Cattani turbo smart Operator's Handbook Manual
Cattani Turbo Smart is a universal aspirator with a maximum flow of 105 m³/h and a maximum head of 2000 mm H2O, suitable for continuous operation. It is protected against the ingress of liquids and can be used both indoors and outdoors, provided it is sheltered from rain, splashing, humidity, frost, and direct sunshine.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
MANUALE ISTRUZIONI
OPERATOR’S HANDBOOK
MANUEL D’UTILISATION
GEBRAUCHSANWEISUNG
MANUAL DE ISTRUCCIONES
T U R B O
A UNIVERSAL ASPIRATOR
T U R B O
A UNIVERSAL ASPIRATOR
INDICE
Dati generali di funzionamento 50/60 Hz .........................................................................2
Introduzione ......................................................................................................................3
Segnali ed avvisi ...............................................................................................................3
Montaggio e messa in funzione ........................................................................................4
Manutenzione ordinaria ....................................................................................................8
Manutenzione straordinaria ..............................................................................................9
Istruzioni per muoversi e modifi care alcuni parametri nei menù di Turbo-Smart ............10
Istruzioni per settare la comunicazione Zig-bee (wireless) .............................................17
Descrizione allarmi .........................................................................................................22
Avvisi importanti .............................................................................................................23
Trasporto e stoccaggio ...................................................................................................23
Trasporto di apparecchi usati ..........................................................................................23
Illustrazioni......................................................................................................................121
1
2
• Dati generali di funzionamento 50/60 Hz
Aspiratore ad uso dentistico Turbo-Smart
Modello
Tensione nominale
Frequenza nominale
Corrente nominale
Modalità di impiego
Protezione contro l’umidità
Potenza massima assorbita
Portata massima
Prevalenza massima per il servizio continuo
Velocità di rotazione
Pressione sonora versione scarenata
Pressione sonora versione carenata in plastica
Pressione sonora versione carenata per interno
Pressione sonora versione carenata per esterno
Turbo-Smart “A”
230 V
50/60 Hz
7 A
Funzionamento continuo
Comune
1,45 kW
86 m³/h
2000 mm H
2
O
70 hz 85 hz
68,4 dB(A) 69 dB(A)
66,4 dB(A) 67 dB(A)
Turbo-Smart “B”
230 V
50/60 Hz
9 A
Funzionamento continuo
Comune
1,87 kW
105 m³/h
2000 mm H
2
O
70 hz 110 hz
68,4 dB(A) 73,7 dB(A)
48,5 dB(A) 49,5 dB(A) 48,5 dB(A) 52,2 dB(A)
54 dB(A) 55 dB(A)
66,4 dB(A) 72 dB(A)
54 dB(A) 58,7 dB(A)
Corrente alternata
Terra di protezione
Grado di protezione contro i contatti diretti o indiretti
Aperto (sconnessione dalla rete di alimentazione)
Chiuso (connessione alla rete di alimentazione)
IEC 417-5032
IEC 417-5019
IEC 60204-1
IEC 417-5008
IEC 417-5007
Livello di pressione sonora rilevato secondo la norma ISO 3746-1979 (E).
Parametri: r = 1,5 - rumore di fondo: 34 dB(A) - strumento Bruel & Kjaer type 2232.
T U R B O
A UNIVERSAL ASPIRATOR
INDEX
General 50/60 Hz running data.......................................................................................26
Introduction ....................................................................................................................27
Signals and warnings .....................................................................................................27
Installation and starting ..................................................................................................28
Routine maintenance .....................................................................................................32
Extraordinary maintenance ............................................................................................33
Instructions to navigate Turbo-Smart menus and to modify some parameters ............34
Instructions for communication settings Zigbee (wireless) ............................................41
Alarms description .........................................................................................................46
Important notices ...........................................................................................................47
Transport and storage ....................................................................................................47
Transport of second-hand appliances ............................................................................47
Pictures ........................................................................................................................121
25
• General 50/60 Hz running data
Turbo-Smart dental aspirator
Model Turbo-Smart “A” Turbo-Smart “B”
Rated voltage 230 V 230 V
Rated frequency
Rated current
Operating conditions
50/60 Hz
7 A
Continuous operation
Ordinary
50/60 Hz
9 A
Continuous operation
Ordinary Protection against ingress of liquids
Max. absorbed power
Max. fl ow
Max. head for continuous service
Rotation speed
Sound pressure version without box
Sound pressure version with plastic box
1,45 kW
86 m³/h
2000 mm H
2
O
70 Hz 85 Hz
68.4 dB(A) 69 dB(A)
66.4 dB(A) 67 dB(A)
1,87 kW
105 m³/h
2000 mm H
2
O
70 Hz 110 Hz
68.4 dB(A) 73.7 dB(A)
66.4 dB(A) 72 dB(A)
Sound pressure level version with box for indoor installation
Sound pressure level version with box for outdoor installation
48.5dB(A)49.5dB(A) 48.5 dB(A) 52.2 dB(A)
54 dB(A) 55 dB(A) 54 dB(A) 58.7 dB(A)
26
Alternating current
Earthing
Degree of protection against electric shock
Open (disconnected from the main electrical supply)
Closed (connected to the main electrical supply)
IEC 417-5032
IEC 417-5019
IEC 60204-1
IEC 417-5008
IEC 417-5007
Sound pressure level tested according to the standard ISO 3746-1979 (E).
