Instruction Manual

HI 98188

USP Compliant

EC, TDS, NaCl, Resistivity,

Temperature Meter

HI 98188

EC/TDS/NaCl/Resistivity Meter

www.hannainst.com

1

Dear Customer,

Thank you for choosing a Hanna Instruments product.

Please read this instruction manual carefully before using the instruments.

This manual will provide you with the necessary information for correct use of the instruments, as well as a precise idea of their versatility.

If you need additional technical information, do not hesitate to e-mail us at [email protected] or turn to the back cover for our worldwide contact list.

These instruments are in compliance with directives.

The HI 98188 is guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.

Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.

If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.

WARRANTY .......................................................................................................................................... 2

PRELIMINARY EXAMINATION ................................................................................................................. 3

GENERAL DESCRIPTION ......................................................................................................................... 4

FUNCTIONAL DESCRIPTION .................................................................................................................. 5

SPECIFICATIONS .................................................................................................................................. 7

OPERATIONAL GUIDE ........................................................................................................................... 9

AUTORANGING ................................................................................................................................... 14

TEMPERATURE COMPENSATION ........................................................................................................... 15

CONDUCTIVITY VERSUS TEMPERATURE CHART ....................................................................................... 16

USP MEASUREMENT .......................................................................................................................... 17

USP MODE PROCEDURES .................................................................................................................... 21

USER CALIBRATION ............................................................................................................................ 25

GOOD LABORATORY PRACTICE (GLP) ..................................................................................................... 33

SETUP .............................................................................................................................................. 35

LOG ON DEMAND .............................................................................................................................. 47

AUTOLOG ......................................................................................................................................... 50

AutoEnd ............................................................................................................................................ 52

TEMPERATURE CALIBRATION (for technical personnel only) ..................................................................... 53

PC INTERFACE ................................................................................................................................... 56

BATTERIES RECHARGING/REPLACEMENT ................................................................................................ 63

TROUBLESHOOTING GUIDE ................................................................................................................ 65

PROBE MAINTENANCE ........................................................................................................................ 66

ACCESSORIES .................................................................................................................................... 67

2

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Customer Service Center.

Each instrument is supplied with:

• HI 76313 4-ring probe with temperature sensor (4m cable)

• 100 mL Plastic Beaker

• 4 x 1.2 V AA Rechargeable Batteries (inside the instrument)

• HI 710042 Inductive Recharger with power adapter

• HI 7031M 1413 m

S/cm calibration standard (230 ml)

• HI 7035M 111.8 mS/cm calibration standard (230 ml)

• Instruction Manual

• Rugged carrying case

Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

3

The HI 98188 instrument is state-of-the-art, heavy-duty conductivity meter, designed to provide laboratory results and accuracy under harsh industrial conditions.

The USP standard compliance makes the instrument useful for pure water determination.

It is provided with a series of new diagnostic features which add an entirely new dimension to the measurement of conductivity, by allowing the user to dramatically improve the reliability of the measurement:

• 7 memorized standards (0.00 m

S/cm, 84.0 m

S/cm, 1.413 mS/cm,

5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm) for calibration.

• EC calibration up to five calibration points.

• Messages on the graphic LCD for an easy and accurate calibration.

• Diagnostic features to alert the user when the electrode needs cleaning.

• User selectable “Outside Calibration Range” warning.

• User-selectable “calibration time out” to remind when a new calibration is necessary.

Moreover, they offer an extended temperature range from –20 to 120 ºC

(–4 to 248 ºF), using temperature sensor inside EC electrode.

This instrument can also measure in Resistivity, TDS and Salinity ranges.

Three salinity modes are available:

% NaCl, Practical salinity and Natural sea water scale.

Other features include:

• Temperature source selection

• Temperature automatic compensation, linear or non linear user selectable

• Temperature reference selection 15 °C, 20 °C or 25 °C.

• Temperature coefficient set

• Log on demand up to 400 samples

• Auto Log feature up to 1000 records

• Auto Hold feature, to freeze first stable reading on the LCD

• Lock and user setup Fixed range selection

• GLP feature, to view last calibration data for EC, NaCl

• Probe change recognition

• PC interface

• Probe replatinization

4

1

2

3

4

5

6

7

HI 98188

EC/TDS/NaCl/Resistivity Meter

ESC

CAL

HELP

RANGE

SETUP RCL MODE

ON

OFF

GLP LIGHT

13

12

11

10

8 9

1) Liquid Crystal Display (LCD).

2) F1, F2, F3 functional keys.

3) / keys to manually increase/decrease the parameters or to scroll between the parameter list.

4) ESC to leave current mode, exit calibration, setup, help. etc.

5) CAL key, to enter/exit calibration mode.

6) RCL key, to enter/exit view logged data mode.

7) SETUP key, to enter/exit SETUP mode.

8) ON/OFF key, to turn the instrument ON and OFF.

9) GLP key, to display Good Laboratory Practice information.

10) LIGHT key to toggle display backlighting.

11) MODE key to toggle between EC and USP or between Salinity ranges.

12) RANGE key, to switch between EC, Resistivity, TDS, NaCl.

13) Help key to enter/exit contextual help.

5

14) Electrode DIN connector.

15) USB connector.

6

( s

A h o tc w s u a l

0 t v c a l o n u d o 4 0 0 e s u c u it p v m t o ti y

S / c m

1

1

0

0

0

0

0

0 m m S /

S / c m c m

)

E C

R e s si it v ti y

T D S

R a n g e

R e s o l u it o n

A c c u r a c y

R a n g e

R e s o l u it o n

A c c u r a c y

R a n g e

R e s o l u it o n

A c c u r a c y

( ± 0 .

1

0 .

1 0

1

0

.

0

0 0

1

0 0 .

0

1

1

0

.

0

0

0

.

.

0

0

0

0 0 t o t o t o t o t o t o

9

9

9

9

9

.

9 9 9

.

9

9

.

9

9

1

9 .

9

9 9 9

9

0 0

.

0

9 9

.

0

( a u t o r a n g i n g )

µ

µ

µ

S

S /

S /

/ c m c m c m

* m m

S

S / m S

/ c m c

/ c m m

0 1

0

0

.

0

.

0

0 1

1

0

0 .

0 1

.

0 0 1

0 .

0 1

0 .

1

µ

µ

µ m m

S

S

S / c m

S

/

/

/ c m c c m m S / c m m

S / c m

*

± 1 %

µ S / c m o f o r g r e a t e r e

1

)r a d i n g d i g ti w ci h e v e r

1 .

0

1 .

1

1 0

0

0

0

.

0

0 t o 9 9 .

9 t o t o t o

9 .

9 9

9

9

9

9 9 .

9

1

1 0

1

.

.

0

0

0

0

0 t o t o t o

1

9 9

9

0

.

9

0

9

9

.

0 o h m s o h

K

K o h o m s m s h m s

K o h m

M o

M o h m h m s s s *

( a u t o r a n g i n g )

0 .

1

0

0

.

0

1

1

.

1 o h m o

K

K o h m o h m h m s s

0

0

.

1

0

.

1

1

K o

M

M h o m o h s m s h m s *

( ± 1 0

± o h m

1 % s g r o r e a o f

1 t e r e d i

)r a g d i ti n g w h ci h e v e r

1

1

1

1

0

0

.

0

0 0

.

0

.

0

0

0

.

0

0

0

.

0

0

0 t o t o

9

9

9 .

9 9

9 9 .

9 t t t o o o

9

9

4

.

9

0

9

.

0

9

9 9

.

9

0 p p p m p m g g g

/ L

/

/

L

L

( a u t o r a n g i n g )

0 .

0 0

0 .

0 1

1 g / L , p p

0 m ,

.

0 1

0 g

.

1

/ L , p p m

0 .

1 g / L

( ± 0 .

0 5

± p p

1 m

% o r o f

1 g r e a t e r e d i a d g i ti n g w h

)r ci h e v e r

* The 0.001 m

S/cm EC range and 0.1 MOhms Resistivity range are not available with the 4m cable probe.

7

L

S

E

T e m

C C a l bi ar t oi n

C e ll

N a lC c o n s at n t s e ut p

C a l bi ar t oi n

I m s a at l ni p p n d i el e m a yt ar e dr n s ut et er

E C p

T e m p .

or b e s o u cr e d

R e

R s a o n ul g t e oi n

A cc u ar c y

R a n g e

R e s o ul t oi n

S

P

% e ar a

N a tc w a ci a l lC t s re a l cs a ni l ti

:e

:y 0

0

.

