Advertisement
Advertisement
Operator’s Manual
© 2011 BUFFALO FILTER®
595 Commerce Drive
Buffalo, New York USA 14228
(716) 835-7000 or (716) 835-3414 fax
Division of Medtek Devices, Inc.
Customer Service: 800-343-2324 (U.S. Only) – www.buffalofilter.com
1-716-835-7000 (International)
For a period of one (1) year following the date of delivery, BUFFALO FILTER warrants the Viro-
Vac™ against any defects in material or workmanship. BUFFALO FILTER will repair or replace
(at BUFFALO FILTER’S option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by BUFFALO
FILTER. This warranty is void if the product is used in a manner or for purposes other than intended.
© 2011 BUFFALO FILTER®
Buffalo, New York USA 14228
(716) 835-7000 www.buffalofilter.com
Division of Medtek Devices, Inc.
Customer Service: 800-343-2324 (U.S. Only)
1-716-835-7000 (International)
The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages (if any).
Manual Number 901837
Unit Serial Number ______________________
9D93
MEDICAL EQUIPMENT WITH RESPECT TO
ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 60601-1 AND CAN/CSA C22.2 NO. 601.1
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
Buffalo Filter ViroVac™ Operator’s Manual: Page 1
Table of Contents
& List of Illustrations
Title
DESCRIPTION
1.1 Introduction
1.2 Inspection
1.3
1.4
Operational Information
Cautions and Warnings
2.0
2.1
2.2
2.3
System Controls
ViroSafe® Filter Instructions
Set-up and Operation
2.4 Specifications
3.0
3.1
MAINTENANCE
General Maintenance Information
3.2 Cleaning
3.3 Periodic Inspection
3.4 Troubleshooting
4.0
4.1
4.2
CUSTOMER
Equipment Return
Ordering Information
TERMS
Figure
1 Control 14
Buffalo Filter ViroVac™ Operator’s Manual: Page 2
System Description
Section 1.0
1.1 Introduction
BUFFALO FILTER® ViroVac™ Smoke Evacuation Systems are designed to remove smoke and noxious odors produced by surgical smoke during laser, electrosurgical, ultrasonic, argon, plasma and mechanical instrument procedures.
The ViroVac™ Smoke Evacuation Systems have been designed with a high suction, high flow rate centrifugal action pump. The ultra-quiet motor is used to draw the surgical smoke from the surgical site through the vacuum tubing and into the ViroSafe filter where the surgical smoke is processed by a series of filters. A single disposable filter is used to simplify the installation and removal during filter changes. The filter is completely enclosed to protect the healthcare personnel from potential contamination during filter changes. One Buffalo Filter ViroSafe filter contains four different stages within to capture the smoke plume.
The first stage filtration is a prefilter whose function is to trap and remove gross particulate and casual fluid.
The second stage filtration is ULPA grade (Ultra Low Penetration Air) filter whose high-tech patented (U.S. Patent #5874052) design captures particulates and micro-organisms from .1 to .2 microns at an efficiency of 99.999%.
The third stage filtration uses the highest grade virgin activated carbon, especially designed for
Buffalo Filter for the removal and adsorption of odors and toxic gases produced by burning tissues. These harmful gases may constitute a health hazard to healthcare professionals who are subjected to prolonged exposure. The activated carbon used in the ViroVac™ Smoke Evacuation
Systems preferentially removes toxic organic gases rather than water vapor and provides optimal odor removal.
The fourth stage filtration is an expanded foam used to trap activated carbon fines from migrating out of the filter.
The electronic controls on the face panel of the ViroVac™ Smoke Evacuation System has been designed “user friendly” and facilitate unit set up and operation. Please refer to Section 2.0 for
Operating Instructions.
1.2 Inspection
The ViroVac™ Smoke Evacuator has been thoroughly tested and inspected before shipment from the factory. Please check the unit before using it to insure that no damage has occurred in transit.
If damage is evident, please contact BUFFALO FILTER Customer Service at 1-800-343-2324 (US
Only) or (716) 835-7000 (International)
Buffalo Filter ViroVac™ Operator’s Manual: Page 3
In addition, please compare the accessories you receive with the standard accessories list below. If an item is missing, please notify BUFFALO FILTER Customer Service.
