AmeriWater
The Water Purification Specialists
Operation & Maintenance Manual
MediQA
Reverse Osmosis System
MODELS MSP AND MDP
3345 Stop 8 Rd, Dayton, OH 45414 | 800 535 5585 | www.ameriwater.com
Manufactured With Pride in The USA
98-0157 Rev K
TABLE OF CONTENTS
3.2 ELECTRICAL SPECIFICATIONS / CONNECTIONS .....................................................10
1 98-0157 Rev K
2 98-0157 Rev K
9.3 MEDICAL DEVICE TRACEABLE COMPONENTS .........................................................67
9.4 REPLACEMENT PART HISTORY LOG SHEET............................................................68
13.1 DEFAULT SETTINGS LIST FOR SINGLE AND DUAL PASS MediQA ..........................88
13.2 IFU – Updating Mitsubishi FX3G PLC Software with the FX3G-EEPROM-32L Module 90
13.3 IFU – Updating Mitsubishi GOT1000 HMI Software w/Jump Drive .........................93
3 98-0157 Rev K
1.
CONTACT DETAILS
We trust the unit meets all your expectations but in the event of any problems please do not hesitate to contact us as follows:
For all spares and consumables contact:
Customer Service Tel No: 1-800-535-5585
For all service or technical support contact:
Technical Support Tel No: 1-800-535-5585
Web address: www.ameriwater.com
(Or your local authorized AmeriWater dealer)
Useful Telephone Nos.
Tel No………………………………Contact Name:……………………………….
Tel No………………………………Contact Name:……………………………….
4 98-0157 Rev K
2.
INTRODUCTION AND SAFETY INFORMATION
2.1
INDICATIONS FOR USE
The AmeriWater MediQA Reverse Osmosis System is one component of a water treatment system designed to pre-treat and purify potable water using reverse osmosis for making dialysate for hemodialysis applications. The device is intended to be a component in a complete water purification system, and is not a complete water treatment system. It must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well to meet current AAMI and Federal (U.S.) standards. The AmeriWater MediQA is intended for use in water rooms in a hospital, clinic, or dialysis center. The device includes an integrated heat sanitization process.
The MediQA is available in both single pass and double pass models that supply from 5.5 to
12.0 gallons per minute (gpm) of product water. Model MSP2 is a single pass, dual-membrane
RO that produces up to 9 gpm of product water. Model MSP3 is a single pass, 3-membrane RO that produces up to 12 gpm of product water. Model MDP2 is a double pass, 3-membrane RO that produces up to 5.5 gpm of product water. Model MDP4 is a double pass, 5-membrane RO that produces up to 10.0 gpm of product water.
The MediQA is a Medical Device; as such modifications to the device are not permitted. Modifications to the machine by anyone other than AmeriWater
personnel will invalidate the marketing clearance for the device.
This Reverse Osmosis System (RO) contains a preservative solution to prevent microbiological growth and freezing. Discard all product water for at least two hours of operation before placing the RO in service.
When used as a medical device, Federal law restricts this device to sale by or on the order of a physician per 21 CFR 801.109(b)(1)!
2.2
CONTROLS
The processes of the MediQA units are controlled automatically. The MediQA is operated via a touch screen display panel which also displays system performance data. Process water is constantly monitored for:
Water quality (μS/cm) Temperature (°F)
Flowrate (gpm) Pressure (psi)
These parameters can be viewed via the touch screen display. A data logger is provided and the data logger information can be downloaded via a memory stick to a PC.
2.3
ALARMS AND MAINTENANCE
The MediQA incorporates monitoring devices and will report alarms and warnings via the touch screen display. These indicate malfunctions or that preventative action is required. This ensures the unit is maintained and operated correctly. Reference to these alarms and appropriate
actions can be found in Section 7.
5 98-0157 Rev K
2.4
CAUTION AND WARNING STATEMENTS
These instructions provide information on safe working practices. These should be adopted to ensure safe and continuing operation of the equipment. The manual should be read and understood before the equipment is placed into service.
AmeriWater reserves the right to make engineering refinements to the equipment that may not be described herein. Any questions that cannot be answered specifically by these instructions should be addressed to AmeriWater or their agents for a response.
AmeriWater will not accept any responsibility for any equipment supplied or the actions of such equipment or associated system when the customer has made a modification that is considered by AmeriWater to compromise the integrity of the original design philosophy.
If the unit’s performance becomes impaired and any remedial work appears to be outside the scope of this manual, then seek advice from AmeriWater Technical
Support at telephone number 1-800-535-5585. Be prepared to provide the unit’s serial number.
The unit must not be dismantled unless carried out by AmeriWater personnel or authorized trained personnel.
Under no circumstances should the unit be connected to the electrical supply with the front control panel open.
Always refer to the Material Safety Data Sheets before handling any of the recommended cleaning chemicals.
There is the potential for sensitive equipment/devices located in close proximity to the
MediQA unit to be affected by electromagnetic or other interference generated from other units. If affected by interference the relevant equipment/device should be relocated.
The use of mobile phones in close proximity to the MediQA unit should be avoided where possible.
The ‘Caution’ symbol is used throughout this manual to highlight where particular care must be taken to ensure the safety of the operator, and the protection provided by the equipment, is not impaired.
6 98-0157 Rev K
2.5
GENERAL SAFETY INFORMATION
Explanation of symbols and references
Danger
This symbol refers to any immediate dangers that may threaten the safety and life of persons.
Failure to observe these notices will have severe consequences
Warning
Caution on health and safety, including life-threatening injuries.
This symbol refers to a possible danger that threatens the safety and life of persons.
This symbol refers to a possibly hazardous situation.
Failure to observe these references may result in minor injuries and/or damage to property.
This symbol points out important information for working with the system in a proper manner.
This symbol indicate possible hot surface. Touching parts of the machine showing this label should be avoided.
Failure to observe these references may result in malfunctions in the system or impact on the environment.
Additional safety requirements
Country-specific requirements standards and regulations must be observed.
Usage in accordance with intended purpose.
The MediQA units are used to purify water for the purpose of feeding medical devices
(dialysis machines). The water produced must not be used for drinking as it may be harmful to the body or for any other purpose not described in this manual. The units must only be operated in accordance with this operating manual. The units must not be operated unless in proper working order. Any malfunctions must be rectified immediately.
Indoor use only.
Not to be used in an explosive atmosphere.
Refer to environmental conditions Section 3.3.
7 98-0157 Rev K
Operating staff
Only persons who have read and understand these Operation Instructions should be permitted to operate the unit. When operating the unit, it is particularly important to observe the safety information strictly.
Mechanical force
Some parts of the system could be under pressure of up to 300 psi. Always make sure the pressure has dispersed from the unit before repairs and maintenance tasks are carried out.
Bringing the system to a stop in the event of an emergency
• Depress the Emergency stop button
• Turn off the electrical supply and isolate.
• Shut off the water supply.
After remedying the fault:
• Open the water supply.
• Turn on the electrical supply.
• Reset to emergency stop button
• Clear on main screen of touch panel
• Restart/operate the unit via the main touch screen panel.
Safety information for maintenance tasks
The operator must take care to ensure that authorized and qualified professionals who have been sufficiently trained for the task at hand by thoroughly studying the Operating
Instructions perform all maintenance, inspection and assembly tasks. Professionally trained staff must properly perform these tasks.
The system must be shut down and protected from being placed in operation again unintentionally before all repair and maintenance tasks have been completed. It is essential to observe the procedure described in these Operating Instructions for shutting down the system.
Before beginning tasks on the electrical equipment of the system, a check must confirm that power has been disconnected from the corresponding section of the system. In addition, the system must be secured to prevent it from being turned on again unintentionally. Follow proper Lock Out Tag Out (LOTO) procedures.
Disposing of system parts and operating materials
When they need to be discarded, consumables must be disposed of according to local requirements.
8 98-0157 Rev K
Unauthorized conversion and manufacturing replacement parts
Conversion or modification of the system is only permitted with the approval of the manufacturer. The same applies to making changes in the programming for the control system. Original replacement parts and accessories authorized by the manufacturer enhance safety. Use of other parts will void the warranty.
Warranty claims and liability
The buyer has a one year warranty on all equipment and parts, excluding non-durable components (e.g., filter cartridges, reverse osmosis membranes, filter media, consumable chemicals, etc.); provided that the system is not subject to abuse, misuse, alteration, neglect, freezing, accident or negligence; and provided further that the system is not damaged as the result of any unusual force of nature such as, but not limited to, flood, hurricane, tornado, or earthquake.
The warranty covers the replacement of equipment and/or parts only. The warranty does not cover labor charges or travel expenses resulting from the service of equipment. The manufacturer is excused if failure to perform its warranty obligations is the result of strikes, government regulation, materials shortages, or other circumstances beyond its control.
To obtain warranty service, notice must be given to the manufacturer within 30 days of the discovery of the defect.
There are no warranties on the AmeriWater system beyond those specifically described above. All implied warranties, including any implied warranty of merchantability or of fitness for a particular purpose are disclaimed to the extent they might extend beyond the above periods. The sole obligation of the manufacturer under these warranties is to replace or repair the component or part which proves to be defective within the specified time period, and the manufacturer is not liable for consequential or incidental damages. No dealer, agent, representative, or other person is authorized to extend or expand the warranties expressly described above.
Some states do not allow limitations on how long an implied warranty lasts or exclusions or limitations of incidental or consequential damage, so the limitations and exclusions in the warranty may not apply to you. This warranty gives you specific legal rights, and you may also have other rights, which vary from state to state.
Manufacturer: AmeriWater
3345 Stop 8 Road
Dayton, OH 45414
United States of America
Telephone: 1-800-535-5585
Fax: 1-937-461-1988
9 98-0157 Rev K
3.
STANDARD FEATURES
3.1
MEDIQA STANDARD FEATURES
• Flow rate, temperature, pressure and water quality monitoring.
• System status alarms.
• Semi - automated chemical cleaning
• Touch screen user interface.
• Heat sanitization feature
3.2
ELECTRICAL SPECIFICATIONS / CONNECTIONS
3.2.1
Mains supply
See specification.
The earth leakage current of this equipment can exceed 3.5mA (<5mA).
Therefore it is essential that the earth connection is made before the supply is established.
A warning label on the equipment states:
DANGER!
DISCONNECT THE MAINS
SUPPLY BEFORE SERVICE
3.2.2
USB Connection Details
Standard USB port suitable for FAT16 formatted USB drive, 2GB max.
10 98-0157 Rev K
3.3
ENVIRONMENTAL
Applicable to all product variants
Parameter Value
Room storage and operating temperature range. 41-104°F (5 to 40°C)
Relative humidity.
Maximum altitude.
Transport and storage temperature
(limited by RO membranes).
30 to 80%
6562 ft
RFI/EMI radiation
41-104°F (5 to 40°C)
The EMC environment must be within the limits to which the RO has been tested. Care must be taken not to have sources of
RFI/EMI, which are liable to cause electromagnetic disturbance of the unit. If the RO is affected by such disturbance, the source must be suppressed or moved.
3.4
STANDARDS APPLIED
Applicable to all product variants
Type Standard
BS EN 61326:1998/IEC 61326-1 :1997;Class A
Electrical equipment, for measurement, control and laboratory use EMC
EMC
LVD
MDD requirements.
