HIGHLIGHTS DATA SHEET VENTILATION SERVO-i

HIGHLIGHTS DATA SHEET VENTILATION SERVO-i
DATA SHEET
VENTILATION
®
SERVO-i UNIVERSAL
HIGHLIGHTS
®
For all patient categories - available in two
editions
NAVA - improved synchrony and unique
monitoring capabilities
Intuitive user interface
For invasive as well as non invasive ventilation
Designed for cost-efficiency
The Open Lung Tool and Stress Index - tools
for lung protection.
Flexible placement
Modular – interchangeable plug-in units
Enables Heliox delivery
®
Available in MR and Transport editions for
continuity of ventilatory care
| 2 | SERVO-i ® UNIVERSAL | Critical Care |
SERVO-i VENTILATOR FAMILY
®
SERVO-i UNIVERSAL
UBE
SERVO-i Configurations
UEE
Stress Index
®
The SERVO-i Ventilator family consists of four configurations:
®
NIV NAVA
®
Infant
Adult
UBE
UEE
Universal Basic Edition
Universal Extended Edition
Standard Configuration
Options
Not applicable
NAVA
Heliox
Nasal CPAP
NIV PC
NIV PS
Bi-Vent
Y Sensor Measuring
CO2 Analyzer
Nebulizer
Alarm output connector
®
The SERVO-i Ventilator family addresses the very different
Open Lung Tool
requirements of neonatal, pediatric and adult needs from a single
Automode
SIMV (PRVC) + PS
PRVC
VS
SIMV (VC) + PS
VC
SIMV (PC) + PS
PC
PS/CPAP
All patient category software
ventilation platform. All four configurations are the same ventilator,
equipped with different functions. They can be customized and
upgraded with different options for future needs. (SERVO-i Adult
and SERVO-i Infant can also be upgraded to SERVO-i Universal.)
The same ventilator can be used at the bedside, during transport*
and in the MR-room* facilitating training, operation and
maintenance, increasing efficiency and flexibility.
SERVO-i Universal is an intensive care ventilator for all patient
®
categories and is available in two editions - Basic and Extended,
the latter featuring a wide range of modes and functions as
standard.
*An agreement with MAQUET must be signed, see conditions in
the SERVO-i MR declaration (Order no. 66 71 670) and the
Interhospital Transport Declaration (Order no. 66 64 721)
respectively.
KEY TO ABBREVIATIONS
NAVA
NIV
SIMV
PRVC
VS
VC
PS
PC
CPAP
Neurally Adjusted Ventilatory Assist
Non-invasive ventilation
Synchronized Intermittent Mandatory Ventilation
Pressure Regulated Volume Control
Volume Support
Volume Control
Pressure Support
Pressure Control
Continuous Positive Airway Pressure
| Critical Care | SERVO-i
®
UNIVERSAL | 3 |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
The system - General
Non-operating conditions
The device complies with
requirements of Medical Device
Directive 93/42/EEC.
Impact:
Peak acceleration: 15 g.
Pulse duration: 6 ms.
Number of impacts: 1000.
Classification:
Class I equipment. According
to IEC/EN 60 601-1.
Storage temperature:
–25 to +60º C (–13 to 140º F)
Storage Relative Humidity:
< 95%
Standards:
IEC/EN 60 601-1 (Type B).
IEC/EN 60 601-2-12.
EN 794-1.
Storage Atmospheric Pressure: 470 to 1060 hPa
Noise level:
<50 dBA, measured at 0.3 m
distance.
Power supply
IP classification
IP 20
Electromagnetic compatibility (EMC):
Emission:
According to IEC/EN 60601-1-2
(Edition 2:2001).
Immunity- Extended test to
30 v/m:
According to IEC/EN 60601-1-2.
The 'EMC Declaration, Information to the Responsible
Organization' is available from MAQUET.
Patient range:
Weight:
Power supply, automatic range 100 – 120 V AC ±10%, 50 –
selection:
60 Hz, or 220 – 240 V AC ±10%,
50 – 60 Hz.