0

0

.

1

.

0

%

1

.

0

0 1

0

0

.

0 ot ot

4

8 ot

2

0 .

.

4

0

0 0

0

0 0 .

0

(

( P p p

%

)t

S U )

-

± 1 %

2 0 .

0 ot fo er a d

1 2 0 .

0 ni g

° C

0 .

1 ° C

A cc u ar c y ± 0 .

2

0 .

0 1 0

° C ot

( e

1 0 cx ul

.

0 0 0 d ni g p or b e re or

( 0

A u

.

0 0 ot m a

µ

1 2 t ci

S

.

8 c/ m

8 u p

, m

8 4

S ot

.

0

5 c/ m ,

µ p o

S

8 ni

0 c/ st m

.

0

, w m i

1

S

.

ht

4 1

7

3 c/ m , m m e m

1 1

S

1 .

c/

8 ro m zi

, e d m

5

S

.

0 s

0 c/ at n d a m ) m S

)r dr s c/ m , u s e c o n

M a d

( u c

.x

w i it

1 ht iv p o yt

H I ni t c a l

7 0 7 bi o n

3 yl ar t s n oi ni at n f

% d ro a o ar n g e dr ht

;) re ar n g e s

U S P c o m p l ai n t

R

T e m e .f

p .

C o m p .

T e m p .

T e m p .

to

C o e .f

L

T D S o g

F a o n tc ro d e m a n d ol g g ni g

H I 7 6 3 1 3

A u ot m a t ci

( 8 p ni f or m

M a s n e n u a l s

D I N , ro e ni s n t yr di e

4 m c a b ht e p el ) or b

N o T ,C L ni e a ,r n o n

1 5 , l ni

2 0 , e a r

2 5

I S O / D I S

° C

7 8 8 8

;e s dt ;

1 , 2 , 5 ,

0 .

0 0 ot 1 0 .

0 0 % / ° C

0 .

4 0

4 0 0 s a ot m

1 .

0 0 p el s

1

(

0 , m

5 ,

1 5 a x

,

1

1 0

0

3

,

0

0

0

,

3 0 s

6 a

0 s e

, m p

1

,c

2 el

0 ,

)s

1 8 0

U p ot 1 0 m ni

M e m ro zi e d p

M e m o d a s e u s er m e n t or if el s

B a t et yr yt p e / il ef

A u ot ar n g ,e A u ot E n d , L O C K a n d f xi e d ar n g e

R e c h a

P

A u ot

D

C i

W e

E n m iv gi e ni t or n h t s gr

O F F re oi n m af re n c e s e n s( t te at b el ) U s re s

1 0

4

0 x h

1 .

2 V n o d c b

1 3 a c lk

0 0 gi h m t

A h

2 5 h

A

H I 7 1 0 0 4 2 ni d u tc vi e er c h a gr cc w i u m ht u b al t a c lk ro /s gi h t re ( ni c ul d e d ) e el c at b l :e 5 , 1 0 , 3 0 , 6 0 m ni u et s ro d si a b el d

O p ot -

U si o

S B al et d

2 2 6 .

5 x 9 5 x 5 2 m m ( 8 .

9 x 3 .

7 5 x 2 ” )

5 2 5

I P 6 7 g

8

INITIAL PREPARATION

The instrument is supplied complete with rechargeable batteries. Proceed with a complete charging process before starting (see page 63).

To prepare the instrument for field measurements close the serial communication socket with proper stopper (to ensure waterproof protection).

Connect the EC probe to the DIN connector on the top of the instrument. Tighten the thread ring. Make sure the probe sleeve is properly inserted.

Turn the instrument ON by pressing ON/OFF key.

At start-up the display will show the Hanna logo for a few seconds followed by the percentage indication of the remaining battery charge and the “Loading

Log...” message, then enters the measurement mode.

The auto-off feature turns the instrument off after a set period (default 30 min) with no button pressed to save battery life. To set another period or to disable this feature, see SETUP menu on page 35. The instrument continues to monitor the inputs and memorize readings if the automatic logging is enabled and started. To stop autologging press StopLog key or simply power off the instrument by pressing ON/OFF key.

The auto-off backlight feature turns the backlight off after a set period (default

1 min) with no buttons pressed. To set another period or to disable this feature, see SETUP menu on page 35.

MEASUREMENTS

Immerse the probe into the solution to be tested. The sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.

If needed, press RANGE repeatedly until the desired range (EC, Resistivity,

TDS, Salt) is selected on the LCD.

Allow for the reading to stabilize. The main LCD line displays the measurement in the selected range, while the temperature is displayed on the lower LCD line.

9

EC range

The conductivity range is from 0 to 400 m

S/cm . The actual conductivity range

(the uncompensated conductivity) is up to 1000 mS/cm. The instrument will display conductivity readings up to 1000 mS.

Note: The symbol in front of the temperature reading means that the temperature can be entered by the user (Manual option selected in

Setup, or temperature out of range).

Resistivity range

The reciprocal of the conductivity of a material is the resistivity.

TDS range

A conductivity measured value can be corrected to a total disolved solids value using the TDS factor.

Salinity

The salinity is derived from the conductivity of a sample.

Salt % range

The percent of salinity in a sample is depending of the sample and on the salinity coefficient.

10

From practical reasons, the salinity of a solution is derived from the salinity of the seawater. Two methods of calculating the salinity from the conductivity are suported:

• Natural seawater scale

• Practical salinity scale

Natural sea water scale (UNESCO 1966)

According to the definition, salinity of a sample in ppt is calculated using the following formula: r

T

C where:

R

T

- coefficient;

T

(sample) - uncompensated conductivity at T °C;

C(35;15)= 42914 m

S/cm - the corresponding conductivity of KCl solution containing a mass of 32.4356 g KCl / 1 Kg solution;

- temperature compensation polynom.

Note: The formula can be applied for temperatures between 10 °C and 31 °C.

To access this range press Mode while in Salinity range until the sea water scale [SW] is displayed.

Practical salinity scale

This is a practical scale based on the precise measurement of the electrical conductivity of a range of solution of known salinity.

The relationship derived from the scale relate salinity, conductivity, temperature and pressure and use a solution with a salinity of 35 ‰ as datum point.

This is taken to have a conductivity of 42.914 % mS/cm of 15 °C and atmospheric pressure.

11

According to the definition, salinity of a sample in psu (practical salinity units) is calculated using the following formula:

C

R

T

T

- coefficient;

(sample) - uncompensated conductivity at T °C;

C(35,15)= 42.914

m

S/cm - the corresponding conductivity of KCl solution containing a mass of 32.4356 g KCl / 1 Kg solution; r

T a

0 a

1 a

2 a

3 a

4 a

5

= 0.008

= 0.008

- temperature compensation polynom

= -0.1692

= 25.3851

= 14.0941

= -7.0261

= 2.7081

b b b b b

0 b

1

2

3

4

5

=0.0005

= -0.0056

= -0.0066

= -0.0375

= 0.0636

= -0.0144

c

0 c

1

= 0.0005

X = 400R

T

Y = 100R

T f (T)= (T-15)/[1+0.0162(T-15)]

Note:

• The formula can be applied for salinity values between

0 and 42 psu.

• The formula can be applied for temperatures between

-2 °C and 35 °C.

To reach this range press Mode while in Salinity range until the practical salinity scale [PSU] is displayed.

12

Notes:

• If the meter displays the top of the range blinking, the reading is out of range.

• If the stability indicator “ ” blinks, the reading is unstable.

• Make sure the meter is calibrated before taking measurements.

• If measurements are taken successively in different samples, for accurate readings it is recommended to rinse the probe thoroughly with deionized water before immersing it into the samples.

• TDS reading is obtained by multiplying the EC reading by the TDS factor, which has a default value of 0.50. It is possible to change the TDS factor in the 0.40 to 1.00 range by entering setup mode.

The probes designed to work with this instrument have an internal ID identificatior. Every time the instrument detects probe changing, it warns the user to update the cell constant according with the new probe and then calibrate the EC range.

13

The EC, Resistivity and TDS scales are autoranging. The meter automatically sets the scale with the highest possible resolution.

By pressing Lock, the autoranging feature is disabled and the current range is frozen on the LCD.

The “Range: Locked” message is displayed. To restore the autoranging option press “AutoRng” functional key again.