BSEN 61000-3-2:1995, Incorporating Amendments 1 & 2.
Mains Harmonic Emissions.
BS EN 61000-3-2:1995, Incorporating Amendment 1
Mains Flicker Emissions.
BS EN 61010-1:2001 , Incorporating Amendment 1
Safety requirements for electrical equipment, for measurement control and laboratory use.
BS EN ISO 13485:2003
Medical Devices Quality management System
11 98-0157 Rev K
3.5
SPECIFICATION DATA
Single Pass MediQA Dual Pass MediQA
Catalog Number
Performance
Permeate flow rate
@ 77°F**
Dimensions
Height
Width
Depth
Weights gpm
Inches
Inches
Inches
9.0
79
43
64
12.0
79
43
64
5.5
79
43
64
10.0
79
43
76
Working
Shipping*
Services-Feed Water
Pressure
Free Chlorine
Temperature
TDS
Total Hardness lbs lbs
1310
1050
1375
1100
1660
1370
2285
1970
30/80 psi
Following activated carbon, free chlorine level must be < 0.1 ppm
For best performance feed water temperature should be 41°F to 95°F
Maximum feed water total dissolved solids 1000 ppm
Softened feed water with a total hardness of < 4 ppm as CaCO
3
S.D.I.
Fluoride
Silt density index following final 5 micron filtration should be < 5
< 1.2 ppm for “Hi Recovery Mode”
Services - Power
Supply
Max Current Draw
V/ph/Hz
Amps
All units require 3-phase supply 208/230V,with Neutral, 60 Hz
44 44
* Un-crated weight.
** Variations in feed water temperature can affect output by up to +3% per degree Fahrenheit.
44 59
12 98-0157 Rev K
3.6
PERFORMANCE PARAMETERS
Catalog Number
Product Flow Specification (GPM)
Product Flow Range (GPM)
Stage 1 Concentrate Flow (GPM)
Stage 2 Concentrate Flow (GPM)
Stage 1 Recovery (%)
Stage 2 Recovery (%)
MediQA Performance Parameters
MSP2
NA
Single Pass
Flow Rates
9
7.2-10.8
Nominal: 5
3.8-9
NA
Nominal: 65
50-70
MSP3
12
9.6-14.4
Nominal: 6.5
5-12
NA
Nominal: 65
50-70
NA
MDP2
Dual Pass
6
4.8-7.2
Nominal: 5
3.8-9
1-3
Nominal: 65
50-70
Nominal: 80
75-85
MDP4
10
8-12
Nominal: 6.5
5-12
1-3
Nominal: 65
50-70
Nominal: 80
75-85
Pressure
Stage 1 Pump Pressure (PSI)
Stage 2 Pump Pressure (PSI)
Feed Water Pressure (PSI)
Interstage Pressure [Bar]
(PSI)
140-180
NA
30-80
NA
140-180
NA
30-80
NA
140-180
140-180
30-80
2-5
(30-70)
140-180
140-180
30-80
2-5
(30-70)
Temperature
Feed Water Temperatures (F) 40-90 40-90 40-90 40-90
Conductivity
Output Conductivity
>94% Incoming
Water
>94% Incoming
Water
>94% Incoming
Water
>94% Incoming
Water
Based on Standard Recovery Mode
Each MediQA should perform within these parameters. Pressures may vary depending on the age and condition of the membrane.
*The ideal Feed water temperature is 77 degrees Fahrenheit. This will be the optimal operation temperature of the system.
The DOW heat disinfection membranes are manufactured with a ±20% tolerance for flow rate. This is factored into the Product Flow Range category.
Output conductivity is based on the quality of the feed water. The output conductivity should be greater than 94% of the incoming feed water.
13 98-0157 Rev K
4.
OPERATIONAL OVERVIEW
The MediQA is available as a single pass or dual pass reverse osmosis (RO) unit to be used for purification of potable water for renal dialysis applications.
Potable water enters the MediQA through an inlet solenoid valve filling the feed water tank. A high pressure pump delivers water from the feed water tank through the 1st stage RO module set (may comprise of one or more modules), with each module containing a high performance membrane.
The water entering the RO module is split into two flows. The water which passes through the RO module (membrane) is known as permeate and is purified water. The rest of the water, rejected by the membrane, passes out of the RO module as a second flow stream. The rejected water contains an increased level of dissolved salts and is known as concentrate. A portion of the concentrate will be recovered and directed back into the tank for re-processing.
In the single pass system, Hi Recovery Mode allows a higher percentage of reject water to be recovered. The default target recovery percentage is 65%.
In the dual pass system, process water (permeate) from the 1st stage is pressurized by a second high pressure pump and fed to the second stage RO module set (may comprise of one or more RO modules).
In the dual pass (two RO stages) MediQA, a portion of the concentrate (reject water) from the 1st stage RO module set is discharged to drain while the balance is returned to the tank. Concentrate from the 2nd stage RO module set is returned to the feed water tank for reprocessing.
Permeate from the 2nd stage RO module set is fed via a manifold to the distribution loop. Permeate returning from the loop is fed back into the feed tank.
The feed and permeate water flows are monitored at various points in the process to verify temperature, conductivity, and flow.
This data is displayed on a touch screen panel to give instant feedback of water quality and process activity.
4.1
MEDIQA BASIC OPERATION
When TIMER or CONTINUOUS is activated by touch buttons on the touch screen display the MediQA initially performs an AutoFlush followed by an AutoRinse routine, once completed purified water will be available (supplied through connection pipe work) to the distribution loop or attached medical device. This feature of the system prevents
14 98-0157 Rev K
inadvertent supply of poor water quality to the loop or attached device (typically dialysis machines). System operation can be switched OFF by pressing the OFF button on the touch screen display. An AutoFlush is performed when the unit is switched off, if it has been in operation for more than two hours.
Purification of the incoming softened water supply is achieved by employing reverse osmosis membranes technology.
The MediQA can be set to “POWER ON STANDBY” for occasions when there are long periods (typically days during shut down periods) without use. In standby mode the
MediQA is factory set to perform 10 minutes of operation in every 2 hours, these values are adjustable. This enables water to be circulated through the MediQA and distribution loop, in order to maintain water quality.
The MediQA has two modes of operation, TIMER and CONTINUOUS. If TIMER is selected an internal time clock activates operation, determined by the timer clock settings. When not scheduled to operate, the device will turn on every 2 hours for 10 minutes of operation to maintain the quality of the water in the distribution loop. If
CONTINUOUS is selected the MediQA will supply water to the attached distribution loop
/ medical devices on a 24/7 basis.
The single pass MediQA units have the capability to run in HI RECOVERY MODE, which uses up to 40% less water than STANDARD RECOVERY MODE. HI RECOVERY MODE is
the default setting. See Section 6.4.1 for navigation instructions.
The MediQA also features routines to enable chemical cleaning of its RO membranes
(but not the connected distribution loop).
In addition to the above, the MediQA incorporates process elements to enable heat disinfection of the reverse osmosis membranes and its pipe work. Heat disinfection can be activated manually using the RO Heat Disinfection button on the touch screen display or automatically if timer clock settings are implemented.
Regular cleaning and sanitization of the unit is recommended to maximize the life of the
RO membranes and ensure high performance.
Once CHEMICAL CLEAN has been selected the process is semi-automatic and requires only the provision of specified cleaning and sanitizing chemical / agents. The touch screen display will give prompts at appropriate stages of the routine.
Frequency of heat disinfection will depend upon usage and application demands.
Monitoring of the system for bacterial content should be conducted at regular intervals to determine the optimum frequency for heat disinfection.
15 98-0157 Rev K
4.2
SEQUENCE OF OPERATIONS
Concentrate Throttle Valve
Product to Loop Valve
Recirculation Throttle Valve
Loop Return Valve
Concentrate to Drain Valve
4.2.1
Introduction
There are two modes of operation, Hi Recovery Mode and Standard Recovery
Mode. Hi Recovery Mode is the default and can be toggled to Standard Recovery
Mode on the RO Supply screen (see Section 6.4.1). The MediQA system uses a
series of automated 2 and 3 way stainless steel valves to direct the water flow through the unit during each of the various stages of operation. These valves are marked with yellow indicators that represent the openings of the valve. The yellow markers are used as a visual indicator of the position of the ball valve as it relates to the flow of the water through the valve. The following pictures illustrate the valves during operation.
Water exits valve
Water enters valve
16 98-0157 Rev K
4.2.2
High Recovery Mode
This mode of operation is designed to limit water usage while still meeting water quality standards. Hi Recovery Mode uses up to 40% less water than Standard
Recovery Mode. The mode is comprised of Flush, Rinse, and Supply.
FLUSH
Before the MediQA provides water to the loop for dialysis use, the system goes through a flush. This isolates the water from the loop and dumps all of the water down the drain for a preset time. The following figures show the valve positioning during this process. Positioning of the valves is controlled via the PLC.
17
Figure 1
Concentrate to Drain Valve
Figure 2
Loop Return Valve
Figure 3
Product to Loop Valve
RINSE
After the flush cycle is completed, the RO will go into a rinse. During rinse the unit continues to run the pumps at a high pressure. Water is flushed down the drain similar to the flush of the membranes. The position of the valves during this process is the same as Figures 1-3.
SUPPLY
After Flush and Rinse cycles are completed, the MediQA will go into Supply.
Supply is when the RO begins providing water to the loop for use with the
Dialysis machines. This is the process the RO will be in for a majority of its usage. During the supply mode, valve position is critical as product flow rate is directly dependent on the valves. Water returning from the loop is redirected to the MediQA tank via the loop return valve shown in Figure 5. The following figures show the expected position of these valves during normal operation and are instructed to be fully open via the PLC. This mode is split into two parts to conserve water, reject to tank and water to drain. During this period of operation, the reject water will be sent to drain at a target of 65% recovery.
Reject to Tank:
Concentrate and permeate water will recirculate to the feed water break tank until the permeate quality reaches the HI RECOVERY QUALITY HIGH LIMIT or
98-0157 Rev K
BREAK TANK TEMPERATURE HIGH LIMIT set-points. When these set-points are reached the water to drain period will begin and the RO will dump water to drain. During times of low water usage in the RO distribution loop, the feed water quality in the break tank will remain close to incoming water quality. See
Section 6.4.1 for more information.
Figure 4 Figure 5 Figure 6
Concentrate to Drain Valve Loop Return Valve
Water to Drain:
Reject water is dumped to the drain until the permeate quality reaches the HI RECOVERY QUALITY
LOW LIMIT or BREAK TANK TEMPERATURE LOW
LIMIT set-point. Valve position will remain the same for Loop Return and Product to Loop Valves. The
Concentrate to Drain Valve will rotate as shown in
Figure 7.
Product to Loop Valve
4.2.3
Standard Recovery Mode
Figure 7
Concentrate to Drain Valve
Standard Recovery Mode is also comprised of Flush, Rinse, and Supply mode.
Flush and Rinse operations remain the same, see Figures 1 through 3. However, when operating in Standard Recovery Mode, there is no Reject to Tank step of
Supply. There will be more water usage due to less water being recovered during this period. See Figures 4-6 for valve positioning.
4.2.4
Heatsan Fill Procedure
The MediQA can be called on to provide water to the Heatsan. Be sure that MINT cable is connected between MediQA and Heatsan for fill procedure to function.