Plug-in battery module:
Battery backup:
Two battery modules are
delivered with the ventilator. Up
to six battery modules can be
included.
Battery capacity:
Rechargable, 12 V, 3.5 Ah each.
Recharge time:
Approximately 3 h/battery.
Battery backup time:
At least 3 h, when using six
batteries.
Adult, all ventilation modes:
10 - 250 kg
Infant, invasive ventilation:
0.5 - 30 kg
External 12 V DC:
12.0 V – 15.0 V DC, 10 A
Infant, NAVA + NIV NAVA:
0.5 - 30 kg
Max power consumption:
Infant, NIV PS+PC:
3 - 30 kg
Infant, Nasal CPAP:
0.5 - 10 kg
At 100 – 120 V: 2 A, 190 VA,
140 W.
At 220 – 240 V: 1 A, 190 VA,
140 W.
Operating conditions
Operating temperature:
+10 to +40ºC
Relative humidity:
15 to 95% non-condensing
Atmospheric pressure:
660 to 1060 hPa
Lowest pressure in breathing
system:
–400 cmH2O
| 4 | SERVO-i ® UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
The ventilator - General
Dimensions:
User interface
(See dimensional drawings page
15)
User Interface:
W 355 x D 53 x H 295 mm
Patient Unit:
W 300 x D 205 x H 415 mm
Weight:
Approximately 5 kg
Attachment:
Can be attached to the mobile
cart, a table, rail or pole (15 –
30 mm diameter).
Weight:
Approximately 20 kg (Patient
Unit 15 kg, User Interface 5 kg)
Method of triggering:
Flow, pressure and Edi
(optional)
Type:
TFT-LCD module
Size:
31 cm (12.1”) diagonal
Approximately 115 cmH2O
Viewing area:
246.0 x 184.5 mm
Max. operating pressure:
Screen
Bias flow:
Adult:
2 l/min
Infant:
0.5 l/min
Gas supply
Inlet gas pressure Air/O2:
Pressure drop:
Max. 6 cmH2O at a flow of 1 l/s
Internal compressible factor:
Max. 0.1 ml/cmH2O
Gas delivery system:
Microprocessor controlled valve
Inspiratory flow range:
200 – 650 kPa / 2.0 – 6.5 bar /
29 – 94 PSI
Connection standards available: AGA, DISS, NIST, or French
standard.
Unavailable gas/loss of gas
pressure:
Inspiratory channel
The flow from an unavailable
gas (air or O2) is automatically
compensated for so that the
patient gets the preset volume
and pressure.
Adult:
0 to 3.3 l/s
Infant:
0 to 0.55 l/s
Expiratory channel
Pressure drop:
Max. 3 cmH2O at a flow of 1 l/s
Internal compressible factor:
Max. 0.1 ml/cmH2O
PEEP regulation:
Microprocessor controlled valve
Patient system gas connectors
Rise time, expiratory flow
measurement:
<12 ms for 10 – 90 % response
at flow of 0.05 – 3.2 l/s
Conical fittings:
Male 22 mm / female 15 mm. In
accordance with ISO 5356-1.
Expiratory flow range:
0 to 3.2 l/s
Gas exhaust port:
Male 30 mm cone
| Critical Care | SERVO-i
®
UNIVERSAL | 5 |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Alarms
Alarms
Airway pressure (upper):
Battery:
Adult, invasive ventilation:
16 - 120 cmH2O
Adult, Non Invasive
Ventilation:
16 - 40 cmH2O
Infant, Invasive ventilation:
16 - 90 cmH2O
Infant, Non Invasive
Ventilation
16 - 40 cmH2O
Expired minute volume (Upper
alarm limit):
Adult:
0.5 – 60 l/min
Infant:
0.01 – 30 l/min
Expired minute volume (Lower
alarm limit):
Adult:
0.5 – 40 l/min
Infant:
0.01 – 20 l/min
Limited battery capacity:
10 min.
No battery capacity: less than
3 min.
Low battery voltage.
