Karl Kaps SOM 62 COLD LIGHT MOTORIZED Stereo microscope User's Manual
SOM 62 COLD LIGHT MOTORIZED is a universal instrument for operative and diagnostic medical use. It provides optimal illumination and magnification of the attendance area with the possibility of manual variation of magnification, focus and image field.
Advertisement
Advertisement
User's Manual SOM 62 COLD LIGHT MOTORIZED
SOM 62 COLD LIGHT
MOTORIZED
Stereo - Operation - and Diagnosis Microscope
—
15 V / 150 W
nn ds
Karl Kaps GmbH & Co. KG
73 06441 / 80704-0
Fax 06441 / 85985
MARKS LICENCE Postfach 1225, D-35608 Asslar
Licence No.: Europastrafse, D-35614 Asslar
093623, page 1 - VDE
www. kaps-optik.de
109666 F - VDE EMV — e-mail: serviceQkaps-optik.de
This microscope is manufactured according to the EU direction 93 / 42 / EWG
Products of medicine (MPG)
KAPS 01.10.1997 Page 1
VOGL D сана DUMM DZ LULL LIGHT MOTORIZED
OF CONTANTS
LIST
1.0 Security indications....... seu vec eu 40801 0 206001 a 50000 0 roce eed
2.0 Condition of material at time of SUPPIY..20 000000 00 000 sun ua sas à 3
3.0 Mounting of roller stand, H-form and column (Fig.1).............4
4.0 Mounting of appliance (19.2). ............... * »recicocoorenmaonoooeoD
4.1 Special indication.....oeceeeeorceosncosoococcarereavocoecoarooa.B
4.2 Additional load.......eocoeocoeeroocooooocooo 1.000 0002000000 ...6
4.3 Measures, weight (Fig.3).........ee.. +. 0000000 00 410 т все весе еб
5.0 Close electrical connections (Fig.4) (Fig.5).....ce.co0ceomonavaoa 78
>.) Adjustment. of: mains-voltage.............eeee.eecocacoooo. 7,8
5.2 Electrical CONNECtions........eeecrocosoccacecanmercanmeoe»o 778
5.3 Changing of fuse (after removal of the defect)......... ..7,8
6.0 Control elements. (Fig.6) (Fig. 7)..... 00661 000000 8 + se 0s+ +1 29,10
7.0 Operating instructions (Fig.8) (Fig.9)......... a 0000000 a 06 11,12
7.1 On-/off switch (Fig.8).........e.eoéeoosoeroeorae, e sscoorecacall
7.2 Change-over switch lamp I/lamp II (Fig.8)........eeeoeo sec. ll
7.3 Changing the motorized speeds (Fig.8)........ .e............l11
N a ae a a aa a ea a ea aan ..............Jl11
7.5 Brightness control (Fig.9)........ cere nooo. .e..........l12
8.0 Instructions for using the microscope......... ..............l2-14 -
8.1 Adjusting the eye distance (Fig.10) (Fig.11)...........12,13
8.2 Changing the magnification (Fig.12)..... .............. 13,14
8.3 How to focuse the object (Fig.12)....... ...............l13,14
8.4 Eyepieces with reticle (help for fine focussing
and when using television or photo systems) (Fig.12)......14
9.0 Breakdown of the cold light ‘illumination (Fig.13)..... sererooao 15
9.1 Breakdown of the cold "light illumination during operation.15
9.2 Changing the xenon reflector lamp later on....... .e,n.....15
10.0 Changing the binocular tube (Fig. 14).........o. ................J16
11.0 Changing the objectives/eyepieces (Fig.15)... cee nn ...........16
12.0 Schedule for microscope magnification and
object field measures.............. es a sance 00 su 000 В ‚ + 17
13.0 Objective quick changer (Fig.16).................. ...eooreccaoo. 18
14.0 Compensation of:weight when using accessories (Fig.17).........19
15.0 Mounting of beam splitter (Fig.17)............ DE ‚19
16.0 Observation tube monocular, observation tube binocular
and assistance tube 90° (Fig.18)...... +00 eu 0001000000 a 00 ....20
17.0 Photo tube with 35mm camera,
TV-tube with video camera (CCD) (Fig.193)......... oo. 0043000 € 21
18.0 Photographing without flash....eií=coocrcocosaoooooa e eccoocacaronoao 22
19.0 How to get a sharp image on the TV screen (Fig.20)..........22,23
20.0 Assistance and observation setting 180° (Fig.21)...............24
21.0 Mounting and operation of couplings (Fig.22)......... .esecc..25,26
22.0 Microscope carrier (2 gears) :(Fig.23)... ee... . 0. 0.0.0.0.0. 0.00. 20. 9 0 0. 0.00 26
23.0 Handles and microscope ‘guides’ (Fig:24).. tess ses esses acne ....27
24.0 Mounting and using the slit lamp (Fig.25)..... + .coscordcciocaaoo.20
25.0. Mounting ‘and: using ‘the XY-coupling: (Fig.26)...:.................29'
26.0 Accessories (survey with order no.)...... 6 a sa nca au 00 ...........30
27.0 Desinfection and sterilization (Е19.27).................. .e......31
28.0 Cleaning and вегу1се.................... e acacoo 00 6 60000000 6 ..32
29.0 Technical data.......ñecooeseccdocovorededa La. 0000000 A 1.000 ne 008 32
29.1 Conformity declaration...........e.eeneo00e0oeo0adredo, cessed
KAPS 01.03.1996 page 2
User's Manual SOM 62 COLD LIGHT MOTORIZED
1.0
1.1
1.2
2.0
SECURITY INDICATIONS, INTENDED USE
This microscope is manufactured according to the security indications of electrical equipment and
complies with the indicated IEC Puplications (IEC 601 - 1 / EN 60601 -1 / VDE 0750 part 1). This
instrument left KAPS Company in perfect condition.
To guarantee the normal and safety use of the microscope it is necessary to read this user's manual
attentively.
This microscope is allowed to use in that way only as described in this manual.
For service and repair only authorized persons are allowed to operate.
Spare parts and accessories must be admitted by the manufacturer. If there are any questions, please
contact our representatives.
Please substitute burned out fuses by fuses of the same type only (rated voltage, rated current,
switch - off characteristics).
The microscope belongs to safety class |, use mains plug and mains socket both with protective earth
conductors only.
Be sure that inlets and outlets of ventilation system for cooling the housing are open (not covered).
The microscope is built for use in dry rooms only. Take care that no fluids can enter the microscope
components.
Microscope is protected against overheating. A defect cooling system (fan is defect, inlets or outlets of
cooling system are unintentionally covered) will be indicated by blinking of a red light emitting diode (3),
Fig. 8, page 11. The protecting system reduces brightness of illumination automatically in that way that
a surgeon can finish his activities and microscope will not be damaged. After the user's activities are
being finished, cases of disturbances must be eliminated immediately, if this is not possible please advice
service department.
Very important: For identification, service people must know the serial number of the microscope!
Please read it from the type labei (4), Fig. 8, page 11.
Intended use |
The SCM 62 COLD LIGHT MOTORIZED is a universal instrument for operative and diagnostic
medical use. If is useful for optimal illumination and magnification of the attendance area with the possibility
of the manual variation of magnification, focus and image field.
The use of the SOM 62 COLD LIGHT MOTORIZED is only allowed to authorized people and medical
applications as described in this instruction handbook. If this instrument is in intended use, there is no
contact between patient and SOM 62.
