Buchi FatExtractor E-500 Operation Manual
Advertisement
Advertisement
FatExtractor E-500
Operation Manual
Imprint
Product Identification:
Operation Manual (Original) FatExtractor E-500
11593950
Publication date: 01.2019
BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
E-Mail: [email protected]
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experience, especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for competitive purposes, nor made available to third parties. The manufacture of any component with the aid of this manual without prior written agreement is also prohibited.
BÜCHI Labortechnik AG Table of contents
Table of contents
Operation Manual FatExtractor E-500 iii
Table of contents BÜCHI Labortechnik AG
iv Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Table of contents
Operation Manual FatExtractor E-500 v
Table of contents BÜCHI Labortechnik AG
vi Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
1
About this document | 1
About this document
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the instructions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions resulting from not following this operation manual.
If you have any questions after reading this operation manual: u Contact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
1.1
Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device. There are four danger levels, each identifiable by the signal word used.
Signal word
DANGER
WARNING
CAUTION
NOTICE
Meaning
Indicates a danger with a high level of risk which could result in death or serious injury if not prevented.
Indicates a danger with a medium level of risk which could result in death or serious injury if not prevented.
Indicates a danger with a low level of risk which could result in minor or medium-severity injury if not prevented.
Indicates a danger that could result in damage to property.
1.2
1.2.1
Symbols
The following symbols are displayed in this operation manual or on the device:
Warning symbols
Symbol Meaning
General warning
Hot surface
Breakable items
Hand injuries
1.2.2
Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
Operation Manual FatExtractor E-500 7/84
1 | About this document BÜCHI Labortechnik AG
R This character draws attention to a requirement that must be met before the instructions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-up
Window
Tab
Dialog
[ Button ]
[ Field names ]
[ Menu / Menu item ]
Status
Signal
Explanation
Software Windows are marked-up like this.
Tabs are marked-up like this.
Dialogs are marked-up like this.
Buttons are marked-up like this.
Field names are marked-up like this.
Menus or menu items are marked-up like this.
Status is marked-up like this.
Signals are marked-up like this.
1.3
Trademarks
Product names and registered or unregistered trademarks that are used in this document are used only for identification and remain the property of the owner in each case.
8/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
2 Safety
2.1
2.2
2.3
Safety | 2
Proper use
The FatExtractor E-500 is designed for the extraction (solid-liquid extraction) of samples with known solvents and the reduction or drying of the extract.
The FatExtractor E-500 can be used in laboratories for the following tasks:
Soxhlet Extraction
Hot Extraction
Economic Continuous Extraction
The FatExtractor E-500 can be used with following solvents:
Chloroform
Diethyl ether
Hexane
Petroleum ether
Other solvents only with the prior written approval of BÜCHI Labortechnik AG.
Use other than intended
Use of any kind other than that described in Chapter 2.1 "Proper use", page 9 and
any application that does not comply with the technical specifications constitutes use other than that intended.
In particular, the following applications are not permissible:
Use of the instrument in areas which require explosion-safe apparatus.
Processing samples which could be caused to explode or ignite by impact, friction, heat or sparks (e.g. explosives).
Use of the instrument with samples that react with solvent.
Use of the instrument with solvents containing peroxides.
Use of the instrument with other than original BUCHI glassware.
Use of the instrument in overpressure situations.
Use of solvents without the prior written approval of BÜCHI Labortechnik AG.
Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the applicable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess the risks associated with the use of the device.
Operation Manual FatExtractor E-500 9/84
2 | Safety
2.4
BÜCHI Labortechnik AG
Operator
The operator (generally the laboratory manager) is responsible for the following aspects:
The device must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the operations described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to the manufacturer ([email protected]).
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair measures.
Location of warning signs on the product
10/84
Hot surfaces Hand injuries
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
2.5
Protection devices
1 2
Safety | 2
2.6
2.6.1
2.6.2
2.6.3
Fig. 1: Protection devices
1 Protection shield 2 Protection shield
Residual risks
The device has been developed and manufactured using the latest technological advances. Nevertheless, risks to persons, property or the environment can arise if the device is used incorrectly.
Appropriate warnings in this manual serve to alert the user to these residual dangers.
Hot surfaces
The surfaces of the device can become very hot. If touched they can cause skin burns.
u Do not touch hot surfaces or else wear suitable protective gloves.
Empty beakers on heating plates
Empty beakers on active heating plates lead to very hot surfaces and can damage the instrument u u
Remove empty beakers from the instrument.
Deactivate heating positions not used.
Mechanical damages
If a device is damaged, sharp edges, moving parts or exposed electrical wires can cause injuries.
u u u
Check device regularly for visible damage.
Make sure that the locking mechanism of the protection shield is working properly.
If faults occur:
Switch off the device immediately.
Unplug the power cord.
Inform the operator.
Cool the device before taking any further action.
u Do not continue to use devices that are damaged.
Operation Manual FatExtractor E-500 11/84
2 | Safety
2.6.4
2.6.5
2.6.6
2.7
2.8
BÜCHI Labortechnik AG
Glass breakage
Broken glass can cause severe cuts.
Not visible cracks on glassware can lead to breakage.
Minor damage to the ground joints impairs the sealing effect and lead to loss of solvent during the extraction process.
u u u u u u u
Wear protective gloves when handling broken glassware.
Handle the glass components carefully and do not drop them.
Always place the glassware in a suitable holder when they are not in use.
Prevent any mechanical impact.
Prevent thermal shocks on glassware.
Always visually inspect glass components for damage every time they are used.
Do not continue to use glass components that are damaged.
Glass damages during extraction process
Glass damages during the extraction process can cause instrument damages caused by solvents, coolant and fire.
u If fault occurs:
Switch off the device immediately.
Unplug the power cord.
Inform the operator.
Let the temperature of the instrument decrease under observation until it is the same as the ambient.
Overfilled extraction beakers
Overfilled extraction beakers can cause spills and lead to instrument damage or fire.
u u
Do not fill more than 170 mL in the extraction beakers.
Do not fill more than 280 mL in the extraction beakers LSV.
Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may arise.
u u
Always wear appropriate personal protective equipment such as safety goggles, protective clothing and gloves.
Make sure that the personal protective equipment meets the requirements of the safety data sheets for all chemicals used.
Modifications
Unauthorized modifications may impair safety and lead to accidents.
u u
Use only genuine BUCHI accessories, spare parts and consumables.
Technical modifications to the device or accessories should only be carried out with the prior written approval of BÜCHI Labortechnik AG and only by authorized
BUCHI technicians.
BUCHI accepts no liability whatsoever for damage arising as a result of unauthorized modifications.
12/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
3 Product description
3.1
3.1.1
Product description | 3
Description of function
The FatExtractor E-500 is designed to carry out the following solid-liquid extraction methods:
Soxhlet Extraction (see Chapter 3.1.1 "Soxhlet Extraction", page 13)
Hot Extraction (see Chapter 3.1.2 "Hot Extraction", page 14)
Economic Continuous Extraction (see Chapter 3.1.3 "Economic Continuous Extraction", page 14)
Soxhlet Extraction
Step 1 extraction
The sample is located in the extraction chamber.
The beaker contains the solvent.
The solvent is heated, vapor rises up to the condenser, condenses and drops into the extraction chamber with the sample.
The magnetic valve is closed, the solvent is collected up to the optical sensor and extracts the analyte.
When the optical sensor is reached, the magnetic valve opens and the solvent containing the analyte flows back into the beaker.
Step 2 rinsing
The solvent is heated, vapor rises up to the condenser, condenses and drops into the extraction chamber with the sample.
The magnetic valve is open, the solvent flows back into beaker, the solvent is not collected.
This flushes traces of fat into the beaker.
Step 3 drying
The solvent is heated, vapor rises up to the condenser, condenses and flows into tank.
The analyte remains in the beaker.
Operation Manual FatExtractor E-500 13/84
3 | Product description
3.1.2
3.1.3
BÜCHI Labortechnik AG
Hot Extraction
Step 1 extraction
The sample is located in the beaker.
The beaker contains the solvent.
The solvent is heated, vapor rises up to the condenser, condenses and drops into the beaker with the sample.
Step 2 rinsing
The solvent in the beaker is heated up and evaporated.
The vapor rises up to the condenser.
The condensed solvent flows into the beaker with the sample.
The tank bottle valve opens periodically and condensed solvent flows in the tank bottle.
The solvent level reduce.
Step 3 drying
The solvent in the beaker is heated up and evaporated.
The vapor rises up to the condenser.
The tank bottle valve is open and condensed solvent flows in the tank bottle.
The analyte remains in the beaker.
Economic Continuous Extraction
Step 1 extraction
The sample is located in the extraction chamber.
The beaker contains the solvent.
The solvent is heated, vapor rises up around the sample to the condenser, condenses and drops back into extraction chamber through the sample into the beaker.
Step 2 drying
The solvent is heated, vapor rises up around the sample to the condenser, condenses and flows into tank.
The analyte remains in the beaker.
14/84
3.1.4
Description of the Analyte Protection (option)
The Analyte Protection protects the analyte from overheating.