Standard Accessories:
• Operator’s Manual
• Power Cord
• Pneumatic Footswitch
Please contact BUFFALO FILTER Customer Service to purchase the following accessories:
• Replacement Filters
• Remote Switch Activator
• Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories
1.3 Operational Information
The operational information contained in this section is intended for the customer review of regulatory issues. The information pertains to the use of the products both domestically and internationally:
1. The BUFFALO FILTER ViroVac™ Smoke Evacuation System(s) complies with IEC60601.1
electrical specifications in the following systems:
100/120 VAC 50/60 Hz, 220/240 VAC 50/60 Hz
2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I
3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part
4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1
5. Method of sterilization or disinfection recommended by BUFFALO FILTER (UL 60601-1,
Clause 5.4):
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
6. Degree of safety of application in the presence of flammable anesthetic mixture with air or
with oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable
7. Mode of operation (UL 60601-1, Clause 5.6): Continuous
8. Upon request, BUFFALO FILTER will provide the following:
Service and Repair Instructions, including Circuit Diagrams and Parts List
9. The fuses on the circuit board are to be serviced by an authorized BUFFALO FILTER techni-
cian as follows:
100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo)
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo)
Buffalo Filter ViroVac™ Operator’s Manual: Page 4
10. This equipment needs special precautions regarding ElectroMagnetic Compatibility and
needs to be installed according to EMC information found in this manual.
11. This equipment utilizes mobile RF communications equipment that can affect medical elec-
trical equipment.
12. This equipment has been tested and found to comply with the limits for a Class A digital de-
vice, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be required to correct the
interference at their expense.
13. This equipment operates in the following radio frequency specifications:
RX modulation: Pulse-width coded, AM 100% modulation
TX Frequencies: Manchester encoded,
A = fc =/- 423.75kHz, B = fc +/- 484.29kHz
Low bit: transition A to B
High bit: transition B to A
The ViroVac™ Smoke Evacuation System(s) and all filters are not intended for contact with
patients.
1.4 Cautions and Warnings
Please note that all Cautions and Warnings should be read and understood before any use of this equipment.
Attention, consult Accompanying Documents (Manual)
1.4.1 WARNINGS:
• Read this manual thoroughly, and be familiar with its contents prior to using this equipment.
• Test this equipment prior to a surgical procedure. This product was thoroughly tested at the fac-
tory before shipment.
• Disconnect the unit from the electrical outlet prior to inspecting system components.
• The ViroVac™ system is only intended and suitable for the applications that are mentioned in
the operating instructions.
Buffalo Filter ViroVac™ Operator’s Manual: Page 5
• The smoke evacuator produces a strong vacuum. Adjust the airflow and the position of the
inlet end of the wand or tubing to prevent patient injury and to prevent suction of surgical
materials and surgical specimens.
• If the smoke evacuator is activated while the airflow is set to a high speed, it may produce a
sudden, strong suction action. Check the airflow setting before activating the smoke evacu-
ator to prevent patient injury and to prevent suction of surgical materials and surgical
specimens.
• To maximize patient safety, the tubing or wand should not come into direct contact with
tissue. Otherwise, patient injury may result.
• The BUFFALO FILTER ViroSafe filters and single-use accessories are completely disposable.
Please dispose of according to your local codes or regulations and hospital policy. These filters
may be disposed of or incinerated, whichever is appropriate for your institution.
• Care should be taken to route the power cord, foot pedal, smoke evacuation tubing, and remote
switch activator cable as to not cause a tripping hazard or crimping of cords.
• Do not operate this device in the presence of flammable or explosive gases.
• This equipment is intended for use by healthcare professionals only. This equipment may cause
radio interference or may disrupt the operation of nearby equipment. It may be necessary to
take mitigation measures, such as re-orienting or relocating the ViroVac™ or shielding the loca-
tion.
• The use of ACCESSORIES other than those specified by BUFFALO FILTER, or sold by BUF
FALO FILTER as replacement parts for internal components, may result in increased emissions
or decreased immunity of the ViroVac™.