SECURITY INDICATIONS DURING THE APPLICATION
Make all works and adjustments at the column before operation, never change something over or
balance above operation field.
Never change adjustments of the microscope during the operation!
Carry out reconstructions and transportations only when swivel arm is blocked.
Lamp change: First test that the {amp is cold!
WARNING FOR EYE SURGERY!
Protect patient through short exposure time and low brightness adjustments, because too intensive
light can injure retina.
CONDITION OF MATERIAL AT TIME OF SUPPLY
The appliance is delivered in 4 assembied groups:
Mobile supporting base
Column
Swivel arm and suspension arm with electrical power supply
Microscope head
Fastening aids are enclosed in delivery.
KAPS 01.10.1997 Page 3
- A A ak a
user's Manual SOM 62 COLD LIGHT MOTORIZED
3.0 MOUNTING OF ROLLER STAND, H-FORM AND COLUMN (Fig.1)
e Locking of roller stand:
Press both round knobs (1) down on roller stand (2) by foot.
Loosening is done by both pedals (3).
- e Insert column (4) into roller stand. Before fastening straighten
handles (5) to roller stand. -
e Screw in screw (6). Fasten screw with enclosed hexagon socket
wrench until column clamps.
E
(Fig.1)
KAPS 01.03.1996 page 4
USEL 5 Manual SUP DZ GULD LIGHT MOTORIZED
4.0 MOUNTING OF APPLIANCE (Fig.2)
® Unscrew safety cap (3) from vertical guide cylinder (1) of
column (2).
Take care that clamp, cross handle (4) is open.
Insert swivel arm. (5) (on-/off switch (6) upwards). DE
Screw on safety cap (3) again. e
Unscrew safety cap (7) from guide cylinder (8) of microscope
carrier (9). |
Take care that clamp, cross handle (10) is open. =
e Press locking‘pin:(11) and‘ insert. guide. cylinder of microscope. - =
carrier to its stop. Release locking pin; it locks and thus -
prevents the drop out of microscope. head (12).
® Screw on.safety cap (7) by hand on overlying threaded coupling
of guide cylinder.
e Control position of counterweights (13), if necessary, put in
counterweights to their stop.
e Insert fibre optic light guide (14) with screwed on adapter (15)
into cold light adapter (16) at the microscope head to its stop.
e Control position of the other side of fibre optic light guide
with mount (17), if necessary, put in fibre optic light guide in
lamp insert (18) again to its stop.
a | rotary motion appr. 400°
(4) (3) © with stop
©
18
+
NUIT”
(NY Po Y / >
UDI
esse (12) a <
(Fig.2)
KAPS 01.03.1996 | page 5
ee AA E
User's Manual SOM 62 COLD LIGHT MOTORIZED
4.1 Special indication:
e The joint which connects swivel arm and suspension arm is fit
out with a stop to prevent
the inner assembled cable.
e The total rotation of suspension arm to the swivel arm is more
than whole circuit (appr. 400*).
e By :reaching :the.stop do not. rotate further with: power: to prevent.
a tear off.
4.2 ADDITIONAL LOAD.
e The load capacity and tilting stability 1 is balanced.with
. Components. of.our product. assortiment.
e Please don't attach additional load.
_4.3 MEASURES, WEIGHT (Fig.3)
Roller stand, H-form
Height of appliance
(suspension arm horizontal)
Total weight
length 700 mm
width 640 mm
1600 mm
appr. 100 kg
1240
- 605 425mm]
а N
т N,
i A 0
; 7
- / +
VÍ
170
©
3
(Fig.3)
KAPS 01.03.1996
page 6
5.0 CLOSE ELECTRICAL CONNECTIONS (Fig.4) (Fig.5)
10
Yo ©
(e
0
(Ета. 4)
KAPS 01.03.1996 page 7
User's Manual SOM 62 COLD LIGHT MOTORIZED
5.1
5.3
(2)
ADJUSTMENT OF MAINS VOLTAGE
The appliance is adjusted for 230 V mains voltage at delivery and can be switched over to
115 V if necessary.
It is only allowed to switch on the appliance, if the mains voltage at place of use is
corresponding with voltage indication at voltage selector switch (1).
Service information:
If 115 V mains voltage is required, change voltage selector switch (1).
The mains input fuses T 1 A (2) are automatically correctly dimensioned by special
control.
ELECTRICAL CONNECTIONS
Insert plug (2) in mains socket.
It is allowed to connect mains plug of the appliance with mains socket both with protective
earth conductors only.
Connect plug (3) from fine focussing unit (4) with plug socket (5). Secure with knurled ring
(6). Attention: The nose of the plug must fit to the groove of plug socket.
Connect piug (7) from foot switch panel (8) with plug socket in the swivel arm (9).
Secure with screws.
CHANGING OF FUSE (after removal of the defect)
Fuses in fuse holder:
T1A (2 mains input fuses for lamp supply during mains voltage 230 V/ 115 V)
Changing of fuse: Draw out plug (2). Insert screw driver into slit of fuse holder (10), screw out
to the left. Remove cap and replace fuse placed in it. Insert fuse with cap again, screw in to
the right.
Attention: It is only allowed to change the fuses against fuses of the same type (rated
voltage, rated current, switch off characteristics).
(Fig. 13)
KAPS 01.10.1997 Page 8
6.0 CONTROL ELEMENTS (Fig.6) (Fig.7)
5 rollers to move the appliance (not showed)
2 brakes to lock the appliance (1)
2 pedals to loosen the brakes (2)
Rotating knob for adjustment of the swivel arm setting (3) _
Rotating knob for adjustment of rotary motion of the suspension -
arm (4)
Clamping lever to lock the vertical adjustment of the suspension
arm ( 5 ) луч
Safety knob:to prevent the microscope: head from. dropping out. .. =
while mounting (6) N
Rotating knob to prevent the microscope head from rotating (7)
Safety cap to secure the microscope head (8)
Safety cap to secure the swivel arm (9) :
Clamping lever to lock the inclination of microscope head (10)
Rotating knob for motorized adjustment of magnification (11)
Foot-operated switch for motorized adjustment of magnification (12 |
Lever to swivel in the green filter (13) (if ordered)
Lever for changing the diameter of illuminated field (14)
On-/off switch (15)
Change-over switch lamp I/lamp ITI (16) A
Knob for brightness control (17) |
Safety knob to change the binocular tube (18)
Knob to adjust the eye distance (19) (on inclinable tube 0-60”,
straight or inclined tube 60°) |
Apperture lever to adjust the depth of field (20)
Rotating knob for manual fine focussing (21).
Foot-operated switch for motorized fine focussing (22)
Rotating knob to change the motorized speed of ZOOM (23)
Rotating knob to change the motorized speed of fine focussing
unit (24)
(Fig.6)
KAPS 01.03.1996 page 9
_-— — SARA GS war wn IATL VEAL ODD)
® Rotating knob to change the motorized speed of XY-coupling (25)
(only if using XY-coupling)
e Foot switch panel (26) (only if using XY-coupling)
e Foot-operated switch for fine adjustment of microscope head in
XY-direction (27) (only if using XY-coupling)
(Fig.7)
KAPS 01.03.1996 page 10
User's Manual SOM 62 COLD LIGHT MOTORIZED
OPERATING INSTRUCTIONS (Fig.8) (Fig.9)
0
-1 ON-/OFF SWITCH (Fig.8)
e On-/off switch (1) is located on the upper side of the swivel
arm. After the appliance is switched on, the green light emitting
diode (2) indicates when microscope is ready for operating. The
xenon lamp burns :and the ventilator runs.
e Attention error indication:
Red light emitting diode (3) is blinking, see page 3, 1.0 security
~ indication: temperature limit... . Co -
Inform after-sales: service immediately...