Parameters: r = 1,5 – background noise: 34 dB (A) – instrument Bruel & Kjær type 2232
• Introduction
• Signals and warnings
• Introduction
This booklet is intended to illustrate the installation and starting of the appliance.
It also informs of possible dangers and about the precautions to be taken for accident prevention.
This manual should be always available for consultation during unpacking, use, installation and starting of Turbo-Smart.
Our updated manuals are available on the web site www.cattani.it.
We recommend their consultation especially for updates about safety .
• Signals and warnings
• Electrical shock risk: also 230 V can be lethal.
• Biological danger, risk of infections from epidemic diseases.
• General danger sign.
• Personal protections for heavy works.
• Personal protections against biological danger.
• High temperature.
• Keep the room free from fl ammable, corrosive or explosive material.
• Compulsory direction of fl ow or of rotation.
Warning signs cannot always fully express danger warnings, therefore it is necessary that the user reads the warnings and keeps them in due consideration.
Failure to observe a danger sign or warning may harm operators or patients.
Safety devices must not be removed. Appliances or their functioning must never be modifi ed.
Despite all our efforts, it is still possible that danger warnings are not exhaustive: we apologise to the users and kindly request them to care for all danger sources that might have passed unnoticed and to inform us accordingly.
27
• Installation and starting
28
Before unpacking the appliance, check the warning shockwatch on the carton. In case of it being red or the carton being damaged, accept the material reserving the right to examine the machine.
Unpack the appliance following the instructions shown on the package. The carton is recyclable. Dispose of it in compliance with regulations in force.
The machine installation must be carried out by a specialist equipped with the necessary tools. Install the appliance in a clean location, far from heat sources, humidity and dust.
Turbo-Smart can be installed outdoors (on a balcony, in veranda or gardens), provided that it is sheltered from rain, splashing, humidity, frost and direct sunshine.
For outdoor installation we recommend the use of our special designed box fi tted with double isolating roof, antifreeze and ventilation systems (both fi tted with fi xed thermostat for automatic temperature control).
In the plant room temperature can range from a minimum of
+ 5 °C to +35 °C max.
Turbo-Smart fi tted with box, for indoors our outdoors installation, can be supplied with antifreeze device. In case the plant room must be ventilated or air-conditioned, we suggest to contact a thermo-technician for a personalized project. The plant room must be closed to patients and extraneous people. If such a room is not available, machines must be protected by a suitable cover, which must not be easy to remove. Use protections and danger warning boards to prevent accidental risk from electrical shocks or the possibility (unlikely but not excludible) of fi re, explosion and contaminating air or liquid leakage. Use indoors and outdoors boxes designed and produced by the manufacturer of the machines only.
Keep the plant room free from fl ammable material. Make sure that there is no possibility for gas leakages. Do not connect damaged appliances to the mains. Do not use extension leads, multiple plugs or sockets. Before connecting the machine to the mains, ascertain that the feeding line is complying with the regulations C.E.I. 64-8 and that a thermal switch and a residual current operated circuitbreaker (class A or B) (16A) according to the regulations
EN 61008-1 are present. Light coloured, wooden, linoleum, rubber or marble fl oors can change colour or get marked if they are kept in contact with rubber vibration-proof devices
(1) . Therefore, it is necessary to use a rubber sheet or some other suitable material to isolate vibration-proof devices from the fl oor.