The autoranging mode is also disabled by selecting a “fixed range” in the

Setup menu. While in fixed range mode the instrument will display the readings with the fixed resolution. Maximum 6 digits can be displayed. The top of the fixed range is displayed blinking when the reading exceeds this value.

To disable fixed range mode enter Setup and select autoranging mode.

Note: Autoranging is automatically restored if the range is changed, if the calibration mode is entered, or if the meter is turned off and back on again.

14

Two selectable temparature sources are available: reading directly from the sensor inside the probe or manual entry.

Three options of compensating temperature are available:

Linear Temperature Compensation: The conductivity of a solution with a specific electrolyte concentration changes with temperature. The relationship of the change in conductivity as a function of temperature is described by a solution’s temperature coefficient. This coefficient varies with each solution and is user selectable (see Setup mode).

Non Linear Temperature Compensation: for natural water measurements.

The conductivity of natural water shows strong non-linear temperature behaviour.

A polinomial relationship is used to improve the accuracy of the calculated results.

Note: Conductivity measurements of natural water can only be performed at temperature ranging from 0 to 36 °C. Otherwise the “Out T range” message will be displayed blinking.

No Temperature Compensation (No TC): The temperature shown on the

LCD is not taken into account.

To select the desired option enter Setup menu (see page 35).

If the temperature is out of the -20 °C - 120 °C range the instrument will do no temperature compensation.

15

CONDUCTIVITY VERSUS

TEMPERATURE CHART

The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion.

The conductivity invariably increases with increasing temperature.

It is affected by the type and number of ions in the solution and by the viscosity of the solution itself. Both parameters are temperature dependent. The dependency of conductivity on temperature is expressed as a relative change per

Celsius degree at a particular temperature, commonly as percent per ºC.

The following table lists the temperature dependence of the HANNA calibration buffers.

2 7

2 8

2 5

2 6

2 3

2 4

2 1

2 2

2 9

3 0

3 1

1 7

1 8

1 9

2 0

5

0

1 0

1 5

1 6

Cº Fº

7 1

7 3

7 5

7 7

2.

4.

8.

6.

2.

4.

6.

8.

6 8

6 9

2.

4.

6.

8.

6 4

6 6

6 0

6 2

5 0

5 9

3 2

4 1

8 2

8 4

7 8

8 0

8

8 6

7 8.

(

H

H 7I 0 3 0

µ S

8I 0 c/

3 0 m ) (

H

H 7I 0 3 1

µ S

8I 0 c/

3 m

1

)

1

1

1

1

2

2

3

3

6

8

1

3

4

8

3

7 0

0

0

0

1

1

1

1

1

1

2

2

6

9

1

3

7

1

5

9 0

0

0

0

1

1

1

1

1

3

3

4

4

6

8

1

3

2

7

2

7 0

0

0

0

7 1 5 0

8 2 2 0

9 3 3 0

0 4 8 0 1

1

1

1

0

0

1

1

7

9

1

4

2

5

9

3 0

0

0

0

1

4 1

4 1

4 1

4

1

4

6

9 4

3

0

7

1

3 1

3 1

3 1

3

0

3

5

8 6

5

2

9

1

1

5

1

5 1

5

7

2

4 8

5

1

1

1 1

2 1

2 1

2

9

2

5

7 8

9

5

1

7

8

0

9

7

2

6

6

1

1 1

1

4

7 3

7

0

(

H

H 7I 0 3 3

µ S

8I 0 c/

3 m

3

)

8 6

8 7

8 2

8 4

7 9

8 1

7 6

7 8

9 2

9 4

8 9

9 0

7 3

7 4

7 0

7 1

6 7

6 8

6 4

6 5

(

H

H 7I 0 3 4

µ S

8I 0 c/

3 m

4

) (

H

H 7I 0 3 5

µ S

8I 0 c/

3 m

5

)

7

7

0

8 3

0 0

0

8

8 1

3

3

0

0

0 0

0

0

0

7

7

7

7

2

4

5

6

4

0

2

5

0

0

0

0 0

0

0

0

8

8

9

8

8

4

6

8

0

9

3

2

0

0

0

0

0 0

0

0

0

4

5

5

8

3

9

5

3

5

6

4

0

0

0

0 0

0

0

0

6

6

6

6

7

8

9

1

2

5

8

3

0

0

0

0 0

0

0

0

1

1

0

1

1 1

1

0 9

3

1 8

5 7

8

8

0

0

0 0

0 0

0

0

1

9

0

1

0 1

0

0 2

5

4 0

7 9

1

9

0

0

0 0

0 0

0

0

1

1

2

1

1

1 8 0 0

1

1

1

1

2

1

3

7

9 7

9

7

0

0

0 0

0

0

6

8

7

5

4

3

2

4

2

5

1

0

0

0

0 0

0

0

0

9

9

9

4

6

8

4

3

2

0

0

0 0

0

0

0 2 0 0

(

H

H 7I 0 3 9

µ S

8I 0 c/

3 9 m )

5 2 8 6

5 3 8 3

5 4 7 9

5 5 7 5

4 9 0 2

5 0 0 0

5 0 9 6

5 1 9 0

4 5 2 3

4 6 1 7

4 7 1 1

4 8 0 5

4 1 5 5

4 2 4 5

4 3 3 7

4 4 2 9

2 7 6 0

3 1 8 0

3 6 1 5

4 0 6 3

16

Pharmaceutical laboratories working in the US market are obliged to respect the regulations set down by the US Pharmacopoeia. The 5 th

supplement of

USP24-NF19 lays down the rules for checking the quality of pure or fully deionized water used for the production of injection products.

The conductivity of water provides information on its chemical composition. It is therefore logical that makes this the main parameter to measure.

The conductivity of a water is a measure of the ion mobility through this water.

The conductivity partly depends on the pH, the temperature and the amount of atmospheric carbon dioxide, witch has been dissolved in water to form ions

(intrinsec conductivity). The conductivity also depends on the chloride, sodium and ammonium ions initially present in water (extraneous conductivity).

The conductivity (intrinsec and extraneous) of the water is measured at first stage and compared to values listed in a table to evaluate if the studied water is suitable or not for use in pharmaceutical applications. If the sample fails stage 1, additional tests have to be performed (Stage 2 and 3) in order to determine if the excessive conductivity value is due to intrinsec factors or extraneous ions.

USP Requirements

Automatic temperature correction must not be used.

Instrument specifications

Minimum resolution of 0.1 m

S/cm on the lowest range. Excluding the cell accuracy, the instrument accuracy must be ±0.1 m

S/cm.

Meter calibration

It is accomplished by replacing the conductivity cell with precision resolution traceable to primary standards (accurate to ±0.1 % of the stated value) or an equivalently accurate resistance device.

Cell calibration

Meter conductivity must be measured accurately using calibrated instrumentation. The conductivity cell constant must be known in ± 2 %.

Before starting water analyze calibrate the EC lowest range or set the probe cell constant (enter the value written in the calibration certificate delivered with the probe).

Stage 1

Determine the temperature and conductivity of the water.

• Rinse the probe carefully with deionized water. Check that the 4 rings and the temperature sensor are immersed in the sample and that no air bubbles

17

are trapped. Connect the probe to the meter enter USP mode and press the

Stage 1 key. The instrument will perform a temperature and conductivity measurement (using a non-temperature corrected conductivity reading).

• Using the Stage 1 temperature and conductivity requirement table the corresponding conductivity limit at that temperature is determined.

• If the measured conductivity is no greater than the table value the water meets the requirements of the test for conductivity. If the conductivity is higher than the table value, proceed with stage 2.

Stage 1 table

Temperature and conductivity requirements *

(for the non-temperature compensated conductivity measurements only)

(*) Values from USP - NF Fifth Supplement

1 8.

1 9.

2 1.

1 4.

1 5.

1 7.

C o n d u c

µ S it c/ m iv yt

0 6.

0 8.

0 9.

1 0.

1 1.

1 3.

2 7.

2 7.

2 7.

2 2.

2 4.

2 5.

2 7.

2 9.

3 1.

4 0

4 5

5 0

2 5

3 0

3 5

T e m p e r

° C ta u r e

0

5

0

1 0

1 5

2 0

7 5

8 0

8 5

6 5

7 0

5 5

6 0

9 0

1 0 0

Physical Tests / Water Conductivity (645) 3465-3467

18

Stage 2

Determine the influence of CO

2

Note: Stability criteria <0.5 %/min corresponds in fact to a change in conductivity <0.02 m

S/cm per minute (equivalent to 0.1 mS/cm per

5 minutes), as required in the Stage 2 USP document.