NOTE
The presence of beaker and solvent is always detected wether the Analyte Protection is activated or deactivated.
The Analyte Protection triggers under the following situations:
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
Presence of beaker
Presence of solvent
(Level check)
Extraction step
Rinse step
Dry step
X
X
X
X
X
Soxhlet
Extraction
X
X
Hot
Extraction
Product description | 3
Economic
Continuous
Extraction
X
X
X
X
(1. Interval)
-
X
X
X
Operation Manual FatExtractor E-500 15/84
3 | Product description
3.2
3.2.1
Configuration
Front view
1 2 3
BÜCHI Labortechnik AG
4
11
10
Fig. 2: Front view
1 Protection shield
3 Chamber rack
5 Solvent tank
7 Installation board
9 Heating plate
11 Interface
(Pro)
9 8 7 6
2 Condenser rack
4 Condenser solvent tank
6 Beaker rack
8 Analyte Protection sensor
(Option)
10 On/Off master switch
5
16/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
3.2.2
Rear view
1
Product description | 3
3.2.3
Fig. 3: Rear view
1 Upper connections
See Chapter 3.2.3 "Upper connections on the rear side", page 17
2
Upper connections on the rear side
1 2
2 Lower connections
See Chapter 3.2.4 "Lower connections on the rear side", page 18
3 4
Fig. 4: Connections on the top
1 Standard BUCHI communication port
(COM)
2 Standard BUCHI communication port
(COM)
3 LAN port 4 Port for the cooling water valve
Operation Manual FatExtractor E-500 17/84
3 | Product description
3.2.4
Lower connections on the rear side
1 2 3 4 5 6 7
BÜCHI Labortechnik AG
12 11
Fig. 5: Connections on the buttom
1 Coolant water solvent tank in
3 Fuse
5 Fuse
7 Coolant water out
9 Coolant water in
11 Coolant water solvent tank out
10 9 8
2 Solvent in
4 Power supply connection
6 Inert gas connection
(option)
8 Exhaust
10 Type plate
12 Solvent out
18/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
3.2.5
Installation board
13
12
11
10
1 2 3
Product description | 3
6
7
4
5
9 8
Fig. 6: Installation board
1 For UniversalExtractor E-800
3 For UniversalExtractor E-800
5 Condenser rack Economic
Continuous Extraction
7 Chamber rack Economic Continuous
Extraction
9 Beaker rack Soxhlet Extraction /
Economic Continuous Extraction
11 Sensor plugs
13 Lift
2 Condenser rack Soxhlet Extraction
4 Chamber rack Soxhlet Extraction
6 Condenser rack Hot Extraction
8 Beaker rack Hot Extraction
10 Solvent collector tube
12 Attachment for tube
Operation Manual FatExtractor E-500 19/84
3 | Product description
3.2.6
Interface
1 2 3 4
BÜCHI Labortechnik AG
Fig. 7: Interface
1 Stop button
3 Function buttons
2 Screen
4 Navigation control
20/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
3.2.7
Pro Interface
1 2 3
Product description | 3
3.3
Fig. 8: Interface Pro
1 Stop button
3 Navigation control
2 Touch-screen display
Type plate
The type plate identifies the instrument. The type plate is located at the rear of the instrument.
Operation Manual FatExtractor E-500 21/84
3 | Product description BÜCHI Labortechnik AG
3.4
3.5
3.5.1
1 2 3
11
10
9
8
7
6
Buchi Labortechnik AG
9230 Flawil / Switzerland
Type:
SN:
Volt:
Freq.:
Power: xxx
10xxxxxxxx xxx-xxx VAC xx Hz xxxxW Built 20xx
5
Fig. 9: Type plate
1 Symbol for "electronics recycling"
3 Approvals
5 Year of manufacture
7 Frequency
9 Serial number
11 Company name and address
4
2 Product code
4 Symbol for "Do not dispose of as household waste"
6 Power consumption maximum
8 Input voltage range
10 Instrument name
Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and delivery note.
Technical data
FatExtractor E-500
Specification
Power consumption
Connection voltage
Fuse
Frequency
Overvoltage category
IP Code
Pollution degree
Protection Class
Dimensions (W x D x H)
Weight
Total heating power
(rated)
Total heating power
(maximum)
E-500
1300 W
100 - 240 ± 10 % VAC
10 A
50 / 60 Hz
II
IP20
2
1
638 x 595 x 613 mm
42 kg
600 W
1200 W
22/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
3.5.2
3.5.3
Specification
Hose connection
Allowed water pressure
( nominal value)
Allowed water pressure
(maximum)
Number of extraction positions
Solvent tank volume
Ambient conditions
For indoor use only.
Max. altitude above sea level
Ambient temperature
Maximum relative humidity
Storage temperature
Materials
Component
Housing
Exhaust
Protection shield
Glass ware
Sealings glassware
Solvent tank
Solvent tank valve
Tubing
Product description | 3
E-500
6 / 9 mm
6 bar
8 bar
6
2 L
2000 m
5 - 40°C (25°C)
80% for temperatures up to 31°C max. 45 °C
Materials of construction
Steel 1.4301/304 with powder coating
Aluminium
PE-UHMW 1000
PMMA GS
Borosilikat 3.3
FKM
Borosilikat 3.3
PTFE / FFKM
FEP
Operation Manual FatExtractor E-500 23/84
4 | Transport and storage
4 Transport and storage
4.1
BÜCHI Labortechnik AG
Transport
NOTICE
Risk of breakage due to incorrect transportation
Make sure that the instrument is fully dismantled.
Pack every instrument components properly to prevent breakage. Use the original packaging whenever possible.
Avoid sharp movements during transit.
4.2
4.3
u u u
After transporting, check the instrument and all glass components for damage.
Damage that has occurred in transit should be reported to the carrier.
Keep packaging for future transportation.
Storage
u u u
Make sure that the ambient conditions are complied with (see Chapter 3.5 "Technical data", page 22).
Wherever possible, store the device in its original packaging.
After storage, check the device, all glass components, seals and tubing for damage and replace if necessary.
Lifting the instrument
WARNING
Danger due to incorrect transportation
The possible consequences are crushing injuries, cuts and breakages.
u u
The instrument should be transported by two persons at the same time.
Lift the instrument at the points indicated.
NOTICE
Dragging the instrument can damage the feet of the instrument.
u Lift the instrument when positioning or re-locating.
24/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Transport and storage | 4
Fig. 10: Lifting the instrument u Lift the instrument – this requires two persons each lifting at the points indicated on the left and right side of the instrument.
Operation Manual FatExtractor E-500 25/84
5 | Installation
5
5.1
BÜCHI Labortechnik AG
Installation
Before installation
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
u Climatize the instrument after transportation.
5.2
5.3
Installation site
The installation site must meet the following requirements:
Firm, level surface
1 m minimum space requirement above the instrument or inside a fume hood.
Take into account the maximum product dimensions and weight.
Do not expose the instrument to any external thermal loads, such as direct solar radiation or other ignition source.
Do not place any flammable devices or substances close to the instrument, as they could ignite due to the heat from the heating plates.
NOTE
Make sure that the power supply can be disconnected at any time in an emergency.
Establishing electrical connections
NOTE
Observe the regulatory provisions when connecting the instrument to power supply.
4 Use external main switches (e.g., emergency off) in accordance with the standards IEC 60947-1 and IEC 60947-3.
4 Use additional electrical safety features (e.g., residual-current circuit breakers) to comply with local laws and regulations.
The power supply must fulfill the following conditions:
1. Provide the mains voltage and frequency specified on the type plate of the instrument.
2. Be designed for the load imposed by the instruments connected.
3. Be equipped with suitable fuses and electrical safety features.
4. Be equipped with a proper grounding.
26/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Installation | 5
NOTICE
Risk of property damage and diminished performance due to use of unsuitable power cables.
The power supply cables supplied with the product by BUCHI precisely match the requirements of the device. If other power cables that do not meet those requirements are used, the device may be damaged and/or its performance diminished.
u u
Use only the power supply cables supplied with the product or ordered separately from BUCHI.
If using any other power supply cables, make sure that they match the specifications on the type plate.
5.4
u u u u
Make sure that all connected devices are grounded.
Make sure that the power plug is freely accessible at all times.
Insert the power cable into the connection labeled Power IN on the back of the instrument.
Insert the power plug into the power socket.
Securing against earthquakes
The instrument has an earthquake fixing point to protect the device against falling.
1
5.5
Fig. 11: Securing against earthquakes
1 Lashing mount u Tie the lashing mount to a fixed point using strong cord or a wire.
Connecting the coolant hose
There are two possibilities for coolant supply:
Cooling water valve Order no. 031356
BUCHI Recirculating Chiller F-305 or F-308
Operation Manual FatExtractor E-500 27/84
5 | Installation
5.6
BÜCHI Labortechnik AG
R The coolant connection complies with the
specified parameters. See Chapter 3.5 "Technical data", page 22
R Make sure that the instrument is not connected to the power supply.
u u u u
Install the inlet tube onto the connection marked CW IN .
Attach the inlet tube in place with a hose clip.