HANDLE
WITH CARE
WARNING
SHOCK W ATCH
R
RED INDICATES ROUGH HANDLING
IF RED, NOTE ON BILL OF LADING
INSPECTION MAY BE WARRANTED www
.shockwatch.com
TOLL
FREE 1-800-527-9497
MODEL L-55 (37g)
0
10
20
30
40
50
30
20
10
20
10
0
10
20
30
50
40
30
1
• Installation
Before connecting the aspirator to the piping of the centralized system, ascertain that aspiration piping is clean as heavy debris could damage the appliances.
Connect the PVC light grey aspiration tube (2b) (supplied with the machine) to the 50 mm Ø tube-holder (2) (“aspirated fl uid inlet”).
The other end of the same tube must be connected to the aspiration piping (3) coming from the surgeries (page 30).
The black heat resistant exhaust air pipe (4b) , fi tted with a metal spiral, must be connected to the 50 mm Ø tube-holder
(4) (“exhaust air outlet”).
Connect the other end of the pipe to the antibacterial fi lter
(5) , passing preferably through a silencer (5a) supplied with the aspirator. The hot air coming from the antibacterial fi lter must be conveyed outside.
Connect the 18 mm Ø tube-holder (6) to the liquid drain pipe.
In the version fi tted with Hydrocyclone the aspirator is draining liquids by gravity, therefore these can never be drained towards the top. The 10 mm Ø tube-holder (8) (“emergency drain”) must be connected to the liquid drain pipe.
The piping connecting the machine to the aspiration and draining system should be fl exible to damper the small vibrations produced by the aspirator.
The aspiration piping should be run in the fl oor and at a point near the aspirator it should rise about 30 cm to reach the tube-holder (2) (fi g. A and B, page 121).
If Turbo-Smart is installed on a lower fl oor than the dental units, the aspiration piping must not enter the centrifugal separator perpendicularly.
Rather, place a few metres of piping horizontally on the same level of the machine and then install it at the centrifugal separator inlet which is on a higher level (use a fl exible pipe) (fi g.B, page
121). The aspirated fl uid is reaspirated by Turbo-Smart.
After the installation, connect the power cable to the mains according to the regulation EN 61008-1.
Finally connect the low voltage users line: the clamps 19 and 20 of the circuit AC80-
07 (fi g. C “AC80-07 wiring diagram”, page 122) to the users line (12) (det.12, page 30 and fi g. F, page 125) which is feeding all the dental units of the same system with connections in parallel.
Ascertain that the contact on the dental unit is clean (free contact without voltage).
2b
6
2
3
8
4
20
4
2
6
5
4b
5a
29
30
• Installation in parallel
It is advisable to install in parallel only machines with the same fl ow and head values. With the installation of 2 or 3 Turbo-Smart units in parallel (fi g. F) the total fl ow is doubled or tripled, provided that the diameter of the main piping is increased of 10 mm for each additional aspirator unit. In the same way the diameter of the exhaust air pipe must be increased. Turbo-Smart is supplied complete with all the accessories which are necessary for the installation in parallel, therefore no one-way valves, supplementary control panels or peripheral units are needed. After connecting 2 or 3 Turbo-Smart units in parallel to the same main piping, connect in parallel the users wires (12) coming from the different dental units. Installing Turbo-Smart units in parallel, pay attention not to invert the small cables of the connection no.19 with the small cables of the connection no.20 (fi g. C, page 122). With multiple aspirators in parallel it can occur that one unit stops and that the operators do not notice it.
In order to bring this to the clinic staff’s attention, the connections 8 and 9 of the control panel (clean contact) (fi g. C, page 122) allow a remote alarm signal to be connected in a room under frequent control. More Turbo-Smart units in parallel can offer better performance and allow the clinic to save energy when they are working simultaneously, independent of the suction demand. Disconnecting one or more aspirators doesn’t result in a saving of energy and also decreases the suction power of the other machines.
12
3
Fig. F
Draw. F
Starting, fi nal test and users instruction
Install and connect the aspirator, select the position ON on the switch, which will light up after one of the dental units has started working. At this point aspiration will start.To check if Turbo-Smart is working correctly, it is advisable to carry out the dynamic tests (see fi g. D, page 123 and the
Turbo-Smart working diagram fi g. E, page 124). Users must be instructed on the aspirator use and routine maintenance of new, not used, and therefore not yet contaminated, machines. Show users how to follow the Turbo-Smart working phases on the display, to interpret danger warnings and to carry out routine maintenance using Puli-Jet plus new
(A) - (by means of Pulse-Cleaner) (B) - and Antifoaming
Tablets (C) .