• Transfer a sufficient amount of water (100 mL or more) to a thermostatic vessel, and stir the test specimen. Adjust the temperature and maintain it at 25 ±1 ºC. We recommend using a thermostatic bath.

• Rinse the cell carefully with deionized water. Place the conductivity probe on the Sample and dip it in the thermostatic vessel containing the sample.

Check that the 4 rings are immersed in the solution.

• Press Step 2.

• When the conductivity value is stable, if the conductivity is not greater than

2.1 m

S/cm the water meets the requirements of the test, for conductivity.

If the conductivity is greater than 2.1 m

S/cm proceed with stage 3.

Stage 3

Determine the combined effect of the CO

2

and pH.

Use a Hanna Instruments pH meter.

Take care that the instrument be calibrated in at least two points using pH 4.01 and pH 7.01 Hanna buffers.

Perform the following test within approximately 5 minutes of the conductivity determination while maintaining the sample temperature at 25±1 ºC.

• Add the Saturated KCl Solution to the same (0.3 mL per 100 mL of the test specimen), and determine the pH to the nearest 0.1 pH unit.

• Press Stage 3 key, then enter the corresponding pH reading.

• Referring to the Stage 3 pH and conductivity requirement table the instrument determines the conductivity limit at the measured pH value. If the measured conductivity is no greater than the conductivity requirements for the pH determined the water meets the requirements of the test for conductivity. If the measured conductivity is greater than this value or the pH is outside of the range of 5.0 to 7.0, the water does not meet the requirements of the test for conductivity.

19

p H

6 7.

6 8.

6 9.

7 0.

6 4.

6 5.

6 6.

6 0.

6 1.

6 2.

6 3.

5 7.

5 8.

5 9.

5 4.

5 5.

5 6.

5 0.

5 1.

5 2.

5 3.

Stage 3 pH and conductivity requirements

(For atmosphere and temperature equilibrated samples only)

2 3.

2 2.

2 1.

2 4.

2 4.

2 5.

2 4.

2 6.

3 1.

3 8.

4 6.

2 5.

2 4.

2 4.

3 0.

2 8.

2 6.

C o n d

µ S u c c/ it m iv yt

4 7.

4 1.

3 6.

3 3.

20

Press Mode key while in EC range to enter USP mode.

The instrument will display the USP main screen.

Press Stage 1 to start with first stage evaluation.

Press Stage 2 to start with second stage evaluation.

If Stage 1 is pressed a tutorial screen is displayed.

Use the Arrow keys to scroll the tutorial message.

Press Continue to skip the tutorial message and enter EC measuring mode.

If the temperature source is manual entry, press USP Fact. to select USP factor to be changed (the is displayed before the 100% value). To increase the accuracy of the analyze decrease the USP factor value.

Wait until reading is stable.

If the sample reading met the USP stage 1 criteria, the “USP Met” message will be displayed.

Press Report to view report.

21

Press Log to memorise USP Stage 1 report.

The record number and the amount of free log space in % is displayed for few seconds.

Note: If the log space is full enter view logged data mode by pressing RCL key and free log space by deleting previously memorized records.

If the sample reading didn’t meet the USP stage 1 criteria, the

“USP Not Met” message is displayed.

Press ESC to return to the USP main screen.

Press Report to view USP report.

Press Log to memorise the Stage 1 report.

Press Stage 2 to enter USP stage 2 mode.

The USP Stage 2 tutorial screen is displayed.

Use the Arrow keys to scroll the tutorial message.

Press Continue to skip the tutorial message and enter USP Stage 2 measuring mode.

22

Wait until the reading is stable (about 5 minutes).

Note: If the input record has an instability higher than 1 m

S, stability period will be reset. The completion time bar will remain empty.

The “USP Met” message will be displayed if the USP stage 2 criteria is reached.

Press Report to view the USP report.

Press key to scroll the report screens.

Press Log to memorise report.

Notes:• The record number will be the same as at the Stage 1 report

(informations of the same analyse).

• If the log space is full enter view logged data mode by pressing

RCL key and free log space by deleting previously memorized records.

The “USP Not Met” message is displayed if the USP stage 2 criteria is not met.

Press Report to view report.

Press key to scroll the report screens. Press Log to memorise report.

Press Stage 3 to enter in stage 3 water analyze.

The USP Stage 3 tutorial is displayed.

23

Press Continue to enter USP Stage 3 analyze.

The instrument will display the sample pH setting mode.

Use a calibrated pH meter to read the pH value of the sample.

Use ARROW keys to set the value to that red from the pH meter.

Press Accept to confirm the pH setting.

The USP Stage 3 report will be displayed.

The report will include all stages information.

Press key to scroll the report pages .

Press Log to memorise the report.

Press ESC to return to USP main screen.

Notes:• The Stage 3 report will include the information about Stage 1 and

Stage 2.

• If the log space is full enter view logged data mode by pressing

RCL key and free log space by deleting previously memorized records.

24

To enter User Calibration screen press CAL key while in EC or Salinity range.

From EC range

Press the corresponding functional key to enter:

• EC user calibration.

• Probe replatinization.

• Temperature user calibration.

From Salinity % range

Press the corresponding functional key to enter:

• Salinity % user calibration.

• Temperature user calibration.

25

It is recommended to calibrate the instrument frequently, especially if high accuracy is required.

The EC range should be recalibrated:

• Whenever the EC electrode is replaced.

• At least once a week.

• Before USP measurement.

• After testing aggressive chemicals.

• When calibration alarm time out is expired - “CAL DUE” blinks (if feature is enabled in SETUP).

• If “Outside Cal Range” message blinks during EC measurement (the measurement range is not covered by current calibration, if feature is enabled in SETUP).

Note: TDS and Resistivity readings are automatically derived from the EC reading and no specific calibration is needed.

PROCEDURE

HI 98188 instrument offers a choice of 7 memorized standards (0.00 m

S/cm,

84.0 m

S/cm, 1.413 mS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and

111.8 mS/cm).

For accurate EC measurements, it is recommended to perform a calibration in maximum allowed points. However, at least a two-point calibration is suggested

(offset calibration in 0.00 m

S and one in any other calibration standard).

The instrument will automatically recognize the standards and skip the standard used during calibration.

• If a five-point calibration is performed, one of the points must be 0.0 m

S (offset).

• Pour small quantities of selected standard solutions into clean beakers. For accurate calibration use two beakers for each standard solution, the first one for rinsing the electrode and the second one for calibration.

• Remove the protective cap and rinse the electrode with some of the standard solution to be used for the first calibration point.

• It is better to perform the first point calibration as offset calibration.

• Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.

• For offset (zero) calibration, just leave the dry probe in the air.

26

• From EC range press CAL to enter calibration screen.

• Leave the probe in the air and press EC. The instrument will display the measured EC on the LCD, first expected standard and the temperature reading.

• If necessary, press the ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer, CFM functional key is displayed.

• Press CFM to confirm first point.

• The calibrated value and the second expected standard value is then displayed on the LCD.

• After the first calibration point is confirmed, immerse the EC electrode into the second standard solution and stir gently. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve. The instrument will automatically detect the standard after few seconds the reading is stable.

• If necessary, press the ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected standard, the CFM

27

functional key is displayed.

• Press CFM to confirm calibration.

• The calibrated value and the third expected standard value will be displayed.

• After the second calibration point is confirmed, immerse the EC electrode into the third standard solution and stir gently. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.

The instrument will automatically detect the standard value.

• If necessary, press the ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected standard, the CFM functional key is displayed.

28

• Press CFM to confirm calibration.

• The calibrated value and the fourth expected standard value will be displayed.

• After the third calibration point is confirmed, immerse the EC electrode into the fourth standard solution and stir gently. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.

• The instrument will automatically detect the standard value.

• If necessary, press the ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected standard, the CFM functional key is displayed.

• Press CFM to confirm calibration.

• The calibrated value and the fourth expected standard will be displayed.

29

• After the fourth calibration point is confirmed, immerse the EC electrode into the fifth standard solution and stir gently.

• If necessary, press the ARROW keys to select a different standard value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected standard, the CFM functional key is displayed.

• Press CFM to confirm calibration.

• The instrument stores the calibration values and returns to normal measurement mode.

FOUR, THREE or TWO-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.

• Press CAL or ESC after the appropriate accepted calibration point. The instruments will return to measurement mode and will memorize the calibration data.

ERROR SCREENS

Wrong standard

The calibration cannot be confirmed.