18 98-0157 Rev K
Figure 8
Concentrate to Drain Valve
Figure 9
Loop Return Valve
Figure 10
Product to Loop Valve
19 98-0157 Rev K
4.2.5
Cleaning Procedure
CHEMICAL CLEAN
The MediQA is equipped with the ability to perform a chemical cleaning of itself once chemicals have been added in the tank. This consists of 4 main steps,
circulation, tank drain, high pressure rinse, and low pressure rinse.
Valve positions vary slightly for each step in the cleaning process. The valves are shown in the following figures.
Recirculation:
During the recirculation phase of the chemical clean process, the valves redirect all of the water back into the MediQA tank. The pumps turn on at a low speed to keep the pressure in the system below the alarm set point.
20
Figure 11
Concentrate to Drain Valve
Figure 12
Loop Return Valve
Tank Drain, High/ Low Pressure Rinse:
During the tank drain, high pressure rinse or low pressure rinse, the concentrate drain valve shown in
Figure 13
Product to Loop Valve
Figure 14 opens fully allowing all of the residule chemicals in the system to go to the drain. The tank will drain completely before the rinse process occurs. Once the rinse process occurs, the fill line opens allowing the tank to refill with clean water. During these 3 phases all other valves remain in the same position as the recirculation procedure.
Figure 14
Concentrate to Drain Valve
98-0157 Rev K
4.2.6
Disinfection Procedure
HEAT DISINFECTION
The MediQA is capable of a heat disinfection of its plumbing and membranes.
During the heat disinfection process the RO circulates water within itself at
185˚F for 30 minutes (adjustable). The valves on the unit are programmed to act in similar fashion to the valves during a chemical disinfection of the unit.
Once the hold period of the unit is met, the unit undergoes a cool down process where it calls for fresh cold water and dumps the hot water in the unit down the drain. The following pictures will show the valve positioning during preheat, hold, and cool down phases.
Preheat/Hold Period:
During the pre-heat and hold period, the RO disables its ability to call for water from its water supply. The water in the unit is circulated throughout the plumbing and membranes until it reaches 185 ˚F. Once the temperature is reached, the water is circulated for 30 minutes (adjustable). The valve positioning remains the same for both phases of the heat disinfection.
Figure 15
Concentrate to Drain Valve
Figure 16
Loop Return Valve
Figure 17
Product to Loop Valve
Cool Down:
During cool down of the MediQA after the heat disinfection, the concentrate to drain valve opens periodically to allow the hot water in the RO to go to the drain.
This allows cool water to refill the tank, gradually cooling the unit down to its operational temperature. All other valves remain in the same position throughout the entire heat disinfect procedure.
21 98-0157 Rev K
Figure 18
Concentrate to Drain Valve
Figure 19
Loop Return Valve
Figure 20
Product to Loop Valve
COOLDOWN OF HEATSAN AFTER HEAT DISENFECTION OF THE LOOP
When the Heatsan unit performs a disinfection of the loop, the unit must be cooled down to allow cold water from the RO to be supplied to the Dialysis machines. By connecting the MediQA to the Heatsan via the MINT cable, the
Heatsan unit can call for a fill from the MediQA when it needs to be cooled down.
When this occurs, the MediQA will position the valves so water is pushed through the loop to gradually cool the unit. Once this procedure is completed, the
MediQA is free to go back into normal operation. The loop return valve, shown in
Figure 22, is positioned to redirect water back to the Heatsan unit. This keeps the hot water from entering the tank of the MediQA.
Figure 21
Concentrate to Drain Valve
Figure 22
Loop Return Valve
Figure 23
Product to Loop Valve
22 98-0157 Rev K
5.
INSTALLATION
5.1
ENVIRONMENT
The unit should be installed in a clean and dry indoor, non-hazardous, ventilated
environment (see section 3.3).
5.2
UNPACKING
Two people are required for this operation. Manual handling requirements should be observed.
The MediQA is shipped on a wooden pallet. Before removing it, ensure the
MediQA is as close to the final location as possible on a smooth level surface.
Take great care as the weight of the machine (depending on the model) is
1050 to 1970 lbs.
Move the MediQA into its final position (the ground must be level).
Ensure the following items have been provided.
• MediQA Unit
• Key for control panel
• Operators’ Manual
5.3
CONNECTIONS
5.3.1
Electrical Supply
An electrical supply (see specification section 3.5) should be made to the
control panel of the MediQA using a suitable armored cable. The cable should be fitted using the gland provided and terminated to the Isolator connections L1, L2, L3, N and Earth.
5.3.2
Water connections
Softened potable water supply: 1.25” NPT female thread
Overflow drain connection: 1.5” NPT female thread
All other connections: 1” NPT female thread
5.3.3
Drainage
A suitable, unrestricted, drain is required, capable of handling the discharge flow of the MediQA (see specification). The drain should be capable of accepting
194°F (90°C) water.
23 98-0157 Rev K
5.4
COMMISSIONING
ONLY BY CERTIFIED AMERIWATER REPRESENTATIVES.
This Reverse Osmosis System (RO) contains a preservative solution to prevent microbiological growth and freezing. Discard all product water for at least two hours of operation before placing the RO in service.
5.4.1
Pretreatment Lockout
The MediQA can utilize pretreatment lockout. Pretreatment lockout will prevent the device from operating while the pretreatment is going through regeneration.
The MediQA is shipped from the factory with a jumper installed between wire terminal X24 and wire terminal 1. To enable the pretreatment lockout functionality, remove the jumper wire between wire terminal X24 and wire terminal 1. After removing the jumper wire, wire the normally closed side of the pretreatment lockout switch or switches (located in the pretreatment controller) as shown below.
Note: This diagram can be adapted for any number of devices so long as when any one opens, the connection to the MediQA is opened.
Failure to remove the jumper between X24-1 will prevent pretreatment lockout. This will allow the MediQA to operate at all times, regardless of the status of the pretreatment.
24
5.4.2
Heatsan Water Loss Lockout
Heatsan water loss lockout will prevent the MediQA from operating while the water loss alarm is active on the Heatsan. This feature stops the flow of water to allow a leak to be dealt with. Once the leak is fixed, the alarm can be cleared on the Heatsan and confirmed on the MediQA. This feature is enabled by the MINT cable connection between the Heatsan and the MediQA.
98-0157 Rev K
5.4.3
Alarm Panel Connections
The MediQA utilizes fail safe connections for the Alarm Panel. The alarm output is normally closed, so that when there is an alarm, this opens. AmeriWater supplies an alarm panel that can be set to either normally open or normally closed contacts. As shipped, the alarm panel will look for a closure to indicate an alarm condition. See the manual for the AmeriWater alarm panel for detailed instructions on changing the input type.
To connect the MediQA to the alarm panel, run a signal wire from Y15-55 to the input on the alarm panel for RO alarm. When the MediQA goes into alarm, the connection will open.
5.4.4
Installation Kit Details
Each MediQA will be shipped with an installation kit.
The temperature probe will need to be installed into the top port of the cross.
The compression fitting will need to be loosened (do not remove) to allow the probe to slide into place. When installed properly, the ridge on the probe will be flush with the top of the fitting. Secure this connection to be tight by hand, then use a wrench to tighten 1-2 turns.
The check valve must be oriented to allow flow into the loop return port of the
RO, but no flow from the RO to the loop.
25
Back
Pressure
Regulator
98-0157 Rev K
Portions of the kit are shipped loose to allow flexibility based on space constraints at installation.
To adjust the back pressure on the loop, loosen the lock nut on the bottom of the back pressure regulator. You can then adjust the bolt to the right to increase or left to decrease the pressure to the desired range. Once set, secure the lock nut to prevent inadvertent adjustment.
5.5
Initial Startup
The MediQA should not be started until the entire system is installed and ready for water. Pretreatment devices should already be in service and filter cartridges installed. Post treatment equipment will be exposed to water the first time, insure all union fittings are tight. Insure all electrical connections are safe and secure.
The MediQA has been factory calibrated and tested. No calibrations are required during the RO startup.
Turn main power supply to the MediQA on at the disconnect or breaker switch.
Turn the MediQA on at the disconnect switch on the RO controller. The HMI screen should boot displaying the main screen to the RO.
Installer shall verify polarity for pump(s) rotation at the main power supply before proceeding with RO startup. The rotational direction can be found on the name plate on the pump(s). Navigate to the CALIBRATION MENU > OVERRIDES and toggle
“RO1 PUMP” on then off, view the rotation of the green light on the face of the pump. Do the same with “RO2 PUMP” if installing a Dual Pass.
26 98-0157 Rev K
Remove the lid to the MediQA tank, and open the tank drain valve on the bottom of the tank.
Slowly open the feed water supply valve to the MediQA. Watch the tank as it fills, water maybe cloudy and contain piping debris. Flush debris from the break tank thru the overflow until water is clear and free of debris. May need to close feed water supply valve periodically to allow tank to empty.
Tank Drain Valve Feed Water Supply Valve
27
Once water is clear close the tank drain valve and install the tank lid.
Navigate SETTINGS MENU > RO SUPPLY and increase the Flush and Rinse time to
10 minutes each. Level 3 User Login will be required for this step. Also, click the
ENTER button under MEDIQA OPERATION MODE. Hold the disable button for 2 seconds to put the unit into Standard Recovery Mode.
98-0157 Rev K
28
Open the bypass to drain valves located at the end of the loop after the product recovery kit, before the loop return to MediQA. This will allow water to run to drain to flush any construction debris from the loop.
Open Valve
Close Valve
Navigate to the OPERATIONS MENU and start the RO by pressing the CONTINUOUS button. The RO will begin to flush then rinse to the drain.
98-0157 Rev K
Warning
This Reverse Osmosis System (RO) contains a preservative solution to prevent microbiological growth and freezing. Discard all product water for at least two hours of operation before placing the RO in service.
After the flush and rinse stages finish, the unit will pause and the permeate valve will shift and the RO will go into Supply mode. Water will begin to flow through the distribution loop.
After 2 hours minutes close the bypass to drain valve. Allow the RO to continue to operate.
Monitor the status of the RO on the Activity Details screen under the Main Menu.
During this time there are a few things that need to be noted and checked. Note the steady state conductivity level of the permeate water by looking at the OUTPUT
CONDUCTIVITY. Also, note the steady range of values for TANK TEMPERATURE.
Verify the measured recovery is at or near 65% on the RO Parameters screen
(CALIBRATION MENU > RO PARAMETERS). Verify permeate flow on screen matches the return flow on the product recovery kit flowmeter.
Product Recovery
Kit Flowmeter
29
If the flow values do not match then on the permeate relief valve, adjust the amount of unused permeate returns to the break tank. Adjust the permeate relief valve on the MediQA until the flow meter matches the screen. Careful not to increase the pressure at the endotoxin filters past 80 psi.
98-0157 Rev K
Navigate SETTINGS MENU > RO SUPPLY > MEDIQA OPERATION MODE ENTER and hold ENABLE for 2 seconds to return to Hi Recovery Mode upon completion of the 2 hours. Hi Recovery Quality is default to a low limit of 6 µS and a high limit of 10 µS during RO Supply. Break Tank Temperature is default to a low limit of 77 ° F and a high limit of 85 ° F. If the value for conductivity or tank temperature recorded during standard recovery operation is higher than the limits set above, adjust the defaults to fit.