Working
The aspirating unit (9) (through pipe 10 ) creates vacuum inside the centrifugal separator (11) . The fl uid coming from the dental units enters the centrifugal separator (11) from the collecting pipe (2) . The centrifugal separator separates air from liquids: air is exhausted outside through the pipe
(4) whereas liquids (in the version without amalgam separator) are drained to the sewage through the waste pipe connected to the tube-holder (6) . The centrifugal separator
(11) is starting before the aspirating unit (9) , this allows to drain the liquids which might collect inside the centrifugal separator before aspiration starts. Moreover, when the machine is switched off a timer (adjustable) is keeping the motor running for min.10’’ - max. 120’’.
Amalgam Separator ISO 18
On request, Turbo-Smart can be supplied with the amalgam separator “Hydrocyclone ISO 18” having its separate manual.
2
6
4
B
C
A
11
9
10
31
32
• Routine maintenance
Routine maintenance must be entrusted to specially instructed surgery staff.
• We recommend to pay special attention to all danger signals and to use goggles, gloves and disposable overalls for personal protection.
Daily
• Check for any possible alarm on the display. In case of alarms, contact the technician.
Every evening
• At the end of every working day aspirate a solution of Puli-Jet plus new disinfectant with anti-scale agent (A) using
Pulse Cleaner (B) ;
• Disconnect the machine from the mains before any intervention.
Daily
• Place the Antifoam Tablets on the dental unit fi lters.
Periodically
• Keep the aspirator fi lter clean.
Occasionally
• Make sure that the aspirator ventilation is not obstructed;
• Keep the plant room free from anything not related to the machines, especially from fl ammable material. Make sure that there is no possibility for the formation of corrosive, fl ammable and explosive mixtures.
ALLARME:
TEMPERATURA ELEVATA
B A
• Extraordinary maintenance
Extraordinary maintenance must be entrusted to a trained technician in possession of original spare parts:
• pay special attention to danger signals. Use goggles, gloves and disposable overalls for personal protection;
• check that routine maintenance has been duly carried out and make sure that Magnolia products are used;
• before any intervention carry out a series of washes with
Eco-Jet 2 disinfectant (E) , then wait 10 minutes for a complete disinfectant action.
Every 12 months
• Check maximum detected temperatures and all alarms.
Intervene accordingly;
• check the aspirator noise level (max. 72 dB according to the regulation 3047 (E);
• if necessary, remove dust from the circuit AC80 inside electronic components using a blast of dry air not exceeding 2 bar pressure. By means of a 6 bar blast of air clean also the small holes on the frontal cover of the aspirating unit (15) ;
• check the conditions of plastic hoses, in particular the hoses under pressure connecting the centrifugal separator and the Hydrocyclone ISO. We suggest to replace these hoses every 12-18 months.
Every 18-24 months
• Check the working conditions of the centrifugal separator
(11) and re-circulation valve (14) .
Every 10.000 working hours
• Every time a component containing a rubber part (“O” ring, gasket or seal) is disassembled, replace the rubber part;
• replace the motors bearings;
• where routine maintenance is not carried out properly or unsuitable products are used, train the staff and inform the person in charge.
B
ALLARME:
TEMPERATURA ELEVATA
D
E
15
11
14
33
• Instructions to navigate Turbo-Smart
menus and to modify some parameters
Main menus
When Turbo-Smart is switched on, the display shows the
Cattani logo for 10 seconds, after which time the main menu appears.
Main Menu “A1”
This menu shows some parameters such as: cycles, suction activation time, temperature, presence/absence of the amalgam container and system software release.
Control Menu “A2”
The display shows the number of times Turbo-Smart has been switched on by means of the main switch (Power
Cycles), the total hours Turbo-Smart has had power supplied (Uptime), the number of times suction has been activated (Work Cycles), the total running hours of the aspirator (Work Time), the average working hours of the blower
(Aspirator) and the number of times the control panel cooling fan has been activated (Fan Cycles).
Events Menu “A3”
The menu shows the last 10 events or alarms that have occurred to Turbo-Smart. Alarms are indicated with a number; refer to the table at page 46 to identify them.
Control Menu “A2”
34
COUNTERS – ODOM . A2
POWER CYCLES 000000 Number of times the aspirator has been switched on by means of the main switch.
UPTIME [ h]
WORK CYCLES
000000 Total hours Turbo-Smart has had power supplied.
000000 Number of times the aspirator has been activated by the dental unit.
WORK TIME [ h]
ASPIRATOR [ h]
FAN CYCLES
000000 Number of real running hours (motors running).
000000 Average working hours of the aspiration motor (UNI-
JET 75).