Install the drain tube onto the connection marked CW OUT .
Attach the drain tube in place with a hose clip.
Connecting the solvent tubes to the condenser tank bottle
R Make sure that the instrument is not connected to the power supply.
u u
Attach the inlet tube onto the connection marked TANK IN .
Attach the drain tube onto the connection marked TANK OUT (SOLVENT) .
u Attach the solvent connection nuts onto the connection marked OUT and IN (SOLVENT) .
u u
Attach the inlet tube onto the connection marked OUT .
Attach the drain tube onto the connection marked IN (SOLVENT) .
28/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
5.7
Installation | 5
Connecting the coolant tubes to the condenser tank bottle
R Make sure that the instrument is not connected to the power supply.
u u u u
Install the inlet tube onto the connection marked CW TANK OUT .
Attach the inlet tube in place with a hose clip.
Install the drain tube onto the connection marked CW TANK IN .
Attach the drain tube in place with a hose clip.
u u
Attach the inlet tube onto the connection marked CW IN .
Attach the drain tube onto the connection marked CW OUT .
5.8
Preparing the tank bottle
NOTICE
Risk of broken parts by using tools
Using tools for preparing the instrument can break parts.
u Tighten all parts with your hands.
u Switch the On/Off master switch to Off.
u Put the bottle cap on the bottle.
Operation Manual FatExtractor E-500 29/84
5 | Installation u Put the tank bottle in the instrument.
BÜCHI Labortechnik AG u Attach the ball joint clamp.
u Tighten the knurled nut at the ball joint clamp.
30/84
5.9
5.9.1
Preparing the instrument for Soxhlet Extraction
Installations for Soxhlet Extraction
NOTE
Condenser rack, chamber rack and beaker rack are already installed according to the ordered method.
For Soxhlet Extraction use the following positions:
Device
Beaker rack
Chamber rack
Condenser rack
Position on the installation board lower
Nr. 4
Nr. 2
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
Preparing the instrument u Switch the On/Off master switch to Off.
u Attach the condenser rack to the instrument with screws.
u
Prepare the condenser. See Chapter 5.13
"Preparing the condenser", page 39
u Attach the chamber rack to the instrument with screws.
u Attach the beaker rack to the instrument with screws.
u Prepare the extraction glass chamber Soxhlet.
See Chapter 5.9.3 "Preparing the chamber for a Soxhlet Extraction", page 32
u Install the protection shields. See Chap-
ter 5.12.1 "Installing two protection shields
u Select the Soxhlet method on the interface.
See Chapter 5.14 "Selecting a configuration", page 41
Installation | 5
Operation Manual FatExtractor E-500 31/84
5 | Installation
5.9.2
Preparing the extraction glass chamber Soxhlet
NOTICE
Risk of broken parts by using tools
Using tools for preparing the instrument can break parts.
u Tighten all parts with your hands.
BÜCHI Labortechnik AG u Attach the valve to the extraction glass chamber at the point indicated.
u Attach the sensor to the extraction glass chamber.
u Attach the sealing to the extraction glass chamber.
The extraction glass chamber is prepared.
32/84
5.9.3
Preparing the chamber for a Soxhlet Extraction u Prepare the extraction glass chamber Soxhlet.
See Chapter 5.9.2 "Preparing the extraction glass chamber Soxhlet", page 32
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG u Pull the chamber rack.
u Put the extraction glass chamber in the chamber rack.
u Turn the extraction glass chamber 90° counterclockwise.
Installation | 5 u Make sure, that the extraction glass chamber is installed correctly.
u Connect the valve with the connection on the installation board.
Operation Manual FatExtractor E-500 33/84
5 | Installation u Push the chamber rack in the instrument.
BÜCHI Labortechnik AG
34/84
5.10
Preparing the instrument for Hot Extraction
5.10.1
Installations for Hot Extraction
NOTE
Condenser rack, chamber rack and beaker rack are already installed according to the ordered method.
For Hot Extraction use the following positions:
Device
Beaker rack
Condenser rack
Position on the installation board upper
Nr. 6
Preparing the instrument u Switch the On/Off master switch to Off.
u Attach the condenser rack to the instrument with screws.
u
Prepare the condenser. See Chapter 5.13
"Preparing the condenser", page 39
u Attach the beaker rack to the instrument with screws.
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Installation | 5 u Install the protection shields. See Chap-
ter 5.12.2 "Installing one protection shield (Hot and Economic Continuous Extraction only)", page 38
u Select the Hot Extraction method on the inter-
face. See Chapter 5.14 "Selecting a configuration", page 41
5.11
Preparing the instrument for Economic Continuous Extraction
5.11.1
Installations for Economic Continuous Extraction
NOTE
Condenser rack, chamber rack and beaker rack are already installed according to the ordered method.
For Economic Continuous Extraction use the following positions:
Device
Beaker rack
Chamber rack
Condenser rack
Position on the installation board lower
Nr. 7
Nr. 5
Preparing the instrument u Switch the On/Off master switch to Off.
u Attach the condenser rack to the instrument with screws.
u
Prepare the condenser. See Chapter 5.13
"Preparing the condenser", page 39
u Attach the chamber rack to the instrument with screws.
Operation Manual FatExtractor E-500 35/84
5 | Installation u Attach the beaker rack to the instrument with screws.
BÜCHI Labortechnik AG u Prepare the Economic Continuous Extraction
chamber. See Chapter 5.11.3 "Preparing the
Economic Continuous Extraction chamber", page 36
u Install the protection shields. See Chap-
ter 5.12.2 "Installing one protection shield (Hot and Economic Continuous Extraction only)", page 38
u Select the Economic Continuous Extraction
method on the interface. See Chapter 5.14
"Selecting a configuration", page 41
5.11.2
Preparing the extraction glass chamber for an Economic Continuous Extraction u Attach the sealing to the extraction glass chamber.
36/84
5.11.3
Preparing the Economic Continuous Extraction chamber u Prepare the extraction glass chamber. See
u Pull the chamber rack.
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG u Put the extraction glass chamber in the instrument.
u Push the chamber rack in the instrument.
Installation | 5
5.12
Installing the protection shields
5.12.1
Installing two protection shields (Soxhlet only) u Make sure, that both protections shields fit together.
u Hold both protection shields on both sides at the point indicated.
Operation Manual FatExtractor E-500 37/84
5 | Installation u Hold both protection shields on both sides at the point indicated.
BÜCHI Labortechnik AG u Put the rear protection shield on both sides into the rear slot at the same time.
u Put the front protection shield on both sides into the front slot at the same time.
5.12.2
Installing one protection shield (Hot and Economic Continuous Extraction only) u Hold the protection shield on both sides at the point indicated.
38/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG u Put the protection shield on both sides into the front slot at the same time.
Installation | 5
5.13
Preparing the condenser
NOTICE
Risk of broken parts by using tools
Using tools for preparing the instrument can break parts.
u Tighten all parts with your hands.
u Attach the solvent connection nut to the condenser.
u Connect the solvent collector tube.
Operation Manual FatExtractor E-500 39/84
5 | Installation u Put the condenser in the condenser rack.
BÜCHI Labortechnik AG u Secure the condenser with the flange lock.
u Turn the condenser.
u Connect the solvent collector tube to the installation board
40/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG u Connect the condensers.
u Establish a condenser circuit.
u Make sure that the condenser circuit
5.14
Selecting a configuration
Navigation path
➔ ➔ [ Settings ]
Precondition:
R Switch the On/Off master switch to On.
u u
Navigate to the [ Settings ] submenu via the navigation path.
Tap the action [ Select configuration ] .
ð The display shows a dialog with the selectable configurations.
u Tap the configuration according to the installed glass setup.
ð The configuration is selected.
ð The dialog box closes.
5.15
Disassembling the condenser circuit
u Disconnect the indicated condenser connection.
Installation | 5
Operation Manual FatExtractor E-500 41/84
5 | Installation u Make sure, that there is no coolant in the condenser circuit.
u Disconnect the other condenser connections.
BÜCHI Labortechnik AG u Remove the flange lock.
u Pull the condenser out of the condenser rack.
u Disconnect the solvent collector hose.
u Disconnect the solvent collector tube from the installation board.
42/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
6 Operating the interface
CAUTION
Risk of injury from glass splinters
Sharp objects can damage the display.
u Keep sharp objects away from the display.
Operating the interface | 6
6.1
6.2
Layout of the interface
4
5
2
3
6
1
2
3
4
5
6
7
Fig. 12: Interface
No.
1
Description
Status bar
Menu bar
Content area
Function bar
Function buttons
Stop button
7 Navigation control
Function
Shows the actually used solvent and extraction method.
Shows symbols representing the menus.
Shows current settings, submenus or actions depending on current operation.
Shows functions that can be performed according to the current operation.
Pressing a function button performs the assigned function on the function bar.
The extraction stops.
The valves close.
The heaters switch off.
The coolant keep running 5 min.
Used for navigating the user interface.
Pressing the control performs the assigned function on the function bar.
Function bar
The function bar shows available functions according to the current operation.
The functions on the function bar are executed by tapping the relevant function buttons.