• This equipment should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use in necessary, the ViroVac™ should be observed to verify normal opera-
tion in the configuration in which it will be used.
• Refer routine servicing to qualified biomedical technical personnel.
• Changes or modifications not expressly approved by Buffalo Filter could void the user’s author
ity to operate the equipment.
The warranty on this product is void if any of these warnings are disregarded.
Buffalo Filter ViroVac™ Operator’s Manual: Page 6
1.4.2 CAUTIONS:
• Federal law (United States of America) restricts this device to be used by, or on the order of a
physician.
• Do not block either the tubing or the filter. If either becomes occluded or significantly restrict-
ed, the motor/blower may overheat and cause the unit to fail.
• Using any other filter or accessory not supplied by BUFFALO FILTER may cause damage and/
or cause the system to be inoperable and may void the warranty.
• Care must be exercised in the installation of hoses, adapters and suction canisters. Failure to
follow the procedures outlined in this manual may result in overheating of the motor and may
void the unit warranty.
• This device is not intended for evacuation of fluid. If fluid is expected to be aspirated to the
ViroSafe Filter, fluid collection devices must be installed with the vacuum hose assembly. Fail-
ure to install a fluid collection device could cause filter blockage and electrical damage.
• The ViroSafe Filter should be changed according to the life of the filter. The ViroSafe Filter,
used with the ViroVac™ Smoke Evacuation System(s), should not be used for more the time
specified for each filter. Failure to change the filter may result in decreased efficiency and
contamination of the electric motor, vacuum pump, and sound absorbing media within the
unit.
• Do not block either the tubing or the filter during operation. An occlusion or significant restric
tion may cause the motor to overheat and the unit to stop working.
• The installation of this equipment must be on a flat, level surface with all four (4) feet in contact
with that surface at all times. Failure to properly install the unit may cause reduced perfor-
mance, damage and/or cause the system to be inoperable and may void the warranty.
• The ambient temperature during operation must be kept between 50°F to 104°F (10°C to
40°C)
• The relative humidity during operation must be kept between 10% to 75%.
• An atmospheric pressure range of 700 hPA to 1,060 hPa.
• Storage environmental ambient temperature 14°F to140°F (-10°C to 60°C).
• Storage environmental relative humidity 10% to 75%.
Buffalo Filter ViroVac™ Operator’s Manual: Page 7
There are no user serviceable components in the ViroVac™ Smoke Evacuation System(s). Refer service to qualified service personnel.
Use only with the power cord provided and always plug into a grounded outlet.
Buffalo Filter ViroVac™ Operator’s Manual: Page 8
Buffalo Filter ViroVac™ Operator’s Manual: Page 9
Operating Instructions
Section 2.0
2.1 System Controls
The electronic system controls on the ViroVac™ Smoke Evacuation System(s) are easy to understand and simple to use. The membrane control panel contains the suction on/off switch, suction power adjustment, filter life indicator, and service indicator light. See Figure 1.
Note: Please be sure to read all instructions before installing accessories or operating this equipment. Failure to do so may result in damage to the unit and/or personal injury.
• SUCTION ON & Standby
There is one ON/OFF button on the ViroVac™ Smoke Evacuation System(s). The suction ON switch on the electronic membrane control panel is located in the upper right hand corner of the membrane panel. To power up the machine, connect the supplied power cord to a grounded outlet and the appliance inlet on the back of the smoke evacuation system. Once power has been applied, the yellow standby LED will illuminate on the keypad. Press the ON/OFF button on the membrane panel to illuminate the “fan running” green LED indicating active suction. Place the unit in “standby” by pressing the suction ON membrane switch to illuminate yellow STANDBY
LED. Turn the system main power off by unplugging the power cord from the appliance inlet on the unit or the receptacle in the wall.
• SUCTION CONTROL
(Membrane Control Panel)
The amount of suction may be adjusted by pressing the suction control button. Each time the suction control button is depressed, the motor speed in increased. Once the suction has reached the maximum level, depressing the button again will return suction level to lowest setting. The suction control should be set at the lowest practical setting to completely remove the surgical smoke from the operative site. Each time the arrow button is pressed, the suction will change to a different flow setting (low / medium / high).