The EC reading is not close to the selected standard. Select another standard using the ARROW keys.

30

CLEAR CALIBRATION

Press Clear functional key when displayed to clear old calibrations.

All old calibrations, are cleared and the instrument continues calibration. The points confirmed in current calibration are kept.

Note: If Clear calibration is invoked during the first calibration point, the instrument returns to measurement mode.

REPLACE CALIBRATION STANDARD

Each time a standard is confirmed the new calibration parameters replace the old calibration parameters of the corresponding standard.

If the current standard has no correspondance in the existing stored calibration structure and this is not full, the current standard is added to the existing stored calibration (up to 5 standards).

If the existing stored calibration is full (five calibration points) after confirming the calibration point, the instrument will ask wich buffer will be replaced by current standard.

Press ARROW keys to select another standard to be replaced.

Press CFM to confirm the standard that will be replaced.

Press CAL or ESC to leave replace mode. In this case, the standard will not be memorised.

Note: The replaced standard is not removed from calibration list and it can be selected for the next calibration points.

NaCl CALIBRATION

NaCl calibration is a one-point procedure in 100.0% NaCl solution. Use the HI 7037L calibration solution (sea water solution) as a 100% NaCl standard solution.

• To enter NaCl calibration select the Salinity % range and press CAL.

31

• The instrument enters the Salinity calibration screen.

• Press Salt. The measured % NaCl, the temperature and the 100% NaCl standard are displayed.

• Rinse the probe with some of the calibration solution or deionized water.

Immerse the probe into HI 7037L solution. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable, the CFM functional key is displayed.

Press CFM to confirm calibration.

• The instrument returns to measurement mode.

Notes:

• If the uncalibrated reading is too far from the expected value, calibration is not recognized. The “Wrong” message will be displayed.

• The meter uses 1.90 %/°C temperature compensation factor during calibration. If the “Temperature Coef” in the Setup menu has been set to a different value, when exiting calibration mode, the displayed value might be different from the nominal standard value.

32

GLP is a set of functions that allows storage and retrieval of data regarding the maintenance and status of the electrode.

All data regarding EC and NaCl calibration is stored for the user to review when necessary.

EXPIRED CALIBRATION

The instrument is provided with a real time clock (RTC), in order to monitor the time elapsed since the last EC/NaCl calibration.

The real time clock is reset every time the instrument is calibrated and the

“expired calibration” status is triggered when the instrument detects a calibration time out. The “CAL” “DUE” tags will start blinking to warn the user that the instrument should be recalibrated.

The calibration time out can be set (see SETUP for details, page 35) from 1 to

7 days or can be disabled.

For example, if a 4 days time out has been selected, the instrument will issue the alarm exactly 4 days after the last calibration.

However, if at any moment the expiration value is changed (e.g. to 5 days), then the alarm will be immediately recalculated and appear 5 days after the last calibration.

Notes: • When the instrument is not calibrated or calibration is cleared

(default values loaded) there is no “expired calibration”, and the display always shows the “CAL” “DUE” tags blinking.

• When an abnormal condition in the RTC is detected, the instrument forces the “expired calibration” status.

LAST EC CALIBRATION DATA

The last pH calibration data is stored automatically after a successful calibration.

To view the EC calibration data, press GLP when the instrument is in the EC measurement mode.

33

The instrument will display a lot of data including calibration standards, offset, time and date, etc.

Use the Arrow keys to select the offset or calibration standards, in order to view new information.

To see more information press More.

• More information regarding offset.

• More information regarding standards.

Notes: • Standards displayed in video inverse mode are from previous

calibrations.

• “No user calibration” message is displayed if all calibrations are cleared or the instrument was not calibrated in the EC range.

LAST NaCl % CALIBRATION DATA

Last NaCl calibration data is stored automatically after a successful calibration.

To view the NaCl calibration data, press GLP key while in NaCl measurement mode.

The instrument will display the NaCl information: calibration date, time and offset.

34

Setup mode allows viewing and modifying the measurement parameters.

These are general SETUP parameters for all the ranges and range specific parameters.

The following table lists the general SETUP parameters, their valid range and the factory default settings.

New Description

Select profile Add /View or select a profile

Valid value Default

Log interval Time for automatic 5, 10, 30 s Disabled

logging 1, 2, 5, 10, 15, 30, 60, 120, 180 min (Log on

demand)

Backlight Backlight level

Contrast Contrast level

0 to 8 4

0 to 20 10

Auto Light Off Time until backlight 1, 5, 10, 30 min 1

is ON

Auto Power Off Time after the instrument Disabled 30

is powered OFF 5, 10, 30, 60 min

Date/Time

Time Format

01.01.2006 to 12.31.2099

00:00 to 23:59

AM/PM or 24 hours

01.01.2006

24 hours

Date Format DD/MM/YYYY

MM/DD/YYYY

YYYY/MM/DD

YYYY-MM-DD

Mon DD, YYYY

DD-Mon-YYYY

YYYY-Mon-DD

Language Message display language Up to 4 languages

YYYY/MM/DD

English

Beep ON Beeper Status Enabled or Disabled

Instrument ID Instrument identification 0000 to 9999

Disabled

0000

Baud Rate Serial Communication 600, 1200, 1800, 9600 9600

Meter Displays general information informations

35

The following table lists the specific range parameters:

Item

Calibr.

Description

Number of days after

Timeout Calibration warning

(EC, NaCl) is displayed

Valid value Default

Disable, 1 to 7 days Disable

Out cal Display warning if the Enable/Disable Disable range check reading is too far from the

(EC range calibration points only)

Temperature Temperature source entry mode

Probe/Manual Probe

Temperature No TC, Linear No TC compensation Non Linear mode

Automatic Range select Fix one Automatic, Fix one

specific range resolution of

EC or resistivity range

Cell constant Manual set of the cell constant

0.010 to 10.000

1.000

Temperature Set the coeficient coeficient for linear temperature

compensation

0.00 to 10.00 %/°C 1.00 %/°C

Temperature Reference temperature 15 °C, 20 °C, 25 °C

Ref

Temperature unit

25 °C

°C or °F °C

TDS factor 0.40 to 1.00 0.50

36

GENERAL PARAMETER SCREENS

Select Profile

Focus on the Select Profile item.

Press Select.

The list with memorised profiles is displayed.

Press Add to add a new profile to the list (max 10).

Use the Arrow keys to focus on the desired profile.

Press Select to select the profile and exit to Setup.

Press View to view profile information.

The profile information includes the date and the time when the profile was added, information about calibration on EC and NaCl range, the setup cell constant, information about temperature setting reference temperature, temperature compensation mode, temperature coefficient, temperature source and information about fixed ranges.

If calibration exists, the GLP functional key is displayed. Press GLP EC to view the corresponding GLP EC information.

Press GLP NaCl to view the corresponding GLP NaCl information.

Note: If the EC or NaCl calibration was not performed while the current profile is set, or calibration was cleared, the corresponding GLP key is not displayed.

37

Press Delete to delete selected profile. The Delete key is displayed only if more than one profile is in the list.

Press Accept to confirm the deletion or Cancel to cancel and return to the previous screen.

Press ESC to return to profile list screen.

Logging interval

Focus on the

Logging interval item.

Press Modify.

Use the Arrow keys to select the logging interval. If “Disabled” option is selected the Autolog feature is disabled and Log on demand is enabled.

Press Accept to confirm the value.

Press ESC to exit without saving.

Backlight

Focus on the Backlight item.

38

Contrast

Focus on the

Contrast item.

Press Modify.

Use / keys to change contrast then press Accept to confirm.

Press ESC to leave without changing.

Auto Light Off

Focus on the Auto Light Off item.

Press 5, 10 or 30 to change settings.

Press one of the functional key to change the option.

Auto Power Off

Focus on the Auto Power Off item.

39

Press Modify.

Press ARROW keys to select internal interval then press Accept.

Press ESC to leave without changing.

Date/Time

Focus on the Date/Time item.

Press Modify.

Use / keys to select item. Use ARROW keys to change focused values.

Press Accept to confirm new setting, or ESC to leave without changing.

Time Format

Focus on the Time Format item.

Press functional key to change the option.

Date Format

Focus on the Date Format item.

40

Press Modify.

Use ARROW keys to select date format then press Accept.

Press ESC to leave without changing.

Language

Focus on the Language item.

Use the desired functional key to change the option. Wait until new language is loaded.

If any language can be loaded, the instrument will work in safe mode. In this mode all messages are displayed in English and Help is not available.