Check the post treatment and distribution loop for leaks.
Once unit has successfully been placed into service, it is recommended to heat disinfect the RO unit prior to disinfecting the loop.
Installation timing may effect at what stage a heat disinfection is conducted on the
RO. May decide to program an automatic heat disinfect.
Make sure the pretreatment is not programmed to backwash or regenerate during the MediQA heat cycle. May be necessary to unplug the pretreatment during this first initial heat disinfect.
30 98-0157 Rev K
To manually begin a heat disinfect from the OPERATION MENU, press RO HEAT
DISINFECT. The controller will ask you to confirm, press YES. The MediQA will automatically complete the heat disinfect cycle.
31
Upon completion of the RO heat disinfect, verify the unit reached the disinfect temperature. Go to MAIN MENU > ACTIVITY LOG and scroll through the data until you see the tank temperature in RED numbers. Insure the unit held 185°F for 30 minutes.
Before programming the MediQA, consult the facility manager for operating and maintenance schedules that require water to complete.
Warning: The supply schedule and heat disinfect schedule may not interfere with pretreatment backwash/regeneration schedule.
To complete set up, navigate SETTINGS MENU > RO SUPPLY > SUPPLY TIME
SCHEDULE ENTER to get to the screen shown below. Enter the times necessary for the RO to supply water each day, using 24:00 clock. If there are days no operation is necessary then enter a start and end time of 00:00. When in TIMER mode the
machine will operate on this schedule. See Section 6.4.1 for details about this
screen. See Section 6.1.1 for information about CONTINUOUS versus TIMER mode.
98-0157 Rev K
Once in the RO SUPPLY screen, reset the flush and rinse times to 2 minutes each.
It is recommended to start the RO at least 30 minutes before staff arrives.
This will allow the staff to take water system readings immediately upon arrival instead of waiting the required 15 minutes.
Navigate SETTINGS MENU > RO HEAT DISINFECT > RO HEAT DISINFECT
SCHEDULE ENTER and set the desired day to heat disinfect the RO.
32
Once programming is complete, return to the OPERATION MENU and press TIMER.
Unit will now operate on the schedule. If it is wished to start the RO immediately within the schedule press TRIGGER. RO will start automatically and continue to operate until the set end time.
98-0157 Rev K
6.
OPERATOR INTERFACE
The MediQA is controlled and operated by the touch screen mounted on the end of the machine. Some of the screens differ in appearance between the
DPH and SPH models based on the parameters necessary for operation.
Function buttons
Main selection buttons
HMI override switch
Emergency stop
Main power isolator
33
6.1
OPERATIONS MENU
The OPERATION MENU is displayed when the MediQA is switched ON or is displayed if the touch screen has not been used for 30 minutes.
The OPERATION MENU features eight function buttons. Pressing any one of these selects the named function. Only one function may be selected at any one time. To switch between functions the OFF button must first be pressed.
Some functions, when switched off using the OFF button, will need to complete some process cycles. It will not be possible to select another function until these process cycles have been completed.
98-0157 Rev K
6.1.1
OPERATION MENU screen
On all screens the title of the screen is shown in the top left hand corner, below this the current MediQA operating status is shown (STANDBY in the example shown above).
In addition there is a colored border which is common to most screens. The color indicates the current status as follows:
Blue – Function operating to specification.
Amber – Warning active
Red – Alarm active
“Warning” alarms are advisory; they do not stop system operation. They indicate that some element of the system needs attention.
“Alarms” stop the system. The system will only restart once the problem giving rise to the Alarm has been corrected.
The top right of the screen displays the date and time and below this a completion time for the current operation (only available for certain functions).
Figure 24
34
TIMER button
TIMER provides timed operation for the MediQA. Operation only commences when reaching times programmed in the RO SUPPLY TIMER screen. If timer button is pressed part way through a timer period, pressing the TRIGGER button will start operation.
CONTINOUS button
Starts system operation whether times have or haven’t been programmed to the RO SUPPLY TIMER screen. Operation is continuous until OFF is selected. Since starting and stopping incorporate essential flushing of the
98-0157 Rev K
35 membrane, CONTINUOUS operation should not be used for extended periods or damage to the membranes may occur.
POWER ON STANDBY button
Energizes the system but does not start operation. The system will perform a
flush and rinse for a predetermined time, see Section 13.1.
OFF button
Deselects and switches off selected functions. The button will illuminate red while ending certain functions.
RO HEAT DISINFECT button
Starts a manual heat sanitization of the MediQA
Caution: Supply to the distribution loop will cease while RO heat disinfect is in process.
CHEMICAL CLEAN button
Starts a manual chemical clean of the MediQA
Supply to the distribution loop will cease while chemical clean is in process. Chemical clean process cannot be aborted once started.
During the chemical clean process you will prompted and required to press buttons to acknowledge certain key stages in order to complete the process (see section 6).
(i) ADD CLEANING CHEMICAL
(ii) TEST FOR RESIDUE CHEMICALS
(iii) PASS WATER QUALITY
EXTRA PURGE button
Flushes purified water through the MediQA.
EMERGENCY STOP RESET button
Only appears if the emergency stop button on the control panel has been pressed. To clear the emergency stop, pull out (reset the emergency stop button) and press the EMERGENCY STOP RESET button on the screen.
OPERATION MENU button operation.
When powered up, the OFF button will be illuminated red. To select another function, press the OFF button (this will turn gray) and then press the desired function button. The function button will illuminate green to show it has been selected.
To switch off a function, press the OFF button, this will illuminate red and the function button will turn gray.
98-0157 Rev K
If both the OFF button and the function button remain illuminated, the function selected is completing shut down processes, when these are complete the function button will turn gray. It will not be possible to select other functions until only the red OFF button is illuminated red.
If a manual RO HEAT DISINFECT is selected it can be aborted by pressing the OFF button.
MAIN MENU button
Switches screen to Main Menu
STATUS MONITOR button
Switches screen to status monitor (see Figure 26)
SETTING MENU button
Switches screen to Settings Menu (see Figure 32)
CALIBRATION MENU button
Only accessible by AmeriWater technicians.
6.1.2
Operation Periods
MediQA Operation Periods
Period
Standby
Supply
Flush
Rinse
Refresh
Explanation
The MediQA can be set to “POWER ON STANDBY” for occasions when there are long periods (typically days during shut down periods) without use. In standby mode the MediQA performs 10 minutes of operation in every 2 hours. The duration and interval are user settable. This enables water to be circulated through the MediQA and distribution loop, in order to maintain water quality.
Supply is when the RO begins providing water to the loop for use with the
Dialysis machines. This is the process the RO will be in for a majority of its usage. Can be operated in continuous or a set schedule.
Before the MediQA provides water to the loop for dialysis use, the system goes through a flush. This isolates the water from the loop and dumps all of the water down the drain for a preset time.
After the flush cycle is completed, the RO will go into a rinse. During rinse the unit continues to run the pumps at a high pressure. Water is flushed down the drain similar to the flush of the membranes.
Loop has exceeded the temperature set point and is refreshing the water to lower the temperature.
RO Heat Disinfect The MediQA is capable of a heat disinfection of its plumbing and membranes.
During the heat disinfection process the RO circulates water within itself at
185˚F.
Temp Hold Once the temperature is reached in the RO Heat Disinfect period, the water is
circulated for a set period. Hold time is user settable, see Section 13.1 for
default.
36 98-0157 Rev K
Cool Down
Clean
Clean Recirc
Clean Rinse
RO Heat Disinfect enters this mode to cool down break tank and membranes at the end of the RO Disinfect cycle. Cool down temperature is user settable,
The MediQA is equipped with the ability to perform a chemical cleaning of itself once chemicals have been added in the tank. Recirculation time is user
settable, see Section 13.1 for default.
During the recirculation phase of the chemical clean process, all of the water is redirected back into the MediQA tank. Clean time is user settable, see
After the clean recirc cycle is completed, the RO will go into a clean rinse to purge the chemicals from the system.
HMI override key switch is turned, the system goes into continuous mode. Override
Remote ON
Loop Heatsan
The MediQA is in slave mode and the master MediQA is telling it to operate.
Emergency Stop System operation has been halted via the Emergency Stop button on the control panel.
System OFF
Heatsan is performing a heat disinfection by cycling hot water through the loop. MediQA locked out from operating.
System operation has been switched OFF via the button on the touch screen display. An AutoFlush is performed when the unit is switched off, if it has been in operation for more than two hours.
Standby HS Over Heatsan disinfection just completed and the unit is returning to standby.
Heatsan Fill The MediQA has been called on to provide water to the Heatsan.
6.2
MAIN MENU
Accessed by pressing MAIN MENU button on the bottom of any screen.
Figure 25
6.2.1
VIEWER button
Takes you to the viewer screen shown below.
37 98-0157 Rev K
Figure 26
The VIEWER screen displays a flow schematic and highlights with colored indicators the status of specific process elements.
Red – OFF
Green – ON
Tank level switch are shown
Yellow – OPEN (water level has not triggered float switch)
Green - CLOSED (water level has closed float switch)
MAIN MENU BUTTON Returns to the main menu screen.
6.2.2
ACTIVITY DETAILS button
Takes you to the activity details screen shown below.
Figure 27
38
The ACTIVITY DETAILS screen shows specific performance data relating to system processing elements. It also displays the number of hours the MediQA has been running or has been in standby. These can be reset to zero by pressing
the ZERO button on the RO PARAMETERS screen (see Section 6.5.2). Needs
reset after 999 days.
98-0157 Rev K
6.2.3
ACTIVITY LOG button
Takes you to the ACTIVITY LOG screen shown below.
39
Figure 28
The activity log displays data recorded to the MediQA internal data logger.
Records are displayed in chronological order. Scroll buttons are provided to move through the data shown on screen. Pressing the down button goes to the most recent data, while the up button moves back in time, one page at a time.
This data may be downloaded to a data bar / memory stick. Insert a memory stick into the USB port found on the touch screen mounting arm. Press the
WRITE TO USB button, wait 5 seconds to press the REMOVE USB button. Data is recorded to the memory stick in the form of a .csv file. This can be opened using
Microsoft Excel. Files are saved to the memory stick root directory named
SAMP01. The file name is SA followed by the machine serial number. USB stick must be formatted as FAT16 (2GB max) to be recognized by the touch screen.
6.2.4
STATUS MONITOR button
Takes you to the STATUS MONITOR screen shown below.
Figure 29
98-0157 Rev K
The STATUS MONITOR shows, via bar graphs, an overview of the system. This can be used as guide to when preventative maintenance may be required.
In addition it shows the status of certain system functions (ACTIVITY LOG, RO
HEAT DISINFECT….). The status of which is shown with colored indicators and text stating the function condition. A short cut button to access relevant details appears next to each indicator.
Green – Function operating to specification.
Amber – Warning active
Red – Alarm active
“Warning” alarms are advisory; they do not affect system operation. They indicate that some element of the system needs attention.
“Alarms” stop the system. The system will only restart once the problem giving rise to the Alarm has been corrected.
6.2.5
ALARMS button
Takes you to the alarm screen shown below.