Operation Manual FatExtractor E-500 43/84
6 | Operating the interface
6.3
Symbol Description
[ Back ]
O K [ Confirm ]
DELETE
[ Delete ]
S AV E [ Save ]
M E N U [ Menu ]
S TA R T
STOP
[ Start ]
[ Stop ]
E S C [ Escape ]
D O W N [ Down ]
U P [ Up ]
C O P Y [ Copy ]
SKIP [ Skip ]
VIEW [ View ]
ABORT
[ Abort ]
BÜCHI Labortechnik AG
Meaning
The display reverts to the previous view.
Confirms an entry.
Deletes the selected method or step.
Saves the setting.
Allows selection of a menu from the menu bar using the navigation control.
Starts the extraction process.
Aborts extraction on all extraction positions.
Cancels the actual entry.
The lift moves down.
The lift moves up.
Copies the selected method.
Skips the actual extraction step.
Shows the parameters of the running extraction.
Aborts extraction on the selected extraction position.
Menu bar
The menus are represented by symbols on the menu bar. Navigation through the menus is by input controls.
The following menus are available:
Menu symbol Meaning
[ Home ] menu
Submenu/Actions
Process control parameters
[ Method ] menu
[ Configuration ] menu
Editing and saving extraction methods
Change settings
Service menu
System information
44/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
Menu symbol
6.3.1
Home menu
1
Meaning
[ Solvent ] menu
Operating the interface | 6
Submenu/Actions
Solvent library
Editing and saving solvents
2 3
6.3.2
6.3.3
4
5
Fig. 13: Interface
1 Time remaining
3 Status
5 Extraction cycles
(Soxhlet Extraction only)
2
4
Method
Extraction positions
Method menu
The [ Method ]
menu allows saving methods. See Chapter 8 "Editing a method", page 54
Configuration menu
On the [ Configuration ] menu, you can enter a variety of settings and retrieve information.
Submenu Settings
The submenu [ Settings ] contains system settings for the instrument.
Action
[ Language ]
[ Beeper ]
[ Display brightness ]
[ Configuration ]
[ Max. Cycle time ]
(Soxhlet Extraction only)
Option
Choice of display language on the interface
Off/On
Enter setting
Choice of extraction method
Enter value
Explanation
The following languages are available:
English/German/French/Italian/
Spanish/Japanese/Chinese
Setting for audible signal in response to input controls.
Display brightness level in %: 0 -
100
The following configurations are available:
SOX (Soxhlet Extraction) / HE (Hot
Extraction) / ECE (Economic
Continuous Extraction)
Maximum cycle time in minutes: 0 -
240
Operation Manual FatExtractor E-500 45/84
6 | Operating the interface BÜCHI Labortechnik AG
Action
[ Mobile connection
QR code ]
[ Mobile connection password ]
[ Network ]
[ Delete app connection ]
[ Demo Mode ]
Option
View
View
Enter value
Explanation
The control unit shows a QR code for reading by the Extraction Reports app.
The control panel shows a password for entry in the Extraction Reports app.
The following parameters can be edited:
Device name/DHCP/MAC address/
System IP address/Subnet mask/
Gateway
Confirmation question Resets external connections to the instrument.
On/Off Simulate an extraction.
Submenu Service
WARNING
Incorrect use of the actor settings.
Incorrect use of the actors settings can cause instrument damages and injuries.
u Carry out the actor setting as described.
Action
[ Sensors ]
Option
View
Explanation
All sensor signals can be observed:
Flow senors
Level sensors
Analyte protection (option)
Protection shield
Valves
Lift current
Line voltage
Line frequency
46/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
6.3.4
Operating the interface | 6
Action
[ Actors ]
Option
Change settings
Explanation
The following settings can be changed:
Valves:
Soxhlet valves See Chapter 13.2
"Draining solvent from Extraction glass chamber Soxhlet", page 76
"Draining solvent from the instrument", page 74
Cooling water valve On/Off
Heater On/Off
Pump On/Off
Lift current
Beeper On/Off
Chiller On/Off (option)
Submenu System information
The submenu [ System information ] contains details of the connected devices and information on network connection diagnosis.
Action
[ Extraction unit ]
[ Interface ]
Option
View
View
Explanation
The following information on the
Extraction Unit is available:
Serial Number
Firmware version
Operating hours
Temperature electronics
Code
Line voltage
Line frequency
The following information on the interface is available:
Serial Number
Firmware version
Operating hours
Board temperature
Code
30 V Operating voltage
5 V Operating voltage
Solvent menu
The solvent menu contains the available solvents for the extraction methods.
The solvent menu allows editing individual solvents. See Chapter 10 "Editing a solvent", page 69
6.4
Status bar
Shows the actually used solvent and configuration.
Operation Manual FatExtractor E-500 47/84
7 | Operating the interface pro
7 Operating the interface pro
CAUTION
Risk of injury from glass splinters
Sharp objects can damage the display.
u Keep sharp objects away from the display.
7.1
Layout of the interface
1
2
3
BÜCHI Labortechnik AG
48/84
4
2
3
4
5
6
Fig. 14: Interface Pro
No.
1
Description
Status bar
Menu bar
Content area
Function bar
5 Stop button
6 Navigation control
Function
Shows the currently used solvent and extraction method.
Shows symbols representing the menus.
Shows current settings, submenus or actions depending on the current operation.
Shows functions that can be performed according to the current operation.
The extraction stops.
The valves close.
The heaters switch off.
The coolant keep running 5 min.
Used for navigating the user interface.
Pressing the control performs the assigned function on the function bar.
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
7.2
Operating the interface pro | 7
Function bar
The function bar shows available functions according to the current operation.
The functions on the function bar are executed by tapping the relevant function buttons.
Symbol
O K
Description
[ Back ]
[ Confirm ]
Meaning
The display reverts to the previous view.
Confirms an entry.
Deletes the selected method or step.
DELETE
[ Delete ]
S AV E [ Save ]
M E N U [ Menu ]
Saves the setting.
S TA R T [ Start ]
Allows selection of a menu from the menu bar using the navigation control.
Starts the extraction process.
STOP [ Stop ]
QR Code
[ QR Code ]
Aborts extraction on all extraction positions.
The display shows a QR Code which contains the data of all extraction parameters.
The lift moves up.
U P [ Up ]
D O W N [ Down ] The lift moves down.
E S C [ Escape ]
C O P Y [ Copy ]
Cancels the actual entry.
Copies the selected method.
7.3
Menu bar
The menus are represented by symbols on the menu bar. Navigation through the menus is by input controls.
The following menus are available:
Menu symbol Meaning
[ Home ] menu
Submenu/Actions
Process control parameters
Operation Manual FatExtractor E-500
[ Method ] menu
[ Configuration ] menu
Editing and saving extraction methods
Change settings
Service menu
System information
49/84
7 | Operating the interface pro
Menu symbol
7.3.1
Home menu
1
Meaning
[ Solvent ] menu
2 3
BÜCHI Labortechnik AG
Submenu/Actions
Solvent library
Editing and saving solvents
50/84
7
8
9
4
5
6
Fig. 15: Home menu
1
3
Time remaining
Status
5 Extraction position
7 Rinse time and heating level
9 Control buttons
2 Method
4 Extraction cycles
(Soxhlet Extraction only)
6
8
Extraction time and heating level
Drying time and heating level
On the home menu the following control buttons are available:
Button Meaning
Skips the actual step at the selected position.
Skips the actual step for all positions.
Aborts extraction on the selected extraction position.
Editing the running method.
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
7.3.2
7.3.3
Operating the interface pro | 7
Method menu
The [ Method ]
menu allows saving methods. See Chapter 8 "Editing a method", page 54
Configuration menu
On the [ Configuration ] menu, you can enter a variety of settings and retrieve information.
Submenu Settings
The submenu [ Settings ] contains system settings for the instrument.
Action
[ Language ]
[ Date ]
[ Time ]
[ Beeper ]
[ Display brightness ]
[ Configuration ]
Option
Choice of display language on the interface
Date input
Time input
Off/On
Enter setting
Choice of extraction method
Explanation
The following languages are available:
English/German/French/Italian/
Spanish/Japanese/Chinese
Enter in sequence: Day, month, year. Apply the settings by pressing
[ OK ] .
Enter in sequence: Minutes, hours.
Apply the settings by pressing
[ OK ] .
Setting for audible signal in response to input controls.
Display brightness level in %: 0 -
100
The following configurations are available:
SOX (Soxhlet Extraction) / HE (Hot
Extraction) / ECE (Economic
Continuous Extraction)
Maximum cycle time in minutes: 0 -
240
[ Max. Cycle time ]
(Soxhlet Extraction only)
[ Mobile connection
QR code ]
[ Mobile connection password ]
[ Network ]
[ Delete app connection ]
[ Demo Mode ]
Enter value
View
View
Enter value
The control unit shows a QR code for reading by the Extraction Reports app.
The control panel shows a password for entry in the Extraction Reports app.
The following parameters can be edited:
Device name/DHCP/MAC address/
System IP address/Subnet mask/
Gateway
Confirmation question Resets external connections to the instrument.