Specify number of appliance, see type label (4).
e. To reduce. the burning life of lamp not unnecessarily, switch off
the appliance if the microscope-is:.not used for longer time... -
-
: |
7.2 CHANGE-OVER SWITCH LAMP I/LAMP II (Fig.8)
e In case of breakdown of lamp I (6), switch over to spare lamp II |
(7) by switch (5), see symbol (8).
3 CHANGING THE MOTORIZED SPEEDS (Fig.8)
.e Turn knob (9) for changing the motorized speed of ZOOM.
e Turn knob (10) for changing the motorized speed of fine focussing
unit.
e Turn knob (11) for changing the motorized speed of XY-coupling
(only if using XY-coupling).
7.4 BRAKES
® Adjust all brakes so that proper function of instrument is
guaranteed. |
(Fig.8)
KAPS 01.03.1996 . page 11
LA
User's Manual SOM 62 COLD LIGHT MOTORIZED
7.5 BRIGHTNESS CONTROL (Fia.9)
® A change of light intensity is made by rotating knob (1).
(Fig.9)
8.0 INSTRUCTIONS FOR USING THE MICROSCOPE
8.1 ADJUST THE EYE DISTANCE (Fig.10) (Fig.11)
® MICROSCOPE WITH STRAIGHT -OR- INCLINED TUBE (Fig.10)
Turn the microscope in working position.
e Straight tube (1) and inclined tube 60° (2): Bring eyepieces (3)
to eye distance by turning lateral knob (4).
® Inclined tube 45° (5): Bring eyepieces (6) by lateral movement
of eyepiece tube to eye distance (grip prism housing (7)).
(Fig.10)
KAPS 01.03.1996 page 12
User's Manual SOM 62 COLD LIGHT MOTORIZED
e¢ MICROSCOPE WITH INCLINABLE TUBE 0-60° (Fig.11)
Turn microscope in working position. Bring eyepieces (1) by
rotating the knob (2) to eye distance (points (3) for memory).
e Loosen clamping with knob (4). Choose best angle of viewing
(points (5) every 15° for memory).
(Fig.11)
8.2 CHANGING THE MAGNIFICATION (Fig.12).. .
® Adjust the highest magnification with rotating knob (1) of the
- magnification changer ZOOM motorized (2), respectively with
foot-operated switch for motorized adjustment of magnification '
(see page 9, 6.0 Control elements, (12) Fig.6).
e The magnification will be changed steplessly.
8.3 HOW TO FOCUSE THE OBJECT (Fig.12)
e Before starting the focussing, the fine focussing gear at fine
focussing unit (6) has to be brought in medium: position. Note
the marking points (8) on the right side of the microscope head.
e Adjusting the eyepieces (before focussing): |
Observer normal sighted: eyecup ejected (3).
e Bring "0" (zero) of the diopter scale of the eyepieces (4) to
match with the index line (5).
e Adjust the highest magnification at magnification changer ZOOM
motorized (2). |
e Approach microscope to the object till it is focussed, improve
focus by foot switch panel with foot-operated switch (see page 9,
6.0 Control elements, (22) Fig.6) for motorized adjustment: of :
fine focussing unit (6), respectively with rotating knob (7) for
manual fine focussing.
se When changing magnification, the image now.always will stay. well.
focussed.
® Spectacle wearers, slide back the eyecups (9) of the eyepieces.
"0" (zero) stayes matched with the index line also. The procedure
of focussing the image is the same as described before. After:
getting a well focussed image, observer maybe makes up his mind
to look with bare eyes through the eyepieces (only possible with
spherical eye deficiency at infinity). This is very easy
realizable: Without changing the distance between microscope and
object he can focus the image now by turning the diopter scale
of eyepieces, eyecups will be ejected again. With the found
position of diopter scale he can also start a new focussing
adjustment every time without wearing spectacles.
KAPS 01.03.1996 | page 13
User's Manual SOM 62 COLD LIGHT MOTORIZED
8.4 EYEPIECES WITH RETICLE (help for fine focussing and when usin
television or photo systems Fig.12
e Only one of the eyepieces must contain a reticle (magnification
must be identical).
e Focus the cross of the reticle by turning the diopterring (10) =
of the eyepiece, then: focus. the object with highest magnification - т
(procedure as described before). -
e Observe with or without spectacles.
(Fig.12)
KAPS 01.03.1996 page 14
User's Manual SOM 62 COLD LIGHT MOTORIZED
9.0
9.1
BREAKDOWN OF THE COLD LIGHT ILLUMINATION (Fig.13)
BREAKDOWN OF THE COLD LIGHT ILLUMINATION DURING OPERATION
If first xenon reflector lamp (1) is burned out switch over to spare lamp (2) by switch (3).
Change fibre optic light guide with mount (4) to spare lamp (exchange against plug (5)).
Attention: Before drawing out and changing fibre optic light guide (4) against plug (5) the
swivel arm (6) and the suspension arm (7) must be brought in showed position, note marks (8).
CHANGING THE XENON REFLECTOR LAMP LATER ON
Change only lamp of the same type!
Switch off appliance and wait until the lamp is cooled down.
| oosen knurled screw (9), lift lamp insert (10) and pull out u-profile with plug (11), then
remove completely lamp insert.
Push out the defect xenon reflector lamp (1) respectively (2) against resistance of holding
spring (12) (in direction of arrow) and replace it.
Attention: The nose (13) of the lamp must fit to the groove (14) of lamp insert.
Put in u-profile with plug (11) while gripping bulb (15) of xenon reflector lamp with clean cloth
or packing material. Do not hold the lamp at the reflector.
Integrate u-profile (11) only lengthwise (only lamps with boss), otherwise the ventilation is
hampered.
Tighten lamp insert with knurled screws (9) again.
(Fig. 13)
KAPS 01.10.1997 Page 15
ME
Ad REARATACI.L PINAL Ad Юй АА II. NFILLI EINER ELT
Un fal Vk le
10.0 CHANGING THE BINOCULAR TUBE (Fig.la)
e The change of binocular tubes is made at all versions in the same
way.
e Press pin with ball head (1), remove inclined tube 45° (2),
inclined tube 60° (3), straight tube (4) or inclinable tube -
0-60° (5) of magnification changer ZOOM motorized (6). ;
e When setting up tube - pin must be pressed down: - the two lugs. 7
should lock into the nuts (7) of the magnification changer ZOOM
motorized. Take care that tube lays plane. Release the pin. |.
(Fig.14)
11.0 CHANGING THE OBJECTIVES/EYEPIECES (Fig.15)
e The objectives (1) have screw bases. Unlock them by turning to a
the left and lock them by turning to the right. | |
e Eyepieces (2) are inserted, draw out for changing only. ©
e Other objectives/eyepieces can be selected by choice.
Refer to magnification chart 12.0.