End-tidal CO2 (upper and lower 0.5 – 20%.
4 – 100 mm Hg.
limit):
0.5 – 14 kPa.
Leakage out of range in NIV:
Yes. Described in the User’s
manual.
Technical:
Yes. Described in the User’s
manual.
Autoset (alarm limits)
specification:
Invasive ventilation, controlled
modes only.
High airway pressure:
Mean peak pressure
+10 cmH2O or at least
35 cmH2O.
Upper minute volume:
It is possible to permanently silence this alarm (an option with
Infant version only)
Expiratory minute volume
+ 50%.
Lower minute volume:
Expiratory minute volume
– 50%.
No patient effort (Apnea) alarm
Upper respiratory frequency: Breathing frequency + 40%.
Adult:
15 – 45 s
Lower respiratory frequency: Breathing frequency – 40%.
Infant
2 – 45 s
Automatic return to support mode on patient triggering
High end expiratory pressure: Mean end expiratory pressure
+ 5 cmH2O.
No consistent patient effort:
Yes, described in User's manual
Low end expiratory pressure: Mean end expiratory pressure
– 3 cmH2O.
Respiratory frequency:
1 – 160 breaths/min
High end expiratory pressure:
0 – 55 cmH2O
Low end expiratory pressure:
0 – 47 cmH2O.
Note. Setting the alarm to 0
(zero) is equal to alarm off.
High continuous pressure:
Set PEEP level + 15 cmH2O
exceeded for more than 15
seconds.
O2 concentration:
Set value ±5 vol% or 18 vol%
Gas supply:
Below 200 kPa / 2.0 bar /
29 PSI and above 650 kPa /
6.5 bar / 94 PSI
Upper end tidal carbon
End tidal carbon dioxide
dioxide concentration (etCO2): concentration + 25%.
Lower end tidal carbon
End tidal carbon dioxide
dioxide concentration (etCO2): concentration – 25%.
| 6 | SERVO-i ® UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Ventilation Modes – Invasive ventilation
Ventilation modes – Non-invasive Ventilation (optional)
Controlled ventilation:
NIV PC
PC
VC
PRVC
NIV PS
Can be configured with
alternative flow patterns
- VC with flow adaptation,
- VC without flow adaptation,
- VC with decelerating flow
Optional with Universal Basic
Edition
Nasal CPAP
Ventilation modes - NAVA (optional)
NAVA
Assist via endotracheal tube or
tracheostomy
Supported ventilation:
NIV NAVA
PS/CPAP
VS
SIMV (VC) + PS:
Comes with the corresponding
controlled ventilation mode
SIMV (PC) + PS:
Comes with the corresponding
controlled ventilation mode
SIMV (PRVC) + PS
Comes with the corresponding
controlled ventilation mode
(optional with Universal Basic
Edition)
Automode
Pressure controlled ventilation
on two independently adjustable
levels, allowing unrestricted
spontaneous breathing on both
levels. (Optional)
Control mode: VC <—> Support
mode: VS
Control mode: PC <—> Support
mode: PS
Control mode: PRVC <—>
Support mode: VS
Optional with Universal Basic
Edition
Neurally Adjusted Ventilatory
Assist via non-invasive patient
Optional with Universal Basic
Edition
Combined ventilation:
Bi-Vent
Neurally Adjusted Ventilatory
interfaces
Waveform and loop presentations
Real time waveforms - up to 4 waveforms can be displayed
simultaneously:
Pressure
Flow
Volume
CO2
Requires SERVO-i CO2 Analyzer
option
Edi
Requires SERVO-i NAVA option
Loops:
Volume / Pressure*
*A reference loop and three
overlaying loops can be
displayed.