Figure 30
40
When triggered, alarms and warnings are detailed in the table. Alarms can be deactivated by correcting the indicated fault and pressing the ALARM RESET button. Alarms will remain present until the fault is corrected. If the fault is not corrected the alarm cannot be reset with the ALARM RESET button.
The HISTORY button drills down to the ALARM LOG screen.
98-0157 Rev K
6.2.6
ALARM LOG button
Takes you to the alarm log screen shown below.
Figure 31
Alarm log retains all alarms that have been triggered.
Use scroll buttons to move through the data. First press CURSOR ON and when completed CURSOR OFF.
6.2.7
DIAGNOSTICS button
Takes you to the diagnostics screen shown below.
Figure 32
The DIAGNOSTICS screen details active system elements.
Active elements are highlighted with green markers in each of the indicator boxes.
41 98-0157 Rev K
6.3
OPERATION MENU screen
Figure 33
Note: The yellow and red buttons only appear when a given process has been activated.
6.4
SETTINGS MENU screen
The SETTINGS MENU allows key operating parameters and system alarms to be set. When the button is pressed you will be prompted for a password.
Takes you to the set-up menu screen shown on the following page.
The password to enter the SETTINGS MENU is
01844.
Punch in the number and press ENT. If you make an error press ESC.
Figure 34
42
Figure 35
98-0157 Rev K
6.4.1
RO SUPPLY button
Screen shown below for DPH.
Figure 36
Screen shown below for SPH.
Figure 37
43
AUTO RINSE MIN defines the duration of initial rinse. If unit has not rinsed to quality set-point, system will rinse for an additional minute until max time has been reached. If unit has not reached the quality after max time elapsed an alarm is logged.
AUTO FLUSH defines the duration of flush.
TARGET RECOVERY RO1 sets target water recovery of stage 1 RO modules.
TARGET RECOVERY RO2 sets target water recovery of stage 2 RO modules.
SUPPLY HEATING defines whether the tank heater is to be used to warm the incoming water supply. When enabled, the heater is turned on at 48°F (9 °C) and off at 52°F (11°C).
98-0157 Rev K
ENTER under SUPPLY TIME SCHEDULE takes you to the screen shown below and enables automatic operating times to be set. System operation is activated when
TIMER is selected in the OPERATION MENU screen.
Figure 38
System operation to programmed times is activated when TIMER is selected in the OPERATION MENU screen.
Timed operation for the MediQA can be set by programming pre-set times.
To change pre-set times, press the highlighted blue boxes enter the time (24 hour clock) on the pop up key pad and press enter.
Enter a time of 00:00 to prevent operation on any particular day.
ENTER under MEDIQA OPERATION MODE takes you to the screen shown below and allows the user to switch between recovery modes. HI RECOVERY MODE is the default and can be disabled by holding the DISABLE button for two seconds.
Must have level 3 access to change the mode of operation.
44 98-0157 Rev K
Figure 39
HI RECOVERY QUALITY defines the high and low limits of permeate conductivity that the water will maintain. The reject water will recirculate to the tank until the high limit of water quality is reached, when the concentrate to drain valve will open. The concentrate to drain valve will close when the low limit is reached.
BREAK TANK TEMPERATURE defines the high and low limits of the temperature that water will be maintained. The drain valve is opened when the temperature reaches the high limit and closes when the low limit is reached.
45 98-0157 Rev K
6.4.2
RO CLEAN button
Takes you to the RO CLEAN screen shown below.
(See Section 8 before operating this function)
Figure 40
Note: The yellow and red buttons only appear when a given process has been activated.
RECIRC PERIOD define here the duration cleaning chemical is circulated around the system.
RINSE PRESSURE PERIOD LOW defines the duration of a low pressure rinse.
RINSE PRESSURE PERIOD HIGH defines the duration of a high pressure rinse.
CLEANING REMINDER can be set to determine when the next clean should be completed. When the counter has counted down to zero a warning will be triggered to the ALARM LOG box and indicated on the status screen. This reminder will automatically reset after a clean has been carried out.
The TIME REMAINING status bar activates when a CHEMICAL CLEAN is started.
This displays the time remaining to complete each of the clean cycle elements.
YELLOW prompt boxes pop up to instruct when to perform given activities during the clean cycle.
46 98-0157 Rev K
6.4.3
RO HEAT DISINFECT button
Takes you to the RO HEAT DISINFECT screen shown below.
Here specific functions for the RO heat disinfect can be set.
Figure 41
TANK TEMP determines the maximum desired tank temperature during a heat disinfection cycle (max 185°F or 85°C).
TANK RETURN TEMP determines the actual disinfection temperature, maximum
185°F (85°C). Tank return temp should always be set 2°F below tank temp.
HOLD PERIOD is the time water is circulated around the system once it has reached the TANK RETURN TEMP i.e. the disinfection temperature.
TIME REMAINING shows time remaining to complete the hold period.
RO HEAT DISINFECT TIME SCHEDULE takes you to the screen shown below and enables automatic operating times to be set. System operation is activated when
TIMER is selected in the OPERATION MENU screen.
47
Figure 42
98-0157 Rev K
Heat disinfection will be postponed if these times conflict with RO SUPPLY TIMER settings (Figure 35). RO HEAT DISINFECTION will commence when SUPPLY ends.
To change pre-set times, press the highlighted blue boxes, enter the time (24 hour clock) on the pop up key pad and press enter. Enter a time of 00:00 to prevent operation on any particular day.
6.4.4
ACTIVITY LOG button
Takes you to the activity log setup screen shown below.
48
Figure 43
Setting a time to the ACTIVITY LOG CYCLE TIME determines the period by which data is stored to the activity log and data logger.
6.4.5
ALARMS button
Takes you to the alarm setup screen shown below.
Figure 44
98-0157 Rev K
The set points on this screen are triggers for system warnings and alarms. If a warning is triggered it will be displayed and the blue bars surrounding the screen will turn yellow. If an alarm is triggered it will be displayed, the bars will turn red, and the system will cease operation. The warning or alarm is triggered if the value falls below the LOW FLOW set point or above the TANK QUALITY, OUTPUT
QUALITY, or HIGH FLOW set points.
6.4.6
CLOCK button
Takes you to the clock setup screen shown below.
Figure 45
49
Clock set enables the system clock to be set. To adjust the clock press ACCESS
TIME SETTING enter the desired time into the blue boxes adjacent to SYSTEM
CLOCK TIME and press SET SYSTEM CLOCK to set the time.
The time can be advanced 1 hour by pressing SUMMER TIME ADVANCE 1 HOUR or reduced by 1 hour by pressing WINTER TIME SET BACK 1 HOUR.
98-0157 Rev K
6.5
CALIBRATION MENU
The Calibration Menu allows access to the calibration settings and alarm set points for the unit. This menu is used for initial setup and fine tuning of the unit.
Figure 46
The Calibration Menu is password protected and requires a level 2 login (____).
The menu should only be accessed by trained personnel and AmeriWater Techs.
Figure 47
50 98-0157 Rev K
6.5.1
FLOW METERS button
The Flow Meter menu allows access to the calibration settings.
Figure 48
OFFSET and GAIN are calibration parameters. Offset adds or subtracts a constant value to the signal. Gain multiplies the signal by a constant factor.
6.5.2
RO PARAMETERS button
In the RO parameters menu, the target percent recovery for the RO can be set.
The default is 65% for stage 1 and 85% for stage 2 (if the unit is a dual pass).
The run time clock can also be reset from this menu.
Figure 49
51 98-0157 Rev K
6.5.3
ALARM SETTINGS button
The Alarm Settings menu allows access to the alarm set points for various parameters of the unit. The set points should be set before the unit is put into service and will only require alteration in special case situations.
Figure 50
The set points on this screen are triggers for system warnings and alarms. If a warning is triggered it will be displayed and the blue bars surrounding the screen will turn yellow. If an alarm is triggered it will be displayed, the bars will turn red, and the system will cease operation. The warning or alarm is triggered if the value exceeds the Loop, Tank, Permeate, or Tank Return Over Temp set points.
6.5.4
LINE CELLS button
The Line Cell menu allows the user to access the calibrated numbers used to determine the conductivity of the water in the unit. This should not be altered after initial set up of the system.
Figure 51
52
OFFSET and GAIN are calibration parameters. Offset adds or subtracts a constant value to the signal. Gain multiplies the signal by a constant factor.
6.5.5
RO HEAT DISINFECT button
98-0157 Rev K
The RO Heat Disinfect menu allows access to the PUMP RUN SLOW and COOL
DOWN END TEMP.
53
Figure 52
PUMP RUN SLOW TEMP is the set point at which the pump on the system will run at a lower pressure than normal. This is required during a heat disinfection of the system because the membranes cannot handle high pressures with water over 113 degrees F at >45 psi of pressure.
COOL DOWN END TEMP is the temperature at which the cool down cycle will end after a heat disinfection of the RO. At this temperature, the RO will be able to supply water to the loop.
6.5.6
CHEMICAL CLEAN button
The Chemical Clean menu allows access to the CHEM CLEAN RINSE set point.
This set point controls the amount of time the unit will perform a rinse to ensure all chemicals introduced to the RO are gone before supplying water to the loop.
Figure 53
98-0157 Rev K
6.5.7
OVERRIDES button
The Overrides menu allows various components of the RO to be turned on while the unit is in standby. This is mainly used for testing or diagnostics purposes for the system. The overrides can only be used while the unit is not in operation. To reset any manual override use the OVERRIDE RESET button.
Figure 54
54
6.5.8
DIAGNOSTICS button
Takes you to the diagnostics screen shown below.
Figure 55
The DIAGNOSTICS screen details active system elements.
Active elements are highlighted with green markers in each of the indicator boxes.
98-0157 Rev K
6.5.9
SYSTEM button
The system menu allows access to the software versions currently installed in the
MediQA.
Figure 56
INSTALL FACTORY DEFAULTS will erase all current settings on the unit and
replace them with the factory default values listed in Section 13.1. This button
should not be pressed unless a critical problem with the system occurs. Requires level 6 access (restricted to AmeriWater service personnel) and 2 second hold to activate.
55 98-0157 Rev K
7.
FAULTS AND FAULT FINDING
ALARMS will stop machine operation. It is not possible to restart the machine without addressing the fault.