On/Off Simulate an extraction.
Operation Manual FatExtractor E-500 51/84
7 | Operating the interface pro BÜCHI Labortechnik AG
Submenu Service
WARNING
Incorrect use of the actor settings.
Incorrect use of the actors settings can cause instrument damages and injuries.
u Carry out the actor setting as described.
52/84
Action
[ Sensors ]
[ Actors ]
Option
View
Change settings
Explanation
All sensor signals can be observed:
Flow senors
Level sensors
Analyte protection (option)
Protection shield
Valves
Lift current
Line voltage
Line frequency
The following settings can be changed:
Valves:
Soxhlet valves See Chapter 13.2
"Draining solvent from Extraction glass chamber Soxhlet", page 76
"Draining solvent from the instrument", page 74
Cooling water valve On/Off
Heater On/Off
Pump On/Off
Lift current
Beeper On/Off
Chiller On/Off (option)
Submenu System Information
The submenu [ System information ] contains details of the connected devices and information on network connection diagnosis.
Action
[ Extraction unit ]
Option
View
Explanation
The following information on the
Extraction Unit is available:
Serial Number
Firmware version
Operating hours
Temperature electronics
Code
Line voltage
Line frequency
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
7.3.4
Operating the interface pro | 7
Action
[ Interface ]
Option
View
Explanation
The following information on the interface is available:
Serial Number
Firmware version
Operating Hours
Temperature electronic
Code
Operation voltage 30 V
Operation voltage 5V
Solvent menu
The solvent menu contains the available solvents for the extraction methods.
The solvent menu allows editing individual solvents. See Chapter 10 "Editing a solvent", page 69
7.4
Status bar
Shows the actually used solvent and configuration.
Operation Manual FatExtractor E-500 53/84
8 | Editing a method
8
BÜCHI Labortechnik AG
Editing a method
This chapter describes how to edit a method.
NOTE
Depending on the extraction method you edit the display shows the available settings for this extraction method.
8.1
8.2
8.3
Creating a new method
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Select the submenu [ Define method ] with the navigation control.
ð The display highlights the selection in green.
u Tap the function [ OK ] on the function bar.
ð The new method is created.
ð The display shows the view Method .
u Tap the function [ Save ] on the function bar.
ð The method is saved.
Changing the name of a method
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u
Tap the function [ Edit ] on the function bar.
Select the action [ Method name ] with the navigation control.
ð The display shows a dialog with an alphanumeric input box.
u u
Enter a name for the method.
Tap the function [ Save ] on the function bar.
ð The new name is saved.
ð The dialog box closes.
Setting the solvent of a method
The following four solvents are selectable:
Chloroform
Diethyl ether
Hexane
Petroleum ether
The control sets the heating level for the selected solvent automatically.
Navigation path
➔
54/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
8.4
Editing a method | 8 u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u
Tap the function [ Edit ] on the function bar.
Select the action [ Solvent ] with the navigation control.
ð The display shows a dialog with the selectable solvents.
u u u
Select the solvent you wish to use with the navigation control.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The solvent is saved.
ð The dialog box closes.
Setting the analyt protection
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u
Tap the function [ Edit ] on the function bar.
Tap the action [ Enable analyt protection ] .
8.5
Setting the extraction time of a method
The extraction time and the number of extraction cycles are interdependent.
If the number of extraction cycles is set to zero, the extraction will finish after the selected time.
If the extraction time is set to zero, the extraction will finish after the number of extraction cycles is reached.
If the extraction time and a number of extraction cycles is set, the extraction will finish once both settings are fulfilled.
The duration of the extraction step is selectable from 0 to 99 h.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Extraction time ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Operation Manual FatExtractor E-500 55/84
8 | Editing a method
8.6
8.7
8.8
BÜCHI Labortechnik AG u Tap the function [ Save ] on the function bar.
ð The value is saved.
ð The dialog box closes.
Setting the extraction cycles of a method (Soxhlet Extraction only)
The extraction time and the number of extraction cycles are interdependent.
If the number of extraction cycles is set to zero, the extraction will finish after the selected time.
If the extraction time is set to zero, the extraction will finish after the number of extraction cycles is reached.
If the extraction time and a number of extraction cycles is set, the extraction will finish once both settings are fulfilled.
The number of extraction cycles is selectable from 0 to 5940.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Extraction cycle ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The number of extraction cycles is saved.
Setting the extraction heating level of a method
The default setting for the heating level setting is depending on the selected solvent.
Change the heating level to compensate ambient conditions.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Extraction heating level ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The heating level is saved.
Setting the Soxhlet valve open time of a method (Soxhlet only)
The Soxhlet valve open time depends on the position of the level sensor and the sample structure.
Set the Soxhlet valve opening time so that the Extraction glass chamber Soxhlet is completely emptied.
56/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Editing a method | 8
The following open times are available:
Open time
[ short ] The level sensor is at the bottom of the Extraction glass chamber
Soxhlet .
The sample structure is easily releasing the solvent.
Hydrolysed sample in the glass sample tubes are extracted.
[ mid ]
[ long ]
The level sensor is at the middle of the Extraction glass chamber
Soxhlet .
The sample structure is easily releasing the solvent.
The level sensor is at the top of the Extraction glass chamber Soxhlet .
The sample structure is slowly releasing the solvent.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u
Tap the function [ Edit ] on the function bar.
Select the action [ Sox valve open time ] with the navigation control.
ð The display shows a dialog with the selectable values.
u u
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The Soxhlet valve open time is saved.
8.9
Setting the rinse time of a method (Soxhlet and Hot Extraction only)
The duration of the rinse time is selectable from 0 to 5940 min.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Rinse time ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The time is saved.
8.10
Setting the rinse heating level of a method (Soxhlet and Hot
Extraction only)
The default setting for the rinse heating level is depending on the selected solvent.
Change the rinse heating level to compensate fluctuations in ambient conditions.
Operation Manual FatExtractor E-500 57/84
8 | Editing a method BÜCHI Labortechnik AG
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Rinse heating level ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The heating level is saved.
8.11
Setting the number of drains for a method (Hot Extraction only)
During each drain interval, a defined amount of solvent is drained into the solvent tank.
The level in the beaker is lowered to finally rinse the sample without touching the extract.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u
Tap the function [ Edit ] on the function bar.
Select the action [ Drain interval ] with the navigation control.
ð The display shows a dialog with the selectable values.
u u u
Select the value you wish to use with the navigation control.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The value is saved.
ð The dialog box closes.
8.12
Setting Smart-Drying (Soxhlet and Economic Continuous
Extraction only)
Smart-Drying automatically sets the parameters for the drying step.
The following parameters are set automatically:
Rinse heating level
Drying time
Drying heating level
Navigation path
➔ u Navigate to the [ Method ] menu via the navigation path.
58/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Editing a method | 8 u Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u
Tap the function [ Edit ] on the function bar.
Tap the action [ Smart-Drying ] .
ð The function is enabled.
8.13
Setting the drying time of a method
The duration of the drying time of a method depends on the solvent volume.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Drying time ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The time is saved.
8.14
Setting the drying heating level of a method
The default setting for the drying heating level is depending on the selected solvent.
Change the drying heating level to compensate fluctuations in ambient conditions.
Navigation path
➔ u u
Navigate to the [ Method ] menu via the navigation path.
Tap the name of the method that you wish to edit.
ð The display shows the view Method .
u u u u u
Tap the function [ Edit ] on the function bar.
Select the action [ Drying heating level ] with the navigation control.
Turn the navigation control until the required value.
Tap the function [ OK ] on the function bar.
Tap the function [ Save ] on the function bar.
ð The heating level is saved.
Operation Manual FatExtractor E-500 59/84
9 | Performing an extraction
9 Performing an extraction
9.1
BÜCHI Labortechnik AG
Adding solvent through the condenser
CAUTION
Not correctly installed condenser
A not correctly installed condenser can cause skin burns or a building fire u u u
A beaker is placed under the position.
An extraction glass chamber is installed at the position (Soxhlet and Economic
Continuous Extraction only).
The lift is in the lower position.
NOTE
Boiling retardation
4 Make sure that the solvent is at ambient temperature.
u Put a funnel on the condenser.
u Add solvent with a beaker.
60/84
9.2
Putting the glass sample tube with frit in the extraction glass chamber
NOTE
Instead of glass sample tube with frit it is possible to insert a thimble.
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG u Put the holder for the glass sample tube with frit into the extraction chamber.
Performing an extraction | 9 u Put the glass sample tube with frit into the holder in the extraction chamber.
9.3
9.3.1
Performing a Soxhlet Extraction
Adjusting the optical sensor on the extraction glass chamber Soxhlet u Adjusting the optical sensor on the extraction glass chamber Soxhlet.
White line to the center of the upper sand layer.
Sand
Celite®
+ sample
Sand
9.3.2
Preparing the instrument for a Soxhlet Extraction
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation", page 26
R Make sure that the configuration Soxhlet Extraction is selected.
R Make sure that there are no extraction beakers in the instrument.
R Make sure that no defective sealings or glass parts are used.