(Fig. 15)
KAPS 01.03.1996
user's Manual SOM 62 COLD L1GHT MOTORIZED
12.0 SCHEDULE FOR MICROSCOPE MAGNIFICATION AND OBJECT FIELD MEASURES
SCHEDULE FOR MICROSCOPE MAGNIFICATION AND OBJEC
T FIELD MEASURES
Objective | Eyepieces | Eypieces
fin mm with Factor sign on magnification changer with
corresponds tube * tube
approxi- f= 159 0,4 0,63 1* 1,6 2,5 32 f=125
matley to mm | | mm
Me Noel > Total magnification / Field of view 2 in mm —
-- 3,2 / 34,7] 5,0 / 22,0) 7,9 / 13,7 | 12,6 / 8,7 | 19,7 / 5,51 25,3 / 4,3 6,3 x
6,3 x 4,0 / 281) 6,3/ 17,9 10,0 / 11,3-|-16,0:/- 7,0 | 2507 45] 3207 3,6] —
| -- 5,0 / 31,4 | 8,0 / 20,01 12,7 1 12,6 | 20,4 / 7,9 | 31,8 / 5,0| 408 / 3,91 10 «x
100 10 x | 6,4 / 23,6 | 10,0 / 17,0 | 15,9 / 10,8 | 25,5 / 6,6 | 39,8 / 3,81 51,0 / 2,91 125 x
12,5 х 8,0 / 22,01 12,5 / 14,8 119,9 / 9,3 | 31,8 / 5,8 | 49,7 / 3,51 63,6 / 2,7| 16 x
16 x | 10,2 / 18,8 | 16,0 / 12,0] 25,5 / 7,5 | 40,8 / 4,7 | 63,6 / 3,0| 81,4 / 2,31 20 x
20 x |12,8 / 14,2120,0 / 9,0|31,8 / 5,61 51,0 / 3,8 | 80,0 / 2,31 102,4 / 1,8 --
= 211560] 33/350] 53/221] 84/140] 13,17 901 17,0 7 6,7 6,3 x
63x | 2,6 / 42,0] 4,2 / 28,21 6,7 / 18,0 | 10,6 / 11,1| 16,7 / 7,21 21,4 7 5,3 --
-- 3,4 / 51,0| 5,3 / 32,5 | 8,4 / 20,5 {13,6 / 12,7 | 21,2 / 81| 272 / 5,9! 10 «x
150 10 x | 42/4001 6,7 / 26,1 | 10,6 / 16,4 | 16,9 / 10,2 | 26,5 / 6,51 33,9 / 44] 125 x
125 x | 5,3 / 35,1| 8,4 / 23,0 | 13,2 / 14,5 121,2 / 8,8 | 33,1 / 5,8 | 42,4 / 411 16 x
16 x | 6,8 / 31,5 110,7 / 17,91 17,0 / 11,0 127,2 / 7,0] 42,4 / 451 54,3 / 35| 20 x
20 x | 8,5 / 26,0 | 13,4 / 16,9 | 21,2 / 10,8 | 34,0 / 6,5 | 53,0 / 4,1| 67,8 / 2,7 --
= 1,8 / 62,5] 2,8 / 41,51 4,5 / 26,0| 7,2 / 16,3| 11,3 710,5 | 14,4 7 7,7 6,3 x
6,3 x 2,3 / 47,0| 3,6 / 32,5| 5,7 / 16,3| 9,2 / 13,0| 14,3 / 8,6| 18,2 / 6,2 --
- 2,9 / 59,8| 4,6 / 38,3| 7,3 / 23,6/| 11,7 / 14,9 | 18,2 / 9,5 | 23,3 / 6,9! 10 x
175 10 x | 37/450| 5,7 /295| 9,1 / 16,6 | 14,5 / 11,5 22,8 / 7,6 | 292 / 5,11 125 x
125 x | 4,6 / 40,3| 7,2 / 26,5|11,4 / 16,6 | 18,2 / 10,21 28,4 / 6,7 | 364 / 48| 16 x
16 x 5,8 / 31,0] 9,2/20,0|14,6 / 13,0123,3 / 7,7| 36,4 / 5,01 46,6 / 41] 20 x
20 «x 7,3 / 30,0 | 11,5 / 19,0 | 18,2 / 12,0 123,2 / 7,5| 455 / 47) 582 / 341 -
— 1,6 / 70,0] 2,5 / 47,21 3,8 / 29,21 6,3 / 18,4 Г 9,8 7 120} 125 / 8,9 6,3 х
6,3 x | 2,0 / 58,01 3,2 / 36,0| 5,0 / 23,0 | 8,1 / 15,0 | 12,6 / 10,0 | 16,0 / 7,1 —
- 2,6 1650] 4,0 / 43,3] 6,4 / 26,9 | 10,2 / 16,9 | 16,0 / 10,9 | 20,5 / 7,9] 10 «x
200 10 x 3,2 / 52,01 5,0 / 342! 8,0 / 22/0/4128 / 13,9|20,0 / 8,51 25,6 / 58| 125 x -
125 x | 40 / 47,0] 6,3 / 31,3] 10,0 / 19,7 | 16,0 / 12,2/25,0 / 78| 320 / 5,5! 16 x
16 x 5,1 / 37,51 8,1 / 23,8 | 12,8 / 15,0120,5 / 93/320/ 6,0/ 41,0 / 47| 20 x
20 x | 6,4 / 34,0| 10,0 / 22,0| 16,0 / 13,7 [25,6 / 8,7|40,0 / 5,41 512 / 3,5 -
= 1,3 / 90,0] 2,0 / 58,5] 3,2 / 37,01 5,0 / 23,0 | 7,9 / 15,0 10,2 / 11,0 6,3 x
6,3 x 1,6 / 71,2j 2,5 / 47,0| 4,0 / 28,7 | 6,5 / 18,0 | 10,0 / 12,01 12,8 / 8,9 _
- 2,1 / 85,0] 3,2 / 54,0] 5,1 / 33,51 8,2 / 21,0 12,8 / 13,2 | 16,4 / 98] 10 x
250 10 x 26 / 61,01 4,0 / 4041 6,4 / 26,3] 10,2 / 16,8 | 16,0 / 10,7|20,5 / 7,3| 125 x
125 x 3,2 / 560} 50/3741 6,0 / 23,5|12,8 / 14,8 [20,0 / 9,4125,6 / 68| 16 x
16 x | 4,17/47,2| 6,5 / 30,0| 10,2 / 19,01 16,4 / 11,8125,0 / 731328 / 59) 20 x
20 x 5,1 / 42,5| 8,1 / 27,31 12,8 / 47,0! 20,5 / 10,5 132,0 / 7,0[41,0 / 4,3 —
-- 1,2 / 97,3 | 1,8 / 62,0! 2,9 / 40,0] 4,6 / 25,21 7,2 / 16,1] 93 / 121 6,3 x
6,3 x 1,5 / 79,0] 2,3 / 50,0| 36 / 32,21 5,8 / 20,3] 9,1 / 13,0111,55 / 98 -
- 1,9 / 90,0} 2,9 / 60,01 4,6 /37,0| 7,4 / 23,0/11,5 / 14,9 | 14,7 / 1081 10 x
275 10 x | 2,3 / 71,0| 3,6 / 46,51 5,8 / 29,01 9,3 / 18,0 | 14,5 / 11,6 118,6 / 8,11 125 x
12,5 x 2,9 / 62,11 4,6 / 41,4] 7,2 / 25,8 | 11,6 / 16,0/18,0 / 10,2|23,0 / 7,6| 16 «x
16 x 3,7 / 51,0| 58 / 32,21 9,3 / 21,0|14,8 / 12,2 [23,1 / 60/295/ 65| 20 x
20 x | 46/445| 7,3 / 29,2|11,6 / 18,9 | 18,5 / 12,0129,0 / 7,5137,1 / 48 -
- 1,0/105,0 | 1,7 / 68,51 2,6 / 43,5| 42 7 27,3] 66 / 175] 83 / 134 6,3 x
6,3 x 1,3 / 84,5] 2,1 / 55,0| 3,4 / 35,0| 5,3 / 22,0/ 8,3 / 1431109 / 1068) —
_ 1,7 / 93,0} 2,7 / 63,0] 4,2 / 40,0| 6,8 / 25,0|10,6 / 16,01 13,6 / 11,8] 10 x
300 10 x | 2,1 / 7651 3,3 / 50,01 5,3 / 32,01 8,5 / 20,0 | 13,3 / 12,8 117,0 / 8,8! 12,5 x
12,5 x 2,7 1 68,8| 42 / 44,8| 6,6 / 28,1 | 10,6 / 17,41 16,6 / 11,0121,2 / 90| 16 x
16 x 3,4 / 55,01 5,3 / 35,0] 8,5 / 21,5|13,6 / 13,2121,2 / 8,5/272/ 70| 20 x
20 x 4,2 / 49,8| 6,7 / 32,5) 10,6 / 20,5 | 17,0 / 12,5126,5 / 8,1|339/ 5,3 -
- 08 /137,0j 1,2 / 85,0] 2,0 / 56,0] 3,2 / 36,0] 49 / 23,01 64 / 186 6,3 x
6,3 x 1,0 /112,5| 1,6 / 71,5| 2,5 / 45,3] 4,0 / 29,2| 6,3 / 18,9| 8,0 / 14,0 ~
- 1,3 /126,0| 2,0 / 82,51 3,2 / 52,5] 5,1 / 33,0| 8,0 / 21,2110,2 / 1571 10 x