000000 Number of times the control panel cooling fan has been activated.
Secondary Menus
Press I to enter the Secondary Menus.
Pressing the key takes you through the other menus.
Drive Status
This menu can be accessed without any access password.
The display shows useful information about the Turbo-
Smart functioning.
User Parameters
Use the password 0000123000 to enter this menu.
From this menu the Vacuum Set Point and the unit’s displayed Language can be set, as well the Activation Code can be entered to upgrade version A to version B.
System Parameters Setup
The password 0000456000 must be entered to gain access to this menu. From this menu Off Delay Time and other technical parameters can be set.
Factory Settings Restricted Access
This menu cannot be accessed by the user. It can be entered by the manufacturer only.
35
Entering Access Passwords
The Drive Status menu is the only menu that can be accessed without the use of an access password.
To enter the User Parameters menu the access password is 0000123000.
To enter the System Parameters Setup menu the access password is 0000456000.
36
From the CATTANI S.p.A. menu, press and then .
The Access Password screen Access Password
0000000000 should then be displayed.
Press the I Enter Key which will make a cursor appear on the last zero to the right.
Press until the cursor is fl ashing on the 6th zero. Press
until number 1 appears.
Press to move to the next zero and then press until number 2 appears. Press to move to the next zero and then press until number 3 appears.
Press the I Enter Key to confi rm the password. The cursor will disappear.
Press to return to the Cattani S.p.A. menu.
It is now possible to modify the parameters of the User
Parameters menu.
Repeat the same process and enter the password
0000456000 to gain access to the System Parameters Setup menu.
Drive Status menu
This menu can be accessed without any access password.
It displays various technical parameters regarding the realtime working conditions of Turbo-Smart. Listed below is the more commonly required information.
Blower Output Frequency
(max. setting for version A: 85 Hz max. setting for version B: 110 Hz)
Blower Output Voltage
(max. setting: 220V)
Blower Overall Current
(max. setting for version A: 5,5 A max. setting for version B: 7,5 A)
Pump (Centrifuge) Output Frequency
(fi xed setting for versions A and B: 55 Hz)
Pump (Centrifuge) Output Voltage
(max. setting: 220V)
Pump (Centrifuge) Overall Bus Current
(max. setting for version A: 5,5 A max. setting for version B: 7 A)
37
38
System Temperature
Maximum Detected Temperature
(+58 °C: Alarm and Stop of the aspirator)
Vacuum
(max. setting: 200 mbar)
User Parameters menu
To access this menu you are required to enter the access password 0000123000 (see instructions on page 36).
Vacuum Set Point
This is the fi gure to which the unit will limit the vacuum.
Press the I Enter Key to make the cursor appear. The fi gure can be changed using the arrow keys . Press the
I
Enter Key again to confi rm and continue.
Language
It is possible to choose either English (1) or Italian (0) for the displayed language.
Press the I Enter Key to enable the cursor and use the arrow keys to select 0 or 1.
Press the I Enter Key again to confi rm the change and continue.
Generated Code
This code is generated by Cattani S.p.A. Each unit is allocated a unique code. This code needs to be supplied to Cattani to purchase the password (Activation Code) to upgrade version
A (2 chair unit) to version B (4 chair unit).
Activation Code
Once purchased from Cattani S.p.A., this code is entered at this point to upgrade version A (2 chair unit) to version
B (4 chair unit).
Press the I Enter Key to make the cursor appear.
Use the keys to move the cursor. Enter the Activation
Code using the keys to increase or decrease the figure.
Press the I Enter Key again to confirm, then press to go back to the User Parameters menu.
System Parameters Setup menu
To access this menu and/or make any changes, you are required to enter the access password 0000456000 (see instructions on page 36).
Vacuum set point
(Max.setting: 200 mbar)
To change any parameters in this menu:
Scroll with through to display the required parameter.
Press the I Enter Key to enable the cursor. The fi gure can be changed using the arrow keys .
Once the fi gure has been set to the desired fi gure, press the
Enter Key again to confi rm and continue.
39
40
Blower Current Limit
(Max. setting for A: 5,5 A
Max. setting for B: 7,5 A)
Turbo-Smart Off Delay Time
(Max. setting: 300 sec.
Min. setting: 10 sec.)
Pump (Centrifuge) Off Delay Time
(Max. setting: 30 sec.
Min. setting: 5 sec.)