R Make sure that the tank bottle is empty.
Operation Manual FatExtractor E-500 61/84
9 | Performing an extraction
9.3.3
9.3.4
BÜCHI Labortechnik AG u Switch the On/Off master switch to On.
ð The instrument is starting up.
Selecting a method
Navigation path
➔ u u
Navigate to the [ Home ] menu via the navigation path.
Tap the method.
ð The display shows a dialog with the selectable configurations.
u u
Select the method you wish to use.
Tap the function [ OK ] on the function bar.
ð The method is selected.
Selecting the extraction positions
The extraction positions can show the following status:
Status Meaning
The extraction position is activated.
The extraction position will be heated.
The extraction position is not activated.
The extraction position will not be heated.
The extraction for this extraction position is aborted.
9.3.5
Navigation path
➔ u u u
Navigate to the [ Home ] menu via the navigation path.
Select the extraction position with the navigation control.
Tap the function [ OK ] on the function bar.
ð The status of the extraction position is changed.
Starting a Soxhlet Extraction
NOTE
Boiling retardation
4 Make sure that the solvent is at ambient temperature.
Navigation path
➔
Precondition:
R The instrument has been prepared.
R A method is selected.
R The sample is prepared (in the glass sample tube or in the paper thimbles).
u u u
Switch on the connected chiller.
Pull the chamber rack out of the instrument.
Place the prepared glass sample tube or paper thimble with the corresponding holder in the extraction chamber.
62/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
9.3.6
Performing an extraction | 9 u u u u u u u
Push the chamber rack back in the instrument.
Fill the beaker with organic solvent.
Put the beaker on the corresponding position on the heating plate.
Close the protection shield.
Navigate to the [ Home ] menu via the navigation path.
Tap the function [ Down ] on the function bar.
ð The lift moves down.
u u u
Wait until the lift is in the lower position.
Make sure that all components of the glass setup are aligned.
Tap the function [ Start ] on the function bar.
ð The display shows the [ Home ] menu with a black background.
ð The instrument is starting the selected method.
u Make sure that there are no leaks.
Ending a Soxhlet Extraction
CAUTION
Risk of hot parts and glassware.
Minor or moderate burns can result.
u u
Do not touch hot parts or surfaces.
Use pliers to remove the glassware.
9.3.7
9.4
9.4.1
Navigation path
➔
Precondition:
R The [ Start ] menu is showing the status Finished .
u Tap the function [ Up ] on the function bar.
ð The lift moves up.
u u u
Open the protection shield.
Remove the beaker for further processing.
Empty the solvent tank. See Chapter 12.2 "Removing the tank bottle", page 73
Shutting down the instrument
Navigation path
➔
Precondition:
R The extraction process has ended.
u Switch the On/Off master switch to Off.
Performing a Hot Extraction
Preparing the instrument for a Hot Extraction
Precondition:
Operation Manual FatExtractor E-500 63/84
9 | Performing an extraction
9.4.2
9.4.3
BÜCHI Labortechnik AG
R
All commissioning operations have been completed. See Chapter 5 "Installation", page 26
R Make sure that the configuration Hot Extraction is selected.
R Make sure that there are no extraction beakers in the instrument.
R Make sure that no defective sealings or glass parts are used.
R Make sure that the tank bottle is empty.
u Switch the On/Off master switch to On.
ð The instrument is starting up.
Selecting a method
Navigation path
➔ u u
Navigate to the [ Home ] menu via the navigation path.
Tap the method.
ð The display shows a dialog with the selectable configurations.
u u
Select the method you wish to use.
Tap the function [ OK ] on the function bar.
ð The method is selected.
Selecting the extraction positions
The extraction positions can show the following status:
Status Meaning
The extraction position is activated.
The extraction position will be heated.
The extraction position is not activated.
The extraction position will not be heated.
The extraction for this extraction position is aborted.
9.4.4
Navigation path
➔ u u u
Navigate to the [ Home ] menu via the navigation path.
Select the extraction position with the navigation control.
Tap the function [ OK ] on the function bar.
ð The status of the extraction position is changed.
Starting a Hot Extraction
Navigation path
➔
Precondition:
R The instrument has been prepared.
R A method is selected.
R The sample is prepared (in the glass sample tube or in the paper thimbles).
u u
Switch on the connected chiller.
Place the prepared glass sample tube or paper thimble with the corresponding holder in the beaker.
64/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
9.4.5
Performing an extraction | 9 u u u u
Put the beaker on the corresponding position on the heating plate.
Close the protection shield.
Navigate to the [ Home ] menu via the navigation path.
Tap the function [ Down ] on the function bar.
ð The lift moves down.
u u u u
Wait until the lift is in the lower position.
Make sure that all components of the glass setup are aligned.
Fill the solvent via the condenser. See Chapter 9.1 "Adding solvent through the condenser", page 60
Tap the function [ Start ] on the function bar.
ð The display shows the [ Home ] menu with a black background.
ð The instrument is starting the selected method.
u Make sure that there are no leaks.
Ending a Hot Extraction
CAUTION
Risk of hot parts and glassware.
Minor or moderate burns can result.
u u
Do not touch hot parts or surfaces.
Use pliers to remove the glassware.
9.4.6
9.5
9.5.1
Navigation path
➔
Precondition:
R The [ Start ] menu is showing the status Finished .
u Tap the function [ Up ] on the function bar.
ð The lift moves up.
u u u
Open the protection shield.
Remove the beaker for further processing.
Empty the solvent tank. See Chapter 12.2 "Removing the tank bottle", page 73
Shutting down the instrument
Navigation path
➔
Precondition:
R The extraction process has ended.
u Switch the On/Off master switch to Off.
Performing an Economic Continuous Extraction
Preparing the instrument for an Economic Continuous Extraction
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation", page 26
Operation Manual FatExtractor E-500 65/84
9 | Performing an extraction
9.5.2
9.5.3
BÜCHI Labortechnik AG
R Make sure that the configuration Economic Continuous Extraction is selected.
R Make sure that there are no extraction beakers in the instrument.
R Make sure that the tank bottle is empty.
R Make sure that no defective sealings or glass parts are used.
u Switch the On/Off master switch to On.
ð The instrument is starting up.
Selecting a method
Navigation path
➔ u u
Navigate to the [ Home ] menu via the navigation path.
Tap the method.
ð The display shows a dialog with the selectable configurations.
u u
Select the method you wish to use.
Tap the function [ OK ] on the function bar.
ð The method is selected.
Selecting the extraction positions
The extraction positions can show the following status:
Status Meaning
The extraction position is activated.
The extraction position will be heated.
The extraction position is not activated.
The extraction position will not be heated.
The extraction for this extraction position is aborted.
9.5.4
Navigation path
➔ u u u
Navigate to the [ Home ] menu via the navigation path.
Select the extraction position with the navigation control.
Tap the function [ OK ] on the function bar.
ð The status of the extraction position is changed.
Starting an Economic Continuous Extraction
NOTE
Boiling retardation
4 Make sure that the solvent is at ambient temperature.
Navigation path
➔
Precondition:
R The instrument has been prepared.
R A method is selected.
R The sample is prepared (in the glass sample tube or in the paper thimbles).
66/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
9.5.5
Performing an extraction | 9 u u u u u u u
Switch on the connected chiller.
Place the prepared glass sample tube or paper thimble with the corresponding holder in the extraction chamber.
Fill the beaker with organic solvent.
Put the beaker on the corresponding position on the heating plate.
Close the protection shield.
Navigate to the [ Home ] menu via the navigation path.
Tap the function [ Down ] on the function bar.
ð The lift moves down.
u u u
Wait until the lift is in the lower position.
Make sure that all components of the glass setup are aligned.
Tap the function [ Start ] on the function bar.
ð The display shows the [ Home ] menu with a black background.
ð The instrument is starting the selected method.
u Make sure that there are no leaks.
Ending an Economic Continuous Extraction
CAUTION
Risk of hot parts and glassware.
Minor or moderate burns can result.
u u
Do not touch hot parts or surfaces.
Use pliers to remove the glassware.
9.5.6
9.6
9.6.1
Navigation path
➔
Precondition:
R The [ Home ] menu is showing the status Finished .
u Tap the function [ Up ] on the function bar.
ð The lift moves up.
u u u
Open the protection shield.
Remove the beaker for further processing.
Empty the solvent tank. See Chapter 12.2 "Removing the tank bottle", page 73
Shutting down the instrument
Navigation path
➔
Precondition:
R The extraction process has ended.
u Switch the On/Off master switch to Off.
Exporting extraction parameters
Exporting extraction parameters without cloud connection
Precondition:
Operation Manual FatExtractor E-500 67/84
9 | Performing an extraction
9.6.2
BÜCHI Labortechnik AG
R The App is installed on a mobile device.
R An extraction is finished.
u Tap the button [ QR Code ] on the function bar.
ð The display shows a QR code.
u Scan the QR code on the display with the camera function of the App.
Exporting extraction parameters with cloud connection
Precondition:
R The App is installed on a mobile device.
R An extraction is finished.
u The extraction parameters are automatically transferred to the App.