Alarm
Description
Trigger Review and Check Corrective Action
ALARMS
RO1 Pump Over
Pressure Alarm
1) Check valve status
2) Perform chemical clean,
RO2 Pump Over
Pressure Alarm
(DPH only)
RO1 Pump Fault
Alarm
RO2 Pump Fault
Alarm (DPH only)
Temperature
High Alarm
Pump pressure transducer
Pump pressure transducer
Loop Over
Temp Alarm
Set-point
Exceeded
1) RO1 Membranes are blocked
2) Concentrate valve on RO1 closed
3) Pressure switch faulty
1) RO2 Membranes are blocked
2) Concentrate valve on RO2 closed
3) Pressure switch faulty
1) Check and inspect pump RO1
1) Check and inspect pump RO2
1) Check valve status
2) Perform chemical clean,
1) Contact AmeriWater service
1) Contact AmeriWater service
1) Check incoming water supply
Output Quality
Alarm
Tank Quality
Alarm
Water Loss
Alarm
Over
Temperature
Fault Alarm
Emergency Stop
Activated Alarm
PLC Analogue
Card Error Alarm
PLC to PCB
Comms Error
Alarm
Output High
Pressure Alarm
Output Quality
Alarm Setpoint
Exceeded
Tank Quality
Alarm Setpoint
Exceeded
Base Tank
Float Switch
Tank temperature probe
E Stop
PLC
Delivery High
Pressure
Switch
1) Supply to loop over limit - check setting in CALIBRATION MENU > ALARM
SETUP screen LOOP OVER TEMP ALARM
2) Softened Potable incoming water supply too hot
1) Rinse period exceeded max time RO
PARAMETERS SCREEN AUTO RINSE
2) Loop supply quality over range
3) Loop supply line cell error
1) Tank line cell out of range RO SUPPLY screen TANK QUALITY settings
2) Tank supply quality over range RO
SUPPLY screen LOOP QUALITY settings
1) Tank empty - check water supply to machine
2) Check tank ball valve operational
3) Check tank float switches
4) Check inlet solenoid valve
1) Tank over temperature check
2) Check loop return temp
3) Check loop supply temp
1) Reset E stop on Control Panel
2) reset E stop on touch screen
Contact AmeriWater
Contact AmeriWater
1) Check if valve in the loop is closed
2) Check Permeate M. Valve is open
3) Verify Permeate relief valve is working
1) Check line cell
2) Perform chemical clean
1) Drain tank refill and perform extra rinse
2)Check line cell
Contact AmeriWater
Contact AmeriWater
1) Increase setting on pressure switch using adjustment screw
56 98-0157 Rev K
Loop Return
Valve Position
Alarm
Permeate Valve
Position Alarm
Heater Over
Temperature
Alarm
Feed Water
Alarm
Loop Return
M. Valve
Permeate M.
Valve
Thermostat
1) Loop Return M. Valve is out of position
2) Limit switch failure
1) Valve is out of position
1) Adjustment for thermostat
2) Tank over temperature
Output Quality
Rinse Alarm
Output Flow
Rate High Alarm
Output Flow
Rate Low Alarm
Float Switch
Delivery
Permeate Line
Cell
High Flow
Alarm Setpoint
Exceeded
Low Flow
Alarm Setpoint
Exceeded
Heatsan
Alarm
1) Check Tank level is actually empty
2) Check feed water supply
3) Verify inlet solenoid valve is working
4) Check strength of flow
5) Check if diffuser is plugged
1) Use handheld conductivity meter to check calibration
2) Check membrane conductivity levels
1) Check calibration of RO
1) Check calibration of RO
1) Replace membrane
Heatsan Unit -
Water Loss
Alarm Detected
Warnings are advisories. The machine will not stop but action should be taken to correct the fault.
Warning
Description
Trigger Review and Check Corrective Action
Time Date Not
Set Warning
Empty schedule fields
1) Leakage
2) Control valve failure
WARNINGS
1) RO Supply Schedule and RO Heat
Disinfect Schedule have no times set for
Mon thru Sunday setups. An Operator put the RO unit into Timer Mode with no
1) Program times to schedule
2) Run in Continuous Mode
Operation Reset
Error Warning
Feed Tank Low
Level Warning
Level Sensor
Error Warning
Function button
Float Switch
(second from bottom)
Float switches have moved out of sequence schedules set.
1) Function button depressed more than
3 times in 3 minutes.
1) Tank empty
2) Check water supply to machine
3) Check valve positions
4) Check for leaks
1) Float valve in tank faulty
2) Float valve in tank "hung up"
3) Check PLC for X6 - High Level; X5 -
Mid Level; X4 - Low Level; X3 - Base
Level
1) Alarm Reset
1) Check float switches.
57 98-0157 Rev K
High Pressure
RO1 Warning
(DPH only)
High Pressure
RO2 Warning
(DPH only)
Output High
Pressure
Warning (DPH only)
High Water
Temperature
Warning
Output Quality
Warning
Output Quality out of Range
Warning
Output Line Cell
Error Warning
Tank Quality
Warning
Tank Quality out of Range
Warning
Tank Line Cell
Error Warning
Loop Supply
Sampling Error
Warning (DPH only)
Loop Flow Rate
High Warning
(SPH only)
Pressure
Switch (PS2)
X7
Pressure
Switch (PS3)
X11
Pressure
Switch (PS1)
X12
1) RO1 Membranes are blocked / failed
2) Concentrate valve on RO1 closed
3) Check pressure switch set-point
1) RO2 Membranes are blocked / failed
2) Concentrate valve on RO1 closed
3) Check pressure switch set-point
1) Permeate valve relief set too high
2) Check pressure switch set-point
Loop Over
Temp
Warning Setpoint
Exceeded
Output
Quality
Warning Setpoint
Exceeded
Permeate
Conductivity
Sensor
1) Check set point
2) Check loop temperature
1) Check valve status
2) Perform chemical clean
3) Adjust set-point
1) Check valve status
2) Perform chemical clean
3) Adjust set-point
1) Adjust valve Max 30Psi
2) Adjust set-point
1) Call AmeriWater
1) Check conductivity with myron-l meter 1) Check line cell
2) Perform chemical clean
3) Replace membrane
Permeate
Conductivity
Sensor
Tank Quality
Warning Setpoint
Exceeded
Tank
Conductivity
Sensor
Tank
Conductivity
Sensor
Permeate
Conductivity
Sensor
High Flow
Warning Setpoint
Exceeded
1) Check set point
2) Check sensor and wiring
3) Check conductivity board
4) Check cables between conductivity board
1) Check set-point
2) Check sensor and wiring
3) Check conductivity board
4) Check cables between conductivity board
1) Check conductivity with Myron-l meter
2) Check water softener and any pretreatment tanks
3) Check tank sensor and wiring
1) Check set-point
2) Check sensor and wiring
3) Check conductivity board
4) Check cables between conductivity board
1) Check set-point
2) Check sensor and wiring
3) Check conductivity board
4) Check cables between conductivity board
1) Loop supply line cell error
2) Perform heat disinfection
1) Check line cell
2) Perform chemical clean
1) Check line cell
2) Perform chemical clean
1) Check line cell
2) Perform chemical cleaning
1) Check line cell
2) Perform chemical clean
1) Check line cell
2) Perform chemical clean
1) Check line cell
2) Replace membrane
1) Check set-point
2) Check loop flowmeter reading on
Activity Details screen
3) Check wiring to flowmeter
58 98-0157 Rev K
Loop Flow Rate
Low Warning
(SPH only)
Heater Contactor
Fault Warning
Loop Return
Valve Position
Warning
Permeate Valve
Position Warning
Low Flow
Warning Setpoint
Exceeded
Heater
Contactor not closed
Loop Return
Valve not closed
Permeate
Valve not closed
4) With no HD machines running, check loop return flow meter and compare to screen reading
1) Check set-point
2) Check loop flowmeter reading on
Activity Details screen
3) Check wiring to flowmeter
4) With no HD machines running, check loop return flow meter and compare to screen reading
1) Check the controller to validate that output Y4 has fired.
2) Check input X10
3) Check contactor for closure
1) Check the controller to validate that output Y23 has fired.
2) Check input X22
3) Check valve for closure
1) Check the controller to validate that output Y14 has fired.
2) Check input X13
3) Check valve for closure
1) review chemical clean clock setting
1) Replace as necessary
1) Replace as necessary
1) Replace as necessary
Chemical Clean
Overdue
Warning
RO Heat
Disinfect 3 Failed
Attempts
Warning
Pretreatment
Lockout Enabled
X24
Clean
Reminder
Set-point
Heat
Disinfect not carried out
RO1 Quality
Over Range
Warning (DPH only)
Output Line Cell
Error Warning
(DPH only)
Leak Detected
Warning
Temperature
Rise Warning
1) Review alarm log
2) Check RO Heat Disinfect is not manually aborted.
1) Conduct chemical clean
1) Complete manual RO Heat
Disinfect
Stage 1
Permeate
Conductivity
Sensor
Stage 1
Permeate
Conductivity
Sensor
Conductivity
Board
Break Tank
Temperature
Probe
1) Check input for illumination and wiring
2) Check head on unit that is backwashing
1) Check set-point
2) Check sensor and wiring
3) Check conductivity board
4) Check cables between conductivity board
1) Check set-point
2) Check sensor and wiring
3) Check conductivity board
4) Check cables between conductivity board
1) Check connection between board and
PLC
2) Check conductivity sensors and wiring
3) Check plug-in locations
1) After 30 min break tank temp is lower than previous reading during RO Heat
Disinfect. Check sensor and wiring
2) Check sensor is inserted properly
1) Check line cell
2) Perform chemical clean
1) Check line cell
2) Perform chemical clean
59 98-0157 Rev K
3) Check temperature card for power and wiring
Temperature
Hold Low
Warning
Break Tank
Temperature
Probe is lower than
Tank Return
Probe
1) Check the temp in the tank with a calibrated measuring device
2) Check the return temperature
3) Check the wiring to the temperature expansion card
4) Check probe is inserted all the way
Note: The pumps will run slow during the RO Heat Disinfection and chemical clean process. If the pumps are running slow in normal operation, ensure that the water temperature is below 86 °F.
Note: If the permeate to the loop is over the alarm set-point, the MediQA will attempt to rectify this by closing the permeate valve and performing a rinse. If this does not bring the quality within the allowable range, the RO will shut down and you will have a permeate quality alarm.
60 98-0157 Rev K
8.
SANITIZING & CLEANING
It is recommended that your MediQA unit be chemically cleaned at regular intervals to ensure performance is maintained. The RO membranes should be chemically cleaned when there is a significant decrease in product water flowrate or an increase in product conductivity up to output quality warning limit.
It is recommended that RO heat disinfection is completed at least weekly in order to protect against possible bio-film formulation.
8.1
EXTERIOR SURFACE CLEANING
When cleaning the exterior surfaces of the device, it is recommended that you use a soft non-marking cloth dampened with water. Do not use chemical cleaning agents. If it is necessary to use a chemical cleaning agent, contact
AmeriWater for permission prior to use.
8.2
CHEMICAL CLEANING
Test equipment required.
• Hand held conductivity meter or chemical test strips
• Approved cleaning chemical
The quantity of chemical required is calculated according to the volume of water in the MediQA, taking care not to exceed the maximum concentration stated.
STIR SOLUTION WELL WHEN USING AMERICLEAN A AND B POWDER.
MediQA water volumes – when full
00MSP2-230 = 30 gallons
00MSP3-230 = 32 gallons
Chemical Volumes:
Unit
00MSP2-230
00MSP3-230
Water Volume
(Gal)
30
32
AmeriClean A
Liquid
P/N: 95810140
(ml)
2839
3028
00MDP2-230 = 32 gallons
00MDP4-230 = 37 gallons
AmeriClean B
Liquid
P/N: 95810141
(ml)
2839
3028
AmeriClean A
Powder
P/N: 37-0004
(ml)
AmeriClean B
Powder
P/N: 37-0005
(ml)
800 (3.0 lbs.) 1110 (3.0 lbs.)
800 (3.0 lbs.) 1110 (3.0 lbs.)
00MDP2-230
00MDP4-230
32
37
3028
3502
3028
3502
800 (3.0 lbs.) 1110 (3.0 lbs.)
1064 (4.0 lbs.) 1488 (4.0 lbs.)