400 10 x 1,6 / 95,7 | 2,5 / 63,0| 4,0 / 42,0] 6,4 / 26,2 10,0 / 17,0 112,8 / 11,8 12,5 x
12,5 x 2,0 / 91,0] 3,2 / 59,0| 5,0 / 37,41 8,0 / 23,1112,5 / 14,8 | 16,0 / 11.0| 16 x
16 x 2,6 / 76,0 | 4,0 / 47,0| 6,0 / 28,3 | 10,2 / 17,5116,0 / 11,5120,5 / 94! 20 «x
20 x | 32/7620] 5017430] 8,0 / 26,7 | 12,7 / 16,6 | 19,9 / 10,6 125,5 / 7.1 —
* To read 1 if microscope used without magnification changer.
KAPS 01.03.1996
page 17
USEL 5 Manual SUM 64 CULD LIGHT MOTORIZED
13.0 OBJECTIVE QUICK CHANGER (Fig.16)
e Screw in the complete objective quick changer (1) instead of a
normal objective into the stereo microscope until stop.
e If the objective quick changer (1) is not in the position showed
at the picture it is necessary to adjust it once.
® Loosen. .a. little . bit. 3 threaded pins (2) with enclosed hexagon * п
socket screw key, width across flats 1,5 mm. -
Bring the. complete objective quick .changer (1) .into the showed
(or in desired) position and tighten the 3 threaded pins (2) =
again, a
e Lift up by hand the "objective in mount" (3) being in the optical
axis of the microscope until stop (see arrow I) and bolt it with
а little rotating move to the right. (see arrow II).
e Quick changing: Unbolt the objective which is in working position
by rotating to the left and pull it down. Bring the second
objective (4) into the optical axis of the microscope by turning
the lower part of the objective quick changer until it locks
(180°). Lift up and bolt to the right.
e A non-reflecting image of object is only possible in this working
position.
(1) Unterteil
lower part
4
(Fig.16)
KAPS 01.03.1996 | page 18
User's Manual SOM 62 COLD LIGHT MOTORIZED
14.0 COMPENSATION OF WEIGHT WHEN USING ACCESSORIES (Fig.17)
e After supplementary accessories are mounted, the additional load
of suspension arm (1) must be compensated by pulling off one or
two counterweights (2).
® When dismantling put on counterweights at first. Then take off =
supplementary. accessories... :
e Before accessories will be removed, at first counterweights must -
be put on the arm so that it does not tip up.
15.0 MOUNTING OF BEAM SPLITTER (Fig.17)
e Press pin.with ball.head (3), remove binocular tube (4) of
magnification changer ZOOM motorized (5).
e When setting up beam splitter bilateral (6) or semilateral (7)
the two lugs of beam splitter should lock into the nuts (8) of
the magnification changer ZOOM motorized. Take care that beam
splitter lays plane.
® Fasten beam splitter with knurled screw (9) by hand.
® When setting up tube - pin with ball head (3) must be pressed
down - the two lugs of tube should lock into the nuts (10) of
beam splitter. Take care that tube lays plane. |
Release the pin.
(Fig.17)
KAPS 01.03.1996
user's manual sum oz COLD LIGHT MOTORIZED
16.0 OBSERVATION TUBE MONOCULAR, OBSERVATION TUBE BINOCULAR AND
ASSISTANCE TUBE 90° (Fiqg.18)
® Before adapting the tubes, remove cover from beam splitter (1)
by turning the threaded ring (2) in direction "open".
e If tubes (observation tube monocular (3), observation tube =
“binocular; (4), assistance tube.90° (5)) will be .inserted, take... . |
care that guide blades (7) hit grooves (6). 8
After. tube being inserted tighten threaded ring (2). о Do
Both superior parts of observation tubes (3) and (4) can be {=
_Swivelled.:laterally-by the observer. Hold: superior pipe with one — |
hand, press catch bolt (8) with ball of the thumb, move pipe in
position as requested and set free catch bolt. again. os
e Image erection: after swivelling for all tubes will be. done by.
turning the knurled ring (9).
e Observation tube binocular (4) can be combined with straight or
inclined tubes, if necessary. Assistance tube 90* mostiy will be
used with binocular tube 45° (10). For the observer these tubes
can be brought in the most confortable position by swivelling
after loosening knurled screw (11). After finishing, fasten -
knurled screw again (image erection is necessary once more).
e¢ Eyepieces (12) used with these accessories must have the same
magnification factor as eyepieces (13) used with the microscope.
(Fig.18)
KAPS 01.03.1996 page 20
17.0
я .
‘Mount: 35mm: camera: (8) -on:.photo- tube: (3) “(bayonet-lock).
user's Manual SOM 62 COLD LIGHT MOTORIZED
PHOTO TUBE WITH 35MM CAMERA, TV-TUBE WITH VIDEO CAMERA (CCD)
(Fig. 19)
Before adapting the photo tube or TV-tube, remove cover from beam
splitter (1) by turning the threaded ring (2) in direction "open".
If tubes (photo tube (3), TV-tube f=54mm (4) or TV-tube £=80mm
..and.f=145mn..(5).) are: inserted; .take care:that. guide blades: (7) -
hit grooves (6).
To find .the.best working position user. can insert tubes. all. 90%...
After. tube being inserted. tighten threaded ring (2)...