68/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
10 Editing a solvent
Editing a solvent | 10
10.1
Creating a new solvent
Navigation path
➔ u u
Navigate to the [ Solvent ] menu via the navigation path.
Tap the submenu [ Create own ] .
ð The new solvent is created.
10.2
Changing the name of a solvent (own created solvents only)
Navigation path
➔ u u
Navigate to the [ Solvent ] menu via the navigation path.
Tap the name of the solvent you wish to edit.
ð The display highlights the selection in green.
u Tap the function [ OK ] on the function bar.
ð The display shows the view Solvent .
u Tap the action [ Name ] .
ð The display shows a dialog with an alphanumeric input box.
u u
Enter a name for the solvent.
Tap the function [ Save ] on the function bar.
ð The new name is saved.
ð The dialog box closes.
Operation Manual FatExtractor E-500 69/84
11 | Mobile connection
11 Mobile connection
BÜCHI Labortechnik AG
11.1
Requirements for local network settings
u The following port has to be enabled in the firewall settings on the internet gateway:
TCP (HTTPS) traffics through remote port 443 u In order to use the BUCHI Cloud a DNS server must be configured on the instrument.
NOTE
If there is no DNS server available enter the IP address for the BUCHI Cloud connection manually.
NOTE
If there is no DHCP server available enter the IP address, gateway subnet mask and
DNS server manually.
11.2
Preparing the instrument for app using
NOTICE! Do not unplug the LAN cable while the device is connected to the
BUCHI Cloud Services u Connect the instrument with the LAN network.
Navigation path
➔ ➔ [ Settings ] ➔ [ Network ] u u
Navigate to the action [ Network ] .
Enable the function [ DHCP ] .
ð The instrument is prepared.
11.3
Connecting the instrument and app using a QR code
u Navigate to the action [ Mobile connection QR code ] .
ð The display shows a QR code.
u Scan the QR code with the camera function of the App.
70/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Mobile connection | 11
11.4
Connecting the instrument and app using a password
u Navigate to the action [ Mobile connection password ] .
ð The display shows a password.
u u
Enter the password in the app.
Tap [ OK ]
Operation Manual FatExtractor E-500 71/84
12 | Cleaning and servicing BÜCHI Labortechnik AG
12 Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this section.
Any servicing and repair work which involves opening up the casing may only be carried out by BUCHI service technicians.
u Use only genuine BUCHI consumables and spare parts in order to ensure correct operation of the device and preserve the warranty.
12.1
Regular maintenance work
CAUTION
Hot surfaces.
Skin burns from hot surfaces.
u Let the instrument cool down sufficiently before carrying out any maintenance work.
Component
Solvent tank
Glass parts
Sealings
Heating plate
Hoses and hose connections
Housing
Warning symbols
Action u Empty the solvent tank. See Chap-
ter 12.2 "Removing the tank bottle", page 73
Frequency
After every extraction u u
Check all glass parts for defects.
If defective, replace defect glass parts
Before every extraction
Daily NOTICE! Do not clean in dishwasher.
u u
Check all sealings for scratches and other defects.
If defective or damaged, replace the sealing.
u u u
Wipe down the heating plate with a damp cloth.
If heavily soiled, use ethanol or a mild detergent.
If cracked or broken, contact BUCHI customer service
Weekly
Weekly u u
Check the hoses and hose connections for defects (cracks, brittle areas).
If defective, replace defect hoses.
u u
Wipe down the housing with a damp cloth.
If heavily soiled, use ethanol or a mild detergent.
u u
Check that the warning symbols on the instrument are legible.
If they are dirty, clean them.
Weekly
Weekly
72/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG Cleaning and servicing | 12
Component Action
Protection shields NOTICE! Do not use Acetone u u u
Wipe down the protection shields with a damp cloth.
Make sure that they click into place. If not, replace the gliding elements. See
Chapter 13.3 "Replacing the gliding elements", page 76
If defective or damaged, replace them.
Reflector foil for analyte protection
Interface
NOTICE! Do not use any other detergent u u u u
Check the reflector foil for dirt.
If dirty, wipe down the reflector foil with a damp cloth.
Make sure that there are no extraction beakers in the instrument.
Switch the On/Off master switch to
On.
u Wipe down the display with a damp cloth.
Frequency
Weekly
Weekly
Monthly
12.2
Removing the tank bottle
u Switch the On/Off master switch to Off.
u Open the knurled nut at the ball joint clamp.
Operation Manual FatExtractor E-500 73/84
12 | Cleaning and servicing u Remove the ball joint clamp.
BÜCHI Labortechnik AG u Push the solvent tank up.
u Tilt the solvent tank at 20° and remove it.
74/84
12.3
Draining solvent from the instrument
Navigation path
➔ ➔ [ Service ] ➔ [ Actors ] ➔
[ Valves ] u Navigate to [ Valves ] via the navigation path.
ð The display shows the view Valves .
u u u u
Select the position of the tank valve you wish to open.
Open the tank valve.
Wait until the all solvent is in the tank bottle.
Close the tank valve.
Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
13 Help with faults
Help with faults | 13
13.1
Troubleshooting
Problem
Flange lock is not inserted completely
Instrument does not work
Possible cause
The condenser is not in the right position.
Instrument is not connected to the power supply.
Action u Move the condenser flange a little bit until the lock falls into the wholes.
u Make sure that the power supply is connected and switched on.
Method cannot be started
Method stops
Solvent tank is full
Protection shield is open
Protection shield lifted
Tank bottle full u u
Cooling water flow is too low u Check the cooling water connection and all hoses for bends and remove them, if necessary
Cooling water sensor is defective (may be the coolant is contaminated with algae, lime, etc.) u Contact BUCHI Costumer
Service.
u
Empty the solvent tank.
Close the protection shield.
Close the protection shield u Empty the tank bottle. See
Chapter 12.2 "Removing the tank bottle", page 73
Loss of solvent Seals wrong, deformed or damaged
Incorrect positioning of the beaker/ glassware and seals
Boiling retardation Incorrect positioning of the beaker u Exchange the seals.
u Contact the BUCHI customer service u Put the beaker directly onto the heating plate
Too long waiting time between preparation of the instrument and start u Start the instrument quickly after solvent was added
No boiling Incorrect positioning of the beaker
Beaker position is not active u Place the beaker directly onto the heating plate u Activate the beaker position
Heater power is set too low
Not enough solvent u u
Adapt the heating power to the ambient conditions
(temperature, level above sea).
Choose correct solvent u Add solvent. See Chap-
ter 9.1 "Adding solvent through the condenser", page 60
Operation Manual FatExtractor E-500 75/84
13 | Help with faults BÜCHI Labortechnik AG
Problem Possible cause
Elution of Celite® Too much organic solvent for extraction
Action u Stick to the standard application parameters
Sand/Celite® layers mixed up u Clean the extraction chambers
13.2
Draining solvent from Extraction glass chamber Soxhlet
Navigation path
➔ ➔ [ Service ] ➔ [ Actors ] ➔
[ Valves ]
Precondition:
R A Soxhlet Extraction was aborted
R A beaker is placed under the extraction glass chamber Soxhlet position.
u Navigate to [ Valves ] via the navigation path.
ð The display shows the view Valves .
u u u u
Select the position of the Soxhlet valve you wish to open.
Open the Soxhlet valve.
Wait until the extraction glass chamber is empty.
Close Soxhlet valve.
13.3
Replacing the gliding elements
Positions of the magnet
Replacing procedure u Attach the back part of the gliding elements to the protection shield.
76/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
NOTICE! Insert the magnet at the positions indicated.
u Insert the magnet.
u Attach the front part of the gliding elements to the protection shield.
Help with faults | 13
Operation Manual FatExtractor E-500 77/84
14 | Taking out of service and disposal
14 Taking out of service and disposal
BÜCHI Labortechnik AG
14.1
Taking out of service
u u u u
Remove all solvents and coolants.
Clean the instrument.
Switch off the instrument and disconnect it from the mains power supply.
Remove all tubing and communication cables from the device.
14.2
Disposal
The operator is responsible for proper disposal of the instrument.
u u
When disposing of equipment observe the local regulations and statutory requirements regarding waste disposal.
When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter 3.5 "Technical data", page 22
14.3
Returning the instrument
Before returning the instrument, contact the BÜCHI Labortechnik AG Service Department.
https://www.buchi.com/contact
78/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
15 Appendix
15.1
Schematics
15.1.1
Condenser circuit
Condenser
Condenser
Condenser
Condenser
Condenser
Condenser
Condenser solvent tank
CW IN
Condenser solvent tank
CW OUT
Fig. 16: Condenser circuit
Appendix | 15
Connection
CW IN
Connection
CW OUT
Water in
Water out
Flow meter
(inside instrument)
Connection
CW Tank OUT
Connection
CW Tank IN
Operation Manual FatExtractor E-500 79/84
15 | Appendix
15.1.2
Solvent collector schematic
Solvent collector
BÜCHI Labortechnik AG
Solvent collector
Solvent collector
Exhaust
Solvent collector
Solvent collector
Tank pump
Solvent collector
Condenser solvent tank
IN (solvent)
Connection
Tank OUT solvent
Condenser solvent tank
OUT
Fig. 17: Solvent circuit
Connection
Tank IN
15.2
Spare parts and accessories
Use only genuine BUCHI consumables and spare parts in order to ensure correct, safe and reliable operation of the system.