Flow / Volume*
*Displayed simultaneously with
Open Lung Tool graphical
trends, if requested.
| Critical Care | SERVO-i
®
UNIVERSAL | 7 |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Displayed
value
Yes
Trended
value*
Yes
Monitoring
End Expiratory Flow:
Displayed
value
Yes
Trended
value*
Yes
Spontaneous breaths per
minute (RRsp):
No
Yes
Static Compliance:
Yes
Yes
Dynamic Compliance:
Yes
Yes
Peak Airway Pressure:
Yes
Yes
Stress Index
Yes
Yes
Mean Airway Pressure:
Yes
Yes
Inspiratory Resistance
Yes
Yes
Pause Airway Pressure:
Yes
Yes
Expiratory Resistance
Yes
Yes
End Expiratory Pressure:
Yes
Yes
Elastance
Yes
Yes
CPAP Pressure:
Yes
Yes
Time Constant
Yes
No
Inspired Tidal Volume:
Yes
Yes
P0.1:
Yes
Yes
Expired Tidal Volume:
Yes
Yes
Work of Breathing patient:
Yes
Yes
Inspired Minute Volume:
Yes
Yes
Work of Breathing ventilator
Yes
Yes
Expired Minute Volume:
Yes
Yes
Shallow Breathing Index (SBI)
Yes
Yes
Leakage fraction in NIV (%):
Yes
Yes
Supply pressure (O2 and air):
Yes
No
Ti/Ttot:
Yes
No
Battery remaining time:
Yes
No
I:E ratio:
Yes
No
Barometric pressure:
Yes
No
Total PEEP:
Yes
No
Edi peak:
Yes
Yes
Edi min:
Yes
Yes
O2 Concentration (measured):
Yes
Yes
CO2 End tidal concentration
(etCO2):
Yes
Yes
CO2 Minute elimination (CO2):
Yes
Yes
Tidal CO2 elimination (VTCO2): Yes
Yes
Monitoring
Breathing frequency:
MVe sp / MVe:
Yes
No
Spontaneous Exp. Minute
Volume (MVe sp):
Yes
Yes
*Stored trend values for up to 24 hours
Log function
Event log:
Alarms.
Ventilator settings.
Apnea periods.
Immediate functions.
Service log:
Technical alarms.
Test results.
Preventive maintenance.
Service history.
Configuration log.
| 8 | SERVO-i ® UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Parameter settings
Setting range:
Parameter settings
Setting range:
Parameter:
Infant
Adult
Parameter:
Infant
Adult
Inspiratory tidal volume (ml):
Inspiratory minute volume
(l/min):
2 – 350
100 – 4000
O2 concentration (%):
21 – 100
21 – 100
0.3 – 20
0.5 – 60
I:E ratio:
1:10 – 4:1
1:10 – 4:1
TInsp (s):
0.1 – 5
Apnea, time to alarm (s):
5 – 45
0.1 – 5
15 – 45
Automode Trigger timeout (s):
3 – 15
NAVA level (cmH2O/µV):
0 - 15
0 - 15
7 – 12
Edi trigger sensitivity (µV):
0.1 – 2.0
PC/PS above PEEP (cmH2O)
0 – (80 -PEEP) 0 – (120
-PEEP)
0.1 – 2.0
TPause (s):
PC/PS above PEEP in NIV
(cmH2O)
0 – (32 - PEEP) 0 – (32 - PEEP)
PEEP (cmH2O)
0 – 50
0 – 50
PEEP in NIV (cmH2O)
2 – 20
2 – 20
CPAP pressure (cmH2O):
2 – 20
CMV frequency (breaths/min):
0 – 1.5
0 – 1.5
TPause (% of breath cycle time): 0 – 30
0 – 30
Flow trigger sensitivity level
(fraction of bias flow):
0 – 100%
0 – 100%
Press. trigg sensitivity (cmH2O): -20 – 0
-20 – 0
–
Insp. rise time (% of breath
cycle time):
0 – 20
0 – 20
4 – 150
4 – 150
Insp. rise time (s):
0 – 0.2
0 – 0.4
SIMV frequency (breaths/min):
1 – 60
1 – 60
Insp. cycle off (% of peak flow): 1 – 70
1 – 70
Breath cycle time, SIMV (s):
0.5 – 15
1 – 15
(PEEP + 1) – 50 (PEEP + 1) – 50
Insp. cycle off in NIV (% of peak 10 – 70
flow):
10 – 70
PHigh (cmH2O)
THigh (s):
0.2 – 10
0.2 – 10
Nebulizer time (min):
5 – 30
5 – 30
TPEEP (s):
0.2 – 10
0.2 – 10
PS above PHigh (cmH2O):
0 – (80 - PHigh) 0 – (120 PHigh)
Backup settings
Setting range:
Parameter:
Infant
Adult
Inspiratory tidal volume (ml):
2 – 350
100 – 4000
PC above PEEP (cmH2O)
5 – (80 -PEEP) 5 – (120
-PEEP)
PC above PEEP in NIV
5 – (32 -PEEP) 5 – (32 -PEEP)
CMV frequency (breaths/min):
4 – 150
4 – 150
I:E ratio:
1:10 – 4:1
1:10 – 4:1
TInsp (s):
0.1 – 5
0.1 – 5
| Critical Care | SERVO-i
®
UNIVERSAL | 9 |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Parameter:
Setting range
Communication/Interface
Oxygen breaths:
100% for 1 minute
Start breath:
Initiation of 1 breath (In SIMV
mode initiation of 1 mandatory
breath)
Screen picture transfer
(optional)
Pause hold:
Insp. or exp (0 – 30 seconds)
Alarm silence/reset:
2 minute silence and reset of
latched alarms
Compliance compensation:
On/Off
Adult
65 l/min
Automode (optional with
Universal Basic Edition):
Automode On/Off
Infant, NIV:
25 l/min
Infant, Nasal CPAP:
15 l/min
Nebulizer (optional)
5 - 30 min./Continuous/Off
Backup ventilation
Backup On/Off
Suction Support
Pre oxygenation time:
Max. 2 min
Post oxygenation time:
Max. 1 min
Suction phase time:
No maximum level
Adjustable oxygen level:
21 – 100 %
Saving of data
Recording of current waveform 20 seconds of data will be
and parameter values:
recorded (10 seconds before
and 10 seconds after
activation).
Communication/Interface
Serial port:
RS-232C - isolated. For data
communication via the
Communication Interface
Emulator (CIE).
Second serial port (optional)
See information above.
Alarm output connector (optional)
Connector:
4-pole Modular connector
Ratings:
Max 40 V DC, Max 500 mA,
Max 20 W
Network connection (optional): MIB (Medical Information Bus)
monitor connection
Data transfer (optional):
Via Ventilation Record Card
Via Ventilation Record Card
Non-invasive Ventilation (optional)
Max. leakage compensation
level
Leakage overrange detection:
Automatic
Disconnect detection:
Automatic
Disconnect flow:
Configurable
Low
7.5 l/min
High
- 40 l/min (Adult)
- 15 l/min (Infant)
Disabled
Deactivates disconnect
detection
Connect detection:
Manual, or automatic via bias
flow
| 10 | SERVO-i ® UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Open Lung Tool (OLT) (optional with Universal Basic Edition)
Three simultaneous graphical
trends, presented
breath-by-breath:
1.
2.
3.
EIP and PEEP (End
Inspiratory Pressure and
Positive End Expiratory
pressure).
VTi and VTe (Inspiratory
and Expiratory Tidal
volume).
Stress Index (optional)
Presented values:
PEEP (Positive End Expiratory
pressure).
VTe (Expiratory Tidal volume).
Stored values:
C dyn i and VTCO2*
(Dynamic inspiratory
Compliance [= VTi/(EIP – PEEP)]
and Tidal CO2 elimination*).
Cursor function:
Zoom function:
Time marking:
A cursor can be moved with the
main rotary dial or via the touch
screen. When moving it along
the graph, the numeric
parameter values valid for that
actual moment will be shown.
Time resolution of the x-axis
can be selected in five different
steps.
Hours and minutes (when values
are measured).
120 values stored in graph
Stored trend values up to 24
hours
Cursor function:
A cursor can be moved with the
main rotary dial. When moving
it along the graph, the numeric
parameter values valid for that
actual moment will be shown.