AmeriClean A/B Liquid 1:40; AmeriWater recommends 950 ml per 25 Gal.
AmeriClean A Powder 1 lbs. (266 ml) to 10 gal water; AmeriClean B Powder 1 lbs. (372 ml) to 10 gal water. Case will include 10-1 lbs. bags of cleaner.
Recommended cleaning regimes:
Iron, Bacteria, Scale: AmeriClean A.
For hard water areas even on systems operating on a softened water supply it is recommended that the unit is chemically cleaned using AmeriClean A.
61 98-0157 Rev K
For soft water areas, cleaning using AmeriClean B, organic removal formulation is recommended in place of AmeriClean A.
Warning
Failure to test sample may lead to chemicals entering the
distribution loop.
Please read the appropriate material safety data sheets and always wear the recommended personnel protective equipment before
handling any of the chemicals.
Danger
Caution
Warning
(i) Warn all end users that chemical cleaning is taking place.
(ii) Stop dialysis, check with Renal technicians.
(iii) Isolate distribution loop main or other systems by closing
appropriate manual isolation valves.
Wear personal protective equipment when conducting the chemical
clean process EYE WEAR, CLOTHING, GLOVES.
Method:
Chemical cleaning of the unit can be initiated from the OPERATION MENU of the touch screen.
Stop the MediQA by pressing the OFF button on the OPERATION MENU. Depress
CHEMICAL CLEAN button
Wait for prompt to ADD CLEANING CHEMICAL. Remove the lid of the tank and add chemical to tank. Slowly add the chemical in the amount listed in the above table. IF USING POWDER, AGITATE THE SOLUTION UNTIL ALL OF THE
POWDER HAS DISSOLVED PRIOR TO ADDING TO BREAK TANK. Replace the tank lid.
Depress ADD CLEANING CHEMICAL button. Once added, the cleaning chemical will be automatically circulated around the MediQA.
On completion of the automatic circulation and drain down / flush, a prompt
TEST FOR RESIDUE CHEMICALS is displayed.
A permeate water sample must be taken and tested for any presence of cleaning chemical. Check with renal clinical / technicians that water quality is acceptable for use.
If acceptable for use depress TEST FOR RESIDUE CHEMICALS once the sample is found to be satisfactory.
62 98-0157 Rev K
A final prompt is displayed PASS WATER QUALITY. Depress to accept water quality. This acknowledges acceptance that the samples have passed water quality test (see above). If water does not meet standards additional rinses are preformed until it does.
If any alarm messages are displayed during the clean then refer to the fault finding section in this manual.
8.3
APPROVED CHEMICALS
AmeriWater supplies a comprehensive range of chemicals for management and cleaning of RO membranes and distribution loops. Clean with AmeriClean B first then A.
The following chemicals are approved for use on the MediQA:
AmeriClean A
This is a low pH RO membrane cleaner to remove hardness scale and iron oxide deposits.
To test for the presence of AmeriClean A in permeate, check that the conductivity is less than 30μS/cm and pH is at or near 7.
AmeriClean B
This is an alkaline liquid cleaner containing organic sequestrant, inorganic sequestrant, detergent and emulsifiers. AmeriClean B is used for removing organics, Silica, inorganic colloids, and biological material.
To test for the presence of AmeriClean B in permeate, check that the conductivity is less than 30 μS/cm and pH is at or near 7.
NOTE: Always refer to Safety Data Sheets before handling any of the cleaning chemical products. Wear the safety equipment recommended.
Follow the instructions on the pack.
8.4
RO HEAT DISINFECTION PROCEDURE
A heat disinfection routine is provided to kill waterborne bacteria. Heat disinfection, when completed regularly, reduces the potential for harmful biofilms to generate within the machine pipe work and system components.
The MediQA unit can be set up to complete automatic heat disinfections or heat disinfections can be started manually.
63 98-0157 Rev K
8.4.1
Manual RO Heat Disinfection
From the OPERATION MENU select RO HEAT DISINFECTION. Heat disinfection
settings must be established in the SETTINGS MENU (see Section 0).
Manual heat disinfections can be aborted by pressing the emergency stop, pressing the OFF button, or switching the main isolator to off.
When the system is restored (switched back ON via isolator or resetting
Emergency stop) the system may contain HOT water (subject to the duration of OFF time). When either CONTINUOUS or TIMER operation is selected the system will first perform a cool down routine before going into service (i.e. supply of water to the distribution loop).
8.4.2
Automatic RO Heat Disinfection
Heat disinfections can be started automatically when times have been
programmed to the RO heat disinfection timer schedule (see Section 6.4.3).
Automatic heat disinfections will only commence if the unit is running in TIMER operation.
64 98-0157 Rev K
9.
MAINTENANCE
There are no user replaceable parts on the MediQA.
The MediQA is a medical device; as such modifications to the device are not permitted. Modifications to the machine by anyone other than AmeriWater
personnel will invalidate marketing clearance for the device.
For all service enquiries contact AmeriWater at:
Tel No: 1-937-461-8833
Fax No: 1-937-461-1988
9.1
THERMOSTAT TESTING
AmeriWater recommends that the thermostat on the heating element be tested annually by qualified personnel. To conduct this testing, you will need access to the device when no patients are under treatment.
To begin, remove the lid from the MediQA storage tank. Gain access to the calibration
menu and log in (see Section 6.5). Access the OVERRIDES menu and open the fill
solenoid valve. Allow the tank to fill to between 50-75% and close the fill valve. Ensure that the heater is completely submerged.
Place a known good thermometer into the tank away from the heater. Turn on the heater on the OVERRIDES page.
During this testing, the temperature of the water will be sufficient to
Caution create steam, which is hot enough to scald. Ensure that you will be able to determine the temperature without reaching into the tank.
Verify that the heater shuts off between 185 – 210° F.
If the thermometer on the heater does not shut off at the prescribed temperature, contact AmeriWater for guidance.
Navigate to the OVERRIDES screen and open the fill solenoid valve and allow the tank to completely fill. Close the fill solenoid. Doing this will reduce the temperature of the water. Open the drain valve and allow the tank to completely drain. Close the drain valve and remove the thermometer from the tank and replace the lid.
65 98-0157 Rev K
9.2
FAULTY ITEMS
Should a faulty item be identified please notify our service department of the item part number (see table below).
Description
Heater
Pump*
Membranes
Flush solenoid
Inlet sol. 1”
Concentrate return motorized valve
AmeriWater
Part Number
Qty
38-0001 1
Model dependent 1
22-8041
59-0034
911-059-0004
51-0028
1
1
1
1
User replacemen
t code
AT
AT
AT
AT
AT
AT
Batch controlle
d spare
No
No
No
No
No
No
Permeate motorized valve
Loop return motorized valve
Pressure gauge
Level switch
Pressure switch
Pressure transducer
Flow transducer
Feed line cell
Permeate line cell
Temp transducer
51-0027
51-0031
430001
See below
See below
See below
See below
See below
See below
See below
1
1
1
1
1
1
1
1
1
1
AT
AT
AT
P
P
P
P
P
P
P
No
No
No
No
No
Yes
Yes
Yes
Yes
No
* The pumps by model are listed in the table below.
Model
00MSP2-230
00MSP3-230
00MDP2-230
00MDP4-230
Pass #1
CRNE 5-20
CRNE 5-20
CRNE 5-15
CRNE 5-24
Pass #2
N/A
N/A
CRNE 3-17
CRNE 3-25
66 98-0157 Rev K
9.3
MEDICAL DEVICE TRACEABLE COMPONENTS
Description
Control Panel
Tank Line Cell
Pressure Switch
Pressure Switch
Product Line Cell
Tank / Tank Return Temp
Sensor
Permeate Temp Probe
Loop Return Temp Probe
Tank Float Switch
Conductivity Board
Pressure Transducer
Permeate Flow Meter
Concentrate Flow Meter
Recirculation Flow Meter
AmeriWater
Part Number
Contact AmeriWater
69-0030
68-0008
68-0006
69-0029
39-0004
39-0005
39-0007
67-0012
69-0098
68-0009
41-0027
41-0026
041-9007
User replacement
code
AT
AT
AT
AT
AT
AT
AT
AT
AT
AT
AT
AT
AT
AT
Batch controlled
spare
Yes
Yes
No
No
Yes
No
No
No
No
Yes
Yes
Yes
Yes
Yes
AT: AmeriWater technician level (AmeriWater trained Technician: special tools required plus additional instructions outside of the scope of this manual. Only AmeriWater trained technicians must replace parts with this assigned code for health and safety reasons. These parts are critical part of the system.)
The MediQA is classified as a medical device. Any person replacing a batch controlled spare, as identified above must, as a matter of due diligence, request from AmeriWater, that it is supplied with a clearly identifiable batch number. Also the authorized person who will install the component must quote the serial number of the machine in which it will be installed at order placement.
The replacement of the batch controlled spare must be recorded using the log sheet (below).
This must be done in order to maintain the integrity of the machine throughout its operational life in accordance with medical device regulations.
If batch controlled parts are not installed, or are not installed by approved AmeriWater
Technicians, then the responsibility for any subsequent incidents linked to the replaced part or affected parts rest with the buyer / hospital / maintenance staff or service provider.
Failure to install AmeriWater supplied replacement parts will invalidate the warranty.
67 98-0157 Rev K
9.4
REPLACEMENT PART HISTORY LOG SHEET
MediQA Replacement Parts Record
Unit Serial
No.
Location
Failed Part
Date
Replaced
Description
Batch
Control
Y/N
AMW*
CRN No.
Return to AMW
Y/N
Reason for replacement
New Part
New Part
Batch No.
AMW
Tech
Initials
68 98-0157 Rev K
10.
FLOW DIAGRAMS
SPH UNITS
69 98-0157 Rev K
DPH UNITS
70 98-0157 Rev K
11.
WIRING SCHEMATIC
SPH UNITS
71 98-0157 Rev K
72 98-0157 Rev K
73 98-0157 Rev K
74 98-0157 Rev K
75 98-0157 Rev K
76 98-0157 Rev K
77 98-0157 Rev K
78 98-0157 Rev K
DPH Units
79 98-0157 Rev K
80 98-0157 Rev K
81 98-0157 Rev K
82 98-0157 Rev K
83 98-0157 Rev K
84 98-0157 Rev K
85 98-0157 Rev K
86 98-0157 Rev K
12.
COMPONENT IDENTIFICATION
Touch Screen Display
Feed Water Tank
Emergency Stop
Electrical Panel
RO Modules (membranes in housings)
87
High Pressure Pump
98-0157 Rev K
13.