Beep On

Focus on Beep On item.

Press the displayed functional key to enable/disable key.

When enabled, beep sounds as a short beep every time a key is pressed or when the calibration icon can be confirmed.

A long beep alert that the pressed key is not active or a wrong condition is detected while in calibration.

Instrument ID

Focus on the

Instrument ID item.

41

Press Modify.

Use ARROW keys to change the instrument ID.

Press Accept to confirm or ESC to exit without saving.

Baud Rate

Focus on the

Baud Rate item.

Press Modify.

Use ARROW keys to select the desired communication baud. Press Accept to confirm or ESC to exit.

Meter information

Focus on the Meter Information item.

Press Select.

42

The meter informations are displayed:

-firmware version

-language version

-mV and temperature factory calibration time/date

-battery capacity

RANGE SPECIFIC PARAMETERS

Calibration Timeout

Focus on the

Calibration Timeout item.

Press Modify.

Use ARROW keys to set desired value.

Press Accept to confirm or ESC to return without saving.

Note: If enabled “CAL DUE” warning will be displayed, the set number of days after calibration is over passed.

Out of calibration range check

Focus on

Out of cal range check item.

43

Press the corresponding functional key in order to enable/disable this feature.

If enabled, a warning message is displayed when the EC reading is too far from the EC calibration points.

Temperature source

Focus on the Temperature source item.

Press the displayed functional key in order to change the option.

Select Probe in order to take the temperature automatic with the temperature sensor inside the electrode.

Select Manual in order to set the temperature using the ARROW keys.

Temperature compensation

Focus on Temperature compensation item.

Press one of the selected functional keys to change the option.

Select No TC in order to display actual conductivity. (no temperature compensation)

Select Linear in order to automatically compensate conductivity using the set temperature coefficient.

Select Non Linear temperature compensation for natural water measurements, using the natural water compensation equation.

Range Select

Focus on the Range Select item.

44

Press Modify to select ranging mode.

Use the Arrow keys to change selection.

Press Accept to confirm or ESC to exit without saving. If Automatic is selected the instrument changes the range automatically according with the input.

If one of the range is selected all the readings are displayed on the corresponding range.

The readings will be displayed with maximum 6 digits. If the reading exceeds the maximum number of digits for the fixed range, the maximum value is displayed blinking.

Note: The Range select parameter can be set only in EC and resistivity range.

Cell constant

Focus on the Cell constant item.

Press Modify to change the cell constant value.

Use Arrow Keys to change the cell constant value.

Press Accept to confirm or ESC to exit without changing.

Out of Cal-Range Warning

Focus on the Out of Cal-Range Warning item.

45

Press the displayed functional key in order to change option.

If enabled, the “Out Cal Range” message will be displayed if the EC reading is not within the calibration range.

Temperature Coefficient

Focus on the Temperature Coefficient item.

Press Modify in order to set the temperature coefficient.

Use the Arrow keys to change the value.

Press Accept to confirm or ESC to exit without changing.

Temperature Reference

Focus on the Temperature Reference item.

Press the corresponding functional key to select the desired reference temperature.

Temperature unit

Focus on the Temperature unit item.

Press the displayed functional key in order to change the temperature unit.

46

This feature allows the user to log up to 400 readings. All logged data can be transferred to a PC through the USB port.

LOGGING THE CURRENT DATA

To store the current reading into memory, press LOG while in measurement mode.

The instrument will display for few seconds the record number and the amount of the free log space.

If the LOG space is full, the “Log space is full” message will be displayed for few seconds when Log key is invoked.

Enter View Logged Data mode and delete records in order to free log space.

VIEW LOG ON DEMAND DATA

Press RCL to enter View Logged Data mode.

Press Man Log to enter the log on demand range selection.

47

Use arrow keys to focus on the desired range then press View.

The list of records corresponding to the selected range is displayed.

If no data were logged on the current range, the instrument will display

“No Records” message.

Use ARROW keys to scroll between the records from the list.

Press Delete All to enter Delete All screen.

Press Delete to enter Delete records screen.

Press More to view more information of the focused record.

If More is pressed.

Use Pg Up or PgDown keys to toggle the complete information screens.

Use ARROW keys to scroll between memorised records.

If Delete is pressed.

Use ARROW key to focus the record to be deleted and then press CFM.

Press ESC to exit.

If Delete All is pressed the instrument asks for confirmation.

Press CFM to confirm or ESC to exit without deleting.

For USP range, the instrument will display the sample ID, analyze time and date:

Use ARROW keys to select the desired record.

Press Delete to enter delete one record mode.

48

Press Delete All to enter delete all mode.

Press More to view the complete record information.

Press PgDown or PgUp to scroll the record screens.

49

This feature allows the user to log up to 1000 readings.

All logged data can be transferred to PC through USP port.

The memory space is organized in lots of records. A lot can contain from 1 to

1000 records. The maximum available lots number is 100.

START AUTOLOG

Set the desired logging interval in the Setup menu.

If 5, 10, 30 s or 1 min interval is selected, the AutoPower Off feature is disabled (each sample memorized restarts the autopower off counter). For other intervals, the instrument will enter a sleeping mode. During this mode the instrument continues to monitor the inputs and memorize the reading at the set interval. To exit sleeping mode, simply press any key (exception ON/

OFF).

Note: While in sleeping mode the instrument cannot be powered off by pressing ON/OFF key. Leave sleeping mode and then press ON/OFF key to power off the instrument.

To start autolog from the measurement screen press StartLog key. The lot number and the amount of memory space is displayed for few seconds.

If the autolog space is full or the lots number exceeds 100 the

“Auto log space full” message is displayed.

VIEW AUTOLOG DATA

Press RCL key to enter View Logged Data mode.

50

Press AutoLog to enter the automatic log range selection.

Use arrow keys to focus on the desired range, then press View.

The list of lots corresponding to the selected range is displayed.

If no data were logged on the current range, the instruments will display

“No Records!”.

Use ARROW keys to scroll the lot list. Press Delete or Delete All to delete the lot or all lots. The confirmation is asked as at individual record delete.

Press More to view more information.

The list of records for the specific lot is displayed.

Press More to view the complete record information.

Press PgDown or PgUp to scroll the information screens.

51

To freeze the first stable reading on the LCD press AutoEnd while the instrument is in measurement mode.

The ”Wait” symbol will blink until the reading is stable.

When the reading is stable, “Hold” icon will be displayed.

Press Continue in order to enter continuous reading mode.

52

TEMPERATURE CALIBRATION

(for technical personnel only)

All the instruments are factory calibrated for temperature.

Hanna’s temperature probes are interchangeable and no temperature calibration is needed when they are replaced.

If the temperature measurements are inaccurate, temperature recalibration should be performed.

For an accurate recalibration, contact your dealer or the nearest Hanna

Customer Service Center, or follow the instructions below.

Press “T” functional key to enter the temperature calibration mode while in EC or Salt calibration screens .

It is better to perform a 2 points calibration.

The calibration can be performed in any two points that have at least 25 ºC distance between. It is recommended that the first point be near 0 ºC and the second point near50 ºC.

• Prepare a vessel containing ice and water and another one containing hot water (at approximately 50 ºC or 122 ºF). Place insulation material around the vessels to minimize temperature changes.

• Use a calibrated thermometer with a resolution of 0.1 ºC as a reference thermometer. Connect the HI 76313 EC probe with built in temperature sensor to the appropriate socket.

• Immerse the HI 76313 probe into the vessel with ice and water as close as possible to the reference thermometer. Allow a few seconds for the probe to stabilize.

• Use the ARROW keys to set the calibration point value to that of ice and water mixture, measured by the reference thermometer. Set the calibration point value to that measured in the reference thermometer.

53

• When the reading is stable and close to the selected calibration point, the

CFM functional key is displayed.

• Press CFM to confirm.

• The second expected calibrated point is displayed.

• Immerse the probe into the second vessel as close as possible to the reference thermometer. Allow a few seconds for the probe to stabilize.

• Use the ARROW keys to set the calibration point value to that measured by the reference thermometer.

• When the reading is stable and close to the selected calibration point,

CFM functional key is displayed.

• Press CFM to confirm. The instrument returns to measurement mode.

54

Note: •If the reading is not close to the selected calibration point or the difference between first selected point and second selected point is less than 25 ºC, “Wrong” message will blink.

• If the WRONG source is the difference between calibration points increase the temperature of the vessel with hot water in order to be acceptable.