• FOOTSWITCH / REMOTE SWITCH ACTIVATOR
(Membrane Control Panel)
The ViroVac™ Smoke Evacuation System also comes equipped with a pneumatic footswitch.
A footswitch or a Remote Switch activator may be added to any system by simply plugging in a
BUFFALO FILTER activation accessory into the appropriate jack on the front of the unit. When the footswitch is plugged in, the unit may be turned on or off by depressing the footswitch pedal once for each operation. For directions on using the Remote Switch Activator, please see instructions that accompany that product.
Buffalo Filter ViroVac™ Operator’s Manual: Page 10
• FILTER LIFE INDICATOR
(Membrane Control Panel)
The filter life indicator on the membrane control panel provides a visual indication of the status of the life of the filter in use. The filter life indicator for the ViroVac™ Smoke Evacuation System will automatically adjust according to the flow setting selected.
Low Flow Setting = 35 hours of filter life
Medium Flow Setting = 24 hours of filter life
High Flow Setting = 18 hours of filter life
The ViroVac™ Filter Life Indicator is factory set. All filter life timing is automatic.
Reading the Filter Life Indicator:
Install an unused ViroSafe Filter into the system per the installation instructions contained in this operator’s manual. When the system is turned on, the filter life indicator with light up the leftmost GREEN LED, indicating 100% filter life. The indicator will progress thru subsequent
GREEN LEDs, to an AMBER LED as time elapses and begin flashing RED to indicate that the filter has expended its useful life and requires replacement.
When the maximum filter life is consumed and the smoke evacuator is not powered off for greater than six (6) hours or the main power is disconnected – a new filter is required to acti vate the smoke evacuator and make operational.
• FUSES
(circuit board)
Two 10 AMP fuses (8 AMP for 220/240 ViroVac™ systems) are located on the circuit board within the housing of the system. It electrically protects both the system and the operator from damage or injury. If the system is overheated or if there is an electrical surge in the electrical system, fuses will break and the system will not operate.
When the Service light illuminates, please contact BUFFALO FILTER Customer Service for system service instructions.
Buffalo Filter ViroVac™ Operator’s Manual: Page 11
2.2 ViroSafe Filter Instructions
Note: Before installing or removing any filter, be sure that the system is turned off.
Filter Installation Instructions:
The installation of the ViroSafe Filter into the BUFFALO FILTER ViroVac™ Smoke Evacuation
System(s) is quick and simple.
1. Remove the ViroSafe Filter from the shipping box and discard any protective wrapping. Ex-
amine all filters for damage during shipping and storage. Do not install any filter with visible
signs of structural damage.
2. Insert the ViroSafe Filter into the filter receptacle. Be sure that the filter is seated completely
against the bottom of the filter chamber and clip is fully engaged.
WARNING: This device is not intended for evacuation of fluid. If fluid is expected to be aspirated using the ViroSafe Filter or the BUFFALO FILTER ViroVac™ system, fluid collection devices must be installed with the vacuum hose assembly. Failure to install a fluid collection device may cause filter blockage and/or electrical damage.
Filter Removal Instructions:
1. After the ViroSafe Filter has been exhausted and requires changing, turn the smoke
evacuation system off and disconnect any accessory tubing attached to the filter.
2. Depress the filter tab and pull the ViroSafe Filter from the smoke evacuation system and dis-
pose of in accordance with hospital policy. The ViroSafe Filter may be disposed of or inciner-
ated.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions
for maintenance and installation of a new ViroSafe Filter.
CAUTION: Using any other filter or accessory not supplied by BUFFALO FILTER may cause damage to the system and/or cause the system to be inoperable and may void the warranty.
WARNING: The ViroSafe Filter should be changed when the Filter Life Indicator shows red flashing LED (replace). Failure to change this filter may result in decreased efficiency and contamination of the electric motor, vacuum pump, and sound absorbing media within the system, or non-operation of smoke evacuator.
Buffalo Filter ViroVac™ Operator’s Manual: Page 12
2.3 Set-Up and Operation
The operation of the ViroVac™ Smoke Evacuation System is as follows:
1. Install the ViroSafe Filter.
2. Attach unit power cord to the receptacle on the rear of the system. Plug the pronged power
cord into an appropriate grounded power outlet. Route power cord in such a way as to mini-
mize potential trip hazard by users or patients or pinch hazard that could cause electric shock.
