A-34C:A34B Bulletin 8/19/09 3:17 PM Page 1
Bulletin A-34C
Series DPGA & DPGW Digital Pressure Gages
Specifications - Installation and Operating Instructions
STANDARD RANGES
CERAMIC RANGES
3-11/16
[93.66]
4-9/64
[105.17]
The Series DPGA & DPGW Digital Pressure Gages have a precise +/- 1.0% full scale accuracy. The 4 digit digital display will reduce the potential for errors in readings by eliminating parallax error commonly produced with analog gages.
The DPGA & DPGW are packaged in an ABS plastic case. The unit is battery powered and has an auto-shut off to conserve battery life. A two button key pad allows easy access to features without the need to work through complex menus. These features include, auto zero and conversion of the pressure units.
Readings may be converted to various engineering units. See reverse page for available units.
OPERATING INSTRUCTIONS
For complete instructions on the two button operation see the flow chart provided on back.
On/Off: Press ON/OFF to toggle on and off. Device will automatically shut off after 20 minutes if not used.
Zero: Zeros the display - Push & hold for 2 seconds to zero the display. If after pressing zero the display does not re-zero, the unit is no longer meeting specified accuracy and should be sent back to the factory for calibration. Do not use the zero button when pressure is applied.
Units: To change units, press both On/Off and Zero buttons simultaneously.
Overpressure Indicator: The LCD will flash “OFL”, if the pressure applied is over 105% F.S.
INSTALLATION
When installing gage always use 1˝ hex at the base of the housing to tighten the gage to a mating fitting. Do not apply wrench to housing.
MAINTENANCE
Battery Removal: Remove the screw on backplate. Lift backplate off by hand. To reassemble replace backplate and screw.
Replace batteries per polarity indicators.
A “LOW BAT” descriptor indication will appear on the display when batteries need to be replaced.
ø2-43/64
[67.87]
1-1/2 [38.10]
1-1/4 NPT
SPECIFICATIONS
Service: DPGA: Air and compatible gases; DPGW: Liquids and compatible gases.
Wetted Materials:
DPGA: 316L SS, Silicone sensor;
DPGW: Type 316L SS; Ceramic Ranges: Type 316L SS, ceramic, fluoroelastomer.
Housing Materials: ABS plastic.
Accuracy: ±1.0% F.S. (Includes linearity, hysteresis, repeatability).
Pressure Limits: 2X pressure range. Vacuum range max.
pressure is 30 psig.
Temperature Limits: 30 to 120°F (-1 to 49°C).
Thermal Effect: 0.05% FS/°F.
Size: 2.62˝ O.D. x 1.52˝ deep.
Process Connections: 1/4˝ male NPT.
Display: 4-digit LCD (.425˝ H x .234˝ W digits).
Power Requirements: 9 volt alkaline battery. Battery included but not connected.
Auto Shut-off: 20 minute auto shut-off.
Weight: 5.6 oz (160 g).
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057 www.dwyer-inst.com
e-mail: [email protected]
A-34C:A34B Bulletin 8/19/09 3:17 PM Page 2
DPGA & DPGW BUTTON OPERATION
To Turn On & Off:
Press to turn unit on.
Press again to turn unit off.
To Zero Display:
To zero display, disconnect gage from any pressure source. Press and hold zero button until the LCD displays
“- - -”, then release.
To Change Units:
To change units, disconnect gage from any pressure source. Press both & buttons simultaneously until desired units are displayed.
Model
DPGA-00
DPGA-01
DPGA-02
DPGA-03
DPGA-04
DPGA-05
DPGA-06
DPGA-07
DPGA-08
DPGA-09
DPGA-10
DPGA-11
DPGA Models
Range psi
30˝ Hg to 0 (vac)
0 to 20˝ w.c.
0 to 1 psi
0 to 2 psi
-14.70
0.722
1.000
2.000
0 to 5 psi
0 to 15 psi
0 to 30 psi
0 to 50 psi
0 to 100 psi
0 to 200 psi
0 to 300 psi
0 to 500 psi
5.000
15.00
30.00
50.00
100.0
200.0
300.0
500.0
kg/cm 2
-1.033
.0508
.0703
.1406
.3515
1.055
2.109
3.515
7.03
14.06
21.09
35.15
bar
-1.013
.0498
.0689
.1379
.3447
1.034
2.068
3.447
6.89
13.79
20.68
34.47
in Hg
-29.93
1.471
2.036
4.072
10.18
30.54
61.1
101.8
203.6
407.2
611
1018
Pressure Ranges ft wc
-33.94
1.667
2.307
4.614
11.53
34.60
69.2
115.3
230.7
461.3
692
1153 kPa
-101.4
4.980
6.89
13.79
34.47
103.4
206.8
344.7
689
1379
2068
3447 oz/in 2
-235.2
11.55
16.00
32.00
80.0
240.0
480.0
800
1600
3200
4800 in wc
-407.3
20.00
27.68
55.4
138.4
415.2
830
1384
2768 mbar
-1013
49.80
68.9
137.9
344.7
1034
2068
3447 cm wc
-1034
50.8
70.3
140.6
351.5
1055
2109
3515 mm Hg
-761
37.37
51.7
103.4
258.6
776
1551
2586
0.01
0.02
0.1
0.1
0.1
0.2
Resolution psi
0.01
0.001
0.001
0.001
0.002
0.01
DPGW Models
Model
DPGW-00
DPGW-00C
DPGW-04
DPGW-05
DPGW-06
DPGW-07
DPGW-08
DPGW-09
DPGW-10
DPGW-11
Range
30˝ Hg to 0 (vac)
30˝ Hg to 0 (vac)
0 to 5 psi psi
-14.70
-14.70
5.000
0 to 15 psi
0 to 30 psi
0 to 50 psi
0 to 100 psi
0 to 200 psi
0 to 300 psi
0 to 500 psi
15.00
30.00
50.00
100.0
200.0
300.0
500.0
kg/cm 2
-1.033
-1.033
.3515
1.055
2.109
3.515
7.03
14.06
21.09
35.15
bar
-1.013
-1.013
.3447
1.034
2.068
3.447
6.89
13.79
20.68
34.47
in Hg
-29.93
-29.93
10.18
30.54
61.1
101.8
203.6
407.2
611
1018
Pressure Ranges ft wc
-33.94
-33.94
11.53
34.60
69.2
115.3
230.7
461.3
692
1153 kPa
-101.4
-101.4
34.47
103.4
206.8
344.7
689
1379
2068
3447 oz/in 2
-235.2
-235.2
80.0
240.0
480.0
800
1600
3200
4800 in wc
-407.3
-407.3
138.4
415.2
830
1384
2768 mbar
-1013
-1013
344.7
1034
2068
3447 cm wc
-1034
-1034
351.5
1055
2109
3515 mm Hg
-761
-761
258.6
776
1551
2586
Resolution psi
0.01
0.01
0.002
0.01
0.01
0.02
0.1
0.1
0.1
0.2
Note: Gages with a “C” in the range have a ceramic sensor.
©Copyright 2009 Dwyer Instruments, Inc. Printed in U.S.A. 8/09 FR# 443416-00 Rev. 1
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057 www.dwyer-inst.com
e-mail: [email protected]