Screw on to stop video camera (9) on TV-tube (4) or (5) (C-moun
thread)... Ш a | |
.Erect the inverted. image: |
Turn threaded ring (10) or (11) in direction "open", turn video
camera till image at screen has position of object.
Tighten threaded ring (10) or (11) again.
Diaphragm setting:
If diaphragm will be closed more and more, the depth of field is:
growing. If illumination is powerfull enough, diaphragm can be
used with smallest opening continuously.
To open diaphragm of TV-tube f=54mm (4) move diaphragm ring (12)
in the same direction as it is written on threaded ring (10).
To open diaphragm of TV-tube f=80mm and f=145mm (5) move the
diaphragm ring (13) in the same direction as it is written on the
threaded ring (2) of beam splitter (1).
(Fig.19)
KAPS 01.03.1996 page 21
18.0
User's Manual SOM 62 COLD LIGHT MOTORIZED
PHOTOGRAPHING WITHOUT FLASH in combination with KAPS stereo
microscopes (photo equipment is already mounted, see 17.0)
Use artificial light film (if possible), at least 400 ASA.
It is helpful if camera shutter works automatically with changing
brightness.
For. correct focussing it:is: absolutely. necessary: to change:one: 4.
of the normal eyepieces against one with the same magnification
but ‘built-in reticle (VS-type with double cross). =. .... .
- This. reticle..is. adjusted.with great. accuracy .in the. correct
~ »Position where-the intermediate image= of: the object appears. -
If the customer uses a correct adapter for the camera, the
exposure plane of the film has the same optical distance.
Now: focus. the double cross of the.reticle by turning: the .
19.0
eyepiece diopterring (observation with spectacle or without).
The normal eyepieces diopter scale should be brought in the same
position. Bring microscope illumination to maximum and open
diaphragm of microscope. Start focussing your object with highest
microscope magnification, after changing magnification.the image
will stay focussed. |
The image in the camera will be focussed too.
Release photo camera by wire releaser (length appr. 30-50 cm)
or by selftimer.
Note:
If the customer uses camera with automatic shutter, it is very
important for making perfect exposures to cover the eyepiece of
the range finder against entering: light, use black tape or
similar, To get best results it is sometimes helpful to use
additional exposure compensation (part of most modern cameras).
In this case the automatic operating shutter speed can be caused
by factors to move more or less quicker. |
He also can use normal shutter speed, 1/2 to 1/30 sec or shorter,
he has to find out the best shutter speed when he is starting
photographing with microscope the first time only (test
exposures).
HOW TO GET A SHARP IMAGE ON THE TV SCREEN (Fig.20)
Very important: The best results will be realized if one of the
used eyepieces will be changed against one with a built-in cross
hair reticle.
Start with focussing the cross hair in the eyepiece by turning
the diopter ring. Adjust biggest factor at magnification changer.
Approach microscope of the object until it is focussed, you now
see a sharp image in the microscope.
If TV camera was delivered by KAPS Company, the camera is
adjusted and the screen now shows-a sharp image: too. If a not
adjusted system is used, the chip in the camera must be moved
a little bit (see camera manual) till you get a sharp image.
If you are not successful, maybe a distance ring between -camera-
and TV-tube is necessary.
If all is well adjusted as described before, you see a sharp
image in the microscope and at the screen even though
magnification will be changed. : В
If TV system is not delivered by KAPS Company, maybe it is
necessary to correct image sharpness on the screen although
the object in the microscope is well focussed.
Correction 1:
The chip position in the TV camera must be changed (without
moving the microscope). Most TV cameras have a possibility of
adjustment (see camera manual).
KAPS
01.03.1996 page 22
a EE
USEL 5 Manual bDUM DZ COLD LIGHT MOTORIZED
e Correction 2:
If correction 1 is not possible, the whole TV camera on TV-tube
must be moved:
IV-tube f=80mm:
Open diaphragm, reduce illumination if necessary. Mount TV camera
(upright position), loosen threaded ring (1) so that the 3 =
threaded: pins. (2:).become- visible,: 1loosen-threaded-pins: (use. =
hexagon socket wrench 1,5mm) by turning around 1/2 rotation each, -
-Cturn around.threaded ring (3) .up or down till a sharp image: on
~ Screen appears, . tighten threaded pins (2)..again, turn TV camera - |
~ till .image:oniscreen: has: position:of object: Tighten threaded: L
ring (1) again. | -
TV=tube f=54mm and £f£=145mm: | В
The threaded.pins..(2). are. visible without moving threaded: ring::
(1). Procedure is the same as described for TV-tube f=80mm.
For erecting the inverted image, loosen threaded ring (1), turn
around TV camera till image on screen has position of object,
tighten threaded ring again.
{= 80:
Tr >
open >
1 )
|
I |
/ — `
900
© ©
—
N
Ne j
(Fig.20)
KAPS 01.03.1996 page 23
user's Manual SOM 52 COLD LIGHT MOTORIZED
ASSISTANCE AND OBSERVATION SETTING 180° (Fig.21)
e
>
e The mounting of assistance and observation setting 180° (1) is
made in the same way as the mounting of beam splitter, see 15.0.
e Adaption with ZOOM magnification unit is made directly.
The adapter h=38mm (2) is necessary in combination with
magnification unit 3fold and. 5fold.
‚eo The mounting of adapter h=38mm is made as the mounting of beam
‚ - splitter, see 15.0. : - os
e To use all these accessories it is necessary that the axle of
. objectivesis brought in vertical position -(3): If there ‘a. —
centering and inclinable coupling 0-90° is in use, the angle
must be changed. to 0° (objective is .looking.to ;the table). . : В
A possibly used 45° coupling must. be taken. off.“ - |
e At the inclined side of assistance and observation setting 180°
preferably a binocular straight tube is to be used, at the side
of the microscope a binocular inclined tube 60°.
(Fig.21)
KAPS 01.03.1996 page 24
user's Manual sUM 52 COLD LIGHT MOTORIZED
21.0 MOUNTING AND OPERATION OF COUPLINGS (Fig.22)
(1) 45° coupling
(2) centering coupling
(3) centering and inclinable coupling 0-90°
(4) 90° coupling |
These. 4 versions.of couplings will: be mounted in- the. same way at: + |
the suspension arm (5).
Draw out microscope head. (6) (dismantling of microscope head in. …
reverse order as described at 4.0).
oe Insert one of the couplings .(1l), (2),.(3) or (4).
For that purpose unscrew safety cap (7) from guide cylinder (8)
= of the respective coupling. Take. care that clamp,. cross handle .
- (9). is open. Press locking pin. (10) .and- insert guide cylinder-of.
coupling to its stop. Release locking pin: it locks and thus
prevents the drop out of coupling. Screw on safety cap (7) on
overlying thread of guide cylinder by hand again.
e Insert microscope head (6).
For that purpose unscrew safety cap (11) from guide cylinder (12)
of microscope carrier (13). Take care that clamp, cross handle
(14), is open. Press locking pin at coupling and insert guide
cylinder of microscope carrier to its stop. Release locking pin;
it locks and thus prevents the drop out of microscope head. Screw
on safety cap (11) on overlying thread of guide cylinder by hand
again.
(Fig.22)
KAPS 01.03.1996 page 25
user's Manual SOM 62 COLD LIGHT MOTORIZED
- “centering -and: incliriable coupling: 0-90° (3).
e Operation: The different couplings are necessary for positioning
of different angles of the microscope carrier.