Time marking:
Hours and minutes (when values
are measured).
* Requires SERVO-i CO2
Analyzer option.
Values stored breath by breath: Up to 21.600 breaths.
SI (Stress Index)
NAVA (optional)
Standards:
IEC/EN 60601-1
(Type CF, defibrillator proof)
Size:
Edi module:
154 x 90 x 21 mm
Edi catheter cable:
Length: 2 m
Weight - Edi module:
0.25 kg
Power source - Edi module
supply voltage:
Powered from SERVO-i. <3 W
at 12 V (normal operation)
Parameters:
Edi waveform
Edi leads waveforms
NAVA estimated pressure
waveform (Pest)
Edi catheters:
6 Fr, length 49 cm
6 Fr, length 50 cm
8 Fr, length 100 cm
8 Fr, length 125 cm
12 Fr, length 125 cm
16 Fr, length 125 cm
| Critical Care | SERVO-i
®
UNIVERSAL | 11 |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
SERVO-i CO2 Analyzer (optional)
SERVO-i CO2 Analyzer – Performance
Standard:
Measuring method:
Mainstream, dual-wavelength,
non-dispersive infrared.
Parameters:
Capnogram.
CO2 End tidal concentration
(etCO2).
EN 864, ISO 9918.
IEC/EN 60601-1
(Type BF, defibrillator proof)
Size:
CO2 Analyzer Module:
154 x 90 x 43 mm
Sensor:
32.0 x 42.4 x 21.6 mm
CO2 Minute elimination ( CO2).
Tidal CO2 elimination (VTCO2).
Measuring range:
0 to 100 mm Hg CO2 partial
pressure.
0 to 13.3 kPa CO2 partial
pressure.
0 to 13.2% CO2 volume (at a
barometric pressure of
1013 hPa).
Step response time:
<25 ms (10 to 90% step
response).
Warm-up time:
30 s to initial CO2 indication,
max. 5 min to full specification.
Oxygen concentration
compensation:
Automatic. Values supplied from
the SERVO-i Ventilator System.
CO2 Analyzer Module supply Powered from the SERVO-i.
voltage:
Barometric pressure
compensation:
Automatic. Values supplied from
the SERVO-i Ventilator System.
Power consumption:
Digitizing rate:
87 Hz
Weight:
CO2 Analyzer Module:
0.45 kg
Sensor:
18 g
Airway adapter:
10 g
Connectors and cables:
CO2 Analyzer Module:
15-pole D-sub female
connector.
Sensor:
20-pole, 2.4 m cable.
Power source:
Sensor:
8 W at 12 V, during warm up.
6.5 W at 12 V, during normal
operation.
Powered from the CO2 Analyzer
Module.
Airway adapter dead space:
Adult:
<5 ml
Infant:
<0.5 ml
| 12 | SERVO-i ® UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Y Sensor measuring (optional)
Servo Ultra Nebulizer (optional)
Size:
Y Sensor Module:
154 x 90 x 43 mm
Y sensor adult:
Length 84 mm
Y sensor infant:
Length 51 mm
Weight:
Y Sensor Module:
0.4 kg
Y sensor adult:
10.5 g
Y sensor infant:
7.5 g
Weight:
Approx. 125 g
Dimensions:
H 105 mm x L 108 mm x
W 60 mm
Connection cable length:
2.0 m
Nebulizer T-piece connections: Inlet/outlet: 22/15 mm
outer/inner diameter and 22 mm
inner diameter, ISO standard.
Pediatric patient tubes:
Adapters 22/10 mm outer
diameter and 15/10 mm outer
diameter.
Sensor material:
Makrolon polycarbonate.
Internal volume:
Tubing:
2.0 m. Medical grade PVC.
Ultrasonic generator frequency: 2.4 MHz
Power source – Y Sensor
Module supply voltage:
Powered from the SERVO-i.
<5 W at 12 V (normal operation).
Particle size (water):
Y Sensor measuring – Performance
Measuring method:
Fixed orifice, differential
pressure.