3. Optional: Attach either Remote Switch Activator or Footswitch plug into appropriate jack on
rear or side of machine. Route footswitch or RSA cord in such a way as to minimize potential
trip hazard by users or patients or pinch hazard that could result in unreliable operation.
4. Ensure that the evacuation tubing is fully seated in the inlet of the filter. Route tubing in such a
way as to minimize potential trip hazard by users or patients.
5. Activate the power unit by:
a. Pressing the suction ON/OFF switch on the membrane panel,
b. Depressing and releasing the footswitch (if connected), or
c. Activating either CUT or COAG on the Electrosurgical Pencil (if Remote Activator is con-
nected).
6. Adjust the suction level to the desired setting by pressing the UP arrow button while the unit is
activated. Noise created by the smoke excavation power unit may be minimized by selecting
the lowest vacuum setting that effectively clears the operative field of surgical smoke.
7. Deactivate the unit by:
a. Pressing the suction ON/OFF switch on the membrane panel,
b. Depressing and releasing the footswitch (if connected), or
c. Releasing either the CUT or COAG button on the Electrosurgical Pencil (If Remote Activa-
tor is connected).
8. Replace the ViroSafe Filter when the Filter Life Remaining Scale FLASHES RED (0% Life Re-
maining). Failure to change the filter will affect the performance of the system.
Buffalo Filter ViroVac™ Operator’s Manual: Page 13
Figure 1
Unit Control Panel
Buffalo Filter ViroVac™ Operator’s Manual: Page 14
Buffalo Filter ViroVac™ Operator’s Manual: Page 15
Buffalo Filter ViroVac™ Operator’s Manual: Page 16
Buffalo Filter ViroVac™ Operator’s Manual: Page 17
Buffalo Filter ViroVac™ Operator’s Manual: Page 18
Buffalo Filter ViroVac™ Operator’s Manual: Page 19
Buffalo Filter ViroVac™ Operator’s Manual: Page 20
Maintenance
Section 3.0
3.1 General Maintenance Information
This section contains information for ordinary upkeep of the BUFFALO FILTER ViroVac™ Smoke
Evacuation System. While the system has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed by a qualified Biomedical Technician to ensure continued safe and effective operation.
3.2 Cleaning
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water.
Wipe dry with a clean cloth. Do not steam sterilize.
3.3 Periodic Inspection
The ViroVac™ Smoke Evacuation System should be visually inspected at least every year. This inspection should include checks for:
• Damage to the power cord.
• Damage to the power plug or power inlet module.
• Proper mating, cleanliness and absence of damage to the filter inlet.
• Obvious external or internal damage to the system.
3.4 Troubleshooting the System – see below.
Buffalo Filter ViroVac™ Operator’s Manual: Page 21
Customer Service
Section 4.0
4.1 Equipment Return
For the quickest response to your service needs, please follow these procedures:
Step 1: Write down model and the serial number of the ViroVac™ Smoke Evacuation System.
Step 2: Call Customer Service at the toll free or local number listed and describe the problem.
Step 3: If the problem cannot be resolved over the phone and the equipment must be returned for repair, you must obtain a “Return Material Authorization” (RMA) number from Customer
Service before returning the system.
Step 4: If you have the original packing for your ViroVac™ Smoke Evacuation System, use it to properly return your unit. If you do not have the original packing material, ask Customer Service for advice on how to pack the unit for return shipment.
Step 5: Freight for all returned goods should be prepaid by the shipper. Address will be supplied by Customer Service.
4.2 Ordering Information
To reorder, obtain replacement parts or to return a unit for service, call Customer Service at:
800-343-2324
OR
(716) 835-7000
or contact your authorized BUFFALO FILTER Distributor/Representative.