In case of 45° coupling (1) the microscope carrier is placed with
an angle of 45° in the coupling. The position of microscope -
carrier in case of 90° coupling (4) is parallel to the table.
The centering and inclinable coupling 0-90° (3) offers the
adjustment: from :0* : to. 902; .by steps: every: 15” .. The: centering: -
coupling (2) allows to adjust the microscope so that the sighted
-center is.not going.out of range,..if rotating (depends-on x
foregoing adjustment).. This.function also.is realized.by the .
_MICROSCOPE. CARRIER (2 GEARS).-(Fi
“The microscope carrier (2 gears) (1) is only available in
connection with fine focussing (2).
This version of microscope carrier must be ordered together with
the microscope (retrofit only in KAPS Company).
The insertion of microscope head (3) into the suspension arm (4)
is made as described at 4.0, page 5.
To adjust the microscope precisely around the vertical axle,
rotate knob (5), around the horizontal axle, rotate knob (6).
Before using knob (5) the guide cylinder (7) must be clamped (8).
\
\
\
| Va
SAN
(Fig.23)
KAPS 01.03.1996 page 26
user's Manual SUM 62 COLD LIGHT MOTORIZED
23.0 HANDLES AND MICROSCOPE GUIDES (Fig.24) _
(1) handle order no. 102.186-000
(2) microscope guide order no. 102.188-000
(3) T-handle order no. 162.860-000
(4) handle | order no. 102.184-000 =
(5) microscope guide . order no. 102.182-000 =
e The adaption of these handles or guides at the microscope head
is made all in the same way.
e Fasten handle or microscope. guide. with 2 screws (6) with hexagon 5
socket wrench. .(2,5mm). CE
e The T-handle (4) and microscope guide (5) has adjustable handles
(7); for adjusting draw out laterally. After bringing handles in …
desired position, they will be engaged automatically after
setting free.
(8) handle, sterilizable order no. 102.186-100
(9) microscope guide, sterilizable order no. 102.188-100
e Fasten handle or microscope guide with 2 knurled screws (10) at
' the microscope head.
e Take off handle or microscope guide for sterilization, both
. knurled screws cannot be lost.
(Fig.24)
KAPS 01.03.1996 | ‚ page 27
User's manual sum ol COLD LIGHT MOTORIZED
— mounted: adapter (see 4.0, page 5) will be inserted in the lamp -
The holding device is fitted with two guide pins (5) wherein the
-slit lamp (4) has been to insert till stop. Arrest with clamping
The second: fibre optic light guide .(7) will be ‘inserted tilk stop
MOUNTING AND USING THE SLIT LAMP (Fig.25)
Mounting: Holding device for slit lamp (2) is fixed on microscope
objective (1) in special mount. Instead of a normal objective
screw in this objective mount in the microscope. |
The holding device for slit lamp (2) is rotating around the lens
axis. Rotating.can be arrested with:clamping:screw-(3). °°"
SCrew (6).
in the slit lamp with plane side (light exit shows no fibre
structures) .. Arrest. with knurled screw (8)... The. other: end..of . . ._ ..
fibre optic. light guide (light. exit. shows. fibre. structures). with. ..
insert at the side of spare lamp.
The fibre optic light guide will be guided in cable holder
movable. Pay attention that the rotating move of the slit lamp
around the objective will not be hindered by fibre optic light: »
guide.
Setting in motion: By pressing change-over switch lamp I/lamp II
(see 7.2, page 11) by switched on appliance, microscope or slit
illumination can be choosen.
Slit width to operate steplessly with lever (9) from 0 to full
opening. The possible rotation of the slit with lever (10) around
the optical axle of slit lamp should be arrested with clamping
screw (11).
By not using the slit lamp it can be removed very easyly. By
inserting to holding device the fixed adjustment position will
always be exactly obtained.
—
TIT
(Fig. 25)
KAPS 01.03.1996 page 28
o
+
<
KAPS
(2) on. overlying.
MOUNTING AND USING THE XY-COUPLING (Fig.26)
The XY-coupling (1) must be ordered together with the microscope
(retrofit only in KAPS Company).
Unscrew safety cap (2) from guide cylinder (3) of XY-coupling (1).
Take care that clamp, cross handle (4), is open.
Press locking pin (5) and insert guide cylinder of XY-coupling to
its stop. Release. locking pin: it. locks. and thus prevents the.
drop out of XY-coupling.
Screw on safety cap
thread of guide
cylinder by hand
again.
Insert microscope
head (6). For that
purpose unscrew
safety cap (7) from
guide cylinder (8)
of microscope carrier
(9). Take care that
clamp, cross handle
(10), is open. Press
locking pin (11) at
XY-coupling and
insert guide cylinder
of microscope carrier
to its stop. Release
locking pin; it locks
and thus prevents the
drop out of microscope
head. Screw on safety
cap (7) on overlying
thread of guide
cylinder by hand
again.
Connect plug (12)
from XY-coupling (1)
with plug socket (13) (8) | in
and secure with |
knurled ring (14).
Attention: The nose
of the plug must fit
to the groove of plug
socket. o )
The XY-coupling allows (9)-
the motorized displace-
ment of the microscope
head in X- and
Y-direction (54mm).
Operation by foot switch
panel (see page 9 and 10,
6.0 Control elements,
Fig.6 (26) and (27)).
When switching on the
appliance the XY-
coupling goes to the
center automatically.
Autocentering of
XY-coupling from every
position by touching
the button (15). (Fig.26)
01.03.1996 page 29
User's Manual SUM 04 CULDY LIGHT MUTURIZED
26.0 ACCESSORIES (Survey with Order No.)
Order No.
101.310-000
101.315-000
162.670-100
101.295-000
101.,320-000
>. 101.330-000-
. 101.360-100
101.328-000
162. 290-200 -
102.187-000
102.189-000
102.170-000
108.120-000
112.370-000
101.945-000
101.355-000
101.930-000
101.110-000
101,955-000
101.950-000
101.970-000
101.800-000
101.850-000
101.860~000
101.410-000
101.430-000
101.440-000
101.450-000
101.460-000
101.510-000
101.530-000
101.540-000
101.550-000
101.560-000
101.610-000
101.620-000
101.680-000
101.630-000
101.640-000
101.650-000
101.660-000
101.670-000
102.181-000
102.191-000
101.900-000
101.910-000
102.186-000
102.186-100
102.188-000
102.188-100
162.860-000
102.184-000
102.182-000
112.580-100
- EYEPIECE
OBJECTIVE
BEAM SPLITTER, bilateral
BEAM SPLITTER,
semilateral
ASSISTANCE AND
OBSERVATION SETTING 180°
ADAPTER h= 38mm
PHOTOTUBE *
OBSERVATION TUBE ‘MONOCULAR
OBSERVATION TUBE BINOCULAR
CAMERA HOUSING
FINE: FOCUSSING ©. ZOOM: MOTORIZED
45° COUPLING
CENTERING COUPLING
CENTERING AND
INCLINABLE COUPLING 0-90°
90° COUPLING
MICROSCOPE CARRIER (2 gears)
(retrofit only in KAPS Company)
BINOCULAR STRAIGHT TUBE f=125mm --
BINOCULAR STRAIGHT TUBE f=159mm
BINOCULAR INCLINED TUBE 45° f=125mm
BINOCULAR INCLINED TUBE 45° f=159mm
BINOCULAR INCLINED TUBE 60° f=159mm
BINO. INCLINABLE TUBE 0-60° f=159mm
ASSISTANCE TUBE 90”
TV-TUBE WITH DIAPHRAGM f=145mm
TV-TUBE WITH DIAPHRAGM. f= 80mm
TV-TUBE WITH DIAPHRAGM f= 54mm
VIDEO CAMERA C-mount connection
EYEPIECE WITH RETICLE WF 6,3xVS
WITH RETICLE. WF 10 xVS.