Control Panel Fan Off Delay Time
(Default setting: 300 sec.
Max. setting 1800 sec.)
Running Options
This enables or disables the amalgam level sensor.
Set to 0: without amalgam separator
Set to 1: with amalgam separator
Drive Commands
Factory use only.
• Instructions for communication settings
(wireless)
Set the network address of the machine
(min 100) (max 110).
Enter the Zig-bee communication key (ETRX2 USB) in the USB port of the computer and launch the display program.
Install the SMART SYSTEM MONITOR program
(supplied by Cattani) in the computer.
Once the program is launched, the company’s logo will appear for 10 seconds followed by the main menu.
Click on the SERVICE menu so that you can access the device search page.
41
Click on available device search, the program will start the search and the operation may require a few seconds.
When completed, you will be able to view the various machines available and their generated code; you may need to repeat this operation a number of times.
Once the program has detected the available devices, you will have to go back to home page in order to view the parameters.
Use the dropdown menu to choose which
Turbo Smart you want to view.
the functional parameters of the Turbo Smart will then be displayed.
42
Clicking on the SERVICE icon gives access to three different menus, an open one and
2 protected by password.
To enter the USER PARAMETERS menu, enter the password 123000 and press OK.
In this menu you can change the language, set the desired level of pressure and switch the machine from aspirator for 2 dental units to aspirator for 4 dental units, thanks to an activation code.
To enter the SYSTEM DATA SETTINGS menu, enter the password 456000 and press OK.
43
In this technical menu you can view and modify all the parameters of the Turbo Smart.
The INVERTER STATUS menu, which does not require an access password, displays the functional parameters of the machine while it is in operation.
From the main menu, click on the email icon to enter data such as the address of the doctor or technician to which alarm communications must be sent.
In case of malfunction, an alarm will be shown on the display and an email sent to the technician in charge.
44
Any alarms generated during the day are displayed with an error code and a description in the ALERTS box.
The full history of all the alarms occurred in the machine during the entire period of operation can be viewed in the EVENT
LOG folder.
The operation cycles of the Turbo Smart are saved in the ODO COUNTERS menu.
45
46
• Alarms description
Alarm code
DESCRIPTION SOLUTION
AC80 AC80-07
00 0
I14 2
I15
I00
I01
I02
I03
I04
I05
I07
S08
S09
S10
S11
I16
I17
3
32
33
34
35
36
37
39
40
41
42
43
48
49
Microcontroller memory alarm Contact the technician
Amalgam level > 95%
Amalgam level > 100%
Microcontroller memory alarm
Replace the amalgam container as soon as possible
Replace the amalgam container
Contact the technician
Short-circuit due to one of the two motors
Short-circuit before the motors’ starting
Condensers have not been charged
Identify where the short-circuit is coming from and eliminate it
Contact the technician (the PC-Board is likely to be damaged)
Contact the technician (replace the PC-Board)
Overtemperature (temperature has exceeded the max. allowed limit)
Ventilate the plant room
Blower overcurrent (the blower has exceeded the current limit)
Check the operation of the blower (seized or excessive friction)
Check the mains voltage (max. 260V) Condensers overvoltage (the condensers have exceeded the max. voltage limit)
Centrifuge short-circuit Eliminate the short-circuit
Replace the PC-Board PC-Board short-circuit on the centrifuge output
Centrifuge instant overcurrent (the centrifuge has exceeded the current limit)
Centrifuge time delayed overcurrent (the centrifuge has exceeded the current limit)
The vacuum detecting pipe is disconnected
Centrifuge thermal switch switched-on
Eliminate siphons in the piping or check the centrifuge operation (seized or excessive friction)
Eliminate siphons in the piping or check the centrifuge operation (seized or excessive friction)
Connect the vacuum detecting pipe to the centrifuge
Ventilate the room
• Important notices
• Transport and storage
• Transport of second-hand appliances
• Important notices
• Appliances are guaranteed for one year from the date of sale, provided that the warranty slip is returned to the manufacturer with date of sale, retailer’s and customer’s name.
• Warranty and manufacturer’s liability cease in case appliances are treated with products which are unsuitable or different from those recommended by the manufacturer and also in case appliances are improperly used or tampered with operations of any kind carried out by people who are not authorized by the manufacturer.
• The manufacturer, concessionaires, agents and authorized technicians are at customers’ disposal for advice and assistance and to supply literature, spare parts and anything useful.