• If the WRONG source is the temperature reading change the probe and restart calibration.

If calibration cannot be performed contact Hanna Service.

•For one point calibration press ESC after first point was confirmed.

55

Data transmission from the instrument to the PC can be done with the HI 92000

Windows

®

compatible software (optional). HI 92000 also offers graphing and on-line help feature.

Data can be exported to the most popular spreadsheet programs for further analysis.

To connect your instrument to a PC, use an USB cable connector. Make sure that your instrument is switched off and plug one connector to the instrument

USB socket and the other to the serial or USB port of your PC.

Note: If you are not using Hanna Instruments HI 92000 software, please see the following instructions.

SENDING COMMANDS FROM PC

It is also possible to remotely control the instrument with any terminal program.

Use an USB cable in according with the model to connect the instrument to a

PC, start the terminal program and set the communication options as follows: 8,

N, 1, no flow control.

COMMAND TYPES

To send a command to the instrument follow the next scheme:

<command prefix> <command

> <CR> where: <command prefix> is a selectable 16 ASCII character

<command> is the command code.

Note: Either small or capital letters can be used.

SIMPLE COMMANDS

KF1

KF2

UPC

DWC

RCL

SET

CLR

KF3

RNG

MOD

CAL

Is equivalent to pressing functional key 1

Is equivalent to pressing functional key 2

Is equivalent to pressing functional key 3

Is equivalent to pressing RANGE

Is equivalent to pressing MODE

Is equivalent to pressing CAL

Is equivalent to pressing the UP arrow key

Is equivalent to pressing the DOWN arrow key

Is equivalent to pressing RCL

Is equivalent to pressing SETUP

Is equivalent to pressing CLR

OFF Is equivalent to pressing OFF

CHR xx Change the instrument range according with the parameter

56

value (xx):

• xx=10 EC range

• xx=11 Resistivity range

• xx=12 TDS range

• xx=13 USP range

• xx=14 NaCl % range

• xx=15 Salinity, Sea water range

• xx=16 Salinity, PSUrange

The instrument will answer for these commands with:

<STX> <answer > <ETX> where: <STX> is 02 ASCII code character (start of text)

<ETX> is 03 ASCII code character (end of text)

<answer>:

<ACK> is 06 ASCII code character (recognized command)

<NAK> is 21 ASCII code character (unrecognized command)

<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER

The instrument will answer for these commands with:

<STX> <answer> <checksum> <ETX> where the checksum is the bytes sum of the answer string sent as 2 ASCII characters.

All the answer messages are with ASCII characters.

RAS Causes the instrument to send a complete set of readings in according with the current range:

• EC and temperature readings or EC range.

• Resistivity, EC and temperature readings on Resistivity

• TDS, EC and temperature readings on TDS range

• Salinity, EC, and temperature or Salinity range.

The answer string contains:

• Meter mode (2 chars):

• xx=10 EC range

• xx=11 Resistivity range

• xx=12 TDS range

• xx=13 USP range

• xx=14 NaCl % range

• xx=15 Salinity, Sea water range

• xx=16 Salinity, PSU range

57

• Meter status (2 chars of status byte): represents a 8 bit hexadecimal encoding.

• 0x10 - temperature probe is connected

• 0x20 - autolog in progress

• 0x01 - new GLP data available

• 0x02 - new SETUP parameter

• 0x04 - out of calibration range

• 0x08 - the meter is in autoend point mode

• 0x30 - the instrument is in fixed range

• Reading status (2 chars): R - in range, O - over range,

U - under range. First character corresponds to the primary reading. Second character corresponds secondary reading.

• Primary reading (corresponding to the selected range) -

10 ASCII chars, including sign, decimal point and unit.

• Secondary reading (only when primary reading is not EC)

- 10 ASCII chars, including sign, decimal point and unit

(if primary reading is not conductivity).

MDR

• Temperature reading - 8 ASCII chars, with sign, decimal point, always in ºC.

Requests the instrument model name and firmware code (16 ASCII chars).

GLPxx Requests the calibration data record for “xx” profile number.

The answer string contains:

• GLP status (1 char): represents a 4 bit hexadecimal encoding.

• 0x01 - EC calibration available

• 0x02 - NaCl calibration available

• EC calibration data (if available), which contains:

• calibration time (12 char)

• the number of calibration point (1 char)

For each calibration point:

• the standard value (10 chars)

• offset factor or cell constant (with sign and decimal point)

• time when was calibrated (12 chars)

• salinity calibration

• calibration time (12 charrs)

• salinity coefficient (10 chars)

PRFxx Request profile “xx” information

The answer string contains:

58

PAR

10 - EC range

11 - Resistivity range

12 - TDS range

13 - USP range

14 - NaCl % range

15 - Salinity Sea Water range

16 - Salinity Sea Water, PSU range

• creation time (12 chars)

• temperature source (1 char): 0 - probe, 1 - manual entry

• temperature compensation: 0 - NOTC, 1 - linear TC, 2 - non linear TC

• beep status (1 char):0- off, 1 - on

• temperature unit (1 char): 0 - ºC, 1 - ºF

• backlight (2 chars)

• contrast (2chars)

• auto light off (3chars)

• auto power off (3chars)

• baud rate (5 chars)

• instrument ID (4chars)

• EC alarm time out interval (2 chars)

• NaCl alarm time out interval (2 chars)

• Logging interval (5chars)

• Calibration range check (1 char): 0 - off; 1 - on

• TDS factor (6 chars)

• Temperature coefficient (6 chars)

• Reference temperature (6 chars)

• Cable correction resistor (5 chars)

• Cable correction capacitance (4 chars)

• EC fixed range number (1 char)

• Resistivity fixed range number (1 char)

• TDS fixed range number (1 char)

• Language ID (3 chars)

• Cell constant (7 chars)

Requests the setup parameters setting.

The answer string contains:

• number of profiles (2chars)

• current profile ID (2 chars)

• number of languages (2 chars)

59

NSLx

ULS

Requests the number of logged samples.

The command parameter (1 char):

• E - the request is for EC range

• R - the request is for Resistivity range

• T - the request is for TDS range

• N - the requests is for NaCl range

• U - the requests is for USP range

The answer string is the number of samples (4 chars)

Requests information about the status of the USP log reports.

The answer string contains:

• the number of reports (3 chars)

LLS

For every USP report the following information are sent:

• report ID (3 chars)

• log time (12 chars)

Requests information about the number of lots

The answer string contains:

• the number of lots (3 chars)

For every lot the following information are sent:

• lot ID (3 chars)

• log time (12 chars)

• lot type (1 char): 0 - EC; 1 - Resistivity;

2 - TDS; 3 - NaCl

GLDxxx Requests all the records for the lot with ID = xxx

The answer string contains:

Lot header data:

• logging interval (1 s) (5 chars)

• temperature source (1char)

• temperature compensation mode (1 char)

• reference temperature (2 chars)

• temperature coefficient (6 chars)

• offset factor (6 chars)

• TDS factor (6 chars) (only for TDS lots)

• salinity coefficient (7 chars) (only for salinity lots)

• profile ID (4 chars)

• profile creation time (12 chars)

• lot start time (12 chars)

Lot record data:

60

• EC value (8 chars)

• EC unit (1 char): 0 - m

S; 1 - mS

• EC overrange flag (1 char): R, U, O

• temperature reading, in ºC (8 chars)

Next data is sent if the lot is not for EC range.