APPENDIX
13.1
DEFAULT SETTINGS LIST FOR SINGLE AND DUAL PASS MediQA
SINGLE PASS:
Setting
Auto Flush
Auto Rinse Min
Auto Rinse Max
Target Recovery Stage 1
Standby Refresh Time
Standby Refresh Interval
Hi Recovery Quality Low
Hi Recovery Quality High
Break Tank Temp Low
12
80
Break Tank Temp High 90
Ro Heat Disinfection
Setting
Tank Temp
Tank Return Temp
Hold period
Setting
Clean Reminder
1
MediQA Settings SPH
RO Supply
Default Initial Setup User Setting Units min
Setting
Tank Quality Warning
1
10
65
10
120
8
Default Initial Setup User Setting Units
185
180
30
RO Clean min min
% min min uS uS
°F
°F
°F
°F min
Default Initial Setup User Setting Units
999 days
Tank Quality Alarm
Output Quality Warning
Output Quality Alarm
Low Flow Warning
Low Flow Alarm
High Flow Warning
High Flow Alarm
Setting
Concentrate offset
Concentrate Gain
Permeate Offset
Permeate Gain
Setting
Loop Over Temp Warning
Recirc Period
Rinse Pressure Period Low
Rinse Pressure Period High
25
40
15 min min min
Loop Over Temp Alarm
RO Heat Disinfect Over Temp
Tank
RO Heat Disinfect Over Temp permeate 2
RO Heat Disinfect Over Temp
Tank return
Alarm Setup
Default Initial Setup User Setting
1100
1200
30
100
1
2
15
18
Flow Regulation
Units uS
Default Initial Setup User Setting Units
200
4.3
350
130
Engineers Alarm Setup uS uS uS gpm gpm gpm gpm
Default Initial Setup User Setting Units
86 °F
95 °F
194
183
183
°F
°F
°F
Setting
Tank Offset
Tank Gain
Output Offset
Output Gain
Setting
RO1 Target Recovery
Dead Band
Line Cell Calibration
Default Initial Setup User Setting Units
0
90
0
95
RO Parameters
Default Initial Setup User Setting Units
65
20
%
%
Setting
Engineers RO Heat Disinfect Setup
Cool Down End Temp
Pump Run Slow Temp
Default Initial Setup User Setting
86
90
Engineers Chem Clean Setup
Units
°F
°F
Setting
Chem Clean Rinse
Default Initial Setup User Setting Units
45 min
88 98-0157 Rev K
DUAL PASS:
Setting
Auto Flush
Auto Rinse Min
Auto Rinse Max
Target Recovery Stage 1
Target Recovery Stage 2
Standby Refresh Time
Standby Refresh Interval
Setting
Tank Temp
Tank Return Temp
Hold period
Setting
Clean Reminder
Recirc Period
Rinse Pressure Period Low
Rinse Pressure Period High
Setting
Tank Offset
Tank Gain
Stage 1 Offset
Stage 1 Gain
Output Offset
Output Gain
Setting
RO1 Target Recovery
RO2 Target Recovery
Dead Band
RO1 Flush Valve Position
Setting
Tank Quality Warning
Tank Quality Alarm
Output Quality Warning
Output Quality Alarm
MediQA Settings DPH
RO Supply
Default Initial Setup User Setting Units
1
1
10 min min min
Setting
R01 Concentrate offset
R01 Concentrate Gain
R02 Concentrate Offset
65
85
10
%
% min
R02 Concentrate Gain
R02 Permeate Offset
R02 Permeate Gain
120
Ro Heat Disinfection min Interstage Pressure Offset
Interstage Pressure Gain
Default Initial Setup User Setting Units RO1 Concnetrate Valve Limits Min
185
180
240
°F
°F
RO1 Concnetrate Valve Limits Max
RO2 Concnetrate Valve Limits Min min RO2 Concnetrate Valve Limits Max
Flow Regulation
Default Initial Setup User Setting Units
217.3
1.132
217
3.487
316
5.6
0
1
15
75
15
75
RO Clean
Default Initial Setup User Setting Units
1
25
40
15
Line Cell Calibration days min
Default Initial Setup User Setting Units
0
90
5
1.5
0
0.95
RO Parameters min min
Default Initial Setup User Setting Units
75
85
20
75
Alarm Setup
%
%
%
%
Default Initial Setup User Setting Units
1100 uS
1200
30
100 uS uS uS
Concentration Valve Settings
Setting
Conc Control Valve 1 RO Heat
Disinfection
Conc Control Valve 2 RO Heat
Default Initial Setup User Setting Units
0
0
Disinfection
Conc Control Valve 1 Chem Clean
Conc Control Valve 2 Chem Clean
0
0
Setting
Engineers Alarm Setup
Default Initial Setup User Setting Units
Loop Over temp Warning
Loop Over temp Alarm
RO Heat Disinfect Overtemp Tank
RO Heat Disinfection Over Temp
Permeate 2
RO Heat Disinfection Over Temp
Tank return
R02 Pressure setpoint Low
86
95
194
183
183
RO2 Pressure setpoint High
Loop Supply Flow Setpoint Low
14
725
1
Loop Supply Flow Setpoint High
Low Pressure Alarm Delay
18
0
Engineers RO Heat Disinfection Setup
°F
°F
°F
°F
°F psi psi gpm gpm hrs
Setting
Cool Down End Temp
Pump Run Slow Temp
Default Initial Setup User Setting Units
86 °F
90
Engineers Chem Clean Setup
°F
Setting
Chem Clean Rinse
Default Initial Setup User Setting Units
45 min
‘Default’ refers to the settings that will be entered if the “Install Factory Defaults” button is pressed
‘Initial Setup’ refers to the settings that are entered in the system when it leaves the factory.
‘User Settings’ refer to the settings on the unit when it is put into service. This should be filled out when the unit is put into service and should be updated any time settings change
89 98-0157 Rev K
13.2
IFU – Updating Mitsubishi FX3G PLC Software with the FX3G-
EEPROM-32L Module
(IFU # 98-9078A)
BACKGROUND:
The Mitsubishi FX3G-EEPROM-32L module offers a means of updating the Mitsubishi FX3G PLC software without the need of a PC, programming cabling or special software.
PROCEDURE FOR SOFTWARE UPDATING:
WARNING: This procedure requires that the electrical panel door be open while power is on. Always use caution while working in live electrical panels.
Refer to the ‘NAMES AND FUNCTIONS OF PARTS’ section of this document for pictorial references.
1.
Navigate to the unit’s SOFTWARE VERSIONS HMI screen. Write down current PLC software version.
MediQA:
CALIBRATION MENU (Password Required)
SYSTEM – “SYSTEM VERSION”
Heatsan:
CURRENT SETUP (Password Required)
SYSTEM SETUP
FACTORY SETUP
NEXT PAGE (until FACTORY SETUP PAGE 5 is reached)
SOFTWARE VERSIONS – “PLC”
2.
Turn the panel power OFF via the power disconnect switch.
3.
Verify that the PLC Power Status LED is OFF.
4.
Place the PLC RUN/STOP switch in the STOP position.
5.
Remove plastic cover from expansion board on the PLC.
6.
Gently install the FX3G-EEPROM-32L module into the Optional Equipment Connector.
7.
With the electrical panel door open, turn the panel power ON.
8.
Verify that the PLC Power Status LED is ON.
9.
Verify the PLC Run Status LED is OFF.
10.
Raise the FX3G-EEPROM-32L module’s front cover.
11.
Press the [WR] key one time. The [WR] LED will illuminate.
12.
Press the [WR] key again. The [WR] LED will blink. Writing is executing. Depending on program size, the program writing can take several seconds to complete. The [WR] LED will blink as long as data is written. When the [WR] LED stops blinking the writing is complete.
13.
When writing is complete, turn the panel power OFF.
14.
Verify that the PLC Power Status LED is OFF.
15.
Remove the FX3G-EEPROM-32L module from the expansion board by gently pulling on the front cover.
16.
Install the plastic cover back on to the expansion board.
17.
Place the PLC RUN/STOP switch in the RUN position.
18.
Turn the panel power ON.
19.
Verify that the PLC Power Status LED is ON.
90 98-0157 Rev K
20.
Verify the PLC Run Status LED is ON.
21.
Shut and secure the panel front door.
22.
Navigate to the unit’s SOFTWARE VERSIONS HMI screen. Verify that the PLC software version matches the version upgrade documentation.
NAMES AND FUNCTIONS OF PARTS:
PLC
[4]
[3]
[1]
[2]
[1] PLC Power Status LED:
[2] PLC Run Status LED:
[3] RUN/STOP switch:
Illuminated green while power is on the PLC.
Illuminated green while the PLC is running.
Starts or stops PLC code execution.
[4] Optional Equipment Connector: FX3G-EEPROM-32L module connection point.
FX3G-EEPROM-32L module
91 98-0157 Rev K
[1]
[3]
[2]
[1] Front Cover: Can be used to remove module from PLC.
[2] [WR] Pushbutton: Initiates module writing to PLC.
[3] [WR] LED: Writing indicator.
92 98-0157 Rev K
13.3
IFU – Updating Mitsubishi GOT1000 HMI Software w/Jump Drive
(IFU # 98-9083A)
BACKGROUND:
The Mitsubishi GOT1000 HMI has a standard USB port on the back of the unit that will allow the transfer of a project file from a jump drive. For field upgrades, this procedure is an easy way for a customer to upload a program without the use of a laptop, proper HMI software, or the correct communication cable. A customer can perform the upgrade, without an onsite field service tech from
AmeriWater Inc. being present. The USB port for the HMI is currently brought out to the front of the control panel as an access point.
REQUIRED ITEMS:
Flash Drive With Current project release of the HMI program ready for loading
PROCEDURE FOR SOFTWARE UPDATING:
1.
Locate the USB port that is accessible from the front of the control panel.
93 98-0157 Rev K
2.
Plug the Jump Drive into the USB port
3.
Press the Upper Left Corner of the Touch Screen. The Screen should change to the following:
94 98-0157 Rev K
4.
Press the Icon for Program/data control. The screen will switch to the following:
5.
Press the P icon for Project Information. The Screen will switch to the following:
95 98-0157 Rev K
6.
Press E: USB Drive. The screen will switch to the following:
96
7.
Press the Download Button. The screen will switch to the following:
98-0157 Rev K
8.
Press OK Button. The screen will switch to the following:
9.
When the download is complete the screen will automatically switch to the following:
97 98-0157 Rev K
10.
Press the OK button. The screen will change to the following:
11.
The screen is now rebooting and will be back in operation with an updated program. Pull the jump drive from the USB port and store it as a backup.
98 98-0157 Rev K
WARNING
This product can expose you to chemicals such as vinyl chloride (used in the production of PVC) or Nickel (used in the production of stainless steel), that are known to the State of California to cause cancer. For more information go to www.P65Warnings.ca.gov.
Dear Valued Customer,
California Proposition 65 (Prop 65) is the Safe Water and Toxic Enforcement Act of 1986. The
State of California began enforcing amendments to California Prop 65 at the end of August
2018. Prop 65 requires manufacturers to provide a clear and reasonable warning to residents of
California about chemicals used in products that they purchase that are included on the Prop 65
Chemical List. The chemicals included on the list are chemicals that are known to the State of
California to cause cancer, birth defects, or other reproductive harm. One such chemical is Vinyl
Chloride, a compound used to produce Polyvinyl Chloride (PVC). The AmeriWater system you have purchased may contain PVC or stainless steel parts.
While warnings are only required in the State of California, AmeriWater has initiated the use of
Prop 65 labeling for all products to ensure compliance with California regulations. Please note that the above warning does not necessarily mean that the product that you have purchased is unsafe. Products that have been cleared for market by FDA have been determined to be safe and effective by the United States Food and Drug Administration. The warning is simply a requirement by the State of California. If you wish to obtain additional information, please visit: p65warnings.ca.gov. You may also contact your AmeriWater representative if you have any questions.
Thank you for your understanding and we look forward to continuing to serve you.