absorbance mode, the data should be close to 0 AU and in transmission mode the data should be close to 100%.
Typically, reference measurements for dissolved samples are made on the cell filled with the solvent used. The absorbance properties of the cell and the solvent also influence the reference data. In wavelength ranges where your solvent or cell is absorbing, the noise of the baseline spectrum is high. Reliable sample data cannot be expected in these areas.
Therefore new reference or blank measurements are required when:
• you change your measurement cell or its orientation relative to the measurement position.
• you use a different solvent or event a different lot of the same solvent.
• the time between your reference measurement and the sample measurement get too long.
• the time conditions are depending on the aging rate of your lamps and possible changes of your environmental conditions—typically the last blank measurement should not be older than half an hour.
To ensure the highest accuracy of sample measurement, the blank measurement should be taken immediately prior to the actual sample measurement.
110 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Index
A absorbance
high
absorbance maximum
accessing lamps
accurate results
acetic acid
acetone
acetonitrile
acquisition date
time
active item
active position
Agilent Bootp Service
Agilent Cary 8454 UV-Visible
Agilent ChemStation analysis panel
family
graphical window
log on
measurement session
method
mode
offline session
online session
password
preview sizes
print preview
printer display
samples
session
session online
workstation
ambient light
analysis
analysis panel
analysis setup calibration
concentration unit
data type
display
prompt for sample information
wavelength
analytes
analytical task
analyze
apertured cells or cuvettes
apertures
application specific
B background correction
Beer’s law
benzene
blank
blank push button
bootp service
bubble formation
busy
C caffeine
calculate
calculation results
calibration
coefficients
curve
carbon disulfide
carbon tetrachloride
CE compliance
cell changer
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual cell path length
chloroform
cleaning
cleaning cells
collimated beam
colloidal dispersions
command
common solvents
compartment for sample
concentration
concentration range
connection network
connector
GPIO
multicell transport
remote
RS232
correction for stray light
current context
current task
cut-off filter
cut-off wavelength
cyclohexane
cyclopentane
111
Index
D data
absorbance
access
archieve
clear
clear math. results
clear standards
evaluation
file extension
file name
file selector box
format
load
local storage
network transfer
removing
replace
retrieve
save
save samples as
saving selected data
storage
degassed
derivative
derivative spectroscopy
description of instrument
determining purity
deuterium lamp
development of an analytical method
dialog fixed wavelength(s) parameters
method options & information
dimethyl formamide
dimethyl sulfoxide
diode array
door to access lamps
dust
E enhancing sensitivity
entrapment of bubbles
Environmental Conditions and Power
Requirements
equation
ethyl acetate
ethyl ether
examples
extinction coefficient
F false light
filter for stray-light correction
fixed wavelength
flow cell
flow cells
flow test
front view of spectrophotometer
G glass cells
glycerol
GPIO connector
grating
H handling cells
highest precision measurements
holographic grating
homogeneity
HPLC grade water
I
ICES/NMB-001
identity
indicator
installation
installation qualification
112 instrument construction
description
electronic assemblies
layout
mechanical assemblies
warm up
instrument panel
instrument session
internal reference
IP address
isopropyl alcohol
L lamps
access through door
deuterium
LAN tungsten
crossover cable
LAN interface
Jet Direct
Talk2Lab
lens
lens tissues
levels of operation
light beam
line power
input socket
switch
liquid samples
logged
M main application window
maintenance
making measurements
manager level
maximum
measure standard
standard information
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Index measure push buttons
measurement blank
noise
reference
sample
sample information
menu
message line
methanol
method
calibrated
current
information
last used
load
load method
modified
name
options & information
parameter
parameters
print preview
printing
report
retrieve
save method as
store
methyl formate
minimum
mode
advanced
color calculations
combined reports
dissolution testing
execute advanced method
kinetics
last used
multibath dissolution testing
standard
switch
thermal denaturation
verification and diagnostics
mouse cursor
multicell transport
8-cell
connector
m-xylene
N
N55011/CISPR11
n-butyl alcohol
network administrator
connection
local
n-hexane
nominal spectral slit width
O offline
online
open sample area
operating system
operation
operation levels
operator level
operator name
optical filter
optical specifications of cells
optical surfaces
optical system
optimization
outliers
P paper orientation
size
parallelism
parameter dialog
particulate matter
passivating new cells
path length setup
PC
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual peak find
performance verification
photochemical reactions
photodegradation
photodiode array
photosensitive substances
pipette
plasma discharge
plastic door
plastic sample cells
poor linearity
poor photometric accuracy
Precautions
printer
configured
processed spectra
processing
spectral
standards
pump peristaltic
purity
push buttons
blank
sample
standard
stop
pyridine
Q quantification
quantitative analysis of samples
ready for analysis
quartz sample cells
R radiation source
rapid absorbance changes
ratio
raw data
113
Index rear view of spectrophotometer
recalculation
recommended cells
reference
Regulatory Information
remote connector
report results
result
results
precise
routine work
RS232C connector
S sample
compartment
push button
sample cell
sample data
sample information
sample/result table
sampling device
sampling interval
sampling system
manual
security lever
self test
histories
start
working conditions
sensitivity
session data analysis only
instrument control
set of parameters
settings
setup dialog
sheet-metal door
shine-through aperture
shutter
114 signal to noise ratio
single beam instrument
sipper flow test
parameter
sipper system
sipper/sampler system
slit
slit width
software general purpose
solute convection
solution
solution noise
solutions
determining purity
enhancing sensitivity
solvent
Solvent Information
solvent suitability
Solvents
solvents
source lens
source of radiation
spare parts
spectral acquisition
spectral operation
spectral processing
spectral raw data
spectrograph
lens
slit
spectrophotometer
front view
rear view
spectroscopy system
spectrum
standard
external
standard cells
standard push button
standard single-cell cell holder
standards
current
minimum required number
number of
status
status indicator
stirring
stirring module
stop push button
stoppered sample cell
stray-light correction
sulfuric acid
T task fixed wavelength
orientation
quantification
quantitative analysis
ratio/equation
spectrum/peaks
TCP/IP protocol
temperature control
test kit
thermostattable cell holder
three point drop line
toluene
tool bar
transmittance
trimethylpentane
2,2,4-trimethylpentane
tungsten lamp
tutorial basic principles
derivative spectroscopy
U understanding
Agilent ChemStation processing
used wavelength
useful wavelength range of solvents
user definable equation
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
Index user interface elements
UV grade water
UV-Visible spectroscopy basics
UV-Visible system
V valley find
view
calibration
results
samples
standards
viscous solutions
volatile solvents
volume
W warning no results present!
Warning Symbols
warning symbols
water
wavelength
wavelength reproducibility
wedge shaped cells
weight
width of slit
window
sample/results table
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 115
Index
116 Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual
www.agilent.com
In This Book
To be able to use your new
Agilent Cary 8454 UV-Visible spectroscopy system quickly, this book gives you step-wise procedures and examples for basic operations and tasks.
Agilent Technologies 2002, 2003-2008, 2011,
2013, 2014, 2016
09/16