WITH RETICLE WF 12,5xVS
WITH RETICLE WF 16 xVS
WITH RETICLE WF 20 xVS
WE 6,3xV
WF 10 xV
WE 12,5xV
WF 16 xV
WE 20 xV
f=100mm
f=150mm
f=175mm
f=200mm
f=250mm
£=275mm
f=300mm
f=400mm
QUICK CHANGER
(for magnification unit 3fold)
EYEPIECE
EYEPIECE
EYEPIECE
EYEPIECE
EYEPIECE
EYEPIECE
EYEPIECE
EYEPIECE
OBJECTIVE
OBJECTIVE
OBJECTIVE
OBJECTIVE
OBJECTIVE
OBJECTIVE
OBJECTIVE
OBJECTIVE
‘OBJECTIVE QUICK CHANGER
(for magnification unit 5fold, ZOOM)
SLIT LAMP 15° f=175mm
SLIT LAMP 15° f=200mm
HANDLE |
HANDLE, STERILIZABLE
MICROSCOPE GUIDE |
MICROSCOPE GUIDE, STERILIZABLE
T-HANDLE
HANDLE
MICROSCOPE GUIDE
XY -COUPLING
(retrofit only in KAPS Company)
(1)
(4)
(2)
(3)
(5)
(5)
(5)
(5)
(4)
(9)
(2)
(1)
(1)
(4)
(1)
(8)
(2)
(9)
(3)
(4)
(5)
(1)
Fig.17,
Fig.17,
Fig.21,
Fig.21,
Fig.19,
Fig.18,
Fig.18
Fig.19,
Fig:23,
Fig. 22,
Fig.22,
Fig.22,
Fig.22,
Fig.23,
Fig.14,
Fig.l4,
Fig.l4,
Fig.14,
Fig.18,
Fig.19,
Fig.1l9,
Fig.19,
Fig.19,
Fig.15,
Fig.15,
Fig.16,
Fig.25,
Fig, 24,
Fig. 24,
Fig.24,
Fig.24,
Fig. 24,
Fig.24,
Fig.24,
Fig. 26,
page
page
page
page
page
page
‚ раде
page
“page:
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
page
ADDITIONAL ACCESSORIES ARE AVAILABLE ON REQUEST, SEE PRICE LIST
19
19
24
24
21
20
20
21
26:
25
25
25
25
26
16
16
16
20
21
21
21
21
16
16
18
28
27
27
27
27
27
27
27
29
KAPS 01.03.1996
page 30
27.0 DESINFECTION AND STERILIZATION (Fig.27)
e For diagnosis:
Moisten smooth cotton cloth with antiseptic fluid (for example
Sagrotan-P). When required clean often touched parts, like
rotating knobs, handles and so on. |
There are also rubber sterilization caps for the rotating knobs of |
fine focussing .(1), magnification unit. (2) and brightness: control |
(3) on offer. -
Sterilization caps: | В
(4) order no. 101.733-000 | [=
(5) order no. 101.732-000 E
(6) order no. 101.731-000
The handle 102.186-100 and the microscope guide 102.188-100
(see 23.0 Handles and microscope guides (Fig.24), page 27) are
also sterilizable by heat (Autoklav).
e Use in surgery:
Very popular are sterilized plastic covers for one time use, |
prepared for accessories and openings for eyepieces and
objective.
Instead of this covers from plastic, texture covers for several
times sterilization are in use, too.
^^
Je
NV
A
(Fig.27)
KAPS 01.03.1996 | page 31
User's Manual SOM 62 COLD LIGHT MOTORIZED
28.0
29.0
29.1
CLEANING AND SERVICE
How to clean optical surfaces:
Remove coarse dirt particle with a clean dry hair penell from optical outer surfaces
(objective, eyepiece - eye lens). Before this, clean the penell in pure alcohol or ether
and let it dry.
Then moisten smooth cotton cloth with pure alcohol and wipe in a circle from the
middle of the lens outside. With dry cotton cloth polish the breathed lens. (You can
also use cleaning kits which are suitable for spectacles.)
The antireflex coatings are very resistant as they are hot evaporated. if you clean as
above described, the coatings will not be damaged.
How to clean lacquered parts:
Moisten clean smooth cotton cloth with water including a few drops of liquid soap
and wipe over the dirty parts. For remaining spots moisten the cloth with pure alcohol or
gasoline for cleaning and wipe carefully.
Service: Service as required, inform after - sales service.
TECHNICAL DATA
model:
measures:
mains supply:
variation of mains voltage:
mains frequency:
type of connection:
mains protection by fuses:
power input:
type of protection
against electrical shock:
mode of operation:
ambient temperature:
relative humidity:
weight:
xenon reflector lamp:
mains control:
lamp:
brightness range:
temperature limitation:
SOM 62 COLD LIGHT MOTORIZED,
order no. 162-400-200
see page 6
115 V/ 230 V, to switch over
max. + /- 10 %
50 / 60 Hz
single - phase, shrouded contact plug
2x T1Awith115V/230V
max. 150 VA
1, protective earth conductor
continuous operation
10° Cto 40° C
max. 75 %
appr. 100 kg
15 V/ 150 W, KAPS ident no. 700-509
+/- 41%
30 to 100 % lamp performance
protecting system by reducing lamp
performance
SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE
DECLARATION OF CONFORMITY WITH EU RULES
This appliance is manufactured in accordance with the specifications of the following
European directions:
93 / 42 EWG
Products of medicine
89/336 / EWG, 92/31 / EWG Emission of electromagnetic disturbances and
73/231 EWG
immunity against electromagnetic disturbances
Electrical production facilitys for use within definated
voltage ranges
For judgement the following harmonized standards are used:
DIN EN 60601 - 1 (VDE 0750 part 1):
A2: 1995
IEC 601 - 1: 1988 - 12
IEC 601 -1/A1: 1991 - 11
IEC 601 - 1 / A2: 1995 - 03
1996 - 03 EN 60601 - 1: 1990 + A1: 1993 +
Transport and storage: a), b) and c) as in EN 60601 - 1
KAPS 01.10.1997
Page 32
">
Download
Advertisement
Key features
Stereo microscope
Cold light illumination
Motorized zoom
Motorized fine focus
Foot-operated controls
Adjustable eye distance
Changeable magnification
Changeable objectives
Accessories compatible
Frequently asked questions
To adjust the eye distance, turn the lateral knob located on the eyepiece tube. For inclined tubes, you may need to move the eyepiece tube laterally to achieve the desired eye distance.
You can change the magnification by using the rotating knob on the magnification changer ZOOM motorized or by using the foot-operated switch for motorized adjustment of magnification.
Before focusing, ensure the fine focusing gear on the fine focusing unit is in the middle position. Adjust the eyepieces by bringing the "0" on the diopter scale of the eyepieces to match the index line. Then, approach the microscope to the object until it is focused and fine-tune the focus using the foot-operated switch for motorized adjustment of the fine focusing unit or the rotating knob for manual fine focusing.