BUFFALO FILTER ViroVac™ Smoke Evacuation System versions available:
• 100/120 VAC 50/60 Hz
• 220/240 VAC 50/60 Hz
Available accessories:
• ViroSafe Filters
• Suction Canister
• Remote Switch Activator
• Hoses & Tubing
• Reducer Fittings
• Electrosurgical Pencil Adapters
• Electrosurgical Smoke Pencil
Buffalo Filter ViroVac™ Operator’s Manual: Page 22
SPECIFICATIONS:
Terms & Warranty
Section 5.0
WARRANTY*:
Specifications are subject to change without notice.
SHIPMENT OF ORDER:
Buffalo Filter will try to accommodate individual customer requests for shipping method. Buffalo Filter reserves the right to decide shipping method on prepaid orders. Care is exercised in the checking and packaging of all merchandise to avoid error, but should discrepancies arise, claims should be made within 24 hours after delivery.
Buffalo Filter’s responsibility ceases with the safe delivery to the carrier at our dock. If the merchandise is damaged in transit, a claim must be made to the carrier involved. Buffalo
Filter will assist customers in pursuing these claims.
RETURN OF MATERIAL:
Buffalo Filter warrants that the filter system manufactured by Buffalo Filter shall be free from defects in material and workmanship.
Products are warranted only to the extent that
Buffalo Filter will replace without charge any filter systems proved to have defects within one (1) year of the date of delivery for P/N
VV120 & VV220 and provided Buffalo Filter has been given the opportunity to inspect the system alleged to be defective and the installation or use thereof. No warranty is included for incidental or consequential damages of any nature arising from any defect. The warranty above is the only warranty made by Buffalo Filter and is expressly in lieu of all other warranties, expressed or implied, including, without limitation, the warranties of merchantability and fitness for a particular purpose. All warranties implied by any course of dealing or usage between parties are expressly excluded.
Return merchandise must have a pre-authorized return number from Buffalo Filter and be marked with this number prior to returning. Transportation costs must be prepaid by the shipper and all risks of loss and damage of goods are the responsibility of the shipper.
Unauthorized returns will be refused. Include a copy of the packing papers and/or invoice with the return. Exchange will be of an equivalent dollar value of returned merchandise less a restocking and handling fee on new, unused, unopened equipment or disposables.
CONFIDENTIAL INFORMATION:
The information, drawings, plans, and specifications being furnished by Buffalo Filter have been developed at Buffalo Filter’s expense and shall not be used or disclosed by purchaser for any purpose other than to install, operate, and maintain the system supplied.
CONSEQUENTIAL DAMAGES/LIMITS
OF LIABILITY:
EXCEPTIONS:
1. Defective merchandise may be returned for replacement only. Please contact Buffalo
Filter Customer Service before shipping back merchandise.
2. Incorrectly shipped merchandise is exempt from restocking fees. Please contact Buffalo
Filter customer service before shipping back merchandise.
Buffalo Filter shall not in any case whatsoever be liable for special, incidental, indirect or consequential damages of any kind. In no case shall Buffalo Filter’s liability exceed the amount paid Buffalo Filter by purchaser for the specific system giving rise to the liability. Purchaser agrees to indemnify and hold
Buffalo Filter harmless from and against all liabilities, claims, and demands of third parties of any kind relating to the system and its use.
Buffalo Filter ViroVac™ Operator’s Manual: Page 23
ENTIRE AGREEMENT:
Purchaser by acceptance of Buffalo Filter’s offer does acknowledge and agree to the terms and conditions contained herein. All matters involving the validity, interpretation and application of this agreement shall be controlled by the laws of New York State. Using any filter not manufactured by Buffalo Filter may cause damage to the systems and will be cause for voiding the warranty.
JURISDICTION:
Purchaser hereby consents to the jurisdiction of the New York Courts with respect to any controversy or dispute arising out of this agreement or the merchandise sold hereunder.
* An Extended Warranty Plan is available for purchase for the ViroVac Smoke Evacu ation System. Please contact Buffalo Filter
Customer Service for more information.
Buffalo Filter ViroVac™ Operator’s Manual: Page 24
© 2010 BUFFALO FILTER®
595 Commerce Drive
Buffalo, New York USA 14228
(716) 835-7000 or (716) 835-3414 fax
Division of Medtek Devices, Inc.
Customer Service: 800-343-2324 (U.S. Only) – www.buffalofilter.com
1-716-835-7000 (International)
Advertisement