• Resistivity or TDS or Salinity (8 chars)

• Reading unit:

• Overage flag (1 char)

• Resistivity: 0 - ohm; 1 - Kohm; 2 - Mohm

• TDS: 0 - ppm; 1 - g/L

• Salinity: 0 - %; 1 - ppt; 2 - PSU

LODxxx Requests the log information for record number “xxx” on range “r”

• “r” is E for EC, R for resistivity, T for TDS and N for salinity

The answering string contains:

• The logged range (2 chars): 10 - EC, 11 - Resistivity,

12 - TDS, 13 - USP, 14 - NaCl %, 15 - Salinity Sea Water,

16 - Salinity PSU

For EC, Resistivity, TDS and salinity ranges:

• EC reading (8 chars)

• EC unit (1 char)

• Temperature source (1 char)

• Temperature compensation mode (1 char)

• Reference temperature (2 chars)

• Temperature coefficient (6 chars)

• Cell constant (7 chars)

• The nearest calibration standard (7 chars)

• The calibration standard unit (1 char)

• Offset factor (6 chars)

• Temperature reading (8 chars)

• Resistivity or TDS or salinity reading (8 chars)

• Unit (1 char)

• TDS factor (6 chars) or salinity coefficient (7 chars)

• Log time (12 chars)

61

For USP records:

• Record ID (3 chars)

• Stage 1 status (1 char): 0 - not verified; 1 - USP met;

2 - USP not met

• Stage 2 status (1 char): see Stage 1

• Stage 3 status (1 char): see Stage 1

• Stage 1 data (if exist):

• EC reading (8 chars)

• EC unit (1 char)

• temperature reading in ºC

• Stage 2 data (if exists) - same as Stage 1

• Stage 3 data (if exists)

• pH value (5 chars)

•USP factor (3 chars)

•Cell constant (1 char)

Note: “Err3” is sent if record does not exist

“Err4” is sent if the identification range character is not recognized

“Err5” is sent if autolog is in progress

“Err6” is sent if the request range is not available

“Err7” is sent if the instrument is logging

“Err8” is sent if the instrument is not in the measuring mode

“Err9” is sent if the battery low condition is detected and the communication cannot be done.

62

BATTERIES

RECHARGING/REPLACEMENT

The instrument is ready with rechargeable batteries inside.

First time you start working with the instrument or when the rechargeable batteries are changed with new ones, perform the following procedure:

• Work with the instrument until the rechargeable batteries are fully discharged

• Perform a complete charging cycle (about 16hours).

Repeat this procedure 3 times.

It is recommended to recharge the rechargeable batteries as soon as the battery indicator does not show full scale before starting in field measurements or every time you finish your work with your instrument.

HI 98185 E Meter

If the battery capacity is less than 20 % the serial communication and the backlight feature are not available.

To replace the rechargeable batteries, follow the next steps:

• Turn OFF the instrument.

• Unscrew the screw from the bottom side of the instrument.

• Remove the battery holder and the old batteries.

• Insert four new 1.2V AA 1300 mAh NiMH rechargeable batteries in the battery compartment while paying attention to the correct polarity.

• Push the battery holder and tighten the screws.

63

To recharge the rechargeable batteries, follow the next steps:

• Connect the 12Vdc power adapter to the main line and to the main line of thebattery recharger. The front LED will turn ON.

• Place the instrument in the battery recharger case. A charging animation will be displayed if the battery capacity is less than 100 %.

HI 98185

Meter

• The complete charging process takes about 16 hours.

Notes: • As the charging process is performed at low current, the instrument can be left on the recharger more than 16 hours, without damaging the rechargeable batteries.

• It is recommended to turn off the instrument while recharging the batteries. The measurements can be affected by the recharging process.

• Batteries recharging must only take place in a non hazardous area, using the HI 710042 inductive recharger.

WARNING:

Do not replace the rechargeable batteries with alkaline batteries.

Do not put ever the instrument with alkaline batteries inside on the recharger.

The manufacturer don’t assume any obligation for malfunctioning appeared as a result of using alkaline batteries.

Note: The instrument is provided with the BEPS (Battery Error Prevention

System) feature, which automatically turns the instrument off when the batteries level is too low to ensure reliable readings.

64

R e a d a n d lb n

T h e to p er ni ss s k

S Y ni ni at ni g tr ts

.g

M P T O ni g d o w n f ul c n( ut io

D fo si p al y ht e s h o ar n g e w s

M S ta es se

.) ht er a d e u p ot p ni g

P R O B L E M

E C p oc n n or b e ce et .d

n to p or p re yl

R e a d ar n g .e

ni g o tu fo

B or k e n E C p or b .e

nI es tr ht e

S O L U p or b

T

.e

I O N

R ce

C h e kc la bi ht ar et e ht e s a m p m e el si et ;r w i ht ar n g ni

.e

C if h e ex d kc ht e m e a ro ht e ol ar n g e kc e .d

s u si ar b el n to

R e p al ec ht e p or b .e

M te re f ia sl er ro a g vi d ni g se

.s

af ot c lu yt la bi ar et tA ts ra ut p ht e m te re d si p al at g s p sy re m

H a n n a a n e n lt

L o g o

.y

M te re s h u st o .ff

" E s at tr rr o u

"r

.p

m e ss a g e ro

O n e b ol kc e fo

.d

ht e k e sy si

D

A e u a ot d e n a b el a off

:d cc u m ef a ni lu ut er ht si ta si ro ,s c a s a tf re

,e m s le fo n o n te ce u s .e

re et d s h u st p o re oi d ff nI et nr la re or .r

ur m w h e e n tn

O N / O F d o

.F

se nI iti la zi ta oi n re or .r

C o n

H a n at n

C e n et .r

a tc y o ru d e la nI ts ur m e n st re

S ta a n y re

R er p ce h a al ec gr e a a cc u m u cc u m u al al ot ;r ot sr

P er ss O N / O F F .

C o n at a n y tc y o

H a n n a ru

S d e re

C e n et .r

iv la ec re ro

P

O er

N / ss

O a

F n

F d f h o ro a dl b a ht cc

.c

n es e u m ro oc n u d n al si e ot oc n n e

.r

tc o n e tc d o o tu w n

2 0 a n d iv ec ro

65

Rinse the probe with clean water after measurements. If more cleaning is required, remove the probe sleeve and clean the probe with a cloth or a nonabrasive detergent. Make sure to reinsert the sleeve onto the probe properly and in the right direction. After cleaning the probe, recalibrate the instrument.

The platinum rings are sustained with glass. Take great care while handling the probe.

The coating of platinum black on the electrodes of the conductivity cell should be inspected before and after use for signs of flaking or loss of material. If the coating seems poor, cleaning and replatinization is required.

To perform probe replatinization, enter calibration screen from EC range by pressing CAL key.

Put the probe into a beaker with replatinization solution and press Start.

66

The process can be interrupted by pressing Stop or ESC key.

The replatinization process takes about 5 minutes.

Remove the probe from the replatinization solution and rinse it with deionizated water.

EC CALIBRATION SOLUTIONS

HI 70033C 84 m

S/cm, 20 ml sachet, 25 pcs.

HI 70031C 1413 m

S/cm, 20 ml sachet, 25 pcs.

HI 70039C 5000 m

S/cm, 20 ml sachet, 25 pcs.

HI 70030C 12880 m

S/cm, 20 ml sachet, 25 pcs.

HI 6033 84 m

S/cm, 500 ml, bottle

HI 6031 1413 m

S/cm, 500 ml, bottle

HI 7039L 5000 m

S/cm, 500 ml, bottle

HI 7030L 12880 m

S/cm, 500 ml, bottle

HI 7034L 80000 m

S/cm, 500 ml, bottle

HI 7035L 114800 m

S/cm, 500 ml, bottle

HI 7037L 100% NaCl, 500 ml, bottle

RECOMMENDATIONS FOR USERS

Before using these products, make sure they are entirely suitable for the environment in which they are used.

Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to follow all necessary steps to correct interferences.

The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.

During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges.

Any variation introduced by the user to the supplied equipment may degrade the instruments’

EMC performance.

To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vcc.

To avoid damage or burns, do not perform any measurement in microwave ovens.

67

SALES AND TECHNICAL SERVICE CONTACTS

Australia:

Tel. (03) 9769.0666 • Fax (03) 9769.0699

China:

Tel. (10) 88570068 • Fax (10) 88570060

Egypt:

Tel. & Fax (02) 2758.683

Germany:

Tel. (07851) 9129-0 • Fax (07851) 9129-99

Greece:

Tel. (210) 823.5192 • Fax (210) 884.0210

Indonesia:

Tel. (21) 4584.2941 • Fax (21) 4584.2942

Japan:

Tel. (03) 3258.9565 • Fax (03) 3258.9567

Korea:

Tel. (02) 2278.5147 • Fax (02) 2264.1729

Malaysia:

Tel. (603) 5638.9940 • Fax (603) 5638.9829

Singapore:

Tel. 6296.7118 • Fax 6291.6906

South Africa:

Tel. (011) 615.6076 • Fax (011) 615.8582

Taiwan:

Tel. 886.2.2739.3014 • Fax 886.2.2739.2983

Thailand:

Tel. 66.2619.0708 • Fax 66.2619.0061

United Kingdom:

Tel. (01525) 850.855 • Fax (01525) 853.668

USA:

Tel. (401) 765.7500 • Fax (401) 765.7575

For e-mail contacts and a complete list of Sales and

Technical offices, please see

www.hannainst.com.