NOTE
Any modifications of spare parts or assemblies are only allowed with the prior written permission of BUCHI.
15.2.1
Spare parts
Extraction glass chamber Soxhlet
Order no.
11062496
Image
Extraction glass chamber Soxhlet LSV 11062497
80/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
Extraction glass chamber ECE
Set of beakers (2pcs)
Set of beakers HE (2pcs)
Condenser E-500 cpl.
Condenser tank bottle
Glass sample tubes with frit, 6 pcs.
Glass sample tubes LSV with frit, 6 pcs.
Tank adapter
Tank bottle 2 L
Order no.
11062499
Turning disk
Allows for turning the instrument for easier access.
11067985
Set of gliding elements including magnets, 10 pcs.
11067827
Soxhlet assembly cpl.
One part constiting of magnetic valve and level sensor for extraction glass chamber Soxhlet
11067065
11067497
11067814
11064590
11065983
11067474
11067475
11067063
11065966
Appendix | 15
Image
Operation Manual FatExtractor E-500 81/84
15 | Appendix
Cooling water valve. 24VAC
Valve opens cooling water feed during distillation.
Meant to be used with a vacuum controller/interface.
Order no.
031356
BÜCHI Labortechnik AG
Image
15.2.2
Conversion kits
The FatExtractor E-500 can be converted into another configuration with an easy change of glass assembly. Choose the needed conversion kit according to the matrix.
Conversion kit from Soxhlet Extraction and Economic Continuous Extraction to Hot Extraction incl.: Beaker rack, tubing set Hot Extraction , beaker carrier, beaker (6)
Conversion kit from Soxhlet Extraction to Economic Continuous Extraction incl.: tubing set ECE, extraction glass chamber ECE (6)
Conversion kit from Hot Extraction to Soxhlet Extraction incl.: chamber rack (2), protection shield (2), tubing, beaker (6), extraction glass chamber Soxhlet (6)
Conversion kit from Hot Extraction to Economic Continuous Extraction incl. chamber rack (2), protection shield (2), tubing set ECE, beaker (6), extraction glass chamber ECE (6)
Conversion kit from Economic Continuous Extraction to Soxhlet Extraction incl: protection shield (2), tubing SOX, soxhlet unit cpl. (6), extraction glass chamber Soxhlet (6)
Conversion kit HE to Soxhlet LSV
Includes 6 Extraction glass chamber Soxhlet LSV (11062497), 6 beaker (3 x
11067474), 6 Soxhlet assemblies cpl. (11067065) and a chamber rack
(11067077), set of draining tubes for Soxhlet (11067478)
Conversion kit ECE to Soxhlet LSV
Includes 6 Extraction glass chamber Soxhlet LSV (11062497), 6 Soxhlet assemblies cpl. (11067065) and a set of draining tubes for Soxhlet (11067478)
Order no.
11068487
11068488
11068489
11068491
11068492
11068490
11068493
15.2.3
Accessories
Holder for glass sample tubes, stainless steel
Order no.
11067219
Image
Holder for glass sample tubes, PTFE 11067220
Holder for extraction thimbles (diameter 25 - 43 mm)
11068443
82/84 Operation Manual FatExtractor E-500
BÜCHI Labortechnik AG
Extraction beaker carrier
Allows to carry 6 beakers (11067474)
Order no.
11067042
Extraction beaker HE carrier
Allows to carry 6 beakers HE (11067475)
11067493
Set condenser insulations E-500, 6 pcs.
The insulation of the condensers prevent condensing water and is recommended in high humidity environment
Set insulation cooling water hoses
The insulation of the water hoses prevent condensing water and is recommended in high humidity environment.
Support solvent supply
Allows to fix the tubes of solvent dispensers to the condensers for convenient solvent addition.
11069078
11069079
11068306
15.2.4
Consumables
Quartz sand 0.3 - 0.9 mm, 2.5 kg
Celite® 545, 1 kg
15.2.5
Holder for extraction thimbles
Set of holders for glass sample tubes with frit, PTFE, 3 pcs.
Set of holders for LSV glass sample tubes, PTFE, 3 pcs.
Holders for thimbles d25, PTFE, 3 pcs.
Holders for thimbles d33, PTFE, 3 pcs.
Holders for thimbles d43, PTFE, 3 pcs.
15.2.6
Extraction thimbles
Extraction thimbles 25 x 100 mm, 25 pcs.
Order no.
018105
Appendix | 15
Image
Order no.
037689
11068920
Order no.
11067485
11067486
11067488
11067490
11067491
Image
Extraction thimbles 33 x 94 mm, 25 pcs.
11058983
Extraction thimbles, Set. 25 pcs, 43 x 118 mm, cellulose
For Soxhlet extraction unit.
018106
Operation Manual FatExtractor E-500 83/84
Quality in your hands
BUCHI Affiliates:
Europe
Switzerland/Austria
BÜCHI Labortechnik AG
CH – 9230 Flawil
T +41 71 394 63 63
F +41 71 394 64 64 [email protected]
www.buchi.com
Italy
BUCHI Italia s.r.l.
IT – 20010 Cornaredo (MI)
T +39 02 824 50 11
F +39 02 575 12 855 [email protected]
www.buchi.com/it-it
Benelux
BÜCHI Labortechnik GmbH
Branch Office Benelux
NL – 3342 GT Hendrik-Ido-Ambacht
T +31 78 684 94 29
F +31 78 684 94 30 [email protected]
www.buchi.com/bx-en
Russia
BUCHI Russia/CIS
Russia 127287 Moscow
T +7 495 36 36 495 [email protected]
www.buchi.com/ru-ru
France
BUCHI Sarl
FR – 94656 Rungis Cedex
T +33 1 56 70 62 50
F +33 1 46 86 00 31 [email protected]
www.buchi.com/fr-fr
United Kingdom
BUCHI UK Ltd.
GB – Oldham OL9 9QL
T +44 161 633 1000
F +44 161 633 1007 [email protected]
www.buchi.com/gb-en
Germany
BÜCHI Labortechnik GmbH
DE – 45127 Essen
T +800 414 0 414 0 (Toll Free)
T +49 201 747 49 0
F +49 201 747 49 20 [email protected]
www.buchi.com/de-de
Germany
BÜCHI NIR-Online
DE – 69190 Walldorf
T +49 6227 73 26 60
F +49 6227 73 26 70 [email protected]
www.nir-online.de
America
Brazil
BUCHI Brasil Ltda.
BR – Valinhos SP 13271-200
T +55 19 3849 1201
F +55 19 3849 2907 [email protected]
www.buchi.com/br-pt
USA/Canada
BUCHI Corporation
US – New Castle, DE 19720
T +1 877 692 8244 (Toll Free)
T +1 302 652 3000
F +1 302 652 8777 [email protected]
www.buchi.com/us-en
Asia
China
BUCHI China
CN – 200233 Shanghai
T +86 21 6280 3366
F +86 21 5230 8821 [email protected]
www.buchi.com/cn-zh
India
BUCHI India Private Ltd.
IN – Mumbai 400 055
T +91 22 667 75400
F +91 22 667 18986 [email protected] www.buchi.com/in-en
Indonesia
PT. BUCHI Indonesia
ID – Tangerang 15321
T +62 21 537 62 16
F +62 21 537 62 17 [email protected]
www.buchi.com/id-in
Japan
Nihon BUCHI K.K.
JP – Tokyo 110-0008
T +81 3 3821 4777
F +81 3 3821 4555 [email protected]
www.buchi.com/jp-ja
Korea
BUCHI Korea Inc.
KR – Seoul 153-782
T +82 2 6718 7500
F +82 2 6718 7599 [email protected]
www.buchi.com/kr-ko
Malaysia
BUCHI Malaysia Sdn. Bhd.
MY – 47301 Petaling Jaya,
Selangor
T +60 3 7832 0310
F +60 3 7832 0309 [email protected]
www.buchi.com/my-en
BUCHI Support Centers:
South East Asia
BUCHI (Thailand) Ltd.
TH-Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54 [email protected]
www.buchi.com/th-th
Middle East
BÜCHI Labortechnik AG
UAE – Dubai
T +971 4 313 2860
F +971 4 313 2861 [email protected]
www.buchi.com
Singapore
BUCHI Singapore Pte. Ltd.
SG – Singapore 609919
T +65 6565 1175
F +65 6566 7047 [email protected]
www.buchi.com/sg-en
Latin America
BUCHI Latinoamérica
S. de R.L. de C.V.
MX – Mexico City
T +52 55 9001 5386 [email protected]
www.buchi.com/es-es
Thailand
BUCHI (Thailand) Ltd.
TH – Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54 [email protected]
www.buchi.com/th-th
We are represented by more than 100 distribution partners worldwide.
Find your local representative at: www.buchi.com