• The manufacturer reserves the right to modify the products for improvements, for technical, normative and functional reasons or for problems due to the availability of products or semifi nished products, without prior notice.
• Our updated manuals are available on the web site www.cattani.it
. We recommend they are consulted especially for updates about safety .
• Turbo-Smart is a EEE device, therefore it is subject to the WEEE (Waste of Electrical and
Electronic Equipment) regulations.
• Transport and storage
• Packed appliances can be transported and stored at a temperature ranging from -10 °C to +
60 °C.
• Packages must be kept away from water and splashing and cannot tolerate humidity >70%.
• Packages with the same weight can be stored in piles of three only.
• Transport of second-hand appliances
• Before packing, cleanse and disinfect with Puli-Jet plus new (refer to the paragraphs “Signals and warnings” and “Routine maintenance“).
• Close with polyethylene plugs all machine inlets and outlets.
• Place the machine into a polyethylene bag, seal and pack it in 3-layer corrugated board.
47
48
• SCHEMA DI MONTAGGIO TURBO-SMART
• INSTALLATION LAYOUT
MONTAGGIO AL PIANO
INSTALLATION ON THE SAME FLOOR OF
THE DENTAL UNITS
Fig. A
Draw. A all’aspiratore to the aspirator
MONTAGGIO SOTTOSTANTE
INSTALLATION ON A LOWER FLOOR THAN THE
ONE OF THE DENTAL UNITS
Fig. B
Draw.B
Min 500 mm
Min 1000 mm all’aspiratore to the aspirator
121
• COLLEGAMENTI CIRCUITO INVERTER
AC80-07
• INVERTER AC80-07 CONNECTIONS
circuito display display circuit fusibili 16 A
16 A fuse interruttore generale main switch
SOFFIANTE
BLOWER fi ltro antidisturbo main fi lter
SEPARATORE
SEPARATOR scheda elettronica main board
Fig. C
Draw. C
122
• PROVE DI DIAGNOSI SUI TURBO-SMART
• DIAGNOSTIC TESTS ON TURBO-SMART
Per verifi care il corretto funzionamento del Turbo-Smart, si possono effettuare alcuni test di tipo dinamico, di seguito esposti.
Si dovrà eseguire il test a macchina in funzione e con l’ingresso d’aspirazione libero, non collegato alla tubazione. L’altro test dovrà essere eseguito con l’ingresso d’aspirazione chiuso.
Here below you can fi nd the description of some dynamic tests useful to check the correct working of
Turbo-Smart.
One test must be carried out on running machines with open aspiration inlet not connected to the piping, the other one with closed aspiration inlet.
Modello
Turbo-Smart
Turbo-Smart model
Versione A
Version A
Frequenza raggiunta
Reached frequency
[Hz]
85
Versione B
Version B
110
Prevalenza raggiunta
Reached head
[mbar]
40 ÷ 50
I totale
Total I
[A]
3 ÷ 4,5
60 ÷ 70 6 ÷ 7,5
I separatore centrifugo
Centrifugal separator I
[A]
0,2 ÷ 0,5
0,2 ÷ 0,5
Condizione
Aspiration inlet
Ingresso aspirazione aperto in aria
Open
Ingresso aspirazione aperto in aria
Open
Versione A/B
Version A/B
60 ÷ 70 195 ÷ 200 4 ÷ 5 0,2 ÷ 0,5 Ingresso aspirazione chiuso
Closed
Fig. D
Draw. D
123
• DIAGRAMMA DI LAVORO DI TURBO-SMART
• TURBO SMART WORKING DIAGRAM
ASPIRAZIONE mbar
ASPIRATION mbar
250
2-VERSIONE "A" e "B"'
2-VERSION “A” and “B”
190
150
1- VERSIONE "A" e "B"'
1-VERSION “A” and “B”
110
50
0
0 40
(666)
60
(1000)
80
(1330)
120
(2000)
PORTATA m3/h
(l/min)
1- CURVA ASPIRAZIONE TURBO-SMART VERSIONE "A" e "B" ASSORBIMENTO IN AMPERE Da 4 a 7,5 '
FLOW RATE m3/h
FREQUENZA (Hz) Da 85 a 110. PORTATA MAX PER POSTO LAVORO 20 m3/h
PREVALENZA CONSIGLIATA PER INTERCETTATORE DI SPRUZZI.
1 - TURBO-SMART ASPIRATION CURVE VERSION “A” AND “B” ABSORPTION IN AMPERE from 4 to 7.5’.