Parameters:
Airway pressure.
Airway flow.
Inspiratory and expiratory
volumes.
Infant:
Mass Median Diameter (MMD)
= approximately 4.0 µm,
measured distally in
endotracheal tube 8 mm inner
diameter.
Output from nebulizer (water) – Min. 0.1 ml water/min at 0.1 l
minimum water flux:
gas flow/s.
Min. 0.3 ml water/min at 0.5 l
gas flow/s.
Min. 0.5 ml water/min at 1.0 l
gas flow/s (Adult only).
Buffer liquid:
Sterile water
2 to 180 l/min
Max. medication temperature:
55º C (131º F)
0.125 to 40 l/min
Volume, medication cup:
Max. 10 ml
Noise level:
Max. 50 dBA, measured at
0.3 m distance.
Measuring range:
Adult:
60 ml
Airway adapter dead space:
Adult:
< 9.0 ml
Infant:
< 0.45 ml
| Critical Care | SERVO-i
®
UNIVERSAL | 13 |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Aeroneb Nebulizer Systems (optional)
SERVO-i Holder (optional)
Weight - Aeroneb module:
Weight:
3.5 kg
Dimensions:
H 352 mm x L 247 mm x
W 159 mm
(see dimensional drawing
below)
Approx. 216 g
Dimensions - Aeroneb module: H 105 mm x L 108 mm x
W 60 mm
Connection cable length:
2.0 m
Particle size (water):
1 - 5 mass median aerodynamic
diameter (MMAD)
Flow rate:
>2 ml/min
Volume - nebulizer unit
Residual volume:
SERVO-i Shelf Base (optional)
Pro - 10 ml
Weight:
1.2 kg
Solo - 6 ml
Dimensions:
H 29 mm x L 205 mm x
W 159 mm
(see dimensional drawing
below)
10 %
SERVO-i Mobile Cart (optional)
Gas cylinder restrainer (optional)
Weight:
20 kg
Max load:
Dimensions:
H 1015 mm x L 640 mm x
W 560 mm
(see dimensional drawing
below)
SERVO-i Drawer kit (optional)
Weight:
4.5 kg
Dimensions:
H 240 mm x L 210 mm x
W 300 mm
2 x 5-liter bottles
SERVO-i IV Pole (optional)
Max load (total):
6 kg
Gas trolley (optional)
Max load:
2x10 kg bottles
Docking:
Dockable to SERVO-i Mobile
Cart.
Dockable to a separate wall
clamp.
| 14 | SERVO-i ® UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Compressor Mini (optional)
Note
See separate data sheet.
For inaccuracies and more detailed technical specifications please
refer to the User’s manual.
Service
Regular maintenance:
Once every 12 months or at
least after 5000 operating hours.
ORDERING INFORMATION
SERVO-i, Ventilator and accessories: See separate information:
“SERVO-i, Version 6.1 — System Flow Chart” (Order no:
66 70 102).
| Critical Care | SERVO-i
TECHNICAL SPECIFICATIONS
®
SERVO-i UNIVERSAL
Dimensional drawings
SERVO-i on Mobile Cart
SERVO-i (patient unit) on SERVO-i Holder
SERVO-i Shelf Base
SERVO-i Holder
®
UNIVERSAL | 15 |
© Maquet Critical Care AB 2011. All rights reserved. • MAQUET reserves the right to modify the design and specifications contained herein without prior notice.
Order No. 66 70 105 • Printed in Sweden • 111116 • Rev: 12 English • Data Sheet
Maquet Critical Care AB
Röntgenvägen 2
SE-171 54 Solna, Sweden
Phone: +46 (0) 8 730 73 00
www.maquet.com
For local contact:
Please visit our website
www.maquet.com
GETINGE GROUP is a leading global provider of products and
systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. We operate under the three
brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh
focuses on patient mobility and wound management solutions.
GETINGE provides solutions for infection control within healthcare
and contamination prevention within life sciences. MAQUET
specializes in solutions, therapies and products for surgical
interventions, interventional cardiology and intensive care.