CYCLE (Hz) From 85 to 110. MAX. FLOW RATE FOR EACH WORKING PLACE 20 m3/h
MAX.HEAD SUGGESTED FOR SPRAY INTERCEPTOR.
2- CURVA ASPIRAZIONE TURBO-SMART VERSIONE "A" e "B" ASSORBIMENTO IN AMPERE Da 5.5 a 7,5 '
FREQUENZA (Hz) Da 80 a 100. PORTATA MAX PER POSTO LAVORO 10 m3/h
PREVALENZA CONSIGLIATA PER ASPIRAZIONE CHIRURGICA.
2 - TURBO-SMART ASPIRATION CURVE VERSION “A” AND “B” ABSORPTION IN AMPERE from 5.5 to 7.5’.
CYCLE (Hz) From 80 to 100. MAX. FLOW RATE FOR EACH WORKING PLACE 10 m3/h
MAX.HEAD SUGGESTED FOR SURGICAL ASPIRATION
Fig. E
Draw.E
124
TURBO-SMART (versione “B”)
TURBO-SMART (version “B”)
Fig. F
Draw.F
TURBO-SMART (versione “A”)
TURBO-SMART (version “A”)
125
• ESPLOSO TURBO SMART
• SPLIT-UP DRAWING
126
Fig. G
Draw. G
Cod. 023785 CENTRALINO ELETTRICO
ELECTRIC CONTROL PANEL
Cod. 025400 SEPARATORE CENTRIFUGO
CENTRIFUGAL
Cod. 015646 SOFFIANTE
BLOWER
Cod. 043380 IDROCICLONE
HYDROCYCLONE
043380
• DIMENSIONI TURBO SMART
• TURBO SMART DIMENSION
Fig. H
Draw.H
127
ITALIAN PATENTS OR PATENT APPLICATIONS:
CATTANI: 1201707 - 1234828 - 1259318 - 1.187.187 - 1253460 - 233634 - 2337706
-1294904
ESAM: 1225173 - 1253783 - 0791751
FOREIGN PATENTS OR PATENT APPLICATIONS:
CATTANI: AU 546.143 - US 4,386,910 - US 4,787,846 - US 5,039,405 - US 5,002,486
AU 580839 - US 4,684,345 - US 5,330,641 - AT 0040181 - CH 0040181 - DE 0040181
FR 0040181 - GB 0040181 - LU 0040181 - SE 0040181 - CH 0211808 - DE 0211808
FR 0211808 - GB 0211808 - SE 0211808 - DE 0335061 - ES 0335061 - FR 0335061
GB 0335061 - AT 0557251 - DE 0557251 - ES 0557251 - FR 0557251 - GB 0557251
DE 0638295 - DK 0638295 - ES 0638295 - FR 0638295 - GB 0638295 - NL 0638295
SE 0638295 - US 6,083,306 - US 6,090,286 - US 6,022,216
ESAM: US 4,948,334 - DE 0351372 - ES 0351372 - FR 0351372 - GB 0351372
EP 0791751 - US 5,779,443 - CH 0791751 - DE 0791751 - ES 0791751 - FR 0791751
GB 0791751 - PT 0791751 - AU 93321 - ES 107358 - FR 222.394/395
PENDING PATENT
CATTANI: IT M098A000019 - IT M098A000119 - EP 99830010.7 - EP 99830011.5
EP 99830250.9 - EP 00830491.7 - IT M099A000165 - US 09/624,182
6/A Via Natta. 43122 Parma, Italy
10’21’48” EST - 44’50’46” NORD
Ph. +39.0521.607604
Fax +39.0521.607628 (Sales Dept.)
Fax +39.0521.607855 (Purchasing Dept.)
Fax +39.0521.399966 (Accounting Dept.) www.cattani.it - e-mail: [email protected]
Company with Quality System Certifi ed according to
UNI EN ISO 9001:2008 - UNI EN ISO 13485:2012
advertisement
Related manuals
advertisement
Table of contents
- 4 Dati generali di funzionamento 50/60 Hz
- 5 Introduzione
- 5 Segnali ed avvisi
- 6 Montaggio e messa in funzione
- 10 Manutenzione ordinaria
- 12 Istruzioni per muoversi e modifi care alcuni parametri nei menù di Turbo-Smart
- 19 Istruzioni per settare la comunicazione Zig-bee (wireless)
- 24 Descrizione allarmi
- 25 Avvisi importanti
- 25 Trasporto e stoccaggio
- 25 Trasporto di apparecchi usati